Contents
Working Principle
Composition
Installation
Adjustment
Maintenance
Upgrade
Troubleshooting
Exam
Principle
1. Working principle
WBC Testing -
WBC Test
Testing
Optical system
Semi-conductor Laser:
Focusing
Long life Testing
Stability
Collimator
Low consumption
Focusing light split
Collimator :
Sheath-flow
Precise angle control cell
Light source
Eliminating stray light
Accurate result
Principle
MO EO
GH LY
Cell internal structure
Principle
Hemoglobin Channel-Colorimetry
The diluted sample is mixed with lyse reagent, RBC dissolve, releas
ing hemoglobin.
Hemoglobin combined with lyse reagent to form hemoglobin complex.
The hemoglobin complex is irradiated by monochromatic light with a w
avelength of 546nm .
Phototube is used to receive transmission light, and convert the light sig
nal into voltage signal.
By comparing with the voltage produced by background transmission li
ght intensity before sample adding (only diluent exists), the hemoglobin
concentration can be obtained(HGB), unit is g / L.
4DIFF Channel :
Reagent :
BF-FDO Reagent : stain, firming agent
BF-FDT Reagent: dissolve RBCs and specialize different types of WB
Cs ;
Diluent : rinsing , offering sheath
BASO Channel :
Reagent:
SLS-I Reagent : dissolve RBCs, platelets,combined with hemoglobin a
nd remove cell membrane of WBCs exclude basophil; keep Baso main
tain the original form.
Diluent : rinsing
Principle
Working principle
Principle
RBC pulse
RBC low
limited
PLT
pulse
PLT low
limited
noise
Principle
Reagent :
Diluent : diluting, rinsing, offering conductive
environment, cells equal volume processing
Principle
Product Name Spec. When When power When runs When runs
switch off (capillary (whole
on blood,CBC+DIFF blood,CBC+DIFF
Mode) Mode)
2.Composition
1
Front cover
2 Counting key
3 Mode key
4 Indicator
5 STAT rack
6 Tube rack
7 Auto loader
Composition
8 Fan
9 BF-SLS-I Inlet
10 BF-FDO Inlet
11 BF-FDT Inlet
12 BF-PK Diluent
13 Waste outlet
Composition
14 Optical Assembly
15 Sampler Assembly
16 Chamber
17 Pressure regulator
18 Sample syringe
19 Sheath liquid syringe
20 Valve Assembly
Four Syringes Assem
21
bly
Composition
22 Right door
23 Net port
24 Connector
25 Power filter
Composition
Composition
Main Board
Function :
Coordinate with other system module
Communicate with computer through net port
Provide RJ45 and RS232 port
Provide SPI for data acquisition
Composition
Driver board :
Function :
Serial communication ( main board , AD board and
temp control board )
Drive the step motor ( sample motor 、 diluent motor 、 Lyse mot
or 、 test motor 、 X axis motor 、 Y axis motor )
Drive the valve and pump ( 27 valves 、 1 positive pressure pump
、 2 vacuum pumps )
Indicator control ( power indicator, status indicator, alarm indicator
)
Reagent supply control
Peripheral sensor control (8 float sensors,5 photo sensors,
5 switchs )
Composition
Whole blood 11 ul
FDO 1 ml
FDT 0.2ml
Diluent 3ml
The ratio of
The ratio of diluent:1:30000 for
diluent:1:583 for WBC/
RBC and PLT
BASO and HGB
Pipeline
Sample20ul
Dispense 0.2mlFDT
Sample probe lift up an
d discharge residual 3ul
sample into swab
Aspirate 1 ml
mixed sample
Sample probe aspirate 5 Sample probe dispense
0ul diluent sample from 50ul diluent sample in R
Test in flowcell WBC/HGB bath BC/PLT bath
3.Installation
1.Intallation requirement
Space Requirement
Ensure enough space for instrument maintenance.
The space between the wall and the right & left side of the
instrument ≥50cm
The space between the wall and the back side of the instrum
ent≥50cm
The space between the front of the analyzer and other equip
ment ≥ 100cm;
Ensure enough space under and above the instrument for dil
uent, reagent and waste solution collecting devices.
BF6500&BF6800 automatic
hematology analyzer
Power Requirement
Power : 100~240VAC 50/60Hz
Fuse : 250V 4A ;
Environmental Requirement
Working Temperature : 15℃-30℃ ;
Relative Humidity :≤ 75 % ;
Atmospheric Pressure : 75kPa-106kPa ;
The instrument should be protected from dust, mechanical vibration,
significant noise and power interference.
It is recommended that the electromagnetic environment assessment of the
laboratory should be conducted prior to test.
Do not use this instrument in close proximity to sources of strong electroma
gnetic radiation, as these may interfere with the proper working.
