B240P/B240 installation

--ITSD

V201810
CONTENTS

01
PART ONE
Preparation work

02
PART TWO
Unpacking and installation

03
PART THREE
Adjustment
01
PART ONE
Preparation work
01
PART ONE
Preparation work
Space Requirement

Biossays 240 Biossays 240P


Weight:66 Kg Weight:68 Kg

FOCUS MAKES PROFESSION P4


01
PART ONE
Preparation work
Environment Requirement

1. well-ventilated and no dust, mechanical vibration, big noise and power supply interference
2. Atmospheric pressure: 86KPa - 106 KPa
3. No corrosive and inflammable gases.
4. Temperature of working environment : 10℃ - 30℃
5. Relative humidity: ≤70%
6. Keep away from brush-type motor, flickering fluorescent lamp and electrical contact type
equipment that are frequently switched off and on.
7. Avoid direct sunlight or direct influence of heat and wind sources.
8. During normal working, the loudest noise 1m away from the instrument shall be lower than 60dB.

FOCUS MAKES PROFESSION P5


01
PART ONE
Preparation work
Power Requirement
1. Application power supply: AC 100-240V,50/60Hz.
2. Rated power: 600VA.
3. Circuit breaker type: UL certified-250V-6A-slow break type
4. The instrument shall be powered by a well-earthed socket to ensure reliable work.

Water Requirement

1. Electrical conductivity of purified water shall be at least CLSI grade two and ≤1μs/cm.
2. Water supply shall purify water for at least 5L/h.
3. Temperature of supplied water shall be 5 ℃-32 ℃.

FOCUS MAKES PROFESSION P6


01
PART ONE
Preparation work
Computer Requirement
Operation system: Windows 7 /Windows 10
Computer : CPU ≥ 3.2GHz
HardDisk ≥ 500G
Memory ≥ 4G
RS-232 interface, LAN interface, USB interface must be available

FOCUS MAKES PROFESSION P7


02
PART TWO
Unpacking and installation
02
PART TWO
Unpacking and installation
When you receive the machine from Snibe …

With computer (3 boxes)

1: analyzer

2: computer and monitor

3: accessories 请在此添加图片说明

Without computer (2 boxes)


请在此添加图片说明
1: analyzer

2: accessories

In case of any damage to the instrument, it is required to take pictures and


contact SNIBE for further assistance.

FOCUS MAKES PROFESSION P9


02
PART TWO
Unpacking and installation
1. Unscrew all the fasteners fixed the package and move out the analyzer

请在此添加图片说明

请在此添加图片说明

FOCUS MAKES PROFESSION P10


02
PART TWO
Unpacking and installation
2. Remove the plastic membrane and foam in the component.

请在此添加图片说明

请在此添加图片说明

FOCUS MAKES PROFESSION P11


02
PART TWO
Unpacking and installation
3. Check the packing list and accessories

请在此添加图片说明

请在此添加图片说明

Tools and tanks Computer and monitor

FOCUS MAKES PROFESSION P12


02
PART TWO
Unpacking and installation
Packing list –Only analyzer

请在此添加图片说明

请在此添加图片说明

FOCUS MAKES PROFESSION P13


02
PART TWO
Unpacking and installation
Packing list –Analyzer with computer

请在此添加图片说明

FOCUS MAKES PROFESSION P14


02
PART TWO
Unpacking and installation
Packing list –Analyzer with ISE

请在此添加图片说明

请在此添加图片说明

FOCUS MAKES PROFESSION P15


02
PART TWO
Unpacking and installation
Packing list –Analyzer with ISE and computer

请在此添加图片说明

请在此添加图片说明

FOCUS MAKES PROFESSION P16


02
PART TWO
Unpacking and installation
Accessories

DB9F1F cable

RS-232 cable

Reagent bottle Standard cup


cap

请在此添加图片说明

Shielded network
cable

Sample tube
adaptor Reagent bottle(30ml/10ml)

FOCUS MAKES PROFESSION P17


02
PART TWO
Unpacking and installation
Accessories

wrench

Allen key

BC1200 needle
cleaning unit
semi Permanent cuvettes

请在此添加图片说明
Philips screwdriver

cable for download


data

power cable liquid detection units

FOCUS MAKES PROFESSION P18


02
PART TWO
Unpacking and installation
4. Install reaction cuvettes

tighten the
8 fixing
请在此添加图片说明 screws

请在此添加图片说明

FOCUS MAKES PROFESSION P19


02
PART TWO
Unpacking and installation
5. Connect the power, computer and analyzer

LAN cable connection


请在此添加图片说明 between computer and
analyzer

请在此添加图片说明

FOCUS MAKES PROFESSION P20


02
PART TWO
Unpacking and installation
6. Connect the consumables system

请在此添加图片说明

请在此添加图片说明

Connect the liquid detection unit following the order and colors. Don’t mix them .

FOCUS MAKES PROFESSION P21


02
PART TWO
Unpacking and installation
7. Prepare wash liquid and purified water

请在此添加图片说明

请在此添加图片说明

wash liquid tank(3.5L)/alkaline wash (2L) purified water tank (10L)

Note:
1. Dilute the alkaline wash with the purified water in the ratio 1:8.
2. Before use , displace the wash liquid for some time to make sure no bubbles in the wash liquid tank.

