You are on page 1of 16

Adv Ther (2021) 38:2114–2129

https://doi.org/10.1007/s12325-021-01692-z

REVIEW

A Review of Robotic and OCT-Aided Systems


for Vitreoretinal Surgery
Elan Z. Ahronovich . Nabil Simaan . Karen M. Joos

Received: January 18, 2021 / Accepted: February 27, 2021 / Published online: April 3, 2021
Ó The Author(s) 2021

ABSTRACT systems, and (4) magnetic guidance robots. This


review covers the improvements and the
The introduction of the intraocular vitrectomy remaining needs for safe, cost-effective clinical
instrument by Machemer et al. has led to deployment of robotic systems in vitreoretinal
remarkable advancements in vitreoretinal sur- surgery.
gery enabling the limitations of human physi-
ologic capabilities to be reached. To overcome
the barriers of perception, tremor, and dexter- Keywords: Image-guided surgery; Medical
ity, robotic technologies have been investigated robotics; Micromanipulator; Ophthalmic
with current advancements nearing the feasi- surgery; Ophthalmology; Optical coherence
bility for clinical use. There are four categories tomography; Telemanipulation; Vitreoretinal
of robotic systems that have emerged through surgery
the research: (1) handheld instruments with
intrinsic robotic assistance, (2) hand-on-hand
robotic systems, (3) teleoperated robotic

E. Z. Ahronovich
Advanced Robotics and Mechanism Applications
(ARMA) Laboratory, Department of Mechanical
Engineering, Vanderbilt University, Nashville, TN
37235, USA

N. Simaan
Advanced Robotics and Mechanism Applications
(ARMA) Laboratory, Department of Mechanical
Engineering, Department of Computer Science,
Vanderbilt University, Nashville, TN 37235, USA
e-mail: nabil.simaan@vanderbilt.edu

K. M. Joos (&)
Vanderbilt Eye Institute, Vanderbilt University
Medical Center, Nashville, TN 37232, USA
e-mail: karen.joos@vumc.org

K. M. Joos
Department of Biomedical Engineering, Vanderbilt
University, Nashville, TN 37235, USA
Adv Ther (2021) 38:2114–2129 2115

INTRODUCTION
Key Summary Points
Needs and Challenges in Ophthalmic
To overcome the barriers of perception, Surgery
tremor, and dexterity in vitreoretinal
surgery, robotic technologies have been Despite vast advances in vitreoretinal surgery
investigated with current advancements since Machemer et al. [1], it presents challenges
nearing the feasibility for clinical use. to surgeons in terms of precision, perception,
There are four categories of robotic and manipulation dexterity. A typical setup
systems that have emerged through the during these procedures involves multiport
research: (1) handheld instruments with access into the vitreous cavity with thin tools
intrinsic robotic assistance, (2) hand-on- (e.g., picks, graspers, light source). Visualization
hand robotic systems, (3) teleoperated of the retina through a surgical microscope is
robotic systems, and (4) magnetic achieved through the pupil with adding a
guidance robots. focusing lens on or above the cornea. Surgeons
have to stabilize the eye while operating one to
The future of vitreoretinal surgery may two instruments within the vitreoretinal space.
include some of the robotic systems or Surgical tools generally lack distal dexterity
implementations of technology owing to their small size (generally less than
introduced in the development of the 900 lm in diameter) and they have to be
robots. maneuvered under deficient perception condi-
tions. For example, the visualization of the
Limitations of surgical robots include
anatomy is limited through a dilated iris, espe-
barriers presented by sensor-actuation lag
cially if dilation is poor. Tool shadows are hard
that can be limited by using high
to perceive because of complex lighting condi-
sampling frequencies. Also, heavy
tions with chandelier illumination or a small
computational demands from visual
moving endoilluminator light held in the sur-
feedback technologies currently make
geon’s second hand and the retinal anatomy
real-time integration challenging.
presents semitransparent features (e.g., retinal
Clinical surgical robotics will likely use membrane) that can be difficult to see. Added to
technologies such as optical coherence these challenges is the difficulty of measuring
tomography and tool tip force tool tip interaction forces because of the inter-
measurements to add accuracy. ference from tool–trocar friction forces as con-
cluded by Jagtap et al. [2]. Further, Jensen et al.
[3] showed that the delicate anatomy of the
retina can apply reaction forces less than
7.5 mN for 77% of the duration of a vitreoreti-
nal procedure which was perceived by the sur-
DIGITAL FEATURES geon only 20% of the time. This review
describes current retinal surgical limitations,
This article is published with digital features,
the parallel developments of optical coherence
including a summary slide, to facilitate under-
tomography (OCT) and surgical robotics, and
standing of the article. To view digital features
the intersection of the two technologies which
for this article go to https://doi.org/10.6084/
have the potential to revolutionize patient care.
m9.figshare.14125301.
Success of many vitreoretinal procedures
depends on safe manipulation of the delicate
anatomy of the retina. For example, epiretinal
membrane peeling requires careful peeling of a
layer that can be on average 61 ± 28 lm thick
2116 Adv Ther (2021) 38:2114–2129

