Attachment A
Detecting and Resolving Cuvette Overflow
A) Definition of Cuvette Overflow
After each measurement, the reaction fluid in a cuvette is aspirated, and the cuvette is washed
and dried. In the case of a system malfunction, fluid may spill over onto the outside of the
cuvette and affect the next measurement in that cuvette or a neighboring cuvette, potentially
leading to an incorrect result (both falsely high and falsely low).
Our internal investigation has indicated that even a minor cuvette overflow can affect assay
results and reaction monitors for at least the following 3 hours.
FSN-14563
Beckman Coulter United Kingdom Limited Telephone 01494 441181
Oakley Court Fax 01494 429291
Kingsmead Business Park e-mail infouk@beckman.com
London Road High Wycombe Website www.beckmancoulter.co.uk
Bucks HP11 1JU Registered in England no 640961 Registered Office as shown
Page 1
Attachment A
Detecting and Resolving Cuvette Overflow
C) Confirming a Cuvette Overflow
If you identify an abnormal reaction monitor for an Overflow Indicator Assay, an instrument
malfunction may have occurred. To confirm whether a cuvette overflow occurred, perform the
following checks:
• Verify the reaction monitors of other samples run prior to your first abnormal sample. If
these reaction monitors are also abnormal, assume that a cuvette overflow has occurred.
• Examine the tops of the reaction cuvettes, following the instruction in your User Guide.
The tops of the cuvettes should appear frosty or white. If they are dark or black, remove
the cuvettes from the wheel to determine if an overflow has occurred.
If you have confirmed a cuvette overflow occurred, proceed to Section D.
If your prior reaction monitors for Overflow Indicator Assays are normal, and the tops of the
reaction cuvettes are frosty or white, then the reason for the abnormal reaction monitor is still
unknown and standard trouble-shooting actions are required, as described in the system User
Guide.
The following instrument malfunctions have been identified as the primary causes of cuvette
overflows:
• Blocked or partially blocked cuvette wash station nozzles
• Valve failure
• Misalignment of wash station components
• Disconnected wash station tubing
• Deterioration of wash station o-rings
To identify a cause, check the wash station for nozzle blockages, valve failures, misalignment,
disconnected tubing or worn o-rings.
As always, daily maintenance is essential to identify hardware deterioration before it affects
system performance. Many cuvette overflow incidents are the result of incorrect handling of
the cuvette Wash Station during user maintenance. When parts are removed or replaced as
part of user maintenance, you must ensure they are handled and repositioned correctly. Refer
to the Maintenance section of your User Guide for correct instructions and ensure that users
who perform maintenance are sufficiently trained to perform those maintenance tasks.
If the reason for cuvette overflow is not identified, contact Beckman Coulter for service.
If you have identified the cause and repaired the instrument, perform the Overflow Recovery
steps describe in your User Guide, including passing the photocal, to achieve standard
operation of the system.
After proper system operation is verified, rerun all samples between
the abnormal Overflow Indicator Assay and the last normal Overflow
Indicator Assay.
