You are on page 1of 386

TECHNICAL DOCUMENTATION - PSS® 100 SERIES

PSS® 100 Series

PA90 Series - Lung Demand Valves

Plus and PSS® Series - Lung Demand Valves

Panorama Nova P Facepiece

Cylinder Valves

A900 Test Kit

Testor

Dräger Safety UK Limited, Ullswater Close, Kitty Brewster Estate, Blyth, Northumberland NE24 4RG. Telephone: (01670) 352891. Fax: (01670) 356266
PSS® 100 SERIES 1

Index
Section 1 Instructions for Use

Section 2 Introduction and Safety Warnings

Section 3 Product description

Section 4 Operating Principle

Section 5 Servicing Instructions

Section 6 Test Procedures

Section 7 Fault Location

Section 8 Accessories

Section 9 Spare Parts Lists and Tool List

tm 1287 - June 2005


PSS® 100 SERIES 2

tm 1287- June 2005


PSS® 100 SERIES 3

Section 1 Instructions for Use

tm 1287 - June 2005


PSS® 100 SERIES 4

PSS® 100 - with Mechanical Pressure Gauge 3337446

PSS®100 - with Bodyguard (Tally) 3337448

tm 1287- June 2005


PSS® 100 SERIES 5

Section 2 Introduction and Safety Warnings

tm 1287 - June 2005


PSS® 100 SERIES 6

Introduction
The Dräger PSS® 100 Series of Self Contained Compressed Air Breathing Equipment
(C.A.B.E.) meets the requirements EC Council Directive 89/686/EEC, BSEN 1233, and
conforms to European Standard EN137.

PSS® 100 Series is CE Mark Approved and issued with EC Type Examination Certificate.

Base versions in this series are as follows:

PSS® 100

PSS® 100 E

PSS® 100 ET

The following versions of the Bodyguard electronic monitoring unit are approved against the
above European Standard and available for use with the Dräger PSS 100 Series of Self
Contained Compressed Air Breathing Equipment (C.A.B.E.).

l Bodyguard - Type E ‘Button’ activated


l Bodyguard - Type ET ‘Tally’ activated

The units are Intrinsically Safe, conforming to EMC (radio interference requirements) and to
JCDD 38.

The PSS® 100 Series of Self Contained Compressed Air Breathing Equipment (C.A.B.E.)
provides the wearer with respiratory protection when working in a contaminated or oxygen
deficient life threatening environment.

The advanced design gives greater performance, comfort and ease of use. The revolutionary
articulated carrying system improves weight distribution and the patented slide and pivot
mechanism permits freedom of movement and increased manoeuvrability.

Variants of the PSS® 100 Series use the same adjustable articulated carrying system and
high performance first stage pressure reducer but differ in pneumatic configuration. All variants
are however fully compatible with the Dräger range of lung demand valves, facepieces and
compressed air cylinders.

tm 1287- June 2005


PSS® 100 SERIES 7

PSS 100

l For Firefighting professionals.


l Moulded composite backplate with adjustable and articulated carrying system
- Colour: Black - Anti-static.
l Professional high temperature carrying system harness with comfort padding.
l Camlock cylinder strap
l Mechanically activated pressure gauge and whistle warning unit.
l Medium pressure hose, from first stage pressure reducer, fitted with female
quick release coupling.
l Demand valve fitted with male adaptor for connection to female quick release
coupling of medium pressure hose of first stage pressure reducer.

0525

tm 1287 - June 2005


PSS® 100 SERIES 8

PSS®100 E

l For Firefighting professionals.


l Moulded composite backplate with adjustable and articulated carrying system
- Colour: Black - Anti-static.
l Professional high temperature carrying system harness with comfort padding.
l Camlock cylinder strap.
l Bodyguard - Button Activated.
l Medium pressure hose, from first stage pressure reducer, fitted with female
quick release coupling.
l Demand valve fitted with male adaptor for connection to female quick release
coupling of medium pressure hose of first stage pressure reducer.
Fig. 2
0592

tm 1287- June 2005


PSS® 100 SERIES 9

PSS® 100 ET

l For Firefighting professionals.


l Moulded composite backplate with adjustable and articulated carrying system
- Colour: Black - Anti-static.
l Professional high temperature carrying system harness with comfort padding.
l Camlock cylinder strap
l Bodyguard - Tally Activated.
l Medium pressure hose, from first stage pressure reducer, fitted with female
quick release coupling.
l Demand valve fitted with male adaptor for connection to female quick release
coupling of medium pressure hose of first stage pressure reducer.
Fig. 3
0526

tm 1287 - June 2005


PSS® 100 SERIES 10

Safety Warnings
l Explain - Instructions for Use supplied with this equipment must be adhered to.

l Maintain - respiratory protection equipment as outlined in Dräger Limited service


schedules and maintenance instructions.

Contact Dräger Limited, Blyth for details of Service Contracts and Service Training
Courses.

l Train - potential wearers in how to wear, and use equipment and to understand its
limitations.

Contact Dräger Limited, Blyth for details of Wearer Training Courses.

tm 1287- June 2005


PSS® 100 SERIES 11

Section 3 Product Description

tm 1287 - June 2005


PSS® 100 SERIES 12

Product Description
The PSS® 100 Series of compressed air respiratory protection equipment consists of three
main variants, (PSS® 100, PSS® 100 E and PSS® 100 ET).

Each variant incorporates an adjustable and articulated carrying system for supporting single
or twin compressed air cylinders.

PSS® 100
Attached to the high performance first stage pressure reducer of the
PSS® 100 is the hose assembly of the mechanically activated pressure gauge and whistle
warning unit.

PSS® 100 E and PSS® 100 ET


Attached to the high performance first stage pressure reducer of the PSS® 100 E and PSS®
100 ET is the pressure switch and hose assembly of the Bodyguard multi-function Electronic
Monitoring Unit.

The Bodyguard is a robust and compact instrument, designed to replace traditional mechanical
pressure gauge, whistle warning unit and Distress Signal Unit (DSU).

Operated by either ‘Button’ - E, or ‘Tally’ - ET, and the system pressure of respiratory protection
equipment, this unique multifunction instrument incorporates pre-programmed features for
continuous monitoring of:

l Air pressure of compressed air cylinder.


l Time remaining to whistle alarm setting (TTW).
l Movement Sensor - Distress Signal Unit (DSU).
l Temperature.
l Battery Life.

Available for Bodyguard is a ‘snap on’ I.R. Link and Windows based Software Package for
programming of additional monitoring Options, such as user I.D. and datalogging with
downloading of datalogged parameters.

PSS® 100 Series Carrying System


The carrying system consists of three elements:

l Adjustable Shoulder Yoke.


l Backplate.
l Articulated Waistbelt Assembly.

Backplate
Manufactured from moulded carbon composite material, the three elements are anti-static,
lightweight and durable.

The ergonomic design of the adjustable and articulated carrying system provides the wearer
with unrestricted freedom of movement, stability, a high degree of comfort, and reduced
wearer fatigue.

tm 1287- June 2005


PSS® 100 SERIES 13

A slide and lock arrangement between the backplate and shoulder harness yoke allows
adjustment of the overall length of the carrying system to suit torso length of wearer. Three
preset length settings available are - short, medium, and long.

A universal flexible joint located in the centre of the waistbelt assembly connects to a vertical
slide arrangement at the base of the backplate.

Positioned at the base of the wearer’s spine, (lumbar region), the flexible joint and slide
arrangement combine to provide a natural pivot point about which the backplate and shoulder
yoke move in response to upper body movements of the wearer.

A rubber shock buffer fitted to the base of the backplate provides additional impact protection.
Side handles, incorporated into the backplate moulding, allow carrying of equipment even
with compressed air cylinder attached.

A mounting arrangement, at the base of the backplate, locates the pressure reducer-pneumatics
assembly.

Harness
Harness arrangement contain:

l Shoulder padding with slide lock buckles mounted to shoulder yoke.


l Pair of adjustable (pull-down) shoulder straps.
l Waist padding with slide lock buckles mounted either side of the articulated waistbelt
assembly
l Adjustable (pull forward) waistbelt straps with push-in quick release buckle set.

Padding and adjustable straps remove easily for maintenance and cleaning.

Pneumatic Arrangement
The pneumatic arrangement incorporates a pressure reducer with a threaded handwheel for
connection to the outlet port of the valve of a compressed air cylinder (200bar - 300bar).
TM
Variants that are fitted with ChargAir utilise a 300bar only pressure reducer incorporating a
threaded handwheel for connection only to the outlet port of a 300bar compressed air cylinder
TM
valve. Assembled to the second high pressure outlet port, from the reducer is the ChargAir
supply hose withTMquick release male coupling. Connection of a special secondary supply
hose to ChargAir allows refilling of cylinder(s) from an independent high pressure (maximum
300bar) secondary breathing air supply source, while wearer is breathing from equipment.

A bolt and locknut arrangement at the base of the backplate secures the reducer to backplate.
A degree of radial movement of the reducer on the mounting bolt assists in alignment and
assembly of the reducer handwheel to the outlet port of the cylinder valve.

Fitted to dual pressure outlet port of reducer is either:

PSS® 100
A unique dual function hose assembly incorporating an in-line whistle warning unit, (high
pressure activated - medium pressure operated), and high pressure contents gauge for
monitoring air pressure in compressed air cylinder.

tm 1287 - June 2005


PSS® 100 SERIES 14

PSS® 100 E and PSS 100 ET


Connected to the second high pressure outlet port is the pressure sensor of the hose to the
Bodyguard.

Fitted to second outlet port (medium pressure) is a medium pressure hose with quick release
female coupling for attaching a separate lung demand valve.

This port also incorporates the pressure relief valve, (fitted to nozzle of medium pressure
hose), and can also be used for connecting an accessory, such as a decontamination hose,
buddy breathing attachment or airline belt manifold.

Hose loops secure the dual pressure supply hose, or hose to the Bodyguard, and medium
pressure hose to the backplate and shoulder harness.

Available for twin cylinder fitting is a Twin Pack Kit. Contact Dräger for details.

The reducer-pneumatic arrangement, harness padding and adjusting straps easily remove
from equipment for maintenance and cleaning.

Lung Demand Valves


The lung demand valve is a separate unit fitted with medium pressure hose and quick release
male coupling.

Functional variants are:

l Type A Positive Pressure with Push-In connector to facepiece.


l Type AE Positive Pressure with M45 Screw-In connector to facepiece.
l Type N Normal Demand with M40 Screw-In connector to facepiece.

Refer to the appropriate Training Manual for details of the selected lung demand valve.

tm 1287- June 2005


PSS® 100 SERIES 15

Section 4 Operating Principle

tm 1287 - June 2005


PSS® 100 SERIES 16

Operating Principle
Pneumatic Arrangement
Pressure Reducer
The high performance first stage pressure reducer is factory set and sealed and must not be
tampered with as this invalidates the conditions of warranty of Dräger Limited.

Service life of the reducer is six years. Dräger recommend that at the end of this period the
reducer be returned to Dräger Service, Blyth, for overhaul under the Repair Exchange Scheme
(REX).

Details of REX Scheme available on request.

Functioning at inlet pressures of 200bar or 300bar, (300bar only with ChargAirTM), the
reducer provides a controlled outlet medium pressure to the lung demand valve.

Incorporated in medium pressure outlet port is a safety pressure relief valve.

PSS® 100
Pneumatic Arrangement
A preset and sealed mechanism between the two outlet ports of the reducer controls the
activation of the whistle warning unit. The mechanism activates when the available air remaining
in compressed air cylinder has reduced to a preset pressure.

Operation - refer to Figure 1, Figure 2, Figure 3 and Figure 4


On opening the valve of the compressed air cylinder the high pressure air flows through a
sintered bronze filter in reducer connector. High pressure air entering the reducer closes the
whistle activation mechanism, pressurises the capillary tube in dual pressure hose activating
pressure gauge, passes into the piston bore, past the piston seat and pres-et plunger and
into the medium pressure chamber.

As pressure increases in the medium pressure chamber it forces the piston to move axially
against the compression spring until the piston seat contacts the end face of plunger - sealing
off high pressure.

On inhalation, from lung demand valve, the air flows from the medium pressure chamber and
as pressure reduces, the compression spring retracts the piston moving the piston seat from
the end face of the plunger. (Fig. 2.)

The high pressure air then begins to enter the medium pressure chamber repeating the cycle.

Consumption of air from the cylinder(s) reduces the pressure and at a pre-set pressure the
compression spring in the whistle activation mechanism retracts the high pressure plunger
from its sealing seat releasing medium pressure air from the medium pressure chamber into
the whistle activation chamber. (Fig. 3.) From the activation chamber a controlled airflow
enters the bore of the outer rubber hose of the dual pressure hose assembly operating the
whistle warning unit.

tm 1287- June 2005


PSS® 100 SERIES 17

Pressure Reducer

High
Pressure

Medium
Pressure

H.P. to gauge High Pressure

Fig. 1
0660

tm 1287 - June 2005


PSS® 100 SERIES 18

Medium Pressure Flow to Demand Valve


High Pressure above 60bar - Whistle not activated

M.P. to Demand Valve Medium


Pressure

High
Pressure

Medium
Pressure

H.P. to Gauge High Pressure

Fig. 2
0661

tm 1287- June 2005


PSS® 100 SERIES 19

Medium Pressure Flow to Demand Valve


High Pressure below 60bar Whistle activated

M.P. to Demand Valve

High
Pressure

Medium
Pressure

L.P. to Whistle

H.P. to Gauge

Fig. 3
0662

tm 1287 - June 2005


PSS® 100 SERIES 20

Reducer Failure
Pressure Relief Valve - ‘Open’

Air venting from


relief valve

High
Pressure

Medium
Pressure

Fig. 4
0664

tm 1287- June 2005


PSS® 100 SERIES 21

Dual Function High and Medium Pressure Hose Assembly


- Refer to Figure 5

The dual pressure hose supplies medium pressure air from the reducer to the in line whistle
warning unit and high pressure air to the cylinder contents gauge.

The hose construction although appearing similar to medium pressure hoses, differs in its
internal construction. Inside the outer rubber hose assembly is a fine bore copper capillary
tube that is coil wound around a multi-strand length of phosphor bronze wire. The capillary
tube and wire are then brazed to each of the end fittings of the capillary tube assembly. High
pressure air flows through the bore of the capillary tube to the pressure gauge and not
through the rubber hose.

At a pre-set pressure the whistle activation mechanism releases medium pressure through
the secondary end fitting, into bore of outer rubber hose assembly then into the whistle tube
operating the whistle warning unit located at base of gauge.

Whistle Warning Unit


- Refer to Figure 5

Constructed as part of the dual pressure hose assembly, and located at base of pressure
gauge, the in-line whistle warning unit provides an audible alarm signal. The alarm begins
sounding when the available air remaining in the compressed air cylinder has reduced to a
pre-set pressure.

The wearer should regularly monitor the gauge pressure to determine the remaining available
air and leave the contaminated area Before activation of whistle, however on hearing the
alarm the wearer Must immediately leave the contaminated area for a safe environment.

The pre-set pressure setting of the Whistle Warning Unit can be adjusted by the user to meet
special requirements.

Operation
See descriptions of Pressure Reducer and Whistle Warning Unit, and Dual Pressure Assembly

High Pressure Contents Gauge


- Refer to Figure 5

The content gauge is a Bourdon Tube type constructed from a stainless steel pressed case
and having a 3mm thick lens window. The gauge conforms to the technical and accuracy
requirements of European Standard EN137.

The gauge dial is luminescent yellow having radial black graduation marks at 10bar intervals
from 0 to 350bar. At 50bar intervals graduation marks are red. A red radial arc highlights the
segment from 0 to 50bar.

tm 1287 - June 2005


PSS® 100 SERIES 22

Dual Function Hose - Gauge and Whistle

Gauge

Gauge
Cover

Whistle

H.P. Capillary

Phosphor
Bronze
Wire

Fig. 5
0177

tm 1287- June 2005


PSS® 100 SERIES 23

PSS® 100 E and PSS100 ET


Pneumatic Arrangement
Connected to the dual pressure outlet port of the reducer is the pressure switch and hose
assembly of the Bodyguard multi-function Electronic Monitoring Unit.

The Bodyguard has basic monitoring and alarm options programmed into the instrument
before despatch from Dräger.

Refer to Instructions for Use supplied with the Bodyguard.

Operation - refer to Figure 6, Figure 2 and Figure 4


On opening the valve of compressed air cylinder the high pressure air flows through a sintered
bronze filter in the connector. Increasing air pressure makes the pressure switch of the
Bodyguard active - turning ‘On’ the unit.

Air passes into the piston bore of reducer, past the piston seat and pre-set plunger and into
the medium pressure chamber of the reducer.

As pressure increases in medium pressure chamber, it forces the piston to move axially against
the compression spring until the piston seat contacts the end face of the plunger - sealing off
high pressure.

On inhalation from the lung demand valve, air flows from the medium pressure chamber and
as pressure reduces the compression spring retracts the piston moving the piston seat from
the end face of the plunger. (Fig. 2). High pressure air begins to enter the medium pressure
chamber repeating the cycle.

The pre-programmed features of the Bodyguard continually monitor performance of the


equipment.

Refer to Instructions for Use supplied with PSS 100 E and PSS 100 ET.

tm 1287 - June 2005


PSS® 100 SERIES 24

Pressure Reducer -
PSS® 100 E and PSS® 100 ET

Medium
Pressure

High
Pressure

Bodyguard High
Pressure Switch Pressure

Fig. 6
0665

tm 1287- June 2005


PSS® 100 SERIES 25

Bodyguard - Hose Assembly - refer to Figure 7 and Figure 8

The hose assembly of the Bodyguard electronic monitoring unit, although appearing similar
to the medium pressure hoses, provides a flexible protective sleeve encasing the electronic
wiring from the pressure sensor to the electronic monitoring unit.

PSS® 100, PSS® 100 E and PSS® 100 ET


Common features of the above equipment variants are:

Hose Assemblies
Medium Pressure Hoses - refer to Figure 9
Flexible and anti-kinking rubber hose lengths combine to provide medium pressure hose
assemblies for the different reducer-pneumatic arrangements and associated lung demand
valves.

Pressure relief valve spring, O-Ring retainer and O-Ring locate onto the nozzle end fitting of
the medium pressure hose that connects to medium pressure outlet port of the reducer.

The O-Ring of the relief valve also functions as a seal for the hose to reducer connection.
TM
ChargAir Hose (Optional) - refer to Figure 10 TM
Assembled to the second high pressure outlet port only of the 300bar reducer, the ChargAir
hose allows for the refilling of equipment cylinder(s) from an independent high pressure,
(300bar maximum), secondary breathing air supply source, while the wearer is breathing
from equipment.
TM
The ChargAir hose assembly is constructed from three main elements:

l Reducer connection - Banjo assembly incorporating a high pressure non-return


valve.
l Connection hose - a tightly interwoven stainless steel wire braiding forms the external
construction of this hose. Inside the stainless steel braiding is a P.T.F.E.
(Polytetrafluoroethylene) tube through which high pressure air flows. Swivel end
fittings connect the hose to the banjo and quick release coupling.
l Secondary supply connection - high pressure male quick release coupling.

Lung Demand Valves

Functional variants of the lung demand valves are:

l Type A Positive Pressure with Push-In connector to facepiece.


l Type AE Positive Pressure with M45 Screw-In connector to facepiece.
l Type N Normal Demand with M40 Screw-In connector to facepiece.

Refer to the appropriate Training Manual for details of the selected Lung Demand Valve.

tm 1287 - June 2005


PSS® 100 SERIES 26

PSS® 100 ET

Tally

Pressure
Sensor

l PSS® 500 ET ‘Tally’ Operated


‘Tally’ fitted
Fig. 7
0456

tm 1287- June 2005


PSS® 100 SERIES 27

PSS® 100 E PSS® 100 ET

Pressure
Sensor

l PSS® 100 E ‘Button’ Operated


l PSS® 100 ET ‘Tally’ removed Fig. 8
0454

tm 1287 - June 2005


PSS® 100 SERIES 28

Medium Pressure Hose


With quick release female coupling

Rubber Hose

Demand Valve

Connected to Demand Valve


Fig. 9
0181

tm 1287- June 2005


PSS® 100 SERIES 29

Cylinder refill
high pressure ‘in’

High Pressure - from reducer non- Non-return


return valve ‘closed’ valve ‘open’ Fig. 10
0315/0316

tm 1287 - June 2005


PSS® 100 SERIES 30

tm 1287- June 2005


PSS® 100 SERIES 31

Section 5 Servicing Instructions

tm 1287 - June 2005


PSS® 100 SERIES 32

Section Description
5.1 Contents

5.2 Service and Test Intervals

5.3 Cleaning, Disinfecting and Drying

5.3.1 General

5.3.2 Cleaning and Disinfecting Pneumatic Assembly

5.4 Replacing Sintered Filter

5.5 Replacing HP O-Ring - Pressure Reducer Connector

5.6 Medium Pressure Hose

5.6.1 Removing M.P. Hose from Reducer

5.6.2 Fitting M.P. Hose to Reducer

5.7 Removing and Fitting Pressure Gauge

5.8 Removing and Fitting Dual Pressure Hose

5.9 Replacing Dual Pressure Hose Inlet Filter Elements

5.10 Replacing Dual Pressure Hose Whistle Jet Filter

5.11 Dual Pressure Hose Components

5.11.1 Dis-assembly of Hose Components

5.11.2 Assembly of Hose Components

5.12 Whistle Warning Unit - Re-setting Alarm Activation Point

5.13 Removing and Fitting Bodyguard (E and ET)

5.13.1 Removing Bodyguard

5.13.2 Fitting Bodyguard

5.14 Removal of Harness from Backplate

5.14.1 Removal of Shoulder Harness Adjusting Straps

5.14.2 Removal of Shoulder Pads

tm 1287- June 2005


PSS® 100 SERIES 33

5.14.3 Removal of Waistbelt and Waistbelt Pad

5.15 Assembly of Harness to Backplate

5.15.1 Assembly of Waistbelt and Waistbelt Pad

5.15.2 Assembly of Shoulder Pads

5.15.3 Assembly of Shoulder Adjusting Pads

5.16 Removal of Cylinder Strap from Backplate

5.16.1 Reassembly of Cylinder Strap to Backplate

5.17 Removal of Yoke from Backplate

5.17.1 Assembly of Yoke to Backplate

5.18 Removal of Waistbelt Pivot Plate from Backplate

5.18.1 Assembly of Waistbelt Pivot Plate to Backplate

5.19 Removal of Pivot Plate and Components

5.19.1 Assembling Pivot Plate Components

5.20 Removal of Yoke Adjuster from Backplate

5.20.1 Assembly of Yoke Adjuster to Backplate

tm 1287 - June 2005


PSS® 100 SERIES 34

5.2 Service and Test Intervals


Carry out regular Inspection, Testing and Servicing of equipment in accordance
with this table. Record all data in equipment Log Book. These instructions also
apply to non-used (in Storage) equipment.
Description After Ever y Ever y Ever y
Use Month Year 6 Yrs
Complete Clean and Disinfect m
Equipment Visual Inspection m u
Functional and Leak Testing as m u
defined in Instructions for Use
Flow and Static Tests as per m
Manufacturer's instructions
Bodyguard Clean m
Visual Inspection m u
Functional and Leak Testing as m u m
defined in Instructions for Use
Pressure reading check m
Demand Valve Clean and disinfect as necessar y mn

Pressure Medium Pressure Check m


Reducer PSS 100, PSS 100 E, PSS 100 ET - m
Replace Sintered Filter
Replace H.P. Connector O-Ring m
Basic Overhaul m
(Repair Exchange Scheme)
Cylinder Charge to correct pressure m
Charged Pressure Check. mu
Check Test Date stamped on Cylinder

Re-Cer tification - Cylinder Pressure


Test (According to National
Standards)
Cylinder Valve Overhaul as necessar y, or at time of
Cylinder Re-Cer tification

l Dräger Recommendation - UK.


l COSHH
l Type A - Push-in demand valve apply grease (3331247) around connector O-ring as
required.

tm 1287- June 2005


PSS® 100 SERIES 35

5.3 Cleaning, Disinfecting and Drying


5.3.1 General
Dräger recommend that contaminated and dirty components or assemblies are
carefully cleaned and disinfected then thoroughly dried after use, or as and when
considered necessary by the user.

To ensure correct operational condition of the equipment use only the cleaning and
disinfecting solutions recommended by Dräger. The use of any other product will
invalidate the Dräger warranty and guarantee.

Important Note:
If the equipment is fitted with an integral lung demand it is important to refer
also to the cleaning, disinfecting and drying section of the relevant training
manual.

When using baths to contain cleaning and disinfecting solutions the immersed
components and assemblies must only be agitated manually.

It is important to note that Dräger Do Not recommend the use of any form of
mechanical, electrical or ultrasonic agitation of this product. If the user decides
however to use ultrasonic cleaning baths then it is important that the instructions
supplied by the manufacturer of the bath are strictly followed.

Refer to the Training Manuals for any ancillary units, e.g. Lung Demand Valve.

Safety Note: Refer to manufacturers’ usage instructions when using cleaning and
disinfecting agents. It is important that attention be paid to concentration and
reaction times. Do Not use organic solvents such as Acetone, Alcohol, White
Spirit, Trichloroethylene or similar.

Dräger recommends:

1. Cleaning:
Dräger Safety Wash
1 Litre Safety Wash and Dispenser 3380164
1 Litre Safety Wash Refill 3380165
5 Litre Safety Wash and Dispenser 3380166
5 Litre Safety Wash Refill 3380167

Sekusept
4 Bottles @ 2 litres 7904071

Do Not exceed a temperature of 30 degrees Celsius.

Note: Rinse off cleaning solution in clean water before disinfecting.

tm 1287 - June 2005


1167

PSS® 100 SERIES 36

2. Disinfecting:
Incidur
1 can 6 litres 7904072
1 can 30 litres 7904073

Wipex Cloths (for manual disinfecting)


Pack of 50 3380375

Do Not exceed a temperature of 30 degrees Celsius.

Note: Rinse off disinfecting solution in clean water before drying.

3. Rinsing and Drying


Remove cleaning and disinfecting solutions by rinsing in clean running water, followed
by drying. For pneumatic assembly see Section 5.3.2.
Rinsing - Do Not exceed a temperature of 30 degrees Celsius.
Drying - Do Not exceed a temperature of 60 degrees Celsius.

5.3.2 Cleaning and Disinfecting Pneumatic Assembly


Important Note: If the equipment is fitted with an integral lung demand valve it is important
to refer also to the cleaning, disinfecting and drying section of the relevant training
manual.

Follow these instructions to remove the pressure reducer and hose assembly from
the backplate for cleaning and disinfecting.

Note: Do Not immerse pressure reducer or pressure sensor of electronic monitoring unit
(Bodyguard) in cleaning or disinfecting solutions. Use clean cloth, moistened in cleaning
fluid, to remove dirt/contaminants.

Tools required:
10mm AF Spanner 3336309
5mm Hexagon socket key 3335738

1 Unfasten hose loops on shoulder 2 1


straps and backplate to release
hoses.

Refer to Figure 1.

2 Remove reducer from mounting of


backplate.

Nut and Socket Head Screw


Using 10mm AF spanner and 5mm
AF hexagon socket key remove
locknut (1), socket set screw (2),
and washers. Scrap locknut, Do Not
re-use, replace with new. Fig. 1
0173

tm 1287- June 2005


PSS® 100 SERIES 37

Mounting Pin
Remove split pin (3) from the end
of reducer mounting pin (4). Push
and remove the mounting pin from 3
mounting block.

3 Remove the reducer assembly then


remove the sleeve (5) from hole in
the flange of the reducer body.
Detach mounting block (6) and 4
mounting plate (7) from backplate. 6

4 Use a clean cloth moistened in


cleaning fluid to remove dir t/
contaminants from the pneumatic 5 7
assembly. When clean, remove Fig. 1
cleaning agent residue with clean 0173
cloth moistened with clean water.
5 After cleaning, disinfect components by wiping with Dräger Wipex Cloths, (Refer to
Wipex Cloth Instructions), or use clean cloth moistened with disinfecting solution.
(Refer to Disinfecting Fluid Instructions).

Note: When drying components Do Not exceed a temperature of 60 degrees Celsius.

6 Re-assemble pressure reducer and hose assembly to backplate, taking care not to
damage sleeve (5). Reverse above procedure.

5.4 Replacing Sintered Filter


Tools required:
Screwdriver 3331246

Refer to Figure 2.

1 Using the screwdriver unscrew


sintered filter from bore of cylinder
connector of the pressure reducer.
Discard sintered filter.

2 Check threaded bore in connector


is clean and not damaged then insert
new sintered filter into connector
bore. Using screwdriver tighten
sintered filter (hand tight) into
connector bore. Fig. 2
0193

tm 1287 - June 2005


PSS® 100 SERIES 38

5.5 Replacing HP O-Ring - Pressure Reducer Connector


Tools required:
Screwdriver 3331246
O-Ring Lifter R21402
O-Ring Assembly Tool 3310465

Refer to Figure 3.

1 Using lifter tool (R21402) remove


defective O-Ring from the groove in
the end face of connector. Scrap the
O-Ring.

2 Check that the groove and connector


are clean and undamaged. Insert
and locate tapered pin of assembly
tool into bore of connector. Moisten
O-Ring with water and locate onto
tapered pin. Using female
component of assembly tool, push
O-Ring into groove in connector. Fig. 3
Remove Assembly tool.
0178

5.6 Medium Pressure Hose


If equipment is fitted with an integral lung demand it is important to refer also to the
relevant training manual for the lung demand valve regarding instructions for removal
and fitting of the medium pressure hose.

5.6.1 Removing MP Hose from Reducer


Safety Note: Before proceeding with the removal of the medium pressure hose,
turn ‘OFF’ cylinder valve, vent pressure from the system, unscrew reducer
handwheel from the cylinder valve and remove cylinder from backplate.

Tools Required:
2mm AF Hexagon Socket Key 3337900

1 Unfasten hose loops of the backplate and if necessary from shoulder strap to release
hose from equipment.

tm 1287- June 2005


PSS® 100 SERIES 39

Refer to Figure 4.

2 Using 2mm AF hexagon socket key,


unscrew (anti-clockwise) and remove
two screws (1), remove retaining cap
(2), and retention staple (3). Grip and
pull hose to remove pressure switch
from pressure reducer.

Note: The component parts of the relief 4


valve, i.e. spring (4), O-Ring retainer
(5) and O-Ring (6), will be attached
to the nozzle end fitting of the medium 1
pressure hose.
2 3 6 5
3 Remove O-Ring retainer and O- Fig. 4
Ring. spring, and if fitted, the anti- 0183
vibration strap. Inspect the hose and
if necessary replace.
1 2 3
5.6.2 Fitting MP Hose to Reducer
Tools Required:
2mm AF Hexagon Socket Key
3337900

Refer to Figure 5

1 If required, assemble anti-vibration


strap to the hose with the hook of
anti-vibration strap facing hose
nozzle. Fit spring (1), O-Ring
retainer (2) and new O-Ring (3), to Fig. 5
new hose, ensuring O-Ring locates 0184
into O -Ring retainer (2).

2 Check that the reducer outlet bore is clean and not damaged then insert hose end
fitting into the outlet port. Hold against the compression spring then insert retention
staple (3) through holes in reducer body ensuring correct location into groove in hose
end fitting. Check hose securely retained by pulling hose away from reducer body.

3 Locate groove of retaining cap (2) over the protruding portion of the retention staple
- align screw holes then insert screws (1) and using 2mm AF hexagon socket key
tighten the screws.

4 Connect cylinder, and if necessary the lung demand valve, to the pressure reducer
and apply pressure to the equipment. Close the cylinder valve and carry out Leak Test.

5 Following satisfactory Leak Test secure the hose to backplate, and if necessary to
the shoulder strap, using the hose loop.

tm 1287 - June 2005


PSS® 100 SERIES 40

5.7 Removing and Fitting Pressure Gauge


Tools required:
1.5mm AF Hexagon Key 3331237
Hexagon Adaptor - 2mm 3310600
Torque Key 3310593
Loctite 542 3333622
O-Ring Assembly Tool 3337901

Safety Note: Vent air pressure from pneumatic system before proceeding with disassembly.
Close cylinder valve (if fitted) and vent system.

1 Disconnect dual pressure hose from


hose loops of shoulder strap.

Refer to Figure 6.
2
2 Using the 2.0mm hexagon adaptor 3
and torque key, remove screw (1).
1
Unfold the rubber cover (2) from
around the gauge then slide cover
along the hose. To prevent rotation
of the whistle tube during removal
of the gauge immediately re-insert
the screw (1) into the threaded radial
hole of the whistle tube.

Important Note: Do Not attempt to remove Fig. 6


1149
the set screw (3) with the screw
(1) removed.

3 Using 1.5 AF hexagon socket key


unscrew and remove socket set
screw (3). Grip body of gauge and
unscrew gauge (anti-clockwise)
from end of hose. Fit new O-Ring
(4) using O-Ring assembly tool. See
Fig. 7.

Note: If fitting new gauge check O-Ring (4)


is fitted and not damaged.

4 Check bore of whistle tube is clean


and not damaged. Grip the gauge
then carefully screw (clockwise) into 4
the end of hose until rotation stops -
Do Not tighten. Fig. 7
0175

tm 1287- June 2005


PSS® 100 SERIES 41

5 Carefully unscrew gauge (approximately 45 degrees anti-clockwise) until the face of


the gauge is 90 degrees to flute in whistle tube and axial slot in the shaft of the
gauge is in line with threaded radial hole in whistle tube. Locate new socket set
screw onto 1.5 AF hexagon socket key and screw set screw into whistle tube - ‘nip’
tight.

Note: End of set screw should not protrude past the outer diameter of the whistle tube.

6 Remove screw (1). Slide the rubber


cover along hose and fold around
gauge body. See Fig. 8.

Note: Flute in whistle tube should be visible


and aligned with the hole in the side
of the rubber cover.

7 Hold screw (1) to the 2.0mm 1


adaptor, apply one drop of Loctite
542 to the thread of the screw then
re-insert the screw through the cross
hole in the rubber cover and into the
threaded radial hole of the whistle
tube and tighten the screw to a
torque of 0.7Nm. Fig. 8
1250

8 Before the adhesive sets connect cylinder, and if necessary the lung demand valve,
to the pressure reducer and apply pressure to the equipment. Close the cylinder
valve and carry out Leak Test. Following satisfactory leak test immediately begin to
slowly vent pressure until the whistle activates.

5.8 Removing and Fitting Dual Pressure Hose


Tools Required:
2mm AF Hexagon Socket Key 3337900

Safety Note: Before proceeding with the replacement of hose, turn ‘OFF’ the cylinder
valve, vent pressure from system then unscrew reducer handwheel from cylinder
valve and remove cylinder from backplate.

1 Unfasten hose loops of backplate and left shoulder strap to release hose from
equipment.

tm 1287 - June 2005


PSS® 100 SERIES 42

Refer to Figure 9.

2 Using 2mm AF hexagon socket key 1


unscrew (anti-clockwise) and
remove the two screws (1), remove 2
retaining cap (2) and retention staple
(3). Grip and pull hose (4) to remove 3
from outlet port of reducer.

Fig. 9
0170

Refer to Figure 10. 5 8 6

3 Inspect O-Rings (5) and (6), sealing


ring (7), and replace if necessary.
To replace the three filter elements
(8) refer to Section 5.9.

Note: If fitting new hose check O-Rings (5)


and (6) and sealing ring (7) are fitted
and not damaged.

4 Check the outlet bore of the reducer


7
is clean and undamaged then insert
hose end fitting into outlet port - Fig. 10
insert retention staple through holes 0171
in reducer body ensuring correct
location into groove in hose end
fitting. Locate groove of retaining
cap over the protruding portion of
retention staple - align screw holes
- insert the two screws and using
2mm AF hexagon socket key tighten
screws.

Note: Check hose securely retained by


pulling hose away from reducer body.

5 Carry out High Pressure Leak Test.

tm 1287- June 2005


PSS® 100 SERIES 43

5.9 Replacing Dual Pressure Hose Inlet Filter Elements


Tools Required:
2mm AF Hexagon Socket Key 3337900
Screwdriver 3336222

Safety Note: Before proceeding with replacement of filter, turn ‘Off’ the cylinder valve,
vent pressure from system, unscrew the reducer handwheel from the cylinder
valve and remove the cylinder from the backplate.

Refer to Fig. 11
1
1 Unfasten hose loops of backplate 2
and left shoulder strap to release
hose from equipment.
3
2 Using 2mm AF hexagon socket key
unscrew (anti-clockwise) and 4
remove two screws (1), remove
retaining cap (2), and retention
staple (3). Grip and pull hose (4) to
remove from outlet port of reducer.

Refer to Figure 12.


Fig. 11
3 Using screwdriver 3336222 remove 0170

circlip (1) and washer (2). Remove


and scrap the three filter elements
(3). Inspect O-Ring (4) and replace 4 3 1
if necessary.

4 Assemble three new filter elements


refit the washer then insert the
circlip into groove in nozzle to
secure O-Ring, filter, washer and
capillary tube.

5 Check reducer outlet bore is clean


and undamaged then insert hose
end fitting into outlet port. Insert
2
retention staple through holes in
reducer body ensuring correct Fig. 12
location into the groove in hose end 0171

fitting. Locate groove of retaining


cap over the protruding portion of retention staple - align screw holes - insert the
two screws and using 2mm AF hexagon socket key tighten screws.

Note: Check hose securely retained by pulling hose away from reducer body.

5 Carry out High Pressure Leak Test.

tm 1287 - June 2005


PSS® 100 SERIES 44

5.10 Replacing Dual Pressure Hose Whistle Jet Filter


Refer to Section 5.11, i.e. Dual Pressure Hose Components.

5.11 Dual Pressure Hose Components


5.11.1 Dis-assembly of Hose Components
Tools Required:
2mm AF Hexagon Socket Key 3337900
1.5 AF Hexagon Socket Key 3331237
O-Ring Assembly Tool 3337901
Screwdriver 3336222

Safety Note: Before proceeding with the removal of hose from reducer - turn ‘Off’ cylinder
valve, vent pressure from system then unscrew the reducer handwheel from the
cylinder valve and remove the cylinder from backplate.

1 Disconnect dual pressure hose from


hose loops of shoulder strap. 1

Refer to Figure 13. 2

2 Using 2mm AF hexagon socket key, 3


unscrew (anti-clockwise) and 4
remove two screws (1), remove
retaining cap (2), and retention
staple (3). Grip and pull hose (4) to
remove from outlet port of reducer.

Fig. 13
0170

Refer to Figure 14.

3 Using the 2.0mm hexagon key 1


remove screw (1). Unfold the rubber
cover (2) from around gauge body
and slide cover along the hose. To 3
prevent rotation of the whistle tube 2
during removal of the gauge re-insert
the screw (1) into the threaded radial
hole of the whistle tube.

Important Note: Do Not attempt to remove


the set screw (3) with the screw
(1) removed.
Fig. 14
1149

tm 1287- June 2005


PSS® 100 SERIES 45

4 Using 1.5 AF hexagon socket key


unscrew and remove socket set
screw (3). Grip the gauge and
unscrew (anti-clockwise) from the
end of the hose. Inspect O-Ring (4)
and if necessary fit new O-Ring
using O-Ring assembly tool. See
Fig. 15.

Note: If fitting new gauge check O-Ring (4)


is fitted and not damaged.

Refer to Figure 16

5 Using screwdriver (3336222) 4


remove circlip (1). Remove washer Fig. 15
(2), the three filter elements (3) and 0175
O-ring (4). Scrap the filter elements.

6 Push end (5) of the capillary tube


assembly into the hose in direction ïB
of arrow (A). Remove O-Ring (6).
Opposite end of capillary tube will
protrude from whistle tube. Grip end
(5a) of capillary tube assembly and 3 2
fully withdraw capillary tube in
direction of arrow (B). 5a

Refer to Figure 17.


ïA
8 Using 2.0 AF hexagon socket key
unscrew and remove socket set 4 6 1 5
screw (1). Carefully screw the Fig. 16
gauge (clockwise) two revo-lutions 0176
into end of whistle flute. Grip the
gauge then pull and withdraw the
whistle jet assembly (3) from whistle
flute (2). Remove the face filter (4) 2 3
from whistle jet (3). Scrap the filter
and replace with new. Remove the
gauge from the whistle jet assembly.

Note: This completes dis-assembly of dual


pressure hose. Inspect all
components for damage and replace
with new components where
necessary.

1 4
Fig. 17
0191

tm 1287 - June 2005


PSS® 100 SERIES 46

5.11.2 Assembly of Hose Components

Tools Required:
2mm AF Hexagon Socket Key 3337900
1.5 AF Hexagon Socket Key 3331237
Hexagon Adaptor (2mm) 3310600
Torque Key 3310593
Loctite 542 3333622
O-Ring Assembly Tool 3337901

Important Note: Check all components are clean and undamaged. Replace with new
components where necessary.

Refer to Figure 18.

1 Assemble filter (4) and, if necessary 2 5 3


a new O-Ring (5) to the whistle jet
(3). Insert the whistle jet (3) into
whistle flute (2) aligning the plain
radial hole in whistle jet with
threaded hole (X). Locate screw (1)
onto 2.0 AF hexagon key and screw
fully into whistle tube and plain radial
hole in whistle jet. Do Not apply
Loctite to the screw, or tighten the
screw.
X 1 4
Refer to Figure 19.
Fig. 18
2 If necessary, assemble new O-Ring 0191

(1) and back-up ring (2) into the


groove in end (A) of the capillary
tube assembly. Insert end (B) of the 2 1
capillary tube into the bore of the
whistle tube and keeping the hose
straight, to ensure smooth assembly
of capillary, push the capillary tube
through the bore of hose until end A
(B) protrudes from opposite end
fitting of the hose.
B
3 If necessary fit new O-Ring to the
gauge using O-Ring assembly tool.
Carefully screw the gauge
(clockwise) into end of hose until
rotation stops - Do Not tighten. Fig. 19
0179

tm 1287- June 2005


PSS® 100 SERIES 47

Refer to Figure 20

4 If necessary, locate new O-Ring (7) 5 3 2 1


to recess in the end face of the hose
fitting - use washer (4) to press and
locate O-Ring into recess - remove
washer. Assemble O-Ring (6), three
new filter elements (5), then washer
(4). Insert circlip (3) into groove in
end of capillary tube to secure O-
Ring, filter elements, washer, and 4
capillary tube. Assemble back up
ring (2) and O-Ring (1).
6 7
5 Carefully unscrew gauge
(approximately 45 degrees anti- Fig. 20
clockwise) until the face of the
0171a
gauge is 90 degrees to flute in the
whistle tube and slot in the gauge
shaft is in line with threaded radial
hole in whistle tube. Locate new
socket set screw onto 1.5 AF
hexagon socket key and screw set
screw into whistle tube - ‘nip’ tight.

Note: End of the set screw should not


protrude past the outside diameter of
whistle tube.

6 Assemble and slide the rubber cover


along hose.

Refer to Figure 21.

7 Check reducer outlet bore is clean


and undamaged then insert hose
end fitting into outlet port. Insert the 1
retention staple through holes in
reducer body ensuring correct 2
location into groove in hose end
fitting. Locate groove of retaining 3
cap over the protruding portion of 4
retention staple - align screw holes
- insert the two screws and using
2mm AF hexagon socket key tighten
screws.

Note: Check hose securely retained by


pulling hose away from reducer body.
Fig. 21
0170

tm 1287 - June 2005


PSS® 100 SERIES 48

8 Using the 2.0mm hexagon key


remove screw (1). Slide the rubber
cover along hose and fold around
gauge body. See Fig. 22.

Note: The flute in the whistle tube should


be visible and aligned with the hole
in side of rubber cover.

9 Hold screw (1) to the 2.0mm 1


adaptor, apply one drop of Loctite
542 to the thread of the screw then
re-insert the screw through the cross
hole in the rubber cover and into the
threaded radial hole of the whistle
tube and then using the 2mm Fig. 22
adaptor and torque driver, tighten the
1250
screw to a torque of 0.7Nm.

10 Before the adhesive sets connect


cylinder, and if necessary the lung
demand valve, to the pressure
reducer and apply pressure to the
equipment. Close the cylinder valve
and carry out Leak Test. Following
satisfactory leak test immediately
begin to slowly vent pressure until
the whistle activates.

5.12 Whistle Warning Unit - Resetting Alarm Activation Point


Tools Required
3mm AF Hexagon Socket Key 3335737

1 If necessary attach the lung demand valve to the equipment. Activate the positive
pressure mechanism (lever or button) to switch ‘Off’ positive pressure.

2 Open cylinder valve slowly, but fully, to pressurise pneumatic system. Check reading
of pressure gauge. Close cylinder valve and vent equipment slowly as follows and
note pressure at which whistle activates;

Positive Pressure Lung Demand Valves - Cover outlet port of demand valve with
palm of hand. Press the centre of the rubber cover of the demand valve and slowly
vent equipment by carefully lifting palm of hand to maintain slow steady decrease in
pressure.

Normal Demand Lung Demand Valves - Slowly vent equipment by carefully pressing
the centre of the rubber cover of the demand valve maintaining slow steady decrease
in pressure.

tm 1287- June 2005


PSS® 100 SERIES 49

Note: If whistle does not sound at required


pressure reset as follows:

Refer to Figure 23.

3 Remove anti-tamper plug (1) from


the adjusting cap situated in the
reducer body between the dual
pressure hose and medium pressure
hose connections.

Fig. 23
0182

Important Note: Do Not turn adjuster unless the pressure is vented from the system.

Using a 3mm hexagon key in the socket of the adjusting cap:

l rotate the adjusting cap clockwise to increase the pressure at which whistle
activates.
l rotate the adjusting cap anti-clockwise to decrease pressure at which whistle
activates.

Note: As a guide - one quarter turn of the adjusting cap is equivalent to a pressure change of
approximately 2bar to 3bar. Do Not turn adjusting cap with the pressure ‘On’.

4 Repeat instructions 2 to 3.

5 If necessary repeat instructions until the required whistle activation setting is achieved.

5.13 Removing and Fitting Bodyguard (E and ET)


5.13.1 Removing Bodyguard
Tools Required:
2mm AF Hexagon Socket Key 3337900

Safety Note: Before proceeding with the removal of the Bodyguard hose from the reducer,
turn ‘Off’ the cylinder valve, vent pressure from system then unscrew the reducer
handwheel from cylinder valve and remove the cylinder from backplate.

1 Unfasten hose loops of the backplate and left shoulder strap to release the Bodyguard
hose from the equipment.

tm 1287 - June 2005


PSS® 100 SERIES 50

Refer to Figure 24.

2 Using 2mm AF hexagon socket key, 1


unscrew (anti-clockwise) and
remove the two screws (1), remove 2
retaining cap (2) then retention
staple (3). Grip and pull the hose to 3
remove pressure switch from the
reducer.

Fig. 24
0170

5.13.2 Fitting Bodyguard


Tools Required:
2mm AF Hexagon Socket Key 3337900

Refer to Figure 25.

Note: Check O-Rings and sealing ring are 1


fitted and not damaged. If necessary 2
fit new sealing components.

1 Check outlet port of reducer is clean 3


and not damaged.

2 Insert the pressure switch end fitting


of the Bodyguard into the port of
the reducer - insert retention staple
(3) through holes in reducer body
ensuring correct location into groove
of end fitting of pressure switch.
Check hose securely retained by Fig. 25
pulling hose away from reducer. 0170

3 Locate groove of retaining cap (2) over the protruding portion of the retention staple
- align screw holes then insert screws (1) and using 2mm AF hexagon socket key
tighten the screws.

4 Connect cylinder, and if necessary the lung demand valve, to the pressure reducer
and apply pressure to the equipment. Close the cylinder valve

tm 1287- June 2005


PSS® 100 SERIES 51

5.14 Removal of Harness from Backplate

2 1

5 4

1 Shoulder Adjusting Strap


2 Shoulder Strap and Pad
3 Cylinder Strap
4 Waistbelt Pad
5 Waistbelt
6 Waistbelt Adjusting Strap

Fig. 26
0525

tm 1287 - June 2005


PSS® 100 SERIES 52

Note: Close cylinder valve and vent air from system. Open hose loops of shoulder straps to
release hoses. If necessary, remove cylinder.

5.14.1 Removal of Shoulder Harness Adjusting Straps (1 Fig. 26)

Refer to Figure 27.

1 From rear of the backplate - push


the end of each shoulder adjusting
strap towards the anchor peg at the
base of the backplate and remove
the peg from the strap then the strap
from the backplate. Retain the
anchor peg.

ê
Fig. 27
0764

Refer to Figure 28.

2 Thread end of the shoulder adjusting


strap through buckle.
ç
3 Repeat procedure to remove the
second adjusting strap.

4 Clean and replace as necessary.

Fig. 28
0748

tm 1287- June 2005


PSS® 100 SERIES 53

5.14.2 Removal of Shoulder Pads (2 Fig. 26)


Refer to Figure 29. 1
1 Open shoulder pad hose loops to
release hoses.

2 Thread the end of each shoulder pad


strap through the buckle (1) and
slots in arm of the yoke.

3 Clean or replace as necessary.

Fig. 29
0749/0766

5:14:3 Removal of Waistbelt and


Waistbelt Pad 2 3
(4,5 +6 Fig. 26)

Refer to Figure 30.

1 Pull the waistbelt padding (1) away


from the centre of the pivot plate (2)
ç
to reveal the anchor peg location.
Push the waistbelt strap (3) towards ç
the anchor peg and remove anchor
peg (4). Pull the waistbelt out of 1
ç

the anchor peg location. 4

Fig. 30
0765

Refer to Figure 31.


3
2 Pull waistbelt (1) through loops of
waistbelt pad (2) and remove
waistbelt adjusting strap (3) from
buckle (4).

3 Repeat procedure for both sides of ç


waistbelt.

4 Clean or replace as necessary.

1 2 4
Fig. 31
0767

tm 1287 - June 2005


PSS® 100 SERIES 54

5.15 Assembly of Harness to Backplate


5.15.1 Assembly of Waistbelt and Waistbelt Pad

Note: RH Male Buckle/LH Female Buckle.

Refer to Figure 32.

1 Thread front half of the waistbelt (1) 3


through the metal buckle (2) then
slide assembled waistbelt (3) through
loops of waistbelt padding (4).

1 2 4
Fig. 32
0767

I Refer to Figure 33.


2 3
2 With pad (1) pulled back from the
pivot plate (2) - place end of
waistbelt (3) through anchor peg
location and insert anchor peg (4).

3 Slide the waistbelt pad back over the ç


pivot plate. Repeat for both sides of
the waistbelt. ç
4 Bring waistbelt strap and buckle
1
ç

together and check that the 4


orientation is correct for male/female
buckle.
Fig. 33
0765

tm 1287- June 2005


PSS® 100 SERIES 55

5.15.2 Assembly of Shoulder Pads


(2 Fig. 26)
ç
Refer to Figure 34.

1 LH + RH shoulder pads. Thread end


of shoulder strap carefully through
slots in arm of yoke assembly and
buckle (1). Repeat for both sides
of harness. 1
2 Reassemble hoses to shoulder pads
and close loops to secure. See Fig.
35.

5.15.3 Assembly of Shoulder Fig. 34


Adjusting Straps (1 Fig. 26) 0749/0766

Refer to Figure 35. 2 1

è
è
ç

Fig. 35
0748

1 Thread shoulder adjusting strap (1) 2 1


through metal buckle (2).

Refer to Figure 36. 3


2 Push free end of the shoulder
adjusting strap (1) through anchor
peg location at base of the backplate
(2) and push the anchor peg (3) into
position - ensure that strap is not
twisted. Pull on shoulder adjusting
strap to seat anchor peg securely
into position. Repeat for both sides Fig. 36
of harness.
0764

tm 1287 - June 2005


PSS® 100 SERIES 56

5.16 Removal of Cylinder Strap


from Backplate
Refer to Figure 37. 1 è
1 Remove cylinder from equipment.

2 Open hose loop to release hoses the


push end of the cylinder strap (1)
è
2
through the slot in the backplate (2)
and remove pin (3).

3 With the pin removed - pull the


cylinder strap through backplate. 3
Repeat for both sides. ê
Fig. 37
Refer to Figure 38 0769

4 Thread strap end through cylinder


strap buckle.

5 Cylinder strap is now free for


cleaning or replacement.

5.16.1 Reassembly of Cylinder Strap


to Backplate
Tools required:
Cylinder strap insertion tool
3337765

Fig. 38
0656

Refer to Figure 39.

1 Thread cylinder strap correctly


through buckle.

Fig. 39
0656

tm 1287- June 2005


PSS® 100 SERIES 57

Refer to Figure 40.

2 Using tool 3337765 - thread the


cylinder strap carefully through slot 2
in the top of the backplate. Secure
into position using pin. Repeat for è
both sides of cylinder strap.
Camlock buckle strap into RH slot
of backplate ç
1
Note: For correct orientation - double loop
towards centre of backplate.

3 Reassemble hose to hose loop (4) 4


3
Fig. 40. Close loop. é
Fig. 40
4 Assemble cylinder to backplate as
described in the Instructions for 0769

Use. Check action of the Camlock


and security of the cylinder on the
backplate.

