Professional Documents
Culture Documents
4140 A. Introduction
Determination of common inorganic anions such as fluo- (IC). Operating costs are significantly less than those of ion
ride, chloride, bromide, nitrite, nitrate, orthophosphate, and chromatography. Capillary ion electrophoresis can detect all
sulfate is a significant component of water quality analysis. anions present in the sample matrix, providing an anionic “fin-
Instrumental techniques that can determine multiple analytes gerprint.”
in a single analysis [i.e., ion chromatography (Section 4110) Anion selectivity of capillary ion electrophoresis is different
and capillary ion electrophoresis] offer significant time and from that of IC and eliminates many of the difficulties present in
operating cost savings over traditional single-analyte wet the early portion of an IC chromatogram. For example, sample
chemical analysis. matrix neutral organics, water, and cations do not interfere with
Capillary ion electrophoresis is rapid (complete analysis in anion analysis, and fluoride is well resolved from monovalent
less than 5 min) and provides additional anion information (i.e., organic acids. Sample preparation typically is dilution with re-
organic acids) not available with isocratic ion chromatography agent water and removal of suspended solids by filtration. If
necessary, hydrophobic sample components such as oil and
* Approved by Standard Methods Committee, 2004. Editorial revisions, 2011. grease can be removed with the use of HPLC solid-phase ex-
Joint Task Group: 20th Edition—Roy-Keith Smith (chair), James Krol, Yuefeng Xie. traction cartridges without biasing anion concentrations.
1. General Discussion purification system performance and TOC greater than 0.1 mg/L.
The addition of 5 mg formate/L in the mixed working anion
a. Principle: A buffered aqueous electrolyte solution con- standard, and to sample where identification of fluoride is in
taining a UV-absorbing anion salt (sodium chromate) and question, aids in the correct identification of fluoride.
an electroosmotic flow modifier (OFM) is used to fill a Generally, a high concentration of any one ion may interfere
75-m-ID silica capillary. An electric field is generated by with resolution of analyte anions in close proximity. Dilution in
applying 15 kV of applied voltage using a negative power reagent water usually is helpful. Modifications in the electrolyte
supply; this defines the detector end of the capillary as the formulation can overcome resolution problems but require indi-
anode. Sample is introduced at the cathodic end of the cap- vidual validation for precision and bias. This method is capable
illary and anions are separated on the basis of their differ- of interference-free resolution of a 1:100 differential of Br⫺ to
ences in mobility in the electric field as they migrate through Cl⫺, and NO2⫺ and NO3⫺ to SO42⫺, and 1:1000 differential of
the capillary. Cations migrate in the opposite direction and are Cl⫺ and SO42⫺.
not detected. Water and neutral organics are not attracted Dissolved ferric iron in the mg/L range gives a low bias for
towards the anode; they migrate after the anions and thus do PO4. However, transition metals do not precipitate with chro-
not interfere with anion analysis. Anions are detected as they mate because of the alkaline electrolyte pH.
displace charge-for-charge the UV-absorbing electrolyte an-
c. Minimum detectable concentrations: The minimum de-
ion (chromate), causing a net decrease in UV absorbance in
tectable concentration for an anion is a function of sample
the analyte anion zone compared to the background electro-
size. Generally, for a 30-s sampling time, the minimum de-
lyte. Detector polarity is reversed to provide positive mv
tectable concentrations are 0.1 mg/L (Figure 4140:2). Accord-
response to the data system (Figure 4140:1). As in chroma-
tography, the analytes are identified by their migration time ing to the method for calculating MDL given in Section
and quantitated by using time-corrected peak area relative to 1030C, the calculated detection levels are below 0.1 mg/L.
standards. After the analytes of interest are detected, the These detection levels can be compromised by analyte im-
capillary is purged with fresh electrolyte, eliminating the purities in the electrolyte.
remainder of the sample matrix before the next analysis. d. Limitations: Samples with high ionic strength may show a
b. Interferences: Any anion that has a migration time similar decrease in analyte migration time. This variable is addressed by
to the analytes of interest can be considered an interference. This using normalized migration time with respect to a reference
method has been designed to minimize potential interference peak, chloride, for identification, and using time-corrected area
typically found in environmental waters, groundwater, drinking for quantitation. With electrophoresis, published data indicate
water, and wastewater. that analyte peak area is a function of migration time. At high
Formate is a common potential interference with fluoride; it is analyte anion concentrations, peak shape becomes asymmetrical;
a common impurity in reagent water, has a migration time this phenomenon is typical and is different from that observed in
similar to that of fluoride, and is an indicator of loss of water ion chromatography.
https://doi.org/10.2105/SMWW.2882.073 1
INORGANIC ANIONS BY CAPILLARY ION ELECTROPHORESIS (4140)/CIE with Indirect UV Detection
Figure 4140:1. Electropherogram of the inorganic anions and typically found organic acids using capillary ion electrophoresis and chromate electrolyte.
