Professional Documents
Culture Documents
Please read and understand the service manual carefully to ensure correct servicing of the Analyzer.
After reading, please keep the service manual properly and place it at an accessible position for lookup at any
time.
Model: FUS-1000
Intellectual Property
Dirui Industrial Co., Ltd. (hereinafter referred to as Dirui Company) reserves the intellectual property of the
service manual and related products.
Without the written approval of Dirui Company, any individuals or organizations shall not copy, alter or
translate any parts of the manual.
Statement
Dirui Company declares that it will be responsible for the safety, reliability and performances of the product
only if all following requirements are met.
(1) The assembly, readjustment, improvement and repair of the Analyzer are undertaken by personnel
approved by Dirui Company.
Charge-free services:
Products within the scope of warranty of Dirui Company can enjoy charge-free services.
Chargeable services:
(1) Services for products beyond the range of the warranty of Dirui Company.
(2) Repair of products caused by following factors, even within the warranty period.
Artificial damages, improper use, network voltage beyond the regulated range, irresistible natural disasters,
accessories or consumables not approved by Dirui Company, or repair of the product by personnel not authorized by
Dirui Company.
Change/ Refund
Please follow steps below for change/ refund from Dirui Company.
(1) Get the right of change/ refund: Contact the after-sales service personnel of Dirui Company, inform the
serial number of the product, indicate the product model and describe the reason of refund in brief.
(2) Transportation expense: The user shall bear the transportation expenses (including customs fees) to Dirui
Company.
Address: 3333 Yiju Road, New & High Tech. Development Zone, Changchun Post Code: 130103
Domestic Customer Service Hotline: 400 811 6695 400 811 6605
The readers of the service manual are Dirui Company and the maintenance personnel authorized by Dirui
Company.
Symbols
1. Symbols
Symbol Meaning
Laser
Be careful, burns
AC
Lot number
Life
Serial No.
Manufacturing date
Manufacturer
The equipment complies with the requirements of European Union for in vitro diagnostic medical
devices.
Classification No.
On (power)
Off (power)
Protective earthing
Symbol Meaning
Mark indicating the use of restricted hazardous substances in electronic and electrical products. It
means the product contains some substances whose use is restricted (but the content is within the
allowable range), the product can be used without hazards in an environment-friendly use period
(7 years) but it shall be put to recycling system after the expiry of environment-friendly use
period.
The symbols above are also applicable to the Analyzer, reagent, QC object and calibration object.
2. Identifications
Laser radiation
Warning Do not look straight at the light beam
Maximum output:
Laser wavelength:
Class 2 laser product
(1)
Class 1 laser
product
(2)
Warning
● Power supply shall be disconnected before
the fuse inspection.
●Fuse of same specification shall be used for
fuse replacement.
Fuse
(3)
Warning
1. Do not open when
operating.
2. Do not touch when
operating.
(4)
Opening by
specialized person
only
(5)
Safety Precautions
● If the hospital or organization in charge of the instrument can not formulate a satisfactory servicing/
repair plan, the instrument may not operate normally and personal health may be affected.
● Combustible gas (such as anesthetic) or combustible liquid (such as ethyl alcohol) shall not be used near
the product to prevent explosion.
● Power supply shall be cut off when the product is repaired. Repair of the instrument when it is turned on
may cause electric shock and damages of electrical components.
● Please connect the Analyzer to a socket with an independent fuse and protective switch. If the Analyzer
shares a fuse and protective switch with other equipment (such as life support equipment), trip may occur in
case of a fault, over current or impact current at the moment of startup.
● The clothes, hair and hands of the servicing personnel shall be away from moving components such as
sampling probe to prevent pinching or piercing.
● As the part with warning signs may have mechanical action, human body injuries such as pinching and
stabbing may be caused in normal operation, disassembly or servicing.
● The operator is obligated to follow national and local regulations on discharge and treatment of expired
reagent, waste liquid, waste sample and consumables.
● The reagent will irritate eyes, skin and mucosa. The operator shall abide by laboratory operation safety
regulations and wear personal protective appliances (such as laboratory clothes and gloves) when touching
reagent related articles in the laboratory.
● Once the reagent touches with skin, use a large amount of water to flush the skin, or send the injured to a
doctor if necessary. Once the reagent touches with eyes, use a large amount of water to flush the eyes and
send the injured to a doctor.
● The Analyzer shall be used by professional medical examination personnel or trained doctors, nurses or
testers.
● The samples, quality control objects, calibration objects and waste liquid have potential biological
infectivity. The servicing person shall abide by laboratory operation safety regulations and wear personal
protective appliances (such as laboratory clothes and gloves) when touching and handling relevant articles in
the laboratory.
● As all components and surface of the Analyzer have potential infectivity, safety protection measures shall
be taken in operation and servicing.
● Do not touch the tip of the sampling probe as the tip is sharp and it may have blood sample, quality control
object and calibration object that have potential biological infectivity.
Contents
III
Service Manual of FUS-1000 Urinalysis Hybrid
1
5
4 2
3
1
Service Manual of FUS-1000 Urinalysis Hybrid
6
1
5
4
3
12
11 2
3
4
10 5
9 6
8 7
Fig. 1-2-3 Rear view (without shield) of the instrument
2
Service Manual of FUS-1000 Urinalysis Hybrid
Sheath
Fig. 1-3-1
The sheath flow technology, high-speed camera technology and artificial intelligent recognition technology used
for the Analyzer are described as below.
(1) Sheath flow technology: The sheath used by the Analyzer in testing process is isotonic and agranular solution
with buffer function. The sheath can ensure the sediment in urine sample solution always flows independently as a
single layer. As flow eytometry is applied, it can be ensured each sediment will flow in front of the microscope lens
and high-speed camera within the focus of microscope lens and be taken with pictures. Besides, as the urine flows in
diffusion, aggregation of sediment can be effectively prevented.
(2) High-speed camera technology: The urine sample wrapped in the sheath flows to the flow cell and passes the
microscope lens in a form of flat laminar flow. Its thickness and position are just within the focus of microscope lens.
According to the sheath flow type imaging principle, all particles will directly align to the lens in its maximum
section area when they pass the lens. When the field of the microscope is lighted up by light source, all passing
sediment will be instantly shot. In a certain time period, the high-speed camera will take 2500 images with sediment
of all samples, As shown in Fig. 1-3-2.
3
Service Manual of FUS-1000 Urinalysis Hybrid
Fig. 1-3-2
(3) Artificial intelligent recognition technology: The automatic sediment recognition software and highly
trained intelligent recognition technology can extract the images of sediment particles quickly and identify and
classify them according to the shape, texture and characteristics in frequency domain of the shot particles. The
Analyzer can classify these particles into 12 classifications: Red blood cell (RBC), white blood cell (WBC), white
blood cell cluster (WBCC), squamous epithelial cell (SQEP), non-squamous epithelial cell (NSE), hyaline cast
(HYAL), pathologic cast (UNCC), bacteria (BACT), microzyme (BYST), unclassified crystallization (UNCX),
sperm (SPRM) and mucous strands (MUCS).
(4) After the automatic sediment recognition software makes classification, the quantity of images of shot
particles and the volume of scanned urine sample will be used to calculate the concentration of sediment. The result
can be represented by the number contained in per microliter or the number contained in per high power field/ low
power field.
4
Service Manual of FUS-1000 Urinalysis Hybrid
5
Service Manual of FUS-1000 Urinalysis Hybrid
If the operating environment or power supply of the Analyzer does not meet requirements above, the
accuracy and precision of test results of the Analyzer may be affected, the Analyzer may be damaged or
personal injury may be caused.
6
Service Manual of FUS-1000 Urinalysis Hybrid
2.2 Unpacking
2.2.1 Unpacking steps
After the arrival of the Analyzer, please carefully check the physical damage of the Analyzer package, if the package
is damaged, please contact with Dirui Company or our local agent. If no outer damage is found, please unpack
following the steps below:
(1)Put the package upright as the arrow indicates.
(2)Open the Analyzer case and accessories case, and check materials inside based on packing list, if any goods
missed, please contact with Dirui Company or our local agent.
(3)Please check the appearance of the Analyzer, if damaged, please contact with Dirui Industrial Co immediately.
2.2.2 Handling methods
(1)Handle directly in short distance under the stable condition.
(2)The Analyzer should be maintained upright when handling and transporting.
(3)Try to avoid vibration when handling.
After the installation of the sample-feeding unit, remove the wire ties on the host used for fastening the
sample-aspirating probe and testing head; otherwise, the Analyzer failure may be caused.
Connector 2
Connector 1
Fig. 2-3-1
7
Service Manual of FUS-1000 Urinalysis Hybrid
Fig. 2-3-2
(2)Connection of computer
Connect the display, mouse, keyboard and power cable of computer, and connect one end of the accompanying
communication cable (with DB-9 holes at the two ends) to the “RS-232” port on the left rear cover plate of the
Analyzer with another end connected to the serial communication port of the computer host.
