Professional Documents
Culture Documents
Cylinder Valves
Testor
Dräger Safety UK Limited, Ullswater Close, Kitty Brewster Estate, Blyth, Northumberland NE24 4RG. Telephone: (01670) 352891. Fax: (01670) 356266
PSS® 100 SERIES 1
Index
Section 1 Instructions for Use
Section 8 Accessories
Introduction
The Dräger PSS® 100 Series of Self Contained Compressed Air Breathing Equipment
(C.A.B.E.) meets the requirements EC Council Directive 89/686/EEC, BSEN 1233, and
conforms to European Standard EN137.
PSS® 100 Series is CE Mark Approved and issued with EC Type Examination Certificate.
PSS® 100
PSS® 100 E
PSS® 100 ET
The following versions of the Bodyguard electronic monitoring unit are approved against the
above European Standard and available for use with the Dräger PSS 100 Series of Self
Contained Compressed Air Breathing Equipment (C.A.B.E.).
The units are Intrinsically Safe, conforming to EMC (radio interference requirements) and to
JCDD 38.
The PSS® 100 Series of Self Contained Compressed Air Breathing Equipment (C.A.B.E.)
provides the wearer with respiratory protection when working in a contaminated or oxygen
deficient life threatening environment.
The advanced design gives greater performance, comfort and ease of use. The revolutionary
articulated carrying system improves weight distribution and the patented slide and pivot
mechanism permits freedom of movement and increased manoeuvrability.
Variants of the PSS® 100 Series use the same adjustable articulated carrying system and
high performance first stage pressure reducer but differ in pneumatic configuration. All variants
are however fully compatible with the Dräger range of lung demand valves, facepieces and
compressed air cylinders.
PSS 100
0525
PSS®100 E
PSS® 100 ET
Safety Warnings
l Explain - Instructions for Use supplied with this equipment must be adhered to.
Contact Dräger Limited, Blyth for details of Service Contracts and Service Training
Courses.
l Train - potential wearers in how to wear, and use equipment and to understand its
limitations.
Product Description
The PSS® 100 Series of compressed air respiratory protection equipment consists of three
main variants, (PSS® 100, PSS® 100 E and PSS® 100 ET).
Each variant incorporates an adjustable and articulated carrying system for supporting single
or twin compressed air cylinders.
PSS® 100
Attached to the high performance first stage pressure reducer of the
PSS® 100 is the hose assembly of the mechanically activated pressure gauge and whistle
warning unit.
The Bodyguard is a robust and compact instrument, designed to replace traditional mechanical
pressure gauge, whistle warning unit and Distress Signal Unit (DSU).
Operated by either ‘Button’ - E, or ‘Tally’ - ET, and the system pressure of respiratory protection
equipment, this unique multifunction instrument incorporates pre-programmed features for
continuous monitoring of:
Available for Bodyguard is a ‘snap on’ I.R. Link and Windows based Software Package for
programming of additional monitoring Options, such as user I.D. and datalogging with
downloading of datalogged parameters.
Backplate
Manufactured from moulded carbon composite material, the three elements are anti-static,
lightweight and durable.
The ergonomic design of the adjustable and articulated carrying system provides the wearer
with unrestricted freedom of movement, stability, a high degree of comfort, and reduced
wearer fatigue.
A slide and lock arrangement between the backplate and shoulder harness yoke allows
adjustment of the overall length of the carrying system to suit torso length of wearer. Three
preset length settings available are - short, medium, and long.
A universal flexible joint located in the centre of the waistbelt assembly connects to a vertical
slide arrangement at the base of the backplate.
Positioned at the base of the wearer’s spine, (lumbar region), the flexible joint and slide
arrangement combine to provide a natural pivot point about which the backplate and shoulder
yoke move in response to upper body movements of the wearer.
A rubber shock buffer fitted to the base of the backplate provides additional impact protection.
Side handles, incorporated into the backplate moulding, allow carrying of equipment even
with compressed air cylinder attached.
A mounting arrangement, at the base of the backplate, locates the pressure reducer-pneumatics
assembly.
Harness
Harness arrangement contain:
Padding and adjustable straps remove easily for maintenance and cleaning.
Pneumatic Arrangement
The pneumatic arrangement incorporates a pressure reducer with a threaded handwheel for
connection to the outlet port of the valve of a compressed air cylinder (200bar - 300bar).
TM
Variants that are fitted with ChargAir utilise a 300bar only pressure reducer incorporating a
threaded handwheel for connection only to the outlet port of a 300bar compressed air cylinder
TM
valve. Assembled to the second high pressure outlet port, from the reducer is the ChargAir
supply hose withTMquick release male coupling. Connection of a special secondary supply
hose to ChargAir allows refilling of cylinder(s) from an independent high pressure (maximum
300bar) secondary breathing air supply source, while wearer is breathing from equipment.
A bolt and locknut arrangement at the base of the backplate secures the reducer to backplate.
A degree of radial movement of the reducer on the mounting bolt assists in alignment and
assembly of the reducer handwheel to the outlet port of the cylinder valve.
PSS® 100
A unique dual function hose assembly incorporating an in-line whistle warning unit, (high
pressure activated - medium pressure operated), and high pressure contents gauge for
monitoring air pressure in compressed air cylinder.
Fitted to second outlet port (medium pressure) is a medium pressure hose with quick release
female coupling for attaching a separate lung demand valve.
This port also incorporates the pressure relief valve, (fitted to nozzle of medium pressure
hose), and can also be used for connecting an accessory, such as a decontamination hose,
buddy breathing attachment or airline belt manifold.
Hose loops secure the dual pressure supply hose, or hose to the Bodyguard, and medium
pressure hose to the backplate and shoulder harness.
Available for twin cylinder fitting is a Twin Pack Kit. Contact Dräger for details.
The reducer-pneumatic arrangement, harness padding and adjusting straps easily remove
from equipment for maintenance and cleaning.
Refer to the appropriate Training Manual for details of the selected lung demand valve.
Operating Principle
Pneumatic Arrangement
Pressure Reducer
The high performance first stage pressure reducer is factory set and sealed and must not be
tampered with as this invalidates the conditions of warranty of Dräger Limited.
Service life of the reducer is six years. Dräger recommend that at the end of this period the
reducer be returned to Dräger Service, Blyth, for overhaul under the Repair Exchange Scheme
(REX).
Functioning at inlet pressures of 200bar or 300bar, (300bar only with ChargAirTM), the
reducer provides a controlled outlet medium pressure to the lung demand valve.
PSS® 100
Pneumatic Arrangement
A preset and sealed mechanism between the two outlet ports of the reducer controls the
activation of the whistle warning unit. The mechanism activates when the available air remaining
in compressed air cylinder has reduced to a preset pressure.
As pressure increases in the medium pressure chamber it forces the piston to move axially
against the compression spring until the piston seat contacts the end face of plunger - sealing
off high pressure.
On inhalation, from lung demand valve, the air flows from the medium pressure chamber and
as pressure reduces, the compression spring retracts the piston moving the piston seat from
the end face of the plunger. (Fig. 2.)
The high pressure air then begins to enter the medium pressure chamber repeating the cycle.
Consumption of air from the cylinder(s) reduces the pressure and at a pre-set pressure the
compression spring in the whistle activation mechanism retracts the high pressure plunger
from its sealing seat releasing medium pressure air from the medium pressure chamber into
the whistle activation chamber. (Fig. 3.) From the activation chamber a controlled airflow
enters the bore of the outer rubber hose of the dual pressure hose assembly operating the
whistle warning unit.
Pressure Reducer
High
Pressure
Medium
Pressure
Fig. 1
0660
High
Pressure
Medium
Pressure
Fig. 2
0661
High
Pressure
Medium
Pressure
L.P. to Whistle
H.P. to Gauge
Fig. 3
0662
Reducer Failure
Pressure Relief Valve - ‘Open’
High
Pressure
Medium
Pressure
Fig. 4
0664
The dual pressure hose supplies medium pressure air from the reducer to the in line whistle
warning unit and high pressure air to the cylinder contents gauge.
The hose construction although appearing similar to medium pressure hoses, differs in its
internal construction. Inside the outer rubber hose assembly is a fine bore copper capillary
tube that is coil wound around a multi-strand length of phosphor bronze wire. The capillary
tube and wire are then brazed to each of the end fittings of the capillary tube assembly. High
pressure air flows through the bore of the capillary tube to the pressure gauge and not
through the rubber hose.
At a pre-set pressure the whistle activation mechanism releases medium pressure through
the secondary end fitting, into bore of outer rubber hose assembly then into the whistle tube
operating the whistle warning unit located at base of gauge.
Constructed as part of the dual pressure hose assembly, and located at base of pressure
gauge, the in-line whistle warning unit provides an audible alarm signal. The alarm begins
sounding when the available air remaining in the compressed air cylinder has reduced to a
pre-set pressure.
The wearer should regularly monitor the gauge pressure to determine the remaining available
air and leave the contaminated area Before activation of whistle, however on hearing the
alarm the wearer Must immediately leave the contaminated area for a safe environment.
The pre-set pressure setting of the Whistle Warning Unit can be adjusted by the user to meet
special requirements.
Operation
See descriptions of Pressure Reducer and Whistle Warning Unit, and Dual Pressure Assembly
The content gauge is a Bourdon Tube type constructed from a stainless steel pressed case
and having a 3mm thick lens window. The gauge conforms to the technical and accuracy
requirements of European Standard EN137.
The gauge dial is luminescent yellow having radial black graduation marks at 10bar intervals
from 0 to 350bar. At 50bar intervals graduation marks are red. A red radial arc highlights the
segment from 0 to 50bar.
Gauge
Gauge
Cover
Whistle
H.P. Capillary
Phosphor
Bronze
Wire
Fig. 5
0177
The Bodyguard has basic monitoring and alarm options programmed into the instrument
before despatch from Dräger.
Air passes into the piston bore of reducer, past the piston seat and pre-set plunger and into
the medium pressure chamber of the reducer.
As pressure increases in medium pressure chamber, it forces the piston to move axially against
the compression spring until the piston seat contacts the end face of the plunger - sealing off
high pressure.
On inhalation from the lung demand valve, air flows from the medium pressure chamber and
as pressure reduces the compression spring retracts the piston moving the piston seat from
the end face of the plunger. (Fig. 2). High pressure air begins to enter the medium pressure
chamber repeating the cycle.
Refer to Instructions for Use supplied with PSS 100 E and PSS 100 ET.
Pressure Reducer -
PSS® 100 E and PSS® 100 ET
Medium
Pressure
High
Pressure
Bodyguard High
Pressure Switch Pressure
Fig. 6
0665
The hose assembly of the Bodyguard electronic monitoring unit, although appearing similar
to the medium pressure hoses, provides a flexible protective sleeve encasing the electronic
wiring from the pressure sensor to the electronic monitoring unit.
Hose Assemblies
Medium Pressure Hoses - refer to Figure 9
Flexible and anti-kinking rubber hose lengths combine to provide medium pressure hose
assemblies for the different reducer-pneumatic arrangements and associated lung demand
valves.
Pressure relief valve spring, O-Ring retainer and O-Ring locate onto the nozzle end fitting of
the medium pressure hose that connects to medium pressure outlet port of the reducer.
The O-Ring of the relief valve also functions as a seal for the hose to reducer connection.
TM
ChargAir Hose (Optional) - refer to Figure 10 TM
Assembled to the second high pressure outlet port only of the 300bar reducer, the ChargAir
hose allows for the refilling of equipment cylinder(s) from an independent high pressure,
(300bar maximum), secondary breathing air supply source, while the wearer is breathing
from equipment.
TM
The ChargAir hose assembly is constructed from three main elements:
Refer to the appropriate Training Manual for details of the selected Lung Demand Valve.
PSS® 100 ET
Tally
Pressure
Sensor
Pressure
Sensor
Rubber Hose
Demand Valve
Cylinder refill
high pressure ‘in’
Section Description
5.1 Contents
5.3.1 General
To ensure correct operational condition of the equipment use only the cleaning and
disinfecting solutions recommended by Dräger. The use of any other product will
invalidate the Dräger warranty and guarantee.
Important Note:
If the equipment is fitted with an integral lung demand it is important to refer
also to the cleaning, disinfecting and drying section of the relevant training
manual.
When using baths to contain cleaning and disinfecting solutions the immersed
components and assemblies must only be agitated manually.
It is important to note that Dräger Do Not recommend the use of any form of
mechanical, electrical or ultrasonic agitation of this product. If the user decides
however to use ultrasonic cleaning baths then it is important that the instructions
supplied by the manufacturer of the bath are strictly followed.
Refer to the Training Manuals for any ancillary units, e.g. Lung Demand Valve.
Safety Note: Refer to manufacturers’ usage instructions when using cleaning and
disinfecting agents. It is important that attention be paid to concentration and
reaction times. Do Not use organic solvents such as Acetone, Alcohol, White
Spirit, Trichloroethylene or similar.
Dräger recommends:
1. Cleaning:
Dräger Safety Wash
1 Litre Safety Wash and Dispenser 3380164
1 Litre Safety Wash Refill 3380165
5 Litre Safety Wash and Dispenser 3380166
5 Litre Safety Wash Refill 3380167
Sekusept
4 Bottles @ 2 litres 7904071
2. Disinfecting:
Incidur
1 can 6 litres 7904072
1 can 30 litres 7904073
Follow these instructions to remove the pressure reducer and hose assembly from
the backplate for cleaning and disinfecting.
Note: Do Not immerse pressure reducer or pressure sensor of electronic monitoring unit
(Bodyguard) in cleaning or disinfecting solutions. Use clean cloth, moistened in cleaning
fluid, to remove dirt/contaminants.
Tools required:
10mm AF Spanner 3336309
5mm Hexagon socket key 3335738
Refer to Figure 1.
Mounting Pin
Remove split pin (3) from the end
of reducer mounting pin (4). Push
and remove the mounting pin from 3
mounting block.
6 Re-assemble pressure reducer and hose assembly to backplate, taking care not to
damage sleeve (5). Reverse above procedure.
Refer to Figure 2.
Refer to Figure 3.
Tools Required:
2mm AF Hexagon Socket Key 3337900
1 Unfasten hose loops of the backplate and if necessary from shoulder strap to release
hose from equipment.
Refer to Figure 4.
Refer to Figure 5
2 Check that the reducer outlet bore is clean and not damaged then insert hose end
fitting into the outlet port. Hold against the compression spring then insert retention
staple (3) through holes in reducer body ensuring correct location into groove in hose
end fitting. Check hose securely retained by pulling hose away from reducer body.
3 Locate groove of retaining cap (2) over the protruding portion of the retention staple
- align screw holes then insert screws (1) and using 2mm AF hexagon socket key
tighten the screws.
4 Connect cylinder, and if necessary the lung demand valve, to the pressure reducer
and apply pressure to the equipment. Close the cylinder valve and carry out Leak Test.
5 Following satisfactory Leak Test secure the hose to backplate, and if necessary to
the shoulder strap, using the hose loop.
Safety Note: Vent air pressure from pneumatic system before proceeding with disassembly.
Close cylinder valve (if fitted) and vent system.
Refer to Figure 6.
2
2 Using the 2.0mm hexagon adaptor 3
and torque key, remove screw (1).
1
Unfold the rubber cover (2) from
around the gauge then slide cover
along the hose. To prevent rotation
of the whistle tube during removal
of the gauge immediately re-insert
the screw (1) into the threaded radial
hole of the whistle tube.
Note: End of set screw should not protrude past the outer diameter of the whistle tube.
8 Before the adhesive sets connect cylinder, and if necessary the lung demand valve,
to the pressure reducer and apply pressure to the equipment. Close the cylinder
valve and carry out Leak Test. Following satisfactory leak test immediately begin to
slowly vent pressure until the whistle activates.
Safety Note: Before proceeding with the replacement of hose, turn ‘OFF’ the cylinder
valve, vent pressure from system then unscrew reducer handwheel from cylinder
valve and remove cylinder from backplate.
1 Unfasten hose loops of backplate and left shoulder strap to release hose from
equipment.
Refer to Figure 9.
Fig. 9
0170
Safety Note: Before proceeding with replacement of filter, turn ‘Off’ the cylinder valve,
vent pressure from system, unscrew the reducer handwheel from the cylinder
valve and remove the cylinder from the backplate.
Refer to Fig. 11
1
1 Unfasten hose loops of backplate 2
and left shoulder strap to release
hose from equipment.
3
2 Using 2mm AF hexagon socket key
unscrew (anti-clockwise) and 4
remove two screws (1), remove
retaining cap (2), and retention
staple (3). Grip and pull hose (4) to
remove from outlet port of reducer.
Note: Check hose securely retained by pulling hose away from reducer body.
Safety Note: Before proceeding with the removal of hose from reducer - turn ‘Off’ cylinder
valve, vent pressure from system then unscrew the reducer handwheel from the
cylinder valve and remove the cylinder from backplate.
Fig. 13
0170
Refer to Figure 16
1 4
Fig. 17
0191
Tools Required:
2mm AF Hexagon Socket Key 3337900
1.5 AF Hexagon Socket Key 3331237
Hexagon Adaptor (2mm) 3310600
Torque Key 3310593
Loctite 542 3333622
O-Ring Assembly Tool 3337901
Important Note: Check all components are clean and undamaged. Replace with new
components where necessary.
Refer to Figure 20
1 If necessary attach the lung demand valve to the equipment. Activate the positive
pressure mechanism (lever or button) to switch ‘Off’ positive pressure.
2 Open cylinder valve slowly, but fully, to pressurise pneumatic system. Check reading
of pressure gauge. Close cylinder valve and vent equipment slowly as follows and
note pressure at which whistle activates;
Positive Pressure Lung Demand Valves - Cover outlet port of demand valve with
palm of hand. Press the centre of the rubber cover of the demand valve and slowly
vent equipment by carefully lifting palm of hand to maintain slow steady decrease in
pressure.
Normal Demand Lung Demand Valves - Slowly vent equipment by carefully pressing
the centre of the rubber cover of the demand valve maintaining slow steady decrease
in pressure.
Fig. 23
0182
Important Note: Do Not turn adjuster unless the pressure is vented from the system.
l rotate the adjusting cap clockwise to increase the pressure at which whistle
activates.
l rotate the adjusting cap anti-clockwise to decrease pressure at which whistle
activates.
Note: As a guide - one quarter turn of the adjusting cap is equivalent to a pressure change of
approximately 2bar to 3bar. Do Not turn adjusting cap with the pressure ‘On’.
4 Repeat instructions 2 to 3.
5 If necessary repeat instructions until the required whistle activation setting is achieved.
Safety Note: Before proceeding with the removal of the Bodyguard hose from the reducer,
turn ‘Off’ the cylinder valve, vent pressure from system then unscrew the reducer
handwheel from cylinder valve and remove the cylinder from backplate.
1 Unfasten hose loops of the backplate and left shoulder strap to release the Bodyguard
hose from the equipment.
Fig. 24
0170
3 Locate groove of retaining cap (2) over the protruding portion of the retention staple
- align screw holes then insert screws (1) and using 2mm AF hexagon socket key
tighten the screws.
4 Connect cylinder, and if necessary the lung demand valve, to the pressure reducer
and apply pressure to the equipment. Close the cylinder valve
2 1
5 4
Fig. 26
0525
Note: Close cylinder valve and vent air from system. Open hose loops of shoulder straps to
release hoses. If necessary, remove cylinder.
ê
Fig. 27
0764
Fig. 28
0748
Fig. 29
0749/0766
Fig. 30
0765
1 2 4
Fig. 31
0767
1 2 4
Fig. 32
0767
è
è
ç
Fig. 35
0748
Fig. 38
0656
Fig. 39
0656
Fig. 41
0501
Fig. 42
0750
2 Slide pivot plate along slot checking the security of the assembly and for unrestricted
freedom of movement.
3 Orient correctly and if necessary reassemble the waistbelt and waistbelt pad.
See Section 5.15.1.
Fig. 43
0774
Fig. 44
0772
Fig. 45
0773
2
Fig. 46
0771
Fig. 47
0774
2 Remove the rubber shock buffer from the opening at the base of the yoke. Turn the
yoke adjuster through 90 degrees and slide the yoke to opening and align the adjuster
with the opening - remove the yoke.
1 Fig. 48
1022
1 2 3 4 5
Fig. 49
1192
Fig. 52
0501
Section Contents
6.3.3 High Pressure Leak and Whistle Warning Unit Test - Bodyguard E and ET
For detailed instructions of the following additional Functional Tests, refer to the
appropriate Test Equipment Manual, (i.e. Test Box, A900 Test Unit, Quaestor).
Type N
l Balanced Piston Leak Test
l Demand Valve Opening Pressure
Note: Check the Test Date, stamped on the cylinder by last the Testing Station. If past the
Test Date Do Not use cylinder.
1 If cylinder assembled to backplate, close cylinder valve and vent pressure from the
pneumatic system. Unhook anti-vibration strap (if fitted) from the reducer handwheel
then loosen and extend cylinder strap. Unscrew reducer handwheel from cylinder
valve and rotate cylinder through 90° positioning cylinder towards top of backplate
until valve is approximately 10cm from reducer handwheel.
2 Screw handwheel of high pressure test gauge into outlet port of cylinder valve then
close relief valve on test gauge. Open cylinder valve slowly, and fully. Check reading
on test gauge.
Test Parameters
The cylinder contents pressure must be at least at 80% of maximum working pressure
(indicated on cylinder label) before attempting to perform functional tests of
equipment. If pressure is low then recharge cylinder.
3 Close cylinder valve then open the relief valve of the high pressure test gauge to vent
pressure. Remove high pressure test gauge from cylinder valve.
4 Re-assemble the cylinder to the reducer. Hook anti-vibration strap (if fitted) to the
reducer handwheel. Tighten cylinder strap.
1 Activate the reset function of positive pressure demand valves to switch ‘Off’ positive
pressure. Open the cylinder valve slowly, but fully, to pressurise the pneumatic system.
Check pressure reading on pressure gauge.
Note: PA90LDVs: Do Not press centre of rubber cover while pressing reset lever. If an
immediate leak indicated from demand valve - press centre of rubber cover to release
positive pressure mechanism - press reset lever/button to lift positive pressure
mechanism. Repeat action two or three times to eliminate leak.
2 Close the cylinder valve and again observe the reading on the pressure gauge.
Test Parameter
Gauge reading shall not decrease more than 10bar in one minute.
