Professional Documents
Culture Documents
IM 01E22A01-02EN
IM 01E22A01-02EN
6th Edition
1
ADMAG TI Series
AXG Magnetic Flowmeter
Maintenance Manual
IM 01E22A01-02EN 6th Edition
Contents
1. Introduction....................................................................................................2
1.1 For Safe Use of Product....................................................................................... 4
1.2 Warranty................................................................................................................. 5
1.3 Regular Inspection Items..................................................................................... 5
2. Model and Specifications Check.................................................................6
3. Changing Direction of Cable Entry..............................................................7
4. Removing and Installing Adapters for Hygienic Types.............................8
5. Transmitter Maintenance (Integral Flowmeter and Remote
Transmitter)................................................................................................. 10
5.1 Changing Direction of Display Unit.................................................................. 10
5.2 Maintenance of the Display............................................................................... 11
5.3 Removing and Installing of Display Board...................................................... 11
5.4 Parameter Backup/Restore/Duplicate.............................................................. 12
6. Hardware Switch Setting........................................................................... 13
6.1 Integral Type and AXG4A Remote Transmitter................................................ 13
6.2 AXG1A Remote Transmitter............................................................................... 17
7. Excitation Coil and Insulation Resistance Check (Remote Sensor)...............18
7.1 Removing or Installing the Cover (Remote Sensor)....................................... 18
7.2 Excitation Coil Check......................................................................................... 18
7.3 Insulation Resistance Check............................................................................. 19
8. Device Health Diagnosis by Verification Function................................. 20
8.1 Parameter of Verification Function................................................................... 20
8.2 Execution of Verification Function................................................................... 22
8.3 Confirmation of Verification Function.............................................................. 23
9. Troubleshooting......................................................................................... 24
9.1 No Display............................................................................................................ 24
9.2 Unstable Zero Point............................................................................................ 26
9.3 Disagreement Between Displayed Flow Rate and Actual Flow Rate............ 28
Revision Information
1. Introduction
This manual provides the basic guidelines for The following symbols are used in the product and
maintenance procedures of ADMAG TI (Total the manual to indicate the accompanying safety
precautions:
Insight) Series AXG magnetic flowmeters.
Protective grounding terminal
For the items which are not covered in this manual,
read the applicable user's manuals listed in "Table Functional grounding terminal (This terminal should
not be used as a protective grounding terminal.)
1.1 Manual and General Specifications List" in
Alternating current
ADMAG TI Series AXG Magnetic Flowmeter Direct current
Installation Manual (IM 01E22A01-01EN). Caution
For specifications or model codes, read ADMAG This symbol indicates that the operator must read an
explanation in the user’s manual in order to avoid the
TI Series AXG Magnetic Flowmeter General risk of injury or death of personnel or damage to the
Specifications (GS 01E22A01-01EN). product.
Website address: http://www.yokogawa.com/fld/doc/
These manuals can be downloaded from the • For the protection and safe use of the product
website of YOKOGAWA or purchased from the and the system in which this product is
YOKOGAWA representatives. incorporated, be sure to follow the instructions
and precautions on safety that is stated in
Precautions Related to the Protection, user's manual whenever you handle the
Safety, and Alteration of the Product product. Take special note that if you handle
The following safety symbol marks are used in this the product in a manner that violated these
manual and product. instructions, the protection functionality of the
product may be damaged or impaired. In such
cases, YOKOGAWA does not guarantee the
WARNING quality, performance, function, and safety of
A WARNING sign denotes a hazard. It calls product.
attention to procedure, practice, condition or the • Should use the parts specified by YOKOGAWA
like, which, if not correctly performed or adhered when replacing. Contact YOKOGAWA’s service
to, could result in injury or death of personnel. office for fuse replacement.
• Do not modify this product.
• YOKOGAWA will not be liable for malfunctions
or damage resulting from any modification
CAUTION made to this product by the customer.
• The product should be disposed of in
A CAUTION sign denotes a hazard. It calls accordance with local and national legislation/
attention to procedure, practice, condition or the regulations.
like, which, if not correctly performed or adhered
to, could result in damage to or destruction of
part or all of the product.