[4] while avoiding trauma to the underlying a microscope and due to limited tactile feed-
retinal anatomy. Treatment of retinal detach- back. Typically, humans rely on a tactile
ments requires maneuvers using picks and response as a form of confirmation of contact-
miniature graspers/cutters while avoiding exac- ing an intended target when performing a task
erbating the retinal detachment and avoiding such as surgery. However, the delicate intraoc-
inadvertent touching of the lens or causing ular anatomy does not produce reaction forces
retinal hemorrhage. Previous experimental great enough to overcome the friction between
characterizations of physiologic tremors illus- a surgical tool and trocar to reach a level per-
trate the challenge of accomplishing such pre- ceptible to humans. With this lack of tactile
cise tasks. For example, the average root mean feedback, surgeons are forced to rely heavily on
square (rms) amplitude of tremor with a four- visual cues to discern tool proximity and con-
subject user study ranged between 14 and tact of the surgical tools with the anatomy [3].
142 lm [5] when holding a tool still and However, intraocular visualization via a surgical
between 59 and 341 lm when actuating a microscope has limits of useful depth percep-
microsurgical grasper. In another example tion and the visible field of view is restricted by
Singh and Riviere [6] tracked the tool motion the dilated iris. In addition to manipulating
during epiretinal membrane peeling and repor- instruments inside the eye, surgeons may also
ted the rms amplitude of tremor at 38 lm for a tilt the eye under the microscope. In some cases,
single-subject study. These reported magnitudes the target anatomy is difficult to visualize using
of tool tremor are large enough to make white-light imaging. For example, epiretinal
microretinal procedures exceedingly challeng- membranes are mostly transparent which stan-
ing. To overcome tremor, four approaches using dard light microscopes are incapable of visual-
robotics were considered in the literature. In the izing without the addition of a steroid
first method, a handheld miniature robotic suspension or indocyanine green (ICG) to stain
platform was used for tremor cancellation of the membranes. Surgeons may also use endo-
tracked instruments. Examples of this approach scopes for auxiliary peripheral visualization;
include Riviere et al.’s Micron [7], Song et al.’s however, these scopes have limited resolutions
SMART OCT-based device [8], and Cheon et al.’s compared to current microscopes. To address
OCT-guided depth-locking handheld microin- the challenge of depth perception, previous
jector [9]. In the second method, a hand-on- investigations have augmented instruments
hand approach is used where the surgical with OCT probes. For example, Balicki et al. [18]
instrument is held by a robot and the surgeon’s used an A-scan OCT probe to maintain the
hand. Forces by the surgeon’s hand are used to distance of a robotic-controlled tool tip from
command the robot (tool) motion while also retinal anatomy. Yu et al. [19, 20] used forceps
providing tremor filtration, Taylor et al.’s integrated with a B-mode OCT probe for depth
Steady-Hand eye robot [10] for example. A third perception feedback. In Yu et al. [19] it was
approach using telemanipulation with a sur- shown that OCT feedback improved depth
geon controlling a robotically guided surgical perception and success of approaching a surface
tool via a control station detached from the and peeling a surface membrane.
robot was initially explored by Charles [11], Wei In addition to precision and perception
et al. [12], Yu et al. [13], and Meenink et al. [14]. challenges, there are challenges due to rigid
Finally, the fourth approach using extraocular instrumentation offering limited maneuver-
magnetic fields achieves manipulation of two ability. Vitreoretinal surgical tools are generally
types of intraocular robots. Kummer et al. [15] slender instruments with rigid shafts. These
demonstrated intraocular microcapsule robots instruments are constrained to the traditional
and Charreyron et al. [16, 17] demonstrated four-DOF (degrees of freedom) motions avail-
steerable magnetic-tipped catheters for drug able to minimally invasive instruments (tilt in
injection delivery and retinal vein cannulation. two directions and rotation about and transla-
Vitreoretinal surgery is also complicated by tion along the longitudinal axis of the tool). As
perception barriers owing to operating through a result, the tool tip dexterity of these
Adv Ther (2021) 38:2114–2129 2117

instruments is quite limited. One can reach a limiting vitreoretinal surgery and discuss
particular site, but with limited control of tool prospective areas of development to further
tip orientation. Surgeons have to carry out sur- lessen these constraints.
gical maneuvers of lifting membranes with
picks and graspers despite their limited distal tip
dexterity and contend with the need for METHODS TO IMPROVE
bimanual manipulation in order to stabilize the VISUALIZATION: OPTICAL
eye while operating tools inside it. Ikuta et al. COHERENCE TOMOGRAPHY (OCT)
[21] proposed the use of manual active bending
forceps, but the current clinical repertoire of OCT is a standard diagnostic and surgical
surgical forceps still remains predominantly planning ophthalmic tool that can develop
without active distal bending. The four scenar- cross-sectional images of tissue using light
ios of ocular and intraocular manipulation have reflectance. Dayani et al. used a handheld
been considered [22] with an emphasis on device during planned surgical procedure
quantifying the possible benefits of instrumen- interruptions [27]. Binder et al. first used a
tation with intraocular dexterity. It was shown microscope-mounted unit following surgical
that adding a single DOF of bending sideways manipulations [28]. The Duke [29–38], Cleve-
can increase orientational dexterity, compared land Clinic/Case Western Reserve [36, 39–47],
to rigid instruments, by 31.6% and 57.7% for Vanderbilt [48], and international groups
translational and rotational manipulation, [28, 49–52] developed improvements for the
respectively. Several tools and robotic instru- microscope-mounted intraoperative OCT sys-
ments have been considered to overcome the tems with commercial US Food and Drug
problem of intraocular dexterity. In Simaan Administration (FDA)-approved systems avail-
et al. [23] the concept of intraocular dexterity able for the operating room [53].
tools using continuum bending cannulas was Non-OCT surgical intraocular endoscopes
introduced and later implemented by Wei et al. are FDA-approved [54–60], but there is not yet
[12, 24] and Yu et al. [13]. He et al. [25] intro- an approved facile ophthalmic OCT probe to
duced a prototype of a 0.9 mm handheld con- enable peripheral retina visualization as well as
tinuum robot offering intraocular dexterity that bypass corneal and lenticular opacities that may
was later integrated with a robotic platform by hinder direct central visualization [61]. Iftimia
Song et al. [26]. et al. [62] developed an A-scan 250 lm OCT
The aforementioned challenges of tremor, probe for one-dimensional measurement of
limited visual and tactile perception, and tool tissue.
tip dexterity can be alleviated in several ways. Balicki et al. [18] reported an intraocular
For example, active tools can limit the effects of common path A-scan OCT probe. A 20-gauge
human tremor to increase surgical precision by coplanar probe was developed which success-
filtering sensor input to produce tremorless fully guided the depth of mid-infrared laser
actuator control. Visual perception can be aug- incisions of the retina [63]. The OCT-imaging
mented using OCT to obtain cross-sectional component alone was housed within a 25-gauge
imaging of tissue yielding a richer set of infor- tube which was readily amenable to imaging
mation of target anatomy. Using sensors that through the more recent 23-gauge and 25-gauge
can detect forces imperceptible to humans trocars preferred for contemporary retinal surg-
enhances a surgeon’s effective tactile percep- eries [64]. Addition of a needle [65] or forceps
tion. For higher dexterity, continuum segment [19] increased the size only to 23 gauge. This
tools offer higher ranges of motion with OCT probe has been added to robotic platforms
manipulation directly by a surgeon or attached described in the following section.
to robotic systems for greater levels of manipu- Ray et al. presented a custom mount that
lability and accuracy. In the following we dis- attached the Bioptigen handheld probe to an
cuss some of the tools that have been developed ophthalmic surgical microscope [66]. OCT
to address the three main areas outlined above images from 24 patients undergoing macular
2118 Adv Ther (2021) 38:2114–2129