5.17 Removal of Yoke from Backplate


1 Remove shoulder pads. See Section
5.14.2.

Refer to Figure 41.

2 Remove the rubber shock buffer


from the opening ‘A’ of the yoke.
Turn the yoke adjuster (1) through
90 degrees and slide the yoke to 1
opening ‘A’ and align the adjuster
with the opening - remove the yoke.
A

Fig. 41
0501

tm 1287 - June 2005


PSS® 100 SERIES 58

5.17.1 Assembly of Yoke to Backplate


1 Place yoke into top position of the slot in the backplate aligning and locating the
yoke adjuster with diamond shaped hole in yoke. Rotate the adjuster until the yoke
can slide down the slot in the backplate and when the preferred working position is
selected - turn yoke adjuster through 90 degrees to lock yoke into place. Assemble
the rubber shock buffer into the opening ‘A’ of the yoke. Reassemble shoulder
pads. See Section 5.15.2.

5.18 Removal of Waistbelt Pivot Plate from Backplate


Tools required:
2.5mm AF Hexagon Socket Key 3310563

1 If necessary - remove waistbelt and waistbelt pad. See Section 5.14.3

Refer to Figure 42.

2 Using 2.5mm hexagon socket key 1


unscrew screws from the pivot plate
retaining block. While holding the
pivot plate assembly in position
carefully remove screws (1), 2
retaining block (2) and spring plate
(3) from slot in the backplate. Retain
components for reassembly. 3
Replace any worn or damaged
components. Pivot plate assembly
can now be removed from backplate.

Fig. 42
0750

5.18.1 Assembly of Waistbelt Pivot Plate to Backplate


1 Place the pivot plate assembly into slot the in the backplate. Hold in position and
place spring plate and retaining block over the boss of the pivot plate. Insert screws
and tighten using a 2.5mm hexagon socket key.

2 Slide pivot plate along slot checking the security of the assembly and for unrestricted
freedom of movement.

3 Orient correctly and if necessary reassemble the waistbelt and waistbelt pad.
See Section 5.15.1.

tm 1287- June 2005


PSS® 100 SERIES 59

5.19 Removal of Pivot Plate and Components


Tools required:
2.5mm AF Hexagon Socket Key 3310563

1 Remove pivot plate assembly. See Section 5.18.

Refer to Figure 43. 1


2 Squeeze sides of the pivot bumper
(1) to reveal pivot pin (2) then using
a screwdriver or similar object, push
the pivot pin out of the assembly
releasing component parts.

Fig. 43
0774

Refer to Figure 44.

3 Remove nuts (1) from the upper


pivot (2).

4 Inspect and replace any worn or


damaged components.
2

Fig. 44
0772

tm 1287 - June 2005


PSS® 100 SERIES 60

5.19.1 Assembling Pivot Plate


Components
1 Insert new nuts (1) into position on
rear of upper pivot (2) Fig. 44.

Note: Assemble as shown in Fig. 44.

Refer to Figure 45.


2
2 Carefully position the pivot bumper
(1) over the boss of the upper 1
pivot (2).

Fig. 45
0773

Refer to Figure 46.


1
3 While holding the lower pivot (1) in
position on the pivot plate (2) - place
the upper pivot assembly (3) over
the boss of lower pivot.

4 While holding assembly together -


pull back pivot bumper (1) to reveal
pivot pin location and insert pivot pin
(2) into position. Release pivot 3
bumper. See Fig. 47.

2
Fig. 46
0771

5 Check the pivot assembly for


security and unrestricted freedom of 1
movement.

6 Reassemble waistbelt pivot plate to


backplate.
See Section 5.19.1.

Fig. 47
0774

tm 1287- June 2005


PSS® 100 SERIES 61

5.20 Removal of Yoke Adjuster from Backplate


Tools required:
Torque Tool (6Nm) 3310662
Yoke Nut - Torque 3310649
Yoke Screw Tool 3310650

1 If fitted remove the cylinder from the backplate.

2 Remove the rubber shock buffer from the opening at the base of the yoke. Turn the
yoke adjuster through 90 degrees and slide the yoke to opening and align the adjuster
with the opening - remove the yoke.

Refer to Figure 48.

3 Remove cylinder cradle (1) to


access the yoke adjuster screw (2).
2

1 Fig. 48
1022

Refer to Figure 49.

4 Using the yoke screw tool and the


yoke nut tool unscrew and remove
components. Inspect components
and replace as necessary.

1 2 3 4 5

Fig. 49
1192

tm 1287 - June 2005


PSS® 100 SERIES 62

5.20.1 Assembly of Yoke Adjuster to Backplate


Tools required:
Torque Tool (6Nm) 3310662
Yoke Nut - Torque 3310649
Yoke Screw Tool 3310650
Loctite 221 Adhesive 3335722

Refer to Figure 50.

1 Locate yoke screw (5) into the hole


in the backplate with head of screw
located into recess. Assemble
washer (4) to the yoke screw.
1 2 3 4 5
2 Assemble the two disc springs (2)
to the yoke nut (1) ensuring that the Fig. 50
disc springs are orientated correctly 1192

as shown in Fig. 51.

Assemble the yoke adjuster (3) to


the yoke nut (1). Apply Loctite 225
to at least 5mm of thread of the yoke
screw (5). While holding the yoke
screw in position, and using the yoke
screw tool, screw the yoke nut
assembly to the yoke screw. Using
yoke screw tool, the yoke nut tool,
and the torque tool (6Nm), finally
tighten assembly.

3 Reassemble cylinder cradle to the


backplate.
Fig. 51
Refer to Figure 52. 1193

4 Place the yoke into the top position


of the slot in the backplate aligning
and locating the yoke adjuster with
diamond shaped hole in yoke.
Rotate the adjuster until the yoke can
slide down the slot in the backplate
and when the preferred working
position is selected - turn yoke
adjuster through 90 degrees to lock
yoke into place. Assemble the
rubber shock buffer into the opening A
‘A’ of the yoke.

Fig. 52
0501

tm 1287- June 2005


PSS® 100 SERIES 63

Section 6 Test Procedures

tm 1287 - June 2005


PSS® 100 SERIES 64

Section Contents

6.1 Test Procedures

6.2 Cylinder Pressure Check

6.3 Equipment Tests

6.3.1 High Pressure Leak Test - Mechanical Gauge

6.3.2 Whistle Warning Unit Test - Mechanical Gauge and Whistle

6.3.3 High Pressure Leak and Whistle Warning Unit Test - Bodyguard E and ET

6.3.4 Failed Test - Optional Procedure - Bodyguard E and ET

6.3.5 Full Flow Test

6.3.6 Medium Pressure Test - Pressure Reducer


TM
6.3.7 High Pressure Leak Test - ChargAir

tm 1287- June 2005


PSS® 100 SERIES 65

6.1 Test Procedures


Standard test procedures for the PSS® 100 Series of Self Contained Breathing
Equipment (S.C.B.A.) are as follows:

l Cylinder Pressure Check


l High Pressure Leak Test - Mechanical Gauge
l Whistle Warning Unit Test
l High Pressure Leak Test - Bodyguard E and ET
l Whistle Warning Unit Test - Bodyguard E and ET
l Full Flow Test
l Medium Pressure Test - Pressure Reducer

For detailed instructions of the following additional Functional Tests, refer to the
appropriate Test Equipment Manual, (i.e. Test Box, A900 Test Unit, Quaestor).

Lung Demand Valves


Type A and AE
l Balanced Piston Leak Test
l Positive Pressure Balance Test
l Static Pressure Test
l Positive Pressure Switch Over

Type N
l Balanced Piston Leak Test
l Demand Valve Opening Pressure

6.2 Cylinder Pressure Check


Tools Required
High Pressure Test Gauge 3331209

Note: Check the Test Date, stamped on the cylinder by last the Testing Station. If past the
Test Date Do Not use cylinder.

1 If cylinder assembled to backplate, close cylinder valve and vent pressure from the
pneumatic system. Unhook anti-vibration strap (if fitted) from the reducer handwheel
then loosen and extend cylinder strap. Unscrew reducer handwheel from cylinder
valve and rotate cylinder through 90° positioning cylinder towards top of backplate
until valve is approximately 10cm from reducer handwheel.

2 Screw handwheel of high pressure test gauge into outlet port of cylinder valve then
close relief valve on test gauge. Open cylinder valve slowly, and fully. Check reading
on test gauge.

Test Parameters
The cylinder contents pressure must be at least at 80% of maximum working pressure
(indicated on cylinder label) before attempting to perform functional tests of
equipment. If pressure is low then recharge cylinder.

tm 1287 - June 2005


PSS® 100 SERIES 66

3 Close cylinder valve then open the relief valve of the high pressure test gauge to vent
pressure. Remove high pressure test gauge from cylinder valve.

4 Re-assemble the cylinder to the reducer. Hook anti-vibration strap (if fitted) to the
reducer handwheel. Tighten cylinder strap.

5 Proceed to Section 6.3.

6.3 Equipment Tests


Before performing operational tests on equipment, ensure cylinder has been pressure
tested as detailed in Section 6.2. If cylinder pressure low, recharge to required
pressure.

6.3.1 High Pressure Leak Test - Mechanical Gauge


Note: If the equipment has a quick release coupling on the medium pressure hose then
attach demand valve to the female coupling.

1 Activate the reset function of positive pressure demand valves to switch ‘Off’ positive
pressure. Open the cylinder valve slowly, but fully, to pressurise the pneumatic system.
Check pressure reading on pressure gauge.

Note: PA90LDVs: Do Not press centre of rubber cover while pressing reset lever. If an
immediate leak indicated from demand valve - press centre of rubber cover to release
positive pressure mechanism - press reset lever/button to lift positive pressure
mechanism. Repeat action two or three times to eliminate leak.

2 Close the cylinder valve and again observe the reading on the pressure gauge.

Test Parameter
Gauge reading shall not decrease more than 10bar in one minute.

3 If equipment ‘Passes’ the test proceed to Section 6.3.2.

If equipment ‘Fails’ to conform to the test parameter, vent pressure from the system,
investigate source of the leak, rectify and repeat the test. Until the leak is rectified
and the equipment conforms to the requirement of the test parameter, Do Not carry
out any further test.

6.3.2 Whistle Warning Unit Test - Mechanical Gauge and Whistle


Specification
The whistle warning unit is set by Dräger to activate at a pre-set pressure of 55bar
+/- 5bar (60bar to 50bar).

4 Positive Pressure Lung Demand Valves - Cover outlet port of the demand valve
with palm of hand. Press the centre of the rubber cover of the demand valve to
activate positive pressure. Slowly vent equipment by carefully lifting the palm of the
hand to maintain a slow and steady decrease in pressure.

tm 1287- June 2005


PSS® 100 SERIES 67

Normal Demand Lung Demand Valves - Slowly vent equipment by carefully pressing
the centre of rubber cover of the demand valve maintaining a slow and steady decrease
in pressure.

As pressure decreases, observe the pressure gauge and note pressure at which
whistle begins to sound.

Test Parameter
The whistle warning unit should activate an audible signal at the pre-set pressure of
55bar +/- 5bar (60bar to 50bar).

5 a. If whistle setting incorrect, reset as detailed in Section 5.12.


b. If whistle setting satisfactory, go to Section 6.3.5.

6.3.3 High Pressure Leak Test - Bodyguard E and ET


Before performing operational tests on equipment, ensure cylinder has been pressure
tested as detailed in Section 6.2. If cylinder pressure is low, recharge to required
working pressure.

1 Activate the reset function of positive pressure demand valves to switch ‘Off’ positive
pressure. Cylinder valve ‘Closed’.

Note: PA90LDVs: Do Not press centre of rubber cover while pressing reset lever. If an
immediate leak indicated from demand valve - press centre of rubber cover to release
positive pressure mechanism - press reset lever/button to lift positive pressure
mechanism. Repeat action two or three times to eliminate leak.

2 Tally Operated - Remove Tally.


Button Operated - Press LH button

Bodyguard will begin self check sequence.

3 When the leak test icon (Fig. 1) is displayed - press the LH button. Bar graph
segments sweep to the right and the ‘open valve’ icon will be displayed. See Fig. 2.

General Note: If during the self check sequence the LH button is not pressed when the
leak test icon (Fig. 1) is displayed then wait till end of sequence when ‘0’ bar is
displayed. Press and hold down LH button until reverse arrow icon displayed (‘a’
Fig. 6) then release button - Bodyguard will then restart the self check sequence.
Repeat from 3.

4 When the ‘open valve’ icon is displayed ‘Open’ the cylinder valve fully to pressurise
system.

5 On sensing pressure, the display changes to Fig. 3 - ‘actual pressure displayed’,


alternating ‘close valve’ (a) and ‘press RH button’ (b) icons.

tm 1287 - June 2005


PSS® 100 SERIES 68

6 As soon as the valve is fully open then ‘Immediately’ begin to ‘close’ the cylinder
valve. Bar segments will sweep to the right and when only two or one segments
remain then ‘press’ the RH button to start the leak test timing sequence. Display will
change to Fig. 4.

Important Note: During the leak test sequence the pressure reading may reduce by up to
10bar. This may be due to settling of the pressure reducer.

7 During the leak test timing sequence the bar graph segments will sweep to the right
and at the end of the timing sequence either:
‘Passed’ icon (a) displayed (Fig. 5) - equipment ‘Passed’ test. Proceed to instruction
8.
or
‘Failed’ icon (b) displayed (Fig. 5) - equipment ‘Failed’ test. See Note below.

Note: If ‘Failed’ icon (b) is displayed during leak test timing sequence then to determine
whether the failure is due to a slow to settle reducer or an high pressure leak - Do Not
vent system of pressure but ‘immediately’ Go to ‘Failed Test’ - Optional Procedure -
Section 6.3.4.)

8 Continue to test Whistle Warning Signal - Venting Procedure.

9 Venting Procedure
Positive Pressure Lung Demand Valves - Cover the outlet port of the demand valve
with palm of hand. Press centre of the rubber cover of the demand valve to activate
positive pressure. Slowly vent equipment by carefully lifting palm of the hand to
maintain a slow and steady decrease in pressure.

Normal Demand Lung Demand Valves - Slowly vent equipment by carefully pressing
centre of the rubber cover of the demand valve maintaining a slow and steady decrease
in pressure.

10 During venting observe the display. The whistle alarm signal (rapid ‘Bleep’) should
become audible at the pre-set pressure level (55bar +/- 5bar - 60bar to 50bar) and
continue down to 10bar. Continue slowly venting until the pressure and time displays
indicate ‘0’ bar.

Note: Using a ‘snap on’ I.R. Link and Widows based Software Package Option, the pressure
setting can be adjusted. Contact Dräger for details.

11 Tally Operated - Refit ‘Tally’. Bodyguard will switch ‘Off’.


Button Operated - Bodyguard will switch ‘Off’ if no further action is taken within 15
seconds. Alternatively, press and hold down LH button - display sequences icons
(a) and (b) Fig. 6. When icon (b) is displayed, release button, Bodyguard switched
‘Off’.

12 Following satisfactory whistle alarm check, ensure positive pressure mechanism of


the lung demand valve is switched ‘Off’.

tm 1287- June 2005


PSS® 100 SERIES 69

Fig. 1 Fig. 4
0475 0483

Fig. 2 Fig. 5
0481 0484

Fig. 3 Fig. 6
0485 0591

tm 1287 - June 2005


PSS® 100 SERIES 70

6.3.4 ‘Failed Test’ - Optional Procedure


1 Following the display of ‘Failed’ icon (b) Fig. 5, during leak test timing sequence,
observe display of Bodyguard until the self check sequence automatically restarts
and when leak test icon (Fig. 1) displayed - press LH button. Bar graph segments
sweep to the right and the ‘open valve’ icon is displayed. See Fig. 3.
General Note: If during the self check sequence the LH button is not pressed, when the leak
test icon is displayed, wait till end of sequence. Press and hold down LH button until
reverse arrow icon displayed (‘a’ Fig. 7) then release button - Bodyguard will then
restart the self check sequence.
2 Again ‘open’ the cylinder valve fully to boost pressure in system. Display changes to
Fig. 3 - actual pressure displayed, alternating ‘close valve’ (a) and ‘press RH button’
(b) icons.
3 As soon as the valve is fully open then ‘Immediately’ begin to ‘close’ the cylinder
valve. The Bar segments will sweep to the right and when only two or one segments
remain then ‘press’ the RH button to start the leak test timing sequence. Display
changes to Fig. 4.
4 During the leak test timing sequence the bar graph segments will sweep to the right
and at end of timing sequence either:
If ‘Passed’ icon (a) Fig. 5, displayed at end of timing sequence - equipment ‘Passed’
test, continue to test Whistle Warning Signal - Venting Procedure instruction 5.
or
If ‘Failed’ icon (b) Fig. 5 is again displayed during leak test timing sequence - ‘Close’
the valve, vent pressure from system and switch ‘Off’ Bodyguard. Investigate source
of leak, rectify and repeat High Pressure Leak and Whistle Warning Test.
5 Venting Procedure
Positive Pressure Lung Demand Valves - Cover outlet port of the demand valve with
palm of hand. Press centre of the rubber cover of the demand valve to activate
positive pressure. Slowly vent equipment by carefully lifting palm of hand to maintain
a slow and steady decrease in pressure.

Normal Demand Lung Demand Valves - Slowly vent equipment by carefully pressing
centre of rubber cover of the demand valve maintaining a slow and steady decrease in
pressure.

6 During venting observe the display - the whistle alarm signal (rapid ‘Bleep’) should
become audible at the pre-set pressure level (see Test Parameters - Section 6.4)
and continue down to 10bar. Continue slowly venting until pressure and time displays
indicate ‘0’bar.

7 Tally Operated - Refit ‘Tally’. Bodyguard will switch ‘Off’.


Button Operated - Bodyguard will switch ‘Off’ if no further action is taken within 15
seconds. Alternatively, press and hold down LH button - display sequences icons (a)
and (b) Fig. 6. When icon (b) is displayed, release button, Bodyguard switched ‘Off’.

8 Following satisfactory whistle alarm check, ensure positive pressure mechanism of


the lung demand valve is switched ‘Off’.

tm 1287- June 2005


PSS® 100 SERIES 71

6.3.5 Full Flow Test


Safety Note: During full flow venting check Do Not direct outlet of demand valve onto
face, eyes, or skin.

1 Activate the reset function of positive pressure demand valves to switch ‘Off’ positive
pressure. Open the cylinder valve slowly but fully to pressurise the pneumatic system.

Note: PA90LDVs: Do Not press centre of rubber cover while pressing reset lever. If an
immediate leak indicated from demand valve - press centre of rubber cover to release
positive pressure mechanism - press reset lever/button to lift positive pressure
mechanism. Repeat action two or three times to eliminate leak.

2 Press centre of the rubber cover of the demand valve. On positive pressure types
this makes positive pressure active. See Safety Note.

Test Parameter
An unobstructed air flow should vent to atmosphere from the outlet of the demand
valve. See Safety Note. This test should take a minimum of 3 - 5 seconds.

3 Positive Pressure Lung Demand Valves - To switch ‘off’ air flow and positive pressure
- activate the positive pressure ‘Off’ mechanism.

4 Close the cylinder valve then vent pressure from the system.

6.3.6 Medium Pressure Test


Tools Required:
Medium Pressure Test Gauge 3331771

Note: Certified equipment incorporate a pre-set and sealed pressure reducer assembly. The
Dräger guarantee is void should original seal caps be tampered with, removed, or
broken. The guarantee of correct operational condition of the pressure reducer is only
valid if Dräger service, and re-seals the reducer.

Demand Valves with Male Quick Release Coupling

1 Detach demand valve from the quick release coupling of the medium pressure hose.
Open the cylinder valve slowly and fully.

2 Insert the test gauge into female quick release coupling of the medium pressure
hose from the reducer. Check the pressure reading on the test gauge.

Test Parameters
At a system inlet pressure (cylinder pressure) of >100bar the test gauge reading
should indicate a medium pressure of between 6bar to 9bar after 5 seconds. After
one minute it is acceptable for the medium pressure to rise to 9.7bar.

4 Following the test ‘Close’ the cylinder valve and disconnect the test gauge from the
quick release coupling.

tm 1287 - June 2005


PSS® 100 SERIES 72

5 Activate the reset function of positive pressure demand valves to switch ‘Off’ positive
pressure and connect demand valve to the female quick release coupling. Press the
centre of rubber cover of demand valve venting pressure from the system. Following
venting activate the reset function of positive pressure demand valves to switch ‘Off’
positive pressure.

Note: If reducer ‘fails’ the medium pressure test then return the reducer, with hoses and any
accessories removed, to Dräger Service for overhaul under the Repair Exchange Scheme
(REX). Details of REX Scheme are available on request.

Demand Valves attached to Reducer


Tools Required:
2mm AF Hexagon Socket Key 3337900
Medium Pressure Test Gauge 3331771
Medium Pressure Hose Assembly 3337650

1 Ensure cylinder valve is ‘closed’ then


vent pressure from the system.

2 Slide the anti-vibration strap (if


fitted) along the medium pressure
hose and away from reducer.
Refer to Figure 7.

3 Using 2mm AF hexagon socket key,


unscrew (anti-clockwise) and 1
remove the two screws (1), remove
retaining cap (2), then retention
staple (3). Grip and pull the hose
(4) of the demand valve and remove 4
from outlet port of the reducer.
2 3 Fig. 7
Note: The relief valve components - spring 0183
(5), O ring retainer (6) and O ring
(7) will remain attached to nozzle.
Do Not remove components.

tm 1287- June 2005


PSS® 100 SERIES 73

Refer to Figure 8.

4 Using 2mm socket key, unscrew 1


(anti-clockwise) and remove two
screws (1) from test hose manifold,
remove retaining cap (2) and 2
retention staple (3). Insert end fitting
of medium pressure hose of demand 3
valve into manifold port of test hose,
holding against compression spring
- insert retention staple (3) through
holes in manifold ensuring correct
location into groove in hose end
fitting.

Note: Check that the hose is securely Fig. 8


retained by pulling the hose away from
0306
manifold.

5 Locate groove of the retaining cap


over protruding portion of retention
staple - align holes - insert screws
and using 2mm socket key, tighten
screws.

Refer to Figure 9.

6 Insert non return valve (1) into bore


of test hose manifold. Ensure O ring
(2) is fitted then insert connector into
outlet port of reducer. Insert retention
staple (3) into holes in reducer body
ensuring correct location into groove
of manifold connector.

5 4 3 1 2
Fig. 9
0307

Note: Check that the manifold is securely retained by pulling manifold away from reducer body.

7 Locate groove of retaining cap (4) over protruding portion of retention staple - align
screw holes - insert screws (5) and using 2mm socket key tighten the screws.

8 Activate the reset function of positive pressure demand valves to switch ‘Off’ positive
pressure then ‘open’ cylinder valve slowly, and fully.

9 Insert test gauge into female quick release coupling on medium pressure test hose
from the reducer. Check the pressure reading on the test gauge.

tm 1287 - June 2005


PSS® 100 SERIES 74

Test Parameters
At a system inlet pressure (cylinder pressure) of >100bar the test gauge reading
should indicate a medium pressure of between 6bar to 9bar after 5 seconds. After
one minute it is acceptable for the medium pressure to rise to 9.7bar.

9 Following the test ‘Close’ the cylinder valve. Press centre of the rubber cover of the
demand valve venting pressure from system. Following venting activate the reset
function of positive pressure demand valves to switch ‘Off’ positive pressure.

10 Disconnect test gauge from quick release coupling.

11 Remove test hose from reducer port - remove demand valve hose from manifold of
test hose, and reassemble demand valve hose to the outlet port of reducer. Carry
out High Pressure Leak Test.

Note: If reducer ‘fails’ medium pressure test then return the reducer, (hoses and any accessories
TM
removed, except ChargAir hose), to Draeger Service for overhaul under the Repair
TM
ExchangeTMScheme (REX). Reducers (300bar) with ChargAir must be returned with
ChargAir hose fitted. Details of REX Scheme are available on request.
TM
6.3.7 High Pressure Leak Test - ChargAir
TM TM
ChargAir Filling Hose and Pressure Reducer with ChargAir

Tools Required
Filling Hose Assembly with Gauge 3336641
Second Fully Charged Valved Cylinder Assembly (300bar)

Note: If the equipment has a quick release coupling on the medium pressure hose then
attach demand valve to the female coupling.
TM
1 Screw the handwheel of the ChargAir Filling Hose into the cylinder valve of the
second (fully charged) valved cylinder assembly (300bar). Open slowly, but fully,
the cylinder valve to pressurise the Filling Hose. Check pressure reading on pressure
gauge.

2 Close cylinder valve of second valved cylinder assembly and observe pressure gauge.

Test Parameter
Gauge reading shall not decrease more than 10bar in one minute.

Note: If Filling Hose fails to conform to instruction 3, investigate source of leak, rectify and
repeat test. Until leak is rectified, Do Not carry out any further test.

3 Carry out High Pressure leak test of equipment as detailed in Section 6.3.1.

4 Close cylinder valve of equipment then activate the reset function of positive pressure
demand valves to switch ‘Off’ positive pressure.

tm 1287- June 2005


PSS® 100 SERIES 75

5 Remove protective
TM
caps from male
ChargAir coupling TM of equipment
and female ChargAir coupling of
filling hose. Push male and female
couplings together until a loud ‘click’
is heard indicating a secure
connection.
2
Note: A ‘dark green’ ring (1) visible behind
the ‘grey’ sleeve (2) of the female 1
coupling indicates a secure
connection. See Fig. 10.

6 Open the cylinder valve of second


valved cylinder assembly and
observe the pressure gauge of the Fig. 10
filling hose and pressure gauge of 0090
the equipment. Following pressure
equalisation both gauges should
indicate the same pressure.

7 Close cylinder valve of second valved cylinder assembly and observe the pressure
gauge of the filling hose and the pressure gauge of the equipment.

Test Parameter
Gauge reading shall not decrease more than 10bar in one minute.

8 If equipment fails to conform to test parameters investigate source of leak, rectify


and repeat test. Until leak is rectified, Do Not carry out any further test.
TM
Note: If reducer assembly ‘fails’ high pressure test due to faultTMof ChargAir return reducer,
(hoses and any accessories removed, except ChargAir hose), to Dräger Service for
overhaul TMunder the Repair Exchange Scheme (REX). Reducers (300bar) with
TM
ChargAir must be returned with ChargAir hose fitted. Details of REX Scheme are
available on request.

9 Check that the cylinder valve of the second valved cylinder assembly is ‘closed’.
TM
Disconnect ChargAir male and female coupling by pulling ‘grey’ sleeve of female
coupling away from male coupling.

Note: As couplings separate, a ‘hiss’ or ‘pop’ may be heard as non-return valves in both
couplings seal ‘off’ high pressure.

10 Assemble protective cap of filling hose to female coupling; press centre of cap to
vent pressure. Unscrew filling hose handwheel from cylinder valve of second valved
cylinder assembly. Press centre of the rubber cover of the demand valve, venting
pressure from equipment. Reset demand valve to positive pressure ‘Off’ position.

11 Test completed.

tm 1287 - June 2005


PSS® 100 SERIES 76

tm 1287- June 2005


PSS® 100 SERIES 77

7 Fault Location

tm 1287 - June 2005


PSS® 100 SERIES 78

Section Contents
7 Fault Location

7.1 Pressure Reducer

7.2 Demand Valve

7.3 Whistle Warning Unit - Mechanical

7.4 Facepiece

7.5 Bodyguard - E and ET

tm 1287- June 2005


PSS® 100 SERIES 79

7.1 Pressure Reducer


Certified equipment incorporates Dräger pre-set and sealed pressure reducer
assemblies.

The Dräger guarantee is void should original seal caps be tampered with, removed,
or broken. Guarantee of correct operational condition of the pressure reducer is
only valid if Dräger service, and re-seals the reducer. Return faulty units (hoses and
any accessories removed) to Dräger Service, Blyth.

Service life is six years. Dräger recommends that at the end of this period the reducer
(hoses and any accessories detached) be returned to Dräger Service, Blyth, for
overhaul under the Repair Exchange Scheme (REX). Details of REX Scheme are
available on request.

FAULT CAUSE REMEDY

Leak from pressure relief 1 Damaged O-ring Replace O-ring


valve of medium pressure 2 Weak Spring Replace spring
hose 3 Damaged O-ring Replace retainer
retainer

Leak from pressure relief Reducer failure Remove hoses from reducer
valve of medium pressure and return reducer to
hose. Relief valve Dräger. (REX)
components O.K.

High Medium Pressure Reducer out of Remove hoses from reducer


specification and return reducer to Drâger.
REX)

Low Medium Pressure Reducer out of Remove hoses from


specification reducer and return reducer
to Drâger.(REX)

Medium pressure High High Pressure Leak Remove hoses from


after one minute reducer and return reducer
Continuous creep to Drâger.(REX)

tm 1287 - June 2005


PSS® 100 SERIES 80

7.2 Demand Valve


For detailed instructions for Inspection, Testing and Servicing, refer to the appropriate
Lung Demand Valve Service/Training Manual.

Fault Cause Remedy


Inward/Outward leak when Damaged/faulty Replace O-Ring
in facepiece por t connection O-Ring

Damaged/faulty facepiece Replace facepiece por t


por t
Leak - positive pressure 1 Banjo seals leaking Replace seals
'off'

2 Balanced piston unit Replace balanced piston


leaking unit
High static pressure Balanced piston lever Replace springs
springs too strong

Low static pressure Balanced piston lever Replace springs


springs too weak
Demand valve fails to 1 Bent/damaged balanced Replaced balanced piston
switch automatically to piston lever unit
positive pressure mode
2 Bent/damaged retaining Replaced balanced piston
spring plate unit
No supplementar y supply Bent/damaged balanced Replaced balanced piston
piston lever unit

tm 1287- June 2005


PSS® 100 SERIES 81

7.3 Whistle Warning Unit - Mechanical

Fault Cause Remedy


Whistle sounds at Incorrect whistle setting Re-set whistle (Section
wrong pressure 5.14.)
Whistle sounds 1 Damaged HP piston seat 1, 2 or 3 - Remove
continuously hoses from reducer and
return to Dräger (REX)

2 Piston seal leaks


3 Plunger housing seal leak
(MP)
Poor sounding 1 Blocked whistle tube flute Clean or replace whistle
whistle and/or jet tube
2 Blocked filter/filters Replace filter/filters

3 Split or per forated rubber Replace hose


hose of dual pressure hose
(leak)
Whistle does not 1 Blocked whistle tube flute Clean or replace whistle
sound and/or jet tube
2 Faulty activation mechanism Remove hoses from
reducer and return
reducer to Dräger (REX)

3 Blocked filter/filters Replace filter/filters

tm 1287 - June 2005


PSS® 100 SERIES 82

7.4 Facepiece
For detailed instructions for Inspection, Testing and Servicing, refer to the appropriate
Facepiece Service/Training Manual.

Fault Cause Remedy


Whistle sounds at Incorrect whistle setting Re-set whistle (Section
wrong pressure 5.14.)
Whistle sounds 1 Damaged HP piston seat 1, 2 or 3 - Remove
continuously hoses from reducer and
return to Dräger (REX)
2 Piston seal leaks
3 Plunger housing seal leak
(MP)

Poor sounding 1 Blocked whistle tube flute Clean or replace whistle


whistle and/or jet tube
2 Blocked filter/filters Replace filter/filters

3 Split or per forated rubber Replace hose


hose of dual pressure hose
(leak)

Whistle does not 1 Blocked whistle tube flute Clean or replace whistle
sound and/or jet tube
2 Faulty activation mechanism Remove hoses from
reducer and return
reducer to Dräger (REX)
3 Blocked filter/filters Replace filter/filters

tm 1287- June 2005


PSS® 100 SERIES 83

7.5 Bodyguard E and ET

FAULT or FAULT ICON CAUSE REMEDY

Low Battery Replace battery. Refer to


Instructions for Use for
correct procedure and
battery specification

High Pressure Leak Rectify cause of leak and


Re-Test

Display showing Fault Instrument Failure Disconnect from equipment


Codes suffixed with X and return Bodyguard to
(See Example Fig. 1) Dräger Ser vice

Bodyguard will not Low Cylinder Pressure. Recharge cylinder to


switch 'ON' (Below 10bar) maximum working pressure

Low Battery Replace battery.


Refer to Instructions for
Use for correct procedure
and battery specification

Unknown Disconnect from equipment


and return Bodyguard to
Dräger Ser vice

Other Fault Codes


S; T; P; H; v; C; r
0684

tm 1287 - June 2005


PSS® 100 SERIES 84

tm 1287- June 2005


PSS® 100 SERIES 85

Section 8 Accessories

tm 1287 - June 2005


PSS® 100 SERIES 86

Section Description
8.1 Contents

8.2 Decontamination Hose and Secondary Hose Accessories

8.2.1 Fitting Decontamination or Secondary Hose Assembly

8.3 Belt Manifold Accessory

8.3.1 Fitting Belt Manifold Assembly

tm 1287- June 2005


PSS® 100 SERIES 87

8.2 Decontamination Hose and Secondary Hose Accessories.


Description and Intended Use
The Decontamination Hose Accessory and Secondary Hose Accessory connect to
the medium pressure outlet of approved Plus and PSS® 100 Series type pressure
reducers.

The hose configurations provide a connection for the equipment existing lung demand
valve plus a second connection, fitted with either a male or female quick release
coupling.

Male Coupling - Decontamination Hose


For connecting:

1. Secondary air supply hose fitted with female quick release coupling.
(Decontamination Procedure).

Female Coupling - Secondary Hose


For connecting:

1. Second lung demand valve and hose assembly. (Buddy Breather).

2. Secondary air supply hose fitted with male quick release coupling.
(Decontamination Procedure).

Safety Warning: When using Secondary Hose Assembly with a secondary air supply hose,
for Decontamination Procedure, inlet air pressure must not exceed 10bar. Do
Not connect high pressure air to female coupling.

Full Technical Specification available from Dräger.

Decontamination Hose and Secondary Hose

0225/0226

tm 1287 - June 2005


PSS® 100 SERIES 88

8.2.1 Fitting Decontamination or Secondary Hose Assembly


Tools Required
2mm AF Socket Key 3337900

1 Close cylinder valve of equipment


and vent system of air. Unscrew
reducer handwheel from cylinder
valve. Loosen cylinder strap, slide
cylinder away from reducer and
remove cylinder from equipment.

Refer to Figure 1.

2 Using 2mm socket key, unscrew


(anti-clockwise) and remove two 1
screws (1) from reducer body
remove retaining cap (2) and 4
retention staple (3). Grip and pull
medium pressure hose (4) pull to
remove from outlet port of reducer. 2 3 Fig. 1
0228
Note: Component parts of relief valve, i.e.
spring, O ring retainer and O ring, 1
remain attached to nozzle of medium
pressure hose (4). Do Not remove
components. 2

Refer to Figure 2. 3
3 Using 2mm socket key, unscrew
(anti-clockwise) and remove two
screws (1) from accessory hose 4
manifold, remove retaining cap (2)
and retention staple (3). Insert end
fitting of medium pressure hose into
manifold port - push and hold against
compression spring - insert retention Fig. 2
staple (3) through holes in manifold. 0306
Fully insert retention staple ensuring
correct location into groove in hose
end fitting.

Note: Check hose securely retained by pulling hose away from manifold.

4 Locate groove of retaining cap (2) over protruding portion of retention staple - align
holes - insert both screws (1) and using 2mm socket key, tighten the screws.

tm 1287- June 2005


PSS® 100 SERIES 89

Refer to Figure 3.

5 Insert non-return valve (1) into bore


of accessor y hose manifold.
Ensuring O ring (2) is fitted then
insert manifold connector into outlet
port of reducer. Insert retention
staple (3) into holes in reducer body.
Fully insert retention staple ensuring
correct location into groove of
manifold connector.

Note: Check manifold securely retained by


pulling manifold away from reducer
body. 5 4 3 1 2
Fig. 3
6 Locate groove of retaining cap (4) 0307
over protruding portion of retention
staple - align screw holes - insert
both screws (5) and using 2mm
socket key, tighten screws.

7 Position the manifold and locate the


accessory hose through the loop of
the waistbelt. See Fig. 4.

8 Assemble compressed air cylinder


to equipment and carry out Leak Test
as detailed in relevant Instructions
for Use. Also see Section 6.3.

Fig. 4
0622

Order List
Secondary Hose 3337650
Decontamination Hose 3337652

8.3 Belt Manifold Accessory


Description and Intended Use
The Belt Manifold Assembly connects to the medium pressure outlet of approved
Plus and PSS® 100 Series pressure reducers.

This configuration provides a connection for equipment existing lung demand valve
plus two additional connections:

tm 1287 - June 2005


PSS® 100 SERIES 90

Refer to Fig. 5

1 Male quick release coupling for


fitting of airline extension hose.
(Independent source of air).
2 Female quick release coupling
for fitting ofairline extension hose to
second breathing equipment wearer.

With equipment cylinder valve turned


‘Off’, and male coupling of belt
manifold connected to an
independent source of air, (e.g.
Dräger Airpak system or works
airline), equipment functions as an
airline working set. Fig. 5
0623

On failure, interruption or turning ‘off’ of independent air supply, the wearer opens
the equipment cylinder valve, disconnects independent air supply from belt manifold
and either, a. continues working using available air from equipment cylinder, or b.
goes on to a safe area.

Safety Warning: When using independent source of air the inlet pressure must not exceed
10bar. Duration of equipment commences from time of opening cylinder valve
and disconnecting airline supply. Time remaining from equipment is dependent
on volume of air remaining in cylinder.

Details of equipment variants, accessories, and configurations of approved


independent air supply source, (e.g. Dräger Airpak), are available from Dräger on
request.

Technical Data
Airline Operating Pressures and Air Quality
Independent source of air should conform to the following parameters when using
compressed air respiratory protection equipment fitted with the Belt Manifold
Accessory.

l Connected to male coupling - One User - 6bar to 10bar with a continuous


air flow of at least 550 Litres/minute.

l Connected to male coupling and second man connected to female coupling


- Two Users - 7bar to 10bar with a continuous air flow of at least 550 Litres/
minute.

Air quality should conform to requirements of the latest edition of EN12021.

Safety Note: Check independent air supply complies with air quality, pressure, and flow
requirements (see Technical Data).

tm 1287- June 2005


PSS® 100 SERIES 91

8.3.1 Fitting Belt Manifold Assembly

Tools Required
2mm AF Socket Key 3337900
Posi-Driv Screwdriver (No.2) 3331245

1 Close cylinder valve of equipment


and vent system of air. Unscrew
reducer handwheel from cylinder
valve. Loosen cylinder strap, slide
cylinder away from reducer and
remove cylinder from equipment.

Refer to Figure 6.

2 Using 2mm socket key, unscrew 4


(anti-clockwise) and remove two
screws (1) from reducer body
remove retaining cap (2) and
retention staple (3). Grip and pull
medium pressure hose (4) pull to
remove from outlet port of reducer. 1 2 3 Fig. 6
0228
Note: Component parts of relief valve, i.e.
spring, O ring retainer and O ring,
remain attached to nozzle of medium 2
pressure hose (4). Do Not remove
components.

Refer to Figure 7

3 Using posi-driv screwdriver, remove


two screws (1), then nuts (2)
separating retaining block halve (3) 3
from belt manifold. See Fig. 7.
1
4 Locate hose of belt manifold through 4
waistbelt loops and position
manifold as shown in Fig. 8. Fig. 7
0172

tm 1287 - June 2005


PSS® 100 SERIES 92

Refer to Fig. 8

5 Using retaining block halve, both


screws and nuts secure the belt
manifold to the waistbelt.

Refer to Figure 9.

6 Using 2mm socket key, unscrew


(anti-clockwise) and remove two Fig. 8
screws (1) from manifold, remove 0623

retaining cap (2) and retention staple


(3). Insert end fitting of medium 1
pressure hose (4) into manifold port
and hold against compression spring
- insert retention staple (3) through 2
holes in manifold. Fully inser t
retention staple ensuring correct 3
location into groove in hose end
fitting. 4
Note: Check hose securely retained by
pulling hose away from manifold.

7 Locate groove of retaining cap (2)


over protruding portion of retention
staple - align holes - insert the Fig. 9
screws (1) and using 2mm socket 0306

key, tighten screws.

Refer to Fig. 10.

8 Insert non-return valve (1) into bore


of manifold. Ensuring O ring (2) is
fitted then insert manifold connector
into outlet port of reducer. Insert
retention staple (3) into holes in
reducer body. Fully insert retention
staple ensuring correct location into
groove of manifold connector.

Note: Check manifold securely retained by 1 2 3 4


pulling manifold away from reducer
body. Fig. 10
0307

tm 1287- June 2005


PSS® 100 SERIES 93

9 Locate groove of retaining cap (4) over protruding portion of retention staple - align
screw holes - insert both screws (5) and using 2mm socket key, tighten screws.

10 Assemble compressed air cylinder to equipment and carry out Leak Test as detailed
in relevant Instructions for Use.

11 Following satisfactory Leak Test, carry out the following Flow Test.

Medium Pressure Flow Test


12 Check cylinder valve closed. Switch ‘Off’ the positive pressure of the lung demand
valve.

13 Connect female quick release coupling, on hose from independent air supply, to
male coupling on belt manifold. Turn ‘On’ independent air supply.

Note: If required connect second man equipment to female coupling of belt manifold.

14 Press centre of rubber cover of lung demand valve/s. See Safety Warning.

Note: Keep the centre of rubber cover of normal demand, N type lung demand valve,
depressed. Positive pressure types will activate to positive pressure. An unobstructed
air flow should vent from outlet of the demand valve. This test should take a minimum
of 3 to 5 seconds.

Safety Warning: Do Not direct air flow from outlet of demand valve onto face, eyes or
skin.

15 Switch ‘Off’ positive pressure of lung demand valve.

16 Turn ‘Off’ the independent air supply and disconnect the supply hose from the male
coupling of the belt manifold.

Note: If second man equipment attached then remove from female coupling of belt manifold.

Order Code
Belt Manifold Assembly 3337600

tm 1287 - June 2005


PSS® 100 SERIES 94

tm 1287- June 2005


PSS® 100 SERIES 95

Section 9 Spare Parts Lists and Tool List

tm 1287 - June 2005


PSS® 100 SERIES 96

Spare Parts Lists


Number Title
1281.5 High Pressure Test Gauge

1282.5 Medium Pressure Test Gauge

1287 PSS® 100 Series

1287.1 PSS® 100 Series - Pneumatic Arrangement

1287.2 PSS® 100 Series - Pneumatic arrangement with ChargAirTM

1285.13 Whistle Tube and Gauge Hose

1285.17 Decontamination Hose (Female)

1285.18 Decontamination Hose (Male)

1289.1 Electronic Monitoring Unit - Bodyguard

Tool List
Number Title
1287.3 Tool List

tm 1287- June 2005


PA90 SERIES - LUNG DEMAND VALVE 1

Index

Section 1 Instructions for Use

Section 2 Introduction and Safety Warnings

Section 3 Product Description

Section 4 Operating Principle

Section 5 Servicing Instructions

Section 6 Test Procedures

Section 7 Fault Location

Section 8 Spare Parts List and Tool List

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 2

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 3

Section 1 Instructions for Use

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 4

Positive Pressure Lung Demand Valve – Type A Series 3337392

Positive Pressure Lung Demand Valve – Type AE Series 3337393

Normal Demand Lung Demand Valve – Type N Series 3337394

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 5

Section 2 Introduction and Safety Warnings

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 6

Introduction
The PA90 Series of Lung Demand Valves are approved for use with approved Dräger
Compressed Air Breathing Equipment (C.A.B.E.) and Facepieces.

Refer to Illustration

The combination of Lung Demand Valve, C.A.B.E. and Facepiece provides the wearer of the
equipment with respiratory protection when working in or escaping from a contaminated or
oxygen deficient life-threatening environment.

Safety Warnings

l Explain - Instructions for Use, supplied with the equipment must be adhered to.

l Maintain - the respiratory protection equipment must be maintained as outlined in the


Dräger Limited service schedules and maintenance instructions.

Contact Dräger Limited, Blyth for details of Service Contracts and Service Training
Courses.

l Train potential wearers in how to wear and use equipment and to understand its
limitations.

Contact Dräger Limited, Blyth for details of Wearer Training Courses.

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 7

Positive Pressure - Type A Illustrated

1241

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 8

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 9

Section 3 Product Description

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 10

Product Description
The variants of the PA90 Series Lung Demand Valve are as follows:

l Type A - Positive Pressure with Push-In connector to facepiece.

l Type AE - Positive Pressure with M45 Screw-In connector to facepiece.

l Type N - Normal Demand with M40 (Rd40) Screw-In connector to facepiece.

The high performance Lung Demand Valves incorporate a precision factory set and sealed
balanced piston unit that has unlimited service life.

The Dräger Warranty and Guarantee is void should any attempt be made by customers to
make adjustment to the balanced piston unit.

Detachable variants of the lung demand valves have a medium pressure hose with a quick
release male coupling to connect the demand valve to the female coupling of the medium
pressure hose from the pressure reducer of the appropriate C.A.B.E.

Fixed variants of the lung demand valves incorporate a medium pressure hose having an end
fitting design for connecting directly to the medium pressure outlet port of the pressure
reducer of the appropriate C.A.B.E.

The housing of the lung demand valves are manufactured from a moulded composite glass
filled material that provides a high degree of strength and durability.

Positive pressure types have a manually operated re-set lever to switch ‘OFF’ the positive
pressure function - this must be pressed before opening the valve of compressed air cylinder.
First breath inhalation by the wearer will switch ‘ON’ the positive pressure.

All variants include a supplementary air supply facility that is made active by pressing the
centre of the protective rubber cover at the front of the lung demand valve. Lung demand
valves fitted to short duration C.A.B.E. and escape units will not normally include a
supplementary air supply facility.

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 11

Section 4 Operating Principle

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 12

Section Description
4.1 Positive Pressure Lung Demand Valves A and AE

4.2 Normal Demand Lung Demand Valves - N

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 13

4.1 Positive Pressure Lung Demand Valves - A and AE


Refer to Figure 1.

Before turning ‘On’ the high pressure air supply the positive pressure mechanism
must be switched ‘Off’ by pressing the re-set lever on the side of the lung demand
valve to lift and lock the main pivot lever and diaphragm.

On opening the cylinder valve, medium pressure air flows from the reducer through
a hose and swivel banjo then into main bore of balanced piston block.

Increasing air pressure moves the balanced piston axially, forcing the piston cone
against the O-Ring seal located in the block - sealing ‘Off’ the airflow from outlet of
lung demand valve.

Refer to Figure 2.

When the wearer begins to inhale this creates a negative pressure inside the facepiece
and lung demand valve. Due to the negative pressure the ambient air pressure acts
against the diaphragm. As the diaphragm deflects it then presses against the main
hinged lever of the balanced piston unit resulting in the release of the positive pressure
‘Off’ mechanism.

Increasing deflection of the diaphragm, during inhalation, pivots the balanced piston
lever forcing it against the end cap of the piston - moving the piston axially and lifting
the piston cone from its seating and releasing air from the lung demand valve into
wearer’s facepiece. Two extension springs attached to the balanced piston block
and the main lever act on the balanced piston, lifting the sealing cone and maintaining
a positive pressure (above ambient) inside the facepiece.

When the wearer stops inhaling then begins to exhale, the diaphragm, pivot lever
and sealing cone retract.

The cycle repeats as the wearer begins to inhale.

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 14

Balanced Piston
‘Closed’

Medium Pressure
(M.P.)

Fig. 1
0008

Balanced Piston
‘Open’

M.P. Flow
to Facepiece

Fig. 2
0078

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 15

4.2 Normal Demand Lung Demand Valves - N


Refer to Figure 3

On opening cylinder valve, medium pressure air flows from the reducer, through a hose
and swivel banjo then into main bore of balanced piston block.

Increasing air pressure moves the balanced piston axially, forcing the piston cone against
the O-Ring seal located in the block - sealing ‘Off’ the airflow from outlet of lung
demand valve.

Refer to Figure 4

When the wearer begins to inhale this creates a negative pressure inside the facepiece
and lung demand valve. Due to the negative pressure the ambient air pressure acts
against the diaphragm then the main hinged lever of the balanced piston unit.

Increasing deflection of the diaphragm, during inhalation, pivots the balanced piston
lever forcing it against the end cap of the piston - moving the piston axially and lifting
the piston cone from its seating and releasing air from the lung demand valve into
wearer’s facepiece.

When the wearer stops inhaling then begins to exhale, the diaphragm, pivot lever and
sealing cone retract.

The cycle repeats as the wearer begins to inhale.

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 16

Balanced Piston
‘Closed’

Medium Pressure
(M.P.)

Fig. 3
0249

Balanced Piston
‘Open’

M.P. Flow
to Facepiece

Fig. 4
0250

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 17

Section 5 Servicing Instructions

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 18

Section Description
5.1 Service and Test Intervals

5.2 Cleaning, Disinfecting and Drying

5.2.1 Important Information

5.2.2 Cleaning and Disinfecting

5.3 Replacing Diaphragm

5.4 Replacing Medium Pressure Hoses

5.4.1 Removing MP Hose (Elbow Connector)

5.4.2 Removing MP (Straight Connector)

5.4.3 Assembly of MP Hose (Elbow Connector) to Demand Valve

5.4.4 Assembly of MP Hose (Straight Connector) to Demand Valve

5.5 The Balanced Piston Unit

5.5.1 Removing Balanced Piston Unit

5.5.1.1 Replacing/Refitting Return Spring – Positive Pressure

5.5.2 Replacing Springs of Balanced Piston Units – Positive Pressure

5.5.3 Assembly of Balanced Piston Unit to Demand Valve

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 19

5.1 Service and Test Intervals


Carry out Inspection, Service and Testing of equipment according to this table.
Record all necessary data in the Log Book for the equipment. These instructions
apply also to non-used in storage equipment.

Important Note: This table is intended as a general guide. Reference must also be made
to the Inspection, Service and Test Intervals specified in the appropriate section
of the relevant C.A.B.E. Training manual.

Description After Ever y Ever y Ever y


Use Month Year 6
Years
Complete Clean and Disinfect m
Equipment
Visual Inspection m u
Functional and Leak Testing as defined m u
in Instructions for Use
Flow and Static Tests as per m
Manufacturers Instructions
Demand Valve Clean and Disinfect as necessar y mn
Pressure Medium Pressure Check m
Reducer
Replace Sintered Filter m
Replace H.P. Connector O-Ring m
Basic Overhaul m
(Repair Exchange Scheme)
Cylinder Charge to Correct Pressure m
Charged Pressure Check. Check Test mu
Date Stamped on Cylinder
Re-Cer tification - Cylinder Pressure
Test
(According to National Standards)
Overhaul as necessar y, or at time of
Cylinder Valve
Cylinder Re-Cer tification

m Dräger Recommendation - UK.


u COSHH
n Type A - Push-In demand valve – apply grease (3331247) around connector O-ring.

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 20

5.2 Cleaning, Disinfecting and Drying


5.2.1 Important Information
Dräger recommend that contaminated and dirty components or assemblies are
carefully cleaned and disinfected then thoroughly dried after use, or as and when
considered necessary by the user.

To ensure correct operational condition of the equipment use the cleaning and
disinfecting solutions recommended by Dräger. The use of any other product will
invalidate the Dräger warranty and guarantee.

When using baths to contain cleaning and disinfecting solutions the immersed
components and assemblies must only be agitated manually.

It is important to note that Dräger Do Not recommend the use of any form of
mechanical, electrical or ultrasonic agitation of this product. If the user decides
however to use ultrasonic cleaning baths then it is important that the instructions
supplied by the manufacturer of the bath are strictly followed.

Refer to Important Notes.

Important Notes:
1. The diaphragm of the lung demand valves must not be cleaned using ultrasonic
techniques, as this will have a detrimental affect on the centre-bonded area of
the diaphragm. Clean and disinfect the diaphragm manually.

2. If Ultrasonic cleaning is to be used on the lung demand valve (diaphragm must


be removed) this will eventually remove lubricant from the critical moving parts.
Duration in the bath combined with the ultrasonic energy of the machine will
affect the time for this to occur. Following cleaning, disinfecting and drying carry
out functional tests. Re-lubricate moving parts as required.

3. Immersing the lung demand valve and manual agitation will also eventually remove
lubricants although of course this will occur over a much longer period. Following
cleaning, disinfecting and drying it is recommended that functional tests be
carried out. Re-lubricate moving parts as required.

Safety Note: When using cleaning and disinfecting agents refer to the relevant
manufacturers’ usage instructions. It is important that attention be paid to
concentration and reaction times. Do Not use organic solvents such as Acetone,
Alcohol, White Spirit, Trichloroethylene or similar.

Dräger recommends:

1. Cleaning:
Safety Wash
1 Litre Dispenser 3380164
1 Litre Refill 3380165
5 Litre Dispenser 3380166
5 Litre Refill 3380167
Sekusept
4 bottles @ 2 litres 7904071

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 21

Do Not exceed a temperature of 30 degrees Celsius.

Note: Rinse off cleaning solution in clean water before disinfecting.