IMPORTANT
An IMPORTANT sign denotes that attention
is required to avoid damage to the product or
system failure.
NOTE
A NOTE sign denotes information necessary
for essential understanding of operation and
features.
IM 01E22A01-02EN
<1. Introduction> 3
Regarding This User’s Manual Trademarks:
• This manual should be provided to the end • HART is a registered trademark of FieldComm
user. Group.
• The contents of this manual are subject to • Modbus is a registered trademark of AEG
change without prior notice. Schneider.
• All rights reserved. No part of this manual • FOUNDATION is a registered trademark of
may be reproduced in any form without FieldComm Group.
YOKOGAWA’s written permission. • PROFIBUS is a registered trademark of
• YOKOGAWA makes no warranty of any kind Profibus Nutzerorganisation e.v., Karlsruhe,
with regard to this manual, including, but not Germany.
limited to, implied warranty of merchantability • "EtherNet/IP", "CIP" and "ODVA" are
and fitness for a particular purpose. trademarks or registered as trademarks of
• If any question arises or errors are found, or if ODVA, Inc.
any information is missing from this manual, • All the brands or names of Yokogawa
inform the nearest YOKOGAWA sales office. Electric’s products used in this manual are
• The specifications covered by this manual are either trademarks or registered trademarks of
limited to those for the standard type under the Yokogawa Electric Corporation.
specified model number break-down and do not • All other company and product names
cover custom-made product. mentioned in this manual are trade names,
• Note that changes in the specifications, trademarks or registered trademarks of their
construction, or component parts of the respective companies.
product may not immediately be reflected in • In this manual, trademarks or registered
this manual at the time of change, provided trademarks are not marked with ™ or ®.
that postponement of revisions will not cause • 3-A is a registered trademark of 3-A Sanitary
difficulty to the user from a functional or Standard,Inc
performance standpoint.
• This manual is intended for the following
personnel;
Engineers responsible for installation and wiring
of the product.
Personnel responsible for normal daily
operation of the product.
• To ensure correct use, read this manual and the
applicable manuals thoroughly before starting
operation. Read the general specifications for
its specification.
IM 01E22A01-02EN
<1. Introduction> 4
1.1 For Safe Use of Product
WARNING
For the protection and safe use of the product and
the system in which this product is incorporated, • Install an external switch or circuit breaker as a
be sure to follow the instructions and precautions means to turn the power off (capacitance: 15A,
on safety that is stated in user's manual whenever conforming to IEC60947-1, IEC60947-2 and
you handle the product. Take special note that if IEC60947-3).
you handle the product in a manner that violated Locate this switch either near the product or in
these instructions, the protection functionality of other places facilitating easy operation. Affix
the product may be damaged or impaired. In such a "Power Off Equipment" label to this external
cases, YOKOGAWA shall not be liable for any switch or circuit breaker.
indirect or consequential loss incurred by either • When opening and closing the cover, be sure
using or not being able to use the product. to handle the cover carefully so that there are
Maintenance no damage and foreign matter adhesion at its
threads and O-ring.
Keep checking their condition and clean the
WARNING
threads in case of adhering the foreign matter.
• Only personnel authorized by Yokogawa Replace the cover in case the treads receive
Electric Corporation can repair the equipment. damages.
• The magnetic flowmeter is a heavy product. Replace the O-ring if there is any scarring or
Be careful that no damage is caused personnel transformation.
through accidentally dropping it, or by exerting And apply silicone based grease at the O-ring in
excessive force on the magnetic flowmeter. case of the shortage and exhaustion of grease.
When moving the magnetic flowmeter, always • If dirt, dust or other substances surfaces on the
use a trolley and have at least two people carry it. glass of display cover, wipe them clean with a
• When the magnetic flowmeter is processing soft dry cloth.
hot fluids, the product itself may become • For sanitary type, be careful not to accumulate
extremely hot. Take sufficient care not to get water in holes and dents.
burnt. If accumulated, clean regularly.
• Where the fluid being processed is a toxic • Wiring work should be done adequate wire,
substance, avoid contact with the fluid and sleeve crimp and torque force. Use terminal
avoid inhaling any residual gas, even after the with insulating cover for the power supply
product has been taken off the piping line for wiring and protective grounding wiring. Do
maintenance and so forth. not pull the wires too much strongly in order
• Do not apply excessive weight, for example, a to prevent electric shocks caused by their
person stepping on the magnetic flowmeter. damage.