hole or epiretinal membrane surgery were ana- ROBOTIC SYSTEMS


lyzed with subsequent quantitative measure-
FOR VITREORETINAL PROCEDURES
ment of geometry and retinal thickness
providing insight into the anatomical changes
To address the host of complications with vit-
in the retina resulting from macular surgery and
reoretinal surgery the literature presents several
verifying surgery completion. The feasibility
types of robotic systems with an array of fea-
and safety of microscope-mounted OCT in
tures offering surgical advantages. These robotic
prospectively and retrospectively enrolled eyes
systems fall into four categories distinguished
during several ophthalmic surgeries was repor-
by their interaction with the surgeon: (1)
ted, and it enhanced surgeons’ understanding
handheld, (2) telemanipulated, (3) hand-on-
of the underlying anatomy in more than 40% of
hand, and (4) magnetically controlled systems.
the cases during lamellar keratoplasty and reti-
nal membrane peeling [67–69].
Microscope-integrated systems have been Handheld Systems for Vitreoretinal
developed which combine the OCT and surgical Procedures
microscope optical paths to enable imaging
simultaneously with surgical maneuvers Handheld robotic surgical tools have been
[30, 31, 41, 42, 70, 71]. The Duke research pro- explored to address the challenges of physio-
totype [30] was clinically evaluated in a study logic tremor and force perception with minimal
involving eight patients undergoing surgery for disruption to the surgical workflow. One
macular holes, epiretinal membranes, and vit- example of a handheld robotic surgical device is
reomacular traction [32]. The results confirmed Micron [7, 76]. Micron is a vitreoretinal surgical
the ability to observe surgically induced chan- tool designed to sense a surgeon’s tremor and
ges in retinal contour and macular hole con- distinguish those movements from intentional
figuration. The ability to acquire OCT images motion. It leverages the effects of constructive
simultaneously through the microscope over- and destructive interaction of wave signals to
came a major limitation of a separate external filter the user’s tremor from the tool tip. The
large imaging probe by eliminating the need for device senses the user’s movements and iden-
frequent pauses during surgery. Commercial tifies tremor as any input signal within the
systems are now available with the first being 8–12 Hz frequency band as determined else-
the Zeiss RESCAN 700. Increases in imaging where [5]. To stabilize the surgical tool, piezo-
speed combined with improved computation electric actuators direct Micron’s tool tip in a
using graphics processing units (GPUs) have direction opposite and equal in magnitude to
enabled real-time 3D [72] and 4D [73, 74] the tremor. Becker et al. [76] (Fig. 1i) improved
intraoperative OCT. This provides improved on the target acquisition of Micron by adding
feedback on instrument position. Real-time image guidance. Two cameras were attached
adjustments of the OCT focus to maintain par- directly to a surgical microscope as a stereo pair
focality with the surgical microscope at differ- for registering the tool tip’s location and to
ent axial positions and zoom levels is possible measure the tool’s lateral displacement relative
[41]. Improved visualization includes heads-up to target vessels. The displacement data is used
display (HUD) technology that adds OCT visu- as an additional feedback signal in combination
alization into the microscope ocular view [45] with the user’s movements to achieve 63%
to project OCT cross-sections [42] onto the success in experimental vessel cannulation as
surgical field. Carrasco-Zevallos et al. demon- shown in Fig. 1ii.
strated volumetric 4D OCT data for real-time Force sensing was introduced by Gonenc
surgical feedback [73, 75]. Others are examining et al. [77] using fiber Bragg grating (FBG) strain
displaying images on virtual reality (VR) plat- sensors to enable force sensing at the tool tip
forms and gradually OCT has been added to following the design from Iordachita et al.’s [78]
several robotic systems as reported in the fol- vitreoretinal tool with a 0.25 mN resolution
lowing section. (5.6 9 10-5 lb). The sensors are located at the
Adv Ther (2021) 38:2114–2129 2119

Fig. 1 i An active surgical tool, Micron, that senses a user’s manipulation with the inclusion of visual feedback to
tremor during manual microsurgeries and cancels the Micron (Figures reproduced with permission from Becker
effects of tremor on tool tip trajectory using piezoelectric et al. [76])
actuators for procedures such as retinal vein cannulation. ii
The improvement of tool tip trajectory during manual

tool’s tip to isolate retinal forces from the scle- Telemanipulation Robotic Systems
rotomy interaction forces, thereby inhibiting for Vitreoretinal Procedures
the user from imposing damaging forces on the
retina. Yang et al. [79] optimized the Micron’s Wei et al. [12, 22] presented a multiplatform
design to allow six DOFs with a 4-mm-diameter robotic system in Fig. 2 that can manipulate the
hemispherical workspace by using a parallel eyeball and offer intraocular dexterity via a
actuator architecture offering more robust tre- bending continuum robot capable of deploying
mor control. Yang et al. [80] used the six-DOF microstents or grippers. The system allows for a
Micron to demonstrate the advantage of using software-controlled remote center of motion
tremor stabilization in acquiring clear B-mode (RCM). The robotic arms of this system have
and C-mode OCT image acquisition and pre- been demonstrated to enable deployment of
sented a precursor of a clinical tool capable of microstents in chorioallantoic chick mem-
tremor filtration paired with the visual feedback branes [24] and were integrated with OCT for
capabilities of OCT. control feedback [20] (Fig. 3).
The Integrated Robotic Intraocular Snake or Yu et al. [19] developed surgical forceps
IRIS, developed by He et al. [25], is a robotic shown in Fig. 4 integrated with a B-mode for-
surgical tool prototype offering surgeons ward-imaging OCT probe enabling real-time
intraocular dexterity that is meant to be a intraocular imaging to improve accuracy for
handheld device or a mountable attachment to membrane peeling procedures. The forceps were
a robotic platform. IRIS was designed to match made with a 25-gauge stainless steel (SS) tube
the sizing of 20-gauge ophthalmic surgical tools within a 23-gauge SS tube. The outer tube slides
with a 0.9 mm outer diameter. The continuum along the 25-gauge tube forcing the opening
segment of the IRIS is 10 mm long and has two and closing of the forceps. The group showed
rotational DOFs each with ± 45° of bending. that integration of the custom OCT forceps with
The linear actuators of the IRIS exhibited large the robot manipulator improved accuracy and
backlash and low actuation resolution which reduced the number of attempts needed to
limited the realizable precision of the final accomplish a membrane peeling procedure.
design. Their results also emphasized the importance of
2120 Adv Ther (2021) 38:2114–2129

Fig. 2 A two-arm parallel robot used for vitreoretinal operations that allow eye maneuvering and intraocular dexterity
(Figure courtesy of Nabil Simaan)

the increased DOFs enabled with two separate


actuators. This six-DOF robot creates a highly
dexterous system eliminating a surgeon’s tre-
mor input and minimizing the effect of user
fatigue; however, challenges associated with
intraocular maneuverability are not addressed.
The Preceyes Surgical Robotic System was
developed for microintraocular procedures like
retinal vein cannulation and internal limiting
membrane peeling. The Preceyes system has a
motion controller that the surgeon uses to
Fig. 3 A nine-DOF robot with parallel actuation platform command surgical tool tip position. The surgi-
carrying a stenting robot capable of maneuvering with cal tool is attached to a parallelogram manipu-
precision better than 5 lm (Photo courtesy of Nabil lator that enables operation around an RCM. By
Simaan) setting a virtual point, at the sclerotomy,
around which a tool will rotate provides
proximity of the OCT image monitor to the advantages for the user. First, minimized inter-
surgeon to minimize head and eye movements action forces between the surgical tool and
of the user. sclerotomy mitigate any scleral trauma; second,
Nasseri et al. developed a six-DOF miniature the orientation of the orbit remains unaffected,
robot [81] (Fig. 5). This robot creates linear and which maintains line of site to the surgeon. The
rotational motion using two parallel prismatic group has also integrated the system with
joints. There are two advantages of such a par- external OCT imaging to establish tool tip
allel mechanism. First the overall stiffness is boundaries aiding the user in tool manipulation
greater than what would be possible with seri- and preventing inadvertent retinal contact or
ally linked actuators. The second advantage is puncture [82].
Adv Ther (2021) 38:2114–2129 2121

