Professional Documents
Culture Documents
Participant Guide
Alinity c Global Field Service Training | Internal Use Only |80001189-101 iii
Revision History
Revision History
Document Control
Revision Date Content or Pages Revised, Added, or Deleted
Number
Any product information in training materials should be used in conjunction with the latest
version of the Operations Manual, Operations Manual addendum, or Product Information
Letter. If discrepancies in information exist within training materials or any other materials,
the latest version of the Operations Manual, Operations Manual addendum, or Product
Information Letter takes precedence.
All samples (printouts, graphics, displays or screens, etc.) are for information and illustration
purposes only and shall not be used for clinical or maintenance evaluations. Data shown in
sample printouts and screens do not reflect actual patient names or test results.
Each person assumes full responsibility and all risks arising from use of the Information. The
Information is presented “AS IS” and may include technical inaccuracies or typographical
errors. Abbott Laboratories reserves the right to make additions, deletions, or modifications to
the Information at any time without any prior notification.
All Abbott Laboratories product names and trademarks are owned by or licensed to Abbott
Laboratories, its subsidiaries or affiliates. No use of any Abbott trademark, trade name, trade
dress, or product name may be made without the prior written authorization of Abbott
Laboratories, except to identify the product or services of Abbott Laboratories. All other
trademarks, brands, product names, and trade names are the property of their respective
companies. All rights reserved. Except as permitted above, no license or right, expressed or
implied, is granted to any person under any patent, trademark, or other proprietary right of
Abbott Laboratories.
This guide was developed and produced by Global Service Training, Dallas, TX.
Course Overview
Objectives
Review training guide notations, class goals, and required pre-
learning activities
Module Topics
Introduction
Training Overview
Goals
Prerequisite Activities
Materials
Service Cadence
Achieving Service Excellence
Effective Troubleshooting
KM Service Documents
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 Intro-1
Course Overview Module Topics
Training Overview
Time: 30 mins
Instrument Description
The Alinity ci-series provides full integration of clinical chemistry and
immunoassay analysis. The Alinity ci-series includes a clinical chemistry
module and an immunoassay module, each performing all sample
processing activities, and a system control module to provide a single
common interface. Each system can be customized by combining
one or more sample processing modules, in multiple configurations,
with a system control module to form a single workstation.
Intro-2 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Module Topics Course Overview
Goals
Upon completion of this training program, you will be able to:
Describe the role each major functional area plays in the Alinity c
clinical chemistry process
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 Intro-3
Course Overview Module Topics
Prerequisite Activities
In preparation for class, you must complete pre-training prior to
the class. This material is mandatory. It is designed to provide a
baseline level of knowledge and will immediately begin building
your understanding of the Alinity c. The material will not be covered
in class. It ensures all learners are at the same level of preparedness
for classroom activities and supports maximum hands-on instrument
troubleshooting during the training.
Intro-4 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Module Topics Course Overview
Materials
The information for the Alinity c Field Service Training will be presented
by instructors using the following materials:
Alinity c Global Field Service Training Program
MyGSS or Alinity c Service Documentation
Alinity ci-series Operations Manual (Online Help)
Alinity immunoassay and Clinical Chemistry System Operations
Manual
Alinity ci-series Assay-specific Package Inserts
Knowledge Management troubleshooting database
Remote diagnostics tools: AbbottLink and LOG-ICAL
Ticket Code Tools (TCT) Database
Technical Service Bulletins (TSB)
Instrument Service Advisories (ISA)
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 Intro-5
Course Overview Module Topics
Typography
Black Text Black text will appear in both Facilitator Guide and Participant Guide
Citation Book titles and other references will be green in color
Activity: Activities are provided throughout the guide to give participants an opportunity to
practice the concepts and techniques discussed.
Intro-6 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Module Topics Course Overview
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 Intro-7
Course Overview Service Cadence
Service Cadence
Time: 60 min Daily Service Cadence is integrated with course technical learning.
The following topics are necessary as part of the Service Cadence
expectations for service product support. A brief overview of how it is
integral to service and support are highlighted in this section.
AST Introduction
Intro-8 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Service Cadence Course Overview
The ASE model uses the following with the AST Tools to prevent
instrument failures, lab down-time and to proactively support the
customer.
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 Intro-9
Course Overview Service Cadence
PRE-VISIT ACTIVITIES
Be Ready - Review
ONeview: Instrument Health Notification/POM
Alert; related tickets
Key customer metrics [note red customer KPIs
for all calls]
Knowledge Management check, ensure
accuracy
Receiving
Initiate contact and plan timely intervention
(Planned Service)
Assure customer of proactive approach with
PHM and ALWAYS ON
DURING VISIT
Understanding:
Demonstrate Active Sense Technology –
provide visual
Share Instrument Health Notification / POM Alert or and PHM related information
Provide more detailed visual
Helping:
Perform Proactive service call or PHM ticket
Utilize Knowledge Management
Utilize Chat or FaceTime with SME if needed
Perform Total Call (address reasons for red KPIs in ONeview)
Keeping:
Conversation with the Customer: Account health
Ensure AbbottLink connectivity
Discuss proactive actions for red customer KPIs (high CPY in ONeview)
Provide Service Report
Follow Up:
Follow up with customer next day
Utilize ASE Skills
Continue to monitor instrument to confirm resolution
Intro-10 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Service Cadence Course Overview
Effective Troubleshooting
Effective troubleshooting and problem resolution requires four logical
STEPs.
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 Intro-11
Course Overview Service Cadence
Actions: Open the GSR database from MyGSS AST Tool table, access
the GSR Reports and complete each activity below:
1) 6)
2) 7)
3) 8)
4) 9)
5) 10)
4. Locate the Parts Viewer report, select the following options to filter
the data, and record the top 10 reasons for a service request.
Intro-12 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Service Cadence Course Overview
2) 7)
3) 8)
4) 9)
5) 10)
7. Review the Avg. Unit Cost (standard) column and place a check-
mark next to a part if the cost is >$1000 US global goal to indicate a
high cost part.
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 Intro-13
Course Overview Service Cadence
KM Service Documents
Troubleshooting:
Diagrams
Power Supplies
Card Cage and Miscellaneous Boards
LEDs and Test points
Board silkscreen actual layout
Intro-14 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Service Cadence Course Overview
Time: 30 mins
1. What is the Card cage back plane connector for the ICT Pre Amp
Board?
Answer: ________________________________________________________
Answer: ________________________________________________________
3. What is the M&D number and description that will tests the Optics
trigger sensor and displays the trigger sensor value.
Answer:
_________________________________________________________________
4. a). What is the R&R number for the Reagent Positioner Assembly?
Answer: ____________________________________________________________
Answer:
___________________________________________________________
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 Intro-15
Instrument Overview and Basic Operations
Objectives
Module Topics
Instrument Overview
Instrument Statuses
RSM Status
System User Access Levels
System Inventory
Required consumables
Inventory management
Required Accessories
Assay Processing
Detection Technologies
Assay Processing sequence
Assay File Installation
Assay Calibration
Assay Quality Control
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 1-1
Instrument Overview and Basic Operations
Assay Precision
Assay Processing Activity
AST Activity
1-2 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Instrument Overview and Basic Operations
Instrument Overview
Time: 60 mins Delivery Method: Review functional areas, components and
subsystems. Participants should be in front of the analyzer for this
section.
Topic Reference
Functional This topic was covered in the pre-learning materials.
Areas System control module (SCM)
Reagent and sample manager (RSM)
Processing module (Proactive Procedures)
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 1-3
Instrument Overview and Basic Operations
1-4 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Instrument Overview and Basic Operations
Instrument Statuses
Time:10 mins Delivery method: class discussion and demonstration in front of
instrument.
Status Description
Offline Indicates when the processing module is not communicating with
the system control module (SCM).
Stopped Indicates when the processing module is turned on and is
communicating with the SCM.
Initializing Indicates a temporary status that occurs when Start or Run on
the Home screen is selected
Warming Indicates when the processing module has completed
initialization but has not reached the required temperature.
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 1-5
Instrument Overview and Basic Operations
Three status indicators (blue, green, and amber) are located above each reagent and
sample manager (RSM) position to indicate the status of sample processing and when racks
and cartridges can be accessed.
Status Description
Indicators off No rack or cartridge is inserted in the position.
Green (steady) The rack or cartridge is inserted but is not in process.
The rack or cartridge can be accessed
Amber (steady) The rack or cartridge is in process.
The rack or cartridge cannot be accessed.
Green (blinking) Processing is completed. The rack or cartridge can be accessed.
Amber (blinking): Unloading of a cartridge or vial rack is in process.
The position is reserved and is unavailable to load a rack or
cartridge can be accessed.
1-6 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Instrument Overview and Basic Operations
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 1-7
Instrument Overview and Basic Operations
System Inventory
Required Consumables
Required consumables are replenishable items that are needed for
Time:30 mins
sample processing.
Consumables Description
Assay Reagents One or more cartridges that contain all the necessary reagent
components for an Alinity ci-series assay.
One reagent assay kit – contains 1 reagent cartridge bottle
Two reagent assays kit – contains 2 reagent cartridge bottles
Cartridges contain a two-dimensional bar code which includes
the following information:
Reagent identifier
Reagent serial number
Number of tests (test size)
Expiration date
Onboard stability time
ICT Module Contains the sodium (Na+ ), potassium (K+ ), chloride (Cl- ), and
reference electrodes.
1-8 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Instrument Overview and Basic Operations
Consumables Description
Bulk Solutions Liquid solutions that are provided in large quantities for use
during sample processing.
Solution Description
ICT Reference A mid-concentration standard solution that
Solution (1L bottle) is aspirated and analyzed by the ICT module
before and after each sample. The solution
provides a reference potential that is used in
result calculation.
Alkaline Wash Used by the cuvette washer to clean the
(0.5L bottle) cuvettes after sample analysis.