(3)Connection of camera communication ports
Connect one end of the accompanying network communication cable to the Camera port on the rear cover plate of the
Analyzer, with another end connected to the special network port (superior posterior part of the host, independent
network card in standard configuration, as shown in Fig. 2-3-3,) of the computer host.
Fig. 2-3-3
Notes:
The computer has two network ports on the back. The independent network card located at upper part is
used for the communication with the Analyzer and the network port on the lower mainboard is used for LIS
system communication.
(4)Connection of detergent liquid level sensor and detergent tube
Insert the plug of the accompanying detergent liquid level sensor into the “Detergent sensor port” on the right rear
cover plate of the Analyzer. Connect one end of detergent tube with the “Detergent port” on the right rear cover plate
of the Analyzer and put the detergent level sensor and the other end of the detergent tube into the detergent tank.
(5)Connection of sheath level sensor and sheath tube
Insert the plug of the accompanying sheath level sensor into the “Sheath sensor port” on the right rear cover plate of
the Analyzer. Connect one end of sheath tube with the “Sheath port” on the right rear cover plate of the Analyzer and
put the sheath level sensor and the other end of the sheath tube into the sheath tank.
The liquid level sensor and the sheath tube shall be inserted simultaneously. Do not touch with hands to
prevent polluting the sheath and affecting the test value accuracy.
8
Service Manual of FUS-1000 Urinalysis Hybrid
The difference in height between the waste liquid tank and Analyzer should be more than 0.5m.
(7)Connection of printer
Please correctly connect printer with the USB cable of the computer host and check:
a)Whether the printer driver is installed.
b)Specification of printing paper used in the printer.
(8)The installation of external bar code reader
Connect another end of accompanying bar code reader with the “USB” port on the computer host.
The bar code reader will emit light harmful to people’s eyes; please do not look straight at the light when the
Analyzer is working.
(9)Connection of power cable
Insert one end of power cable into Power socket, with another end connected with receptacle to ensure reliable
grounding.
● The receptacle connected with the power cable shall be reliably grounded.
● The receptacle connected with power cable shall be placed near the Analyzer and easily disconnected.
● To ensure the reliable operation of the Analyzer, sharing the same receptacle with high-power electrical
appliances (such as air conditioner, refrigerator and oven) should be avoided.
9
Service Manual of FUS-1000 Urinalysis Hybrid
Fig. 2-4-1
After inputting correct installation password (installation password is provided by Dirui Industrial Co., Ltd.), click
and the interface is as shown in Fig. 2-4-2:
Fig. 2-4-2
Click on the interface (Fig. 2-4-2) directly, and the interface as shown in Fig. 2-4-3 will be displayed.
10
Service Manual of FUS-1000 Urinalysis Hybrid
Fig. 2-4-3
Select the language when the program is operating on the interface (Fig. 2-4-3) and then click , and the
interface is as shown in Fig. 2-4-4:
Fig. 2-4-4
Click on the interface (Fig. 2-4-4) directly, and the interface as shown in Fig. 2-4-5 will be
displayed.
11
Service Manual of FUS-1000 Urinalysis Hybrid
Fig. 2-4-5
Click on the interface (Fig. 2-4-5) and the installation of online software FUS-1000 is finished.
Fig. 2-5-1
2.6 Login
Turn on the power switch of the Analyzer first and then log in the application software.
2.6.1 Login
After the software installation, double-click the icon of Analyzer application software (hereinafter referred
to as the software), or find the software in “Program” window and click “Start” to enter the “System login” window,
and the interface is as shown in Fig. 2-6-1:
12
Service Manual of FUS-1000 Urinalysis Hybrid
Fig. 2-6-1
Input user name and password, the initial user name of Analyzer is Admin, and the initial password is 1, if wrong user
name and password are input, login failure will appear, and the interface is as shown in Fig. 2-6-2:
Fig. 2-6-2
If the error repeats for 3 times, the software will show “Login failure for three times!”, click the button
in the prompt box, and the program will automatically exit.
Input the correct user name and password in the login window (Fig. 2-6-1), and click or Enter to enter
the main window of the software, and the interface is as shown in Fig. 2-6-3:
Fig. 2-6-3
13
Service Manual of FUS-1000 Urinalysis Hybrid
Click in the main key area on the interface (Fig. 2-6-3), and the interface is as shown in Fig. 2-6-4:
Fig. 2-6-4
Click in the prompt box to go back to the login interface (Fig. 2-6-1) to switch user.
2.6.3 Exit system
Under the standby mode, click in the main key area to enter the shutdown dialog box interface, as
shown in Fig. 2-6-5.
Fig. 2-6-5
Click the button, and the Analyzer executes rinse and exit; click the button to exit the
system; click the button, and the Analyzer is started again; click the button, and close the
shutdown window to return to the original interface.
● If the operator is to leave the Analyzer for a long time, it is recommended to log off the software to prevent
information modification. Users are recommended to back up the database regularly in case of accidental
data loss.
● After the initial user name and password are input and first-time login is finished, set the user name,
14
Service Manual of FUS-1000 Urinalysis Hybrid
Fig. 2-7-1
Among them, is an alarm icon. When an alarm occurs, the alarm icon will appear at the right side of the main
key area. The operator can click the icon to enter the alarm information window and check the content and solutions.
The icon will disappear when alarm information is deleted, and will not appear again until the next alarm occurs.
(2)State bar
It is located at the bottom of the screen, and the interface is shown in Fig. 2-7-2 below.
Fig. 2-7-2
: Analyzer state, real-time display of all state of the Analyzer, including standby mode,
sample test, shutdown, sample preparation and etc.
: Show the current user name and authority of the software;the user can add or delete
intab “User information” of the “Management” window.
: Display the communication status between the Analyzer and online software in
realtime.
: Display the communication status between Analyzer and LIS system in real time.
15
Service Manual of FUS-1000 Urinalysis Hybrid
Fig. 2-7-3
● After the program is installed for the first time, the strip type is FUS-12MA II by default.
● If the program is not installed for the first time, the type set before the program uninstallation will be
displayed. If the strip type is set as FUS-10 II before software uninstallation, the strip type will be displayed as
FUS-10 II when logging in after re-installation or under standby mode after turn-on and resetting.
16
Service Manual of FUS-1000 Urinalysis Hybrid
5 2
4
3
Fig. 3-1-1 Refractometer unit
17
Service Manual of FUS-1000 Urinalysis Hybrid
18
Service Manual of FUS-1000 Urinalysis Hybrid
10
2
3
4
5
7
Fig. 3-3-1 Syringe pump unit
19
Service Manual of FUS-1000 Urinalysis Hybrid
20
Service Manual of FUS-1000 Urinalysis Hybrid
Fig. 3-4-1 Rinsing bath unit
No. Name No. Name
21
Service Manual of FUS-1000 Urinalysis Hybrid
8 2
3
6
4
5
Fig. 3-5-1 Strip-selecting unit
No. Name No. Name
22
Service Manual of FUS-1000 Urinalysis Hybrid
4
2
3
Fig. 3-6-1 Strip-conveying unit
3 Bracket assembly
23
Service Manual of FUS-1000 Urinalysis Hybrid
4
3
Fig. 3-7-1 Probe unit
2 Horizontal resetting optocoupler (inner end) 5 Horizontal resetting optocoupler (outer end)
24
Service Manual of FUS-1000 Urinalysis Hybrid
motion assembly slides horizontally and reaches right above the zero point of dripping, and the probe descends
vertically to the positioning position to start dripping; the probe lifts vertically to the positioning point and slides
horizontally till reaching right above the rinsing bath, then it descends vertically to the positioning point to realize
rinsing purpose; the probe lifts vertically to the positioning point, i.e., the starting position, at this moment, the whole
work process is completed.
25
Service Manual of FUS-1000 Urinalysis Hybrid
1 2 4
7 6 5
Fig. 3-8-1 Sample feeding unit
26
Service Manual of FUS-1000 Urinalysis Hybrid
Sample preparation
Rack-withdrawing assembly
starts to withdraw racks
3.9 Valve group
3.9.1 Diagram of valve group 1
1
3
4
3 Valve plate 1
27
Service Manual of FUS-1000 Urinalysis Hybrid
2
3
4
10 1
9 2
8 3
4
7 5
6
28
Service Manual of FUS-1000 Urinalysis Hybrid
3
4
29
Service Manual of FUS-1000 Urinalysis Hybrid
Refractometer
Sheath
flow
cell
Rinsing
bath
component
Detergent tank
Sheath tank
semi-finished product
Waste liquid
Fig. 4-2-1 Fluid circuit and gas circuit of the analyzer
30
Service Manual of FUS-1000 Urinalysis Hybrid
4.3 Description of work process of the gas circuit and fluid circuit of the Analyzer
4.3.1 Chemistry analysis process
The chemistry test process of the fluid circuit system covers sample mixing, sample aspirating, sample dripping and
rinsing of fluid circuit system.
