Professional Documents
Culture Documents
RU
I
IN
DU
ST
RIA
L
CO
.,
LT
D
The user manual is applicable to Automatic Hematology Analyzer (model: BF-6900CRP/ BF-6960CRP) and
all figures concerned are described based on the Automatic Hematology Analyzer (model: BF-6960CRP,
hereinafter referred to as the Analyzer).
DESCRIPTION
Dear users, thanks for buying the Automatic Hematology Analyzer (model: BF-6900CRP/ BF-6960CRP).
Please read the manual carefully before operation as incorrect operation may affect the test results of the Analyzer or
cause personal injury.
After reading, please reserve the manual properly for reference at any time.
Manufacturer Address:
D
LT
95 Yunhe Street, New & High Tech. Development Zone, Changchun, Jilin 130012, the People’s Republic of China
.,
Production Address:
CO
3333 Yiju Road, New & High Tech. Development Zone Changchun, Jilin 130103, the People’s Republic of China
AL
Website: http://www.dirui.com.cn
DU
E-mail: dirui@dirui.com.cn
IN
Fax: 0431-85173354
DI
● The Analyzer shall be used by professional medical examination personnel or trained doctors, nurses or testers.
● As the Analyzer has biological and chemical risks, the operator shall be trained and use personal protective
appliance to reduce the risk.
● Only trained operators are allowed to conduct dangerous operations, such as moving parts.
● The hospital or inspection institution shall prepare a service plan and carry out servicing and maintenance in strict
accordance with the service plan, or the Analyzer may have faults.
● The Analyzer shall be controlled with a special software designated by the company. Installation of other software
or hardware on the computer may affect the normal operation of the Analyzer. Please do not operate other software
D
during the operation of the Analyzer.
LT
● Please refer to the instructions of reagent for the use and storage of the reagent and ensure the reagent is used
.,
within the life of the reagent shown in the instructions.
CO
● Do not use reagent beyond its life. After its seal is damaged, prevent dust, dirty matters or bacteria entering it.
AL
● Do not use any organic solvents such as turpentine and benzene to rinse the outer part of the Analyzer as it may
I
cause the color or shape changes of the Analyzer. Use soft or wet cloth to clean the Analyzer and use diluted
TR
● Under an environment with low transportation or storage temperature or relative humidity greater than 70%, the
IN
Analyzer shall be turned on for testing only after it is stored in a normal working environment for 24 hours.
● The Analyzer shall be provided with an independent power source. If it shares one power socket with other
I
RU
electrical equipment, the electromagnetic interference may affect the test results of the Analyzer.
DI
● Do not pull or insert the plug with wet hands as it may cause electric shock.
● Damaged power cable and connection cable shall not be used. The power cables and wires shall not be trodden,
twisted or pulled as it may cause a fire.
● The input voltage shall meet the requirements of the Analyzer and the fuse of specified specification shall be used.
● Ensure the switch of the Analyzer is at the [O] position before connecting the power cable.
● Do not touch the moving components when the Analyzer is operating to prevent accidents.
● When the power for the Analyzer is connected, servicing personnel not authorized shall not open the left and right
doors and upper cover.
● Please use the Analyzer under conditions regulated in the manual. If not, the Analyzer may not operate normally,
the test results may not be reliable, the components of the Analyzer may be damaged and personal injuries may be
caused.
● The protection measures provided for the Analyzer may become invalid if the Analyzer is not used according to
the manual.
D
LT
WARNING
.,
CO
● The operator is obligated to follow national and local regulations on discharge and treatment of expired reagent,
AL
● The waste liquid and the consumables shall be correctly treated. As the blood in the waste liquid may be polluted
TR
by pathogen, please treat the waste liquid and the consumables of the Analyzer according to regulations about
S
DU
● Do not touch the sampling probe as the blood sample, quality control object and calibration object on the sampling
probe have potential biological infectivity. During the sample aspirating process of the sampling probe, prevent
I
touching the test tube wall and probe tip as it may cause bleeding. Besides, a certain distance shall be kept from the
RU
aspiration probe tip to the bottom of container, or the accuracy of the aspirated liquid volume may be affected.
DI
Dirui Company declares that it will be responsible for the safety, reliability and performances of the Analyzer only
if all following requirements are met.
(1)The installation, commissioning and servicing of the Analyzer are undertaken by professional personnel of Dirui
Company.
No further notice will be provided in case of any changes to the software interface.
D
LT
.,
CO
I AL
S TR
DU
IN
I
RU
DI
User Manual
Contents
D
1.5 Structure of the instrument ......................................................................................................................... 1-11
LT
1.5.1 Front view of the Analyzer .............................................................................................................................................. 1-11
1.5.2 Rear view of the Analyzer ............................................................................................................................................... 1-11
.,
1.5.3 Left view of the Analyzer ................................................................................................................................................ 1-12
CO
1.6 External device of the Analyzer................................................................................................................... 1-12
1.7 Symbol ........................................................................................................................................................... 1-12
AL
I
User Manual
D
Chapter 4 Calibration.....................................................................................................................4-1
LT
4.1 Overview .......................................................................................................................................................... 4-1
.,
4.2 Calibration frequency .................................................................................................................................... 4-1
CO
4.3 Calibration method......................................................................................................................................... 4-1
4.4 Calibrator calibration .................................................................................................................................... 4-1
4.4.1 Input of calibrator reference value..................................................................................................................................... 4-2
AL
II
User Manual
D
Chapter 7 Routine operation .........................................................................................................7-1
LT
7.1 Overview .......................................................................................................................................................... 7-1
.,
7.2 Preparation for operation .............................................................................................................................. 7-1
CO
7.3 Turn-on ............................................................................................................................................................ 7-2
7.4 Daily QC .......................................................................................................................................................... 7-2
7.5 Sample preparation ........................................................................................................................................ 7-2
AL
III
User Manual
D
9.3.4 HGB verification ............................................................................................................................................................... 9-9
LT
9.3.5 Mechanical detection ......................................................................................................................................................... 9-9
9.3.6 Instrument registration .................................................................................................................................................... 9-13
.,
9.3.7 Data backup ..................................................................................................................................................................... 9-13
CO
9.3.8 Counter ............................................................................................................................................................................ 9-14
9.4 Replacement of vulnerable parts ................................................................................................................. 9-15
9.4.1 Replacement of sampling probe ...................................................................................................................................... 9-15
AL
IV
User Manual
1.1 Overview
The Automatic Hematology Analyzer (model: BF-6900CRP/ BF-6960CRP) is designed and produced based on the
idea of accurate measurement, simple operation and low material consumption, conforming to users' requirements.
The Analyzer can provide quantitative analysis results of 27 parameters.
Applicable scope of the Analyzer: It can count the number of red blood cells and platelets in blood sample with
impedance method, test the hemoglobin concentration with colorimetric method, get the total number and have
five-classification of the white blood cells with semiconductor laser flow cytometry, test the c-reactive protein
concentration with latex immunoturbidimetric assay and calculate hemocyte related parameters.
Contraindications: None.
D
(3)Relative humidity: Not exceeding 70%.
LT
(4)Atmospheric pressure: 75kPa~106kPa.
.,
(5)Altitude: Not greater than 2000m.
(6)No frost, condensation, water seepage, rain or solar exposure. CO
1.3 Analyzer parameters
AL
Model
Indicators
I
BF-6900CRP BF-6960CRP
TR
WBC, BAS#, NEU#, EOS#, LYM#, WBC, BAS#, NEU#, EOS#, LYM#, MON#,
S
MON#, BAS%, NEU%, EOS%, LYM%, BAS%, NEU%, EOS%, LYM%, MON%,
DU
MON%, RBC, HGB, MCV, MCH, RBC, HGB, MCV, MCH, MCHC,
Test items
MCHC, RDW-CV, RDW-SD, HCT, RDW-CV, RDW-SD, HCT, PLT, MPV,
PLT, MPV, PDW, PCT, P-LCR, P-LCC, PDW, PCT, P-LCR, P-LCC, FR-CRP,
IN
Parameters 2 histograms, prompt and alarm functions histograms, prompt and alarm functions in
RU
Sample feeding Manual or automatic feeding (sample Manual or automatic feeding (sample feeder
way feeder is optional) is optional)
Sample storage
At least 200000 At least 500000
volume
Sample bar code Automatic recognition or manual input Automatic recognition or manual input
6 types, including 1 type of diluent and 5 6 types, including 1 type of diluent and 5
Reagent type
types of lyse types of lyse
1-1
User Manual
Model
Indicators
BF-6900CRP BF-6960CRP
D
With sample feeder: With sample feeder:
LT
531mm×556mm×679mm 531mm×556mm×679mm
Whole
(H×W×L); (H×W×L);
.,
machine Dimensions
Without sample feeder: Without sample feeder:
system
531mm×380mm×520mm 531mm×380mm×520mm
(H×W×L) CO (H×W×L)
Power
260VA 260VA
consumption
AL
The Analyzer applies the electrical impedance method to test the number and volume distribution of red blood cell
and platelet. Colorimetric method is used to measure the concentration of hemoglobin. Semiconductor laser flow
S
cytometry is used to obtain the total number of white blood cells and count the white blood cell of five
DU
classifications. Latex immunoturbidimetric assay is used to measure the concentration of C-reactive protein. On this
basis, it calculates the results of other parameters.
IN
automatic sample feeding. The closed sample feeding supports whole blood mode, micro-whole blood mode and
RU
pre-dilution mode, while the automatic sample feeding supports whole blood mode.
DI
The instrument supports the whole blood, micro-whole blood and pre-dilution modes. Under whole blood mode or
micro-whole blood mode, the Analyzer will aspirate 25μL (CBC+DIFF mode) or 20μL (CBC mode) whole blood
sample; 30μL (CBC+DIFF+CRP mode), 28μL (CBC+CRP mode) or 18μL (CRP mode) whole blood sample.
Under pre-dilution test mode, the operator shall mix the 20μL peripheral blood sample with 180μL dilute outside the
Analyzer to form a diluted sample (dilution ratio: 1:10), and then send the diluted sample to the Analyzer for
sampling. And then, the Analyzer will aspirate 80μL (CBC+DIFF mode), 65μL (CBC mode), 150μL
(CBC+DIFF+CRP mode), 85μL (CRP mode) or 135μL (CBC+CRP mode) diluted sample.
1.4.2 Sample dilution
After the sample to be tested is collected with a sampling probe, it will then be dispensed to C-reactive protein
detector, WBC classification detector and RBC count cell according to test demands. And then, in parallel dilution
processes, test sample used for WBC classification measurement, WBC count/ hemoglobin measurement, red blood
cell/ platelet measurement and C-reactive protein measurement are respectively formed with the action of different
reagents.
For different test samples, the Analyzer provides three test modes, whole blood test mode, micro whole blood mode
and pre-dilution (peripheral blood) test mode.
1-2
User Manual
Whole-blood sample
6μL
Dispense BF-5D
diluent 2500μL
Pre-dilution
Aspirate sample 45μL
Dilution ratio1:418
D
Fig. 1-4-1 Red blood cell/ platelet dilution process
LT
(2)White blood cell/ hemoglobin dilution process
.,
CO
I AL
S TR
DU
Dispense BF-FDO I
lyse 500μL
Dispense BF-FDT I
lyse 120μL
1:104
1-3
User Manual
D
LT
Whole-blood sample
.,
20μL
CO
BF-5D diluent 180μL
AL
after mixing
RU
1-4
User Manual
D
LT
.,
CO
I AL
S TR
DU
IN
I
RU
1-5
User Manual
D
LT
.,
Fig. 1-4-8 C-reactive protein dilution process
blood cell will be injected to a conical flow chamber fulfilled with diluent through nozzle. Wrapped by sheath flow
formed by diluent, the cells will flow through the center of flow chamber singly in row and line, as shown in the
I
figure below:
S TR
DU
IN
I
RU
DI
1-6
User Manual
BF-6900CRP has one DIFF scattergram, which is obtained from low-angle forward scattered light and
middle-angle forward scattered light.
BF-6960CRP has three DIFF scattergrams, which are obtained from low-angle forward scattered light,
D
middle-angle forward scattered light and high-angle forward scattered light respectively.
LT
From the BASO channel scattergram, the percentage of white blood cell, basophil and eosinophil in total number of
white blood cell can be obtained.
.,
1.4.3.2 White blood cell parameters
CO
Through the analysis of BASO channel scattergram and its BASO region, the number of WBC and basophil
(BASO#) is got and then the percentage of basophil (BASO%) can be calculated. The Analyzer gets the percentage
of lymphocyte (LYM%), neutrophil (NEU%), monocyte (MON%) and eosinophil (EOS%) through the analysis of
AL
DIFF channel scattergram and the LYM zone, NEU zone, MON zone and EOS zone in it. Based on the number of
white blood cells acquired with electrical impedance method, the number of lymphocytes (LYM#), neutrophils
I
(NEU#), monocytes (MON#) and eosinophils (EOS#) is obtained. The counting unit of the cells above is 109/L.
TR
WBC=Total number of all particles in the BASO channel, except from ghost region.
(2)Basophil
IN
BAS #
BAS% 100%
WBC
DI
(4)Percentage of lymphocyte
(5)Percentage of neutrophil
(6)Percentage of monocyte
(7)Percentage of eosinophil
1-7
User Manual
(8)Lymphocyte
(9)Neutrophil
(10)Monocyte
(11)Eosinophil
D
the hemoglobin and lyse will generate hemoglobin compound. On one end of the colorimeter cell, LED luminotron
LT
passes a monochromatic light with wavelength of 540nm to reflect the hemoglobin compound solution. On the other
end, it passes phototube to receive transmission light and converts the light intensity signal to voltage signal after
.,
amplification. Through the comparison with the voltage generated by the transmission light intensity before the
CO
sample is dispensed to the colorimeter cell (only diluent in the colorimeter cell), the hemoglobin concentration
(HGB) of sample can be obtained and the unit is g/L. The measurement and calculation are completed by the
Analyzer automatically and the results will be displayed in the analysis results area on the counting interface.
AL
The Analyzer uses the electrical impedance method to count red blood cells/ platelets (as shown in Fig. 1-4-11). The
red blood cell/ platelet sample will enter RBC test unit after secondary dilution. The test unit has a small hole for test.
IN
At the two sides of the hole, there is a pair of positive and negative electrode to connect constant current power
supply. As the cells have the features of poor conductor, the DC resistance between electrodes will change when the
cells in diluted sample pass the small detection hole under constant negative pressure and pulse signal in proportion
I
to the cell volume and size will be formed at the two ends of electrode. When cells pass the small hole continuously,
RU
a string of electric pulses will be generated at two ends of the electrode. The number of pulses and the number of
DI
cells passing the small hole are equivalent and the pulse magnitude is in direct proportion to the volume of cell.
1-8
User Manual
After the collected electric pulse is amplified, it is compared with the channel voltage threshold corresponding to the
volume range of normal red blood cell/ platelet. And then, the number of electric pulses of pulse amplitude in red
blood cell/ platelet channel will be calculated. The collected electric pulses are classified according to different
channel voltage threshold and the number of electric pulses in red blood cell/ platelet channel is just the number of
red blood cells/ platelets. The number of cells in each channel divided according to the pulse voltage range
determines the volume distribution of cells. The abscissa shows the volume of cell and the ordinate shows the 2D
diagram of cell quantity, i.e. the histogram reflecting the cell group distribution.
1.4.5.2 Red blood cell parameters
(1)Number of red blood cells
The Analyzer gets the number of red blood cells (RBC) by the direct measurement of the number of electric pulses
of red blood cell, and the unit is 1012/L.
RBC n 1012 / L
D
concentration
LT
With the formulas below, the red blood cell hematocrit (HCT) (unit: %), mean red blood cell hemoglobin content
(MCH) (unit: pg) and mean red blood cell hemoglobin concentration (MCHC) (unit: g/L) can be calculated.
.,
RBC MCV
HCT
RBC hematocrit 10 CO
HGB
MCH =
AL
HGB
I
MCHC 100
TR
The unit of RBC is 1012/L, the unit of MCV is fL and the unit of HGB is g/L.
S
DU
The deviation limit of red blood cell distribution width (RDW-SD) is the width at peak 20% in histogram of red
RU
blood cell distribution and the unit is fL, as shown in the figure.
DI
1-9
User Manual
D
(5)Proportion of large platelet (P-LCR)
LT
The proportion of large platelet can be obtained from the platelet histogram. It is the proportion of large platelet in
.,
platelet.
(6)Large platelet counts (P-LCC)
CO
The count of large platelet can be obtained from the platelet histogram. It is the count of large platelets in platelet.
(7)Platelet distribution histogram
AL
The Analyzer can provide a platelet volume distribution graph while providing the platelet counting results. The
graph that shows the distribution of cells is called platelet distribution histogram. On the histogram, the abscissa
I
shows the volume of platelet (unit: fL) and the ordinate shows the relative number of platelet (unit: 109/L). After
TR
each counting, the platelet distribution histogram can be checked in the analysis results area on counting interface or
S
BF-CRP lyse mixes with whole-blood sample to treat hemocyte and remove the disturbance of cell particles in
blood to immunoturbidimetric test. The C-reactive protein reagent mixes with the sample after hemolysis and the
I
antibody labeling latex microsphere in C-reactive protein reagent has agglutination reaction with the C-reactive
RU
protein in sample. As a result, the turbidity of the solution increases and the concentration of C-reactive protein can
be obtained through immunoturbidimetry.
DI
1-10
User Manual
D
LT
.,
CO
1 Front door 2 Display screen 3 Emergency position 4 Count key 5 Sample feeder
AL
1 Upper cover 2 Right side door assembly 3 Back plate 4 Power vent 5 Waste liquid level sensor port
6 Diluent port 7 Waste liquid port 8 Left side door assembly 9 Power vent
10 Power switch/ power socket/ power filter
Fig. 1-5-2 Rear view of the Analyzer
1-11
User Manual
D
LT
.,
1 USB interface 2 USB interface 3 RJ45 network interface 4 Test switch 5 Door lock
CO
Fig. 1-5-3 Left view of the Analyzer
(1)Printer: It is connected with the instrument directly and it prints reports through the software of the instrument.