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 1-9
Instrument Overview and Basic Operations
Item Description
Sample Cups Topic was covered in pre-training materials.
1400μL disposable containers that hold
samples for testing.
Volume graduation marks at 125μL, 500μL,
and 1400μL eliminate the need to pipette
with precision.
Can be placed in sample tubes with
bar code labels to facilitate positive
identification.
Reference: See Alinity ci-series Operations Manual > Section 1:
Use or function > Required consumables for important details
regarding storage and use of consumables.
1-10 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Instrument Overview and Basic Operations
Inventory Management
Time: 10 mins System inventory is viewed and managed using the following screens:
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 1-11
Instrument Overview and Basic Operations
Required Accessories
Time: 15 mins Required accessories are parts that are needed for sample processing
on the Alinity c. Required accessories include the following:
Item Description
Racks and Trays Racks: Used to transport specimens, calibrators, and controls to
the sample pipettor by the RSM
Vial rack: Bar-coded for identification with the
letter V. Holds six open vials of calibrators or
controls for immediate use. The rack cannot be
stored in the reagent carousel.
Onboard vial rack: Bar-coded for identification
with the letter U. Holds six vials of calibrators
or controls. The rack is stored in the reagent
carousel. The carousel position will display as
purple in the Reagents screen. Onboard racks
can be assigned by the operator.
Sample rack: Holds six primary tubes, aliquot
tubes, or sample cups. Any combination of
tubes can be used in the rack.
Sample Gauge: used to verify the amount in
an aliquot tube exceeds 8mm.
Trays: Used to hold multiple racks of samples,
calibrators, and controls and to hold reagent
cartridges. Trays of racks and cartridges are loaded on the
reagent and sample manager. Each tray holds a maximum of
five racks or cartridges.
1-12 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Maintenance and Diagnostic Procedures Instrument Overview and Basic Operations
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 1-13
Instrument Overview and Basic Operations Maintenance and Diagnostic Procedures
Procedure Key
Time: 15 mins Provides access to the front and rear processing center covers of the
Alinity ci-series processing modules and the RSM by overriding the
interlocks for the processing module and SCM covers.
Procedure key location for SCM Procedure key location for Alinity c
Procedure Key overrides all the Processing Module, RSM and SCM
covers
Procedure keys are shipped insert into the SCM module and can
be move for use to any module.
1-14 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Maintenance and Diagnostic Procedures Instrument Overview and Basic Operations
Backup
Time: 10 mins
.
Automatic
Manual
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 1-15
Instrument Overview and Basic Operations Maintenance and Diagnostic Procedures
Time: 60 mins
Purpose: To familiarize participants with performing a manual
instrument backup.
1-16 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Maintenance and Diagnostic Procedures Instrument Overview and Basic Operations
Assay Processing
Detection Technologies
Time: 10 mins Photometric Technology
Measurements occur either when all the reactant is depleted and the
reaction is stable (end-point assays) or when the reactant reaches a
stable rate (rate assays).
Potentiometric Technology
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 1-17
Instrument Overview and Basic Operations Maintenance and Diagnostic Procedures
1-18 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Maintenance and Diagnostic Procedures Instrument Overview and Basic Operations
Position(s) Description
1 The sample pipettor dispenses sample into the cuvette at the
starting position.
2 The reaction carousel rotates counterclockwise approximately ¼
turn (47 cuvette positions). The cuvette containing sample is now
at the first reagent dispense position. Reagent pipettor 1 dispenses
reagent 1 in the cuvette.
3 The reaction carousel rotates one cycle (¼ turn) to the first mixing
position where the mixer unit (mixer 1) mixes the sample and
reagent 1.
4 The reaction carousel rotates one cycle to this position at which no
activity takes place.
5 The reaction carousel has completed 4 cycles. Each cycle rotates
the reaction carousel 47 cuvette positions (¼ turn). There are 187
cuvettes in the reaction carousel. After 4 cycles the cuvette is now
one position beyond the original starting position (4x47=188).
6–153 The reaction carousel continues to rotate and the reaction mixture
incubates. The photometer takes absorbance readings every
time the cuvette passes the photometric position. The optics take
absorbance readings every time the cuvette passes the light path
for a total of 38 read times.
154–182 The cuvette washer removes the reaction mixture to waste and
cleans the cuvette with Alkaline Wash, Acid Wash, and DI water.
Then the cuvette washer dispenses DI water into the cuvette for a
water blank measurement to ensure cuvette integrity. Finally, the
cuvette washer aspirates the water and dries the cuvette.
183–188 The clean cuvette waits for new sample dispense.
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 1-19
Instrument Overview and Basic Operations Maintenance and Diagnostic Procedures
Position(s) Description
5 If onboard dilution is required, the sample pipettor aspirates the
diluted sample and dispenses the sample into the new cuvette that
is currently at position 1.
40 For an ICT sample, the ICT probe aspirates the diluted sample into
the ICT unit.
75 If the reaction requires a second reagent, reagent pipettor 2
dispenses reagent 2 into the cuvette.
76 The mixer 2 mixes the second reagent with the sample and reagent
mixture.
1-20 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Maintenance and Diagnostic Procedures Instrument Overview and Basic Operations
The Assay tab also provides for assay configuration tasks, such as:
Configure assay parameters
Configure assay panels
Configure a calibrator set
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 1-21
Instrument Overview and Basic Operations Maintenance and Diagnostic Procedures
Assay Calibration
Time: 10 mins Calibration is the process used to analyze calibrators of a known
concentration, to record system response values, and to plot
measured values against the known concentration.
Before performing sample processing, calibrate all assays.
Assay calibrations require assay-specific calibrators for analysis.
Acceptable calibration curves are valid when a QC analysis is run
and data are within specifications.
Mandatory or optional
Calibration Data
1-22 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Maintenance and Diagnostic Procedures Instrument Overview and Basic Operations
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 1-23
Instrument Overview and Basic Operations Maintenance and Diagnostic Procedures
QC Summary Screen:
Levey-Jennings Graph:
Stored Calibrations
Assay calibrations are stored by the system after the run is completed
and displays as one of the following status types:
Active: values are within specifications and the system software will
calculate patient and QC results from this curve 1 calibration curve is
stored per lot number
1-24 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Maintenance and Diagnostic Procedures Instrument Overview and Basic Operations
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 1-25
Instrument Overview and Basic Operations Maintenance and Diagnostic Procedures
Assay Precision
Time: 10 mins Precision indicates the reproducibility or how well an assay or a system
provides the same result when a given sample is tested repeatedly
over time.
1-26 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Maintenance and Diagnostic Procedures Instrument Overview and Basic Operations
Materials Needed:
Reagents, Calibrators, Controls
Sample Cups
Racks
Bulk solutions
Daily Maintenance Cartridges
Smart Wash Solutions
Service Tips:
Toshiba manufacturing requires all new systems to pass the PVT
acceptance criteria prior to shipping. Performing the PVT during
new system installation verifies that issues have not occurred as a
result of shipping and/or the installation process.
Daily Maintenance procedure initialized the DAQ Board during
system installation and DAQ Board replacement.
Note: Some actions may not be required for this training activity,
but should be reviewed in class.
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 1-27
Instrument Overview and Basic Operations Maintenance and Diagnostic Procedures
Verify calibration
Note: Skip steps for creating PVT panels for assay, as this will be
done on Day 4.
Activity: QC Analysis
1-28 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Maintenance and Diagnostic Procedures Instrument Overview and Basic Operations
Service Tips:
Ensure samples and controls are devoid of foam and bubbles;
remove any if present.
Sample cups contain 3 approximate volume lines (125μL, 500μL,
and 1400μL).
Ensure reagents have sufficient volume to perform requested
orders.
Place the processing module and RSM into Running status prior to
ordering samples to improve efficiency.
Calibrations must be reordered if an error sends them to the
Exceptions.
When viewing Results screens, select column header name to
organize data alphanumerically and easily find/sort data prior to
printing.
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 1-29
Instrument Overview and Basic Operations Maintenance and Diagnostic Procedures
__________________________________________________________
__________________________________________________________
__________________________________________________________
Delete the control configured above.
1-30 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Maintenance and Diagnostic Procedures Instrument Overview and Basic Operations
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 1-31
Instrument Overview and Basic Operations Maintenance and Diagnostic Procedures
1-32 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Maintenance and Diagnostic Procedures Instrument Overview and Basic Operations
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 1-33
Instrument Overview and Basic Operations Maintenance and Diagnostic Procedures
Component
Description
Number
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
1-34 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Power Distribution
2 Power Distribution
Objectives
Use service tools and diagnostic tools to interpret data and isolate
root causes of power supply and power distribution failures
Module Topics
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 2-1
Power Distribution Power Distribution Overview
The main power supply for the SCM is located behind the lower access
panel in the rear of the SCM. It provides power to the RSM and key
SCM components housed inside the SCM.
The main power supply for the c-series Processing Module is located
behind the power supply panel in rear of the instrument and contains
six internal DC power supplies (with variable resistors) and a power
distribution side panel.
Power Requirements
Time: 5 mins Review pre-site checklist information in the classroom.
2-2 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Power Distribution Overview Power Distribution
Component Function
SCM Main Located at the rear of the SCM
Power Supply Provides power to both the SCM and the RSM components
Self-contained, auto-selecting power supply for frequency and
voltage requirements
Input voltage range: 90-264VAC
Input frequency range: 47–63Hz
Provides 12VDC, 24VDC, 42VDC output
Fuses are replaceable and have various ratings indicated on the
fuse and by the tip color
Fuse indicator wire will pop out to identify an over current condition
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 2-3
Power Distribution Processing Module Power Components
Topic Reference
c-series Main Receives the incoming 200–220VAC line voltage and routes it to
Power Supply the primary side of the Power Transformer and Power Supply 7
(PS7).