4.3.1.1 Sample mixing
Tube of refractometer
Refractometer
Test tube
Detergent semi-finished
Detergent tank
product
The pipeline unit realizes the sample mixing through the sample aspirating and dispensing of probe. The probe moves
to the sample aspirating position first and then starts sample aspirating and dispensing and mixing.
(1) The probe moves to sample aspirating position;
(2) Open valve V3, V4, V5, V6;
(3) M1 motor moves downwards and drives the syringe to have quantitative sample aspirating;
(4) M1 motor moves upwards and drives the syringe to have quantitative sample dispensing;
(5) Close valve V3, V4, V5, V6;
31
Service Manual of FUS-1000 Urinalysis Hybrid
Refractometer
Test tube
Detergent semi-finished
Detergent tank
product
The pipeline unit aspirates the sample in test tube to pipeline through the probe. The probe moves to the sample
aspirating position first and then starts sample aspirating after sample mixing.
(1) Open valve V3, V4, V5, V6;
(2) M1 motor moves downwards and drives the syringe to have quantitative sample aspirating;
32
Service Manual of FUS-1000 Urinalysis Hybrid
Refractometer
Strip
Detergent semi-finished
Detergent tank
product
The pipeline unit dispenses the sample to strip through the probe. After sample aspirating, the probe moves to the
sample dripping position. At this moment, the strip has been prepared at the sample dripping position. The probe
drips sample on strips in sequence.
(1) The probe moves to sample dripping position;
(2) The probe moves to the first strip block position, M1 motor moves upwards and drives the syringe to have
quantitative sample dispensing;
(3) The probe moves to the second strip block position, M1 motor moves upwards and drives the syringe to have
quantitative sample dispensing;
(4) The probe moves to the third strip block position, M1 motor moves upwards and drives the syringe to have
quantitative sample dispensing;
(5) The probe finishes the sample dispensing on strips for test of all items according to the strip type and software
setting.
33
Service Manual of FUS-1000 Urinalysis Hybrid
Refractometer
Rinsing
bath
component
Detergent semi-finished
product
product
After the probe finishes sample dispensing to strip block, it moves to the rinsing bath for the rinsing of probe inner
wall and outer wall and tubing.
(1) The probe moves to the rinsing bath;
(2) Open valve V2 and V6, M1 motor moves downwards and the syringe has quantitative sample supplementing;
(3) Close valve V2 and V6, open valve V3, V4 and V5, M1 motor moves upwards for the rinsing of probe inner wall
and tubing;
(4) Close valve V3, V4 and V5, M1 motor moves upwards for the rinsing of probe outer wall;
(5) Start up P1 pump and then shut down it after it aspirates waste liquid for a certain time.
(6) Probe resetting.
34
Service Manual of FUS-1000 Urinalysis Hybrid
Sheath
flow
cell
Rinsing
bath
component
Detergent tank
Sheath tank
semi-finished product
Waste liquid
The fluid circuit system transmits samples to sheath flow cell. In the sheath flow cell, the sample flow and sheath flow
form laminar flow under the pushing of syringe and the optical system will take picture of the laminar flow. The
laminar flow includes quickly pushed laminar flow, stable laminar flow, laminar flow for picture taking and end
laminar flow.
(1) Quickly pushed laminar flow: Open valve V4, M2 motor and M3 motor will quickly move upwards and dispense
the sample and sheath simultaneously to the sheath flow cell;
(2) Stable laminar flow: Keep valve V4 open, M2 motor and M3 motor will slowly move upwards simultaneously
and dispense the sample and sheath simultaneously to the sheath flow cell to form laminar flow;
(3) Laminar flow for picture taking: Keep valve V4 open, M2 motor and M3 motor will slowly move upwards
simultaneously and dispense the sample and sheath simultaneously to the sheath flow cell to form laminar flow and
then the optical system is started up for picture taking;
(4) End laminar flow: Keep valve V4 open, M2 motor and M3 motor will slowly move upwards simultaneously and
dispense the sample and sheath simultaneously to the sheath flow cell to form laminar flow and then the optical
system will be shut down to stop picture taking.
35
Service Manual of FUS-1000 Urinalysis Hybrid
Sheath
flow
cell
Rinsing
bath
component
semi-finished product
Waste liquid
After finishing the chemistry test and morphology test, the fluid circuit system will be rinsed, including the rinsing of
probe inner wall and outer wall, rinsing of pipeline and solenoid valve and rinsing of sheath flow cell.
(1) After the fluid circuit system finishes sample dripping and laminar flow, the probe moves to the rinsing bath;
(2) Forward flushing of sheath flow cell: Open V4 valve and M2 motor will move upwards.
(3) Backflushing of sheath flow cell: Close V4 valve and open V5 valve and M2 motor will move upwards.
(4) Bubble exhausting of sheath flow cell: Open valve V4, V9 and V10 and M2 motor will move upwards.
(5) Rinsing of probe inner wall, tubing and solenoid valve: Close valve V9 and V10, open valve V3, V4 and V5, M1
motor will move upwards.
(6) Rinsing of probe outer wall: Close valve V3, V4 and V5, M1 motor moves upwards for the rinsing of probe outer
wall;
(7) Start up P1 pump and then shut down it after it aspirates waste liquid for a certain time.
(8) Reset probe and syringe motor.
36
Service Manual of FUS-1000 Urinalysis Hybrid
3 2 1
3 2 1
2Green+24 2Green+24 2Green+24
G0106 2Shielded 3WhiteL J5 2Green+24
N0101 L0101 J4 3WhiteL 4 3 Yellow 3Yellow
G0104 4 P0105 J0105
1RedH 1RedH DF1B-4DS-2.5RC control board G0105
green green
DF1B-4DS-2.5RC 2Shielded 2Shielded ground ground
3WhiteL 3WhiteL
N L 4 4
220V power
DF1B-4EP-2.5RC DF1B-4ES-2.5RC Sample
socket feeding unit
Yellow Green
B06-P13 B01-P23
1BlackGND
DC0101+24V Green 2Green+24 1BlackGND
DC0101-GND Black 2Green+24 1BlackGND P0106
3 2 1
2Green+24
3Yellow green
G0104 Yellow Green
37
Service Manual of FUS-1000 Urinalysis Hybrid
1White 1White
2Black 2Black 2Black Twisted pair
3Blue J11 WeldingSW03
NO
2Red+5V
3BlackGND PSW06 JSW06
4Red+5V 1BlackGND 1BlackGND Touch switch/ small
5BlackGND 2Red+5V 2Blue+5V SW06
C
DF1B-2EP-2.5RC
Borrowed FUS2000 J10 8Red+5V DF1B-2ES-2.5RC
9Blue
NC
10White+5V
USOP2 JUS OP2 11BlackGND Twisted pair PSW07 JSW07 Twisted pair
Welding B02-P08 12Red+5V Rec laiming tray full
13Blue 1BlackGND 1BlackGND
1Red White 1BlackGND 2Red+5V 2Red+5V dete ction touc h switch
C
correlation 16Red+5V
DF1B-2ES-2.5RC DF1B-2EP-2.5RC
optocoupler
NC
Fig. 5-3-1 Electrical wiring diagram of sample feeding unit control board
38
Service Manual of FUS-1000 Urinalysis Hybrid
SW02
JSW02 PSW02
DF1B-2ES-2.5RC DF1B-2EP-2.5RC
Twisted pair
B01-P15 PM11 JM11 Syringe pump
STOP Twisted pair 1Black 1Black
1RedOUT1_A- 1Red 1Red
DF1B-6EP-2.5RC DF1B-6ES-2.5RC
V09 2 1
JV09 PV09
1Black Twisted
B01-P29 PPS20 JPS20
1Red bubble optocoupler
1Brown White
2Brown White 2Red pair 1Red 1Red
2Black 2Black Main control unit
2White Four- 3White
V10 3White
Own c able DF1B-2ES-2.5RC DF1B-2EP-2.5RC J29 3Black
4Blue stranded 4Blue 4Green Own bubble optocoupler
2 1
JV10 PV10 Twisted PS20
1Brown White 1Black 1 B01-P19 wire DF1B-4EP-2.5RC DF1B-4ES-2.5RC cable
