Professional Documents
Culture Documents
Polymer Surfaces—Problems
and Troubleshooting
KATRIN HOFFMANN, RENATE MIX, UTE RESCH-GENGER, AND JOERG F. FRIEDRICH
Federal Institute for Materials Research and Testing (BAM), Div. I.5, Berlin, Germany
Key words: fluorescence spectroscopy; polymer functionalization; surface labeling; pyrylium label
FIGURE 1. Labeling reactions. (A) Labeling of OH groups at a PP surface with a dansyl fluorophore
by using a diisocyanate spacer. (B) Labeling of surface amino groups with the blue Py-1 chromophore,
yielding a red PP conjugate. (In color in Annals online.)
strategy, NH 2 -terminated PP surfaces have been gen- sensitive technique.6 Second, further complications
erated.12,15 can arise from background emission either from the
polymer support itself and/or from adsorbed (non-
Labeling of Surface Functionalities linked) dye molecules. Furthermore, the environmen-
Different synthetic concepts have been tested for tal dependence of the spectroscopic properties of most
fluorophore labeling of surface functionalities,13,15 dyes and matrix–dye or dye–dye interactions result-
as well as to introduce spacer molecules.16 Con- ing in fluorescence quenching must be taken into ac-
ventional isothiocyanate labels or amino-terminated count.20 All these problems raise qualms about the
labels like dansyl fluorophores were covalently at- reliability of quantification relying on fluorescence
tached by means of spacer molecules to OH func- measurements after labeling reactions. Despite these
tions at the polymer surface by strategies adopted from, limitations, the unique sensitivity of fluorometry with
for example, well-investigated polyurethane chemistry a limit of detection on the order of approximately
(FIG. 1A).13 For the fluorometric characterization of 10−12 mol/cm2 for fluorophores on polymer surfaces,8
amino-functionalized PP surfaces, the chromogenic in conjunction with its comparative ease of use, renders
and fluorogenic pyrylium dye Py-1 has been used, fluorescence techniques as attractive tools for analyz-
which was recently introduced for the derivatization ing organically modified polymer surfaces, especially
of primary amino groups (FIG. 1B).17,18 For each series when only monitoring of changes in surface group con-
of fluorescence measurements,19 blank or so-called ref- centration is needed.
erence samples were prepared by reaction of the non-
functionalized polymer film with the respective flu- Effect of Spectral Correction:
orescent label by using the same procedure as that Comparability of Data
used to covalently attach fluorophores to the surface- In general and independent of the type of fluores-
functionalized films. cent samples (e.g., fluorophores in solution or fluores-
cent labels attached to polymer surfaces), the fluores-
Problems Inherent to Fluorescence cence signal is affected by both the fluorescent analyte
Measurements at Surfaces and instrumental effects. Relevant properties of ana-
Fluorescence techniques in conjunction with fluo- lytes are the fluorophore’s absorptance at the excita-
rescent labels can be exploited to characterize sur- tion wavelength and its fluorescence quantum yield.
face functionalities at low concentration levels in the These quantities, as well as the spectral shape and po-
range between approximately 109 (on glass4 ) and 1012 sition of the fluorophore’s absorption, excitation, and
(on polymer supports8 ) mol/cm2 . The reliable use of emission spectra, can strongly depend on the polarity
these techniques, however, requires overcoming some of the dye’s microenvironment, that is, on the solvent
challenges.7 First, the observed fluorescence signal re- or the matrix (e.g., polarity, proticity, pH, and viscos-
sults from the bulk material because commonly used ity). This dependence complicates quantification from
steady-state fluorescence spectroscopy is not a surface- measurements of relative fluorescence intensities. From
30 Annals of the New York Academy of Sciences
troscopically addressed. Astonishingly, also for PP–Py- quantitative fluorescence measurements on solid sup-
1 conjugates, the integral emission decreased with the ports are still hampered by all the above-mentioned
number of extraction cycles (FIG. 6, solid circles),12 complicating factors, the derivatization with the chro-
although only amino-conjugated pyrylium molecules mogenic and fluorogenic label strongly improves the
contribute to the fluorescence signal. relative fluorometric characterization of surface amino
However, the much slower extraction kinetics indi- groups directly at solid supports.