The instrument supports the following printer:
I
(2)Bar code reader: It is connected with the instrument directly and it can input the bar code information quickly.
S
DU
1.7 Symbol
I
Table 1-7-1
RU
Symbols Meaning
DI
Biohazard, reminding the user to pay attention; otherwise there is risk of potential
bio-infectivity
ALTERNATING CURRENT
BATCH CODE
USE BY
1-12
User Manual
Symbols Meaning
SERIAL NUMBER
DATE OF MANUFACTURE
MANUFACTURER
D
CATALOGUE NUMBER
LT
.,
Do not contact the sampling probe when the instrument is working.
CO
"ON" (POWER)
AL
"OFF" (POWER)
I
TR
PROTECTIVE EARTH
S
DU
The symbol of the crossed out wheeled bin indicates that the product (electrical and electronic
equipment) should not be placed in municipal waste. Please check local regulations for
IN
Temperature limit
DI
Humidity limitation
USB
network interface
1-13
User Manual
The symbols above are also applicable to the Analyzer, reagent, QC object and calibration object.
1.8 Identification
(1)
D
LT
.,
CO
I AL
S TR
DU
IN
I
RU
DI
1-14
User Manual
Chapter 2 Installation
To ensure the normal operation of the Analyzer after the installation, the Analyzer shall be installed by the
authorized operators of the manufacturer at the delivery.
D
(3)The distance from the front instrument to other instruments shall not be shorter than 100cm.
LT
(4)Sufficient space shall be guaranteed on the operation platform or under the Analyzer for placing collecting
devices of diluent, reagent and waste liquid.
.,
2.1.2 Power requirements
(1)Supply voltage: 100-240V~, 50/60Hz. CO
(2)Power consumption: 260VA.
AL
into the same receptacle with electrical equipment with a heavy load such as air conditioner, refrigerator and oven.
TR
d)The Analyzer shall be put in a dust-free environment without mechanical vibration, source of large noise and
RU
power interference.
e)It is recommended to evaluate the electromagnetic environment in laboratory before operating the Analyzer.
DI
f)Do not have the Analyzer closing to the interference source of high electromagnetic, lest it may interfere with
the normal operation of the Analyzer.
g)Do not put the Analyzer near brush motor, scintillant fluorescent lamp and electrical contact equipment often
used.
h)The Analyzer shall be prevented from direct sunlight exposure and not be placed near heat and wind sources.
i)The environment shall be well-ventilated and the ventilating device shall be used if necessary. But the Analyzer
should be protected from the direct airflow, otherwise, the test accuracy may be affected.
If the operating environment or power supply of the Analyzer do not meet the requirements above, the
accuracy and precision of test results of the Analyzer may be affected, the Analyzer may be damaged or
personal injury may be caused.
(2)Safety conditions
a)Indoor use;
b)Altitude shall not exceed 2000m.
2-1
User Manual
2.2 Unpacking
2.2.1 Unpacking steps
After the instrument arrives, please open the case according to the following steps:
(1)Make the anti-inclination sign on the packing case upright. Verify that the package is complete, and the
appearance of the package does not allow sharp penetration injury and heavy extrusion leading to deformation
which causes the loss of protective function of the packaging case. If there is any damage, please contact the
manufacturer or local agent.
D
(2)Take out the host and accessories; check whether the host and accessories are complete according to the packing
LT
list. If there is any missing, please contact the manufacturer or local agent.
.,
(3)Check the appearance of the system carefully. If there is any damage, please contact the manufacturer or local
agent.
2.2.2 Handling method CO
(1)Use transport machine such as utility trolley for the stable transportation of short distance.
AL
(2)Prevent the aspiration probe from other objects and being damaged during handling and transportation processes.
(3)Keep the Analyzer vertical, not tilting or side laying when moving and carrying.
I
a)Unscrew three fixing screws for the right door, remove the right door of the instrument, and then unscrew the
IN
right fixing screw for the front door, as shown in the following figure:
I
RU
DI
Fig. 2-2-1
b)Open the left door of the instrument and unscrew the left fixing screw of the front door, as shown below:
2-2
User Manual
Fig. 2-2-2
c)Open the front door of the instrument as shown in the figure:
D
LT
.,
CO
I AL
S TR
DU
IN
Fig. 2-2-3
(2)Remove the rubber plug of the reaction cell as shown in the figure:
I
RU
DI
Fig. 2-2-4
(3)Remove the probe fastener: unscrew the two cross recessed pan head combination screws in the figure below and
remove the probe fastener, as shown in the figure:
2-3
User Manual
Fig. 2-2-5
(4)If a sample feeder is selected, the mixing unit fastener needs to be removed: open the front door, and screw down
the cross recessed pan head combination screwfor fixing the isolation sleeve.
Remove theisolation sleeve; then screw down the two cross recessed pan head combination screws that fix the
D
LT
fastener and remove the fastener, as shown in the following figure:
.,
CO
I AL
S TR
DU
Fig. 2-2-6
IN
Do not disassemble or install the instrument except for the normal maintenance.
DI
2-4
User Manual
D
LT
1 Waste liquid level sensor port 2 BF-5D diluent port 3 Waste liquid port
.,
Fig. 2-3-1 Connection of the instrument and reagent
CO
I AL
S TR
DU
IN
I
RU
DI
1 BF- FDOⅠ lyse port 2 BF-FDTⅠ lyse port 3 BF-CRP lyse port
4 BF-FBH lyse port 5 C-reactive protein reagent port
Fig. 2-3-2 Connection with reagent inside the instrument
(1)Connection with lyse, diluent and waste liquid
Open the left side door of the instrument and place the bottles of BF-FDT I lyse, BF-FDO I lyse, BF-FBH lyse,
BF-CRP lyse and C-reactive protein reagent inside the instrument and connect them in accordance with signs in Fig.
2-3-2.
Place the BF-5D diluent tank and waste liquid tank of the instrument under the operation platform and connect them
in accordance with signs in Fig. 2-3-1.
● The height difference between the waste liquid tank and the waste liquid port of the instrument shall not be
2-5
User Manual
Connect "the network interface" of the computer with " " (3 in Fig. 1-5-3) on the right panel of the Analyzer.
(4)Connection of power cable
Plug one end of the accompanying power cable into the power socket on the right panel of the Analyzer (10 in Fig.
1-5-2), and plug the host power cable, display power cable and printer power cable.
D
LT
● The receptacle connected with the power cable shall be reliably grounded.
● Do not put the instrument at a place where the disconnecting device is hard to be operated.
.,
(5)Connection of bar code reader
Connect another end of accompanying bar code reader with any "
CO " interface on the instrument.
AL
The bar code reader will emit light harmful to people's eyes, please do not look straight at the light when the
I
Analyzer is working.
TR
(6)Connection of printer
S
Use a data cable to connect the printer and the " " interface of the Analyzer correctly and check the
DU
Plug one end of the printer power cable to the supply socket of the printer and plug the other end to the socket.
I
In the transportation process, to prevent damage to the aspiration probe, moving mechanisms of the instrument have
RU
been fixed when the instrument is delivered. Before the instrument is powered on, the fixing screws and tie shall be
removed first, or the aspiration probe may be damaged.
DI
2-6
User Manual
D
LT
.,
CO
I AL
Fig. 2-4-1
TR
In the login dialog box, input the correct user name and password, the initial user name of the instrument is admin,
and the initial password is 1, if a wrong user name and password are input, login failure will appear, and the
S
Fig. 2-4-2
The user name and the password are composed of lower and upper case English letters or figures (1-12
characters).
After logging in, the instrument enters the initialization interface to initialize. It performs version verification,
transmission gain, mechanical reset, sending mode, reagent residue detection, pipeline rinsing, background
detection, temperature detection and blank test.
● The blank result obtained during the blank test is beyond the range set by the software and the instrument
will perform the blank test again. If the result is still not qualified after tested 3 times, "blank test error" will
be prompted.
● The instrument will not have High valve, Low value or Suspicious alarms for the blank test results.
● When the power-on self-check alarm triggers, the instrument enters the main interface. Click the red
2-7
User Manual
exclamation mark to check corresponding alarm information. After the alarm is cleared, the instrument
performs self-check again.
After the instrument finishes the self-check, enter the main window of the software, as shown in the figure below:
D
LT
.,
CO
I AL
Fig. 2-4-3
TR
2.4.2.1 Menu
DU
Click the [Menu] to pop up the menu as shown in the figure. The functions in the menu bar are described as follows:
IN
I
RU
DI
Fig. 2-4-4
Spl. Reg.: Have sample registration and work list input.
Calibration: Calibrate the instrument.
Q.C.: L-J/Xbar, Xbar-R, X-B quality control and CRP quality control.
Maint.: Check the status of the instrument and have maintenance and check of the instrument.
2-8
User Manual
Fig. 2-4-5
(1)Total number of locations: Display the current sample serial number and total sample size
(2)Current status: Display the analysis mode of the sample.
(3)Test status: Indicate the current status of the instrument.
D
2.4.2.3 Status indication area
LT
On the right side of Fig. 2-4-5 is the status indication area. The area displays in sequence the X-B QC switch status,
.,
U disk connection status, LIS system connection status, and printer status, described as follows:
(1)X-B QC switch status:
CO
: The X-B quality control is on.
AL
(2)USB status:
S TR
Fig. 2-4-6
(1)Spl. analysis: Finish sample test and the analysis and display of test results on the interface.
(2)History: The instrument will automatically save the analysis results in the database after executing sample
analyzing each time. The operator can query all sample results stored in the database.
(3)QC: Enter the QC interface to realize the basic setting of QC and QC counting.
2-9
User Manual
(4)Add diluent: Dispense diluent for the preparation of pre-dilution sample under the pre-dilution mode.
(5)Rinse: Manually clean equipment.
(6)Print: Print the reports of selected samples.
(7)Fault information area:
In case of failures of the instrument, relevant failure level information will be displayed in the area. At the moment,
click the area to pop up a failure display dialog box, as shown in the figure below:
D
LT
.,
CO
I AL
S TR
DU
Fig. 2-4-7
After clicking relevant failure information, the detailed solution to the failure will be displayed in the "Alarm help"
IN
and the user can have simple troubleshooting according the solution.
2.4.2.5 Info prompt area
I
RU
DI
Fig. 2-4-8
(1)Interface location: display the current interface.
(2)Prompt message: instrument abnormal info display area.
(3)Auto-rinse times: whether the instruments reaches the number of times for rinse.
When the instrument has abnormal info, it will be displayed in the area. Click the prompt message and the following
prompt will pop up:
2-10
User Manual
D
LT
.,
Fig. 2-4-9
CO
2.5 Gas & liquid path diagrams
I AL
S TR
DU
IN
I
RU
DI
Fig. 2-5-1
2-11
User Manual
D
LT
.,
Fig. 2-5-2
CO
I AL
TR
S
DU
IN
I
RU
DI
Fig. 2-5-3
2-12
User Manual
D
LT
.,
CO
I AL
Fig. 2-5-4
TR
S
DU
IN
I
RU
DI
2-13
DI
RU
I
IN
DU
S
2-14
TR
I
User Manual
AL
CO
.,
LT
D
User Manual
3.1 Overview
The system parameters of the instrument have had initialization setting in the factory. The interface shown at the
first turn-on of the instrument is the default interface. To meet different demands in actual use, the software sets two
identities, user and administrator. Users with different identities can reset different parameters.
D
LT
.,
CO
I AL
S TR
DU
IN
I
Fig. 3-2-1
RU
(1)Next No.: Enter the number of the next sample to be analyzed in the input box.
DI
(2)Next bar code: What is input in the box is the barcode number of the next manual analysis sample.
The sample number can only be entered in figure or "-", and its beginning and ending must be numeric and
up to 12 digits.
(3)Analysis mode: There are totally four modes, among which the [WB], [Micro-WB] and [Pre-dilute] modes are
manual analysis modes and [Auto-WB] mode is an automatic analysis mode.
The automatic-whole blood mode can only be used during the assembly of the sample feeder.
(4)Test mode: there are totally 5 modes;
"CBC" mode: count only, with no classification of the white blood cells. The count result comprises of histograms
of the RBC and PLT and other parameters.
"CBC+DIFF" mode: in addition to counting, it also has five differential classifications of the white blood cells,
including 25 parameters and scattergrams and histograms.
“CBC+CRP”: on the basis of the “CBC” mode, the CRP parameters of the sample are analyzed at the same time.
3-1
User Manual
“CBC+DIFF+CRP” mode: it analyzes the CRP parameter of sample on the basis of “CBC+DIFF” mode.
“CRP” mode: it only analyzes the CRP parameter of sample.
(5)Tube number and rack number: If the [Auto-WB] mode is selected and [Auto-identify] is not selected, the [Rack
No.] and [Tube No.] will change to an editable state and the rack number and tube number can be reset in the box
following [Rack No.] and [Tube No.].
(6)Use worklist: use the manually entered or downloaded worklist for testing.
(7)Two-way LIS: when bi-directional LIS is set, after the testing is started, the instrument will obtain a worklist
from LIS terminal automatically and then conduct test. After the testing, the worklist information and test results
will be transmitted to LIS terminal.
(8)Auto-COM.: After connection to the LIS, automatically transmit the test results after the testing.
If no barcode reader is connected, the “Two-way LIS” option will not be displayed on the interface.
D
LT
.,
CO
I AL
S TR
DU
IN
I
RU
DI
Fig. 3-3-1
Click the corresponding input box of the unit following each test item and select a different unit as the test unit
according to actual demand.
After inputting, click [Save] to confirm, and when "Saved successfully" is prompted, press [OK] on the interface.
Click [DEF] to restore the unit to the default settings.
Unit of each test item: several units for the setting of each test item of the instrument are available for the user to
choose and specific setting unit is as shown in Table 3-3-1:
3-2
User Manual
Table 3-3-1
103/μL ***.**
WBC
2
10 /μL ****.*
/nL ***.**
106/μL **.**
RBC
104/μL ****
/pL **.**
D
LT
g/L *** Default unit
.,
HGB g/dL **.*
mmol/L **.*
fL
CO
***.* Default unit
MCV, RDW-SD
AL
μm3 ***.*
I
MCH
fmol ***
S
mmol/L ***.*
% **.*
HCT
DI
103/μL ****
PLT, P-LCC
4
10 /μL ***.*
/nL ****
3-3
User Manual
After the unit of the parameter is changed, the data format of the test result also changes.
D
LT
.,
CO
I AL
S TR
DU
IN
I
RU
DI
Fig. 3-4-1
Input of reference value lower and upper limits:
Click the corresponding input box of lower and upper limits following each test item and directly input the lower
and upper limits of corresponding test item.
After the limits are input, press [Save] for confirmation. When a prompt information of "Saved successfully" pops
up, click [OK] on the interface to save.
Click [DEF] to restore the default lower and upper reference values.
3-4
User Manual
D
LT
.,
Fig. 3-5-1
CO
Select [RBC/PLT] and the interface is as shown in the figure below:
I AL
S TR
DU
IN
I
RU
DI
Fig. 3-5-2
Select [Suspicious mark], as shown in the figure below:
3-5
User Manual
D
LT
.,
Fig. 3-5-3
CO
On this interface, the user can set the prompt range of WBC, RBC/ PLT and the abnormal alarm information of
AL
suspicious marks.
I
After the values are input, click [Save] to save the settings or click [DEF] to restore the default value.
TR
3.6 QC setting
S
3.6.1 QC method
DU
3-6
User Manual
D
LT
.,
Fig. 3-6-1
CO
The user can select the QC mode. After saving the settings, the user can click "QC" on the shortcut key interface to
AL
Fig. 3-6-2
In the figure above, users can select the L-J QC calculation method and range:
3-7
User Manual
D
LT
.,
CO
I AL
S TR
DU
Fig. 3-6-3
IN
In the figure above, users can select the Xbar QC calculation method and range:
Calculation and range selection of deviation:
I
(1)If [Absolute value] is selected as the calculation method, the deviation limit input will be displayed in the form of
RU
an absolute value, and 2SD or 3SD will be displayed on [Range] as deviation limit.
DI
(2)If [Percentage] is selected as the calculation method, the deviation limit input will be displayed in the form of a
percentage, and 2CV or 3CV will be displayed on [Range] as deviation limit.
3.6.4 X-B setting
Select [X-B] and the interface is as shown in the figure below:
3-8
User Manual
D
LT
.,
Fig. 3-6-4
CO
The user can choose X-B QC switch and set sample volume.
AL
Fig. 3-6-5
In the figure above, users can select the CRP QC calculation method and range:
Calculation and range selection of deviation:
3-9
User Manual
(1)If [Absolute value] is selected as the calculation method, the deviation limit input will be displayed in the form of
an absolute value, and 2SD or 3SD will be displayed on [Range] as deviation limit.
(2)If [Percentage] is selected as the calculation method, the deviation limit input will be displayed in the form of a
percentage, and 2CV or 3CV will be displayed on [Range] as deviation limit.
D
LT
.,
CO
I AL
S TR
DU
Fig. 3-7-1
IN
(1)Department: Click [Add] to add department information in the popup interface as shown in the figure below:
I
RU
DI
3-10
User Manual
D
LT
.,
Fig. 3-7-2
CO
Select the row where the department information to be deleted is, click [Delete] in the figure to pop up the prompt as
AL
Fig. 3-7-3
RU
3-11
User Manual
D
LT
.,
Fig. 3-7-4
CO
Click [Add] to add doctor information in the popup interface as shown in the figure below:
I AL
S TR
DU
IN
I
RU
DI
Fig. 3-7-5
Select the row where the doctor information to be deleted is, click [Delete] in the figure to pop up the prompt as
shown in the figure below:
3-12
User Manual
Fig. 3-7-6
Click [OK] to delete corresponding doctor information.