DF1B-4DS-2.5RC
V07 2Brown White 2Red pair 2
3 2 1 3BlackGND
Own c able DF1B-2ES-2.5RC DF1B-2EP-2.5RC 4Red24V
JV07 PV07 5
1Black 1Black 6
V08 2Red 2Red 7Red24V
3 2 1
DF1B-2ES-2.5RC DF1B-2EP-2.5RC
8BlackGND J19
Own c able Twisted pair 9BlackGND
JV08 PV08 10Red24V
1Black 1Black 11Red24V P0903 J0903 PPS15 JPS15 Probe motor vertical
V05 3 2 1
2Red 2Red DF1B-5EP-2.5RC DF1B-5ES-2.5RC resetting
12BlackGND
13BlackGND
B01-P11 DF1B-4EP-2.5RC DF1B-4ES-2.5RC
1Red
optocoupler
DF1B-2ES-2.5RC DF1B-2EP-2.5RC 1Red 1Red 1Red
Own c able 14Red24V 1Red+5V 2Black 2Black 2Black 2Black
1Black
JV05 PV051Black Twisted pair 2BlackGND PS15
Four- 3White 3White 3White 3White
2Red 2Red DF1B-14DS-2.5RC 3WhiteOPT2 4Blue 4Green 4Green 4Green
4BlueGND stranded Own
Own c able DF1B-2ES-2.5RC DF1B-2EP-2.5RC Twisted pair J11 5Red+5V wire PPS13 JPS13 cable
6BlackGND
7WhiteOPT1 1Red 1Red
2Black 2Black Probe motor horizontal resetting
8BlueGND
V06 3 2 1
DF1B-8DS-2.5RC
Four-
stranded
3White
4Blue
3White
4Green
PS13 optocoupler (inner end)
Probe
Own c able
JV06 PV06 1Black
1Black pair
Twisted wire DF1B-4EP-2.5RC DF1B-4ES-2.5RC
PPS04 JPS04
optocoupler
V04 2Red 2Red
B01-P10
Optical Focusing
unit motor resetting
DF1B-4EP-2.5RC DF1B-4ES-2.5RC
3 2 1
1Red 1Red
Own c able DF1B-2ES-2.5RC DF1B-2EP-2.5RC 1Red+5V 2Black
2BlackGND 2Black PS04
JV04 PV04 B01-P18 3White 3White
V03 1Brown White 1Black Twisted 3WhiteOPT6 Four- 4Green optocoupler
2Brown White 2Red 1Red24V 4BlueGND stranded 4Blue
pair Own c able
3 2 1
Own c able 2BlackGND J10 5Red+5V
wire PPS17 JPS17
DF1B-2ES-2.5RC DF1B-2EP-2.5RC 3Red24V 6BlackGND 1Red 1Red
V02 PV03 1Black 4BlackGND 7WhiteOPT7 2Black 2Black PS17
1Black JV03 5BlackGND 8BlueGND 3White
3 2 1
2Red 2Red
Twisted 6Red24V J18 Four- 4Blue
3White
4Green
7Red24V DF1B-8DS-2.5RC stranded Probe motor horizontal resetting
Own c able DF1B-2ES-2.5RC DF1B-2EP-2.5RC pair 8BlackGND Own c able
PV02 1Black optocoupler (outer end)
V01 3 2 1 1BlackJV02 9BlackGND wire DF1B-4EP-2.5RC DF1B-4ES-2.5RC
2Red 2Red 10Red24V
Twisted PPS19 JPS19
Own c able DF1B-2ES-2.5RC DF1B-2EP-2.5RC
pair DF1B-10DS-2.5RC
1BlackJV01 PV01 1Black B01-P09 DF1B-4EP-2.5RC DF1B-4ES-2.5RC
1Red 1Red
2Red 2Red Twisted 1Red+5V 2Black Sam ple m otor resetting
2Black
Own c able
2BlackGND
Four- 3White 3White PS19 optocoupler
DF1B-2ES-2.5RC DF1B-2EP-2.5RC pair 3WhiteOPT4 4Green
4Blue
4BlueGND stranded Own
B01-P21
New waste +Red 1Red
JP01 PP01
1Red
1Red J9 5Red+5V wire PPS05
1Red
JPS05
1Red cable
Sam ple aspira ting
liquid pump
P1
-Black
2Black 2Black
2Black
3 J21
6BlackGND
7WhiteOPT3
8BlueGND
Four-
2Black
3White
2Black
3White
PS05 motor resetting Syringe
4 optocoupler
P01 DF1B-2ES-2.5RC DF1B-2EP-2.5RC
Twisted pair
DF1B-4DS-2.5RC DF1B-8DS-2.5RC
stranded
wire
4Blue 4Green
Own c able pump
Twisted pair DF1B-4EP-2.5RC DF1B-4ES-2.5RC
PPS14 JPS14 optocoupler
JS03 PS03 B01-P25 B01-P08 DF1B-4EP-2.5RC DF1B-4ES-2.5RC
1Green 1Blue 1Blue 1Red 1Red Sam ple aspirating motor lower
Hydraulic 2Black
3White
2Black
3White Four-
2Black
3White J25
1Red+5V
2BlackGND
3WhiteOPT8 Four-
2Black
3White
2Black
3White
PS14
optocoupler
4Red 4Red 4Red 4Blue 4Green Own
S03 stranded 4BlueGND stranded
Own c able J8 5Red+5V cable
wire DF1B-4DS-2.5RC wire PPS18 JPS18
1Red
6BlackGND
DF1B-4ES-2.5RC DF1B-4EP-2.5RC 7WhiteOPT6 1Red
2Black Sheath motor
8BlueGND 2Black PS18
Four- 3White 3White optocoupler
B01-P22 DF1B-8DS-2.5RC 4Blue 4Green
Own
JFAN01 PFAN01 stranded
1Red 1Red 1Red+24V wire DF1B-4EP-2.5RC DF1B-4ES-2.5RC cable
2Black 2Black 2BlackGND J22
FAN01 Twisted
DF1B-2ES-2.5RC DF1B-2EP-2.5RC
pair
DF1B-4DS-2.5RC
B01-P01
1Red+5V 3Red+5V
2Blue-signal 1Blue-signal
S01
3Black-GND
Thre e-
2Black-GND J1
stranded wire DF1B-4DS-2.5RC
39
Service Manual of FUS-1000 Urinalysis Hybrid
PS19
B12-P07-PS1119 optocoupler
Strip-placing
1RedVCC zero-position
2BlackGND optocoupler
3WhiteOPT5
4GreenGND
J7 5RedVCC
6BlackGND
7WhiteOPT6
8GreenGND Own c able
PPS19 B12-P02 J2 DF1B-8DS-2.5RC
1White 1Blue
Humidity sensor 2Red 2Black Own c able PS18
board
3Blue
4Black 3White optocoupler
Strip-selecting
4Red
zero-position
DF1B-4DS-2.5RC
B12-P08-M1114 optocoupler
DF11-4DS-2C 1Blue
4C*24AWG(30/0.1B)
2Red
J8 3Yellow
4Brown
M14 motor
DF1B-4DS-2.5RC
Own c able
DF1B-4ES-2.5RC DF1B-4EP-2.5RC
JPS08 PPS08
1Red 1Blue
Strip-pushing 2Black 2White
3White 3Red
motor zero-position PS08 4Green 4Black B06-P1-0611S
optocoupler optocoupler 1BlueSLOT1_IN
DF1B-4ES-2.5RC DF1B-4EP-2.5RC
JPS11 PPS11 2WhiteGND
1Red 1Blue 3RedSLOT1_OUT
2Black 2White 4BlackGND Chemistry test unit
Test motor zero PS11
3White 3Red 5BlueSLOT2_IN
position optocoupler optocoupler 4Green 4Black 6WhiteGND
7RedSLOT2_OUT
DF1B-4ES-2.5RC DF1B-4EP-2.5RC
8BlackGND J5 Borrowed
Horizontal motor
9BlueSLOT3_IN
10WhiteGND
J1
1Red 1Blue
PS09 2Black 2White 11RedSLOT3_OUT
zero position optocoupler 3White 3Red 12BlackGND
optocoupler 4Green 4Black 13Red
JPS09 PPS09 15Blue
16Black
1Red
PS10 PS10 strip amount 2Blue
optocoupler judging optocoupler 3Black DF1B-16DS-2.5RC
DF1B-4EP-2.5RC DF1B-4ES-2.5RC
1 B06-P9-0601M 1Blue 1Red
1White 2White 2Blue
2Blue 3Red 3Green
J9 3Red 4Black 4Black 0601 motor Horizontal motor
4Black
PM05 JM05
B06-P6-0701S DF1B-4DS-2.5RC
1Red
2Red White DF1B-4EP-2.5RC DF1B-4ES-2.5RC
IrED01 transmitting tube
3Green
4Green White
J6
B06-P10-0602M 1Blue 1Red
1Blue 2White 2Blue
IrRD01 rece iving tube 2White 3Red 3Green
J10 3Red 4Black 4Black
DF1B-4DS-2.5RC 4Black
PM07 JM07 0602 motor Vertical motor
IrRD01 receiving tube UL3265 DF1B-4DS-2.5RC
DF1B-4EP-2.5RC DF1B-4ES-2.5RC
B06-P11-0603M 1Blue 1Red
1Blue 2White 2Blue
B06-P2-0612S 2White 3Red 3Green
1Blue
J11 3Red 4Black 4Black
4Black
2White PM06 JM06
3Red J2 0603 motor Tes t motor
P0660 DF1B-4DS-2.5RC
4Black
DF1B-4ES-2.5RC DF1B-4EP-2.5RC
Vertical motor zero 1Red 1Blue
position optocoupler 2Black 2White DF1B-4DS-2.5RC
3White 3Red DF1B-4EP-2.5RC DF1B-4ES-2.5RC
4Green 4Black B06-P12-0604M 1Blue 1Red
1White 2White 2Blue
JPS10 PPS10 2Blue 3Red 3Green
J12 3Red 4Black 4Black
0604 motor
Strip-pushing
4Black PM04 JM04 Motor
DF1B-4DS-2.5RC
40
Service Manual of FUS-1000 Urinalysis Hybrid
S02
Model: DS18B20
1Red+5V
3Black-GND
2Blue-signal
J0802
1Red+5V
2Blue-signal
3Black-GND
DF1B-3ES-2.5R
Fan02 60JFan0802
1Red
2Black
Own cable
DF1B-2ES-2.5RC
B10-P02
B 1Black 1Black GND
R 2Red 3Red +5V J2
APCD-850-01-C3-A\ turbidity DF1B-3S-2.5R
laser tube
Turbidity meter light source
APCD-850 board
B10-P04
1Black(Shielded wire) 1Black(Shielded wire)GND
2White 2WhiteBWP
J4 Refr actome ter
BPW24R-photodiode B10-P07
DF1B-3S-2.5R sensor
1RedPSD1 1RedPSD1
2Shielded wire 2Shielded wire
H800-LASER board
B11-P06 J7 3WhitePSD2 3WhitePSD2
DF1B-3S-2.5R PSD
photoelectric
J6 PSD-08-27-H800-05
receiving tube
PSD101
H-800\ refractometer color sensor module
(assembly)
41
Service Manual of FUS-1000 Urinalysis Hybrid
6.1.1.2 Replacement of sample aspirating motor resetting optocoupler on syringe pump unit
42
Service Manual of FUS-1000 Urinalysis Hybrid
(1) As shown in the figure above, remove the pan-head cross screws M3×6 with a cross screwdriver and take out the
sample aspirating motor resetting optocoupler - wire JPS05;
(2) Install a new sample-aspirating motor resetting optocoupler to the original position in a reversed sequence.
