Professional Documents
Culture Documents
5510 A. Introduction
1. General Discussion shown to interfere include fatty acids, phenols, surfactants, pro-
teinaceous materials, and DOC leached from cellulose.
a. Principle: AHS are concentrated by column chromatogra- c. Minimum detectable concentration: Estimated limit of de-
phy on diethylaminoethyl (DEAE) cellulose and measured as tection is 1.1 mg/L using a 50-mL water sample. The detection
dissolved organic carbon (DOC). AHS are weak organic acids limit can be decreased by increasing sample volume. The major
that bind to anion-exchange materials, such as DEAE cellulose, limitation is blank contamination.
at neutral pH values. The method is based on the assumption that d. Standard substance: Eliminate documentation of false neg-
AHS are the major dissolved organic acids present. atives by analyses of a sample of known humic concentration at
b. Interferences: Any carbonaceous nonhumic materials that regular intervals (at least once per batch of samples).
are concentrated and isolated by the chromatographic method e. Quality control (QC): The QC practices considered to be an
will interfere (false positive response). Substances that have been integral part of each method are summarized in Table 5020:I.
https://doi.org/10.2105/SMWW.2882.106 1
AQUATIC HUMIC SUBSTANCES (5510)/Diethylaminoethyl (DEAE) Method
https://doi.org/10.2105/SMWW.2882.106 2
AQUATIC HUMIC SUBSTANCES (5510)/XAD Method
1. General Discussion Wet dry column with methanol. When the air has been dis-
placed, pump distilled water through the column until the efflu-
a. Principle: AHS are concentrated by column chromatogra- ent concentration of DOC decreases to 0.5 mg/L (approximately
phy on XAD resin and measured as dissolved organic carbon 20 bed volumes). An alternative procedure can also be used.1
(DOC). Acidification of AHS decreases polarity, allowing par- c. Chromatography: Preclean column with three cycles of
tition into the nonpolar XAD matrix. The method is based on the 0.1N NaOH and 0.1N HCl just before pumping sample into
assumption that AHS are the major dissolved organic acids column. Leave column saturated with 0.1N HCl. Acidify sample
present. to pH 2.0 with concentrated HCl, and pump it onto the column
b. Interferences: Any carbonaceous nonhumic materials that at rate of 1.0 mL/min. Save column effluent for DOC analysis.
are concentrated and isolated by the chromatographic method Significant concentrations of DOC in the effluent can indicate
will interfere. This includes fatty acids, phenols, surfactants, that the column was overloaded and that a smaller sample
proteinaceous materials, and DOC leached from the resin, chro- volume should be used. Colored organic acids adsorb to the top
matography pump, or tubing. of the column. Back-elute (reverse flow) the column with
c. Minimum detectable concentration: Estimated limit of de- 0.1N NaOH at 0.2 mL/min and collect eluate in a graduated,
tection is 1.4 mg/L using a 50-mL water sample. The detection conical test tube until it becomes colorless (about 2 mL). Acidify
limit can be decreased by increasing sample volume. The major with conc H3PO4 to a pH of 2 or less (about 2 to 3 drops) and
limitation is blank contaminations. remove dissolved carbon dioxide (inorganic carbon) by purging
d. Quality control (QC): The QC practices considered to be with nitrogen for 10 min. Avoid exposure of alkaline samples to
an integral part of each method are summarized in Table air (i.e. acidify immediately) to minimize contamination with
5020:I. CO2. Determine volume and DOC of acidified column effluent.
After eluting and collecting AHS from the column with back-
2. Apparatus elution using 0.1N NaOH, continue rinsing with about 20 bed-
volumes of the basic solution. Rinse with water for about 20 bed
See 5510B.2a, c, and d. In addition, the following are re- volumes. Repeat the triplicate acid/base column precleaning
quired: procedure described above, then reuse the column to analyze a
a. Glass column, 0.2 ⫻ 25 cm with silanized glass wool. replicate sample. Process two portions of water by the same
b. Pump, with inert internal parts and tubing, capable of flow procedure to serve as controls.
rates of 0.2 to 1.0 mL/min.* The XAD column may be reused to analyze subsequent sam-
c. TFE tubing, 0.2 cm ID. ples and controls if the triplicate acid/base precleaning procedure
d. Extraction apparatus, Soxhlet. is repeated immediately before analysis of each replicate. Re-
place the column if recovery is poor or the resin becomes
3. Reagents
discolored.
In addition to reagents 5510B.3a, c, e, g, and i: 5. Calculation
a. XAD resin,† approximately 250-m size.
b. Hexane.
Calculate the concentration of AHS as given in 5510B.5.
c. Methanol.
d. Acetonitrile.
6. Precision and Bias
4. Procedure
For seven single-operator analyses, the relative standard de-
viation of triplicate samples (about 10 mg/L as AHS) ranged
a. Sample collection and preservation: See 5510B.4a.
from 0.9 to 20.7% with an average of 5.4% (n ⫽ 7).
b. Preparation of XAD resin: Clean resin by successive wash-
For seven single-operator analyses, recoveries ranged from
ing with 0.1N NaOH for 5 d. Extract resin sequentially in a
15.1 to 71.0% with an average of 51.6% and a relative standard
Soxhlet extractor with hexane, methanol, acetonitrile, and meth-
deviation of 35.1%.
anol, for 24 h each. Pack clean resin into a 0.2- ⫻ 25-cm glass
column that has a 2-mm length of glass wool in one end. After
7. Reference
filling, cap column with another 2-mm length of glass wool.
1. STANDLEY, L.J. & L.A. KAPLAN. 1998. Isolation and analysis of
* Pump parts may be of stainless steel or TFE. lignin-derived phenols in aquatic humic substances: Improvements
† XAD-8. XAD is a trademark of Rohm and Haas Co., Philadelphia, PA. on the procedures. Org. Geochem. 28(11):689.
https://doi.org/10.2105/SMWW.2882.106 3