• Wiring and maintenance of the magnetic • This product employs the parts which are
flowmeter must be performed by expert affected by a function damage caused by static
engineer or skilled personnel. No operator electricity. Thus, you should do the antistatic
shall be permitted to perform procedures work using an anti-static wrist band for it and
relating to wiring and maintenance. be careful to avoid touching each electrical
• Do not open the cover in wet weather or humid parts and circuitry directly.
environment. When the cover is open, stated • When connecting the wiring, check that the
enclosure protection is not applicable. supply voltage is within the range of the
• Ensure that the power supply is off in order to voltage specified for this product before
prevent electric shocks. connecting the power cable. In addition, check
• When opening the cover, wait for more than 20 that no voltage is applied to the power cable
minutes after turning off the power. Only expert before connecting the wiring.
engineer or skilled personnel are permitted to • To prevent electric shocks, ensure the
open the cover. electrical wiring cover is completely attached
• Ensure the protective grounding terminal is after the wiring work.
grounded before turning on the power. • To prevent electric shocks, do not impress over
rated voltage to each input/output terminals.
IM 01E22A01-02EN
<1. Introduction> 5
1.2 Warranty
WARNING • The warranty shall cover the period noted on
• To prevent electric shocks, do not remove the quotation presented to the purchaser at the
safety cover (Read chapter 5,6). time of purchase. Problems occurred during the
• If there is any unused cable entry, use the warranty period shall basically be repaired free
of charge.
blanking plug to cover which comes with this
• In case of problems, the customer should
product or which is supplied by YOKOGAWA.
contact the YOKOGAWA representative from
The blanking plug should be fastened into the
which the product was purchased, or the
unused cable entry without any mistake. If not,
nearest YOKOGAWA office.
stated enclosure protection is not applicable.
• If a problem arises with this product, inform
• Maintenance of this flowmeter should be us of the nature of the problem and the
implemented in a maintenance service shop circumstances under which it developed,
where the necessity tools and environment including the model specification and
condition are provided. serial number. Any diagrams, data and
The necessity of this environmental condition other information you can include in your
is that ambient temperature is 5 to 40°C communication will also be helpful.
(the maximum relative humidity is 80 % for • Responsible party for repair cost for the
temperature 5 to 31°C, and decreasing linearly problems shall be determined by YOKOGAWA
to 50 % relative humidity at 40°C). based on our investigation.
• Be sure to enable the write protect function • The Purchaser shall bear the responsibility for
to prevent the overwriting of parameters after repair costs, even during the warranty period, if
finishing parameter setting. the malfunction is due to:
In rare cases, the IR switches may respond - Improper and/or inadequate maintenance by
unexpectedly to water drops or extraneous the purchaser.
substances sticking on the surface of display - Failure or damage due to improper handling,
panel, due to the operating principal. The use or storage which is out of design
possibility of malfunction arises after rain or conditions.
cleaning operation near the place where the - Use of the product in question in a location
flowmeter is installed. Turning on and off the not conforming to the standards specified
flashlight etc. towards the IR switch may also by YOKOGAWA, or due to improper
maintenance of the installation location.
be a cause of malfunction.
- Failure or damage due to modification or
Read Chapter 6 for the hardware write protect
repair by any party except YOKOGAWA or
function, and the user's manual of applicable
an approved representative of YOKOGAWA.
communication type for the software write
- Malfunction or damage from improper
protect function.
relocation of the product in question after
• For explosion protection type products, confirm delivery.
that it is "no explosive atmosphere". - Reason of force majeure such as fires,
• Use our designated parts for replacement earthquakes, storms/floods, thunder/
parts. For detailed information, contact our lightening, or other natural disasters, or
sales or service office and inform specifications disturbances, riots, warfare, or radioactive
of the product. contamination.
