Endoscopy series product Disinfection and sterilization V1.

High temperature
Low
and pressure / Ethylene
Equipment/instrument temperature Glutaraldehyde 备注
pressure steam oxide (ET0)
plasma
method
Camera control unit、 All instruments are
forbidden to be
light source、
prohibit prohibit prohibit prohibit sterilized, otherwise the
Insufflator、Fluid equipment will be
Irrigation Pump damaged
In order to extend the
HD3/EC3 Camera
prohibit prohibit prohibit prohibit service life, it is
head
strongly recommended
that the camera and
Not guide beam be sterile
fibre-optical apply apply prohibit
recommended and cover the sterile
sleeve for use
Apply to
Not Not
Hard endoscopes recommend Johnson /
recommended recommended
equipment
Pneumoperitoneum
tube for wom Not Not
recommend prohibit /
Insufflator heating /no recommended recommended
heating
infusion tube for wom Not Not
recommend prohibit /
Fluid Irrigation Pump recommended recommended

A silicone component on an instrument or mirror can not be sterilized with low temperature plasma;
Use of low temperature plasma sterilization of plastic or rubber components, easy to fade the surface.

Precautions for disinfection and sterilization:


1) Do not exceed the time specified in the disinfection and sterilization procedures;
2) Attention should be paid to the proportion of liquid disinfection (glutaraldehyde 2%);
3) The plasma cannot be sterilized by the silicone hose;
4) Endoscopes and instruments that are sterilized by pressure steam should be used after natural cooling;
5) When the pressure steam sterilization, the equipment shall not touch the inner wall of the disinfection container;
6) There is no endoscope labeled "autoclavable" that can not be sterilized by high temperature and high pressure;
7) The endoscope can not be changed frequently and sterilized so as to reduce the life span