If equipment ‘Fails’ to conform to the test parameter, vent pressure from the system,
investigate source of the leak, rectify and repeat the test. Until the leak is rectified
and the equipment conforms to the requirement of the test parameter, Do Not carry
out any further test.
4 Positive Pressure Lung Demand Valves - Cover outlet port of the demand valve
with palm of hand. Press the centre of the rubber cover of the demand valve to
activate positive pressure. Slowly vent equipment by carefully lifting the palm of the
hand to maintain a slow and steady decrease in pressure.
Normal Demand Lung Demand Valves - Slowly vent equipment by carefully pressing
the centre of rubber cover of the demand valve maintaining a slow and steady decrease
in pressure.
As pressure decreases, observe the pressure gauge and note pressure at which
whistle begins to sound.
Test Parameter
The whistle warning unit should activate an audible signal at the pre-set pressure of
55bar +/- 5bar (60bar to 50bar).
1 Activate the reset function of positive pressure demand valves to switch ‘Off’ positive
pressure. Cylinder valve ‘Closed’.
Note: PA90LDVs: Do Not press centre of rubber cover while pressing reset lever. If an
immediate leak indicated from demand valve - press centre of rubber cover to release
positive pressure mechanism - press reset lever/button to lift positive pressure
mechanism. Repeat action two or three times to eliminate leak.
3 When the leak test icon (Fig. 1) is displayed - press the LH button. Bar graph
segments sweep to the right and the ‘open valve’ icon will be displayed. See Fig. 2.
General Note: If during the self check sequence the LH button is not pressed when the
leak test icon (Fig. 1) is displayed then wait till end of sequence when ‘0’ bar is
displayed. Press and hold down LH button until reverse arrow icon displayed (‘a’
Fig. 6) then release button - Bodyguard will then restart the self check sequence.
Repeat from 3.
4 When the ‘open valve’ icon is displayed ‘Open’ the cylinder valve fully to pressurise
system.
6 As soon as the valve is fully open then ‘Immediately’ begin to ‘close’ the cylinder
valve. Bar segments will sweep to the right and when only two or one segments
remain then ‘press’ the RH button to start the leak test timing sequence. Display will
change to Fig. 4.
Important Note: During the leak test sequence the pressure reading may reduce by up to
10bar. This may be due to settling of the pressure reducer.
7 During the leak test timing sequence the bar graph segments will sweep to the right
and at the end of the timing sequence either:
‘Passed’ icon (a) displayed (Fig. 5) - equipment ‘Passed’ test. Proceed to instruction
8.
or
‘Failed’ icon (b) displayed (Fig. 5) - equipment ‘Failed’ test. See Note below.
Note: If ‘Failed’ icon (b) is displayed during leak test timing sequence then to determine
whether the failure is due to a slow to settle reducer or an high pressure leak - Do Not
vent system of pressure but ‘immediately’ Go to ‘Failed Test’ - Optional Procedure -
Section 6.3.4.)
9 Venting Procedure
Positive Pressure Lung Demand Valves - Cover the outlet port of the demand valve
with palm of hand. Press centre of the rubber cover of the demand valve to activate
positive pressure. Slowly vent equipment by carefully lifting palm of the hand to
maintain a slow and steady decrease in pressure.
Normal Demand Lung Demand Valves - Slowly vent equipment by carefully pressing
centre of the rubber cover of the demand valve maintaining a slow and steady decrease
in pressure.
10 During venting observe the display. The whistle alarm signal (rapid ‘Bleep’) should
become audible at the pre-set pressure level (55bar +/- 5bar - 60bar to 50bar) and
continue down to 10bar. Continue slowly venting until the pressure and time displays
indicate ‘0’ bar.
Note: Using a ‘snap on’ I.R. Link and Widows based Software Package Option, the pressure
setting can be adjusted. Contact Dräger for details.
Fig. 1 Fig. 4
0475 0483
Fig. 2 Fig. 5
0481 0484
Fig. 3 Fig. 6
0485 0591
Normal Demand Lung Demand Valves - Slowly vent equipment by carefully pressing
centre of rubber cover of the demand valve maintaining a slow and steady decrease in
pressure.
6 During venting observe the display - the whistle alarm signal (rapid ‘Bleep’) should
become audible at the pre-set pressure level (see Test Parameters - Section 6.4)
and continue down to 10bar. Continue slowly venting until pressure and time displays
indicate ‘0’bar.
1 Activate the reset function of positive pressure demand valves to switch ‘Off’ positive
pressure. Open the cylinder valve slowly but fully to pressurise the pneumatic system.
Note: PA90LDVs: Do Not press centre of rubber cover while pressing reset lever. If an
immediate leak indicated from demand valve - press centre of rubber cover to release
positive pressure mechanism - press reset lever/button to lift positive pressure
mechanism. Repeat action two or three times to eliminate leak.
2 Press centre of the rubber cover of the demand valve. On positive pressure types
this makes positive pressure active. See Safety Note.
Test Parameter
An unobstructed air flow should vent to atmosphere from the outlet of the demand
valve. See Safety Note. This test should take a minimum of 3 - 5 seconds.
3 Positive Pressure Lung Demand Valves - To switch ‘off’ air flow and positive pressure
- activate the positive pressure ‘Off’ mechanism.
4 Close the cylinder valve then vent pressure from the system.
Note: Certified equipment incorporate a pre-set and sealed pressure reducer assembly. The
Dräger guarantee is void should original seal caps be tampered with, removed, or
broken. The guarantee of correct operational condition of the pressure reducer is only
valid if Dräger service, and re-seals the reducer.
1 Detach demand valve from the quick release coupling of the medium pressure hose.
Open the cylinder valve slowly and fully.
2 Insert the test gauge into female quick release coupling of the medium pressure
hose from the reducer. Check the pressure reading on the test gauge.
Test Parameters
At a system inlet pressure (cylinder pressure) of >100bar the test gauge reading
should indicate a medium pressure of between 6bar to 9bar after 5 seconds. After
one minute it is acceptable for the medium pressure to rise to 9.7bar.
4 Following the test ‘Close’ the cylinder valve and disconnect the test gauge from the
quick release coupling.
5 Activate the reset function of positive pressure demand valves to switch ‘Off’ positive
pressure and connect demand valve to the female quick release coupling. Press the
centre of rubber cover of demand valve venting pressure from the system. Following
venting activate the reset function of positive pressure demand valves to switch ‘Off’
positive pressure.
Note: If reducer ‘fails’ the medium pressure test then return the reducer, with hoses and any
accessories removed, to Dräger Service for overhaul under the Repair Exchange Scheme
(REX). Details of REX Scheme are available on request.
Refer to Figure 8.
Refer to Figure 9.
5 4 3 1 2
Fig. 9
0307
Note: Check that the manifold is securely retained by pulling manifold away from reducer body.
7 Locate groove of retaining cap (4) over protruding portion of retention staple - align
screw holes - insert screws (5) and using 2mm socket key tighten the screws.
8 Activate the reset function of positive pressure demand valves to switch ‘Off’ positive
pressure then ‘open’ cylinder valve slowly, and fully.
9 Insert test gauge into female quick release coupling on medium pressure test hose
from the reducer. Check the pressure reading on the test gauge.
Test Parameters
At a system inlet pressure (cylinder pressure) of >100bar the test gauge reading
should indicate a medium pressure of between 6bar to 9bar after 5 seconds. After
one minute it is acceptable for the medium pressure to rise to 9.7bar.
9 Following the test ‘Close’ the cylinder valve. Press centre of the rubber cover of the
demand valve venting pressure from system. Following venting activate the reset
function of positive pressure demand valves to switch ‘Off’ positive pressure.
11 Remove test hose from reducer port - remove demand valve hose from manifold of
test hose, and reassemble demand valve hose to the outlet port of reducer. Carry
out High Pressure Leak Test.
Note: If reducer ‘fails’ medium pressure test then return the reducer, (hoses and any accessories
TM
removed, except ChargAir hose), to Draeger Service for overhaul under the Repair
TM
ExchangeTMScheme (REX). Reducers (300bar) with ChargAir must be returned with
ChargAir hose fitted. Details of REX Scheme are available on request.
TM
6.3.7 High Pressure Leak Test - ChargAir
TM TM
ChargAir Filling Hose and Pressure Reducer with ChargAir
Tools Required
Filling Hose Assembly with Gauge 3336641
Second Fully Charged Valved Cylinder Assembly (300bar)
Note: If the equipment has a quick release coupling on the medium pressure hose then
attach demand valve to the female coupling.
TM
1 Screw the handwheel of the ChargAir Filling Hose into the cylinder valve of the
second (fully charged) valved cylinder assembly (300bar). Open slowly, but fully,
the cylinder valve to pressurise the Filling Hose. Check pressure reading on pressure
gauge.
2 Close cylinder valve of second valved cylinder assembly and observe pressure gauge.
Test Parameter
Gauge reading shall not decrease more than 10bar in one minute.
Note: If Filling Hose fails to conform to instruction 3, investigate source of leak, rectify and
repeat test. Until leak is rectified, Do Not carry out any further test.
3 Carry out High Pressure leak test of equipment as detailed in Section 6.3.1.
4 Close cylinder valve of equipment then activate the reset function of positive pressure
demand valves to switch ‘Off’ positive pressure.
5 Remove protective
TM
caps from male
ChargAir coupling TM of equipment
and female ChargAir coupling of
filling hose. Push male and female
couplings together until a loud ‘click’
is heard indicating a secure
connection.
2
Note: A ‘dark green’ ring (1) visible behind
the ‘grey’ sleeve (2) of the female 1
coupling indicates a secure
connection. See Fig. 10.
7 Close cylinder valve of second valved cylinder assembly and observe the pressure
gauge of the filling hose and the pressure gauge of the equipment.
Test Parameter
Gauge reading shall not decrease more than 10bar in one minute.
9 Check that the cylinder valve of the second valved cylinder assembly is ‘closed’.
TM
Disconnect ChargAir male and female coupling by pulling ‘grey’ sleeve of female
coupling away from male coupling.
Note: As couplings separate, a ‘hiss’ or ‘pop’ may be heard as non-return valves in both
couplings seal ‘off’ high pressure.
10 Assemble protective cap of filling hose to female coupling; press centre of cap to
vent pressure. Unscrew filling hose handwheel from cylinder valve of second valved
cylinder assembly. Press centre of the rubber cover of the demand valve, venting
pressure from equipment. Reset demand valve to positive pressure ‘Off’ position.
11 Test completed.
7 Fault Location
Section Contents
7 Fault Location
7.4 Facepiece
The Dräger guarantee is void should original seal caps be tampered with, removed,
or broken. Guarantee of correct operational condition of the pressure reducer is
only valid if Dräger service, and re-seals the reducer. Return faulty units (hoses and
any accessories removed) to Dräger Service, Blyth.
Service life is six years. Dräger recommends that at the end of this period the reducer
(hoses and any accessories detached) be returned to Dräger Service, Blyth, for
overhaul under the Repair Exchange Scheme (REX). Details of REX Scheme are
available on request.
Leak from pressure relief Reducer failure Remove hoses from reducer
valve of medium pressure and return reducer to
hose. Relief valve Dräger. (REX)
components O.K.
7.4 Facepiece
For detailed instructions for Inspection, Testing and Servicing, refer to the appropriate
Facepiece Service/Training Manual.
Whistle does not 1 Blocked whistle tube flute Clean or replace whistle
sound and/or jet tube
2 Faulty activation mechanism Remove hoses from
reducer and return
reducer to Dräger (REX)
3 Blocked filter/filters Replace filter/filters
Section 8 Accessories
Section Description
8.1 Contents
The hose configurations provide a connection for the equipment existing lung demand
valve plus a second connection, fitted with either a male or female quick release
coupling.
1. Secondary air supply hose fitted with female quick release coupling.
(Decontamination Procedure).
2. Secondary air supply hose fitted with male quick release coupling.
(Decontamination Procedure).
Safety Warning: When using Secondary Hose Assembly with a secondary air supply hose,
for Decontamination Procedure, inlet air pressure must not exceed 10bar. Do
Not connect high pressure air to female coupling.
0225/0226
Refer to Figure 1.
Refer to Figure 2. 3
3 Using 2mm socket key, unscrew
(anti-clockwise) and remove two
screws (1) from accessory hose 4
manifold, remove retaining cap (2)
and retention staple (3). Insert end
fitting of medium pressure hose into
manifold port - push and hold against
compression spring - insert retention Fig. 2
staple (3) through holes in manifold. 0306
Fully insert retention staple ensuring
correct location into groove in hose
end fitting.
Note: Check hose securely retained by pulling hose away from manifold.
4 Locate groove of retaining cap (2) over protruding portion of retention staple - align
holes - insert both screws (1) and using 2mm socket key, tighten the screws.
Refer to Figure 3.
Fig. 4
0622
Order List
Secondary Hose 3337650
Decontamination Hose 3337652
This configuration provides a connection for equipment existing lung demand valve
plus two additional connections:
Refer to Fig. 5
On failure, interruption or turning ‘off’ of independent air supply, the wearer opens
the equipment cylinder valve, disconnects independent air supply from belt manifold
and either, a. continues working using available air from equipment cylinder, or b.
goes on to a safe area.
Safety Warning: When using independent source of air the inlet pressure must not exceed
10bar. Duration of equipment commences from time of opening cylinder valve
and disconnecting airline supply. Time remaining from equipment is dependent
on volume of air remaining in cylinder.
Technical Data
Airline Operating Pressures and Air Quality
Independent source of air should conform to the following parameters when using
compressed air respiratory protection equipment fitted with the Belt Manifold
Accessory.
Safety Note: Check independent air supply complies with air quality, pressure, and flow
requirements (see Technical Data).
Tools Required
2mm AF Socket Key 3337900
Posi-Driv Screwdriver (No.2) 3331245
Refer to Figure 6.
Refer to Figure 7
Refer to Fig. 8
Refer to Figure 9.
9 Locate groove of retaining cap (4) over protruding portion of retention staple - align
screw holes - insert both screws (5) and using 2mm socket key, tighten screws.
10 Assemble compressed air cylinder to equipment and carry out Leak Test as detailed
in relevant Instructions for Use.
11 Following satisfactory Leak Test, carry out the following Flow Test.
13 Connect female quick release coupling, on hose from independent air supply, to
male coupling on belt manifold. Turn ‘On’ independent air supply.
Note: If required connect second man equipment to female coupling of belt manifold.
14 Press centre of rubber cover of lung demand valve/s. See Safety Warning.
Note: Keep the centre of rubber cover of normal demand, N type lung demand valve,
depressed. Positive pressure types will activate to positive pressure. An unobstructed
air flow should vent from outlet of the demand valve. This test should take a minimum
of 3 to 5 seconds.
Safety Warning: Do Not direct air flow from outlet of demand valve onto face, eyes or
skin.
16 Turn ‘Off’ the independent air supply and disconnect the supply hose from the male
coupling of the belt manifold.
Note: If second man equipment attached then remove from female coupling of belt manifold.
Order Code
Belt Manifold Assembly 3337600
Tool List
Number Title
1287.3 Tool List
Index
Introduction
The PA90 Series of Lung Demand Valves are approved for use with approved Dräger
Compressed Air Breathing Equipment (C.A.B.E.) and Facepieces.
Refer to Illustration
The combination of Lung Demand Valve, C.A.B.E. and Facepiece provides the wearer of the
equipment with respiratory protection when working in or escaping from a contaminated or
oxygen deficient life-threatening environment.
Safety Warnings
l Explain - Instructions for Use, supplied with the equipment must be adhered to.
Contact Dräger Limited, Blyth for details of Service Contracts and Service Training
Courses.
l Train potential wearers in how to wear and use equipment and to understand its
limitations.
1241
Product Description
The variants of the PA90 Series Lung Demand Valve are as follows:
The high performance Lung Demand Valves incorporate a precision factory set and sealed
balanced piston unit that has unlimited service life.
The Dräger Warranty and Guarantee is void should any attempt be made by customers to
make adjustment to the balanced piston unit.
Detachable variants of the lung demand valves have a medium pressure hose with a quick
release male coupling to connect the demand valve to the female coupling of the medium
pressure hose from the pressure reducer of the appropriate C.A.B.E.
Fixed variants of the lung demand valves incorporate a medium pressure hose having an end
fitting design for connecting directly to the medium pressure outlet port of the pressure
reducer of the appropriate C.A.B.E.
The housing of the lung demand valves are manufactured from a moulded composite glass
filled material that provides a high degree of strength and durability.
Positive pressure types have a manually operated re-set lever to switch ‘OFF’ the positive
pressure function - this must be pressed before opening the valve of compressed air cylinder.
First breath inhalation by the wearer will switch ‘ON’ the positive pressure.
All variants include a supplementary air supply facility that is made active by pressing the
centre of the protective rubber cover at the front of the lung demand valve. Lung demand
valves fitted to short duration C.A.B.E. and escape units will not normally include a
supplementary air supply facility.
Section Description
4.1 Positive Pressure Lung Demand Valves A and AE
Before turning ‘On’ the high pressure air supply the positive pressure mechanism
must be switched ‘Off’ by pressing the re-set lever on the side of the lung demand
valve to lift and lock the main pivot lever and diaphragm.
On opening the cylinder valve, medium pressure air flows from the reducer through
a hose and swivel banjo then into main bore of balanced piston block.
Increasing air pressure moves the balanced piston axially, forcing the piston cone
against the O-Ring seal located in the block - sealing ‘Off’ the airflow from outlet of
lung demand valve.
Refer to Figure 2.
When the wearer begins to inhale this creates a negative pressure inside the facepiece
and lung demand valve. Due to the negative pressure the ambient air pressure acts
against the diaphragm. As the diaphragm deflects it then presses against the main
hinged lever of the balanced piston unit resulting in the release of the positive pressure
‘Off’ mechanism.
Increasing deflection of the diaphragm, during inhalation, pivots the balanced piston
lever forcing it against the end cap of the piston - moving the piston axially and lifting
the piston cone from its seating and releasing air from the lung demand valve into
wearer’s facepiece. Two extension springs attached to the balanced piston block
and the main lever act on the balanced piston, lifting the sealing cone and maintaining
a positive pressure (above ambient) inside the facepiece.
When the wearer stops inhaling then begins to exhale, the diaphragm, pivot lever
and sealing cone retract.
Balanced Piston
‘Closed’
Medium Pressure
(M.P.)
Fig. 1
0008
Balanced Piston
‘Open’
M.P. Flow
to Facepiece
Fig. 2
0078
On opening cylinder valve, medium pressure air flows from the reducer, through a hose
and swivel banjo then into main bore of balanced piston block.
Increasing air pressure moves the balanced piston axially, forcing the piston cone against
the O-Ring seal located in the block - sealing ‘Off’ the airflow from outlet of lung
demand valve.
Refer to Figure 4
When the wearer begins to inhale this creates a negative pressure inside the facepiece
and lung demand valve. Due to the negative pressure the ambient air pressure acts
against the diaphragm then the main hinged lever of the balanced piston unit.
Increasing deflection of the diaphragm, during inhalation, pivots the balanced piston
lever forcing it against the end cap of the piston - moving the piston axially and lifting
the piston cone from its seating and releasing air from the lung demand valve into
wearer’s facepiece.
When the wearer stops inhaling then begins to exhale, the diaphragm, pivot lever and
sealing cone retract.
Balanced Piston
‘Closed’
Medium Pressure
(M.P.)
Fig. 3
0249
Balanced Piston
‘Open’
M.P. Flow
to Facepiece
Fig. 4
0250
Section Description
5.1 Service and Test Intervals
Important Note: This table is intended as a general guide. Reference must also be made
to the Inspection, Service and Test Intervals specified in the appropriate section
of the relevant C.A.B.E. Training manual.
To ensure correct operational condition of the equipment use the cleaning and
disinfecting solutions recommended by Dräger. The use of any other product will
invalidate the Dräger warranty and guarantee.
When using baths to contain cleaning and disinfecting solutions the immersed
components and assemblies must only be agitated manually.
It is important to note that Dräger Do Not recommend the use of any form of
mechanical, electrical or ultrasonic agitation of this product. If the user decides
however to use ultrasonic cleaning baths then it is important that the instructions
supplied by the manufacturer of the bath are strictly followed.
Important Notes:
1. The diaphragm of the lung demand valves must not be cleaned using ultrasonic
techniques, as this will have a detrimental affect on the centre-bonded area of
the diaphragm. Clean and disinfect the diaphragm manually.
3. Immersing the lung demand valve and manual agitation will also eventually remove
lubricants although of course this will occur over a much longer period. Following
cleaning, disinfecting and drying it is recommended that functional tests be
carried out. Re-lubricate moving parts as required.
Safety Note: When using cleaning and disinfecting agents refer to the relevant
manufacturers’ usage instructions. It is important that attention be paid to
concentration and reaction times. Do Not use organic solvents such as Acetone,
Alcohol, White Spirit, Trichloroethylene or similar.
Dräger recommends:
1. Cleaning:
Safety Wash
1 Litre Dispenser 3380164
1 Litre Refill 3380165
5 Litre Dispenser 3380166
5 Litre Refill 3380167
Sekusept
4 bottles @ 2 litres 7904071
2. Disinfecting:
Wipex Cloths (Pack of 50) 3380375
Note: A Detach demand valves (fitted with male coupling) from female coupling of medium
pressure hose of the C.A.B.E.
Note: B If necessary, demand valves attached directly to the reducer should be removed at the
hose connection to the reducer. For details of removal refer to the appropriate section
of the relevant C.A.B.E. Training Manual.
Tools required:
Plate Spanner R26817
1 If required, detach the demand valve
from the equipment.
û
2 Press the reset lever of positive
pressure type demand valves. Do Not
press centre of rubber cover while
pressing the reset lever or force lever
against the stop. See Fig. 1.
Refer to Figure 1.
(PA90A Illustrated.)
ü Fig. 1
0032/0251
4 Re-assemble the rubber cover (1) over the front of the lung demand valve to protect
balanced piston mechanism. Immerse and manually agitate the demand valve,
diaphragm and bayonet cap into cleaning fluid. (Refer to cleaning fluid instructions.)
5 After cleaning, rinse the components in clean water then immerse and manually
agitate the demand valve, diaphragm and bayonet cap into disinfecting fluid. (Refer
to disinfecting fluid instructions.) After disinfecting, remove rubber cover and rinse
components in clean water, taking care not to damage balanced piston lever
mechanism.
6 Press reset lever of the positive pressure demand valves to close balanced piston.
7 Connect the medium pressure hose of the demand valve to medium pressure air
supply. Turn ‘ON’ air supply then gently press the balanced piston lever to blow out
fluid residue.