2. Disinfecting:
Wipex Cloths (Pack of 50) 3380375

Incidur – 6 litre container 7904072


Incidur - 30 litre container 7904073

Do Not exceed a temperature of 30 degrees Celsius.

Note: Rinse off disinfecting solution in clean water before drying.

3. Rinsing and Drying


Remove cleaning and disinfecting solutions by rinsing in clean running water, followed
by drying.
Rinsing - Do Not exceed a temperature of 30 degrees Celsius.
Drying - Do Not exceed a temperature of 60 degrees Celsius.

5.2.2 Cleaning and Disinfecting


When the user considers it necessary to internally clean and disinfect the lung demand
valve then these instructions should be followed:

Note: A Detach demand valves (fitted with male coupling) from female coupling of medium
pressure hose of the C.A.B.E.

Note: B If necessary, demand valves attached directly to the reducer should be removed at the
hose connection to the reducer. For details of removal refer to the appropriate section
of the relevant C.A.B.E. Training Manual.

Tools required:
Plate Spanner R26817
1 If required, detach the demand valve
from the equipment.

û
2 Press the reset lever of positive
pressure type demand valves. Do Not
press centre of rubber cover while
pressing the reset lever or force lever
against the stop. See Fig. 1.

Refer to Figure 1.
(PA90A Illustrated.)

ü Fig. 1
0032/0251

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 22

3 Remove the rubber cover (1) then


using plate spanner R26817
unscrew and remove the bayonet cap
(2). Press reset lever (3) - positive
pressure type only. Carefully unhook
diaphragm (4) from the lever (5) of
the balanced piston assembly.

Important Notes: The diaphragm of


the lung demand valves must not
be cleaned using ultrasonic
techniques, as this will have a
detrimental affect on the centre-
bonded area of the diaphragm. 3 5 4 2 1
Clean and disinfect the diaphragm
manually. Fig. 2
0028

4 Re-assemble the rubber cover (1) over the front of the lung demand valve to protect
balanced piston mechanism. Immerse and manually agitate the demand valve,
diaphragm and bayonet cap into cleaning fluid. (Refer to cleaning fluid instructions.)

5 After cleaning, rinse the components in clean water then immerse and manually
agitate the demand valve, diaphragm and bayonet cap into disinfecting fluid. (Refer
to disinfecting fluid instructions.) After disinfecting, remove rubber cover and rinse
components in clean water, taking care not to damage balanced piston lever
mechanism.

6 Press reset lever of the positive pressure demand valves to close balanced piston.

7 Connect the medium pressure hose of the demand valve to medium pressure air
supply. Turn ‘ON’ air supply then gently press the balanced piston lever to blow out
fluid residue.

Safety Note: Wear suitable eye protection.

8 Press reset lever of the positive pressure demand valves to close the balanced piston.
Turn ‘OFF’ and vent the air supply then disconnect medium pressure hose of demand
valve from air supply. Thoroughly ‘dry’ components.

Note: Do Not exceed a temperature of 60 degrees Celsius when drying.

9 After drying insert the balanced piston lever into hole in centre boss of the diaphragm
(chamfered side of boss towards lever) then carefully locate outer bead of diaphragm
into groove in demand valve body. See Fig. 3 for correct orientation of the diaphragm.

10 Moisten lip of the bayonet cap with clean water - refit bayonet cap to the lung and
using plate spanner R26817 tighten the cap to the demand valve.

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 23

Note: Re-check orientation of diaphragm.


See Fig. 3.

11 Re-assemble rubber cover over the


front of the demand valve then press
the reset lever of positive pressure
type demand valves.

12 Connect the demand valve to


equipment and carry out Leak Test
and Functional Test.

5.3 Replacing Diaphragm


Tools required:
Plate Spanner R26817 Fig. 3
0029
Refer to Figure 4 - (PA90A Illustrated)

1 Remove the rubber cover (1) then


using plate spanner R26817
unscrew and remove the bayonet cap
(2). Press reset lever (3) - positive
pressure types only. Carefully
unhook diaphragm (4) from lever (5)
of balanced piston assembly. Scrap
the diaphragm.

2 Insert the balanced piston lever into


hole in centre boss of the new
diaphragm (chamfered side of boss
towards lever) then carefully locate 3 5 4 2 1
outer bead of diaphragm into groove
in demand valve body. See Fig. 5 Fig. 4
for correct orientation of diaphragm.
0028

3 Moisten the lip of the bayonet cap


with clean water - refit the bayonet
cap to lung and using the plate
spanner R26817 tighten the cap to
the demand valve.
Note: Re-check orientation of diaphragm.
See Fig. 5.

4 Re-assemble the rubber cover over


the front of the demand valve then
press the reset lever of positive
pressure type demand valves.

5 Connect the demand valve to the


equipment and carry out Leak Test Fig. 5
and Functional Test.
0029

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 24

5.4 Replacing Medium Pressure Hoses


Fixed variants of the lung demand valve incorporate a medium pressure hose fitted
directly to the medium pressure outlet port of the pressure reducer. These demand
valves should also be removed from the reducer. For details of removal of hose from
the reducer refer to the appropriate section of the relevant C.A.B.E. Training Manual.

5.4.1 Removing MP Hose (Elbow Connector)


Tools Required:
10mm/11mm AF Open Ended Spanner 3336309
14mm AF Open Ended Spanner 3336308

1 Disconnect demand valve from the


female quick release coupling of the
equipment.

Refer to Figure 6.
1 2
2 Using 14mm spanner hold the banjo
(1) across flats on banjo flange and
using the 11mm spanner unscrew
the elbow (2) of the hose from the
banjo. Inspect the hose and if
defective, scrap and replace.

3 Inspect the threaded bore of the


banjo. If necessary clean and remove
all traces of adhesive from the
thread.

Fig. 6
0040

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 25

5.4.2 Removing MP Hose (Straight Connector)


Tools Required:
Fine Point Circlip Pliers 3333968
Screwdriver 3336222
Hexagon Adaptor – 2mm (Locking End Cap) 3310600
Torque Key (Locking End Cap) 3310593

Refer to Figure 7.

1 Disconnect the demand valve from


the equipment and remove the end
cap (1). 4
Plastic End Cap - Insert screwdriver
(3336222) into radial slot of end cap 3
and remove end cap. Scrap the end
cap, replace with new. 2

Locking End Cap – Using 2mm


adaptor and torque key unscrew the
locking cap screw enough to enable
1
the end cap to be removed. Fig. 7
1239
2 Using fine point circlip pliers, remove
circlip (2) from groove in the banjo.
Remove washer (3). Scrap the
circlip - Do Not re-use.

Note: On positive pressure demand valves


Do Not detach reset lever (4). If the
lever becomes detached, reassemble
the lever to the demand valve as
detailed in Section 5.5.1.1.

3 Push end of the banjo and extract


banjo and hose assembly (1) Fig. 8
from demand valve body. Check O- 2 1
Ring (2) is on the banjo, if not,
remove the O-Ring from bore of the Fig. 8
demand valve housing.
1238

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 26

5.4.3 Assembly of MP Hose (Elbow Connector) to Demand Valve


Tools Required:
10mm/11mm AF Open Ended Spanner 3336309
14mm AF Open Ended Spanner 3336308
Loctite 638 3335389

1 Inspect the threaded bore of the banjo and if necessary clean and remove all traces
of adhesive from the thread. Check that the O ring seal of the new hose is not
damaged and is fitted correctly to the elbow. If necessary replace the O ring.

2 Screw the elbow of the new hose into the banjo by one revolution - apply one drop
of Loctite 638 to thread of the elbow then finally screw the elbow into the banjo.
Using 14mm spanner, hold the banjo across flats of the banjo flange and using the
11mm spanner tighten elbow to the banjo.

Note: Do Not pressurise pneumatic system for 30 to 40 minutes to allow Loctite to set. Fixed
variants of the lung demand valve should be fitted directly to the medium pressure
outlet port of the pressure reducer. For details of fitting of hose to the reducer refer to
the appropriate section of the relevant C.A.B.E. Training Manual.

3 Carry out Leak Test and Functional Test after 30 to 40 minutes.

5.4.4 Assembly of MP Hose (Straight Connector) to Demand Valve


Tools Required:
Molykote 111 grease 3331247
Circlip Assembly Tool 3335972
Hexagon Adaptor – 2mm (Locking End Cap) 3310600
Torque Key (Locking End Cap) 3310121

1 Inspect and if necessary assemble


new O-Rings into the grooves of the
banjo (1) Fig 9. Before assembly of
the banjo to the demand valve apply
a coating of Molykote 111 grease
around the outside diameter of O-
Rings.

2 Carefully insert new banjo and hose


assembly into bore of demand valve
body, aligning the banjo through
bore of balanced piston block and 2 1
taking care not to detach the reset
lever (2) of positive pressure
demand valves.
Fig. 9
1238

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 27

Note: If reset lever becomes detached,


reassemble lever to demand valve as 1
detailed in Section 5.5.1.1.

4 Assemble washer (1) Fig. 10. Using


the assembly tool 3335972
assemble the circlip (2) into groove
of banjo to secure banjo to demand
valve body. Following assembly of
the circlip check that the banjo
swivels freely.

5 Fit new End Cap into bore of the


banjo.
2
Plastic End Cap - Insert new plastic Fig. 10
end cap into bore of banjo. 1120

Locking End Cap – Insert locking


end cap into bore of the banjo and
using 2mm adaptor and torque key
tighten end cap to banjo – torque to
0.7Nm.

Note: Fixed variants of the lung demand


valve should be fitted directly to the
medium pressure outlet port of the
pressure reducer. For details of fitting
of hose to the reducer refer to the
appropriate section of the relevant
C.A.B.E. Training Manual.

6 Carry out Leak Test and Functional


Test.

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 28

5.5 The Balanced Piston Unit


PA90 Series lung demand valves incorporate a pre-set and sealed balanced piston
unit. The Dräger guarantee is void should original seal, in sealing cone, be tampered
with, or broken. Guarantee of correct operational condition of the balanced piston
unit is only valid if serviced, and re-sealed, by Dräger.

Replace with new factory set, or Repair Exchange balanced piston unit should
problems occur (other than a requirement to replace springs on positive pressure
type units).

Balanced Piston Units will only be accepted for Repair Exchange after disassembly
from demand valve, i.e. Balanced Piston Assembly only.

For details of the Terms and Conditions of the Repair Exchange Scheme contact
Dräger Service - Blyth.

Replace complete balanced piston unit, or springs, as follows:

5.5.1 Removing Balanced Piston Unit from Demand Valve


Tools required:
Plate Spanner R26817
Fine Point Circlip Pliers 3333968
Posi-Drive Screwdriver 3331245
Screwdriver 3336222
Hexagon Adaptor – 2mm (Locking End Cap) 3310600
Torque Key (Locking End Cap) 3310593
Molykote 111 Grease 3331247
Loctite 245 3335722
Connector Tool (PA90AE) 3310497
Connector Tool (PA90N) 3310496
Circlip Assembly Tool 3335972

1 Remove the rubber cover (1) then


using plate spanner R26817
unscrew and remove the bayonet cap
(2). Press reset lever (3) - positive
pressure types only. Carefully
unhook diaphragm (4) from lever (5)
of balanced piston assembly.

3 5 4 2 1
Fig. 11
0028

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 29

Refer to Figure 12.

2 Using posi-drive screwdriver


unscrew retaining screw (1) until it 2 3
is withdrawn from space in the
castellation of connector (2). 1 4
Note: Do Not attempt to remove push-in or
screw-in connectors with the retaining
screw in the castellation of the 2
connector.
3
3 Remove connector. 1 4

Fig. 12
0041/0127

Push-in Type - A
Unscrew push-in type connector (anti-clockwise) from demand valve body and remove
anti-injector pipe (3).

Screw-in Type - AE
Use connector tool (3310497) to unscrew screw type connector and handwheel
assembly from demand valve body. Remove anti-injector pipe.

Screw-in Type - N
Screw and tighten M40 handwheel to connector tool 3310496. Turn handwheel and
connector tool anti-clockwise to unscrew screw type connector and handwheel
assembly from demand valve body. Remove anti-injector pipe.

4 Remove end cap (1) Fig. 13.

Plastic End Cap - Insert screwdriver


(3336222) into radial slot of end cap
and remove end cap. Scrap the end 4
cap, replace with new.

Locking End Cap – Using 2mm 3


adaptor and torque key unscrew the 2
locking cap screw enough to enable
the end cap to be removed.
1
5 Using fine point circlip pliers, remove
circlip (2), from groove in banjo.
Remove washer (3). Discard the Fig. 13
circlip - Do Not re-use.
0042

Note: On positive pressure demand valves Do Not detach reset lever (4). If lever becomes
detached then reassemble lever to the demand valve as detailed in Section 5.5.1.1.

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 30

6 Push end of the banjo and extract


banjo and hose assembly (1) from
demand valve body. See Fig. 14.
Check O-Ring (2) is on the banjo, if
not, remove the O-Ring from bore
of the demand valve housing.

7 Remove balanced piston unit from


inside demand valve body and if
necessary return balanced piston
unit to Dräger Ser vice for
replacement under the Repair 2 1
Exchange Scheme.

8 Clean old grease from the banjo,


inspect and if necessary replace the Fig.14
O-Rings. 0043

5.5.1.1 Replacing/Refitting Return Spring Positive Pressure Demand Valves


Note: Using a finger inside the demand valve body press the end of the positive pressure arm (1)
Fig. 15 to prevent any axial movement.

1 Rotate the positive pressure arm (1)


until the radial extension arm points
towards outlet port of the demand 3 2 1
valve body.

2 Insert return spring (2) into the


housing recess with the radial arm
of the spring located into the radial
slot in demand valve body.

3 Hook the axial arm of the return


spring (2) into the radial slot in the
lever (3) - not onto hexagon - then
carefully rotate the lever
anticlockwise, against spring
Fig. 15
tension, until the first hexagon flat
aligns with hexagon on the positive 0044

pressure arm. Press the lever onto


the positive pressure arm.

4 Rotate the lever anticlockwise, carefully lifting the stop of the lever to the opposite
side of the rib on the body of demand valve. At this position press the lever onto the
positive pressure arm.

5 Check assembly and carefully press lever (3) against spring tension - lever should
return freely under spring load.

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 31

5.5.2 Replacing Springs of Balanced Piston Unit - Positive Pressure


Tools required:
Fine Point Tweezers 3333326

1 Remove Balanced Piston Unit from


Demand Valve. Repeat instructions
1 to 9 in Section 5.5.1.
2
Refer to Figure 16.

2 Unhook closed loop of each spring


(1) from locating lugs of balanced 3
piston lever (2), then unhook open
loop of each spring from pin of 5 1(x2)
mounting block (3). Scrap the 4
springs. Do Not re-use springs.

3 Using fine point tweezers assemble


new springs to balanced piston Fig. 16
block assembly as follows:
0047

Hook open loop, of each spring


around the spigots of the mounting
block (3). Using fine point tweezers
hold the closed loop of each spring
then extend each spring and hook
closed loop onto locating lugs of
balanced piston lever (2). Do Not
over stretch the springs.

Note: Position the spring on the plastic form


side (4), under the pivot pin (5) as
shown in Fig. 16.

4 To re-assemble the Balanced Piston


Unit to Demand Valve, repeat
operation instructions 1 to 16 in
Section 5.5.3.

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 32

5.5.3 Assembly of Balanced Piston


Unit to Demand Valve
1 Inspect and if necessary assemble
new O-Rings into the grooves of the
banjo. Before assembly of the banjo
to the demand valve apply a coating
of Molykote 111 grease around the
outside diameter of the O-Rings.

Safety Note: Carefully read instructions on


Molykote 111 container regarding
the handling of silicone grease.

2 Insert new balanced piston unit into


Fig. 17
demand valve body and locate onto
0045
internal location boss of demand
valve body. See Fig. 17. Hold
balanced piston in position by
pressing with thumb against the end
of seal cone.

3 Carefully insert banjo and hose


assembly (1) Fig. 18 into bore of
demand valve body, aligning the
banjo through bore of balanced
piston block and taking care not to
detach the reset lever (2) of positive
pressure demand valves. 1

Note: If reset lever (2) becomes detached,


reassemble to the demand valve as
detailed in Section 5.5.1.1. 2 Fig. 18
0043

4 Assemble washer (1) Fig. 19. Using


the assembly tool 3335972 1
assemble the circlip (2) into groove
of banjo to secure banjo to demand
valve body. Following assembly of
the circlip check that the banjo
swivels freely.

5 Position and hold demand valve with


outlet port in vertical position. Insert
the anti-injector pipe (1) into the port
and locate onto the balanced piston
block. See Fig. 20.

2
Fig. 19
0046

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 33

6 Assemble connector.

Push-in Type - A
Assemble new O-Ring (2) Fig. 20
to connector. Apply Loctite 245 to
connector thread and screw
connector (3) into demand valve
body. Tighten until anti-injector pipe
is secure. Do Not overtighten.

Screw-in Type - AE
Check handwheel freely rotates on
connector handwheel assembly and
that retaining circlip is secure.
Assemble new O-Ring to connector
handwheel assembly. Apply Loctite Fig. 20
245 to connector thread. Using tool
0041/0127
(3310497), screw connector
handwheel assembly into demand
valve body until anti-injector pipe is
secure. Do Not overtighten.

Screw-in Type - N
Check handwheel freely rotates on connector handwheel assembly and that retaining
circlip is secure. Assemble new O-Ring to connector handwheel assembly. Apply
Loctite 245 to connector thread. Screw and tighten handwheel into tool 3310496
and screw connector handwheel assembly into demand valve body until anti-injector
pipe is secure. Do Not overtighten.

8 Carefully turn connector anti-clockwise till first space, in castellations on end face of
connector, is in line with retaining screw hole. Fully tighten the retaining screw to
secure the connector to the demand valve body making sure the screw enters space
in the castellation.

Note: Remove assembly tools from handwheel connectors of AE and N type demand valves.

9 Check again that the banjo swivels freely.

10 Positive Pressure Demand Valves


Press the reset lever, checking that spring loaded lever of the balanced piston lifts
and, on releasing the reset the lever the spring loaded lever holds in position. Gently
press the end of balanced piston lever checking it releases from locking plate and
retracts freely under spring load. Repeat check. Press reset lever to lift and hold the
balanced piston lever.

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 34

11 Insert the balanced piston lever into


hole in centre boss of the new
diaphragm (chamfered side of boss
towards lever) then carefully locate
outer bead of diaphragm into groove
in demand valve body. See Fig. 21
for correct orientation of diaphragm.

12 Moisten the lip of the bayonet cap


with clean water - refit the bayonet
cap to lung demand valve and using
plate spanner R26817 tighten the
cap.

Note: Re-check orientation of diaphragm.


See Fig. 21. Fig. 21
0029

13 Positive Pressure Demand Valves


Press the reset lever, checking that diaphragm lifts and then on releasing the reset
lever that the diaphragm holds in position. Do Not press the centre of the diaphragm
when pressing reset lever, or force lever against stop. Press the centre of diaphragm
gently, checking it releases and retracts under spring load. Repeat check.

15 Fit new End Cap to banjo.

Plastic End Cap - Insert new plastic end cap into bore of banjo.

Locking End Cap – Insert locking end into bore of banjo and using 2mm adaptor
and torque key tighten end cap to banjo – torque to 0.7Nm.

16 Assemble rubber cover to the demand valve then carry out Leak Test and Functional
Test.

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 35

Section 6 Test Procedures

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 36

6.1 Test Procedures


The PA90 Series of Lung Demand Valve are approved for use with approved Dräger
Compressed Air Breathing Equipment (C.A.B.E.).

Standard test procedures for C.A.B.E. are as follows and are covered in the relevant
Training Manual for the C.A.B.E.

l Cylinder Pressure Check and Valve Leak Test


l High Pressure Leak Test
l Whistle Warning Unit Test
l Full Flow Test
l Medium Pressure Test - Pressure Reducer

For detailed instructions of the following Functional Tests relating to the testing of
the lung demand valve and associated ancillary items refer to the appropriate Test
Equipment Manual.

Type A and AE
l Balanced Piston Leak Test
l Positive Pressure Balance Test
l Static Pressure Test
l Positive Pressure Switch Over

Type N
l Balanced Piston Leak Test
l Demand Valve Opening Pressure

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 37

Section 7 Fault Location

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 38

7.1 Demand Valve


The high performance PA90 Series lung demand valves incorporate a pre-set and
sealed balanced piston unit that has unlimited service life.

The Dräger Warranty and Guarantee is void should any attempt be made by customers
to make adjustment to the balanced piston unit. Guarantee of correct operational
condition of the balanced piston unit is only valid if serviced, and re-sealed, by
Dräger.

Replace with new factory set or Repair Exchange balanced piston unit if problems
occur (other than a requirement to replace springs on positive pressure type units).
Balanced Piston Units will only be accepted for Repair Exchange after disassembly
from demand valve, i.e. Balanced Piston Assembly only.

For details of Terms and Conditions of Repair Exchange Scheme contact Dräger
Service - Blyth.

The following Fault – Cause – Remedy chart outlines a list of possible faults that may
occur.

FAULT CAUSE REMEDY

Leak from pressure relief 1 Damaged O-ring Replace O-ring


valve of medium pressure 2 Weak Spring Replace spring
hose 3 Damaged O-ring Replace retainer
retainer

Leak from pressure Reducer failure Remove hoses from


relief valve of medium reducer and return reducer
pressure hose. Relief to Dräger. (REX)
valve components O.K.

High Medium Pressure Reducer out of Remove hoses from


specification reducer and return reducer
to Drâger. REX)

Low Medium Pressure Reducer out of Remove hoses from


specification reducer and return reducer
to Drâger. (REX)

Medium pressure High High Pressure Leak Remove hoses from


after one minute reducer and return reducer
Continuous creep to Drâger. (REX)

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 39

Section 8 Spare Parts Lists and Tool List

tm 12842 - January 2004


PA90 SERIES - LUNG DEMAND VALVE 40

Spare Parts Lists

Number Title

1284.2 PA90 Series Lung Demand Valve

1285.16 Tool List

tm 12842 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 1

Index

Section 1 Instructions for Use

Section 2 Introduction and Safety Warnings

Section 3 Product Description

Section 4 Operating Principle

Section 5 Servicing Instructions

Section 6 Test Procedures

Section 7 Fault Location

Section 8 Spare Parts Lists and Tool List

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 2

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 3

Section 1 Instructions for Use

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 4

Positive Pressure Lung Demand Valve - Type A Series 3338727

Positive Pressure Lung Demand Valve - Type AE Series 3338726

Normal Demand Lung Demand Valve - Type N Series 3338725

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 5

Section 2 Introduction and Safety Warnings

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 6

Introduction
The Plus and PSS Series range of Lung Demand Valves are approved for use with approved
Dräger Compressed Air Breathing Equipment (C.A.B.E.) and Facepieces.

The combination of Lung Demand Valve, C.A.B.E. and Facepiece provides the wearer of the
equipment with respiratory protection when working in or escaping from a contaminated or
oxygen deficient life-threatening environment.

For some markets the PSS prefix denotes ‘Short Hose LDV’.

Safety Warnings
l Explain - Instructions for Use supplied with the equipment must be adhered to.

l Maintain - the respiratory protection equipment must be maintained as outlined in


Dräger Limited service schedules and maintenance instructions.

Contact Dräger Limited, Blyth for details of Service Contracts and Service Training
Courses.

l Train - potential wearers must be trained in how to wear, and use the equipment and to
understand its limitations.

Contact Dräger Limited, Blyth for details of Wearer Training Courses.

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 7

Positive Pressure - Type A Illustrated

1100

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 8

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 9

Section 3 Product Description

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 10

Product Description
The variants of the Plus and PSS Series Lung Demand Valve are as follows:

l Type A - Positive Pressure with Push-In connector to facepiece.

l Type AE - Positive Pressure with M45 Screw-In connector to facepiece.

l Type N - Normal Demand with M40 (Rd40) Screw-In connector to facepiece.

The high performance Lung Demand Valves incorporate a precision factory set Air Transfer
Block Assembly containing a pre-set and sealed balanced piston unit. These items have
unlimited service life.

The Dräger Warranty and Guarantee is void should any attempt be made by customers to
make adjustment to the balanced piston unit or to the associated Injector Pipe.

Re-adjustment of the main lever setting height will be required following the fitting of a
replacement balanced piston unit to an existing Air Transfer Block, or the fitting of a replacement
Air Transfer Block to an existing balanced piston unit.

The threaded handwheel connector of the screw-in variants (types AE and N) incorporate a
swivel feature (patent pending) and when fitted to the facepiece this feature allows the demand
valve a degree of radial movement during head and body movement of the wearer. The swivel
feature is not however suitable for constant rotation.

Detachable variants of the lung demand valves have a medium pressure hose with a quick
release male coupling to connect the demand valve to the female coupling of the medium
pressure hose from the pressure reducer of the appropriate C.A.B.E. (PSS LDV Series are
short hose configuration.)

Fixed variants of the lung demand valves incorporate a medium pressure hose having an end
fitting design for connecting directly to the medium pressure outlet port of the pressure
reducer of the appropriate C.A.B.E.

The housing of the lung demand valves are manufactured from a moulded composite glass
filled material that provides a high degree of strength and durability.

Positive pressure types have a manually operated re-set button to switch ‘OFF’ the positive
pressure function - this must be pressed before opening the valve of compressed air cylinder.
First breath inhalation by the wearer will switch ‘ON’ the positive pressure.

All variants include a supplementary air supply facility that is made active by pressing the
centre of the protective rubber cover at the front of the lung demand valve. Lung demand
valves fitted to short duration C.A.B.E. and escape units will not normally include a
supplementary air supply facility.

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 11

Section 4 Operating Principle

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 12

Operation
Positive Pressure Lung Demand Valves - Types A and AE
Refer to Figure 1.

Before turning ‘On’ the high pressure air supply the positive pressure mechanism must be
switched ‘Off’. This is achieved by pressing the re-set button at the top of the lung demand
valve to lift and lock the main pivot lever and diaphragm. The conical compression spring, in
the main chamber of the balanced the piston unit, acts against the piston flange resulting in
the piston cone being pressed against the sealing rim of the piston housing - sealing ‘Off’ the
outlet of the balanced the piston unit.

On opening the cylinder valve, medium pressure air flows from the reducer, through the
medium pressure hose of the lung demand valve, into the air transfer block then into the main
chamber of balanced the piston unit.

Refer to Figure 2.

When the wearer begins to inhale a negative pressure is created inside the facepiece and
lung demand valve. Due to the negative pressure the ambient air pressure acts against the
diaphragm. As the diaphragm deflects it then presses against the main hinged lever of the
balanced piston unit resulting in the release of the positive pressure ‘Off’ mechanism.

Increasing deflection of the diaphragm, during inhalation, pivots the cam of the main lever
against the secondary lever. The secondary lever then pivots and presses against the end cap
of the piston - moving the piston axially against the conical spring - lifting the piston cone from
its seating and releasing air from the lung demand valve into wearer’s facepiece. A compression
spring attached to the front cover of the lung demand valve acts on the diaphragm, then the
pivot levers and the conical compression spring, lifting the sealing cone and maintaining a
positive pressure (above ambient) inside the facepiece.

When the wearer stops inhaling then begins to exhale, the diaphragm, pivot levers and sealing
cone retract.

This cycle repeats as the wearer begins to inhale.

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 13

Balanced Piston
‘Closed’

Medium Pressure
(M.P.)

Fig. 1
0583

Balanced Piston
‘Open’

M.P. Flow
to Facepiece

Fig. 2
0587

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 14

Normal Demand Lung Demand Valves - Type N


Refer to Figure 3

The conical compression spring, in the main chamber of the balanced the piston unit, acts
against the piston flange resulting in the piston cone being pressed against the sealing rim of
the piston housing - sealing ‘Off’ the outlet of the balanced the piston unit.

On opening the cylinder valve, medium pressure air flows from the reducer, through the
medium pressure hose of the lung demand valve, into the air transfer block then into the main
chamber of balanced the piston unit.

Refer to Figure 4

When the wearer begins to inhale a negative pressure is created inside the facepiece and
lung demand valve. Due to the negative pressure the ambient air pressure acts against the
diaphragm, then the main hinged lever of the balanced piston unit.

Increasing deflection of the diaphragm, during inhalation, pivots the cam of the main lever
against the secondary lever. The secondary lever then pivots and presses against the end cap
of the piston - moving the piston axially against the conical spring - lifting the piston cone from
its seating and releasing air from the lung demand valve into wearer’s facepiece.

When the wearer stops inhaling then begins to exhale, the diaphragm, pivot levers and sealing
cone retract.

This cycle repeats as the wearer begins to inhale.

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 15

Balanced Piston
‘Closed’

Medium Pressure
(M.P.)

Fig. 3
1169

Balanced Piston
‘Open’

M.P. Flow
to Facepiece

Fig. 4
1170

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 16

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 17

Section 5 Servicing Instructions

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 18

Section Description
5.1 Contents

5.1 Service and Test Intervals

5.2 Cleaning, Disinfecting and Drying

5.2.1 Important Information

5.2.2 Cleaning and Disinfecting

5.3 Replacing Diaphragm

5.4 Replacing Positive Pressure Spring

5.5 Replacing Medium Pressure Hose

5.5.1 Removing Medium Pressure Hose

5.5.2 Assembly of Medium Pressure Hose

5.6 Replacing Injector Pipe

5.7 Replacing Air Transfer Block Assembly

5.7.1 Removing Air Transfer Block Assembly

5.7.2 Fitting Air Transfer Block Assembly – Lever Height Setting

5.8 Lubrication of Connector Assemblies - M40 and M45

5.8.1 Removing Connector Assemblies from LDV - M40 and M45

5.8.2 Disassembly of Connectors - M40 and M45

5.8.3 Assembly of Connector - M40

5.8.4 Assembly of Connector - M45

5.8.5 Assembly of Connector Assemblies to LDV - M40 and M45

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 19

5.1 Service and Test Intervals


Carry out Inspection, Service and Testing of equipment according to this table.
Record all necessary data in the logbook for the equipment. These instructions
apply also to non-used in storage equipment.

Important Note: This table is intended as a general guide. Reference must also be made
to the Inspection, Service and Test Intervals specified in the appropriate section
of the relevant C.A.B.E. Training manual.

Description After Ever y Ever y Ever y


Use Month Year 6
Years
Complete Clean and Disinfect m
Equipment
Visual Inspection m u

Functional and Leak Testing as defined m u


in Instructions for Use
Flow and Static Tests as per m
Manufacturers Instructions
Demand Valve Clean and Disinfect as necessar y mn
Pressure Medium Pressure Check m
Reducer
Replace Sintered Filter m
Replace H.P. Connector O-Ring m
Basic Overhaul m
(Repair Exchange Scheme)
Cylinder Charge to Correct Pressure m
Charged Pressure Check. mu
Check Test Date Stamped on Cylinder
Re-Cer tification - Cylinder Pressure
Test
(According to National Standards)
Overhaul as necessar y, or at time of
Cylinder Valve
Cylinder Re-Cer tification

¦ Dräger Recommendation - UK.


u COSHH
n Type A - Push-In demand valve - apply grease (3331241) around the connector
O-ring.

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 20

5.2 Cleaning, Disinfecting and Drying


5.2.1 Important Information
Dräger recommend that contaminated and dirty components or assemblies are
carefully cleaned and disinfected then thoroughly dried after use, or as and when
considered necessary by the user.

To ensure correct operational condition of the equipment use the cleaning and
disinfecting solutions recommended by Dräger. The use of any other product will
invalidate the Dräger warranty and guarantee.

When using baths to contain cleaning and disinfecting solutions the immersed
components and assemblies must only be agitated manually. Any form of mechanical
agitation, e.g. ultrasonic, washing machines, dishwashers e.t.c. must be approved
and covered by a written agreement with Dräger.

Note: Cleaning solutions will eventually remove lubricants from moving parts. Following any
cleaning, disinfecting and drying carry out functional tests. Re-lubricate moving parts
as required.

Safety Note: When using cleaning and disinfecting agents refer to the relevant
manufacturers’ usage instructions. It is important that attention be paid to
concentration and reaction times. Do Not use organic solvents such as Acetone,
Alcohol, White Spirit, Trichloroethylene or similar.

Dräger recommends the following:

1. Cleaning:
Safety Wash
1 Litre Dispenser 3380164
1 Litre Refill 3380165
5 Litre Dispenser 3380166
5 Litre Refill 3380167
Sekusept
4 bottels @ 2 litres 7904071

These products must be used with cold water however Never exceed a temperature
of 30 degrees Celsius.

Note: Rinse off cleaning solution in clean water before disinfecting.

2. Disinfecting:
Wipex Cloths (for manual disinfecting)
Pack of 50 3380375
Incidur (6 litre container) 7904072
Incidur (30 litre container) 7904073

This product must be used with cold water however Never exceed a temperature of
30 degrees Celsius.

Note: Rinse off disinfecting solution in clean water before drying.

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 21

3. Rinsing and Drying


Remove cleaning and disinfecting solutions by rinsing in clean running water, followed
by drying.

Do Not exceed a temperature of 60 degrees Celsius when drying components.


Immediately remove from the drying cabinet when dry. Never exceed a period of
30 minutes.

5.2.2 Cleaning and Disinfecting


External cleaning of the lung demand valve may be carried out as required using a
cloth dampened in the above cleaning solution.

Important Note: When cleaning using a dampened cloth, take care to prevent any fluid
residue entering the demand valve.

When however the user considers it necessary to internally disinfect the lung demand
valve then these instructions should be followed:

Note: 1 Detach demand valves (fitted with male coupling) from the female coupling of the
medium pressure hose of C.A.B.E.

Note: 2 If necessary, demand valves attached directly to the reducer should be removed at the
hose connection to the reducer. For details of removal refer to the appropriate section
of the relevant C.A.B.E. Training Manual.

Tools Required
Plate Spanner R26817
Silicone Spray - SILKOSPRAY® (500ml) 1563343

Refer to Figure 1 – (Type A Illustrated.)


1 6 2 4 3 6
1 If required, detach demand valve
from equipment.

2 Fold rubber cover (1) from front of


lung demand valve. Using plate
spanner inserted into radial slots of
bayonet cap, turn bayonet cap (2)
counter-clockwise and remove cap
from body of demand valve.

Note: Types A and AE - the positive pressure


spring remains attached to the bayonet 5
cap. Do Not remove spring. Take care Fig. 1
not to stretch or distort the spring. 0584

3 Using the thumb and forefinger carefully grip the centre plate of the diaphragm (3),
tilt centre plate and lift the diaphragm from the body of the lung demand valve.
Remove slip ring (4).

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 22

4 Fold rubber cover back over the front of the body of the demand valve to cover and
protect the main pivot lever (5). Immerse and manually agitate demand valve,
diaphragm, slip ring and bayonet cap in cleaning fluid. (Refer to cleaning fluid
instructions.)

Note: Types A and AE - Take care not to stretch or distort the positive pressure spring.

5 After cleaning, rinse components in clean water then immerse and manually agitate
demand valve, diaphragm and bayonet cap in disinfecting fluid. (Refer to disinfecting
fluid instructions.) After disinfecting, fold rubber cover away from the front of lung
demand valve and rinse all components in clean water, taking care not to damage
main pivot lever.

6 Press re-set button (6) of type A and AE demand valves to lift the main lever and to
close the balanced piston.

7 Connect the medium pressure hose of the demand valve to a medium pressure air
supply. Turn ‘ON’ air supply then gently press main pivot lever to blow out fluid
residue.

Safety Note: Wear suitable eye protection.

8 Press the re-set button of type A and AE demand valves to lift the main lever and to
close the balanced piston. Turn ‘OFF’ and vent air supply then disconnect medium
pressure hose of the demand valve from air supply. Thoroughly ‘dry’ components.

Note: Do Not exceed a temperature of 60 degrees Celsius when drying.

Refer to Figure 2.

9 Lever Lubrication - Hold the


demand valve as shown in Fig. 2,
with the main lever (1) positioned
to achieve maximum access to the
contact points of the main lever and
the secondary lever (2). Direct and

è
locate the nozzle outlet of the
silicone spray canister inside the
demand valve body and apply two
short spray blasts of lubricant in the
direction indicated by the arrow - 1
Fig. 2.
2 Fig. 2
1386

Safety Note: Refer to the Caution and Warning information on the SILKOSPRAY® canister.

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 23

10 Carefully locate outer bead of diaphragm into groove in the body of the demand
valve. Locate tapered side of slip ring into recess in bead of diaphragm.

Types A and AE - Locate positive pressure spring of bayonet cap into the recess in
centre plate of diaphragm.

Using plate spanner, inserted into radial slots of the bayonet cap, lock bayonet cap
to body then fold rubber cover back over the front of the demand valve.

11 Press re-set button of type A and AE demand valves. Connect demand valve to
equipment and carry out Leak Test and Functional Test.

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 24

5.3 Replacing Diaphragm


Tools Required
Plate Spanner R26817

Refer to Figure 3 - Type A Demand


Valve Illustrated. 1 6 2 4 3 6
1 Fold rubber cover (1) from front of
lung demand valve. Using plate
spanner, inserted into radial slots of
bayonet cap, turn bayonet cap (2)
counter-clockwise and remove from
body of demand valve.

Note: Types A and AE - the positive pressure


spring remains attached to the
bayonet cap. Do Not remove spring. 5
Take care not to stretch or distort the
spring. Fig. 3
0584

2 Using the thumb and forefinger carefully grip the centre plate of the diaphragm (3),
tilt centre plate and lift the diaphragm from the body of the lung demand valve.
Remove slip ring (4). Discard defective diaphragm.

3 Carefully locate outer bead of new diaphragm into groove in the body of the demand
valve. Locate tapered side of slip ring into recess in bead of diaphragm.

Types A and AE - Locate positive pressure spring of bayonet cap into the recess in
centre plate of diaphragm.

Using plate spanner, inserted into radial slots of the bayonet cap, lock bayonet cap
to body then fold rubber cover back over the front of the demand valve.

4 Press re-set button of type A and AE demand valves. Connect demand valve to
equipment and carry out Leak Test and Functional Test.

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 25

5.4 Replacing Positive Pressure Spring - Types A and AE


Tools Required
Plate Spanner R26817

Refer to Figure 4 - Type A Demand


Valve Illustrated.

1 Fold rubber cover (1) from front of 1 2 3


lung demand valve. Using plate
spanner, inserted into radial slots of
bayonet cap, turn bayonet cap (2)
counter-clockwise and remove from
body of demand valve.

Note: The positive pressure spring (3)


remains attached to the bayonet cap.

Fig. 4
1133

2 When looking inside the bayonet cap and with the axial retaining leg of the spring at
the 11 o’clock position, unhook the first coil of the spring from the retaining lug at
the 2 o’clock position, then from the retaining lug at the 6 o’clock position. Turning
the spring slightly anti-clockwise removes the coil from the final retaining lug (10
o’clock) allowing the axial retaining leg of the spring to be removed from the hole in
the bayonet cap. Discard the defective spring.

3 Fit new spring as follows. When looking inside the bayonet cap and with the retention
hole for the axial retaining leg of the spring at the 11 o’clock position, insert the axial
retaining leg of the spring into the retention hole in the bayonet cap. Hook the first
coil of the spring under the retaining lug at the 10 o’clock position - then under the
retaining lug at the 6 o’clock position - then under the retaining lug at the 2 o’clock
position.

Note: Ensure only one coil diameter of the spring is located under each of the three retaining
lugs.

4 Check that the diaphragm and slip ring are still correctly located in the body of the
lung demand valve then locate positive pressure spring of bayonet cap into the
recess in centre plate of diaphragm. Using plate spanner, inserted into radial slots of
the bayonet cap, lock bayonet cap to body then fold rubber cover back over the front
of the demand valve.

5 Press re-set button and then connect demand valve to equipment and carry out Leak
Test and Functional Test.

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 26

5.5 Replacing Medium Pressure Hose


5.5.1 Removing Medium Pressure Hose
Tools required
Screwdriver 3336222
Plate Spanner R26817
Retractor Tool 3310663

1 Disconnect demand valve from female quick release coupling of equipment.

Note: Fixed variants of the lung demand valves incorporate a medium pressure hose fitted
directly to the medium pressure outlet port of the pressure reducer. These demand
valves should be removed from the reducer. For details of removal of hose from the
reducer refer to the appropriate section of the relevant C.A.B.E. Training Manual.

Refer to Figure 5.
(Type A illustrated)
1 2 4 3
2 Fold the rubber cover (1) from front
of lung demand valve. Using plate
spanner inserted into radial slots of
bayonet cap, turn bayonet cap (2)
counter-clockwise and remove from
body of demand valve.

Note: The positive pressure spring remains


attached to the bayonet cap. Do Not
remove spring. Take care not to
stretch or distort the spring.

3 Using the thumb and forefinger Fig. 5


carefully grip the centre plate of the 0584

diaphragm (3), tilt centre plate and


lift the diaphragm from the body of
the lung demand valve. Remove slip
ring (4). 1 3

Refer to Figure 6.

Important Note: Spacer (1) must be


removed before attempting to
remove Clip (2).

2
Fig. 6
1134

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 27

4 Insert the hook end of the retractor


tool into the hole in the side lugs of
the spacer (1) - pull and remove 1 3
spacer. While keeping the hose fully
inserted into the air transfer block
(3), slide the clip (2) towards the air
transfer block to release clip from
hole in body. Insert the hook end of
the retractor tool under the extended
pin of the clip (2) then pull to remove
the clip from end fitting of the hose.
Grip and pull the hose to remove the
tailpiece from the body of the
demand valve. Discard defective
hose. 2
Fig. 6
Note: Discard spacer (1) and clip (2) –
1134
Do Not re-use.

5.5.2 Assembly of Medium


Pressure Hose 1 2

1 Check that O ring seals (1) and (2)


Fig. 7 of tailpiece are fitted correctly
and not damaged. If necessary
replace O-rings. Apply thin coating
of Molykote 111 grease around
outside diameter of the O-rings.

Safety Note: Read instructions on Molykote


111 container regarding the safe
handling of silicone grease.
Fig. 7
Refer to Figure 8.
1135

2 Fully insert tailpiece through hole in 3


body of lung demand valve and into
hole in air transfer block (3) then
push and assemble a new clip (2)
over the end fitting of hose - retaining 2
the hose in position. Whilst keeping
the end fitting of the hose fully
inserted - slide the clip (2) away from
the air transfer block - insert a new
spacer (1) ensuring that the lugs of
the spacer locate around the
extended pin of the clip (2). See Fig.
8. Pull on the hose to check that it is
secure. 1
Fig. 8
1136

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 28

3 Carefully locate outer bead of diaphragm into the body of the demand valve. Locate
tapered side of slip ring into recess in bead of diaphragm then locate positive pressure
spring of bayonet cap into the recess in centre plate of diaphragm.

Using plate spanner, inserted into radial slots of the bayonet cap, lock bayonet cap
to body then re-fit and fold rubber cover over the front of the demand valve. Press re-
set button to switch ‘Off’ positive pressure.

Note: Fixed variants of the lung demand valves should be fitted directly to the medium pressure
outlet port of the pressure reducer. For details of fitting of hose to the reducer refer to
the appropriate section of the relevant C.A.B.E. Training Manual.

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 29

5.6 Replacing Injector Pipe - A, AE, and N


Important Note: The Injector Pipes are pre-
set units for use only with the
correct variant of lung demand valve
- A, AE or N. Do Not mix - fit only
the correct Injector Pipe to the
appropriate lung demand valve. For
the purpose of ease of identification
each type of Injector Pipe is colour
coded - refer to Fig. 9.

Any attempts to rotate or adjust the


cone end (1) Fig. 9 of the injector
pipe assembly will invalidate the
Dräger warranty and guarantee. If
required to replace the injector 2 1
pipe always replace with a new Type A - 1 - Black 2 - Black
factory set assembly. Type AE - 1 - White 2- Black
Type N - 1 - Black 2 - White Fig 9
Tools required 1137

Plate Spanner R26817


No. 2 Posi-Driv Screwdriver 3331245

AE and N
Assembly Tool – (Connector O ring) 3310664
Connector Tool 3310679

1 Refer to Figure 10.


(Type A illustrated.) 1 2 4 3
2 Fold rubber cover (1) from front of
lung demand valve and remove from
lung demand valve. Using plate
spanner, inserted into radial slots of
bayonet cap, turn bayonet cap (2)
counter-clockwise and remove from
body of demand valve.

Fig. 10
0584

Note: Types A and AE - the positive pressure spring remains attached to the bayonet cap. Do
Not remove spring. Take care not to stretch or distort the spring.

3 Using the thumb and forefinger carefully grip the centre plate of the diaphragm (3),
tilt centre plate and lift the diaphragm from the body of the lung demand valve.
Remove slip ring (4).

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 30

Refer to Fig. 11.

4 Using posi-driv screwdriver unscrew


retaining screw (1) until the locking
arm (2) is able to be withdrawn from
castellation of connector (3).

5 Remove connector as follows.

Note: Do Not attempt to remove push-in or


screw-in type connectors with locking
arm located in the castellation of
connector. Refer to instruction 4.

Push-in Type - A 1 2 4 3
Grip and unscrew push-in type Fig.11
connector (anti-clockwise) from 1138
demand valve body. Remove injector
pipe (4).

Screw-in Type - AE and N


Using connector tool (3310679) -
inser t pins of tool into swivel
connector - rotate to align slots of
fixed end of connector and fully
inser t pins - unscrew (anti-
clockwise) to remove screw type
connector and handwheel assembly
from demand valve body. Remove
injector pipe (4).

6 Locate and correctly position the new


injector pipe as shown in Fig. 12.

Important Note: Fit only the correct Injector


Pipe to the appropriate lung
demand valve.

7 Hold body of lung demand valve with


the injector pipe in the vertical
position then assemble connector as
follows.

Push-in Type - A
Inspect and if necessary fit new O-
ring to connector then screw Fig. 12
connector into demand valve body.
Tighten until flange of the connector 1139

contacts the body of the demand


valve. Do Not overtighten.

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 31

Screw-in Types - AE and N


Check handwheel freely rotates on
connector assembly. If necessary
use assembly tool to assemble new
O-ring to connector handwheel
assembly then using tool (3310679)
screw connector handwheel
assembly into demand valve body
until secure. Do Not overtighten.

Refer to Figure 13.

8 When viewing inside body of the


lung demand valve - carefully turn the
connector anti-clockwise until the
first castellation tab of the connector Fig. 13
is visible and centred between the 1232
internal slot in the demand valve
body. Lift the locking arm and locate
the fingers of the locking arm either
side of the castellation preventing
rotation of the connector. Ensure
correct location then carefully tighten
screw to secure the locking arm. Do
Not overtighten.

9 Reassemble diaphragm - check that


the outer bead of diaphragm is
located into the body of the demand
valve and that the slip ring is correctly
located into recess in bead of
diaphragm.

Types A and AE - Locate positive


pressure spring of bayonet cap into
the recess in centre plate of
diaphragm. Using plate spanner,
inserted into radial slots of the
bayonet cap, lock bayonet cap to
body then fold rubber cover back
over the front of the demand valve.

10 Press re-set button of types A and


AE demand valves. Connect demand
valve to equipment and carry out
Leak Test and Functional Test.

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 32

5.7 Air Transfer Block and Balanced Piston Assembly


Important Note: The Air Transfer Block and Balanced Piston Assembly is a precision pre-
set unit that incorporates a factory set Air Transfer Block Assembly containing a
pre-set and sealed balanced piston unit. These items have unlimited service life.

The Dräger guarantee is void should any attempt be made by customers to make
adjustment to the balanced piston unit or the associated Injector Pipe.

Following the fitting of a replacement balanced piston unit to an Air Transfer


Block, or the fitting of a replacement Air Transfer Block to an existing balanced
piston unit will require re-adjustment of the main lever setting height.

5.7.1 Removing Air Transfer Block and Balanced Piston Assembly


Tools required
Plate Spanner R26817
Retractor Tool 3310663
No. 2 Posi-Drive Screwdriver 3331245
Ball End Setting Tool (1.5mm) 3310676

AE and N
Assembly Tool 3310664
Connector Tool 3310679

1 If required disconnect demand valve


from female quick release coupling
of equipment.

Refer to Figure 14.

2 Fold rubber cover (1) from front of


lung demand valve. Using plate
spanner inserted into radial slots of
bayonet cap, turn bayonet cap (2)
counter-clockwise and remove from
body of demand valve. 1 2 4 3

Note: Types A and AE - the positive pressure


spring remains attached to the
bayonet cap. Do Not remove spring.
Take care not to stretch or distort the
spring.

3 Using the thumb and forefinger


carefully grip the centre plate of the
diaphragm (3), tilt centre plate and
lift the diaphragm from the body of
the lung demand valve. Remove slip
ring (4). Fig. 14
0584

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 33

Refer to Figure 15.

Important Note: Spacer (1) must be 1 3


removed before attempting to
remove Clip (2).

4 Insert the hook end of the retractor


tool into the hole in the side lugs of
the spacer (1) - pull and remove
spacer. While keeping the hose fully
inserted into the air transfer block
(3), slide the clip (2) towards the air
transfer block to release clip from
hole in body. Insert the hook end of
the retractor tool under the extended 2
pin of the clip (2) then pull to remove Fig. 15
the clip from end fitting of the hose. 1134
Grip and pull the hose to remove the
tailpiece from the body of the
demand valve.

Note: Scrap spacer (1) and clip (2) –


Do Not re-use.

Refer to Fig. 16.

5 Using posi-drive screwdriver


unscrew retaining screw (1) until the
locking arm (2) is able to be
withdrawn from castellation of
connector (3).

6 Remove connector as follows.

Note: Do Not attempt to remove push-in or


screw-in type connectors with locking
arm located in castellation of
connector. 1 2 4 3
Fig. 16
1140
Push-in Type - A
Unscrew push-in type connector (anti-clockwise) from demand valve body then remove
injector pipe (4).

Screw-in Types - AE and N


Use connector tool (3310679) - insert pins of tool into swivel connector - rotate to
align slots of fixed end of connector and insert pins - unscrew (anti-clockwise) to
remove screw type connector and handwheel assembly from demand valve body.
Remove injector pipe (4).

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 34

7 Remove Air Transfer Block and


Balanced Piston Assembly (1)
3
Fig. 17 as follows:

Types A and AE - Press re-set button


to switch ‘off’ positive pressure.

Types A, AE and N - Using two


fingers either side of the air transfer
block push the Air Transfer Block
Assembly from the outlet of the lung
demand valve.

2 1
Fig. 17
1141

8 Refer to Fig. 18. To remove the


balanced piston unit from the air
transfer block grip the air transfer
block and as shown insert ball end
setting tool through slot in the main
lever and against the end face of the
balanced piston housing. Press in
direction of arrow to push and extract 3 1
the balanced piston unit.
As necessary scrap either the air
transfer block or balanced piston
unit.

9 Check that the pivot lever (1) is in


the correct position (see
2 Fig. 18
Fig. 18) then re-assemble new air
1222
transfer block or new balanced
piston unit as required. Inspect O
rings (2) and (3) and replace with
new if necessary.

Important Note: Checking and possible re-adjustment of the main lever setting height will
be required following the fitting of a replacement balanced piston unit to an
existing Air Transfer Block, or the fitting of a replacement Air Transfer Block to
an existing balanced piston unit.

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 35

5.7.2 Fitting Air Transfer Block Assembly – Lever Height Setting


Tools required
Plate Spanner R26817
No. 2 Posi-Drive Screwdriver 3331245
Molykote 111 Grease 3331241
Ball End Setting Tool (1.5mm) 3310676
Height Gauge 3310675

AE and N
Assembly Tool 3310664
Connector Tool 3310679

Refer to Figure 19.


3
1 Type A and AE demand valves -
Press re-set lever to lift and lock the
lever in the positive pressure ‘off’
position.

2 Insert the main pivot lever (2)


through the outlet of the body of the
lung demand and locate then push
the housing of the air transfer block
assembly (1) into the bore ensuring
the flange of the housing is fully
inserted. 2 1

Types A and AE demand valves:- Fig. 19


while holding the housing of the air 1141

transfer block assembly (1) into the


bore - check the function of the
positive pressure mechanism as
follows - press the main pivot lever
to release the positive pressure
mechanism then press re-set button
(3) to switch ‘off’ positive pressure
mechanism. If satisfactory proceed
as follows.

3 Locate and position the correct


injector pipe as shown in Fig. 20
then hold body of lung demand valve
with the injector pipe in the vertical
position and assemble the connector
as follows. Fig. 20
1139

Push-in Type - A
Inspect and if necessary fit new O-ring to connector then screw connector into
demand valve body. Tighten until flange of connector contacts the body of the demand
valve. Do Not overtighten.

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 36

Screw-in Types - AE and N


Check handwheel freely rotates on
connector assembly. If necessary
use assembly tool to assemble new
O-ring to connector handwheel
assembly then using tool (3310679)
screw connector handwheel
assembly into demand valve body
until secure. Do Not overtighten.

Refer to Figure 21.

4 When viewing inside body of the


lung demand valve - carefully turn the
connector anti-clockwise until the
first castellation tab of the connector Fig. 21
is visible and centred between the
1232
internal slot. Lift the locking arm and
locate the fingers of the locking arm
either side of the castellation
preventing rotation of the connector. 1 2
Ensure correct location then
carefully tighten screw to secure the
locking arm. Do Not over-tighten.