It should be placed far from the constant ON-OFF electrical devices like bru
sh-type motor, fluorescent lamp.
It should be placed far from heat and wind source, sunlight, brush-type mot
or, flickering fluorescent light and electrical contact equipment.
A well-ventilated place is a must.
BF6500&BF6800 automatic
hematology analyzer
2. Unpacking
Check the packaging of the instrument. Contact Dirui or
local distributor for any physical damage.
3.Intallation step
1. Connecting
a. Lyse, Diluent, Waste Solution Connecting
b. Liquid Level Sensor Connecting
c.Computer Connecting
d.Power Wire Connecting
e.Barcode Reader Connecting
f.Printer Connecting
BF6500&BF6800 automatic
hematology analyzer
BF6500&BF6800 automatic
hematology analyzer
Warning:
Safty screws should be removed before power o
n, otherwise analyzer may be damaged.
BF6500&BF6800 automatic
hematology analyzer
BF6500&BF6800 automatic
hematology analyzer
BF6500&BF6800 automatic
hematology analyzer
4.setup
Login the software and find the instrument IP as picture .
Set the computer IP as 172.16.100.XXX to get the communication .
BF6500&BF6800 automatic
hematology analyzer
4.Adjustment
1. Sample probe
The nozzle of the probe should face to the probe base board and at
the same level with outlet of swab
If necessary ,the vertical position sensor of probe could be adjust
also .
BF6500&BF6800 automatic
hematology analyzer
vertical position
sensor of probe
BF6500&BF6800 automatic
hematology analyzer
Tube should
Aline with stay at the
each other center of
mechanical
hand center
BF6500&BF6800 automatic
hematology analyzer
Excursion in
Excursion in
backward
forward
direction
direction
BF6500&BF6800 automatic
hematology analyzer
Screw
Screw
BF6500&BF6800 automatic
hematology analyzer
Excursion
Excursion
BF6500&BF6800 automatic
hematology analyzer
6.Preesure adjustment
Preesure Regulator
Exhaust Knob
BF6500&BF6800 automatic
hematology analyzer
5.Maintenance
1 .Regular Maintenance
Daily Maintenance:
Rinsing Equipment should be carry out every 24 hours .
Click the Icon to run it
Weekly Maintenance:
If rinsing equipment is conducted everyday, the “Soaking with detergent
“ should be carry out every week .
Enter service –Maitain—soaking with detergent
Then run WBC Pool ,RBC pool ,DIFF pool one by one
BF6500&BF6800 automatic
hematology analyzer
BF6500&BF6800 automatic
hematology analyzer
Monthly Maintenance:
The rinse Swab should be carry out every month
Enter service –Maitain—rinse—rinse Swab
Then run rinse Swab
BF6500&BF6800 automatic
hematology analyzer
BF6500&BF6800 automatic
hematology analyzer
Necessary Maintenance
1.If live the analyzer vacantly more than 2 weeks ,the package clean
and empty should be carry out .
2.If the analyzer give alarm as “aperture clog” please carry out
aperture as follow
3.If the analyzer does not startup for a long time ,please prime the
pipe before use
6.Upgrading
1. Control Board Upgrading
Control Board upgrading: Open service interface of BF-
6800 information management system, click 【 Servic
e 】 ---> 【 Status 】 ---> 【 System Version 】 , open c
ontrol panel interface, as figure shows:
BF6500&BF6800 automatic
hematology analyzer
BF6500&BF6800 automatic
hematology analyzer
Click OK, select the control panel document need to be upgraded, the follo
wing interface pops up, as figure shows.
BF6500&BF6800 automatic
hematology analyzer
Click Open, control panel start to upgrade, the following box pops up, as
figure 8-4 shows
BF6500&BF6800 automatic
hematology analyzer
2 .Software Upgrading
Note : Exit the program before BF-6800 information management syste
m upgrading.
Click Upgrade.exe in program installation directory, as figure 8-6 shows.
BF6500&BF6800 automatic
hematology analyzer
BF6500&BF6800 automatic
hematology analyzer
BF6500&BF6800 automatic
hematology analyzer
BF6500&BF6800 automatic
hematology analyzer
BF6500&BF6800 automatic
hematology analyzer
BF6500&BF6800 automatic
hematology analyzer
BF6500&BF6800 automatic
hematology analyzer
BF6500&BF6800 automatic
hematology analyzer
BF6500&BF6800 automatic
hematology analyzer
7.Troublshooting
1. 4 Diff scattergram does not appear after test or bad scattergram
Analysis :
1.Mixed sample can not reach flow cell
2. Not enough reagent reach 4 diff bath or expired reagent
3.Dirty or blocked
BF6500&BF6800 automatic
hematology analyzer
Solution :