Professional Documents
Culture Documents
RU
I
IN
DU
ST
RIA
L
CO
.,
LT
D
The user manual is applicable to Automatic Hematology Analyzer (model: BF-6900/ BF-6960) and all figures
concerned are described based on the Automatic Hematology Analyzer (model: BF-6960, hereinafter
referred to as the Analyzer).
DESCRIPTION
Dear users, thanks for buying the Automatic Hematology Analyzer (model: BF-6900/ BF-6960).
Please read the manual carefully before operation as incorrect operation may affect the test results of the Analyzer or
cause personal injury.
After reading, please reserve the manual properly for reference at any time.
Manufacturer Address:
D
LT
95 Yunhe Street, New & High Tech. Development Zone, Changchun, Jilin 130012, the People’s Republic of China
.,
Production Address:
CO
3333 Yiju Road, New & High Tech. Development Zone Changchun, Jilin 130103, the People’s Republic of China
AL
Website: http://www.dirui.com.cn
DU
E-mail: dirui@dirui.com.cn
IN
Fax: 0431-85173354
DI
● The Analyzer shall be used by professional medical examination personnel or trained doctors, nurses or testers.
● As the Analyzer has biological and chemical risks, the operator shall be trained and use personal protective
appliance to reduce the risk.
● Only trained operators are allowed to conduct dangerous operations, such as moving parts.
● The hospital or inspection institution shall prepare a service plan and carry out servicing and maintenance in strict
accordance with the service plan, or the Analyzer may have faults.
● The Analyzer shall be controlled with a special software designated by the company. Installation of other software
or hardware on the computer may affect the normal operation of the Analyzer. Please do not operate other software
D
during the operation of the Analyzer.
LT
● Please refer to the instructions of reagent for the use and storage of the reagent and ensure the reagent is used
.,
within the life of the reagent shown in the instructions.
CO
● Do not use reagent beyond its life. After its seal is damaged, prevent dust, dirty matters or bacteria entering it.
AL
● Do not use any organic solvents such as turpentine and benzene to rinse the outer part of the Analyzer as it may
I
cause the color or shape changes of the Analyzer. Use soft or wet cloth to clean the Analyzer and use diluted
TR
● Under an environment with low transportation or storage temperature or relative humidity greater than 70%, the
IN
Analyzer shall be turned on for testing only after it is stored in a normal working environment for 24 hours.
● The Analyzer shall be provided with an independent power source. If it shares one power socket with other
I
RU
electrical equipment, the electromagnetic interference may affect the test results of the Analyzer.
DI
● Do not pull or insert the plug with wet hands as it may cause electric shock.
● Damaged power cable and connection cable shall not be used. The power cables and wires shall not be trodden,
twisted or pulled as it may cause a fire.
● The input voltage shall meet the requirements of the Analyzer and the fuse of specified specification shall be used.
● Ensure the switch of the Analyzer is at the [O] position before connecting the power cable.
● Do not touch the moving components when the Analyzer is operating to prevent accidents.
● When the power for the Analyzer is connected, servicing personnel not authorized shall not open the left and right
doors and upper cover.
● Please use the Analyzer under conditions regulated in the manual. If not, the Analyzer may not operate normally,
the test results may not be reliable, the components of the Analyzer may be damaged and personal injuries may be
caused.
● The protection measures provided for the Analyzer may become invalid if the Analyzer is not used according to
the manual.
D
LT
WARNING
.,
CO
● The operator is obligated to follow national and local regulations on discharge and treatment of expired reagent,
AL
● The waste liquid and the consumables shall be correctly treated. As the blood in the waste liquid may be polluted
TR
by pathogen, please treat the waste liquid and the consumables of the Analyzer according to regulations about
S
DU
● Do not touch the sampling probe as the blood sample, quality control object and calibration object on the sampling
probe have potential biological infectivity. During the sample aspirating process of the sampling probe, prevent
I
touching the test tube wall and probe tip as it may cause bleeding. Besides, a certain distance shall be kept from the
RU
aspiration probe tip to the bottom of container, or the accuracy of the aspirated liquid volume may be affected.
DI
Dirui Company declares that it will be responsible for the safety, reliability and performances of the Analyzer only
if all following requirements are met.
(1)The installation, commissioning and servicing of the Analyzer are undertaken by professional personnel of Dirui
Company.
No further notice will be provided in case of any changes to the software interface.