IM 01E22A01-02EN
<2. Model and Specifications Check> 6
F0201.ai
F0202.ai
Figure 2.1 Name Plate (AXG Integral Flowmeter) Figure 2.2 Name Plate (AXG Remote Sensor)
Figure 2.3 Name Plate (AXG4A Remote Transmitter) Figure 2.4 Name Plate (AXG1A Remote Transmitter)
F0205.ai
IM 01E22A01-02EN
<3. Changing Direction of Cable Entry> 7
CAUTION
Do not loosen the screw at the opposite side of
+90 degrees +180 degrees
cable entry (see the figure below).
F0302.ai
IM 01E22A01-02EN
<4. Removing and Installing Adapters for Hygienic Types> 8
F0401.ai
IM 01E22A01-02EN
<4. Removing and Installing Adapters for Hygienic Types> 9
CAUTION
• The lining of Hygienic type uses fluorocarbon
PFA. For the property of fluorocarbon PFA,
it is possible that the adapter mounting bolts
may loosen as time passes, so retighten them
regularly. Tighten the adapter mounting bolts in
diagonal order step by step.
• In case of leakage from adapter connection
point, retighten the adapter mounting bolts.
If leakage doesn’t stop even if they are
retightened, replace the gasket between mini-
flange of flowmeter and adapter.
• It is recommended to replace the gasket
periodically. The replacement period needs to
be decided with consideration for the frequency
of cleaning cycles, the cleaning temperature and
the fluid temperature.
• Use our designated parts for replacement parts.
For detailed information, contact our sales or
service office and inform specifications of the
product.
IM 01E22A01-02EN
<5. Transmitter Maintenance (Integral Flowmeter and Remote Sensor)> 10
Mounting screw
(two screws)
(1)
IMPORTANT
To preserve the safety, do not touch the electrical
circuit and cable of shaded area.
(1)
IM 01E22A01-02EN
<5. Transmitter Maintenance (Integral Flowmeter and Remote Sensor)> 11
(3) Installing the Cover 5.3 Removing and Installing of
• Install the cover to the flowmeter by rotating
the cover clockwise.
Display Board
• Tighten cover locking screw (1) (See Figure
(1) Removing the Display unit side cover
5.1.1 and Figure 5.1.2) counterclockwise
Read Section 5.1 and remove the cover.
using a hexagonal wrench (nominal size 3) to
lock the cover. (2) Removing the Display Board
IMPORTANT Hold the display unit with your hand and loosen
the two mounting screws.
Be sure to screw the cover firmly into the housing
without any space between them. Safety cover
Display board
WARNING
• For explosion protection type products, confirm
that it is "no explosive atmosphere".
• Before doing the maintenance work, be sure Connector
to turn off the power supply of the magnetic
Mounting screw (two screws)
flowmeter. F0504.ai
CAUTION Connector
insertion port
WARNING
Do not use the insertion port of display board for
other purposes, because some areas inside the
connector insertion port have high voltages.
Do not insert except the connector into the
insertion port of display board.
IM 01E22A01-02EN
<5. Transmitter Maintenance (Integral Flowmeter and Remote Sensor)> 12
(4) Installing the Display Board 5.4 Parameter Backup/Restore/
• Insert the connector into connector insertion
port of display board.
Duplicate
• Taking care not to entangle the cables of A backup of parameter setting can be stored to
connector, confirm the installation direction, the built-in memory on the display unit
and fix the display board with two screws. (display board). The backup be restored to
Torque of indicator fixing screw: the device and also be duplicated to other device
0.6±0.05 N·m with the same model, suffix code, hardware and
software revision through the display board with
(5) Installing the Display unit side cover
the backup. For the remote type, It is necessary to
Read Section 5.1 and install the cover.
set the device information of remote sensor into the
parameters of remote transmitter.
WARNING Read the user’s manual of applicable
communication type for checking device information
• When opening and closing the cover, be sure
from parameters and for functions in detail.
to handle the cover carefully so that there are
Display Board of
no damage and foreign matter adhesion at AXG 1
Duplicate
its threads and O-ring. Keep checking their
condition and clean the threads in case of
adhering the foreign matter.
Backup Restore
• Replace the cover in case the treads receive
damages.