(3)Cost type: Select [Cost type], the interface shown in the figure below:
D
LT
.,
CO
I AL
S TR
DU
IN
I
RU
Fig. 3-7-7
Click [Add] to add charge information in the popup interface as shown in the figure below:
DI
3-13
User Manual
D
LT
.,
Fig. 3-7-8
CO
Select the row where the charge information to be deleted is, click [Delete] in the figure to pop up the prompt as
AL
Fig. 3-7-9
RU
When adding the department, doctor, and cost type, it is not allowed to input special characters.
3-14
User Manual
D
LT
.,
Fig. 3-8-1
CO
(1)Add user: Click [Add user] in the figure to add user information as shown in the figure below:
I AL
S TR
DU
IN
I
RU
DI
Fig. 3-8-2
Fill in the user, name, password, confirm password and select permission according to the corresponding prompt
and click [Save] to complete the operation.
3-15
User Manual
Fig. 3-8-3
Click [OK] to delete corresponding user information.
D
LT
The currently logged in user cannot delete itself.
.,
(3)Change password: In Fig. 3-8-1, select the row of the user information that you want to change your password
and click [Modify PWD], as shown in the figure below:
CO
I AL
S TR
DU
IN
I
RU
DI
Fig. 3-8-4
In the figure, enter the old password and new password respectively, and click [Save] to save the modified
password.
3-16
User Manual
Fig. 3-8-5
Click [OK] to reset the password. After the reset is successful, the new password is the same as the user name.
D
LT
.,
CO
I AL
S TR
DU
IN
I
RU
DI
Fig. 3-9-1
After [Time], you can set the instrument sleep time in [60,120] minutes.
3.9.2 Automatic rinsing setting
Click [Menu] - [Settings] - [Machine] and select [Auto-rinse], as shown in the figure below:
3-17
User Manual
D
LT
.,
Fig. 3-9-2
CO
Input the number of automatic rinsing intervals 10-200 in the input box following "Count number" under Auto-rinse
AL
times.
Input the number of automatic rinsing intervals 1000-5000 in the input box following "Count number" under
I
Click [Menu] - [Settings] - [Machine] and select [Blocked hole], as shown in the figure below:
DU
IN
I
RU
DI
Fig. 3-9-3
3-18
User Manual
Set if re-test is required after hole blockage and if hole blockage data is to be transmitted. After the setting
completion, click [Save] at the lower right corner. When the message “Saved successfully” is prompted, click [OK].
3.9.4 Other
Click [Menu] - [Settings] - [Machine] and select [Other], as shown in the figure below:
D
LT
.,
CO
I AL
TR
Fig. 3-9-4
S
Select whether to support 1.5mL centrifuge tube from the pull-down menu. The default is Not Support.
DU
3-19
User Manual
D
LT
.,
Fig. 3-10-1
CO
The date display format includes the following three types:
AL
yyyy-MM-dd, dd-MM-yyyy and MM-dd-yyyy; select a common date format; after that, click [Save]; when "Saved
I
When the time is modified, it is synchronized directly for saving without requiring clicking [Save].
S
DU
The changed date format will be displayed at all the time display positions (such as sending time and
sampling time).
IN
Click [Menu] - [Settings] - [Software] - [Print] and the interface is as shown in the figure below:
RU
DI
3-20
User Manual
D
LT
.,
Fig. 3-10-2
CO
(1)Auto print: select "Yes" or "No" to decide whether to automatically print the report when the test results are
AL
returned.
I
(2)Printer: select the model of connected external printer from the pull-down menu.
TR
(3)Printing template: select the printing template from the pull-down menu: "Horizontal version" or "Vertical
version";
S
(4)Print pictures: select "Yes" or "No" from the pull-down menu to decide whether to print the histogram.
DU
(5)Print format: select a piece of appropriate paper from the pull-down menu.
IN
3-21
User Manual
D
LT
.,
Fig. 3-10-3
CO
After the setting of the IP and port of the LIS system and the IP and subnet mask of the equipment network is
AL
completed, click [Save]. When the message “Saved successfully” is prompted, click [OK].
I
LIS connection: after the network is set, connect the LIS server.
TR
Fig. 3-10-4
3-22
User Manual
After selecting the corresponding language category and clicking [Save], the message “Saved successfully and the
instrument will restart” will pop up on the screen. After clicking [OK], the software will restart and the language
will be switched successfully.
D
LT
.,
CO
I AL
S TR
DU
IN
Fig. 3-10-5
Start sample number: the initial sample number to test each time after start.
I
Abnormal value icon setting: The user can select high and low value flags, including the use of "H, L" and "↑, ↓".
RU
After clicking [Save], when the message “Saved successfully” is prompted, click [OK].
DI
3-23
User Manual
D
LT
.,
Fig. 3-11-1
CO
HS-CRP: whether to test hypersensitive CRP; the default is off, not to test.
AL
CRP calibration: input the corresponding value in the “Sensitivity check value” input box; the input range is
I
0~999999.
TR
“Sensitivity check value” is used in the CRP multi-point calibration sensitivity check. For details refer to 4.8.
After clicking [Save], when “Saving succeeded” is prompted, click [OK].
S
DU
3-24
User Manual
Chapter 4 Calibration
4.1 Overview
The instrument is calibrated to ensure the accuracy of analysis result. The instrument must be calibrated before the
measured data can be used as valid data.
If the calibration fails, check whether the input parameters, calibrator used and reagents are correct, or
operate according to the alarm message and relevant prompts.
D
LT
4.3 Calibration method
.,
The instrument has three calibration modes: manual calibration, calibrator calibration and fresh blood calibration.
In calibrator calibration and fresh blood calibration, all calibration-related math calculations are performed
CO
automatically by the instrument. The calibration coefficient obtained after the calibration is saved on the “Manual”
interface.
AL
Preparation before calibration: check according to the following steps before the calibration. In case of any problem,
do not perform calibration. Contact the After-sales Service Department of the manufacturer of agent.
I
(1)Check the instrument and reagent to ensure that the reagent volume is sufficient to complete the whole calibration
TR
4-1
User Manual
D
LT
.,
Fig. 4-4-1
CO
4.4.1 Input of calibrator reference value
AL
Click the "Ref." under each item, directly input the reference value of each parameter (refer to the user manual for
I
a)If the instrument has a sample feeder, put a test tube with mixed calibrator into the STAT chamber.
b)If the instrument does not have a sample feeder, put the test tube with mixed calibrator below the sampling
IN
other operating interfaces. The test status of current calibration will be displayed in the status column at the bottom.
RU
(3)After the test is completed, the interface is unlocked and the result is displayed in the test result list. The current
record is selected by default "√".
DI
If a certain parameter in the obtained count results exceeds the valid range, the prompt will be popped up as shown
in the figure below:
Fig. 4-4-2
The current calibration count results are not saved and the current record is not selected by default. Count the
calibration again to overwrite the result.
(4)When the calibration is not completed (the effective calibration count does not reach 5 times), the software will
prompt the user to switch to the other interface as shown below:
4-2
User Manual
Fig. 4-4-3
When the effective calibration count reaches 5 times, the instrument will automatically calculate the average Mean,
CV%, and the new calibration coefficient for all the calibration data selected by default “√” according to the
formula.
Operators can select some groups of data to calculate the calibration coefficients, but the calibration coefficients can
only be calculated after selecting at least 5 groups of data. Each time the operator click the check box “√” to select or
cancel the “√” data, the calibration coefficient is refreshed and displayed in time.
D
Calibrate calibration is only performed in the whole blood mode.
LT
4.4.3 Save calibration coefficient
.,
Click [Save] in Fig. 4-4-1. The followings may occur:
CO
(1)When the effective calibration count does not reach 5 times, the prompt is as shown in the figure below:
I AL
S TR
DU
IN
Fig. 4-4-4
(2)When the CV% value of a calibration parameter exceeds the repeatability index of the instrument, the prompt is
I
Fig. 4-4-5
(3)When the new calibration coefficient of a certain parameter is not between 75% and 125%, the prompt is as
shown in the figure below:
4-3
User Manual
Fig. 4-4-6
In cases 2 and 3, the operator needs to check whether the input reference value is correct and start calibration
counting again.
(4)If the calculated calibration coefficients are all between 75% and 125%, and the CV% values of all the calibration
parameters do not exceed the repeatability index of the instrument, the prompts are as shown in the figure below:
D
LT
.,
Fig. 4-4-7
CO
Click [OK], and the calibration coefficient will be automatically saved as the calibration coefficient on the
AL
Fig. 4-4-8
4-4
User Manual
D
LT
.,
CO
I AL
S TR
DU
IN
I
RU
Fig. 4-5-1
(1)Select the analysis mode for whole blood or pre-dilution.
DI
(2)Select the number at the "Sample ID" before each test of fresh blood samples.
(3)Put a test tube:
a)If the instrument has a sample feeder, put a test tube with mixed fresh blood into the STAT chamber.
b)If the instrument does not have a sample feeder, put the test tube with mixed fresh blood below the sampling
probe to ensure the sampling probe can aspirate the calibrator.
(4)Press "Count" on the instrument to start calibration test. Then interface will be locked and the user cannot go to
other operating interfaces. The test status of current calibration will be displayed in the status column at the bottom.
(5)After the test is completed, the interface is unlocked and the result is displayed in the test result list. The current
record is selected by default "√".
If a certain parameter in the obtained count results exceeds the valid range, the prompt will pop up as shown in the
figure below:
4-5
User Manual
Fig. 4-5-2
The current calibration count results are not saved and the current record is not selected by default. Count the
calibration again to overwrite the result.
(6)When the effective calibration count of each portion of fresh blood reaches 5 times, the instrument will
automatically calculate the average Mean, CV%, and the new calibration coefficient for all the calibration data
selected by default "√" according to the formula. Operators can select some groups of data to calculate the
calibration coefficients, but the calibration coefficients can only be calculated after selecting at least 5 groups of data.
Each time the operator click the check box "√" to select or cancel the "√" data, the calibration coefficient is refreshed
and displayed in time.
D
(7)If the calibration coefficients calculated of the current number of fresh blood samples are between 75% and 125%,
LT
and the CV% values of all calibration parameters do not exceed the repeatability index of the instrument. Select the
"Sample number" again and test the remaining fresh blood samples.
.,
4.5.3 Save calibration coefficient
CO
After the instrument has tested 3 or more fresh blood samples and obtained valid calibration coefficients, click
[Calculate Coef.] to display the content as shown in the figure below:
I AL
S TR
DU
IN
I
RU
DI
Fig. 4-5-3
When less than 3 portions are selected, the average calibration coefficient will not be calculated. When 3 or more
portions are selected, the average calibration coefficient will be automatically calculated. Click [Save] to save the
current calibration coefficient.
When there are less than 3 portions of fresh blood for test, click [Calculate Coef.] and a prompt as shown in Fig.
4-5-4 will appear:
4-6
User Manual
Fig. 4-5-4
After selecting the calibration coefficient of more than 3 fresh blood samples, click [Save] to pop up the dialog box
as shown in the figure below:
D
LT
.,
Fig. 4-5-5
CO
Click [OK], and the calibration coefficient will be automatically saved as the calibration coefficient on the "Manual"
interface.
After calibrator calibration or fresh blood calibration is executed and correctly saved, the obtained calibration
I
Fig. 4-6-1
4-7
User Manual
When fine adjustment of calibration coefficient is required, click the corresponding row of each item under
[Cal.Coef.], directly input the calibration coefficient and click [Save]. When a dialog box of "Saved successfully"
pops up on the screen, click [OK].
D
LT
.,
CO
I AL
S TR
Fig. 4-7-1
DU
(2)Prepare two concentrations of calibrators and enter the corresponding reference values in the "Ref." edit boxes of
the calibrators.
IN
(3)Aspirate calibrator:
a)If the instrument is equipped with a sample feeder, put the prepared first concentration calibrator into the
I
emergency chamber, click the "Count" button on the sample feeder, and the instrument automatically performs
RU
b)If the instrument is not equipped with a sample feeder, place the prepared first concentration calibrator under
the aspiration probe so that the aspiration probe can aspirate the calibrator. Click the "Count" button to perform
calibration count 4 times.
(4)After the counting process of the first concentration calibrator is completed, the second concentration calibrator
calibration count is performed according to step (3).
After the second concentration calibrator is counted, the system prompts "Scaling successfully" to save and display
the new calibration curve.
Fig. 4-7-2
4-8
User Manual
(5)If automatic calibration is not completed, the display is as shown below when the interface is switched:
Fig. 4-7-3
Click "OK" to exit CRP 2-point calibration and relevant data will not be saved.
● If the instrument is equipped with a sample feeder, a tube adapter is required for calibration.
● Each calibrator must be counted in order from lowest to highest concentration.
● When executing CRP two-point calibration, blank concentration and a concentration calibrator can be
D
used, or two concentrations of calibrators other than blank can be selected.
LT
● If the calibrator count results do not increase progressively with concentrations, the system prompts "The
current calibration data is invalid"; and the calibration operation should be performed again.
.,
4.8 CRP multi-point calibration CO
(1)Go to the [Calibration] menu, click [CRP multiple-point calibration], and the interface below will appear:
I AL
S TR
DU
IN
I
RU
DI
Fig. 4-8-1
(2)Prepare calibrators of five concentrations according to the user manual for calibrators, and successively input the
reference value for each calibrator in the order from low concentration to high concentration in the "Ref." edit box of
the calibrator.
a)If the instrument is equipped with a sample feeder, click the "Count" button on the sample feeder, and the
instrument will automatically perform blank count 4 times.
4-9
User Manual
b)If the instrument is equipped with a sample feeder, click the "Count" button to perform blank count 4 times.
(3)After the blank count is completed, prepare the calibrators according to the order of the calibrator concentration
from low to high, and perform calibration counting:
a)If the instrument is equipped with a sample feeder, put the prepared calibrator into the emergency chamber,
click the "Count" button on the sample feeder, and the instrument automatically performs calibration count 4
times.
b)If the instrument is not equipped with a sample feeder, place the prepared calibrator under the aspirating probe
so that the aspirating probe can aspirate the calibrator. Click the "Count" button to perform calibration count 4
times.
(4)The prepared calibrators are used for calibration count in order of concentration from low to high according to
step (3).
After the counting process of the highest concentration calibrator is completed, the system prompts "Scaling
successfully" to save and display the new calibration curve.
D
LT
.,
Fig. 4-8-2
CO
(5)If automatic calibration is not completed, the display is as shown below when the interface is switched:
I AL
S TR
DU
IN
Fig. 4-8-3
I
RU
Click "OK" to exit automatic calibration and relevant data will not be saved.
(6)If the difference in reactivity between the highest concentration calibrator and the blank concentration calibrator
DI
is less than the set value of the instrument sensitivity check, the display is as shown:
Fig. 4-8-4
Click [Yes] to save the current calibration curve; click [Cancel], the current calibration curve is not saved.
● If the instrument is equipped with a sample feeder, a tube adapter is required for calibration.
● Each calibrator must be counted in order from lowest to highest concentration.
4-10
User Manual
● If the calibrator count results do not increase progressively with concentrations, the system prompts "The
current calibration data is invalid"; and the calibration operation should be performed again.
D
LT
.,
CO
I AL
S TR
Fig. 4-9-1
DU
(2)Blank counting
IN
a)If the instrument has a sample feeder, click Count on the sample feeder and the instrument will automatically
execute blank counting for 4 times.
b)If the instrument does not have a sample feeder, click "Count" and execute blank counting for 4 times.
I
RU
(3)After the blank counting is completed, a prompt of "Scaling successfully" is given by the system and the new
calibration curve is saved.
DI
Fig. 4-9-2
(4)If automatic calibration is not completed, the display is as shown below when the interface is switched:
4-11
User Manual
Fig. 4-9-3
Click "OK" to exit CRP blank calibration and relevant data will not be saved.
D
LT
.,
CO
I AL
S TR
DU
IN
I
RU
DI
Fig. 4-10-1
When fine adjustment of calibration coefficient is required, click the corresponding line of each item under
[Concentration] or [Reaction degree] and directly input the value. Click [Save] to save the data successfully.
4-12
User Manual
D
LT
.,
Fig. 4-11-1
CO
The calibration date, calibration person, calibration method and calibration mode are displayed in the order of the
AL
list.
I
Condition query: the operator can also screen calibration history according to the calibration method.
TR
4-13
DI
RU
I
IN
DU
S
4-14
TR
I
User Manual
AL
CO
.,
LT
D
User Manual
Chapter 5 QC
5.1 Overview
There may be a certain degree of error in the long-term use of the instrument, and the existence of error is likely to
lead to incorrect or unreliable analysis results. The quality control program provides an effective method of
detecting possible errors, and the quality control test allows the operator to test the accuracy and precision of daily
sample analysis results.
To ensure the reliability of the sample analysis results, it is suggested that the operators conduct quality control on
each instrument once with QC materials at low, middle and high levels respectively.
The instrument must be properly calibrated to perform a quality control test.
The instrument provides five quality control methods, i.e. L-J QC, QC, Xbar QC, X-B floating mean method QC,
X-R QC and CRP QC.
● The user must use the QC materials specified by the manufacturer, otherwise it will lead to inaccurate QC
results.
D
● The use of QC materials is detailed in its instructions.
LT
● If a quality control test is conducted in the event of an instrument failure, the results obtained are
.,
unreliable.
● If QC is out of control, check if the input parameters, QC products used and reagents are correct, or
operate according to the alarm message and relevant prompts. CO
5.2 L-J QC
AL
Operators can conduct QC for 24 parameters under L-J QC. The instrument provides 12 QC files in total to save QC
I
parameters and results. Each QC file can automatically save up to 31 groups of QC results. When QC is conducted
TR
for more than 31 times, the new QC results will overwrite the previous results.
