Software

MB17 - MB22
ENG - REV. 1
SYMBOLS
The following symbols are used to emphasize important safety and service information:

WARNING – Identifies danger points or conditions and the correct procedures that
must be followed to avoid injury.

CAUTION – Identifies actions that could damage the device/component.

HOT SURFACES – Hot items, take care to avoid burns.

COMPONENTS SENSITIVE TO ELECTROSTATIC CHARGES


Components on electronic boards are sensitive to electrostatic charges conducted by
the human body. Take precautions before handling sensitive components and boards
to avoid malfunction.

CAUTION – Don’t use any glue.

Do not apply any lubricant. Use water instead.

Moisten with water only.

NOTE Provides important additional information.

SAFETY ADVICE

Service should be performed by qualified technicians only. Observe all local and national
directives related to service and safety standards.

Before performing any repairing or service interventions on Lina MB sterilizers, ALWAYS


DISCONNECT the electric power supply cable to avoid injury.
If for any reason you are unable to remove the power supply or if you have to walk away
from the unit, always place a “WORK IN PROGRESS” sign, to warn other people that might
walk up to the sterilizer.

2-b LINA MB - Software - Rev. 1


PRODUCT HISTORY AND SOFTWARE UPGRADE
SYMBOLS ....................................................................................................................................... b

SAFETY ADVICE ............................................................................................................................ b

PRODUCT HISTORY AND SOFTWARE UPGRADE ........................................................................c


SUMMARY OF SOFTWARE RELEASES SINCE PRODUCT LAUNCH ........................................... 1
AVAILABLE LANGUAGES .............................................................................................................. 2
HOW TO VERIFY THE SOFTWARE VERSION ON THE STERILIZER............................................ 3
SOFTWARE UPDATE ..................................................................................................................... 3
SOFTWARE INSTALLATION .......................................................................................................... 3
PROCEDURE FOR SOFTWARE UPDATE / INSTALLATION .......................................................... 4
REPROGRAMMING AFTER UPDATE / INSTALL ........................................................................... 8
SELECTION OF THE HEATER AND OF THE DOOR LOCKING SYSTEM .................................... 10
a) Heater ................................................................................................................................ 10
b) Door locking system ........................................................................................................... 10
MENU LEVELS ............................................................................................................................. 11

LINA MB - Software - Rev. 1 c


SUMMARY OF SOFTWARE RELEASES SINCE PRODUCT LAUNCH

Descripti
Software

from S/N

TechInfo
Number
Starting
Release
version

name

on
1 001.06 102698 - Product launch
106103 - Modified water injection pattern during PP phases (LINA
2 001.07
(plus 106178) MB17 only)
- Danish translation improvement.
- In case of a flat CPU-board battery the sterilizer displays
the “set date and time” screen during power on (instead of
3 001.08 106170 the alarm E040).
- Bug fixed: The firmware would not start when the CPU-
board battery was flat (firmware upgrade possible even
with flat battery).
- Fixed possible heating element overvoltage during
4 001.09 106175
component test mode.
- In the LABEL PRINTER menu added the “offset X” and
“offset Y” options for label centering purpose.
- Drying time set resolution changed to 15 s.
- Sleep mode (display turned off and related sound) man-
aged also during all the waiting status; ”open the door”
waiting, cycle/alarm final message, loop/delayed cycle
waiting
- (service only):
- component test: bug fixed on door locking
5 001.10 107351
- component test: EV6, EV7 added
- component test: “messages” menu added
- (communication protocol) V00 FVS=4, value ID 0x09 =
cycle identifier (only), ID 0x26 = cycle name identifier
- (CB4 only) fixed bug on blackout management: in the
previous revision the cycle name loose in case of black
out alarm code E010 (converted to “no cycle”).
- fixed bug: random false “File save error “ provided if
PC/logger warning activated
6 001.11 107453 - Bug fixed: false E182
- added CZ with specific configuration
7 001.12 107640
- added Romanian language
- New installer (single program)
8 001.13 107932 - Turkish messages updated
- MS version available
9 001.14 108066 - Bug fixed: false
10 001.15 108182 - added Russian language
11 001.16 108198 - bux fixed: Expiry time on labels
12 001.17 108753 - added Albanian, Hungarian, Greek languages
001.18 - added Croatian language

LINA MB - Software - Rev. 1 1


AVAILABLE LANGUAGES
Lina MB sterilizers may be provided with different language packs depending on the area of destination.
Languages highlighted with an asterisk (*) can be displayed only if the user interface S4006230 is fitted.

