MB17 - MB22
ENG - REV. 1
SYMBOLS
The following symbols are used to emphasize important safety and service information:
WARNING – Identifies danger points or conditions and the correct procedures that
must be followed to avoid injury.
SAFETY ADVICE
Service should be performed by qualified technicians only. Observe all local and national
directives related to service and safety standards.
Descripti
Software
from S/N
TechInfo
Number
Starting
Release
version
name
on
1 001.06 102698 - Product launch
106103 - Modified water injection pattern during PP phases (LINA
2 001.07
(plus 106178) MB17 only)
- Danish translation improvement.
- In case of a flat CPU-board battery the sterilizer displays
the “set date and time” screen during power on (instead of
3 001.08 106170 the alarm E040).
- Bug fixed: The firmware would not start when the CPU-
board battery was flat (firmware upgrade possible even
with flat battery).
- Fixed possible heating element overvoltage during
4 001.09 106175
component test mode.
- In the LABEL PRINTER menu added the “offset X” and
“offset Y” options for label centering purpose.
- Drying time set resolution changed to 15 s.
- Sleep mode (display turned off and related sound) man-
aged also during all the waiting status; ”open the door”
waiting, cycle/alarm final message, loop/delayed cycle
waiting
- (service only):
- component test: bug fixed on door locking
5 001.10 107351
- component test: EV6, EV7 added
- component test: “messages” menu added
- (communication protocol) V00 FVS=4, value ID 0x09 =
cycle identifier (only), ID 0x26 = cycle name identifier
- (CB4 only) fixed bug on blackout management: in the
previous revision the cycle name loose in case of black
out alarm code E010 (converted to “no cycle”).
- fixed bug: random false “File save error “ provided if
PC/logger warning activated
6 001.11 107453 - Bug fixed: false E182
- added CZ with specific configuration
7 001.12 107640
- added Romanian language
- New installer (single program)
8 001.13 107932 - Turkish messages updated
- MS version available
9 001.14 108066 - Bug fixed: false
10 001.15 108182 - added Russian language
11 001.16 108198 - bux fixed: Expiry time on labels
12 001.17 108753 - added Albanian, Hungarian, Greek languages
001.18 - added Croatian language
ATTENTION!
It is strongly recommended to update all sterilizers in the marketplace with the latest software
version.
This will provide users with the opportunity to enjoy the full benefit of new developments and
improvements virtually free of charge. Please make use of this opportunity whenever you visit a
doctor’s office, during regular service and maintenance calls, and whenever you have a sterilizer in
your shop for repair.
SOFTWARE UPDATE
UPDATE is the standard procedure to provide the sterilizer with the latest software version.
This procedure should be followed whenever sterilizers are serviced, when a more recent software version is
available.
This procedure keeps all the most important settings, data and parameters stored in the sterilizer memory
but could alter other minor settings. For more details please refer to paragraph “REPROGRAMMING AFTER
UPDATE/INSTALL”.
ATTENTION!
The software UPDATE procedure must NOT be used if the CPU-board was replaced
during the service. If a new CPU-bard was placed, use the INSTALL procedure
instead!
SOFTWARE INSTALLATION
INSTALL is the special procedure that must be followed when the sterilizer CPU-board was replaced
during service.
The INSTALL procedure will complwetely erase and re-program the CPU-board memory.
ATTENTION!
Both the INSTALL and the UPDATE procedures will erase the sterilizer cycle
history. Take care to print or store the cycle reports on the USB memory (using the
Multiport data logger) if required by the user.
STEP DESCRIPTION
If you want to update the sterilizer with a new software release, select UPDATE and then jump to
7
step 14 (at the end of the grey block)
If a new PCB has been installed and must be completely reprogrammed, you have to activate the
INSTALL procedure
NOTE: INSTALL erases completely the parameters, data and history stored in the memory of the
8 main board and reprograms it in a “as new” state. This procedure must be used if a new PCB has
been installed, or if you install a used PCB having in memory parameters, data and history that do
not belong to the sterilizer you are about to install on.
If you want to preserve the data stored in memory and just update the software version, use
UPDATE.
11
The MODEL, SUB-MODEL and
SERIAL NUMBER selections are
now accessible
Select the serial port you have used for the connection between the sterilizer and the PC. If you are
using a USB-serial converter, please remember the following:
- always install the converter with its appropriate drivers;
14 - restart the computer before using the serial communication
- after restarting, the port number could be not the same as when installed;
- the communication line must be installed BEFORE launching the WHInstaller.bat file
- generally (but not always) the port to be used it is the last available in the pop-up list.
The baud rate is the speed of the data communication. The pre-set value is 57600 bytes per
15 second and is supported by most PCs. Higher speeds may be selected, if supported. If the
communication doesn’t take place or if some errors occur, reduce the speed until it works.
16 Click on CONTINUE
The following parameters (mostly related to optional accessories or custom settings) will be lost when run-
ning the INSTALL procedure, but might be lost or altered even when performing an UPDATE.
Whichever procedure you have followed, check and restore the parameters according to the original configu-
ration / setup of the sterilizer.
Changing the sterilizer serial number AFTER the installation procedure or at any
time during the product life will erase the cycle hystory and will restore the device
default settings (any special functionalities enabled with Activation Codes will be
lost).
a) Heater
You must check the electric supply and the number of electric wires of the heater, then refer to the table be-
low.
The ohmic value is also shown in the table, but an ohmic test is not necessary.
To measure the ohmic values, the heating element must be disconnected and isolated from the sterilizer in-
ternal wiring, otherwise the values could be altered.
Measure the ohmic value between the common wire (the wire coming from the thermal overload thermostat)
and the other wires.
Part number
Construction of the body Selection
label
No label
Stainless steel plates, or S5200020 OP9 - S520006X
assembled by welding or S5200060 Welded Stainless Steel
or G0090020
LEGENDA
Colour Min. access level Password
User None
Advanced user 000123
Service 200100
Hardware key None