Professional Documents
Culture Documents
net/publication/223533988
CITATIONS READS
64 926
3 authors, including:
Some of the authors of this publication are also working on these related projects:
All content following this page was uploaded by Sinan D Haliyo on 25 January 2019.
Mechatronics
journal homepage: www.elsevier.com/locate/mechatronics
a r t i c l e i n f o a b s t r a c t
Article history: Actual endoscopes and boroscopes, widely used in industry and in minimal invasive surgery, have con-
Received 12 March 2009 siderable limitations, mainly due to their low number of degrees of freedom and their manual operation.
Accepted 1 December 2009 Two different solutions for the electrical actuation of articulated endoscopes are presented in this paper.
The technical constraints for this kind of application are very limited space for the actuators and high per-
formance in terms of torque and angular reach. The first solution classically consists in a 2 d.o.f. structure
Keywords: steered by two pairs of antagonist shape memory alloy (SMA) wires. The sizing and preload determina-
Active endoscope
tion for those actuators follow an original analytical approach. The second solution consists in a multi-
Shape memory alloys
Minimal invasive surgery
d.o.f. structure actuated by thin NiTi springs mounted in an antagonist configuration and directly
Sliding mode control integrated in the structure of the endoscope. The geometry of the springs is obtained by optimization
Genetic algorithms through genetic algorithms and finite elements method. Experiments show good adequacy between real
behaviour and numerical model and also validate the approach.
This study is also enhanced by a control scheme specifically developed for SMA actuators in an antag-
onist configuration. It is based on a first order sliding mode scheme, which has the advantage of a great
structural simplicity. The experimental results show that this solution can reach a good compromise
between the dynamic behaviour of the actuator, its energy consumption and the structural lifetime of
the endoscope.
Ó 2009 Elsevier Ltd. All rights reserved.
0957-4158/$ - see front matter Ó 2009 Elsevier Ltd. All rights reserved.
doi:10.1016/j.mechatronics.2009.12.001
252 V.D. Sars et al. / Mechatronics 20 (2010) 251–264
Fig. 1. An existing 1 d.o.f. endoscope and its application to stomach wall inspection.
the optimal design of, respectively 2 d.o.f. and multi-d.o.f. SMA dri- catheter with three inflatable balloons able to move itself inside
ven endoscopes. In Section 5, we propose a new approach for the a tube in peristalsis mode. In [6], a rolling stents systems mounted
tunning of general 1st order sliding mode controller applied to on the head allows the progression through the intestines. [7] pre-
SMA driven endoscopes. Finally, conclusion and perspectives are sents different systems for ‘‘inchworm’’ locomotion alongside with
drawn in Section 6. in vivo experiments on pigs. These solutions do not resolve the
progression and curvature control problems in case of cavity explo-
ration where the device cannot rely on the environment.
2. State of the art On the contrary, several works deal with active structures,
capable of maintaining their configuration without contacting the
There are several solutions found in the literature for an actively environment. Several 1 or 2 degrees of freedom devices are found
controlled endoscope. Most of them address the field of minimal in that category. Some of them are actuated through electrical mo-
invasive surgery. Generally, these systems are dedicated to a par- tors that pull the cables in place of the operator [8,9]. But these
ticular type of medical intervention. For example, in [3], authors solutions are somehow cumbersome and are not well-suited for
propose an autonomous poly-articulated endoscope for a coronary portable instruments. The other 3 d.o.f. active endoscopes that
anastomosis, where the kinematic and the actuation of the device can be found in the literature are driven using SMA wires
are optimized for needle insertion. An other active endoscope, [10,11], or helicoidal springs [12]. In this case, the SMA wires or
especially designed for the treatment of an aortic aneurysm, is pre- springs are embedded into the instruments body and disposed
sented in [4]. It has a serial/parallel structure of pneumatic actua- antagonist two by two so that a pair of actuators can control the
tors for precise positioning of an endovascular stent. These systems device movement in a given flexion plane. However, the maximal
are obviously geared towards very specific operations and there- bending angles these devices can reach are generally significantly
fore cannot be used for general endoscopy. below the angles the existing manually controlled endoscopes
Other systems are designed for automatic progression in the reach. We show in Section 3 that this problem can be solved by a
body (arteries, intestines). For example, in [5], authors present a consistent choice of the SMA wires dimensions and preload.
However according to the needs stated above, and especially to 4. The large number of joint axis to be controlled in case of ‘‘snake-
allow to work in areas difficult to access, several ‘‘snake-like” de- like” devices implies to consider only simple control solutions,
vices with a great number of mobilities, each controlled indepen- easily repeatable, with as few sensor equipment and tuning
dently, were also developed. Most of these systems also rely on parameters as possible.