6.1.1.3 Replacement method of other syringe and optocoupler is same with that in 6.1.1.1 and 6.1.1.2.
43
Service Manual of FUS-1000 Urinalysis Hybrid
6.1.2.3 Replacement of horizontal resetting optocoupler (inner end) on probe transverse motion assembly
44
Service Manual of FUS-1000 Urinalysis Hybrid
6.1.2.4 Replacement of horizontal resetting optocoupler (outer end) on probe transverse motion assembly
45
Service Manual of FUS-1000 Urinalysis Hybrid
6.1.2.6 Replacement of sample aspirating air bubble optocoupler on probe longitudinal motion assembly
46
Service Manual of FUS-1000 Urinalysis Hybrid
6.1.3 Replacement of rinsing and dyeing unit solenoid valve connector and O-ring
47
Service Manual of FUS-1000 Urinalysis Hybrid
1
4
2 3
(3) Dismantle the strip-selecting unit circuit board cover and pull out the harness of the strip-selecting unit that
connects the machine, as shown in the figure below:
CAN
cable
Power
Cable
48
Service Manual of FUS-1000 Urinalysis Hybrid
1
2
(6) Use an Allen wrench to unscrew the seven M4×8 socket head cap screws that fix the strip-conveying unit, as
shown in the figure below:
1
2
2
(7) Pull out the harness of the strip-conveying unit that connects the machine.
(8) Take off the strip-conveying unit from the instrument at left side, as shown in the figure below:
49
Service Manual of FUS-1000 Urinalysis Hybrid
50
Service Manual of FUS-1000 Urinalysis Hybrid
(3) Use a cross screwdriver (medium) to remove four pan-head cross screws that fix the circuit board and then take
off the circuit board.
(4) Unscrew the two table fixing screws on the strip table manually and then take off the strip table.
(5) Use a cross screwdriver (medium) to remove four cross recessed pan-head screws that fix the transmitting tube
and receiving tube, and take off the transmitting tube and receiving tube.
(6) Use a cross screwdriver (medium) to remove two cross recessed pan-head screws that fix the second-time strip
judging optocoupler, and then take off the second-time strip judging optocoupler.
(7) Install new circuit board, strip table, transmitting tube and receiving tube and second-time strip judging
optocoupler in a reversed sequence. After the assembly of the new strip table, test BD again.
51
Service Manual of FUS-1000 Urinalysis Hybrid
52
Service Manual of FUS-1000 Urinalysis Hybrid
53
Service Manual of FUS-1000 Urinalysis Hybrid
(3) Take off the control cable connected with solenoid valve;
(4) Use a screwdriver (medium) to take off the unit;
(5) Refer to the exploded drawing to remove the parts.
6.1.6.2 Replacement of valve group 2
(1) Dismantle the rear cover and top cover;
(2) Dismantle tube 12, tube 13, tube 24, tube 25 and tube 26 on valve plate 2;
(3) Open the cable clamp to remove the control cable;
(4) Use a screwdriver (medium) to take off the assembly;
(5) Refer to the exploded drawing to remove the parts on valve group 2.
6.1.6.3 Replacement of valve group 3
(1) Dismantle the rear cover;
(2) Dismantle the syringe unit (after pulling out the control cable and relevant tubing);
(3) Use a screwdriver (medium) to dismantle the optocoupler press block;
(4) Dismantle tube 3, tube 5, tube 6, tube 15, tube 16, tube 20, tube 21, tube 17, tube 22 and tube 23 on valve plate 3;
(5) Remove the control cable;
(6) Use a screwdriver (medium) to take off the assembly;
(7) Refer to the exploded drawing to remove the parts on valve group 3.
1
54
Service Manual of FUS-1000 Urinalysis Hybrid
55
Service Manual of FUS-1000 Urinalysis Hybrid
1
1
2
56
Service Manual of FUS-1000 Urinalysis Hybrid
1
2 3
57
Service Manual of FUS-1000 Urinalysis Hybrid
58
Service Manual of FUS-1000 Urinalysis Hybrid
Rack-feeding assembly
Stepper motor
59
Service Manual of FUS-1000 Urinalysis Hybrid
Rack-conveying
assembly
60
Service Manual of FUS-1000 Urinalysis Hybrid
Shifting fork
Timing pulley
Groove
optocoupler
Motor
61
Service Manual of FUS-1000 Urinalysis Hybrid
Rack-withdrawing
assembly
Swinging fork
Timing pulley
Groove optocoupler
Stepper motor
62
Service Manual of FUS-1000 Urinalysis Hybrid
Attention
Please refer to Compatibility Configuration Table of FUS-1000 Software Version Information to confirm the
compatibility between each program version to be upgraded; if a version is not compatible, it may cause the
equipment to be disabled.
When updating the program, the equipment should be in standby/stop status. If the equipment is in operation, it can
be stopped by pressing the "Stop" button on the equipment panel.
Notes: when the multiprograms need to be upgraded, it is recommended to first upgrade the unit program,
and then upgrade the master control program.
Program upgrading steps:
(1) Run the "FUS-1000 debug software" program and select the "lower computer upgrading" button in the title bar;
(2) Then, "lower computer upgrading software of DIRUI" will pop up and please click the "Exit" of FUS-1000
program;
(3) First choose the "Serial port No." for communication with the instrument (the default communication port of
instrument is COM1), choose 115200 baud rate, and then click "Open the serial port", and then the triangle logo
indicating position becomes red, and then click the "Add files" button;
(4) Then, the following dialog box pops up, and choose the file to be upgraded (with the suffix of .s19) to open;
(5) Then the following interface pops up, and then check the corresponding box of the triangular symbol and click the
"Start upgrading";
(6) Then the following dialog box pops up, click "Yes", and then the program starts upgrading;
(7) The progress bar renews continuously during program upgrading.
(8) After the completion of upgrading, the upgrading result window pops up to prompt that the upgrading is
completed, then click "OK" and the upgrading is completed. Click the "Remove files"; for upgrading other programs,
repeat the step (3)-(6).
63
Service Manual of FUS-1000 Urinalysis Hybrid
● Don't spill water, reagent, detergent or other solutions on the machine or electric parts of the Analyzer in
case of any damages.
● Don’t touch probe during operation of the Analyzer in case of infection or injury.
● During operation, the operators shall take preventive measures like wear protective gloves, glasses, working
suits in case of infection due to touch of polluted area or solution or skin injury due to contact with corrosive
liquid. The operator shall rinse with water and take disinfection measures after contact with the polluted or
corrosive liquid due to carelessness.
● If the Analyzer is installed with accessories which are not provided or recommended by the manufacturer
or the Analyzer is used otherwise specified by the manufacturer, the relating protection may be weakened.
● During maintenance process, please check whether there are hazards caused by inefficiency of hoses or
parts filled with solution.
Click in the main menu area and enter in the system maintenance interface as shown below:
64
Service Manual of FUS-1000 Urinalysis Hybrid
Fig. 7-2-1
● It is not allowed to suddenly suspend the reset process. Other operation is allowable only after standby
mode is displayed.