8 Press reset lever of the positive pressure demand valves to close the balanced piston.
Turn ‘OFF’ and vent the air supply then disconnect medium pressure hose of demand
valve from air supply. Thoroughly ‘dry’ components.
9 After drying insert the balanced piston lever into hole in centre boss of the diaphragm
(chamfered side of boss towards lever) then carefully locate outer bead of diaphragm
into groove in demand valve body. See Fig. 3 for correct orientation of the diaphragm.
10 Moisten lip of the bayonet cap with clean water - refit bayonet cap to the lung and
using plate spanner R26817 tighten the cap to the demand valve.
Refer to Figure 6.
1 2
2 Using 14mm spanner hold the banjo
(1) across flats on banjo flange and
using the 11mm spanner unscrew
the elbow (2) of the hose from the
banjo. Inspect the hose and if
defective, scrap and replace.
Fig. 6
0040
Refer to Figure 7.
1 Inspect the threaded bore of the banjo and if necessary clean and remove all traces
of adhesive from the thread. Check that the O ring seal of the new hose is not
damaged and is fitted correctly to the elbow. If necessary replace the O ring.
2 Screw the elbow of the new hose into the banjo by one revolution - apply one drop
of Loctite 638 to thread of the elbow then finally screw the elbow into the banjo.
Using 14mm spanner, hold the banjo across flats of the banjo flange and using the
11mm spanner tighten elbow to the banjo.
Note: Do Not pressurise pneumatic system for 30 to 40 minutes to allow Loctite to set. Fixed
variants of the lung demand valve should be fitted directly to the medium pressure
outlet port of the pressure reducer. For details of fitting of hose to the reducer refer to
the appropriate section of the relevant C.A.B.E. Training Manual.
Replace with new factory set, or Repair Exchange balanced piston unit should
problems occur (other than a requirement to replace springs on positive pressure
type units).
Balanced Piston Units will only be accepted for Repair Exchange after disassembly
from demand valve, i.e. Balanced Piston Assembly only.
For details of the Terms and Conditions of the Repair Exchange Scheme contact
Dräger Service - Blyth.
3 5 4 2 1
Fig. 11
0028
Fig. 12
0041/0127
Push-in Type - A
Unscrew push-in type connector (anti-clockwise) from demand valve body and remove
anti-injector pipe (3).
Screw-in Type - AE
Use connector tool (3310497) to unscrew screw type connector and handwheel
assembly from demand valve body. Remove anti-injector pipe.
Screw-in Type - N
Screw and tighten M40 handwheel to connector tool 3310496. Turn handwheel and
connector tool anti-clockwise to unscrew screw type connector and handwheel
assembly from demand valve body. Remove anti-injector pipe.
Note: On positive pressure demand valves Do Not detach reset lever (4). If lever becomes
detached then reassemble lever to the demand valve as detailed in Section 5.5.1.1.
4 Rotate the lever anticlockwise, carefully lifting the stop of the lever to the opposite
side of the rib on the body of demand valve. At this position press the lever onto the
positive pressure arm.
5 Check assembly and carefully press lever (3) against spring tension - lever should
return freely under spring load.
2
Fig. 19
0046
6 Assemble connector.
Push-in Type - A
Assemble new O-Ring (2) Fig. 20
to connector. Apply Loctite 245 to
connector thread and screw
connector (3) into demand valve
body. Tighten until anti-injector pipe
is secure. Do Not overtighten.
Screw-in Type - AE
Check handwheel freely rotates on
connector handwheel assembly and
that retaining circlip is secure.
Assemble new O-Ring to connector
handwheel assembly. Apply Loctite Fig. 20
245 to connector thread. Using tool
0041/0127
(3310497), screw connector
handwheel assembly into demand
valve body until anti-injector pipe is
secure. Do Not overtighten.
Screw-in Type - N
Check handwheel freely rotates on connector handwheel assembly and that retaining
circlip is secure. Assemble new O-Ring to connector handwheel assembly. Apply
Loctite 245 to connector thread. Screw and tighten handwheel into tool 3310496
and screw connector handwheel assembly into demand valve body until anti-injector
pipe is secure. Do Not overtighten.
8 Carefully turn connector anti-clockwise till first space, in castellations on end face of
connector, is in line with retaining screw hole. Fully tighten the retaining screw to
secure the connector to the demand valve body making sure the screw enters space
in the castellation.
Note: Remove assembly tools from handwheel connectors of AE and N type demand valves.
Plastic End Cap - Insert new plastic end cap into bore of banjo.
Locking End Cap – Insert locking end into bore of banjo and using 2mm adaptor
and torque key tighten end cap to banjo – torque to 0.7Nm.
16 Assemble rubber cover to the demand valve then carry out Leak Test and Functional
Test.
Standard test procedures for C.A.B.E. are as follows and are covered in the relevant
Training Manual for the C.A.B.E.
For detailed instructions of the following Functional Tests relating to the testing of
the lung demand valve and associated ancillary items refer to the appropriate Test
Equipment Manual.
Type A and AE
l Balanced Piston Leak Test
l Positive Pressure Balance Test
l Static Pressure Test
l Positive Pressure Switch Over
Type N
l Balanced Piston Leak Test
l Demand Valve Opening Pressure
The Dräger Warranty and Guarantee is void should any attempt be made by customers
to make adjustment to the balanced piston unit. Guarantee of correct operational
condition of the balanced piston unit is only valid if serviced, and re-sealed, by
Dräger.
Replace with new factory set or Repair Exchange balanced piston unit if problems
occur (other than a requirement to replace springs on positive pressure type units).
Balanced Piston Units will only be accepted for Repair Exchange after disassembly
from demand valve, i.e. Balanced Piston Assembly only.
For details of Terms and Conditions of Repair Exchange Scheme contact Dräger
Service - Blyth.
The following Fault – Cause – Remedy chart outlines a list of possible faults that may
occur.
Number Title
Index
Introduction
The Plus and PSS Series range of Lung Demand Valves are approved for use with approved
Dräger Compressed Air Breathing Equipment (C.A.B.E.) and Facepieces.
The combination of Lung Demand Valve, C.A.B.E. and Facepiece provides the wearer of the
equipment with respiratory protection when working in or escaping from a contaminated or
oxygen deficient life-threatening environment.
For some markets the PSS prefix denotes ‘Short Hose LDV’.
Safety Warnings
l Explain - Instructions for Use supplied with the equipment must be adhered to.
Contact Dräger Limited, Blyth for details of Service Contracts and Service Training
Courses.
l Train - potential wearers must be trained in how to wear, and use the equipment and to
understand its limitations.
1100
Product Description
The variants of the Plus and PSS Series Lung Demand Valve are as follows:
The high performance Lung Demand Valves incorporate a precision factory set Air Transfer
Block Assembly containing a pre-set and sealed balanced piston unit. These items have
unlimited service life.
The Dräger Warranty and Guarantee is void should any attempt be made by customers to
make adjustment to the balanced piston unit or to the associated Injector Pipe.
Re-adjustment of the main lever setting height will be required following the fitting of a
replacement balanced piston unit to an existing Air Transfer Block, or the fitting of a replacement
Air Transfer Block to an existing balanced piston unit.
The threaded handwheel connector of the screw-in variants (types AE and N) incorporate a
swivel feature (patent pending) and when fitted to the facepiece this feature allows the demand
valve a degree of radial movement during head and body movement of the wearer. The swivel
feature is not however suitable for constant rotation.
Detachable variants of the lung demand valves have a medium pressure hose with a quick
release male coupling to connect the demand valve to the female coupling of the medium
pressure hose from the pressure reducer of the appropriate C.A.B.E. (PSS LDV Series are
short hose configuration.)
Fixed variants of the lung demand valves incorporate a medium pressure hose having an end
fitting design for connecting directly to the medium pressure outlet port of the pressure
reducer of the appropriate C.A.B.E.
The housing of the lung demand valves are manufactured from a moulded composite glass
filled material that provides a high degree of strength and durability.
Positive pressure types have a manually operated re-set button to switch ‘OFF’ the positive
pressure function - this must be pressed before opening the valve of compressed air cylinder.
First breath inhalation by the wearer will switch ‘ON’ the positive pressure.
All variants include a supplementary air supply facility that is made active by pressing the
centre of the protective rubber cover at the front of the lung demand valve. Lung demand
valves fitted to short duration C.A.B.E. and escape units will not normally include a
supplementary air supply facility.
Operation
Positive Pressure Lung Demand Valves - Types A and AE
Refer to Figure 1.
Before turning ‘On’ the high pressure air supply the positive pressure mechanism must be
switched ‘Off’. This is achieved by pressing the re-set button at the top of the lung demand
valve to lift and lock the main pivot lever and diaphragm. The conical compression spring, in
the main chamber of the balanced the piston unit, acts against the piston flange resulting in
the piston cone being pressed against the sealing rim of the piston housing - sealing ‘Off’ the
outlet of the balanced the piston unit.
On opening the cylinder valve, medium pressure air flows from the reducer, through the
medium pressure hose of the lung demand valve, into the air transfer block then into the main
chamber of balanced the piston unit.
Refer to Figure 2.
When the wearer begins to inhale a negative pressure is created inside the facepiece and
lung demand valve. Due to the negative pressure the ambient air pressure acts against the
diaphragm. As the diaphragm deflects it then presses against the main hinged lever of the
balanced piston unit resulting in the release of the positive pressure ‘Off’ mechanism.
Increasing deflection of the diaphragm, during inhalation, pivots the cam of the main lever
against the secondary lever. The secondary lever then pivots and presses against the end cap
of the piston - moving the piston axially against the conical spring - lifting the piston cone from
its seating and releasing air from the lung demand valve into wearer’s facepiece. A compression
spring attached to the front cover of the lung demand valve acts on the diaphragm, then the
pivot levers and the conical compression spring, lifting the sealing cone and maintaining a
positive pressure (above ambient) inside the facepiece.
When the wearer stops inhaling then begins to exhale, the diaphragm, pivot levers and sealing
cone retract.
Balanced Piston
‘Closed’
Medium Pressure
(M.P.)
Fig. 1
0583
Balanced Piston
‘Open’
M.P. Flow
to Facepiece
Fig. 2
0587
The conical compression spring, in the main chamber of the balanced the piston unit, acts
against the piston flange resulting in the piston cone being pressed against the sealing rim of
the piston housing - sealing ‘Off’ the outlet of the balanced the piston unit.
On opening the cylinder valve, medium pressure air flows from the reducer, through the
medium pressure hose of the lung demand valve, into the air transfer block then into the main
chamber of balanced the piston unit.
Refer to Figure 4
When the wearer begins to inhale a negative pressure is created inside the facepiece and
lung demand valve. Due to the negative pressure the ambient air pressure acts against the
diaphragm, then the main hinged lever of the balanced piston unit.
Increasing deflection of the diaphragm, during inhalation, pivots the cam of the main lever
against the secondary lever. The secondary lever then pivots and presses against the end cap
of the piston - moving the piston axially against the conical spring - lifting the piston cone from
its seating and releasing air from the lung demand valve into wearer’s facepiece.
When the wearer stops inhaling then begins to exhale, the diaphragm, pivot levers and sealing
cone retract.
Balanced Piston
‘Closed’
Medium Pressure
(M.P.)
Fig. 3
1169
Balanced Piston
‘Open’
M.P. Flow
to Facepiece
Fig. 4
1170
Section Description
5.1 Contents
Important Note: This table is intended as a general guide. Reference must also be made
to the Inspection, Service and Test Intervals specified in the appropriate section
of the relevant C.A.B.E. Training manual.
To ensure correct operational condition of the equipment use the cleaning and
disinfecting solutions recommended by Dräger. The use of any other product will
invalidate the Dräger warranty and guarantee.
When using baths to contain cleaning and disinfecting solutions the immersed
components and assemblies must only be agitated manually. Any form of mechanical
agitation, e.g. ultrasonic, washing machines, dishwashers e.t.c. must be approved
and covered by a written agreement with Dräger.
Note: Cleaning solutions will eventually remove lubricants from moving parts. Following any
cleaning, disinfecting and drying carry out functional tests. Re-lubricate moving parts
as required.
Safety Note: When using cleaning and disinfecting agents refer to the relevant
manufacturers’ usage instructions. It is important that attention be paid to
concentration and reaction times. Do Not use organic solvents such as Acetone,
Alcohol, White Spirit, Trichloroethylene or similar.
1. Cleaning:
Safety Wash
1 Litre Dispenser 3380164
1 Litre Refill 3380165
5 Litre Dispenser 3380166
5 Litre Refill 3380167
Sekusept
4 bottels @ 2 litres 7904071
These products must be used with cold water however Never exceed a temperature
of 30 degrees Celsius.
2. Disinfecting:
Wipex Cloths (for manual disinfecting)
Pack of 50 3380375
Incidur (6 litre container) 7904072
Incidur (30 litre container) 7904073
This product must be used with cold water however Never exceed a temperature of
30 degrees Celsius.
Important Note: When cleaning using a dampened cloth, take care to prevent any fluid
residue entering the demand valve.
When however the user considers it necessary to internally disinfect the lung demand
valve then these instructions should be followed:
Note: 1 Detach demand valves (fitted with male coupling) from the female coupling of the
medium pressure hose of C.A.B.E.
Note: 2 If necessary, demand valves attached directly to the reducer should be removed at the
hose connection to the reducer. For details of removal refer to the appropriate section
of the relevant C.A.B.E. Training Manual.
Tools Required
Plate Spanner R26817
Silicone Spray - SILKOSPRAY® (500ml) 1563343
3 Using the thumb and forefinger carefully grip the centre plate of the diaphragm (3),
tilt centre plate and lift the diaphragm from the body of the lung demand valve.
Remove slip ring (4).
4 Fold rubber cover back over the front of the body of the demand valve to cover and
protect the main pivot lever (5). Immerse and manually agitate demand valve,
diaphragm, slip ring and bayonet cap in cleaning fluid. (Refer to cleaning fluid
instructions.)
Note: Types A and AE - Take care not to stretch or distort the positive pressure spring.
5 After cleaning, rinse components in clean water then immerse and manually agitate
demand valve, diaphragm and bayonet cap in disinfecting fluid. (Refer to disinfecting
fluid instructions.) After disinfecting, fold rubber cover away from the front of lung
demand valve and rinse all components in clean water, taking care not to damage
main pivot lever.
6 Press re-set button (6) of type A and AE demand valves to lift the main lever and to
close the balanced piston.
7 Connect the medium pressure hose of the demand valve to a medium pressure air
supply. Turn ‘ON’ air supply then gently press main pivot lever to blow out fluid
residue.
8 Press the re-set button of type A and AE demand valves to lift the main lever and to
close the balanced piston. Turn ‘OFF’ and vent air supply then disconnect medium
pressure hose of the demand valve from air supply. Thoroughly ‘dry’ components.
Refer to Figure 2.
è
locate the nozzle outlet of the
silicone spray canister inside the
demand valve body and apply two
short spray blasts of lubricant in the
direction indicated by the arrow - 1
Fig. 2.
2 Fig. 2
1386
Safety Note: Refer to the Caution and Warning information on the SILKOSPRAY® canister.
10 Carefully locate outer bead of diaphragm into groove in the body of the demand
valve. Locate tapered side of slip ring into recess in bead of diaphragm.
Types A and AE - Locate positive pressure spring of bayonet cap into the recess in
centre plate of diaphragm.
Using plate spanner, inserted into radial slots of the bayonet cap, lock bayonet cap
to body then fold rubber cover back over the front of the demand valve.
11 Press re-set button of type A and AE demand valves. Connect demand valve to
equipment and carry out Leak Test and Functional Test.
2 Using the thumb and forefinger carefully grip the centre plate of the diaphragm (3),
tilt centre plate and lift the diaphragm from the body of the lung demand valve.
Remove slip ring (4). Discard defective diaphragm.
3 Carefully locate outer bead of new diaphragm into groove in the body of the demand
valve. Locate tapered side of slip ring into recess in bead of diaphragm.
Types A and AE - Locate positive pressure spring of bayonet cap into the recess in
centre plate of diaphragm.
Using plate spanner, inserted into radial slots of the bayonet cap, lock bayonet cap
to body then fold rubber cover back over the front of the demand valve.
4 Press re-set button of type A and AE demand valves. Connect demand valve to
equipment and carry out Leak Test and Functional Test.
Fig. 4
1133
2 When looking inside the bayonet cap and with the axial retaining leg of the spring at
the 11 o’clock position, unhook the first coil of the spring from the retaining lug at
the 2 o’clock position, then from the retaining lug at the 6 o’clock position. Turning
the spring slightly anti-clockwise removes the coil from the final retaining lug (10
o’clock) allowing the axial retaining leg of the spring to be removed from the hole in
the bayonet cap. Discard the defective spring.
3 Fit new spring as follows. When looking inside the bayonet cap and with the retention
hole for the axial retaining leg of the spring at the 11 o’clock position, insert the axial
retaining leg of the spring into the retention hole in the bayonet cap. Hook the first
coil of the spring under the retaining lug at the 10 o’clock position - then under the
retaining lug at the 6 o’clock position - then under the retaining lug at the 2 o’clock
position.
Note: Ensure only one coil diameter of the spring is located under each of the three retaining
lugs.
4 Check that the diaphragm and slip ring are still correctly located in the body of the
lung demand valve then locate positive pressure spring of bayonet cap into the
recess in centre plate of diaphragm. Using plate spanner, inserted into radial slots of
the bayonet cap, lock bayonet cap to body then fold rubber cover back over the front
of the demand valve.
5 Press re-set button and then connect demand valve to equipment and carry out Leak
Test and Functional Test.
Note: Fixed variants of the lung demand valves incorporate a medium pressure hose fitted
directly to the medium pressure outlet port of the pressure reducer. These demand
valves should be removed from the reducer. For details of removal of hose from the
reducer refer to the appropriate section of the relevant C.A.B.E. Training Manual.
Refer to Figure 5.
(Type A illustrated)
1 2 4 3
2 Fold the rubber cover (1) from front
of lung demand valve. Using plate
spanner inserted into radial slots of
bayonet cap, turn bayonet cap (2)
counter-clockwise and remove from
body of demand valve.
Refer to Figure 6.
2
Fig. 6
1134
3 Carefully locate outer bead of diaphragm into the body of the demand valve. Locate
tapered side of slip ring into recess in bead of diaphragm then locate positive pressure
spring of bayonet cap into the recess in centre plate of diaphragm.
Using plate spanner, inserted into radial slots of the bayonet cap, lock bayonet cap
to body then re-fit and fold rubber cover over the front of the demand valve. Press re-
set button to switch ‘Off’ positive pressure.
Note: Fixed variants of the lung demand valves should be fitted directly to the medium pressure
outlet port of the pressure reducer. For details of fitting of hose to the reducer refer to
the appropriate section of the relevant C.A.B.E. Training Manual.
AE and N
Assembly Tool – (Connector O ring) 3310664
Connector Tool 3310679
Fig. 10
0584
Note: Types A and AE - the positive pressure spring remains attached to the bayonet cap. Do
Not remove spring. Take care not to stretch or distort the spring.
3 Using the thumb and forefinger carefully grip the centre plate of the diaphragm (3),
tilt centre plate and lift the diaphragm from the body of the lung demand valve.
Remove slip ring (4).
Push-in Type - A 1 2 4 3
Grip and unscrew push-in type Fig.11
connector (anti-clockwise) from 1138
demand valve body. Remove injector
pipe (4).
Push-in Type - A
Inspect and if necessary fit new O-
ring to connector then screw Fig. 12
connector into demand valve body.
Tighten until flange of the connector 1139
The Dräger guarantee is void should any attempt be made by customers to make
adjustment to the balanced piston unit or the associated Injector Pipe.
AE and N
Assembly Tool 3310664
Connector Tool 3310679
2 1
Fig. 17
1141
Important Note: Checking and possible re-adjustment of the main lever setting height will
be required following the fitting of a replacement balanced piston unit to an
existing Air Transfer Block, or the fitting of a replacement Air Transfer Block to
an existing balanced piston unit.
AE and N
Assembly Tool 3310664
Connector Tool 3310679
Push-in Type - A
Inspect and if necessary fit new O-ring to connector then screw connector into
demand valve body. Tighten until flange of connector contacts the body of the demand
valve. Do Not overtighten.
Important Note: If a replacement balanced piston unit has been fitted to an existing Air
Transfer Block, or a replacement Air Transfer Block has been fitted to an existing
balanced piston unit then checking and possible re-adjustment of the main lever
height will be required. Check height setting as follows.
follows:
Insert the ball end setting tool through the slot at the pivot end of the main lever and
insert end of tool into the slot of the adjusting set screw in the secondary lever.
Rotate the set screw clockwise to increase height and anti-clockwise to reduce
height of the main lever. Following each adjustment remove the tool and press down
carefully then release the main lever and check that when the main lever is released
that the two fingers of the lever are positioned between the two engraved setting-
lines on the face of the gauge plate. Re-adjust as necessary and when set remove
the Height Gauge.
9 Carefully locate outer bead of diaphragm into the body of the demand valve. Locate
tapered side of slip ring into recess in bead of diaphragm.
Types A and AE - Locate positive pressure spring of bayonet cap into the recess in
centre plate of diaphragm.
Using plate spanner, inserted into radial slots of the bayonet cap, lock bayonet cap
to body then fold rubber cover back over the front of the demand valve.
10 Press re-set button of type A and AE demand valves. Connect demand valve to quick
release coupling of equipment and carry out Leak Test and Functional Test.
Note: Fixed variants of the lung demand valves should be fitted directly to the medium pressure
outlet port of the pressure reducer. For details of fitting of hose to the reducer refer to
the appropriate section of the relevant C.A.B.E. Training Manual.
Tools required
Plate Spanner R26817
No. 2 Posi-Driv Screwdriver 3331245
Connector Tool 3310679
Use connector tool (3310679) - insert pins of tool into the swivel connector and
rotate to align slots of fixed end of connector and insert pins - unscrew (anti-clockwise)
to remove screw type connector and handwheel assembly from demand valve body.
Remove injector pipe (4).
1 2 9 4 7 6 10 5
1 2 3 4 9 10 8
Fig. 27
1285007
1 Using the O ring lifter remove and discard O ring (1) and using suitable screwdriver
open gap in circlip (2) and remove the circlip.
3 Carefully pull apart connector - back (5) and (8) from connector - front (9). Retrieve
any loose ball bearings. Remove M45 handwheel and sleeve.