5 Check that O ring seals (1) and (2)


Fig. 22 of tailpiece are fitted correctly
and not damaged. If necessary
replace O-rings. Apply thin coating
of Molykote 111 grease around
outside diameter of the O-rings.

Safety Note: Read instructions on Molykote


Fig. 22
111 container regarding the safe
1135
handling of silicone grease.
3
Refer to Figure 23.

6 Fully insert tailpiece through hole in


body of lung demand valve and into 2
hole in air transfer block (3) then push
and assemble a new clip (2) over the
end fitting of hose - retaining the hose
in position. Whilst keeping the end
fitting of the hose fully inserted - slide
the clip (2) away from the air transfer
block - insert a new spacer (1)
ensuring that the lugs of the spacer 1
locate around the extended pin of the
clip (2). See Fig. 23. Pull on the hose
to check that it is secure. Fig. 23

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 37

Important Note: If a replacement balanced piston unit has been fitted to an existing Air
Transfer Block, or a replacement Air Transfer Block has been fitted to an existing
balanced piston unit then checking and possible re-adjustment of the main lever
height will be required. Check height setting as follows.

7 Type A and AE demand valves: -


press the end of the main pivot lever
to release the positive pressure
mechanism.

8 Position and hold the body of the


demand valve as shown in Fig. 24
then insert the Height Gauge as
shown. Press down carefully then
release the main lever ensuring it
moves freely between the gauge
plate. Check that when the main
lever is released that the two fingers
of the lever are positioned between
the two engraved setting-lines on the
face of the gauge plate. If the lever Fig. 24
is too high or too low re-set as 1223

follows:

Insert the ball end setting tool through the slot at the pivot end of the main lever and
insert end of tool into the slot of the adjusting set screw in the secondary lever.
Rotate the set screw clockwise to increase height and anti-clockwise to reduce
height of the main lever. Following each adjustment remove the tool and press down
carefully then release the main lever and check that when the main lever is released
that the two fingers of the lever are positioned between the two engraved setting-
lines on the face of the gauge plate. Re-adjust as necessary and when set remove
the Height Gauge.

9 Carefully locate outer bead of diaphragm into the body of the demand valve. Locate
tapered side of slip ring into recess in bead of diaphragm.

Types A and AE - Locate positive pressure spring of bayonet cap into the recess in
centre plate of diaphragm.

Using plate spanner, inserted into radial slots of the bayonet cap, lock bayonet cap
to body then fold rubber cover back over the front of the demand valve.

10 Press re-set button of type A and AE demand valves. Connect demand valve to quick
release coupling of equipment and carry out Leak Test and Functional Test.

Note: Fixed variants of the lung demand valves should be fitted directly to the medium pressure
outlet port of the pressure reducer. For details of fitting of hose to the reducer refer to
the appropriate section of the relevant C.A.B.E. Training Manual.

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 38

5.8 Lubrication of Connector Assemblies - M40 and M45


Important Note: Immersing of components will eventually remove lubricants. Following
cleaning, disinfecting and drying it is recommended that functional tests be
carried out. Re-lubricate as required. Refer to Page 5:3.

Tools required
Plate Spanner R26817
No. 2 Posi-Driv Screwdriver 3331245
Connector Tool 3310679

5.8.1 Removing Connector Assemblies from LDV- M40 and M45


1 If required disconnect demand valve from the female quick release coupling of
equipment.
Refer to Figure 25. 1 2 4 3

2 Fold back the rubber cover (1) then


using plate spanner inserted into radial
slots of bayonet cap, turn bayonet cap
(2) counter-clockwise and remove
from body of demand valve. Using the
thumb and forefinger carefully grip the
centre plate of the diaphragm (3), tilt
centre plate and lift the diaphragm
from the body of the lung demand
valve. Remove slip ring (4).

Note: Type M45 - the positive pressure spring Fig. 25


remains attached to the bayonet cap. 0619
Do Not remove spring. Take care not
to stretch or distort the spring.

Refer to Fig. 26.

3 Using posi-driv screwdriver unscrew


retaining screw (1) until the locking
arm (2) is able to be withdrawn from
castellation of connector (3).

4 Remove connector as follows.

Note: Do Not attempt to remove connectors 4 3


with the locking arm located in the 1 2
castellation of the connector. Fig. 26
1186

Use connector tool (3310679) - insert pins of tool into the swivel connector and
rotate to align slots of fixed end of connector and insert pins - unscrew (anti-clockwise)
to remove screw type connector and handwheel assembly from demand valve body.
Remove injector pipe (4).

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 39

5.8.2 Disassembly of Connectors - M40 and M45


Tools required
Screwdriver
O ring Lifter R21402

Refer to Figure 27.

1 2 9 4 7 6 10 5

1 2 3 4 9 10 8
Fig. 27
1285007

1 Using the O ring lifter remove and discard O ring (1) and using suitable screwdriver
open gap in circlip (2) and remove the circlip.

2 M40 - Remove handwheel (3). Retrieve loose ball bearings (4).


M45 - Place the connector - back (5) on flat surface. Hold down connector (5) with
thumbs and using fingers lift handwheel (6) and sleeve (7) until the ball bearings (4)
are visible. Remove and retrieve ball bearings.

3 Carefully pull apart connector - back (5) and (8) from connector - front (9). Retrieve
any loose ball bearings. Remove M45 handwheel and sleeve.

4 Using the O ring lifter remove and discard quad ring (10) from connectors (5) and
(8).

5 Clean and inspect all components before re-assembly. Replace as necessary.

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 40

5.8.3 Assembly of Connector - M40


Tools required M40
Assembly Tool 3310664
Molykote 111 Grease 3331241 10 8

Refer to Figure 28.

1 Assemble new quad ring (10) to


connector - back (8) and place
flange of connector on a flat surface.
Apply a thin film of Molykote grease
around the outside of the quad ring
and into the groove for the ball
bearings. Apply a thin film of grease 1 2 3 4 9
into the bore of connector (9) then Fig. 28
push and assemble connector (9) to 1165
connector (8).

2 Insert a portion of Molykote grease


into each of the three radial holes in
the connector - front (9) then insert
a ball bearing (4) into each hole. The 1
grease will assist in holding the ball
bearings in position.

3 Place the handwheel (3) over the


connectors then assemble circlip (2)
to secure the assembly.

Refer to Figure 29.

4 Using assembly tool assemble new


O ring (1) to the connector
assembly. Assembly of connector
completed - assemble to lung
demand valve Section 5.9.5. Fig. 29
1167

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 41

5.8.4 Assembly of Connector - M45


Tools required
Assembly Tool 3310664
Molykote 111 Grease 3331241

Refer to Figure 30.


M45
1 Assemble new quad ring (10) to
connector - back (5) and place 6 10 5
flange of connector on a flat surface.
Apply a thin film of Molykote grease
around the outside of the quad ring
and into the groove for the ball
bearings.

2 Assemble the sleeve (7) over the


connector - front (9) sliding the
larger diameter over the thread.
Place the handwheel (6) over the
sleeve (7) and again push back 1 2 4 9 7
towards thread of connector. From Fig. 30
inside the bore of the connector (9) 1166

insert a portion of Molykote grease


into each of the three radial holes in
the connector then insert a ball
bearing (4) into each hole. The
grease will assist in holding the ball
bearings in position. 1
3 Place the handwheel and connector
sub-assembly over the connector (5)
then push and assemble to
connector (5). Assemble circlip (2)
to secure the assembly.

Refer to Figure 31.

4 Using assembly tool assemble new


O ring (1) to the connector
assembly. Assembly of connector
completed - assemble to lung
demand valve.
Fig. 31
Section 5.9.5.
1168

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 42

5.8.5 Assembly of Connector Assemblies to LDV - M40 and M45

Important Note: If the Injector Pipe has been


removed then fit only the correct
Injector Pipe to the appropriate
lung demand valve. For the
purpose of identification each type
of Injector Pipe is colour coded -
refer to Fig. 32. Any attempts to
rotate or adjust the cone end (1)
will invalidate the Dräger warranty
and guarantee. If it is a
requirement to replace the injector
pipe then always replace with a
new factory set assembly.

2 1
Type A - 1 - Black 2 - Black
Type AE - 1 - White 2- Black
Type N - 1 - Black 2 - White Fig. 32
1137

Tools required
Assembly Tool 3310664
Connector Tool 3310679
Plate Spanner R26817
No. 2 Posi-Driv Screwdriver 3331245

1 Locate and position the correct


injector pipe as shown in Fig. 33
then hold the body of the lung
demand valve with the injector pipe
in the vertical position and assemble
connector as follows.

2 Check handwheel freely rotates on


connector assembly. If necessary
use assembly tool to assemble new
O-ring to connector handwheel
assembly then using tool (3310679)
screw connector handwheel
assembly into demand valve body
until secure. Do Not overtighten.
Fig. 33
1139

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 43

3 When viewing inside body of the


lung demand valve - carefully turn the
connector anti-clockwise until the
first castellation tab of the connector
is visible and centred between the
internal slot. Lift the locking arm (2)
and locate the fingers of the locking
arm either side of the castellation
preventing rotation of the connector.
See Fig. 34. Ensure correct location
then carefully tighten screw (1) to
secure the locking arm. Do Not over- 2
tighten.

4 Carefully locate outer bead of 1


diaphragm into the body of the Fig 34
demand valve. Locate tapered side 1232
of slip ring into recess in bead of
diaphragm.

Types A and AE - Locate positive


pressure spring of bayonet cap into
the recess in centre plate of
diaphragm.

Using plate spanner, inserted into


radial slots of the bayonet cap, lock
bayonet cap to body then fold rubber
cover back over the front of the
demand valve.

5 Press re-set button of type AE


demand valve. Connect demand
valve to quick release coupling of
equipment and carry out Leak Test
and Functional Test.

Note: Fixed variants of the lung demand


valves should be fitted directly to the
medium pressure outlet port of the
pressure reducer. For details of fitting
of hose to the reducer refer to the
appropriate section of the relevant
C.A.B.E. Training Manual.

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 44

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 45

Section 6 Test Procedures

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 46

6 Test Procedures
The Plus and PSS Series range of Lung Demand Valves are approved for use with
approved Dräger Compressed Air Breathing Equipment (C.A.B.E.)

Standard test procedures for C.A.B.E. are as follows and are covered in the relevant
Training Manual for the C.A.B.E.

l Cylinder Pressure Check and Valve Leak Test


l High Pressure Leak Test
l Whistle Warning Unit Test
l Full Flow Test
l Medium Pressure Test - Pressure Reducer

For detailed instructions of the following Functional Tests relating to the testing of
the lung demand valve and associated ancillary items refer to the appropriate Test
Equipment Manual.

Type A and AE
l Balanced Piston Leak Test
l Positive Pressure Balance Test
l Static Pressure Test
l Positive Pressure Switch Over

Type N
l Balanced Piston Leak Test
l Demand Valve Opening Pressure

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 47

Section 7 Fault Location

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 48

7.1 Demand Valve


The high performance Plus and PSS Series Lung Demand Valves incorporate a
precision factory set Air Transfer Block Assembly containing a pre-set and sealed
balanced piston unit. These items have unlimited service life.

The Dräger warranty and guarantee is void should any attempt be made by customers
to make adjustment to the balanced piston unit or the associated Injector Pipe.

Following the fitting of a replacement balanced piston unit to an existing Air Transfer
Block, or the fitting of a replacement Air Transfer Block to an existing balanced
piston unit will require re-adjustment of the main lever setting height.

The following Fault - Cause - Remedy chart outlines a list of possible faults that may
occur.

Fault Cause Remedy


Inward/outward 1 Damaged/faulty connection O-Ring Replace O-Ring
leak when in 2 Damaged/faulty facepiece por t Replace facepiece por t
facepiece por t
Leak - Positive 1 O-Ring leak - tailpiece Replace O-Rings
pressure 'Off' 2 Balanced piston unit leak a. Replace external O-Rings.
b. Replace Balanced Piston
Unit.
3 Per forated hose Replace hose
Leak - Positive 1 Incorrectly fitted or per forated Re-assemble or replace as
pressure 'On' diaphragm. necessar y.
2 Damaged/per forated switch off Replace lung demand valve
cover body.
A or AE - swivel connector leak Replace O-Ring
High static 1 PP spring not located in centre of Locate spring correctly.
pressure diaphragm.
2 PP spring too strong - stretched Replace spring.
3 Lever setting height too high Reset lever height.
Low static pressure 1 PP spring too weak. Replace spring.
2 Lever setting height too low. Reset lever height.
Demand valve fails Faulty switch 'Off' mechanism Replace air transfer block
to switch assembly
automatically to
positive pressure
mode
No supplementary Incorrect lever setting height Reset lever height.
supply

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 49

Section 8 Spare Parts and Tool List

tm 1285.001 - January 2004


PLUS AND PSS® SERIES - LUNG DEMAND VALVE 50

Contents
1285.001 Lung Demand Valve - Push-In - Type A

1285.002 Lung Demand Valve - Screw-In - Type M45

1285.003 Lung Demand Valve - Screw-In - Type M40 (Rd40)

1285.005 Air Transfer Block and Balanced Piston Assemblies

1285.006 Lung Demand Valve - Screw-In - Type M40 (Rd40) - Straight Connector

1285.007 Connector Assemblies - M40 and M45

1285.008 Tool List

tm 1285.001 - January 2004


PANORAMA NOVA P 1

Index

Section 1 Instructions for Use

Section 2 Introduction and Safety Warnings

Section 3 Product Description

Section 4. Operating Principle

Section 5 Servicing Instructions

Section 6 Test Procedures

Section 7 Fault Location

Section 8 Spare Parts Lists and Tool List

tm 1412p - January 2004


PANORAMA NOVA P 2

tm 1412p - January 2004


PANORAMA NOVA P 3

Section 1 Instructions for Use

tm 1412p - January 2004


PANORAMA NOVA P 4

Refer to the Instructions for Use supplied with the Facepiece

tm 1412p - January 2004


PANORAMA NOVA P 5

Section 2 Introduction and Safety Warnings

tm 1412p - January 2004


PANORAMA NOVA P 6

Introduction
The Dräger Panorama Nova P facepiece conforms to the requirements of European Standards
EN136 and EN137-10 and is designed for use with approved and certified compressed air
breathing equipment and airline breathing systems. The facepiece is marked with the CE
symbol.

Two versions of the Panorama Nova P facepiece are available;

Panorama Nova P - Front Port - Figure 1

Front port assembly with integral exhalation valve and plug-in port connection for a push-in
type positive pressure lung demand valve.

Other facepieces are available with either M45 (positive pressure - screw in) or M40 (normal
demand - screw in) front port assemblies.

Panorama Nova P - Side Port - Figure 2

Front port assembly with integral exhalation valve. The plug-in port is sealed off with a blanking
piece. Fitted as standard is a plug-in side port connection for a push-in type positive pressure
lung demand valve.

When worn in combination with approved and certified compressed air breathing equipment,
or airline breathing systems, the Panorama Nova P full face facepieces provide the wearer
with full face, eyes and respiratory protection when working in a contaminated or oxygen
deficient gaseous atmosphere.

Using the most up to date materials and advanced component design these facepieces provide
the wearer with greater performance, comfort and ease of use.

Safety Warnings

l Explain - Instructions for Use supplied with this equipment, must be adhered to.

l Maintain - equipment as outlined in Dräger Limited service schedules and maintenance


instructions.

Contact Dräger Limited, Blyth for details of Service Contracts and Service Training
Courses.

l Train - potential wearers in how to wear, and use equipment and to understand its
limitations.

Contact Dräger Limited, Blyth for details of wearer Training Courses.

tm 1412p - January 2004


PANORAMA NOVA P 7

Panorama Nova P
Front Port Facepiece

Fig. 1
0086

tm 1412p - January 2004


PANORAMA NOVA P 8

Panorama Nova P
Side Port Facepiece

Fig. 2
0003

tm 1412p - January 2004


PANORAMA NOVA P 9

Section 3 Product Description

tm 1412p - January 2004


PANORAMA NOVA P 10

Product Description
The main component parts of the Panorama Nova P full face facepiece is as follows:

l Moulded rubber facepiece.


l Adjustable head harness.
l Full vision visor.
l Inner orinasal mask.
l Panorama Nova P - Front Port - Front port assembly with integral exhalation valve and
plug-in port connection.
l Panorama Nova P - Side Port - Front port assembly with integrated exhalation valve and
plug-in port connection (as front port facepiece) sealed off with a blanking piece.
l Neck strap.

Moulded Rubber Facepiece


The ergonomically designed moulded rubber facepiece provides a high degree of safety,
comfort and stability regardless of facial size, or contours.

Rubber facepiece options are:

EPDM - i.e. Ethylene-Propylene-Diene-India Rubber. Colour:- Black.


or
SI - i.e. Silicone. Colour:- Yellow.

The integral double reflex sealing lips, around the facepiece rubber, follow the facial contours
of the wearer maintaining a twofold safety feature to the faceseal.

A grooved moulded bead around the front of the facepiece rubber ensures a positive location
for the full vision visor. The opening for the front port assembly has a moulded flange for
location of the clamping ring.

The side port facepiece also incorporates a moulded flange around the side port for location
of the clamping ring.

Head Harness
The Neoprene head harness provides a safe and comfortable means of securing the facepiece
to the wearer.

Five straps with adjustable buckles radiate from a central plate which positions at the back of
the head. One strap locates centrally over the head and connects to the top of the facepiece.
The remaining straps radiate either side of the head, one pair located at eye level (top straps)
and the second pair at cheek level (lower straps).

Visor
The shape of the visor allows an undistorted 180 degree field of vision. A Polycarbonate visor
is fitted as standard and provides excellent impact and heat resistant properties. A grooved
moulded bead around the front of the facepiece moulding provides positive location for the visor
which is secured in position by two clamp frame halves and screws. Visors are easily replaceable.

Acrylic, or Laminated Glass (Triplex) visors are also available for purchase as optional extras.

tm 1412p - January 2004


PANORAMA NOVA P 11

Inner Orinasal Mask


The ergonomically designed moulded Neoprene rubber inner orinasal mask provides a high
degree of safety, comfort and stability regardless of facial size or contours.

Located inside the facepiece, the orinasal mask locates into a groove around the speech
diaphragm retaining ring in the front port assembly. Two matched inhalation valve assemblies
are fitted either side of the mask close to the nasal contact cavity. Each inhalation valve
assembly consists of a moulded valve seat and flexible rubber valve disc.

Front Port Assembly

Panorama Nova P - Front Port


The push-in front port housing incorporates the following component parts:

l Speech diaphragm.
l Exhalation valve disc.
l Exhalation valve spring
l Exhalation valve spring holder.
l Exhalation valve cover.
l Locking mechanism for plug-in connector of demand valve. (Inhalation port.)

A locking ring secures the complete housing assembly to the rubber facepiece. The front port
locking mechanism incorporates a ‘U’ shaped spring that locates into a groove in the demand
valve male connector boss to secure the demand valve into the port. Pressing a ‘push button’ on
the top of the port retracts the spring allowing the demand valve to be removed.

Panorama Nova P - Side Port


The front port housing incorporates the following component parts:

l Speech diaphragm.
l Exhalation valve disc.
l Exhalation valve spring and spring holder assembly.
l Exhalation valve cover.
l Blanking piece. (Inhalation port.)

A locking ring secures the complete housing assembly to the rubber facepiece. The front
inhalation port is sealed off with a plug-in blanking piece.

Side Port Assembly (Panorama Nova P - Side Port Only)


A locking ring secures the complete side port housing assembly to the rubber facepiece. The
push-in side port assembly incorporates a ‘U’ shaped spring that locates into a groove in the
male connector boss of the demand valve to secure the demand valve into the port. Pressing a
‘push button’ on the top of the port retracts the spring allowing the demand valve to be removed.

Supplied as optional extras are screw-in connectors for fitting to the facepiece to accommodate
screw-in demand valves.

Neck Strap
Attached to either side of the facepiece, the neck strap incorporates a hanging clip allowing
the wearer to carry the facepiece on the front of the chest in the ready for use position.

tm 1412p - January 2004


PANORAMA NOVA P 12

tm 1412p - January 2004


PANORAMA NOVA P 13

Section 4 Operating Principle

tm 1412p - January 2004


PANORAMA NOVA P 14

Operating Principle
The illustrations used to describe the operating principle show the PA90 Series lung demand
valve.

Refer to Figures 1 and

Figure 2 for the Panorama Nova P – Side Port.

Refer to Figure 3 and

Figure 4 for the Panorama Nova P - Front Port.

On opening compressed air cylinder valve medium pressure air (6bar to 9bar) flows from the
reducer through a hose and into the main bore of balanced piston block of the lung demand
valve.

Increasing air pressure moves the demand valve piston axially, forcing the piston cone against
the O-Ring located in the balanced piston block - resulting in air flow being sealed ‘OFF’ from
lung demand valve outlet.

After putting on the facepiece, with the demand valve fitted, and when the wearer inhales a
negative pressure is created inside the facepiece and lung demand valve. Due to the negative
pressure and the ambient air pressure acting against the diaphragm and hinged balanced
piston lever, the positive pressure ‘OFF’ mechanism is released. Increasing deflection of the
diaphragm pivots the balanced piston lever forcing it against the piston end cap, moving the
piston axially and lifting the piston cone from its seating. This releases medium pressure air
from the lung demand valve into the wearer’s facepiece. The air flows into the main cavity of
the facepiece and across the inside of the visor providing a demisting function. Air then flows
from the main facepiece cavity and through the two inhalation valve discs of the orinasal mask,
to the wearer.

When the wearer stops inhaling the balanced pressure inside the facepiece is between 5mmwg
to 45mmwg (0.5mbar to 4.5mbar). When the wearer begins to exhale the facepiece cavity
pressure increases closing the orinasal inhalation valves and lifting the preset spring loaded
exhalation valve disc from its sealing seat allowing the exhaled air to exhaust to atmosphere.
Following exhalation the spring loaded exhalation valve disc closes onto the sealing seat.

The exhalation valve lift off pressure is preset between:

Panorama Nova P - Front Port - 45mmwg and 57mmwg (4.5mbar to 5.7mbar).


Panorama Nova P - Side Port Port - 58mmwg and 70mmwg (5.8mbar to 7.0mbar).

Air pressure in the bore of the piston block increases during exhalation, moving the piston
axially and retracting the pivot lever and diaphragm. Two tension springs, attached to the
hinged lever of the demand valve, act on the hinged lever maintaining positive pressure inside
the facepiece during the complete breathing cycle. This cycle repeats as the wearer begins to
inhale.

tm 1412p - January 2004


PANORAMA NOVA P 15

Fig. 1
0009

tm 1412p - January 2004


PANORAMA NOVA P 16

Fig. 2
0010

tm 1412p - January 2004


PANORAMA NOVA P 17

Inhalation
Fig. 3
0087

tm 1412p - January 2004


PANORAMA NOVA P 18

Exhalation
Fig 4
0088

tm 1412p - January 2004


PANORAMA NOVA P 19

Section 5 Servicing Instructions

tm 1412p - January 2004


PANORAMA NOVA P 20

Section Description
5.1 Contents

5.2 Service and Test Intervals

5.3 Cleaning, Disinfecting and Drying

5.4 Replacing Exhalation Valve Spring

5.4.1 Panorama P - Front Port

5.4.2 Panorama P - Side Port

5.5 Replacing Exhalation Valve Disc

5.6 Replacing Inhalation Valve Discs (Inner Orinasal Mask)

5.7 Replacing Speech Diaphragm O Ring

5.8 Replacing Speech Diaphragm

5.9 Replacing Visor

5.10 Replacing Front Port Assembly

5.11 Replacing Side Port Assembly

5.12 Replacing Head Strap Buckles

tm 1412p - January 2004


PANORAMA NOVA P 21

5.2 Service and Test Intervals


Carry out regular Inspection, Testing and Servicing of equipment in accordance with
this table. Record all data in Equipment Log Book. These instructions also apply to
non-used (in Storage) equipment.

Description After Ever y Ever y


Use Month Year
Facepiece Cleaning m m
Complete Disinfecting m m
Visual Inspection m m
Functional and Leak Testing as m m
defined in Instructions for Use
Exhalation Valve Check - Visual m m m(1)
Replace when necessar y
Orinasal Valve Check Visual m m
Discs Replace when necessar y
Speech Check - Visual m(2)
Diaphragm Replace when necessar y
Speech Check - Visual m(2) m(2)
Diaphragm O-Ring Replace when necessar y

1. Annual Test Procedure – Refer to the relevant Test Instruction Manual.

2. Only applicable if the facepiece is used in areas subject to aggressive media, i.e. chlorine,
ammonia, etc.

tm 1412p - January 2004


PANORAMA NOVA P 22

5.3 Cleaning, Disinfecting and Drying


5.3.1 General
Carefully clean, disinfect and thoroughly dry contaminated dirty components after
use, or when necessary.

If using baths to contain cleaning and disinfecting solutions, immersed components


and assemblies must be agitated manually. Dräger recommends that no form of
mechanical, electrical or ultrasonic agitation be used.

Safety Note: Refer to manufacturers’ usage instructions when using cleaning and
disinfecting agents. It is important that attention be paid to concentration and
reaction times. Do Not use organic solvents such as Acetone, Alcohol, White
Spirit, Trichloroethylene or similar.

Dräger recommends:

For Cleaning:
Dräger Safety Wash
1 Litre Safety Wash and Dispenser 3380164
1 Litre Safety Wash Refill 3380165
5 Litre Safety Wash and Dispenser 3380166
5 Litre Safety Wash Refill 3380167

Sekusept
4 Bottles @ 2 litres 7904071

Do Not exceed a temperature of 30 degrees Celsius.

For Disinfecting:
Wipex Cloths
Pack of 50 3380375

Incidur (6 litre container) 7904072


Incidur (30 litre container) 7904073

Do Not exceed a temperature of 30 degrees Celsius.

Special Note: Silicone material adsorbs cleaning and disinfecting agents. Adsorption
increases with each cleaning procedure. Following the use, and long term effects
of the use of such agents, some skin irritation may occur on contact. When
cleaning and disinfecting silicone facepieces observe the following:

l Do Not use cleaning or disinfecting agents containing phenolderivatives,


aldehydes or iodine.
l Only use agents that confirm, in writing, that “daily use over a period of months
does not cause any irritation of the skin”.
l Strictly comply with minimum concentration and reaction times quoted in
instructions for use supplied with cleaning and disinfecting agents.
l Carry out rinsing in clean water for a minimum of 10 minutes.

tm 1412p - January 2004


PANORAMA NOVA P 23

Rinsing and Drying


Remove cleaning and disinfecting solutions by rinsing in clean running water followed
by drying.

Rinsing - Do Not exceed a temperature of 30 degrees Celsius.


Drying - Do Not exceed a temperature of 60 degrees Celsius. Do Not dry rubber
parts on heaters. Keep out of direct sunlight.

5.4 Replacing Exhalation Valve Spring


5.4.1 Panorama Nova P - Front Port

1 Refer to Fig. 1.

Unclip and remove the front cover


(1) of the exhalation valve.

Fig. 1
0048

2 Refer to Fig. 2.

Press the exhalation valve spring


holder inwards at one end then
remove the holder and spring
assembly.

3 Disassemble spring from holder and


scrap the spring. Do Not re-use.

Fig. 2
0049

tm 1412p - January 2004


PANORAMA NOVA P 24

4 Refer to Fig.3.

Assemble a new spring to the holder


ensuring that the radial leg of the 2
spring is correctly located.

Note: Insert the first coil of the spring (2)


into the outside groove in the centre
boss of the spring holder, ensuring
the radial leg of the spring locates
into the radial slot.

5 Inser t the spring and holder


assembly so that the holder location
lugs engage into the slots in the
housing and the spring locates Fig. 3
centrally to the metal support disc
0089
of the silicone valve disc assembly.

6 Confirm correct location by checking


holder has a small side to side
movement in housing then refit the
front port cover.

7 Carry out Functional Tests.


See Section 6.

5.4.2 Panorama Nova P - Side Port


1 Unclip and remove the front cover of the exhalation valve.

2 Press the exhalation valve spring holder inwards at one end then remove the holder
and spring assembly. Scrap the defective spring and holder assembly. Do Not re-
use.

3 Insert the new spring and holder assembly so that the holder location lugs engage
into the slots in the housing and the spring locates centrally to the metal support
disc of the silicone valve disc assembly.

4 Confirm correct location by checking holder has a small side to side movement in
housing then refit the front port cover.

5 Carry out Functional Tests. See Section 6.

tm 1412p - January 2004


PANORAMA NOVA P 25

5:5 Replacing Exhalation Valve Disc


1 Unclip and remove the front cover of the exhalation valve.

2 Press the exhalation valve spring holder inwards at one end then remove the holder
and spring assembly.

3 Refer to Fig. 4.

Grip the centre nipple of the


exhalation valve disc and pull out the
valve disc assembly (1) from the
guide. Scrap the defective valve disc 1
assembly.

4 Insert a new valve disc assembly.


2
Note: Before fitting the valve disc assembly
check that sealing rim(2) in housing
is clean and not damaged and that
valve stem slides freely in guide bore.
I.e. with disc in horizontal position and Fig. 4
the stem vertical it should fall freely 0050

onto the sealing rim under its own


weight.

5 Insert the spring and holder assembly so that the holder location lugs engage into
the slots in the housing and the spring locates centrally to the metal support disc of
the silicone valve disc assembly.

6 Confirm correct location by checking holder has a small side to side movement in
housing then refit the front port cover.

7 Carry out Functional Tests. See Section 6.

5.6 Replacing Inhalation Valve


Discs (Inner Orinasal Mask)
1 Refer to Fig. 5.

Grip and remove the inner orinasal


mask assembly (1) from inside the
facepiece.

1
Fig. 5
0051

tm 1412p - January 2004


PANORAMA NOVA P 26

2 Refer to Fig. 6.
2 2
Remove defective rubber valve discs
(1) from valve seats (2). Scrap the
defective rubber valve discs.

Note: If replacing complete inhalation valve


assemblies use as pairs from same
spares pack.

Do Not mix valve discs assemblies


from different packs. 1 1

3 Moisten the stem of each new valve


disc, (1) Fig. 6, with water. Insert
the stem, from inside orinasal mask Fig. 6
(from sealing rim side) into bore of 0052

valve seat (2) and pull the stem


through from the other side until the
stem engages. Repeat on second
valve.

4 Locate the main bore of the orinasal


mask into the groove of the
diaphragm retaining ring then rotate
the orinasal mask until the moulded
central marking aligns with central
marking on the front port of the
facepiece. See Fig. 7. Insert the
flange of exhalation bore into the
orifice in the exhalation valve port.

Fig. 7
0053

5.7 Replacing Speech Diaphragm O-Ring


Tools Required:
Plate Spanner R26817
O Ring Lifter R21402

1 Grip and remove the inner orinasal mask assembly from inside the facepiece.

tm 1412p - January 2004


PANORAMA NOVA P 27

2 Refer to Fig. 8.

Using the plate spanner (large pin


spacing) unscrew (anti-clockwise),
and remove the (1). Re-orientate the
pin spanner then carefully insert
narrow pin spacing into the
perforations of the protection disc
of the diaphragm. Rotate the speech
diaphragm (2) to release, then
remove the diaphragm from the
housing. 1

Note: Do Not damage perforated protection 2


disc or diaphragm membrane.
Fig. 8
4 Refer to Fig. 9. 0054

Using the lifter tool (R21402)


carefully remove the defective O ring 1 2
(1) and scrap the defective O ring.
Check the O ring location groove is
not damaged and insert new O ring
(1). Locate the speech diaphragm
(2) (convex perforated protection
side towards inside of facepiece)
into housing.

5 Refer to Fig. 10.

Screw the diaphragm retaining ring


(1) into housing and using the plate
spanner (large pin spacing) tighten Fig. 9
to secure and seal diaphragm. 0056

6 Locate the main bore of the orinasal


mask into the groove of the 1
diaphragm retaining ring then rotate
the orinasal mask until the moulded
central marking aligns with central
marking on the front port of the
facepiece. Insert the flange of
exhalation bore into the orifice in the
exhalation valve port.

7 Carry out Leak Test. See Section 6.

Fig. 10
0057

tm 1412p - January 2004


PANORAMA NOVA P 28

5.8 Replacing Speech Diaphragm


Tools Required:
Plate Spanner R26817
O Ring Lifter R21402

1 Grip and remove the inner orinasal mask assembly from inside the facepiece.

2 Using the plate spanner (large pin spacing) unscrew (anti-clockwise), and remove
the (1). Re-orientate the pin spanner then carefully insert narrow pin spacing into
the perforations of the protection disc of the diaphragm. Rotate the speech diaphragm
(2) to release, then remove the diaphragm from the housing. Scrap the defective
diaphragm.

3 Check that the O ring is not damaged. If suspect replace the O ring.

4 Locate the new speech diaphragm (convex perforated protection side towards inside
of facepiece) into housing. Screw the diaphragm retaining ring into housing and
using the plate spanner (large pin spacing) tighten to secure and seal diaphragm.

5 Locate the main bore of the orinasal mask into the groove of the diaphragm retaining
ring then rotate the orinasal mask until the moulded central marking aligns with
central marking on the front port of the facepiece. Insert the flange of exhalation
bore into the orifice in the exhalation valve port.

6 Carry out Leak Test. See Section 6.

5.9 Replacing Visor


Tools Required:
7mm/8mm Open Ended Spanner 3331244
Screwdriver (5.5 x 1) 3331246
Screwdriver (Posidriv) 3331245

1 Refer to Fig. 11
1 1
Using the spanner and screwdriver
(3331246), or posidriv screwdriver
(3331245), remove the two
clamping screws (1) and locknuts
(2). Scrap the locknuts. Do Not re-
use.

2 2

Fig. 11
0058

tm 1412p - January 2004


PANORAMA NOVA P 29

2 Prise the clamping frames apart,


(Fig. 12). Remove the upper frame
then twist and remove the lower
frame from the support plate, if fitted,
and the rubber bead of the
facepiece.

3 Remove the visor from the groove in


the rubber bead of the facepiece.
Scrap the defective visor.

4 Select a new visor ensuring it is of


the correct material. Check the
identification marking at the top
centre position of the visor.
Fig. 12
l PC stands for Polycarbonate. 0060

l L stands for
laminated safety glass.

5 Assemble the new visor by first


locating the top of the visor
(designation marking to top) into the
groove in upper portion of the
facepiece bead - then into the lower
portion. Align the central markings
0.5mm to
on visor with the centre moulding
3.0mm
seam of the rubber facepiece then
locate the remainder of rubber bead
around the visor.

Fig. 13
0061

6 To assist assembly of the clamp frame halves, moisten the outside of facepiece
bead, and the inside groove of clamp frame halves with water. Press and locate the
upper clamp centrally over the bead. Locate the lower clamp centrally into the support
plate, if fitted, and over bead. Align centre marking on each clamp with centre
moulding mark of facepiece rubber.

7 Insert the screws through the bosses of the clamp frames and fit new locknuts.
Tighten equally until the gap between bosses of clamp frames is between 0.5mm to
3.0mm. See Fig. 13.

Note: The end of the screws should be at least level with the end face of the locknut.

8 Carry out Leak Test. See Section 6.

tm 1412p - January 2004


PANORAMA NOVA P 30

5.10 Replacing Front Port Assembly


Tools Required:
Screwdriver
3331246
Oetiker Pliers R53239

1 Grip and remove the inner orinasal mask assembly from inside the facepiece.

2 Refer to Fig. 14.

Using a screwdriver in position as


indicated by the arrow, lift and open
the clamp. Remove the clamp and
slip ring and scrap.

3 Pull and remove the front por t


assembly from the body of the
facepiece and discard the defective
front port housing. Check that the
rubber flange of the facepiece is not
damaged.

4 Insert a new front port assembly into


the body of the facepiece, aligning
the central moulding seam of front
Fig. 14
por t housing with the centre
moulding seam of the facepiece 0062

rubber. Check that moulded flange


of facepiece locates correctly around
the front port housing.

5 Assemble a new slip ring and a new


clamp around the moulded flange
and over the support plate, in centre
of lower visor clamp. Insert and
engage the hooks of the clip into
middle two holes of the clip.
See Fig. 15.

Fig. 15
0063

tm 1412p - January 2004


PANORAMA NOVA P 31

6 Using Oetiker pliers, (R53239),


squeeze the nose of clip to secure
and seal the front port assembly to
the facepiece. See Fig. 16.

Note: Squeeze nose of clip by 3mm to


4mm.

7 Carry out Leak and Functional Tests.


See Section 6.

Fig. 16
0064

5.11 Replacing Side Port Assembly


Tools Required:
Screwdriver 3331246

1 Using screwdriver remove the screw from the side port clamp ring and remove the
clamp ring.

2 Pull and remove the side port assembly from the body of the facepiece and scrap
the defective assembly Check that the rubber flange of the facepiece is not damaged.

3 Insert a new side port assembly into


the body of the facepiece then
aligning the push button as shown
in Fig. 17. Check that moulded
flange of facepiece locates correctly
around the housing.

4 Assemble clamp ring around the


side por t flange aligning the
bosses of the clamp ring as shown
in Fig. 17. Insert the retaining
screw and using the screwdriver
tighten the clamp ring until the gap
between clamp ring bosses is
between 2mm to 3mm.

5 Carry out Leak Test. 2-3mm


See Section 6.

Fig. 17
0065

tm 1412p - January 2004


PANORAMA NOVA P 32

5.12 Replacing Head Strap Buckles

1 Refer to Fig. 18 for the correct


assembly of the headstrap buckles
to the straps of the head harness.

Fig. 18
0066

tm 1412p - January 2004


PANORAMA NOVA P 33

Section 6 Test Procedures

tm 1412p - January 2004


PANORAMA NOVA P 34

6 Test Procedures
Refer to the appropriate Test Equipment Manual, (i.e. A900 Test Unit, Testor, Test
Box, Quaestor) for the standard test procedures for the Panorama P facepiece.

Test the facepiece in combination with approved and certified compressed air
breathing equipment, or airline breathing systems, and in line with the test procedures
specified in the appropriate equipment Service Manuals.

tm 1412p - January 2004


PANORAMA NOVA P 35

Section 7 Fault Location

tm 1412p - January 2004


PANORAMA NOVA P 36

7 Fault Location

FAULT CAUSE REMEDY

Inward Leak 1 Exhalation valve leaking Clean and/or replace


silicone disc
Inspect exhalation valve
seat in front por t for
damage.
2 Loose connections Clean and replace
Tighten connections
3 Face seals leaking Inspect for damage
4 Facepiece rubber body Replace facepiece rubber
per forated (pin holes, body
cracking)

Will not 1 Dir ty exhalation valve disc Clean valve disc and
hold/maintain positive and/or sealing rim sealing rim
pressure 2 Weak exhalation valve spring Replace exhalation valve
spring assembly

Unable to maintain 1 Strap ratchets worn Replace headstrap


headstrap adjustment 2 Buckle rollers incorrectly Assemble correctly
fitted

tm 1412p - January 2004


PANORAMA NOVA P 37

Section 8 Spare Parts Lists and Tool List

tm 1412p - January 2004


PANORAMA NOVA P 38

Spare Parts Lists and Tool List


Number Title

1412.052 Panorama Nova P - Front Port

1281.7 Tool List - Facepieces

tm 1412p - January 2004


CYLINDER VALVES 1

Index

Section 1 Instructions for Use

Section 2 Introduction and Safety Warnings

Section 3 Product Description

Section 4 Operating Principle

Section 5 Servicing Instructions

Section 6 Test Procedures

Section 7 Fault Location

Section 8 Spare Parts Lists and Tool List

tm 9822 - January 2004


CYLINDER VALVES 2

tm 9822 - January 2004


CYLINDER VALVES 3

Section 1 Instructions for Use

tm 9822 - January 2004


CYLINDER VALVES 4

Refer to Instructions for Use supplied with the equipment for use with the compressed air
valved cylinder assemblies.

tm 9822 - January 2004


CYLINDER VALVES 5

Section 2 Introduction and Safety Warnings

tm 9822 - January 2004


CYLINDER VALVES 6

Introduction
The Dräger Cross Flow, Right Angled Flow and In Line Cylinder Valves conform to the
following standards:

Valve Outlet Connections


l British Standard BS 341
l German Standard DIN 477
l European Standard EN 144-2

Valved Cylinder Assembly


Compressed Air Valved Cylinder Assemblies meet the requirements of EC Council Directive
89/686/EEC and when used in combination with approved and certified compressed air
breathing equipment conform to the following standards:

l European Standard EN 137 - C.A.B.E. Working Sets.


l European Standard EN 402 - C.A.B.E. Escape Equipment.
l European Standard EN 139 - C.A.B.E. Airline Equipment.

All valve types are designed with or without ratchet operation and are available in either
200bar or 300bar pressure versions, with the valve to cylinder thread connections suitable
for use with a wide range of approved and certified compressed air containers, (cylinders).

Compressed Air Valved Cylinder Assemblies must be charged (pressurised) only with breathable
quality air meeting the air quality standard as defined in EN 12021. Refer also to BS 4275 for
guidance on selection, use and maintenance of respiratory protection equipment. Cylinders used
in combination with approved and certified compressed air breathing equipment provide the
breathing equipment wearer with an independent source of breathable quality air. This combined
system provides respiratory protection in a contaminated, or oxygen deficient gaseous atmosphere.

Duration available is dependant on capacity (volume) of cylinder selected and of course the
breathing work rate of the wearer.

Safety Warnings
l Explain - Instructions for Use for equipment used with compressed air valved cylinder
assemblies, must be adhered to.

l Maintain - equipment and valved cylinder assemblies as outlined in Dräger Limited


service schedules and maintenance instructions. Compressed Air Valved Cylinder
Assemblies must be charged with breathable quality air meeting air quality standard as
defined in EN 12021.

Contact Dräger Limited, Blyth for details of Service Contracts and Service Training
Courses.

l Train - potential wearers in how to wear, and use equipment an to understand its
limitations.

Contact Dräger Limited, Blyth for details of wearer Training Courses.

tm 9822 - January 2004


CYLINDER VALVES 7

Cross-Flow Non-Ratchet Valve

Fig. 1
0080

tm 9822 - January 2004


CYLINDER VALVES 8

Right Angle Flow - Non-Ratchet

Fig. 2
1304

tm 9822 - January 2004


CYLINDER VALVES 9

In-Line Ratchet Valve

Fig. 3
0004

tm 9822 - January 2004


CYLINDER VALVES 10

tm 9822 - January 2004


CYLINDER VALVES 11

Section 3 Product Description

tm 9822 - January 2004


CYLINDER VALVES 12

Product Description
The main component parts of the Cross Flow, Right Angled Flow, and In Line Cylinder Valves
are as follows:

Type 1 - Non-Ratchet
l Valve body.
l Valve head.
l Spindle.
l Spindle bearing.
l Spindle bearing seal.
l Control knob (handwheel).

Type 2 - with Ratchet


l Valve body.
l Valve head.
l Spindle.
l Spindle bearing.
l Spindle bearing seals.
l Ratchet cap.
l Ratchet control knob (handwheel).

Valve Body
The precision machined valve bodies are manufactured from a Brass (Cu Zn 40 Pb2 F37)
pressing and are available in 200bar or 300bar pressure versions.

The cylinder thread connection variants are as follows:

l BS to BS341
l DIN to DIN477 and European Standard EN 144-2
l M18x1.5 carbon fibre cylinder specification HSE AL FW1 and FW2
l M25x2 carbon fibre cylinder specification HSE AL FW1 and FW2

The high pressure outlet port connection of the valve is to specification G5/8 as per DIN477:
Part 6. Other connections are available on request.

Valve Head
Phosphor Bronze to PB1 BS 2874, the machined valve head incorporates a square drive
socket for locating to the mating (male) drive square on the end of the spindle. Retained in
the end face of the valve head is a ‘Torlon’ insert with precision machined face used to seal
onto the high pressure sealing rim in the valve body.

Spindle
Stainless Steel to DIN 671, the machined spindle incorporates a male drive square for locating
into the mating (female) square socket in the end of the valve head. Two designs of spindle
are available - ratchet or non ratchet.

Spindle Bearing
High Tensile Brass to BS 2874, the threaded spindle bearing is torqued into the valve body to
a value of 40Nm to 41Nm (30ft lbs).

tm 9822 - January 2004


CYLINDER VALVES 13

Spindle Bearing Seals - Non Ratchet Valve


Nylon 66 sealing washer.

Spindle Bearing Seals - Ratchet Valve


Supplied in a kit consisting of the following components:

l Pair of Polyacetal (Natural) sealing washers.


l One Nitrile O ring.

The O ring locates between the two sealing washers.

Non Ratchet Control Knob (Handwheel)


Moulded in Latamid 66 25% Glass Filled Nylon.

Ratchet Cap
Moulded in Latamid 66 25% Glass Filled Nylon.

Ratchet Control Knob (Handwheel)


Moulded in Latamid 66 25% Glass Filled Nylon.

tm 9822 - January 2004


CYLINDER VALVES 14

tm 9822 - January 2004


CYLINDER VALVES 15

Section 4 Operating Principle

tm 9822 - January 2004


CYLINDER VALVES 16

Section Description
4.1 Non Ratchet Cylinder Valves

4.2 Ratchet Cylinder Valves

tm 9822 - January 2004


CYLINDER VALVES 17

Operating Principle
4.1 Non Ratchet Cylinder Valves
Refer to Figures 1-4

When valve is in the ‘closed’ position, the nylon insert in the valve head seals onto
the sealing rim of the valve body preventing high pressure air being discharged from
the cylinder.

Note: Do Not fully open cylinder valve unless valve connected to compressed air breathing
equipment or high pressure test gauge.

Slowly turning the handwheel of the valve anticlockwise, rotates the spindle and the
threaded valve head. As the valve head rotates it moves away from the sealing rim of
the valve body sliding up the square drive on the end of the spindle. The released
high pressure air from the cylinder discharges to the compressed air breathing
equipment attached to the valve outlet. To ensure no restriction of air to breathing
equipment and to maximise the air flow from the cylinder the valve should always be
fully opened.

Slowly turning the handwheel clockwise, rotates the spindle and threaded valve
head. As the valve head rotates it moves towards the sealing rim of the valve body
sliding down the square drive on the end of the spindle. When the nylon insert
contacts the sealing rim of the valve body this closes the valve preventing high
pressure air being discharged from the cylinder. Closing the valve requires minimal
force - use fingers and thumb- Do Not use excessive force, or tools to seal off
(close) the valve.

Note: Right Angled Flow valves operate as described above.

tm 9822 - January 2004


CYLINDER VALVES 18

Non-Ratchet In-Line
Cross Flow

Closed Closed

Fig. 1 Fig.2
0081 0013

Open Open

Fig. 3 Fig. 4
0082 1092

tm 9822 - January 2004


CYLINDER VALVES 19

4.2 Ratchet Cylinder Valves


Refer to Figures 5-8

When valve is in ‘closed’ position, the nylon insert in the valve head seals onto the
sealing rim of the valve body preventing high pressure air being discharged from the
cylinder.

Note: Do Not fully open cylinder valve unless valve connected to compressed air breathing
equipment or high pressure test gauge.

During anticlockwise rotation of the ratchet handwheel of the valve to ‘open’ the
valve, the control knob intermittently lifts and emits an audible ‘click’ as it moves over
the ratchet projections of the ratchet cap. Slowly turning the ratchet handwheel
rotates the spindle and threaded valve head. As the valve head rotates it moves away
from the sealing rim of the valve body sliding up the square drive on the end of the
spindle. The released high pressure air from the cylinder discharges to the
compressed air breathing equipment attached to the valve outlet. To ensure no
restriction of air to breathing equipment and to maximise the air flow from the cylinder
the valve should always be fully opened.

To close the valve, pull the ratchet handwheel away from the valve to lift and over-
ride the ratchet feature during rotation of the handwheel. Repeated pulling and
clockwise turning of the handwheel rotates the spindle and threaded valve head. As
the valve head rotates it moves towards the sealing rim of the valve body sliding
down the square drive on the end of the spindle. When the nylon insert in the valve
head contacts the sealing rim of the valve body this closes the valve preventing high
pressure air being discharged from the cylinder. Closing the valve requires minimal
force - use fingers and thumb - Do Not use excessive force, or tools, to seal off
(close) the valve.

Note:Right Angled Flow valves operate as described above.

tm 9822 - January 2004


CYLINDER VALVES 20

Ratchet In-Line
Cross Flow
Closed Closed

Fig. 5 Fig. 6
0083 0011

Open Open

Fig. 7 Fig. 8
0084 0014

tm 9822 - January 2004


CYLINDER VALVES 21

Section 5 Servicing Instructions

tm 9822 - January 2004


CYLINDER VALVES 22

Section Description
5.1 Service and Test Intervals

5.2 Cleaning

5.3 Charging Cylinders

5.4 Disassembly of Valve Components

5.5 Assembly of Valve Components

tm 9822 - January 2004


CYLINDER VALVES 23

5.1 Service and Test Intervals


Carry out regular Inspection, Testing and Servicing of the equipment in accordance
with this table. Record all data in the equipment Log Book. These instructions also
apply to non-used (in Storage) equipment.

Description After Ever y


Use Month
Valved Clean and Visually Inspect m
Cylinder
Assembly
Charge Cylinder to correct pressure m m
Charged pressure check (HP Check) m
Safety 'pressure test' and recer tification (according
to National Standards for cylinder type)
Cylinder Basic Overhaul (when necessar y or at time of
Valve Cylinder Re-Cer tification

5.2 Cleaning
After use, carefully clean, thoroughly dry and inspect contaminated dirty valved cylinder
assemblies, followed by charging.

Safety Note: Check Test Date, stamped on cylinder by last Testing Station. Do Not use
cylinder if past the Test Date. Arrange for recertification of the cylinder.

Refer to manufacturers’ usage instructions when using cleaning agents. It is


important that attention be paid to concentration and reaction times. Do Not use
organic solvents such as Acetone, Alcohol, White Spirit, Trichloroethylene or
similar.

Dräger recommends:

For Cleaning:
Dräger Safety Wash
1 Litre Safety Wash and Dispenser 3380164
1 Litre Safety Wash Refill 3380165
5 Litre Safety Wash and Dispenser 3380166
5 Litre Safety Wash Refill 3380167

Sekusept
4 Bottles @ 2 litres 7904071

tm 9822 - January 2004


CYLINDER VALVES 24

5.3 Charging Cylinders


After use, charge the cylinder assemblies with breathable quality air meeting the air
quality standard as defined in EN 12021. (Refer also to BS 4275). Charge the
cylinder to the correct rated pressure as stamped on the cylinder. Do Not overcharge.

Safety Note: Only charge cylinders which:

l Conform to National Standards.


l Are stamped with the ‘test date’ and ‘test mark’ of approved the Testing Authority
and have not exceeded the test interval indicated on the cylinder.
l Are in good condition, showing no signs of damage, corrosion or moisture
(water droplets) in the threaded port connection of the valve.

1 Screw the male adaptor of compressor charging hose into female outlet port of valve.

2 Follow ‘Charging Instructions’ as specified by manufacturer of compressor.

Note: Charge cylinder to correct pressure as stamped on cylinder. Do Not over


charge.Charging can induce an increase in air and cylinder temperature resulting in an
incomplete charge. To minimise this affect Dräger recommend a charging rate of 27bar/
minute. Following charging re-check the cylinder pressure at ambient temperature.
Boost charge if necessary.

5.4 Disassembly of Valve Components


Tools Required:
Peg Spanner 3331240
18mm/19mm Open Ended Spanner 3332188

Safety Note: Fully vent (discharge) air from the cylinder before disassembly of the
valve components. Do Not direct the discharging air stream towards the eyes,
face or skin.

Refer to Figure 1 - Cross Flow Valve and Right Angled Flow.

Refer to Figure 2 - In Line Valve.

1 Using the peg spanner, located into radial slots or into holes in face of flange nut (1
or 1a), unscrew the flange nut and remove spring (2 or 2a).

2 Remove handwheel - non ratchet (3), ratchet (3a), from the end of the spindle.

3 Ratchet Type:
Remove the nylon spacer (4) and lift the ratchet cap (5) from hexagon of spindle
bearing (6a). During disassembly of the ratchet cap if damage occurs to any of the
three retaining lugs of the ratchet cap, scrap the cap.

4 Using 18mm AF spanner unscrew, (anticlockwise), and remove the bearing (6 or


6a) from the valve body.

tm 9822 - January 2004


CYLINDER VALVES 25

Orientation
of 8
Seals
7a
8 11

10
11
7 11 10
Ratchet 11
6a
9
5
6 9 4

Non-Ratchet
3
3a

2 2a
1a
1
Fig. 1
0085

tm 9822 - January 2004


CYLINDER VALVES 26

Orientation 8
of
Seals 7
10
10
9 9
10
10
6a

5
4
3
3a

2
2
1
1a
Ratchet Type Non-Ratchet Type

Fig. 2
0314

tm 9822 - January 2004


CYLINDER VALVES 27

5 Locate the handwheel (3 or 3a) onto the spindle (7 or 7a), and use the handwheel
to rotate the spindle anti-clockwise to unscrew and remove the valve head (8).

