You are on page 1of 3

ELECTRO SURGICAL UNIT

Brand LAMIDEY NOURY


Type SURGILEC OPTIMA Electro Surgery Unit on Trolley
Origin FRANCE

Technical Data of Surgilec Optima


SPECIAL FEATURES
Display Friendly user interface with an LCD touch screen
Program Library of more than 80 programs for all surgical specialties
Personal Program Saving more than 1000 personal programs
1. Digestive Endoscopy : Mucosectomy, Mucosectomy2, Pedicle Polyp, Sphinctererctomy. 2.Laparoscopy : Broncial
Endoscopy, General Surgery, Gynecology, Urology; 3. Open Surgery : ENT Surgery, General Surgery, Gynecology,
Progam by Specialities Neurosurgery, Orthopedicsurgery, Plastic Surgery, Thoracsic Cardiovascular, Urology
4. Resection : Gynecology (Bipolar Hysterorsection, Monopolar Hysteroresection ); Urology (Bipolar TURP,Bipolar
TURV, Monopolar TURP, Monopolar TUR P)
Output 4 High Frequency output : 2 Monopolar and 2 Bipolar
Non Contact Coagulation Fulguration / Spray coagulation, without contact
Team Coag simultaneous coagulation from 2 monopolar outputs
Monopolar and Bipolar Currents A real hemostatic cut with a white coagulation without charring and resection without bleeding
Automatic activation and stop of the bipolar coagulation, by sensing the tissues contact impedance on the forceps tips.
Automatic Bipolar Coagulation
3 available activation delays
The principle of independent self-running security is a general principle rulling the conception of OPTIMA. When
Smart Security Management security problem detected, OPTIMA start immediately current off without letting the information going through a
central system which many give wrong information.
INPUT FEATURES
Main supply 230 V 50 or 60 Hz
Safety Class 1 type CF
Maximum power 1200 VA (600 W)
OUTPUT FEATURES
Waveform sinus
Frequency Emitted 450 KHz
Modulation Frequency 21 KHz (pulsed modulaton for coagulation and haemostatic cutting)
MONOPOLAR CUTS
FORCED PURE CUT : 370 W / 450 Ω, forced cut without haemostatic effect
Forced Cut FORCED BLEND CUT 1 : 340 W / 450 Ω, forced cut a light haemostatic effect
FORCED BLEND CUT 2 : 330 W / 450 Ω, forced cut deep haemostatic effect
PURE CUT: 360 W / 250 Ω without haemostatic effect
Autocut BLEND CUT 1 : 330 W / 250 Ω, light haemostatic effect
BLEND CUT 2 : 320 W / 250 Ω, deep haemostatic effect
UROCUT P : 350 W / 450 Ω for transurethral resection of the prostate (TURP)
Cut Under Liquid (use glycine irrigation) (2) UROCUT B : 350 W / 450 Ω for transurethral resection of the bladder (TURB)
HYSTEROCUT : 320 W / 250 Ω autocut for hysteroscopic resection (TCRE)
ENDOCUT 1 : 200 W / 250 Ω : for gastrointestinal endoscopy (polypectomies)
Endocut (2) ENDOCUT 2 : 200 W / 250 Ω : for gastrointestinal endoscopy (polypectomies)
SPHINCTEROCUT : 200 W / 450 Ω : pulsed cut for sphincterotomy
MONOPOLAR COAGULATIONS
Soft coagulation 170 W / 250 Ω, soft contact coagulation for vascular surgery, pediatric and micro surgery
Desiccation 210 W / 450 Ω, forced contact coagulation for all surgeries
Fulguration 100 W / 900 Ω, non-contact coagulation (spray coagulation) for hemostasis of large area
Team Coag 100 W / 900 Ω,simultaneous coagulation from 2 monopolar outputs
BIPOLAR CUT
Pure cut 200 W / 150 Ω cut for the use of bipolar scissors
BIPOLAR COAGULATIONS
Soft Coagulation 100 W / 25 Ω soft coagulation, for neurosurgery and laroscopic
Forced Coagulation (1) 100 % / 150 Ω for all tissues (vascularized and adipose tissues)
ERGONOMY
User Interface Color LCD touch screen
Digital display of the maximum power (Watt max.), except for bipolar forced coagulation displayed in % of maximum
HF Power Display
power
Home View (Get visual information at a glance, to all parameters at selectionned effects sets on the central
electrosurgery unit : Cutting, Coaglation, Menu, Patient Plate Indicator, Volume Control, Bipolar Automatic, Footswitch
Display Indicator
switching). Focus View (Screen display a window with specific informations corresponding to surgeon’s choice made
by touching the “HOME VIEW’ screen display)
Adjustments 1 W / step on the touch screen
ELECTRO SURGICAL UNIT