D
LT
.,
CO
I AL
S TR
DU
IN
I
RU
DI
User Manual
Contents
D
1.5.1 Front view of the Analyzer .............................................................................................................................................. 1-10
LT
1.5.2 Rear view of the Analyzer ............................................................................................................................................... 1-10
1.5.3 Left view of the Analyzer ................................................................................................................................................ 1-11
.,
1.6 External device of the Analyzer................................................................................................................... 1-11
CO
1.7 Symbol ........................................................................................................................................................... 1-11
1.8 Identification ................................................................................................................................................. 1-13
AL
I
User Manual
D
4.2 Calibration frequency .................................................................................................................................... 4-1
LT
4.3 Calibration method......................................................................................................................................... 4-1
4.4 Calibrator calibration .................................................................................................................................... 4-1
.,
4.4.1 Input of calibrator reference value..................................................................................................................................... 4-2
CO
4.4.2 Calibration counting .......................................................................................................................................................... 4-2
4.4.3 Save calibration coefficient ............................................................................................................................................... 4-3
AL
II
User Manual
D
7.6.2 View diagram .................................................................................................................................................................... 7-6
LT
7.6.3 View of research parameters ............................................................................................................................................. 7-7
7.7 Analysis of pre-dilution samples.................................................................................................................... 7-8
.,
7.8 Sample analysis under automatic whole-blood mode .................................................................................. 7-9
CO
7.8.1 Bar code and pasting requirements .................................................................................................................................... 7-9
7.8.2 Change of mode ................................................................................................................................................................ 7-9
AL
7.8.3 Sample analysis steps ...................................................................................................................................................... 7-10
7.9 Analysis of emergency samples.................................................................................................................... 7-11
I
III
User Manual
D
LT
11.1 Reagent registration ................................................................................................................................... 11-1
11.2 Reagent setting ............................................................................................................................................ 11-3
.,
Chapter 12 Alarm information and handling ............................................................................12-1
CO
12.1 Overview ...................................................................................................................................................... 12-1
12.2 Alarm information and troubleshooting................................................................................................... 12-2
AL
IV
User Manual
1.1 Overview
The Automatic Hematology Analyzer (model: BF-6900/ BF-6960) is designed and produced based on the idea of
accurate measurement, simple operation and low material consumption, conforming to users' requirements. The
Analyzer can provide quantitative analysis results of 25 parameters.
Applicable scope of the Analyzer: It can count the number of red blood cells and platelets in blood sample with
impedance method, test the hemoglobin concentration with colorimetric method, get the total number and have
five-classification of the white blood cells with semiconductor laser flow cytometry and calculate hemocyte related
parameters.
Contraindications: None.
D
(3)Relative humidity: Not exceeding 70%.
LT
(4)Atmospheric pressure: 75kPa~106kPa.
.,
(5)Altitude: Not greater than 2000m.
(6)No frost, condensation, water seepage, rain or solar exposure. CO
1.3 Analyzer parameters
AL
Model
Indicators
I
BF-6900 BF-6960
TR
WBC, BAS#, NEU#, EOS#, LYM#, WBC, BAS#, NEU#, EOS#, LYM#, MON#,
S
MON#, BAS%, NEU%, EOS%, LYM%, BAS%, NEU%, EOS%, LYM%, MON%,
DU
Test items MON%, RBC, HGB, MCV, MCH, RBC, HGB, MCV, MCH, MCHC,
MCHC, RDW-CV, RDW-SD, HCT, RDW-CV, RDW-SD, HCT, PLT, MPV,
IN
PLT, MPV, PDW, PCT, P-LCR, P-LCC, PDW, PCT, P-LCR, P-LCC,
Parameters 2 histograms, prompt and alarm functions histograms, prompt and alarm functions in
RU
Sample feeding Manual or automatic feeding (sample Manual or automatic feeding (sample feeder
way feeder is optional) is optional)
Sample storage
At least 200000 At least 500000
volume
Sample bar code Automatic recognition or manual input Automatic recognition or manual input
4 types, including 1 type of diluent and 3 4 types, including 1 type of diluent and 3
Reagent type
types of lyse types of lyse
1-1
User Manual
Model
Indicators
BF-6900 BF-6960
D
Without sample feeder: Without sample feeder:
system
LT
531mm×380mm×520mm 531mm×380mm×520mm
(H×W×L) (H×W×L)
.,
Power
240VA 240VA
consumption
CO
1.4 Working principle
AL
The Analyzer applies the electrical impedance method to test the number and volume distribution of red blood cell
and platelet. Colorimetric method is used to measure the concentration of hemoglobin. Semiconductor laser flow
I
cytometry is used to obtain the total number of white blood cells and count the white blood cell of five
TR
With different configuration, the instrument provides two sample feeding ways, closed sample feeding and
automatic sample feeding. The closed sample feeding supports whole blood mode, micro-whole blood mode and
IN
pre-dilution mode, while the automatic sample feeding supports whole blood mode.