6 Clean and inspect all components. Scrap any damaged or suspect components and
replace with new.

5.5 Assembly of Valve Components


Tools Required:
Peg Spanner 3331240
18mm/19mm Open Ended Spanner 3332188
Molykote 111 Grease 3331247
18mm AF socket and Torque Wrench (40Nm)

Note: All components must be clean and undamaged. Inspect prior to assembly.

1 Apply a small amount of grease (3331247) onto the thread of the valve head (8).

Important Note: Do Not allow grease to contact the ‘Torlon’ sealing face of the valve
head.

2 Apply a small amount of grease (3331247) onto seal location diameter of spindle (7
or 7a) and locate the first sealing washer (11) then the O-ring (10) and then the
second sealing washer (11), or sealing washer (9) onto spindle. (See Figs. 1 and
2 for correct orientation of sealing components.).

3 Locate the valve head (8) onto the square drive of the spindle (7 or 7a), and using
the spindle, screw the valve head into valve body. Rotate clockwise until the valve
head contacts the sealing rim in valve body.

4 Locate the spindle bearing (6 or 6a) over the spindle and screw the bearing into the
valve body. The spindle bearing should screw fully into the valve body by hand,
without restriction.

5 Using 18mm AF adaptor and torque wrench, tighten the spindle bearing to valve
body to a torque of 40Nm to 41Nm (30ft lbs).

6 Ratchet Type:
During disassembly of the ratchet cap if damage had occurred to any of the three
retaining lugs of the ratchet cap, scrap the cap and use a new ratchet cap. Locate
and press the ratchet cap (5) onto hexagon of spindle bearing (6a) and locate nylon
spacer (4) into the recess in the end face cap.

7 Locate the handwheel (3 or 3a) onto spindle ensuring correct location. Insert the
spring (2 or 2a) into the recess in handwheel and using the peg spanner screw the
flange nut (1 or 1a) onto the end of the spindle. Do Not over tighten.

8 To check rotation is smooth and unrestricted, operate valve as follows:

tm 9822 - January 2004


CYLINDER VALVES 28

Non Ratchet Valve


Slowly turn the handwheel anticlockwise until rotation stops i.e. valve fully open
valve. Do Not use excessive force or tools to open the valve. Slowly turn handwheel
clockwise until rotation stops i.e. valve closed. Closing the valve requires minimal
force - fingers and thumb - Do Not use excessive force or tools.

Ratchet Valve
Slowly turn the handwheel anticlockwise until rotation stops i.e. valve fully open. Do
Not use excessive force or tools to open the valve. During rotation of handwheel, to
‘open’ the valve, the handwheel intermittently lifts and emits an audible ‘click’ as it
moves over the ratchet projections of the ratchet cap.

To close the valve, pull the handwheel away from the valve to lift and over-ride the
ratchet feature. Pull and turn the handwheel clockwise until rotation stops i.e. valve
closed. Closing the valve requires minimal force - fingers and thumb - Do Not use
excessive force, or tools.

10 Charge the cylinder assemblies with breathable quality air meeting air quality standard
as defined in EN 12021. (Refer also to BS 4275). Charge cylinder to correct
pressure as stamped on cylinder.
Do Not overcharge. Refer to Section 5.4.

tm 9822 - January 2004


CYLINDER VALVES 29

Section 6 Test Procedures

tm 9822 - January 2004


CYLINDER VALVES 30

Section Description
6.1 Cylinder Pressure Check

6.2 Air Quality Test

tm 9822 - January 2004


CYLINDER VALVES 31

6.1 Cylinder Pressure Check


Following charging of the cylinder after use, and at monthly intervals if cylinder has
not been in use, carry out cylinder high pressure test as follows:

Tools Required
High Pressure Test Gauge 3331209

Note: Check Test Date, stamped on cylinder by last Testing Station.


Do Not use cylinder if past the Test Date. Arrange for recertification of cylinder.

1 Screw the handwheel of the check gauge into the outlet port of cylinder valve and
close the relief valve of the check gauge.

2 Open the cylinder valve slowly, but fully. Check reading of check gauge.

Note: If the cylinder pressure low, charge to required pressure (stamped on neck or shoulder
of cylinder). If the cylinder is to be used as an air source when performing functional
tests of associated compressed air breathing equipment, then the cylinder pressure
must be at least 80% of maximum working pressure (stamped on neck or shoulder of
cylinder, i.e., 200bar cylinder at least 160bar - 300bar cylinder, at least 240bar).

3 Close the cylinder valve. Open the relief valve of the check gauge to vent pressure
then remove the check gauge from the cylinder valve.

4 Pass the fully charged and checked cylinder to ‘Ready for Use Area’.

6.2 Air Quality Test


Personnel responsible for the respiratory protection equipment should examine and
test the quality of compressed air to ensure conformity to the air quality standard as
defined in EN 12021. (Refer also to BS 4275).

Breathable air should meet the following standards of purity:


l If not specified otherwise, in any event all contaminants should be kept to as
low a level as possible and shall be far below the national occupational exposure
limits (OEL) for substances hazardous to health.
l The air shall be without significant odour or taste.

For the evaluation testing of breathing air quality for compressors and compressed air
cylinders Dräger recommends using the Dräger High Pressure Aerotest - which connects
directly to a charging compressor outlet or the outlet port of a compressed air cylinder
valve, and uses Dräger Detector Tubes for evaluation and measurement of the following:

Note: Air supplied to the wearer should not contain impurities in excess of limits stated:

l Carbon Dioxide 500 parts per million (500 ml/m3)


l Carbon Monoxide 15 parts per million (15 ml/m3)
l Oil (Mist and Vapour) 0.5 mg/m3
l Water Vapour Content of air supplied by the compressor for 300bar or 200bar
cylinders should not exceed 25 mg/m3

tm 9822 - January 2004


CYLINDER VALVES 32

Dräger High Pressure Aerotest


Part Number - D 20700

Dräger Tubes Part Number


Carbon Dioxide 100/a-P (100 -3,000ppm) 6728521
Carbon Monoxide 5/a-P (2.5-150ppm) 6728511
Oil (Mist and Vapour) 10/a-P (0.1-10 mg/m3) 6728371
Water Vapour 5/a-P (2-450 mg/m3) 6728531
Nitrous Fumes 0.5/a (0.25-5ppm) CH29401

tm 9822 - January 2004


CYLINDER VALVES 33

Section 7 Fault Location

tm 9822 - January 2004


CYLINDER VALVES 34

7 Fault Location
FAULT CAUSE REMEDY
Leak from valve 1 Valve head leaking Replace valve head and seals
por t 2 Sealing rim in valve body Return to Dräger (Replace
damaged cylinder valve)
Leak through 1 Bent spindle Replace spindle
control knob 2 Damaged spindle seals Replace spindle seals
(handwheel)
Will not 1 Valve head leaking Replace valve head and seals
hold/maintain 2 Sealing rim in valve body Return to Dräger
pressure damaged (Replace cylinder valve)
Stiff handwheel 1 Damaged valve head Lubricate or change valve head
2 Bent spindle Replace spindle
3 Damaged spindle bearing Replace spindle bearing

tm 9822 - January 2004


CYLINDER VALVES 35

Section 8 Spare Parts Lists and Tool List

tm 9822 - January 2004


CYLINDER VALVES 36

Spare Parts Lists

Number Title

9822.1 Cross Flow Cylinder Valves - Ratchet and Non Ratchet

9822.8 In Line Cylinder Valves - Ratchet and Non Ratchet

9822.10 Right Angled Flow - Ratchet

9822.14 Right Angled Flow - Non Ratchet

tm 9822 - January 2004


A900 TEST KIT 1

Index

Section 1 Introduction and Safety Warnings

Section 2 A900 Test Kit

Section 3 Setting Up and Checking Test Equipment

Section 4 Test Procedures

Section 5 Fault - Cause - Remedy

te 1412 - January 2004


A900 TEST KIT 2

te 1412 - January 2004


A900 TEST KIT 3

Section 1 Introduction and Safety Warnings

te 1412 - January 2004


A900 TEST KIT 4

Introduction
The following Functional Tests detailed in this manual cover the A900 Test Unit procedures
for the testing of Positive Pressure (Push-In) Lung Demand Valves which are used in conjunction
with the following facepieces,

Panorama Nova P Side Port


Panorama Nova P Front Port
Futura P

l Balanced Piston Leak Test


l Static Positive Pressure Balance Test
l Whistle Activation Test
l Exhalation Valve Inward Leak Test
l Negative Pressure Leak Test - Facepiece

For detailed instructions of the following additional tests refer to Test Procedure Section of
the appropriate Service/Training Manual for the C.A.B.A. in use.

l Cylinder Pressure Check


l High Pressure Leak Test
l Full Flow Test
l Whistle Warning Unit Test
l Medium Pressure Test - Pressure Reducer

Safety Warnings
l Explain - strictly follow instructions as outlined in the Instructions for Use, supplied with
the equipment.

l Maintain - respiratory protection equipment in line with Draeger Limited service schedules
and maintenance instructions.

Contact Draeger Limited, Blyth for details of Service Contracts and Service Training
Courses.

l Train - potential wearers in how to use, and wear equipment and to understand its
limitations.

Contact Draeger Limited, Blyth for details of Wearer Training Courses.

te 1412 - January 2004


A900 TEST KIT 5

Section 2 A900 Test Kit

te 1412 - January 2004


A900 TEST KIT 6

2.1 Component Parts of Equipment


The tests detailed in this manual require the following ancillary component parts.

Refer to Figure 1.

7 D 8 E 1

4
12

9
2 11 6 3 10

Fig. 1
0067

Item Description Order Code


No.

1 Manometer Assembly 3331288


2 Connector - Demand Valve 3335335
3 Connector - Exhalation Valve (Panorama Nova) 3331295
4 ‘Y’ Piece Tube Assembly 3332658
5 Male Blank (Connector 2) 3335421
6 Female Blank (Connector 3) 3332657
7 Aspirator 6801933
8 Clamp 3331287
9 Facepiece Former (Bladder) 3331290
10 Connector 3331298
11 Connector (Front or Side Port) 3335422
12 Female Blank (Connector 11) 3335420
13 Manometer Fluid (Not Illustrated) 3331969

te 1412 - January 2004


A900 TEST KIT 7

Section 3 Setting Up and Checking Test Equipment

te 1412 - January 2004


A900 TEST KIT 8

3 Setting Up and Checking Test Equipment


Refer to Figure 1.

7 D 8 E 1

4
12

9
2 11 6 3 10

Fig. 1
0067
1 Set up the manometer (1) on a clean flat surface allowing enough working space for
the test equipment and the breathing equipment.

2 Open the valves A, B and C of the manometer.

Note: The valves are in the ‘open’ position when the valve lever is pointing in the direction of
the valve outlet. Valves shown in Figure 1 are in the ‘Closed’ position. Do Not move or
transport manometer when valves are in ‘Open’ position.

3 Level the manometer by rotating the black knob at the base of the manometer until
the bubble in spirit level is between the two graduation marks. By rotating the knurled
collar at base of the valve B, it is possible to finely adjust the fluid column to the zero
point of the manometer scale. If necessary top up the manometer with manometer
fluid through valve B until the top of the fluid column is aligned with the zero point of
the manometer scale.

4 Panorama Nova P
Assemble connectors (2) and (3) to each of the long tubes on ‘Y’ piece (4). Connect
‘Y’ piece to end of tube D then assemble male blank (5) into connector (2) and
female blank (6) onto connector (3).

te 1412 - January 2004


A900 TEST KIT 9

Futura P
Assemble connectors (2) and (11) to each of the long tubes on ‘Y’ piece (4). Connect
‘Y’ piece to end of tube D then assemble male blank (5) into connector (2) and
female blank (12) onto connector (11).

5 Locate the free end of tube E to valve B.

6 Assemble the long tube of the aspirator (7) to valve C then Close valve C. Squeeze
the aspirator bulb expelling all air from inside the bulb. On releasing the aspirator
bulb the bulb should remain collapsed.

7 Carefully open valve C allowing the fluid to move slowly down the manometer scale.
The aspirator bulb will begin to slowly expand. Quickly close valve C when the fluid
level is between 60mm to 70mm/wg. Remove the aspirator from valve C.

Check fluid level in manometer tube for 30 seconds.

Note: Fluid level must remain static. If fluid level rises up manometer tube this indicates a
leak. Check seals and hose connections and repeat leak test. If unable to eliminate the
leak, close the valves and return the manometer unit to Draeger Limited - Service
Department.

8 Following satisfactory leak test - Open valve C.

te 1412 - January 2004


A900 TEST KIT 10

te 1412 - January 2004


A900 TEST KIT 11

Section 4 Test Procedures

te 1412 - January 2004


A900 TEST KIT 12

Section Contents
4 Test Procedures

4.1 Balanced Piston Leak Test,


Static Positive Pressure Balance Test
Whistle Test

4.2 Failed Static Pressure Test

4.2.1 No Audible Leak

4.2.2 Audible Leak

4.3 Facepiece Testing

4.3.1 Negative Pressure Leak Test

te 1412 - January 2004


A900 TEST KIT 13

4 Test Procedures

The following instructions cover the Functional Test procedures for Positive Pressure
(Push-in) Lung Demand Valves fitted with either:

l medium pressure hose with Quick Release Coupling


l medium pressure hose connected directly to pressure reducer.

Refer to Test Procedure Section of the appropriate Service/Training Manual for


additional tests.

4.1 Balanced Piston Leak Test


Static Positive Pressure Balance Test, Whistle Test

Specification
Allowable Leak 0 to 40mm/wg in less than 5 seconds
Static Balance 5 to 45mm/wg. (No audible leak)
Whistle Activation 55Bar +/- 5Bar

1 Set up and check test equipment as detailed in Section 3.

2 Open valves A, B and C.

3 Panorama Nova P
Remove blank (5) from connector (2) and female blank (6) from connector (3).

Futura P
Remove blank (5) from connector (2) and female blank (12) from connector (11).

4 Locate the free end of tube E to valve A.

5 Demand Valve with Quick Release Coupling


Connect the male coupling of the lung demand valve to the quick release coupling
of the medium pressure hose of the equipment.

6 Panorama Nova P
Remove inner mask assembly from the facepiece.

Futura P
Inflate the bladder (9), fit the facepiece and then leak test to the bladder as described
in Section 4.3.1.

7 Panorama Nova P
Moisten the exhalation valve then insert connector (3) into bore of exhalation valve
from inside the facepiece.

Futura P
Moisten the exhalation valve then insert connector (11) into the front port of the facepiece.

8 Press the reset feature of the lung demand valve to switch ‘off’ positive pressure
then open the cylinder valve slowly, but fully, to pressurise pneumatic system.

te 1412 - January 2004


A900 TEST KIT 14

9 Insert outlet of the lung demand valve into the female connector (2).

Balanced Piston Leak Test

10 Close valve C and ‘start’ stop watch.

Leak Test: Check fluid level in manometer tube. The fluid must not move from the
zero line to the 40mm/wg line in less than 5 seconds.

Note: If the demand valve Fails the leak test, check for possible cause and if necessary
replace the seals of the demand valve, reassemble demand valve and re-test. Should
the demand valve again Fail the leak test, replace balanced piston unit. Re-test.

A detected leak, but still within specification limits, will however affect the Static Positive
Pressure Balance Test. This is acceptable. During the Static Positive Pressure Balance
Test it is therefore important to take the initial reading.

11 If leak test satisfactory continue as follows:

12 Close valve A then press the centre of the rubber cover of the demand valve to
switch to positive pressure.

Static Positive Pressure Balance Test

13 Open valve A.

Static Test - Check fluid level in the manometer tube. The fluid should come to rest at any
point between 5mm and 45mm/wg lines. See Note. There should be no audible
leaks.

Note: If the demand valve Fails the Static Pressure Test, open valve C then press the reset of
the demand valve to switch ‘off’ positive pressure. Refer to Section 4:2.

14 If the static pressure test is satisfactory then continue as follows.

Whistle Activation Test (55bar +/- 5bar)

15 Close the cylinder valve. Carefully open valve C to slowly vent air pressure from the
system. Observe the pressure gauge of the equipment and as the pressure decreases
note the pressure at which the whistle begins to sound.

Note: Should the whistle not sound at the required pressure, reset the whistle as detailed
in the relevant section of the Training/Service Manual for the equipment on test.

te 1412 - January 2004


A900 TEST KIT 15

4.2 Failed Static Pressure Test


4.2.1 No Audible Leak
If there is no audible leak, but the demand valve Fails the Static Pressure Test, (high or
low static) then replace the positive pressure spring(s) of the demand valve. Reassemble
the demand valve and re-test from Section 4.1. Should the demand valve again Fail
Static Pressure Test, replace balanced piston unit. Re-test from Section 4.1.

4.2.2 Audible Leak


An audible leak indicates either a fault with the demand valve and/or facepiece
exhalation valve. To identify the fault, test the demand valve and facepiece separately
as follows:

Testing Demand Valve

1 Panorama Nova P
Remove connector (3) from the front port of the facepiece and assemble female
blank (6) to connector.

Futura P
Remove connector (11) from the front port of the facepiece and assemble female
blank (12) to connector.

2 Close valve A then open valve C.

3 Press the centre of the rubber cover of the demand valve to switch to positive pressure.

4 Close valve C then Open valve A.

Note: If audible leak occurs;

a. Close valve A.
b. Check security of the diaphragm, re-test.
c. Check and replace seals of the demand valve, reassemble demand valve then Re-
test from section 4.1.

If no audible leak proceed as follows:

Testing Facepiece

1 Open valve C.

2 Remove female connector (2) from the demand valve then insert male blank (5) into
the connector.

3 Panorama Nova P
Remove female blank (6) from connector (3).

Futura P
Remove female blank (12) from connector (11).

te 1412 - January 2004


A900 TEST KIT 16

4 Panorama Nova P
Moisten the exhalation valve and insert connector (3) into bore of the exhalation
valve from inside the facepiece.

Futura P
Moisten the exhalation valve and insert connector (11) into the front port of the
facepiece.

5 Disconnect tube E from valve A and locate free end of the tube E onto valve B.

6 Assemble long tube, of aspirator (7) to valve C.

7 Close valve C. Squeeze aspirator bulb expelling all air from inside the bulb. On
releasing aspirator bulb the bulb should remain collapsed.

8 Carefully open valve C allowing the fluid to move slowly down the manometer scale.
The aspirator bulb will slowly expand. Quickly close valve C when the fluid level in
the manometer tube is between 60mm/wg to 70mm/wg. Remove the aspirator
from valve C.

Note: Manometer fluid level must not rise from 60mm/wg line to 30mm/wg line in less than
one minute.

If the leak is more than specification, remove and visually inspect the silicone disc and
valve seat in the exhalation valve assembly. Clean the valve seat and clean or replace
the silicone disc assembly as necessary. Re-test.

If the exhalation valve passes the inward leak test, (remembering that Static Positive
Pressure Test ‘Failed’) then the exhalation valve spring assembly is not within
specification. Replace the exhalation valve spring assembly (Side Port), or spring (Front
Port), and re-test from Section 4.1.

te 1412 - January 2004


A900 TEST KIT 17

4.3 Facepiece Testing


4.3.1 Negative Pressure Leak Test
The following negative pressure leak test must be carried out after:

l Replacing speech amplifier diaphragm.


l Disassembly - Assembly of front port assembly.
l Replacing visor in Facepiece.
l Replacing Facepiece rubber.

1 Set up and check the test equipment as detailed in Section 3.

2 Open valves A, B and C.

3 Open the clamp (8) and position the clamp over the protective sleeve of the inlet
tube of the bladder (9).

4 Insert connector (10) into the short tube of the aspirator (7) and assemble connector
(10) into inlet tube of bladder. Inflate the bladder and seal off by screwing the clamp
onto the inlet tube of the bladder then detach the aspirator from the bladder.

5 Panorama Nova P
Fit female blank (6) into connector (3). Remove connector (2) from the ‘Y’ piece
tube and replace with connector (11). Insert connector (11) into female blank (12).

Futura P
Fit female blank (12) into connector (11).

6 Leak test equipment as detailed in Section 3, Instructions 1 to 10.

Note: Leak test is necessary to check integrity of the connector (12).

7 Remove blank (12) from connector (11).

8 Assemble the free end of tube E to valve B.

9 Assemble the long tube of the aspirator (7) to valve C.

10 Fit the facepiece over the bladder. Check that the facepiece is a good fit and that the
seal and headstrap is tight and secure.

11 Insert connector (11) into:


l Front Port Mask - front port of facepiece.
l Side Port Mask - side port of facepiece.

12 Gently squeeze the aspirator bulb until the fluid level in the manometer tube is
between 60mm/wg to 70mm/wg then close valve C.

Note: Manometer fluid level must not rise from 60mm/wg line to 30mm/wg line in less than
one minute.

te 1412 - January 2004


A900 TEST KIT 18

If the leak is more than specification, check all connections, inflate the bladder a little
more, re-adjust head straps and re-test.

13 Following a satisfactory leak test, Open valve C and remove connector (11) from the
facepiece.

14 Panorama Nova P
Reassemble inner mask to facepiece.

15 Test completed.

te 1412 - January 2004


A900 TEST KIT 19

Section 5 Fault - Cause - Remedy

te 1412 - January 2004


A900 TEST KIT 20

5.1 Demand Valve


Note: The following table is to be used only as a guide. Reference should be made to the
relevant lung demand valve manual.

FAULT CAUSE REMEDY

Balanced Piston Leak Test ; 1 O-Rings leaking Replace O-Rings


Pressure rise from zero to
40mm/wg in less than 5 seconds 2 Balanced Piston Unit Replace Balanced
Leaking Piston Unit

High Static Pressure>45mm/wg< Spring(s) too strong Replace Spring(s)

Low Static Pressure>5mm/wg< Spring(s) too weak Replace Spring(s)

Demand Valve fails to switch to 1 Faulty switch 'Off' Refer to LDV Manual
positive pressure mode

No Supplementary Supply Refer to LDV Manual

5.2 Facepiece
Note: For further Facepiece Test Procedures refer to the relevant Facepiece Service/Training
Manual.

FAULT CAUSE REMEDY


Inward Leak 1 Exhalation valve leaking Clean and/or replace silicone
disc Inspect exhalation valve
seat in front por t for damage.
Clean and/or replace

2 Loose connections Tighten connections


3 Face seals leaking Inspect for damage
4 Facepiece rubber body Replace facepiece rubber
per forated (pin holes, cracking) body

Will not hold/ Weak exhalation valve spring Replace exhalation valve
maintain positive spring assembly
pressure
Unable to maintain 1 Strap ratchets worn Replace headstrap
headstrap 2 Buckle rollers incorrectly fitted Assemble correctly
adjustment

te 1412 - January 2004


TESTOR 1

Index

Section 1 Introduction and Safety Warnings

Section 2 Test Procedures

Section 3 Test Hoses

Section 4 Fault Location

te 16142 - May 2005


TESTOR 2

te 16142 - May 2005


TESTOR 3

1 Introduction and Safety Warnings

te 16142 - May 2005


TESTOR 4

1.1 Introduction
The Dräger Testor test equipment provides an efficient means of testing the Functional
features of the Dräger range of compressed air respiratory protection equipment.

For details of the following reference should be made to the Instructions for Use
supplied with the Testor Unit:

l Cleaning and Storage.


l Technical Data.
l Maintenance Intervals.
l Trouble Shooting.

It is recommended that the Testor Instructions for Use and this manual be read and
fully understood before proceeding with any Functional Testing of compressed air
respiratory protection equipment.

1.2 For Your Safety


Approved Dräger compressed air respiratory protection equipment incorporate a
Dräger pre-set and sealed pressure reducer. A range of Lung Demand Valves,
Facepieces and associated ancillary units are approved for use with the equipment.

The Dräger guarantee is void should original seal caps of the pressure reducer, or
any sealed component assemblies of the ancillary units be tampered with removed
or broken.
l Use of compressed air respiratory protection equipment requires wearer and service
training, knowledge and observance of the instructions contained in the Training/
Service Manual and the Instructions for Use supplied with equipment.
l Use of equipment requires compliance with National Regulations, Laws and
Standards governing the use of respiratory equipment in the country of use.
l Use equipment only for the purpose specified in the Instructions for Use supplied
with equipment, or as confirmed in writing by Dräger.
l Trained and competent personnel should inspect and service equipment at regular
intervals and a record kept of such inspections and servicing.
l Dräger recommends a service contract be obtained from your Dräger Branch or
Agent.
l Contact Dräger for details of Service Contracts and Service Training Courses.
l Notify Dräger in the event of component fault or failure.
l Use only original Dräger Spare Parts for service and maintenance.

te 16142 - May 2005


TESTOR 5

2 Test Procedures

te 16142 - May 2005


TESTOR 6

Section Description
2 Test Procedures

2.1 Leak Testing - Testor

2.2 Testing - Breathing Equipment

2.3 Testing Equipment – Fault Finding

te 16142 - May 2005


TESTOR 7

2 Test procedures
Standard test procedures for compressed air respiratory protection equipment are
as follows:

l Cylinder Pressure Check


l High Pressure Leak Test
l Whistle Warning Unit Test e.t.c.

Details of these tests are covered in the Functional Test Procedure Section of the
relevant Service/Training Manual for the equipment.

Having completed the Functional Test Procedures as defined in the relevant Service/
Training Manual, and if the integrity of the lung demand valve, facepiece or set has
been broken due to repair and/or replacement of components, the following additional
tests should be carried out using the Testor Test Equipment.

l Full Flow Test


l Medium Pressure Test
l Facepiece Leak Test
l Facepiece Exhalation Valve Lift Off Test
l Balanced Piston Leak Test
l Positive Pressure Switch Over
l Static Pressure Test

Refer to Section 2.2 for detailed instructions for the Functional Tests.

These functional tests should also be carried out in line with the Dräger
recommendations as outlined in the Maintenance and Test Interval Chart of the
Instructions for Use and the Test Procedure Section of the appropriate Training/
Service Manual for the equipment to be tested.

It is recommended that this manual be read and fully understood before proceeding
with any Functional Testing of compressed air respiratory protection equipment.

te 16142 - May 2005


TESTOR 8

2.1 Leak Testing Testor


Before starting any test or series of tests of compressed air respiratory protection
equipment it is recommended that the Testor Unit be leak tested as described in the
following instructions.

Tools Required
Adaptor (Rd40 to A) R50028
Blanking Plug 3335421

Safety Note: Connect only breathing quality air to the Testor Unit from a supply pressure
source of maximum pressure 10bar.

1 Disconnect the lung demand valve


from the female coupling of the
medium pressure hose of the
equipment pressure reducer and
connect the separate lung demand
valve to the female quick release
connector located at the left side of
the Testor. Refer to Fig. 1

Connect the female coupling of the


medium pressure hose of the
equipment to the male connector
located at the right hand side of the
Testor Unit. Refer to Fig. 2 Fig. 1
1272
Note: If the equipment to be tested
incorporates a lung demand valve that
is connected directly to the pressure
reducer then it is necessary to
assemble the following test hoses as
required. Refer to Section 3.

lPA90 Pressure Reducer 3339620


lPlus Pressure Reducer 3337650

Important Note: If using a test hose (lung


demand valve connected) activate
the lung demand valve to switch
‘Off’ the positive pressure.

2 Set the timer to one minute. Fully Fig. 2


1271

‘open’ the cylinder valve (or independent air supply source). Allow the needle of the
medium pressure gauge of the Testor to stabilise.

3 ‘Close’ the cylinder valve then ‘start’ the timer. Observe the high pressure gauge of
the equipment – the gauge reading should not drop more than 10bar in one minute.
During the timing of the pressure leak test – check for audible leak.

te 16142 - May 2005


TESTOR 9

Refer to Fig 3

4 Reset the timer to one minute. Fully ‘open’ the cylinder valve. Carefully push the
centre lever forward towards the ‘+’ sign and very carefully inflate the rubber head to
approximately average head size.

Safety Note: Do Not over inflate test head.

5 ‘Start’ the timer. At the end of the timing sequence the test head should show no
signs of reduction in size (deflation).

6 Reset the timer to 30 seconds. With the test head still inflated, moisten the end of
the finger or thumb with water and place and hold over the hole in the mouth fitting
of the test head. Carefully push the RH lever forward towards the ‘+’ sign until the
mbar gauge indicates +20mbar. If an over pressure reading occurs when pushing
lever forward then pull carefully on the LH lever away from the ‘-’ sign until the
pressure reading reduces to +20mbar.

‘Start’ the timer. The gauge reading should remain at +20mbar for 30 seconds.
After 30 seconds remove thumb from mouth fitting of test head – the needle of the
mbar gauge should return to ‘0’. Press the timer ‘Stop’ button.

7 Pull and hold the centre lever towards the ‘-’ sign to deflate the rubber test head.

mbar Medium Pressure High Pressure


Gauge Gauge Timer Air Source

Mouthpiece LH Centre RH Male Connector


Fig. 3
1269

te 16142 - May 2005


TESTOR 10

8 Pull and hold the RH lever back away from ‘+’ sign and insert male blanking plug
into port at rear of test head. Release the RH lever.

Note: Do Not insert the blanking plug into port unless RH lever is held back - otherwise a
positive overpressure will be shown on the mbar gauge.

9 Carefully push LH lever forward towards the ‘-’ sign until the needle of the mbar
gauge rotates anti-clockwise and indicates a negative pressure of -10mbar. Release
the LH lever then ‘Start’ the timer – the gauge reading should remain at -10mbar for
30 seconds.

Note: If an under pressure occurs when pushing the lever forward then carefully pull the lever
away from the ‘-’ sign until -10mbar is achieved, then start leak test as above.

10 Pull and hold the RH lever away from the ‘+’ sign and remove the blanking plug from
the port at rear of the test head. Release the RH lever. The needle of the mbar gauge
should return to ‘0’.

11 ‘Close’ the cylinder valve then press and hold the LH lever towards the ‘-‘ sign to
vent pressure from system. When vented activate the lung demand valve to switch
off positive pressure. Leak testing of the Testor now completed.

12 Proceed to test the compressed air respiratory protection equipment


– see Section 2.2.

mbar Medium Pressure High Pressure


Gauge Gauge Timer Air Source

Mouthpiece LH Centre RH Male Connector


Fig. 3
1269

te 16142 - May 2005


TESTOR 11

2.2 Testing the Breathing Apparatus


This procedure covers the functional testing requirement of the complete assembly
of the compressed air respiratory protection equipment, i.e. pressure reducer, lung
demand valve and facepiece.

2.2.1 High Pressure Leak and Whistle Warning Unit Test


Important Note:
The following is a general procedural sequence – reference must be made to
the appropriate Pre-Operational Checks section of the Instructions for Use or
the Test Procedure section of the Training/Service Manual for the equipment.

Note: If equipment has a quick release coupling on medium pressure hose then attach the
lung demand valve to the coupling.

1 A and AE - Activate the lung demand valve to switch ‘Off’ positive pressure.

2 Open cylinder valve slowly but fully to pressurise the pneumatic system. Check the
pressure reading of the high pressure gauge. Close the cylinder valve and observe
high pressure gauge.

Cylinder content should be at least 80% of maximum charge, i.e. 160bar for 200
units or 240bar for 300 units.

3 High Pressure Leak Test - The high pressure gauge reading should not drop more
than 10bar in one minute. During the timing of the high pressure leak test check for
audible leak(s).

If the equipment fails to conform to test parameters, vent the pressure from the
system, investigate source of the leak, rectify and repeat test. Until the leak is rectified,
Do Not carry out any further test. If test is satisfactory proceed to Instruction 4.

4 Whistle Warning Unit Test - While slowly venting pressure from the system observe
the high pressure gauge. As the pressure decreases check that the whistle activates
an audible alarm signal at between 60bar to 50bar.

Vent pressure from system as follows:

A and AE - Cover the outlet port of the lung demand valve with palm of the hand then
press the centre of the rubber cover of the demand valve to switch ‘On’ positive
pressure. Slowly vent pressure by carefully lifting palm of hand to maintain a slow
steady decrease in pressure.

N - Slowly vent equipment by carefully pressing centre of the rubber cover of the
demand valve maintaining a slow steady decrease in pressure.

Note: Before proceeding to test the equipment with the Testor ensure that the Testor Unit is
satisfactory and has ‘Passed’ the leak test as described in Section 2.1.

te 16142 - May 2005


TESTOR 12

Tools Required
Adaptor (Rd40 to A) R50028
Blanking Plug 3335421

Important Note: If the equipment to be tested incorporates a lung demand valve that
is connected directly to the pressure reducer then remove the demand valve
from the reducer and assemble the following test hose as required.
Refer to Section 3.

5 Connect the pressure reducer to a fully charged cylinder assembly (or an independent
high pressure air source) then connect the female coupling of the medium pressure
hose of the equipment to the male connector located at the right hand side of the
Testor.

6 Remove the lung demand valve from the inhalation port of the facepiece. Activate the
lung demand valve to switch ‘Off’ positive pressure. Connect the male coupling of
the hose of separate lung demand valve units to the female quick release connector
located at the left side of the Testor.

7 ‘Open’ the valve of the high-pressure air source.

mbar Medium Pressure High Pressure


Gauge Gauge Timer Air Source

Mouthpiece LH Centre RH Male Connector


Fig. 3
1269

te 16142 - May 2005


TESTOR 13

8 Full Flow Test - Press the centre of the rubber cover of the lung demand valve. A and
AE types will switch ‘On’ to positive pressure. Check that an unobstructed airflow
vents from the lung demand valve. See Safety Note.

Safety Note: During the full flow test Do Not direct the outlet of the lung demand valve
onto the face or skin.

This test should take a minimum of 3 – 5 seconds. Following the test re-activate the
lung demand valve (A and AE) to switch ‘Off’ positive pressure then immediately
observe the medium pressure gauge.

Medium Pressure Test - At an inlet pressure of >100bar, the medium pressure


gauge should indicate a pressure of between 6bar to 9bar after 5 seconds. After
one minute it is acceptable for the medium pressure to rise to 9.7bar.

During the timing of the medium pressure test - check for audible leak(s). If a fault is
highlighted refer to Section 4.

mbar Medium Pressure High Pressure


Gauge Gauge Timer Air Source

Mouthpiece LH Centre RH Male Connector


Fig. 3
1269

te 16142 - May 2005


TESTOR 14

10 Remove the inner (orinasal) mask.


Fully extend the straps of the head
harness of the facepiece. Moisten
the inner face sealing lip and
exhalation valve of the facepiece with
clean water. Place facepiece over
rubber test head – pulling the
harness down the back of the head
and positioning the head harness
centrally with back of the head.
While holding down the harness
towards the back of the head, tighten
each of the lower straps evenly
towards back of head and repeat
with the upper side straps. See Fig.
4. If necessary adjust the centre Fig. 4
strap as required. 1273

11 Carefully push the centre lever forward towards the ‘+’ sign to very carefully inflate
the rubber head to achieve complete contact around the face seal of the facepiece.

Safety Note: Do Not over inflate test head.

12 Insert the blanking plug into the inlet (inhalation) port of the facepiece. Ensure a
secure assembly.

13 Facepiece Negative Pressure Leak Test - Set the timer to one minute. Carefully
push the LH lever forward towards the ‘-’ sign until the needle of the mbar gauge
rotates anti-clockwise and indicates a negative pressure of (minus) -6mbar. Release
the LH lever.

Immediately ‘Start’ the timer. The needle of the mbar gauge should not move from
(minus) -6mbar to -3mbar in less than one minute. If a fault is highlighted refer to
Section 2.3.1.1.

14 Exhalation Valve Lift Off Test - Push and hold the RH lever forward towards the ‘+’
sign. The needle of the mbar gauge rotates clockwise and will stop when the exhalation
valve, in the front port of the facepiece, lifts from its sealing rim. Keep the RH lever
pushed forward - the needle of the gauge should remain stable between the following
values,

Front Port Facepiece - +4.5mbar and +5.7mbar.


Side Port Facepiece - +5.8mbar and +7.0mbar

Release the RH lever. If a fault is highlighted then refer to Section 2.3.1.2.

15 Remove the blanking plug from the inhalation port of the facepiece then connect the
lung demand valve to the inhalation port ensuring a secure connection.

te 16142 - May 2005


TESTOR 15

16 Balanced Piston Leak Test – Pull the RH lever away from the ‘+’ sign to zero (‘0’)
the mbar gauge. Release the RH lever. The needle of the mbar gauge should not
move from ‘0’ to +4mbar in less than 5 seconds.

Note: If a fault is highlighted this indicates a faulty Balanced Piston Unit – stop the test –
service lung demand valve then re-test.

17 Static Pressure Test - Push and hold the LH lever forward towards the ‘-’ sign.
The lung demand valve should activate to positive pressure mode (audible activation).
Release the LH lever. The mbar gauge should indicate a positive pressure reading
of between +1mbar to +3.9mbar. No audible leak.

Note: If a fault is highlighted activate the lung demand valve to switch ‘Off’ positive pressure
– remove the lung demand valve from the facepiece. Proceed to Section 2.3.2.

18 If test satisfactory - ‘Close’ the valve of the high pressure air source.

Whistle Warning Unit Test - Carefully push the LH lever towards the ‘-’ sign to
slowly vent pressure from the system. During slow venting observe the high pressure
gauge of the equipment and check that the whistle activates an audible signal at
between 60bar to 50bar.

Note: As soon as the whistle activates an audible signal – release the LH lever to allow the
whistle to vent the air.

mbar Medium Pressure High Pressure


Gauge Gauge Timer Air Source

Mouthpiece LH Centre RH Male Connector


Fig. 3
1269

te 16142 - May 2005


TESTOR 16

19 Remove the lung demand valve from the facepiece. Pull and hold the centre lever
towards the ‘-’ sign to deflate the test head to average size and until the mbar and
medium pressure gauges indicates ‘0’.

20 Release and extend the side straps of the head harness of the facepiece and remove
the facepiece from the rubber test head. Refit the inner (orinasal) mask.

21 If the equipment ‘Passes’ the tests - disconnect female coupling of the medium
pressure hose from the coupling of the Testor. If the reducer is fitted to an independent
air source grip firmly the outer diameter of the handwheel of the pressure reducer
and unscrew (anti-clockwise) from the high pressure air source.

22 Disconnect the hose coupling of the separate lung demand valve hose from the left
hand female coupling of the Testor and connect to the female coupling of the medium
pressure hose of the pressure reducer. Activate the lung demand valve to switch
‘Off’ positive pressure then refit the lung demand valve to the facepiece ensuring a
secure connection.

Note: If the equipment tested has a continuous hose from the lung demand valve to the
pressure reducer, remove the test hoses as required. Refer to Section 3. Carry out a
high pressure leak test of the equipment.

23 Following satisfactory Tests, complete the test details in the Log Book for the
compressed air respiratory equipment. Store the equipment in designated area,
‘Ready for Use’.

te 16142 - May 2005


TESTOR 17

2.3 Testing Equipment – Fault Finding


The following test sequence should be carried out following any ‘Failure’ identified
during Functional Tests as described in Section 2.2. Reference should be also
made to the Fault Location Section 4 and the appropriate Training/Service Manual.

2.3.1 Facepiece
This fault analysis is for guidance – reference must be made to the Fault Location
Section 4 and the appropriate Training/Service Manual for the facepiece.

2.3.1.1 Facepiece Negative Pressure Leak Test


This fault analysis is for guidance – reference must be made to the Fault Location
Section 4 and the appropriate Training/Service Manual for the facepiece.

l Check integrity of fit of facepiece to rubber test head.


Re-fit facepiece and re-test.

l Damage O ring on blanking plug. Inspect.


Fit new O ring and re-test.

l Check exhalation valve rubber disc is clean and not damaged. Inspect.
Wash and clean - Re-fit and re-test.

l Check sealing rim for exhalation valve rubber disc is clean and not damaged. Inspect.
Wash and clean - re-test. If damaged – replace port assembly – re-test.

l Speech diaphragm locking ring not tight. Inspect.


Tighten/replace as necessary and re-test.

l Speech diaphragm and/or sealing ring damaged. Inspect.


Replace as necessary and re-test.

2.3.1.2 Exhalation Valve Lift Off Test


l High or Low ‘Lift Off’ value.
High value - Strong spring.
Low value - Weak spring.
Inspect and fit new spring.

If an initial high reading is indicated, but then settles to within the test parameter, this
indicates a ‘sticky’ exhalation valve disc and/or sealing rim. Remove – clean – rinse
in clean water – re-fit and re-test.

te 16142 - May 2005


TESTOR 18

2.3.2 Lung Demand Valve


1 Pull the centre lever of the Testor towards the ‘-’ sign to reset the internal shuttle
valve of the Testor Unit then release the lever. Pull and hold the RH lever away from
‘+’ sign then insert the lung demand valve into the adaptor port at the rear of the test
head. Release the RH lever. The needle of the mbar gauge should return to ‘0’.

2 Leak Test Lung Demand Valve – Carefully push the RH lever forward towards ‘+’
sign. The needle of the mbar gauge rotates clockwise. Set the needle of the gauge
to +7mbar. Release the RH lever.

Note: If an over pressure occurs when pushing the RH lever forward then carefully pull the
LH lever away from the ‘-’ sign until +7mbar is achieved, then start leak test as follows.

Needle of the mbar gauge should not move from +7mbar to +6mbar in less than 15
seconds. If a failure occurs refer to Section 4. Following the test pull the RH lever
back away from ‘+’ sign until the needle of the mbar gauge indicates ‘0’. Release the
RH lever.

4 Static Pressure Test - Push and hold the LH lever forward towards the ‘-’ sign. The lung
demand valve should activate to positive pressure mode (audible activation). Release the
LH lever. The mbar gauge should indicate a positive pressure reading of between +1mbar
to +3.9mbar. If a failure occurs refer to Section 4. Vent pressure from system.

Important Note: If any failures occur during these tests – service the equipment as
necessary. Refer to the appropriate Training/Service Manual. Retest equipment
with the facepiece.

mbar Medium Pressure High Pressure


Gauge Gauge Timer Air Source

Mouthpiece LH Centre RH Male Connector


Fig. 3
1269

te 16142 - May 2005


TESTOR 19

3 Test Hoses

te 16142 - May 2005


TESTOR 20

Section Description
3 Test Hoses

3.1 PA90 Series Pressure Reducer

3.2 PA90Plus Series Pressure Reducer

te 16142 - May 2005


TESTOR 21

3 Test Hoses
If the equipment to be tested has a continuous hose from the pressure reducer to
the lung demand valve then to perform full functional testing using the Testor the
hose must be disassemble from the reducer and replaced with the following test
hoses as required.

l PA90 Pressure Reducer 3339620


l Plus Pressure Reducer 3337650

3.1 PA90 Series Pressure Reducer


Tools Required:
5mm AF Hexagon Socket Key 3335738
Medium Pressure Test Hose 3339620

1 Ensure the cylinder valve is ‘Closed’ and the pressure vented from the system.

2 Using the socket key unscrew and remove the blanking plug from the secondary
medium pressure port in the end face of the pressure reducer.

3 Inspect test hose and ensure that the O ring is fitted to the end fitting and is not
damaged. Screw the end fitting of the hose into the secondary medium pressure
port in the end face of the pressure reducer – hand tight.

4 Proceed to either Section 2.1 – Leak Testing Testor, or Section 2.2 – Testing the
Breathing Apparatus.

5 Following satisfactory testing of the equipment, remove the test hose, inspect O
ring of the blanking plug and re-fit plug to secondary port. Tighten using socket key.
Carry out a high pressure leak test of the equipment.

3.2 Plus Series Pressure Reducer


Tools Required
2mm AF Socket Key 3337900
Medium Pressure Test Hose 3337650

1 ‘Close’ the cylinder valve of the equipment and ensure the system is vented. Unscrew
reducer handwheel from cylinder valve. Loosen cylinder strap, slide cylinder away
from reducer and remove cylinder from equipment.

te 16142 - May 2005


TESTOR 22

Refer to Figure 1.

2 Using 2mm socket key unscrew


(anti-clockwise) and remove two
screws (1) from reducer body,
remove retaining cap (2) and
retention staple (3). Grip and pull
the lung demand valve hose (4) to
remove the hose from the outlet port
of the reducer.
4
Note: Component parts of relief valve, i.e.
spring, O ring retainer and O ring,
remain attached to nozzle of the lung
demand valve hose. Do Not remove
components. 1 2 3 Fig. 1
Refer to Figure 2. 0228

3 Using 2mm socket key, unscrew


1
(anti-clockwise) and remove two
screws (1) from the manifold block
of the test hose, remove retaining 2
cap (2) and retention staple (3).
Insert the end fitting of the lung 3
demand valve hose into manifold
port - push and hold against the
compression spring - insert retention
staple (3) through holes in manifold.
Fully insert retention staple ensuring
correct location into groove in hose
end fitting.
Fig. 2
Note: Check hose securely retained by
pulling hose away from manifold. 0306

4 Locate groove of retaining cap over


protruding portion of retention staple
- align holes - insert screws and
using 2mm socket key, tighten
screws.

Refer to Fig. 3

5 Ensuring O ring is fitted then insert


the connector of the manifold into
the outlet port of the reducer. Insert
retention staple into holes in reducer
body. Fully insert retention staple
ensuring correct location into groove
of manifold connector. Fig. 3
1270

te 16142 - May 2005


TESTOR 23

Note: Check manifold securely retained by pulling manifold away from reducer body.

6 Locate groove of the retaining cap over the protruding portion of retention staple -
align screw holes - insert screws and using 2mm socket key, tighten screws.

7 Proceed to either Section 2.1 – Leak Testing Testor, or Section 2.2 – Testing the
Breathing Apparatus.

8 Following satisfactory testing of the equipment, remove the test hose from the reducer.
Remove the lung demand valve from the manifold of the test hose and re-assemble
to the pressure reducer. Carry out a high pressure leak test of the equipment.

te 16142 - May 2005


TESTOR 24

te 16142 - May 2005


TESTOR 25

4 Fault Location

te 16142 - May 2005


TESTOR 26

Section Desription
4.1 Pressure Reducer

4.2 Lung Demand Valve

4.3 Facepiece

4.4 Whistle Warning Unit

te 16142 - May 2005


TESTOR 27

4.1 Pressure Reducer


Compressed air respiratory protection equipment manufactured by Dräger incorporate
a pre-set and sealed pressure reducer assembly. The Dräger warranty conditions
and guarantee is void should original seal caps be tampered with or removed. Correct
operational condition of the pressure reducer is guaranteed only if the unit is serviced
and re-sealed by Dräger.

Return faulty units (less hoses and any accessories) to Dräger Service.

Refer to the relevant Training/Service Manual, for full details of Service/Maintenance


recommendations with regard to the product in use.

Important Note: This table is for guidance only – reference must be made to the Fault
Location section of the appropriate Training/Service Manual.

Fault Cause Remedy


High Pressure 1 Damaged valve plunger/seat Overhaul Valve - Return to
Leak Valve Dräger Service
2 Gauge Seal Replace O-Ring
Leak from pressure 1 Damaged O-Ring Replace O-Ring
relief valve 2 Weak Spring Replace Spring
3 Damaged O-Ring Retainer Replace Retainer
Leak from pressure Reducer Failure Remove hose and replace
relief valve - reducer
components o.k. Return suspect reducer to
Dräger Ser vice
High or Low Reducer out of specification Remove hose and replace
Medium Pressure reducer
Return suspect reducer to
Dräger Ser vice
High Pressure Damaged HP O-Ring Replace O-Ring
Leak at connection Damaged Metal Components Refer to Dräger Service

te 16142 - May 2005


TESTOR 28

4.2 Lung Demand Valve


It is important to refer to the relevant Training/Service Manual, for full details of
Service/Maintenance recommendations with regard to the product in use.

Important Note: This table is for guidance only – reference must be made to the Fault
Location section of the appropriate Training/Service Manual.

FAULT CAUSE REMEDY

Inward/outward leak when 1 Damaged/faulty connection Replace O-ring


in facepiece por t O-ring

2 Damaged bore in front por t of Replace Front Por t


facepiece
Leak Test - Positive 1 Banjo Seals Leaking Replace Seals
pressure 'off'
2 Balanced Piston Unit Leaking Replace Balanced
Piston Unit

High Static pressure Balanced piston lever springs Replace Springs


too strong

Low Static pressure Balanced piston lever springs Replace Springs


too weak
Demand Valve fails to 1 Bent/damaged Balanced Replace Balanced
switch automatically to Piston Lever Piston Unit
positive pressure mode

2 Bent/damaged retaining spring Replace Balanced


plate Piston Unit

te 16142 - May 2005


TESTOR 29

4.3 Facepiece
It is important to refer to the relevant Training/Service Manual, for full details of
Service/Maintenance recommendations with regard to the product in use.

Important Note: This table is for guidance only – reference must be made to the Fault
Location section of the appropriate Training/Service Manual.

FAULT CAUSE REMEDY

Inward Leak 1 Exhalation valve leaking Clean and/or replace silicone disc.
Inspect exhalation valve seat in front
por t for damage.
Clean and/or replace.

2 Loose connections Tighten connections

3 Face seals leaking Inspect for damage

4 Facepiece rubber body Replace facepiece rubber body


per forated (pin holes,
cracking)

Will not Weak exhalation valve Replace exhalation valve spring


hold/maintain spring assembly.
positive pressure

Unable to maintain 1 Strap ratchets worn Replace headstrap


headstrap
adjustment
2 Buckle rollers Assemble correctly
incorrectly fitted

te 16142 - May 2005


TESTOR 30

4.4 Whistle Warning Unit


It is important to refer to the relevant Training/Service Manual, for full details of
Service/Maintenance recommendations with regard to the product in use.

Important Note: This table is for guidance only – reference must be made to the Fault
Location section of the appropriate Training/Service Manual.

