You are on page 1of 7

Fluorescence Measurements on Functionalized

Polymer Surfaces—Problems
and Troubleshooting
KATRIN HOFFMANN, RENATE MIX, UTE RESCH-GENGER, AND JOERG F. FRIEDRICH
Federal Institute for Materials Research and Testing (BAM), Div. I.5, Berlin, Germany

Plasma-chemically tailor-made polymer surfaces are of ever-increasing importance to control


surface properties in material science, as well as for (bio)analytical and biomedical applications.
For the characterization of such systems, sensitive fluorescence techniques are attractive tools. To
underline the potential and drawbacks of these strategies, this article addresses different problems
that complicate fluorometric analysis. To overcome some of these limitations, such as nonspecific
adsorption of unreacted fluorescent probes, we discuss potential troubleshooting, including the
use of a chromogenic and fluorogenic pyrylium dye for the detection of amino functionalities at
polypropylene surfaces.

Key words: fluorescence spectroscopy; polymer functionalization; surface labeling; pyrylium label

Introduction surface functions is complicated by different factors,8,11


including nonspecific adsorption,8 an inhomogeneous
Functionalization of polymer surfaces with OH, dye distribution or penetration of dye molecules into
NH 2 , or CHO groups by, for example, plasma chem- polymers,12–14 dye–dye and dye–matrix interactions,
ical modification allows control of the hydrophilicity, and the sensitivity of the fluorophore’s spectroscopic
as well as the adsorption and wetting properties of properties to the microenvironment.25 Because the
polymeric surfaces and provides the basis for the at- potential of fluorometry is still under debate, this
tachment of bio- and sensor molecules.1,2 Crucial for report discusses the possibilities of and general re-
all applications of these tailor-made polymeric materi- quirements on the fluorometric characterization of
als is the characterization of the type and the density of dye-labeled surface functionalities, exemplary for
reactive functional groups at the surface. Commonly plasma-functionalized polypropylene (PP) films. The
used surface-sensitive techniques such as X-ray photo- results obtained with a traditional dansyl label that
electron spectroscopy and attenuated total reflection does not reveal binding-induced changes of its spec-
Fourier transform infrared spectroscopy are expen- troscopic properties with those of a chromogenic and
sive, provide only the elemental composition of a 5- fluorogenic pyrylium dye are compared.
to 7-nm-thick surface layer, or are limited to a surface
layer with a thickness of 1.6–2.5 µm3 ; they often fail Results and Discussion
at low concentrations of surface functionalities. Thus,
the application of straightforward labeling techniques Plasma Chemical Modification
in combination with simple fluorometric methods be- An elegant approach to introduce functional groups
comes attractive.4 into chemically inert polymer surfaces is the appli-
Despite the obvious potential of fluorometry, there cation of plasma processes, thus forming anchoring
have been only a few examples reported for the charac- sites for chemical grafting of different labels or other
terization of surface functionalities on (bio)analytically functional molecules. For the presented spectrofluoro-
relevant supports.4–10 Fluorometric characterization of metric characterization of PP films (100-µm thickness;
Goodfellow, Cambridge, U.K.), we used two different
routes for surface modification. The first approach is
generating 10–14 OH groups/100 C atoms on the
Address for correspondence: Dr. Katrin Hoffmann, Federal Institute PP surface by means of an oxygen plasma, followed
for Materials Research and Testing (BAM) I.5 Bioanalytics, Richard-
Willstätter-Str. 11, D-12489 Berlin-Adlershof, Germany. Voice: +030-
by a wet-chemical reduction process. Another promis-
8104-5878; fax: +030-8104-5005. ing approach is the plasma-initiated polymerization
katrin.hoffmann@bam.de of monomers carrying functional groups. With this
Ann. N.Y. Acad. Sci. 1130: 28–34 (2008). 
C 2008 New York Academy of Sciences.
doi: 10.1196/annals.1430.015 28
Hoffmann et al.: Fluorescence Spectroscopy on Polymer Surfaces 29

FIGURE 1. Labeling reactions. (A) Labeling of OH groups at a PP surface with a dansyl fluorophore
by using a diisocyanate spacer. (B) Labeling of surface amino groups with the blue Py-1 chromophore,
yielding a red PP conjugate. (In color in Annals online.)

