FLEXIT MultiSmart multi-shot drillhole survey

system with StoreIT Data Pad


SmartTool Operator Checklist
To run a single survey
1 Switch on StoreIT Data Pad—press O| . Start up screen shows
briefly with unit serial number, followed by
First survey... press NEW
2 Check/change survey settings
Scroll SETTINGS to Hole Ref Hole Ref ABnn
Use +_ _ OK to set
Scroll SETTINGS to Start Depth Start Depth 0 m
Use + _ to set start depth
Scroll SETTINGS to Depth Interval Depth Interval M m For backlighting press
Use + _ to set depth interval
Scroll SETTINGS to Survey Direction Survey Direction Now:In
Use + _ to set In or Out
Scroll SETTINGS to Time Interval Time Interval T Secs
Use + _ to set time interval
Scroll SETTINGS to Time Set Time hh:mm:ss
Use + _ OK to set time
Scroll SETTINGS to Date Set Date dd:mm:yy

Use + _ OK to set date


3 When settings complete
Press and hold DATA down First survey... press NEW

4 Start SensIT
Bring SensIT close to StoreIT
Press and hold NEW down Depth (start) m Press OK

5 Load SensIT into running gear


Run SensIT to start depth
Press and hold OK down Measuring, wait...y secs
Depth (2nd) m Press OK
Move SensIT to next depth
Press and hold OK down Measuring, wait...y secs
Depth (3rd) m Press OK
Repeat for as many stations as needed

6 End survey Depth (last+1) m Press OK


Scroll SETTINGS to find… End Survey? Press OK
Press and hold OK down Next Survey...Press NEW
Scroll SETTINGS to find… End Session? Press OK

7 Recover SensIT and download data to StoreIT


Bring SensIT close to StoreIT
Press and hold OK down Contacting tool...
Downloading data
Record Number xxx
8 Turn StoreIT OFF First Survey... Press New
Press O|
Continued overleaf 
9 Change current depth to repeat earlier station or to change interval
Depth (x) m Press OK

Scroll SETTINGS to find … Current Depth (x) m


Press + _ to select new depth Current depth (new) m
Press DATA Depth (new) m Press OK

Run SensIT to depth shown


Press and hold OK down Measuring, wait… y secs

10 Change depth interval (in addition to above) Depth (x) m Press OK


Scroll SETTINGS to find … Depth Interval (x) m
Press + _ to select new interval Depth Interval (new) m
Press and hold DATA down Depth (new) m Press OK
Run SensIT to depth shown
Press and hold OK down Measuring Wait... secs

To run additional surveys before uploading


11 End first survey Depth (last+1) m Press OK
Scroll SETTINGS to find … End Survey? Press OK
Press and hold OK down Next Survey... Press NEW

12 Check/change second survey settings

Scroll SETTINGS to Hole Ref Hole Ref ABnn


Use + _ OK to set hole reference
Scroll SETTINGS to Start Depth Start Depth xx m
Use + _ to set start depth
Scroll SETTINGS to Depth Interval Depth Interval yy m
Use + _ to set depth interval
Scroll SETTINGS to Survey Direction Survey Direction Now: In
Use + _ to set survey direction

13 When settings are complete

Version 2.0/Aug 2001©/RWB-RB. http://www.flexit.se/


Press and hold DATA down Next Survey... Press NEW
Press and hold NEW down. Continue as Steps 5-8

Download StoreIT to a PC to calculate and display surveys


You may run several surveys in one or more sessions before downloading StoreIT data to a
PC. To download, connect TransIT to your PC. Open MeasureIT software on your PC.
Press O| to switch ON your StoreIT data pad

Choose DOWNLOAD from first menu of MeasureIT and follow on screen instructions.
See User Guide for further MeasureIT information.
Press O| to switch OFF First Survey... Press NEW

Notes
1 If SensIT is too far from StoreIT and cannot communicate Tool Not Responding
2 To switch StoreIT to single-shot mode

Scroll SETTINGS to find … Single-Shot mode?