Starting from Language pack


Language ID
release std
English 1 001.06 X
Italian 2 001.06 X
Spanish 3 001.06 X
French 4 001.06 X
German 5 001.06 X
Polish 6 001.06 X
Dutch 7 001.06 X
Swedish 8 001.06 X
Finnish 9 001.06 X
Danish 10 001.06 X
Portuguese 11 001.06 X
Norwegian 12 001.06 X
Turkish 14 001.10 X
Czech 17 001.10 X
Bulgarian 15 001.11 X(*)
Romanian 23 001.12 X
Russian 16 001.15 X(*)
Albanian 24 001.17 X
Greek 19 001.17 X (*)
Croatian 21 001.18 X

2 LINA MB - Software - Rev. 1


HOW TO VERIFY THE SOFTWARE VERSION ON THE STERILIZER

Use the UP, DOWN and OK buttons to browse


the menu, choosing the following options in
sequence: MENU – DEVICE INFO

Scroll the parameters by pressing DOWN until


the software version is displayed.

ATTENTION!

It is strongly recommended to update all sterilizers in the marketplace with the latest software
version.

This will provide users with the opportunity to enjoy the full benefit of new developments and
improvements virtually free of charge. Please make use of this opportunity whenever you visit a
doctor’s office, during regular service and maintenance calls, and whenever you have a sterilizer in
your shop for repair.

SOFTWARE UPDATE
UPDATE is the standard procedure to provide the sterilizer with the latest software version.
This procedure should be followed whenever sterilizers are serviced, when a more recent software version is
available.

This procedure keeps all the most important settings, data and parameters stored in the sterilizer memory
but could alter other minor settings. For more details please refer to paragraph “REPROGRAMMING AFTER
UPDATE/INSTALL”.

ATTENTION!
The software UPDATE procedure must NOT be used if the CPU-board was replaced
during the service. If a new CPU-bard was placed, use the INSTALL procedure
instead!

SOFTWARE INSTALLATION
INSTALL is the special procedure that must be followed when the sterilizer CPU-board was replaced
during service.
The INSTALL procedure will complwetely erase and re-program the CPU-board memory.

LINA MB - Software - Rev. 1 3


PROCEDURE FOR SOFTWARE UPDATE / INSTALLATION
The procedure requires:
• a PC with a free serial port (if the PC doesn’have a serial port, you will need a USB-Serial
converter: W&H provides such converter, ref. n. A7010040);
• a NULL-MODEM DB9 Female-Female serial cable (ref. A7010030);
• the latest version of the installation / update program. This is usually sent in a ZIPped folder.
Unzip it before starting the procedure.

ATTENTION!
Both the INSTALL and the UPDATE procedures will erase the sterilizer cycle
history. Take care to print or store the cycle reports on the USB memory (using the
Multiport data logger) if required by the user.

STEP DESCRIPTION

1 Switch OFF the sterilizer.

Connect the sterilizer to the serial port of the PC using the


NULL-MODEM DB9 Female-Female serial cable.

Hold the “OK” and the “BACK” buttons pressed


simultaneously and at the same time switch ON the
sterilizer.
3
Release the buttons: a “SERVICE MODE” message
appears on the display.
Launch the WHInstaller.bat file on the PC
4
NOTE: DO NOT LAUNCH the EXE file as you will not obtain the expected result.
If you get an alarm message like “Unable to find the runtime to run this application”, launch (by
5
double - clicking) the vcredist_x86.exe file, install the runtime and then repeat from step 1.