strategy, NH 2 -terminated PP surfaces have been gen- sensitive technique.6 Second, further complications
erated.12,15 can arise from background emission either from the
polymer support itself and/or from adsorbed (non-
Labeling of Surface Functionalities linked) dye molecules. Furthermore, the environmen-
Different synthetic concepts have been tested for tal dependence of the spectroscopic properties of most
fluorophore labeling of surface functionalities,13,15 dyes and matrix–dye or dye–dye interactions result-
as well as to introduce spacer molecules.16 Con- ing in fluorescence quenching must be taken into ac-
ventional isothiocyanate labels or amino-terminated count.20 All these problems raise qualms about the
labels like dansyl fluorophores were covalently at- reliability of quantification relying on fluorescence
tached by means of spacer molecules to OH func- measurements after labeling reactions. Despite these
tions at the polymer surface by strategies adopted from, limitations, the unique sensitivity of fluorometry with
for example, well-investigated polyurethane chemistry a limit of detection on the order of approximately
(FIG. 1A).13 For the fluorometric characterization of 10−12 mol/cm2 for fluorophores on polymer surfaces,8
amino-functionalized PP surfaces, the chromogenic in conjunction with its comparative ease of use, renders
and fluorogenic pyrylium dye Py-1 has been used, fluorescence techniques as attractive tools for analyz-
which was recently introduced for the derivatization ing organically modified polymer surfaces, especially
of primary amino groups (FIG. 1B).17,18 For each series when only monitoring of changes in surface group con-
of fluorescence measurements,19 blank or so-called ref- centration is needed.
erence samples were prepared by reaction of the non-
functionalized polymer film with the respective flu- Effect of Spectral Correction:
orescent label by using the same procedure as that Comparability of Data
used to covalently attach fluorophores to the surface- In general and independent of the type of fluores-
functionalized films. cent samples (e.g., fluorophores in solution or fluores-
cent labels attached to polymer surfaces), the fluores-
Problems Inherent to Fluorescence cence signal is affected by both the fluorescent analyte
Measurements at Surfaces and instrumental effects. Relevant properties of ana-
Fluorescence techniques in conjunction with fluo- lytes are the fluorophore’s absorptance at the excita-
rescent labels can be exploited to characterize sur- tion wavelength and its fluorescence quantum yield.
face functionalities at low concentration levels in the These quantities, as well as the spectral shape and po-
range between approximately 109 (on glass4 ) and 1012 sition of the fluorophore’s absorption, excitation, and
(on polymer supports8 ) mol/cm2 . The reliable use of emission spectra, can strongly depend on the polarity
these techniques, however, requires overcoming some of the dye’s microenvironment, that is, on the solvent
challenges.7 First, the observed fluorescence signal re- or the matrix (e.g., polarity, proticity, pH, and viscos-
sults from the bulk material because commonly used ity). This dependence complicates quantification from
steady-state fluorescence spectroscopy is not a surface- measurements of relative fluorescence intensities. From
30 Annals of the New York Academy of Sciences

TABLE 1. Selected spectroscopic data (spectrally cor-


rected) of PP-DNS-Hy in different solvents
ET a / Stokes
Solvent kJ mol−1 ν̃a b s /cm−1 ν̃e m /cm−1 shiftb /cm−1

1.4-Dioxane 150.5 30 000 21 790 8210


DMFc 183.4 30 140 21 890 8250
DMSOd 188.4 30 090 22 055 8035
Propanol 203.2 30 010 21 990 8020
Ethanol 216.9 30 140 21 840 8300
Water 263.8 29 930 21 680 8250
Dry (air) —e 29 850 21 770 8080
a
E T , Dimroth and Reichardt’s polarity scale of solvents.23
b
The Stokes shift is calculated from the energetic difference
between the longest wavelength maximum of the fluorescence
FIGURE 2. Comparison of the measured uncorrected excitation spectra (ν̃a b s ) and the corresponding emission maxima
(gray) and the spectrally corrected (black) excitation (ν̃e m ).
c
(dashed lines) and emission (solid lines) spectra of a labeled DMF, dimethyl formamide.
d
PP film. The PP–Py-1 conjugate was excited at 503 nm, and DMSO, dimethyl sulfoxide.
e
the excitation spectrum was recorded at 602 nm. —, none.

