Professional Documents
Culture Documents
Service Manual
Shenzhen Landwind Industry Co., Ltd. ©2010.All rights reserved.This manual and the
transmitted the information contained in this manual without the prior wiitten permission of
Shenzhen Landwind Industry Co., Ltd.The manual is intended for users who are authorized
to use such manual as a part of the product purchased from Shenzhen Landwind Industry
merchantability and fitness for for a particular purpose.Although every effort has been made
to ensure the accuracy of the information contained in this manual, Shenzhen Landwind
Industry Co., Ltd.assumes no liabilities for any errors or omissions and reserves the right to
modify the product to improve its reliablity,function or desigh without further notice.
Shenzhen Landwind Industry Co., Ltd.may modify or improve the product or program
Statement
Shenzhen Landwind Industry Co., Ltd. has the right to explain the manual.
Authorized EC Representative
II
Foreword
Safty Symbols
The following table explains the safety symbols used in this manual.
Symbols Description
This symbol means to note. What the symbol indicates may damage
your analyzer or lead to unreliable test results.
This symbol means biohazard. What the symbol indicates may cause
biological contamination.
This symbol means electric shock. What this symbol indicates may get
you electrically shocked.
This symbol means corrosion. What this symbol indicates may cause
corrosive damage.
This symbol means high temperature. What the symbol indicates may
cause personal injury.
This symbol means strong light. What the symbol indicates may hurt
your eyes.
This symbol means personal injury. What this symbol indicates may
cause personal injury.
Labels
The following table explains the safety symbols used in this manual.
Labels Description
III
In Vitro diagnostic equipment
~ Alternating current
Power on
Power off
Graphics
All graphics in this manual are for illustration only and must not be used for any other
purposes.
IV
Contents
Foreword................................................................................................................................III
Contents..................................................................................................................................V
Chapter 1 Precautions on Safety and Use .................................................................................1
1.1 Safety Precautions .................................................................................................1
1.2 Precautions on Use................................................................................................2
Chapter 2 Installation ........................................................................................................................6
2.1 Installer.....................................................................................................................6
2.2 Installation Requirements......................................................................................6
2.3 Connecting Purified Water Bucket .......................................................................7
2.4 Connecting Waste Bucket .....................................................................................7
2.5 Installing or Removing Reagent/Sample Disk....................................................8
2.6 Installing or Removing Sample Tubes.................................................................8
2.7 Installing or Removing Reagent Bottles..............................................................9
2.8 Installing or Removing Cuvettes ..........................................................................9
2.9 Installing Software ................................................................................................10
Chapter 3 System Description.........................................................................................................15
3.1 Analyzing Unit .......................................................................................................15
3.1.1 Overall Structure ...........................................................................................15
3.1.2 Top View.........................................................................................................15
3.1.3 Rear View ......................................................................................................16
3.1.4 Left View ........................................................................................................16
3.1.5 Right View......................................................................................................17
3.2 Functional Modules ..............................................................................................17
3.2.1 Reagent/Sample Dispenser Assembly......................................................17
3.2.2 Mixer Assembly.............................................................................................18
3.2.3 Auto washing System ..................................................................................19
3.2.4 Reaction Disk ................................................................................................19
3.2.5 Reaction Compartment................................................................................20
3.2.6 Reagent/Sample Disk ..................................................................................20
3.2.7 Reagent Refrigerator ...................................................................................21
3.2.8 Acid/Alkaline Wash Position .......................................................................22
3.2.9 Optical System..............................................................................................22
3.3 Analysis Procedure ..............................................................................................23
3.3.1 Actions............................................................................................................23
3.3.2 Action Positions ............................................................................................24
3.3.3 Testing Flow ..................................................................................................24
3.3.4 Measuring Point............................................................................................24
Chapter 4 Board Description ......................................................................................................26
4.1 Main Board ............................................................................................................26
4.1.1 Overview ........................................................................................................26
4.1.2 Main Unit........................................................................................................27
4.1.3 Reaction Disk and Photometric Unit..........................................................28
V
4.1.4 Probe Unit......................................................................................................29
4.1.5 Mixer Unit.......................................................................................................29
4.1.6 Temperature Unit ..........................................................................................30
4.1.7 Power Supply Assembly ..............................................................................30
4.2 Rear Board ............................................................................................................31
4.3 Power Driving Board ............................................................................................31
4.3.1 Overview ........................................................................................................31
4.3.2 Driving Stepper Motor ..................................................................................31
4.3.3 Driving Pumps, Valves, Mixer Rotor, Light Source and Reaction Disk
Heater 32
4.3.4 Powering Drive Boards ................................................................................32
4.4 AD Collection Board.............................................................................................32
4.4.1 Overview ........................................................................................................32
4.4.2 Eight Photometric Conversion Circuits......................................................33
4.4.3 Circuit of AD Collection Board ....................................................................33
4.5 Fluid Level Detection Board................................................................................34
4.5.1 Overview ........................................................................................................34
4.5.2 Structure of the Probe ..................................................................................34
4.6 Power Supply Assembly......................................................................................35
4.6.1 Overview ........................................................................................................35
4.6.2 Description of Power Supply Assembly.....................................................35
Chapter 5 Function Overview of Fluid Path .............................................................................36
5.1 Fluid Path Diagram...............................................................................................36
5.2 Function Overview................................................................................................36
Chapter 6 Component Disassembly and Assembly................................................................37
6.1 Replacing Reagent/Sample Probe ...........................................................................37
6.2 Replacing Mixer......................................................................................................37
6.3 Unclogging Reagent/Sample Probe ........................................................................37
6.4 Replacing Syringe ...................................................................................................38
6.5 Replacing Lamp ......................................................................................................38
Chapter 7 Interface and Operations ..........................................................................................40
7.1 Screen Structure ...................................................................................................40
7.2 System Main Function .........................................................................................42
7.2.1 Sample ...........................................................................................................42
7.2.1.1 Patient’s information registry form ................................................................44
7.2.2 Rgt/Cal ...........................................................................................................45
7.2.2.1 Calibrator setting .............................................................................................46
7.2.2.2 Request calibration .........................................................................................48
7.2.3 QC...................................................................................................................51
7.2.3.1 Requesting QC test.........................................................................................52
7.2.4 Status..............................................................................................................54
7.2.4.1 Sample Disk.....................................................................................................54
7.2.4.2 Reagent Disk ...................................................................................................55
7.2.4.3 Reaction Disk................................................................................................57
VI
7.2.4 Result search ................................................................................................57
7.2.5 Parameters ....................................................................................................61
7.2.6 Setup ..............................................................................................................70
7.2.7 Maint...............................................................................................................79
7.2.8 Starting Analysis ...........................................................................................81
7.2.9 Stopping Analysis .........................................................................................82
7.2.10 Shutdown .......................................................................................................83
7.2.11 Emergence Exit.............................................................................................83
Chapter 8 Troubleshooting..........................................................................................................84
8.1 Overview.................................................................................................................84
8.2 Alarm Messages ......................................................................................................84
8.2.1 List of Alarm Messages...................................................................................84
Chapter 9 Appendices ...............................................................................................................102
9.1 Terminology .........................................................................................................102
9.1.1 AD Value ......................................................................................................102
9.1.2 Dark Current................................................................................................102
9.1.3 Water Blank .................................................................................................102
9.1.4 Measuring Point..........................................................................................102
9.1.5 Absorbance .................................................................................................102
9.1.6 Reaction Curve ...........................................................................................102
9.1.7 Response.....................................................................................................103
9.1.8 Calibration....................................................................................................103
9.1.9 Calibration Curve........................................................................................103
9.1.10 Calibration Parameter ................................................................................103
9.2 System Specification..........................................................................................103
9.2.1 Analyzing System .......................................................................................103
9.2.2 Optical System............................................................................................104
9.2.3 Sampling and Mixing System ...................................................................104
9.2.4 Temperature Control System ....................................................................104
9.2.5 Operating System.......................................................................................104
9.2.6 Whole System .............................................................................................105
9.3 Transport and Storage .......................................................................................105
VII
Chapter 1 Precautions on Safety and Use
The following sections introduce the warning symbols used on the LW C100plus.