Shape Memory Alloys for their actuation (Fig. 2). In [13], the endo-
scope presented has 13 mm in diameter and is comprised of five Various types of control schemes for SMA based actuators have
segments actuated by thin SMA solenoids. The radius of curvature been proposed in the literature (for a large survey of existing solu-
of the device is interesting though the generated forces are weak. A tions, one should refer to [17]). They are generally of three types:
comparable system, having 2 mm in diameter is presented in [14]. (1) Linear controllers [18,19], which are simple but poorly compat-
A spine-like structure is presented in [15]. It is comprised of binary ible with the non-linear and hysteretic behaviour of SMA. (2) Mod-
actuated simple elements, using transverse SMA blades. An impor- el based controllers [20,21], which require a difficult off-line
tant number of elements is needed for a decent curvature. Finally, identification of parameters such as the convection coefficient,
in [16] authors present a thin structure (0.9 mm in diameter), actu- the specific heat. (3) Variable structure controllers such as on–off
ated by NiTi plate springs. This solution allows for a better integra- controllers [22] or sliding mode controllers [23,24], which combine
tion of the structure but the produced motion is of weak parametric simplicity and relative insensitivity in respect to model
amplitude. Moreover, in most of these prototypes, the internal uncertainties. In Section 5, an original efficient approach for the
diameter for passing tools or electrical connexions or optical fib- tuning of 1st order sliding mode controllers applied to SMA based
bers is generally too small. In this paper, we present an approach actuation systems is proposed and then experimentally validated
combining genetic algorithm and finite elements method for the on a typical 2 d.o.f. active endoscope.
design of an original distributed actuation system suited for mul-
ti-d.o.f. endoscopes actuation (Section 4).
Finally, control of SMA based active endoscopes requires a par- 3. Design of a 2 d.o.f. SMA based active endoscope
ticular approach taking into account the specificities of SMA mate-
rials as well as the severe constraints encountered in endoscopy, 3.1. System description
especially in the case of medical applications:
This section describes the structural design and dimensioning of
1. SMA springs or wires that are mounted in an antagonist config- a 2 d.o.f. active endoscope actuated by four antagonist SMA wires.
uration can sometime pull again each others and generate This device (Fig. 3) is composed of an articulated head (made by
internal efforts without producing any motion. This is the case Fort Imaging Systems Company), a 32 cm long tubular body con-
when they are activated simultaneously or simply because they taining the SMA wires, a stretcher for the preloading of the SMA
are identically pre-stressed. Unnecessary internal efforts have wires and a handle.
to be kept small in order to avoid important structural con- The mobile head of the endoscope is composed of a serial
straints and energy waste. arrangement of six stainless steel rings whose external diameter
2. SMA based actuators have very poor efficiency. The major part is 7 mm (Fig. 4). The rings are linked to each other through rotoid
of the injected energy is released as heat and contributes to joints whose axes are alternatively oriented by 90°. The internal
temperature elevation. The energy consumption has thus to diameter of the rings is 5.6 mm. The maximal bending angle is
be minimized by limiting unnecessary or chaotic activation of about 100° in all directions and the minimal curvature radius is
SMA actuators. less than 20 mm.
3. In common endoscopic applications, manually driven endo- The mobile head is classically actuated by stretching on four
scopes can bend their field of vision from 0° (frontal vision) to cables mounted in a two by two antagonist configuration (Fig. 4).
90° (lateral vision) in less than 1 s. An SMA driven endoscope Hence, the left and right (resp. up and down) cables produce bend-
should exhibit a comparable bandwidth although a well known ing in the horizontal (resp. vertical) plane. Stretching force is cre-
drawback of SMA based actuators is their lack of rapidity espe- ated by contracting SMA wires attached to the cables. The
cially when their cooling relies on natural convection only. contraction of the SMA wires is obtained by increasing their
Fig. 3. Two d.o.f. active endoscope. Fig. 4. Mobile head of the 2 d.o.f. endoscope.
254 V.D. Sars et al. / Mechatronics 20 (2010) 251–264
3.3. Prestrain
The maximal bending angle that the device can achieve de-
pends on the prestrain of SMA actuators. This prestrain stand for
the strain that the SMA wires undergo when the mechanical struc-
ture to be actuated is in its original position and the SMA wires are
not activated.
As an example, Fig. 8 shows the forces the right and left SMA
wires would apply to the structure (with respect to the structure
Fig. 8. Force applied by the SMA wires with respect to their prestrain values.