The instrument supports the whole blood, micro-whole blood and pre-dilution modes. Under whole blood mode or
I
micro-whole blood mode, the Analyzer will aspirate 25μL (CBC+DIFF mode) or 20μL (CBC mode) whole blood
RU
sample.
Under pre-dilution test mode, the operator shall mix the 20μL peripheral blood sample with 180μL dilute outside the
DI
Analyzer to form a diluted sample (dilution ratio: 1:10), and then send the diluted sample to the Analyzer for
sampling. And then, the Analyzer will aspirate 80μL (CBC+DIFF mode) or 65μL (CBC mode) diluted sample.
1.4.2 Sample dilution
After the sample to be tested is collected with a sampling probe, it will then be dispensed to WBC classification
detector and RBC count cell according to test demands. And then, in parallel dilution processes, test sample used for
WBC classification measurement, WBC count/ hemoglobin measurement, red blood cell/ platelet measurement are
respectively formed with the action of different reagents.
For different test samples, the Analyzer provides three test modes, whole blood test mode, micro whole blood mode
and pre-dilution (peripheral blood) test mode.
1.4.2.1 Whole-blood mode
(1)Red blood cell/ platelet dilution process
1-2
User Manual
Whole-blood sample
6μL
Dispense BF-5D
diluent 2500μL
Pre-dilution
Aspirate sample 45μL
Dilution ratio1:417.65
D
LT
.,
CO
I AL
TR
1-3
User Manual
Whole-blood sample
20μL
D
sample with dilution ratio of
LT
about 1:28928
.,
Fig. 1-4-4 Red blood cell/ platelet dilution process
(2)White blood cell/ hemoglobin dilution process CO
I AL
S TR
DU
IN
I
RU
DI
1-4
User Manual
D
Fig. 1-4-6 White blood cell classification dilution process
LT
1.4.3 White blood cell measurement
.,
1.4.3.1 Laser flow cytometry
When a certain amount of blood cells is aspirated and cytochemically stained with a specific amount of reagent, the
CO
blood cell will be injected to a conical flow chamber fulfilled with diluent through nozzle. Wrapped by sheath flow
formed by diluent, the cells will flow through the center of flow chamber singly in row and line, as shown in the
figure below:
I AL
S TR
DU
IN
I
RU
DI
1-5
User Manual
BF-6900 has one DIFF scattergram, which is obtained from low-angle forward scattered light and
middle-angle forward scattered light.
BF-6960 has three DIFF scattergrams, which are obtained from low-angle forward scattered light,
D
middle-angle forward scattered light and middle-angle forward scattered light respectively.
LT
From the BASO channel scattergram, the percentage of white blood cell, basophil and eosinophil in total number of
white blood cell can be obtained.
.,
1.4.3.2 White blood cell parameters
CO
Through the analysis of BASO channel scattergram and its BASO region, the number of WBC and basophil
(BASO#) is got and then the percentage of basophil (BASO%) can be calculated. The Analyzer gets the percentage
of lymphocyte (LYM%), neutrophil (NEU%), monocyte (MON%) and eosinophil (EOS%) through the analysis of
AL
DIFF channel scattergram and the LYM zone, NEU zone, MON zone and EOS zone in it. Based on the number of
white blood cells acquired with electrical impedance method, the number of lymphocytes (LYM#), neutrophils
I
(NEU#), monocytes (MON#) and eosinophils (EOS#) is obtained. The counting unit of the cells above is 109/L.
TR
WBC=Total number of all particles in the BASO channel, except from ghost region.