AXG 1 AXG 2
• Replace the O-ring if there is any scarring
or transformation. And apply silicone based
grease at the O-ring in case of the shortage
and exhaustion of grease. F0506.ai
• Tighten while confirming that the cover rotates Device Information Duplicate Data Restore Data
Option Board ID M M
smoothly.
Transmitter Serial No M
Sensor Serial No M
Main Board Software Rev. M M
Sensor Board Software
M M
Rev.
Display Board Software
M M
Rev.
Model (Note) M M
Communication and I/O
M M
Code
Note: Only between AXG type and AXG type can be duplicated
or restored the data.
IM 01E22A01-02EN
<6. Hardware Switch Setting> 13
WARNING
To prevent electric shock and maintain
performance, do not remove the safety cover.
NOTE
The hardware switches are adjacent. Special care
should be taken when making switch settings.
IM 01E22A01-02EN
<6. Hardware Switch Setting> 14
● Setting of Burnout Switch ● Setting of Simulation Switch
(Except FOUNDATION fieldbus, PROFIBUS PA, (Only for FOUNDATION fieldbus, PROFIBUS
EtherNet/IP) PA)
The burnout function sets the direction of current The simulation function simulates the input of a
output in situations where the CPU has become function block and lets it operate as if the data was
damaged. Upon shipment from the factory, the received from the transducer block. It is possible to
burnout direction is set to High (i.e., > 21.6 mA); conduct testing for the downstream function blocks
however, in cases where the optional codes C1 or or alarm processes.
C2 have been specified, the output direction will be A SIMULATE_ENABLE switch is mounted in
set to Low (i.e., < 2.4 mA). the transmitter amplifier. This is to prevent the
Modification of the burnout direction must be carried accidental operation of this function. When this
out using the burnout switch (SW1-1) (See Figure is switched on, simulation is enabled. (See table
6.1). below)
For FOUNDATION fieldbus type, if REMOTE LOOP
Table 6.1
Burnout switch (SW1-1) TEST SWITCH is written to the SIM_ENABLE_
Position of Burnout Burnout MSG parameter (index 2922) of the maintenance
Description
Switch Direction Output transducer block, it activates the same operation as
1 2 When optional code when the above switch is turned ON. Note that this
C1 or C2 is not
H
L
ON
OFF
High > 21.6 mA
specified, the setting parameter value is lost when the power is turned
is “High”. OFF.
When optional
1 2
code C1 or C2 is
In simulation enabled status, an alarm is generated
H ON Low < 2.4 mA
L OFF specified, the setting from the resource block, and other device alarms
is “Low”. will be masked. The simulation must be disabled
immediately after using this function.
● Setting of Write Protect Switch
(Except FOUNDATION fieldbus, PROFIBUS PA)
Table 6.3 Simulation Switch (SW1-1)
The write protect function is to prevent the
Position of
overwriting of parameters. Switch
Simulation Function
Write protection can be carried out using either
1 2
the write protection switch (SW1-2) (See Figure ON:
H ON
6.1) or software function with parameter setting. If L OFF
Simulation function enabled.
Position of
Write Protect Function
Switch
1 2
OFF (Factory setting)
H ON
L OFF
Parameter can be overwritten.
1 2
ON
H ON
L OFF
Parameter can not be overwritten.
IM 01E22A01-02EN
<6. Hardware Switch Setting> 15
● Setting of Write Lock Switch • Address Switch (Position 7)
(Only for FOUNDATION fieldbus, PROFIBUS By using the address switch (Position 7), select
PA) the device address to be used from the device
The transmitter is provided with a write lock function address specified by either the hardware switch
to restrict write operations to blocks and prevent or the parameter setting.
inadvertent writing of parameter data.
For FOUNDATION fieldbus: Table 6.5 Address switch (Position 7)
To enable this function, use the write lock switch Position of Switch Description
(Hard W Lock) or the WRITE_LOCK (index 1034) 7 6 5 4 3 2 1 0
The device address set by Address Switch
(Soft W Lock). To enable write lock function, set 1
(Position 0 to 6) is used.
0
parameter in Feature Selection of Resource 7 6 5 4 3 2 1 0
The device address set by parameter
Block. For parameter in details, read section 5.12 1 setting is used.
of the user’s manual for FOUNDATION Fieldbus 0 Factory setting.
communication type.