Contains LED’s for each DC power voltage (illuminated indicates
ON)
Contains potentiometers for adjusting DC power
Table continued…
2-4 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Processing Module Power Components Power Distribution
Topic Reference
c-series Main Power
Power Supply Voltage and Function Adjustable
Supply
(continued)
PS1 +5VDC Card Cage Backplane/CPU Yes
Bd/Logic
+3.3VDC Card Cage Backplane/CPU
Bd/CMOS
±12VDC Pressure Sensor
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 2-5
Power Distribution Processing Module Power Components
2-6 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Processing Module Power Components Power Distribution
Topic Reference
Reagent Power Power
Supplies Voltage and Function Adjustable
Supply
PS7 Receives 200VAC from the c-series Yes
main power supply
Generates 60VDC for the Servo
Motor Drive (drives the Reagent
Carousel).
PS8 Receives 100VAC from the c-series Yes
main power supply
Generates 12VDC for the
thermoelectric devices (TED).
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 2-7
Power Distribution Processing Module Power Components
2-8 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Processing Module Power Components Power Distribution
Topic Reference
Transformer Transformer is tapped to allow selection of the corresponding
incoming voltage: 200VAC, 220VAC, 230VAC and 240VAC
Provides 100VAC to the:
6 internal Power Supplies
SIO Driver Board for the Vacuum Pump
Thermal Bath Controller Board for the water bath heater
Contains a voltage selector cable or jumper that is factory set
and may require a different setting if incoming voltage is different.
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 2-9
Power Distribution Processing Module Power Components
2-10 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Processing Module Power Components Power Distribution
Note: *Ensure module status Is offline and remains off least 1 minute prior to
turning them back ON to prevent errors.
#8035 related
to 5V power for
Module Load
Control 15V and
l. Turn ON front
PM Power
the supply error 5V Power Good
but the sensor Switch
failed. m. Tap Start
#5119 RSH
Module Load
e. Turn ON front
when module
Control 15V is at Stopped
and 5V Power SCM Power status
Good but Switch
the sensor
f. Turn ON front
failed.
PM Power
Turn ON front PM Switch*
Power Switch*
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 2-11
Power Distribution Processing Module Power Components
When cycling PS, wait until all PS LEDs are OFF plus a count to ten
before powering ON the main breaker to prevent errors.
SCM Control board monitors voltage signals for power quality and if
an error condition is detected will disable distribution.
SCM PS Fan Service Due LED provides status for periodic service
replacement:
Steady ON – Fan Good
Blinking – Replacement required within 30 days
OFF – Fan failure
If the RSM servo motor is manually moved with the power on then
the holding current is removed from the motor to prevent damage.
Note: Turn off the system power to move any motor assembly to
prevent damage to the motor and assemblies with belts.
When the front Power Switch is OFF on the processing module the
following will remain ON:
Reagent carousel cooler
Reagent container access covers
Located at the transfer door and LLS access areas
Reagent positioner door present sensor
Power ON the following in the order listed to avoid errors when the
main SCM Power supply is switched to OFF
1) SCM back switch
2) UIC
3) SCM front switch (wait for Home Screen to display before
powering ON)
4) PM (wait for Home Screen to display before powering
ON)
2-12 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Power Distribution Activities Power Distribution
Time: 30 mins
Purpose: To perform service procedures related to power distribution.
Service Tips:
5. Remove the Thermistor heater Bracket, pull forward and set aside.
6. Loosen the 2 screws (do not remove them) holding the Voltage
Selection Terminal Housing and pull it forward to make the
corresponding AC Voltage selection.
10. Reinstall the Thermistor Heater Bracket and the Cooling Unit.
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 2-13
Power Distribution Power Distribution Activities
Service Tips:
2-14 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Power Distribution Activities Power Distribution
Time: 15 mins
Abbott: This tools activity is for Abbott personnel only. Non-Abbott
participants should partner with an Abbott employee.
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 2-15
Power Distribution Power Distribution Activities
Time: 15 mins
Abbott: This tools activity is for Abbott personnel only. Non-Abbott
participants should partner with an Abbott employee.
2-16 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Power Distribution Activities Power Distribution
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 2-17
Electronic Circuitry
3 Electronic Circuitry
Objectives
Use service tools and diagnostic tools to interpret data and isolate
root causes of electronic circuitry failures
Module Topics
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 3-1
Electronic Circuitry Electronic Circuitry Overview
3-2 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Electronic Circuitry Overview Electronic Circuitry
Component Function
SCM Control Board Located in the upper rear
section of the SCM
Receives power from the SCM
main power supply
1 board in each instrument,
regardless of processing
modules connected in a
workcell (instrument)
Functions include:
RSM Transport motor X, Theta
and Z drive
RSM X Drive module is a high
power module application.
Connects up to Four Load/
Unload boards
Uses the Load/Unload board
to communicate to the
sample positioner
Lid safety interlock monitoring
X servo drive power relay
Power quality monitoring
Service bypass for RSM
interlock
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 3-3
Electronic Circuitry Electronic Circuitry Overview
Component Function
RSM Computer Located in the Rear of the SCM
(Also referred to as SCM Embedded below Ethernet Switch B
Computer) Acts as the RSM CPU and
interfaces to the UIC via a
Ethernet switch
Receives 12VDC voltage from
the SCM main power supply
Firmware updates occur during
system software installation
automatically
Ethernet Switches 2 network switches located
inside the SCM.
Handle all of the network
communication between
the UIC and the connected
modules.
3-4 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Electronic Circuitry Overview Electronic Circuitry
Component Function
User Interface Computer (UIC) Located in the SCM
Unaffected by the use of the
SCM Power Switch
Receives 120-220VAC from the
SCM main power supply
Provides a common centralized
user interface for the:
Processing related
components
RSM components
Host computer
Contains the Windows
Operating software (OS) and
Alinity ci-series Software
Utilized for remote diagnostics
capabilities and system boot-up
observation
System backup can be auto-
programmed to capture the
following files:
Module calibrations: System
calibration files, robo.cal,
and ish.cal files
System configurations:
default host and system
parameters
Module database: System
database file, M&D
procedures, assay files
and calibrations, results,
and module and SCM
configuration
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 3-5
Electronic Circuitry Electronic Circuitry Overview
Component Function
RSM Load/Unload Located at the left end of each RSM module section
Control Board Provides power and detects
status of RSM 10 and 15 Slot
Sensor Boards through a single
connector
Powered by the processing
module
Receives 24 VDC from the
processing module power
supply This is true for IA and
CC.
Monitors the processing module front lid sensor status for control
of the RSM Transport
Detection of the Priority Switch and indication of the slot in the
load area to be treated as priority
Requires a jumper setting for module identification during
installation
RSM 2-Axis Driver Small extension of the RSM Load/Unload Control board used
Board to isolate the rack positioner motor drivers from the rest of the
controls and indicators
Located at the left end of each RSM load area
Takes motor commands from the Load Control board and
converts them to high power motor drives for the two rack
positioner motors Attaches directly to a connector on the Load/
Unload Control Board
Board contains fuses, LED’s and switches for diagnostics
RSM 10 and 15 Slot Located above the sample rack and reagent cartridge slots on
Indicator Boards the loading area
Driven by the RSM Load/Unload Control board according to
inputs from the RSM sensor boards which detect presence of
cartridges and sample racks
Slot 1-10 use the 10 slot indicator board
Slot 11-25 use the 15 slot indicator board
Board contains LEDs for diagnostics
3-6 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Electronic Circuitry Overview Electronic Circuitry
Component Function
RSM 10 and 15 Slot Located under the loading area
Sensor Boards Two sensors used to detect:
Rack or cartridge: optical sensor
Tray: magnetic sensor
Slot 1-10 use the 10 slot indicator board
Slot 11-25 use the 15 slot indicator board
Board contains LEDs for diagnostics
RSM Transport Located on the RSM transport
Interface Board Communications with the SCM Control board
Provides transport functions as follows
Communications with the sample and reagent bar code
reader
RSM X home sensor
Processing module X home sensor
Theta and Z home sensor
Signal pass-through for:
Theta and Z motor drive and encoder
Transport Arm Sensor board sensors
RSM Transport Arm Serves as a mount for the pick or place sensor
or End Sensor Board Interface for cartridge or sample rack sensor (SICK)
A single connector conveys electronic signals from both sensors
to the RSM Transport Interface Board
Board contains LED’s for diagnostics
Located on RSM transport arm
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 3-7
Electronic Circuitry Electronic Circuitry Overview
Component Function
CPU Contains the c-series firmware and provides communication
with the SCM
A backup of the robotics information is stored in the CPU board
flash memory
Controls the movement of the reagent carousel servo motor by
monitoring sensor status and sending control data to the Servo
Motor Driver.
SM AC/DC Controls the:
Controller 12VDC valves and solenoids
100VAC vacuum pump
Three WMU 24VDC pumps
RD30 DC Pump (External Probe/Mixer and cuvette wash)
Internal Probe wash
Water bath circulation RD20 Pump
Three CAB’s 24VDC Transfer pumps
Controls the three reagent Transport 24VDC Reagent Transport
motors
Reagent transport Y motor
Reagent transport Z motor
Reagent positioner open/close motor
Stepper Motor Controls 2-phase DC stepper motor, 5-phase DC stepper motor
Controller (SMC) Sends control data to the Stepper Motor Driver Boards (SMD-Q1,
SMD-Q2)
3-8 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Electronic Circuitry Overview Electronic Circuitry
Component Function
DAQ Amplifies and converts the optical and ICT analog signals to a
digital format and sends data to the SCM
Uses stored robotics step table offsets alignment information in
SRAM memory
Board contains a non-replaceable battery continually charged
when instrument is ON
Continued...
