Product Training: ADAMS HA-8180V

Mariann Amador
Product Manager
Arkray Global Business, Inc.(Miami)
ADAMS
ADAMS
AA utomatic
utomatic
D
D iabetes
iabetes
AA dvanced
dvanced
MM onitoring
onitoring
SS ystem
ystem
V

Operation

4
Start - Up

Press the green


button on the lower
left hand side of the
instrument.
Make sure to lift the
cover.
Start - Up
Maximum time is 30
minutes for start up

Column Status

Reagent Status
Start Up – Ready for Measurement
Calibration
Calibration materials

For HbA1c
Calibration HbA1c
Calibration HbA1c
Calibration HbA1c

Low level
High level
Control Materials

For HbA1c

Level 1 – 4 vials
Level 2 – 4 vials Assay Values
Control HbA1c

1:91 working dilution =


0.25 mL of reconstituted vial + 22.5 mL of DILUENT 80
Control HbA1c
Take out from
refrigerator
Reconstitute with 250uL
of Distilled water

Wait 10 min Remove cap Replace cap Wait 15 min

NOTE:
After 15 minutes, shake Reconstituted material
And invert vial to mix is good for 8 days
Add 5uL of reconstituted refrigerated
material to sample cup

Add 450 uL of Diluent 80


Sample
Put in control rack
cup
and perform QC
Program the Assay Values in HA-8180V
Program the Assay Values in HA-8180V
Control Measurement
Control Measurement
V

Maintenance

21
User maintenance
1. Piercing nozzle cleaning

Wipe off dirt from the outside of the


piercing nozzle with cotton swabs.
For stubborn dirt, use gauze moistened
with purified water.
2. Automatic tube washing

71237
Washing solution for tubes

If the content is 7mL or less, cleaning cannot be done


thoroughly. Please note that control values can be affected when
the flow path gets dirty.
Wash rack (Blue rack)
3. Sample tube spinning unit cleaning

Wipe off dirt from the parts listed


below using gauze moistened with
purified water.
4. Dilution container and washing
container cleaning

Wipe off dirt from the back of the Blot up any liquid inside the
dilution container cover. dilution container and washing
• Use cotton swabs or gauze moistened container, and wipe off any dirt.
with purified water. • Use cotton swabs or gauze
• Be careful in particular not to leave moistened with purified water.
dirt on the sensors.
5. Washing the Air Filter
6. Nozzle washing block cleaning
Piercing nozzle & O-ring replacement

O-ring

Wipe the bottom of the


nozzle washing block clean.
7. Mesh filter replacement for eluent and
hemolysis washing solution nozzles
8. Drain pinch valve tube replacement
9. Part cleaning for dilution container and
washing container
Replacing Consumables – Eluent Packs
Replacing Consumables – Eluent Packs
Replacing Consumables – Eluent Packs
Replacing Consumables – Eluent Packs
Replacing Consumables – Hemolysis Washing
Solution Bottle
Replacing Consumables – Hemolysis Washing
Solution Bottle
Replacing Consumables – Hemolysis Washing
Solution Bottle
Replacing Consumables – Column
Replacing Consumables – Column
Replacing Consumables – Column
Replacing Consumables – Column
Replacing Consumables – Column
Replacing Consumables – Column
Replacing Consumables – Printer Paper
Service engineer maintenance

Washing the optical unit cell (Can be performed by the user)

Cell washing kit for semiautomatic washing


* Compatible with the HA60

80386 Cell washing kit


Replacement of Plunger seal replacement for Fluid pump

•Increase in plunger seal life due


to shortened measurement
time.
10637 10641
Seal-Replacing jig Plunger seal
Washing the optical unit cell

Inaccurate measurement results may be obtained if the optical


unit cell is contaminated. Wash the optical unit cell if necessary.

-Recommended frequency : Every 12000 measurements


(About 1 year)

Preparation
・ Cell Washing Kit (code :80386)···································1pc
・ Sodium hypochlorite solution (approx.0.75%) ··············4mL
Operation procedure
・ Refer to the package insert that come with the Cell washing
kit.
Replacement of Plunger seal (1/2)
If the plunger seal on the fluid pump becomes worn and fluid leaks
occur, the analyzer can no longer maintain the described flow rate or
produce accurate measurement results. The plunger seal inside the
pump head must be replaced if necessary.

-Recommended frequency : Every 60,000 measurements.


(About 5 years)
Preparation
・ Plunger seal (Product code : 10641)·····································1 pc
・ Seal-replacing jig (Product code : 10637) ·····························1 pc
・ Hexagon wrench·································································1 pc
・ 6-8 Double open end wrench ···············································1 pc
・ Column sealing screw···························································2 pc
Refer to the Service manual
Chapter7.Maintenace
Replacement of Plunger seal (2/2)
① ②

③ ④ ⑤

Refer to the Service manual


Chapter7.Maintenace
HA-8180V Maintenance
Laboratory Name: Month: Year:

Serial number: Reviewed by: Date:

Routine Maintenance
Item Day 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31

Liquid waste disposal


T urn off instrument
Initial

Weekly maintenance
Item m/d, intial m/d, intial m/d, intial m/d, intial m/d, intial
Piercing nozzle cleaning
Automatic tube washing

Monthly maintenance Replacement of supplies


Item m/d, intial m/d, intial Item m/d initial md/initial
Sample tube spinning unit cleaning Eluent A pack replacement
Dilution container and washing container cleaning Eluent B pack replacement
Eluent CV pack replacement
Every 3 months maintenance Hemolysis washing solution
Item m/d, intial m/d, intial Column replacement (every 5000)
Nozzle washing block cleaning Piercing nozzle (every 60,000)
Printer paper
Every 6 months maintenance
Item m/d, intial m/d, intial
Mesh filter replacement for eluent and
hemolysis washing solution nozzles
Drain pinch valve tube replacement

Every 12 months maintenance


Item m/d, intial m/d, intial
Part cleaning for dilution container and
washing container
Speed you can feel,
        quality you can trust

ARKRAY HbA1c devices will


continue to evolve