a much more pronounced dependence of their spec-


the instrument side, the excitation light intensity at troscopic features on microenvironment.23 Their sol-
sample position and the wavelength- and polarization- vatochromic behavior can be used to study the accessi-
dependent spectral responsivity of the emission chan- bility of small molecules24 and to characterize local po-
nel affect the shape and the intensity of the mea- larity effects, which mainly determine the properties of
sured fluorescence spectra. This effect is highlighted in surface-modified polymers.3,25,26 To illustrate such ef-
FIGURE 2, which compares the measured instrument- fects for dansyl chromophores bound to PP surfaces, we
specific fluorescence spectrum of the pyrylium dye determined the microenvironment-dependent spectral
Py-1, bound to plasma-chemically-generated surface properties of the surface-attached fluorophore. For that
amino groups of PP films, and its analogue, corrected purpose, labeled PP films were immersed for 15 min
for such instrument-specific effects. Accordingly, the in the solvents listed in TABLE 1, and the fluorescence
comparison of fluorescence spectra measured at dif- spectra of the nondried films were recorded. Measur-
ferent instruments or at different times requires the able, even though minor, solvent-dependent spectro-
removal of wavelength-dependent instrumental effects scopic properties, shown in FIGURE 3, reveal the re-
from the raw emission data by spectral correction sponse of the surface-attached dansyl chromophores
procedures.21,22 to changes in the polarity of the microenviron-
Spectral correction becomes particularly important ment.
if different polarizer settings are used. For example, With the exception of polarity-probing stud-
neglecting the polarization dependence of the spectral ies of specific fluorophore–solvent interactions, the
responsivity of the emission channel can even result in environment-dependent emission properties must be
an apparent polarization dependence of the measured taken into account. Thus, intensive washing, and in
spectra.22 particular drying, procedures are required to guaran-
tee reproducible and comparable fluorescence data.
Sensitivity of the Fluorophore’s Spectroscopic Despite careful sample preparation steps, the fluo-
Properties to Microenvironment Including rometric detection of surface functionalities on solid
Surface Attachment supports is often hampered by an enhanced back-
Almost all reactive dyes intended for the detection ground signal due to nonspecifically adsorbed dye
of functional groups at surfaces are xanthene dyes molecules for fluorophores lacking binding-induced
(e.g., fluoresceins, rhodamines). These chromophores spectroscopic changes. Hence, more suitable are so-
reveal a locally excited state–type emission, the spec- phisticated labels revealing changes in their absorp-
troscopic properties of which are less sensitive to the tion and fluorescence properties upon surface attach-
fluorophore’s microenvironment. Charge transfer dyes ment.6,8,27 This approach enables a clear distinction
such as dansyl chromophores, which are also com- between covalently linked and free, only adsorbed,
monly used for fluorophore labeling, however, reveal reporter molecules. Excellent examples for this design
Hoffmann et al.: Fluorescence Spectroscopy on Polymer Surfaces 31

FIGURE 3. Steady-state fluorescence emission spectra of a plasma-chemically-modified and dansyl-


labeled PP film impregnated with several solvents (left) and (right) shift of the emission maximum of the
film as a function of Dimroth and Reichardt’s polarity scale E T .23