Ignorance of these symbols may get you into the risks of death or injury. The symbols are
STRONG LIGHT
1 The light transmitted by the bar code reader may hurt your eyes.
2 The light transmitted by the lamp may hurt your eyes. Do not stare at the lamp
when the system is in operation.
FLAMABILITY
1 Do not use flammable substances around the system, such as alcohol and ether.
EXPLOSION
1 Do not use explosive substances around the system.
HIGH TEMPERATURE
1 Before replacing the lamp, please switch off the Main Power and then wait at
least 30 minutes till the lamp cooling down.
2 Do not touch the print head or the metal part near it; otherwise you may get
burned.
ELECTRIC SHOCK
1 When the Main Power is ON, the panels, side cover or rear cover of the analyzer
are not allowed to be opened unless by qualified personnel.
2 Please make sure that no liquid spillage spatters on the panels. In case liquid
goes inside analyzer, just turn off the power supply and contact Landwind as soon
as possible.
PERSONAL INJURY
1 The sharp components of the analyzer can cause puncture wounds. To prevent
injury, please exercise caution when working around the reagent probe, sample
probe, mixer and steel pipe of the wash unit.
2 When the analyzer is in operation, do not touch such moving parts as sample
probe, reagent probe, mixer, wash unit, fan, and etc.
BIOHAZARD
1 All test samples, such as calibrator and control, are infectious. If necessary,
please wear gloves when working with these samples.
2 All waste liquids should be considered infectious. Please wear gloves when
working with waste liquids.
3 Those components that have been attached with the reagent probe, sample
probe, mixer, wash unit, waste tubing and waste tank are infectious. Please wear
gloves when working with them.
CORROSION
1 The wash solution and certain reagents are strongly acid or alkaline. While using
them, exercise caution and prevent direct contact with your hands or clothes. In
case your hands or clothes contact the wash solution or reagents, wash them off
immediately with soap and clean water.
2 In case they spill into your eyes, rinse them with much water and consult an
oculist.
INTENDED USE
1 The LW C100plus is a fully-automated and computer-controlled chemistry analyzer
designed for in vitro quantitative determination of clinical chemistry parameters in
serum, plasma, urine in medical institutions.
2 To draw a clinical conclusion, please also refer to the patient’s clinical symptoms
and other test results.
OPERATOR
1 The LW C100plus is to be operated only by personnel trained and authorized by
Landwind or Landwind-authorized distributors.
ENVIRONMENT
1 Please install and operate the system in an environment specified by this manual.
Installing and operating the system in other environment may lead to unreliable
results and even equipment damage.
2 To relocate the system, please contact Landwind or your local distributor.
ELECTROMAGNETIC NOISE
1 Electromagnetic noise may interfere with operations of the system, or even lead to
unreliable test results and misoperation. Do not use such devices as electric drill,
mobile phones or radio transmitters near the system.
2 The system may generate electromagnetic wave during operation. Do not install or
use electrosensitive equipment near the system.
GROUNDING
1 Make sure the power socket is grounded correctly. Improper grounding may lead to
electric shock.
2 The grounding resistance must be lower than 10Ω. Improper grounding may lead to
unreliable test results and risk o f enclosure leakage and even electric shock.
LIQUID LEAKAGE
1 Please check the tubing connectors carefully before starting tests. Leakage may
cause inaccurate aspiration or dispensing.
2 Do not place reagents or samples on the panels of the analyzer; otherwise liquid
spray or leakage may occur.
PROBE BLOCKAGE
1 Check the reagents and samples carefully for insoluble matters, such as fibrin and
protein; otherwise the reagent probe or sample probe may be blocked.
WATER
1 The water used should meet the requirements of the CAP Type II water.
SYSTEM USE
1 Operate the system strictly as instructed by this manual. Inappropriate use of the
system may lead to unreliable test results or even equipment damage or personal
injury.
2 Before using the system for the first time, run the calibration program and QC
program to make sure the analyzer is in proper state.
3 Be sure to run the QC program every time you use the system, otherwise the result
may be unreliable.
4 Do not open the cover of the reaction disk when the analyzer is in operation
5 The operating software installed on the operation unit is specified by
Landwind.Please do not install any other software or hardware,otherwise it may
affect the system..Do not run other software when the system is in operation.
6 Do not use this computer for other purposes. Inappropriate use of it will probably
introduce computer virus, which may spread through floppy disks, software or
network, into the system.
SYSTEM MAINTENANCE
1 Maintain the system strictly as instructed by this manual. Inappropriate
maintenance may lead to unreliable results or equipment damage or personal
injury.
2 Replacement of such major parts as photometer, probe, mixing bar and syringe
plunger assembly must be followed by a calibration.
3 In the case that the analyzer discontinue use because of error or other reason and
needs repair or disposal,please contact Landwind,meanwhile
Take out the reagent bottle in the analyzer and store it in accordance with the
reagent kit. for example,put the reagent back in the refrigeratory to keep it
from spoilage.
SAMPLES
1 Use serum samples that are completely separated from blood clots or urine
samples that are free from suspended matter. If fibrin exists in the serum samples
or suspended matter exists in the urine samples, the probe may be blocked.
2 Medicines, anticoagulants or preservative in the samples may lead to unreliable
results.
3 Hemolysis, jaundice or chylomicron in the samples may lead to unreliable test
results.
4 Store the samples properly. Improper storage may change the compositions of the
samples and lead to unreliable results.
5 Sample volatilization may lead to unreliable results. Do not leave the sample open
for a long period.
6 The system has a specific requirement on the sample volume. Refer to this manual
for proper sample volume.
7 Load the sample to proper tube position on the sample disk before the analysis
begins; otherwise you will not obtain correct results.
PARAMETERS
1 Incorrect analysis parameters may cause erroneous test results. Please consult
Landwind or your reagent supplier for correct parameters.