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 3-9
Electronic Circuitry Electronic Circuitry Overview
Component Function
Pressure Monitoring Has a digital processor that monitors the electrical output signals
(PM) from the pressure monitoring sensors during the aspiration,
dispense, and wash processes on all probes
Component Function
SIO Driver Board Controls voltage to drive the following:
12VDC drive voltage to activate the valves and solenoids
100VAC to the vacuum pump
24VDC to the RD30 Water Pump (external probe/cuvette/
mixer wash) & Internal Probe Wash
Stepper Motor Driver Receives control data from the Stepper Motor Controller (SMC)
(2) 2-phase Directs power to 24VDC stepper motors
Driver Boards The 5-phase motors receive control signals via SMD-Q1.
SMD-Q1
SMD-Q2 The 5-phase stepper motors perform smaller movements
than the 2-phase stepper motors and are used on
(3) 5-phase assemblies that require precise movement control.
Driver Boards
LLS Board Used to detect contact with liquid surface by the sample and
reagent probes
Note: All LLS pipettor circuit board LEDs should be lit when
the system is on.
3-10 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Electronic Circuitry Overview Electronic Circuitry
Component Function
Temperature Bath The water bath temperature is monitored and controlled by the
Controller (TBC) TBC board
The TBC monitors the status of the water bath temperature
sensor and drives a heater and pump to maintain water bath
temperature at 37°C
Miscellaneous Boards
Time: 10 mins
Component Function
CNN Board Serves as a power supply and distribution/interface for signals
between the SMC and SM AC/DC controller boards and sensors
mounted within the instrument.
Contains 2 DC-DC converters that generate the following
voltages:
+5VDC: Voltage supply of photo sensors and Servo Motor
Driver
±15VDC: Voltage supply of LLS board
Pipettor LLS Located on the head of the pipettors
Pre-Amp Board Amplifies the sample and reagent probes LLS signal
Optics Pre-Amp Amplifies the signal of the photo-diode array of the optics
Board assembly and transmits the signal to the DAQ for processing; this
board is part of the Optics assembly
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 3-11
Electronic Circuitry Electronic Circuitry Overview
Component Function
Cooler Controller Controls the Cooler Unit operation
Board Monitors the internal air temperature of the Reagent Supply
Center.
Sends temperature information to the CPU Board
Controls the two reagent carousel lid heaters
The heaters are placed around the reagent aspiration holes
to prevent condensation from forming around the opening
Monitors the lid sensors for the Reagent Positioner Assembly and
the rear service access cover.
Monitors the Reagent Coolant Bottle float sensor
Shut Board Located to the right of the card cage
Receives 24VDC from power supply
Functions as a safety switch to stop the processing module
operation when the front and rear covers are open.
Shut board is only found on the Total Wire Diagram service
documents currently.
ICT Pre-Amp Board Receives and amplifies signals from the ICT module and
transmits signals to DAQ board
I-ref Level Detect Connects to the IREF Cup electrode sensors to:
Board Detect the presence of IREF solution in the IREF Cup
Verifies IREF solution is aspirated out of the cup by the ICT probe.
3-12 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Electronic Circuitry Overview Electronic Circuitry
Time: 20 mins Materials: 15-Pin VGA Male to VGA Male Cable (6 ft.)
2. Connect a VGA cable to the VGA port on the UIC monitor and the
SCM Embedded Computer VGA port.
3. Switch the monitor to view VGA by using the switch located at the
bottom of the monitor.
Note: Change the view to check the Alinity software screen goes
to Stopped.
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 3-13
Electronic Circuitry Electronic Circuitry Overview
Time: 20 mins Purpose: Learn board-related diagnostic procedures for servicing the
electronic circuitry.
Activity: Remove the following boards and identify the key board
components listed.
LEDs
Battery
RAM
LAN Port
a. Which LAN port connection contains a cable? __________
System
Utilities
3-14 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Electronic Circuitry Overview Electronic Circuitry
Note: The log file without a trailing number is the most current Log.
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 3-15
Electronic Circuitry Electronic Circuitry Overview
Activity: AbbottLink
3-16 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Electronic Circuitry Overview Electronic Circuitry
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 3-17
Electronic Circuitry Electronic Circuitry Overview
BOARD FUNCTION
2. Where is the robotic step table offset data stored and what M&D
allows you to view and edit this information?
Answer:
3-18 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Robotics
4 Robotics
Objectives
Use diagnostic tools and interpret data to isolate the root cause of
reagent management-related errors
Module Topics
Robotics Overview
Summary – Robotics
Review – Robotics
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 4-1
Robotics Robotics Overview
Robotics Overview
Time: 35 mins
4-2 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Robotics Overview Robotics
Topic Description
Robotics Stepper motor controller board contains a generic factory set of
Calibration robotic step table data in firmware default position step numbers for
Offsets robotic positions
Instrument-specific adjustments are made by completing robotic
calibration M&Ds
Calibrations provide offset data for firmware position data
After successful completion of M&D procedures, data are
automatically updated on the CPU and DAQ boards
M&D 4804 Save/Restore DAQ Data provides options to view,
modify, save, restore and print the offset step table information.
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 4-3
Robotics Robotics Overview
Topic Description
Motor Assembly Optical sensors and flags are used to determine the home,
Position Sensors rotational, or vertical position of a motor driven device
Many robotic assemblies locate the Home position by utilizing a
pitch flag on the motor drive shaft and a home flag on the device
that is being driven
When the leading edge of the home flag enters the home sensor:
The software starts monitoring the pitch sensor status.
The motor continues to run until the leading edge of the pitch
flag activates the pitch sensor.
When the leading edge of the pitch flag activates the pitch sensor:
The motor will stop. The pitch sensor stops the motor.
The software verifies that the home sensor is still active.
The home flag should be approximately centered in the sensor.
4-4 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Processing Center Robotics Components Robotics
The 5-phase Stepper Motors are driven by 5-phase SMDs and the
control signals are provided via SMD-Q1. 5-Phase Motor Drivers:
Reaction Carousel Motor
Sample Pipettor Rotate Motor
Sample Pipettor Up/Down Motor
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 4-5
Robotics Processing Center Robotics Components
Component Function
Reaction Carousel Uses a 24VDC 5-phase motor to positions cuvettes for sample
processing
Uses the following components:
Optical home and pitch sensors
Guide rollers use tension to hold carousel
4 rollers – 1 adjustable guide roller, 3 fixed
Adjustable guide roller is located by the optics assembly
and is used to adjust the optics beam height for low
reaction volume assays.
Optical Trigger Sensor monitors positional flags on cuvette
segments
Cuvette segment vertical flags signal proper position of
cuvette for optical read
Adjusted during M&D 4001 Optics Trigger Sensor Check
Reaction carousel is aligned using P-4008 reaction Carousel
alignment
24VDC 5-phase motor
Cuvette Washer Consists of:
Eight nozzles, mounted on a platform, that are used for an
eight-step cuvette washing process.
2 Optical sensors are used to tell the Stepper Motor Controller
Board (SMC) when the elevator is in the Up and Down
positions
24VDC 2-phase motor
Manual alignment performed via P-4009 Cuvette Washer
Alignment
Uses alignment Tool engraved with “WASH”
4-6 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Processing Center Robotics Components Robotics
Component Function
Cuvette Washer Washes and dries the cuvettes
head and Nozzles Front row of tubing = aspirate lines
Back row tubing = dispense lines
Water dispense from nozzles 1, 4, 5
Dispense is in an umbrella spray pattern
Nozzles Functions
1 Aspirates HCW from cuvette via HCW
pump
Dispenses degassed reagent grade type II
water into cuvette
2 Delivers a 1.3% mixture of alkaline and
water from the Wash Solution Pump.
After the alkaline solution is dispensed
into the cuvette, it is aspirated through a
vacuum pump.
3 Delivers a 0.9% mixture of acid and water
from the Wash Solution Pump.
After the acid solution is dispensed into the
cuvette, it is aspirated through a vacuum
pump.
4 and 5 Dispenses water into the cuvette
Aspirates water through a vacuum pump.
6 Provides degassed reagent grade type II
water from wash solution pump to fill the
cuvette to a certain level for blank read
(cuvette integrity check)
7 Performs an aspiration of the water blank.
8 Provides final cuvette drying via:
A porous cuvette drying tip
Aspiration of residual water via vacuum
pump (VP1)
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 4-7
Robotics Processing Center Robotics Components
Component Function
Mixers Assembly Consists of two main sections :
Lower mechanical section :
Linear slide, provides vertical drive to raise and lower the upper
assembly.
Two optical sensors : detects when the assembly is in the correct
up or down position
24 VDC two-phase motor
Upper section:
2 customer replaceable Mixers blades.
Uses the piezoelectric effect.
Frequency applied depends on the mixer position in the
reaction carousel or mixer wash cup
120Hz is used for faster movement/shorter travel when
down in the cuvette or wash cup
70 Hz is used for slower movement to remove water from
the mixer in the upper position at the wash cup
Optical sensor located in front of wash cup to detect blade
movement
4-8 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Processing Center Robotics Components Robotics
Component Function
ICT Assembly The upper section contains:
ICT module and holder
Two water lines to allow 37°C water to recirculate and
indirectly stabilize the temperature of the ICT module
Optical sensors to indicate when assembly is in the up/down or
rotational home position.
Uses two 24VDC two-phase stepper motors to move horizontally
and vertically.
Manual alignment performed using robotic alignment blocks:
“ELC” labeled robotic alignment block used for the cuvette
alignment position
“ELC-POD” labeled alignment block used for the IREF Cup
alignment position
Sample Pipettor (S) Belts are used to drive the assembly in horizontal and vertical
movement
Drive mechanisms use optical home and pitch sensors to tell the
Stepper Motor Controller Board (SMC) when the pipettor is in the
home position.
Crash guard sensor located in the pipettor head is used to detect
obstacles during downward movement.
Adjusted using M&D 4101Crash Sensor Alignment
Two 24VDC 5 phase stepper motors for theta an Z movement
controlled by SMD-Q1
Uses a calibration alignment tool found in the customer
maintenance kit to perform automated calibration M&D 4102
Sample Pipettor Calibration
Calibrates to positions:
Sample wash cup
Sample positioner
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 4-9
Robotics Processing Center Robotics Components
Component Function
Reagent Pipettors Belts are used to drive the assembly in horizontal and vertical
(R1, R2) movement
24VDC 2-phase stepper motors controlled by SMD-Q1
Crash guard sensor located in the pipettor head is used to detect
obstacles during downward movement.