(2)Basophil
IN
BAS #
BAS% 100%
WBC
DI
(4)Percentage of lymphocyte
(5)Percentage of neutrophil
(6)Percentage of monocyte
(7)Percentage of eosinophil
1-6
User Manual
(8)Lymphocyte
(9)Neutrophil
(10)Monocyte
(11)Eosinophil
D
the hemoglobin and lyse will generate hemoglobin compound. On one end of the colorimeter cell, LED luminotron
LT
passes a monochromatic light with wavelength of 540nm to reflect the hemoglobin compound solution. On the other
end, it passes phototube to receive transmission light and converts the light intensity signal to voltage signal after
.,
amplification. Through the comparison with the voltage generated by the transmission light intensity before the
CO
sample is dispensed to the colorimeter cell (only diluent in the colorimeter cell), the hemoglobin concentration
(HGB) of sample can be obtained and the unit is g/L. The measurement and calculation are completed by the
Analyzer automatically and the results will be displayed in the analysis results area on the counting interface.
AL
The Analyzer uses the electrical impedance method to count red blood cells/ platelets (as shown in Fig. 1-4-11). The
red blood cell/ platelet sample will enter RBC test unit after secondary dilution. The test unit has a small hole for test.
IN
At the two sides of the hole, there is a pair of positive and negative electrode to connect constant current power
supply. As the cells have the features of poor conductor, the DC resistance between electrodes will change when the
cells in diluted sample pass the small detection hole under constant negative pressure and pulse signal in proportion
I
to the cell volume and size will be formed at the two ends of electrode. When cells pass the small hole continuously,
RU
a string of electric pulses will be generated at two ends of the electrode. The number of pulses and the number of
DI
cells passing the small hole are equivalent and the pulse magnitude is in direct proportion to the volume of cell.
1-7
User Manual
After the collected electric pulse is amplified, it is compared with the channel voltage threshold corresponding to the
volume range of normal red blood cell/ platelet. And then, the number of electric pulses of pulse amplitude in red
blood cell/ platelet channel will be calculated. The collected electric pulses are classified according to different
channel voltage threshold and the number of electric pulses in red blood cell/ platelet channel is just the number of
red blood cells/ platelets. The number of cells in each channel divided according to the pulse voltage range
determines the volume distribution of cells. The abscissa shows the volume of cell and the ordinate shows the 2D
diagram of cell quantity, i.e. the histogram reflecting the cell group distribution.
1.4.5.2 Red blood cell parameters
(1)Number of red blood cells
The Analyzer gets the number of red blood cells (RBC) by the direct measurement of the number of electric pulses
of red blood cell, and the unit is 1012/L.
RBC n 1012 / L
D
concentration
LT
With the formulas below, the red blood cell hematocrit (HCT) (unit: %), mean red blood cell hemoglobin content
(MCH) (unit: pg) and mean red blood cell hemoglobin concentration (MCHC) (unit: g/L) can be calculated.
.,
RBC MCV
HCT
RBC hematocrit 10 CO
HGB
MCH =
AL
HGB
I
MCHC 100
TR
The unit of RBC is 1012/L, the unit of MCV is fL and the unit of HGB is g/L.
S
DU
The deviation limit of red blood cell distribution width (RDW-SD) is the width at peak 20% in histogram of red
RU
blood cell distribution and the unit is fL, as shown in the figure.
DI
1-8
User Manual
D
(5)Proportion of large platelet (P-LCR)
LT
The proportion of large platelet can be obtained from the platelet histogram. It is the proportion of large platelet in
.,
platelet.
(6)Large platelet counts (P-LCC)
CO
The count of large platelet can be obtained from the platelet histogram. It is the count of large platelets in platelet.
(7)Platelet distribution histogram
AL
The Analyzer can provide a platelet volume distribution graph while providing the platelet counting results. The
graph that shows the distribution of cells is called platelet distribution histogram. On the histogram, the abscissa
I
shows the volume of platelet (unit: fL) and the ordinate shows the relative number of platelet (unit: 109/L). After
TR
each counting, the platelet distribution histogram can be checked in the analysis results area on counting interface or
S
1.4.6 Rinsing
The Analyzer will rinse the components where sample flows automatically in each counting period to ensure no
IN
1-9
User Manual
D
LT
1 Front door 2 Display screen 3 Emergency position 4 Count key 5 Sample feeder
.,
Fig. 1-5-1 Front view of the Analyzer
1 Upper cover 2 Right side door assembly 3 Back plate 4 Power vent 5 Waste liquid level sensor port
6 Diluent port 7 Waste liquid port 8 Left side door assembly 9 Power vent
10 Power switch/ power socket/ power filter
Fig. 1-5-2 Rear view of the Analyzer
1-10
User Manual
1 USB interface 2 USB interface 3 RJ45 network interface 4 Test switch 5 Door lock
D
Fig. 1-5-3 Left view of the Analyzer
LT
1.6 External device of the Analyzer
.,
(1)Printer: It is connected with the instrument directly and it prints reports through the software of the instrument.