4 LINA MB - Software - Rev. 1


STEP DESCRIPTION

A dialog window will


appear
6
(see image on side)

If you want to update the sterilizer with a new software release, select UPDATE and then jump to
7
step 14 (at the end of the grey block)
If a new PCB has been installed and must be completely reprogrammed, you have to activate the
INSTALL procedure

NOTE: INSTALL erases completely the parameters, data and history stored in the memory of the
8 main board and reprograms it in a “as new” state. This procedure must be used if a new PCB has
been installed, or if you install a used PCB having in memory parameters, data and history that do
not belong to the sterilizer you are about to install on.
If you want to preserve the data stored in memory and just update the software version, use
UPDATE.

Before launching INSTALL,


you must make the option
available.
9
Select:
FILE
ENTER SPECIAL CODE

LINA MB - Software - Rev. 1 5


STEP DESCRIPTION

Enter the code 200100 into


10
the box

Now the option INSTALL is


available: select INSTALL

11
The MODEL, SUB-MODEL and
SERIAL NUMBER selections are
now accessible

Select the model and the sub-model of the sterilizer.


12
Enter the serial number of the sterilizer in the appropriate field.

If the OTHER SETTINGS fields


become accessible, this means
that the serial number or model
you entered has multiple
possibilities for the heater and for
the door locking system. Check
the paragraph “SELECTION OF
13
THE HEATER AND OF THE
DOOR LOCKING SYSTEM” in
the further pages to individuate
the type of components that are
actually installed in the sterilizer,
and choose the appropriate
options in the selection pop-ups.

Select the serial port you have used for the connection between the sterilizer and the PC. If you are
using a USB-serial converter, please remember the following:
- always install the converter with its appropriate drivers;
14 - restart the computer before using the serial communication
- after restarting, the port number could be not the same as when installed;
- the communication line must be installed BEFORE launching the WHInstaller.bat file
- generally (but not always) the port to be used it is the last available in the pop-up list.
The baud rate is the speed of the data communication. The pre-set value is 57600 bytes per
15 second and is supported by most PCs. Higher speeds may be selected, if supported. If the
communication doesn’t take place or if some errors occur, reduce the speed until it works.
16 Click on CONTINUE

6 LINA MB - Software - Rev. 1


STEP DESCRIPTION
A wait message will be shown on the display of the sterilizer (if the SERVICE MODE message
persists, this means that there is no communication).
17 During the procedure, single-step progress bars and a general progress bar are continuously
updated.
After the procedure is completed, a message “UPDATE DONE !” appears.
18 Switch OFF the sterilizer.
Only if you are running the INSTALL procedure, switch ON the sterilizer again and access the
menu to introduce the following information:
• Cycle counter
19 • Test cycle counter
• Maintenance counters for door seal, dust filter, bacteriological filter
• Maintenance counters for 4000 and 20000 cycles servicing
• All “Activation codes” previously introduced to enable extra functions.
For both the UPDATE and for the INSTALL procedure, access the menu and enter the parameters
20
listed in the table of the paragraph “Reprogramming after UPDATE / INSTALL”.

LINA MB - Software - Rev. 1 7


REPROGRAMMING AFTER UPDATE / INSTALL

The following parameters (mostly related to optional accessories or custom settings) will be lost when run-
ning the INSTALL procedure, but might be lost or altered even when performing an UPDATE.
Whichever procedure you have followed, check and restore the parameters according to the original configu-
ration / setup of the sterilizer.

Some of these parameters could be automatically configured as a part of the


NOTE
country default settings.

PARAMETER When to set Setting


Cycle - setup
All cycle custom settings Consult user or check an old
If different from the default drying time
(e.g., the drying time) cycle report if available.
Setup
Language If different from the default User setting
Date and time Could get lost if the battery was flat Current date and time
Consult user or check an old
User Name Always
cycle report if available.
Sleep mode If different from the default User setting
If an accessory is connected to the According to the connected
Serial port
serial port accessory
If a report printer is connected (either Set model and baud rate
Printer settings directly or through the Multiport data compatible with the installed
logger) printer
Preheating If the user wants to disable it Never
If a warning is required on chamber
Hot surface warning Yes
temperatures above 70°C
If the Multiport data logger is con-
PC/Logger warning nected and equipped with the USB Yes
pen drive
Cycle reports – autom.
If a report printer is connected User choice
printing
Cycle reports – HTM2010
If a report printer is connected User choice
option
Label Printer – model If a label printer is connected Printer model
Cycle labels If a label printer is connected User setting
The last lot number used (ask
the user; in case of doubt set
If a label printer is connected and lot
Lot labels – set counter a value reasonably higher
labels are used
than the last used in order to
avoid duplicates)
If a label printer is connected and lot
Lot labels – label field User setting
labels are used
Configuration
If sterilizer is connected to automated
Aut. Water supply Yes
water supply from a demineralizer.

