Document This document is not intended to replace the information in your User’s
Disclaimers Guides, Quick Response Guide or other product documentation.
Information in the User’s Guide and Quick Response Guide
supersedes information in any other manual.
Warnings Read all product manuals and consult with Beckman Coulter trained
and personnel before attempting to operate the instrument.
Precautions
Beckman Coulter, Inc. urges its customers and employees to comply
with all national health and safety standards such as the use of barrier
protection. This may include but is not limited to, protective eyewear,
gloves, suitable laboratory attire when operating or maintaining this or
any other automated laboratory equipment.
Intention for This document is not intended to replace the information in your
Use User’s Guide and Quick Response Guide. Information in the
User’s Guide supersedes information in any other manual.
Supplies Required:
Open the front and rear main cover
Cotton swabs on each analyzer unit
Disposable
transfer pipette
10% Bleach Manually turn the mix bar units so
(Beckman
the mix bars are away from the
Sodium
wash wells
Hypochlorite
(5%) diluted Select Replace Mix Bar. Make the
1:10. Prepare following selections:
by adding 10 Use a transfer pipette to dispense
All units are selected by default,
parts bleach to 10% bleach solution into each
deselect units not required
90 parts wash well. Use a separate cotton
DI water)
Select The Second Mixer
swab to clean each wash well.
At Times, enter 1, select OK
Avoid splashing and clean up
spills immediately
Press the green DIAG button to
initialize the R2 mix bar unit
From the Home screen, select
the Analyzer Maintenance
jump button Select Prime Washing Line. Make
the following selections:
All units are selected by default,
Place a in the deselect units not required
Analyzer Maintenance check box At Times, enter 1, select OK
Select Replace Mix Bar. Make the Press the green DIAG button. The
following selections: analyzer will prime DI water through
All units are selected by default, the sample and reagent probes, mix
deselect units not required bars, and the wash nozzle. Inspect
Select The First Mixer the wash wells for proper drainage
At Times, enter 1, select OK
Yes
Yes No
Do you have
Continue on the
another wash nozzle
to remove?
next page