For PROFIBUS PA: • Address Switch (Position 0 to 6)
To enable this function, use the write lock By using the address switch (Position 0 to 6), the
switch (Hard W Lock) or the WRITE_LOCKING device address is set.
(Physical Block Relative index 18). For parameter For Modbus;
in details, read section 5.12 of the user’s manual Setting range: 1 to 127
for PROFIBUS PA communication type. If the address switch is set to 0, the device
address is automatically converted to 1.
Table 6.4 Write Lock Switch (SW1-2) For PROFIBUS PA;
Position of
Setting range: 0 to 126
Write Lock Function
Switch Setting example:
1 2
If only the address switch “position 6” is set to
OFF:
H ON Write Lock function disabled. 1, the resulting the device address is 64.
L OFF (It depends on the setting in Feature Selection.) (1 * 26 + 0 * 25 + 0 * 24 + 0 * 23 + 0 * 22 + 0 * 21
+ 0 * 20)
1 2 ON:
H ON Write Lock function enabled.
L OFF (It depends on the setting in Feature Selection.) For the product with EtherNet/IP communication,
the device IP address can be set using either the
address switch (ADDRESS) (See Figure 6.1) or
NOTE software parameters.
The Address Switch (ADDRESS) is used to set the
• If the hardware switch is set to “ON”, the 4th octet (8 bits) of the IP address. The 1st, 2nd
condition of preventing parameter overwriting and 3rd octets of the IP address are set by software
keeps until the switch is set to “OFF”. parameters. The 4th octet of the address which
• For the software write protect function, read is set by this switch (ADDRESS) is enabled when
the user’s manual of applicable communication the Enable hardware switch (SW3-1) is set to ON.
type. When the Enable hardware switch (SW3-1) is set
to OFF, the 4th octet of the IP address set by this
● Setting of Address Switch (ADDRESS) switch (ADDRESS) is disabled, and the 4th octet of
This switch is only available for the product the address set by software parameter is enabled.
with Modbus, PROFIBUS PA, or EtherNet/IP
communication. Table 6.6 Address switch (Position 7)
For the product with Modbus or PROFIBUS PA
Position of Switch Description
communication, it is necessary to set the device 7 6 5 4 3 2 1 0 Setting range: 1 to 244
address. 1 If all switches are set to "OFF" or "ON",
0 the device IP address (4th octet) is
The device address can be set using either the ADDRESS automatically converted to 210.
address switch (ADDRESS) (See Figure 6.1) or
software function with parameter setting. Setting example: If only the address switch “position
7” is set to 1, the device IP address (4th octet) is
128. (1 * 27 +0 * 26 + 0 * 25 + 0 * 24 + 0 * 23 + 0 * 22 +
0 * 21 + 0 * 20)
IM 01E22A01-02EN
<6. Hardware Switch Setting> 16
● Setting of Line Termination Switch (SW2) ● Setting Enable Hardware Address Switch
This switch is only available for the product with (SW3-1) (Only for EtherNet/IP)
Modbus communication. This switch is used to enable or disable the Address
Line terminations of two ends on the bus are Switch (ADDRESS) and DHCP Switch (SW3-2).
required to communicate Modbus.
One termination mode can be set using the line Table 6.9 Enable hardware address switch (SW3-1)
termination switch (SW2) (See Figure 6.1). Position of
Description
Switch
Table 6.7 Line termination switch (SW2) 1 2
Master
D R
Pull Up
R R LT:Line Termination
D D
ADMAG Slave n
F0602.ai
IM 01E22A01-02EN
<6. Hardware Switch Setting> 17
6.2 AXG1A Remote Transmitter (1) Setting of Burnout Switch
The burnout function sets the direction of current
IMPORTANT output in situations where the CPU has become
• Removing and installing the cover are necessary damaged. Upon shipment from the manufacturing
for setting the hardware switches. Perform factory, the burnout direction is set to High (i.e.,
removing and installing the cover as described >21.6 mA); however, in cases where the optional
below. When opening the cover, wait for more codes C1 or C2 have been specified, the output
than 20 minutes after turning off the power. This direction will be set to Low (i.e., <2.4 mA).
work must be carried out by the trained personnel Modification of the burnout direction must be carried
having knowledge of safety standard. out using the burnout switch (i.e., SW1-1) (See
• To preserve the safety, do not touch the electrical Figure 6.4).
circuit and the cables except the setting switches.