of both the excitation and emission spectra,12 as illus-


trated in FIGURE 4.
Also, the formation of the PP–Py-1 conjugate is ac-
companied by a strong increase in emission intensity
(note the noisier signals in FIG. 4 of the normalized
emission spectra of the unmodified reference film).12
The drastic changes in energy and intensity of the
absorption and emission bands elegantly circumvent
the simultaneous excitation of nonspecifically adsorbed
dye molecules, thereby enhancing the signal-to-noise
ratio and reducing labor-intensive washing steps.
Reproducibility of Fluorescence Measurements
at Solid Supports
Despite improved fluorophore labeling and de-
FIGURE 4. Hypsochromic shifts of the excitation tection approaches, fluorescence studies on solid
(dashed lines) and the emission (solid lines) spectra of substrates are prone to reduced measurement repro-
Py-1 after its reaction with amino-terminated PP surfaces. ducibility compared with measurements of transpar-
The emission of the PP–Py-1 conjugate (symbols) was ex- ent dilute solutions taken using cuvettes. For the latter,
cited at 503 nm, and the fluorescence of the nonreacted the reproducibility is commonly better than 2%, as
PP-adsorbate (line) at 621 nm; the corresponding excitation
revealed in FIGURE 5A.21
spectra were recorded at 602 and 665 nm, respectively.
For measurements of solid films with a routine flu-
orometer, however, a significantly lower reproducibil-
concept are amino-sensitive pyrylium dyes, such as ity has been found.21 Use of traditional dansyl labels
Py-1.17,18 Because of the reaction of amino groups yields an uncertainty of 15% despite tedious wash-
in, for example, proteins, the Py-1 absorption and ing and sample preparation procedures (FIG. 5B).13
emission maxima shift hypsochromically and the fluo- This uncertainty arises, most likely, from errors due to
rescence quantum yield of the very weakly emissive sample positioning in front-face measurement geome-
pyrylium dye strongly increases.17,18 In our studies on tries and inhomogeneous functionalization/labeling,
amino-functionalized solid PP supports, the labeling as well as from sample-related parameters such as
reaction with Py-1 also induced significant blueshifts scattering and reflections and difficulties of eliminating
32 Annals of the New York Academy of Sciences

FIGURE 5. Reproducibility of fluorescence measurements in solution compared with those on solid


polymer supports. Standard deviations (error bars) of repeatedly recorded emission spectra (uncorrected,
gray solid lines) of an organic dye in solution (A), of dansyl-labeled PP (excitation at 350 nm; B), and of
an amino-modified and Py-1–derivatized PP film (excitation at 503 nm, C).

background fluorescence.22 To evaluate the effect of


the type of fluorescent label on the reproducibility of
emission measurements on polymer films, we anal-
ogously determined the standard deviation from the
variation of the signal intensities of repeated emission
records for amino-functionalized PP derivatized with
the chromogenic label Py-1. With this labeling strat-
egy, the standard deviation could be reduced to ap-
proximately 7%, as illustrated in FIGURE 5C. This im-
proved measurement uncertainty for Py-1–derivatized
films is most likely owing to the strong binding-induced
spectroscopic changes. For the excitation wavelength
of approximately 500 nm used, the surface emission is
always recorded against an almost dark background FIGURE 6. Monoexponentially fitted decrease in rela-
even if nonreacted, adsorbed fluorophores or dye tive integral fluorescence emission I t /I 0 of dansyl-labeled
molecules diffused into the polymer are still present, PP (-◦-) and an amino-modified PP film derivatized with Py-1
because the free pyrylium dye remains silent under (-•-) after subsequent Soxhlet extractions with ethanol.
these conditions.12,17

Nonspecific Adsorption of Fluorophores loss of the initial fluorescence signal of approximately


Fluorescence measurements of functionalized poly- 50% for plasma-chemically-modified PP films reacted
mer supports or other polymeric and porous materials with dansyl labels (FIG. 6, open circles) as exemplary de-
typically require extensive washing and extraction pro- termined by Soxhlet extraction. For the chromogenic
cedures to remove nonreacted dye molecules, particu- and fluorogenic Py-1 fluorophore, however, we ex-
larly for conventional labels. The importance of such pected a constant signal independent of extraction
purification procedures follows immediately from the steps because here exclusively the reacted dye is spec-
Hoffmann et al.: Fluorescence Spectroscopy on Polymer Surfaces 33