4 Using the O ring lifter remove and discard quad ring (10) from connectors (5) and
(8).
2 1
Type A - 1 - Black 2 - Black
Type AE - 1 - White 2- Black
Type N - 1 - Black 2 - White Fig. 32
1137
Tools required
Assembly Tool 3310664
Connector Tool 3310679
Plate Spanner R26817
No. 2 Posi-Driv Screwdriver 3331245
6 Test Procedures
The Plus and PSS Series range of Lung Demand Valves are approved for use with
approved Dräger Compressed Air Breathing Equipment (C.A.B.E.)
Standard test procedures for C.A.B.E. are as follows and are covered in the relevant
Training Manual for the C.A.B.E.
For detailed instructions of the following Functional Tests relating to the testing of
the lung demand valve and associated ancillary items refer to the appropriate Test
Equipment Manual.
Type A and AE
l Balanced Piston Leak Test
l Positive Pressure Balance Test
l Static Pressure Test
l Positive Pressure Switch Over
Type N
l Balanced Piston Leak Test
l Demand Valve Opening Pressure
The Dräger warranty and guarantee is void should any attempt be made by customers
to make adjustment to the balanced piston unit or the associated Injector Pipe.
Following the fitting of a replacement balanced piston unit to an existing Air Transfer
Block, or the fitting of a replacement Air Transfer Block to an existing balanced
piston unit will require re-adjustment of the main lever setting height.
The following Fault - Cause - Remedy chart outlines a list of possible faults that may
occur.
Contents
1285.001 Lung Demand Valve - Push-In - Type A
1285.006 Lung Demand Valve - Screw-In - Type M40 (Rd40) - Straight Connector
Index
Introduction
The Dräger Panorama Nova P facepiece conforms to the requirements of European Standards
EN136 and EN137-10 and is designed for use with approved and certified compressed air
breathing equipment and airline breathing systems. The facepiece is marked with the CE
symbol.
Front port assembly with integral exhalation valve and plug-in port connection for a push-in
type positive pressure lung demand valve.
Other facepieces are available with either M45 (positive pressure - screw in) or M40 (normal
demand - screw in) front port assemblies.
Front port assembly with integral exhalation valve. The plug-in port is sealed off with a blanking
piece. Fitted as standard is a plug-in side port connection for a push-in type positive pressure
lung demand valve.
When worn in combination with approved and certified compressed air breathing equipment,
or airline breathing systems, the Panorama Nova P full face facepieces provide the wearer
with full face, eyes and respiratory protection when working in a contaminated or oxygen
deficient gaseous atmosphere.
Using the most up to date materials and advanced component design these facepieces provide
the wearer with greater performance, comfort and ease of use.
Safety Warnings
l Explain - Instructions for Use supplied with this equipment, must be adhered to.
Contact Dräger Limited, Blyth for details of Service Contracts and Service Training
Courses.
l Train - potential wearers in how to wear, and use equipment and to understand its
limitations.
Panorama Nova P
Front Port Facepiece
Fig. 1
0086
Panorama Nova P
Side Port Facepiece
Fig. 2
0003
Product Description
The main component parts of the Panorama Nova P full face facepiece is as follows:
The integral double reflex sealing lips, around the facepiece rubber, follow the facial contours
of the wearer maintaining a twofold safety feature to the faceseal.
A grooved moulded bead around the front of the facepiece rubber ensures a positive location
for the full vision visor. The opening for the front port assembly has a moulded flange for
location of the clamping ring.
The side port facepiece also incorporates a moulded flange around the side port for location
of the clamping ring.
Head Harness
The Neoprene head harness provides a safe and comfortable means of securing the facepiece
to the wearer.
Five straps with adjustable buckles radiate from a central plate which positions at the back of
the head. One strap locates centrally over the head and connects to the top of the facepiece.
The remaining straps radiate either side of the head, one pair located at eye level (top straps)
and the second pair at cheek level (lower straps).
Visor
The shape of the visor allows an undistorted 180 degree field of vision. A Polycarbonate visor
is fitted as standard and provides excellent impact and heat resistant properties. A grooved
moulded bead around the front of the facepiece moulding provides positive location for the visor
which is secured in position by two clamp frame halves and screws. Visors are easily replaceable.
Acrylic, or Laminated Glass (Triplex) visors are also available for purchase as optional extras.
Located inside the facepiece, the orinasal mask locates into a groove around the speech
diaphragm retaining ring in the front port assembly. Two matched inhalation valve assemblies
are fitted either side of the mask close to the nasal contact cavity. Each inhalation valve
assembly consists of a moulded valve seat and flexible rubber valve disc.
l Speech diaphragm.
l Exhalation valve disc.
l Exhalation valve spring
l Exhalation valve spring holder.
l Exhalation valve cover.
l Locking mechanism for plug-in connector of demand valve. (Inhalation port.)
A locking ring secures the complete housing assembly to the rubber facepiece. The front port
locking mechanism incorporates a ‘U’ shaped spring that locates into a groove in the demand
valve male connector boss to secure the demand valve into the port. Pressing a ‘push button’ on
the top of the port retracts the spring allowing the demand valve to be removed.
l Speech diaphragm.
l Exhalation valve disc.
l Exhalation valve spring and spring holder assembly.
l Exhalation valve cover.
l Blanking piece. (Inhalation port.)
A locking ring secures the complete housing assembly to the rubber facepiece. The front
inhalation port is sealed off with a plug-in blanking piece.
Supplied as optional extras are screw-in connectors for fitting to the facepiece to accommodate
screw-in demand valves.
Neck Strap
Attached to either side of the facepiece, the neck strap incorporates a hanging clip allowing
the wearer to carry the facepiece on the front of the chest in the ready for use position.
Operating Principle
The illustrations used to describe the operating principle show the PA90 Series lung demand
valve.
On opening compressed air cylinder valve medium pressure air (6bar to 9bar) flows from the
reducer through a hose and into the main bore of balanced piston block of the lung demand
valve.
Increasing air pressure moves the demand valve piston axially, forcing the piston cone against
the O-Ring located in the balanced piston block - resulting in air flow being sealed ‘OFF’ from
lung demand valve outlet.
After putting on the facepiece, with the demand valve fitted, and when the wearer inhales a
negative pressure is created inside the facepiece and lung demand valve. Due to the negative
pressure and the ambient air pressure acting against the diaphragm and hinged balanced
piston lever, the positive pressure ‘OFF’ mechanism is released. Increasing deflection of the
diaphragm pivots the balanced piston lever forcing it against the piston end cap, moving the
piston axially and lifting the piston cone from its seating. This releases medium pressure air
from the lung demand valve into the wearer’s facepiece. The air flows into the main cavity of
the facepiece and across the inside of the visor providing a demisting function. Air then flows
from the main facepiece cavity and through the two inhalation valve discs of the orinasal mask,
to the wearer.
When the wearer stops inhaling the balanced pressure inside the facepiece is between 5mmwg
to 45mmwg (0.5mbar to 4.5mbar). When the wearer begins to exhale the facepiece cavity
pressure increases closing the orinasal inhalation valves and lifting the preset spring loaded
exhalation valve disc from its sealing seat allowing the exhaled air to exhaust to atmosphere.
Following exhalation the spring loaded exhalation valve disc closes onto the sealing seat.
Air pressure in the bore of the piston block increases during exhalation, moving the piston
axially and retracting the pivot lever and diaphragm. Two tension springs, attached to the
hinged lever of the demand valve, act on the hinged lever maintaining positive pressure inside
the facepiece during the complete breathing cycle. This cycle repeats as the wearer begins to
inhale.
Fig. 1
0009
Fig. 2
0010
Inhalation
Fig. 3
0087
Exhalation
Fig 4
0088
Section Description
5.1 Contents
2. Only applicable if the facepiece is used in areas subject to aggressive media, i.e. chlorine,
ammonia, etc.
Safety Note: Refer to manufacturers’ usage instructions when using cleaning and
disinfecting agents. It is important that attention be paid to concentration and
reaction times. Do Not use organic solvents such as Acetone, Alcohol, White
Spirit, Trichloroethylene or similar.
Dräger recommends:
For Cleaning:
Dräger Safety Wash
1 Litre Safety Wash and Dispenser 3380164
1 Litre Safety Wash Refill 3380165
5 Litre Safety Wash and Dispenser 3380166
5 Litre Safety Wash Refill 3380167
Sekusept
4 Bottles @ 2 litres 7904071
For Disinfecting:
Wipex Cloths
Pack of 50 3380375
Special Note: Silicone material adsorbs cleaning and disinfecting agents. Adsorption
increases with each cleaning procedure. Following the use, and long term effects
of the use of such agents, some skin irritation may occur on contact. When
cleaning and disinfecting silicone facepieces observe the following:
1 Refer to Fig. 1.
Fig. 1
0048
2 Refer to Fig. 2.
Fig. 2
0049
4 Refer to Fig.3.
2 Press the exhalation valve spring holder inwards at one end then remove the holder
and spring assembly. Scrap the defective spring and holder assembly. Do Not re-
use.
3 Insert the new spring and holder assembly so that the holder location lugs engage
into the slots in the housing and the spring locates centrally to the metal support
disc of the silicone valve disc assembly.
4 Confirm correct location by checking holder has a small side to side movement in
housing then refit the front port cover.
2 Press the exhalation valve spring holder inwards at one end then remove the holder
and spring assembly.
3 Refer to Fig. 4.
5 Insert the spring and holder assembly so that the holder location lugs engage into
the slots in the housing and the spring locates centrally to the metal support disc of
the silicone valve disc assembly.
6 Confirm correct location by checking holder has a small side to side movement in
housing then refit the front port cover.
1
Fig. 5
0051
2 Refer to Fig. 6.
2 2
Remove defective rubber valve discs
(1) from valve seats (2). Scrap the
defective rubber valve discs.
Fig. 7
0053
1 Grip and remove the inner orinasal mask assembly from inside the facepiece.
2 Refer to Fig. 8.
Fig. 10
0057
1 Grip and remove the inner orinasal mask assembly from inside the facepiece.
2 Using the plate spanner (large pin spacing) unscrew (anti-clockwise), and remove
the (1). Re-orientate the pin spanner then carefully insert narrow pin spacing into
the perforations of the protection disc of the diaphragm. Rotate the speech diaphragm
(2) to release, then remove the diaphragm from the housing. Scrap the defective
diaphragm.
3 Check that the O ring is not damaged. If suspect replace the O ring.
4 Locate the new speech diaphragm (convex perforated protection side towards inside
of facepiece) into housing. Screw the diaphragm retaining ring into housing and
using the plate spanner (large pin spacing) tighten to secure and seal diaphragm.
5 Locate the main bore of the orinasal mask into the groove of the diaphragm retaining
ring then rotate the orinasal mask until the moulded central marking aligns with
central marking on the front port of the facepiece. Insert the flange of exhalation
bore into the orifice in the exhalation valve port.
1 Refer to Fig. 11
1 1
Using the spanner and screwdriver
(3331246), or posidriv screwdriver
(3331245), remove the two
clamping screws (1) and locknuts
(2). Scrap the locknuts. Do Not re-
use.
2 2
Fig. 11
0058
l L stands for
laminated safety glass.
Fig. 13
0061
6 To assist assembly of the clamp frame halves, moisten the outside of facepiece
bead, and the inside groove of clamp frame halves with water. Press and locate the
upper clamp centrally over the bead. Locate the lower clamp centrally into the support
plate, if fitted, and over bead. Align centre marking on each clamp with centre
moulding mark of facepiece rubber.
7 Insert the screws through the bosses of the clamp frames and fit new locknuts.
Tighten equally until the gap between bosses of clamp frames is between 0.5mm to
3.0mm. See Fig. 13.
Note: The end of the screws should be at least level with the end face of the locknut.
1 Grip and remove the inner orinasal mask assembly from inside the facepiece.
Fig. 15
0063
Fig. 16
0064
1 Using screwdriver remove the screw from the side port clamp ring and remove the
clamp ring.
2 Pull and remove the side port assembly from the body of the facepiece and scrap
the defective assembly Check that the rubber flange of the facepiece is not damaged.
Fig. 17
0065
Fig. 18
0066
6 Test Procedures
Refer to the appropriate Test Equipment Manual, (i.e. A900 Test Unit, Testor, Test
Box, Quaestor) for the standard test procedures for the Panorama P facepiece.
Test the facepiece in combination with approved and certified compressed air
breathing equipment, or airline breathing systems, and in line with the test procedures
specified in the appropriate equipment Service Manuals.
7 Fault Location
Will not 1 Dir ty exhalation valve disc Clean valve disc and
hold/maintain positive and/or sealing rim sealing rim
pressure 2 Weak exhalation valve spring Replace exhalation valve
spring assembly
Index
Refer to Instructions for Use supplied with the equipment for use with the compressed air
valved cylinder assemblies.
Introduction
The Dräger Cross Flow, Right Angled Flow and In Line Cylinder Valves conform to the
following standards:
All valve types are designed with or without ratchet operation and are available in either
200bar or 300bar pressure versions, with the valve to cylinder thread connections suitable
for use with a wide range of approved and certified compressed air containers, (cylinders).
Compressed Air Valved Cylinder Assemblies must be charged (pressurised) only with breathable
quality air meeting the air quality standard as defined in EN 12021. Refer also to BS 4275 for
guidance on selection, use and maintenance of respiratory protection equipment. Cylinders used
in combination with approved and certified compressed air breathing equipment provide the
breathing equipment wearer with an independent source of breathable quality air. This combined
system provides respiratory protection in a contaminated, or oxygen deficient gaseous atmosphere.
Duration available is dependant on capacity (volume) of cylinder selected and of course the
breathing work rate of the wearer.
Safety Warnings
l Explain - Instructions for Use for equipment used with compressed air valved cylinder
assemblies, must be adhered to.
Contact Dräger Limited, Blyth for details of Service Contracts and Service Training
Courses.
l Train - potential wearers in how to wear, and use equipment an to understand its
limitations.
Fig. 1
0080
Fig. 2
1304
Fig. 3
0004
Product Description
The main component parts of the Cross Flow, Right Angled Flow, and In Line Cylinder Valves
are as follows:
Type 1 - Non-Ratchet
l Valve body.
l Valve head.
l Spindle.
l Spindle bearing.
l Spindle bearing seal.
l Control knob (handwheel).
Valve Body
The precision machined valve bodies are manufactured from a Brass (Cu Zn 40 Pb2 F37)
pressing and are available in 200bar or 300bar pressure versions.
l BS to BS341
l DIN to DIN477 and European Standard EN 144-2
l M18x1.5 carbon fibre cylinder specification HSE AL FW1 and FW2
l M25x2 carbon fibre cylinder specification HSE AL FW1 and FW2
The high pressure outlet port connection of the valve is to specification G5/8 as per DIN477:
Part 6. Other connections are available on request.
Valve Head
Phosphor Bronze to PB1 BS 2874, the machined valve head incorporates a square drive
socket for locating to the mating (male) drive square on the end of the spindle. Retained in
the end face of the valve head is a ‘Torlon’ insert with precision machined face used to seal
onto the high pressure sealing rim in the valve body.
Spindle
Stainless Steel to DIN 671, the machined spindle incorporates a male drive square for locating
into the mating (female) square socket in the end of the valve head. Two designs of spindle
are available - ratchet or non ratchet.
Spindle Bearing
High Tensile Brass to BS 2874, the threaded spindle bearing is torqued into the valve body to
a value of 40Nm to 41Nm (30ft lbs).
Ratchet Cap
Moulded in Latamid 66 25% Glass Filled Nylon.
Section Description
4.1 Non Ratchet Cylinder Valves
Operating Principle
4.1 Non Ratchet Cylinder Valves
Refer to Figures 1-4
When valve is in the ‘closed’ position, the nylon insert in the valve head seals onto
the sealing rim of the valve body preventing high pressure air being discharged from
the cylinder.
Note: Do Not fully open cylinder valve unless valve connected to compressed air breathing
equipment or high pressure test gauge.
Slowly turning the handwheel of the valve anticlockwise, rotates the spindle and the
threaded valve head. As the valve head rotates it moves away from the sealing rim of
the valve body sliding up the square drive on the end of the spindle. The released
high pressure air from the cylinder discharges to the compressed air breathing
equipment attached to the valve outlet. To ensure no restriction of air to breathing
equipment and to maximise the air flow from the cylinder the valve should always be
fully opened.
Slowly turning the handwheel clockwise, rotates the spindle and threaded valve
head. As the valve head rotates it moves towards the sealing rim of the valve body
sliding down the square drive on the end of the spindle. When the nylon insert
contacts the sealing rim of the valve body this closes the valve preventing high
pressure air being discharged from the cylinder. Closing the valve requires minimal
force - use fingers and thumb- Do Not use excessive force, or tools to seal off
(close) the valve.
Non-Ratchet In-Line
Cross Flow
Closed Closed
Fig. 1 Fig.2
0081 0013
Open Open
Fig. 3 Fig. 4
0082 1092
When valve is in ‘closed’ position, the nylon insert in the valve head seals onto the
sealing rim of the valve body preventing high pressure air being discharged from the
cylinder.
Note: Do Not fully open cylinder valve unless valve connected to compressed air breathing
equipment or high pressure test gauge.
During anticlockwise rotation of the ratchet handwheel of the valve to ‘open’ the
valve, the control knob intermittently lifts and emits an audible ‘click’ as it moves over
the ratchet projections of the ratchet cap. Slowly turning the ratchet handwheel
rotates the spindle and threaded valve head. As the valve head rotates it moves away
from the sealing rim of the valve body sliding up the square drive on the end of the
spindle. The released high pressure air from the cylinder discharges to the
compressed air breathing equipment attached to the valve outlet. To ensure no
restriction of air to breathing equipment and to maximise the air flow from the cylinder
the valve should always be fully opened.
To close the valve, pull the ratchet handwheel away from the valve to lift and over-
ride the ratchet feature during rotation of the handwheel. Repeated pulling and
clockwise turning of the handwheel rotates the spindle and threaded valve head. As
the valve head rotates it moves towards the sealing rim of the valve body sliding
down the square drive on the end of the spindle. When the nylon insert in the valve
head contacts the sealing rim of the valve body this closes the valve preventing high
pressure air being discharged from the cylinder. Closing the valve requires minimal
force - use fingers and thumb - Do Not use excessive force, or tools, to seal off
(close) the valve.
Ratchet In-Line
Cross Flow
Closed Closed
Fig. 5 Fig. 6
0083 0011
Open Open
Fig. 7 Fig. 8
0084 0014
Section Description
5.1 Service and Test Intervals
5.2 Cleaning
5.2 Cleaning
After use, carefully clean, thoroughly dry and inspect contaminated dirty valved cylinder
assemblies, followed by charging.
Safety Note: Check Test Date, stamped on cylinder by last Testing Station. Do Not use
cylinder if past the Test Date. Arrange for recertification of the cylinder.
Dräger recommends:
For Cleaning:
Dräger Safety Wash
1 Litre Safety Wash and Dispenser 3380164
1 Litre Safety Wash Refill 3380165
5 Litre Safety Wash and Dispenser 3380166
5 Litre Safety Wash Refill 3380167
Sekusept
4 Bottles @ 2 litres 7904071
1 Screw the male adaptor of compressor charging hose into female outlet port of valve.
Safety Note: Fully vent (discharge) air from the cylinder before disassembly of the
valve components. Do Not direct the discharging air stream towards the eyes,
face or skin.
1 Using the peg spanner, located into radial slots or into holes in face of flange nut (1
or 1a), unscrew the flange nut and remove spring (2 or 2a).
2 Remove handwheel - non ratchet (3), ratchet (3a), from the end of the spindle.
3 Ratchet Type:
Remove the nylon spacer (4) and lift the ratchet cap (5) from hexagon of spindle
bearing (6a). During disassembly of the ratchet cap if damage occurs to any of the
three retaining lugs of the ratchet cap, scrap the cap.
Orientation
of 8
Seals
7a
8 11
10
11
7 11 10
Ratchet 11
6a
9
5
6 9 4
Non-Ratchet
3
3a
2 2a
1a
1
Fig. 1
0085
Orientation 8
of
Seals 7
10
10
9 9
10
10
6a
5
4
3
3a
2
2
1
1a
Ratchet Type Non-Ratchet Type
Fig. 2
0314
5 Locate the handwheel (3 or 3a) onto the spindle (7 or 7a), and use the handwheel
to rotate the spindle anti-clockwise to unscrew and remove the valve head (8).
6 Clean and inspect all components. Scrap any damaged or suspect components and
replace with new.
Note: All components must be clean and undamaged. Inspect prior to assembly.
1 Apply a small amount of grease (3331247) onto the thread of the valve head (8).
Important Note: Do Not allow grease to contact the ‘Torlon’ sealing face of the valve
head.
2 Apply a small amount of grease (3331247) onto seal location diameter of spindle (7
or 7a) and locate the first sealing washer (11) then the O-ring (10) and then the
second sealing washer (11), or sealing washer (9) onto spindle. (See Figs. 1 and
2 for correct orientation of sealing components.).
3 Locate the valve head (8) onto the square drive of the spindle (7 or 7a), and using
the spindle, screw the valve head into valve body. Rotate clockwise until the valve
head contacts the sealing rim in valve body.
4 Locate the spindle bearing (6 or 6a) over the spindle and screw the bearing into the
valve body. The spindle bearing should screw fully into the valve body by hand,
without restriction.
5 Using 18mm AF adaptor and torque wrench, tighten the spindle bearing to valve
body to a torque of 40Nm to 41Nm (30ft lbs).
6 Ratchet Type:
During disassembly of the ratchet cap if damage had occurred to any of the three
retaining lugs of the ratchet cap, scrap the cap and use a new ratchet cap. Locate
and press the ratchet cap (5) onto hexagon of spindle bearing (6a) and locate nylon
spacer (4) into the recess in the end face cap.
7 Locate the handwheel (3 or 3a) onto spindle ensuring correct location. Insert the
spring (2 or 2a) into the recess in handwheel and using the peg spanner screw the
flange nut (1 or 1a) onto the end of the spindle. Do Not over tighten.
Ratchet Valve
Slowly turn the handwheel anticlockwise until rotation stops i.e. valve fully open. Do
Not use excessive force or tools to open the valve. During rotation of handwheel, to
‘open’ the valve, the handwheel intermittently lifts and emits an audible ‘click’ as it
moves over the ratchet projections of the ratchet cap.