Table 6.11
Burnout switch (SW1-1)
(1) Remove the cover (See Figure 6.3). Position of Burnout Burnout
Description
Switch Direction Output
L H When optional code
1
C1 or C2 is not
2
High >21.6 mA
specified, the setting
OFF ON is “High”.
L H When optional
1
code C1 or C2 is
2 Low <2.4 mA
specified, the setting
OFF ON is “Low”.
Table 6.12
Write protect switch (SW1-2)
Position of
Write Protect Function
Switch
L H
1 OFF (Factory setting)
2 Parameter can be overwritten.
OFF ON
L H
F0604.ai 1 ON
2 Parameter can not be overwritten.
Figure 6.4 Hardware switch (SW1-1, SW1-2)
OFF ON
IM 01E22A01-02EN
<7. Excitation Coil and Insulation Resistance Check (Remote Sensor)> 18
(1)
F0701.ai
IM 01E22A01-02EN
<7. Excitation Coil and Insulation Resistance Check (Remote Sensor)> 19
7.3 Insulation Resistance Check
Check the insulation resistances in the remote
sensor in accordance with the tables below. If any
of them falls below the values listed in the tables,
consult YOKOGAWA’s sales or service offices.
If the insulation resistance cannot be restored,
replacement or repair of the sensor is necessary.
Coil Circuit
Checking is possible even if the pipe is filled with
fluid.
Terminal Test Voltage Specification
Excitation current 500 V DC (Use an 1 MΩ or more
terminal (EX1) and insulation tester or
Common terminal the equivalent.)
(C)
Signal Circuit
Before checking, be sure to empty and dry the
interior of the pipe, checking that there is no
adhesive material.
Terminal Test Voltage Specification
Common terminal 500 V DC (Use an 100 MΩ or more
(C) and Signal insulation tester or (between each
terminal (A) the equivalent.) terminal)
Common terminal
(C) and Signal
terminal (B)
IMPORTANT
In the case of the explosion protection type (Suffix
Code expect for 000: non-explosion protection
type), Signal Circuit can't conduct Insulation
Resistance Check.
CAUTION
In case of the following types, remove the wiring
connection on the transmitter side and measure
resistance at the cable terminals.
• Submersible Type
• For District Heating and Cooling or
Condensation-proof (Optional code: DHC)
IM 01E22A01-02EN
<8. Device Health Diagnosis by Verification Function> 20
IM 01E22A01-02EN
<8. Device Health Diagnosis by Verification Function> 21
Display Description
Device setup ► Diag/Service ► Specifies the fluid status to execute the
(1)
Verification ► Mode verification function.*1
Device setup ► Diag/Service ► Specifies the execution of the verification
(2)
Verification ► Execute function.*2
Device setup ► Diag/Service ►
(3) Specifies the diagnosis result display time.*3
Verification ► VF No
Device setup ► Diag/Service ►
(4) Verification ► Result ► Displays the diagnosis result.*4
Failed/Passed
Device setup ► Diag/Service ►
(5) Verification ► Result ► Displays the magnetic circuit diagnosis result.*4
Magnetic circuit
Device setup ► Diag/Service ►
(6) Displays the excitation circuit diagnosis result.*4
Verification ► Result ► Excite circuit
Device setup ► Diag/Service ►
(7) Displays the calculation circuit diagnosis result.*4
Verification ► Result ► Calc circuit
Device setup ► Diag/Service ►
(8) Displays the device alarm diagnosis result.*4
Verification ► Result ► Device status
Device setup ► Diag/Service ►
(9) Verification ► Result ► Displays the wiring misconnection check result.*4
Connect status
*1: From the table below, select the fluid status required to execute the verification function.
Selection
Description
Display
No flow Executes the verification function with the fluid not flowing.
Flow Executes the verification function with the fluid flowing.
*2: From the table below, select whether to execute the verification function.
Selection
Description
Display
Not execute Does not execute the verification function.