troscopically addressed. Astonishingly, also for PP–Py- quantitative fluorescence measurements on solid sup-
1 conjugates, the integral emission decreased with the ports are still hampered by all the above-mentioned
number of extraction cycles (FIG. 6, solid circles),12 complicating factors, the derivatization with the chro-
although only amino-conjugated pyrylium molecules mogenic and fluorogenic label strongly improves the
contribute to the fluorescence signal. relative fluorometric characterization of surface amino
However, the much slower extraction kinetics indi- groups directly at solid supports.
cates a different reaction course for the degradation
of the Py-1 fluorescence compared with that of the
dansyl label. The measured effect suggests either the Acknowledgment
presence of an additional unspecific source of amino
functionalities aside from the polymer-attached amino Technical assistance by R. Decker and financial
groups, which is implausible, or certain instabilities of support from Federal Ministry of Economics and La-
the Py-1–reacted amino surface functionalities. The bor (BMWA) are gratefully acknowledged. We are in-
latter is possibly not related to fluorophore labeling debted to Prof. O. S. Wolfbeis and Actif Motiv GmbH
but rather to different routes for plasma-chemically- for the generous supply of Py-1.
induced surface functionalization. Dansyl labeling was
performed on OH functions at the polymer back-
bone, whereas Py-1 modification was carried out on Conflict of Interest
NH 2 functions of plasma-deposited ppAAm layers.
The authors declare no conflicts of interest.
Because low-molecular-weight polymer fragments are
not resistant to organic solvents,6 the decrease in rel-
ative emission to about 60% of the initial fluores-
cence intensity is ascribed to some chemical instabil- References
ity of the plasma-deposited ppAAm layers, where Py-
1–labeled low-molecular-weight oligomeric structures 1. CHU, P.K., J.Y. CHEN, L.P. WANG & N. HUANG. 2002.
are partially detached from the polymeric substrate. Plasma-surface modification of biomaterials. Mater. Sci.
Engineering, R 36: 142–206.
This assumption is supported by additional investiga-
2. KANG, J., L. DING, F. LÜ, et al. 2006. Dansyl-based fluores-
tions by means of surface-sensitive X-ray photoelectron cent film sensor for nitroaromatics in aqueous solution. J.
spectroscopy.12 Phys. D: Appl. Phys. 39: 5097–5102.
3. NITSCHKE, M., A. HOLLÄNDER, F. MEHDORN, et al. 1996.
Infrared spectroscopic characterization of plasma-treated
Conclusion polyethylene. J. Appl. Polym. Sci. 59: 119–124.
4. XING, Y. & E. BORGUET. 2007. Specificity and sensitivity
of fluorescence labeling of surface species. Langmuir 23:
Fluorometric analysis of functionalized surfaces 684–688.
labeled with organic chromophores is a promising, 5. NASHAT, A.H., M. MORONNE & M. FERRARI. 1998. De-
simple, and sensitive technique for monitoring plasma- tection of functional groups and antibodies on microfabri-
chemically-introduced OH and NH 2 groups at PP sup- cated surfaces by confocal microscopy. Biotechnol. Bioeng.
ports. The reliability of the data requires proper con- 60: 137–146.
sideration of various factors that influence fluorescence 6. HOLLÄNDER, A. 2004. Labelling techniques for the chemical
analysis of polymer surfaces. Surf. Interf. Anal. 36: 1023–
measurements. Here, the choice of an appropriate
1026.
fluorescent label also plays a significant role. The cor- 7. OEHR, C. 2003. Plasma surface modification of polymers
rection of the emission spectra for the spectral charac- for biomedical use. Nucl. Instr. And Meth. in Phys. Res. B
teristics of the instrument is essential to obtaining com- 208: 40–47.
parable, instrument-independent results.22 To improve 8. IVANOV, V.B., J. BEHNISCH, A. HOLLÄNDER, et al. 1996.
the reproducibility of fluorescence data, one must ac- Determination of functional groups on polymer sur-
count for the sensitivity of the fluorophore’s spectro- faces using fluorescence labeling. Surf. Interf. Anal. 24:
257–262.
scopic properties to the microenvironment, and careful
9. HENNEUSE-BOXUS, C., A. DE RO, P. BERTRAND & J.
sample preparation, washing, and particularly drying MARCHAND-BRYAERT. 2000. Covalent attachment of fluo-
procedures become mandatory. These procedures are rescence probes on the PEEK–OH film surface. Polymer
especially important for labeling protocols that use flu- 41: 2339–2348.
orescent reporters that do not reveal binding-induced 10. HOLMES-FARLEY, S.R. & G.M. WHITESIDES. 1986. Fluores-
spectral properties, where nonreacted dye molecules cence properties of dansyl groups covalently bonded to the
also contribute to the fluorescence signal. Even though surface of oxidatively functionalized low-density polyethy-
lene film. Langmuir 2: 266–281.
34 Annals of the New York Academy of Sciences