FSN-14563
Beckman Coulter United Kingdom Limited Telephone 01494 441181
Oakley Court Fax 01494 429291
Kingsmead Business Park e-mail infouk@beckman.com
London Road High Wycombe Website www.beckmancoulter.co.uk
Bucks HP11 1JU Registered in England no 640961 Registered Office as shown
Page 2
Attachment A – Section E
Reaction Monitor Libraries
E) Reaction Monitor Libraries
This section provides examples of normal and abnormal reaction monitors for Overflow
Indicator Assays on specific AU Systems. In assessing the reaction monitors for the
Overflow Indicator Assays, please refer to the appropriate section (E1- E5) for your particular
system, specifically:
FSN-14563
Beckman Coulter United Kingdom Limited Telephone 01494 441181
Oakley Court Fax 01494 429291
Kingsmead Business Park e-mail infouk@beckman.com
London Road High Wycombe Website www.beckmancoulter.co.uk
Bucks HP11 1JU Registered in England no 640961 Registered Office as shown
Page 3
Attachment A- Section E2
AU 680 Reaction Monitor Library
Section E2: AU680 Reaction Monitor Library:
FSN-14563
Beckman Coulter United Kingdom Limited Telephone 01494 441181
Oakley Court Fax 01494 429291
Kingsmead Business Park e-mail infouk@beckman.com
London Road High Wycombe Website www.beckmancoulter.co.uk
Bucks HP11 1JU Registered in England no 640961 Registered Office as shown
Section E2 - Page 4
Attachment A- Section E2
AU 680 Reaction Monitor Library
AU 680 Creatinine (OSR6x78) Normal
FSN-14563
Beckman Coulter United Kingdom Limited Telephone 01494 441181
Oakley Court Fax 01494 429291
Kingsmead Business Park e-mail infouk@beckman.com
London Road High Wycombe Website www.beckmancoulter.co.uk
Bucks HP11 1JU Registered in England no 640961 Registered Office as shown
Section E2 - Page 5
Attachment A- Section E2
AU 680 Reaction Monitor Library
AU680 Creatinine (OSR6x78) Abnormal
FSN-14563
Beckman Coulter United Kingdom Limited Telephone 01494 441181
Oakley Court Fax 01494 429291
Kingsmead Business Park e-mail infouk@beckman.com
London Road High Wycombe Website www.beckmancoulter.co.uk
Bucks HP11 1JU Registered in England no 640961 Registered Office as shown
Section E2 - Page 6
Attachment A- Section E2
AU 680 Reaction Monitor Library
AU680 C-Reactive Protein (OSR6x99) Normal
FSN-14563
Beckman Coulter United Kingdom Limited Telephone 01494 441181
Oakley Court Fax 01494 429291
Kingsmead Business Park e-mail infouk@beckman.com
London Road High Wycombe Website www.beckmancoulter.co.uk
Bucks HP11 1JU Registered in England no 640961 Registered Office as shown
Section E2 - Page 7
Attachment A- Section E2
AU 680 Reaction Monitor Library
AU680 C-Reactive Protein (OSR6x99) Abnormal
FSN-14563
Beckman Coulter United Kingdom Limited Telephone 01494 441181
Oakley Court Fax 01494 429291
Kingsmead Business Park e-mail infouk@beckman.com
London Road High Wycombe Website www.beckmancoulter.co.uk
Bucks HP11 1JU Registered in England no 640961 Registered Office as shown
Section E2 - Page 8
Attachment A- Section E2
AU 680 Reaction Monitor Library
AU680 Iron (OSR6x86) Normal
FSN-14563
Beckman Coulter United Kingdom Limited Telephone 01494 441181
Oakley Court Fax 01494 429291
Kingsmead Business Park e-mail infouk@beckman.com
London Road High Wycombe Website www.beckmancoulter.co.uk
Bucks HP11 1JU Registered in England no 640961 Registered Office as shown
Section E2 - Page 9
Attachment A- Section E2
AU 680 Reaction Monitor Library
AU680 Iron (OSR6x86) Abnormal
FSN-14563
Beckman Coulter United Kingdom Limited Telephone 01494 441181
Oakley Court Fax 01494 429291
Kingsmead Business Park e-mail infouk@beckman.com
London Road High Wycombe Website www.beckmancoulter.co.uk
Bucks HP11 1JU Registered in England no 640961 Registered Office as shown
Section E2 - Page 10
Attachment A- Section E2
AU 680 Reaction Monitor Library
AU680 Direct Bilirubin Color (OSR6x11) Normal
FSN-14563
Beckman Coulter United Kingdom Limited Telephone 01494 441181