Fig. 14. Three solutions for integrating the plate springs into the structure.
Fig. 18. Geometric parameters of the spring used in the genetic algorithm.
In order to define the SMA springs geometric parameters which SMA springs in an antagonist configuration two by two. At each
correspond to an optimal actuator structure a genetic algorithm generation, the population of actuators is renewed by keeping
based method was used. the most fitting individuals and creating new ones by crossover
Fig. 18 shows the geometric parameters of an SMA spring. These and mutation.
properties were chosen as the alleles characterising the popula- The software used for this part of the work is based on GALib
tions of actuators, each actuator being constituted of four identical 2.45, a freeware library available at http://www.lancet.mit.edu/
258 V.D. Sars et al. / Mechatronics 20 (2010) 251–264
Table 1
Optimized geometrical parameters.
Specs Value A prototype actuator was produced as shown in Fig. 24. This
prototype is comprised of two articulated rings and four integrated
Torque at 0° 30.5 mN m
Torque at 9° 10.1 mN m
SMA springs. The springs are mounted two by two on each side of
Maximum stress 543 MPa the joint axis. They are positioned and insulated from the metallic
Rotation range (loaded) 11.5° structure by four bases made of peek material. Different electrical
Rotation range (unloaded) 13.5° connexions complete the assembly.
From the electrical point of view, springs situated at the same
side of the link are serially connected to each other using flexible
copper blades, which also supply the connection to external wires.
Epoxy resin, with a high shear strength is used for fixing all parts.
Note that included springs do not come from the optimization
presented above but were designed for a preliminary evaluation
(Fig. 25). These springs were produced using laser cutting. A micro-
scopic observation of the produced springs has shown a good sur-
face quality (Fig. 26).
The performances of this actuator were experimentally studied.
As shown in Fig. 27, the maximum rotation of the unloaded proto-
type is 10.5° in both directions. The setup shown in Fig. 28 was
used to evaluate its load capacity. The torque measured at 0° rota-
tion is 17 mN m. Table 3 shows a comparison between these mea-
Fig. 23. Maximum reachable angle for the optimized actuator. sured performances and the results of a numerical simulation for
this actuator. As we can see, the differences are relatively small:
5.5% for the torque and 16% for the rotation. Hence, the simulation
the nominal loading line. The intersection of the two plots indi- tool is validated and theoretical results obtained for the optimized
cates that the maximum reachable angle is at 11.5°. springs can be considered trustworthy.
260 V.D. Sars et al. / Mechatronics 20 (2010) 251–264
5. Control solution for SMA based active endoscopes A sliding curve equation was also defined in the plane e_ vs. e:
Table 3
Comparison of experimental and theoretical performances.
1. if e > 0 and e_ > 0 then left SMA is heated (point A in Fig. 32)
2. if e > 0 and e_ < 0 then
(a) if Sðe; e_ Þ > 0 then left SMA is heated (point B in Fig. 32)
(b) if Sðe; e_ Þ < 0 then nothing (point C in Fig. 32)
3. if e < 0 and e_ < 0 then right SMA is heated (point D in Fig. 32)
4. if e < 0 and e_ > 0 then
(a) if Sðe; e_ Þ < 0 then right SMA is heated (point E in Fig. 32)
(b) if Sðe; e_ Þ > 0 then nothing (point F in Fig. 32)
In the area containing points A and B, the joint angle error is po-
sitive thus, the left SMA actuator is activated in order to produce a
positive joint torque and a joint acceleration in the same sense.
This ensures that e_ (in case of area A) or e (in case of area B) de-
creases (curves a and b, respectively in Fig. 33). For the same rea- Fig. 31. Sliding mode control: set of experimental sliding lines.
son, the right SMA actuator is activated as the system is located in
the areas of points D and E.
In the two areas corresponding to points C and F, e_ and e have the position of these areas with respect to the sliding line shows
opposite signs, resulting in a decrease in magnitude of e. Moreover, that the error decreasing velocity is higher than the targeted
The smallest value, K = 0.8 s1 always lead to a very slow trajec-
tory during which the sliding line was crossed very often (see
Fig. 34 corresponding to 70°).
On the contrary, for K = 100 s1 the trajectory in the e vs. e_ plane
was a spiral which converged rapidly to the zero point (Fig. 35).
This trajectory shown large oscillations around the desired position
(e = 0). This result is consistent with the fact that for very large val-
ues of K, the C and F neutral areas in the e_ vs. e plane nearly vanish,
leading to an elementary two states controller and imposing the
actuator to be continuously activated with respect to the sign of
the position error e.