To close the valve, pull the handwheel away from the valve to lift and over-ride the
ratchet feature. Pull and turn the handwheel clockwise until rotation stops i.e. valve
closed. Closing the valve requires minimal force - fingers and thumb - Do Not use
excessive force, or tools.
10 Charge the cylinder assemblies with breathable quality air meeting air quality standard
as defined in EN 12021. (Refer also to BS 4275). Charge cylinder to correct
pressure as stamped on cylinder.
Do Not overcharge. Refer to Section 5.4.
Section Description
6.1 Cylinder Pressure Check
Tools Required
High Pressure Test Gauge 3331209
1 Screw the handwheel of the check gauge into the outlet port of cylinder valve and
close the relief valve of the check gauge.
2 Open the cylinder valve slowly, but fully. Check reading of check gauge.
Note: If the cylinder pressure low, charge to required pressure (stamped on neck or shoulder
of cylinder). If the cylinder is to be used as an air source when performing functional
tests of associated compressed air breathing equipment, then the cylinder pressure
must be at least 80% of maximum working pressure (stamped on neck or shoulder of
cylinder, i.e., 200bar cylinder at least 160bar - 300bar cylinder, at least 240bar).
3 Close the cylinder valve. Open the relief valve of the check gauge to vent pressure
then remove the check gauge from the cylinder valve.
4 Pass the fully charged and checked cylinder to ‘Ready for Use Area’.
For the evaluation testing of breathing air quality for compressors and compressed air
cylinders Dräger recommends using the Dräger High Pressure Aerotest - which connects
directly to a charging compressor outlet or the outlet port of a compressed air cylinder
valve, and uses Dräger Detector Tubes for evaluation and measurement of the following:
Note: Air supplied to the wearer should not contain impurities in excess of limits stated:
7 Fault Location
FAULT CAUSE REMEDY
Leak from valve 1 Valve head leaking Replace valve head and seals
por t 2 Sealing rim in valve body Return to Dräger (Replace
damaged cylinder valve)
Leak through 1 Bent spindle Replace spindle
control knob 2 Damaged spindle seals Replace spindle seals
(handwheel)
Will not 1 Valve head leaking Replace valve head and seals
hold/maintain 2 Sealing rim in valve body Return to Dräger
pressure damaged (Replace cylinder valve)
Stiff handwheel 1 Damaged valve head Lubricate or change valve head
2 Bent spindle Replace spindle
3 Damaged spindle bearing Replace spindle bearing
Number Title
Index
Introduction
The following Functional Tests detailed in this manual cover the A900 Test Unit procedures
for the testing of Positive Pressure (Push-In) Lung Demand Valves which are used in conjunction
with the following facepieces,
For detailed instructions of the following additional tests refer to Test Procedure Section of
the appropriate Service/Training Manual for the C.A.B.A. in use.
Safety Warnings
l Explain - strictly follow instructions as outlined in the Instructions for Use, supplied with
the equipment.
l Maintain - respiratory protection equipment in line with Draeger Limited service schedules
and maintenance instructions.
Contact Draeger Limited, Blyth for details of Service Contracts and Service Training
Courses.
l Train - potential wearers in how to use, and wear equipment and to understand its
limitations.
Refer to Figure 1.
7 D 8 E 1
4
12
9
2 11 6 3 10
Fig. 1
0067
7 D 8 E 1
4
12
9
2 11 6 3 10
Fig. 1
0067
1 Set up the manometer (1) on a clean flat surface allowing enough working space for
the test equipment and the breathing equipment.
Note: The valves are in the ‘open’ position when the valve lever is pointing in the direction of
the valve outlet. Valves shown in Figure 1 are in the ‘Closed’ position. Do Not move or
transport manometer when valves are in ‘Open’ position.
3 Level the manometer by rotating the black knob at the base of the manometer until
the bubble in spirit level is between the two graduation marks. By rotating the knurled
collar at base of the valve B, it is possible to finely adjust the fluid column to the zero
point of the manometer scale. If necessary top up the manometer with manometer
fluid through valve B until the top of the fluid column is aligned with the zero point of
the manometer scale.
4 Panorama Nova P
Assemble connectors (2) and (3) to each of the long tubes on ‘Y’ piece (4). Connect
‘Y’ piece to end of tube D then assemble male blank (5) into connector (2) and
female blank (6) onto connector (3).
Futura P
Assemble connectors (2) and (11) to each of the long tubes on ‘Y’ piece (4). Connect
‘Y’ piece to end of tube D then assemble male blank (5) into connector (2) and
female blank (12) onto connector (11).
6 Assemble the long tube of the aspirator (7) to valve C then Close valve C. Squeeze
the aspirator bulb expelling all air from inside the bulb. On releasing the aspirator
bulb the bulb should remain collapsed.
7 Carefully open valve C allowing the fluid to move slowly down the manometer scale.
The aspirator bulb will begin to slowly expand. Quickly close valve C when the fluid
level is between 60mm to 70mm/wg. Remove the aspirator from valve C.
Note: Fluid level must remain static. If fluid level rises up manometer tube this indicates a
leak. Check seals and hose connections and repeat leak test. If unable to eliminate the
leak, close the valves and return the manometer unit to Draeger Limited - Service
Department.
Section Contents
4 Test Procedures
4 Test Procedures
The following instructions cover the Functional Test procedures for Positive Pressure
(Push-in) Lung Demand Valves fitted with either:
Specification
Allowable Leak 0 to 40mm/wg in less than 5 seconds
Static Balance 5 to 45mm/wg. (No audible leak)
Whistle Activation 55Bar +/- 5Bar
3 Panorama Nova P
Remove blank (5) from connector (2) and female blank (6) from connector (3).
Futura P
Remove blank (5) from connector (2) and female blank (12) from connector (11).
6 Panorama Nova P
Remove inner mask assembly from the facepiece.
Futura P
Inflate the bladder (9), fit the facepiece and then leak test to the bladder as described
in Section 4.3.1.
7 Panorama Nova P
Moisten the exhalation valve then insert connector (3) into bore of exhalation valve
from inside the facepiece.
Futura P
Moisten the exhalation valve then insert connector (11) into the front port of the facepiece.
8 Press the reset feature of the lung demand valve to switch ‘off’ positive pressure
then open the cylinder valve slowly, but fully, to pressurise pneumatic system.
9 Insert outlet of the lung demand valve into the female connector (2).
Leak Test: Check fluid level in manometer tube. The fluid must not move from the
zero line to the 40mm/wg line in less than 5 seconds.
Note: If the demand valve Fails the leak test, check for possible cause and if necessary
replace the seals of the demand valve, reassemble demand valve and re-test. Should
the demand valve again Fail the leak test, replace balanced piston unit. Re-test.
A detected leak, but still within specification limits, will however affect the Static Positive
Pressure Balance Test. This is acceptable. During the Static Positive Pressure Balance
Test it is therefore important to take the initial reading.
12 Close valve A then press the centre of the rubber cover of the demand valve to
switch to positive pressure.
13 Open valve A.
Static Test - Check fluid level in the manometer tube. The fluid should come to rest at any
point between 5mm and 45mm/wg lines. See Note. There should be no audible
leaks.
Note: If the demand valve Fails the Static Pressure Test, open valve C then press the reset of
the demand valve to switch ‘off’ positive pressure. Refer to Section 4:2.
15 Close the cylinder valve. Carefully open valve C to slowly vent air pressure from the
system. Observe the pressure gauge of the equipment and as the pressure decreases
note the pressure at which the whistle begins to sound.
Note: Should the whistle not sound at the required pressure, reset the whistle as detailed
in the relevant section of the Training/Service Manual for the equipment on test.
1 Panorama Nova P
Remove connector (3) from the front port of the facepiece and assemble female
blank (6) to connector.
Futura P
Remove connector (11) from the front port of the facepiece and assemble female
blank (12) to connector.
3 Press the centre of the rubber cover of the demand valve to switch to positive pressure.
a. Close valve A.
b. Check security of the diaphragm, re-test.
c. Check and replace seals of the demand valve, reassemble demand valve then Re-
test from section 4.1.
Testing Facepiece
1 Open valve C.
2 Remove female connector (2) from the demand valve then insert male blank (5) into
the connector.
3 Panorama Nova P
Remove female blank (6) from connector (3).
Futura P
Remove female blank (12) from connector (11).
4 Panorama Nova P
Moisten the exhalation valve and insert connector (3) into bore of the exhalation
valve from inside the facepiece.
Futura P
Moisten the exhalation valve and insert connector (11) into the front port of the
facepiece.
5 Disconnect tube E from valve A and locate free end of the tube E onto valve B.
7 Close valve C. Squeeze aspirator bulb expelling all air from inside the bulb. On
releasing aspirator bulb the bulb should remain collapsed.
8 Carefully open valve C allowing the fluid to move slowly down the manometer scale.
The aspirator bulb will slowly expand. Quickly close valve C when the fluid level in
the manometer tube is between 60mm/wg to 70mm/wg. Remove the aspirator
from valve C.
Note: Manometer fluid level must not rise from 60mm/wg line to 30mm/wg line in less than
one minute.
If the leak is more than specification, remove and visually inspect the silicone disc and
valve seat in the exhalation valve assembly. Clean the valve seat and clean or replace
the silicone disc assembly as necessary. Re-test.
If the exhalation valve passes the inward leak test, (remembering that Static Positive
Pressure Test ‘Failed’) then the exhalation valve spring assembly is not within
specification. Replace the exhalation valve spring assembly (Side Port), or spring (Front
Port), and re-test from Section 4.1.
3 Open the clamp (8) and position the clamp over the protective sleeve of the inlet
tube of the bladder (9).
4 Insert connector (10) into the short tube of the aspirator (7) and assemble connector
(10) into inlet tube of bladder. Inflate the bladder and seal off by screwing the clamp
onto the inlet tube of the bladder then detach the aspirator from the bladder.
5 Panorama Nova P
Fit female blank (6) into connector (3). Remove connector (2) from the ‘Y’ piece
tube and replace with connector (11). Insert connector (11) into female blank (12).
Futura P
Fit female blank (12) into connector (11).
10 Fit the facepiece over the bladder. Check that the facepiece is a good fit and that the
seal and headstrap is tight and secure.
12 Gently squeeze the aspirator bulb until the fluid level in the manometer tube is
between 60mm/wg to 70mm/wg then close valve C.
Note: Manometer fluid level must not rise from 60mm/wg line to 30mm/wg line in less than
one minute.
If the leak is more than specification, check all connections, inflate the bladder a little
more, re-adjust head straps and re-test.
13 Following a satisfactory leak test, Open valve C and remove connector (11) from the
facepiece.
14 Panorama Nova P
Reassemble inner mask to facepiece.
15 Test completed.
Demand Valve fails to switch to 1 Faulty switch 'Off' Refer to LDV Manual
positive pressure mode
5.2 Facepiece
Note: For further Facepiece Test Procedures refer to the relevant Facepiece Service/Training
Manual.
Will not hold/ Weak exhalation valve spring Replace exhalation valve
maintain positive spring assembly
pressure
Unable to maintain 1 Strap ratchets worn Replace headstrap
headstrap 2 Buckle rollers incorrectly fitted Assemble correctly
adjustment
Index
1.1 Introduction
The Dräger Testor test equipment provides an efficient means of testing the Functional
features of the Dräger range of compressed air respiratory protection equipment.
For details of the following reference should be made to the Instructions for Use
supplied with the Testor Unit:
It is recommended that the Testor Instructions for Use and this manual be read and
fully understood before proceeding with any Functional Testing of compressed air
respiratory protection equipment.
The Dräger guarantee is void should original seal caps of the pressure reducer, or
any sealed component assemblies of the ancillary units be tampered with removed
or broken.
l Use of compressed air respiratory protection equipment requires wearer and service
training, knowledge and observance of the instructions contained in the Training/
Service Manual and the Instructions for Use supplied with equipment.
l Use of equipment requires compliance with National Regulations, Laws and
Standards governing the use of respiratory equipment in the country of use.
l Use equipment only for the purpose specified in the Instructions for Use supplied
with equipment, or as confirmed in writing by Dräger.
l Trained and competent personnel should inspect and service equipment at regular
intervals and a record kept of such inspections and servicing.
l Dräger recommends a service contract be obtained from your Dräger Branch or
Agent.
l Contact Dräger for details of Service Contracts and Service Training Courses.
l Notify Dräger in the event of component fault or failure.
l Use only original Dräger Spare Parts for service and maintenance.
2 Test Procedures
Section Description
2 Test Procedures
2 Test procedures
Standard test procedures for compressed air respiratory protection equipment are
as follows:
Details of these tests are covered in the Functional Test Procedure Section of the
relevant Service/Training Manual for the equipment.
Having completed the Functional Test Procedures as defined in the relevant Service/
Training Manual, and if the integrity of the lung demand valve, facepiece or set has
been broken due to repair and/or replacement of components, the following additional
tests should be carried out using the Testor Test Equipment.
Refer to Section 2.2 for detailed instructions for the Functional Tests.
These functional tests should also be carried out in line with the Dräger
recommendations as outlined in the Maintenance and Test Interval Chart of the
Instructions for Use and the Test Procedure Section of the appropriate Training/
Service Manual for the equipment to be tested.
It is recommended that this manual be read and fully understood before proceeding
with any Functional Testing of compressed air respiratory protection equipment.
Tools Required
Adaptor (Rd40 to A) R50028
Blanking Plug 3335421
Safety Note: Connect only breathing quality air to the Testor Unit from a supply pressure
source of maximum pressure 10bar.
‘open’ the cylinder valve (or independent air supply source). Allow the needle of the
medium pressure gauge of the Testor to stabilise.
3 ‘Close’ the cylinder valve then ‘start’ the timer. Observe the high pressure gauge of
the equipment – the gauge reading should not drop more than 10bar in one minute.
During the timing of the pressure leak test – check for audible leak.
Refer to Fig 3
4 Reset the timer to one minute. Fully ‘open’ the cylinder valve. Carefully push the
centre lever forward towards the ‘+’ sign and very carefully inflate the rubber head to
approximately average head size.
5 ‘Start’ the timer. At the end of the timing sequence the test head should show no
signs of reduction in size (deflation).
6 Reset the timer to 30 seconds. With the test head still inflated, moisten the end of
the finger or thumb with water and place and hold over the hole in the mouth fitting
of the test head. Carefully push the RH lever forward towards the ‘+’ sign until the
mbar gauge indicates +20mbar. If an over pressure reading occurs when pushing
lever forward then pull carefully on the LH lever away from the ‘-’ sign until the
pressure reading reduces to +20mbar.
‘Start’ the timer. The gauge reading should remain at +20mbar for 30 seconds.
After 30 seconds remove thumb from mouth fitting of test head – the needle of the
mbar gauge should return to ‘0’. Press the timer ‘Stop’ button.
7 Pull and hold the centre lever towards the ‘-’ sign to deflate the rubber test head.
8 Pull and hold the RH lever back away from ‘+’ sign and insert male blanking plug
into port at rear of test head. Release the RH lever.
Note: Do Not insert the blanking plug into port unless RH lever is held back - otherwise a
positive overpressure will be shown on the mbar gauge.
9 Carefully push LH lever forward towards the ‘-’ sign until the needle of the mbar
gauge rotates anti-clockwise and indicates a negative pressure of -10mbar. Release
the LH lever then ‘Start’ the timer – the gauge reading should remain at -10mbar for
30 seconds.
Note: If an under pressure occurs when pushing the lever forward then carefully pull the lever
away from the ‘-’ sign until -10mbar is achieved, then start leak test as above.
10 Pull and hold the RH lever away from the ‘+’ sign and remove the blanking plug from
the port at rear of the test head. Release the RH lever. The needle of the mbar gauge
should return to ‘0’.
11 ‘Close’ the cylinder valve then press and hold the LH lever towards the ‘-‘ sign to
vent pressure from system. When vented activate the lung demand valve to switch
off positive pressure. Leak testing of the Testor now completed.
Note: If equipment has a quick release coupling on medium pressure hose then attach the
lung demand valve to the coupling.
1 A and AE - Activate the lung demand valve to switch ‘Off’ positive pressure.
2 Open cylinder valve slowly but fully to pressurise the pneumatic system. Check the
pressure reading of the high pressure gauge. Close the cylinder valve and observe
high pressure gauge.
Cylinder content should be at least 80% of maximum charge, i.e. 160bar for 200
units or 240bar for 300 units.
3 High Pressure Leak Test - The high pressure gauge reading should not drop more
than 10bar in one minute. During the timing of the high pressure leak test check for
audible leak(s).
If the equipment fails to conform to test parameters, vent the pressure from the
system, investigate source of the leak, rectify and repeat test. Until the leak is rectified,
Do Not carry out any further test. If test is satisfactory proceed to Instruction 4.
4 Whistle Warning Unit Test - While slowly venting pressure from the system observe
the high pressure gauge. As the pressure decreases check that the whistle activates
an audible alarm signal at between 60bar to 50bar.
A and AE - Cover the outlet port of the lung demand valve with palm of the hand then
press the centre of the rubber cover of the demand valve to switch ‘On’ positive
pressure. Slowly vent pressure by carefully lifting palm of hand to maintain a slow
steady decrease in pressure.
N - Slowly vent equipment by carefully pressing centre of the rubber cover of the
demand valve maintaining a slow steady decrease in pressure.
Note: Before proceeding to test the equipment with the Testor ensure that the Testor Unit is
satisfactory and has ‘Passed’ the leak test as described in Section 2.1.
Tools Required
Adaptor (Rd40 to A) R50028
Blanking Plug 3335421
Important Note: If the equipment to be tested incorporates a lung demand valve that
is connected directly to the pressure reducer then remove the demand valve
from the reducer and assemble the following test hose as required.
Refer to Section 3.
5 Connect the pressure reducer to a fully charged cylinder assembly (or an independent
high pressure air source) then connect the female coupling of the medium pressure
hose of the equipment to the male connector located at the right hand side of the
Testor.
6 Remove the lung demand valve from the inhalation port of the facepiece. Activate the
lung demand valve to switch ‘Off’ positive pressure. Connect the male coupling of
the hose of separate lung demand valve units to the female quick release connector
located at the left side of the Testor.
8 Full Flow Test - Press the centre of the rubber cover of the lung demand valve. A and
AE types will switch ‘On’ to positive pressure. Check that an unobstructed airflow
vents from the lung demand valve. See Safety Note.
Safety Note: During the full flow test Do Not direct the outlet of the lung demand valve
onto the face or skin.
This test should take a minimum of 3 – 5 seconds. Following the test re-activate the
lung demand valve (A and AE) to switch ‘Off’ positive pressure then immediately
observe the medium pressure gauge.
During the timing of the medium pressure test - check for audible leak(s). If a fault is
highlighted refer to Section 4.
11 Carefully push the centre lever forward towards the ‘+’ sign to very carefully inflate
the rubber head to achieve complete contact around the face seal of the facepiece.
12 Insert the blanking plug into the inlet (inhalation) port of the facepiece. Ensure a
secure assembly.
13 Facepiece Negative Pressure Leak Test - Set the timer to one minute. Carefully
push the LH lever forward towards the ‘-’ sign until the needle of the mbar gauge
rotates anti-clockwise and indicates a negative pressure of (minus) -6mbar. Release
the LH lever.
Immediately ‘Start’ the timer. The needle of the mbar gauge should not move from
(minus) -6mbar to -3mbar in less than one minute. If a fault is highlighted refer to
Section 2.3.1.1.
14 Exhalation Valve Lift Off Test - Push and hold the RH lever forward towards the ‘+’
sign. The needle of the mbar gauge rotates clockwise and will stop when the exhalation
valve, in the front port of the facepiece, lifts from its sealing rim. Keep the RH lever
pushed forward - the needle of the gauge should remain stable between the following
values,
15 Remove the blanking plug from the inhalation port of the facepiece then connect the
lung demand valve to the inhalation port ensuring a secure connection.
16 Balanced Piston Leak Test – Pull the RH lever away from the ‘+’ sign to zero (‘0’)
the mbar gauge. Release the RH lever. The needle of the mbar gauge should not
move from ‘0’ to +4mbar in less than 5 seconds.
Note: If a fault is highlighted this indicates a faulty Balanced Piston Unit – stop the test –
service lung demand valve then re-test.
17 Static Pressure Test - Push and hold the LH lever forward towards the ‘-’ sign.
The lung demand valve should activate to positive pressure mode (audible activation).
Release the LH lever. The mbar gauge should indicate a positive pressure reading
of between +1mbar to +3.9mbar. No audible leak.
Note: If a fault is highlighted activate the lung demand valve to switch ‘Off’ positive pressure
– remove the lung demand valve from the facepiece. Proceed to Section 2.3.2.
18 If test satisfactory - ‘Close’ the valve of the high pressure air source.
Whistle Warning Unit Test - Carefully push the LH lever towards the ‘-’ sign to
slowly vent pressure from the system. During slow venting observe the high pressure
gauge of the equipment and check that the whistle activates an audible signal at
between 60bar to 50bar.
Note: As soon as the whistle activates an audible signal – release the LH lever to allow the
whistle to vent the air.
19 Remove the lung demand valve from the facepiece. Pull and hold the centre lever
towards the ‘-’ sign to deflate the test head to average size and until the mbar and
medium pressure gauges indicates ‘0’.
20 Release and extend the side straps of the head harness of the facepiece and remove
the facepiece from the rubber test head. Refit the inner (orinasal) mask.
21 If the equipment ‘Passes’ the tests - disconnect female coupling of the medium
pressure hose from the coupling of the Testor. If the reducer is fitted to an independent
air source grip firmly the outer diameter of the handwheel of the pressure reducer
and unscrew (anti-clockwise) from the high pressure air source.
22 Disconnect the hose coupling of the separate lung demand valve hose from the left
hand female coupling of the Testor and connect to the female coupling of the medium
pressure hose of the pressure reducer. Activate the lung demand valve to switch
‘Off’ positive pressure then refit the lung demand valve to the facepiece ensuring a
secure connection.
Note: If the equipment tested has a continuous hose from the lung demand valve to the
pressure reducer, remove the test hoses as required. Refer to Section 3. Carry out a
high pressure leak test of the equipment.
23 Following satisfactory Tests, complete the test details in the Log Book for the
compressed air respiratory equipment. Store the equipment in designated area,
‘Ready for Use’.
2.3.1 Facepiece
This fault analysis is for guidance – reference must be made to the Fault Location
Section 4 and the appropriate Training/Service Manual for the facepiece.
l Check exhalation valve rubber disc is clean and not damaged. Inspect.
Wash and clean - Re-fit and re-test.
l Check sealing rim for exhalation valve rubber disc is clean and not damaged. Inspect.