Fig. 4 OCT-forceps with OCT fiber embedded in the 25-gauge stainless steel tube (SST). External actuation causes the
23-gauge SST to slide axially on the 25-gauge SST causing opening–closing of the forceps (Photo courtesy of Karen Joos)

tracks can be independently positioned with


two separate actuators to allow six DOFs of
surgical tool manipulation. Each tool is kine-
matically constrained to a fixed RCM defined by
the device geometry. The IRISS was tested on
cadaver porcine eyes to demonstrate retinal
vein cannulation and for cataract extraction.
The group also tested the feasibility of using
OCT for calibrating the RCM point and con-
cluded that RCM alignment using visible red
dot lasers may introduce deviations too large for
full autonomy.
Del Giudice et al. [86, 87] developed con-
tinuum robots for multiscale motion (CREM)
Fig. 5 A hybrid parallel-serial surgical cooperative robot demonstrating a novel concept for teleoperated
capable of microscale motion using piezo actuators robot actuation for surgeries requiring micro-
(Figure reproduced with permission from Nasseri et al. scale motion like microvascular reconstruction
[81]) and image-based (OCT) diagnosis. The CREM
robot is capable of maneuvering tools within
both macro- and micro-workspaces with a
Gijbels et al. [83] developed a telemanipu-
positional resolution of 1 lm. It was shown that
lated robotic system to aid in retinal vein can-
3D OCT images may be obtained by using the
nulation and epiretinal membrane peeling and
robot’s micromotion capability while carrying a
achieved successful in vivo human retinal vein
B-mode OCT probe, which was an adaptation
cannulation [84]. By enabling axial translation
from Shen et al. [88]. In addition to validating
as in Gijbels et al. [83], the size of tooling
3D OCT on a cadaveric porcine retina, closed-
around the eye was minimized and the robot
loop control and OCT-guided visual servoing at
workspace for maneuvering around the micro-
the micromotion scale was also demonstrated
scope viewing cone was maximized. Wilson
for targeting a needle into a microchannel.
et al. [85] also developed a unique telemanipu-
While this system was not miniaturized to
lated intraocular robotic interventional surgical
operate within the eye, the same design concept
system (IRISS) with two manipulators that
may be used for high-precision and low-cost
mount and travel on semicircular tracks. The
2122 Adv Ther (2021) 38:2114–2129

robotic devices for manipulating needles within tool at a fixed position even if the surgeon lets
the eye with OCT feedback. go of it.
The Steady-Hand Robot [10] and Steady-
Hand-on-Hand Robotic Systems Hand Robot 2 [89], shown in Fig. 6, were
for Vitreoretinal Procedures developed to augment a surgeon’s capabilities
with retinal and other microsurgeries in mind.
By minimizing positioning error of a tool tip The Steady-Hand is a cooperative robotic sys-
caused by human tremor, some active surgical tem, where the surgeon and a robotic actuator
tools enable microsurgeries that are impossible simultaneously control a surgical tool. The sur-
with traditional surgical tools. Maneuvering geon manipulates surgical tools in the same
handheld robotic tools for microsurgeries still fashion as traditional tools with the robot con-
requires a tremendous level of skill, however, as troller reading force signals from the surgeon’s
the level of tremor filtration is limited by the hand movements to drive the robot. The robot
stroke magnitude of the device’s actuators. Like is capable of producing smooth, natural motion
telemanipulated robots, hand-on-hand robots profiles that a surgeon would typically use dur-
offer a greater advantage of tremor filtration by ing retinal procedures while eliminating the
leveraging the mechanical stiffness of robotic extraneous tool movements that result from
systems to drive surgical tools in tandem with tremor.
the surgeon. The first realization of the Steady-Hand from
Hand-on-hand robotic systems allow the Taylor et al. [10] improved success of needle
surgeon to drive a tool mounted on a robotic insertion into a hole 150 lm in diameter by
platform using force input commands. This 36% [90] when compared to manual needle
approach has several advantages in terms of insertion. Balicki et al. [18] used a custom
reduced cost and ease of clinical deployment. In 25-gauge surgical pick with integrated OCT for
addition, the robot can be used for tremor fil- epiretinal membrane peeling. With the OCT
tering, for position recall, and for reducing surgical pick providing visual feedback, the
surgeon fatigue since the robot can hold the Steady-Hand was able to maintain a specified
distance of the surgical tool tip from retinal

Fig. 6 i An early iteration of a cooperative surgical robot, the Steady-Hand Robot 2, or Eye Robot 2 (ER2). (Fig-
the Steady-Hand Robot with robotic platform and surgical ure reproduced with permission from Üneri et al. [10])
tool attached to a six-DOF force sensor for robot control
(Photo courtesy of Russell H. Taylor and Iulian I.
Iordachita). ii An iteration of the Steady-Hand Robot,
Adv Ther (2021) 38:2114–2129 2123

tissue to within 10 lm of a desired 150 lm. The peeling motion that minimizes resistance to
OCT imaging enabled identification of struc- limit membrane tearing.
tures beyond surface layers as targets that can be
used to guide tool puncturing tasks while lim- Magnetically Controlled Robot Systems
iting puncture depths. Üneri et al. [89] evolved
the Steady-Hand by including a force sensor Finally, other approaches using magnetically
attached to the surgical tool. The additional controlled microrobots have been explored over
sensor provides applied tool forces as feedback the past decade. These systems utilize an
data to the robot controller to limit maximum extraocular magnetic field to control robotic
forces applied to intraocular tissue. The force microcapsules within the eye for procedures like
feedback data aids in guiding surgeons to avoid retinal vein cannulation and localized drug
unintended destructive contact and, like the delivery using drug-eluting microcapsules.
OCT feedback, is used to maintain tool posi- Kummer et al. [15] used a magnetic field system
tioning with respect to intraocular anatomy. called the OctoMag (Fig. 7) to guide a micro-
The combined inputs from the OCT imaging capsule robot in five DOFs, i.e., three degrees of
and force sensor optimize tool trajectory, for positional control and two orientational
instance, while maintaining tool angle during a degrees. One of the advantages of magnetic
systems is achieving high levels of intraocular
dexterity and maneuverability without physical
attachment to the extraocular space [15]. Min-
imizing attachments between the extraocular
and intraocular space eliminates the eye
manipulability constraints imposed by pars
plana sclerotomy surgical tools and maximizes
the degrees of manipulation available for the
line of site of intraocular anatomy. These
advantages, however, come at the expense of
very complex magnetic field generators that
encompass a large portion of the space around
the patient’s head.
Charreyron et al. [16, 17] used the OctoMag
magnetic field system to drive a magnetic tip
microcannula for delivery of gene therapy
injections to subretinal tissue. The group
developed semiautonomous control that auto-
matically aligns the tool magnetic field to keep
the tool tip perpendicular to the retinal surface
at a target site identified by the surgeon. The
user, while following tool manipulation
through the microscope, determines when the
tool tip is optimally placed for injection. Mag-
netic microcannulas maintain the advantages
of intraocular dexterity that characterize
microcapsule robots by relying solely on the
Fig. 7 The extraocular magnetic field generating system, magnetic field for actuation. These magnetic
OctoMag, capable of guiding magnetic drug-eluting manipulators also offer a potential safety
microcapsules and magnetic tip microcannula in five advantage over traditional rigid surgical tools
DOFs, i.e., three degrees of positional control and two because of their limited rigidity and limitations
orientational degrees (Figure courtesy of Bradley Nelson of achievable forces. The group explored the
and MagnebotiX AG, Zurich, Switzerland)
2124 Adv Ther (2021) 38:2114–2129