Chapter 2 Installation
2.1 Installer
The LW C100plus should be installed by Landwind or its authorized agent only. Proper
environment and proper room is needed for installation. If the analyzer needs to be
relocated, please contact Landwind or your local agent.
Item Requirements
The bearing ground should be level with gradient less than 1/200. It should
be hard enough to bear weight.
Installation site The installation site should be free of dust, corrosive and flammable gas, no
heat and wind source, neither mechanical vibration.
It shall not be exposed in direct sun light and should be well ventilated.
Wall
Sewer
Within Above
2000mm Within 3000mm 500mm
Space Operation
665mm Analyzing Unit Unit
1009mm
Environment 10℃-30℃
Temperature
Relative ≤70% with no condensation.
humidity
Atmospheric 86.0 kPa-106.0kPa
pressure
Well-ventilated to the outside with the analyzer not exposed directly to the
Ventilation
wind source.
Electromagnetic Away from brush-type motors and electrical-contacts equipments which are
wave powered on and off frequently.
Diameter of the particle in water shall be less than microns.
Resistivity of water shall be greater than 0.5MΩcm.
Water
The colony count shall be less than 10cfu/ml.
The dissolved silicon shall be less than 0.1mg/L.
Waste liquid Waste liquid shall be disposed of compliantly with local regulations.
NOTE:The purified water bucket must be placed around the analyzer to ensure that
the purified water inlet tube absorb the water in the bucket.
NOTE: The waste bucket should be lower than the analyzer waste liquid connector.
2. To remove reagent/sample disk,hold the handle and pull the disk upward to remove it.
Partial
Enlarged
Drawing
WARNING
1 Before installing/removing reagent/sample disk,make sure that all moving parts
have been stoped,such as probe 、mixer、reaction disk and reagent/sample disk.
BIOHAZARD
1 Wear gloves and lab coat and, if necessary, goggles.
CAUTION
1 Make sure that the cover of the reagent/sample disk is closed,otherwise cooling
effect of the refrigerator will be degraded and the sample/reagent probe may be
demaged.
2 The reagent/sample disk and the reagent/sample compartment may be
contaminated when being used.If samples or reagents spill on the
reagent/sampled disk or in the reagent/sample comparement,wipe them with cloth
soaked with water or disinfector after placing the Power to OFF.
2.To remove sample tubes,grab the tube and pull it upward to remove it from the tube
holder.
WARNING
1 Before installing/removing sample tubes,make sure the probe and
reagent/sample disk have been stopped. Do not use sample tubes which is
beyond the specific specification.
BIOHAZARD
1 Wear gloves and lab coat and, if necessary, goggles.
2.To remove the reagent bottles,grab the bottle and pull it upward to remove it from the
bottle holder.
WARNING
1 Before installing/removing reagent bottles,make sure the probe and
reagent/sample disk have been stopped. Do not use reagent bottles which is
beyond the specific specification.
BIOHAZARD
1 Wear gloves and lab coat and, if necessary, goggles.
2.To remove cuvettes, open the reaction disk cover, grab the cuvette segment by the
middle of its buckle,pull it upward to remove it form the cuvette holder.
WARNING
1 Before installing/removing reagent bottles,make sure the probe 、mixer and
reagent/sample disk have been stopped.
BIOHAZARD
1 Wear gloves and lab coat and, if necessary, goggles.
NOTE
1 Do not touch the optical surface at the lower part of the cuvette when installing
the cuvette ,otherwise it may affect the cuvette’s color comparison and the
accuracy of the test result.
2 The optical surface should not contact the cuvette holder when installing the
cuvette,otherwise it may scratch the covette and affect the accuracy of the test
result.
When the below interface appears,click “Exit”,then a dialog box will popup to request
And in the below figures,select “I have read and accept the license terms.”and click
“Install ”,see the figure below:
When the below screen appears,click “I accept the licencing terms and conditions” and
click “Next”;
In the following screen that appears,Click “Install”, “Next” unti the below screen
shows.Input the Name and the Company,then click “Next”;
When the above figure shows,click “Next” and the below figure will be
desplayed.Select ”Windows Authentication Mode” or “Mixed Mode”(FOR Mixed
Mode,you need to enter the password) and click “Next”’
In the following screen that appears,Click “Next”,”Install” “Next” and ”Finished” unti the
below screen shows. Click “Next”.
(4)When the installtion above is completed,the software has already been successfully
installed. Click the icon in the desktop and you can start the software
3 2
4
2
3
4
5
1
6
7
Reagen/sample Probe
b. Specifications
c. Actions
The probe moves up and down at the following positions in the order presented
below.
Mixer
The mixer moves up and down, rotates at the following positions in the order presented
below.
Wash Probe
a. Functions
Wash cuvettes automatically.
b. Actions
The wash probe moves up and down,finishes the actions of draining the reaction liquid
and filling in the water.
a. Functions
The reaction disk holds cuvettes, in which the sample and reagent react at a constant
temperature of 37℃ and then measured directly by colorimetry.
b. Specifications
Cuvettes: 81
c. Actions
a. Functions
b. Specifications
Fluctuation: ±0.1℃
a. Functions
The sample disk holds reagent bottles/sample tubes, and rotates to carry each of
them to the specified position for reagent/sample aspirating.
b. Specifications
Round disk, with three circles, hold 40 sample positions, and 40 reagent positions.
NOTE
(1) The No.39 reagent position is specified for the wash solution only. Please use only
the wash solution supplied by Landwind to wash cuvettes, Otherwise, you may not
obtain expected results.
(2) The wash solution supplied by Landwind is WASH WI.
c. Actions
a. Functions
The reagent refrigerator holds the reagent disk and the sample disk and provides a
refrigerated environment for the reagents.
b. Specifications
a. Functions
The acid/alkaline wash position is for acid/alkaline wash solution which is used to
wash probe and mixer.
b. Specifications
NOTE
Landwind recommends you to use the following acid/alkaline wash solution:
(1)acid wash solution:0.1mol/L HCl;
(2)alkaline wash solution: sodium hypochlorite solution containing 0.5% active
chlorine
NOTE
Landwind recommends you to use the acid/alkaline wash solution alternately,eg:use
the acid wash solution one day ,and use the alkaline wash solution the next day.
NOTE
Please use the acid/alkaline wash solution recommended by Landwind, Otherwise,
you may not obtain expected results.
The photometer measures the absorbance of the reaction liquid in the cuvette while
the reaction disk is rotating.
b. Specifications
This section introduces the analysis procedure of the LW C100plus by actions, action
positions, testing flow and measuring points.
3.3.1 Actions
The LW C100plus finishes all the tests by performing the following steps repeatedly:
The reaction disk rotates to carry the cuvettes to the first set of wash probes and
The reaction disk rotates 87 cuvettes positions (during rotating the absorbance of the
period.
During the 1st and the 81st period, R1, sample and R2 are dispensed and the reaction
In the 81st period, the cuvettes are carried to the first set of wash probes and then
periods, each of which takes 24 seconds in fast mode and 36 seconds in routine mode. See the
figure below.