The instrument supports the following printer:
HP LaserJet Pro 400 M401d / HP LaserJet Pro 400 M403d.
CO
(2)Bar code reader: It is connected with the instrument directly and it can input the bar code information quickly.
I AL
1.7 Symbol
S
DU
Table 1-7-1
IN
Symbols Meaning
I
Biohazard, reminding the user to pay attention; otherwise there is risk of potential
RU
bio-infectivity
DI
ALTERNATING CURRENT
BATCH CODE
USE BY
SERIAL NUMBER
DATE OF MANUFACTURE
1-11
User Manual
Symbols Meaning
MANUFACTURER
CATALOGUE NUMBER
D
LT
"ON" (POWER)
.,
"OFF" (POWER)
CO
PROTECTIVE EARTH
AL
The symbol of the crossed out wheeled bin indicates that the product (electrical and electronic
I
equipment) should not be placed in municipal waste. Please check local regulations for
TR
Temperature limit
IN
I
Humidity limitation
RU
DI
USB
network interface
The symbols above are also applicable to the Analyzer, reagent, QC object and calibration object.
1-12
User Manual
1.8 Identification
(1)
D
LT
.,
CO
I AL
TR
S
DU
IN
I
RU
DI
1-13
DI
RU
I
IN
DU
S
1-14
TR
I
User Manual
AL
CO
.,
LT
D
User Manual
Chapter 2 Installation
To ensure the normal operation of the Analyzer after the installation, the Analyzer shall be installed by the
authorized operators of the manufacturer at the delivery.
D
(3)The distance from the front instrument to other instruments shall not be shorter than 100cm.
LT
(4)Sufficient space shall be guaranteed on the operation platform or under the Analyzer for placing collecting
devices of diluent, reagent and waste liquid.
.,
2.1.2 Power requirements
(1)Supply voltage: 100-240V~, 50/60Hz. CO
(2)Power consumption: 240VA.
AL
into the same receptacle with electrical equipment with a heavy load such as air conditioner, refrigerator and oven.
TR
d)The Analyzer shall be put in a dust-free environment without mechanical vibration, source of large noise and
RU
power interference.
e)It is recommended to evaluate the electromagnetic environment in laboratory before operating the Analyzer.
DI
f)Do not have the Analyzer closing to the interference source of high electromagnetic, lest it may interfere with
the normal operation of the Analyzer.
g)Do not put the Analyzer near brush motor, scintillant fluorescent lamp and electrical contact equipment often
used.
h)The Analyzer shall be prevented from direct sunlight exposure and not be placed near heat and wind sources.
i)The environment shall be well-ventilated and the ventilating device shall be used if necessary. But the Analyzer
should be protected from the direct airflow, otherwise, the test accuracy may be affected.
If the operating environment or power supply of the Analyzer do not meet the requirements above, the
accuracy and precision of test results of the Analyzer may be affected, the Analyzer may be damaged or
personal injury may be caused.
(2)Safety conditions
a)Indoor use;
b)Altitude shall not exceed 2000m.
2-1
User Manual
2.2 Unpacking
2.2.1 Unpacking steps
After the instrument arrives, please open the case according to the following steps:
(1)Make the anti-inclination sign on the packing case upright. Verify that the package is complete,
and the appearance of the package does not allow sharp penetration injury and heavy extrusion
leading to deformation which causes the loss of protective function of the packaging case. If there is
any damage, please contact the manufacturer or local agent.
D
LT
(2)Take out the host and accessories; check whether the host and accessories are complete according
to the packing list. If there is any missing, please contact the manufacturer or local agent.
.,
(3)Check the appearance of the system carefully. If there is any damage, please contact the manufacturer or local
agent.
2.2.2 Handling method
CO
AL
(1)Use transport machine such as utility trolley for the stable transportation of short distance.
(2)Prevent the aspiration probe from other objects and being damaged during handling and transportation processes.
I
(3)Keep the Analyzer vertical, not tilting or side laying when moving and carrying.