Brand LAMIDEY NOURY


Type SURGILEC OPTIMA Electro Surgery Unit on Trolley
Origin FRANCE

Technical Data of Surgilec Optima


Memorization Storage of the setting parameters: selections of the cut and coagulation modes, powers adjustments
Initial Programs 80 programs classified by surgical specialties
Memorization of the surgeons’ personal adjustments (possible memorization of several thousand personal
Personal Programs
adjustments)
Visual Demo Available video assist for assembly and disassembly Thermocut 10, Thermocut 5 and Resector
Langguages Available various langguages : France, English, Italia etc
Update USB port for software update
2 outputs for international plugging for hand switches with a 3-pin in-line plug, or for cable with one 4 mm diameter
MONOPOLAR OUTPUTS
male plug
BIPOLAR OUTPUTS 2 outputs for plugging cables with 2-pin 4 mm diameter male plug (interaxial spacing 28.5 mm)

Automatic activation and stop of the bipolar coagulation, by sensing the tissues contact impedance on the forceps tips.
Automatic Bipolar Coagulation
3 available activation delays

1 socket for controls in : • Monopolar and bipolar cut • Monopolar and bipolar coagulation. Available facility switching
Twin Footswitches
twinfootswithes
Single Footswitches 2 sockets for controls in : • Bipolar coagulation •Vessel sealing / THERMOCISION™
Hand Control Monopolar cut and coagulation control on handswitch
Dimentions 429 x 190 x 523 mm- 14 kg
SAFETIES
Monitoring of the skin contact with a single use adhesive split plate
Audible and visual alarm in case of a contact failure or a short circuit, or an electrical discontinuity with inhibition of
Neutral Electrode Safety
the HF power
Compatible with adult and pediatric plates
Neonatal Plates Safety Limitation of monopolar output power up to 50 W, when used with neonatal adhesive double plate
Single Plate Monitoring of the electrical continuity between the unit and the neutral electrode
Monitoring of the emitted HF power, with an automatic emission shutdown in case of an excessive HF power (IEC
HF Power Control
60601-2-2)
HF Leakage Currents Ihf < 100 mA / 200 Ω (< 130 mA / 200 Ω in fulguration mode)
> Class 1, CF type (Ilf < 1 μA in Normal Conditions). >Defibrillation proof applied parts.
Electric Protection
>Short circuit protection > 10 s
Thermal Safety Forced ventilation
HF Voltage Regulated (automatic limitation of the voltage)
Automatic stop after one minute of continuous activation.
Limitation of the Activation Time
The normal working is restored after the interruption of the control
Auto check When powered on, the unit checks all the connected accessories and internal circuits
DELIVERED ACCESSORIES
Single footswitch waterlight pedal for bipolar control 1
Twin footswitches waterlight pedal 1
Main supply cord 1
Trolley 1
Standar Monopolar and Bipolar
Monopolar cable 4m lenght, male4mm / fem 4mm diameter 1
Cable and plug for single use plates lenght 5 m 1
Adult disposable adhesive split plates 10
Child adhesive split plates 10
Baby adhesive split plates 10
TACTILEC Handle with 2 fingerswitches (3 plugs) & cable 4m length 1
Set of 8 standard electrodes 1
Straight Bipolar Forceps 1
Bayonet Bipolar Forceps 1
Bipolar cable lg 4m plug BCP / connector EU 1