There are 71 measuring points during the whole reaction process, and the interval
between two is 24s in fast mode and 36s in routine mode. See the figure below.
Communication: The mixer unit receives commands from the main unit via a serial port.
Control function: The mixer unit sends control signals of the vertical stepper motor, the horizontal stepper
motor and the wash motor to the power driving board.
Signal detection: The mixer unit detects the signal of the following sensors, which include: 1) vertical
initial position sensor; 2) horizontal initial position sensor; and 3) coder position sensor.
LW C200/220 背板结构示意图
24V/1.5A, 24V/1A
Clock
Coupler
The stepper motor is driven by two circuits; one is the driving chip, and the other is the logic chip, which
receives a control signal from the main board and then converts it into rotation direction and microstep driving
phasor. Two signals are input from the main board:
4.3.3 Driving Pumps, Valves, Mixer Rotor, Light Source and Reaction
Disk Heater
Figure 5-1 Driving pumps, valves, mixer motor, DC motor and reagent preheating assembly
The reagent preheater, mixer DC motor, pumps and solenoid valves are each controlled by an IRF540
switch. (Note: The reagent preheater, mixer motor and other elements are respectively powered by 24V, 6V
and 12V.)
4.4.1 Overview
(1) Functional block diagram
Photometric data
AD converter
conversion circuit
Main board
8 circuits
Control signal
Photometric Signal processing
conversion circuit
AD collection board
AD collection: The photometric signals output from the eight detection circuits are filtered and amplified,
and then gated through multiple channels.
Photometric conversion: The eight conversion circuits convert the light intensity signal transmitted
through the reaction cuvette into voltage signal, and then send it to the input terminal of the AD converter.
The AD conversion board receives control signal from the main board, collects the processed eight
photometric signals, and then sends the AD values of light intensity to the main board for processing.
Flat cable
Conversion circuit Primary Secondary
Amplifier Multiplex
filter filter
The dotted frame shows the structure of the eight-channel signal processing circuit. The input photometric
signal is filtered through two stages and amplified on the AD collection board. Manual gain adjustment is
applied on the amplifying circuit in order to make the output values of all signals reach the maximum. As the
light source is attenuated to decrease the power and the amplitude of output signals, manual gain adjustment
may help make the output values of all signals reach the maximum.
(2) Multiplexing and AD collection
The eight processed photometric signals are transmitted to the input end of the multiple selector and then
routed to corresponding channel for AD conversion. The AD converter is of 16-bit for resolution.
(3) Power supply of AD collection board
The AD collection board is powered by an analog power supply of ±15V and a digital power supply of 5V,
which are all introduced via the main board.
The 5V of the AD977AR is stabilized linearly from the 15V power supply of the AD collection board,
which is transmitted via a five-wire shield cable.
change of phase. The phase detector then outputs an error voltage, which is then filtered by the low-pass filter.
When obtaining the change, the band-pass filter shapes it and then sends it to the CPU of the main board. The
fluid level detection board then outputs a positive pulse signal.
(4) Power supply of fluid level detection board
The fluid level detection board is powered by 12V, which is provided by the main board and converted to
5V and 9V. 9V is used to power the phase-locked loop and 5V to power other elements on the board.
NOTE: Biohazard
The reagent/sample probe should be considered infectious. If necessary,
please wear gloves when working around it.
NOTE: Biohazard
The mixer should be considered infectious. Please wear gloves when working
around it.
NOTE: Biohazard
The sample probe and reagent probe should be considered infectious. If
necessary, please wear gloves when working around it.
If you need to minimize the window after login into the system,perss the “Win” and “D” key.
The user interface of the LW C100plus auto chemistry analyzer includes toolbar,status bar
and function display area.See the figure below.
Status Bar
Temperature and Lamp
Function Buttons
Status Bar
Shoutcut
Function Display Area
Buttons
Area
Status Bar
Display the information such as system status(eg: standby or testing),testing periods,reaction
disk temperature or system time and so on.
Function Buttons
Display system’s every function button.
Sample To request sample test (including batch request) and QC testing.Support to enter
patient’s information,release sample positon and sample No.,ect.
Rgt/Cal To to set calibrator information and reagent position,request calibration test and
reagent blank test,search test results.
QC: To set controls information,request QC test,search test result and print the results.
Status Display the related information of sample disk,reagent disk and reaction disk.
For sample disk,you can view sample info,change sample position,release sample
position,view the reaction curve,etc.
For reagent disk,you can set reagent position,view reagent info, check the reagent
inventory ,etc.
For reaction disk,you can view the status of reaction disk,the cuvette info and testing
item list,etc.
Result To search sample result,calibration result,QC result, and reagent blank.
Select “sample”,input the search conditions,then you can view the patient’s sample test
result,select “reac curve” and the curve of the sample test result will be displayed.
Select “calibration” ,input the search conditions,then you can view the calibration
result and calibration curve.
Select “QC”, input the search conditions,then you can view the QC reqult and QCplot.
Select “reagent blank”, input the search conditions, then you can view the trend
graphics of eagent blank and reaction curve.
Stat Including workload statistics,fees statistics and so on
Para: Set related parameters of test, calculation test, profile and test order..
Setup Including setup of system,dictionary,hospital,user and print setup.
Maint Including maintanance,log,temperature curve,archive database and re-login.
Shoutcut Buttons Area
Start Start all tests that has been requested.
Stop Stop test.
Shutdown Exit the software by following the shutdown procedure.
Emergent Exit the software directly
V To view software name and version.
A To go to the screen “Log” and view failure message.
0 Clear error message.
Temperature and Lamp Status Bar
(1)Explanation of fields
Requesting a sample
1 Set the sample disk, tube position and sample type.
2 If the sample needs to be analyzed upon emergent needs, select the STAT check box.
3 Enter the sample No. in the Sample Number field.
4 Select desired items in the Item Test Selection Area, and select desired profiles in the
Profile Selection Area. Clicking once means to select, and clicking again means to
deselect it. If you want to set up profiles, click the Profile Setting button and then operate
in the displayed window.
5 Click Request.
Requesting samples in batch
1 Set the sample disk, tube position and sample type.
2 Enter the start sample No., and then enter the number of samples you want to request.
3 Select desired items in the Item Test Selection Area, and select desired profiles in the
Profile Selection Area. Clicking once means to select, and clicking again means to
deselect it.
4 Click Request.
NOTE
1. In case of requesting samples in batch, the sequence No. and tube position on
the sample
disk are determined increasingly from the start sample No. and position.
2. Batch request and single-sample request can be performed simultaneously.
(1)Explanation of fields
7.2.2 Rgt/Cal
(1)Explanation of fields
(1)Explanation of fields
Adding a calibrator
1 Select disk number;
2 Select a empty position in the sample disk
3 Enter the calibrator information in corresponding edit boxes
4 To save your settings, click Save; otherwise, click Close;
Setting calibrator concentration
1 Select a calibrator you want to Edit in Calibrators
2 Input the concentration in “conc” in the concentration area corresponding to the item. .
(1)Explanation of fields
Field Description Operation
Sequence number of the sample
Disk disk. There are 5 virtual sample Select it from the pull-down list box.
disks in total.
