• Manual
• Impedance
• Optical
Manual Counting Method
WBC/RBC/Platelets
• Dilutions of Blood were made and Cells Counted
in a Counting Chamber.
Differential
• A Blood Smear is Stained and the Different
4
8 4. Unified Fluidics Circuit
7 5. Perox Optical Assembly
2 3
6. Rbc Optical Assembly
9 7. Reagents, Sheath, Wash &
1 Rinse
8. Vacuum & Pressure Regulator
Knobs
9. Touchpad
Barcode Reader
Waste Removal System
The Touchpad
Unified Fluidics Circuit
UFC
UFC
1. Hgb Colorimeter
1 2
3
4
2
3
5
Reagent Inventory
Advia 120 Methods
Lamp
assembly UFC Filter + Photodiode
Seconds
5. Advia 120 Sheath rinse Readings (Baseline Mean)
Calculating Reported Parameters
• HGB Log (Sample Mean) x Hgb Cal
Factor x 50.0 (Haemoglobin) (Baseline Mean)
• MCH (Mean Corpuscular Haemoglobin)
(HGB) x 10
(RBC)
• MCHC (Mean Corpuscular Haemoglobin
Concentration)
(HGB ) x 1000 ([ RBC x MCV])
Advia 120 Methods
Filter
Sample Scatter
detector
Tungsten Stream
lamp
Dark
Beam
splitter stop
The Perox Method
Scatter signal to measure the
volume of the cells
7 2 nRBC
5 6 3 Platelet Clumps
4
Cell Size
5 Luc
4 6 Monocytes
2 3 7 Neutrophils
2 8
1 8 Eosinophils
Peroxidase Activity X
Advia 120 Methods
No matter
what your shape
or size ....
We can make you
a SPHERE
RBC’s Biconcave Disc
RBC’s Sphered
Isovolumetrically Sphered RBC’s
and Plts
Orientation of Cells is
Not a Variable
Predictable Light Scatter and
Absorption
The Flow Cell
Advia 120 Sheath/rinse Encases the
Sample
Sample stream Concentric Flow Cell Module (CFM)
The RBC Method
Laser Light
Laser Light
Laser
Zero Angle
1. 60 fL Volume marker
3. 28 g/dL HC marker
4. 41 g/dL HC marker
Haemoglobin Concentration