feasibility of autonomous control and also on OCT visual feedback, the robot controller
explored OCT imaging for improved tool tip can bypass the force input from a user to drive
tracking. Their system exhibits 11 degrees of surgical tools to locations identified with the
angular error of the magnetic field in a worst- OCT. By utilizing distance measurement capa-
case scenario which translates to 4.2 mm of bilities of OCT or force guidance with FBG
displacement of a 21-mm-long cannula. For full integrated tool tips, the likelihood of tissue
autonomous control to be clinically realizable, damaging contact is reduced.
precision of magnetic field alignment will be With a system like CREM, tasks requiring
necessary ultra-precision and autonomous intervention
are possible with miniaturization of the tech-
nology. Utilizing equilibrium modulation as an
DISCUSSION actuation technique enables robotics in surgery
to carry out microscale tasks with much finer
Vitreoretinal surgical techniques and available positional adjustments than other robots. Fur-
procedures are in a transitional state due to the ther, enabling these micromanipulation tasks
contributions of advanced visualization tech- does not sacrifice the capabilities in the macro-
niques and the development of robotic surgical workspace enabling a greater number of surgical
devices. The future of vitreoretinal surgery is the tasks.
possible realization of autonomy for an array of Although surgical robots have shown their
procedures available today, but more impor- capabilities in advancing vitreoretinal surgery,
tantly, enabling procedures that are not cur- they are not without limitations. Developing
rently available because of human physiological OCT-guided robotic tools that do not disrupt
limitations. For instance, gene therapy injec- the clinical workflow requires hardware capable
tions require the utmost precision for target of high sampling frequencies allowing for
acquisition. Magnetic field generators guide improved OCT image quality via filtering and
magnetic drug-eluting microrobots enabling multiframe averaging techniques. Registration
intraocular drug injections. Magnetic intraocu- of the OCT probe image frame to the robot
lar robots eliminate the pars plana sclerotomy frame is challenging—especially for systems
manipulation constraints but require a large using an external OCT. Stabilization of the
volume of the workspace surrounding the robotic tool relative to the patient head (or use
patient’s head. An alternative approach for tar- of active eye motion tracking) are key to
geting is OCT imaging which delivers high- enhancing safety. Finally, the ability to achieve
definition images necessary for intraocular fast and safe tool retraction in case of involun-
navigation. When OCT imaging is paired with tary ocular movements or in case of a clinical
robotic platforms as a feedback modality, the emergency will be paramount to safe clinical
combined system can drive tools to specific deployment.
retinal locations with improved accuracy and
reduced dependence upon human surgical
skills. CONCLUSIONS
Some of the robotic systems mentioned are
already in a developmental stage where auton- Clinical surgical robotics will likely use tech-
omy is possible. The Steady-Hand robot, for nologies like OCT and tool tip force measure-
instance, demonstrated autonomous manipu- ments to enhance the surgeon’s perception and
lation of a needle inserted into a 150–250 lm accuracy. These robotic systems will include
hole, improving success of insertion by an powerful control computers, particularly those
average of 31.8% compared to handheld inser- using OCT, with calibration methods and con-
tions. With added OCT imaging, performing trol algorithms accounting for safe anatomical
vitreoretinal tasks expands further to assist in movements during procedures. The future of
manipulation of semitransparent membranes vitreoretinal surgery may include some of the
and target subsurface tissue. Depending solely robotic systems or implementations of
Adv Ther (2021) 38:2114–2129 2125

technology introduced in the development of distribution and reproduction in any medium


the robots discussed in this review. or format, as long as you give appropriate credit
This review article is based on previously to the original author(s) and the source, provide
conducted studies and does not contain any a link to the Creative Commons licence, and
new studies with human participants or animals indicate if changes were made. The images or
performed by any of the authors. other third party material in this article are
included in the article’s Creative Commons
licence, unless indicated otherwise in a credit
ACKNOWLEDGEMENTS line to the material. If material is not included
in the article’s Creative Commons licence and
your intended use is not permitted by statutory
Funding. This work was supported in part by regulation or exceeds the permitted use, you
the National Institute of Health National Eye will need to obtain permission directly from the
Institute (NEI/NIH) grant 1R01EY028133 (KMJ, copyright holder. To view a copy of this licence,
NS), in part by The National Science Foundation visit http://creativecommons.org/licenses/by-
grant CMMI-1537659 (NS, KMJ), in part by the nc/4.0/.
Joseph Ellis Family and William Black Research
Funds (KMJ), in part by the Unrestricted
Departmental Grant to the Vanderbilt Eye
Institute from Research to Prevent Blindness,
REFERENCES
Inc., NY (KMJ). No Rapid Service Fee was
received by the journal for the publication of 1. Machemer R. Vitrectomy, a pars plana approach.
Trans Am Acad Ophthalmol. 1971;75:813–20.
this article.
2. Jagtap AD, Riviere CN. Applied force during vitre-
Authorship. All named authors meet the oretinal microsurgery with handheld instruments.
International Committee of Medical Journal In: The 26th annual international conference of the
Editors (ICMJE) criteria for authorship for this IEEE engineering in medicine and biology society,
vol. 1. IEEE; 2004. pp. 2771–2773.
article, take responsibility for the integrity of
the work as a whole, and have given their 3. Jensen P, Gupta P, De Juan E. Quantification of
approval for this version to be published. microsurgical tactile perception. In: Proceedings of
the first joint BMES/EMBS conference. 1999 IEEE
engineering in medicine and biology 21st annual
Disclosures. Elan Ahronovich declares that conference and the 1999 annual fall meeting of the
he has no conflict of interest. Nabil Simaan and biomedical engineering society, vol. 2. IEEE; 1999.
Karen Joos have received funding from NEI/NIH p. 839.
and NSF.
4. Wilkins JR, Puliafito CA, Hee MR, et al. Character-
ization of epiretinal membranes using optical
Compliance with Ethics Guidelines. This coherence tomography. Ophthalmology.
review article is based on previously conducted 1996;103(12):2142–51.
studies and does not contain any new studies
with human participants or animals performed 5. Riviere CN, Rader RS, Khosla PK. Characteristics of
hand motion of eye surgeons. In: Proceedings of the
by any of the authors. 19th annual international conference of the IEEE
engineering in medicine and biology society ‘Mag-
Data Availability. Data sharing is not nificent Milestones and Emerging Opportunities in
applicable to this article as no datasets were Medical Engineering’ (Cat. No. 97CH36136), vol. 4.
IEEE; 1997. pp. 1690–1693.
generated or analyzed during this review article.
6. Singh S, Riviere C. Physiological tremor amplitude
Open Access. This article is licensed under a during retinal microsurgery. In: Proceedings of the
Creative Commons Attribution-NonCommer- IEEE 28th annual northeast bioengineering confer-
cial 4.0 International License, which permits ence (IEEE Cat. No. 02CH37342). IEEE; 2002. pp.
171–172.
any non-commercial use, sharing, adaptation,
2126 Adv Ther (2021) 38:2114–2129