The number of times to repeat the
Replicates test by using this calibrator.,at most Select it from the pull-down list box.
5 times.
To view the list of calibration that Click to enter the screen of request
Requests
has been requested. calibration.
2 Select the very item that you want to delete and click “delete”.If you need to delete more
Items at the same time,point to the desired item,click once to select it,and click again to cancel it,and then
click “delete”.
Request reagent blank test
2 Select the very item that you want to delete and click “delete”.If you need to delete more
Items at the same time,point to the desired item,use left mouse button to click once to select it,and click
again to cancel it,and then click “delete”.
7.2.3 QC
(1)Explanation of fields
Adding a control
1 Select an empty box in the control name area and input the control name;
2 Selcet or enter the Exp Date and Lot No;
3 To save the control information,click “save”
Setting the target value and SD of the control
1 Select a control form the list in the left side;
2 Input the QC concentration in the target value column and SD in the SD column in
corresponding items
3 To save the information of the target value and SD,click “save”
Editing control information
1 Select the control form the list in the left side;
2 Enter the control information in corresponding edit boxes
3 To save the information,click “save”
Deleting a control
1 Select a control you want to delete form the list in the left side
2 Click the Delete button to delete it.
Test No.(on
the right No.of the profile item Automatically generate,can not be changed.
side)
Delete the QC test that has
Delete Click to delete
been requested.
Requesting QC test
1 Select “QC ” ——“QC ” to enter the screen of requesting QC test;
2 Select Control , Disk and No.
3 Select the test item or the profile item that needs to be requested in the , click once to
select it,and click again to cancel it;
4 Click “request”
Deleting QC test
2 Select the very item that you want to delete and click “delete”.If you need to delete more
Items at the same time,point to the desired item,use the left mouse button to click once to
select it,and click again to cancel it,and then click “delete”.
7.2.4 Status
Including status of sample disk,reagent disk and reaction disk.
To view the testing status of all requested samples on the sample disk.
1.Select the desired sample disk and you can view the testing status of the requested samples on
2.The sample disk is marked with different colour and different shape to distinguish different
sample type and testing status.
3.Select a sample on the sample disk. The Sample Information display area shows the
information of the sample; the Test display area shows all requested items; the Run Time Left for
Selected Smaple area shows the time needed to finish the sample runs.
4.Select a sample and click “refresh” to refresh this sample’s status and inventory.In the situation
of lacking sample,add enough sample and click “refresh”,only after that,you can restart the test of
this sample.
5. By clicking “release”,the sample position will be released. By clicking “clear”,all sample positons
can be released.
6. Click “Reac. curve”,you can view the reaction curve of the test that has been finished.
1.select reagent disk number in “disk” and you can view the reagent position,reagent info and
current reagent corresponding items.
2.The status of the reagent positon in the reagent disk includes empty,distilled water,R1,R2 and
sharing,those status are marked with different colors.The No .39 positon is for wash solution only.
3.There is a circle near the center of every reagent positon,the colour of the circle may be
different,White means default reagent, gray means non-default reagent and red means
insufficient or depleted reagent.
4.In the Test display area which is on the right side of the screen,the faint-yellow means that the
LW C100plus Service Manual 55
Chapter 7 Interface and Operations
reagent positon of the item is not set. The blue means the reagent position is already set.Click the
blue item and a small orange-red circle will appear on the reagent disk to indicate the item’s
position.
NOTE
The reagent specification selected must be consistent with the actual specification,other
NOTE
1. In case of a double-reagent item, you are required to position both R1 and R2 on
the same reagent disk.
2. You are allowed to check reagent inventory only when the system is in standby
status
(1)General operation
Search result
In the figure above,input the search condition and click “search”,then the search result will be
displayed,click “more” then the following screen will displayed.By inputting more condition,you can
view more detailed result.
2.The search result will appear in the “samples” area,and the corresponding test result will be
displayed on the right side of the screen.For those samples of which the status is “complete”,you
can click “rec.curve” to view the reaction curve.
3.If you need to inpu or modify patient’s info,select one sample and input the patient’s info in the
“ demographics” area and then click “save”
4. If you need to view the result in the mode of “by test”,click “by test” and search the result by
following the step1.
Deleting results
1 Select the samples you want to delete in the “Samples” area,and click “Del Sample”,then
all the tests of this sample will be deleted.
2 If you need to delete all samples, point to the desired item,click once to select it,and click
again to cancel it,and then click “delete”.
3 If you need to delete some item’s test result,select the sample in the “Samples”area,then
select the test result in the “Test” area,,and then click “Del Resule”.
Printing results
1 Select the samples you want to print in the “Samples” area, and click “Print”. If you need to
print all samples, point to the desired item,click once to select it,and click again to cancel
it,and then click “Print”.
2 Click “Preview” to preview the sample report
7.2.5 Parameters
Including test,calculation,profile and test order. The following sections introduce the settings
in detail.
7.2.6.1 Test
The test parameter Including Basic,Limits and QC. The following sections introduce the
settings in detail.When setting the test parameter,follow the order”Baic >Limits>QC.
7.2.6.1.1 Basic
(1)Explanation of fields
Number of decimals in the Select it from the pull-down list box.,there are 0,0.0,
Precision
test result. 0.00,0.000 and 0.0000
Unit Unit of the result. Select it from the pull-down list box.
More Set result unit Click to enter the test unit setting window.
Reac Testing method of the Select it from the pull-down list box. The options include
Type current item. endpoint, two-point(fixed-time) and Kinetic
Changing direction of the
Reac Select it from the pull-down list box. The options include
absorbance during the
Direction ascending and descending.
reaction process.
Primary wavelength used Select it from the pull-down list box. The options include
Pri Wave
for testing. 340, 405, 450, 510, 546, 578, 630 and 670.
Sec Secondary wavelength Select it from the pull-down list box. The options include
Wave used for testing. 340, 405, 450, 510, 546, 578, 630 and 670.
Amount of sample
Sample Enter it in the edit box. It should be within 2 to 45 at the
dispensed for normal
Vol increment of 1.
analysis. The unit is μl.
Amount of sample
dispensed for
More
concentrating or dilution
analysis.
Amount of reagent 2
Enter it in the edit box,It should be within 20 to 300ul at
R2 dispensed for testing. The
the increment of 1.
unit is μl.
Enter them in the edit boxes. Enter N in the left, and
enter M in the right one. The N and P you have entered
should meet the following requirements:
(1)single-reagent endpoint method: 10≤N≤P≤48,
where N+4≥P
(2)double-reagent endpoint method: 19≤N≤P≤48,
The Start/end measuring where N+4≥P
Reaction
points for result
Time (3)single-reagent two-point method :10≤N<P≤48
calculation.