Oakley Court Fax 01494 429291
Kingsmead Business Park e-mail infouk@beckman.com
London Road High Wycombe Website www.beckmancoulter.co.uk
Bucks HP11 1JU Registered in England no 640961 Registered Office as shown
Section E2 - Page 11
Attachment A- Section E2
AU 680 Reaction Monitor Library
AU680 Direct Bilirubin Color (OSR6x11) Abnormal
FSN-14563
Beckman Coulter United Kingdom Limited Telephone 01494 441181
Oakley Court Fax 01494 429291
Kingsmead Business Park e-mail infouk@beckman.com
London Road High Wycombe Website www.beckmancoulter.co.uk
Bucks HP11 1JU Registered in England no 640961 Registered Office as shown
Section E2 - Page 12
Attachment A- Section E2
AU 680 Reaction Monitor Library
FSN-14563
Beckman Coulter United Kingdom Limited Telephone 01494 441181
Oakley Court Fax 01494 429291
Kingsmead Business Park e-mail infouk@beckman.com
London Road High Wycombe Website www.beckmancoulter.co.uk
Bucks HP11 1JU Registered in England no 640961 Registered Office as shown
Section E2 - Page 13
Attachment A- Section E2
AU 680 Reaction Monitor Library
AU680 Ceruloplasmin (OSR6x64) Abnormal
FSN-14563
Beckman Coulter United Kingdom Limited Telephone 01494 441181
Oakley Court Fax 01494 429291
Kingsmead Business Park e-mail infouk@beckman.com
London Road High Wycombe Website www.beckmancoulter.co.uk
Bucks HP11 1JU Registered in England no 640961 Registered Office as shown
Section E2 - Page 14
Attachment A- Section E2
AU 680 Reaction Monitor Library
AU680 Magnesium (OSR6x89) Normal
FSN-14563
Beckman Coulter United Kingdom Limited Telephone 01494 441181
Oakley Court Fax 01494 429291
Kingsmead Business Park e-mail infouk@beckman.com
London Road High Wycombe Website www.beckmancoulter.co.uk
Bucks HP11 1JU Registered in England no 640961 Registered Office as shown
Section E2 - Page 15
Attachment A- Section E2
AU 680 Reaction Monitor Library
AU680 Magnesium (OSR6x89) Abnormal
FSN-14563
Beckman Coulter United Kingdom Limited Telephone 01494 441181
Oakley Court Fax 01494 429291
Kingsmead Business Park e-mail infouk@beckman.com
London Road High Wycombe Website www.beckmancoulter.co.uk
Bucks HP11 1JU Registered in England no 640961 Registered Office as shown
Section E2 - Page 16
Attachment A- Section E2
AU 680 Reaction Monitor Library
AU680 Total Bilirubin (OSR6x12) Normal
FSN-14563
Beckman Coulter United Kingdom Limited Telephone 01494 441181
Oakley Court Fax 01494 429291
Kingsmead Business Park e-mail infouk@beckman.com
London Road High Wycombe Website www.beckmancoulter.co.uk
Bucks HP11 1JU Registered in England no 640961 Registered Office as shown
Section E2 - Page 17
Attachment A- Section E2
AU 680 Reaction Monitor Library
AU680 Total Bilirubin (OSR6x12) Abnormal
FSN-14563
Beckman Coulter United Kingdom Limited Telephone 01494 441181
Oakley Court Fax 01494 429291
Kingsmead Business Park e-mail infouk@beckman.com
London Road High Wycombe Website www.beckmancoulter.co.uk
Bucks HP11 1JU Registered in England no 640961 Registered Office as shown
Section E2 - Page 18
Attachment A- Section E2
AU 680 Reaction Monitor Library
AU680 Uric Acid (OSR6x98) Normal
FSN-14563
Beckman Coulter United Kingdom Limited Telephone 01494 441181
Oakley Court Fax 01494 429291
Kingsmead Business Park e-mail infouk@beckman.com
London Road High Wycombe Website www.beckmancoulter.co.uk
Bucks HP11 1JU Registered in England no 640961 Registered Office as shown
Section E2 - Page 19
Attachment A- Section E2
AU 680 Reaction Monitor Library
AU680 Uric Acid (OSR6x98) Abnormal
FSN-14563
Beckman Coulter United Kingdom Limited Telephone 01494 441181
Oakley Court Fax 01494 429291
Kingsmead Business Park e-mail infouk@beckman.com
London Road High Wycombe Website www.beckmancoulter.co.uk
Bucks HP11 1JU Registered in England no 640961 Registered Office as shown
Section E2 - Page 20
Attachment A- Section E2
AU 680 Reaction Monitor Library
AU680 Glucose (OSR6x21) Normal