7. Riviere CN, Ang WT, Khosla PK. Toward active 18. Balicki M, Han JH, Iordachita I, et al. Single fiber
tremor canceling in handheld microsurgical optical coherence tomography microsurgical
instruments. IEEE Trans Robot Autom. 2003;19(5): instruments for computer and robot-assisted retinal
793–800. surgery. In: International conference on medical
image computing and computer-assisted interven-
8. Song C, Gehlbach PL, Kang JU. Active tremor can- tion. Springer; 2009. pp. 108–115.
cellation by a ‘‘smart’’ handheld vitreoretinal
microsurgical tool using swept source optical 19. Yu H, Shen JH, Shah RJ, Simaan N, Joos KM. Eval-
coherence tomography. Opt Express. 2012;20(21): uation of microsurgical tasks with OCT-guided and/
23414–21. or robot-assisted ophthalmic forceps. Biomed Opt
Express. 2015;6(2):457–72.
9. Cheon GW, Huang Y, Kang JU. Active depth-lock-
ing handheld micro-injector based on common- 20. Yu H, Shen JH, Joos KM, Simaan N. Calibration and
path swept source optical coherence tomography. integration of B-mode optical coherence tomogra-
In: Optical fibers and sensors for medical diagnos- phy for assistive control in robotic microsurgery.
tics and treatment applications XV, vol. 9317. IEEE ASME Trans Mechatron. 2016;21(6):2613–23.
International Society for Optics and Photonics;
2015. p. 93170U. 21. Ikuta K, Kato T, Nagata S. Development and
experimental verification of micro active forceps for
10. Taylor R, Jensen P, Whitcomb L, et al. A steady- microsurgery. In: MHS’96 proceedings of the sev-
hand robotic system for microsurgical augmenta- enth international symposium on micro machine
tion. Int J Robot Res. 1999;18(12):1201–10. and human science. IEEE; 1996. pp. 229–234.

11. Charles S. Dexterity enhancement in microsurgery 22. Wei W, Goldman RE, Fine HF, Chang S, Simaan N.
using telemicro-robotics. In: Medicine meets virtual Performance evaluation for multi-arm manipula-
reality II, interactive technology and healthcare: tion of hollow suspended organs. IEEE Trans Robot.
visionary applications for simulation, visualization, 2008;25(1):147–57.
robotics; 1994.
23. Simaan N, Taylor RH, Handa JT. System and
12. Wei W, Goldman R, Simaan N, Fine H, Chang S. method for macro-micro distal dexterity enhance-
Design and theoretical evaluation of microsurgical ment in micro-surgery of the eye. Google Patents;
manipulators for orbital manipulation and 2019. US Patent 10,406,026.
intraocular dexterity. In: Proceedings 2007 IEEE
international conference on robotics and automa- 24. Wei W, Popplewell C, Chang S, Fine HF, Simaan N.
tion. IEEE; 2007. pp. 3389–3395. Enabling technology for microvascular stenting in
ophthalmic surgery. J Med Devices. 2010;4(1):
13. Yu H, Shen JH, Joos KM, Simaan N. Design, cali- 014503.
bration and preliminary testing of a robotic tele-
manipulator for OCT guided retinal surgery. In: 25. He X, Van Geirt V, Gehlbach P, Taylor R, Iordachita
2013 IEEE international conference on robotics and I. IRIS: integrated robotic intraocular snake. In:
automation. IEEE; 2013. pp. 225–231. 2015 IEEE international conference on robotics and
automation (ICRA). IEEE; 2015. pp. 1764–1769.
14. Meenink T, Naus G, de Smet M, Beelen M, Stein-
buch M. Robot assistance for micrometer precision 26. Song J, Gonenc B, Guo J, Iordachita I. Intraocular
in vitreoretinal surgery. Investig Ophthalmol Vis snake integrated with the steady-hand eye robot for
Sci. 2013;54(15):5808–5808. assisted retinal microsurgery. In: 2017 IEEE inter-
national conference on robotics and automation
15. Kummer MP, Abbott JJ, Kratochvil BE, Borer R, (ICRA). IEEE; 2017. pp. 6724–6729.
Sengul A, Nelson BJ. OctoMag: an electromagnetic
system for 5-DOF wireless micromanipulation. IEEE 27. Dayani PN, Maldonado R, Farsiu S, Toth CA.
Trans Robot. 2010;26(6):1006–17. Intraoperative use of handheld spectral domain
optical coherence tomography imaging in macular
16. Charreyron SL, Zeydan B, Nelson BJ. Shared control surgery. Retina. 2009;29(10):1457.
of a magnetic microcatheter for vitreoretinal tar-
geted drug delivery. In: ICRA; 2017. pp. 4843–4848. 28. Binder S, Falkner-Radler CI, Hauger C, Matz H,
Glittenberg C. Feasibility of intrasurgical spectral-
17. Charreyron SL, Boehler Q, Danun A, Mesot A, domain optical coherence tomography. Retina.
Becker M, Nelson BJ. A magnetically navigated 2011;31(7):1332–6.
microcannula for subretinal injections. In: IEEE
transactions on biomedical engineering. 2020. 29. Hahn P, Migacz J, O’Connell R, Izatt JA, Toth CA.
Unprocessed real-time imaging of vitreoretinal
surgical maneuvers using a microscope-integrated
Adv Ther (2021) 38:2114–2129 2127

spectral-domain optical coherence tomography 41. Tao YK, Srivastava SK, Ehlers JP. Microscope-inte-
system. Graefe’s Arch Clin Exp Ophthalmol. grated intraoperative OCT with electrically tunable
2013;251(1):213–20. focus and heads-up display for imaging of oph-
thalmic surgical maneuvers. Biomed Opt Express.
30. Tao YK, Ehlers JP, Toth CA, Izatt JA. Intraoperative 2014;5(6):1877–85.
spectral domain optical coherence tomography for
vitreoretinal surgery. Opt Lett. 2010;35(20):3315–7. 42. Ehlers JP, Srivastava SK, Feiler D, Noonan AI, Rollins
AM, Tao YK. Integrative advances for OCT-guided
31. Pasricha ND, Shieh C, Carrasco-Zevallos OM, et al. ophthalmic surgery and intraoperative OCT:
Real-time microscope-integrated OCT to improve microscope integration, surgical instrumentation,
visualization in DSAEK for advanced bullous ker- and heads-up display surgeon feedback. PLoS One.
atopathy. Cornea. 2015;34(12):1606. 2014;9(8):e105224.