Click [Menu] - [Q.C.] - [L-J], as shown in the figure below:
S
DU
IN
I
RU
DI
Fig. 5-2-1
5-1
User Manual
5.2.1 Setting
5.2.1.1 Lot number information setting
Click [Setting], as shown in the figure below:
D
LT
.,
CO
I AL
Fig. 5-2-2
TR
Click the file number pull-down menu and select the file to be conducted with QC whose range is 1~12.
DU
Input the lot number of the corresponding QC product in the input box according to the User Manual of QC blood.
(3)Life setting:
Input the life of QC product in the input box according to the User Manual of QC.
I
RU
(4)Selection of QC level:
Click the pull-down menu and select a QC level (including high, medium and low levels), and every QC lot number
DI
Lot number and life input shall conform to those noted in the user manual.
5.2.1.2 Take preset value
Click [Use DEF] to input the reference value and deviation limit at corresponding level saved in the system into the
current QC file.
If the reference value and deviation limit at corresponding level are not saved in the system, click [Use DEF] and
then a prompt dialog box below will pop up.
5-2
User Manual
Fig. 5-2-3
After taking the preset value is finished, click [Save] to pop up the "Saved successfully" information, and click
[OK].
5.2.2 QC count
Click [QC count] in Fig. 5-2-2 and display will be as shown below:
D
LT
.,
CO
I AL
S TR
DU
IN
I
RU
Fig. 5-2-4
DI
(1)QC counting
a)If the instrument has a sample feeder, place the test tube with mixed QC object in the STAT chamber and click
Count, then the instrument will start QC counting and analysis. Meanwhile, display the QC test status
successively in the operating status column at the bottom of the interface.
b)If the instrument does not have a sample feeder, place the test tube with mixed QC object below the sampling
probe to make the sampling probe aspirate the sample, then click Count and the instrument will start QC counting
and analysis. Meanwhile, display the QC test status successively in the operating status column at the bottom of
the interface.
(2)After the analysis, the counting results will be displayed on the screen, as shown in the figure below:
5-3
User Manual
D
LT
.,
Fig. 5-2-5
CO
(3)Printing of QC data
AL
Click [Print] to print the current record if single QC result needs to be printed.
I
5.2.3 QC diagram
TR
Click [QC diagram] in Fig. 5-2-5 and display will be as shown below:
S
DU
IN
I
RU
DI
Fig. 5-2-6
(1)QC diagram query: select to view item data.
5-4
User Manual
D
a)Each point in QC chart corresponds to a QC result and the points are connected by line sections.
LT
b)"Green solid circle" means that the QC result is under control, while "Red hollow circle" means that the QC
.,
result is out of control.
(4)If points in the QC diagram are beyond the QC range, deal with them in accordance with the steps below:
CO
a)Check whether the target value and deviation limit of QC material are correct.
b)Whether the background test of the instrument is normal.
AL
c)If the above two points are all normal, calibrate the instrument and then perform QC and count.
I
d)If the instrument remains abnormal after calibration, please contact the after-sales service department of
TR
manufacturer or agent.
(5)Print QC diagram:
S
DU
Operator can calculate Mean (mean), SD (standard deviation) and CV% (coefficient of variation) in QC diagram
RU
5-5
User Manual
D
LT
.,
Fig. 5-2-7
CO
b)Click the selected abnormal QC point and the red cursor line will stop on the abnormal QC point. Click [Delete]
AL
and the selected abnormal QC point will change to "black point". When the preset value is taken for calculation,
the removed QC point (black QC point) is not included.
I
TR
c)After removing the invalid point, click [Save DEF] and the instrument will save the preset value which can be
used to calculate current level into the system.
S
d)If there are less than 3 valid QC points, click [Save DEF], and a prompt box will pop up on the interface as
DU
Fig. 5-2-8
If the operation above has error, click the screen to move the red cursor line to the removed point, and black QC
point will turn to be red or green through clicking [Add], then the removed point will recover to normal.
When calculating preset value, adopt data of valid QC point as reference value and deviation limit.
5.2.4 QC list
Users can select different QC files to browse QC data.
Click [QC list] and select the QC file number to be queried from the pull-down menu of [File No.], as shown in the
figure below:
5-6
User Manual
D
LT
.,
Fig. 5-2-9
CO
(1)Delete all: the operator can delete all QC counting results in the current QC file.
AL
Fig. 5-2-10
Click [OK] to delete all the data.
DI
(2)Delete: the operator can delete a QC counting result in the current QC file.
Select the QC data to be deleted and click [Delete] to pop up the prompt box as shown in the figure below:
Fig. 5-2-11
Click [OK] to delete the selected data.
(3)LIS transmission: when connecting LIS, operators can transmit the QC results in the current QC file to the LIS
software.
5-7
User Manual
(4)View data: select to view item data through page up and down; click
to view item data through page left and right.
5.3 Xbar QC
Xbar QC: adopt the mean of two measurement results as a QC point to be saved in the QC diagram.
Other operations are the same as "5.2 L-J QC".
5.4 X-B QC
X-B floating mean method is put forward by Dr. BrianBull. Through stability of red blood cell parameters such as
MCV, MCH and MCHC, the monitoring and control of instrument performance can be achieved. It is a QC method
without QC object and mainly to monitor the instrument performance like quality control of QC object. The two
methods reflect the analysis performance of the instrument from different aspects, but they cannot replace each
other. X-B QC is recommended when the daily sample volume of the instrument is larger than 100. Accompanying
sample is required to be used in the QC method which, as a result, cannot be used for sample classified according to
disease types. Its reference range is composed of given reference value and upper and lower limit. Observe the
change trend of QC result within the reference range.
The instrument conducts X-B QC to three parameters: MCV, MCH and MCHC, and 20-200 samples can be tested
D
in X-B value analysis for each group which come from the result of normal counting without distinguishing
LT
whole-blood from pre-dilution mode.
.,
As the X-B QC reference value is worked out after statistics and analysis of a large amount of accompanying
samples, the instrument shall be calibrated before test.
5.4.1 Setting CO
Before X-B QC, QC parameter shall be set.
AL
Fig. 5-4-1
Input target value and deviation limit: click the table of corresponding target value and deviation limit of parameters
to input.
Target value: as the reference value differs from region to region, sample quantity is required to reach a certain
number (more than 150 is recommended), then the arithmetic mean value of these samples is selected as the target
5-8
User Manual
D
LT
.,
CO
I AL
S TR
DU
IN
Fig. 5-4-2
(1)Descriptions of QC graph:
I
RU
The X-axis of QC diagram shows the number of QC count result in QC file, while the Y-axis shows QC count result
of each parameter.
DI
As for every parameter, the three data on the left of the QC diagram correspond to three boundaries, representing
respectively, from top to bottom, the upper limit, target value and lower limit of the parameter.
Upper limit: QC material reference value + deviation limit
Target: reference value of QC material
Lower limit: QC material reference value - deviation limit
As for each parameter, the three data on the left of QC graph respectively represent Mean (mean), SD (standard
deviation) and CV% (coefficient of variation).
(2)Descriptions of points in QC diagram:
Each point in QC diagram corresponds to a QC result and the points are connected by line sections.
"Green solid circle" means that the QC result is under control, while "Red hollow circle" means that the QC result is
out of control.
(3)If points in the QC diagram are out of the QC range, dispose following the procedures below:
a)Check whether the target value and deviation limit are correct.
b)Whether the background test of the instrument is normal.
c)If the instrument remains abnormal after calibration, please contact the after-sales service department of
5-9
User Manual
manufacturer or agent.
5.4.3 QC list
Click [QC list], as shown in the figure below:
D
LT
.,
CO
I AL
Fig. 5-4-3
TR
(1)Delete: click QC data following the corresponding number to display the selection status in the row, and click
S
Fig. 5-4-4
Click [OK] and the prompt information "Deleted successfully" will pop up on the interface which means that
selected records are successfully deleted.
5-10
User Manual
Fig. 5-4-5
Click [OK] to delete all the records.
(3)Check the data: select , check the item data through page turning.
(4)LIS transmission: click [Transmit LIS] to transfer the data to the LIS system connected with.
5.5 Xbar-R QC
Xbar QC diagram is mainly used to observe the changes of normally distributed mean, and R QC diagram is used to
observe the changes of dispersion or variation of normal distribution while Xbar-R QC diagram integrated with the
D
two diagrams above is used to observe the changes of normal distribution.
LT
Xbar-R QC diagram reflects the differences between batches and days of day-to-day test, achieving the effective
monitoring and control of system factors during the test, thus it is applicable to laboratory with high-level
.,
automation and large inspection amount.
CO
Xbar-R QC: adopt the mean of two measurement results as a QC point saved in the mean QC diagram; adopt the
difference between two measurement results as a QC point saved in the range QC diagram. Every parameter is
displayed in the mean QC diagram and range QC diagram.
AL
Fig. 5-5-1
5.5.1 Setting
Click [Setting], as shown in the figure below:
5-11
User Manual
D
LT
.,
Fig. 5-5-2
CO
(1)Selection of file No.:
AL
Input the lot number of the corresponding QC product in the input box according to the User Manual of QC blood.
S
(3)Life setting:
DU
Input the life of QC product in the input box according to the User Manual of QC.
(4)Selection of QC level:
IN
Select the QC level (including high, medium and low levels) from the "Level" pull-down menu. Each QC lot
number corresponds to one level.
I
After finishing the setting, click [Save] to pop up the "Saved successfully" information, and click [OK].
RU
DI
Lot number and life input shall conform to those noted in the user manual.
5.5.2 QC count
Click [QC count], as shown in the figure below:
5-12
User Manual
D
LT
.,
Fig. 5-5-3
CO
(1)QC counting
AL
a)If the instrument has a sample feeder, place the test tube with mixed QC object in the STAT chamber and click
I
Count, then the instrument will start QC counting and analysis. Meanwhile, display the QC test status
TR
probe to make the sampling probe aspirate the sample, then click Count and the instrument will start QC counting
DU
and analysis. Meanwhile, display the QC test status successively in the operating status column at the bottom of
the interface.
IN
(2)After the analysis, the counting results will be displayed on the screen.
I
RU
Only when the result is 1, it is forbidden to switch the interface. If it is switched, the test result of result 1 will
not be saved.
DI
(3)After the analysis, the sample probe will reset and get prepared for next counting and the counting results will be
displayed on the screen, as shown in the figure below:
5-13
User Manual
D
LT
.,
Fig. 5-5-4
CO
If the counting result worked out is lower than the lower limit set in the instrument software, it will be displayed in
AL
blue, and if higher than the display range of the instrument, it will be displayed in red.
I
(5)Printing of QC data
Click [Print] to print the current record if a single QC result needs to be printed.
IN
5.5.3 QC diagram
Click [QC diagram], as shown in the figure below:
I
RU
DI
5-14
User Manual
D
LT
.,
Fig. 5-5-5
CO
(1)Relevant calculation description:
AL
a)Mean:
I
Xbar
n
S
Where:
DU
Ra Rb Rc Rd Re Rf Rg Rh Ri Rj
Rbar
RU
n
DI
Where:
Ra~Rj refer to the range of every two measurement results;
n refers to the number of range.
(2)Description of QC diagram interface:
If the number of QC count is less than 3, QC result of each parameter will not be displayed on the right side of QC
diagram.
The X-axis of QC diagram shows the number of QC count result in QC file, while the Y-axis shows QC count result
of each parameter.
Vertical bar is used to mark the count data of the same group, and at most 31 points can be displayed in the QC
diagram for every parameter.
As for each parameter, these three data on the left of the QC diagram correspond to three boundary values,
representing respectively, from top to bottom, the upper limit, target and lower limit of the parameter.
QC diagram of every test item includes mean QC diagram and range QC diagram, take WBC for example, and the
display will be as shown below:
a)Mean diagram (WBC M part is displayed in the figure):
5-15
User Manual
D
LT
b)If the background test is normal, re-execute QC count, and if QC point remains abnormal in QC diagram,
conduct QC count after the calibration of instrument.
.,
c)If the instrument remains abnormal after calibration, please contact the After-sales Service Department of the
manufacturer.
5.5.4 QC list
CO
The user can select different QC files to browse QC data.
AL
Click [QC list] and select the QC file number to be queried from the [File No.] pull-down menu, as shown in the
figure below:
I
S TR
DU
IN
I
RU
DI
Fig. 5-5-6
(1)Delete all: the operator can delete all QC counting results from the current QC file. Click [Delete all] in the figure
above, and then the screen is as shown below:
5-16
User Manual
Fig. 5-5-7
Click [OK] to delete all the data.
(2)Delete: the operator can delete a QC counting result from the current QC file.
Select the QC data to be deleted and click [Delete], as shown in the figure below:
D
LT
.,
Click [OK] to delete the selected data.
Fig. 5-5-8 CO
AL
(3)LIS transmission: operators can transmit the selected QC counting results to the LIS system.
(4)Print: the operator can print the QC count result in the current QC file.
I
5.6 CRP QC
Operators can conduct QC to CRP under CRP QC. 12 QC files are provided by the instrument to facilitate saving
I
QC setting parameter and result. Each QC file can automatically save up to 31 groups of QC results. When QC is
RU
conducted for more than 31 times, the new QC results will overwrite the previous results.
Click [Menu] - [Q.C.] - [CRP], as shown in the figure below:
DI
5-17
User Manual
D
LT
.,
Fig. 5-6-1
CO
5.6.1 Setting
AL
Fig. 5-6-2
(1)Selection of file No.:
Select the QC files numbering 1~12 from the pull-down menu.
5-18
User Manual
Lot number and life input shall conform to those noted in the user manual.
5.6.1.2 Take preset value
Click [Use DEF] to input the reference value and deviation limit at corresponding level saved in the system into
current QC file.
D
LT
If the reference value and deviation limit at corresponding level are not saved in the system, click [Use DEF] and
then a prompt dialog box below will pop up.
.,
CO
I AL
S TR
Fig. 5-6-3
DU
After taking the preset value is finished, click [Save] to pop up the "Saved successfully" information, and click
IN
[OK].
5.6.2 QC count
I
Click "QC count" in Fig. 5-6-2 and display will be as shown below:
RU
DI
5-19
User Manual
D
LT
.,
Fig. 5-6-4
CO
(1)QC counting:
AL
a)If the instrument has a sample feeder, place the test tube with mixed QC object in the STAT chamber and click
I
Count, then the instrument will start QC counting and analysis. Meanwhile, display the QC test status
TR
probe to make the sampling probe aspirate the sample, then click Count and the instrument will start QC counting
DU
and analysis. Meanwhile, display the QC test status successively in the operating status column at the bottom of
the interface.
IN
I
If the instrument is equipped with a sample feeder, a tube adapter is required for QC count.
RU
(2)After the analysis, the counting results will be displayed on the screen, as shown in the figure below:
DI
5-20
User Manual
D
LT
.,
Fig. 5-6-5
CO
(3)Printing of QC data
AL
Click [Print] to print the current record in the shortcut key area if a single QC result needs to be printed.
I
5.6.3 QC diagram
TR
Click "QC diagram" in Fig. 5-6-5 and display will be as shown below:
S
DU
IN
I
RU
DI
Fig. 5-6-6
(1)Description of QC diagram interface:
5-21
User Manual
a)If the number of the QC count is less than 3, QC result of each parameter will not be displayed on the right of
QC graph.
b)No target value or deviation limit is set. The QC diagram does not show the QC results.
c)The X-axis of QC diagram shows the number of QC count results in QC file, while the Y-axis shows QC count
result of each parameter.
d)Vertical bar is used to mark the count data of the same group, and at most 31 points can be displayed in the QC
diagram for each parameter.
e)As for parameter CRP, these three data on the left of the QC diagram correspond to three boundary values,
representing respectively, from top to bottom, the upper limit, target value and lower limit of the parameter.
Upper limit: QC material reference value + deviation limit
Target: reference value of QC material
Lower limit: QC material reference value - deviation limit
f)As parameter CRP, the three data on the left of QC diagram respectively represent Mean (mean), SD (standard
deviation) and CV% (coefficient of variation).
(2)Descriptions of points in QC diagram:
Each point in QC diagram corresponds to a QC result and the points are connected by line sections.
D
"Green solid point" means that the QC result is under control, while "Red hollow point" means that the QC result is
LT
out of control.
.,
(3)If points in the QC diagram are out of the QC range, dispose following the procedures below:
a)Check whether the target value and deviation limit of QC material are correct.
b)Whether the background test of the instrument is normal.
CO
c)If these two points above are normal, re-execute QC counting, and if abnormal QC point remains in QC
AL
manufacturer or agent.
TR
(5)Print QC diagram:
S
Correctly connect the printer, and click [Print] in the figure above to print the report.
IN
Operators can calculate Mean (mean), SD (standard deviation) and CV% (coefficient of variation) in QC diagram
RU
5-22
User Manual
D
LT
.,
Fig. 5-6-7
CO
Click the selected abnormal QC point and the red cursor line will stop on the abnormal QC point. Click [Delete] and
AL
the selected abnormal QC point will change to "black point". When the preset value is taken for calculation, the
removed QC point (black QC point) is not included.
I
TR
After removing the invalid point, click [Save DEF] and the instrument will save the preset value which can be used
to calculate the current level into the system.
S
If there are less than 3 valid QC points, click [Save DEF], and a prompt box will pop up on the interface as shown in
DU
Fig. 5-6-8
If the operation above has error, click the screen to move the red cursor line to the removed point, and black QC
point will turn to be red or green through clicking [Add], then the removed point will recover to normal.
When calculating the preset value, adopt data of valid QC point as reference value and deviation limit.
5.6.4 QC list
The user can select different QC files to browse QC data.
Click [QC list] and select the QC file number to be queried from the pull-down menu of [File No.], as shown in the
figure below:
5-23
User Manual
D
LT
.,
Fig. 5-6-9
CO
(1)Delete all: operators can delete all QC counting results from the current QC file.
AL
Fig. 5-6-10
RU
(2)Delete: the operator can delete a QC counting result from the current QC file.
