Phoenix Medical Systems, INDIA

We Work towards reducing child mortality rates and to improve maternal health
Infant Radiant Warmer NWS101

Trouble Shooting
Power supply board Connector details

Main supply 230 volt AC

Battery connector location JP7


Voltage connector

Signal connector
Heater connector
Control board Connector details

Probe Connector

On/off switch connector

Programming IC

Voltage Connector

Signal Connector
Block Diagram

Buzzer RAM Keys

Analogue section

THERMISTOR Display section

AMPLIFIER MICRO PROCESSOR DISPLAY


LATCH DISPLAY
DRIVER

ANALOGUE
SWITCH
EEPROM

Battery

SMPS TRIAC HEATER

Power supply section & heater section


Wiring Diagram

From the Control


Module
1 2
RED (14cm)
4 way I/P
Johns Plug
3 4

GREEN (14cm)

Observation
Lamp
BLACK (14 cm)
driver
BLUE ( 92cm )

Gray
Gray
BLUE ( 52cm )

LED OBSERVATION
Teflon Wire Teflon Wire LAMP
IR LAMP / HEATER LAMP 600W
Heater Fail
“Ort” and Probe fail
Trouble shooting Probe fail
Trouble shooting Erotic Display
Trouble shooting Probe fail
Trouble Shooting- Observation Lamp
Trouble shooting

Problems Root cause Solution

Check the fuse if burn out,


1.Equipment Not Switching On Input Main Supply and Fuse ,
replace new one.

Check the resistance value of


90+/- 5 Ohms.
2.”HEF” Heater fail Showing in the display IR lamp and Relay
Check Relay resistance and
contacts

Check the probe resistance 3 to 5


3.”Prb” probe fail Probe and 3 pin RMC wire k ohms. Check the continuity from
3 pin RMC to control board

4.System Failure Programming IC Check Programming voltage

5.Seven Segment Not Working Check sevmensegment led Check transistor using multimeter
Analog Section
Heater Section Systematic Diagram
Infant Radiant Warmer OPERATION CONTROL PANEL
Infant Radiant Warmer
Alarms

Manual mode alarm Manual mode alarm

In the manual mode, the mutable alarm is produced every 15 minutes


In the manual mode, the mutable alarm is
produced after 15 minutes. if we set it heater output above 60%.
Infant Radiant Warmer
Alarms

Low temperature Alarms High temperature Alarms Working condition


Infant Radiant Warmer
Alarms

Possible cause : If Probe is disconnected or probe is faulty

Action: Check the probe if it is not good, change the probe or call Service Team
Infant Radiant Warmer
Power Failure System Failure

Trouble : Power Failure


Trouble : System fail light glowing on control panel with continuous alarm
Possible cause: There is no Input Mains supply (220V)
Action: Do not use the equipment in the system failure condition and call the Service
Action: Check whether the equipment is connected to the Main supply Team.
Any Questions
Thank you

PHOENIX MEDICAL SYSTEMS PVT LTD. – Chennai.

PHOENIX drives to work towards


a future where every Newborn
has access to
high -quality Medical Care