SAIT Polytechnic

QUALITY MANAGEMENT SYSTEM MANUAL


REVISION: 2 DATE: 2014/03/24 Section 12 page 1 of 1

CALIBRATION CERTIFICATE
Company SAIT Polytechnic Contact(s) Phone 403 774 4794

CERTIFICATE NUMBER: 566853-00-2 JOB NUMBER: S3369

Date 15/01/13 Date due 16/01/13

Instrument Type: WIKA pressure gauge Serial No: 1091

Calibration Environment: 21°C +/- 3°C Number of Points: 3000 Kpa Max.

Procedure

Verify pressure and temperature, clean and certify

Results were within acceptable limits of pressure +/-2Kpa @ +/- 0.5 °C and +/- 5%

Results

Instrument Details

CALIBRATION REFERENCE INSTRUMENT


Deadweight 062250 09/07/12
S/N:

CERTIFICATE OF REFERENCE
10647
INSTRUMENT:

RECALIBRATION DATE OF REFERENCE


INSTRUMENT:

Signature: ………………………...................................…
K,Kaasewski
SAIT Polytechnic
QUALITY MANAGEMENT SYSTEM MANUAL
REVISION: 2 DATE: 2014/03/24 Section 12 page 1 of 1

Difference between
Instrument Reference Instrument
Target value reference and Off set
Reference Measurement Measurement
instrument

062250/1091 °C/Kpa °C/Kpa °C/Kpa °C/Kpa °C/Kpa

062250/1091 21/3000 21/3000 21/3001 0/1 N/A

062250/1091 16/3000 16/3000 15/3001 1/1 N/A

062250/1091 7/3000 7/3000 7/3001 0/1 N/A

062250/1091 2/3000 2/3000 3/3000 1/0 N/A

062250/1091 -3/3000 -3/3000 -2/3000 1/0 N/A