Digital Ripple & Noise Meter

RM-103
Indispensable tool for DC Power Supply periodic check

Simple
One touch!

No reading error!

i o n o n reading.
No h e s i t a t

http://www.keisoku.co.jp/
Digital Ripple & Noise Meter

RM-103 ●● Automatic measurement possible which is correlated to


oscilloscope reading.
●● Realized “One touch measurement without hesitation” by
digital technology!
●● Automatic “PASS” or “FAIL” judgment!
●● Conforming to JEITA standard.
(JEITA: Japan Electronics and Information Technology
Industries Association,)

It is mandatory to conduct periodic maintenance of the power supplies in industrial equipment used in
power plant or other public services.
As power supply is the key devise in the system and their system will shut off when power supply failed and
will cause serious damage to our life. Normally ripple voltage and noise voltage of the power supply are
measured in the periodic check but those are not that straightforward.

Since the measuring object is “NOISE”, reading waveforms on the oscilloscope is very complicated and good
experience is needed. However digital ripple & noise meter can discriminate them and read them accurately.

Up to now……

It is very difficult to
read ripple voltage
from all mixed
waveform…
Oscilloscope is bulky
and not that user
friendly too…. We need easy way…

Noise? Ripple?

Switching power supply


Oscilloscope

Measurement by oscilloscope
The reading result is heavily relying to the operator.

Panel operation is not simple

Need additional differential probes to measure


do
differently! RM-103 measures five voltages simply!
The output of switching power supply contains various kinds of waveforms and are all combined. With RM-
103, 5 different voltages can be measured easily.

Diagram of Ripple & Noise waveform

A B E
D
C

A: Ripple & noise voltage B: Ripple voltage C: Noise voltage


D: Switching ripple voltage E: AC ripple voltage

*It is possible to judge if the capacitance of the electrolytic capacitor used inside of the switching power supply is in allowable range or not by measuring
D (Switching ripple) or B (Ripple) above.

From now on…

Do by whoever,
it is the same!
One touch simple measurement

Differential probe
Switching power supply DP-100
No hesitation Ripple & Noise Meter
at the site! RM-103

Measurement by RM-103
No reading error due to digital display

One touch simple measurement

Passive type differential probe as a standard accessory


Specification
RM-103 Ripple & Noise Meter
DC Volt. Measurement
Range ± 6.0000V ± 60.000V ± 500.00V
Resolution 0.1mV 1.0mV 10.0mV
Accuracy (*5) ± 0.025% of rdg. ± 0.025% of f.s.
Max. Applied Volt. ± 500V
Measurement Time less than 90ms(Fast Mode) / less than250ms(Slow Mode)
Ripple & Noise Measurement
Ranges 300.0mVp-p 3000.0mVp-p
Resolution 0.1mV 1.0mV
Accuracy (*1,*2,*3) ± 2% of rdg. ± 1% of f.s.
Nominal Bandwidth 100MHz
LF Bandwidth 50Hz~ 2kHz
HF Bandwidth 2kHz~
THRU Bandwidth 50Hz~ 100MHz
Band Pass Filter 20MHz
Ripple Ratio (*3) 0.5%~ 50.0% ( Resolution 0.5% )
Measurement Time (*4) approx.170ms (TYP)
Allowed Input Voltage 4Vp-p
Input Condition
Impedance DC:1MOhm , High Frequency :50 Ohm
Input Cable 50 Ohm Signal Cable RG-58 / Differential Probe DP-100
Interface
GP-IB Conform to IEEE488-1987
PASS/FAIL Judgment PhotoCoupler Output same Common (14 pin Connector)
Start Trigger PhotoCoupler Input same Common (14 pin Connector)
4CH Output Photo Coupler Output for SC-81 Control
Input Connector BNC Connector
Other Function Automatic RipplNoise Ratio Selection Function
General
Input Voltage AC90-264V(50Hz/60Hz)
Power Consumption 20VA
Dimensions , Weight 180(W)x85(H)x300(D) mm , approx.1.7kg
*1 Effective when Ripple Ratio is setup between 0% and 10%. *2 Effective when Frequency Range is between 10kHz and 10MHz. *3 Operates when
Frequency Range is between 10kHz and 100MHz. *4 Measurement Time in the same Measurement Range. *5 Guaranteed for 6 months when
Operational Temp. is 23+-5degree C and Humidity is less than 70degree C.

DM-100A Differential probe
The DP-100 reduces common mode noise drastically then measures the signal between two measurement points accurately.
No power to the probe is required.
Common mode rejection ratio CMRR = 40dB (100MHz)
Maximum input voltage ± 200V DC or AC p-p
Attention ratio 1:1
Frequency bandwidth DC ~ 100MHz
Impedance 50 Ohm ( > 1MHz)
Input capacitance 0.01uF (When terminated by RM-103)

TRC-50F2 HF terminator
When measuring Ripple & noise with an oscilloscope, using this 50 ohm terminator (50 ohm resistor and DC cut capacitor
inclusive) is recommended for accurate measurement because it can resuce Noise Reflection due to impedance unbalance. The
TRC-50F2 conforms JEITA specification.
Oscilloscope Max. Input Voltage DC ± 500V
Impedance 50 Ohm
Frequency bandwidth 1MHz ~ 100MHz
Allowable continuous power 0.25W
Differential probe DP-100A TRC-50F2 Connector BNC
Dimensions 17 φ× 54 (L) mm
※ RM-103 is equipped with a terminator equivalent to TRC-50F2.