TR
a)Unscrew three fixing screws for the right door, remove the right door of the instrument, and then unscrew the
right fixing screw for the front door, as shown in the following figure:
I
RU
DI
Fig. 2-2-1
b)Open the left door of the instrument and unscrew the left fixing screw of the front door, as
shown below:
2-2
User Manual
Fig. 2-2-2
c)Open the front door of the instrument as shown in the figure:
D
LT
.,
CO
I AL
S TR
DU
IN
I
Fig. 2-2-3
RU
(2)Remove the rubber plug of the reaction cell as shown in the figure:
DI
Fig. 2-2-4
(3)Remove the probe fastener: unscrew the two cross recessed pan head combination screws in the figure below and
2-3
User Manual
Fig. 2-2-5
(4)If a sample feeder is selected, the mixing unit fastener needs to be removed: open the front door, and screw down
D
the cross recessed pan head combination screw for fixing the isolation sleeve. Remove the isolation sleeve; then
LT
screw down the two cross recessed pan head combination screws that fix the fastener and remove the fastener, as
shown in the following figure:
.,
CO
I AL
S TR
DU
IN
Fig. 2-2-6
(5)According to the order of c), b) and a) in step (1), fix the instrument front door and right door again.
I
Do not disassemble or install the instrument except for the normal maintenance.
2-4
User Manual
D
1 Waste liquid level sensor port 2 BF-5D diluent port 3 Waste liquid port
LT
Fig. 2-3-1 Connection of the instrument and reagent
.,
CO
I AL
S TR
DU
IN
I
RU
DI
● The height difference between the waste liquid tank and the waste liquid port of the instrument shall not be
less than 0.5m.
● The waste liquid shall be discharged in accordance with the local regulations on the disposal of medical
waste.
● The drainage system shall be in compliance with the local regulations with regard to sewage discharge and
2-5
User Manual
Connect "the network interface" of the computer with " " (3 in Fig. 1-5-3) on the right panel of the Analyzer.
(4)Connection of power cable
Plug one end of the accompanying power cable into the power socket on the right panel of the Analyzer (10 in Fig.
1-5-2), and plug the host power cable, display power cable and printer power cable.
● The receptacle connected with the power cable shall be reliably grounded.
● Do not put the instrument at a place where the disconnecting device is hard to be operated.
(5)Connection of bar code reader
D
LT
Connect another end of accompanying bar code reader with any " " interface on the instrument.
.,
CO
The bar code reader will emit light harmful to people's eyes, please do not look straight at the light when the
Analyzer is working.
(6)Connection of printer
AL
Use a data cable to connect the printer and the interface of the Analyzer correctly and check the
I
TR
In the transportation process, to prevent damage to the aspiration probe, moving mechanisms of the instrument have
IN
been fixed when the instrument is delivered.Before the instrument is powered on, the fixing screws and tie shall be
removed first, or the aspiration probe may be damaged.
I
2-6
User Manual
D
LT
.,
Fig. 2-4-1
CO
In the login dialog box, input the correct user name and password, the initial user name of the instrument is admin,
AL
and the initial password is 1, if a wrong user name and password are input, login failure will appear, and the
interface is as shown in the figure below:
I
S TR
DU
IN
I
RU
Fig. 2-4-2
DI
The user name and the password are composed of lower and upper case English letters or figures (1-12
characters).
After logging in, the instrument enters the initialization interface to initialize. It performs version verification,
transmission gain, mechanical reset, sending mode, reagent residue detection, pipeline rinsing, background
detection, temperature detection and blank test.
● The blank result obtained during the blank test is beyond the range set by the software and the instrument
will perform the blank test again. If the result is still not qualified after tested 3 times, "blank test error" will
be prompted.
● The instrument will not have High valve, Low value or Suspicious alarms for the blank test results.
● When the power-on self-check alarm triggers, the instrument enters the main interface. Click the red
exclamation mark to check corresponding alarm information. After the alarm is cleared, the instrument
performs self-check again.
After the instrument finishes the self-check, enter the main window of the software, as shown in the figure below:
2-7
User Manual
D
LT
.,
Fig. 2-4-3
CO
2.4.2 Description of counting interface
AL
2.4.2.1 Menu
I
Click the [Menu] to pop up the menu as shown in the figure. The functions in the menu bar are described as follows:
S TR
DU
IN
I
RU
DI
Fig. 2-4-4
Spl. Reg.: Have sample registration and work list input.
Calibration: Calibrate the instrument.
Q.C.: L-J/Xbar, Xbar-R, X-B quality control.
Maint.: Check the status of the instrument and have maintenance and check of the instrument.
Settings: Set parameters of the instrument.
Log: Record system operating information and error information.
Reagent: reagent registration and reagent information settings.
2-8
User Manual
Logout: Switch between users and log in software interface with a new user identity.