Select the QC data to be deleted and click [Delete] to pop up the prompt box as shown in the figure below:
Fig. 5-6-11
Click [OK] to delete the selected data.
(3)LIS transmission: when connecting LIS, operators can transmit the QC results in the current QC file to the LIS
software.
(4)Check the data: select , check the item data through page turning.
5-24
User Manual
6.1 Overview
The operator can input the information of the sample to be analyzed before the sample analysis and then click Use
worklist in mode setting and test the sample to be analyzed in sequence.
Click [Menu] - [Sample registration], as shown in the figure below:
D
LT
.,
CO
I AL
S TR
DU
Fig. 6-1-1
IN
6-1
User Manual
D
LT
.,
Fig. 6-2-1
CO
6.2.1 Sample information
AL
(2)Bar code: The default is null and it can be input manually or by scanning.
TR
(3)Analysis mode: The newly added record defaults to the mode currently used by the instrument and cannot be
modified.
S
DU
(4)Test mode: select an appropriate test mode from the pull-down menu.
(5)Sending doc.: Select the doctor from the pull-down menu.
IN
changed. When changing it, note that the sampling time cannot be greater than the submission time, and neither can
RU
Fig. 6-2-2
6-2
User Manual
The reference group is linked with age and sex. After the sex is selected, age is input and age unit is selected, the
reference group will be selected automatically. For the newborn, it is smaller than or equal to 28 days. For children,
it is greater than 28 days but no greater than 13 years. For the ordinary, it is greater than 13 years. If it is greater than
13 years and the sex is male, it will be deemed adult male by default. If it is greater than 13 years and the sex is
female, it will be deemed adult female by default.
(5)Department: Select it from the pull-down menu directly.
(6)Cost type: Select the charge in the pull-down menu.
(7)Remark: user can input detected clinical diagnosis information (can be input directly).
After all the above said information is input, click [Save] to save them.
D
LT
.,
CO
Fig. 6-3-1
AL
Input the sample bar code manually or scan it, then Click [OK]. After it is downloaded successfully, the patient
I
information will be recorded in work list and "Data back succeeded" will display below the bar code and the work
TR
list can be downloaded. The following abnormal situations may occur during download:
(1)When the bar code number is not input, the prompt box as shown in the figure below will pop up:
S
DU
IN
I
RU
DI
Fig. 6-3-2
(2)When the LIS is not connected, the prompt box as shown in the figure below will pop up:
Fig. 6-3-3
(3)When the instrument has 200 work lists, the prompt box as shown in the figure below will pop up:
6-3
User Manual
Fig. 6-3-4
(4)When the bar code information and the LIS terminal do not match, the content shown in the figure below will
display:
D
LT
.,
Fig. 6-3-5
CO
AL
After finishing the download of work lists, click [Cancel] to exit the interface.
After the work list is downloaded, click [Cancel] to exit the interface.
I
TR
Fig. 6-4-1
(2)Delete: the operator can delete a QC counting result in the current QC file.
In Fig. 6-3-1, select the work list to be deleted and then click [Delete], and the interface below will appear:
Fig. 6-4-2
Click [OK] to delete the selected work list.
6-4
User Manual
7.1 Overview
This chapter introduces the whole regular operation process from startup to close of the instrument, which gives
specific instructions on the operation process of the whole-blood sample analysis and pre-dilution (peripheral blood)
sample analysis.
The daily operation process is shown as follows:
D
LT
.,
CO
I AL
TR
Fig. 7-1-1
Before turning on the power switch of the instrument, the operator shall conduct an inspection in accordance with
the following requirements to ensure the system prepared to work.
IN
Check if the pipeline connected with the reagent and waste liquid tank is bent or twisted, and if the connection is
DI
reliable.
Please use matching reagent produced by Dirui Company. The Company will not be responsible for
inaccurate test results caused by failure to use matching reagent.
(3)Inspection of power source
Check if the power cable of the instrument is connected correctly.
(4)Inspection of printer
a)Check if the data cable of the printer is correctly connected to the instrument, and if the power cable is
connected correctly.
b)Check if the printing paper is sufficient, and if the installation is correct.
(5)Inspection of bar code reader
Check if the cable of the external bar code reader is correctly connected to the instrument.
7-1
User Manual
7.3 Turn-on
Turn on the power switch. The screen displays “System Loading...” and the instrument is initialized.
7.4 Daily QC
Before the sample analysis, a QC analysis of the instrument shall be conducted every day to ensure reliable
analysis results of the instrument. Refer to Chapter 6 for specific operation.
● The operator shall use clean EDTA-K2 anticoagulation vacuum blood collecting tubes, silicified
glass/plastic test tubes and 20μL of silicon-boronized glass capillaries.
● Samples used for white blood cell classification or platelet count shall be saved under room temperature,
and shall be analyzed within 8 hours after collection.
● If analysis results of PLT, MCV or five classification of white blood cells are not needed, the samples can be
D
saved in 2℃~8℃ of refrigerator for 24 hours. The refrigerated samples shall be placed under room
LT
temperature for at least 30 minutes before analysis.
● Emulsion reagent used for CRP measurement shall be saved in low-temperature ambient, with the saving
.,
condition of 2℃~8℃. If the room temperature is under 2℃, please take the reagent out of the instrument
and put it into the refrigerator to ensure it is saved under the temperature of 2℃~8℃.
CO
● Samples having been placed for a certain time shall be re-mixed before analyzing.
7.5.1 Test tube types
AL
Vacuum Blood
Emergency Position Both Ok ×
IN
7-2
User Manual
D
LT
.,
CO
I AL
S TR
● Please use the adapter for tests with a micro-blood test tube.
IN
venous-blood sample, and ensure the collected sample volume over 1mL.
RU
(2)Quickly and fully mix the venous blood and anticoagulant in the tube.
DI
Fig. 7-5-1
(2)Click [OK] and it will prompt "Preparing diluent..."
(3)After the diluent is prepared, a diluent dispensing interface will pop up, as shown in figure below:
7-3
User Manual
Fig. 7-5-2
(4)Dispense diluent
a)If the instrument has a sample feeder, take a clean centrifuge tube and put it in the STAT chamber as shown in
the figure, and then press the [Count] key. The instrument will push the STAT chamber to the position below the
sampling probe, the diluent will be dispensed to centrifuge tube automatically. After that, the chamber will get
D
back to its original position and then the centrifuge tube can be removed.
LT
.,
CO
I AL
S TR
DU
Fig. 7-5-3
IN
b)If the instrument does not have a sample feeder, put the centrifuge tube below the sampling probe and click
"Count" to start dispensing.
I
RU
DI
Fig. 7-5-4
(5)After diluent dispensing is finished, collect 20μL of peripheral blood and quickly inject it along the centrifuge
tube wall into the tube filled with BF-5D diluent, and then mix them to finish a preparation process of a pre-dilution
sample.
● Operators can also use a pipette to dispense 180μL of diluent, and inject the collected 20μL of peripheral
blood along the centrifuge tube wall into the tube, and then mix them.
● Avoid dust mixing into the prepared diluent, or it will cause analysis error.
7-4
User Manual
● Fully mixed peripheral blood and diluent shall be placed for 3 minutes, and then be re-mixed for
analyzing.
● Ensure the diluted sample be analyzed within 30 minutes, or it will cause unreliable analysis results.
● Samples having been placed for a certain time shall be re-mixed before analyzing.
● Each laboratory shall conduct a stability assessment of the sample analysis result under pre-dilution mode
in accordance with the specific sample quantity, sampling method and technological level of each lab.
● Set the reference value range of the parameters on [Settings] interface before the test, or there may be an
alarm prompt of incorrect result after the test.
● Conduct the sample analysis directly after selecting work mode, and the defaulted reference range of the
instrument is "Normal". After the analysis, the alarm of the instrument is given in accordance with the
D
reference value range of "Normal" setting.
LT
(1)Click [Mode] on the sample analysis interface, and the interface will pop up as shown below:
.,
CO
I AL
S TR
DU
IN
I
RU
DI
Fig. 7-6-1
If no barcode reader is connected, the “Two-way LIS” option will not be displayed on the interface.
(2)Modification of analysis mode and sample number:
a)Analysis mode:
Under the analysis mode, select [WB] or [Micro-WB] mode. If it is a vacuum blood collecting tube, select [WB].
If it is a micro-blood test tube, select [Micro-WB]. After correct selection, click [Save] in Fig. 7-6-1, and the
operation mode will automatically switch to the [WB] mode.
b)Select mode under "Test mode":
7-5
User Manual
"CBC" mode: count only, with no classification of the white blood cells. The count result comprises histograms
of the RBC and PLT and 15 parameters.
"CBC+DIFF" mode: count, and classify the white blood cells. The count result comprises 25 parameters and
scattergrams and histograms.
“CBC+CRP” mode: count only, and not classify white blood cells. The count result comprises 17 parameters and
scattergrams and histograms.
"CBC+DIFF+CRP" mode: count, and classify the white blood cells. The count result comprises 27 parameters
and scattergrams and histograms.
"CRP" mode: Only count the CRP.
c)Select [Use work list] to test the sample record to be tested in the test work list.
d)Click [Save] to save the setting.
The upper limit of the input sample number is 12 bits, and "-" is valid, but "." is invalid.
(3)Sample test
a)If the instrument has a sample feeder, under the condition prepared for counting, shake the sample tube up and
D
down gently to mix the sample, and then put the sample tube in STAT chamber and press "Count" to start test.
LT
After the test, the software will save the analysis results automatically and display the test results on screen. After
the sample is aspirated, the STAT chamber will get back to initial position to wait for the next test.
.,
CO
I AL
S TR
DU
IN
I
Fig. 7-6-2
RU
b)If the instrument does not have a sample feeder, under the condition prepared for counting, shake the sample
DI
tube up and down gently to mix the sample, and then put the sample tube below the sampling probe and press
"Count" on front panel of the instrument to start test. After the test, the software will save the analysis results
automatically and display the test results on screen. After the sample is aspirated, the probe will return to the
initial position to wait for the next test.
(4)Place the next sample in the STAT chamber position and press "Count" to start the test.
(5)On the sample analysis interface, click "Print" in the shortcut key area to print the report; or select sample record
on query history interface, and click "Print" in the shortcut key area to print the reports of single-sample record or
multi-sample records.
● When the analysis mode is “Micro-WB”, a tube adapter is required for count.
● Disposable vacuum blood collecting tube can be used five times at most, since the fragments produced by
repeated puncturing may lead to inaccurate test results.
● Browsing of the graphs and viewing of the research parameters can be conducted in the analysis process,
but no operation can be performed; the data will be automatically saved in the sample analysis after
analyzing.
7-6
User Manual
● During analysis, if the aperture is plugged or behind the test data there is a"? ", the analysis results are not
reliable. Please conduct "Unblocking" and continue the test.
● If the room temperature exceeds the normal working temperature range of the instrument, analysis results
will not be reliable.
7.6.2 View diagram
The graphs of the test results can be viewed by clicking the DIFF scattergram on the BF-6960CRP sample analysis
or graphic review interface, as shown in the figure below:
D
LT
.,
CO
I AL
S TR
DU
Fig. 7-6-3
IN
The histograms of red blood cell and platelet, and scattergram information of five classifications of WBC.
I
RU
In BF-6900CRP, the DIFF scattergram cannot be clicked and there is no information about the three graphs
at the left side.
DI
7-7
User Manual
D
LT
.,
Fig. 7-6-4
CO
Including BLAST#(blast number), BLAST%(blast percentage), IMM#(immature cells number), IMM%(immature
AL
cells percentage), LEFT#(left shift cells number), LEFT%(left shift cells percentage), ABNLYM#(abnormal
lymphocytes number), ABNLYM%(abnormal lymphocytes percentage), NRBC#(neucleated red blood cells
I
Confirm if the work mode of the sample is "Pre-dilute", if not, switch the current mode to "Pre-dilute" mode.
(1)Click [Mode] on the sample analysis interface, and the interface will pop up as shown below:
IN
I
RU
DI
7-8
User Manual
D
LT
.,
Fig. 7-7-1
CO
AL
If no barcode reader is connected, the “Two-way LIS” option will not be displayed on the interface.
I
TR
7-9
User Manual
D
Fig. 7-8-1(a)
LT
7.8.2 Change of mode
.,
Confirm the analysis mode is automatic whole-blood mode, or set the current mode as “Auto-WB” according to the
following operation. CO
(1)Click [Mode] on the sample analysis interface, and the interface will pop up as shown below:
I AL
S TR
DU
IN
I
RU
DI
Fig. 7-8-2
Select the mode as “Auto-WB” mode.
(2)After the setting, click [Save] to make the operation mode automatically switch to the “Auto-WB” mode.
● If no barcode reader is connected, the “Two-way LIS” option will not be displayed on the interface.
7-10
User Manual
● The upper limit of the input sample number is 12 bits, and "—" is valid, but "." is invalid.
7.8.3 Sample analysis steps
● Set the reference value range of the parameters on [Setting] interface before the test, or there may be an
alarm prompt of incorrect result after the test.
● Conduct the sample analysis directly after selecting work mode, and the defaulted reference range of the
instrument is "Normal". After the analysis, the alarm of the instrument is given in accordance with the
reference value range of "Normal" setting.
● Disposable vacuum blood collecting tube can be used five times at most, since the fragments produced by
repeated puncturing may lead to inaccurate test results.
(1)Click [Start] on the sample analysis interface, and the interface will pop up as shown below:
D
LT
.,
CO
I AL
S TR
DU
IN
I
RU
Fig. 7-8-3
DI
When the instrument is under standby status, put the prepared sample tubes on test tube rack in sequence and put the
test tube rack at the right side of sample feeder (with the side with groove at bottom facing left). Click the [Start] in
Fig. 6-6-1 and the instrument will execute sample analysis in a sequence of the sample position. In the test process,
the operation status at the bottom of the software interface will display as "Under test".
(2)Test result of each sample after the test shall be saved on history query interface. After all the tests are completed,
the test tube rack will automatically move to the left side of the sample feeder, and the operator can remove it.
7-11
User Manual
D
LT
.,
Fig. 7-9-1
CO
a)The defaulted emergency sample ID starts from 0-1.
AL
(2)If it is a vacuum blood collecting tube, select [WB] under analysis mode; if it is a micro-blood test tube, select
[Micro-WB] under analysis mode; if it is pre-dilution sample, select [Pre-dilute] under analysis mode. After a
S
correct selection, click "OK" and the STAT chamber will automatically move to the test position as shown in the
DU
figure below:
IN
I
RU
DI
Fig. 7-9-2
● When the analysis mode is “Micro-WB”, a tube adapter is required for count.
● Emergency samples can only be tested in the analysis mode of "WB" or "Micro-WB" or "Pre-dilute".
Click "STAT" and "Stop STAT" is displayed on the button.
(3)Put the prepared emergency samples at STAT position and press "Count" to start the test. In the test process, the
operation status at the bottom of counting interface displays "In testing" and the sample test results will be saved in
7-12
User Manual
D
LT
.,
CO
I AL
Fig. 7-9-3
S TR
If the room temperature exceeds the normal working temperature range of the instrument, the analysis
DU
(1)When there are prompts of "↑, ↓" or "H", "L" in the counting results, it indicates that the counting results exceed
DI
7-13
User Manual
D
the set reference value
LT
Heterotypic/abnormal There could be heterotypic or atypical Heterotypic or atypical lymphocytes are
lymphocytes lymphocytes bigger than the set reference value
There is red blood There could be incomplete hemolysis of
.,
Dense scatter points in the area between
cells/platelets RBC or immature stage of red blood cells
lymph and shadow cells
agglutination or platelets agglutination
Uneven sizes of red blood cells Uneven sizes of red blood cells RDW-SD>65 or RDW-CV>20
S
7-14
User Manual
7.11 Sleep
When the stop time of the instrument meets the automatic sleep time set by the software, it enters hibernation and
displays the instrument sleep status in the status bar. After the instrument enters hibernation, the screen will sleep at
the same time. Click anywhere on the screen to wake up the screen. If you want to continue the instrument operation,
press the counter key and the system will continue to operate after it wakes up.
D
(1)Internal and external sides of the aspiration probe.
LT
(2)Count cell.
.,
(3)Flow count chamber.
When instrument gives prompt "blocking", execute unblocking operation on [Maint.]→ [Service] → [Maintenance]
CO
interface (equal to executing operation of "Cauterize" and "Back washing" functions).
7.13 Turn-off
I AL
In order to ensure stability of instrument and accuracy of analysis results, after a continual working of 24
TR
hours, operator shall execute shutdown operation according to the following steps.
S
Fig. 7-13-1
(1)If the instrument has a sample feeder, place the probe detergent prepared at the STAT position and click Count to
execute the instrument turn-off process.
(2)If the instrument does not have a sample feeder, place the probe detergent prepared below the sampling probe and
click Count to execute the instrument turn-off process.
Fig. 7-13-2
7-15
User Manual
In the shutdown process, execute the soaking and rinsing of pipeline and count cell. After the two actions are
executed, prompt box shown below will pop up:
Fig. 7-13-3
At this time, the operator can turn off the instrument and set the power switch of the host to “O” to shut down the
instrument.
After shutdown, empty and properly deal with the waste liquid in the waste liquid tank.
D
LT
.,
CO
I AL
S TR
DU
IN
I
RU
DI
7-16
User Manual
8.1 Overview
The instrument will automatically save the analysis results in the database after executing sample analyzing each
time. The instrument can store up to 200000 (BF-6900CRP)/ 500000 (BF-6960CRP) sample analysis results. The
administrator can query all the sample results stored in the database. If sample storage capacity is larger than the
maximum, previous sample results will be overlaid automatically by the software.
Select [History] on the main interface and the interface is shown below:
D
LT
.,
CO
I AL
S TR
DU
IN
Fig. 8-1-1
The results in the sample database are displayed in the form of list by the testing order.
I
RU
If the current screen does not fully display, press at the bottom of the screen to
turn the page.
