Professional Documents
Culture Documents
This content has been downloaded from IOPscience. Please scroll down to see the full text.
(http://iopscience.iop.org/0957-0233/11/3/318)
View the table of contents for this issue, or go to the journal homepage for more
Download details:
IP Address: 153.19.58.31
This content was downloaded on 12/11/2015 at 20:14
An ultra-low-noise electrical-potential
probe for human-body scanning
R J Prance, A Debray, T D Clark, H Prance, M Nock, C J Harland
and A J Clippingdale
Physical Electronics Group, School of Engineering, University of Sussex, Brighton,
Sussex BN1 9QT, UK
E-mail: t.d.clark@sussex.ac.uk and r.j.prance@sussex.ac.uk
Received 2 November 1999, in final form and accepted for publication 13 January 2000
1. Introduction the sensor and the skin. Clearly, the preparation of an array of
in-contact BPSM sensors is time consuming. Furthermore,
In previous work [1–3] we described the use of a high-input- since it is difficult to keep the conducting-paste interfaces
impedance electrical potential probe to make non-contact entirely separate in an array of sensors, cross coupling
electrical measurements of the human body. We introduced between neighbouring sensors (through leakage currents)
this probe precisely to overcome limitations associated with is common. These problems are eliminated completely if
conventional electrocardiographic (ECG) sensors. There displacement-current electrical-potential probes are used. By
were two key features in its design. First, it operated on way of illustration, we have already published an example
displacement, not real charge, current. This eliminated of non-invasive electrical-potential-probe heart imaging,
the need to make direct electrical contact with the body. utilizing an array of 25 elements (plus a separate, independent
For comparison, conventional ECG sensors always need to array of eight elements when required) [1].
maintain this good electrical contact with the body. In its
In this earlier work [1–4] we were able to engineer
practical embodiment [1, 2] the probe was positioned off the
very substantial improvements in the input impedance and
body at a distance of 3 mm, with a body-to-probe capacitance
noise performance of an electrometer-grade operational
of 7 pF. Second, the high input impedance of the probe (a few
amplifier (an OPA 111) by using this in a stable buffer-
times 1012 at 1 Hz) ensured that the surface of the body, as
amplifier configuration with precisely unity gain. The
the source, was not loaded down.
Surface potentials due to heart currents are spatio- basis of these improvements was a combination of well-
temporal in nature. In conventional ECG work a single sensor known guarding techniques and novel feedback circuits.
(plus one or more reference electrodes) is used to record In particular, we were able to make the effective (i.e.
temporal information which changes with its location on the electronically manipulated) input impedance as high as
surface of the body. Clearly, in this electrical technique the 1016 [3], corresponding to an input capacitance close to
functioning of the heart can best be analysed by recording 10−17 F. In this [4], and other, previous designs it was always
all of the relevant potential information on this surface, i.e. necessary to stabilize the operating point of the operational
by constructing a body-surface-potential map (BSPM). In amplifier at zero frequency by providing an internal source of
principle, this entails making a time record of the potential dc bias current within the probe. We were able to achieve this
at each point on the surface. In practice, of course, we are without making any compromises in the input impedance of
constrained to accept finite spatial resolution. Even so, to the probe within its operating bandwidth (typically 0.01 Hz
obtain a satisfactory resolution (say 1 cm to a few centimetres) to 50 kHz). A block diagram of the probe, including the dc
means that we must make use of a large number of separate biasing circuit, is shown in figure 1. With this configuration
sensor probes. If these are conventional in nature (as is it is apparent that the probe is acting as a voltmeter. A typical
the case for BSPM), each sensor must make low-resistance measured noise-voltage spectral plot for this type of probe is
contact with the skin. This low resistance is achieved by using shown in figure 2 for two values [2] (approximately 0.2 and
a high-conductivity electrolytic paste as the interface between 1 pF) of source–probe capacitance.
292
An ultra-low-noise probe
Figure 2. The noise-voltage spectral density of the OPA 111-based probe of figure 1 with input-electrode arrangements giving
source–probe coupling capacitances of approximately 0.2 and 1 pF [4].
Figure 3. A plot of the probe-output noise voltage at 1 Hz versus the bias-stabilization-network resistance R for the OPA 111-based system.