FAULT CAUSE REMEDY


Whistle sounds at Incorrect whistle setting Reset whistle
wrong pressure
Whistle sounds 1 Damaged H.P. piston seat 1,2 or 3 - Remove
continuously 2 Piston seal leak (H.P.) hoses from reducer and
3 Plunger housing seal leak (M.P.) return reducer to
Dräger. (REX)
Poor sounding 1 Blocked whistle tube flute Clean or replace whistle
whistle and/or jet tube.
2 Blocked Filter Replace filter/filters.
3 Split or per forated rubber hose of Replace hose
dual pressure hose leaking
4 Faulty activation Remove hoses from
reducer and return
reducer to Dräger. (REX)
Whistle does not 1 Blocked whistle tube flute and/or jet Clean or replace whistle
sound Blocked Filter tube
2 Split or per forated rubber hose of Replace filter/filters.
3 dual pressure hose (leaking) Replace hose
4 Faulty Activation Remove hoses from
reducer and return
reducer to Dräger.
(REX)

te 16142 - May 2005


PSS 100 Series - PSS 100, PSS 100 D - A, AE, N
Compressed Air Respiratory Protection Equipment Instructions for Use
Approvals Preparation for use Charging Cylinders
PSS 100 Series compressed air respiratory protection equipment conforms to EN137 and meets the u Check secondary air supply outlet thread, the Recharge to the rated pressure indicated on the
requirements of EC Council Directive 89/686/EEC, 9723/EC, BSEN 133. ChargAirTM connector thread and sealing ring label or stamped on the neck or shoulder of cylinder.
are not damaged. Dräger recommend a charge rate of 27bar/minute.
PPE - EC Type examiner: SGS United Kingdom Ltd, Unit 202b Worle Parkway, Weston- u Screw secondary ChargAirTM supply hose Rapid charging will induce an increase in
super-Mare, BS22 0WA, UK. Notified Body No. 0120. CE Mark Approved and issued with EC connector to secondary air supply outlet - ensure temperature resulting in an incomplete charge -
Type Examination Certificate. secure connection. check pressure at ambient and if required ‘Top Up’
u Turn ‘On’ secondary air supply to pressurise charge.
PED - EC Type examiner: Lloyds Register, Hiramford, Middlemarch OfficeVillage, Siskin Drive, Coventry, secondary ChargAir TM hose. Remove
CV 3 4FJ, England. Notified Body No. 0038. The equipment is manufactured in accordance with the protection cap from female quick release It is recommended that a pressure limiting device
requirements of Conformity Assessment Modules B + D of the Pressure Equipment Regulations. A copy 1 12 coupling. is fitted to the charging compressor to prevent
of the Declaration of conformity is available from Dräger on request. 0501 0090 u ChargAirTM supply hose ready for use. over charging of the selected cylinder.

u Repeat the whistle check from the High Safety Warning: Air quality for compressed
For Your Safety Pressure Leak Test.
Note: If high pressure air leak detected either at
connections, or along hose - close secondary air air breathing systems should conform to
PSS 100 Series incorporates Dräger pre-set, and sealed pressure reducer. Drägers’ guarantee is void should u If necessary repeat instructions to achieve supply valve and vent pressure from hose. Replace requirements of EN12021. Only charge
original seal caps be tampered with, removed, or broken. Correct operational condition is only valid if Dräger required whistle activation setting. hose. compressed air pressure vessels which:
service, and re-seals pressure reducer. u Following satisfactory whistle test, ensure
u PSS 100 D (with ChargAirTM) can only be connected to 300bar valved cylinder assemblies. positive pressure mechanism activated (A and u Conform to National Standards.
u Use of equipment requires, knowledge, and compliance with National Regulations, Laws and Standards, Connecting Secondary Air Supply
AE) to switch ‘Off’ positive pressure. u Feature original manufacturers test date
governing the use of respiratory protection equipment in the country of use. and test mark.
u Use of equipment requires wearer training and observance of these Instructions for Use. Connecting Demand Valve to the Note: ChargAirTM connection allows refilling of
u Have not exceeded test date indicated on
u Use equipment only for the purpose specified in this manual, or as confirmed in writing by Draeger. Facepiece cylinder(s), from an independent high pressure
u Only trained competent personnel should inspect and service equipment at regular intervals and a record
2 (300bar) secondary breathing air supply source, label affixed by last testing station.
kept of such inspections and service. 0160 Note: Refer also to the Instructions for Use provided while wearer is breathing from equipment.
u Dräger recommends a service contract be obtained from your Dräger Branch or Agent. with the Lung Demand Valve. If the facepiece and the demand valve are not being Visual Inspection
worn, ensure positive pressure mechanism of lung Check integrity of:
u Contact Dräger for details of Service Contracts and Service Training Courses. Type A demand valve (A and AE) activated to switch ‘Off’ u Carrying system.
u Notify Dräger if there is component fault or failure. u Check facepiece port, and the demand valve O positive pressure. u All straps, buckles and harness.
u Use only original Dräger Spare Parts for service and maintenance. ring, are clean and undamaged. u Valves, connectors and manifold.
u Push demand valve into facepiece port until it u Remove protection cap from ChargAirTM male u Lung Demand Valve.
For Your Safety - ChargAirTM (If Fitted) ‘clicks’ into position. quick release coupling on equipment. u Facepiece.
u Use ChargAirTM only for the purpose specified in this manual, or as confirmed in writing by Draeger. u Check attachment is secure. u Ensure cylinder valve(s) are fully ‘Open’.
u Do Not use ChargAirTM to transfer air from one compressed air breathing equipment to another, or for
second man attachment. Type’s AE and N u Push male and female ChargAirTM couplings Cleaning, Disinfecting, Drying
u Check facepiece port, and sealing ring, are clean together until a loud ‘click’ is heard - indicating Carefully clean, disinfect, and thoroughly dry
u Do Not use any charging hose assemblies, or any other equipment, unless specifically supplied by 3 a secure connection. High pressure air from contaminated dirty components after use.
Dräger for use with ChargAirTM. Only use approved equipment. and undamaged.
0156
u Screw demand valve handwheel into facepiece secondary supply source begins to refill When using baths to contain cleaning and
u Do Not lubricate, or allow oil, grease or other contaminants to contact ChargAirTM fittings or hoses.
port. Handtight. equipment cylinder(s). disinfecting solutions, immersed components and
u Do Not attempt to disassemble or repair ChargAirTM fittings or hoses. Drägers’ guarantee is void should
u Align demand valve (Fig. 6) and tighten. assemblies must be agitated manually. Draeger
ChargAirTM be tampered with, removed, or broken. Correct operational condition is only valid if Dräger
Note: A secure connection has been made when a recommends that no form of mechanical, electrical
service pressure reducer and ChargAirTM assembly.
u Caution: secondary air supply pressure must not exceed maximum rated working pressure (300bar) of Operation ‘dark green’ ring is visible behind the ‘grey’ sleeve or ultrasonic agitation be used.
cylinder(s) being filled. Putting on equipment of the female coupling. See Fig. 12.
Safety Note: Refer to manufacturers’ usage
u Caution: if leak detected while refilling in contaminated, or oxygen deficient gaseous atmosphere, u If necessary set Wearer Height Adjustment.
u If facepiece and demand valve are being worn instructions when using cleaning and
discontinue refilling and immediately leave hazardous area. u Extend waistbelt and shoulder adjusting straps.
- breathe normally during refill. disinfecting agents. It is important that attention
u Caution: it is the users’ responsibility to ensure that a. the secondary ChargAirTM hose connector is Put on equipment - taking weight on shoulders
u Check reading on equipment pressure gauge, be paid to concentration and reaction times.
compatible with outlet connector of independent high pressure (300bar) secondary breathing air supply by pulling on shoulder adjusting straps.
and if fitted, gauge on secondary supply hose, or Do Not use organic solvents, such as Acetone,
source, and b. the connection is correctly, and securely assembled. 4 Do Not tighten.
secondary supply gauge. When both gauges Alcohol, White Spirit, Trichloroethylene or similar.
0157
u Using right and left hand, grip each side of
Extreme Service Conditions waistbelt pad assembly lifting assembly to top indicate the same pressure - secondary supply
Dräger recommends the following:
of vertical slot of backplate. See Fig. 7. pressure and equipment cylinder(s) pressure
In the presence of chemicals; Some substances, e.g. Phosgene, may be absorbed by, or diffuse through,
Fasten waistbelt buckle. Pull ends of waistbelt have equalised.
the material of the silicone diaphragm of the PA90 Series lung demand valve.
adjusting straps forward (Fig. 8) until waistbelt
1. Cleaning
pad secure and comfortable on hip. Tuck strap Note: Dependent on the pressure remaining in u Dräger Safety Wash.
Liability Statement ends between body and waistbelt pad. the equipment cylinder before refill, pressure u Sekusept
Responsibility for reliable function of equipment transfers to the owner or operator when serviced, or u Pull down shoulder adjusting straps until equalisation occurs after approximately 45-60 Do Not exceed a temperature of 30 degrees
repaired by untrained personnel, (not employed or authorised by Dräger) or when used in a manner not equipment is secure and comfortable on hip. Do seconds. Final equalised pressure governed Celsius.
1
conforming to its intended use. Not overtighten. Tuck the strap ends under the by available pressure and volume of air in
waistbelt. secondary air supply. Due to rapid compression Note: Before disinfecting, rinse off cleaning solution
u Extend the facepiece head harness straps, of air in cylinder(s), during refill, a temperature in clean water.
5 leaving centre strap in position, (Fig. 9). increase occurs (approximately 10° C).
Description To set follow these instructions.
Refer to Fig. 1.
0158
Put neck strap over neck, insert neck strap stud 2. Disinfecting
The PSS 100 Series incorporates a unique into hole in centre strap of head harness. Indicated pressure after refill may decrease as u Incidur.
adjustable articulated carrying system. PSS 100 u Carry out High Pressure Leak Test. cylinder(s) cools to ambient temperature. Remaining Do Not exceed a temperature of 30 degrees
u Lift equipment into vertical position.
Series pneumatics use a high performance first working time will be reduced accordingly. Celsius.
u Turn locking knob (90 degrees) to unlock
stage pressure reducer, with mechanically activated Note: Ensure the positive pressure mechanism of
shoulder yoke.
pressure gauge and whistle warning unit, and is the lung demand valve (A and AE) is activated to u Following pressure equalisation - disconnect Note: Before drying, rinse off disinfecting solution
u Slide shoulder yoke, aligning required torso
compatible with a wide range of balanced piston switch ‘Off’ positive pressure. quick release coupling by pulling ‘grey’ sleeve, in clean water.
length with locking knob. Turn locking knob (90
lung demand valves facepieces and compressed
degrees) to lock shoulder yoke. Do Not press u ‘Open’ valve slowly but fully to pressurise of female coupling, away from male coupling.
air cylinders. The cylinder strap, provides secure u Dräger Wipex Cloths.
down on the adjuster while rotating it. system. Check gauge. See Fig. 12.
support for single cylinder.
Details of cleaning and disinfecting agents are
PSS 100 D - The ChargAirTM connection allows Fitting Cylinder 6 Putting on Facepiece Note: As the coupling separates, a ‘hiss’ or ‘pop’
available from Dräger on request.
u Check valve port and reducer handwheel threads may be heard as the non return valves in both
refilling of cylinder(s), from an independent high 0033
pressure (300bar) secondary breathing air supply are undamaged, connector O ring in position, Note: Refer also to the Instructions for Use provided couplings seal off the high pressure air.
and undamaged. with the Facepiece. 3. Rinsing and Drying
source, while wearer is breathing from equipment. Remove cleaning and disinfecting solutions by
u Place backplate horizontall and extend cylinder u Refit protection caps to male and female
strap. Safety Warning: Facial hair, beard stubble, couplings. rinsing in clean water, followed by drying.
Details of equipment variants and approved side whiskers, and the wearing of spectacles
u Slide cylinder through strap, locating valve port u Recheck pressure gauge and continue. Do Not exceed a temperature of 60 degrees
accessories available from Dräger on request. will adversely affect and interfere with Celsius when drying components.
to reducer handwheel.
facepiece seal. Correct fit of facepiece only
Intended Use u Lift unit upright, screw handwheel into valve
port, (handtight), and, if fitted, hook anti-vibration ensured if facepiece seal makes close contact After Use - ChargAirTM Pressure Reducer and Hose
The PSS 100 Series used in approved lung demand with skin. u Turn ‘Off’ the air supply at secondary air supply.
strap to handwheel. (Fig. 2).
u Fit protection cap over end of female quick
Assemblies
valve, facepiece and cylinder combinations, u Place backplate horizontal, take up slack on u Detach neck strap stud from centre strap of head
provides the wearer with respiratory protection when cylinder strap (Fig. 3), pull strap over cylinder to harness. release coupling on secondary supply hose.
Note: Refer to For Your Safety.
working in a contaminated, or oxygen deficient activate Camlock (Fig. 4), secure strap to Velcro. 7 u Spread harness (Fig. 10), place chin into u Vent air pressure from the secondary supply
gaseous atmosphere. 0502
facepiece, position harness over head, locating hose by pressing the centre of the end face of the u For cleaning and disinfecting, remove pressure
Connecting Demand Valve to Equipment harness centre plate with back of head. protection cap. reducer and hose assembly from backplate.
Working duration of equipment is dependent on u Tighten both lower (1), then upper straps (2) u Unscrew secondary ChargAirTM supply hose u Remove the locknut, washers and the reducer
capacity (volume) of cylinder selected, and breathing evenly towards back of head. Fig. 11. Only connector from secondary air supply outlet. retaining screw.
Note: Refer also to the Instructions for Use provided
rate of wearer. tighten centre strap (3) if necessary. u Unfasten hose loops, releasing hoses, and
with the Lung Demand Valve.
PSS 100 D - While wearer is breathing from equipment,
u On achieving face seal positive pressure After Use - Equipment remove reducer and hose assembly.
u Insert male coupling, of demand valve, into demand valves (A and AE) automatically
it is possible to extended working time by connecting activate on first inhalation. Note: Do Not immerse pressure reducer in cleaning
female coupling of medium pressure hose. Check Safety Warning: Do Not remove equipment until
ChargAirTM to a high pressure secondary breathing u Breathe normally.
security of connection. in safe area, clear of hazard. or disinfecting solutions.
air source to refill cylinder/cylinders.
Function Checks
Technical Data Pre Operational Checks 8 u Inhale and hold breath. Unit must balance, i.e. u Release side straps on head harness.
u Using clean cloth, moistened in cleaning, or
disinfecting solution, remove dirt and
High Pressure Connection: High Pressure Leak Test 0471
no audible leak. contaminants from the pneumatic assembly.
u Before opening cylinder valve, ensure positive u Continue breathing. Expired air should flow Note: Ensure the positive pressure mechanism of
200bar or 300bar u Using clean cloth, moistened in clean water,
pressure mechanism of lung demand valve (A easily out of exhalation valve. the lung demand valve (A and AE) is activated to
Standard G5/8 as per DIN 477: Part 6. remove disinfecting residue, followed by drying.
and AE) activated to switch ‘Off’ positive u Press centre of rubber cover of demand valve switch ‘Off’ positive pressure.
Other connections available to National Standards.
pressure. checking supplementary supply. Note: Do Not exceed a temperature of 60 degrees
u Open cylinder valve slowly, but fully, to u ‘Close’ cylinder valve. u Remove the facepiece. ‘Close’ cylinder valve.
Celsius when drying components.
High Pressure Connection - pressurise system. u Breathe normally to vent system. During venting u Unbuckle waistbelt, lift shoulder strap buckles
ChargAirTM (If Fitted) u Close cylinder valve and observe the pressure observe gauge - whistle alarm should sound at to loosen, remove equipment. u Re-assemble reducer and hose assembly to
300Bar maximum. gauge. preset pressure, (See Technical Data). When the backplate and harness.
High Pressure Male Quick Release Coupling. u Gauge reading should not decrease more than gauge indicates zero, hold breath. Facepiece Note: Do Not drop or throw down equipment as
10bar in 1 minute. should hold onto the face indicating positive seal. damage could occur.
Lung Demand Valves
Full Technical Specification available from Dräger. u Continue to Whistle Warning Unit Test. 9 u ‘Open’ the valve slowly but fully to pressurise Refer also to the Instructions for Use provided with
u Press the centre of the rubber cover of the
the system. the Lung Demand Valve.
Whistle Warning Unit Whistle Warning Unit Test demand valve to vent system and ensure
Note: If leak is detected then readjust the head positive pressure mechanism of lung demand
Preset by Dräger to 55bar +/- 5bar (50bar to 60bar). u A and AE - Cover outlet of demand valve with Facepiece
ball of hand, press centre of rubber cover, turning harness and retest. valve activated to switch ‘Off’ positive
pressure. Refer also to the Instructions for Use provided with
Compressed Air Cylinders ‘On’ positive pressure. Slowly vent system by
Use u Pass equipment to Service Department. the Facepiece.
4L to 12L capacity cylinders (200bar or 300bar) carefully lifting ball of hand maintaining a slow
decrease in pressure. u Regularly observe gauge reading. Whistle
available in either steel or composite materials.
Available for twin cylinder mounting is an approved u N - Slowly vent system. Carefully press centre alarm sounds at pressure setting of whistle Routine Maintenance Storage - Ready for Use
conversion kit. Contact Dräger for details. of rubber cover. warning unit.
To be performed after Use. u Fully extend shoulder straps, waistbelt and head
u During venting observe gauge. Whistle should u Go on to safe area, at least when whistle sounds,
by shortest and safest route. See also Maintenance and Test Intervals Chart. harness of the facepiece.
Safety Warning: When using equipment with sound at preset pressure.
twin cylinder configuration, or single 12Litre u Should whistle not sound at required pressure, 10 Refilling Cylinder(s) using u Store the equipment ‘ready for use’ in a cool
whistle may be reset as follows: Removal of cylinder dry environment, free from dust and dirt and
composite, it is essential to consider possible
restriction when working in confined space.
0016
ChargAirTM (If Fitted) away from excessive heat. Protect the
Refer to Fig. 5. Safety Warning: Cylinder valve should be equipment from any possible damage and
3 Note: See For Your Safety - ChargAirTM closed and system vented. ensure that rubber parts are not subjected to
Preparation for Use u Remove anti-tamper plug from adjuster (1) and 2 direct sunlight.
Wearer Height Adjustment using 3mm hexagon socket key: Secondary Air Supply Source u Lift free end of cylinder strap against Camlock to
PSS 100 Series carrying system length can be set a. Turn adjuster clockwise to increase pressure. Any system where compressed breathing air is release, loosen strap. Unhook anti-vibration strap Safety Note: When stored ‘ready for use’, on
to suit torso length of wearer. b. Turn adjuster anti-clockwise to decrease stored can be classified as a secondary air supply (if fitted) from the reducer handwheel and unscrew wall or bulkhead mounting, e.g. Fire Tender,
pressure. 1 source, such as: ensure equipment secured and supported by
handwheel from the cylinder valve.
u Cascade system mounting system.
Position 1.........Short Note: A quarter turn of adjuster is equivalent to a Slide the cylinder away from reducer and remove
u High pressure breathable air compressor with
Position 2.........Medium pressure change of approximately 2bar to 3bar. reservoir from equipment.
Position 3.........Long Do Not turn adjuster with pressure ‘On’. 11 u High pressure compressed air cylinder u Recharge the cylinder.
0018

3337446: 6th Edition February 2003 : Subject to Modification - (A3-D-P)


PSS 100 Series - PSS 100, PSS 100 D - A, AE, N
Compressed Air Respiratory Protection Equipment Instructions for Use

FAULT, CAUSE, REMEDY MAINTENANCE AND TEST INTERVALS


Carry out regular Inspection, Testing and Servicing of Equipment according to this table. Record all data in Equipment Log Book.
These instructions also apply to non-used (in Storage) Equipment.

FAULT POSSIBLE CAUSE REMEDY After Every Every Every


Description
Use Month Year 6 Years
Facepiece leakage Faulty or missing demand valve Fit or replace sealing ring or O-Ring
(Leak test with connector O-Ring (push-in demand Complete Clean and Disinfect !
equipment in valve) Equipment
Visual Inspection ! "
position)
Exhalation valve leakage Clean and replace as necessary Functional and Leak Testing as defined in
! "
Instructions for Use
Speech diaphragm defective Replace
Flow and Static Tests per Manufacturer’s instructions !
Headstraps not tight Tighten Headstraps
Demand Valve Clean and disinfect as necessary !#
Unsatisfactory Speech diaphragm defective Replace Pressure Medium Pressure Check !
communication
Reducer
Replace Sintered Filter !
High pressure leak Check tightness of connections Tighten or replace seals as necessary
Replace H.P. Connector O-ring !
Check hoses Replace as required Basic Overhaul
!
(Repair Exchange Scheme)
Safety relief valve Check relief valve O-Ring Replace as necessary
venting Cylinder Charge to correct pressure !
Pressure reducer defective Return to Dräger Service
Charged Pressure Check. Check Test Date Stamped
!"
Demand valve Banjo O-Rings defective Replace on Cylinder
vents Re-Certification - Cylinder Pressure Test (According
(constant flow) Balanced piston leak Replace
to National Standards).
Diaphragm incorrectly fitted Re-assemble correctly Cylinder Overhaul as necessary or at time of Cylinder
Valve Re-Certification.

m Dräger Recommendation - UK
uCOSHH
n Grease Demand Valve connector O-ring (Use Molykote 111) as required.
(Type A - Push-in demand valve only).

Dräger Safety UK Limited


Dräger, while endeavouring to ensure correctness of statements of fact and Ullswater Close Dräger Safety AG & Co. KGaA
advice contained in this publication; gives no guarantee or warranty in respect Kitty Brewster Estate Revalstraße 1 Telefon: 0049451882-0
thereof, and accepts no liability for any mis-statement or inaccuracy in publication, Blyth D-23560
or for any omission therefrom. Northumberland Tel: (01670) 352891 Lübeck Telex: 26807-0
NE 24 4 RG Fax: (01670) 356266 Germany Internet http://www.draeger.com

3337446: 6th Edition February 2003 : Subject to Modification - (A3-D-P)


PA90AE Lung Demand Valve
Positive Pressure Instructions for Use
Approvals Dräger recommends only the following: uConnect demand valve to equipment
PA90AE Series Lung Demand Valve, is approved for use with Dräger compressed air and perform Pre Operational Checks.
respiratory protection equipment conforming to EN137 and meeting the requirements of 1. Cleaning
EC Council Directive 89/686/EEC, BSEN 133. Equipment is CE Mark Approved and issued uDräger Safety Wash. Replacing Diaphragm

ü
with EC Type Examination Certificate. uSekusept
EC Type examiner: SGS United Kingdom Ltd, Unit 202b Worle Parkway, Weston-super- Do Not exceed a temperature of 30 Refer to Fig. 3.
Mare, BS22 0WA, UK. Notified Body No. 0120. degrees Celsius.
uPress re-set lever (5). Do Not press
For Your Safety Note: Before disinfecting, rinse off cleaning centre of rubber cover while pressing
solution in clean water. re-set lever, or force lever against stop.
uUse of equipment requires, knowledge, and compliance with National Regulations,
See Fig. 1.
Laws and Standards, governing the use of respiratory protection equipment in the 2. Disinfecting uRemove rubber cover (1). Using plate
country of use.

û
uIncidur. spanner (R26817), unscrew bayonet
uUse of equipment requires wearer training and observance of these Instructions for Do Not exceed a temperature of 30 cap (2).
Use and Instructions for Use supplied with the compressed air respiratory protection degrees Celsius. uCarefully unhook diaphragm (3) from
equipment. lever (4) of balanced piston and discard
uUse equipment only for the purpose specified in this manual, or as confirmed in writing Note: Before drying, rinse off disinfecting defective diaphragm.
by Dräger. solution in clean water. uInsert balanced piston lever (4) into hole
uOnly trained competent personnel should inspect and service equipment at regular Fig. 1 in centre boss of new diaphragm (3)
Details of cleaning and disinfecting agents
intervals and a record kept of such inspections and service. 0032/0251 are available from Dräger on request. (chamfered side of boss towards lever)
uDräger recommends a service contract be obtained from your Dräger Branch or Agent. and carefully locate outer bead of
uContact Dräger for details of Service Contracts and Service Training Courses. 3. Rinsing and Drying diaphragm into groove in demand valve
uNotify Dräger if there is component fault or failure. body - note orientation. Fig. 4.
Remove cleaning and disinfecting solutions by
uUse only original Dräger Spare Parts for service and maintenance. uMoisten lip of bayonet cap (2) with clean
rinsing in clean water, followed by drying.
uUse only Dräger Test Equipment for service and maintenance. water - re-fit bayonet cap. Re-check
Do Not exceed a temperature of 60 orientation of diaphragm.
Liability Statement degrees Celsius when drying components. uAssemble rubber cover (1).
uPress re-set lever. Do Not press centre
Responsibility for reliable function of equipment transfers to the owner or operator when To internally clean and disinfect lung of rubber cover while pressing re-set
serviced, or repaired by untrained personnel, (not employed or authorised by Dräger) or demand valve it is important to follow these lever, or force lever against stop. See
when used in a manner not conforming to its intended use. instructions. (see Fig. 3). Fig. 1.
uConnect demand valve to equipment
uPress re-set lever (5). Do Not press
and perform Pre Operational Checks.
centre of rubber cover while pressing
Description Connecting Demand Valve to re-set lever, or force lever against stop.
Fault, Cause, Remedy,
See Fig. 1.
The high performance PA90AE Lung Facepiece uRemove rubber cover (1). Using plate Service and Test Intervals
Demand Valve incorporates a unique pre- uCheck facepiece port, and sealing ring,
set and sealed balanced piston unit. spanner (R26817), unscrew bayonet uSee attached and refer also to
clean and undamaged. cap (2).
Should sealed balanced piston unit be Instructions for Use provided with
uScrew demand valve handwheel into uCarefully unhook diaphragm (3) from compressed air respiratory protection
tampered with, Drägers’ guarantee is void.
facepiece port. Handtight. Fig. 2 0028 lever (4) of balanced piston. equipment.
Lung Demand Valve - Part No. 3339406 is uAlign demand valve (Fig. 2) and tighten. 0033 uAssemble rubber cover (1) to demand
for use with the following compressed air valve body to protect balanced piston
respiratory protection equipment. Operation and Use lever.
PA80/90 Series uConnect lung demand valve to compressed uImmerse, and manually agitate demand
PA90 Series air respiratory protection equipment. valve, diaphragm and bayonet cap in
PA90Plus Series uRefer to Instructions for Use provided cleaning fluid. (Refer to cleaning fluid
and Lung Demand Valve - Part No. 3339405 with compressed air respiratory instructions.)
is for use with the following compressed air uAfter cleaning, rinse components in
protection equipment.
respiratory protection equipment. clean water.
PSS 100 Series uImmerse, and manually agitate
PSS 500 Series
After Use components in disinfecting fluid. (Refer
uSee attached and refer also to to disinfecting fluid instructions.)
In the presence of chemicals; Some Instructions for Use provided with uAfter disinfection, remove rubber cover
substances, e.g. Phosgene, may be compressed air respiratory protection (1) and rinse components in clean water.
absorbed by, or diffuse through, the equipment. Take care not to damage balanced
material of the silicone diaphragm of the piston lever.
PA90 Series lung demand valve. Cleaning, Disinfecting, Drying uConnect hose on demand valve to
medium pressure air supply. Press re-
Carefully clean, disinfect, and thoroughly dry
When lung demand valve fitted with non set lever. Turn ‘ON’ air supply.
contaminated dirty components after use.
silicone diaphragm, fit ‘green’ cover (1 Fig. 5 4 3 2 1 uGently depress balanced piston lever
2) for identification. To ensure correct operational condition of
Fig. 3 (4), blowing out fluid residue.
equipment, use only cleaning and 0762
Safety Note: Wear suitable eye
Preparation for Use disinfecting solutions recommended by protection.
Connecting Demand Valve to Dräger.
Equipment uDisconnect from air supply, dry
Use of other products invalidates Dräger’s components and, if necessary, using
Note: Do Not press centre of rubber cover guarantee. Dräger Wipex Cloths, remove excess
while pressing reset lever, or force lever fluid.
against stop. See Fig. 1. When using baths to contain cleaning and uAfter drying, insert balanced piston lever
disinfecting solutions, immersed (4) into hole in centre boss of diaphragm
uInsert male coupling of demand valve components and assemblies must be (3) (chamfered side of boss towards
into female coupling of equipment agitated manually. Dräger recommends lever) and carefully locate outer bead
medium pressure hose. Check security that no form of mechanical agitation be of diaphragm into groove in demand
of connection. used. valve body - note orientation. Fig. 4.
uRefer to Instructions for Use provided uMoisten lip of bayonet cap (2) with clean
with compressed air respiratory Safety Note: Refer to manufacturers’ usage water - re-fit bayonet cap. Re-check
protection equipment. instructions when using cleaning and orientation of diaphragm.
disinfecting agents. It is important that uAssemble rubber cover (1).
Pre Operational Checks attention be paid to concentration and uPress re-set lever. Do Not press centre
uRefer to Instructions for Use provided reaction times. Do Not use organic of rubber cover while pressing re-set
with compressed air respiratory solvents, such as Acetone, Alcohol, White Fig. 4 lever, or force lever against stop. See
protection equipment. Spirit, Trichloroethylene or similar. Fig. 1.
0029

Order List
Maintenance and Test Intervals Fault - Cause - Remedy
Description Order No
Lung After Every Fault Possible Cause Remedy
Demand Description
Use Month PA90AE Lung Dem and Valv e - Positiv e Pressure - Short Hose 3339405
Valve Demand Valve Banjo O-Rings Defective Replace O-Rings PA90AE Lung Dem and Valv e - Positiv e Pressure - Long Hose 3339406
Clean and Disinfect as m Vents -
necessary Constant Flow
Wipex Cloths (Pack of 50) 3380375

Incidur Disinfectant (6 litres) 7904072


Visual Inspection m u Balanced Piston Leak Replace Balanced
Incidur Disinfectant (30 litres) 7904073
Piston
1 Litre Safety Wash and Dispenser 3380164
Functional and Leak Test as m u Diaphragm incorrectly Reassemble
1 Litre Safety Wash Refill 3380165
defined in Instructions for Use fitted correctly
supplied w ith respiratory 5 Litre Safety Wash and Dispenser 3380166
protection equipm ent Perforated Diaphragm Replace Diaphragm
5 Litre Safety Wash Refill 3380167

Sekusept (4 bottles @ 2 litres) 7904071


m Dräger Recommendation
uCOSHH Diaphragm - PA90 Series Lung Dem and Valv es 3339452

Note: Refer also to Instructions for Use provided with compressed air respiratory Rubber Cov er - Green 3339450
equipment.
Non-Silicon Diaphragm 3339761

Dräger while endeavouring to ensure correctness of statements Dräger Safety UK Limited Dräger Safety AG & Co. KGaA
of fact and advice contained in this publication; gives no Ullswater Close Revalstraße 1 Telefon: 0049451882-0
guarantee or warranty in respect thereof, and accepts no liability Kitty Brewster Estate D-23560 Telefax: 0049451882-2080
Blyth Tel: (01670) 352891 Lübeck
for any mis-statement or inaccuracy in publication, or for any
ommission therefrom. Northumberland NE 24 4 RG Fax: (01670) 356266 Germany Internet http://www.draeger.com

3337393: 6th Edition: January 2003: Subject to Modification - (A3-D-P)


PA90A Lung Demand Valve
Positive Pressure Instructions for Use
Approvals uAssemble rubber cover (1) to demand Fault, Cause, Remedy,
PA90A Series Lung Demand Valve, is approved for use with Dräger compressed air valve body to protect balanced piston Service and Test Intervals
respiratory protection equipment conforming to EN137 and meeting the requirements of lever.
uSee attached and refer also to
EC Council Directive 89/686/EEC, BSEN 133. Equipment is CE Mark Approved and issued uImmerse, and manually agitate demand
Instructions for Use provided with

ü
with EC Type Examination Certificate. valve, diaphragm and bayonet cap in
cleaning fluid. (Refer to cleaning fluid compressed air respiratory protection
EC Type examiner: SGS United Kingdom Ltd, Unit 202b Worle Parkway, Weston-super- equipment.
Mare, BS22 0WA, UK. Notified Body No. 0120. instructions.)
uAfter cleaning, rinse components in clean
water.
For Your Safety uImmerse, and manually agitate
uUse of equipment requires, knowledge, and compliance with National Regulations,
components in disinfecting fluid. (Refer
Laws and Standards, governing the use of respiratory protection equipment in the
to disinfecting fluid instructions.)
country of use.

û
uAfter disinfection, remove rubber cover
uUse of equipment requires wearer training and observance of these Instructions for (1) and rinse components in clean water.
Use and Instructions for Use supplied with the compressed air respiratory protection Take care not to damage balanced
equipment. piston lever.
uUse equipment only for the purpose specified in this manual, or as confirmed in writing uConnect hose on demand valve to
by Dräger.
Abb. 1 medium pressure air supply. Press re-
uOnly trained competent personnel should inspect and service equipment at regular
set lever. Turn ‘ON’ air supply.
intervals and a record kept of such inspections and service. 0032/0251 uGently depress balanced piston lever
uDräger recommends a service contract be obtained from your Dräger Branch or Agent. (4), blowing out fluid residue.
uContact Dräger for details of Service Contracts and Service Training Courses.
uNotify Dräger if there is component fault or failure.
Safety Note: Wear suitable eye
uUse only original Dräger Spare Parts for service and maintenance.
protection.
uUse only Dräger Test Equipment for service and maintenance.
uDisconnect from air supply, dry
Liability Statement components and, if necessary, using
Responsibility for reliable function of equipment transfers to the owner or operator when Dräger Wipex Cloths, remove excess
serviced, or repaired by untrained personnel, (not employed or authorised by Dräger) or
fluid.
when used in a manner not conforming to its intended use.
uAfter drying, insert balanced piston
lever (4) into hole in centre boss of
diaphragm (3) (chamfered side of
Description Connecting Demand Valve to boss towards lever) and carefully
locate outer bead of diaphragm into
The high performance PA90A Lung Facepiece
Demand Valve incorporates a unique pre- groove in demand valve body - note
uCheck facepiece port, and demand
set and sealed balanced piston unit. orientation. Fig. 3.
valve O ring, are clean and undamaged.
Should sealed balanced piston unit be 5 4 3 2 1 uMoisten lip of bayonet cap (2) with clean
tampered with, Drägers’ guarantee is void.
uPush demand valve into facepiece port Abb. 2 water - re-fit bayonet cap. Re-check
until it clicks into position.
0028 orientation of diaphragm.
uCheck attachment is secure. Pull
Lung Demand Valve - Part No. 3339402 is uAssemble rubber cover (1).
demand valve away from facepiece -
for use with the following compressed air uPress re-set lever. Do Not press centre
respiratory protection equipment. there should be no axial movement.
of rubber cover while pressing re-set
PA80/90 Series lever, or force lever against stop. See
PA90 Series Operation and Use Fig. 1.
PA90Plus Series uConnect lung demand valve to compressed uConnect demand valve to equipment
and Lung Demand Valve - Part No. air respiratory protection equipment.
and perform Pre Operational Checks.
3339400 is for use with the following uRefer to Instructions for Use provided
compressed air respiratory protection with compressed air respiratory
equipment. protection equipment. Replacing Diaphragm
PSS 100 Series
PSS 500 Series After Use Refer to Fig. 2.
uSee attached and refer also to uPress re-set lever (5). Do Not press
In the presence of chemicals; Some
Instructions for Use provided with centre of rubber cover while pressing
substances, e.g. Phosgene, may be
compressed air respiratory protection re-set lever, or force lever against stop.
absorbed by, or diffuse through, the
equipment. See Fig. 1.
material of the silicone diaphragm of the
uRemove rubber cover (1). Using plate
PA90 Series lung demand valve.
Cleaning, Disinfecting, Drying Abb. 3 spanner (R26817), unscrew bayonet
When lung demand valve fitted with non Carefully clean, disinfect, and thoroughly dry cap (2).
silicone diaphragm, fit ‘green’ cover (1 Fig. contaminated dirty components after use. 0029
uCarefully unhook diaphragm (3) from
2) for identification. lever (4) of balanced piston and discard
To ensure correct operational condition of Dräger recommends only the following: 3. Rinsing and Drying defective diaphragm.
Remove cleaning and disinfecting
Preparation for Use equipment, use only cleaning and
1. Cleaning solutions by rinsing in clean water,
uInsert balanced piston lever (4) into hole
disinfecting solutions recommended by in centre boss of new diaphragm (3)
Connecting Demand Valve to Equipment Dräger. Use of other products invalidates uDräger Safety Wash. followed by drying. (chamfered side of boss towards lever)
Dräger’s guarantee. uSekusept and carefully locate outer bead of
Note: Do Not press centre of rubber cover Do Not exceed a temperature of 30 Do Not exceed a temperature of 60
while pressing reset lever, or force lever diaphragm into groove in demand valve
degrees Celsius. degrees Celsius when drying components.
against stop. See Fig. 1. When using baths to contain cleaning and body - note orientation. Fig. 3.
disinfecting solutions, immersed uMoisten lip of bayonet cap (2) with clean
Note: Before disinfecting, rinse off cleaning To internally clean and disinfect lung
uInsert male coupling of demand valve components and assemblies must be solution in clean water. water - re-fit bayonet cap. Re-check
demand valve it is important to follow these
into female coupling of equipment agitated manually. Dräger recommends instructions. (see Fig. 2). orientation of diaphragm.
medium pressure hose. Check security that no form of mechanical agitation be 2. Disinfecting uAssemble rubber cover (1).
of connection. used. uIncidur. uPress re-set lever (5). Do Not press uPress re-set lever. Do Not press centre
uRefer to Instructions for Use provided Do Not exceed a temperature of 30 centre of rubber cover while pressing of rubber cover while pressing re-set
with compressed air respiratory Safety Note: Refer to manufacturers’ usage degrees Celsius. re-set lever, or force lever against stop. lever, or force lever against stop. See
protection equipment. instructions when using cleaning and See Fig. 1. Fig. 1.
disinfecting agents. It is important that Note: Before drying, rinse off disinfecting uRemove rubber cover (1). Using plate uConnect demand valve to equipment
Pre Operational Checks attention be paid to concentration and solution in clean water. spanner (R26817), unscrew bayonet and perform Pre Operational Checks.
uRefer to Instructions for Use provided reaction times. Do Not use organic cap (2). Refer to Instructions for Use provided
with compressed air respiratory solvents, such as Acetone, Alcohol, White Details of cleaning and disinfecting agents uCarefully unhook diaphragm (3) from with compressed air respiratory
protection equipment. Spirit, Trichloroethylene or similar. are available from Dräger on request. lever (4) of balanced piston. protection equipment.

Maintenance and Test Intervals Fault - Cause - Remedy Order List


Description Order No
Lung Fault Possible Cause Remedy
After Every
Demand Description PA90A Lung Dem and Valve - Positive Pressure - Short Hose (PSS500) 3339400
Use Month
Valve Demand Valve Banjo O-Rings Defective Replace O-Rings PA90A Lung Dem and Valv e - Positiv e Pressure - Long Hose 3339402
Vents -
Clean and Disinfect as mn Wipex Cloths (Pack of 50) 3380375
Constant Flow
necessary Incidur Disinfectant (6 litres) 7904072

Balanced Piston Leak Replace Balanced Incidur Disinfectant (30 litres) 7904073
Visual Inspection m u
Piston 1 Litre Safety Wash and Dispenser 3380164
Functional and Leak Test as m u
Diaphragm incorrectly Reassemble 1 Litre Safety Wash Refill 3380165
defined in Instructions for
fitted correctly 5 Litre Safety Wash and Dispenser 3380166
Use supplied w ith respiratory
protection equipm ent Perforated Diaphragm Replace Diaphragm 5 Litre Safety Wash Refill 3380167

Sekusept (4 bottles @ 2 litres) 7904071


m Dräger Recommendation
u COSHH Diaphragm - PA90 Series Lung Dem and Valv es 3339452
n Grease connector O-Ring of Demand Valve as required (Use Molykote 111 Grease)
Connector O-Ring - Plus90A Lung Dem and Valve - Positive Pressure R18352
Note: Refer also to Instructions for Use provided with compressed air respiratory equipment.
Rubber Cov er - Green 3339450

Non-Silicon Diaphragm 3339761

Dräger while endeavouring to ensure correctness of statements of Dräger Safety UK Limited Dräger Safety AG & Co. KGaA
fact and advice contained in this publication; gives no guarantee or Ullswater Close Revalstraße 1 Telefon: 0049451882-0
warranty in respect thereof, and accepts no liability for any mis-statement Kitty Brewster Estate Tel: (01670) 352891 D-23560 Telefax: 0049451882-2080
or inaccuracy in publication, or for any omission therefrom. Blyth Lübeck
Northumberland NE 24 4 RG Fax: (01670) 356266 Germany Internet http://www.draeger.com

3337392: 6th Edition: January 2003: Subject to Modification - (A3-D-P)


Plus Series Lung Demand Valve - Plus N
Compressed Air Respiratory Protection Equipment Instructions for Use

Approvals u Fold rubber cover back over front of demand


Plus Series Lung Demand Valves, are approved for use with Dräger compressed air respiratory protection valve body to protect balanced piston lever.
equipment conforming to EN137 and meeting the requirements of EC Council Directive 89/686/EEC, u Immerse, and manually agitate demand valve,
BSEN 133. Equipment is CE Mark Approved and issued with EC Type Examination Certificate. diaphragm, slip ring, and bayonet cap in
EC Type examiner: SGS United Kingdom Ltd, Unit 202b Worle Parkway, Weston-super-Mare, BS22 cleaning fluid. (Refer to cleaning fluid
0WA, UK. Notified Body No. 0120. instructions.)
u After cleaning, rinse components in clean
water.
For Your Safety u Immerse, and manually agitate components
u Use and maintenance of equipment requires knowledge, and compliance with National Regulations, in disinfecting fluid. (Refer to disinfecting fluid
Laws and Standards, governing the use of respiratory protection equipment in the country of use. instructions.)
u Use of equipment requires training and observance of these Instructions for Use and Instructions for u After disinfection, fold rubber cover from front
Use supplied with the compressed air respiratory protection equipment. of demand valve and rinse components in
u Use equipment only for the purpose specified in this manual, or as confirmed in writing by Dräger. clean water. Take care not to damage balanced
u Only trained competent personnel should inspect and service equipment at regular intervals and a piston lever (5).
record kept of such inspections and service. u Connect hose of demand valve to medium
u Dräger recommends a service contract be obtained from your Dräger Branch or Agent. pressure air supply.
u Contact Dräger for details of Service Contracts and Service Training Courses. u Gently press balanced piston lever (5) blowing
u Notify Dräger if there is component fault or failure. out fluid residue.
u Use only original Dräger Spare Parts for service and maintenance.
u Use only Dräger Test Equipment for service and maintenance. Safety Note: Wear suitable eye protection.

Liability Statement u Disconnect from air supply, dry components


Responsibility for reliable function of equipment transfers to the owner or operator when serviced, or and, if necessary remove excess fluid. Do not
repaired by untrained personnel, (not employed or authorised by Dräger) or when used in a manner not leave LDV in drying temperatures for extended
conforming to its intended use. periods, remove as soon as dry.
u After drying, carefully locate outer bead of
diaphragm into groove in demand valve body.
u Carefully locate tapered side of slip ring into
Description After Use recess in bead of diaphragm and refit bayonet
cap. 1 2 3 4 5
The high performance Plus N Normal Demand u Fold rubber cover back over front of demand
Lung Demand Valve incorporates a unique preset Safety Warning: Do Not remove equipment
until in safe area, clear of hazard. valve.
and sealed balanced piston unit. Should sealed u Connect demand valve to equipment and
u Refer to Instructions for Use provided with
balanced piston unit be tampered with, Drägers’ perform Pre Operational Checks as detailed
compressed air respiratory protection
guarantee is void. Correct operational condition in Instructions for Use for equipment.
equipment.
is valid only if Dräger service and reseals the u Following venting of equipment, pass 1
balanced piston unit. equipment to Service Department. Replacing Diaphragm
0593
Plus N Lung Demand Valve is approved for use
with the following compressed air respiratory
Routine Maintenance Refer to Fig. 1.

protection equipment. To be performed after Use. u Fold rubber cover (1) from front of body of
Refer to Instructions for Use provided with demand valve. Grip bayonet cap (2), turn Order List
PA80 upgrade compressed air respiratory protection equipment. anticlockwise and remove bayonet cap.
PA90 upgrade u Using thumb and forefinger carefully grip
centre plate of diaphragm (4) tilt centre plate Description Order Code
PA90Plus series + Upgrades
PSS100 Series
Cleaning, Disinfecting, Drying and lift the diaphragm from the body of lung
Carefully clean, disinfect, and thoroughly dry demand valve. Remove slip ring (3). Discard
PSS500 Series
contaminated dirty components as necessary. diaphragm.
u Carefully locate outer bead of new diaphragm
Contact Dräger for details of additional approvals. To ensure correct operational condition of into groove in demand valve body.
Plus N Lung Dem and Valv e - Short Hose 3350501
equipment covered in this Instruction for Use, u Carefully locate tapered side of slip ring into
Intended Use use only cleaning and disinfecting solutions recess in bead of diaphragm and refit bayonet
Plus N Lung Demand Valve, used with approved recommended by Dräger. Use of other products cap.
compressed air respiratory protection equipment, invalidates the Dräger Warranty. u Fold rubber cover back over front of demand Second Lung Dem and Valv e - Plus N 3350606
facepiece and cylinder combinations, provide the When using baths to contain cleaning and valve.
wearer with respiratory protection when working disinfecting solutions, immersed components u Connect demand valve to equipment and
in a contaminated, or oxygen deficient gaseous and assemblies must only be agitated perform Pre Operational Checks as detailed
atmosphere. manually. Any form of mechanical agitation, in Instructions for Use for equipment. Sekusept (4 bottles @ 2 litres) 7904071
e.g. ultrasonic, washing machines,
Refer to Instructions for Use supplied with the dishwashers etc, must be approved and Fault, Cause, Remedy,
compressed air respiratory protection equipment. covered by a written agreement with Dräger.
Service and Test Intervals Incidur Disinfectant (6 litres container) 7904072
In the presence of chemicals:some substances, Note: Cleaning solutions will eventually remove u Refer to Instructions for Use provided with
e.g. Phosgene, may be absorbed by, or diffuse lubrication from moving parts. Following cleaning, compressed air respiratory protection
through the material of the silicone diaphragm of disinfecting and drying carry out functional tests. equipment.
the Plus Series lung demand valve. When the lung Re-lubricate moving parts as required. Incidur Disinfectant (30 litres container) 7904073
demand valve is fitted with the non-silicone
diaphragm fit ‘green’ cover (1) Fig. 1 for Safety Note: Refer to manufacturers’ usage
identification. instructions when using cleaning and
disinfecting agents. It is important that
attention be paid to concentration and reaction
Diaphragm - Silicone 3338747
In temperatures below freezing: correct times. Do Not use organic solvents, such as
function of the Lung Demand Valve will be impaired Acetone, Alcohol, White Spirit,
should water be trapped, then freezes inside the Trichloroethylene or similar.
Lung Demand Valve. It is important that water is Rubber Cov er - Green 3338799
removed from the Lung Demand Valve and Dräger recommends only the following:
connection hose. Thoroughly dry after any
cleaning operation. 1. Cleaning
u Dräger Safety Wash. Non-Silicone Diaphragm 3338751
Technical Data u Sekusept
Facepiece to Demand Valve
These products must be used with cold water
Connection however never exceed temperature 30 degrees
Normal Demand
TypeN - 40mm Thread to EN148-1
Celsius. Maintenance and Test Intervals
When the demand valve is fitted to the face piece Note: Before disinfecting, rinse off cleaning
the swivel feature of the handwheel allows a solution in clean water. Lung Demand Description After Use Every
degree of radial movement of the demand valve
during head and body movement of the wearer. 2. Disinfecting
Valve Month
The connection is not suitable for constant rotation. u Incidur
Full Technical Specification available from Dräger. Clean and Disinfect as necessary m
This product must be used with cold water
however never exceed temperature 30 degrees
Preparation for Use Celsius. Visual Inspection m u
Connecting Demand Valve to
Note: Before drying, rinse off disinfecting solution
Equipment
This instruction applies to demand valves with
in clean water. Functional and Leak Test as defined in Instructions for Use m u
quick release male coupling on hose. Details of cleaning and disinfecting agents are supplied w ith respiratory protection equipm ent
u Insert male coupling of demand valve into available from Dräger on request.
female coupling of equipment medium
pressure hose. Check security of connection. 3. Rinsing and Drying m Dräger Recommendation
u Refer to Instructions for Use provided with u COSHH
Remove cleaning and disinfecting solutions by
compressed air respiratory protection rinsing in clean water, followed by drying.
equipment.
Note: Refer also to Instructions for Use provided with compressed air respiratory equipment.
Do Not exceed a temperature of 60 degrees
Pre Operational Checks Celsius when drying components. Immediately
u Refer to Instructions for Use provided with remove from the drying cabinet when dry.
compressed air respiratory protection Never exceed a period of 30 minutes.
equipment.
Lung Demand Valve
Important Note: If an audible leak is evident To internally clean and disinfect lung demand valve
from the lung demand valve - Do Not restrict it is important to follow these instructions.
the outlet port.
Refer to Fig. 1.
Connecting Demand Valve to
Facepiece u Fold rubber cover (1) from front of body of
u Check facepiece port and sealing ring, clean demand valve. Grip bayonet cap (2), turn
and undamaged. anticlockwise and remove bayonet cap.
u Screw the demand valve handwheel into u Using thumb and forefinger carefully grip
facepiece port and tighten. centre plate of diaphragm (4), tilt centre plate
and lift the diaphragm from the body of lung
Operation and Use demand valve. Remove slip ring (3).
u Refer to Instructions for Use provided with Note: The diaphragm must not be ultrasonically
compressed air respiratory protection cleaned, as this will affect the bonded area of the
equipment. diaphragm. Clean and disinfect manually.

Draeger Safety UK Limited Dräger Safety AG & Co. KGaA


Dräger, while endeavouring to ensure correctness of statements of fact Ullswater Close Revalstraße 1 Telefon: 0049451882-0
and advice contained in this publication; gives no guarantee or warranty in Kitty Brewster Estate Tel: (01670) 352891 D-23560 Telefax: 0049451882-2080
respect thereof, and accepts no liability for any mis-statement or inaccuracy Blyth
in publication, or for any omission therefrom.
Lübeck
Northumberland NE 24 4 RG Fax: (01670) 356266 Germany Internet http://www.draeger.com

3338725: 10th Edition: August 2004: Subject to Modification (A3-D-P)


PLUS SERIES LUNG DEMAND VALVE - PLUS A
COMPRESSED AIR RESPIRATORY PROTECTION EQUIPMENT Instructions for Use

Approvals u Using thumb and forefinger carefully grip


Plus Series Lung Demand Valves are approved for use with Dräger compressed air respiratory
centre plate of diaphragm (4) then tilt centre 6
plate and lift the diaphragm from the body of
protection equipment conforming to EN137 and meeting the requirements of EC Council Directive
lung demand valve. Remove slip ring (3).
89/686/EEC, BSEN 133. Equipment is CE Mark Approved and issued with EC Type Examination
Certificate.
Note: The diaphragm must not be
ultrasonically cleaned, as this will affect the
EC Type examiner: SGS United Kingdom Ltd, Unit 202b Worle Parkway, Weston-super-Mare,
bonded area of the diaphragm. Clean and
BS22 0WA, UK. Notified Body No. 0120.
disinfect manually.

For Your Safety u Fold rubber cover back over front of demand
u Use and maintenance of equipment requires knowledge, and compliance with National Regulations, valve body to protect balanced piston lever.
Laws and Standards, governing the use of respiratory protection equipment in the country of use. u Immerse, and manually agitate demand valve,
u Use of equipment requires wearer training and observance of these Instructions for Use and diaphragm, slip ring, and bayonet cap in
Instructions for Use supplied with the compressed air respiratory protection equipment. cleaning fluid. (Refer to cleaning fluid
u Use equipment only for the purpose specified in this manual, or as confirmed in writing by Dräger. instructions.)
u Only trained competent personnel should inspect and service equipment at regular intervals and u After cleaning, rinse components in clean
a record kept of such inspections and service. water.
u Dräger recommends a service contract be obtained from your Dräger Branch or Agent. u Immerse and manually agitate components
u Contact Dräger for details of Service Contracts and Service Training Courses. in disinfecting fluid. (Refer to disinfecting fluid
u Notify Dräger if there is component fault or failure. instructions.)
u Use only original Dräger Spare Parts for service and maintenance. u After disinfection fold rubber cover from front
u Use only Dräger Test Equipment for service and maintenance. of demand valve and rinse components in
clean water. Take care not to damage
Liability Statement balanced piston lever (5).
u Connect hose of demand valve to medium
Responsibility for reliable function of equipment transfers to the owner or operator when serviced,
pressure air supply. Press reset button (6)
or repaired by untrained personnel, (not employed or authorised by Dräger) or when used in a
and turn ‘ON’ air supply.
manner not conforming to its intended use.
u Gently press balanced piston lever (5)
blowing out fluid residue.

Safety Note: Wear suitable eye protection.