(4)double-reagent two-point method: 19≤N<P≤48
Input the default lower limit Input the value in the edit box
Ref Low
of the reference range
Input the default upper
High Input the value in the edit box
limit of the reference range
Select it from the pull-down list box. The options include
Linear, Logit-4P, Logit-5P, Exponentional-5P,
Cal Calibration method of the
Polynominal-5P, Spline and K Factor. When selecting K
Method current item.
Factor, you need to enter a specific value in the edit box
next to this field.
Click to display a figure displayed as the following,enter
the value in the edit boxes. cilck “save” to save the
information,click “cancel” to cancel the edit.
Slope a The test result can be Enter a number in the edit box. The default is 1.
calibrated with the
equation of y=ax+b. Enter a number in the edit box. The default is 0.
Where, x is the actual
Offset b result; y is the calibrated
result; a is the slope; and b
is the intercept.
Adding an item
1. Click the Add button.
2. Enter the item information in corresponding edit boxes.
3. To save your settings, click Save; otherwise, click Cancel.
Editing item parameters
1. Select an item you want to Edit in the item list.
2. Enter the item information in corresponding edit boxes.
3.To save your changes, click Save; otherwise, click Cancel.
Printing item parameter
1. 1. Selcet the item of which you want to print the parameter;
2. Click “print”
4.2.6.1.2 Limits
(1)Explanation of fields
(2)General operation
4.2.6.1.3 QC parameter
(1)Explanation of fields
Pringting parameter
1. select the item that you want to print;
2. Click “Print”.
7.2.6.2 Calculation
(1)Explanation of fields
Full Name Full name of the item Enter it in the edit box.
Number of decimals in the Select it from the pull-down list box.,theres are
Precision calculation test result. 0,0.0,0.00,0.000 and 0.0000
Unit of the calculation test Select it from the pull-down list box.
Unit
result.
Click to enter a new interface. Enter it in the edit
More Set more unit
box.
Input the lower limit Input the value in the edit box
Ref Low
of the reference range
Input the upper limit Input the value in the edit box
High
of the reference range
(2)General operation
2. Enter the calculation test information in corresponding edit boxes or select an option from the
pull-down list boxes.
3. Enter the calculation formula by selecting items in the Routine Tests area and selecting values
and operators in the Boolean Calculation Button display area.
1. Select a calculation test you want to Edit in the Calculation Test area.
2. Enter the calculation test information in corresponding edit boxes or select an option from the
pull-down list boxes.
3. To Edit the calculation formula, click the Clear button in the Calculation Button display area and
set the formula again.
1. Select a calculation test you want to delete in the Calculation Test area.
7.2.6.3 Profile
(1)Explanation of fields
Full Name Full name of the profile Enter it in the edit box.
(2)General operation
Adding a profile
1. Click the Add button.
2. Enter the profile information in corresponding edit boxes and select items in the item display
area. Clicking once means to select the item, and clicking again means to deselect it.
3. To save your settings, click Save; otherwise, click Cancel.
(1)General operation
1. Select the item that you want to change the test order form the list.
2. Click “Home” “Up” “Down” or “End” to set the test order. Or input the number in “Move to” to ste
the test order.
3. click “save” to save the test order,otherwise,click “close”.
7.2.6 Setup
Including setup of system,dictionary,hospital,user and print setup. The following sections
introduce the settings in detail.
7.2.7.1 System
(1)Explanation of fields
(2)General operation
2.If you need to modify the information, edit it and then click “save”.
7.2.7.2 Dictionary
(1)General operation
1 . Click the type of the dictionary form the pull-down list,and enter the information in
corresponding edit boxes,then click “save”
2. If you need to add more types,select the type of the dictionary first,and enter the information
in the blank line,after that,click “save” .
3.If you need to delete a type,select the type and click “delete”
4. If you want to modify the type,select the type and then enter the information,after that,click
“save” ,
7.2.7.3 Hospital
Including hospial setting,department setting and doctor setting.The following sections will give
detailed introduction about them.
4.2.7.3.1 Settingup hospital information
(1)Explanation of fields
Contact Person Contact person of the hospital. Enter it in the edit box.
(2)General operation
1. Enter the hospital information in corresponding edit boxes and then click “save” to save the
hospital
Information.
2. If you need to modify the hospital information,edit it in the edit boxes and then click “save”
(1)Explanation of fields
(2)General operation
1. Enter the department information in corresponding edit boxes and then click “save” to save the
department Information.
2.If you need to add a department,input it in the blank line,click “save” after you input the
department information.
3. If you need to delete a department information,select it form the list and then click “delete”.
(1)Explanation of fields
(2)General operation
1. Enter the doctor information in corresponding edit boxes and then click “save” to save the
department information.
2.If you need to add a doctor,input it in the blank line,and then click “save”
3. If you need to delete a doctor information,select it form the list and then click “delete”.
7.2.7.4 User
Including mangement of user ,role and permission.The following sections will give detailed
introduction about them.
4.2.7.4.1 User Management
(1)Explanation of fields
(2)General operation
1. Enter the information in corresponding edit boxes and then click “save” to save the information.
2.If you need to add a user,input it in the blank line,and then click “save”
3. If you need to delete a user,select it form the list and then click “delete”.
4. If you need to modify the information,select the very line that you want to modify and then edit
it,after that,click “save”.
4.2.7.4.1Role Management
By assigning roles to give each role to access to different operations.
(1)General operation
1. Choose the role in the Role area in the left of the screen,and then check the box if you want to
assign the operation authority to the role.Click “+” which is on the left side of the items,you can
have more choice;
2. If you need to modify the authority,click the box once to select it,click again to cancel it,after
that,click “save”.
4.2.7.4.1Permission Management.
By assigning perssion to give each user to access to different operations.
(1)General operation
1. Choose the user in the user area in the left of the screen,and then check the box if you want to
assign the operation authority to the user.
2. If you need to modify the authority,click the box once to select it,click again to cancel it,after
that,click “save”.
(1)General operation
1. There are 7 default report templates,and neither deleting or editing those templates are
allowed.
2. Select one template from the Report Type,then click “add” to add a template,after you finish
editing the template,click “save”.If you need to modify the template,click “edit”,if you need to
delete the template,click “delete”.
(1)General operation
1. All of the items which needs to be print is displayed in the screen,the number in the the Print
Order colume shows the order.Select an item,and click “Home” “Up” “Down” or “End” to set the
print order.
7.2.7 Maint
7.2.8.1 Maintenance
(1)General operation
1. Select “Initialization”and click “Execute” to execute the initialization.You can perform this
function when startingup initialization is failed.
2. Select “Failure Recover” and click “Execute” to execute the failure recover.You can perform this
function when error occurs during the process of testing.
7.2.8.2 Log
(1)General operation
1. In the Log area,choose a fault form the list,and the description of the error,the probable cause
and the corrective action will be displaed on the right of the screen.;
LW C100plus Service Manual 79
Chapter 7 Interface and Operations
3. Click “Filter” and then enter the filter conditions to filter the error information.
4. Select a fault form the Log area and click “delect” to delete this error information.
7.2.8.4 Re-login
Click the “Re-login” button and the following screen will be displayed,enter the User Name and
Password,you can login the system again.