Wash and clean - re-test. If damaged – replace port assembly – re-test.
If an initial high reading is indicated, but then settles to within the test parameter, this
indicates a ‘sticky’ exhalation valve disc and/or sealing rim. Remove – clean – rinse
in clean water – re-fit and re-test.
2 Leak Test Lung Demand Valve – Carefully push the RH lever forward towards ‘+’
sign. The needle of the mbar gauge rotates clockwise. Set the needle of the gauge
to +7mbar. Release the RH lever.
Note: If an over pressure occurs when pushing the RH lever forward then carefully pull the
LH lever away from the ‘-’ sign until +7mbar is achieved, then start leak test as follows.
Needle of the mbar gauge should not move from +7mbar to +6mbar in less than 15
seconds. If a failure occurs refer to Section 4. Following the test pull the RH lever
back away from ‘+’ sign until the needle of the mbar gauge indicates ‘0’. Release the
RH lever.
4 Static Pressure Test - Push and hold the LH lever forward towards the ‘-’ sign. The lung
demand valve should activate to positive pressure mode (audible activation). Release the
LH lever. The mbar gauge should indicate a positive pressure reading of between +1mbar
to +3.9mbar. If a failure occurs refer to Section 4. Vent pressure from system.
Important Note: If any failures occur during these tests – service the equipment as
necessary. Refer to the appropriate Training/Service Manual. Retest equipment
with the facepiece.
3 Test Hoses
Section Description
3 Test Hoses
3 Test Hoses
If the equipment to be tested has a continuous hose from the pressure reducer to
the lung demand valve then to perform full functional testing using the Testor the
hose must be disassemble from the reducer and replaced with the following test
hoses as required.
1 Ensure the cylinder valve is ‘Closed’ and the pressure vented from the system.
2 Using the socket key unscrew and remove the blanking plug from the secondary
medium pressure port in the end face of the pressure reducer.
3 Inspect test hose and ensure that the O ring is fitted to the end fitting and is not
damaged. Screw the end fitting of the hose into the secondary medium pressure
port in the end face of the pressure reducer – hand tight.
4 Proceed to either Section 2.1 – Leak Testing Testor, or Section 2.2 – Testing the
Breathing Apparatus.
5 Following satisfactory testing of the equipment, remove the test hose, inspect O
ring of the blanking plug and re-fit plug to secondary port. Tighten using socket key.
Carry out a high pressure leak test of the equipment.
1 ‘Close’ the cylinder valve of the equipment and ensure the system is vented. Unscrew
reducer handwheel from cylinder valve. Loosen cylinder strap, slide cylinder away
from reducer and remove cylinder from equipment.
Refer to Figure 1.
Refer to Fig. 3
Note: Check manifold securely retained by pulling manifold away from reducer body.
6 Locate groove of the retaining cap over the protruding portion of retention staple -
align screw holes - insert screws and using 2mm socket key, tighten screws.
7 Proceed to either Section 2.1 – Leak Testing Testor, or Section 2.2 – Testing the
Breathing Apparatus.
8 Following satisfactory testing of the equipment, remove the test hose from the reducer.
Remove the lung demand valve from the manifold of the test hose and re-assemble
to the pressure reducer. Carry out a high pressure leak test of the equipment.
4 Fault Location
Section Desription
4.1 Pressure Reducer
4.3 Facepiece
Return faulty units (less hoses and any accessories) to Dräger Service.
Important Note: This table is for guidance only – reference must be made to the Fault
Location section of the appropriate Training/Service Manual.
Important Note: This table is for guidance only – reference must be made to the Fault
Location section of the appropriate Training/Service Manual.
4.3 Facepiece
It is important to refer to the relevant Training/Service Manual, for full details of
Service/Maintenance recommendations with regard to the product in use.
Important Note: This table is for guidance only – reference must be made to the Fault
Location section of the appropriate Training/Service Manual.
Inward Leak 1 Exhalation valve leaking Clean and/or replace silicone disc.
Inspect exhalation valve seat in front
por t for damage.
Clean and/or replace.
Important Note: This table is for guidance only – reference must be made to the Fault
Location section of the appropriate Training/Service Manual.
u Repeat the whistle check from the High Safety Warning: Air quality for compressed
For Your Safety Pressure Leak Test.
Note: If high pressure air leak detected either at
connections, or along hose - close secondary air air breathing systems should conform to
PSS 100 Series incorporates Dräger pre-set, and sealed pressure reducer. Drägers guarantee is void should u If necessary repeat instructions to achieve supply valve and vent pressure from hose. Replace requirements of EN12021. Only charge
original seal caps be tampered with, removed, or broken. Correct operational condition is only valid if Dräger required whistle activation setting. hose. compressed air pressure vessels which:
service, and re-seals pressure reducer. u Following satisfactory whistle test, ensure
u PSS 100 D (with ChargAirTM) can only be connected to 300bar valved cylinder assemblies. positive pressure mechanism activated (A and u Conform to National Standards.
u Use of equipment requires, knowledge, and compliance with National Regulations, Laws and Standards, Connecting Secondary Air Supply
AE) to switch Off positive pressure. u Feature original manufacturers test date
governing the use of respiratory protection equipment in the country of use. and test mark.
u Use of equipment requires wearer training and observance of these Instructions for Use. Connecting Demand Valve to the Note: ChargAirTM connection allows refilling of
u Have not exceeded test date indicated on
u Use equipment only for the purpose specified in this manual, or as confirmed in writing by Draeger. Facepiece cylinder(s), from an independent high pressure
u Only trained competent personnel should inspect and service equipment at regular intervals and a record
2 (300bar) secondary breathing air supply source, label affixed by last testing station.
kept of such inspections and service. 0160 Note: Refer also to the Instructions for Use provided while wearer is breathing from equipment.
u Dräger recommends a service contract be obtained from your Dräger Branch or Agent. with the Lung Demand Valve. If the facepiece and the demand valve are not being Visual Inspection
worn, ensure positive pressure mechanism of lung Check integrity of:
u Contact Dräger for details of Service Contracts and Service Training Courses. Type A demand valve (A and AE) activated to switch Off u Carrying system.
u Notify Dräger if there is component fault or failure. u Check facepiece port, and the demand valve O positive pressure. u All straps, buckles and harness.
u Use only original Dräger Spare Parts for service and maintenance. ring, are clean and undamaged. u Valves, connectors and manifold.
u Push demand valve into facepiece port until it u Remove protection cap from ChargAirTM male u Lung Demand Valve.
For Your Safety - ChargAirTM (If Fitted) clicks into position. quick release coupling on equipment. u Facepiece.
u Use ChargAirTM only for the purpose specified in this manual, or as confirmed in writing by Draeger. u Check attachment is secure. u Ensure cylinder valve(s) are fully Open.
u Do Not use ChargAirTM to transfer air from one compressed air breathing equipment to another, or for
second man attachment. Types AE and N u Push male and female ChargAirTM couplings Cleaning, Disinfecting, Drying
u Check facepiece port, and sealing ring, are clean together until a loud click is heard - indicating Carefully clean, disinfect, and thoroughly dry
u Do Not use any charging hose assemblies, or any other equipment, unless specifically supplied by 3 a secure connection. High pressure air from contaminated dirty components after use.
Dräger for use with ChargAirTM. Only use approved equipment. and undamaged.
0156
u Screw demand valve handwheel into facepiece secondary supply source begins to refill When using baths to contain cleaning and
u Do Not lubricate, or allow oil, grease or other contaminants to contact ChargAirTM fittings or hoses.
port. Handtight. equipment cylinder(s). disinfecting solutions, immersed components and
u Do Not attempt to disassemble or repair ChargAirTM fittings or hoses. Drägers guarantee is void should
u Align demand valve (Fig. 6) and tighten. assemblies must be agitated manually. Draeger
ChargAirTM be tampered with, removed, or broken. Correct operational condition is only valid if Dräger
Note: A secure connection has been made when a recommends that no form of mechanical, electrical
service pressure reducer and ChargAirTM assembly.
u Caution: secondary air supply pressure must not exceed maximum rated working pressure (300bar) of Operation dark green ring is visible behind the grey sleeve or ultrasonic agitation be used.
cylinder(s) being filled. Putting on equipment of the female coupling. See Fig. 12.
Safety Note: Refer to manufacturers usage
u Caution: if leak detected while refilling in contaminated, or oxygen deficient gaseous atmosphere, u If necessary set Wearer Height Adjustment.
u If facepiece and demand valve are being worn instructions when using cleaning and
discontinue refilling and immediately leave hazardous area. u Extend waistbelt and shoulder adjusting straps.
- breathe normally during refill. disinfecting agents. It is important that attention
u Caution: it is the users responsibility to ensure that a. the secondary ChargAirTM hose connector is Put on equipment - taking weight on shoulders
u Check reading on equipment pressure gauge, be paid to concentration and reaction times.
compatible with outlet connector of independent high pressure (300bar) secondary breathing air supply by pulling on shoulder adjusting straps.
and if fitted, gauge on secondary supply hose, or Do Not use organic solvents, such as Acetone,
source, and b. the connection is correctly, and securely assembled. 4 Do Not tighten.
secondary supply gauge. When both gauges Alcohol, White Spirit, Trichloroethylene or similar.
0157
u Using right and left hand, grip each side of
Extreme Service Conditions waistbelt pad assembly lifting assembly to top indicate the same pressure - secondary supply
Dräger recommends the following:
of vertical slot of backplate. See Fig. 7. pressure and equipment cylinder(s) pressure
In the presence of chemicals; Some substances, e.g. Phosgene, may be absorbed by, or diffuse through,
Fasten waistbelt buckle. Pull ends of waistbelt have equalised.
the material of the silicone diaphragm of the PA90 Series lung demand valve.
adjusting straps forward (Fig. 8) until waistbelt
1. Cleaning
pad secure and comfortable on hip. Tuck strap Note: Dependent on the pressure remaining in u Dräger Safety Wash.
Liability Statement ends between body and waistbelt pad. the equipment cylinder before refill, pressure u Sekusept
Responsibility for reliable function of equipment transfers to the owner or operator when serviced, or u Pull down shoulder adjusting straps until equalisation occurs after approximately 45-60 Do Not exceed a temperature of 30 degrees
repaired by untrained personnel, (not employed or authorised by Dräger) or when used in a manner not equipment is secure and comfortable on hip. Do seconds. Final equalised pressure governed Celsius.
1
conforming to its intended use. Not overtighten. Tuck the strap ends under the by available pressure and volume of air in
waistbelt. secondary air supply. Due to rapid compression Note: Before disinfecting, rinse off cleaning solution
u Extend the facepiece head harness straps, of air in cylinder(s), during refill, a temperature in clean water.
5 leaving centre strap in position, (Fig. 9). increase occurs (approximately 10° C).
Description To set follow these instructions.
Refer to Fig. 1.
0158
Put neck strap over neck, insert neck strap stud 2. Disinfecting
The PSS 100 Series incorporates a unique into hole in centre strap of head harness. Indicated pressure after refill may decrease as u Incidur.
adjustable articulated carrying system. PSS 100 u Carry out High Pressure Leak Test. cylinder(s) cools to ambient temperature. Remaining Do Not exceed a temperature of 30 degrees
u Lift equipment into vertical position.
Series pneumatics use a high performance first working time will be reduced accordingly. Celsius.
u Turn locking knob (90 degrees) to unlock
stage pressure reducer, with mechanically activated Note: Ensure the positive pressure mechanism of
shoulder yoke.
pressure gauge and whistle warning unit, and is the lung demand valve (A and AE) is activated to u Following pressure equalisation - disconnect Note: Before drying, rinse off disinfecting solution
u Slide shoulder yoke, aligning required torso
compatible with a wide range of balanced piston switch Off positive pressure. quick release coupling by pulling grey sleeve, in clean water.
length with locking knob. Turn locking knob (90
lung demand valves facepieces and compressed
degrees) to lock shoulder yoke. Do Not press u Open valve slowly but fully to pressurise of female coupling, away from male coupling.
air cylinders. The cylinder strap, provides secure u Dräger Wipex Cloths.
down on the adjuster while rotating it. system. Check gauge. See Fig. 12.
support for single cylinder.
Details of cleaning and disinfecting agents are
PSS 100 D - The ChargAirTM connection allows Fitting Cylinder 6 Putting on Facepiece Note: As the coupling separates, a hiss or pop
available from Dräger on request.
u Check valve port and reducer handwheel threads may be heard as the non return valves in both
refilling of cylinder(s), from an independent high 0033
pressure (300bar) secondary breathing air supply are undamaged, connector O ring in position, Note: Refer also to the Instructions for Use provided couplings seal off the high pressure air.
and undamaged. with the Facepiece. 3. Rinsing and Drying
source, while wearer is breathing from equipment. Remove cleaning and disinfecting solutions by
u Place backplate horizontall and extend cylinder u Refit protection caps to male and female
strap. Safety Warning: Facial hair, beard stubble, couplings. rinsing in clean water, followed by drying.
Details of equipment variants and approved side whiskers, and the wearing of spectacles
u Slide cylinder through strap, locating valve port u Recheck pressure gauge and continue. Do Not exceed a temperature of 60 degrees
accessories available from Dräger on request. will adversely affect and interfere with Celsius when drying components.
to reducer handwheel.
facepiece seal. Correct fit of facepiece only
Intended Use u Lift unit upright, screw handwheel into valve
port, (handtight), and, if fitted, hook anti-vibration ensured if facepiece seal makes close contact After Use - ChargAirTM Pressure Reducer and Hose
The PSS 100 Series used in approved lung demand with skin. u Turn Off the air supply at secondary air supply.
strap to handwheel. (Fig. 2).
u Fit protection cap over end of female quick
Assemblies
valve, facepiece and cylinder combinations, u Place backplate horizontal, take up slack on u Detach neck strap stud from centre strap of head
provides the wearer with respiratory protection when cylinder strap (Fig. 3), pull strap over cylinder to harness. release coupling on secondary supply hose.
Note: Refer to For Your Safety.
working in a contaminated, or oxygen deficient activate Camlock (Fig. 4), secure strap to Velcro. 7 u Spread harness (Fig. 10), place chin into u Vent air pressure from the secondary supply
gaseous atmosphere. 0502
facepiece, position harness over head, locating hose by pressing the centre of the end face of the u For cleaning and disinfecting, remove pressure
Connecting Demand Valve to Equipment harness centre plate with back of head. protection cap. reducer and hose assembly from backplate.
Working duration of equipment is dependent on u Tighten both lower (1), then upper straps (2) u Unscrew secondary ChargAirTM supply hose u Remove the locknut, washers and the reducer
capacity (volume) of cylinder selected, and breathing evenly towards back of head. Fig. 11. Only connector from secondary air supply outlet. retaining screw.
Note: Refer also to the Instructions for Use provided
rate of wearer. tighten centre strap (3) if necessary. u Unfasten hose loops, releasing hoses, and
with the Lung Demand Valve.
PSS 100 D - While wearer is breathing from equipment,
u On achieving face seal positive pressure After Use - Equipment remove reducer and hose assembly.
u Insert male coupling, of demand valve, into demand valves (A and AE) automatically
it is possible to extended working time by connecting activate on first inhalation. Note: Do Not immerse pressure reducer in cleaning
female coupling of medium pressure hose. Check Safety Warning: Do Not remove equipment until
ChargAirTM to a high pressure secondary breathing u Breathe normally.
security of connection. in safe area, clear of hazard. or disinfecting solutions.
air source to refill cylinder/cylinders.
Function Checks
Technical Data Pre Operational Checks 8 u Inhale and hold breath. Unit must balance, i.e. u Release side straps on head harness.
u Using clean cloth, moistened in cleaning, or
disinfecting solution, remove dirt and
High Pressure Connection: High Pressure Leak Test 0471
no audible leak. contaminants from the pneumatic assembly.
u Before opening cylinder valve, ensure positive u Continue breathing. Expired air should flow Note: Ensure the positive pressure mechanism of
200bar or 300bar u Using clean cloth, moistened in clean water,
pressure mechanism of lung demand valve (A easily out of exhalation valve. the lung demand valve (A and AE) is activated to
Standard G5/8 as per DIN 477: Part 6. remove disinfecting residue, followed by drying.
and AE) activated to switch Off positive u Press centre of rubber cover of demand valve switch Off positive pressure.
Other connections available to National Standards.
pressure. checking supplementary supply. Note: Do Not exceed a temperature of 60 degrees
u Open cylinder valve slowly, but fully, to u Close cylinder valve. u Remove the facepiece. Close cylinder valve.
Celsius when drying components.
High Pressure Connection - pressurise system. u Breathe normally to vent system. During venting u Unbuckle waistbelt, lift shoulder strap buckles
ChargAirTM (If Fitted) u Close cylinder valve and observe the pressure observe gauge - whistle alarm should sound at to loosen, remove equipment. u Re-assemble reducer and hose assembly to
300Bar maximum. gauge. preset pressure, (See Technical Data). When the backplate and harness.
High Pressure Male Quick Release Coupling. u Gauge reading should not decrease more than gauge indicates zero, hold breath. Facepiece Note: Do Not drop or throw down equipment as
10bar in 1 minute. should hold onto the face indicating positive seal. damage could occur.
Lung Demand Valves
Full Technical Specification available from Dräger. u Continue to Whistle Warning Unit Test. 9 u Open the valve slowly but fully to pressurise Refer also to the Instructions for Use provided with
u Press the centre of the rubber cover of the
the system. the Lung Demand Valve.
Whistle Warning Unit Whistle Warning Unit Test demand valve to vent system and ensure
Note: If leak is detected then readjust the head positive pressure mechanism of lung demand
Preset by Dräger to 55bar +/- 5bar (50bar to 60bar). u A and AE - Cover outlet of demand valve with Facepiece
ball of hand, press centre of rubber cover, turning harness and retest. valve activated to switch Off positive
pressure. Refer also to the Instructions for Use provided with
Compressed Air Cylinders On positive pressure. Slowly vent system by
Use u Pass equipment to Service Department. the Facepiece.
4L to 12L capacity cylinders (200bar or 300bar) carefully lifting ball of hand maintaining a slow
decrease in pressure. u Regularly observe gauge reading. Whistle
available in either steel or composite materials.
Available for twin cylinder mounting is an approved u N - Slowly vent system. Carefully press centre alarm sounds at pressure setting of whistle Routine Maintenance Storage - Ready for Use
conversion kit. Contact Dräger for details. of rubber cover. warning unit.
To be performed after Use. u Fully extend shoulder straps, waistbelt and head
u During venting observe gauge. Whistle should u Go on to safe area, at least when whistle sounds,
by shortest and safest route. See also Maintenance and Test Intervals Chart. harness of the facepiece.
Safety Warning: When using equipment with sound at preset pressure.
twin cylinder configuration, or single 12Litre u Should whistle not sound at required pressure, 10 Refilling Cylinder(s) using u Store the equipment ready for use in a cool
whistle may be reset as follows: Removal of cylinder dry environment, free from dust and dirt and
composite, it is essential to consider possible
restriction when working in confined space.
0016
ChargAirTM (If Fitted) away from excessive heat. Protect the
Refer to Fig. 5. Safety Warning: Cylinder valve should be equipment from any possible damage and
3 Note: See For Your Safety - ChargAirTM closed and system vented. ensure that rubber parts are not subjected to
Preparation for Use u Remove anti-tamper plug from adjuster (1) and 2 direct sunlight.
Wearer Height Adjustment using 3mm hexagon socket key: Secondary Air Supply Source u Lift free end of cylinder strap against Camlock to
PSS 100 Series carrying system length can be set a. Turn adjuster clockwise to increase pressure. Any system where compressed breathing air is release, loosen strap. Unhook anti-vibration strap Safety Note: When stored ready for use, on
to suit torso length of wearer. b. Turn adjuster anti-clockwise to decrease stored can be classified as a secondary air supply (if fitted) from the reducer handwheel and unscrew wall or bulkhead mounting, e.g. Fire Tender,
pressure. 1 source, such as: ensure equipment secured and supported by
handwheel from the cylinder valve.
u Cascade system mounting system.
Position 1.........Short Note: A quarter turn of adjuster is equivalent to a Slide the cylinder away from reducer and remove
u High pressure breathable air compressor with
Position 2.........Medium pressure change of approximately 2bar to 3bar. reservoir from equipment.
Position 3.........Long Do Not turn adjuster with pressure On. 11 u High pressure compressed air cylinder u Recharge the cylinder.
0018
m Dräger Recommendation - UK
uCOSHH
n Grease Demand Valve connector O-ring (Use Molykote 111) as required.
(Type A - Push-in demand valve only).
ü
with EC Type Examination Certificate. uSekusept
EC Type examiner: SGS United Kingdom Ltd, Unit 202b Worle Parkway, Weston-super- Do Not exceed a temperature of 30 Refer to Fig. 3.
Mare, BS22 0WA, UK. Notified Body No. 0120. degrees Celsius.
uPress re-set lever (5). Do Not press
For Your Safety Note: Before disinfecting, rinse off cleaning centre of rubber cover while pressing
solution in clean water. re-set lever, or force lever against stop.
uUse of equipment requires, knowledge, and compliance with National Regulations,
See Fig. 1.
Laws and Standards, governing the use of respiratory protection equipment in the 2. Disinfecting uRemove rubber cover (1). Using plate
country of use.
û
uIncidur. spanner (R26817), unscrew bayonet
uUse of equipment requires wearer training and observance of these Instructions for Do Not exceed a temperature of 30 cap (2).
Use and Instructions for Use supplied with the compressed air respiratory protection degrees Celsius. uCarefully unhook diaphragm (3) from
equipment. lever (4) of balanced piston and discard
uUse equipment only for the purpose specified in this manual, or as confirmed in writing Note: Before drying, rinse off disinfecting defective diaphragm.
by Dräger. solution in clean water. uInsert balanced piston lever (4) into hole
uOnly trained competent personnel should inspect and service equipment at regular Fig. 1 in centre boss of new diaphragm (3)
Details of cleaning and disinfecting agents
intervals and a record kept of such inspections and service. 0032/0251 are available from Dräger on request. (chamfered side of boss towards lever)
uDräger recommends a service contract be obtained from your Dräger Branch or Agent. and carefully locate outer bead of
uContact Dräger for details of Service Contracts and Service Training Courses. 3. Rinsing and Drying diaphragm into groove in demand valve
uNotify Dräger if there is component fault or failure. body - note orientation. Fig. 4.