Description u Refer to Instructions for Use provided with 2 3 4 5
compressed air respiratory protection u Disconnect from air supply, dry components 1
The high performance PlusA Positive Pressure and, if necessary remove excess fluid.
equipment.
Lung Demand Valve incorporates a unique u After drying, carefully locate outer bead of
preset and sealed balanced piston unit. Should diaphragm into groove in demand valve body.
1
Note: Following venting of equipment press
sealed balanced piston unit be tampered with u Carefully locate tapered side of slip ring into
reset button (6) to switch ‘Off’ positive 0584
Drägers’ guarantee is void. Correct operational recess in bead of diaphragm.
pressure mechanism.
condition is valid only if Dräger service and u Locate spring of bayonet cap in the centre
reseals the balanced piston unit.
u Pass equipment to Service Department. of the diaphragm and refit bayonet cap.
u Fold rubber cover back over front of demand Order List
Plus90A Lung Demand Valve is approved for valve.
use with the following compressed air respiratory Routine Maintenance u Press reset button to switch ‘Off’ positive
protection equipment. To be performed after Use. pressure mechanism. Description Order Code
u Connect demand valve to equipment and
PA80 upgrade Refer to Instructions for Use provided with perform Pre Operational Checks as detailed
PA90 upgrade compressed air respiratory protection equipment. in Instructions for Use for equipment. Plus A Lung Demand Valve - Shor t Hose 3338700
PA90Plus series + Upgrades
PSS 100 Series Cleaning, Disinfecting, Drying
PSS 500 Series Replacing Diaphragm Diaphragm - Silicone 3338747
Carefully clean, disinfect, and thoroughly dry
contaminated dirty components as necessary. Refer to Fig. 1.
Contact Dräger for details of additional approvals.
Connector O-Ring R18352
To ensure correct operational condition of u Fold rubber cover (1) from front of body of
Intended Use equipment covered in this Instruction for Use, demand valve. Grip bayonet cap (2), turn
Plus A Lung Demand Valve, used with approved use only cleaning and disinfecting solutions anticlockwise and remove bayonet cap.
compressed air respirator y protection recommended by Dräger. Use of other products
Rubber Cover - Green 3338799
equipment, facepiece and cylinder invalidates the Dräger Warranty. Note: Positive pressure spring remains
combinations, provide the wearer with respiratory attached to bayonet cap. Do Not remove, Non-Silicone Diaphragm 3338751
protection when working in a contaminated, or When using baths to contain cleaning and stretch or compress spring.
oxygen deficient gaseous atmosphere. disinfecting solutions, immersed components
and assemblies must only be agitated manually. u Using thumb and forefinger carefully grip Sekusept ( 4 bottles @ 2 litres) 7904071
Refer to Instructions for Use supplied with the Any form of mechanical agitation, e.g. centre plate of diaphragm (4) then tilt centre
compressed air respiratory protection equipment. ultrasonic, washing machines, dishwashers etc, plate and lift the diaphragm from the body of
must be approved and covered by a written lung demand valve. Remove slip ring (3). Incidur Disinfectant (6 litres container) 7904072
In the presence of chemicals; some substances, agreement with Dräger. Discard diaphragm.
e.g. Phosgene, may be absorbed by, or diffuse u Carefully locate outer bead of new diaphragm
through the material of the silicone diaphragm Note: Cleaning solutions will eventually into groove in demand valve body. Incidur Disinfectant (30 litres container) 7904073
of the Plus Series lung demand valve. When remove lubrication from moving parts. u Carefully locate tapered side of slip ring into
the lung demand valve is fitted with the non- Following cleaning, disinfecting and drying recess in bead of diaphragm.
silicone diaphragm fit ‘green’ cover (1) Fig. 1 carry out functional tests. Re-lubricate moving u Locate spring of bayonet cap in the centre
for identification. parts as required. of the diaphragm and refit bayonet cap.
u Fold rubber cover back over front of demand
In temperatures below freezing: correct Safety Note: Refer to manufacturers’ usage
function of the Lung Demand Valve will be instructions when using cleaning and
valve.
u Press reset button to switch ‘Off’ positive
Maintenance and Test Intervals
impaired should water be trapped, then freezes disinfecting agents. It is important that attention pressure mechanism.
be paid to concentration and reaction times. u Connect demand valve to equipment and
inside the Lung Demand Valve. It is important
Do Not use organic solvents, such as Acetone, Lung
that water is removed from the Lung Demand perform Pre Operational Checks as detailed Every
Valve and connection hose. Thoroughly dry after Alcohol, White Spirit, Trichloroethylene or in Instructions for Use for equipment. Demand Description After Use
similar. Month
any cleaning operation.
Valve
Fault, Cause, Remedy,
Technical Data Dräger recommends only the following:
Service and Test Intervals
Facepiece to Demand Valve Clean and Disinfect as necessary mn
1. Cleaning u Refer to Instructions for Use provided with
Connection u Dräger Safety Wash. compressed air respiratory protection
Positive Pressure u Sekusept equipment. Visual Inspection m u
Type A - Push-In
These products must be used with cold water Functional and Leak Test as defined in Instructions for Use m u
Full Technical Specification available from Dräger. however never exceed temperature 30 degrees
Celsius. supplied with respiratory protection equipment
Preparation for Use
Connecting Demand Valve to Note: Before disinfecting, rinse off cleaning
solution in clean water.
Equipment m Dräger Recommendation
This instruction applies to demand valves with u COSHH
2. Disinfecting n Grease connector O-Ring of Demand Valve as required (Use Molykote 111 Grease)
quick release male coupling on hose.
u Incidur
u Insert male coupling of demand valve into
female coupling of equipment medium Note: Refer also to Instructions for Use provided with compressed air respiratory equipment.
This product must be used with cold water
pressure hose. Check security of connection.
however never exceed temperature 30 degrees
u Refer to Instructions for Use provided with
Celsius.
compressed air respiratory protection
equipment.
Note: Before drying, rinse off disinfecting
solution in clean water.
Pre Operational Checks
u Refer to Instructions for Use provided with Details of cleaning and disinfecting agents are
compressed air respiratory protection available from Dräger on request.
equipment.
3.Rinsing and Drying
Important Note: If an audible leak is evident Remove cleaning and disinfecting solutions
from the lung demand valve – Do Not restrict by rinsing in clean water, followed by drying.
the outlet port.
Do Not exceed a temperature of 60 degrees
Connecting Demand Valve to Celsius when drying components.
Facepiece Immediately remove from the drying cabinet
u Check facepiece port, and demand valve O when dry. Never exceed a period of 30
ring are clean and undamaged. minutes.
u Push demand valve into facepiece port until
it clicks into position. Lung Demand Valve
u Check attachment is secure. Pull demand To internally clean and disinfect lung demand
valve away from facepiece - there should be valve it is important to follow these instructions.
no axial movement.
Refer to Fig. 1.
Operation and Use u Fold rubber cover (1) from front of body of
u Refer to Instructions for Use provided with demand valve. Grip bayonet cap (2), turn
compressed air respiratory protection anticlockwise and remove bayonet cap.
equipment.
Note: Positive pressure spring remains
After Use attached to the bayonet cap. Do Not remove,
stretch or compress spring. Take care not to
Safety Warning: Do Not remove equipment damage spring during cleaning and
until in safe area, clear of hazard. disinfecting procedures.

3338727 : 7th Edition : January 2005 : Subject to Modification (A3-D-P)


Plus Series Lung Demand Valve - Plus AE - M45 x 3
Compressed Air Respiratory Protection Equipment Instructions for Use

Approvals Note: Positive pressure spring remains attached


Plus Series Lung Demand Valves are approved for use with Dräger compressed air respiratory protection equipment to bayonet cap. Do Not remove, stretch or
conforming to EN137 and meeting the requirements of EC Council Directive 89/686/EEC, BSEN 133. Equipment compress spring. Take care not to damage spring
is CE Mark Approved and issued with EC Type Examination Certificate. EC Type examiner: SGS United Kingdom during cleaning and disinfecting procedures. 6
Ltd, Unit 202b Worle Parkway, Weston-super-Mare, BSS 0WA, UK. Notified Body No. 0120.
u Using thumb and forefinger carefully grip
For Your Safety centre plate of diaphragm (4) then tilt centre
plate and lift the diaphragm from the body of
u Use and maintenance of equipment requires knowledge, and compliance with National Regulations, lung demand valve. Remove slip ring (3).
Laws and Standards, governing the use of respiratory protection equipment in the country of use.
u Use of equipment requires wearer training and observance of these Instructions for Use and Instructions Note: The diaphragm must not be ultrasonically
for Use supplied with the compressed air respiratory protection equipment. cleaned, as this will affect the bonded area of the
u Use equipment only for the purpose specified in this manual, or as confirmed in writing by Dräger. diaphragm.
u Only trained competent personnel should inspect and service equipment at regular intervals and a Clean and disinfect manually.
record kept of such inspections and service.
u Dräger recommends a service contract be obtained from your Dräger Branch or Agent. u Fold rubber cover back over front of demand
u Contact Dräger for details of Service Contracts and Service Training Courses. valve body to protect balanced piston lever.
u Notify Dräger if there is component fault or failure. u Immerse, and manually agitate demand valve,
u Use only original Dräger Spare Parts for service and maintenance. diaphragm, slip ring, and bayonet cap in
u Use only Dräger Test Equipment for service and maintenance. cleaning fluid. (Refer to cleaning fluid
instructions.)
Liability Statement u After cleaning, rinse components in clean
Responsibility for reliable function of equipment transfers to the owner or operator when serviced, or water.
repaired by untrained personnel, (not employed or authorised by Dräger) or when used in a manner not u Immerse, and manually agitate components
conforming to its intended use. in disinfecting fluid. (Refer to disinfecting fluid
instructions.)
u After disinfection fold rubber cover from front
of demand valve and rinse components in
clean water. Take care not to damage balanced
Description After Use piston lever (5).
The high performance Plus AE Positive Pressure u Connect hose of demand valve to medium
Lung Demand Valve incorporates a unique preset Safety Warning: Do Not remove equipment pressure air supply. Press reset button (6) and
and sealed balanced piston unit. Should sealed until in safe area, clear of hazard. turn ‘ON’ air supply. 1 2 3 4 5
balanced piston unit be tampered with Drägers’ u Gently press balanced piston lever (5) blowing
guarantee is void. Correct operational condition u Refer to Instructions for Use provided with out fluid residue.
is valid only if Dräger service and reseals the
balanced piston unit.
compressed air respiratory protection
equipment. Safety Note: Wear suitable eye protection. 1
u Disconnect from air supply, dry components 0619
Plus AE Lung Demand Valve is approved for use Note: Following venting of equipment press reset and, if necessary remove excess fluid.
with the following compressed air respiratory button (6) to switch ‘Off’ positive pressure u After drying, carefully locate outer bead of
protection equipment. mechanism. diaphragm into groove in demand valve body. Order List
u Carefully locate tapered side of slip ring into
PA80 upgrade u Pass equipment to Service Department. recess in bead of diaphragm.
PA90 upgrade u Locate spring of bayonet cap in the centre of Description Order Code
PA90Plus series + Upgrades Routine Maintenance the diaphragm and refit bayonet cap.
PA 100 Series u Fold rubber cover back over front of demand
TTo be performed after Use. valve. Plus AE Lung Dem and Valv e - Short Hose 3338706
PA 500 Series
u Press reset button to switch ‘Off’ positive
Refer to Instructions for Use provided with pressure mechanism.
Contact Dräger for details of additional approvals. compressed air respiratory protection equipment. u Connect demand valve to equipment and Sekusept ( 4 bottles @ 2 litres) 7904071
perform Pre Operational Checks as detailed
Intended Use Cleaning, Disinfecting, Drying in Instructions for Use for equipment.
Plus AE Lung Demand Valve, used with approved Carefully clean, disinfect, and thoroughly dry Incidur Disinfectant (6 litres container) 7904072
compressed air respiratory protection equipment, contaminated dirty components as necessary. Replacing Diaphragm
facepiece and cylinder combinations, provide the
wearer with respiratory protection when working To ensure correct operational condition of Refer to Fig. 1. Incidur Disinfectant (30 litres container) 7904073
in a contaminated, or oxygen deficient gaseous equipment covered in this Instruction for Use,
atmosphere. use only cleaning and disinfecting solutions u Fold rubber cover (1) from front of body of
recommended by Dräger. Use of other products demand valve. Grip bayonet cap (2), turn Diaphragm - Silicone 3338747
Refer to Instructions for Use supplied with the invalidates the Dräger Warranty. anticlockwise and remove bayonet cap.
compressed air respiratory protection equipment.
When using baths to contain cleaning and Note: Positive pressure spring remains attached Rubber Cov er - Green 3338799
In the presence of chemicals; some substances, disinfecting solutions, immersed components to bayonet cap. Do Not remove, stretch or
e.g. Phosgene, may be absorbed by, or diffuse and assemblies must only be agitated compress spring.
through the material of the silicone diaphragm of manually. Any form of mechanical agitation, Non-Silicone Diaphragm 3338751
the Plus Series lung demand valve. When the lung e.g. ultrasonic, washing machines, u Using thumb and forefinger carefully grip
demand valve is fitted with the non-silicone dishwashers etc, must be approved and centre plate of diaphragm (4) then tilt centre
diaphragm fit ‘green’ cover (1) Fig. 1 for covered by a written agreement with Dräger. plate and lift the diaphragm from the body of
identification.
lung demand valve. Remove slip ring (3). Maintenance and Test Intervals
Note: Cleaning solutions will eventually remove Discard diaphragm.
In temperatures below freezing: correct lubrication from moving parts. Following cleaning, u Carefully locate outer bead of new diaphragm
function of the Lung Demand Valve will be impaired disinfecting and drying carry out functional tests. into groove in demand valve body.
Lung
should water be trapped, then freezes inside the Re-lubricate moving parts as required. u Carefully locate tapered side of slip ring into Demand Description After Use Every Month
Lung Demand Valve. It is important that water is
recess in bead of diaphragm. Valve
removed from the Lung Demand Valve and Safety Note: Refer to manufacturers’ usage u Locate spring of bayonet cap in the centre of
connection hose. Thoroughly dry after any instructions when using cleaning and the diaphragm and refit bayonet cap. Clean and Disinfect as necessary m
cleaning operation. disinfecting agents. It is important that u Fold rubber cover back over front of demand
attention be paid to concentration and reaction valve.
Technical Data times. Do Not use organic solvents, such as
Visual Inspection m u
u Press reset button to switch ‘Off’ positive
Facepiece to Demand Valve Acetone, Alcohol, White Spirit, pressure mechanism.
Connections
Trichloroethylene or similar. u Connect demand valve to equipment and Functional and Leak Test as defined in Instructions for m u
perform Pre Operational Checks as detailed Use supplied w ith respiratory protection equipm ent
Positive Pressure Dräger recommends only the following: in Instructions for Use for equipment.
TypeAE - M45 X 3 Thread to EN148-3
When the demand valve is fitted to the facepiece 1. Cleaning
the swivel feature of the handwheel allows a Fault, Cause, Remedy, m Dräger Recommendation
u Dräger Safety Wash.
degree of radial movement of the demand valve u Sekusept Service and Test Intervals u COSHH
during head and body movement of the wearer. u Refer to Instructions for Use provided with
The connection is not suitable for constant rotation. These products must be used with cold water compressed air respiratory protection Note: Refer also to Instructions for Use provided with compressed air respiratory equipment.
however never exceed temperature 30 degrees equipment.
Full Technical Specification available from Dräger. Celsius.

Preparation for Use Note: Before disinfecting, rinse off cleaning


solution in clean water.
Connecting Demand Valve to
Equipment 2. Disinfecting
This instruction applies to demand valves with u Incidur
quick release male coupling on hose.
u Insert male coupling of demand valve into This product must be used with cold water
female coupling of equipment medium however never exceed temperature 30 degrees
pressure hose. Check security of connection. Celsius.
u Refer to Instructions for Use provided with
compressed air respiratory protection Note: Before drying, rinse off disinfecting solution
equipment. in clean water.

Pre Operational Checks Details of cleaning and disinfecting agents are


u Refer to Instructions for Use provided with available from Dräger on request.
compressed air respiratory protection
equipment. 3. Rinsing and Drying
Remove cleaning and disinfecting solutions by
Important Note: If an audible leak is evident rinsing in clean water, followed by drying.
from the lung demand valve – Do Not restrict
the outlet port. Do Not exceed a temperature of 60 degrees
Celsius when drying components. Immediately
remove from the drying cabinet when dry.
Connecting Demand Valve to Never exceed a period of 30 minutes.
Facepiece
u Check facepiece port, and sealing ring, clean Lung Demand Valve
and undamaged. To internally clean and disinfect lung demand valve
u Screw demand valve handwheel into facepiece it is important to follow these instructions.
port. Handtight.
Refer to Fig. 1.
Operation and Use
u Refer to Instructions for Use provided with u Fold rubber cover (1) from front of body of
compressed air respiratory protection demand valve. Grip bayonet cap (2), turn
equipment. anticlockwise and remove bayonet cap.

Draeger Safety UK Limited Dräger Safety AG & Co. KGaA


Dräger, while endeavouring to ensure correctness of statements of fact Ullswater Close Revalstraße 1 Telefon: 0049451882-0
and advice contained in this publication; gives no guarantee or warranty in Kitty Brewster Estate D-23560 Telefax: 0049451882-2080
respect thereof, and accepts no liability for any mis-statement or inaccuracy Blyth Tel: (01670) 352891
in publication, or for any omission therefrom.
Lübeck
Northumberland NE 24 4 RG Fax: (01670) 356266 Germany Internet http://www.draeger.com

3338726: 6th Edition: October 2002: Subject to Modification (A3-D-P)


PSS 100 Series PSS 100 ET, PSS 100 ETD - A, AE and N
Compressed Air Respiratory Protection Equipment Instructions for Use
Approvals u Place backplate horizontal, take up slack Type A
PSS 100 Series compressed air respiratory protection equipment conforms to EN137 and 3 on cylinder strap (Fig. 10), pull strap u Check facepiece port and demand valve
meets the requirements of EC Council Directive 89/686/EEC, 97/23/EC, BSEN 133. over cylinder to activate Camlock (Fig. O ring are clean and undamaged.
PPE - EC Type examiner: SGS United Kingdom Ltd, Unit 202b Worle Parkway, Weston- 2 11), secure strap to Velcro. u Push the demand valve into facepiece
1 port until it ‘clicks’ into position. Check
super-Mare, BS22 0WA, UK. Notified Body No. 0120. CE Mark Approved and issued
with EC Type Examination Certificate. Connecting Demand Valve to the attachment is secure.
1 Equipment Type’s AE and N
PED - EC Type examiner: Lloyds Register, Hiramford, Middlemarch Office Village, Siskin
Drive, Coventry, CV 3 4FJ, England. Notified Body No. 0038. The equipment is u Check the facepiece port and sealing
Note: Refer also to the Instructions for Use ring are clean and undamaged.
manufactured in accordance with the requirements of Conformity Assessment Modules B
+ D of the Pressure Equipment Regulations. A copy of the Declaration of conformity is 1 12 23
provided with the Lung Demand Valve. u Screw demand valve handwheel into
available from Dräger on request. 0585 facepiece port. Handtight. Align the
0477 0018 u Insert male coupling, of demand valve, demand valve (Fig. 18) and tighten.
into female coupling of medium
For Your Safety pressure hose. Check security of
Operation
PSS 100 Series incorporates Dräger pre-set, and sealed pressure reducer. Drägers’ connection.
guarantee is void should original seal caps be tampered with, removed, or broken. Correct Putting on equipment
operational condition is only valid if Dräger service and re-seals pressure reducer. Pre Operational Checks u If necessary set Wearer Height
u PSS 100 ETD Series with ChargAirTM can only be connected to 300bar valved cylinder Adjustment.
High Pressure Leak Test u Extend waistbelt and shoulder adjusting
assemblies. u Ensure positive pressure mechanism of
u Use of equipment requires knowledge and compliance with National Regulations, Laws straps. Put on equipment - taking weight
the lung demand valve (A and AE) is on shoulders by pulling on shoulder
and Standards, governing the use of respiratory protection equipment in the country of use. activated to switch ‘Off’ positive straps. Do Not tighten.
u Use of equipment requires wearer training and observance of these Instructions for Use. 2 13 24 pressure and cylinder valve ‘off’. u Using right and left hand grip each side
u Use equipment only for the purpose specified in this manual, or as confirmed in writing u Remove Tally 1 Fig. 1. Bodyguard will of waistbelt pads lifting assembly to top
0490 0488 0500
by Dräger. begin ‘self check’ display sequence and of vertical slot of backplate. See Fig.
u Only trained competent personnel should inspect and service equipment at regular when Fig. 5 displayed - press LH button. 19. Fasten waistbelt buckle. Pull the
intervals and a record kept of such inspections and service. Bar graph segments sweep to the right ends of waistbelt adjusting straps
u Dräger recommends a service contract be obtained from your Dräger Branch or Agent. and ‘open valve’ icon displayed. forward (Fig. 20) until waistbelt pad
u Contact Dräger for details of Service Contracts and Service Training Courses. See Fig. 14. secure and comfortable on hip. Tuck
u Notify Dräger if there is component fault or failure. strap ends between body and waistbelt
u Open the cylinder valve fully to
u Use only original Dräger Spare Parts for service and maintenance. pad.
pressurise system. Display changes - u Pull down shoulder adjusting straps until
u Use only Dräger Test Equipment for service and maintenance. Fig. 15 - actual pressure displayed, equipment is secure and comfortable
alternating ‘close valve’ and ‘press RH on hip. Do Not overtighten. Tuck strap
For Your Safety - ChargAir (If Fitted) TM
button’ icons. ends under the waistbelt.
u Use ChargAirTM only for the purpose specified in this manual, or as confirmed in writing 3 14 25 u Immediately ‘Close’ the cylinder valve. u Extend the facepiece head harness
by Dräger. 0473 0481
Bar segments will sweep to the right and straps, leaving centre strap in position,
0496
u Do Not use ChargAirTM to transfer air from one compressed air breathing equipment when only two or one segments remain (Fig. 21). Put neck strap over neck,
to another, or for second man attachment. then ‘press’ RH button. Displays insert neck strap stud into hole in centre
u Do Not use any charging hose assemblies, or any other equipment, unless specifically changes to Fig. 16 - starting leak test strap of the head harness.
supplied by Dräger for use with ChargAirTM. Only use approved equipment. timing sequence. u Ensure positive pressure mechanism
u Do Not lubricate, or allow oil, grease or other contaminants to contact ChargAirTM activated to switch ‘Off’ positive
fittings or hoses. Note: During the test timing sequence pressure.
u Do Not attempt to disassemble or repair ChargAirTM fittings or hoses. Drägers’ the pressure reading may reduce by up u Remove Tally. Bodyguard will begin
to 10bar. This may be due to the settling ‘self check’ display sequence.
guarantee is void should ChargAirTM be tampered with, removed, or broken. Correct
operational condition is only valid if Dräger service pressure reducer and ChargAirTM of the pressure reducer.
Safety Warning: Do Not continue if ‘Low
assembly.
Battery’ warning (Fig. 12) or ‘Failed’ icon
u Caution: secondary air supply pressure must not exceed maximum rated working 4 15 26 u Bar graph segments sweep to the right
(X) displayed.
pressure (300bar) of cylinder(s) being filled. 0590 0485 0497 and either:
u Caution: if leak detected while refilling in contaminated, or oxygen deficient gaseous Icon ‘a’ Fig. 17 displayed at end of
u When Fig. 5 displayed - press LH
atmosphere, discontinue refilling and immediately leave hazardous area. timing sequence - equipment ‘Passed’
button. Bar graph segments sweep to
u Caution: it is the users’ responsibility to ensure that a. the secondary ChargAirTM test. Continue to Whistle Alarm Test.
the right and ‘open valve’ icon
hose connector is compatible with outlet connector of independent high pressure or
displayed. See Fig. 14.
(300bar) secondary breathing air supply source, and b. the connection is correctly, Icon ‘b’ Fig. 17 displayed during leak
u Open the cylinder valve fully to
and securely assembled. test timing sequence - equipment
pressurise system. Display changes -
‘Failed’ test. See Note.
Fig. 15 - actual pressure displayed,
Extreme Service Conditions Note: To determine whether the failure is
alternating ‘close valve’ and ‘press RH
In the presence of chemicals; Some substances, e.g. Phosgene, may be absorbed by, button’ icons.
due to a slow to settle reducer or a HP leak
or diffuse through, the material of the silicone diaphragm of the PA90 Series lung demand u Immediately ‘Close’ the cylinder valve.
valve. 5 16 27 – proceed to retest as follows:
Bar segments will sweep to the right and
0483 0499 when only two or one segments remain
0475
u Do Not vent pressure. Observe display
Liability Statement of Bodyguard until the self check
then ‘press’ RH button. Displays
changes to Fig. 16 - starting the leak
Responsibility for reliable function of equipment transfers to the owner or operator when sequence automatically restarts and
test timing sequence.
equipment serviced, or repaired by untrained personnel, (not employed or authorised by when Fig. 5 is displayed – press LH
u Bar graph segments sweep to the right
Dräger) or when used in a manner not conforming to its intended use. button. Bar graph segments sweep to
and Icon ‘a’ Fig. 17 displayed at end of
the right and ‘open valve’ icon
timing sequence - equipment ‘Passed’
displayed. See Fig. 14.
test.
u Open the cylinder valve fully to boost
u Vent pressure from the system refit
Description PSS 100 ETD - While wearer is breathing the pressurise in the system. Display
‘Tally’ - Bodyguard will switch ‘Off’.
from equipment, it is possible to extended changes - Fig. 15 - actual pressure
The PSS 100 Series with incorporates a u Open cylinder valve slowly but fully to
displayed, alternating ‘close valve’ and
unique adjustable articulated carrying system. working time by connecting ChargAirTM to
6 17 28 ‘press RH button’ icons.
pressurise system. Bodyguard will
a high pressure secondary breathing air begin ‘self check’ display sequence.
PSS 100 ET pneumatics use a high source to refill cylinder/cylinders. 0476 0484 0498 u Immediately ‘Close’ the cylinder valve.
At end of sequence Fig. 13 displayed
performance first stage pressure reducer, Bar segments will sweep to the right and
showing the actual system pressure.
multi-function Electronic Monitoring Unit Technical Data when only two or one segments remain
(Bodyguard) and are compatible with a wide ‘press’ RH button. Display changes to
High Pressure Connection: Fig. 16 - starting leak test timing Putting on Facepiece
range of balanced piston lung demand
200bar or 300bar sequence.
valves facepieces and compressed air
Standard G5/8 as per DIN 477: Part 6. u Bar graph segments sweep to the right Note: Refer also to the Instructions for Use
cylinders. The cylinder strap provides secure
support for single cylinder configurations. and either; provided with the Facepiece.
Other connections available to National
Standards. Icon ‘a’ Fig. 17 displayed at the end of
PSS 100 ETD - The ChargAir TM the timing sequence - equipment Safety Warning: Facial hair, beard
connection allows refilling of cylinder(s), ‘Passed’ test. Continue to Whistle stubble, side whiskers, and the wearing
High Pressure Connection - 7 18
from an independent high pressure 29 Alarm Test. of spectacles will adversely affect and
(300bar) secondary breathing air supply ChargAirTM (If Fitted) 0684 0033 0090 or interfere with facepiece seal. Correct fit
source, while wearer is breathing from 300Bar maximum. Icon ‘b’ Fig. 17 again displayed. Vent of facepiece only ensured if facepiece
High Pressure Male Quick Release u Line up battery terminals and insert battery seal makes close contact with skin.
equipment. pressure from system refit ‘Tally’ to
Coupling. into battery compartment. Bodyguard
switch ‘Off’ Bodyguard. Investigate
The Bodyguard is a robust compact will emit single ‘Bleep’ and begin ‘self u Detach neck strap stud from centre
Full Technical Specification available from source of leak, rectify and repeat test.
instrument, replacing traditional mechanical check’. See Bodyguard Self Check strap of head harness. Spread harness
pressure gauge, whistle warning unit and Dräger. sequence, Fig’s 3, 4, 5 and 6. As each (Fig. 22) place chin into facepiece,
Distress Signal Unit (ADSU). Activated by display changes, Bodyguard emits Whistle Alarm Test
position harness over head locating
‘Tally’ (PSS 100 ETD) and system pressure Bodyguard single ‘bleep’. An X icon with ‘Fault Code’ u Icon ‘a’ Fig. 17 displayed at end of the
harness centre plate with back of head.
Pressure Connection - Electronic displayed in Fig.7 indicates Bodyguard timing sequence - equipment has
of respiratory protection equipment, this u Tighten both lower (1) then the upper
Pressure Sensor suitable for connection to has ‘failed self check’. Return ‘Passed’ test.
unique multi-function instrument straps (2) evenly towards back of the
8 u A and AE - Cover outlet of demand
incorporates pre-programmed features for reducer high pressure port of compressed 19 Bodyguard to Dräger Service.
valve with ball of hand, press centre of
head. Fig. 23. Only tighten the centre
continuous monitoring of: air respiratory protection equipment - u Check sealing gasket not damaged. strap (3) if necessary.
200bar or 300bar. 0501 0502
Relocate cover plate, and using three rubber cover, turning ‘On’ positive
u Air pressure of air containment system. u On achieving face seal, positive
Power Supply - Battery Type: screws and 2mm hexagon key pressure. Slowly vent system by
u Time remaining to whistle alarm setting pressure demand valves automatically
9 Volt Alkaline. secure cover plate (0.4Nm to 0.5Nm). carefully lifting ball of hand maintaining
(TTW). activate on inhalation. Breathe normally.
Battery supplied with equipment, but not Do Not overtighten. a slow decrease in pressure.
u Movement Sensor - Distress Signal
fitted. Actual operating time of battery u N - Slowly vent system. Carefully press
Unit (ADSU).
centre of rubber cover. Function Check
u Temperature. dependent on how long instrument has Wearer Height Adjustment
u During venting observe display - the u Inhale and hold breath. Unit must
u Battery Life. been operating, frequency of alarms, PSS 100 Series carrying system length can
whistle alarm (rapid ‘Bleep’) should balance, i.e. no audible leak.
ambient temperature and frequency of use be set to suit torso length of wearer.
sound at the pre-set pressure and u Continue breathing. Expired air should
A ‘Back Light’ feature allows illumination of of back light. With instrument ‘Off’ a small
continue down to 10bar. Continue to flow easily out of exhalation valve.
the liquid crystal display. amount of power still consumed from Position 1.........Short
slowly vent until the pressure and time u Press centre of rubber cover of demand
battery. Remove battery if instrument not Position 2.........Medium
used for a long period.
9 20 Position 3.........Long displays indicate ‘0’ bar. valve checking supple-mentary supply.
Available for the Bodyguard are ‘snap on’ u ‘Close’ cylinder valve.
0160 0471
To set follow these instructions. u A and AE - Cover outlet of demand
I.R. Link and Windows based Software u Breathe normally to vent system.
Display Accuracy - In the presence of RF valve with ball of hand, press centre of
Packages for programming of additional During venting observe Bodyguard.
Interference 150-400 Mhz, Pressure Refer to Fig. 8. rubber cover, turning ‘On’ positive
monitoring Options, such as User I.D. and Whistle alarm should sound at pre-set
Display accuracy +11bar -14bar at full pressure. Slowly vent system by
datalogging with downloading of u Lift equipment into vertical position and pressure. When display indicates zero
scale. (Normal Hand Held radio operating carefully lifting ball of hand maintaining
datalogged parameters. turn locking knob (90 degrees) to pressure, hold breath. Facepiece
frequency 440-480 Mhz). a slow decrease in pressure.
unlock shoulder yoke. u N - Slowly vent system. Carefully press should hold onto the face indicating a
Available from Dräger on request, are Compressed Air Cylinders u Slide shoulder yoke aligning required positive seal.
centre of rubber cover.
details of equipment variants, approved torso length with locking knob. Turn u Open cylinder valve, slowly but fully,
4L to 12L capacity cylinders (200bar or u During venting observe display - the
accessories and I.R. Link with Windows locking knob (90 degrees) to lock to pressurise system. Breathe
300bar) available in either steel or whistle alarm (rapid ‘Bleep’) should
based Software Package Options. shoulder yoke. Do Not press down on normally.
composite materials. Available for twin sound at the pre-set pressure and
cylinder mounting is an approved 10 21 the adjuster while rotating it. continue down to 10bar. Continue to
Intended Use conversion kit. Contact Dräger for details. 0156 0017 slowly vent until the pressure and time Note: If leak detected, readjust the head
The PSS 100 ET Series, used in approved Fitting Cylinder displays indicate ‘0’ bar. harness and retest.
lung demand valve, facepiece and cylinder Safety Warning: When using equipment u Check valve port and reducer u Refit ‘Tally’. Bodyguard will switch ‘Off’.
combinations, provides the wearer with with twin cylinder configuration, or handwheel threads are undamaged, u Following satisfactory whistle alarm u Remove ‘Tally’. Display momentarily
respiratory protection when working in single 12Litre composite, it is essential changes to Fig. 24, emitting short ADSU
connector O ring in position, and check, ensure the positive pressure
to consider possible restriction when alarm check, then to Fig. 25 -
contaminated, or oxygen deficient gaseous undamaged. mechanism (A and AE) is activated to
working in confined space. ‘operational mode’ - showing actual
atmosphere. The Bodyguard provides u Place backplate horizontal and extend switch ‘Off’ positive pressure.
continuous monitoring of the respiratory cylinder strap. pressure and time remaining to whistle
protection equipment, movement of wearer,
Preparation for Use u Slide cylinder through strap, locating Connecting Demand Valve to activation. Pass ‘Tally to controller.
and temperature. Fitting Battery to Bodyguard valve port to reducer handwheel.
Working duration of equipment is u Using 2mm hexagon key, remove three
Facepiece
u Lift the unit upright screw handwheel into Safety Warning: Do Not continue if ‘Low
dependent on capacity (volume) of cylinder screws from battery cover plate and 11 22 valve port (handtight) and if fitted hook anti- Note: Refer also to the Instructions for Use Battery’ warning (Fig. 12 or ‘Failed’ icon
selected, and breathing rate of wearer. carefully remove cover plate. Fig. 2. vibration strap to handwheel. (Fig. 9). provided with the Lung Demand Valve. (X) displayed.
0157 0016

3337448: 10th Edition: April 2004: Subject to Modification


PSS 100 Series PSS 100 ET, PSS 100 ETD - A, AE and N
Compressed Air Respiratory Protection Equipment Instructions for Use
In Use Note: - The ‘main alarm’ can only be switched u Turn ‘On’ secondary air supply to Indicated pressure after refill may decrease Routine Maintenance Cleaning, Disinfecting, u Remove the locknut, washers and the
reducer retaining screw.
u Regularly observe pressure and time
reading of instrument display. Whistle
‘Off’ by refitting the ‘Tally’. pressurise secondary ChargAir TM
hose. Remove protection cap from the
as the cylinder(s) cools to ambient
temperature. Remaining working time will
To be performed after Use. Drying u Unfasten hose loops, releasing hoses,
Panic Button female quick release coupling. be reduced accordingly. Carefully clean, disinfect, and thoroughly dry and remove reducer and hose
alarm sounds at pressure setting of See also Maintenance and Test Intervals
whistle and first three segments of bar u ChargAirTM supply hose ready for use. contaminated dirty components after use. assembly.
Should wearer require help and assistance, Chart.
graph will intermittently illuminate (Red). u Following pressure equalisation - When using baths to contain cleaning and
this functions as a manual Distress Signal
Note: If high pressure air leak detected disconnect quick release coupling by disinfecting solutions, immersed Note: Do Not immerse pressure reducer
u Go on to safe area, at latest when Unit (DSU).
whistle alarm sounds, by the shortest either at connections, or along the hose - pulling ‘grey’ sleeve, of female coupling, Removal of cylinder components and assemblies must be in cleaning or disinfecting solutions.
and safest route. close secondary air supply valve and vent away from male coupling. See Fig. 29. agitated manually. Draeger recommends
u To start ‘alarm’ press button in centre u Using clean cloth, moistened in
pressure from the hose. Replace the hose. Safety Warning: Cylinder valve should that no form of mechanical, electrical or
of instrument. Fig. 28 displayed. cleaning, or disinfecting solution,
Note: As coupling separates, a ‘hiss’ or be closed and system vented. ultrasonic agitation be used.
In Use Functions Connecting Secondary Air ‘pop’ may be heard as the non return valves
remove dirt and contaminants from the
Note: - The ‘main alarm’ can only be pneumatic assembly.
Display - Operating Mode switched ‘Off’ by refitting the ‘Tally’. Supply in both couplings seal off high pressure air. u Lift free end of cylinder strap against Safety Note: Refer to manufacturers’
usage instructions when using cleaning u Using clean cloth, moistened in clean
Figure 25 shows information displayed Camlock to release, loosen strap.
Note: ChargAir connection allows
TM
and disinfecting agents. It is important water, remove disinfecting residue,
during complete breathing cycle. Low Battery u Refit protection caps to the male and Unhook anti-vibration strap (if fitted)
refilling of cylinder(s), from an independent that attention be paid to concentration followed by drying.
System pressure continuously monitored The first low battery level displays the low female couplings. from the reducer handwheel and
and shown digitally (numeric) and as an high pressure (300bar) secondary u Recheck pressure gauge and continue. and reaction times. Do Not use organic
battery ‘icon’ in Fig. 12 and the unit emits a breathing air supply source, while wearer unscrew handwheel from the cylinder Note: Do Not exceed a temperature of 60
analogue bar graph. The lower numeric valve. solvents, such as Acetone, Alcohol,
beep every 5 seconds when in idle or active is breathing from equipment. If facepiece degrees Celsius when drying components.
value displays time remaining (available modes to warn that available power is getting and demand valve are not being worn, After Use - ChargAir TM
Slide the cylinder away from reducer White Spirit, Trichloroethylene or similar.
breathable air), to whistle alarm setting and remove from equipment. u Re-assemble reducer and hose
(TTW). Clock dial ‘icon’ indicates time.
low. ADSU mode can still be activated. ensure positive pressure mechanism of
lung demand valve (A and AE) activated
(If Fitted) u Recharge the cylinder. Dräger recommends the following: assembly to backplate and harness.
to switch ‘Off’ positive pressure. u Turn ‘Off’ the air supply at secondary
When this ‘icon’ is first displayed, it is
Available to the user, during the complete acceptable to safely use the compressed
air supply.
Charging Cylinders 1. Cleaning Lung Demand Valves
breathing cycle, are the following additional u Remove protection cap from the u Fit protection cap over the end of the u Dräger Safety Wash. Refer also to the Instructions for Use
air respiratory equipment for up to 2 hours. Recharge to the rated pressure indicated
functions. ChargAir TM male quick release female quick release coupling of the u Sekusept provided with the Lung Demand Valve.
secondary supply hose. on the label or stamped on the neck or
As an additional safety feature, the instrument coupling of the equipment.. Do Not exceed a temperature of 30
shoulder of cylinder. Dräger recommend a
Back Light will not allow ADSU mode to be activated after u Ensure the cylinder valve fully ‘Open’. u Vent air pressure from secondary
charge rate of 27bar/minute. Rapid degrees Celsius. Facepiece
u To illuminate display, press and release u Push male and female ChargAirTM supply hose by pressing the centre of Refer also to the Instructions for Use
a second low battery level is reached. This the end face of the protection cap. charging will induce an increase in
LH or RH button. Display illuminates for second level is the limit to enable full operational couplings together until a loud ‘click’ is Note: Before disinfecting, rinse off cleaning provided with the Facepiece.
u Unscrew secondary ChargAirTM supply temperature resulting in an incomplete
a predetermined time. use for a minimum of 1 hour achieving heard - indicating a secure connection. solution in clean water.
charge - check pressure at ambient and if
minimum sound levels for all alarms with the High pressure air from secondary hose connector from secondary air
supply outlet. required ‘Top Up’ charge. Storage - Ready for Use
Temperature full alarm signal sounding continuously. supply source begins to refill the 2. Disinfecting u Fully extend shoulder straps, waistbelt
equipment cylinder(s). u Incidur and head harness of facepiece.
u Press then release RH button. Time It is recommended that a pressure limiting
remaining changes to temperature (ºC). When the first limit is reached - After Use - Equipment device is fitted to the charging compressor Do Not exceed a temperature of 30 u Store the equipment ‘ready for use’ in
REPLACE BATTERY Note: A secure connection has been made degrees Celsius. a cool dry environment, free from dust
See Fig. 26. to prevent over charging of the selected
when the ‘dark green’ ring is visible behind Safety Warning: Do Not remove and dirt and away from excessive heat.
u Press then release RH button to change the ‘grey’ sleeve of the female coupling. cylinder.
equipment until in safe area, clear of
display back to ‘time remaining’. Refilling Cylinder(s) using See Fig. 29. hazard.
Note: Before drying, rinse off disinfecting
solution in clean water.
Protect the equipment from any
possible damage and ensure that
See Fig. 25. Safety Warning: Air quality for
ChargAirTM (If Fitted) u If the facepiece and the demand valve u Release side straps of head harness. compressed air breathing systems rubber parts are not subjected to direct
Note: After 5 seconds display changes should conform to requirements of u Dräger Wipex Cloths. sunlight.
Note: See For Your Safety - ChargAirTM are being worn - breathe normally u Ensure the positive pressure
from temperature back to ‘time remaining’ during refill. mechanism (A and AE) is activated to EN12021. Only charge compressed air Safety Note: When stored ‘ready for use’,
if the button is not pressed. pressure vessels which: Details of cleaning and disinfecting agents on wall or bulkhead mounting, e.g. Fire
Secondary Air Supply Source u Check reading on equipment pressure switch ‘Off’ the positive pressure.
are available from Dräger on request. Tender, ensure equipment secured and
Any system where compressed breathing gauge, and if fitted, gauge on secondary u Remove the facepiece. Close the
Motion Sensor supply hose, or secondary supply cylinder valve. u Conform to National Standards. supported by mounting system.
Functions as an Automatic Distress Signal
air is stored can be classified as a
gauge. When both gauges indicate the u Unbuckle waistbelt, lift shoulder strap u Feature original manufacturers test 3. Rinsing and Drying
secondary air supply source, such as: date and test mark. Remove cleaning and disinfecting
Unit (ADSU). same pressure - secondary supply buckles to loosen, remove equipment.
u Cascade system u Have not exceeded test date solutions by rinsing in clean water,
u If no movement detected, a ‘pre-alarm’ pressure and equipment cylinder(s)
u High pressure breathable air indicated on label affixed by last followed by drying.
sounds after 25 seconds - if movement pressure have equalised. Note: Do Not drop or throw down the
compressor with reservoir testing station. Do Not exceed a temperature of 60
detected within a further 8 seconds of equipment as damage could occur.
u High pressure compressed air cylinder degrees Celsius when drying
‘pre-alarm’ sounding - alarm Note: Dependent on pressure remaining
Visual Inspection components.
automatically cancelled. in the equipment cylinder before refill, u Press centre of rubber cover of demand
Preparation for use pressure equalisation occurs after valve to vent system until pressure and Check integrity of:
Note: Do Not attempt to use buttons to u Check secondary air supply outlet
Reducer and Hose Assemblies
approximately 45-60 seconds. Final time displays indicate ‘0’, and ensure
switch ‘Off’ the pre-alarm. thread, ChargAirTM connector thread equalised pressure governed by available positive pressure mechanism activated u Carrying system. Note: Refer to For Your Safety.
and sealing ring are not damaged. pressure and volume of air in secondary to switch ‘Off’ positive pressure. u All straps, buckles and harness.
u If no movement is detected at the end u Screw secondary ChargAirTM supply air supply. Due to rapid compression of the u Refit ‘Tally’. Bodyguard will switch ‘Off’. u Valves, connectors and manifold. u For cleaning and disinfecting, remove
of 8 second ‘pre-alarm’, a higher pitch hose connector to secondary air supply air in cylinder(s), during refill, a temperature u Pass the equipment to the Service u Lung Demand Valve. pressure reducer and hose assembly
‘main alarm’ starts. See Fig. 27. outlet - ensure a secure connection. increase occurs (approximately 10° C). Department. u Facepiece. from backplate.

FAULT, CAUSE REMEDY MAINTENANCE AND TEST INTERVALS


Carry out regular Inspection, Testing and Servicing of Equipment according to this table. Record all data in Equipment
Log Book. These instructions also apply to non-used (in Storage) Equipment.

FAULT POSSIBLE CAUSE REMEDY After Every Every Every


Description
Use Month Year 6 Years
Facepiece leakage Faulty or missing demand valve Fit or replace sealing ring or O-Ring Complete Clean and Disinfect !
(Leak test with connector O-Ring (push-in demand Equipment Visual Inspection ! "
equipment in valve) Functional and Leak Testing as defined in Instructions for Use ! "
position) Flow and Static Tests per Manufacturer’s instructions !
Exhalation valve leakage Clean and replace as necessary
Demand Valve Clean and disinfect as necessary !#

Speech diaphragm defective Replace Pressure Reducer Medium Pressure Check !


Replace Sintered Filter !
Headstraps not tight Tighten Headstraps Replace H.P. Connector O-ring !
Basic Overhaul
Unsatisfactory Speech diaphragm defective Replace !
(Repair Exchange Scheme)
communication Cylinder Charge to correct pressure !
High pressure leak Check tightness of connections Tighten or replace seals as necessary Charged Pressure Check. Check Test Date Stamped on
!"
Cylinder

Check hoses Replace as required Re-Certification - Cylinder Pressure Test (According to National
Standards).
Safety relief valve Check relief valve O-Ring Replace as necessary Cylinder Overhaul as necessary or at time of Cylinder Re-Certification.
venting Valve
Pressure reducer defective Return to Dräger Service
mDräger Recommendation UK
Demand valve Banjo O-Rings defective Replace uCOSHH
vents n Grease Demand Valve connector O-ring (Use Molykote 111) as required. (Type A - Push-In demand valve only).
(constant flow) Balanced piston leak Replace

Diaphragm incorrectly fitted Re-assemble correctly


Bodyguard will not Low Cylinder pressure Recharge cylinder
switch 'ON'
Low Battery Replace battery
Display showing Instrument Failure Disconnect Bodyguard from equipment.
fault codes Return Bodyguard to Dräger Service

Dräger Safety UK Limited


Dräger, while endeavouring to ensure correctness of Ullswater Close Dräger Safety AG & Co. KGaA
statements of fact and advice contained in this publication; Kitty Brewster Estate Revalstraße 1 Telefon: 0049451882-0
gives no guarantee or warranty in respect thereof, and Blyth D-23560 Telefax: 0049451882-2080
accepts no liability for any mis-statement or inaccuracy in Northumberland Tel: (01670) 352891 Lübeck
publication, or for any omission therefrom. Germany Internet http://www.draeger.com
NE 24 4 RG Fax: (01670) 356266

3337448: 10th Edition: April 2004: Subject to Modification


PA90N Lung Demand Valve
Normal Demand Instructions for Use
Approvals When using baths to contain cleaning and u Disconnect from air supply, dry
PA90N Series Lung Demand Valve, is approved for use with Dräger disinfecting solutions, immersed components components and, if necessary, using
compressed air respiratory protection equipment conforming to EN137 and assemblies must be agitated manually. Dräger Wipex Cloths, remove excess
Dräger recommends that no form of mechanical fluid.
and meeting the requirements of EC Council Directive 89/686/EEC, BSEN
133. Equipment is CE Mark Approved and issued with EC Type
agitation be used. uAfter drying, insert balanced piston
lever (4) into hole in centre boss of
Examination Certificate. Safety Note: Refer to manufacturers’ diaphragm (3) (chamfered side of
EC Type examiner: SGS United Kingdom Ltd, Unit 202b Worle Parkway, usage instructions when using cleaning boss towards lever) and carefully
Weston-super-Mare BS22 0WA, UK. Notified Body No. 0120. and disinfecting agents. It is important locate outer bead of diaphragm into
that attention be paid to concentration
groove in demand valve body - note
and reaction times. Do Not use organic
For Your Safety solvents, such as Acetone, Alcohol,
orientation. Fig. 3.
uUse of equipment requires, knowledge, and compliance with National White Spirit, Trichloroethylene or similar. uMoisten lip of bayonet cap (2) with clean
water - re-fit bayonet cap. Re-check
Regulations, Laws and Standards, governing the use of respiratory orientation of diaphragm.
protection equipment in the country of use. Dräger recommends only the following:
uAssemble rubber cover (1).
uUse of equipment requires wearer training and observance of these 1. Cleaning uConnect demand valve to equipment
Instructions for Use and Instructions for Use supplied with the u Dräger Safety Wash. and perform Pre Operational Checks.
compressed air respiratory protection equipment. Fig. 1 uSekusept
uUse equipment only for the purpose specified in this manual, or as 0033 Do Not exceed a temperature of 30 Replacing Diaphragm
degrees Celsius.
confirmed in writing by Dräger.
u Only trained competent personnel should inspect and service Refer to Fig. 2.
Note: Before disinfecting, rinse off cleaning
equipment at regular intervals and a record kept of such inspections solution in clean water.
uRemove rubber cover (1). Using plate
and service. spanner (R26817), unscrew bayonet
uDräger recommends a service contract be obtained from your Dräger 2. Disinfecting
u Incidur. cap (2).
Branch or Agent. Do Not exceed a temperature of 30 uCarefully unhook diaphragm (3) from
uContact Dräger for details of Service Contracts and Service Training degrees Celsius. lever (4) of balanced piston and discard
Courses. defective diaphragm.
uNotify Dräger if there is component fault or failure. Note: Before drying, rinse off disinfecting uInsert balanced piston lever (4) into hole
solution in clean water.
uUse only original Dräger Spare Parts for service and maintenance. in centre boss of new diaphragm (3)
uUse only Dräger Test Equipment for service and maintenance. (chamfered side of boss towards lever)
Details of cleaning and disinfecting agents
and carefully locate outer bead of
are available from Dräger on request.
diaphragm into groove in demand valve
Liability Statement body - note orientation. Fig. 3.
Responsibility for reliable function of equipment transfers to the owner 3. Rinsing and Drying
Remove cleaning and disinfecting solutions by uMoisten lip of bayonet cap (2) with clean
or operator when serviced, or repaired by untrained personnel, (not 4 3 2 1 rinsing in clean water, followed by drying. water - re-fit bayonet cap. Re-check
employed or authorised by Dräger) or when used in a manner not Fig. 2 0028
orientation of diaphragm.
conforming to its intended use. 0006 Do Not exceed a temperature of 60 uAssemble rubber cover (1).
degrees Celsius when drying components. uConnect demand valve to equipment
and perform Pre Operational Checks.
To internally clean and disinfect lung
demand valve it is important to follow these Fault, Cause, Remedy,
Description Preparation for Use instructions. (see Fig. 2). Service and Test Intervals
The high performance PA90N Lung Demand Connecting Demand Valve to u See attached and refer also to
Valve incorporates a unique pre-set and Equipment
uRemove rubber cover (1). Using plate
spanner (R26817), unscrew bayonet Instructions for Use provided with
sealed balanced piston unit. Should sealed uInsert male coupling of demand valve cap (2). compressed air respiratory protection
balanced piston unit be tampered with, into female coupling of equipment equipment.
Drägers’ guarantee is void. uCarefully unhook diaphragm (3) from
medium pressure hose. Check security
lever (4) of balanced piston.
of connection.
Lung Demand Valve - Part No. 3339409 is uRefer to Instructions for Use provided uAssemble rubber cover (1) to demand valve
for use with the following compressed air body to protect balanced piston lever.
with compressed air respiratory
respiratory protection equipment. protection equipment. uImmerse, and manually agitate demand
PA80/90 Series valve, diaphragm and bayonet cap in
PA90 Series Pre Operational Checks cleaning fluid. (Refer to cleaning fluid
PA90Plus Series instructions.)
uRefer to Instructions for Use provided
and Lung Demand Valve - Part No. 3339408
with compressed air respiratory Fig. 3 u After cleaning, rinse components in
is for use with the following compressed air clean water.
respiratory protection equipment. protection equipment. 0029
u Immerse, and manually agitate
PSS 100 Series Connecting Demand Valve to components in disinfecting fluid. (Refer
PSS 500 Series
Facepiece to disinfecting fluid instructions.)
uCheck facepiece port, and sealing ring, uRefer to Instructions for Use provided with Cleaning, Disinfecting, Drying uAfter disinfection, remove rubber cover
In the presence of chemicals; Some
clean and undamaged. compressed air respiratory protection (1) and rinse components in clean water.
substances, e.g. Phosgene, may be Carefully clean, disinfect, and thoroughly dry
uScrew demand valve handwheel into equipment. Take care not to damage balanced
absorbed by, or diffuse through, the contaminated dirty components after use.
facepiece port. Handtight. piston lever.
material of the silicone diaphragm of the
PA90 Series lung demand valve. uAlign demand valve (Fig. 1) and tighten. After Use To ensure correct operational condition of
u Connect hose on demand valve to
medium pressure air supply.
u See attached and refer also to equipment, use only cleaning and
When lung demand valve fitted with non Operation and Use Instructions for Use provided with disinfecting solutions recommended by uGently depress balanced piston lever
silicone diaphragm, fit ‘green’ cover (1 Fig. uConnect lung demand valve to compressed compressed air respiratory protection Dräger. Use of other products invalidates (4), blowing out fluid residue.
2) for identification. air respiratory protection equipment. equipment. Dräger’s guarantee. Safety Note: Wear suitable eye protection.

Maintenance and Test Intervals Fault - Cause - Remedy Order List

Lung After Every Description Order No


Description Fault Possible Cause Remedy
Demand Use Month PA90N Lung Dem and Valv e Norm al Dem and - Short Hose 3339408
Valve Demand Valve Banjo O-Rings Defective Replace O-Rings
Clean and Disinfect as m PA90N Lung Dem and Valv e Norm al Dem and - Long Hose 3339409
Vents -
necessary Constant Flow Wipex Cloths (Pack of 50) 3380375

Visual Inspection m u Balanced Piston Leak Replace Balanced Incidur Disinfectant (6 litres) 7904072
Piston
Incidur Disinfectant (30 litres) 7904073
Functional and Leak Test as m u Diaphragm incorrectly Reassemble
1 Litre Safety Wash and Dispenser 3380164
defined in Instructions for Use fitted correctly
supplied w ith respiratory 1 Litre Safety Wash Refill 3380165
protection equipm ent Perforated Diaphragm Replace Diaphragm
5 Litre Safety Wash and Dispenser 3380166
m Dräger Recommendation
uCOSHH 5 Litre Safety Wash Refill 3380167

Note: Refer also to Instructions for Use provided with compressed air respiratory Sekusept (4 bottles @ 2 litres) 7904071
equipment.
Diaphragm - PA90 Series Lung Dem and Valv es 3339452

Rubber Cov er - Green 3339450

Non-Silicon Diaphragm 3339761

Dräger Safety UK Limited Dräger Safety AG & Co. KGaA


Dräger, while endeavouring to ensure correctness of
statements of fact and advice contained in this publication;
Ullswater Close Revalstraße 1 Telefon: 0049451882-0
gives no guarantee or warranty in respect thereof, and Kitty Brewster Estate D-23560 Telefax: 0049451882-2080
accepts no liability for any mis-statement or inaccuracy in Blyth Tel: (01670) 352891 Lübeck
publication, or for any ommission therefrom. Northumberland NE 24 4 RG Fax: (01670) 356266 Germany Internet http://www.draeger.com

3337394: 7th Edition: August 2004: Subject to Modification - (A3-D-P)


PSS 100 Series Spare Parts List 1287.2
Pneumatic Arrangement
with ChargAirTM DIN (300Bar)
Gauge or Bodyguard
Page 1 of 2

18

17

16
8
15
14

19,20
13

7
6
5
12 12
11
11
10
10
9

4(x4) 3(x2) 2(x2) 1,1a

NPR 0810
DCC 2482
12872
Ordering Details:
No. of parts required = 16; Edition No: 3
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: July 1999
PSS 100 Series Spare Parts List 1287.2
Pneumatic Arrangement
with ChargAirTM DIN (300Bar)
Gauge or Bodyguard
Page 2 of 2

Order Pack.
Item No Description
Code Qty.