(1)General operation
1. Click “Browse” to select path of automatic backup and then click “save” to save the path.After
the archive path is set,the system will backup the database everyday when you run the software
for the first time.
2. If you need to restore auto archive,click “restore auto archive”,when it is finished, restart the
software.
4.2.8.5.2 Manual
(1)General operation
1. Click “Browse” to select path of manual backup ,then click “save” to save the path, and then
click “Manual” to run manual backup.
2. If you need to restore manual archive,click “restore manu archive”,when it is finished,restart
the software.
the test type and sample range, and then click OK to start analysis.
1 If you want to analyze the samples quickly, mark the checkbox in front of Fast Mode. If
not, proceed to the next step.
2 Select a reagent disk (No.1-No.5) and sample disk (No.1-No.4) to run the tests.
3 Input the start number and the end number of the test in the Sample ID Range,or else the
system will start all test.
4 Click OK to start the analysis.
NOTE
The LW C100plus can run the reagent blanks, QC tests and routine samples
simultaneously, which are given with the following priority.
1. Reagent blank tests
2.Calibration tests
3. QC tests
4. Routine samples
NOTE
The High-speed mode is only for single-reagent items.
Click the button ”stop” and a dialog box will be displayed.Click “OK” to stop the analysis,Click
“cancel” to cancel.
7.2.10 Shutdown
When the system is in standby status, you can exit it by selecting the Shutdown icon in the
shortcut buttons area.
If the analyzer is not to be used in a long period (several days),click the optional box to select
Holiday mode first,and then click OK to exit the application
NOTE
(1)If it is shut down in normal mode,the cuvettes will be filled in with purified water.
(2)If is is shut down in Long-leave mode,the water in the cuvette will be drained
out
Chapter 8 Troubleshooting
8.1 Overview
This chapter lists all system alarm messages and there corresponding solutions,please take Corrective Actions when error occurs.If it can not be solved,please contact
Landwind.
Probe home position sensor error, or step missing. Probe fails to stop at Recover failure. If the message appears for 3 times,
Lwc0002001 Probe Home Position Error
home position before finishing the set steps. contact the manufacturer.
Lwc0002002 Probe collision: Collision occurs when probe moves down vertically.
Probe home position sensor error, or step missing. Probe fails to leave home
lwc0002003 Offset from probe home position error
position before finishing the set steps
Liquid level sensor error, or probe tip drops. Probe detects reagent level
lwc0004004 Reagent level detection error
before reaching reagent bottle.
Not enough reagents in position {0} of the Probe attaches reagent bottle bottom within 5 steps due to insufficient Check if reagent is sufficient, and refresh reagent
lwc0004005
current reagent disk. Please refill reagent. inventory.
Reagent in position {0} of the current reagent Liquid level sensor error, or reagent depleted. Probe fails to detect reagent Check if reagent is sufficient, refresh reagent
lwc0004006
disk is exhausted. Running tests is disabled. level when lowering down to reagent bottle. inventory, and rerun.
Probe fails to lower down to given position due to being out of home
lwc0002007 Probe location error
position, and may be broken if continue
Probe is not in home position when dispensing though nothing has been
lwc0002008 Dispense error
aspirated.
lwc0002010 Probe wash error Probe is not in home position when being washed after aspirating.
Probe is not in home position when being washed though nothing has been
lwc0002011
aspirated.
Probe home position sensor error, or step missing. Probe fails to find home
lwc0002012 Probe rotating to home position error
position before finishing the set steps
Probe home position sensor error, or step missing. Probe fails to leave home
lwc0002013 position after rotating anticlockwise for certain steps and then rotating back
clockwise.
Probe wash position coder sensor error. Probe fails to find wash position
lwc0002014 Probe rotating to wash position error
before finishing the set steps due to step missing.
Horizontal coder sensor error, or step missing. Probe fails to rotate to given
lwc0004015 Reagent position error
reagent bottle.
Horizontal coder sensor error, or step missing. Probe fails to rotate to given
lwc0002016 Sample position error
sample tube.
Probe fails to locate before rotating. Reset the probe horizontally and try
lwc0002017 Probe horizontal position error
again.。
Reaction disk coder sensor error. Probe fails to find reaction disk position
lwc0002018 Probe rotating to reaction disk error
before finishing the set steps due to step missing.
Liquid level sensor error, or probe tip drops. Probe detects sample level
lwc0002019 Sample level detection error
before reaching sample tube.
lwc0011020 Turning on solenoid valve error Solenoid valve cannot be turned on to wash the probe interior.
Liquid pump cannot be turned on when kept always on or washing the probe
lwc0011021 Turning on liquid pump error
exterior.
Solenoid valve fails to be turned off when liquid pump cannot be turned on
lwc0011022 off solenoid valve error 0.8 second after solenoid valve is turned on to wash the probe interior and
exterior
Turning off liquid pump and solenoid valve Liquid pump and solenoid valve cannot be turned off when probe washing is
lwc0011023
error finished.
lwc0011024 Turning off solenoid valve error Solenoid valve cannot be turned off when probe washing is finished.
lwc0011025 Turning off liquid pump error Liquid pump cannot be turned off when probe washing is finished.
lwc0011026 Turning on probe wash valves error Wash valves cannot be turned on to wash the probe interior and exterior.
Probe syringe home position sensor error, or step missing. Syringe fails to
lwc0002027 Syringe home position error
stop at home position before finishing the set steps.
An error occurs when probe moves back to home position after aspirating
lwc0002028 Sample aspirating error
sample.
lwc0002029 Probe unit instruction check sum error Check sum of the instruction frame received by the probe unit is incorrect.
Not enough samples in position {0} of the Check if sample is sufficient, and refresh sample
lwc0002030 Probe attaches sample tube bottom within 5 steps due to insufficient sample.
current sample disk. Please refill. inventory.
Sample in position {0} of the current sample Liquid level sensor error, or sample depleted. Probe fails to detect sample Check if sample is sufficient, refresh sample
lwc0002031
disk is exhausted. Running tests is disabled. level when lowering down to sample tube. inventory, and rerun.
Probe reaches to vertical home position early due to sensor error or external
lwc0002032 Probe vertical home position error light interference. Check for light interference. If error remains, contact the
manufacturer.
Probe syringe reaches to vertical home position early due to sensor error or
lwc0002033 Syringe vertical home position error external light interference. Check for light interference. If error remains,
contact the manufacturer.
lwc0002034 Offset from syringe home position error Probe syringe fails to leave vertical home position
lwc0002036 Probe offset from home position Probe cannot rotate horizontally due to being out of home position.
lwc0002038 Probe rotating to S position error: Probe horizontal coder sensor error, or step missing.
lwc0013041 Turning off cooling fan error: Cooling fan cannot be turned off normally.
lwc0013042 Turning on cooling fan error: Cooling fan cannot be turned on normally.
lwc0013043 Turning off vacuum pump error: Vacuum pump cannot be turned off normally.
lwc0013044 Turning on vacuum pump error: Vacuum pump cannot be turned on normally.