Adjusted using M&D 4101Crash Sensor Alignment
Uses a calibration alignment tool found in the customer
maintenance kit to perform automated calibration M&D 4103 R1
Pipettor Calibration and M&D 4104 R2 Pipettor Calibration and:
Reagent carousel targets
Reaction carousel
Pipettors are unique for R1 and R2
4-10 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Processing Center Robotics Components Robotics
Time: 90 mins
Service Tips:
Ensure the proper alignment blocks are used to prevent errors.
Remove power to instrument before moving any pipettor to
prevent mis-alignment of home and pitch sensors.
Use a voltmeter to measure the actual frequency at the mixer
blade cable connection.
Ensure cuvette Washer is in the up position after performing the
alignment procedure to prevent nozzle and cuvette damage.
Use M&D 4802 Robotics Test Tool to manually align the following
positions using the steps below:
Sample Wash Solution Tray positions 1 & 2
R1 Pipettor Water Bath Additive position
To align the probes from the M&D 4802 Robotics Test Tool:
Select
Procedures
Diagnostics tab
Module 1
M&D 4802 Robotics Test Tool
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 4-11
Robotics Processing Center Robotics Components
Perform
Select
Proceed
Select:
Home
Move to Position – choose position for alignment
Sample Pipettor: Sample Wash Solution area 1 and 2
R1 Pipettor: Water Bath Additive position
Continue
Align
Tap Clockwise or Counterclockwise to align the probe to
the target
Return to Previous menu
Down
Vertical Adjustment (to check)
Down or LLS Down
UP
Return to Previous menu
Done
Home
Return
Type 10 to End
Select:
Continue
Proceed
Done
4-12 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Processing Center Robotics Components Robotics
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 4-13
Robotics Reagent Supply Center Components
4-14 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Reagent Supply Center Components Robotics
Component Function
Reagent The carousel is split into 12 segments and have 72 cartridge locations
Carousel (only 70 reagent cartridge positions)
The last segment only has four positions (67 – 70) configured for the
Calibration and Control rack.
Two positions onboard solutions and maintenance cartridges
which are loaded into any carousel position for:
Smart Wash
Water bath Additive
Cleaning Solution
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 4-15
Robotics Reagent Supply Center Components
Component Function
Reagent Provide closure of the Reagent Supply Center
Positioner Position cartridges and vial racks to load in the reagent carousel or
(door) onto the loading area.
Door alternates open and closed for the complete positioning
event
Open Position – cartridge or rack is transferred to the loading
area reagent positioner pocket
Closed Position – cartridge or rack is lifted by the reagent
transport
Open Position – cartridge is lowered into reagent carousel
Closed Position – cartridge or rack is inside the reagent
carousel
Manual access of reagent cartridges from the carousel is available in
the event of a failure.
4-16 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Reagent Supply Center Components Robotics
Component Function
Reagent Moves Reagent Cartridges, Calibrator or Control racks between the
Transport Reagent Transfer pocket and a position on the Reagent Carousel
Motion coordinated with reagent position
SM AC/DC controller board, 24VDC driven by SMD-Q2
Sensor provides an input to the CNN Board
Reagent Transport
Function
Sensors
Cartridge Present Detect presence of a cartridge or rack
Used in the calibration of carousel rotational
alignment to the transport
Reflective sensor. Also known as “SICK”
sensor.
Cartridge Picked Detects proper engagement of the cartridge
lift tab
Optical interrupt sensor mounted on the
Reagent Transport Cartridge Load Board
Latched Roller switch (mechanical sensor) located at
the bottom center of the lifter
Detect proper release of the latch for each
cartridge slot in the carousel.
Calibration Mechanical sensor wired in series with the
latched sensor
Used for aligning the transport to the carousel
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 4-17
Robotics Reagent Supply Center Components
Time: 60 mins
Service Tips:
4-18 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Reagent and Sample Manager Components Robotics
RSM Configurations
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 4-19
Robotics Reagent and Sample Manager Components
Component Function
RSM The following boards are related to the RSM for control and voltage. They
Electronic are described at length the Electronic Circuits module.
Circuit Boards
Note: RSM electronic circuitry was discussed earlier.
4-20 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Reagent and Sample Manager Components Robotics
Component Function
RSM Transport Moves racks and cartridges to and from:
The loading area
The bar code reader to read and identify samples, reagents, and
solutions
Module-specific sample positioner or reagent positioner
Consists of replaceable components:
Roller wheels: two fixed and one spring-loaded wheels that ride on
guide rails to move transport
Theta motor: provides transport arm rotation
Z motor: provides transport arm up/ down movement for pick and
place of racks or cartridges
Transport arm attached to the Z nut with a single screw. The
complete arm can be replaced if it gets damaged
Transport Interface Board: connects power and routes signals for
transport position sensors, Bar Code Reader (BCR) communication,
and motor drives
Motors receive 42VDC from SCM Control Board
A flexible cable chain is used to route cables to interface board
and SCM
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 4-21
Robotics Reagent and Sample Manager Components
Component Function
RSM Transport Sensor Description
Sensors
Transport Arm (2 Present sensor ( reflective): detects if rack/
sensors) cartridge is present at a pick up point also
used with the RSM X calibration target to
establish X corrections for each Processing
Module
Picked sensor ( optical interrupt): detects
if rack or cartridge was properly picked or
placed
Instrument X home Detects home flag located at the far right of
(optical interrupt) any work cell combination
For a single processing module
instrument, the instrument X home flag is
located on the SCM.
For a multiple processing module system,
the system X home flag is moved to the
right side of the farthest right processing
module
Processing module Detects home flag at the left of each
X home (optical processing module and is used to set the X
interrupt) position information within the limits of that
particular module.
Theta home Detects if the arm is in the theta home
position
Z home Detects if arm is fully in the up in the home
position
4-22 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Reagent and Sample Manager Components Robotics
Component Function
RSM Transport Belt moves transport via X-Motor
X-Motor, Belt 42VDC servo motor with power and control from SCM Control Board
and Idler
Pulleys (2) Lubricate both idler pulleys, one on left and right ends of the RSM
Pulley wheel is lubricated with SuperLube oil during installation, during
a PM, and as needed to prevent the belt from riding up and creating
movement errors
Flex IGUS Controls position and minimizes wear for the three cables that connects
Cable Chain power and signals from the SCM to the RSM Transport Interface Board
Position sensors
BCR communication
X-motor
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 4-23
Robotics Reagent and Sample Manager Components
Service Tips:
Follow the service procedure M&Ds in the order written to prevent
errors.
Move Carrier Transport to extreme left of RSM to remove it or
perform maintenance to prevent damage to transport sensors.
Proper Belt positioning is important for proper operation.
M&D 1600 RSM Transport Calibration procedure aligns the Carrier
Transport position with positioner pockets, tray and Carrier positions
(Continued…)
4-24 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Reagent and Sample Manager Components Robotics
Actions: Use your service documents to find the procedures for the
components listed below. Then perform the removal and replacement
of the parts.
Belt
Pulleys - left side only
Note: For training purposes, insert only a few racks/trays for testing
to learn the M&D function.
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 4-25
Robotics Reagent and Sample Manager Components
4-26 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Reagent and Sample Manager Components Robotics
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 4-27
Robotics Reagent and Sample Manager Components
Review – Robotics
Time: 15 mins
Reagent Positioner
Door (CRA Door)
2) What voltages are generated by the CNN board, and describe the
purpose of each voltage?
4) List two symptoms that might suggest the cuvette washer is mis-
aligned.
4-28 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Objectives Fluidics
5 Fluidics
Objectives
Identify the physical location and function of fluidics system
components
Use fluidics diagram to identify how fluid flows through the system
and how waste is removed
Use service tools and diagnostic tools to interpret data and isolate
root causes of fluidic system failures
Module Topics
Fluidics Overview
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 5-1
Fluidics Fluidics Overview
Waste removal
Low concentration waste (LCW) components
High concentration waste (HCW) components
Liquid waste consolidation
Fluidics Overview
Time: 10 mins
5-2 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Fluidics Overview Fluidics
Degassing
Reduces bubbles within the water fluidics areas
Waste removal
High Concentration Waste
Low Concentration Waste
ICT Fluidics
ICT Reference Solution Cup Fill
ICT Aspiration
The fluidics system uses several electronic valves to direct fluid flow
throughout the system for use or moved to waste Valve control: SM
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 5-3
Fluidics Fluidics Overview
Component Function
Bulk Solutions Bulk solutions bottles contain keyed caps and a septum that is pierced
& On-board during load process
Reservoirs Bulk solutions bottles have keyed openings to prevent load errors
Item/Description Function
1. Vent Manifold
2. Bottle Key Piercing
Mechanism and Needle
Assembly
3. Molded Cover
4. Door Hinges
5. Gear Pump/Manifold
Assembly
6. 0.5L On-Board Alk
Bottle & Level
Sensors
7. 0.5L On-Board Acid
Bottle & Level Sensors
8. 1L On-Board ICT Ref
solution & Level Sensors
5-4 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Water Distribution System Fluidics
Review the fluidics diagram for big picture of both Lower and Upper
Fluidics functions and components.
The components are located in the SCM under the SCM power
supply
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 5-5
Fluidics Water Distribution System
Component Function
SCM Regulator Regulates incoming water pressure to 15-20 psi.