8 LINA MB - Software - Rev. 1


PARAMETER When to set Setting
If sterilizer is connected to automated
Ext. water sensor water supply from a demineralizer that No/Yes
checks the water quality
Always keep the internal water quality
Int. water sensor Yes
sensor active.
If more than one temperature sensor According to the number of
Process sensor
is installed sensors installed
Service
All released activation codes

If one or more activation codes were Excluding activation codes


Activation code released to the final user upon spe- released specifically to fix
cific request problems related to a CPU-
board that has just been re-
placed
Device info
Performed cycles
Aborted cycles
Set to the same values as be-
Interrupt. cycles
Always fore the installation/updating
Test cycles
procedure.
H2O non conform
H2O unacceptable

NEVER CHANGE THE SERIAL NUMBER OF THE STERILIZER!

Changing the sterilizer serial number AFTER the installation procedure or at any
time during the product life will erase the cycle hystory and will restore the device
default settings (any special functionalities enabled with Activation Codes will be
lost).

LINA MB - Software - Rev. 1 9


SELECTION OF THE HEATER AND OF THE DOOR LOCKING SYSTEM

a) Heater
You must check the electric supply and the number of electric wires of the heater, then refer to the table be-
low.
The ohmic value is also shown in the table, but an ohmic test is not necessary.

Electric Number of Ohmic value


Selection
supply wires (nominal, cold conditions)

110-125VAC - 7.92 Ώ OP7 - 1600W - 100 - 127V

2 26.5 Ώ OP8 - 2000W - 220-240V

3 39 Ώ / 79 Ώ OP2 - 1333W + 666W - 220-240V


200-240VAC
38 Ώ / 76 Ώ / 76 Ώ
4 OP11 - 2x850W - - 220-240V
38 Ώ / 38 Ώ

To measure the ohmic values, the heating element must be disconnected and isolated from the sterilizer in-
ternal wiring, otherwise the values could be altered.
Measure the ohmic value between the common wire (the wire coming from the thermal overload thermostat)
and the other wires.

b) Door locking system


Look at the construction method of the body of the door locking system. Check also the part number label , if
present on the part.

Part number
Construction of the body Selection
label

No label
Stainless steel plates, or S5200020 OP9 - S520006X
assembled by welding or S5200060 Welded Stainless Steel
or G0090020

Aluminium casting S5200070 OP12 - S520007X - Aluminum Casting

10 LINA MB - Software - Rev. 1


MENU LEVELS

MENU ITEM SUB-MENU ITEMS


Messages
Select
View
Cycle history Print
Print labels
Send HTML
View alarms
Alarm history
Print all
Language
Date and time
User name
Sleep mode
Volume setting
Display contrast
Serial port
Printer settings
Setup Preheating
Hot surface warning
PC/Logger warning
Maint. 4000 warning
Units
Cycle reports
Label printer
Cycle labels
Menu Lot labels
Automatic water supply
External water sensor
Internal water sensor
Configuration
Process sensor
Air detector
Liquid sensor
Current levels
Activation code
Service Set defaults
Set parameters
Component test
Print
Print lot labels
Reprint
Brand
Model
Serial number
Performed cycle
Cycle counters
Device info Service counters
Software rev.
Boot version
Power version
HW key ID
Multiport version

LEGENDA
Colour Min. access level Password
User None
Advanced user 000123
Service 200100
Hardware key None

LINA MB - Software - Rev. 1 11


W&H Sterilization S.r.l Lina MB
Italy, I-24060 Brusaporto (BG), Via Bolgara, 2 Software ENG Rev. 1

t +39/035/66 63 000 f +39/035/50 96 988 Subject to alterations


wh.com