Execute Executes the verification function.
*3: From the table below, select the execution result of the verification function.
Selection
Description
Display
Factory Displays the result obtained upon shipment from the manufacturing factory.
Previous Displays the previous result.
Present Displays the result at this time.
*5: When the alarms in the following table occur or big flow noise is detected while the diagnosis runs, the diagnosis will stop.
Display
010: Main CPU FAIL
014: Snsr bd FAIL
015: Snsr comm ERR
016: AD 1 FAIL[Sig]
017: AD 2 FAIL[Excit]
050: Signal overflow
051: Empty detect
IM 01E22A01-02EN
<8. Device Health Diagnosis by Verification Function> 22
8.2 Execution of Verification Function
In accordance with the following flowchart, execute the verification function.
START
NO
Is the sensor completely full of fluid?
YES
Does any alarm or warning occur?
Unknown or Failed
IM 01E22A01-02EN
A
START
Among the
parameter (5) to (9) in Section 8.1, which
parameter indicates "Failed" or "Unknown"?
What is the diagnosis result What is the diagnosis result What is the diagnosis result What is the diagnosis result What is the diagnosis result
Unknown
of the parameter (5) in Section 8.1? of the parameter (6) in Section 8.1? of the parameter (7) in Section 8.1? of the parameter (8) in Section 8.1? of the parameter (9) in Section 8.1?
(Magnetic circuit) (Excitation circuit) (Calculation circuit) (Device status) (Connect status)
See Flowchart B.
See Flowchart B.
See Flowchart B.
8.3 Confirmation of Verification Function
B
START
In accordance with the following flowchart, confirm the diagnosis result.
NO
F0802.ai
IM 01E22A01-02EN
23
<9. Troubleshooting> 24
9. Troubleshooting
This chapter describes troubleshooting procedures, when the trouble occurs in the magnetic flowmeter.
For AXFA11, read the applicable user’s manual.
9.1 No Display
If nothing is displayed on the display even though the fluid is flowing, check according to the flowchart below.
START
NO
Is the sensor completely full of fluid?
IM 01E22A01-02EN
<9. Troubleshooting> 25
(Continued)
NO
Is the electrode resistance
4 MΩ or less?
Remove the sensor from the piping and take the foreign
YES matter away from the interior of the sensor. Especially the
adhesive materials on the electrode surface and around the
grounding ring must be removed.
YES
Was the problem solved?
NO FINISH
YES
Is foreign matter adhering to the interior
surface and the grounding ring?
NO
Does the insulation resistance
fulfill the specification?
YES
IM 01E22A01-02EN
<9. Troubleshooting> 26
9.2 Unstable Zero Point
if the adjustment value or output value of the zero point is unstable, check according to the flowchart below.
START
NO
Is the sensor completely full of fluid?
YES
Are there bubbles in the fluid?
YES
Is valve or seal part between the
sensor and the piping leaking?
NO
Is the flowmeter properly grounded?
YES
Was the problem solved?
NO
FINISH
IM 01E22A01-02EN
<9. Troubleshooting> 27
(Continued)
YES
Is the fluid likely to adhere to the piping?
NO
YES
Is the fluid likely to adhere to the piping?
NO
IM 01E22A01-02EN
<9. Troubleshooting> 28
9.3 Disagreement Between Displayed Flow Rate and Actual Flow Rate
If the displayed flow rate does not match the actual flow rate, check according to the flowchart below.
START
YES
NO
Is the flowmeter properly grounded?
IM 01E22A01-02EN
<9. Troubleshooting> 29
(Continued)
YES
Was the problem solved?
NO
FINISH
YES
Is the fluid likely to adhere to the piping?
NO
IM 01E22A01-02EN
<9. Troubleshooting> 30
(Continued) (Continued)
YES
Is the fluid likely to adhere to the piping?
Drain the fluid from the sensor and dry the interior fully to
Contact YOKOGAWA service center. restore insulation.
F0907.ai
IM 01E22A01-02EN
31
Revision Information
Title : ADMAG TI Series AXG Magnetic Flowmeter Maintenance Manual
Manual No. : IM 01E22A01-02EN
IM 01E22A01-02EN