Pre-site specification is 25psi minimum input
Water Distribution Two-outlet manifold and valve used to direct regulated water to
Manifold the Water Management Unit (WMU)
Right Port: directs water to standalone system
Left Port: directs water to the second or integrated system
5-6 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Processing Module Fluidics Fluidics
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 5-7
Fluidics Processing Module Fluidics
Component Function
Inlet Water Input for plumbed Reagent Grade Type II Water:
Connection 15–20psi or 0.103hPa–0.172hPa
Inlet Water Connections located underneath left rear system corner
Valve Two-way valve regulates incoming water
ON (open) to FILL open water tank
OFF (closed) to STOP incoming water flow
Flow Restrictor Restricts/slows incoming water flow rate through narrowing diameter
DI Water 9L water reservoir that fills automatically to maintain water availability
Tank & Float and enables water to be distributed throughout the instrument
Switches (3) Water levels are monitored by float switches:
Overflow sensor or float switch: provides alerts of an overflow
condition
High (full) and low float switches: signal control to open/close
incoming On/Off valve
RD30 Water 24VDC motor
Pump Provides water to:
Probes and mixers wash cups
Cuvette Wash Nozzles 1, 4, 5
Electronic valve(s) located at Probe Wash Cup
M&D 4202 Wash Cup Test Procedure used as a tool to measure
water flow to only the Sample Probe wash cup
The water pump is constantly running when the instrument is ON and
water flow is present; this is to maintain circulation and decrease
occurrence of contamination
5-8 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Processing Module Fluidics Fluidics
Component Function
Water Bath Pump supplies degassed water to water bath
Fill Peristaltic A float switch located inside water bath signals SM AC/DC Controller
Pump water bath is full
Utilizes a photo-sensor and encoder
When replacing tubing use, all supplied lubricant on the tubing
Replaceable 12VDC 2-phase motor
Internal Probe SIO driver board provides 24VDC via PS6
Wash 24VDC Supplies degassed water for internal probe cleaning:
Water Pump
Uses a common manifold to move fluid to 3 different water lines
The water flows through a syringe, PM sensor, and probes
Uses a two-way valve for control
Water Bath Filters water entering the water bath as it exits the heater
Filter Filter is replaced during annual PM
Water Bath Allows water bath to empty to LCW during M&D procedures
Drain Valve SM AC/DC Controller Board supplies the control signal SIO Driver
board provides the voltage
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 5-9
Fluidics Processing Module Fluidics
Degasser
Component Function
Degasser A self-contained unit used to remove air from incoming water
Canister contains semi-permeable membrane tubing to remove air
from incoming water
Provides pressure errors :
High pressure error generated when tubing leaks excessive
water into canister
5-10 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Processing Module Fluidics Fluidics
Process Function
The Water Bath fill Peristaltic Pump provides degassed water
into the water bath.
A float switch in the water bath monitors the water level and
Water Bath Fill
sends signal to module when the water bath is full.
The water bath fill Pump turns OFF and the Circulation Pump
and Heater turn ON
The Water Heater and Water Bath Circulation Pump maintain a
constant water bath temperature of 37°C by re-circulating the
water from the bath through the pump, into the Water Heater,
Water Bath through a manifold and back to the water bath in a closed
Circulation loop.
The manifold has a port that provides re-circulated heated
water to the ICT Warming Jacket to keep the ICT module
warm.
Water bath The Water Bath Drain Valve is used to direct the bath water to the
Drain LCW External Waste Port.
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 5-11
Fluidics Processing Module Fluidics
Components Description/Component
Internal Probe When it is necessary to wash the probes internally, the Internal Probe
Wash Wash Pump and the 2- way sample reagent valves turn ON.
Valves are driven by 12VDC provided by the SIO Driver Board
Degassed water is pumped through the syringe blocks, PM Sensors
and out of the sample and reagent probes.
Upon completion of a probe wash cycle, the pump turns OFF and
the sample reagent valves close. The 3- way valve on the sample
line directs a small amount of water to the water bath overflow
compartment to relieve pressure in the line.
5-12 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Processing Module Fluidics Fluidics
Components Description/Component
External Probe The RD30 Water Pump pumps water in a re-circulating closed loop
Wash from the water tank, to a manifold and 3-way valve and back to the
tank.
The Water Pump is constantly running to maintain circulation. When
the valves open, water is directed to the cuvette washer nozzles and
probe and mixer wash cups.
An adjustable needle valve is used to set the sample probe external
wash volume.
The wash volume for the mixers and reagent probes are not
adjustable. The flow rate is based on the pump output and tubing
size.
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 5-13
Fluidics Processing Module Fluidics
Fluidics Activity 1
Activity: Fluidics
Time: 60 mins
Materials Needed:
Service Tips:
Use all lubricant on water bath pump head tubing during
replacement and ensure tubing is installed without being twisted
During Wash Cup test, ensure:
Review M&D instructions prior to starting
Syringe valve cable is disconnected prior to beginning test and
reconnected at the end to avoid errors
Needle valve locking ring is loosened to adjust the valve and
secured after the adjustment
5-14 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Processing Module Fluidics Fluidics
Cuvettes are washed with alkaline and acid wash solutions and
then rinsed.
After the water rinses, it dispenses water for a water blank reading
and then dries the cuvette prior to the next sample.
Process Description
Cuvette The RD 30 Water Pump in the WMU pumps water in a recirculating
Water Wash closed loop from the water tank, to a 3-way valve and back to the
tank.
The Water Pump is constantly running to maintain circulation.
When the 3-way valve opens, water is directed to the cuvette washer
nozzles 1, 4 and 5. An adjustable needle valve (prior to the 3-way
valve) is used to set the wash volume.
Nozzle 6 dispenses degassed water that is used for the water blank
read. The center position 1-ml syringe on the Wash Solutions Pump
provides the degassed water to nozzle 6.
Cuvette The Wash Solution Pump dilutes the alkaline solution to 1.3% and the
AlKaline and acid solution to 0.9% and then distributes the solutions to the nozzles for
Acid Wash the cuvette washing.
Wash solution Pump consists of:
Three 1mL syringes and check valves (blue)
Two micro syringes:
Aspirates and create dilution of Alkaline and Acid solutions
Utilizes 4 valves: ALK1, ALK2, ACD1, ACD2
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 5-15
Fluidics Processing Module Fluidics
Process Description
Cuvette AlKaline On the down stroke of the pump:
and Acid Wash Degassed water from the Degasser Unit is supplied to the two
1mL syringes through check valves mounted at the top of
each syringe.
At the same time, the ALK1 and ACD1 valves turn on (NC is
open, NO is closed) allowing aspiration of straight alkaline
and acid solution from the bulk solution bottles into the micro
syringes.
The 1mL syringes filled with water and the small, measured
amounts of alkaline and acid solutions in the tubing lines
create the proper dilutions to be delivered to the cuvette
washer nozzles.
On the up stroke of the pump:
Valves ALK1 and ACD 1 turn off (NC is closed, NO is open)
and valves ACD2 and ALK2 turn on (NC is open, NO is
closed), allowing the solutions to be pumped into the nozzles
2 and 3 of the cuvette washer.
At the completion of the cycle, valves ALK2 and ACD2 turn
off (NO open, NC closed), allowing any remaining solution
or air to be directed to the low concentration waste tank
through the valves normally open port..
Cuvette Washer A dual headed vacuum pump is used to remove the contents
Waste and Drying of the cuvettes and provide a final drying before the cuvettes
are used again. The vacuum pump has two heads that are
identified as VP1 and VP2.
VP1 is attached directly to the cuvette dryer tip and provides
final drying.
Aspirated water is directed through the vacuum pump to the
LCW tank in the WMU.
VP2 aspirates water, diluted alkaline and acid wash solutions
from the cuvettes into the nozzles through a manifold and
directs it to the LCW tank in the WMU.
5-16 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Waste Removal Fluidics
Waste Removal
Time: 15 mins
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 5-17
Fluidics Waste Removal
Component Function
HCW Pump Peristaltic style pump; different than water bath fill pump tubing
Provides aspiration of waste from cuvettes to optional HCW container
or drain
HCW aspirated by nozzle 1 on the cuvette washer
24VDC 2-phase motor
Controlled by the SMC board and powered by the SMD-2
Uses a photo-sensor and encoder
Replaceable
Ensure all lubricant is used during tubing replacement
ICT Aspiration Aspirates HCW from ICT HCW cup through a check valve to HCW
Pump container or external drain
5-18 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Waste Removal Fluidics
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 5-19
Fluidics Waste Removal
5-20 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Fluidics Activity 2 Fluidics
Fluidics Activity 2
Time: 80 mins
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 5-21
Fluidics Fluidics Activity 2
Time: 15 mins
Actions: Perform one of the following scenarios:
Clamp off VP2 line to induce a multiple cuvette overflow.
Clamp off Cuvette Washer Nozzle drain tubing to vacuum
pump to induce a single cuvette overflow.
Replace the HCW Pump tubing with a slightly damaged (cut)
tubing to induce Cuvette Washer nozzle 1 overflow and leak in
the HCW Pump head.
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
Abbott: The tools actions below are for Abbott personnel only.
Non-Abbott participants should partner with an employee.
Based on the above error code, list the steps to access the KM article
for recommended troubleshooting or related service documents.
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
5-22 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
ICT Fluidics and Sample Progression Fluidics
ICT Overview
Delivery Method: Review and discuss in front of the instrument
Time: 20 mins
ICT methodology uses solid state ion-selective electrodes contained in
a single chip (ICT module) and reduces maintenance.
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 5-23
Fluidics ICT Fluidics and Sample Progression
Component Function
ICT Reference Delivers and removes ICT reference solution (IREF) from ICT reference
(IREF) Solution solution cup (IREF cup)
Pump Pump is driven by a 24VDC 2-phase motor.