32. Hahn P, Migacz J, O’Connell R, et al. Preclinical 43. El-Haddad MT, Tao YK. Automated stereo vision
evaluation and intraoperative human retinal instrument tracking for intraoperative OCT guided
imaging with a high-resolution microscope-inte- anterior segment ophthalmic surgical maneuvers.
grated spectral domain optical coherence tomog- Biomed Opt Express. 2015;6(8):3014–31.
raphy device. Retina. 2013;33(7):1328.
44. Ehlers JP, Han J, Petkovsek D, Kaiser PK, Singh RP,
33. Hahn P, Carrasco-Zevallos O, Cunefare D, et al. Srivastava SK. Membrane peeling-induced retinal
Intrasurgical human retinal imaging with manual alterations on intraoperative OCT in vitreomacular
instrument tracking using a microscope-integrated interface disorders from the PIONEER study.
spectral-domain optical coherence tomography Investig Ophthalmol Vis Sci. 2015;56(12):7324–30.
device. Transl Vis Sci Technol. 2015;4(4):1–1.
45. Ehlers JP, Goshe J, Dupps WJ, et al. Determination
34. Ehlers JP, Tao YK, Farsiu S, Maldonado R, Izatt JA, of feasibility and utility of microscope-integrated
Toth CA. Integration of a spectral domain optical optical coherence tomography during ophthalmic
coherence tomography system into a surgical surgery: the DISCOVER Study RESCAN Results.
microscope for intraoperative imaging. Investig JAMA Ophthalmol. 2015;133(10):1124–32.
Ophthalmol Vis Sci. 2011;52(6):3153–9.
46. Cost B, Goshe JM, Srivastava S, Ehlers JP. Intraop-
35. Hahn P, Migacz J, O’Connell R, Maldonado RS, Izatt erative optical coherence tomography-assisted
JA, Toth CA. The use of optical coherence tomog- Descemet membrane endothelial keratoplasty in
raphy in intraoperative ophthalmic imaging. Oph- the DISCOVER study. Am J Ophthalmol.
thalmic Surg Lasers Imaging Retina. 2011;42(4): 2015;160(3):430–7.
S85–94.
47. Ehlers JP, Uchida A, Srivastava SK. Intraoperative
36. Ehlers J. Intraoperative optical coherence tomog- optical coherence tomography-compatible surgical
raphy: past, present, and future. Eye. 2016;30(2): instruments for real-time image-guided ophthalmic
193–201. surgery. Br J Ophthalmol. 2017;101(10):1306–8.

37. Ehlers JP, Tao YK, Farsiu S, Maldonado R, Izatt JA, 48. Malone JD, El-Haddad MT, Bozic I, et al. Simulta-
Toth CA. Visualization of real-time intraoperative neous multimodal ophthalmic imaging using
maneuvers with a microscope-mounted spectral swept-source spectrally encoded scanning laser
domain optical coherence tomography system. ophthalmoscopy and optical coherence tomogra-
Retina. 2013;33(1):232. phy. Biomed Opt Express. 2017;8(1):193–206.

38. Grewal DS, Bhullar PK, Pasricha ND, et al. Intraop- 49. Lytvynchuk L, Glittenberg C, Binder S. Intraopera-
erative 4D microscope-integrated optical coherence tive spectral domain optical coherence tomogra-
tomography guided 27-gauge transvitreal reti- phy: technology, applications, and future
nochoroidal biopsy for choroidal melanoma. perspectives. In: Meyer C, Saxena S, Sadda S, edi-
Retina. 2017;37(4):796. tors. Spectral domain optical coherence tomogra-
phy in macular diseases. New Delhi: Springer; 2017.
39. Ehlers JP, McNutt S, Dar S, Tao YK, Srivastava SK. pp. 423–443.
Visualisation of contrast-enhanced intraoperative
optical coherence tomography with indocyanine 50. Binder S, Falkner-Radler C, Hauger C, Matz H,
green. Br J Ophthalmol. 2014;98(11):1588–91. Glittenberg C. Clinical applications of intrasurgical
SD-optical coherence tomography. Investig Oph-
40. Ehlers JP, Tao YK, Srivastava SK. The value of thalmol Vis Sci. 2010;51(13):268–268.
intraoperative OCT imaging in vitreoretinal sur-
gery. Curr Opin Ophthalmol. 2014;25(3):221.
2128 Adv Ther (2021) 38:2114–2129

51. Matz H, Binder S, Glittenberg C, et al. Intraopera- coherence tomography probe to guide real-time
tive applications of OCT in ophthalmic surgery. laser ablation. Lasers Surg Med. 2014;46(3):
Biomed Eng. 2012;57(SI-1 Track-P):297–297. 193–202.

52. Maier MM, Nasseri A, Framme C, Bohnacker S, 64. Joos KM, Shen JH. Miniature real-time intraopera-
Becker MD, Heinrich D, et al. Die intraoperative tive forward-imaging optical coherence tomogra-
optische Kohärenztomografie in der Netzhaut- phy probe. Biomed Opt Express. 2013;4(8):1342–50.
Glaskörper-Chirurgie. Aktuelle Erfahrungen und
Ausblick auf künftige Entwicklungsschritte. Klin 65. Joos KM, Shen JH. Preliminary design and evalua-
Monbl Augenheilkd. 2018;235(09):1028–34. tion of a B-scan OCT-guided needle. Photonics.
2014;1:260–266.
53. Carrasco-Zevallos OM, Viehland C, Keller B, et al.
Review of intraoperative optical coherence tomog- 66. Ray R, Barañano DE, Fortun JA, et al. Intraoperative
raphy: technology and applications. Biomed Opt microscope-mounted spectral domain optical
Express. 2017;8(3):1607–37. coherence tomography for evaluation of retinal
anatomy during macular surgery. Ophthalmology.
54. Sasahara M, Kiryu J, Yoshimura N. Endoscope-as- 2011;118(11):2212–7.
sisted transscleral suture fixation to reduce the
incidence of intraocular lens dislocation. J Cataract 67. Branchini LA, Gurley K, Duker JS, Reichel E. Use of
Refract Surg. 2005;31(9):1777–80. handheld intraoperative spectral-domain optical
coherence tomography in a variety of vitreoretinal
55. Tarantola RM, Agarwal A, Lu P, Joos KM. Long-term diseases. Ophthalmic Surg Lasers Imaging Retina.
results of combined endoscope-assisted pars plana 2016;47(1):49–54.
vitrectomy and glaucoma tube shunt surgery.
Retina. 2011;31(2):275. 68. Ehlers JP, Dupps WJ, Kaiser PK, et al. The prospec-
tive intraoperative and perioperative ophthalmic
56. Olsen TW, Pribila JT. Pars plana vitrectomy with imaging with optical coherence tomography (PIO-
endoscope-guided sutured posterior chamber NEER) study: 2-year results. Am J Ophthalmol.
intraocular lens implantation in children and 2014;158(5):999–1007.
adults. Am J Ophthalmol. 2011;151(2):287–96.
69. De Benito-Llopis L, Mehta JS, Angunawela RI, Ang
57. Al Sabti K, Raizada S. Endoscope-assisted pars plana M, Tan DT. Intraoperative anterior segment optical
vitrectomy in severe ocular trauma. Br J Ophthal- coherence tomography: a novel assessment tool
mol. 2012;96(11):1399–403. during deep anterior lamellar keratoplasty. Am J
Ophthalmol. 2014;157(2):334–41.
58. Maeda M, Watanabe M, Ichikawa K.
Goniosynechialysis using an ophthalmic endo- 70. Carrasco-Zevallos O, Keller B, Viehland C, et al. 4D
scope and cataract surgery for primary angle-closure microscope-integrated OCT improves accuracy of
glaucoma. J Glaucoma. 2014;23(3):174–8. ophthalmic surgical maneuvers. In: Ophthalmic
technologies XXVI, vol. 9693. International Society
59. Kawashima S, Kawashima M, Tsubota K. Endo- for Optics and Photonics; 2016. p. 969306.
scopy-guided vitreoretinal surgery. Expert Rev Med
Devices. 2014;11(2):163–8. 71. Todorich B, Shieh C, DeSouza PJ, et al. Impact of
microscope-integrated OCT on ophthalmology
60. Francis BA, Kwon J, Fellman R, et al. Endoscopic resident performance of anterior segment surgical
ophthalmic surgery of the anterior segment. Surv maneuvers in model eyes. Investig Ophthalmol Vis
Ophthalmol. 2014;59(2):217–31. Sci. 2016;57(9):OCT146–53.