11. WANG, L., A.K. GAIGALAS & Y. REIPA. 2005. Optical prop- 19. Fluorescence spectra of labeled polymers were recorded on
erties of AlexaTM 488 and CyTM 5 immobilized on a glass air-dried films between two quartz windows in front-face
surface. BioTechniques 38: 127–138. geometry with a Spectronics Instruments 8100 spectroflu-
12. HOFFMANN, K., R. MIX, U. RESCH-GENGER & J.F. orometer.
FRIEDRICH. 2007. Monitoring of amino functionalities on 20. MIRENDA, M., M.G. LAGORIO & E. SAN ROMAN. 2004. Pho-
plasma-chemically modified polypropylene supports with tophysics on surfaces: determination of absolute fluores-
a chromogenic and fluorogenic pyrylium reporter. Lang- cence quantum yields from reflectance spectra. Langmuir
muir 23: 8411–8416. 20: 3690–3697.
13. HOFFMANN, K., U. RESCH-GENGER, R. MIX & J.F. 21. RESCH-GENGER, U., D. PFEIFER, C. MONTE, et al. 2005.
FRIEDRICH. 2006. Fluorescence spectroscopic studies Traceability in fluorometry: Part II. Spectral fluorescence
on plasma-chemically modified polymer surfaces with standards. J. Fluoresc. 15: 325–346.
fluorophore-labeled functionalities. J. Fluoresc. 16: 441– 22. RESCH-GENGER, U., K. HOFFMANN, W. NIETFELD, et al.
448. 2005. How to improve quality assurance in fluorometry:
14. ZHANG, C., N. LUO & D.E HIRT. 2005. Penetration be- fluorescence-inherent sources of error and suited fluores-
havior and subsurface grafting of dansyl cadaverine and cence standards. J. Fluoresc. 15: 337–362.
polyethylene glycol (PEG) derivatives in poly(ethylene-co- 23. LAKOWICZ, C.J.R. 2006. Principles of Fluorescence Spec-
acrylic acid) (EAA) film. Polymer 46: 9257–9264. troscopy, 3rd ed. Springer Science and Business. New York.
15. MIX, R., K. HOFFMANN, U. RESCH-GENGER, et al. 2007. 24. HOFFMANN, D.A., J.E. ANDERSON & C.W. FRANK. 1995.
Covalent coupling of fluorophores to polymer surface- Solvent partitioning in fluorescent polydimethylsiloxane
bonded functional groups. In Polymer Surface Modifica- networks. J. Mater. Chem. 5: 13–25.
tion, Vol. 4. K.L. MITTAL, Ed.: 171–191. 25. GONZALES-BENITO, J., A. AZNAR & J. BASELGA. 2001. Sol-
16. SHCHEPINOV, M.S., S.C. CASE-GREEN & E.M. SOUTHERN. vent and temperature effects on polymer-coated glass
1997. Steric factors influencing hybridisation of nucleic fibers. Fluorescence of the dansyl moiety. J. Fluoresc. 11:
acids to oligonucleotide arrays. Nucleic Acids Res. 25: 307–314.
1155–1161. 26. TANAKA, M., H. TANAKA & R. PELTON. 2007. Amine-
17. WETZL, B.K., S.M. YARMOLOUK, D.B. CRAIG & O.S. WOLF- derivatized poly(diallyldimethylammonium chloride) from
BEIS. 2004. Chameleon labels for staining and quantifying N -vinylformamide copolymerization. J. Appl. Polym. Sci.
proteins. Angew. Chem. Int. Ed. Engl. 43: 5400–5402. 104: 1068–1075.
18. HOEFELSCHWEIGER, B.K., A. DUERKOP & O.S. WOLFBEIS. 27. UDENFRIEND, S., S. STEIN, P. BÖHLEN, et al. 1972.
2005. Novel type of general protein assay using a chro- Fluorescamine—reagent for assay of amino-acids, pep-
mogenic and fluorogenic amine-reactive probe. Anal. tides, proteins, and primary amines in picomole range.
Biochem. 344: 122–129. Science 178: 871–872.

You might also like