NOTES
Note 1: The output power displays in the modes covered by this note are relative displays. (see the power curves in the technical manual of the device)
Note 2: Required upgrade opptional Lamidey/Others accessories to activated the system
ELECTRO SURGICAL UNIT

Brand LAMIDEY NOURY


Type SURGILEC OPTIMA Electro Surgery Unit on Trolley
Origin FRANCE

Technical Data of Surgilec Optima


POTENTIAL UPGRADE / OPTIONS SUPPORTED (available in E-catalogue, required separate order )
Saline cut under saline irrigation with PLASMA EDGE™ technology:
Saline Bipolar Plasma Resection > TURP/TURB
> Intra-uterine resection / TCRE
> Automatic Defective Electrodes detections
Bipolar Plasma Resection Electrodes Full range of PLASMA EDGE™ compatible electrode : Loop, Needle, Ball, Vaporization, Enucleation
Recognition and traceability of vessel sealing instruments
Reusable Cutting and Sealing forceps :
①THERMOCISION™ forceps : ● Completely reusable; ● Dissection, sealing and cutting in one act

Vessel Sealing System (for Open Surgery Only ②THERMOCUT™ forceps : ● Completely reusable and autoclavable; ● Safe and efficient sealing up to 7 mm diameter
/ for Laparascopy Only) vessels; ● Completely dismountable for a full cleaning ● Single use blade; ● Short length ( 20 cm for open surgery)/
long length ( 33 cm for laparoscopy); ● 5 mm and 10 mm diameter forceps; ● Hand activation cable

③ THERMOCLAMP™ forceps : ● Completely reusable and autoclavable; ● Safe and efficient sealing up to 7 mm
diameter; ● 16, 22, 32 cm length; ● Locking system on the cable; ● Open surgery
POWER SYSTEM READY ON THE UNIT FOR POTENTIAL UPGRADE / OPTIONS SUPPORTED
PLASMA EDGE™ CUTS UNDER LIQUID (2) (NaCl 0,9%)
Minimum : 80-Maximum :120 / 450 Ω, use 0.9% NaCl irrigation for transurethral resection of bladder (TURB) and intra-
Thin Loop (1)
uterine resection (TCRE)

Thick Loop (1) Minimum : 120-Maximum : 160 / 450 Ω, use 0.9% NaCl irrigation for transurethral resection of Prostate (TURP)

Minimum : 100-Maximum : 140 / 450 Ω, use 0.9% NaCl irrigation for transurethral resection of Prostate (TURP) and
Vaporization (1)
intra-uterine resection (TCRE)
PLASMA EDGE™ COAGULATIONS (2)
Minimum : 70-Maximum : 130 / 450 Ω, use 0.9% NaCl irrigation for transurethral resection of Prostate/Bladder
Saline Coagulation (1)
(TURP/B)
Ball (1) Minimum : 50-Maximum: 130 / 450 Ω, use 0.9% NaCl irrigation for transurethral resection of Prostate/Bladder
VESSEL SEALING SYSTEM (2)
Maximum power: 150 W / 30 Ω (1)
-3 level power adjusment for each forceps ( 16 Thin Thermoclamp, 22-32 Thick Thermoclamp, Thermocut 5, Thermocut 10)
-Sealing of vascular walls up to 7 mm of diameter with reusable thermofusion forceps
-The jaws are completely insulated: the thermal spread doesn’t exceed 2 mm
-Reverse action mode for THERMOCUT™ handle: the clamping pressure is superior to 700 mmHg
-THERMOCLAMP™ forceps for open surgery, with a vessel clamping device
-THERMOCUT™ 10 mm diameter forceps for open surgery (20 cm) and laparoscopy (30 cm), with vessel autoclamping system and integrated cut system, straight jaws
width 10 mm
-THERMOCUT™ 5 mm diameter forceps for open surgery (20 cm) and laparoscopy (30 cm), with vessel autoclamping system and integrated cut system
-THERMOCISION™ forceps for dissection, hemostasis and cutting in one act (thyroidectomies)
-Automatic output power pre-adjustment following connected forceps
-THERMOCONTROL™ program: vessel sealing process monitoring, with sound signal at the end of the cycle and diagnosis of vessel sealing incident