1 Safety instructions (note)
Thank you for purchasing the company xenon test program controller.
This operating manual describes the operation of this product is as follows.
(A) Please refer to the operating manual for the final use of the personnel, at any time to
check the place, so that at any time to check.
(B) Please read the operating instructions carefully before using this product.
(C) The operating instructions in detail the various functions of the product, the operation
of the operating instructions are not guaranteed.
(D) To edit, copy, and use the part or all of the contents of this operating manual without
permission.
(E) The contents of the operating instructions can be changed without prior notice or
notice.
(F) The operating instructions although experienced all kinds of efforts to write complete,
but rather will have poor or typo, missing, if found please contact the place of purchase or
the company contact. Thank you very much.
(A) In order to ensure the system maintenance and safety of the product and the connection
of this product, please read the relevant safety precautions (instructions) in this operation
and then use the product.
(B) The company does not take any responsibility for any loss or damage caused by the use
of the contents of the instructions in accordance with the instructions of this operation.
(C) In order to ensure the system maintenance and safety of the products and the use of this
product, it is required to install additional protection devices or safety circuits. It is strictly
prohibited to transform the interior of the product (change) or add.
(D) Please do not arbitrarily break down, repair or transformation, it may cause electric
shock, fire or failure to work.
(E) The replacement of the product parts or consumables please contact the sales
department.
(F) Please do not allow water to flow into this product, it is possible to cause trouble.
(G) Please do not hit this product, it is possible to cause damage or failure of products.
(A) In addition to the company's quality assurance conditions, the company will not be
guaranteed, it is not responsible for any.
(B) In the process of using this product, the company shall not be responsible for any
damage to the user or the third party directly or indirectly, as a result of the defects or
natural disasters that the company is unable to predict.
2
2 Operation and setting
▶ After the initial installation of a good product insert power, showing the start of the
progress screen.
☞ Start screen takes about 15 seconds.
Start screen
Home Screen
3
2.2 Parameter setting
① Digital input keyboard
4
2.2.1 The validity of setting key and setting value
▶ The validity of the key and the setting value of the product is to confirm whether the set
value of the input key is valid or not.
☞ “Bi”: When you click on the basic set button, enter the set value of the normal time
☞ “BBB”: The setting value of the input box is outside the input range.
① Display parameters.
② Display range.
③ Set the value of the display window, beyond the scope of the setting will display the
fault warning information ("Input Error").
④ To stop the input, to return to the original image of the use of the key.
⑤ After storing the input value, return the original picture.
⑥ Enter the decimal point.
⑦ Input (+ / -) symbols used.
⑧ When modifying the input value, the input value can be deleted one by one.
⑨ Delete all of the input values .
5
3. Running state setting
3.1 Home Screen
No Content Describe
Monitor
① Switch to run screen 。
screen
Curve Switch to curve representation and curve storage, USB storage
②
display devices, and export curves.
Operation
③ Switch to function and running mode settings screen.
setting
Program
④ Switch to program settings main screen.
settings
Reserve Switch to set the current time and the appointment time of the
⑤
set screen .
Screen
⑥ Using the prompt tone, the display settings screen.
setting
6
3.2 Fix Run
▶ [3.1 Home Screen] Select Monitor Screen,Can be switched to " Fix Run first run
screen".
▶ [4. Operation setting]Select run mode as " FIX".
▶ The input and operation method of the temperature setting value are as follows.
▶ When you enter the temperature setting, click the red button at the top of the left, which
can be activate [Figure 3-3] Input keyboard for temperature setting value.
7
▶ The input and operation method of humidity setting value are as follows.
▶ When you enter the humidity setting, click the blue button at the bottom of the left,
which can be activate [Figure 3-4] Humidity set value input keyboard
▶ After inputting the temperature and humidity setting value, select the right lower end of
the Run key, can perform the fixed value operation.
8
3.2.2 Fix Run first run screen
9
3.2.3 Fixed value (FIX) run second run screen
▶ The left side of the picture shows the setting value of the condensation temperature,
condensation humidity, radiation intensity and the blackboard temperature, and the output
value of the control.
▶ To set whether to display the data through the tick box.
① Indicates the current temperature, humidity, the command value, the set value, the
control output value.
② Enter the USB storage device screen.
③ Delete current test curve.
④ Movement of time axis.
☞ The sampling time is 30 seconds and can be stored for about 56 days.
10
3.3 Program Run
▶ [3.1 Home Screen] select "Monitor Screen",can be switched to "Pogram Run first run
screen".
▶ [4. Run operation settings] The running mode is selected as the "program ".
▶ Program setting method, please refer to [5.1 program settings].
▶ [Figure 3-9 Program Run first stop screen] in the stop screen click on the right side at the
lower end of the Run button will switch to [Figure 3-10 Program Run first run screen] .
11
3.3.2 Program Run first run screen
12
3.3.4 Program Run third run screen
① Indicates the current temperature, humidity, the command value, the set value, the
control output value.
② Enter the USB storage device screen.
③ Delete current test curve.
④ Movement of time axis.
☞ The sampling time is 30 seconds and can be stored for about 56 days.
13
4. Run operation settings
14
[Figure 4-2] Enter lock set enter password
Enter the lock set enter password: default is 0
15
5. Program Settings
▶ Click the Program Settings button, then the screen to switch to [Figure 5-1 Program
Settings screen].
▶ Program set running related parameters settings screen
No Content Describe
Program
① Switch to program edit screen
Editor
② Loop Set Switch to cycle set mode and partial loop
File
③ Switch to set mode to copy and delete the screen
Management
④ Cycle Set Switch to cycle settings screen
16
5.1 Program editor
17
▶ Click (Program Number) button to display the input box that can be set mode number.
18
5.2 Program loop settings
▶ Set the screen for all or part of the loop function of the set.
[Figure 5-3] The program and part of the loop settings screen
19
5.3 File Management
▶ [5.1 Program Settings] in the input parameters in the program can be copied to other
programs, can also be deleted.
▶ Cannot delete running program number.
▶ Deleted programs cannot be recovered.
20
21
5.4 Cycle Setting
▶ The cycle setting is divided into a ON/OFF operation and a time period setting
operation, in which a set time signal period is set in the program segment setting for the
[5.1 Program Setting]
22
5.5 Wait Settings
▶ The program group runs in the process of waiting for the action, set the temperature
and humidity waiting for the scope and time of the screen.
▶ In this setting the waiting action is applied to the [5.1 Program Setting].
23
5.6 Test name setting
▶ Click (Name) button to display the input keyboard to set the name of the test.
24
6. History curves and storage devices
25
7 Screen display settings
7.1 Screen display settings
① Language selection.
② The entire set of buttons in the monitor screen is displayed.
③ Calibrate touch screen.
④ Set whether to use prompt tone.
☞ Even if the setting is not used, the DI fault occurs when the prompt tone will not be
closed.
⑤ Can be set back light energy saving time.
☞ Energy saving working time setting is to refer to no need to operate the button, how
long the back light lamp is closed.
⑥ Set user password.
26
7.2 DI Fault record view
▶ Display the breakdown of the DI type and date, time of the screen.
▶ Fault records can show up to 30.
☞ If the failure number more than 30, starting from the front to delete fault number, after
the occurrence of the fault will meet at the end.
27
8 Time setting
28