Sample/reagent disk fails to detect coder sensor signal after finishing the set
lwc0006045 Sample/reagent disk coder signal error
steps for a sample/reagent position
Sample/reagent disk fails to detect coder signal after finishing the set steps
lwc0006047 Sample/reagent coder error:
for a sample/reagent position.
Mixer fails to lower down to given position due to being out of home
lwc0008052 Mixer location error:
position, and may be broken if continue.
lwc0008053 Turning on mixer motor error: Mixer motor cannot be turned on normally.
lwc0008054 Turning off mixer motor error: Mixer motor cannot be turned off normally.
Turning off mixer wash valve and mixer motor Mixer exterior wash valve and mixer motor cannot be turned off when
lwc0008056
error: washing and mixing is finished.
Turning off liquid pump(to wash mixer Liquid pump cannot be turned on when kept always on or washing the mixer
lwc0011057
exterior) error: exterior due to circuit error or circuit interference.
Mixer exterior wash valve cannot be turned on for washing and mixing, and
Turning on mixer wash valve and turning off
lwc0008058 an error occurs to turn off the mixer motor.
mixer motor error:
Mixer home position sensor error, or step missing. Mixer fails to find home
Mixer rotating to home position error:
lwc0008059 position before finishing the set steps.
Mixer home position sensor error, or step missing. Mixer fails to leave home
position after rotating anticlockwise for certain steps and then rotating back
lwc0008060
clockwise.
Mixer wash position coder sensor error. Mixer fails to find wash position
Mixer rotating to wash position error:
lwc0008061 before finishing the set steps due to step missing.
Mixer wash position coder sensor error. Mixer fails to find wash position
lwc0008062 before finishing the set steps due to step missing.
Mixer home position sensor error or reaction disk coder sensor error. Mixer
Mixer rotating to reaction disk error: fails to find reaction disk position before finishing the set steps due to step
lwc0008063
missing.
Reaction disk coder sensor error. Mixer fails to find reaction disk position
lwc0008064 before finishing the set steps due to step missing.
Mixer horizontal position error: Mixer fails to locate before rotating. Reset the mixer horizontally and try
lwc0008065
again.
Mixer reaches to vertical home position early due to sensor error or external
Mixer vertical home position error:
lwc0008066 light interference. Check for light interference. If error remains, contact the
manufacturer.
Mixer unit instruction check sum error: Check sum of the instruction frame received by the mixer unit is incorrect.
lwc0008067
Incorrect response frame is returned due to inconsistent check sum and byte
Reaction disk unit instruction check sum error:
lwc0007068 with the calculated check sum.
Turning off vacuum valve error: Vacuum valve cannot be turned off normally.
lwc0011077
Wash probe offset from home position before Wash probe is offset from home position before
lwc0011083 moving: moving to the waste aspirating position.
Wash probe already in home position before Wash probe is already in home position before
lwc0011084 moving: moving up from cuvette bottom for 185 steps.
Peristaltic pump timed out: Waiting and running time of peristaltic pump is out
lwc0011085
of range.
lwc0011088 Turning off liquid pump error: Liquid pump cannot be turned off normally.
Target temperature you entered is greater than 95℃. Contact the manufacturer.
lwc0010094 Target temperature too high:
Turning off cooling fan error: Cooling fan cannot be turned off normally.
lwc0013096
Incorrect channel No.: Reinstall the operating software. If the error remains, contact the
lwc0013103 Invalid channel No. is included in the instruction.
manufacturer.
Main unit instruction check sum error: Check sum of the instruction received by the main
lwc0001104
unit is incorrect.
Light intensity too weak Reinstall the lamp. If the error remains for 3 times, contact the
lwc0013108 Lamp intensity too weak
manufacturer.
lwc0006109 Reagent not enough Check if sufficient reagent is available, and refresh reagent inventory.
Failed to receive correct result frame from the No response frame is returned from the execution
lwc0013110
execution unit. unit 20 seconds after it receives the instruction.
Check if acid and alkaline wash solution is sufficient, and refresh its
lwc0004112 Insufficient acid and alkaline wash solution. inventory.
An error occurs when the sample probe Although no sample is aspirated; however, the
dispenses wash solution to the wash pool and sample probe is away from the initial position and
lwc0002116
is washed. cannot lower down to the wash pool for cleaning.
The specified reagent position is beyond 1 and 40, Restart the operating software. If the error remains, contact the
Incorrect reagent position.
lwc0013117 and therefore cannot be located. manufacturer.
Operating system error: Another operating system is installed but Windows Reinstall with Windows XP and try again.
lwc0101001
XP.
lwc0102001 failed to open serial port{0}. check the serial port connection
No test data received: No test data found or data reading error. Test should be defined first.
lwc0111001
Receiving reagent disk error: An error occurs when receiving the specified reagent Recover failure by checking the reagent disk.
lwc0119001
disk.
Insufficient {1} for test {0}. Disk: {2} Insufficient {1} for test {0}. Disk: {2} Position: {3}
lwc0119002
Position: {3}
Insufficient reagent. Disk: {0} Position: {1}
lwc0119003 Insufficient reagent. Disk: {0} Position: {1}
Auto archive failed. External database selected, or archive path error. Check if a remote database is used or space is included in the archive path.
lwc0107001
lwc0107003 Restoring database error Check if connection string in the database is correct.
Restoring database failed. Shut down Database still connected by other program Disconnect all programs from the database when restoring it.
lwc0107004
programs connected to it and try again.
lwc0109002 Failure
Chapter 9 Appendices
9.1 Terminology
9.1.1 AD Value
The light that reaches the detector generates current, which is amplified and then transferred
to photovoltage (analog signals) after passing through the fixed resistor. The photovoltage is
converted to corresponding numeric value via the AD (analog-digital) converter. The obtained
transmitted. It is equivalent to the circuit background and should be subtracted during absorbance
calculating.
relative concept and based upon the standard absorbance value. In the LW C100plus, the
absorbance of the water blank is 0 and should be subtracted for calculating any other absorbance.
a numeric value. Each measuring point is strictly determined and closely related with one another.
In the LW C100plus, each reaction process is divided into 71 measuring points. The interval
9.1.5 Absorbance
Absorbance is the logarithm of minus quotient between the transmitted light intensity and the
input light intensity. In the LW C100plus, transmitted light intensity is the AD value of a reaction
cuvette filled with distilled water; the final absorbance is the product of the calculated value ×
20000.
9.1.7 Response
Response refers to the absorbance change before and after the reaction, or the absorbance
9.1.8 Calibration
Calibration is a sample measurement, during which one or more samples (or calibrators) with
given concentration (or activity) are measured to get its response. According to the set calibration
method (linear or nonlinear), an optimal curve is fit based on the given concentration and the
calculated response and used to generate an equation. By using the curve, a sample can be
measured to get its response, which is then employed to calculate its concentration or activity.
calibration formula.
Mixer Single mixing bar,with exterior polishing and no adsorption,stiring after sample
or second reagent is dispensed.