● Please execute reset if the alarm information give prompts or the Analyzer has an emergency stop.
In case of start blank test failure or after maintenance of the Analyzer, click on the interface (Fig.
7-2-1) to conduct blank test.
The screen will display “Blank test success” as long as the blank test is acceptable.
The screen will display “Blank test failed” if it does not pass the blank test. “Fill whole pipeline” (see 7.2.4) and
“Rinse flow cell” (see 7.2.5) should be carried out repeatedly. If still failed, please contact customer service personnel
of Dirui Company.
Click on the blank test list and the interface is as shown below:
Fig. 7-2-2
65
Service Manual of FUS-1000 Urinalysis Hybrid
Select the operator, blank test result, and date from the pull-down menu, then click and the print
preview report will pop up.
7.2.3 Empty pipeline
Conduct “Empty Pipeline” before removal of the Analyzer. Click on the interface (Fig. 7-2-1) and
pipelines of the Analyzer are emptied.
Please disassemble the Analyzer into a host and a sample-feeding unit first before transportation.
Conduct “Empty pipeline” before stopping the Analyzer and rinse pipelines with distilled water (replace the sheath
solution in the sheath bottle by distilled water), and finally conduct “Empty pipeline” again.
The solution in sheath bottle should be emptied before conducting “Empty pipeline”.
In case of bubbles in tubing or during replacement of sheath solution, click on the interface (Fig.
7-2-1), and the bubbles in tubing of the Analyzer will be emptied, which can avoid the impact of test results caused by
bubbles.
Before reuse of the Analyzer after idling for a period of time, conduct “Fill whole pipeline”.
When there are bubbles in the pipelines or replacing sheath liquid, click on the interface (Fig.
7-2-1); the bubbles in each pipeline will be exhausted, avoiding test results being affected by bubbles.
When there are bubbles in the pipelines or replacing detergent (concentrated type), click on the
interface (Fig. 7-2-1); the bubbles in each pipeline will be exhausted, avoiding test results being affected by
bubbles.
66
Service Manual of FUS-1000 Urinalysis Hybrid
Fig. 7-2-3
Click on the interface (Fig. 7-2-1) and the tube rack will be pushed to the testing position
automatically, then the probe aspirates about 2mL detergent; clean the flow cell after soaking the flow cell for about
20min (follow the screen tips).
7.2.8 Cleaning of worktable
Click on the interface (Fig. 7-2-1), and the Analyzer enters in the cleaning state. Open the left front
door of the Analyzer, the testing head will run to the side away from the operator, and the stepping gear plate will rise
to the highest point automatically. Clean the worktable and stepping gear plate with absorbent cotton soaked in clean
water to clear all impurities on them, and after that wipe to dry the plates.
● Don’t wipe the white benchmark with absorbent cotton dipped in alcohol in case of color change or scratch
of the benchmark and calibration failure of the Analyzer.
● Don’t use hard brush to clean the surface of the worktable to avoid impact on its service life!
Click on the interface (Fig. 7-2-1) and the probe will be pushed to the emergency position and
aspirates the detergent for 3 times and 2mL for each time to clean the refractometer and turbidimeter.
7.2.10 Rinse tubing
To avoid the impact of residual in flow cell on the accuracy of testing results of the Analyzer, the tubing shall be
cleaned every week.
Click “Rinse tubing” in the “System maintenance” interface, and the prompt box as below will pop up:
Fig. 7-2-4
67
Service Manual of FUS-1000 Urinalysis Hybrid
Load the test tube with around 5mL detergent and put it in the first position of the test tube rack, then put the rack on
the right side of the sample-feeding unit. Click “OK” and the tip “Preparing detergent...” will pop up; then the test
tube will be pushed to the sampling position automatically and the sample-aspirating probe will aspirate about 2mL
detergent, then the hint “About 20 min is necessary for soaking” will pop up.
After cleaning, the system is at the standby state.
Fig. 7-3-1
To avoid the dust on the surface of or inside the rinsing bath that may affect the accuracy of Analyzer, the rinsing bath
should be cleaned every week.
Rinsing method: clean the inner wall of the rinsing bath with swab dipped in alcohol and at the same time wipe the
upper surface of the rinsing bath as shown below.
Fig. 7-3-2
The drainage system shall be in compliance with the local regulations with regard to sewage discharge of
medical institutions.
68
Service Manual of FUS-1000 Urinalysis Hybrid
Fig. 7-5-1
Step 4: Select the “inie” mode through the [+ -] button as the figure below shows, and then press the “MODE”
button.
Fig. 7-5-2
Step 5: When the screen displays red and green numbers, put a test tube rack in the rack conveying area, put a clean
glass test tube without barcode in the tube judging position, and press the [TUNE] button. The screen display is as
shown:
Fig. 7-5-3
Remove the test tube, and press the “TUNE” button. The screen display is as shown in Figure 7-5-4. If the base line
value (green) is not in the given range, press the [+ -] button to adjust to 6000±1000.
Fig. 7-5-4
Step 6: Press the “L/D” button, and change the work mode to the “D” mode.
Step 7: Close the upper cover of the fiber sensor.
Success standard: when placing a test tube in the tube judging position, the indicator lamp on the circuit board
burns; when removing the test tube, the indicator lamp on the circuit board does not burn.
69
Service Manual of FUS-1000 Urinalysis Hybrid
Yellow
indicating lamp
SET key
Fig. 7-5-5
If the problem cannot be solved, please contact the Customer Service Department or the agent of Dirui Company.
Fig. 7-6-1
Fig. 7-7-1
70
Service Manual of FUS-1000 Urinalysis Hybrid
● The barcode reader of Analyzer belongs to Class 2 laser product. Users shall not stare at the laser beam
according to the alarm mark and protect themselves by turning head around or closing eyes.
● Users shall not look at the laser beam constantly in case of damages to their eyes.
● Turn off the Analyzer before wiping the barcode reader.
Embedded handle
Fig. 7-8-1
Fig. 7-8-2
Turn the manual knob of motor in anticlockwise direction to let the slide table adapter board move left till the end as
shown below:
71
Service Manual of FUS-1000 Urinalysis Hybrid
Fig. 7-8-3
Fig. 7-8-4
Wipe the front side of object lens with lens paper, and install it to the original position, as shown below.
Fig. 7-8-5
After wiping, put down the support rod and close the back door of the Analyzer.
Fig. 7-9-1
72
Service Manual of FUS-1000 Urinalysis Hybrid
(2)Firstly open the left front door of the Analyzer, remove the tube connectors on the two sides of the sample filter:
use one hand to pinch the sample filter, use the other hand to rotate the tube connector until it is separated from the
filter.
Fig. 7-9-2
(3)Put a piece of absorbent paper flat on the table, take out one filter net removing tool from the accompanying
accessories bag, take out the sample filter net from the part clamp, remove sample filter end A, and insert the
sample filter removing tool into the hole of sample filter end B to push out the sample filter net and make it fall
onto the absorbent paper, as the figure shows:
Fig. 7-9-3
(4)Take a new sample filter net out from the accompanying accessories bag, and install the new sample filter net to
the sample filter. (Sample filter end B should be kept upright to ensure the sample filter net is completely placed in
the groove. Then install sample filter end A to end B.) Put the sample filter assembly in the part clamp, and connect
the tubing connector to sample filter two ends, as the figure shows:
Fig. 7-9-4
73
Service Manual of FUS-1000 Urinalysis Hybrid
(5)Click the [OK] button on the popup box on the interface of the host computer, and close the left front door to
finish the replacement of the sample filter net.
● When installing the sample filter net, keep sample filter net end B upright to make the filter net located in
the groove. Then install filter net end A to end B. Otherwise the sample filter net will fall off sample filter
end B groove, and get damaged.
● Sample filter end A and sample filter end B should be installed firmly. Otherwise air leak may affect the
test result accuracy.
● To ensure test result accuracy, the sample filter net should be replaced according to the software prompt.
Fig. 7-10-1
Fig. 7-10-2
(3)Take the strip holder out, and place it on the clean, flat worktable.
Be careful not to make the strip holder fall down or crash to avoid deformation or break.
(4)Use absorbent cotton dipped in clear water to clean the strip holder and tooth plate, and ensure that there is no
accumulative matter on the strip holder and tooth plate(clean the edges and corners of the strip holder carefully).
Then wipe dry (no cotton fiber or other foreign matter is left on the edges and corners of the strip holder after
wiping dry).
74
Service Manual of FUS-1000 Urinalysis Hybrid
Fig. 7-10-3
(5)Use a brush to clear the scraps on the strip receiving frame of the strip-selecting unit.
(6)Open the cover of the strip sealed bin, take out the remaining strips from the bin, use the brush to clean the inside of
the strip bin, put the strips back after ensuring that the roll is clean, and close the cover of the strip bin well.
(7)Install the cleaned strip holder to the Analyzer according to the correct order (the front end face of the strip holder is
close to the latter 2 fixing screws of the strip holder).