DI
Fig. 8-2-1
Enter the selection range and click [OK]. The result of entering the sample number is displayed on the current screen
8-1
User Manual
and is selected.
The input sample position shall be within the scope of the sample database, otherwise a warning "the input
beyond the maximal scope" will prompt.
To cancel the selected sample, click [Unselect] in Fig. 8-1-1, then all selected samples will be unselected and the
sample number at the current position will become 0 in the status bar.
8.3 Query
The user can query data in the sample database by selecting different conditions. Click [Query] in the bottom menu
bar in Fig. 8-1-1, and the interface as shown below will pop up:
D
LT
.,
CO
I AL
S TR
DU
IN
I
Fig. 8-3-1
RU
Query conditions: sample ID, analysis mode, test mode, name, case number, department, cost type, test date,
DI
unaudited, unprinted, and not communicated. Query can be made by one of these conditions, or by different
combinations. Moreover, query can be operated by selecting "not checked today" or "not printed today".
8.6 Delete
In Fig. 8-1-1, select the record to be deleted, and click [Delete] in the interface, as shown in the figure below:
8-2
User Manual
Fig. 8-6-1
Click [OK] to delete the selected record.
D
LT
.,
CO
I AL
S TR
DU
IN
I
RU
DI
Fig. 8-7-1
The arrangement of sample database and total number of samples are displayed in the status bar in the form of
"position/total number".
On this interface, users can review and edit information and results and perform histogram adjustment.
8.7.1 Edit information
Click [Edit info] in Fig. 8-7-1, as shown in the figure below:
8-3
User Manual
D
LT
.,
Fig. 8-7-2
CO
Edit the sample information in the corresponding text box and then click [Save] to finish editing. The sample
AL
Fig. 8-7-3
After directly modifying the result value of the corresponding parameter, click [Save] and return to the interface
8-4
User Manual
After modification of the results, the sample records are shown in italics in the graph review and sample
analysis interfaces.
8.7.3 Histogram adjustment
In Fig. 8-7-1, click [Histogram] to enter into the histogram adjustment window, as shown in the figure below:
D
LT
.,
CO
I AL
S TR
DU
Fig. 8-7-4
IN
(1)There are histograms of RBC and PLT on the right side of histogram adjustment window; and every parameter
value is displayed on the left side as well. The RBC and PLT histograms have two vertical lines on both left and
I
right. The vertical line is the classification line, and the two vertical line intervals are the value areas. On the left side
RU
of the histogram it shows the names of all the vertical lines displayed in the radio format. Select the name of the
DI
vertical line to be modified. Click and , and adjust the position of the selected vertical line.
The left parameter value will change with the adjustment line. For example, when adjusting the left classification
line of the RBC histogram, both the RBC parameter value and the RBC related parameter value change.
(2)Click [Save] after adjustment to save the results.
(3)Cancel: Return to the sample analysis interface.
8-5
DI
RU
I
IN
DU
S
8-6
TR
I
User Manual
AL
CO
.,
LT
D
User Manual
9.1 Overview
To guarantee the effective operation and lengthen the service life of the instrument, users shall have routine
maintenance for the instrument according to requirements. Although the instrument will automatically give prompt
to remind users to execute corresponding maintenance after testing a certain amount of samples or continuously
operating for a certain time, and the software system is provided with a special service menu which will show the
measures of routine maintenance or in case of failure, users shall make appropriate maintenance and servicing plan
as per specific laboratory conditions, such as the daily sample volume, operating environment and operating time of
the instrument, to reduce factors that affect analysis accuracy and to guarantee the normal operation of instrument
system.
● Don't spill reagent, sample or waste liquid on the mechanical or electrical parts of the instrument in case of
any damages.
● During operation, operators shall take preventive measures like wearing protective gloves and working
clothes, or they may be infected after touching the polluted area or solution. Please wash with water and take
D
disinfection measures if touching any polluted liquid.
LT
● During rinsing and maintenance, please check whether parts containing liquid lose efficiency as they may
.,
cause hazards.
● The instrument is installed with a laser. Operator shall check the warning labels during maintenance and
CO
servicing process, but not stare forward the light beam directly or through the optical instrument. Please
wear protective glasses when looking straight at laser source.
AL
● Improper maintenance may cause instrument damage. Operators shall maintain the instrument according
to the user manual.
I
● If the instrument is installed with accessories which are not provided or recommended by the
TR
manufacturer or the instrument is used otherwise specified by the manufacturer, the relating protection may
be weakened.
S
DU
● During maintenance process, please check whether there are hazards caused by inefficiency of hoses or
parts filled with solution.
IN
● For the problems without solutions in the User Manual, please contact the after-sales service department
of the manufacturer to obtain suggestions from professional personnel.
I
To maintain the best working state of instruments, scheduled maintenance is necessary, and it is respectively
described as below:
(1)Daily maintenance:
Carry out AC before blood sample analysis every day according to "Chapter 5 Quality control".
If the instrument is not shut down for 24 hours, execute "Rinse tubing" once every day: click
[Menu]→[Maint.]→[Service]→[Rinse] and [Tubing] to execute the operation.
If the instrument is not shut down for 24 hours, execute [Rinse WBC cell], [Rinse RBC cell], [Rinse sheath flow
cell], [Rinse CRP1 cell] and [Rinse CRP2 cell] once a day: click [Menu]→[Maint.]→[Service]→[Rinse] to execute
the operation of each cell.
(2)Weekly maintenance:
If the instrument is turned on or off normally every day, execute "Soak" (WBC cell, RBC cell, sheath flow cell,
CRP1 cell, CRP2 cell) once a week (execute through selecting [Menu]→[Maint.]→[Service]→[Maintenance]).
(3)Monthly maintenance:
If the instrument is shut down or opened normally, execute [Rinse sample-aspirating probe] every month (execute
through selecting [Menu]→[Maint.]→[Service]→[Rinse]).
9-1
User Manual
9.3 Maintenance
D
LT
9.3.1 Basic status
Click [Menu] - [Maint.] - [Basic status], as shown in the figure below:
.,
CO
I AL
S TR
DU
IN
I
RU
DI
Fig. 9-3-1
The followings are the description about the basic state in the figure above:
(1)Temperature: display the ambient temperature, pre-heating cell temperature, CRP refrigeration temperature,
CRP reaction cell temperature, colorimetric ware temperature and the normal range of each temperature.
(2)Pressure: display the positive pressure, negative pressure, the pressure of sheath flow cell and normal pressure
range.
(3)Voltage: display the voltage value in small holes of red blood cell and the normal voltage range.
(4)Current: display the laser current and normal current range.
9-2
User Manual
D
LT
.,
CO
I AL
Fig. 9-3-2
S TR
Version number in the figure is for reference only, the version number of software will change with the
DU
9.3.3 Service
The software provides many maintenance functions to guarantee the accurate and effective operation.
I
Conduct background test after replacing diluent and lyse, and execute sample test when background test result is
within the normal range. Click [Menu] - [Maint.] - [Services] and select the [Replace]/[Fill] tab.
DI
9-3
User Manual
D
LT
.,
Fig. 9-3-3
CO
a)Replacement of BF-5D diluent
AL
If the BF-5D diluent tubing contains bubbles or the BF-5D diluent is polluted or used up and the entire tank of
BF-5D diluent needs to be replaced, click [BF-5D diluent] and it will prompt "Replacing BF-5D diluent...". If the
I
b)If any lyse should be replaced, the steps are the same with the replacement of BF-5D diluent.
S
(2)If to fill the single reagent, follow the steps of Fig. 9-3-4 as detailed below:
DU
IN
I
RU
DI
Fig. 9-3-4
9-4
User Manual
When software gives prompt of no BF-FDO I, the instrument will automatically fill BF-FDO I lyse after
corresponding failure information in "Fault information area" is eliminated.
9.3.3.2 Rinse
Click [Menu] - [Maint.] - [Service] and select the [Rinse] tab, as shown in the figure below:
D
LT
.,
CO
I AL
S TR
DU
IN
I
Fig. 9-3-5
RU
(1)If WBC, HGB or WBC classification related parameters are found abnormal in background test, execute "Rinse
WBC cell" operation. Click [WBC cell] and it will prompt "Rinse WBC cell". The rinsing is finished when the
DI
9-5
User Manual
D
LT
.,
CO
I AL
TR
Fig. 9-3-6
S
(1)Drainage
DU
During the maintenance of pipeline system of the instrument, execute drainage before maintenance to prevent the
spill of liquid, the followings are the specification, taking draining waste liquid disk as example:
IN
a)Draining waste liquid bottle can drain the liquid in waste liquid bottle and tubing. Click [Waste bottle] and it
will prompt "Drain waste liquid bottle". The draining is finished when the progress bar comes to an end.
I
The operation to drain "WBC cell", "RBC cell", "CRP1 cell" and "CRP2 cell" is the same as above.
RU
b)If the instrument will not be used for 3 days- 2 weeks, the tubing should be drained. Click [Tubing] and it will
DI
Fig. 9-3-7
Click [OK] and a prompt box shown below will pop up:
9-6
User Manual
Fig. 9-3-8
After removing the reagent, click [OK] and it will prompt "Drain tubing". The draining is finished when the
progress bar comes to an end. After the drainage of liquid in tubing, place the instrument in a clean and dry place.
c)If the instrument will not be used for over 2 weeks, replace the reagent with distilled water and execute
"Package rinsing" operation. Click [Package rinsing] and it will prompt "Package rinsing". The package rinsing
of tubing is finished when the progress bar comes to an end.
Then, cut off distilled water and execute package draining. The operation steps are the same with steps for
draining tubing. Then, the tubing draining operation is finished. After finishing the two operations above, place
the instrument at a clean position.
D
(2)Aperture
LT
As sundries will be attached in aperture after long time of using, unblock, cauterize and back wash the aperture to
.,
clear the attached sundries.
a)Click [Cauterize] to cauterize the two ends of the aperture with high voltage DC, and meanwhile, pop up the
progress bar "Cauterize". CO
Clear sundries blocked in aperture to prevent and eliminate RBC blocking failure.
AL
b)Click [Back washing] to flush the aperture, and meanwhile, pop up the progress bar "Back washing".
Assistance in cauterization can prevent and eliminate RBC hole blockage.
I
TR
c)Click [Unblocking], a progress bar "Unblocking" pops up, then the instrument will execute cauterization and
back washing. The progress bar ends and the task is complete.
S
(3)Soak
DU
To guarantee the accuracy of the testing result of the instrument, soak WBC cell, RBC cell and sampling probe with
IN
Click [RBC cell] and a dialog box as shown below will pop up:
Fig. 9-3-9
If the instrument has a sample feeder, place the probe detergent at the STAT position and click Count, and the
interface below will appear:
9-7
User Manual
Fig. 9-3-10
After soaking, a dialog as below pops up:
D
Fig. 9-3-11
LT
Click [Next] to start rinsing RBC cell. When the progress bar comes to an end, it shows the end of the process.
.,
CO
I AL
TR
Fig. 9-3-12
S
It the instrument does not have a sample feeder, the prompt box below will pop up.
DU
IN
I
RU
DI
Fig. 9-3-13
Put the probe detergent below the sampling probe according to prompt and the steps are same with that with a
sample feeder.
(2)Soak WBC cell:
Soak WBC cell every two weeks, and the operation steps are same with the soaking of RBC cell.
(3)Soak sample aspirating probe:
Soak rinsing sample-aspirating probe every month, and the operation steps are as below:
Click [Maintenance] - [Sampling probe] to pop up the prompt box as shown below:
9-8
User Manual
Fig. 9-3-14
Place the Probe cleanser at the STAT position and click Count, and the interface below will appear:
D
LT
Fig. 9-3-15
.,
The soaking is finished when the progress bar comes to an end.
9.3.4 HGB verification
CO
Click [Menu] - [Maint.] - [HGB verification], as shown in the figure below:
I AL
S TR
DU
IN
I
RU
DI
Fig. 9-3-16
Click [Verification] in the figure, and HGB verification value will be displayed in the box following "Verification".
9.3.5 Mechanical detection
9.3.5.1 Whole machine
Click [Menu] - [Maint.] - [Mechanical inspection], as shown in the figure below:
9-9
User Manual
D
LT
.,
Fig. 9-3-17
CO
(1)Motor:
AL
When some part of the instrument breaks down, the failure can be judged by motor detection. This program sets the
I
following motors: lifting motor, rotary motor and rinsing motor. Click [Detection] after the corresponding motor
TR
and the test result will be displayed on the screen ("Success" or "Failure").
(2)Valve:
S
If the instrument works abnormally due to valve failure, valve detection can serve as an important method to
DU
Valve status after the test is completed: Yellow indicates successful detection and red indicates failed detection.
The operation process is as follows:
I
Click the corresponding number of the valve (numbers in Fig. 9-3-17 refer to the valves), and the instrument will
RU
automatically detect the valve. Click the valve number and the sound of opening/closing the valve can be heard,
which suggests that the valve is in a normal state.
DI
Click [Detection] after automatic detection, and the instrument will automatically detect the valves in order.
Meanwhile, the valve opening and closing sounds should be able to hear.
(3)Pump:
Click "Positive pressure" or "Negative pressure" to check whether the pump is under normal status.
After detection, click [Sample analysis] at the upper part of the screen to exit mechanical detection interface and
return to main interface.
● Valve detection and positive/negative pressure pump detection cannot be executed at the same time.
● After exiting from the “Mech. test” interface, the Analyzer automatically performs the whole machine
resetting.
9.3.5.2 Sample feeder
Click [Menu] - [Maint.] - [Mech.test] and select the [Sample feeder] tab, as shown in the figure below:
9-10
User Manual
D
LT
.,
Fig. 9-3-18
CO
When some part of moving parts of the sampler breaks down, the failure can be judged by motor detection. The
AL
program of motor detection includes: loading motor, rack-conveying motor, emergency feeder motor, unloading
motor, shaking & lifting motor, pinch motor, bar code scanning motor click [Detection] on the back of
I
corresponding motor, and the test result will be displayed on the screen ("Succeeded" or "Failed").
S TR
DU
The interface will not display the tab if no sample feeder is set.
If no barcode reader is equipped, the interface is as shown below:
IN
I
RU
DI
9-11
User Manual
D
LT
.,
CO
Fig. 9-3-18(a)
AL
9.3.5.3 CRP
Click [Menu] - [Maint.] - [Mech. test] and select the [CRP] tab, as shown in the figure below:
I
S TR
DU
IN
I
RU
DI
Fig. 9-3-19
The detection method is same with that of whole machine.
9-12
User Manual
The bar code reader will emit light harmful to people's eyes, please do not look straight at the light when the
Analyzer is working.
9.3.6 Instrument registration
Click [Menu]-[Maint.]-[Instrument reg.], and the interface is as shown below:
D
LT
.,
CO
I AL
S TR
Fig. 9-3-20
DU
Enter the instrument registration code or click Import (import the registration code in the USB flash drive), enter the
IN
registration code and click Register. Registration success is prompted and the registration is completed.
9.3.7 Data backup
I
9-13
User Manual
D
LT
.,
Fig. 9-3-21
CO
After inserting the U disk in the USB interface behind the instrument, click [Backup] on the interface to export the
AL
If the user needs to restore the exported data, click [Recover] on the interface to export the data in the U disk into the
TR
instrument.
S
DU
Data should be backed up regularly to prevent data loss when hardware or software fails.
IN
9.3.8 Counter
Click [Menu] - [Maint.] - [Counter], as shown in the figure below:
I
RU
DI
9-14
User Manual
D
LT
.,
Fig. 9-3-22
CO
(1)The sampling syringe, rinsing syringe and sheath syringe operation times, as well as the operating time of the
AL
positive pressure pump and negative pressure pump (in unit of s) and probe puncturing times can be viewed.
I
(2)Users can view the operating times of the CRP syringe and the operating time of CRP negative pressure pump (in
TR
unit of s).
During the instrument testing, the probe tip will get worn by the puncturing movement to some degree, so that it is
recommended to replace the sampling probe when the sampling tests are up to 30,000 times as per the methods
below:
I
(1)First unscrew the two screws 1 & 2 of the sampling probe and draw out the sampling probe upward vertically as
RU
9-15
User Manual
Swab
D
Fig. 9-4-1
LT
Screw off the connectors on the upper part of the sampling probe to replace with a new sampling probe as shown in
.,
the following figure:
CO
I AL
S TR
DU
Fig. 9-4-2
Insert the new sampling probe vertically into the swab (as shown in Fig. 9-4-2) and fix the screws 1 & 2 (as shown
IN
in Fig. 9-4-1).
Take the following measures to maintain the Analyzer before maintenance or treatment:
(1)Remove reagent bottles and waste liquid tanks at the back of the Analyzer before handling.
DI
9-16
User Manual
(3)In case of doubts about the compatibility between sanitizer or detergent and parts or materials in equipment,
please consult the manufacturer or its agencies.
Please handle the waste fluid of the Analyzer to prevent potential biological and chemical pollution.
D
LT
.,
CO
I AL
S TR
DU
IN
I
RU
DI
9-17
DI
RU
I
IN
DU
S
9-18
TR
I
User Manual
AL
CO
.,
LT
D
User Manual
D
LT
.,
CO
I AL
TR
Fig. 10-1-1
S
DU
The operator can also select [Operation logs], [Fault logs], and [Other logs] in the interface tab to view.
IN
I
RU
DI
10-1
DI
RU
I
IN
DU
S
10-2
TR
I
User Manual
AL
CO
.,
LT
D
User Manual
D
LT
.,
CO
I AL
TR
Fig. 11-1-1
S
DU
The interface shows the reagent type, reagent volume, and expiration date. If the current reagent becomes invalid, it
will be marked with red in the validity information column.
