AU5800 Monthly Maintenance Job Aids

For Training Purposes Only


These job aids are shortened versions of the procedures found in the source below. The
procedures are written as standalone procedures to ensure they can be performed in
any order. The optional ISE module has references for two flow cells. Information in the
job aid is correct as of the date published. Verify you have the correct information. Part
numbers for supplies required for the maintenance procedures are found in the reference
listed below.

Source: AU5800® Chemistry Analyzer User’s Guide PN A98352AB (October 2012)

AU5800, Customer, Monthly Maintenance Job Aids Page 1 of 9


Rev. A (May 2013) Beckman Coulter Professional Development
Document Disclaimers

Document This document is not intended to replace the information in your User’s
Disclaimers Guides, Quick Response Guide or other product documentation.
Information in the User’s Guide and Quick Response Guide
supersedes information in any other manual.

Warnings Read all product manuals and consult with Beckman Coulter trained
and personnel before attempting to operate the instrument.
Precautions
Beckman Coulter, Inc. urges its customers and employees to comply
with all national health and safety standards such as the use of barrier
protection. This may include but is not limited to, protective eyewear,
gloves, suitable laboratory attire when operating or maintaining this or
any other automated laboratory equipment.

Intention for This document is not intended to replace the information in your
Use User’s Guide and Quick Response Guide. Information in the
User’s Guide supersedes information in any other manual.

Revision Rev. A (May 2013)


Status Software version 4.11

Trademarks AU5800® Chemistry Analyzer

AU5800, Customer, Monthly Maintenance Job Aids Page 2 of 9


Rev. A (May 2013) Beckman Coulter Professional Development
AU5800, Customer, Monthly Maintenance Job Aids Page 3 of 9
Rev. A (May 2013) Beckman Coulter Professional Development
AU5800, Customer, Monthly Maintenance Job Aids Page 4 of 9
Rev. A (May 2013) Beckman Coulter Professional Development
Clean the Mix Bar Wash Wells

Ensure the system is in


Standby Mode

Supplies Required:
Open the front and rear main cover
Cotton swabs on each analyzer unit
Disposable
transfer pipette
10% Bleach Manually turn the mix bar units so
(Beckman
the mix bars are away from the
Sodium
wash wells
Hypochlorite
(5%) diluted Select Replace Mix Bar. Make the
1:10. Prepare following selections:
by adding 10 Use a transfer pipette to dispense
All units are selected by default,
parts bleach to 10% bleach solution into each
deselect units not required
90 parts wash well. Use a separate cotton
DI water)
Select The Second Mixer
swab to clean each wash well.
At Times, enter 1, select OK
Avoid splashing and clean up
spills immediately
Press the green DIAG button to
initialize the R2 mix bar unit
From the Home screen, select
the Analyzer Maintenance
jump button Select Prime Washing Line. Make
the following selections:
All units are selected by default,
Place a in the deselect units not required
Analyzer Maintenance check box At Times, enter 1, select OK

Select Replace Mix Bar. Make the Press the green DIAG button. The
following selections: analyzer will prime DI water through
All units are selected by default, the sample and reagent probes, mix
deselect units not required bars, and the wash nozzle. Inspect
Select The First Mixer the wash wells for proper drainage
At Times, enter 1, select OK

Press the green DIAG button to


initialize the R1/S mix bar unit
No
Are wash wells
draining properly?

Yes

Document you completed Deselect the Analyzer


the procedure on the paper Maintenance check box. Close
maintenance log the front and rear main covers

AU5800, Customer, Monthly Maintenance Job Aids Page 5 of 9


Rev. A (May 2013) Beckman Coulter Professional Development
Clean the Wash Nozzle Unit and Check the Tube Mounting Joints

Ensure the system is in Standby Mode

Open the rear main cover of each analyzer unit


Supplies
Required:

Dry clean From the Home screen, select the


cloth or Analyzer Maintenance jump button
paper
towel
Sonicator
with DI Place a in the
water Analyzer Maintenance check box

Select Replace Wash Nozzle (systems with


multiple units are all selected by default, deselect
units if not required), select OK
Always drain the
fluid from the wash
nozzle before
Press the green DIAG button removing the
tubing to avoid
to drain the liquid from the tubing
spills in the
cuvette wheel

Loosen the 6 grey manifolds and remove them


from their mounting positions. Inspect the 6 o-rings
in the water supply manifold and verify all o-rings are
present, not worn (over stretched), or torn and that
the area is free of dust or detergent crystals. Clean
or replace any o-ring if necessary

Loosen the silver knob holding the wash nozzle unit


and lift the wash nozzle unit up over the
positioning screws
Each wash nozzle
has been aligned
to a specific
Remove the wash nozzle unit along with tubing and analyzer unit.
inspect the joints (flexible tubing joining the metal Ensure the wash
nozzle to the tubing) for cracks. Replace any nozzle is returned
joints as necessary to the correct
analyzer unit

Yes No
Do you have
Continue on the
another wash nozzle
to remove?
next page

AU5800, Customer, Monthly Maintenance Job Aids Page 6 of 9


Rev. A (May 2013) Beckman Coulter Professional Development
AU5800, Customer, Monthly Maintenance Job Aids Page 7 of 9
Rev. A (May 2013) Beckman Coulter Professional Development
AU5800, Customer, Monthly Maintenance Job Aids Page 8 of 9
Rev. A (May 2013) Beckman Coulter Professional Development
AU5800, Customer, Monthly Maintenance Job Aids Page 9 of 9
Rev. A (May 2013) Beckman Coulter Professional Development