FSN-14563
Beckman Coulter United Kingdom Limited Telephone 01494 441181
Oakley Court Fax 01494 429291
Kingsmead Business Park e-mail infouk@beckman.com
London Road High Wycombe Website www.beckmancoulter.co.uk
Bucks HP11 1JU Registered in England no 640961 Registered Office as shown
Section E2 - Page 21
Attachment A- Section E2
AU 680 Reaction Monitor Library
AU680 Glucose (OSR6x21) Abnormal
FSN-14563
Beckman Coulter United Kingdom Limited Telephone 01494 441181
Oakley Court Fax 01494 429291
Kingsmead Business Park e-mail infouk@beckman.com
London Road High Wycombe Website www.beckmancoulter.co.uk
Bucks HP11 1JU Registered in England no 640961 Registered Office as shown
Section E2 - Page 22
Attachment A- Section E2
AU 680 Reaction Monitor Library
AU680 Phosphorus (OSR6x22) Normal
FSN-14563
Beckman Coulter United Kingdom Limited Telephone 01494 441181
Oakley Court Fax 01494 429291
Kingsmead Business Park e-mail infouk@beckman.com
London Road High Wycombe Website www.beckmancoulter.co.uk
Bucks HP11 1JU Registered in England no 640961 Registered Office as shown
Section E2 - Page 23
Attachment A- Section E2
AU 680 Reaction Monitor Library
AU680 Phosphorus (OSR6x22) Abnormal
FSN-14563
Beckman Coulter United Kingdom Limited Telephone 01494 441181
Oakley Court Fax 01494 429291
Kingsmead Business Park e-mail infouk@beckman.com
London Road High Wycombe Website www.beckmancoulter.co.uk
Bucks HP11 1JU Registered in England no 640961 Registered Office as shown
Section E2 - Page 24
Attachment A- Section E2
AU 680 Reaction Monitor Library
AU680 Triglyceride (OSR6x118) Normal
FSN-14563
Beckman Coulter United Kingdom Limited Telephone 01494 441181
Oakley Court Fax 01494 429291
Kingsmead Business Park e-mail infouk@beckman.com
London Road High Wycombe Website www.beckmancoulter.co.uk
Bucks HP11 1JU Registered in England no 640961 Registered Office as shown
Section E2 - Page 25
Attachment A- Section E2
AU 680 Reaction Monitor Library
AU680 Triglyceride (OSR6x118) Abnormal
FSN-14563
Beckman Coulter United Kingdom Limited Telephone 01494 441181
Oakley Court Fax 01494 429291
Kingsmead Business Park e-mail infouk@beckman.com
London Road High Wycombe Website www.beckmancoulter.co.uk
Bucks HP11 1JU Registered in England no 640961 Registered Office as shown
Section E2 - Page 26
Attachment A- Section E2
AU 680 Reaction Monitor Library
AU680 Cholesterol (OSR6x16) Normal
FSN-14563
Beckman Coulter United Kingdom Limited Telephone 01494 441181
Oakley Court Fax 01494 429291
Kingsmead Business Park e-mail infouk@beckman.com
London Road High Wycombe Website www.beckmancoulter.co.uk
Bucks HP11 1JU Registered in England no 640961 Registered Office as shown
Section E2 - Page 27
Attachment A- Section E2
AU 680 Reaction Monitor Library
AU680 Cholesterol (OSR6x16) Abnormal
FSN-14563
Beckman Coulter United Kingdom Limited Telephone 01494 441181
Oakley Court Fax 01494 429291
Kingsmead Business Park e-mail infouk@beckman.com
London Road High Wycombe Website www.beckmancoulter.co.uk
Bucks HP11 1JU Registered in England no 640961 Registered Office as shown
Section E2 - Page 28
Attachment A- Section E2
AU 680 Reaction Monitor Library
AU680 Total Protein (OSR6x32) Normal
FSN-14563
Beckman Coulter United Kingdom Limited Telephone 01494 441181
Oakley Court Fax 01494 429291
Kingsmead Business Park e-mail infouk@beckman.com
London Road High Wycombe Website www.beckmancoulter.co.uk
Bucks HP11 1JU Registered in England no 640961 Registered Office as shown
Section E2 - Page 29
Attachment A- Section E2
AU 680 Reaction Monitor Library
AU680 Total Protein (OSR6x32) Abnormal
FSN-14563
Beckman Coulter United Kingdom Limited Telephone 01494 441181
Oakley Court Fax 01494 429291
Kingsmead Business Park e-mail infouk@beckman.com
London Road High Wycombe Website www.beckmancoulter.co.uk
Bucks HP11 1JU Registered in England no 640961 Registered Office as shown
Section E2 - Page 30