IN
11-1
User Manual
D
LT
.,
Fig. 11-1-2
CO
(1)Input bar code information:
AL
If external barcode reader is not connected, input barcode information manually, and the operation is shown as
I
below:
TR
Click the input box to correctly input the bar code information on the outer case of the reagent packing box (click
[Clear] to empty the error message and enter new message), and after input, the registration will be done
S
Connect the bar code reader correctly (the other end of the bar code reader data cable is connected to the instrument
"USB" port), make the "bar code number" input box in the input state, scan the bar code on the reagent bottle with
the bar code reader to make the bar code is automatically generated in the input box after the "bar code number".
I
When "Registration fails and the format of the bar code number is wrong" is prompted, check the input bar code to
see if it contains special characters.
When "Registration fails and the parsing of the bar code number is wrong" is prompted, check whether the input bar
code is correct.
When "Registration fails and the bar code has been registered" is prompted, repeat registration after replacing the
reagent.
When "Registration fails and the reagent expires", replace the reagent and register it again.
If there are other prompts, contact the manufacturer or agent.
(4)After the reagent registration is successful, replace the corresponding reagent on the interface of [Menu] - [Maint.]
- [Service] - [Replace].
The instrument reagent bar code does not allow cumulative registration. Registering a new bar code will
remove the original reagent margin.
11-2
User Manual
D
LT
.,
CO
AL
Fig. 11-2-1
I
TR
The life of reagent can be modified. After the modification, click [Save] to exit the setting interface and enter the
interface shown in Fig. 11-1-1.
S
If the selected reagent residual is 0, the validity cannot be set, as shown in the figure below:
DU
IN
I
RU
DI
Fig. 11-2-2
11-3
DI
RU
I
IN
DU
S
11-4
TR
I
User Manual
AL
CO
.,
LT
D
User Manual
Analyzing the sample in the event of a failure may result in incorrect analysis. If a fault alarm occurs during
sample analysis, be sure to perform sample analysis after troubleshooting.
12.1 Overview
When the instrument fails, an alarm message will be displayed in the fault information area. In the "Shortcut key
area", click [Fault information area], as shown in the figure below:
D
LT
.,
CO
I AL
S TR
DU
IN
I
RU
Fig. 12-1-1
DI
The fault information is displayed in the order of occurrence of faults: the last fault message is displayed at the top.
After selecting the corresponding fault information, view the solution to the selected fault in the "Alarm help"
column.
Then, press [Clear alarm] to clear the fault information, as shown in the figure below: 0………………………….
12-1
User Manual
D
LT
.,
Fig. 12-1-2
CO
After clearing, click [Close] to close the dialog box. If the alarm is not cleared or a new alarm occurs during the
AL
12-2
User Manual
D
CRP reagent chamber door is not 2. Please execute the clear alarm operation.
02-111-31 Stop
closed 3. Please contact after-sales personnel if it
LT
cannot return to normal or other failures
occur.
.,
02-131-06 Stop SV21 fault
12-3
User Manual
03-183-02 Stop FPGA program loading failed 1. Please execute the clear alarm operation.
D
2. Please contact after-sales personnel if it
LT
cannot return to normal or other failures
03-183-03 Stop FPGA program loading failed occur.
.,
03-201-22 Stop Failed to discharge WBC cell fluid
occur.
TR
occur.
1. Please check whether there is foreign
IN
occur.
1. Please check whether there is foreign
DI
12-4
User Manual
D
2. Please contact after-sales personnel if it
05-101-10 Stop Clamping motor operation failure
LT
cannot return to normal or other failures
occur.
.,
1. Please check whether the test tube rack
was moved.
05-101-11 Stop
Test tube rack position error before
mixing
CO 2. Please execute the clear alarm operation.
3. Please contact after-sales personnel if it
cannot return to normal or other failures
occur.
AL
05-101-12 Stop Clamping motor abnormal closing 1. Please execute the clear alarm operation.
I
05-101-18 Stop Feeding motor operation timeout 2. Please execute the clear alarm operation.
3. Please contact after-sales personnel if it
cannot return to normal or other failures
occur.
I
05-101-19 Stop
timeout
DI
12-5
User Manual
D
05-151-03 Stop Emergency chamber operation failure
LT
05-151-04 Stop Rack conveying failure
.,
1. Please check whether there is foreign
05-151-05 Stop Rack conveying failure matter on the surface of the conveying area.
CO 2. Please execute the clear alarm operation.
3. Please contact after-sales personnel if it
05-151-06 Stop Rack conveying failure cannot return to normal or other failures
AL
occur.
05-151-07 Stop Rack conveying failure
I
TR
occur.
IN
occur.
DI
12-6
User Manual
D
LT
06-101-03 Stop Probe lifting motor failure
.,
06-101-04 Stop Probe rotating motor failure
06-101-08 Stop Sample-aspirating motor sensor fault 1. Please execute the clear alarm operation.
I
12-7
User Manual
D
06-131-13 Stop SV13 valve fault
LT
06-131-14 Stop SV14 valve fault
.,
06-131-15 Stop SV15 valve fault
CO
06-131-16 Stop SV16 valve fault
AL
12-8
User Manual
D
06-92-05 Stop Gas positive pressure is too high
LT
06-92-06 Stop Gas positive pressure is too low
.,
1. Execute the soaking operation of the
occur.
1. Replace with the correct CRP reagent.
IN
occur.
RU
12-9
DI
RU
I
IN
DU
S
12-10
TR
I
User Manual
AL
CO
.,
LT
D
User Manual
13.1 Transportation
The transport of Analyzer is specified in the contract, during which fierce collision and placing together with
corrosive materials shall be prevented.
13.2 Storage
The packed instrument shall be placed in a well-ventilated room with temperature at -40℃~55℃, relative humidity
not more than 93% and where no corrosive gas exists.
D
LT
.,
CO
I AL
S TR
DU
IN
I
RU
DI
13-1
DI
RU
I
IN
DU
S
13-2
TR
I
User Manual
AL
CO
.,
LT
D
User Manual
D
After-sales Service Unit: Dirui Industrial Co., Ltd.
LT
Address: 3333 Yiju Road, New & High Tech. Development Zone, Changchun Post Code: 130103
.,
International Customer Service Hotline: +86 400 808 7597
International Customer Service E-mail: service@dirui.com.cn
CO
Domestic Customer Service Hotline: 400 811 6695 400 811 6605
AL
A-1
DI
RU
I
IN
DU
S
A-2
TR
I
User Manual
AL
CO
.,
LT
D
User Manual
D
<SB>dddd <EB><CR>
LT
<SB>: message start character (1 byte). ASCII character <VT>, i.e. 0x0B.
.,
dddd: data (composed of different length of bytes). This is HL7 data contents of the block. The data can contain any
values of single byte greater than the hexadecimal value 0x1F and carriage return of ASCII code, <CR>.
CO
<EB>: message end character (1 byte). ASCII character <FS>, i.e. `0x1C.
<CR>: carriage return (1 byte). ASCII character <CR>, i.e. 0x0D.
AL
For example:
<SB>MSH|^~\&|BCC3900|X1706900BF0001|||20090419104618||ORU^R01|1|P|2.3.1|||||CHN|UTF-8
I
<CR><EB><CR>
TR
Where:
S
The five characters after MSH are separators among data fields, assemblies and sub-assemblies. Although those
DU
characters are any non-text characters, the characters in the table below are recommended in HL7 standard:
IN
Delimiter value
Data field separator |
I
Component separator ^
RU
Repeating delimiter ~
Quoting character \
B-1
User Manual
Figure B-1 Schematic diagram of communication of test results and quality control data
Note:
The host needs the LIS end to reply with an ACK^R01 message. If it does not receive the ACK^R01
message, it is determined that the result has not been transmitted to the LIS end (refer to IV. Transmission
Example-9, Sample and QC Response).
B. Work order information query
The work order belongs to Order information. You can use related HL7 messages: ORM (General Order Message)
and ORR (General Order Response Message). The communication process is shown in Figure B-2.
D
LT
.,
CO
Figure B-2 Schematic diagram of communication of work order query
AL
Note:
The system that receives the test results and the system that feeds back the work order may be distributed
I
on different servers, so these two functions need to be configured with their service providers’
TR
ORU^R01 message: mainly used for transmission of test results and quality control data.
ORU Observational Results (Unsolicited) Description
IN
MSH message header, required, including communication information such as message number, sending time,
message separator, and encoding method
I
PID patient basic information, including patient name, gender, medical record number, etc.
RU
PV1 patient consultation information, including patient type, department, charge type, etc.
DI
OBR sample information, including sample number, inspector, inspection time, etc.
{OBX} Inspection data items, including inspection parameter results and inspection related data such as work
mode, etc.
Note: The message segment in { } can be repeated one or more times.
ACK^R01 message: acknowledge the received ORU^R01 message.
ACK Acknowledgment description
MSH message header
MSA message acknowledgement, describing whether the communication message was successfully received
ORM^O01 message: General Order message, basically Order-related actions use this message type, such as
creating a new order, canceling an order, etc. Here is the host requesting LIS / HIS to refill the order message.
ORM General Order Message description
MSH message header
General information of {ORC} Order, including the number information of the queried sample
Note: The message segment in { } can be repeated one or more times.
B-2
User Manual
ORR^O02 message: Confirmation of ORM^O01 message. Here, the complete information of order (that is, work
order) is returned.
ORR^O02 General Order Response Message description
MSH message header
MSA message acknowledgement
PID patient information
PV1 patient consultation information
OBR sample information
{OBX} Other sample information data items, including sample work modes, etc.
Note: The message segment in { } can be repeated one or more times.
B.5 HL7 attribute list
The message sections of the agreement are classified as required, optional and repeatable.
MSH – message header: this section is required, including the basic information of HL7 message like the value of
message separator, message type and encoding system, and it is the first section of each HL7 message.
Message example:
D
MSH|^~\&|BF-6900|X1706900BF0001|||20090419104618||ORU^R01|1|P|2.3.1|||||CHN|UNICODE
LT
No. Field name Length Description Examples
.,
Contains the delimiter between the
segment ID and the first real field,
1# Field Separator 1 |
CO
defining the field delimiter (|) for the rest
of the message.
Component separator, repetition
AL
BF-6900
Sender equipment. Instrument code is
4 Sending Facility 180 X1706900BF0001
S
instrument ID.
DU
Receiving
5 180 Application program of receiving end. LIS
Application
IN
Message
YYYY[MM[DD[HH[MM[SS]]]]]), and
RU
B-3
User Manual
D
103 Table value not found Tabular value not found, not used temporarily.
LT
Status code rejected AR
.,
200 Unsupported message type Message type not supported.
201 Unsupported event code COEvent code not supported.
202 Unsupported processing id Handling ID not supported.
AL
Message example:
RU
PID|1||7393670||Liu Jia|||F|||||||||||||||||||||||25^Y
DI
Leng
No. Field name Description Examples
th
Identify different patient message
1 Set ID – PID 10 1
segments.
Patient Identifier
3# 16 Medical record number. 7393670
List
Liu Jia(Chinese name)
5# Patient Name 50 The name of the patient. String
Joan^Jiang(English name)^
F, M and O respectively indicates
Male, send M; female, send F; others, Female;
8 Sex 1
send O. Male;
Other.
Age and age units. Separate the age
from the age unit with ^. Where age is
an integer and length is 3. The age unit
31 Patient Age 5 25^Y
is a string with a length of 1, Y is sent
for age, M is sent for month, D is sent
for day, and H is sent for hour
B-4
User Manual
D
LT
No. Field name Length Description Examples
.,
1 Set ID – OBR 10 Determine different OBR fields. 1
Requester’s medical order number, used as
2
Placer Order
Number
32 CO
the sample barcode number.
Bar code ID number in sample test
1A2165C24B
LJ/X quality control is lot number
Executor order number, used as sample
AL
Filler Order
3 12 number. Sample number in sample test; 20090807011
Number
File number in LJ/X quality control
I
4# 200 1001^CountResults
ID code + name
Requested Request time/date. Sampling time in sample
S
6 26 20090807140600
Date/time test; valid period in LJ/X quality control
DU
Identifier sender
Specimen
14 Received 26 Test sending date/time. 20090807150000
DI
Date/Time
Attending doctor, used as a test doctor.
20 Filler Field 1 50 Dr
Operator in Lj Quality Control
28 Result Copies To 50 Result carbon copy. For reviewers. Dr
B-5
User Manual
D
“N”-Normal
LT
“A”-abnormal
“H”-result is above the upper limit of the
.,
reference range
8 Abnormal Flags 5 “L”-result is below the lower limit of the L
reference range
CO
Note: This field may have abnormal signs or
high and low alarm signs at the same time. At
this time, multiple signs are connected with “~”,
AL
such as “H ~ A”.
This flag is not used during transmission.
I
results.
Custom content, which is used to store reagent
expiry signs and modification signs. The format is
IN
User Defined
13 16 E-active edit results flag E
RU
Access Checks
e-passive editing result flag
T-instrument temperature over-temperature
DI
B-6
User Manual
D
1006 XbarR QC Xbar-R QC count results 1006^XbarRQC
LT
B.7 OBX-3 identifier code definition
.,
Code Name Description Value type OBX-3 FIELD
2001 MODE Analysis mode CO
IS 2001^MODE
2002 MODE_EX Measurement mode IS 2002^MODE_EX
AL
L-J/Xbar/Xbar-R/CRP QC
2005 Level IS 2005^Level
level
S
B-7
User Manual
D