The result for an intermediate value K = 4 s1 is shown in
Fig. 36. In this case, the convergence toward the zero point was
more direct and the obtained trajectory in the e_ vs. e plane is inter-
estingly close to the fixed sliding line.
Note that for any value of K, a limit cycle around the desired
Fig. 33. Sliding mode control: typical trajectories. angular position appeared. However, its magnitude remained al-
ways below 1°. It is found that this limit cycle can be almost com-
pletely suppressed by adding a dead zone to the proposed control
decreasing velocity specified by the slope K. In this case, it is then scheme, like depicted in Fig. 37.
wiser to let the system evolve freely, without activating any SMA The influence of the parameter K is explored through the mea-
actuator until an other area of the e_ vs. e plane is reached. Avoiding surements presented in Fig. 38. The evolution of rising times, fall-
any SMA activation in these areas allows to conserve energy and to ing times, the energy consumption and maximal internal efforts
minimize the stress on the structure.
Practically, trajectories in these areas can be of two types as
illustrated by the curves c1 and c2 in Fig. 33. In the first case, the
system’s passivity and the loss of kinetic energy (mainly due to
friction) make the velocity error e_ decrease in magnitude rapidly
enough to reach the sliding line before the error e is nullified. In
the second case, the system enters an other area before reaching
the sliding line.
This control solution satisfies the last requirement mentioned in
Section 2, as it has a very simple structure with only two parame-
ters to be fixed:
and the induced stress. A compromise choice for K, in the particular [6] Breeddveldm P, van der Kouwe DE, van Gorp MAJ. Locomotion through the
intestine by means of rolling stents. In: Proceedings of the ASME design
case of this SMA wire driven endoscope, is obviously K = 5 s1.
engineering technical conferences and computers and information in
engineering conference, Salt Lake City, USA; 2004.
6. Conclusion [7] Phee L, Menciassi A, Gorini S, Pernorio G, Arena A, Dario P. An innovative
locomotion principle for minirobots moving in the gastrointestinal tract. In:
Proceedings of the IEEE international conference on robotics and automation,
As presented earlier in this paper, solutions for active endos- Washington, USA; 2002. p. 1125–30.
copy in the literature are numerous and of various types but al- [8] Nakamura R, Kobayashi E, Masamune K, Sakuma I, Dohi T, Yahagi N, et al.
Multi-DOF forceps manipulator system for laparoscopic surgery. In:
ways suffer performance limitations. Hence, in this paper we Proceedings of the third international conference on medical image
presented two original approaches for the design of integrated computing and computer-assisted intervention; 2000. p. 653–60.
actuation systems for a 2 d.o.f. endoscope on one hand and for a [9] Peirs J, Van Brussel H, Reynaerts D, De Gersem G. A flexible distal tip with two
degrees of freedom for enhanced dexterity in endoscopic robot surgery. In:
poly-articulated endoscope on the other hand. In each case, we de-
Proceedings of the 13th micromechanics europe workshop; 2002. p. 271–4.
tailed the method for designing the actuation system in an optimal [10] Dario P et al. A miniature steerable end-effector for application in an
way. integrated system for computer-assisted arthroscopy. In: Proceedings of the
IEEE international conference on robotics and automation; 1997. p. 1573–9.
Particularly for a poly-articulated endoscope, the design meth-
[11] Peirs J, Reynaerts D, Van Brussel H. A retrospective evaluation of a sma micro-
od makes use of finite elements and genetic algorithms. This ap- actuation. In: Proceedings of the international conference on new actuators;
proach takes into account desired mechanical performances and 2002. p. 77–80.
volumetric constraints. A prototype actuator was then constructed [12] Maeda S, Abe K, Yamamoto K, Tohyama O, Ito H. Active endoscope with SMA
(Shape Memory Alloy) coil springs. In: Proceedings of the IEEE 9th annual int.
and experimented. The experimental results are in good accor- workshop on micro electro mechanical systems; 1996. p. 290–5.
dance with the results obtained by simulation, in terms of maxi- [13] Chang J et al. Development of endovascular microtools. Proc J Micromech
mum joint rotation and torque. Microeng 2002;12:824–31.