Consists of 2 syringes with orange check valves
Syringe #12
Delivers ICT reference solution through water bath warming ring
and into IREF cup
Syringe #12 should be free of bubbles
Syringe #13
Aspirates remaining solution from IREF cup and moves it into
water bath/waste overflow
Syringe #13 may normally contain bubbles
Syringe tubing connected to incorrect locations (connected in
reverse) will result in IREF cup overflow
IREF solution leaks cause syringe plunger to stick to the platform, know
how to clean it
ICT fluidics are common customer errors
IREF solution Cup Fill and IREF Solution not Aspirated are the common
errors
ICT Warming Warms IREF solution to 37°C before it fills IREF cup
Ring Holes in warming ring may impact voltage readings
Caution: Follow approved procedures for cleaning water bath to prevent
errors, warming ring holes, and ICT result issues.
IREF Cup & Contains preheated ICT reference solution aspirated by ICT probe and
Electrodes measured by ICT module
Contains two electrodes for capacitance level sensing
Ensure spatial arrangement of ICT LLS electrodes remains
unchanged to prevent ICT cup aspiration errors
5-24 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
ICT Fluidics and Sample Progression Fluidics
Component Function
ICT Aspiration Used to aspirate sample or ICT Reference Solution into the ICT Module
Pump for measurement.
Consists of two 1mL syringes
Syringe #11
Aspirates samples or ICT reference solution into ICT module for
measurement and pumps it into ICT waste cup
Syringe #11 should be free of bubbles
Syringe #15
Aspirates waste from water bath waste area and pumps it out
the HCW line
Only syringe #15 has a check valve (orange)
Pump driven by 24VDC two-phase stepper motor.
ICT aspiration or pinch valve (3-way valve) directs the sample flow
from ICT probe to water bath waste area
Note: The ICT pinch tubing will need to be seated in the valve
during the installation process.
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 5-25
Fluidics ICT Fluidics and Sample Progression
Component Function
ICT Module The ICT probe:
Aspirates diluted sample from the cuvette
Aspirates ICT reference solution from the IREF cup into the ICT
module for processing
ICT (integrated chip technology) measure sodium (Na), potassium (K),
and chloride (Cl) simultaneously
Uses solid state ion-selective electrodes contained in a single chip
to reduce required maintenance
The ICT module manufacture date is determined from serial number
on the module as follows:
The table below shows how the ICT serial number breaks down.
100515042
10 05 15 042
Year made Month made Day made Sequence #
5-26 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
ICT Fluidics and Sample Progression Fluidics
Component Function
IREF Level
Detect Board
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 5-27
Fluidics ICT Fluidics and Sample Progression
ICT Diagnostics
Time: 10 mins
5-28 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
ICT Fluidics and Sample Progression Fluidics
Time: 10 mins During ICT sample processing, the following events occur:
2. Sample is analyzed.
ICT unit moves out and down
to position ICT probe in the
cuvette
ICT aspiration pump Syringe
#11 aspirates cuvette
contents into ICT module
ICT module measures sample
System converts
measurements into millivolt
readings used to calculate
sample results.
ICT pre-amp amplifies
signal
DAQ board converts
reads to a digital signal
CPU sends digital data to SCM UIC
SCM UIC calculates the final sample results using stored
calibration data
3. Waste is removed.
ICT aspiration pump Syringe #11 moves sample from ICT module
to HCW compartment
ICT aspiration pump Syringe #15 aspirates liquid waste from
HCW compartment
ICT aspiration pump Syringe #15 moves liquid waste to HCW
tubing
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 5-29
Fluidics ICT Fluidics and Sample Progression
Time: 60 mins
Materials Needed:
Standard Tool Kit
Cotton Swabs
Service Tips:
Only hand-tighten the ICT Module to prevent errors.
Syringes on both ICT pumps (should be free from bubbles on the left
side to prevent errors.
Actions: Remove ICT aspiration tubing from the pinch valve; perform
the following M&Ds
__________________________________________________________
__________________________________________________________
5-30 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
ICT Fluidics and Sample Progression Fluidics
__________________________________________________________
__________________________________________________________
Reinstall pinch valve tubing and perform the following procedure
until the error is resolved:
M&D 5102, Flush ICT Module
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 5-31
Fluidics Liquid Level Sense and Pressure Monitoring
Liquid level sense (LLS) and pressure monitoring (PM) are used to
detect liquid and ensure sample and reagent liquid aspirations occur
without error during assay processing.
5-32 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Liquid Level Sense and Pressure Monitoring Fluidics
LLS signals are monitored for any change after liquid is detected
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 5-33
Fluidics Liquid Level Sense and Pressure Monitoring
5-34 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Liquid Level Sense and Pressure Monitoring Fluidics
Component Function
Pressure Sensor Contains piezoelectric sensor and an internal preamplifier board
Field-replaceable part
Signals from fluid flow are sent to the Pressure Monitor board for
evaluation
Fluid flow is monitored by the PM system during probe aspirations,
dispense, and probe washes for detection of clots, foam, or bubbles
Pressure Pressure is monitored for the sample/reagent probes during:
Monitor Board Sample/Reagent aspiration and dispense
Internal probe wash cycle
PM data collected by the system reflects the fluid pressure in the
probes, tubing, syringes, and manifold
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 5-35
Fluidics Liquid Level Sense and Pressure Monitoring
Syringe Drives
Component Function
Syringe Drives Provide aspiration and dispensing of sample and reagent
Sample (S) 24VDC 2-phase motor drives plunger up/down
Reagent 1 Only reagent syringe drive assemblies are interchangeable
(R1) Sample syringe block and internal components are different than the
Reagent 2 R1 & R2 syringe block
(R2) O-rings and seal tips are replaced during customer Quarterly
maintenance
5-36 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Liquid Level Sense and Pressure Monitoring Fluidics
PM Error Descriptions
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 5-37
Fluidics Liquid Level Sense and Pressure Monitoring
Time: 30 mins
Time: 15 mins
Troubleshooting Scenario
5-38 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Liquid Level Sense and Pressure Monitoring Fluidics
Abbott: The following actions are for Abbott personnel only. Non-
Abbott participants should partner with an Abbott employee.
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
If the R2 pipettor was identified as the cause of the issue, complete the
following using GPPM as a reference tool:
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 5-39
Fluidics Summary – Fluidics
Summary – Fluidics
Time: 10 mins
5-40 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Summary – Fluidics Fluidics
ICT Fluidics
ICT Aspiration ICT Ref Solution Not M&D 5102 Flush ICT Module
Pump Aspirated error M&D 4208 Pump/Valves Test
ICT assay result P-4010 ICT Alignment
imprecision,
calibration failure
ICT assay calibration
failure
ICT Reference Bubbles in syringe M&D 5101 Flush ICT Reference
Solution Pump #12 and/or ICT cup Solution Cup
fill errors M&D 4208 Pump/Valves Test
Erratic ICT results M&D 5804 Replace 1mL Syringes
IREF cup overflow Inspect/replace check valve
ICT Result Unable to calculate ICT module cable failure or loose
Troubleshooting results, ICT Reference connection
Solution voltage drift ICT Preamp board failure
errors
M&D 5101 Flush ICT Reference
Unable to calculate Solution Cup
results, mV reading
outside measurable M&D 5102 Flush ICT Module
range of ICT
assembly
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 5-41
Fluidics Review – Fluidics
Review – Fluidics
Time: 10 mins
Internal Probe
Wash Pump
#12 syringe
5-42 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Optics
6 Optics
Objectives
Use service tools and diagnostic tools to interpret data and isolate
root causes of optic control system failures
Module Topics
Optics activities
Summary - Optics
Review Optics
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 6-1
Optics Optics System Overview
Directs and aligns the light from the source lamp, through the water
bath and cuvette, and then into the optics assembly
The Alinity c is capable of taking monochromatic (single
wavelength) and bi-chromatic (dual wavelength) measurements
at 16 different wavelengths.
6-2 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Optics System Components Optics
Optics Housing
Component Description
With use or age, the lamp heat may cause faint brown
spots that may impact some assay reads and cuvette
integrity issues
3. Convex Lens Focuses the light from a tungsten halogen lamp and passes the
(2) light through the cuvette to measure absorbance changes as
the reaction progresses
A second convex lens focuses the light through the entrance slit
onto the diffraction grating
4. Water Bath Separates and seals lamp and optics assembly from water bath
Lens Replaceable
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 6-3
Optics Optics System Components
Component Description
7. Diffraction Breaks the light beam into 16 wavelengths and reflects the light
Grating onto a photo-diode array
9. Optics Pre- Converts optical signal to a voltage (0-5V), amplifies signal, and
Amp Board sends to DAQ board
10. Data Receives photometric data from the optics preamp board,
acquisition quantifies and converts optical signals to a digital format then
(DAQ) board sends data to the CPU
11. Central Receives DAQ photometric data and sends it directly to the UIC
processing
unit (CPU) Note: Boards were reviewed previously.
board
Cuvette Wipers Remove debris and bubbles from cuvettes prior to optical reads
Replaceable
Optical Trigger Monitors positional flags on cuvette segments and signals DAQ
Sensor board when cuvette is positioned in front of the Optics for a
read
Sensor is replaceable; mounting bracket MUST be reused- NOT
replaceable
Resets Optics wavelength gains and cuvette read offset values
during RUNNING status
6-4 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Optics Photometric Sample Processing Optics
The following describes the optic processes utilized during an assay run
and the components involved.
Note: When reaction carousel is homed and in the 4.5 second lock
step wait, the cuvette spoke blocks the light path.
Note: Nozzle 6 dispenses the degassed water for the cuvette blank
reading.
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 6-5
Optics Optics Photometric Sample Processing
Lamp Monitoring
6-6 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Optics Photometric Sample Processing Optics
As the cuvettes pass between the light source and the optics
assembly, the trigger sensor signals the DAQ board that a cuvette is
in the light path and in position for an optical reading
Light passes through the cuvette onto the diffraction grating in the
optics assembly:
Diffraction grating breaks the beam into 16 wavelengths and
reflects the light onto a photo diode array
Photo diode array detects and measures changes in the light
intensity
Photo diode array signal is sent to the preamp board where it is
converted to a voltage and amplified
Preamp board sends the signal to the DAQ board in the master
card cage
The cuvette washer dispenses DI water into the cuvette for the
water blank measurement
The cuvette washer aspirates the water and dries the cuvette in
preparation for the next sample
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 6-7
Optics Optics Activities
Optics Activities
Time: 60 mins
Materials Needed:
Standard tool kit
Beam Height Alignment Tool
Service Tools:
The Beam height alignment is typically only done on installation or
after replacing identified components.