2101 from median computer to ED 2101^RBC Scattergram.PNG
Histogram.PNG
LT
LIS, the picture format is
PNG
PLT histogram PNG data
.,
Remark: in transmission
PLT
2102 from median computer to ED 2102^PLT Scattergram.PNG
Histogram.PNG
LIS, the picture format is
PNG
CO
BASO scattergram PNG
AL
data
BASO Remark: in transmission
2033 ED 2033^BASO Scattergram.PNG
I
DIFF
2034 from median computer to ED 2034^DIFF Scattergram.PNG
Scattergram.PNG
LIS, the picture format is
IN
PNG
Original data
Remark: used in
2035 Dat1 data ED 2035^Dat1
I
Remark: used in
2036 Dat2 data ED 2036^Dat2
transmission from median
computer to host computer
Original data
Remark: used in
2037 Dat3 data ED 2037^Dat3
transmission from median
computer to host computer
Original data
Remark: used in
2038 Dat4 data ED 2038^Dat4
transmission from median
computer to host computer
B-8
User Manual
D
list[12] = refence.plcrU;
LT
list[13] = refence.plcrsU;
.,
list[14] = refence.pctU;
list[15] = refence.neuCU; CO
list[16] = refence.lymCU;
list[17] = refence.monCU;
AL
list[18] = refence.eosCU;
I
list[19] = refence.basCU;
TR
list[20] = refence.neuPU;
S
list[21] = refence.lymPU;
DU
list[22] = refence.monPU;
list[23] = refence.eosPU;
IN
list[24] = refence.basPU;
list[25] = refence.crpU;
I
RU
0-Whole blood/Closed whole blood 1-Trace whole blood 2-Predilution 3-Automatic whole
Analysis mode
blood
Measurement mode 0- CBC 1-CBC+DIFF 2-CBC+DIFF+CRP 3-CRP 4-CBC+CRP
0 - Normal 1 - Male 2 - Female 3 - Child 4 - Newborn 5 -Self-defined 16- Self-defined 27
Reference group -Self-defined 38 -Self-defined 4
9 -Self-defined 5
QC level 0-high 1-middle 2-low
B-9
User Manual
OBX|2|IS|2002^MODE_EX||1||||||F||<cr>
OBX|3|IS|2003^Ref||0||||||F||<cr>
OBX|4|IS|2004^Note||||||||F||<cr>
OBX|5|NM|2006^V_WBC||0|10^9/L|4-10||||F||<cr>
OBX|6|NM|2007^V_NEU_p||0|%|50-70||||F||<cr>
OBX|7|NM|2008^V_LYM_p||0|%|20-40||||F||<cr>
OBX|8|NM|2009^V_MON_p||0|%|3-8||||F||<cr>
OBX|9|NM|2010^V_EOS_p||0|%|0.5-5||||F||<cr>
OBX|10|NM|2011^V_BAS_p||0|%|0-1||||F||<cr>
OBX|11|NM|2012^V_NEU_c||0|10^9/L|2-7||||F||<cr>
OBX|12|NM|2013^V_LYM_c||0|10^9/L|0.8-4||||F||<cr>
OBX|13|NM|2014^V_MON_c||0|10^9/L|0.12-0.8||||F||<cr>
OBX|14|NM|2015^V_EOS_c||0|10^9/L|0.02-0.5||||F||<cr>
OBX|15|NM|2016^V_BAS_c||0|10^9/L|0-0.1||||F||<cr>
OBX|16|NM|2017^V_RBC||0|10^12/L|3.5-5.5||||F||<cr>
D
LT
OBX|17|NM|2018^V_HGB||1|g/L|110-160||||F||<cr>
OBX|18|NM|2019^V_MCV||0|fL|80-100||||F||<cr>
.,
OBX|19|NM|2020^V_HCT||0|L/L|0.35-0.5||||F||<cr>
OBX|20|NM|2021^V_MCH||0|pg|27-34||||F||<cr>
OBX|21|NM|2022^V_MCHC||0|g/L|320-360||||F||<cr>
CO
OBX|22|NM|2023^V_RDW_SD||0|fL|35-56||||F||<cr>
AL
OBX|23|NM|2024^V_RDW_CV||0|%|11-16||||F||<cr>
I
OBX|24|NM|2025^V_PLT||0|10^9/L|100-300||||F||<cr>
TR
OBX|25|NM|2026^V_MPV||0|fL|7-13||||F||<cr>
S
OBX|26|NM|2027^V_PCT||0|%|0.1-0.28||||F||<cr>
DU
OBX|27|NM|2028^V_PDW||0|fL|15-18||||F||<cr>
IN
OBX|28|NM|2029^V_P_LCR||0|%|13-43||||F||<cr>
OBX|29|NM|2030^V_P_LCC||0|10^9/L|13-129||||F||<cr>
OBX|30|NM|2031^V_CRP||0|mg/L|0-6||||F||<cr>
I
RU
OBX|31|ST|2032^V_HS_CRP||0.00|mg/L|0-6||||F||<cr>
OBX|32|ED|2101^V_RBCScattergram.PNG||……PNG binary data converted into BASE64 coding……||||||F||<cr>
DI
B-10
User Manual
D
LT
2018 V_HGB 2018^V_HGB √ √ √ × √
2019 V_MCV 2019^V_MCV √ √ √ × √
.,
2020 V_HCT 2020^V_HCT √ √ √ × √
2021 V_MCH 2021^V_MCH √ √ CO √ × √
2022 V_MCHC 2022^V_MCHC √ √ √ × √
AL
2023^V_RDW_S
2023 V_RDW_SD √ √ √ × √
D
I
2024^V_RDW_C
2024 V_RDW_CV √ √ √ × √
TR
V
2025 V_PLT 2025^V_PLT √ √ √ × √
S
DU
(2)L-J QC
<SB>MSH|^~&|BF-6900|20180613001|LIS||20110613153445||ORU^R01|5|Q|2.3.1||||||UTF-8<cr>
PID|1||||||||||||||||||||||||||||||^<cr>
PV1|1|||||||||||||||||||<cr>
B-11
User Manual
OBR|1|||1||1002^LJQC|20180420|||20180420113307||||||||||||||||||<cr>
OBX|1|IS|2005^Level||1||||||F||<cr>
OBX|2|NM|2006^V_WBC||465.11|10^9/L|490-510||||F||<cr>
OBX|3|NM|2007^V_NEU_p||0|%|0-0||||F||<cr>
OBX|4|NM|2008^V_LYM_p||0|%|0-0||||F||<cr>
OBX|5|NM|2009^V_MON_p||0|%|0-0||||F||<cr>
OBX|6|NM|2010^V_EOS_p||0|%|0-0||||F||<cr>
OBX|7|NM|2011^V_BAS_p||0|%|0-0||||F||<cr>
OBX|8|NM|2012^V_NEU_c||0|10^9/L|0-0||||F||<cr>
OBX|9|NM|2013^V_LYM_c||0|10^9/L|0-0||||F||<cr>
OBX|10|NM|2014^V_MON_c||0|10^9/L|0-0||||F||<cr>
OBX|11|NM|2015^V_EOS_c||0|10^9/L|0-0||||F||<cr>
OBX|12|NM|2016^V_BAS_c||0|10^9/L|0-0||||F||<cr>
OBX|13|NM|2017^V_RBC||3.94|10^12/L|0-0||||F||<cr>
OBX|14|NM|2018^V_HGB||110|g/L|0-0||||F||<cr>
D
LT
OBX|15|NM|2019^V_MCV||87.6|fL|0-0||||F||<cr>
OBX|16|NM|2020^V_HCT||0.345|L/L|0-0||||F||<cr>
.,
OBX|17|NM|2021^V_MCH||27.9|pg|0-0||||F||<cr>
OBX|18|NM|2022^V_MCHC||319|g/L|0-0||||F||<cr>
OBX|19|NM|2023^V_RDW_SD||51.2|fL|0-0||||F||<cr>
CO
OBX|20|NM|2024^V_RDW_CV||15|%|0-0||||F||<cr>
AL
OBX|21|NM|2025^V_PLT||135|10^9/L|0-0||||F||<cr>
I
OBX|22|NM|2026^V_MPV||12.2|fL|0-0||||F||<cr>
TR
OBX|23|NM|2027^V_PCT||0.165|%|0-0||||F||<cr>
S
OBX|24|NM|2028^V_PDW||19.2|fL|0-0||||F||<cr>
DU
OBX|25|NM|2029^V_P_LCR||42.5|%|0-0||||F||<cr>
IN
<EB><CR>
(3)XBar QC data sending
<SB>MSH|^~&|BF-6900|20180613001|LIS||20110613153445||ORU^R01|5|Q|2.3.1||||||UTF-8<cr>
I
RU
PID|1||||||||||||||||||||||||||||||^<cr>
PV1|1|||||||||||||||||||<cr>
DI
OBR|1|||1||1003^XbarQC|20180420|||20180420113307||||||||||||||||||<cr>
OBX|1|IS|2005^Level||1||||||F||<cr>
OBX|2|NM|2006^V_WBC||465.11|10^9/L|490-510||||F||<cr>
OBX|3|NM|2007^V_NEU_p||0|%|0-0||||F||<cr>
OBX|4|NM|2008^V_LYM_p||0|%|0-0||||F||<cr>
OBX|5|NM|2009^V_MON_p||0|%|0-0||||F||<cr>
OBX|6|NM|2010^V_EOS_p||0|%|0-0||||F||<cr>
OBX|7|NM|2011^V_BAS_p||0|%|0-0||||F||<cr>
OBX|8|NM|2012^V_NEU_c||0|10^9/L|0-0||||F||<cr>
OBX|9|NM|2013^V_LYM_c||0|10^9/L|0-0||||F||<cr>
OBX|10|NM|2014^V_MON_c||0|10^9/L|0-0||||F||<cr>
OBX|11|NM|2015^V_EOS_c||0|10^9/L|0-0||||F||<cr>
OBX|12|NM|2016^V_BAS_c||0|10^9/L|0-0||||F||<cr>
B-12
User Manual
OBX|13|NM|2017^V_RBC||3.94|10^12/L|0-0||||F||<cr>
OBX|14|NM|2018^V_HGB||110|g/L|0-0||||F||<cr>
OBX|15|NM|2019^V_MCV||87.6|fL|0-0||||F||<cr>
OBX|16|NM|2020^V_HCT||0.345|L/L|0-0||||F||<cr>
OBX|17|NM|2021^V_MCH||27.9|pg|0-0||||F||<cr>
OBX|18|NM|2022^V_MCHC||319|g/L|0-0||||F||<cr>
OBX|19|NM|2023^V_RDW_SD||51.2|fL|0-0||||F||<cr>
OBX|20|NM|2024^V_RDW_CV||15|%|0-0||||F||<cr>
OBX|21|NM|2025^V_PLT||135|10^9/L|0-0||||F||<cr>
OBX|22|NM|2026^V_MPV||12.2|fL|0-0||||F||<cr>
OBX|23|NM|2027^V_PCT||0.165|%|0-0||||F||<cr>
OBX|24|NM|2028^V_PDW||19.2|fL|0-0||||F||<cr>
OBX|25|NM|2029^V_P_LCR||42.5|%|0-0||||F||<cr>
<EB><CR>
(4)XBar-R QC data sending
D
LT
<SB>MSH|^~&|BF-6900|20180613001|LIS||20110613153516||ORU^R01|6|Q|2.3.1||||||UTF-8<cr>
PID|1||||||||||||||||||||||||||||||^<cr>
.,
PV1|1|||||||||||||||||||<cr>
CO
OBR|1|||1||1006^XbarRQC|20180420|||20180420124717||||||||||||||||||<cr>
OBX|1|IS|2005^Level||1||||||F||<cr>
OBX|2|NM|2006^V_WBC||739.805^512.27|10^9/L|||||F||<cr>
AL
OBX|3|NM|2007^V_NEU_p||0^0|%|||||F||<cr>
I
OBX|4|NM|2008^V_LYM_p||0^0|%|||||F||<cr>
TR
OBX|5|NM|2009^V_MON_p||0^0|%|||||F||<cr>
S
OBX|6|NM|2010^V_EOS_p||0^0|%|||||F||<cr>
DU
OBX|7|NM|2011^V_BAS_p||0^0|%|||||F||<cr>
IN
OBX|8|NM|2012^V_NEU_c||0^0|10^9/L|||||F||<cr>
OBX|9|NM|2013^V_LYM_c||0^0|10^9/L|||||F||<cr>
OBX|10|NM|2014^V_MON_c||0^0|10^9/L|||||F||<cr>
I
RU
OBX|11|NM|2015^V_EOS_c||0^0|10^9/L|||||F||<cr>
OBX|12|NM|2016^V_BAS_c||0^0|10^9/L|||||F||<cr>
DI
OBX|13|NM|2017^V_RBC||5.79^0.0799999|10^12/L|||||F||<cr>
OBX|14|NM|2018^V_HGB||157.5^3|g/L|||||F||<cr>
OBX|15|NM|2019^V_MCV||80.2^0.200005|fL|||||F||<cr>
OBX|16|NM|2020^V_HCT||0.4645^0.005|L/L|||||F||<cr>
OBX|17|NM|2021^V_MCH||27.15^0.700001|pg|||||F||<cr>
OBX|18|NM|2022^V_MCHC||339^8|g/L|||||F||<cr>
OBX|19|NM|2023^V_RDW_SD||38.4^0|fL|||||F||<cr>
OBX|20|NM|2024^V_RDW_CV||13.3^0|%|||||F||<cr>
OBX|21|NM|2025^V_PLT||241.5^25|10^9/L|||||F||<cr>
OBX|22|NM|2026^V_MPV||11.15^1.3|fL|||||F||<cr>
OBX|23|NM|2027^V_PCT||0.2705^0.057|%|||||F||<cr>
OBX|24|NM|2028^V_PDW||15.2^3.6|fL|||||F||<cr>
OBX|25|NM|2029^V_P_LCR||33.55^8.5|%|||||F||<cr>
B-13
User Manual
<EB><CR>
(5)X-B QC data sending
<SB>MSH|^~&|BF-6900|20180613001|LIS||20110613153529||ORU^R01|7|Q|2.3.1||||||UTF-8<cr>
PID|1||||||||||||||||||||||||||||||^<cr>
PV1|1|||||||||||||||||||<cr>
OBR|1|||||1004^XBQC||||20180426110445||||||||||||||||||<cr>
OBX|1|NM|2019^V_MCV||97.245|fL|60-100||||F||<cr>
OBX|2|NM|2021^V_MCH||30.735|pg|0-87||||F||<cr>
OBX|3|NM|2022^V_MCHC||316.4|g/L|280-320||||F||<cr>
<EB><CR>
(6)CRP QC data sending
<SB>MSH|^~&|BF-6900||LIS||20190822095204||ORU^R01|4|Q|2.3.1||||||UTF-8<cr>
PID|1||||||||||||||||||||||||||||||^<cr>
PV1|1|||||||||||||||||||<cr>
OBR|1|122|1|1005^CRPQC||20190802|20190802134916|||||||||||||||||||||<cr>
D
LT
OBX|1|IS|2005^Level||0||||||F||<cr>
OBX|2|NM|2031^V_CRP||0.03|mg/L|0-2||||F||<cr>
.,
<EB><CR>
(7)Work order application CO
<SB>MSH|^~&|BF-6900|20180613001|LIS||20110613153408||ORM^O01|4|P|2.3.1||||||UTF-8<cr>
ORC|RF||218||IP<cr>
AL
<EB><CR>
I
<SB>MSH|^~\&|LIS||||20180613154025||ORR^O02|4|P^S|2.3.1||||||UTF8<cr>
S
MSA|AA|1||||0<cr>
DU
PID|1||5||T5|||M|||||||||||||||||||||||3^Y<cr>
IN
PV1|1||Orthopedics|||||||||||||||||Medical insurance<cr>
ORC|AF|218|||<cr>
OBR|1|218|5|1001^Count ||20180613153909||||Gu Yisheng||||20180613153919||||||||||||||<cr>
I
RU
OBX|1|IS|2001^MODE||0||||||||<cr>
OBX|2|IS|2002^MODE_EX||0||||||||<cr>
DI
OBX|3|IS|2003^Ref||0||||||||<cr>
OBX|4|ST|2004^Note||test||||||||<cr>
<EB><CR>
(9)Sample and quality control response
<SB>MSH|^~\&|LIS||||20190819160739||ACK^R01|2|P|2.3.1||||||UTF8<cr>
MSA|AA|1||||0<cr>
<EB><CR>
B-14
User Manual
D
Sleep (not used for 2 hours) BF-5D diluent 61mL
LT
Wake BF-5D diluent 85mL
.,
BF-5D diluent 107mL
Manual washing capability
CO
Probe cleanser
BF-5D diluent
4mL
107mL
Automatic rinsing of equipment (setting times of 10-200)
AL
C-1
User Manual
D
BF-FDTⅠ lyse 5mL
LT
BF-5D diluent 9mL
Fill BF-FBH lyse (after drainage)
.,
BF-FBH lyse 10mL
Fill BF-5D diluent (after drainage) BF-5D diluent 159mL
BF-5D diluent
CO 9mL
Replace BF-FDOⅠ lyse
BF-FDOⅠ lyse 10mL
AL
C-2
User Manual
D
Platelet distribution width PDW %
LT
(2)Parameter from calculation
.,
Name Abbreviations Default unit
Percentage of neutrophil NEU%
CO %
Number of lymphocyte LYM# 109/L
AL
C-3
DI
RU
I
IN
DU
S
C-4
TR
I
User Manual
AL
CO
.,
LT
D
User Manual
D
9
LT
(100.1~500.0)×10 /L Not exceeding ±10.0%
12
(0~1.00)×10 /L Not exceeding ±0.03×1012/L
.,
RBC ≥0.990
(1.01~8.00)×1012/L Not exceeding ±3.0%
HGB
(0~70) g/L
(71~240) g/L
CO
Not exceeding ±2.0g/L
Not exceeding ±2.0%
≥0.990
9/
(1001~5000)×10 L Not exceeding ±10.0%
TR
FR-CRP ≥0.990
(10.1~320)mg/L Not exceeding ±15.0%
DU
D.3 Precision
IN
WBC 9
(9.6~15.0)×10 /L ≤2.5%
12
(3.50~5.50)×10 /L ≤1.5%
RBC 12
(5.51~6.50)×10 /L ≤1.5%
(110~160) g/L ≤1.5%
HGB
(161~180) g/L ≤1.5%
9
(100~149)×10 /L ≤5.0%
PLT 9
(150~500)×10 /L ≤4.0%
HCT (30~50)% ≤1.5%
(70~100)fL ≤1.0%
MCV
(101~120)fL ≤1.0%
(2.0~10)mg/L ≤10.0%
FR-CRP
(10.1~320)mg/L ≤4.0%
D.4 Accuracy
D-1
User Manual
D
D.6 Carrying pollution rate
LT
The carrying pollution rate of the Analyzer shall meet the following requirements:
.,
(1)RBC≤0.5%.
(2)WBC≤0.5%.
(3)HGB≤0.5%.
CO
(4)PLT≤0.5%.
AL
(5)FR-CRP≤1.0%.
I
S TR
DU
IN
I
RU
DI
D-2
User Manual
D
Please contact This part can be
LT
CRP syringe Test of 30,000
CRP syringe unit customer service checked/replaced by the
pump samples
personnel manufacturer/agency only
.,
Please contact This part can be
Rinse syringe Test of 30,000
CRP syringe unit customer service checked/replaced by the
pump samples CO
personnel manufacturer/agency only
Sampling probe Sampling unit 100,000 samples 9.4.1
AL
E-1
DI
RU
I
IN
DU
S
E-2
TR
I
User Manual
AL
CO
.,
LT
D
User Manual
Standard requirements
Radiated
electromagnetic IEC 61000-4-3 3V/m, 80MHz~2.0GHz, 80%AM Applicable A
Shell
fields
Rated
power-frequency IEC 61000-4-8 3A/m, 50Hz Applicable A
D
magnetic field
LT
0% for 1 cycle; B
IEC
.,
Voltage sag 40% for 5/6 cycles; Applicable B
61000-4-11
70% for 25/30 cycles C
Voltage
interruption
IEC
61000-4-11
CO
5%, duration: 250/300 cycles Applicable C
AL
AC power
Pulse train IEC 61000-4-4 1kV(5/50ns, 5kHz) Applicable B
supply
I
TR
Radio-frequency
DU
Not
Pulse train IEC 61000-4-4 1kV(5/50ns, 5kHz) --
applicable
I
Radio-frequency Not
IEC 61000-4-6 3V, 150kHz~80MHz, 80%AM --
conduction applicable
Not
Pulse train IEC 61000-4-4 0.5kV(5/50ns, 5kHz) --
applicable
Not
I/O signal Surge IEC 61000-4-5 None --
applicable
Radio-frequency
IEC 61000-4-6 3V, 150kHz~80MHz, 80%AM Applicable A
conduction
Not
Pulse train IEC 61000-4-4 1kV(5/50ns, 5kHz) --
applicable
I/O signal
Not
to main Surge IEC 61000-4-5 None --
applicable
power
Radio-frequency Not
IEC 61000-4-6 3V, 150kHz~80MHz, 80%AM --
conduction applicable
F-1
User Manual
Applicable
Test items EMC standard Test value
or not
Pst value is not greater than 1.0; Plt value is not greater than 0.65;
duration ford(t) value exceeding 3.3% in voltage change period is
Voltage fluctuation
IEC 61000-3-3 nolonger than 500ms; change of relative steady-state voltage (dc) No
and flicker
does not exceed 3.3%; and relative maximum voltage change
(dmax) does not exceed 6%.
Harmonic current It is 1.5 times of the limit value specified in Table 1 of GB/T
IEC 61000-3-2 No
emission 17625.1.
D
LT
● The manufacturer is responsible for providing the consumer/user with the info of electromagnetic
compatibility of the equipment.
.,
● The user is responsible for guaranteeing the environment for electromagnetic compatibility of the
equipment so that it can work normally.
CO
● The equipment meets the specified requirements on emission and immunity in IEC 61326.
● The equipment is designed and tested according to Class A equipment in IEC/CISPR 11. In home
AL
environment, the equipment may cause radio interference, thus preventive measures should be taken.
● It is recommended to evaluate electromagnetic environment before using the equipment.
I
TR
● It is forbidden to use the equipment beside any source of intense radiation, or it may interfere with normal
operation of the equipment.
S
DU
IN
I
RU
DI
F-2
DI
RU
I
IN
DU
ST
RIA
L
CO
.,
LT
D