Remove cleaning and disinfecting solutions by
uUse only original Dräger Spare Parts for service and maintenance. uMoisten lip of bayonet cap (2) with clean
rinsing in clean water, followed by drying.
uUse only Dräger Test Equipment for service and maintenance. water - re-fit bayonet cap. Re-check
Do Not exceed a temperature of 60 orientation of diaphragm.
Liability Statement degrees Celsius when drying components. uAssemble rubber cover (1).
uPress re-set lever. Do Not press centre
Responsibility for reliable function of equipment transfers to the owner or operator when To internally clean and disinfect lung of rubber cover while pressing re-set
serviced, or repaired by untrained personnel, (not employed or authorised by Dräger) or demand valve it is important to follow these lever, or force lever against stop. See
when used in a manner not conforming to its intended use. instructions. (see Fig. 3). Fig. 1.
uConnect demand valve to equipment
uPress re-set lever (5). Do Not press
and perform Pre Operational Checks.
centre of rubber cover while pressing
Description Connecting Demand Valve to re-set lever, or force lever against stop.
Fault, Cause, Remedy,
See Fig. 1.
The high performance PA90AE Lung Facepiece uRemove rubber cover (1). Using plate Service and Test Intervals
Demand Valve incorporates a unique pre- uCheck facepiece port, and sealing ring,
set and sealed balanced piston unit. spanner (R26817), unscrew bayonet uSee attached and refer also to
clean and undamaged. cap (2).
Should sealed balanced piston unit be Instructions for Use provided with
uScrew demand valve handwheel into uCarefully unhook diaphragm (3) from compressed air respiratory protection
tampered with, Drägers guarantee is void.
facepiece port. Handtight. Fig. 2 0028 lever (4) of balanced piston. equipment.
Lung Demand Valve - Part No. 3339406 is uAlign demand valve (Fig. 2) and tighten. 0033 uAssemble rubber cover (1) to demand
for use with the following compressed air valve body to protect balanced piston
respiratory protection equipment. Operation and Use lever.
PA80/90 Series uConnect lung demand valve to compressed uImmerse, and manually agitate demand
PA90 Series air respiratory protection equipment. valve, diaphragm and bayonet cap in
PA90Plus Series uRefer to Instructions for Use provided cleaning fluid. (Refer to cleaning fluid
and Lung Demand Valve - Part No. 3339405 with compressed air respiratory instructions.)
is for use with the following compressed air uAfter cleaning, rinse components in
protection equipment.
respiratory protection equipment. clean water.
PSS 100 Series uImmerse, and manually agitate
PSS 500 Series
After Use components in disinfecting fluid. (Refer
uSee attached and refer also to to disinfecting fluid instructions.)
In the presence of chemicals; Some Instructions for Use provided with uAfter disinfection, remove rubber cover
substances, e.g. Phosgene, may be compressed air respiratory protection (1) and rinse components in clean water.
absorbed by, or diffuse through, the equipment. Take care not to damage balanced
material of the silicone diaphragm of the piston lever.
PA90 Series lung demand valve. Cleaning, Disinfecting, Drying uConnect hose on demand valve to
medium pressure air supply. Press re-
Carefully clean, disinfect, and thoroughly dry
When lung demand valve fitted with non set lever. Turn ON air supply.
contaminated dirty components after use.
silicone diaphragm, fit green cover (1 Fig. 5 4 3 2 1 uGently depress balanced piston lever
2) for identification. To ensure correct operational condition of
Fig. 3 (4), blowing out fluid residue.
equipment, use only cleaning and 0762
Safety Note: Wear suitable eye
Preparation for Use disinfecting solutions recommended by protection.
Connecting Demand Valve to Dräger.
Equipment uDisconnect from air supply, dry
Use of other products invalidates Drägers components and, if necessary, using
Note: Do Not press centre of rubber cover guarantee. Dräger Wipex Cloths, remove excess
while pressing reset lever, or force lever fluid.
against stop. See Fig. 1. When using baths to contain cleaning and uAfter drying, insert balanced piston lever
disinfecting solutions, immersed (4) into hole in centre boss of diaphragm
uInsert male coupling of demand valve components and assemblies must be (3) (chamfered side of boss towards
into female coupling of equipment agitated manually. Dräger recommends lever) and carefully locate outer bead
medium pressure hose. Check security that no form of mechanical agitation be of diaphragm into groove in demand
of connection. used. valve body - note orientation. Fig. 4.
uRefer to Instructions for Use provided uMoisten lip of bayonet cap (2) with clean
with compressed air respiratory Safety Note: Refer to manufacturers usage water - re-fit bayonet cap. Re-check
protection equipment. instructions when using cleaning and orientation of diaphragm.
disinfecting agents. It is important that uAssemble rubber cover (1).
Pre Operational Checks attention be paid to concentration and uPress re-set lever. Do Not press centre
uRefer to Instructions for Use provided reaction times. Do Not use organic of rubber cover while pressing re-set
with compressed air respiratory solvents, such as Acetone, Alcohol, White Fig. 4 lever, or force lever against stop. See
protection equipment. Spirit, Trichloroethylene or similar. Fig. 1.
0029
Order List
Maintenance and Test Intervals Fault - Cause - Remedy
Description Order No
Lung After Every Fault Possible Cause Remedy
Demand Description
Use Month PA90AE Lung Dem and Valv e - Positiv e Pressure - Short Hose 3339405
Valve Demand Valve Banjo O-Rings Defective Replace O-Rings PA90AE Lung Dem and Valv e - Positiv e Pressure - Long Hose 3339406
Clean and Disinfect as m Vents -
necessary Constant Flow
Wipex Cloths (Pack of 50) 3380375
Note: Refer also to Instructions for Use provided with compressed air respiratory Rubber Cov er - Green 3339450
equipment.
Non-Silicon Diaphragm 3339761
Dräger while endeavouring to ensure correctness of statements Dräger Safety UK Limited Dräger Safety AG & Co. KGaA
of fact and advice contained in this publication; gives no Ullswater Close Revalstraße 1 Telefon: 0049451882-0
guarantee or warranty in respect thereof, and accepts no liability Kitty Brewster Estate D-23560 Telefax: 0049451882-2080
Blyth Tel: (01670) 352891 Lübeck
for any mis-statement or inaccuracy in publication, or for any
ommission therefrom. Northumberland NE 24 4 RG Fax: (01670) 356266 Germany Internet http://www.draeger.com
ü
with EC Type Examination Certificate. valve, diaphragm and bayonet cap in
cleaning fluid. (Refer to cleaning fluid compressed air respiratory protection
EC Type examiner: SGS United Kingdom Ltd, Unit 202b Worle Parkway, Weston-super- equipment.
Mare, BS22 0WA, UK. Notified Body No. 0120. instructions.)
uAfter cleaning, rinse components in clean
water.
For Your Safety uImmerse, and manually agitate
uUse of equipment requires, knowledge, and compliance with National Regulations,
components in disinfecting fluid. (Refer
Laws and Standards, governing the use of respiratory protection equipment in the
to disinfecting fluid instructions.)
country of use.
û
uAfter disinfection, remove rubber cover
uUse of equipment requires wearer training and observance of these Instructions for (1) and rinse components in clean water.
Use and Instructions for Use supplied with the compressed air respiratory protection Take care not to damage balanced
equipment. piston lever.
uUse equipment only for the purpose specified in this manual, or as confirmed in writing uConnect hose on demand valve to
by Dräger.
Abb. 1 medium pressure air supply. Press re-
uOnly trained competent personnel should inspect and service equipment at regular
set lever. Turn ON air supply.
intervals and a record kept of such inspections and service. 0032/0251 uGently depress balanced piston lever
uDräger recommends a service contract be obtained from your Dräger Branch or Agent. (4), blowing out fluid residue.
uContact Dräger for details of Service Contracts and Service Training Courses.
uNotify Dräger if there is component fault or failure.
Safety Note: Wear suitable eye
uUse only original Dräger Spare Parts for service and maintenance.
protection.
uUse only Dräger Test Equipment for service and maintenance.
uDisconnect from air supply, dry
Liability Statement components and, if necessary, using
Responsibility for reliable function of equipment transfers to the owner or operator when Dräger Wipex Cloths, remove excess
serviced, or repaired by untrained personnel, (not employed or authorised by Dräger) or
fluid.
when used in a manner not conforming to its intended use.
uAfter drying, insert balanced piston
lever (4) into hole in centre boss of
diaphragm (3) (chamfered side of
Description Connecting Demand Valve to boss towards lever) and carefully
locate outer bead of diaphragm into
The high performance PA90A Lung Facepiece
Demand Valve incorporates a unique pre- groove in demand valve body - note
uCheck facepiece port, and demand
set and sealed balanced piston unit. orientation. Fig. 3.
valve O ring, are clean and undamaged.
Should sealed balanced piston unit be 5 4 3 2 1 uMoisten lip of bayonet cap (2) with clean
tampered with, Drägers guarantee is void.
uPush demand valve into facepiece port Abb. 2 water - re-fit bayonet cap. Re-check
until it clicks into position.
0028 orientation of diaphragm.
uCheck attachment is secure. Pull
Lung Demand Valve - Part No. 3339402 is uAssemble rubber cover (1).
demand valve away from facepiece -
for use with the following compressed air uPress re-set lever. Do Not press centre
respiratory protection equipment. there should be no axial movement.
of rubber cover while pressing re-set
PA80/90 Series lever, or force lever against stop. See
PA90 Series Operation and Use Fig. 1.
PA90Plus Series uConnect lung demand valve to compressed uConnect demand valve to equipment
and Lung Demand Valve - Part No. air respiratory protection equipment.
and perform Pre Operational Checks.
3339400 is for use with the following uRefer to Instructions for Use provided
compressed air respiratory protection with compressed air respiratory
equipment. protection equipment. Replacing Diaphragm
PSS 100 Series
PSS 500 Series After Use Refer to Fig. 2.
uSee attached and refer also to uPress re-set lever (5). Do Not press
In the presence of chemicals; Some
Instructions for Use provided with centre of rubber cover while pressing
substances, e.g. Phosgene, may be
compressed air respiratory protection re-set lever, or force lever against stop.
absorbed by, or diffuse through, the
equipment. See Fig. 1.
material of the silicone diaphragm of the
uRemove rubber cover (1). Using plate
PA90 Series lung demand valve.
Cleaning, Disinfecting, Drying Abb. 3 spanner (R26817), unscrew bayonet
When lung demand valve fitted with non Carefully clean, disinfect, and thoroughly dry cap (2).
silicone diaphragm, fit green cover (1 Fig. contaminated dirty components after use. 0029
uCarefully unhook diaphragm (3) from
2) for identification. lever (4) of balanced piston and discard
To ensure correct operational condition of Dräger recommends only the following: 3. Rinsing and Drying defective diaphragm.
Remove cleaning and disinfecting
Preparation for Use equipment, use only cleaning and
1. Cleaning solutions by rinsing in clean water,
uInsert balanced piston lever (4) into hole
disinfecting solutions recommended by in centre boss of new diaphragm (3)
Connecting Demand Valve to Equipment Dräger. Use of other products invalidates uDräger Safety Wash. followed by drying. (chamfered side of boss towards lever)
Drägers guarantee. uSekusept and carefully locate outer bead of
Note: Do Not press centre of rubber cover Do Not exceed a temperature of 30 Do Not exceed a temperature of 60
while pressing reset lever, or force lever diaphragm into groove in demand valve
degrees Celsius. degrees Celsius when drying components.
against stop. See Fig. 1. When using baths to contain cleaning and body - note orientation. Fig. 3.
disinfecting solutions, immersed uMoisten lip of bayonet cap (2) with clean
Note: Before disinfecting, rinse off cleaning To internally clean and disinfect lung
uInsert male coupling of demand valve components and assemblies must be solution in clean water. water - re-fit bayonet cap. Re-check
demand valve it is important to follow these
into female coupling of equipment agitated manually. Dräger recommends instructions. (see Fig. 2). orientation of diaphragm.
medium pressure hose. Check security that no form of mechanical agitation be 2. Disinfecting uAssemble rubber cover (1).
of connection. used. uIncidur. uPress re-set lever (5). Do Not press uPress re-set lever. Do Not press centre
uRefer to Instructions for Use provided Do Not exceed a temperature of 30 centre of rubber cover while pressing of rubber cover while pressing re-set
with compressed air respiratory Safety Note: Refer to manufacturers usage degrees Celsius. re-set lever, or force lever against stop. lever, or force lever against stop. See
protection equipment. instructions when using cleaning and See Fig. 1. Fig. 1.
disinfecting agents. It is important that Note: Before drying, rinse off disinfecting uRemove rubber cover (1). Using plate uConnect demand valve to equipment
Pre Operational Checks attention be paid to concentration and solution in clean water. spanner (R26817), unscrew bayonet and perform Pre Operational Checks.
uRefer to Instructions for Use provided reaction times. Do Not use organic cap (2). Refer to Instructions for Use provided
with compressed air respiratory solvents, such as Acetone, Alcohol, White Details of cleaning and disinfecting agents uCarefully unhook diaphragm (3) from with compressed air respiratory
protection equipment. Spirit, Trichloroethylene or similar. are available from Dräger on request. lever (4) of balanced piston. protection equipment.
Balanced Piston Leak Replace Balanced Incidur Disinfectant (30 litres) 7904073
Visual Inspection m u
Piston 1 Litre Safety Wash and Dispenser 3380164
Functional and Leak Test as m u
Diaphragm incorrectly Reassemble 1 Litre Safety Wash Refill 3380165
defined in Instructions for
fitted correctly 5 Litre Safety Wash and Dispenser 3380166
Use supplied w ith respiratory
protection equipm ent Perforated Diaphragm Replace Diaphragm 5 Litre Safety Wash Refill 3380167
Dräger while endeavouring to ensure correctness of statements of Dräger Safety UK Limited Dräger Safety AG & Co. KGaA
fact and advice contained in this publication; gives no guarantee or Ullswater Close Revalstraße 1 Telefon: 0049451882-0
warranty in respect thereof, and accepts no liability for any mis-statement Kitty Brewster Estate Tel: (01670) 352891 D-23560 Telefax: 0049451882-2080
or inaccuracy in publication, or for any omission therefrom. Blyth Lübeck
Northumberland NE 24 4 RG Fax: (01670) 356266 Germany Internet http://www.draeger.com
protection equipment. To be performed after Use. u Fold rubber cover (1) from front of body of
Refer to Instructions for Use provided with demand valve. Grip bayonet cap (2), turn Order List
PA80 upgrade compressed air respiratory protection equipment. anticlockwise and remove bayonet cap.
PA90 upgrade u Using thumb and forefinger carefully grip
centre plate of diaphragm (4) tilt centre plate Description Order Code
PA90Plus series + Upgrades
PSS100 Series
Cleaning, Disinfecting, Drying and lift the diaphragm from the body of lung
Carefully clean, disinfect, and thoroughly dry demand valve. Remove slip ring (3). Discard
PSS500 Series
contaminated dirty components as necessary. diaphragm.
u Carefully locate outer bead of new diaphragm
Contact Dräger for details of additional approvals. To ensure correct operational condition of into groove in demand valve body.
Plus N Lung Dem and Valv e - Short Hose 3350501
equipment covered in this Instruction for Use, u Carefully locate tapered side of slip ring into
Intended Use use only cleaning and disinfecting solutions recess in bead of diaphragm and refit bayonet
Plus N Lung Demand Valve, used with approved recommended by Dräger. Use of other products cap.
compressed air respiratory protection equipment, invalidates the Dräger Warranty. u Fold rubber cover back over front of demand Second Lung Dem and Valv e - Plus N 3350606
facepiece and cylinder combinations, provide the When using baths to contain cleaning and valve.
wearer with respiratory protection when working disinfecting solutions, immersed components u Connect demand valve to equipment and
in a contaminated, or oxygen deficient gaseous and assemblies must only be agitated perform Pre Operational Checks as detailed
atmosphere. manually. Any form of mechanical agitation, in Instructions for Use for equipment. Sekusept (4 bottles @ 2 litres) 7904071
e.g. ultrasonic, washing machines,
Refer to Instructions for Use supplied with the dishwashers etc, must be approved and Fault, Cause, Remedy,
compressed air respiratory protection equipment. covered by a written agreement with Dräger.
Service and Test Intervals Incidur Disinfectant (6 litres container) 7904072
In the presence of chemicals:some substances, Note: Cleaning solutions will eventually remove u Refer to Instructions for Use provided with
e.g. Phosgene, may be absorbed by, or diffuse lubrication from moving parts. Following cleaning, compressed air respiratory protection
through the material of the silicone diaphragm of disinfecting and drying carry out functional tests. equipment.
the Plus Series lung demand valve. When the lung Re-lubricate moving parts as required. Incidur Disinfectant (30 litres container) 7904073
demand valve is fitted with the non-silicone
diaphragm fit green cover (1) Fig. 1 for Safety Note: Refer to manufacturers usage
identification. instructions when using cleaning and
disinfecting agents. It is important that
attention be paid to concentration and reaction
Diaphragm - Silicone 3338747
In temperatures below freezing: correct times. Do Not use organic solvents, such as
function of the Lung Demand Valve will be impaired Acetone, Alcohol, White Spirit,
should water be trapped, then freezes inside the Trichloroethylene or similar.
Lung Demand Valve. It is important that water is Rubber Cov er - Green 3338799
removed from the Lung Demand Valve and Dräger recommends only the following:
connection hose. Thoroughly dry after any
cleaning operation. 1. Cleaning
u Dräger Safety Wash. Non-Silicone Diaphragm 3338751
Technical Data u Sekusept
Facepiece to Demand Valve
These products must be used with cold water
Connection however never exceed temperature 30 degrees
Normal Demand
TypeN - 40mm Thread to EN148-1
Celsius. Maintenance and Test Intervals
When the demand valve is fitted to the face piece Note: Before disinfecting, rinse off cleaning
the swivel feature of the handwheel allows a solution in clean water. Lung Demand Description After Use Every
degree of radial movement of the demand valve
during head and body movement of the wearer. 2. Disinfecting
Valve Month
The connection is not suitable for constant rotation. u Incidur
Full Technical Specification available from Dräger. Clean and Disinfect as necessary m
This product must be used with cold water
however never exceed temperature 30 degrees
Preparation for Use Celsius. Visual Inspection m u
Connecting Demand Valve to
Note: Before drying, rinse off disinfecting solution
Equipment
This instruction applies to demand valves with
in clean water. Functional and Leak Test as defined in Instructions for Use m u
quick release male coupling on hose. Details of cleaning and disinfecting agents are supplied w ith respiratory protection equipm ent
u Insert male coupling of demand valve into available from Dräger on request.
female coupling of equipment medium
pressure hose. Check security of connection. 3. Rinsing and Drying m Dräger Recommendation
u Refer to Instructions for Use provided with u COSHH
Remove cleaning and disinfecting solutions by
compressed air respiratory protection rinsing in clean water, followed by drying.
equipment.
Note: Refer also to Instructions for Use provided with compressed air respiratory equipment.
Do Not exceed a temperature of 60 degrees
Pre Operational Checks Celsius when drying components. Immediately
u Refer to Instructions for Use provided with remove from the drying cabinet when dry.
compressed air respiratory protection Never exceed a period of 30 minutes.
equipment.
Lung Demand Valve
Important Note: If an audible leak is evident To internally clean and disinfect lung demand valve
from the lung demand valve - Do Not restrict it is important to follow these instructions.
the outlet port.
Refer to Fig. 1.
Connecting Demand Valve to
Facepiece u Fold rubber cover (1) from front of body of
u Check facepiece port and sealing ring, clean demand valve. Grip bayonet cap (2), turn
and undamaged. anticlockwise and remove bayonet cap.
u Screw the demand valve handwheel into u Using thumb and forefinger carefully grip
facepiece port and tighten. centre plate of diaphragm (4), tilt centre plate
and lift the diaphragm from the body of lung
Operation and Use demand valve. Remove slip ring (3).
u Refer to Instructions for Use provided with Note: The diaphragm must not be ultrasonically
compressed air respiratory protection cleaned, as this will affect the bonded area of the
equipment. diaphragm. Clean and disinfect manually.
For Your Safety u Fold rubber cover back over front of demand
u Use and maintenance of equipment requires knowledge, and compliance with National Regulations, valve body to protect balanced piston lever.
Laws and Standards, governing the use of respiratory protection equipment in the country of use. u Immerse, and manually agitate demand valve,
u Use of equipment requires wearer training and observance of these Instructions for Use and diaphragm, slip ring, and bayonet cap in
Instructions for Use supplied with the compressed air respiratory protection equipment. cleaning fluid. (Refer to cleaning fluid
u Use equipment only for the purpose specified in this manual, or as confirmed in writing by Dräger. instructions.)
u Only trained competent personnel should inspect and service equipment at regular intervals and u After cleaning, rinse components in clean
a record kept of such inspections and service. water.
u Dräger recommends a service contract be obtained from your Dräger Branch or Agent. u Immerse and manually agitate components
u Contact Dräger for details of Service Contracts and Service Training Courses. in disinfecting fluid. (Refer to disinfecting fluid
u Notify Dräger if there is component fault or failure. instructions.)
u Use only original Dräger Spare Parts for service and maintenance. u After disinfection fold rubber cover from front
u Use only Dräger Test Equipment for service and maintenance. of demand valve and rinse components in
clean water. Take care not to damage
Liability Statement balanced piston lever (5).
u Connect hose of demand valve to medium
Responsibility for reliable function of equipment transfers to the owner or operator when serviced,
pressure air supply. Press reset button (6)
or repaired by untrained personnel, (not employed or authorised by Dräger) or when used in a
and turn ‘ON’ air supply.
manner not conforming to its intended use.
u Gently press balanced piston lever (5)
blowing out fluid residue.
Check hoses Replace as required Re-Certification - Cylinder Pressure Test (According to National
Standards).
Safety relief valve Check relief valve O-Ring Replace as necessary Cylinder Overhaul as necessary or at time of Cylinder Re-Certification.
venting Valve
Pressure reducer defective Return to Dräger Service
mDräger Recommendation UK
Demand valve Banjo O-Rings defective Replace uCOSHH
vents n Grease Demand Valve connector O-ring (Use Molykote 111) as required. (Type A - Push-In demand valve only).
