This method bases on the
measurement of changes in
electrical resistance produced
by a blood cell, suspended in a
conductive diluent as it passes
through an aperture of known
dimensions, and this change
will produces a measurable
electrical pulse. The number of
Impedance Method
these pulses indicate the
number of WBCs that have
passed through the aperture.
HGB Measurement Principle
This method bases on the
measurement of emitting a beam
of LED light through the diluted
HGB sample mixed with a
certain amount of the hemolysin,
and it can be measurable at 525-
535 nanometer by a photo diode.
The signal is then amplified and
the voltage is measured and
Colorimetric Method
compared to the blank reference
reading.
RBC/PLT Measurement Principle
This method is an improved
method of the Impedance
method. The diluted sample is
pushed into the conical chamber
by the sample nozzle, and
wrapped by reagent in the
anterior sheath flow through the
center of the aperture. And the
Sheath Flow Impedance Method accuracy and reproducibility of
blood cell counting will be
improved by the sheath flow
technique.
Function of hydraulic module
Syringe
RBC WBC/HGB
Probe & Swab
Detecting Pool Detecting Pool
Isolation Pool Air tank Protection Pool
Air/Liquid RBC Mix Pool Waste
isolation Pool Collection Pool