Shutdown: Turn the instrument off.
2.4.2.2 Public information area
The public information area is shown in the lower part of the counting interface, as shown in figure below
Fig. 2-4-5
(1)Total number of locations: Display the current sample serial number and total sample size
(2)Current status: Display the analysis mode of the sample.
(3)Test status: Indicate the current status of the instrument.
2.4.2.3 Status indication area
On the right side of Fig. 2-4-5 is the status indication area. The area displays in sequence the X-B QC switch status,
U disk connection status, LIS system connection status, and printer status, described as follows:
D
(1)X-B QC switch status:
LT
: The X-B quality control is on.
.,
: The X-B quality control is not turned on.
(2)USB status:
CO
AL
(4)Printer status:
RU
Fig. 2-4-6
(1)Spl. analysis: Finish sample test and the analysis and display of test results on the interface.
(2)History: The instrument will automatically save the analysis results in the database after executing sample
analyzing each time. The operator can query all sample results stored in the database.
(3)QC: Enter the QC interface to realize the basic setting of QC and QC counting.
(4)Add diluent: Dispense diluent for the preparation of pre-dilution sample under the pre-dilution mode.
(5)Rinse: Manually clean equipment.
(6)Print: Print the reports of selected samples.
2-9
User Manual
D
LT
.,
CO
I AL
Fig. 2-4-7
TR
After clicking relevant failure information, the detailed solution to the failure will be displayed in the "Alarm help"
and the user can have simple troubleshooting according the solution.
S
DU
Fig. 2-5-1
2-10
User Manual
D
LT
.,
Fig. 2-5-2
CO
I AL
TR
S
DU
IN
I
RU
DI
Fig. 2-5-3
2-11
User Manual
D
LT
.,
CO
I AL
Fig. 2-5-4
TR
S
DU
IN
I
RU
DI
2-12
User Manual
3.1 Overview
The system parameters of the instrument have had initialization setting in the factory. The interface shown at the
first turn-on of the instrument is the default interface. To meet different demands in actual use, the software sets two
identities, user and administrator. Users with different identities can reset different parameters.
D
LT
.,
CO
I AL
S TR
DU
IN
I
Fig. 3-2-1
RU
(1)Next No.: Enter the number of the next sample to be analyzed in the input box.
DI
(2)Next bar code: What is input in the box is the barcode number of the next manual analysis sample.
The sample number can only be entered in figure or "-", and its beginning and ending must be numeric and
up to 12 digits.
(3)Analysis mode: There are totally four modes, among which the [WB], [Micro-WB] and [Pre-dilute] modes are
manual analysis modes and [Auto-WB] mode is an automatic analysis mode.
The automatic-whole blood mode can only be used during the assembly of the sample feeder.
(4)Test mode: there are totally four modes;
"CBC" mode: count only, with no classification of the white blood cells. The count result comprises of histograms
of the RBC and PLT and other parameters.
"CBC+DIFF" mode: in addition to counting, it also has five differential classifications of the white blood cells,
including 25 parameters and scattergrams and histograms.
(5)Tube number and rack number: If the [Auto-WB] mode is selected and [Auto-identify] is not selected, the [Rack
3-1
User Manual
No.] and [Tube No.] will change to an editable state and the rack number and tube number can be reset in the box
following [Rack No.] and [Tube No.].
(6)Use worklist: use the manually entered or downloaded worklist for testing.
(7)Two-way LIS: when bi-directional LIS is set, after the testing is started, the instrument will obtain a worklist
from LIS terminal automatically and then conduct test. After the testing, the worklist information and test results
will be transmitted to LIS terminal.
(8)Auto-COM.: After connection to the LIS, automatically transmit the test results after the testing.
If a barcode reader is not connected, the “Two-way LIS” option will not be displayed on the interface.
D
LT
.,
CO
I AL
S TR
DU
IN
I
RU
DI
Fig. 3-3-1
Click the corresponding input box of the unit following each test item and select a different unit as the test unit
according to actual demand.
After inputting, click [Save] to confirm, and when "Saved successfully" is prompted, press [OK] on the interface.
Click [DEF] to restore the unit to the default settings.