FOCUS MAKES PROFESSION P22


03
PART THREE
Adjustment
03
PART THREE
Adjustment
1. Install the software
(If purchase the computer from Snibe, this step is not necessary.)

请在此处添加小标题 Note:

1) It is forbidden to install the software


in C disk. D disk is strongly suggested.

2) After initialization, don’t forget to


copy and paste the backup folder
<Component> received from Snibe
to restore the data under the
Biossay 240/240 Plus installation
path.

FOCUS MAKES PROFESSION P24


03
PART THREE
Adjustment
2. Log in service software and check port setting

Server IP : 192.168.1.80

请在此处添加小标题

Server port : 2000

Choose TCP/IP

FOCUS MAKES PROFESSION P25


03
PART THREE
Adjustment
3. Check the IP address of the computer.
(If purchase the computer from Snibe, this step is not necessary.)

The IP address of computer should


be 192.168.1.xx.
The last two numbers of IP
address in computer can’t be the
same as the service software (80).

FOCUS MAKES PROFESSION P26


03
PART THREE
Adjustment
4. Input the SN number in service software.

FOCUS MAKES PROFESSION P27


03
PART THREE
Adjustment
5. Check the alignment of moving part

请在此处添加小标题 请在此处添加小标题

FOCUS MAKES PROFESSION P28


03
PART THREE
Adjustment
6. Prime the tubing system-washer (Only for B240P)

Note:
1. When priming the washer , it should be taken apart into a container .
2. Prime enough time to make sure the water flows continuously.

FOCUS MAKES PROFESSION P29


03
PART THREE
Adjustment
7. Prime the tubing system-Sample & Reagent Probe

Prime enough time to make sure


the water flows continuously.

FOCUS MAKES PROFESSION P30


03
PART THREE
Adjustment
8. Prime the tubing system- Stirring washing hole

请在此处添加小标题

Initialize enough times to make sure that water flows continuously .

FOCUS MAKES PROFESSION P31


03
PART THREE
Adjustment
9. Run the cell blank test

请在此处添加小标题 请在此处添加小标题
Reference range:
The results of all wavelength of the first cuvette
should be around 50000, and the others should
be around ±3500.

FOCUS MAKES PROFESSION P32


03
PART THREE
Adjustment

ISE
请在此处添加小标题 请在此处添加小标题

FOCUS MAKES PROFESSION P33


03
PART THREE
Adjustment
1. Open the cover of ISE module.

Move down the


pipetting arm and
请在此处添加小标题 remove the请在此处添加小标题
cover
from the left side

Loose these
fixing screws

FOCUS MAKES PROFESSION P34


03
PART THREE
Adjustment
2. Fill the electrode filling in electrodes –Reference electrode

请在此处添加小标题

 Charge the Filling solution from the charging hole on the top of
Reference electrode. the Filling solution should reach 2/3 to 3/4
of Electrode cavity volume(~5mL)
 Make sure no bubbles between Filling solution and Electrode
membrane, if any bubbles inside, swing the electrode or flip it
with finger to remove the bubbles
 Make sure the charging hole is not blocked by crystal.

FOCUS MAKES PROFESSION P35


03
PART THREE
Adjustment
2. Fill the electrode filling in electrodes –Ion-selective electrode

请在此处添加小标题

 The filling solution should reach 2/3 to 3/4 of cavity volume, make sure not too less or too much.
 Make sure no bubbles between filling solution and Electrode membrane, if any bubbles inside,
swing the electrode or flip it with finger to remove the bubbles.

FOCUS MAKES PROFESSION P36


03
PART THREE
Adjustment
3. Install electrodes .
 The order of electrodes from left to right: Dummy electrode ,K, Na, Ca, PH, Cl, Ref (Reference electrode) .
 Install the electrodes in the correct order, rotate the knob on the right side of electrode box to seal the tube of electrodes.
 The standard of good sealing of the tube of electrodes: the waste liquid coming out from sample peristaltic pump is
continuous without bifurcated, and it is very smooth.
请在此处添加小标题 请在此处添加小标题

Dummy electrode

FOCUS MAKES PROFESSION P37


03
PART THREE
Adjustment
4. Priming-Method 1.
1) Log in user software; after initialization , click <Replace A Standard Reagent> button in <ISE>
dialog of <Status> interface; follow the tips to finish installation. Then load the B standard
reagent in the same method.
请在此处添加小标题

RFID tag RFID

Click <Replace A Standard Reagent> Read RFID Load reagent

FOCUS MAKES PROFESSION P38


03
PART THREE
Adjustment
4. Priming-Method 1.
2) Input the low level command and prime ISE tubing system.

请在此处添加小标题 请在此处添加小标题
1
3 “07 01 16 “ A standard reagent

“07 01 17” B standard reagent

2
4
Note: It is suggested to prime for 2-3 times for A/B standard reagent separately.

FOCUS MAKES PROFESSION P39


03
PART THREE
Adjustment
4. Priming-Method 2.
Load A and B standard reagents; log in service software; prime ISE tubing system in <ISE> dialog.

请在此处添加小标题
1

2 3

Note: It is suggested to prime for 2-3 times for A/B standard reagent separately.

FOCUS MAKES PROFESSION P40