(8)Slowly turn the strip-selecting unit to the initial position, use the removed 2 screws to tighten the fixing plate to the
strip-selecting support plate, close the right small door, and left front door and right front door of the Analyzer.
Click in the main key area and enter in system maintenance interface, then click
to display general information, as shown below.
Fig. 7-12-1
75
Service Manual of FUS-1000 Urinalysis Hybrid
It is recommended to back up the database regularly in case of accidental data loss. Conduct backup and
recovery of database in standby state.
(1)Database backup
Click on the interface (Fig. 7-12-1), and the interface as shown will be displayed.
Fig. 7-13-1
Fig. 7-13-2
76
Service Manual of FUS-1000 Urinalysis Hybrid
Select log path of backup documents (in the disks other than disk C), otherwise the backup documents will be stored
in subfile “backup” of the default installation folder of the software. The file name of database backup is the current
date, and the suffix is *.bak.
After selecting the log path of documents, click in Fig. 7-13-1, and the dialog box below will pop up:
Fig. 7-13-3
It is allowed to tick “After the backup delete the history data” in the dialog box to select whether to delete the history
data.
Click in the dialog box (Fig.7-13-3) and the screen will display progress bar “backup is in progress”;
after backup, the dialog box below will pop up:
Fig. 7-13-4
Click on the interface (Fig. 7-13-1), select the log path of backup documents, and then select
backup document as per the backup date and time.
Click on the interface (Fig. 7-13-1) and the screen will display progress bar “backup is in progress”;
after backup, the dialog box below will pop up:
Fig. 7-13-5
77
Service Manual of FUS-1000 Urinalysis Hybrid
Click , and the database recovery is finished. After database recovery, the query and printing of data
are available.
Take care that the recovered database will cover the current data.
Slipway
pushing
direction
Slipway
Fig. 7-14-1
Fig. 7-14-2
(2)Use a cotton swab to clean up the outer wall of the flow cell (the side far from the object lens, as shown in Fig.
7-14-2).
78
Service Manual of FUS-1000 Urinalysis Hybrid
(4)Avoid vibration during carrying, and check and adjust the Analyzer after carrying to make sure it is normal before
use.
Please handle the wastes of the Analyzer to prevent potential biological and chemical pollution.
79
Service Manual of FUS-1000 Urinalysis Hybrid
80
Service Manual of FUS-1000 Urinalysis Hybrid
81
Service Manual of FUS-1000 Urinalysis Hybrid
82
Service Manual of FUS-1000 Urinalysis Hybrid
02-08-01-0e Device not reset Test sent when the device is not reset Incorrect operation process, which is not a device fault
1. Hardware abnormality alarm will pop up if the hardware fault is not removed.
Abnormal hardware of system or
02-08-01-0f Hardware fault of lower computer Check the device log or determine the fault position according to the alarm code
abnormal stop of last process
from the upper computer and then conduct repair according to prompt.
1. If it is not caused by viscous sample, check if the sample probe is blocked, if the
Viscous sample causes slow sample sample filter is blocked, if the pressure sensor is well connected and if the output
02-08-01-12 Sample aspirating timeout
aspirating speed. voltage is normal.
2. Check the device operation log and find out the timeout position
Slow operating speed of sample
02-08-01-13 Chemistry test timeout aspirating pump and sample Check the device operation log and find out the timeout position
dispensing probe
Too small syringe pump speed in
02-08-01-15 Sheath flow cell rinsing timeout Check the device operation log and find out the timeout position
configuration file
Too small configured syringe pump
02-08-01-18 Sediment test timeout speed and too large laminar flow Check the device operation log and find out the timeout position
sample volume
83
Service Manual of FUS-1000 Urinalysis Hybrid
84
Service Manual of FUS-1000 Urinalysis Hybrid
85
Service Manual of FUS-1000 Urinalysis Hybrid
Probe unit horizontal motor SPI Poor soldering or damage of motor 1. Have repair soldering or replacement of 30521 chip at U8 position
03-08-07-00
communication failure driven chip U8 2. Replace the main control circuit board if problems above are not solved
1. Ensure the B01-P13 and JM08 harness of probe horizontal motor are securely
1. Loosen harness of horizontal motor and reliably connected
Probe unit horizontal motor X coil 2. Probe horizontal motor damaged 2. Connect a motor of same model on circuit board and click the Forward operation
03-08-07-01
open circuit 3. Circuit board motor driving system of horizontal motor in “Fus1000 probe” column of debug software. If the motor
fault rotates, it means the circuit board has no fault and the motor is damaged, so the
motor shall be replaced. Or, the main control circuit board shall be replaced.
86
Service Manual of FUS-1000 Urinalysis Hybrid
87
Service Manual of FUS-1000 Urinalysis Hybrid
88
Service Manual of FUS-1000 Urinalysis Hybrid
89
Service Manual of FUS-1000 Urinalysis Hybrid
90
Service Manual of FUS-1000 Urinalysis Hybrid
If it alarms during testing, check the air bubble optocoupler in sample aspirating
03-00-15-01 Sample aspirates air or blood sample Use empty test tube for test
tubing and check if its wiring is in good condition.
91
Service Manual of FUS-1000 Urinalysis Hybrid
M1 sheath motor SPI Poor soldering or damage of motor 1. Have repair soldering or replacement of 30521 chip at U11 position
04-08-03-00
communication failure driven chip U11 2. Replace the main control circuit board if problems above are not solved
1. Ensure the B01-P16 and JM10 harness of sheath motor are securely and reliably
connected
1. Loose harness of sheath motor
2. Connect a motor of same model on circuit board and click the M1 upward
2. Sheath motor is damaged
04-08-03-01 M1 sheath motor X coil open circuit operation in “Fus1000 syringe” column of debug software. If the motor rotates, it
3. Circuit board motor driving system
means the circuit board has no fault and M1 syringe pump is damaged, so the
fault
sheath syringe pump shall be replaced. Or, the main control circuit board shall be
replaced.
1. Ensure the B01-P16 and JM10 harness of sheath motor are securely and reliably
connected
1. Loose harness of sheath motor
2. Connect a motor of same model on circuit board and click the M1 upward
2. Sheath motor is damaged
04-08-03-02 M1 sheath motor Y coil open circuit operation in “Fus1000 syringe” column of debug software. If the motor rotates, it
3. Circuit board motor driving system
means the circuit board has no fault and M1 syringe pump is damaged, so the
fault
sheath syringe pump shall be replaced. Or, the main control circuit board shall be
replaced.
1. Loose harness of sheath motor
1. Replace with a syringe pump motor of same model
M1 sheath motor charge pump 2. Sheath motor is damaged
04-08-03-03 2. If the problem is not solved by following step 1 above, replace the main control
abnormality 3. Circuit board motor driving system
circuit board
fault
1. Loose harness of sheath motor
1. Replace with a syringe pump motor of same model
2. Sheath motor is damaged
04-08-03-04 M1 sheath motor overheat 2. If the problem is not solved by following step 1 above, replace the main control
3. Circuit board motor driving system
circuit board
fault
1. Loose harness of sheath motor
1. Replace with a syringe pump motor of same model
2. Sheath motor is damaged
04-08-03-05 M1 sheath motor X coil overcurrent 2. If the problem is not solved by following step 1 above, replace the main control
3. Circuit board motor driving system
circuit board
fault
1. Loose harness of sheath motor
1. Replace with a syringe pump motor of same model
2. Sheath motor is damaged
04-08-03-06 M1 sheath motor Y coil overcurrent 2. If the problem is not solved by following step 1 above, replace the main control
3. Circuit board motor driving system
circuit board
fault
92
Service Manual of FUS-1000 Urinalysis Hybrid
93
Service Manual of FUS-1000 Urinalysis Hybrid
94
Service Manual of FUS-1000 Urinalysis Hybrid
95
Service Manual of FUS-1000 Urinalysis Hybrid
96
Service Manual of FUS-1000 Urinalysis Hybrid
97
Service Manual of FUS-1000 Urinalysis Hybrid
98
Service Manual of FUS-1000 Urinalysis Hybrid
99
Service Manual of FUS-1000 Urinalysis Hybrid
100
Service Manual of FUS-1000 Urinalysis Hybrid
101
Service Manual of FUS-1000 Urinalysis Hybrid
102
Service Manual of FUS-1000 Urinalysis Hybrid
103
Service Manual of FUS-1000 Urinalysis Hybrid
104
Service Manual of FUS-1000 Urinalysis Hybrid
105
Service Manual of FUS-1000 Urinalysis Hybrid
106
Service Manual of FUS-1000 Urinalysis Hybrid
107
Service Manual of FUS-1000 Urinalysis Hybrid
108
Service Manual of FUS-1000 Urinalysis Hybrid
109
Service Manual of FUS-1000 Urinalysis Hybrid
110
Service Manual of FUS-1000 Urinalysis Hybrid
1. Laser burned 1. During the test, check if the laser is on. If not, replace the laser.
0A-04-03-01 Specific gravity laser error
2. Sensor damaged 2. Replace the sensor
1. Laser burned 1. During the test, check if the laser is on. If not, replace the laser.
0A-04-04-02 Turbidity laser error
2. Sensor damaged 2. Replace the sensor
1. During the test, check if the laser lamp source is on. If not, replace the color
1. Color lamp burned
0A-04-05-03 Color sensor error lamp.