The 1012 data point was taken using a glass-encapsulated, carbon-film, resistor.
noise performance, high input impedance, on-chip guarding, The quality of the fabrication of the chip is such that the
superior (unconditional) stability and extended frequency effects of low-frequency fluctuations and drift (thermally or
response. A block diagram of the INA 116, configured as a otherwise induced) are almost exactly balanced out between
charge (coulombmeter) amplifier, is shown in figure 4, with the inputs. This makes the INA 116 a very suitable amplifier
the signal applied to the non-inverting input and the inverting for our purposes.
input grounded. It can be seen here that, although guarding The coulombmeter (INA 116-based) probe was
is applied to both inputs, the inverting input is treated as a constructed using modern surface-mounting techniques.
dummy (i.e. grounded). However, this inverting amplifier, This allowed us to extend and maintain the level of on-
on-chip with its non-inverting dual, plays an essential role. chip guarding through to the input-electrode structure. In
293
R J Prance et al
Figure 4. The block diagram of the INA 116, configured as a charge amplifier, with the signal applied to the non-inverting input and the
inverting input grounded.
practice, this hybrid guarding could be made so accurate probe (figure 4) over the range 1 mHz to 100 kHz. Here,
that, to an extremely high degree, leakage-current paths could it is clear that the response is flat down to 30 mHz, with a
be eliminated. In the actual embodiment a double-sided slight and irregular rise (<2 dB) from 30 to 1 mHz. That
printed-circuit-board layout was selected with the bottom the output remains stable over 15 h (>50 000 s) in figure 6
(reverse) face used as a guard plane while the top face was indicates that this response continues down to at least 20 µHz.
configured as a guarded input-electrode structure. As shown If required, the low-frequency rise below 30 mHz can be
in the layout of figure 5, with the INA 116 amplifier surface truncated by imposing a high-pass filter (3 dB point at 0.1 Hz)
mounted, guard tracks were provided either side of the input on the output of the probe (figure 7(b)). For comparison, we
electrode and maintained over the entire circuit board. This show in figure 8 the frequency response of the OPA 111-
planar design (top-surface tracks and reverse plane) provided based probe (figure 1) acting as a voltmeter [1–4] (with a
a remarkably high degree of guarding and shielding. The few times 1012 input impedance at 1 Hz). From the first-
design was implemented using a software package (IsoPro) order roll off in this response around 100 mHz, it is obvious
and cut using a Quick Circuit 5000 PCB milling machine that the input impedance of this probe contains a resistive
(T-Tech, Inc). The use of a milling machine ensured that component. This is in contrast to the purely capacitive input
the electrical-insulation properties of the circuit board were to the INA 116-based probe configuration of figure 4. Using
not compromised by wet chemical etching. Naively, without a potential-divider-based technique, with a 10 pF series input
this extremely accurate guarding, we would have expected to the probe, we found that its input capacitance was close to
the leakage to be that quoted by the manufacturer ('3 fA). 1 pF (corresponding to an input impedance >1012 at 1 Hz).
However, as we shall see, the actual leakage current is The improvement in noise performance of the INA 116-
negligible on the scale of femto-amperes. based system compared with previous probes is striking. All
noise measurements were carried out using a HP 3562A
3. The probe performance signal analyser in logarithmic resolution mode. A ×10 low-
voltage noise preamplifier was inserted between the probe
In order to demonstrate unambiguously that the INA and the signal analyser. This ensured that the probe noise
116-based probe was functioning as a charge amplifier was well above the noise floor of the HP 3562A. In figure 9
(coulombmeter) a dc voltage was applied to its (non- we show the noise response referred to the input (the noise
inverting) input through a 10 pF capacitor; the response with spectral density) between 1 Hz and 100 kHz. As can be seen,
time was then measured. In figure 6 we show a record the input noise voltage at 1 Hz is 2 µV Hz−1/2 , reducing
of this output taken over 15 h with a stabilized dc voltage to 0.2 µV Hz−1/2 at 100 Hz. At 1 Hz this constitutes a
of 10 V applied to the input. During this time the output factor of 15 improvement over the OPA 111-based voltmeter
from the source was monitored and remained within ±5 mV design, quite sufficient to make almost all of the electrical
at all times. As can be seen, with some fluctuations the activity known to occur in the human body accessible in
output remains that of the input over very long time scales. real time. We note that, although 1 Hz is clearly within
This is the result to be expected if the probe is acting as a the 1/f -noise regime in figure 9, the noise level is still only
coulombmeter, measuring charge rather than voltage, with 10 µV Hz−1/2 at 0.1 Hz (not shown). Quite obviously,
a concomitant frequency response flat down to dc. It is to make use of this noise performance the quasi-dc drift
apparent from figure 6 that this is the case. In figure 7(a) in the probe has to be very small. As we have already
we show the measured frequency response for the INA 116 pointed out, this certainly requires that the guarding on the
294
An ultra-low-noise probe
Figure 5. The printed-circuit-board layout of the INA 116-based probe, designed with IsoPro software and machined on a Quick Circuit
5000 PCB milling machine (copper tracks are shown shaded).
(a)
Figure 6. A plot of the long-time-scale (≈15 h) output-voltage
response of a coulombmeter configuration probe with 10 V of dc
offset applied to its (non-inverting) input.
295
R J Prance et al
Figure 8. The frequency response of the OPA 111-based Figure 11. The quasi-dc drift curve for the INA 116-based probe
electrical-potential probe of figure 1 over the range 0.01 Hz to taken over 26 h; typically the drift is <2.5 mV h−1 .
1 MHz.
296
An ultra-low-noise probe
297