cates a different reaction course for the degradation
of the Py-1 fluorescence compared with that of the
dansyl label. The measured effect suggests either the Acknowledgment
presence of an additional unspecific source of amino
functionalities aside from the polymer-attached amino Technical assistance by R. Decker and financial
groups, which is implausible, or certain instabilities of support from Federal Ministry of Economics and La-
the Py-1–reacted amino surface functionalities. The bor (BMWA) are gratefully acknowledged. We are in-
latter is possibly not related to fluorophore labeling debted to Prof. O. S. Wolfbeis and Actif Motiv GmbH
but rather to different routes for plasma-chemically- for the generous supply of Py-1.
induced surface functionalization. Dansyl labeling was
performed on OH functions at the polymer back-
bone, whereas Py-1 modification was carried out on Conflict of Interest
NH 2 functions of plasma-deposited ppAAm layers.
The authors declare no conflicts of interest.
Because low-molecular-weight polymer fragments are
not resistant to organic solvents,6 the decrease in rel-
ative emission to about 60% of the initial fluores-
cence intensity is ascribed to some chemical instabil- References
ity of the plasma-deposited ppAAm layers, where Py-
1–labeled low-molecular-weight oligomeric structures 1. CHU, P.K., J.Y. CHEN, L.P. WANG & N. HUANG. 2002.
are partially detached from the polymeric substrate. Plasma-surface modification of biomaterials. Mater. Sci.
Engineering, R 36: 142–206.
This assumption is supported by additional investiga-
2. KANG, J., L. DING, F. LÜ, et al. 2006. Dansyl-based fluores-
tions by means of surface-sensitive X-ray photoelectron cent film sensor for nitroaromatics in aqueous solution. J.
spectroscopy.12 Phys. D: Appl. Phys. 39: 5097–5102.
3. NITSCHKE, M., A. HOLLÄNDER, F. MEHDORN, et al. 1996.
Infrared spectroscopic characterization of plasma-treated
Conclusion polyethylene. J. Appl. Polym. Sci. 59: 119–124.
4. XING, Y. & E. BORGUET. 2007. Specificity and sensitivity
of fluorescence labeling of surface species. Langmuir 23:
Fluorometric analysis of functionalized surfaces 684–688.
labeled with organic chromophores is a promising, 5. NASHAT, A.H., M. MORONNE & M. FERRARI. 1998. De-
simple, and sensitive technique for monitoring plasma- tection of functional groups and antibodies on microfabri-
chemically-introduced OH and NH 2 groups at PP sup- cated surfaces by confocal microscopy. Biotechnol. Bioeng.
ports. The reliability of the data requires proper con- 60: 137–146.
sideration of various factors that influence fluorescence 6. HOLLÄNDER, A. 2004. Labelling techniques for the chemical
analysis of polymer surfaces. Surf. Interf. Anal. 36: 1023–
measurements. Here, the choice of an appropriate
1026.
fluorescent label also plays a significant role. The cor- 7. OEHR, C. 2003. Plasma surface modification of polymers
rection of the emission spectra for the spectral charac- for biomedical use. Nucl. Instr. And Meth. in Phys. Res. B
teristics of the instrument is essential to obtaining com- 208: 40–47.
parable, instrument-independent results.22 To improve 8. IVANOV, V.B., J. BEHNISCH, A. HOLLÄNDER, et al. 1996.
the reproducibility of fluorescence data, one must ac- Determination of functional groups on polymer sur-
count for the sensitivity of the fluorophore’s spectro- faces using fluorescence labeling. Surf. Interf. Anal. 24:
257–262.
scopic properties to the microenvironment, and careful
9. HENNEUSE-BOXUS, C., A. DE RO, P. BERTRAND & J.
sample preparation, washing, and particularly drying MARCHAND-BRYAERT. 2000. Covalent attachment of fluo-
procedures become mandatory. These procedures are rescence probes on the PEEK–OH film surface. Polymer
especially important for labeling protocols that use flu- 41: 2339–2348.
orescent reporters that do not reveal binding-induced 10. HOLMES-FARLEY, S.R. & G.M. WHITESIDES. 1986. Fluores-
spectral properties, where nonreacted dye molecules cence properties of dansyl groups covalently bonded to the
also contribute to the fluorescence signal. Even though surface of oxidatively functionalized low-density polyethy-
lene film. Langmuir 2: 266–281.
