Practice Problems Dosimetry & Instrumentation

Each question is worth 2 points.


19 questions = 38 possible points

1. The process whereby energy is transferred from a photon beam to electrons in the
medium is called:
A. Electron Capture
B. Absorption
C. Bremsstrahlung
D. Scatter
2. Match the most appropriate instrument to the procedure in each question:
I. Liquid scintillation counter
II. NaI well counter
III. Geiger-Mueller (GM) counter
IV. Thermoluminescent dosimeter (TLD)
V. Ionization chamber survey meter
A. Used to detect very small levels of radioactivity
B. Particulate contamination survey for Tc-99m or used with a wipe test to
check radioactive shipments
C. Radiation survey of a diagnostic x-ray installation
D. Personnel monitoring
3. Collecting all the negative ions produced by a beam of photons in a small volume of
air, under conditions of electronic equilibrium, is a direct measure of:
A. Dose Equivalent
B. LET
C. Absorbed Dose
D. Exposure
E. Specific Ionization
4. If the temperature is 22ºC and the pressure is 770 mm mercury, the correction to an
ionization chamber reading would be :
A. 1.013
B. 1.0
C. 0.994
D. 0.987
E. 0.974

5. Megavoltage photons can be detected by all of the following except:


A. An Ion Chamber
B. A Geiger Counter
C. A Diode
D. A Thermoluminescent Dosimeter (TLD)
E. A Flow Meter

6. Exposure is
A. The energy absorbed in a given mass of a medium
B. The air kerma of a photon beam
C. Measured in Sv
D. The ionization in a given mass of air
7. Match the following units with the quantities below: (answers may be used more than
once or not at all)
I. Bq
II. Sv
III. C/kg
IV. Gy
V. J
A. Absorbed Dose
B. Activity
C. Exposure
D. Dose Equivalent

8. Ion chamber readings are corrected for temperature and pressure because:
A. The chamber’s calibration factor is stated at 22ºC and 760 mmHg
B. As temperature increases, the gas in the chamber expands, resulting in a
higher collected charge
C. As pressure increases, the gas in the chamber is compressed, resulting in
a lower collected charge
D. All of the above are true

9. Which of the following radiation detectors can be used to calibrate a linac photon
beam?
I. TLD chip
II. Therapy Verification Film
III. Diode
IV. Farmer Ion Chamber
A. 1, 2, 3
B. 1, 3
C. 2, 4

D. 4 only

E. All of the above


10. Factors influencing the relationship of “dose equivalent” to “absorbed dose” for
particulate radiation include:
I. Atomic Mass
II. Electric Charge
III. Kinetic Energy
IV. Linear Energy Transfer (LET)
A. 1,3
B. 2,4
C. 3,4
D. 2,3,4
E. all of the above

11. Ion-recombination can be a problem when using a:


A. Calorimeter
B. Geiger Counter
C. Ionization Chamber
D. TLD

12. An unsealed ion chamber is exposed under identical condition to the same amount of
radiation in cities A and B. It has a lower charge reading in city A than in city B because:
A. The relative humidity might be higher in city B
B. The temperature might be higher in city B
C. The pressure might be lower in city A
D. The stem effect might be more in city B
E. Background radiation could be higher in city B
13. For a cylindrical ion chamber with inner radius, r, the effective point of dose
measurement in water is shifted:
A. Upstream by a distance of 2 r for electron beams
B. Upstream by a distance of r for electron beams
C. Downstream by a distance of 0.75 r for electron beams
D. Downstream by a distance of r for electron beams
E. Upstream by a distance of 0.5 r for electron beams

14. Exposure is defined for ionization produced in:


A. Water
B. Tissue
C. Air
D. Fat

15. One Roentgen corresponds to a charge of:


A. 3.7 x 1010 disintegrations/sec
B. 2.58 x 10-4 coulomb/kg
C. 0.03 esu of electrostatic charge
D. 1 electron volt

16. A free-air ionization chamber is an instrument used to measure:


A. The charge of free electrons in a radiation beam
B. Ionization produced in a well specified volume of air according to
the definition of Roentgen
C. Charge particle equilibrium
D. Number of ionization tracks produced by electrons
17. The most common wall material used for the outer wall of the ionization chamber is:
A. Aluminum
B. Copper
C. Graphite or Nylon
D. Lead

18. Calibration of an ionization chamber means it has been:


A. Exposed to radiation to stabilize
B. Compared against a national standard to establish a correction factor
C. Tested against leakage and electrical shorts
D. Made waterproof

19. An electrometer is an instrument used to measure:


A. Charge, Current
B. Voltage, Resistance
C. Capacitance
D. All of the above