X-Rays
X-Ray Tube
Filament Filament Voltage
0.7 mA
Tube Voltage
140 kV
4
09/11/2020
Barre du Collimateur
Générateur X-Ray
Faisceau
en Cone Ligne du faisceau
5
09/11/2020
6
09/11/2020
P/N: TA 55130-H
Collimator orientation for Collimator orientation for X-ray machines 620, 622, 626, 627
X-ray machine 628
7
09/11/2020
Specification
Carte contrôleur réf: TA55097A
Voltage en Sortie : 140kV +/-2kV
Courant en sortie : 0.7mA +/-0.05mA
mA
VR2
RISE-TIME
VR3
FS1 (mA)
kV
VR4
FS2 (kV)
FS1 800mAT
FS2 5AT (140kV X-ray generator).
8
09/11/2020
140kV
0,7 mA
0
1s
140kV
TP2 X-ray tube voltage measurement.
1V = 40.8kV
100 kV
90 kV
80 kV TP15 X-ray tube current measurement.
70 kV 1V = 0.2mA
0
10 mn 10 mn 10 mn 10 mn 10 mn 10 mn 10 mn
The voltage is rising too rapidly, and the current rising too slowly.
9
09/11/2020
The voltage is rising too slowly, and the current rising too quickly.
A setting such as this is liable to cause a trip.
10
09/11/2020
Collimator orientation for X-ray machine 628 Collimator orientation for X-ray machines 620, 622, 626, 627
11
09/11/2020
Spécifications
Carte contrôleur réf: TA55097A -180 Générateur 180kV
Voltage: 160kV +/-2kV
Courant: 1mA +/-0.05mA
RISE-TIME
VR3
mA
VR2
kV
VR4
FS1 (mA)
FS2 (kV)
12
09/11/2020
Lateral adjustment
screws Gap adjustment
Primary screws
collimator
Primary collimator
clamping screws
P/N: TA55130/2mA
Cooling
Voltage 150kV +/-2kV Fan
Lead Collar
Collimator orientation for X-ray machine 628, 632 Collimator orientation for X-ray machines 620, 622, 626, 627
13
09/11/2020
RISE-TIME
VR3
mA
VR2
kV
VR4
FS1 (mA)
FS2 (kV)
14
09/11/2020
Lateral adjustment
screws Gap adjustment
Primary screws
collimator
2mA X-ray Generator X-ray Control PCB
Primary collimator
clamping screws
Lead Collar
Collimator orientation for X-ray machine 628, 632 Collimator orientation for X-ray machines 620, 622, 626, 627
Spécifications
Carte Contrôleur:TA55097A/2mA
Générateur 180kV
Voltage: 160kV +/-2kV
Fusibles :
Courant: 1mA +/-0.05mA
FS1 T1.6A
Générateur 2mA
FS2 T15A Voltage: 150kV +/-2kV
Courant: 1.8mA +/-0.05mA
mA
VR2
RISE-TIME
VR3
FS1 (mA)
kV
VR4
FS2 (kV)
15
09/11/2020
Voltage on TP2 Real kV Reading on kvma Test Voltage on TP15 Real mA Reading on kvma Test
Box Box
The voltage is rising too rapidly, and the current rising too slowly.
16
09/11/2020
The voltage is rising too slowly, and the current rising too quickly.
A setting such as this is liable to cause a trip.
17
09/11/2020
18
09/11/2020
19
09/11/2020
Center of rotation
Center of rotation
Turning the primary collimator will allow the X-ray fan beam to cross the detector array. Since this crossing is
only a small area, the signal shows only as a localized peak. Turn the primary collimator to the opposite
position. If the peak shifts, this indicates that the centre of rotation is not on the detector array. Slide the X-
ray generator so that the centre of rotation shifts onto the array.
Center of rotation
When the centre of rotation is on the detector array, turning the primary collimator will not shift the centre of the peak in array
response. The general shape of peak will change but the peak should be at the same place vertically. The X-ray generator is
now in the correct place laterally. Rotate the primary collimator to get a good response across the whole array.
20
09/11/2020
TP2 +12Vdc
TP3 +63Vdc
21