(constant flow) Balanced piston leak Replace
Visual Inspection m u Balanced Piston Leak Replace Balanced Incidur Disinfectant (6 litres) 7904072
Piston
Incidur Disinfectant (30 litres) 7904073
Functional and Leak Test as m u Diaphragm incorrectly Reassemble
1 Litre Safety Wash and Dispenser 3380164
defined in Instructions for Use fitted correctly
supplied w ith respiratory 1 Litre Safety Wash Refill 3380165
protection equipm ent Perforated Diaphragm Replace Diaphragm
5 Litre Safety Wash and Dispenser 3380166
m Dräger Recommendation
uCOSHH 5 Litre Safety Wash Refill 3380167
Note: Refer also to Instructions for Use provided with compressed air respiratory Sekusept (4 bottles @ 2 litres) 7904071
equipment.
Diaphragm - PA90 Series Lung Dem and Valv es 3339452
18
17
16
8
15
14
19,20
13
7
6
5
12 12
11
11
10
10
9
NPR 0810
DCC 2482
12872
Ordering Details:
No. of parts required = 16; Edition No: 3
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: July 1999
PSS 100 Series Spare Parts List 1287.2
Pneumatic Arrangement
with ChargAirTM DIN (300Bar)
Gauge or Bodyguard
Page 2 of 2
Order Pack.
Item No Description
Code Qty.
5 O-Ring 3337403 20
7 Spring 3337106 10
10 O-Ring 3337204 20
12 O-Ring 3337314 10
15 Screw 3338959 10
17 O-Ring 3337310 10
Ordering Details:
No. of parts required = 16; Edition No: 3
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: July 1999
PSS® 100 and DrägerMan PSS® 100 Series Spare Parts List 1287.1
Pneumatic Arrangement
Gauge, Bodyguard and DrägerMan
Bodyguard II
Page 1 of 3
21,22
20
19
18 11
17
25,26
23,24
16
10
9
15 15 8
15
14 14 14
13 13 13
12
NPR 1049
NPR 0810
DCC 2774
12871
Ordering Details:
No. of parts required = 16; Edition No: 5
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: June 2000
PSS® 100 and DrägerMan PSS® 100 Series Spare Parts List 1287.1
Pneumatic Arrangement
Gauge, Bodyguard and DrägerMan
Bodyguard II
Page 2 of 3
Order Pack.
Item No Description
Code Qty.
8 O-Ring 3337403 20
10 Spring 3337106 10
13 O-Ring 3337204 20
15 O-Ring 3337314 10
18 Screw 3338959 10
Ordering Details:
No. of parts required = 16; Edition No: 5
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: June 2000
PSS® 100 and DrägerMan PSS® 100 Series Spare Parts List 1287.1
Pneumatic Arrangement
Gauge, Bodyguard and DrägerMan
Bodyguard II
Page 3 of 3
Order P ack.
Item N o D escription
C ode Qty.
20 O-Ri ng 3337310 10
Ordering Details:
No. of parts required = 16; Edition No: 5
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: June 2000
Medium Pressure Test Gauge Spare Parts List 1282.5
Page 1 of 2
3
4
DCC 2784
12825
Ordering Details: Edition No: 3
No. of parts required = 16;
Packaged Quantity = 5; Order Quantity = 20
Subject to change without notice/Änderungen vorbehalten
Date: January 1999
Order in multiples of packaged quantity
Medium Pressure Test Gauge Spare Parts List 1282.5
Page 2 of 2
Order Pack.
Item No Description
Code Qty.
1-4 Low Pressure Test Gauge c/w Quick Release Coupling 3331771
3-5 Low Pressure Test Gauge c/w Broom Wade Coupling 3331214
2 Adaptor 3331900
3 Gauge 3331899
Page 1 of 2
8 9
6 5 4 3 2 1
DCC 5222
12815
Ordering Details:
No. of parts required = 16; Edition No: 3
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: February 2005
High Pressure Check Gauge Spare Parts List 1281.5
Page 2 of 2
Order Pack.
Item No Description
Code Qty.
1 O-Ring D17409 10
2 Connector 3333033
5 O-Ring 3330953 10
7 Washer 3334797 10
8 Gauge 3339388
9 Cover 3339382
Ordering Details:
No. of parts required = 16; Edition No: 3
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: February 2005
PSS 100 Series Spare Parts List 1287
Single Cylinder Strap
Page 1 of 4
35(x4)
26(x2)
27(x2)
20
4
3
25 24 23 22
21
5(x2)
36
28(x2)
18
37,38 29
34
30
31 32
33(x2) 19(x2)
16(x2) 8
9
18 13
14
15(x2)
12
11(x2)
DCC 4081
17(x2) 7 6 10 DCC 2997
DCC 4080
1287
Ordering Details:
No. of parts required = 16; Edition No: 5
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: May 2000
PSS 100 Series Spare Parts List 1287
Single Cylinder Strap
Page 2 of 4
Order Pack.
Item No Description
Code Qty.
1 Backplate 3338268
5 Pin 3338264 2
Ordering Details:
No. of parts required = 16; Edition No: 5
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: May 2000
PSS 100 Series Spare Parts List 1287
Single Cylinder Strap
Page 3 of 4
Order Pack.
Item No Description
Code Qty.
22 Washer 3337312 20
31 Bolt
32 Spacer R50305 10
Ordering Details:
No. of parts required = 16; Edition No: 5
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: May 2000
PSS 100 Series Spare Parts List 1287
Single Cylinder Strap
Page 4 of 4
Order Pack.
Item No Description
Code Qty.
33 Washer
Ordering Details:
No. of parts required = 16; Edition No: 5
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: May 2000
PSS 100 Series Spare Parts List 1287.3
Tool List
Page 1 of 2
1 8
9-11
12,13
14-15
4 16
17
5,6
18-19
DCC 5222
DCC 4966
20,21 Not Illustrated
DCC 4731
12873
Ordering Details:
No. of parts required = 16; Edition No: 4
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: March 2005
PSS 100 Series Spare Parts List 1287.3
Tool List
Page 2 of 2
Order Pack.
Item No Description
Code Qty.
Ordering Details:
No. of parts required = 16; Edition No: 4
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: March 2005
Plus Series Spare Parts List 1285.18
Male Decontamination Hose
3337652 (300mm)
3350588 (400mm)
Page 1 of 2
1
2
9,10
5(x2)
DCC 5360
NPR 0903
128518
Ordering Details:
No. of parts required = 16; Edition No: 4
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: October 2003
Plus Series Spare Parts List 1285.18
Male Decontamination Hose
3337652 (300mm)
3350588 (400mm)
Page 2 of 2
Order Pack.
Item No Description
Code Qty.
2 O-Ring 3337611 10
5 Screw 3337103 10
6 Tailpiece 3335213
Ordering Details:
No. of parts required = 16; Edition No: 4
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: October 2003
PA90Plus Series Spare Parts List 1285.17
Second Man (LDV) Attachment
3337650 (QRC), 3310106 (QRC)
3353559 (QRC), 3339665 (B&W)
Page 1 of 2
1
2
6,7,8
10
DCC 5438
NPR 1379
NPR 1339
ECN 2324
128517
Ordering Details:
No. of parts required = 16; Edition No: 7
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: January 2004
PA90Plus Series Spare Parts List 1285.17
Second Man (LDV) Attachment
3337650 (QRC), 3310106 (QRC)
3353559 (QRC), 3339665 (B&W)
Page 2 of 2
Order Pack.
Item No Description
Code Qty.
2 O-Ring 3337611 10
Ordering Details:
No. of parts required = 16; Edition No: 7
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: January 2004
Whistle Tube and HP Hose Spare Parts List 1285.13
3337316 (985mm)
3353639 (730mm)
3353640 (650mm)
Page 1 of 2
3
7
10
1,1a,1b
4,4a,4b
16 10
11 3
2
12
13 NPR 1414
14 NPR 1413
15 DCC 4332
DCC 4079
DCC 4244
128513
Ordering Details:
No. of parts required = 16; Edition No: 9
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: March 2004
Whistle Tube and HP Hose Spare Parts List 1285.13
3337316 (985mm)
3353639 (730mm)
3353640 (650mm)
Page 2 of 2
Order Pack.
Item No Description
Code Qty.
2 O-Ring 3337204 20
5 Filter 3337340 10
6 O-Ring 3337309 10
8 Screw 3337305 20
11 O-Ring 3337313 10
12 O-Ring 3337314 10
13 Filter 3337350 30
14 Washer 3337312 20
15 Circlip 3337306 20
Ordering Details:
No. of parts required = 16; Edition No: 9
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: March 2004
PSS 100 Series Spare Parts List 1289.1
Electronic Monitoring Unit (Bodyguard)
3338080 - Tally (PSS 100 ET)
3338400 - Button (PSS 100 E)
Page 1 of 2
13
14 11
12
4
3
2
10
15
9(x3) 7
s1289
Ordering Details:
No. of parts required = 16; Edition No: 5
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: October 2000
PSS 100 Series Spare Parts List 1289.1
Electronic Monitoring Unit (Bodyguard)
3338080 - Tally (PSS 100 ET)
3338400 - Button (PSS 100 E)
Page 2 of 2
Item Pack.
Description Order Code
No Qty.
1-14 Electronic Monitoring Unit Tally - PSS 100 ET 3338080
1 Not Available
2 O-Ring 3337314 10
4 O-Ring 3337204 20
15 Battery 3338471
Ordering Details:
No. of parts required = 16; Edition No: 5
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: October 2000
PA90Plus Series - CABE Spare Parts List 1285.16
Basic Customer Tool List
3337915
Page 1 of 2
1 t0016 12 t0034
3 t0006 15 t0014
6 t0004 17 t0032
8 t0007
20 t0034
Page 2 of 2
Order Pack.
Item No Description
Code Qty.
14,
15
13
12
16
17,
18,
17a,
3 18a
1
2a
4
2
7
7
10
8 4
5
11
9
6
36 38 37
DCC 5718
DCC 4416
DCC 2631
12842
Ordering Details:
No. of parts required = 16; Edition No: 13
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: November 2004
PA90 Series Spare Parts List 1284.2
Lung Demand Valve
PA90A, PA90A (German Version)
PA90AE, PA90N
Page 2 of 5
35, 35a
32
31
35,35a
30
32
29
34
27
33
28
27
25
19,20
27
26 23
25
24
21 22 21
12842a
Ordering Details:
No. of parts required = 16; Edition No: 13
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: November 2004
PA90 Series Spare Parts List 1284.2
Lung Demand Valve
PA90A, PA90A (German Version)
PA90AE, PA90N
Page 3 of 5
Pack.
Item No Description Order Code
Qty.
1-6 ,12-14 ,16,17,19, 21-32,37 PA90A PP (Blue Cover) 3339400
1,2a,3,4, 7-9,12,13, 15-17, 21-32,38 PA90AE PP M45 to suit PA80 (Red Cover) 3339406
3 Screw 3339483 10
4 O-Ring 3339459 10
6 O-Ring R18352 5
7 Circlip 3339461 5
12 Diaphragm 3339452
Ordering Details:
No. of parts required = 16; Edition No: 13
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: November 2004
PA90 Series Spare Parts List 1284.2
Lung Demand Valve
PA90A, PA90A (German Version)
PA90AE, PA90N
Page 4 of 5
Pack.
Item No Description Order Code
Qty.
15 Rubber Cover (Red) 3339451
16 Spring 3339419 10
21 Ferrule 20
22 Hose
23 Male Coupling
24 Elbow
25 O-Ring R50117 10
26 Banjo 3339449
27 O-Ring 3339440 10
29 Spring 3339465 5
30 Lever 3339463 2
32 Circlip 3339470 20
34 Washer ND 3339489 20
Ordering Details:
No. of parts required = 16; Edition No: 13
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: November 2004
PA90 Series Spare Parts List 1284.2
Lung Demand Valve
PA90A, PA90A (German Version)
PA90AE, PA90N
Page 5 of 5
Pack.
Item No Description Order Code
Qty.
36 Protection Cap (M40) 3339069 10
Ordering Details:
No. of parts required = 16; Edition No: 13
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: November 2004
Plus and PSS Series LDV Spare Parts List 1285.007
Connector Assemblies
3338791 - M45
3338794 - M40 Page 1 of 2
5 4 2 3(x3) 6 7 1 8
5 4 9 3(x3) 2 1 10
1285007
Ordering Details: Edition No: 1
No. of parts required = 16;
Packaged Quantity = 5; Order Quantity = 20
Subject to change without notice/Änderungen vorbehalten
Date: March 2000
Order in multiples of packaged quantity
Plus and PSS Series LDV Spare Parts List 1285.007
Connector Assemblies
3338791 - M45
3338794 - M40 Page 2 of 2
Order Pack.
Item No Description
Code Qty.
4 Circlip 3338823 5
5 O-Ring 3339459 10
6 Sleeve 3338789
18
17
16
15
9 3
8 2
10
11
1
13
14
4
12
6
19
DCC 4587
DCC 4579
NPR 1001
1285003
Ordering Details:
No. of parts required = 16; Edition No: 3
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: July 2001
PSS Series LDV Spare Parts List 1285.006
Screw-In - M40 (Rd 40)
3338709
Page 2 of 2
Order Pack.
Item No Description
Code Qty.
1 Body 3350204
3 Grommet 3338750 5
6 O-Ring 3339459 10
9 Screw 3350526 10
10 O-Ring 3337215 20
11 O-Ring 3339440 10
13 Spacer 3338733 10
14 Clip 3338737 10
15 Diaphragm 3338747
Ordering Details:
No. of parts required = 16; Edition No: 3
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: July 2001
Plus and PSS Series LDV Spare Parts List 1285.005
Air Transfer Block and
Balanced Piston Unit Assemblies
Page 1 of 2
1,2
4 5 6
DCC 5480
1285005
Ordering Details: Edition No: 1
No. of parts required = 16;
Packaged Quantity = 5; Order Quantity = 20
Subject to change without notice/Änderungen vorbehalten
Date: March 2000
Order in multiples of packaged quantity
Plus and PSS Series LDV Spare Parts List 1285.005
Air Transfer Block and
Balanced Piston Unit Assemblies
Page 2 of 2
Order Pack.
Item No Description
Code Qty.
3 ATB - N
4 O-Ring 3338763 10
6 O-Ring 3338776 10
21,22
20
19
18
9 3
8 2
10
11
1
15
16
4
12,13,14
6
23
NPR 1001
1285003
Ordering Details:
No. of parts required = 16; Edition No: 2
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: April 2001
Plus and PSS Series LDV Spare Parts List 1285.003
Screw-In - M40 (Rd 40)
3350501 3338969 N (H2S)
3338710 3338711
Page 2 of 3
Order Pack.
Item No Description
Code Qty.
1 Body 3350204
3 Grommet 3338750 5
6 O-Ring 3339459 10
9 Screw 3350526 10
10 O-Ring 3337215 20
11 O-Ring 3339440 10
15 Spacer 3338733 10
16 Clip 3338737 10
Ordering Details:
No. of parts required = 16; Edition No: 2
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: April 2001
Plus and PSS Series LDV Spare Parts List 1285.003
Screw-In - M40 (Rd 40)
3350501 3338969 N (H2S)
3338710 3338711
Page 3 of 3
Order Pack.
Item No Description
Code Qty.
Ordering Details:
No. of parts required = 16; Edition No: 2
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: April 2001
Plus and PSS Series LDV Spare Parts List 1285.002
Screw In - M45
3338706 3338968 AE (H2S)
3338707 3338708 Page 1 of 3
21,22
20
19
18
16,17
8 2
7
9
10
1
14
15
4
11,12,13
23
NPR 1458
NPR 1002
DCC 4416
1285002
Ordering Details: Edition No: 4
No. of parts required = 16;
Packaged Quantity = 5; Order Quantity = 20
Subject to change without notice/Änderungen vorbehalten
Date: April 2004
Order in multiples of packaged quantity
Plus and PSS Series LDV Spare Parts List 1285.002
Screw In - M45
3338706 3338968 AE (H2S)
3338707 3338708 Page 2 of 3
Order Pack.
Item No Description
Code Qty.
1 Body 3350204
2 Grommet 3338750 5
5 O-Ring 3339459 10
8 Screw 3350526 10
9 O-Ring 3337215 20
10 O-Ring 3339440 10
14 Spacer 3338733 10
15 Clip 3338737 10
19 Spring 3338768 5
Order Pack.
Item No Description
Code Qty.
24,25
23
22 21
19,20
2
9
8
10
11
17 1
18
4
12,13,14,
15,16
5
NPR 1428
26 NPR 1343
NPR 1003
DCC 4416
1285001
Ordering Details: Edition No: 6
No. of parts required = 16;
Packaged Quantity = 5; Order Quantity = 20
Subject to change without notice/Änderungen vorbehalten
Date: Febuary 2004
Order in multiples of packaged quantity
Plus and PSS Series LDV Spare Parts List 1285.001
Push-In Type A
3338700 3338967 H2S
3338701 3353337 Page 2 of 3
3338702
Order Pack.
Item No Description
Code Qty.
1 Body 3350204
2 Grommet 3338750 5
5 O-Ring 3339459 10
6 Connector - A 3338927
7 O-Ring R18352 10
9 Screw 3350526 10
10 O-Ring 3337215 20
11 O-Ring 3339440 10
17 Clip 3338737 10
18 Spacer 3338733 10
22 Spring 3338768 5
Page 1 of 2
1 t0016 t0036 6
2 t0008 t0044 7
3 t0005 t0045 8
4 t0037 t0046 9
t0014 10
DCC 4935
5 t0034 DCC 4731
Page 2 of 2
Order P ack.
Item N o D escription
C ode Qty.
23 BRIDGE R54133 10
24 CAP R50078 5
25 SPRING R28385 10
26 CAP R28383 10
27 KEY R28384 10
28 CONNECTION PIECE R51490
29 CAP, CPL. R51513 5
30 VALVE DISC R51488 10
31 VALVE SEAT R51489 5
32 PLATE R53179 10
33 SLIDING RING R52184 10
34 CLAMP R52185 10
35 NECK SUPPORT STRAP PVC R51772
35.1 NECK SUPPORT STRAP (NFPA) R54212
36 INHALATION VALVE DISC R54113 10
37 CLAMP R54748
38 ROUND THREAD CONNECTION R54772
39 WASHER RM07946 10
40 HEAD STRAP CPL. R54774
41 CAP BLACK R54441
42 SCREW RING R54442 4
43 FRAME RED R54440 2
44 SCREW 1341642 10
45 SIEVE R55074
30, 46-48 Transition piece ESA R55070
46 VALVE SEAT R55009
47 CONNECTOR R54949
48 O-RING R55014
Internal: Spare parts list for technician
Dräger Service
Safety
Ordering Details:
No. of parts required = 16; Edition No: 1
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: December 1995
Tool List - Facepieces Spare Parts List 1281.7
Push-In (also with Side Port)
M45
M40 (Normal Demand)
Page 2 of 2
Order Pack.
Item No Description
Code Qty.
Ordering Details:
No. of parts required = 16; Edition No: 1
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: December 1995
Cylinder Valve Assembly Spare Parts List 9822.10
Right Hand Angled Flow - Ratchet
3336393-M18-200bar; 3336663-M18-300bar;
3338036-(B.S.) - 300bar
Page 1 of 2
3338038 - (DIN) - 300bar
11
10
3
4
3
1,1a,1b,1c
12
NPR 0435
982210
Ordering Details:
No. of parts required = 16; Edition No: 4
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: January 2000
Cylinder Valve Assembly Spare Parts List 9822.10
Right Hand Angled Flow - Ratchet
3336393-M18-200bar; 3336663-M18-300bar;
3338036-(B.S.) - 300bar
Page 2 of 2
3338038 - (DIN) - 300bar
Order Pack.
Item No Description
Code Qty.
1c, 2-11 Cyl. Valve Assy 300bar - DIN R.H./Angled Flow 3338038
10 Spring 3335639 5
Ordering Details:
No. of parts required = 16; Edition No: 4
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: January 2000
In-Line Cylinder Valves Spare Parts List 9822.8
BS:DIN:M18:M25
Ratchet and Non-Ratchet Types
Page 1 of 3
17,18
1 (BS) 3 (M18)
2(DIN) 4 (M25)
Orientation of
Stem Seals 8
7 6
7
6
6
9
Ratchet Type Non-Ratchet Type
10
11
16
12
15
13
14
14
DCC 4548
98228
Ordering Details:
No. of parts required = 16; Edition No: 7
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: June 2001
In-Line Cylinder Valves Spare Parts List 9822.8
BS:DIN:M18:M25
Ratchet and Non-Ratchet Types
Page 2 of 3
Order P ack.
Item N o D escription
C ode Qty.
4-9, 14-17 C yli nder Valve M25 300bar wi thout Ratchet 3335687
3, 5-9 14-16, 18, C yli nder Valve M18 300bar wi thout Ratchet 3335707
10 Ratchet C ap 3335599
Ordering Details:
No. of parts required = 16; Edition No: 7
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: June 2001
In-Line Cylinder Valves Spare Parts List 9822.8
BS:DIN:M18:M25
Ratchet and Non-Ratchet Types
Page 3 of 3
Order Pack.
Item No Description
Code Qty.
Ordering Details:
No. of parts required = 16; Edition No: 7
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: June 2001
Cross Flow Cylinder Valves Spare Parts List 9822.1
200 and 300bar
BS: DN: and M25 Threads
Ratchet and Non-Ratchet Types
Page 1 of 2
14,15
1-3
4-5
7 7
Ratchet Valve
Orientation of stem seals
9 16 18
17
8
16 16
10
17
16
11
19
20
12 21
22
13
23
24
DCC 4548
98221
Ordering Details:
No. of parts required = 16; Edition No: 10
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: June 2001
Cross Flow Cylinder Valves Spare Parts List 9822.1
200 and 300bar
BS: DN: and M25 Threads
Ratchet and Non-Ratchet Types
Page 2 of 2
Order Pack.
Item No Description
Code Qty.
6 Buffer 3330901 2
Ordering Details:
No. of parts required = 16; Edition No: 10
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: June 2001
Cylinder Valve Assembly Spare Parts List 9822.14
Right Hand Angled Flow - Non Ratchet
Page 1 of 2
11
10
5
6
5
12
1-3 13
DCC 4548
NPR 0676
982214
Ordering Details:
No. of parts required = 16; Edition No: 2
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: June 2001
Cylinder Valve Assembly Spare Parts List 9822.14
Right Hand Angled Flow - Non Ratchet
Page 2 of 2
Order Pack.
Item No Description
Code Qty.
1, 4,12 Cylinder Valve Assembly BS-200Bar (RH) 3337564
10 Spring 3330906 5
Ordering Details:
No. of parts required = 16; Edition No: 2
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: June 2001