Electrolyte: 4.7 mN Na2CrO4/4.0 mM TTAOH/10 mM CHES/0.1 mM calcium gluconate; capillary: 75-m-ID ⫻ 375-m-OD ⫻ 60-cm length,
uncoated silica; voltage: 15 kV using a negative power supply; current: 14 ⫾ 1 A; sampling: hydrostatic at 10 cm for 30 s; detection: indirect
UV with Hg lamp and 254-nm filter.
https://doi.org/10.2105/SMWW.2882.073 2
INORGANIC ANIONS BY CAPILLARY ION ELECTROPHORESIS (4140)/CIE with Indirect UV Detection
https://doi.org/10.2105/SMWW.2882.073 3
INORGANIC ANIONS BY CAPILLARY ION ELECTROPHORESIS (4140)/CIE with Indirect UV Detection
https://doi.org/10.2105/SMWW.2882.073 4
INORGANIC ANIONS BY CAPILLARY ION ELECTROPHORESIS (4140)/CIE with Indirect UV Detection
Figure 4140:4. Linearity calibration curve for chloride, bromide, and Figure 4140:5. Linearity calibration curve for fluoride and o-phosphate.
sulfate. Three data points were used per concentration; Three data points were used per concentration; based on
based on Youden pair design. Youden pair design.
5. Calculation
C ⫽ A⫻F
where:
C ⫽ analyte concentration in original sample, mg/L,
A ⫽ analyte concentration from calibration curve, mg/L, and
F ⫽ scale factor or dilution factor (for a 1:5 sample dilution,
F ⫽ 5).
6. Quality Control
Figure 4140:6. Linearity calibration curve for nitrite and nitrate. Three
The QC practices considered to be an integral part of each data points were used per concentration; based on Youden
pair design.
method are summarized in Table 4020:I.
a. Analytical performance check: Unless analyst has already
demonstrated ability to generate data with acceptable precision and
bias by this method, proceed as follows: Analyze seven replicates of
c. Water blank analysis: At the beginning of every set of anal-
a certified performance evaluation standard containing the analytes
yses run a water blank to demonstrate that the water is free of
of interest. Calculate mean and standard deviation of these data. The
analyte anions. Dissolved bicarbonate will always be observed as a
mean must be within the performance evaluation standard’s 95%
confidence interval. Calculate percent relative standard deviation positive or negative peak having a migration time greater than
(RSD) for these data as (SD ⫻ 100) / mean; % RSD should conform PO43⫺ and does not interfere with the analysis. Any negative peak
to acceptance limit given in Section 1020B. indicates the presence of an anion impurity in the electrolyte; a
b. Calibration verification: Analyze an independent, certified positive peak indicates the presence of an impurity in the reagent
performance evaluation standard at the beginning and end of the water. If this is noted, discard electrolyte and prepare electrolyte and
analyses, or if many samples are analyzed, after every 10 samples. sample dilutions again with water from a different source.
The determined analyte concentration should be within ⫾10% of d. Analyte recovery verification: For each sample matrix an-
the true value, and the migration time of the Cl⫺ reference peak alyzed (e.g., drinking water, surface water, groundwater, or
should be within 5% of the calibrated migration time. If the Cl⫺ wastewater典, analyze duplicate known-addition samples
reference peak differs by more than 5% of the calibrated migration (4140B.3f). Analyte recoveries should conform to acceptance
time, repeat capillary conditioning and recalibrate before proceeding. limits given in Section 1020B.
https://doi.org/10.2105/SMWW.2882.073 5
INORGANIC ANIONS BY CAPILLARY ION ELECTROPHORESIS (4140)/CIE with Indirect UV Detection
Figure 4140:7. Electropherogram of typical drinking water. Figure 4140:9. Electropherogram of typical industrial wastewater dis-
charge, undiluted.
Chloride ⫽ 24.72 mg/L Fluoride ⬍ 0.10 mg/L
Sulfate ⫽ 7.99 mg/L Carbonate & Chloride ⫽ 2.0 mg/L Formate ⬍ 0.05 mg/L
Nitrate ⫽ 0.36 mg/L bicarbonate ⫽ natural Nitrite ⫽ 1.6 mg/L Phosphate ⫽ 12.3 mg/L
Sulfate ⫽ 34.7 mg/L Carbonate &
Nitrate ⫽ 16.5 mg/L bicarbonate ⫽ natural
https://doi.org/10.2105/SMWW.2882.073 6
INORGANIC ANIONS BY CAPILLARY ION ELECTROPHORESIS (4140)/CIE with Indirect UV Detection
TABLE 4140:V. CAPILLARY ION ELECTROPHORESIS KNOWN-ADDITION RECOVERY AND PRECISION OF PERFORMANCE EVALUATION STANDARD
WITH DRINKING WATER
https://doi.org/10.2105/SMWW.2882.073 7
INORGANIC ANIONS BY CAPILLARY ION ELECTROPHORESIS (4140)/CIE with Indirect UV Detection
TABLE 4140:VI. COMPARISON OF CAPILLARY ION ELECTROPHORESIS WITH CHROMATE ELECTROLYTE WITH OTHER METHODS FOR THE
DETERMINATION OF ANIONS
Concentration by Given Method*
mg/L
Sample Wet Chemical or Other
Anion Matrix No. Method† IC‡ CIE
Chloride Effluent 1 — 149 147
2 — 162 161
3 — 152 151
4 — 139 139
5 — 111 110
6 — 109 107
7 — 3.6 3.5
Landfill leachate 1 ⬍1 ND ND
2 190 211 201
https://doi.org/10.2105/SMWW.2882.073 8
INORGANIC ANIONS BY CAPILLARY ION ELECTROPHORESIS (4140)/CIE with Indirect UV Detection
Landfill leachate 1 — ND ND
2 — ND ND
https://doi.org/10.2105/SMWW.2882.073 9