[14] Ikuta K, Tsukamoto M, Hirose S. Shape memory alloy servo actuator system
Additionally, a control scheme was developed specifically for with electric resistance feedback. In: Proceedings of the IEEE international
this kind of application based on Shape Memory Alloys. The pro- conference on robotics and automation; 1988. p. 427–30.
posed solution is a first order sliding mode type variable structure [15] Park K, Esashi M. A multilink active catheter with polymide-based integrated
CMOS interface circuit. IEEE J Microelectromech Syst (JMEMS)
controller. The main advantage of this solution comes from its sim- 1996;8(4):349–56.
plicity and intrinsic insensitivity to modelling inaccuracy. This con- [16] Reynaerts D, Peirs J, Van Brussel H. Shape memory micro-actuation for a
trol scheme was experimentally validated on a real 2 d.o.f. gastro-intestinal intervention system. Sensors Actuat A 1999;77:157–66.
[17] Sreekumar M, Singaperumal M, Nagarajan T, Zoppi M, Molfino R. Recent
endoscope driven by SMA wires. It has been shown that a satisfy-
advances in non-linear control technologies for shape memory alloy actuators.
ing compromise between speed, low energy consumption and low J Zhejiang Univ Sci A 2007;8(5):818–29.
internal stress on the structure could be reached by appropriately [18] Kuribayashi E. A new actuator of a joint mechanism using NiTi alloy wire. Int J
Robotic Res 1986;4:47–58.
choosing the sliding line in the phase plane.
[19] Cruz-Hernandez J, Hayward V. An approach to reduction of histeresis in smart
A prototype with optimized SMA springs is actually in the material. In: Proceedings of the IEEE international conference robotic and
works. In addition to the actuation system, it will also integrate a automation; 1998. p. 1510–5.
Hall effect linear position sensor. Moreover, a molded plastic poly- [20] Benzaoui H et al. Experimental and theoretical study of shape memory alloy
wire actuator. J Intell Mater Syst Struct 1997;8(7):619–29.
mer model of the mechanical structure will be realized in order to [21] Gorbet R, Russel R. A novel differential shape memory alloy actuator for
facilitate the electrical integration and lower cost production of the position control. Robotica 1995;13:423–30.
proposed actuator. In long term, a full endoscope with 12 actuated [22] Grant D. Accurate and rapid control of shape memory alloy actuators. PhD
thesis, McGill University; 1999.
rings will be produced in order to conduct in vivo experiments. [23] Ghosh-Roy R, Moura JT, Olgac N. Implementation of sliding mode control using
the concept of perturbation. Mechatronics 1997;7(8):723–36.
References [24] Elahinia MH, Ashrafiuon H. Nonlinear control of a shape memory alloy
actuated manipulator. J Vib Acoust 2002;124(4):566–75.
[25] Mineta T et al. Batch fabricated flat meandering shape memory alloy actuator
[1] Szewczyk J, Desars V, Bidaud Ph. An active tubular polyarticulated micro-
for active catheter. Sensors Actuat A 2001;88:112–20.
system for flexible endoscope. In: Proceeding of the 7th international
[26] Fatikov S, Rembold U. Microsystems technology and microrobotics. Berlin
symposium on experimental robotics, Hawa; 2000. p. 179–88.
Heidelberg: Springer-Verlag; 1997.
[2] de Sars V, Szewczyk J, Bidaud Ph. Force and position control of a SMA actuated
[27] Reynaerts D, Peirs J, Van Brussel H. Production of shape memory alloys for
endoscope. In: Proceedings of the 14th CISM-IFTOMM symposium on theory
microactuation. J Micromech Microeng 1995;5:150–2.
and practice of robots and manipulators, Udine; 2002.
[28] Kohl M. Shape memory microactuators. New York: Springer; 2004.
[3] Sallé D, Bidaud Ph, Morel G. Optimal design of high dexterity modular MIS
[29] Tung A et al. Design and fabrication of tubular shape memory alloy actuators
instrument for coronary artery bypass grafting. In: Proceedings of the IEEE
for active catheters. In: Proceedings of the IEEE 1st international conference on
international conference on robotics and automation, New Orleans, USA; 2004.
biomedical robotics and biomechatronics; 2006. p. 775–80.
p. 1276–81.
[30] Goldberg D. Genetic algorithms in search, optimization and machine
[4] Bailly Y, Chauvin A, Amirat Y. Control of a high dexterity micro-robot based
learning. Addison Wesley; 1989.
catheter for aortic aneurism treatment. In: Proceedings of the IEEE
[31] Slotine J, Sastry S. Tracking control of non-linear systems using sliding
international conference on robotics and automation, New Orleans, USA;
surfaces, with application to robot manipulators. Proc Int J Control
2004. p. 65–70.
1983:465–92.
[5] Mangnan V, Kingsley D, Quinn R, Chiel H. Development of a peristaltic
endoscope. In: Proceeding of the IEEE international conference on robotics and
automation, washington, USA; 2002. p. 347–52.