Unit of each test item: several units for the setting of each test item of the instrument are available for the user to
choose and specific setting unit is as shown in Table 3-3-1:
Table 3-3-1
103/μL ***.**
WBC
102/μL ****.*
/nL ***.**
3-2
User Manual
106/μL **.**
RBC
104/μL ****
/pL **.**
mmol/L **.*
D
***.*
LT
pg **.* Default unit
MCH
.,
fmol ***
mmol/L ***.*
I
% **.*
S
HCT
L/L *.*** Default unit
DU
103/μL ****
PLT, P-LCC
I
104/μL ***.*
RU
/nL ****
DI
After the unit of the parameter is changed, the data format of the test result also changes.
3-3
User Manual
D
LT
.,
CO
Fig. 3-4-1
AL
Click the corresponding input box of lower and upper limits following each test item and directly input the lower
TR
After the limits are input, press [Save] for confirmation. When a prompt information of "Saved successfully" pops
DU
Click [Menu] - [Settings] - [Abno.flag] and select [WBC], as shown in the figure below:
RU
DI
3-4
User Manual
D
LT
.,
Fig. 3-5-1
CO
Select [RBC/PLT] and the interface is as shown in the figure below:
I AL
S TR
DU
IN
I
RU
DI
Fig. 3-5-2
Select [Suspicious mark], as shown in the figure below:
3-5
User Manual
D
LT
.,
Fig. 3-5-3
CO
On this interface, the user can set the prompt range of WBC, RBC/ PLT and the abnormal alarm information of
AL
suspicious marks.
I
After the values are input, click [Save] to save the settings or click [DEF] to restore the default value.
TR
3.6 QC setting
S
3.6.1 QC method
DU
3-6
User Manual
D
LT
.,
Fig. 3-6-1
CO
The user can select the QC mode. After saving the settings, the user can click "QC" on the shortcut key interface to
AL
Fig. 3-6-2
In the figure above, users can select the L-J QC calculation method and range:
3-7
User Manual
D
LT
.,
CO
I AL
S TR
DU
Fig. 3-6-3
IN
In the figure above, users can select the Xbar QC calculation method and range:
Calculation and range selection of deviation:
I
(1)If [Absolute value] is selected as the calculation method, the deviation limit input will be displayed in the form of
RU
an absolute value, and 2SD or 3SD will be displayed on [Range] as deviation limit.
DI
(2)If [Percentage] is selected as the calculation method, the deviation limit input will be displayed in the form of a
percentage, and 2CV or 3CV will be displayed on [Range] as deviation limit.
3.6.4 X-B setting
Select [X-B] and the interface is as shown in the figure below:
3-8
User Manual
D
LT
.,
Fig. 3-6-4
CO
The user can choose X-B QC switch and set sample volume.
AL
Fig. 3-7-1
(1)Department: Click [Add] to add department information in the popup interface as shown in the figure below:
3-9
User Manual
D
LT
.,
Fig. 3-7-2
CO
Select the row where the department information to be deleted is, click [Delete] in the figure to pop up the prompt as
AL
Fig. 3-7-3
RU
3-10
User Manual
D
LT
.,
Fig. 3-7-4
CO
Click [Add] to add doctor information in the popup interface as shown in the figure below:
I AL
S TR
DU
IN
I
RU
DI
Fig. 3-7-5
Select the row where the doctor information to be deleted is, click [Delete] in the figure to pop up the prompt as
shown in the figure below:
3-11
User Manual
Fig. 3-7-6
Click [OK] to delete corresponding doctor information.
(3)Cost type: Select [Cost type], the interface shown in the figure below:
D
LT
.,
CO
I AL
S TR
DU
IN
I
RU
Fig. 3-7-7
Click [Add] to add charge information in the popup interface as shown in the figure below:
DI
3-12
User Manual
D
LT
.,
Fig. 3-7-8
CO
Select the row where the charge information to be deleted is, click [Delete] in the figure to pop up the prompt as
AL
Fig. 3-7-9
RU
When adding the department, doctor, and cost type, it is not allowed to input special characters.
3-13
User Manual
D
LT
.,
CO
Fig. 3-8-1
AL
(1)Add user: Click [Add user] in the figure to add user information as shown in the figure below:
I
S TR
DU
IN
I
RU
DI
Fig. 3-8-2
Fill in the user, name, password, confirm password and select permission according to the corresponding prompt
and click [Save] to complete the operation.
3-14
User Manual
D
Fig. 3-8-3
LT
Click [OK] to delete corresponding user information.
.,
The currently logged in user cannot delete itself. CO
(3)Change password: In Fig. 3-8-1, select the row of the user information that you want to change your password
AL
Fig. 3-8-4
In the figure, enter the old password and new password respectively, and click [Save] to save the modified
password.
3-15