2. Sensor damaged
2. Replace the sensor
1.Conductivity measuring chip
1. Replace the circuit board
0A-04-06-04 Conductivity error burned
2. Check if the wire connecting the electrode is normal
2. Software BUG
1. Connecting cable of temperature
1. Check J9 connecting cable
0A-04-06-05 Temperature sensor error sensor not connected securely
2. Replace the temperature sensor
2. Temperature sensor damaged
111
Service Manual of FUS-1000 Urinalysis Hybrid
112
Service Manual of FUS-1000 Urinalysis Hybrid
113
Service Manual of FUS-1000 Urinalysis Hybrid
114
Service Manual of FUS-1000 Urinalysis Hybrid
115
Service Manual of FUS-1000 Urinalysis Hybrid
116
Service Manual of FUS-1000 Urinalysis Hybrid
117
Service Manual of FUS-1000 Urinalysis Hybrid
118
Service Manual of FUS-1000 Urinalysis Hybrid
119
Chapter 9 List of Spare Parts
9.1 List of Easily-worn Parts and Consumables
9.1.1 List of Easily-worn Parts
No. Name SAP Code Remarks
9.1.2 List of Consumables
No. Name SAP Code Replacement Period
Prompt for replacement (5
1 Sample Filter 1024330
accessories)
2 Strip Table 1025227 2 years
120
9.3 List of Frequently-used Spare Parts
9.3.1 List of Assembly Spare Parts
No. Name SAP Code Remarks
121
9.3.4 List of Sensor Spare Parts
No. Name SAP Code Remarks
122
9.3.6 List of Timing Belt Spare Parts
No. Name SAP Code Remarks
Strip-selecting unit reversing roll
1 Timing Belt 1008959
motor
2 Timing Belt 1008951 Strip-selecting unit main roll
Strip-conveying unit
3 Timing Belt 1026930
strip-conveying assembly
4 Timing Belt 1008960 Strip-conveying unit test assembly
Strip-conveying strip-pushing
5 Timing Belt 1032127
assembly
Sample feeding unit rack-conveying
6 Timing Belt 1008966
assembly
Sample feeding unit
7 Timing Belt 1008965
rack-withdrawing assembly
Sample feeding unit rack-feeding
8 Timing Belt 1008967
assembly
9 Timing Belt 1025267 Probe unit horizontal mechanism
123
9.3.8 List of Other Spare Parts
No. Name SAP Code Remarks
8 Filter 1026914
9 Fuse 1013295
124
9.4 Unit Exploded Drawing and Spare Parts List
9.4.1 Rack Body Assembly (FUS-1000-01-00)
SAP Code: ——
5 Buffer 1008137
7 Nut 1008657 M4
9 Gasket 1007367
23 Gasket 1007367
125
FUS-1000- 01- 00
45
44 1
43
42 2
41
40 3
39
4
38
37 5
36
35 6
34 7
8
33 9
32
31 10
30 11
29 12
13
28 14
27 15
26 16
17
18
25
19
24
20
23
22 21
126
9.4.1 Rack Body Assembly (FUS-1000-01-00)
SAP Code: ——
28 Magnet 1018581
37 Filter 1026914
127
FUS-1000- 01- 00
45
44 1
43
42 2
41
40 3
39
4
38
37 5
36
35 6
34 7
8
33 9
32
31 10
30 11
29 12
13
28 14
27 15
26 16
17
18
25
19
24
20
23
22 21
128
9.4.2 Morphological Optical Unit (FUS-1000-02-00)
SAP Code: 2009358
13 Fan 2012370
129
FUS-1000-02-00
1
15 2
3
16 4
5
17 6
18 7
19 8
20 9
10
11
21 12
13
14
130
9.4.3 Syringe Pump Unit
9.4.3.1 Syringe Pump Unit (FUS-1000-06-00)
SAP Code: 2009361
131
FUS --1000-06-00
- -
10
2
3
4
5
132
9.4.3.2 Pump Body Assembly 3 (FUS-1000-06-03-00)
SAP Code: ——
133
FUS -1000-06-03-00
- - -
134
9.4.3.3 Pump Body Assembly 2 (FUS-1000-06-02-00)
SAP Code: ——
135
FUS -1000-06-02-00
- - -
136
9.4.3.4 Pump Body Assembly 1 (FUS-1000-06-01-00)
SAP Code: ——
137
FUS -1000-06-01-00
- - -
2
3
138
9.4.4 Rinsing Bath Unit (FUS-1000-09-00)
SAP Code: 2009363
7 O-ring 1008706
139
FUS --1000-09
- --00
1
2
4
12
5
11
10
9 7
8
140
9.4.5 Strip-selecting Unit
9.4.5.1 Strip-selecting Unit Frontside (GMD-S600-05-00)
SAP Code: 2008999
141
GMD-S600-05- 00
3
10
9
9
4
8
7 6
142
9.4.5.2 Strip-selecting Unit Backside (GMD-S600-05-00)
SAP Code: 2008999
143
GMD-S600-05- 00
11
15 12
14 13
144
9.4.6 Strip-conveying Unit
9.4.6.1 Strip-conveying Unit (FUS-1000-03-00)
SAP Code: 2009359
9 Bracket Assembly ——
145
FUS-1000- 03- 00
12
1
2
4
5
11
6
7 10
146
9.4.6.2 Carrier Rack Assembly (FUS-1000-03-04-00)
SAP Code: ——
2 Optocoupler 1013338
147
F US-1000-03-04-00
148
9.4.6.3 Strip-conveying Assembly (FUS-1000-03-03-00)
SAP Code: ——
1 Optocoupler 1013338
5 Optocoupler 1013338
149
F US -1000-03 -03-00
2
5
150
9.4.6.4 Strip-pushing Assembly (FUS-1000-03-02-00)
SAP Code: ——
3 Optocoupler 1013338
151
FUS -1000 - 03- 02 - 00
3
2
152
9.4.7 Probe Unit
9.4.7.1 Probe Unit Assembly (FUS-1000-05-00)
SAP Code: 2009360
153
FUS-1000-05-00
154
9.4.7.2 Probe Transverse Motion Assembly (FUS-1000-05-02-00)
SAP Code: 2012066
1 Bearing 1008856
155
FUS-1000-05-02-00
5
1
2 6
156
9.4.7.3 Probe Longitudinal Motion Assembly (FUS-1000-05-01-00)
SAP Code: 2012065
4 Bearing 1008856
157
FUS-1000-05-01-00
5
1
2
6
7
3
158
9.4.8 Sample Feeding Unit
9.4.8.1 Sample Feeding Unit Assembly (FUS-1000-04-00)
SAP Code: ——
159
FUS-1000-04-00
7
3
160
9.4.8.2 Baseplate Assembly (FUS-1000-04-01-00)
SAP Code: ——
161
FUS-1000-04-01-00
162
9.4.8.3 Bar Code Reader Assembly (FUS-1000-04-04-00)
SAP Code: ——
163
FUS-1000-04-04-00
164
9.4.8.4 Rack-feeding Assembly (FUS-1000-04-05-00)
SAP Code: ——
1 Microswitch 2012223
4 Motor 2012221
165
FUS-1000-04-05-00
166
9.4.8.5 Working Panel Assembly (FUS-1000-04-06-00)
SAP Code: ——
1 Microswitch 2012225
2 Microswitch 2012224
6 Microswitch 2012226
167
FUS-1000-04-06-00
1 6
168
9.4.8.6 Rack-conveying Assembly (FUS-1000-04-03-00)
SAP Code: ——
169
FUS-1000-04-03-00
4
1
170
9.4.8.7 Rack-withdrawing Assembly (FUS-1000-04-02-00)
SAP Code: ——
171
FUS-1000-04-02-00
172
9.5 Liquid Circuit Diagram and Spare Parts List
9.5.1 List of Tube Connector and Filter Parts
1 F1 Sample Filter ——
3 F3 Clamp 1008144
4 F4 T-Connector 1009599
8 F8 Y-Connector 1009549
9 F9 Reducer 1009588
173
4-01 V4-01 Wire JV04-SV04 1030077
174
8 08 Tube 8 1040723 FUS-1000-FD04-120
175
9.5.5 List of Other Electrical Parts and Functional Parts
176
177
9.6 Accessories Case Spare Parts List
No. Name SAP Code Remarks
5 Atlas 1040507
7 Disk 1011784
8 CD Bag 1011783
178
Service Manual of FUS-1000 Urinalysis Hybrid
Date of Chapter of
No. Revision No. Contents of Revision
Revision Revision
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
179