1 Pres. Red./ChargAir™ DIN (300bar). See 1285.15 3337258

1a Pres. Red./ChargAir™DIN (300bar). R.E.X. 3337264

2 Retention Staple 3337101 10

3 Retention Cap 3337102 10

4 Button Head Screw 3337103 10

5 O-Ring 3337403 20

6 O-Ring Retainer 3337105 10

7 Spring 3337106 10

8 Medium Pressure Hose 3338316

9 Anti-Vibration Strap R50792 4

10-15 Gauge Hose and Whistle. See 1285.13 3337316

10 O-Ring 3337204 20

11 Back-Up Ring 3337205 20

12 O-Ring 3337314 10

14 Grub Screw 3337305 20

15 Screw 3338959 10

15,16 Gauge Cover 3337307

17 O-Ring 3337310 10

18 Pressure Gauge 3337303

10-12,19 Bodyguard (ET) Tally. See 1289.1 3338080

10-12,20 Bodyguard (E) Button. See 1289.1 3338400

Ordering Details:
No. of parts required = 16; Edition No: 3
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: July 1999
PSS® 100 and DrägerMan PSS® 100 Series Spare Parts List 1287.1
Pneumatic Arrangement
Gauge, Bodyguard and DrägerMan
Bodyguard II
Page 1 of 3

21,22

20

19
18 11
17
25,26
23,24
16

10
9
15 15 8
15
14 14 14
13 13 13
12

7(x4) 6(x2) 5(x2) 1,1a 2,2a


3,3a, 4,4a

NPR 1049
NPR 0810
DCC 2774
12871
Ordering Details:
No. of parts required = 16; Edition No: 5
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: June 2000
PSS® 100 and DrägerMan PSS® 100 Series Spare Parts List 1287.1
Pneumatic Arrangement
Gauge, Bodyguard and DrägerMan
Bodyguard II
Page 2 of 3

Order Pack.
Item No Description
Code Qty.

1 Pressure Reducer - DIN. See 1285.14 3337200

1a Pressure Reducer - DIN. Repair Exchange 3337260

2 Pressure Reducer - France. See 1285.14 3337251

2a Pressure Reducer - France. Repair Exchange 3337262

3 Pressure Reducer - Japan. See 1285.14 3337252

3a Pressure Reducer - Japan. Repair Exchange 3337263

4 Pressure Reducer - Italy. See 1285.14 3337253

4a Pressure Reducer - Italy. Repair Exchange 3337261

5 Retention Staple 3337101 10

6 Retention Cap 3337102 10

7 Button Head Screw 3337103 10

8 O-Ring 3337403 20

9 O-Ring Retainer 3337105 10

10 Spring 3337106 10

11 Medium Pressure Hose 3338316

12 Anti-Vibration Strap R50792 4

13-21 Gauge and W.W.U. Assembly Cpl. (Bar) 3337300

13-18 Gauge Hose and Whistle. See 1285.13 3337316

13 O-Ring 3337204 20

14 Back-Up Ring 3337205 20

15 O-Ring 3337314 10

17 Grub Screw 3337305 20

18 Screw 3338959 10

18-19 Gauge Cover 3337307

Ordering Details:
No. of parts required = 16; Edition No: 5
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: June 2000
PSS® 100 and DrägerMan PSS® 100 Series Spare Parts List 1287.1
Pneumatic Arrangement
Gauge, Bodyguard and DrägerMan
Bodyguard II
Page 3 of 3

Order P ack.
Item N o D escription
C ode Qty.

20 O-Ri ng 3337310 10

21 Pressure Gauge 3337303

22 Pressure Gauge (Mpa) 3350238

13-15,23 Bodyguard (ET) Tally. See 1289.1 3338080

13-15,24 Bodyguard (E) Button. See 1289.1 3338400

13-15,25 Bodyguard II. See 1289.11 3351426

13-15,26 Bodyguard II - T. See 1289.11 3351425

Ordering Details:
No. of parts required = 16; Edition No: 5
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: June 2000
Medium Pressure Test Gauge Spare Parts List 1282.5

Page 1 of 2

3
4

DCC 2784
12825
Ordering Details: Edition No: 3
No. of parts required = 16;
Packaged Quantity = 5; Order Quantity = 20
Subject to change without notice/Änderungen vorbehalten
Date: January 1999
Order in multiples of packaged quantity
Medium Pressure Test Gauge Spare Parts List 1282.5

Page 2 of 2

Order Pack.
Item No Description
Code Qty.

1-4 Low Pressure Test Gauge c/w Quick Release Coupling 3331771

3-5 Low Pressure Test Gauge c/w Broom Wade Coupling 3331214

1 Male Coupling - Quick Release Coupling 3331740

2 Adaptor 3331900

3 Gauge 3331899

4 Gauge Cover 3330106 5

5 Male Coupling - Broom Wade 3331215

Ordering Details: Edition No: 3


No. of parts required = 16;
Packaged Quantity = 5; Order Quantity = 20
Subject to change without notice/Änderungen vorbehalten
Date: January 1999
Order in multiples of packaged quantity
High Pressure Check Gauge Spare Parts List 1281.5

Page 1 of 2

8 9

6 5 4 3 2 1

DCC 5222
12815
Ordering Details:
No. of parts required = 16; Edition No: 3
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: February 2005
High Pressure Check Gauge Spare Parts List 1281.5

Page 2 of 2

Order Pack.
Item No Description
Code Qty.

1-9 High Pressure Check Gauge 3331209

1 O-Ring D17409 10

2 Connector 3333033

3 Handw heel R27380

4 Rubber Ring R20620

5 O-Ring 3330953 10

6 Relief Valve Assembly 3331254

7 Washer 3334797 10

8 Gauge 3339388

9 Cover 3339382

Ordering Details:
No. of parts required = 16; Edition No: 3
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: February 2005
PSS 100 Series Spare Parts List 1287
Single Cylinder Strap

Page 1 of 4

35(x4)
26(x2)

27(x2)
20
4

3
25 24 23 22
21
5(x2)
36

28(x2)
18
37,38 29
34
30
31 32
33(x2) 19(x2)
16(x2) 8
9

18 13
14
15(x2)
12
11(x2)

DCC 4081
17(x2) 7 6 10 DCC 2997
DCC 4080
1287
Ordering Details:
No. of parts required = 16; Edition No: 5
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: May 2000
PSS 100 Series Spare Parts List 1287
Single Cylinder Strap

Page 2 of 4

Order Pack.
Item No Description
Code Qty.

1-37 PSS 100 3337664

1-36,38 PSS 100D 3337665

1-37 PSS 100 (France) 3338069

1-37 PSS 100 (Japan) 3338070

1-37 PSS 100 (Italy) 3338071

1-37 PSS 100 E 3338168

1-37 PSS 100 E France 3338064

1-37 PSS 100 E (Japan) 3338065

1-37 PSS 100 E (Italy) 3338066

1-37 PSS 100 ET 3338072

1-37 PSS 100 ET (France) 3338074

1-37 PSS 100 ET (Japan) 3338075

1-37 PSS 100 ET (Italy) 3338076

1-36, 38 PSS 100 ETD 3338073

1-36,38 PSS 100 ED 3338169

1 Backplate 3338268

2 Cylinder Cradle 3338269

3 Cylinder Strap (Buckle End) 3338267

4 Cylinder Strap (Loose End) 3337437

5 Pin 3338264 2

6 Backplate Boot 3338289

7 Pivot Plate 3338207

8 Low er Pivot 3338202

9 Pivot Bumper 3338205

Ordering Details:
No. of parts required = 16; Edition No: 5
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: May 2000
PSS 100 Series Spare Parts List 1287
Single Cylinder Strap

Page 3 of 4

Order Pack.
Item No Description
Code Qty.

10 Pivot Pin 3338204

11 Nut (M4) 3331367 10

12 Upper Pivot 3338201

13 Spring Plate 3338247 2

14 Retaining Block 3338203 2

15 Screw (M4) 3338219 10

16 Waistbelt Pad 3338215 2

17 Wing Strap Assembly 3338228 2

18 Adjusting Belt - Pair 3337098

19 Anchor Peg 3338236 10

20 Yoke Sub Assembly 3337434

21 Yoke Screw 3338246 2

22 Washer 3337312 20

23 Yoke Adjuster 3338245 2

24 Disc Spring 3337430 10

25 Yoke Nut 3338248 2

26 Shoulder Pad Assembly 3337425 2

27 Adjusting Straps 3338223 2

28 Anchor Peg 3338016 10

29 Mounting Bracket 3338655 5

30 Reducer Block 3339082

31-34 Rep. Set Bushing R23582

31 Bolt

32 Spacer R50305 10

Ordering Details:
No. of parts required = 16; Edition No: 5
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: May 2000
PSS 100 Series Spare Parts List 1287
Single Cylinder Strap

Page 4 of 4

Order Pack.
Item No Description
Code Qty.

33 Washer

34 Lock Nut 1333062 50

35 Hose Clip 3337088 5

36 Rubber Bung 3340539

37 Pneumatic Arrangement. See 1287.1

38 Pneumatics w ith ChargAir. See 1287.2

Ordering Details:
No. of parts required = 16; Edition No: 5
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: May 2000
PSS 100 Series Spare Parts List 1287.3
Tool List

Page 1 of 2

1 8

9-11

12,13

14-15

4 16

17
5,6

18-19

DCC 5222
DCC 4966
20,21 Not Illustrated
DCC 4731
12873
Ordering Details:
No. of parts required = 16; Edition No: 4
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: March 2005
PSS 100 Series Spare Parts List 1287.3
Tool List

Page 2 of 2

Order Pack.
Item No Description
Code Qty.

1-13 PSS 100 Series - Basic Tool Kit

1 Tool Bag 3332196

2 Plate Spanner R26817

3 Fine Point Pliers 3333968

4 Posi Drive Screw driver No. 2 Point 3331245

5 Screw driver - 3mm DIN 3336222

6 Screw driver 3331246

7 Circlip Assembly Tool 3335972

8 Fine Point Tw eezers 3333326

9 2mm AF Hexagon Socket Key 3337900

10 1.5mm AF Hexagon Socket Key 3331237 10

11 3mm AF Hexagon Socket Key 3335737

12 O-Ring Assembly Tool (Gauge) 3337901 10

13 O-Ring Assembly Tool (HP Connector) 3310465

14 10/11mm AF Spanner 3336309 10

15 14/15mm AF Spanner 3336308

16 O-Ring Lifter R21402

17 M oly kote 111 Grease (100 grams) 3331247

18 Loctite 245 3335722

19 Loctite 638 3335389

20 High Pressure Check Gauge (not illustrated) 3331209

M edium Pressure Check Gauge (not


21 3331771
illustrated)

Ordering Details:
No. of parts required = 16; Edition No: 4
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: March 2005
Plus Series Spare Parts List 1285.18
Male Decontamination Hose
3337652 (300mm)
3350588 (400mm)
Page 1 of 2

1
2

9,10

5(x2)

DCC 5360
NPR 0903
128518
Ordering Details:
No. of parts required = 16; Edition No: 4
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: October 2003
Plus Series Spare Parts List 1285.18
Male Decontamination Hose
3337652 (300mm)
3350588 (400mm)
Page 2 of 2

Order Pack.
Item No Description
Code Qty.

1-9 Decontamination Hose (300mm) 3337652

1-8,10 Decontamination Hose (400mm) 3350588

1 Non Return Valve 3337616 2

2 O-Ring 3337611 10

3 Retention Staple 3337101 10

4 Retention Cap 3337102 10

5 Screw 3337103 10

6 Tailpiece 3335213

7,8 Q.R.C. (NRV) 3331741

8 Protection Cap 3335757

9 300mm Assy. (Not Available)

10 400mm Assy. (Not Available)

Ordering Details:
No. of parts required = 16; Edition No: 4
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: October 2003
PA90Plus Series Spare Parts List 1285.17
Second Man (LDV) Attachment
3337650 (QRC), 3310106 (QRC)
3353559 (QRC), 3339665 (B&W)
Page 1 of 2

1
2

6,7,8

10

DCC 5438
NPR 1379
NPR 1339
ECN 2324
128517
Ordering Details:
No. of parts required = 16; Edition No: 7
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: January 2004
PA90Plus Series Spare Parts List 1285.17
Second Man (LDV) Attachment
3337650 (QRC), 3310106 (QRC)
3353559 (QRC), 3339665 (B&W)
Page 2 of 2

Order Pack.
Item No Description
Code Qty.

1-6 Second Man Attachment, (300mm) - QRC 3337650

1-5, 7 Second Man Attachment, (390mm) - QRC 3350106

1-5, 8,9 Second Man Attachment, (985mm) - QRC 3353559

1-5, 10 Second Man Attachment, Connector (B&W) 3339665

1 Non Return Valve 3337616 2

2 O-Ring 3337611 10

3 Retention Staple 3337101 10

4 Retention Cap 3337102 10

5 Button Head Screw 3337103 10

9 Protection Cap (not illustrated) 3334048 2

Ordering Details:
No. of parts required = 16; Edition No: 7
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: January 2004
Whistle Tube and HP Hose Spare Parts List 1285.13
3337316 (985mm)
3353639 (730mm)
3353640 (650mm)
Page 1 of 2

3
7

10
1,1a,1b

4,4a,4b

16 10

11 3

2
12
13 NPR 1414
14 NPR 1413
15 DCC 4332
DCC 4079
DCC 4244
128513
Ordering Details:
No. of parts required = 16; Edition No: 9
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: March 2004
Whistle Tube and HP Hose Spare Parts List 1285.13
3337316 (985mm)
3353639 (730mm)
3353640 (650mm)
Page 2 of 2

Order Pack.
Item No Description
Code Qty.

1,2-4,5-16 Whistle Tube and Gauge Hose (985mm) 3337316

1-15 Whistle Tube and Gauge Hose (730mm) 3353639

1-15 Whistle Tube and Gauge Hose (650mm) 3353640

1 Hose Assembly (985mm) 3337317

1a Hose Assembly (730mm) 3353908

1b Hose Assembly (650mm) 3353902

2 O-Ring 3337204 20

3 Back-Up Ring 3337205 20

4 Capillary Assembly (985mm) 3337330

4a Capillary Assembly (730mm) 3353913

4b Capillary Assembly (650mm) 3353914

5 Filter 3337340 10

6 O-Ring 3337309 10

7 Whistle Jet 3337302

8 Screw 3337305 20

9 Screw (Gauge Cover w /o hole) 3336438 20

10 Screw (Gauge Cover w ith hole and Sentinel) 3338959 10

11 O-Ring 3337313 10

12 O-Ring 3337314 10

13 Filter 3337350 30

14 Washer 3337312 20

15 Circlip 3337306 20

16 Retaining Washer - Sentinel 3338861 10

Ordering Details:
No. of parts required = 16; Edition No: 9
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: March 2004
PSS 100 Series Spare Parts List 1289.1
Electronic Monitoring Unit (Bodyguard)
3338080 - Tally (PSS 100 ET)
3338400 - Button (PSS 100 E)
Page 1 of 2

13

14 11

12

4
3
2
10

15

9(x3) 7

s1289
Ordering Details:
No. of parts required = 16; Edition No: 5
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: October 2000
PSS 100 Series Spare Parts List 1289.1
Electronic Monitoring Unit (Bodyguard)
3338080 - Tally (PSS 100 ET)
3338400 - Button (PSS 100 E)
Page 2 of 2

Item Pack.
Description Order Code
No Qty.
1-14 Electronic Monitoring Unit Tally - PSS 100 ET 3338080

1-12 Electronic Monitoring Unit Button - PSS 100 E 3338400

1 Not Available

2 O-Ring 3337314 10

3 Back-Up Ring 3337205 20

4 O-Ring 3337204 20

5,6,7 Battery Cover 3338688

6 Label (Not Illustrated) 3338498 5

7 Foam Adhesive Strip 3338469 10

8 Gasket Battery Cover 3338413 2

9 Button Head Screw 3337103 10

10 Rubber Protective Cover 3338408

11 Protective Cover - Display 3338410

12 Rubber Ring 3337667 4

13,14 Tally Assembly 3338407

14 Split Ring 3334578 10

15 Battery 3338471

Ordering Details:
No. of parts required = 16; Edition No: 5
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: October 2000
PA90Plus Series - CABE Spare Parts List 1285.16
Basic Customer Tool List
3337915

Page 1 of 2

1 t0016 12 t0034

2 t0008 13-14 t0002

3 t0006 15 t0014

4,5 t0005 16 t0031

6 t0004 17 t0032

7 t0056 18,19 t0015

8 t0007
20 t0034

21-23 not illustrated DCC 5222


DCC 5485
DCC 5210
9-11 t0023
DCC 4966

Ordering Details: Edition No: 7


No. of parts required = 16;
Packaged Quantity = 5; Order Quantity = 20
Subject to change without notice/Änderungen vorbehalten
Date: March 2005
Order in multiples of packaged quantity
PA90Plus Series - CABE Spare Parts List 1285.16
Basic Customer Tool List
3337915

Page 2 of 2

Order Pack.
Item No Description
Code Qty.

1-14 PA90Plus Series - Basic Tool Kit 3337915

1 Tool Bag 3332196

2 Plate Spanner R26817

3 Fine Point Pliers 3333968

4 Screw driver PosiDriv (No.1) 3310759

5 Posi Driv Screw driver No. '0' Point 3310500

6 Screw driver - 3mm DIN 3336222

7 Circlip Assembly Tool 3335972

8 Fine Point Tw eezers 3333326

9 2mm AF Hexagon Socket Key 3337900 10

10 1.5mm AF Hexagon Socket Key 3331237 10

11 3mm AF Hexagon Socket Key 3335737 10

12 O-Ring Assembly Tool (Gauge) 3337901

13 10/11mm AF Spanner 3336309

14 14/15mm AF Spanner 3336308

15 Molycote 111 Grease 3331247

16 Torque Driver 3310593

17 Torque Adaptor (2mm) 3310600

18 Loctite 245 3335722

19 Loctite 290 3333620

20 O-Ring Assy Tool (HP) 3310465

21 High Pressure Check Gauge 3331209

22 Medium Pressure Check Gauge 3331771

23 Medium Pressure Hose Assembly 3337650

Ordering Details: Edition No: 7


No. of parts required = 16;
Packaged Quantity = 5; Order Quantity = 20
Subject to change without notice/Änderungen vorbehalten
Date: March 2005
Order in multiples of packaged quantity
PA90 Series Spare Parts List 1284.2
Lung Demand Valve
PA90A, PA90A (German Version)
PA90AE, PA90N
Page 1 of 5

14,
15
13

12

16
17,
18,
17a,
3 18a

1
2a

4
2
7
7

10
8 4

5
11

9
6

36 38 37

DCC 5718
DCC 4416
DCC 2631
12842
Ordering Details:
No. of parts required = 16; Edition No: 13
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: November 2004
PA90 Series Spare Parts List 1284.2
Lung Demand Valve
PA90A, PA90A (German Version)
PA90AE, PA90N
Page 2 of 5

35, 35a

32
31
35,35a
30
32
29
34
27
33
28
27

25
19,20
27

26 23

25

24

21 22 21

12842a
Ordering Details:
No. of parts required = 16; Edition No: 13
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: November 2004
PA90 Series Spare Parts List 1284.2
Lung Demand Valve
PA90A, PA90A (German Version)
PA90AE, PA90N
Page 3 of 5

Pack.
Item No Description Order Code
Qty.
1-6 ,12-14 ,16,17,19, 21-32,37 PA90A PP (Blue Cover) 3339400

1-6,12-1416,17, 2032,37 PA90A PP to suit PA80 -Blue Cover 3339402

1-6,12-13, 15-17,19, 21-32,37 PA90A PP (Germ. Version - Red ) 3339404

1-6,12,13, 15-17,20,32,37 PA90A PP to suit PA80 (Germ, Version - Red) 3339407

1,2a,3,4,7-9,12,13,14,16-17,19,21-32,38 PA90AE PP M45 (Red Cover) 3339405

1,2a,3,4, 7-9,12,13, 15-17, 21-32,38 PA90AE PP M45 to suit PA80 (Red Cover) 3339406

1,2a,3,4,710-14,18, 19,21-27, 32-36 PA90N ND M40 DIN (Blue Cover) 3339408

1,2a,3,4,7,10-13,15,18,20-27,32-36 PA90N ND M40 DIN to suit PA80 (Blue) 3339409

1 Demand Valve Body 3339481

2 Anti-Injector Pipe (PA90A) 3339460

2a Anti-Injector Pipe (PA90AE + N) 3339818

3 Screw 3339483 10

4 O-Ring 3339459 10

5,6 Connector (Type A) 3339458

6 O-Ring R18352 5

7  Circlip 3339461 5

8  Handw heel (M45) R50884

9 Handw heel Retainer (M45) 3339469

10 Handw heel (M40) R28007

11 Handw heel Retainer (M40) 3339445

12 Diaphragm 3339452

13 Bayonet Cap 3339454

14 Rubber Cover (Blue) 3339453

Ordering Details:
No. of parts required = 16; Edition No: 13
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: November 2004
PA90 Series Spare Parts List 1284.2
Lung Demand Valve
PA90A, PA90A (German Version)
PA90AE, PA90N
Page 4 of 5

Pack.
Item No Description Order Code
Qty.
15 Rubber Cover (Red) 3339451

16 Spring 3339419 10

16,17 Balanced Piston Assy (PP) 3339410

16,17a Balanced Piston Assy (PP) Rep + Exch. 3339425

18 Balanced Piston Assy (ND) 3339426

18a Balanced Piston Assy (ND) Rep + Exch. 3339427

19,21-25 Hose Assy (PA90 Series) 3339437

20,21-25 Hose Assy (PA90 Series) to suit PA80 3339438

21 Ferrule 20

22 Hose

23 Male Coupling

24 Elbow

25 O-Ring R50117 10

26 Banjo 3339449

27 O-Ring 3339440 10

28 Arm (PP) 3339462

29 Spring 3339465 5

30 Lever 3339463 2

31 Washer (PP) 3339485 20

32 Circlip 3339470 20

33 Banjo End Blank ND 3339447

34 Washer ND 3339489 20

35 End Cap Assy (Locking) 3350048 10

35a End Cap 3339448 20

Ordering Details:
No. of parts required = 16; Edition No: 13
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: November 2004
PA90 Series Spare Parts List 1284.2
Lung Demand Valve
PA90A, PA90A (German Version)
PA90AE, PA90N
Page 5 of 5
Pack.
Item No Description Order Code
Qty.
36 Protection Cap (M40) 3339069 10

37 Protection Cap (Push-In) 3351069 10

38 Protection Cap (M45) R52043 10

Ordering Details:
No. of parts required = 16; Edition No: 13
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: November 2004
Plus and PSS Series LDV Spare Parts List 1285.007
Connector Assemblies
3338791 - M45
3338794 - M40 Page 1 of 2

5 4 2 3(x3) 6 7 1 8

5 4 9 3(x3) 2 1 10

1285007
Ordering Details: Edition No: 1
No. of parts required = 16;
Packaged Quantity = 5; Order Quantity = 20
Subject to change without notice/Änderungen vorbehalten
Date: March 2000
Order in multiples of packaged quantity
Plus and PSS Series LDV Spare Parts List 1285.007
Connector Assemblies
3338791 - M45
3338794 - M40 Page 2 of 2

Order Pack.
Item No Description
Code Qty.

1-8 Connector Assy - M45 3338791

1-5,9,10 Connector Assy - M40 3338794

1 Quad Ring 3350595 10

2 Connector - Front 3350594

3 Ball Bearing 3338792 15

4 Circlip 3338823 5

5 O-Ring 3339459 10

6 Sleeve 3338789

7 Handw heel - M45 R50884

8 Connector - Back (M45) 3338788

9 Handw heel - M40 R28007

10 Connector - Back (M40) 3338793

Ordering Details: Edition No: 1


No. of parts required = 16;
Packaged Quantity = 5; Order Quantity = 20
Subject to change without notice/Änderungen vorbehalten
Date: March 2000
Order in multiples of packaged quantity
PSS Series LDV Spare Parts List 1285.006
Screw-In - M40 (Rd 40)
3338709
Page 1 of 2

18

17

16

15

9 3

8 2
10
11
1
13

14
4

12
6

19

DCC 4587
DCC 4579
NPR 1001
1285003
Ordering Details:
No. of parts required = 16; Edition No: 3
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: July 2001
PSS Series LDV Spare Parts List 1285.006
Screw-In - M40 (Rd 40)
3338709
Page 2 of 2

Order Pack.
Item No Description
Code Qty.

1-19 Plus N LDV - Short Hose - Straight 3350501

1 Body 3350204

2 Button Insert 3338926

3 Grommet 3338750 5

4 ATB Assy. See 1285.005 3350551

5 Injector - Preset 3339685

6 O-Ring 3339459 10

7 Connector Assy. See 1285.007 3338794

8 Locking Arm 3350098 5

9 Screw 3350526 10

10 O-Ring 3337215 20

11 O-Ring 3339440 10

10-12 MP Hose (Short) 3350499

13 Spacer 3338733 10

14 Clip 3338737 10

15 Diaphragm 3338747

16 Slip Ring 3338755 5

17 Bayonet Cap 3338735

18 Rubber Cover - Black 3350677

19 Protection Cap 3339069 10

Ordering Details:
No. of parts required = 16; Edition No: 3
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: July 2001
Plus and PSS Series LDV Spare Parts List 1285.005
Air Transfer Block and
Balanced Piston Unit Assemblies
Page 1 of 2

1,2

4 5 6

DCC 5480
1285005
Ordering Details: Edition No: 1
No. of parts required = 16;
Packaged Quantity = 5; Order Quantity = 20
Subject to change without notice/Änderungen vorbehalten
Date: March 2000
Order in multiples of packaged quantity
Plus and PSS Series LDV Spare Parts List 1285.005
Air Transfer Block and
Balanced Piston Unit Assemblies
Page 2 of 2

Order Pack.
Item No Description
Code Qty.

1,4-6 ATB Assembly - A and AE 3350550

2,4-6 ATB Assembly - A w ith By-Pass 3350552

3-6 ATB Assembly - N 3350551

1 ATB - A and AE 3350553

2 ATB - A w ith By-Pass 3350555

3 ATB - N

4 O-Ring 3338763 10

4-5 Balanced Pston Assembly 3350556

6 O-Ring 3338776 10

Ordering Details: Edition No: 1


No. of parts required = 16;
Packaged Quantity = 5; Order Quantity = 20
Subject to change without notice/Änderungen vorbehalten
Date: March 2000
Order in multiples of packaged quantity
Plus and PSS Series LDV Spare Parts List 1285.003
Screw-In - M40 (Rd 40)
3350501 3338969 N (H2S)
3338710 3338711
Page 1 of 3

21,22

20

19

18

9 3

8 2
10
11
1
15

16
4

12,13,14
6

23

NPR 1001
1285003
Ordering Details:
No. of parts required = 16; Edition No: 2
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: April 2001
Plus and PSS Series LDV Spare Parts List 1285.003
Screw-In - M40 (Rd 40)
3350501 3338969 N (H2S)
3338710 3338711
Page 2 of 3

Order Pack.
Item No Description
Code Qty.

1-12,15-17,19-21,23 Plus and PSS N LDV - Short Hose 3350501

1-11,13,15-17,19-21,23 Plus N LDV - Long Hose 3338710

1,11,14-17,19-21,23 Plus N LDV - w ith MP Connection 3338711

1-12,15,16,18-20,22,23 Plus N LDV - Short Hose (H2S) 3338710

1 Body 3350204

2 Button Insert 3338926

3 Grommet 3338750 5

4 ATB Assy. See 1285.005 3350551

5 Injector - Preset 3339685

6 O-Ring 3339459 10

7 Connector - Assy. See 1285.007 3338794

8 Locking Arm 3350098 5

9 Screw 3350526 10

10 O-Ring 3337215 20

11 O-Ring 3339440 10

10-12 MP Hose (Short) 3350499

10,11,13 MP Hose (Long) 3338716

10,11,14 MP Hose (w ith MP Connection) 3338717

15 Spacer 3338733 10

16 Clip 3338737 10

17 Diaphragm - Silicone 3338747

18 Diaphragm - Polymer (H2S) 3338751

19 Slip Ring 3338755 5

20 Bayonet Cap 3338735

Ordering Details:
No. of parts required = 16; Edition No: 2
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: April 2001
Plus and PSS Series LDV Spare Parts List 1285.003
Screw-In - M40 (Rd 40)
3350501 3338969 N (H2S)
3338710 3338711
Page 3 of 3

Order Pack.
Item No Description
Code Qty.

21 Rubber Cover - Blue 3338741

22 Rubber Cover - Green (H2S) 3338799

23 Protection Cap 3339069 10

Ordering Details:
No. of parts required = 16; Edition No: 2
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: April 2001
Plus and PSS Series LDV Spare Parts List 1285.002
Screw In - M45
3338706 3338968 AE (H2S)
3338707 3338708 Page 1 of 3

21,22

20

19
18

16,17

8 2
7

9
10
1
14

15

4
11,12,13

23
NPR 1458
NPR 1002
DCC 4416
1285002
Ordering Details: Edition No: 4
No. of parts required = 16;
Packaged Quantity = 5; Order Quantity = 20
Subject to change without notice/Änderungen vorbehalten
Date: April 2004
Order in multiples of packaged quantity
Plus and PSS Series LDV Spare Parts List 1285.002
Screw In - M45
3338706 3338968 AE (H2S)
3338707 3338708 Page 2 of 3

Order Pack.
Item No Description
Code Qty.

1-11,14-16,18-21,23 Plus and PSS AE LDV - Short Hose 3338706

1-10,12,14-16,18-21,23 Plus AE LDV - Long Hose 3338707

1,10,13-16,18-21,23 Plus AE LDV - w ith MP Connection 3338708

1-11,14,15,17-20,22,23 Plus AE LDV - Short Hose (H2S) 3338968

1 Body 3350204

2 Grommet 3338750 5

3 ATB Assy. See 1285.005 3350550

4 Injector - Preset 3353962

5 O-Ring 3339459 10

6 Connector - Assy. See 1285.007 3338791

7 Locking Arm 3350098 5

8 Screw 3350526 10

9 O-Ring 3337215 20

10 O-Ring 3339440 10

9-11 MP Hose (Short) 3350499

9,10,12 MP Hose (Long) 3338716

9,10,13 MP Hose (w ith MP Connection 3338717

14 Spacer 3338733 10

15 Clip 3338737 10

16 Diaphragm - Silicone 3338747

17 Diaphragm - Polymer (H2S) 3338751

18 Slip Ring 3338755 5

19 Spring 3338768 5

20 Bayonet Cap 3338735

Ordering Details: Edition No: 4


No. of parts required = 16;
Packaged Quantity = 5; Order Quantity = 20
Subject to change without notice/Änderungen vorbehalten
Date: April 2004
Order in multiples of packaged quantity
Plus and PSS Series LDV Spare Parts List 1285.002
Screw In - M45
3338706 3338968 AE (H2S)
3338707 3338708 Page 3 of 3

Order Pack.
Item No Description
Code Qty.

21 Rubber Cover - Red 3338773

22 Rubber Cover - Green - (H2S) 3338799

23 Protection Cap 3351069 10

Ordering Details: Edition No: 4


No. of parts required = 16;
Packaged Quantity = 5; Order Quantity = 20
Subject to change without notice/Änderungen vorbehalten
Date: April 2004
Order in multiples of packaged quantity
Plus and PSS Series LDV Spare Parts List 1285.001
Push-In Type A
3338700 3338967 H2S
3338701 3353337 Page 1 of 3
3338702

24,25

23

22 21

19,20

2
9
8

10
11
17 1

18

4
12,13,14,
15,16
5

NPR 1428
26 NPR 1343
NPR 1003
DCC 4416
1285001
Ordering Details: Edition No: 6
No. of parts required = 16;
Packaged Quantity = 5; Order Quantity = 20
Subject to change without notice/Änderungen vorbehalten
Date: Febuary 2004
Order in multiples of packaged quantity
Plus and PSS Series LDV Spare Parts List 1285.001
Push-In Type A
3338700 3338967 H2S
3338701 3353337 Page 2 of 3
3338702
Order Pack.
Item No Description
Code Qty.

1-12,17-19,21-24,26 Plus and PSS A LDV - Short Hose 3338700

1-11,13,17-19,21-24,26 Plus A LDV - Long Hose 3338701

1-11,16-19,21-24,26 Plus A LDV - (1.5m) - Integrated 3353755

1-11,14,17-19,21-24,26 Plus A LDV Japan 3350916

1-11,15-19,21-24,26 Plus A LDV - w ith M P Connection 3338702

1-12,17,18,20-23,25,26 Plus A LDV - Short Hose (H2S) 3338967

1 Body 3350204

2 Grommet 3338750 5

3 ATB Assy. See 1285.005 3350550

4 Injector - Preset 3339683

5 O-Ring 3339459 10

6 Connector - A 3338927

7 O-Ring R18352 10

8 Locking Arm 3350098 5

9 Screw 3350526 10

10 O-Ring 3337215 20

11 O-Ring 3339440 10

10-12 M P Hose (Short) 3350499

10,11,13 M P Hose (Long) 3338716

10,11,14 M P Hose (Japan) 3350915

10,11,15 M P Hose (w ith M P Connection) 3338717

10,11,16 M P Hose (1.5m) 3353756

17 Clip 3338737 10

18 Spacer 3338733 10

Ordering Details: Edition No: 6


No. of parts required = 16;
Packaged Quantity = 5; Order Quantity = 20
Subject to change without notice/Änderungen vorbehalten
Date: Febuary 2004
Order in multiples of packaged quantity
Plus and PSS Series LDV Spare Parts List 1285.001
Push-In Type A
3338700 3338967 H2S
3338701 3353337 Page 3 of 3
3338702
Order Pack.
Item No Description
Code Qty.

19 Diaphragm - Silicone 3338747

20 Diaphragm - Poly mer(H2S) 3338751

21 Slip Ring 3338755 5

22 Spring 3338768 5

23 Bay onet Cap 3338735

24 Rubber Cover (Red) 3338773

25 Rubber Cover (Green) - (H2S) 3338799

26 Protection Cap 3351069 10

Ordering Details: Edition No: 6


No. of parts required = 16;
Packaged Quantity = 5; Order Quantity = 20
Subject to change without notice/Änderungen vorbehalten
Date: Febuary 2004
Order in multiples of packaged quantity
Plus and PSS Series LDV Spare Parts List 1285.008
Tool List

Page 1 of 2

1 t0016 t0036 6

2 t0008 t0044 7

3 t0005 t0045 8

4 t0037 t0046 9

t0014 10
DCC 4935
5 t0034 DCC 4731

Ordering Details: Edition No: 3


No. of parts required = 16;
Packaged Quantity = 5; Order Quantity = 20
Subject to change without notice/Änderungen vorbehalten
Date: July 2002
Order in multiples of packaged quantity
Plus and PSS Series LDV Spare Parts List 1285.008
Tool List

Page 2 of 2

Order P ack.
Item N o D escription
C ode Qty.

1 Tool Bag 3332196

2 Plate Spanner R 26817

3 Screwdri ver Posi D ri v (No.1) 3310579

4 O-Ri ng Li fter R 21402 5

5 Assembly Tool 3310664

6 Retractor Tool 3310663

7 Ball End Setti ng Tool (1.5mm) 3310676

8 Hei ght Gauge 3310675

9 C onnector Tool (M40 & M45) 3310566

10 Molykote 111 Grease 3331241

Ordering Details: Edition No: 3


No. of parts required = 16;
Packaged Quantity = 5; Order Quantity = 20
Subject to change without notice/Änderungen vorbehalten
Date: July 2002
Order in multiples of packaged quantity
Internal: Spare parts list for technician
Dräger Service
Safety

Subject to change without

Spare Part List Edition date Spare Part List


PANORAMA NOVA PE/P / ESA 10.09.2003 1412.052

R52972, R52992, R54215, R53070, R54216, R51492 Page: 1


R51792, R54217, R54218, R59019, R59024, R54450 Image: 1
R59065, R55073, R55001, R55005, R54720, R54743, ..., ..., ...
Internal: Spare parts list for technician
Dräger Service
Safety

Subject to change without

Spare Part List Edition date Spare Part List


PANORAMA NOVA PE/P / ESA 10.09.2003 1412.052

R52972, R52992, R54215, R53070, R54216, R51492 Page: 2


R51792, R54217, R54218, R59019, R59024, R54450 Image: 1
R59065, R55073, R55001, R55005, R54720, R54743, ..., ..., ...

POSITION DESIGNATION ORDER PACK.


CODE QTY.
_________________________________________________________________________________________________________________________________________________________________________________________________________

1, 4, 6-10, PANORAMA NOVA P PC R52972


12-13,
15-27,
32-36
2, 4, 6-10, PANORAMA NOVA P TRIPLEX R52992
12-13,
15-27,
32-36
3-4, 6-10, PANORAMA NOVA P PC/C R54215
12-13,
15-27,
32-36
1, 5-9, PANORAMA NOVA SILICONE P PC R53070
11-12,
14-27,
32-34, 36
3, 5, 11-12, PANORAMA NOVA SILICONE P PC/C R54216
14-27,
32-34, 36
1, 5.1, PANORAMA NOVA P PC SIDEPORT RA R59019
6-10,
12-13,
15-27, 30,
32-39
1, 4, 9, 13, Panorama Nova Std. P Escape R59024
15-16,
19-23,
25-27,
32-34, 36,
40-44
Internal: Spare parts list for technician
Dräger Service
Safety

Subject to change without

Spare Part List Edition date Spare Part List


PANORAMA NOVA PE/P / ESA 10.09.2003 1412.052

R52972, R52992, R54215, R53070, R54216, R51492 Page: 3


R51792, R54217, R54218, R59019, R59024, R54450 Image: 1
R59065, R55073, R55001, R55005, R54720, R54743, ..., ..., ...

POSITION DESIGNATION ORDER PACK.


CODE QTY.
_________________________________________________________________________________________________________________________________________________________________________________________________________

1, 4, 9-10, PANORAMA NOVA STANDARD P R54450


12-13,
15-16,
19-23,
25-27,
32-34, 36,
40-44
1, 4, 6, Panorama Nova P-Mon R59065
6.1-10,
12-13,
15-27,
32-36, 7.2
1, 4, 6-10, PANORAMA NOVA PE PC R51492
12-13,
15-19,
21-23,
28-35
2, 4, 6-10, PANORAMA NOVA PE TRIPLEX R51792
12-13,
15-19,
21-23,
28-35
3, 4, 6-10, PANORAMA NOVA PE PC/C R54217
12-13,
15-19,
21-23,
28-35
3.1, 5.2, Panorama Nova SI P NFPA sw R55080
6-9, 11.1,
12, 14.1,
15-27,
32-34, 35.1
Internal: Spare parts list for technician
Dräger Service
Safety

Subject to change without

Spare Part List Edition date Spare Part List


PANORAMA NOVA PE/P / ESA 10.09.2003 1412.052

R52972, R52992, R54215, R53070, R54216, R51492 Page: 4


R51792, R54217, R54218, R59019, R59024, R54450 Image: 1
R59065, R55073, R55001, R55005, R54720, R54743, ..., ..., ...

POSITION DESIGNATION ORDER PACK.


CODE QTY.
_________________________________________________________________________________________________________________________________________________________________________________________________________

3-4, 6-10, Panorama Nova P Amsterdam R55073


12-13,
15-23,
25-27,
32-36, 41,
45
1, 4, 6-10, Panorama Nova EPDM ESA R55001
12-13,
15-19,
21-23,
28-30,
32-35,
46-48
1, 5-9, Panorama Nova Silicone ESA R55005
11-12,
14-19,
21-23,
28-30,
32-34,
46-48
1, 4, 9-10, PANORAMA NOVA STANDARD PE R54720
12-13,
15-16, 19,
21-23,
28-34,
42-44
1 POLYCARBONATE VISOR R50833
2 TRIPLEX LENS R26989
3 PC VISOR ANTI-FOG ANTI-SCRATCH R55071
3.1 LENS, COATED,NFPA 4055619
4 MASK BODY (EPDM)-CE- R53662
4 MASK BODY (EPDM) R53066
5 MASK BODY (SI) -CE- R53663
5 MASK BODY (SI) R50284
Internal: Spare parts list for technician
Dräger Service
Safety

Subject to change without

Spare Part List Edition date Spare Part List


PANORAMA NOVA PE/P / ESA 10.09.2003 1412.052

R52972, R52992, R54215, R53070, R54216, R51492 Page: 5


R51792, R54217, R54218, R59019, R59024, R54450 Image: 1
R59065, R55073, R55001, R55005, R54720, R54743, ..., ..., ...

POSITION DESIGNATION ORDER PACK.


CODE QTY.
_________________________________________________________________________________________________________________________________________________________________________________________________________

5.1 MASK BODY EPDM (CL3) R54944


5.1 MASKBODY SUPRA SILICONE R52692
5.2 MASK BODY SI-BLACK R55081
6-8 REP.SET TENSION FRAME R23117
6 FRAME PN-SUPRA
6.1 STRAP,CPL. R55060
7 RECESSED HEAD SCREW M4X20 10
7.1 AM4X22 DIN 84-A4/051
7.2 DIN 7985-M4X30-A2-H
7.3 SCHRAUBE ISO7045-M4X22-A4-70-H
8 HEXAGON NUT M4DIN985-A4 1336002 20
9 DOUBLE BUTTON RM08910 15
10 HEAD STRAP CR/NR R26279
11 HEADSTRAP,SILICONE R50282
11.1 HEADSTRAP (NFPA) R54724
12 SLIDING BUCKLE R28329 10
13 INNER MASK R52822
14 INNER MASK,SILICON R52832
14.1 INNER MASK (SI/BLACK) R55082
15-16 CONTROL VALVE,CPL. R26686 4
15 CONTROL VALVE PLATE RM08304 10
16 VALVE SEAT RM08303 10
17 SCREW RING R52816 10
18 DIAPHRAGM R51102 4
19 O-RING T10549 10
20 CONNECTION PIECE R50070
21 VALVE DISC CPL. R54170 5
22 SPRING R50075 10
Internal: Spare parts list for technician
Dräger Service
Safety

Subject to change without

Spare Part List Edition date Spare Part List


PANORAMA NOVA PE/P / ESA 10.09.2003 1412.052

R52972, R52992, R54215, R53070, R54216, R51492 Page: 6


R51792, R54217, R54218, R59019, R59024, R54450 Image: 1
R59065, R55073, R55001, R55005, R54720, R54743, ..., ..., ...

POSITION DESIGNATION ORDER PACK.


CODE QTY.
_________________________________________________________________________________________________________________________________________________________________________________________________________

23 BRIDGE R54133 10
24 CAP R50078 5
25 SPRING R28385 10
26 CAP R28383 10
27 KEY R28384 10
28 CONNECTION PIECE R51490
29 CAP, CPL. R51513 5
30 VALVE DISC R51488 10
31 VALVE SEAT R51489 5
32 PLATE R53179 10
33 SLIDING RING R52184 10
34 CLAMP R52185 10
35 NECK SUPPORT STRAP PVC R51772
35.1 NECK SUPPORT STRAP (NFPA) R54212
36 INHALATION VALVE DISC R54113 10
37 CLAMP R54748
38 ROUND THREAD CONNECTION R54772
39 WASHER RM07946 10
40 HEAD STRAP CPL. R54774
41 CAP BLACK R54441
42 SCREW RING R54442 4
43 FRAME RED R54440 2
44 SCREW 1341642 10
45 SIEVE R55074
30, 46-48 Transition piece ESA R55070
46 VALVE SEAT R55009
47 CONNECTOR R54949
48 O-RING R55014
Internal: Spare parts list for technician
Dräger Service
Safety

Subject to change without

Spare Part List Edition date Spare Part List


PANORAMA NOVA PE/P / ESA 10.09.2003 1412.052

R52972, R52992, R54215, R53070, R54216, R51492 Page: 7


R51792, R54217, R54218, R59019, R59024, R54450 Image: 1
R59065, R55073, R55001, R55005, R54720, R54743, ..., ..., ...

POSITION DESIGNATION ORDER PACK.


CODE QTY.
_________________________________________________________________________________________________________________________________________________________________________________________________________

- SPECIAL PIN KEY R55275


WITHOUT ILLUSTRATION
Tool List - Facepieces Spare Parts List 1281.7
Push-In (also with Side Port)
M45
M40 (Normal Demand)
Page 1 of 2

Ordering Details:
No. of parts required = 16; Edition No: 1
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: December 1995
Tool List - Facepieces Spare Parts List 1281.7
Push-In (also with Side Port)
M45
M40 (Normal Demand)
Page 2 of 2

Order Pack.
Item No Description
Code Qty.

1-6 Facepiece Tol Kit 3333696

1 Tool Bag 3332196

2 Plate Spanner R26817

3 7 x 8mm Open Ended Spanner 3331244

4 Screw driver 5.5 x 1 3331246

5 Screw driver (Posidrive) 3331245

6 Oetiker Pliers R53239

Ordering Details:
No. of parts required = 16; Edition No: 1
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: December 1995
Cylinder Valve Assembly Spare Parts List 9822.10
Right Hand Angled Flow - Ratchet
3336393-M18-200bar; 3336663-M18-300bar;
3338036-(B.S.) - 300bar
Page 1 of 2
3338038 - (DIN) - 300bar

11

10

3
4
3

1,1a,1b,1c

12

NPR 0435
982210
Ordering Details:
No. of parts required = 16; Edition No: 4
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: January 2000
Cylinder Valve Assembly Spare Parts List 9822.10
Right Hand Angled Flow - Ratchet
3336393-M18-200bar; 3336663-M18-300bar;
3338036-(B.S.) - 300bar
Page 2 of 2
3338038 - (DIN) - 300bar
Order Pack.
Item No Description
Code Qty.

1-2-12 Cyl. Valve Assy 200bar - M18 R.H./Angled Flow 3336393

1a-2-12 Cyl. Valve Assy 300bar - M18 R.H./Angled Flow 3336663

1b, 2-11 Cyl. Valve Assy 300bar - BS R.H./Angled Flow 3338036

1c, 2-11 Cyl. Valve Assy 300bar - DIN R.H./Angled Flow 3338038

1 Valve Body (M18) 200bar. Not supplied

1a Valve Body (M18) 300bar. Not supplied

1b Valve Body (B.S.) 300bar. Not supplied

1c Valve Body (DIN) 300bar. Not supplied

2-4 Kit - Valve Head 3335864

5 Valve Stem 3335841 5

6 Spindle Bearing 3335842 5

7 Ratchet Cap 3335599 5

8 Nylon Spacer 3339379 10

9 Control Knob w ith Ratchet 3335638 5

10 Spring 3335639 5

11 Flange Nut 3330905 5

12 O-Ring (M18) 3335413 10

Ordering Details:
No. of parts required = 16; Edition No: 4
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: January 2000
In-Line Cylinder Valves Spare Parts List 9822.8
BS:DIN:M18:M25
Ratchet and Non-Ratchet Types
Page 1 of 3

17,18

1 (BS) 3 (M18)
2(DIN) 4 (M25)

Orientation of
Stem Seals 8

7 6
7
6
6

9
Ratchet Type Non-Ratchet Type
10
11
16
12

15
13
14
14

DCC 4548
98228
Ordering Details:
No. of parts required = 16; Edition No: 7
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: June 2001
In-Line Cylinder Valves Spare Parts List 9822.8
BS:DIN:M18:M25
Ratchet and Non-Ratchet Types
Page 2 of 3

Order P ack.
Item N o D escription
C ode Qty.

1, 5-14 C yli nder Valve BS 200bar wi th Ratchet 3335678

1, 5-9, 14-16 C yli nder Valve BS 200bar wi thout Ratchet 3335679

1, 5-14 C yli nder Valve BS 300bar wi th Ratchet 3335680

1, 5-9, 14-16 C yli nder Valve BS 300bar wi thout Ratchet 3335681

2, 5-14 C yli nder Valve D IN 200bar wi th Ratchet 3335682

2, 5-9, 14-16 C yli nder Valve D IN 200bar wi thout Ratchet 3335683

2, 5-14 C yli nder Valve D IN 300bar wi th Ratchet 3335684

2, 5-9, 14-1 C yli nder Valve D IN 300bar wi thout Ratchet 3335685

4-14, 17 C yli nder Valve M25 300bar wi th Ratchet 3335686

4-9, 14-17 C yli nder Valve M25 300bar wi thout Ratchet 3335687

3, 5-14, 18 C yli nder Valve M18 200bar wi th Ratchet 3336151

3, 5-9, 14-16, 18 C yli nder Valve M18 200bar wi th Ratchet 3336152

3,5-14, 18 C yli nder Valve M18 300bar wi th Ratchet 3335706

3, 5-9 14-16, 18, C yli nder Valve M18 300bar wi thout Ratchet 3335707

1 Valve Body (BS) not suppli ed

2 Valve Body (D IN) not suppli ed

3 Valve Body (M18) 200bar or 300bar not supplied

4 Valve Body (M25) 300bar not suppli ed

5-7 Ki t - Valve Head 3335864 5

8 Valve Stem 3335841 5

9 Spi ndle Beari ng 3335842 5

10 Ratchet C ap 3335599

11 Nylon Spacer 3339379 10

12 C ontrol Knob wi th Ratchet 3335638 5

Ordering Details:
No. of parts required = 16; Edition No: 7
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: June 2001
In-Line Cylinder Valves Spare Parts List 9822.8
BS:DIN:M18:M25
Ratchet and Non-Ratchet Types
Page 3 of 3

Order Pack.
Item No Description
Code Qty.

13 Spring - Ratchet Type 3335639 5

14 Flange Nut 3330905 5

15 Spring - Non-Ratchet Type 3330906 5

16 Control Knob - Non-Ratchet 3330907 5

17 O-Ring (M25) 3335414 10

18 O-Ring (M18) 3335413 10

Ordering Details:
No. of parts required = 16; Edition No: 7
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: June 2001
Cross Flow Cylinder Valves Spare Parts List 9822.1
200 and 300bar
BS: DN: and M25 Threads
Ratchet and Non-Ratchet Types
Page 1 of 2

14,15
1-3
4-5

7 7
Ratchet Valve
Orientation of stem seals
9 16 18

17
8
16 16
10
17
16
11
19

20
12 21
22
13

23

24

DCC 4548
98221
Ordering Details:
No. of parts required = 16; Edition No: 10
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: June 2001
Cross Flow Cylinder Valves Spare Parts List 9822.1
200 and 300bar
BS: DN: and M25 Threads
Ratchet and Non-Ratchet Types
Page 2 of 2

Order Pack.
Item No Description
Code Qty.

1,6-13 Cy linder Valve 200bar BS Thread Non-Ratchet 3330891

2,6-13 Cy linder Valve 200bar DIN Thread Non-Ratchet 3330894

3,6-13 Cy linder Valve 300bar BS Thread Non-Ratchet 3330882

2,6,7,16-24 Cy linder Valve 200bar BS Thread Ratchet Ty pe 3335927

4,6-13,14 Cy linder Valve 300bar M 25 Thread Non-Ratchet 3332013

5,6,7,15,16,24 Cy linder Valve 300bar M 18 Thread Ratchet Ty pe 3336157

5,7-13,15 Cy linder Valve 200bar M 18 - Non-Ratchet 3336214

1,2,3,4,5 Valve Body not supplied

6 Buffer 3330901 2

7-8 Spares Set - Valve HEad (Non-Ratchet) 3333356

9 Valve Spindle (Non-Ratchet 3335637 5

10 Bearing (Non-Ratchet) 3330903 5

11 Control Knob (Non-Ratchet) 3330907 5

12 Spring (Non-Ratchet) 3330906 5

13 Flange Nut (Non-Ratchet) 3330905 5

14 O-Ring - M 25 Valve 3335414 10

15 O-Ring - M 18 Valve 3335413 10

7,16,17 Kit - Valve Head 3335864

18 Valve Stem 3335841 5

19 Spindle Bearing 3335842 5

20 Ratchet Cap 3335599 5

21 Ny lon Spacer 3339379 10

22 Control Knob w ith Ratchet 3335638 5

23 Spring - Ratchet Ty pe Valve 3335639 5

24 Flange Nut 3335928 5

Ordering Details:
No. of parts required = 16; Edition No: 10
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: June 2001
Cylinder Valve Assembly Spare Parts List 9822.14
Right Hand Angled Flow - Non Ratchet

Page 1 of 2

11

10

5
6
5

12

1-3 13

DCC 4548
NPR 0676
982214
Ordering Details:
No. of parts required = 16; Edition No: 2
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: June 2001
Cylinder Valve Assembly Spare Parts List 9822.14
Right Hand Angled Flow - Non Ratchet

Page 2 of 2

Order Pack.
Item No Description
Code Qty.
1, 4,12 Cylinder Valve Assembly BS-200Bar (RH) 3337564

2, 4-13 Cylinder Valve Assembly M18-200Bar (RH) 3337767

3-13 Cylinder Valve Assembly M18-300Bar (RH) 3338977

1 BS-200Bar (Not Available)

2 M18-200Bar (Not Available)

3 M18-300Bar (Not Available)

4-6 Kit-Valve Head 3335864

7 Valve Stem 3335841 5

8 Spindle Bearing 3335842 5

9 Control Knob 3330907 5

10 Spring 3330906 5

11 Flange Nut 3330905 5

12 Rubber Buffer 3330901 2

13 O-Ring (M18 Valve) 3335413 10

Ordering Details:
No. of parts required = 16; Edition No: 2
Packaged Quantity = 5; Order Quantity = 20
Order in multiples of packaged quantity Subject to change without notice/Änderungen vorbehalten Date: June 2001

You might also like