Handle lamp with gloved hands.
Cuvettes are only replaceable as a segment.
A run must be generated to obtain cuvette integrity results. The
cuvette integrity results are readings obtained from the last time
the cuvette blank was performed.
_________________________________________________________________
6-8 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Optics Activities Optics
_________________________________________________________________
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 6-9
Optics Optics Activities
Time: 15 mins
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
6-10 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Optics Activities Optics
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
Time: 10 mins
Action: Refer to the Ticket Code Tool to find the best Work Done Code
to represent if the Degasser was replaced to resolve bubbles error.
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 6-11
Optics Optics Hardware Summary
6-12 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Review – Optics Optics
Review – Optics
Time: 10 mins
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 6-13
Temperature Control
7 Temperature Control
Objectives
Use service tools and diagnostic tools to interpret data and isolate
root causes of temperature control system failures
Module Topics
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 7-1
Temperature Control Temperature Control Overview
Temperature Control
Temperature Monitoring
7-2 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Temperature Control Overview Temperature Control
SM AC/DC
Actual
Area Thermistor Controller
Temp Range Temperature
Monitored Location Displayed Temp
Adjustment
Adjustment
Reagent 2–10°C Bottom of N/A N/A
Supply Reagent
Center Carousel area
Water Bath 36.8–37.2°C Bottom of Water VR2 TBC Bd
(see Note below) Bath
Ambient 15–30°C WMU VR1 N/A
Component Function
Ambient Located in the WMU area to monitor internal instrument temperature
Thermistor Temperature specification 15°–30°C
Control: SM AC/DC Controller; VR1 adjustment potentiometer adjusts
display temperature only
Replaceable
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 7-3
Temperature Control Water Bath Temperature Control System Components
The water bath heater and circulation pump are responsible for
circulating heated water (37°C) to the water bath and ICT module
holder/jacket.
Component Function
Heater Located in the Water Management Unit
Heats degassed DI water supplied to the Water Bath to 37°C ±0.2°C
Controlled from Thermal Bath Controller (TBC) board, located in the
driver box
Voltage ramps up and down as heat to maintain bath
temperature
Water bath heater is turned OFF any time the float switch detects
a low water level condition
Replaceable
Water Bath Located at bottom of Water Bath to monitor bath temperatures
Control Temperature specification: 37°C ±0.2°C
Thermistor
Temperatures outside 36.8°–37.2°C signals SM AC/DC Controller
Board to generate errors
Temperature reported to the SM AC/DC Controller; VR2 adjustment
potentiometer adjusts display temperature only
Replaceable
RD20 Pump Located in WMU and provides water circulation
(Water bath Is ON constantly when the water bath float switch indicates full
circulation
pump) 24VDC power from TBC board
Control: SM AC/DC controller board
Replaceable
7-4 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Water Bath Temperature Control System Components Temperature Control
Component Function
Temperature Located in the Driver Box
Bath Controller Controls actual water bath temperature
Board (TBC)
Provides 24VDC to RD20 circulating pump
Provides 100VAC to water bath heating element
Feedback control through thermistor
Contains circuitry to open and close the 24VAC motor drain valve
used to empty the water bath during M&D procedures
Contains two LEDs to indicate status of heater
CPU CMD LED:2
Displays ON/OFF signal status from CPU
ON when signal is output and water bath float switch indicates
full
Heater LED: 1
ON when heater is activated
Dims when temperature is constant and temperature control
stabilized
Contains two switches to adjust the water temperature in the water
bath.
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 7-5
Temperature Control Reagent Supply Center Cooler Unit
Component Function
Thermoelectric Cools reagent supply center assay reagents to 2-10°C
Devices (TEDs) Control: Cooler Controller Board
(8)
Power: PS5 and PS8 provide 12VDC for the TEDs
TEDs are not individually replaceable. Replace entire cooling unit.
7-6 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Reagent Supply Center Cooler Unit Temperature Control
Component Function
Reagent Contains float sensor and signals sent to Cooler Controller Board
Coolant Bottle Coolant is Abbott customized ethylene glycol solution added during
installation, topped off annually, or as needed
Reagent Cover heaters are placed around the two reagent aspiration holes
Carousel Lid Prevents condensation from forming around the opening
Heaters (2)
Minimize probe/reagent contamination due to condensation drips
Monitored by Cooler Controller Board
Replaceable
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 7-7
Temperature Control Reagent Supply Center Cooler Unit
Time: 30 mins
Materials Needed:
Standard Tool Kit
Calibrated thermometer (range + 0.01 degrees)
Service Tips:
Displayed Water Bath and Ambient temperatures are adjustable
on the SM AC/DC Controller.
Water in the bath is circulated when the float switch indicates full.
____________________________________________________________
____________________________________________________________
Discussion: Check understanding and clarify key points.
7-8 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Reagent Supply Center Cooler Unit Temperature Control
Time: 15 mins
Materials needed:
Coolant for demonstration
Coolant kit (tubing & funnel) for demonstration
Service Tips:
Use only the specific Cooler Unit Coolant to prevent errors and
component damage.
Coolant is formulated for Abbott and must be obtained from
Abbott.
The Cooler coolant fluid line should be free from bubbles.
The Reagent Supply temperature is NOT adjustable
Coolant line MUST be free of air bubbles
________________________________________________________________
________________________________________________________________
________________________________________________________________
________________________________________________________________
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 7-9
Temperature Control Temperature – Hardware Summary
7-10 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Review - Temperature Control Temperature Control
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 7-11
Installation and Integration
Objectives
Module Topics
Installation
Review
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 8-1
Installation and Integration
Installation
Time: 5 mins
Installation Documentation
Pre-site inspection
Pre-site installation checklist
8-2 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Installation Activities Installation and Integration
Installation Activities
Time: 40 mins
Actions: Perform the following actions and steps included ISA 210-001
Alinity c Installation Procedure; Operation Qualification- Performance
Verification Testing (PVT) for the following activity.
Difference
Assay High Low (High - Pass/Fail
Low)
AST
ALKP
TP
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 8-3
Installation and Integration Multimodule Integration Installation Activity
Difference
Assay High Low (High - Pass/Fail
Low)
Na
CI
Video: Training videos are intended for class instruction only and
SHOULD NOT BE DISTRIBUTED.
Actions: Open the following procedures listed below and follow along
during the video review:
Alinity c-series Module Installation procedure
Alinity i-series Module Installation procedure
Alinity ci-series multimodule installation and integration
procedure
8-4 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Multimodule Integration Installation Activity Installation and Integration
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 8-5
Installation and Integration Prognostic Health Management and Preventive Maintenance
8-6 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Prognostic Health Management and Preventive Maintenance Installation and Integration
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 8-7
Installation and Integration Prognostic Health Management & Preventive Maintenance Activities
Time: 10 mins Purpose: To be proficient in service skills associated with AST tools,PM
and related procedures.
Action: Using the History and Maintenance Log saved earlier in the
week, load and view log in LOG-ICAL and then answer the following:
________________________________________________________________
________________________________________________________________
5. How could you utilize these logs prior to going on site to perform a
PM?
________________________________________________________________
8-8 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Prognostic Health Management & Preventive Maintenance Activities Installation and Integration
6. Choose the best AST Tool to use and review the overall instrument
and account health Real Time for ASE Proactive Customer Care. Be
prepared to explain your answer. (Choose one)
a. AbbottLink
a. ONeview [*]
a. LOG-ICAL
a. HCI GSR report
Answer:
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 8-9
Installation and Integration Prognostic Health Management & Preventive Maintenance Activities
Review
Time: 5 mins 1) What are the required conditions for PHN Alerts?
8-10 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Lab Practical and Final Exam
Objectives
Module Topics
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 9-1
Lab Practical and Final Exam Overview of Lab Practical and Final Exam
Note: Instructor will observe use of tools and resources and provide
participant feedback, as needed.
Time: 60 mins
9-2 Alinity c Global Field Service Training | Internal Use Only | 80001189-101
Overview of Lab Practical and Final Exam Lab Practical and Final Exam
Key Notes
Alinity c Global Field Service Training | Internal Use Only | 80001189-101 9-3
Appendix - Additional Handouts and Information
Appendix - Additional
A Handouts and Information
Alinity c-series Field Service Training | Global Service Training | 80001189-101 A-1
Appendix - Additional Handouts and Information
Note: The steps above were created for Alinity ci-series Software
version 1.0. Operations steps may vary depending on software
version used. Please refer to the current Alinity ci-series Operations
Manual for additional detail.
A-2 Alinity c-series Field Service Training | Global Service Training | 80001189-101
Appendix - Additional Handouts and Information
Alinity c-series Field Service Training | Global Service Training | 80001189-101 A-3
Appendix - Additional Handouts and Information
Component Illustration
1. Water Bath Filter
2. Low Concentration Waste Tank
3. Inlet Water Valve and Flow
Restrictor
4. DI Water and External Waste
Tubing Connectors
5. Water Bath Fill Peristaltic Pump
6. Water Heater
7. Water Bath Drain Valve
8. External Probe/Mixer/Cuvette
Wash RD30 24 VDC Water Pump
9. Internal Probe Wash 24 VDC
Water Pump with Damper
10. Water Bath Circulation RD20 24
VDC Water Pump
11. DI Water Tank
12. Connector Panel
13. Float Sensors
A-4 Alinity c-series Field Service Training | Global Service Training | 80001189-101
Appendix - Additional Handouts and Information
ng
Alinity c-series Field Service Training | Global Service Training | 80001189-101 A-5