34 Annals of the New York Academy of Sciences
11. WANG, L., A.K. GAIGALAS & Y. REIPA. 2005. Optical prop- 19. Fluorescence spectra of labeled polymers were recorded on
erties of AlexaTM 488 and CyTM 5 immobilized on a glass air-dried films between two quartz windows in front-face
surface. BioTechniques 38: 127–138. geometry with a Spectronics Instruments 8100 spectroflu-
12. HOFFMANN, K., R. MIX, U. RESCH-GENGER & J.F. orometer.
FRIEDRICH. 2007. Monitoring of amino functionalities on 20. MIRENDA, M., M.G. LAGORIO & E. SAN ROMAN. 2004. Pho-
plasma-chemically modified polypropylene supports with tophysics on surfaces: determination of absolute fluores-
a chromogenic and fluorogenic pyrylium reporter. Lang- cence quantum yields from reflectance spectra. Langmuir
muir 23: 8411–8416. 20: 3690–3697.
13. HOFFMANN, K., U. RESCH-GENGER, R. MIX & J.F. 21. RESCH-GENGER, U., D. PFEIFER, C. MONTE, et al. 2005.
FRIEDRICH. 2006. Fluorescence spectroscopic studies Traceability in fluorometry: Part II. Spectral fluorescence
on plasma-chemically modified polymer surfaces with standards. J. Fluoresc. 15: 325–346.
fluorophore-labeled functionalities. J. Fluoresc. 16: 441– 22. RESCH-GENGER, U., K. HOFFMANN, W. NIETFELD, et al.
448. 2005. How to improve quality assurance in fluorometry:
14. ZHANG, C., N. LUO & D.E HIRT. 2005. Penetration be- fluorescence-inherent sources of error and suited fluores-
havior and subsurface grafting of dansyl cadaverine and cence standards. J. Fluoresc. 15: 337–362.
polyethylene glycol (PEG) derivatives in poly(ethylene-co- 23. LAKOWICZ, C.J.R. 2006. Principles of Fluorescence Spec-
acrylic acid) (EAA) film. Polymer 46: 9257–9264. troscopy, 3rd ed. Springer Science and Business. New York.
15. MIX, R., K. HOFFMANN, U. RESCH-GENGER, et al. 2007. 24. HOFFMANN, D.A., J.E. ANDERSON & C.W. FRANK. 1995.
Covalent coupling of fluorophores to polymer surface- Solvent partitioning in fluorescent polydimethylsiloxane
bonded functional groups. In Polymer Surface Modifica- networks. J. Mater. Chem. 5: 13–25.
tion, Vol. 4. K.L. MITTAL, Ed.: 171–191. 25. GONZALES-BENITO, J., A. AZNAR & J. BASELGA. 2001. Sol-
16. SHCHEPINOV, M.S., S.C. CASE-GREEN & E.M. SOUTHERN. vent and temperature effects on polymer-coated glass
1997. Steric factors influencing hybridisation of nucleic fibers. Fluorescence of the dansyl moiety. J. Fluoresc. 11:
acids to oligonucleotide arrays. Nucleic Acids Res. 25: 307–314.
1155–1161. 26. TANAKA, M., H. TANAKA & R. PELTON. 2007. Amine-
17. WETZL, B.K., S.M. YARMOLOUK, D.B. CRAIG & O.S. WOLF- derivatized poly(diallyldimethylammonium chloride) from
BEIS. 2004. Chameleon labels for staining and quantifying N -vinylformamide copolymerization. J. Appl. Polym. Sci.
proteins. Angew. Chem. Int. Ed. Engl. 43: 5400–5402. 104: 1068–1075.
18. HOEFELSCHWEIGER, B.K., A. DUERKOP & O.S. WOLFBEIS. 27. UDENFRIEND, S., S. STEIN, P. BÖHLEN, et al. 1972.
2005. Novel type of general protein assay using a chro- Fluorescamine—reagent for assay of amino-acids, pep-
mogenic and fluorogenic amine-reactive probe. Anal. tides, proteins, and primary amines in picomole range.
Biochem. 344: 122–129. Science 178: 871–872.