61. Li J, Quirk BC, Noble PB, Kirk RW, Sampson DD, 72. Wieser W, Draxinger W, Klein T, Karpf S, Pfeiffer T,
McLaughlin RA. Flexible needle with integrated Huber R. High definition live 3D-OCT in vivo:
optical coherence tomography probe for imaging design and evaluation of a 4D OCT engine with 1
during transbronchial tissue aspiration. J Biomed GVoxel/s. Biomed Opt Express. 2014;5(9):2963–77.
Opt. 2017;22(10):106002.
73. Carrasco-Zevallos O, Keller B, Viehland C, et al. Live
62. Iftimia NV, Bouma BE, Pitman MB, Goldberg B, volumetric (4D) visualization and guidance of
Bressner J, Tearney GJ. A portable, low coherence in vivo human ophthalmic surgery with intraoper-
interferometry based instrument for fine needle ative optical coherence tomography. Sci Rep.
aspiration biopsy guidance. Rev Sci Instrum. 2016;6:31689.
2005;76(6):064301.
74. Viehland C, Keller B, Carrasco-Zevallos OM, et al.
63. Li Z, Shen JH, Kozub JA, Prasad R, Lu P, Joos KM. Enhanced volumetric visualization for real time 4D
Miniature forward-imaging B-scan optical
Adv Ther (2021) 38:2114–2129 2129

intraoperative ophthalmic swept-source OCT. 83. Gijbels A, Vander Poorten EB, Stalmans P, Van
Biomed Opt Express. 2016;7(5):1815–29. Brussel H, Reynaerts D. Design of a teleoperated
robotic system for retinal surgery. In: 2014 IEEE
75. Shen L, Carrasco-Zevallos O, Keller B, et al. Novel international conference on robotics and automa-
microscope-integrated stereoscopic heads-up dis- tion (ICRA). IEEE; 2014. pp. 2357–2363.
play for intrasurgical optical coherence tomogra-
phy. Biomed Opt Express. 2016;7(5):1711–26. 84. Gijbels A, Smits J, Schoevaerdts L, et al. In-human
robot-assisted retinal vein cannulation, a world
76. Becker BC, Voros S, Lobes LA, Handa JT, Hager GD, first. Ann Biomed Eng. 2018;46(10):1676–85.
Riviere CN. Retinal vessel cannulation with an
image-guided handheld robot. In: 2010 Annual 85. Wilson JT, Gerber MJ, Prince SW, et al. Intraocular
international conference of the IEEE engineering in robotic interventional surgical system (IRISS):
medicine and biology. IEEE; 2010. pp. 5420–5423. mechanical design, evaluation, and master–slave
manipulation. Int J Med Robot Comput Assist Surg.
77. Gonenc B, Balicki MA, Handa J, et al. Preliminary 2018;14(1):e1842.
evaluation of a micro-force sensing handheld robot
for vitreoretinal surgery. In: 2012 IEEE/RSJ inter- 86. Del Giudice G, Wang L, Shen JH, Joos K, Simaan N.
national conference on intelligent robots and sys- Continuum robots for multi-scale motion: micro-
tems. IEEE; 2012. pp. 4125–4130. scale motion through equilibrium modulation. In:
2017 IEEE/RSJ international conference on intelli-
78. Iordachita I, Sun Z, Balicki M, et al. A sub-milli- gent robots and systems (IROS). IEEE; 2017. pp.
metric, 0.25 mN resolution fully integrated fiber- 2537–2542.
optic force-sensing tool for retinal microsurgery. Int
J Comput Assist Radiol Surg. 2009;4(4):383–90. 87. Del Giudice G, Orekhov A, Shen J, Joos K, Simaan
N. Investigation of micro-motion kinematics of
79. Yang S, MacLachlan RA, Riviere CN. Design and continuum robots for volumetric OCT and OCT-
analysis of 6 DOF handheld micromanipulator. In: guided visual servoing. IEEE/ASME Trans Mecha-
2012 IEEE international conference on robotics and tron. 2020. https://doi.org/10.1109/TMECH.2020.
automation. IEEE; 2012. pp. 1946–1951. 3043438.

80. Yang S, Balicki M, MacLachlan RA, et al. Optical 88. Shen JH, Kozub J, Prasad R, Joos KM. An intraocular
coherence tomography scanning with a handheld OCT probe. Investig Ophthalmol Vis Sci.
vitreoretinal micromanipulator. In: 2012 Annual 2011;52(14):1326–1326.
international conference of the IEEE Engineering in
Medicine and Biology Society. IEEE; 2012. pp. 89. Üneri A, Balicki MA, Handa J, Gehlbach P, Taylor
948–951. RH, Iordachita I. New steady-hand eye robot with
micro-force sensing for vitreoretinal surgery. In:
81. Nasseri MA, Eder M, Eberts D, et al. Kinematics and 2010 3rd IEEE RAS and EMBS international confer-
dynamics analysis of a hybrid parallel-serial micro- ence on biomedical robotics and biomechatronics.
manipulator designed for biomedical applications. IEEE; 2010. pp. 814–819.
In: 2013 IEEE/ASME international conference on
advanced intelligent mechatronics. IEEE; 2013. pp. 90. Kumar R, Goradia TM, Barnes AC, et al. Perfor-
293–299. mance of robotic augmentation in microsurgery-
scale motions. In: International conference on
82. Maberley DA, Beelen M, Smit J, et al. A comparison medical image computing and computer-assisted
of robotic and manual surgery for internal limiting intervention. Springer; 1999. pp. 1108–1115.
membrane peeling. Graefe’s Arch Clin Exp Oph-
thalmol. 2020:258(4):773–8.

You might also like