Professional Documents
Culture Documents
Surface Scatter 6
High Range Turbidimeter Monitoring high
range turbidity
with greater
reliability
and accuracy
Hach’s patented* surface scatter design extensive range of sampling conditions – is an SS6 built tough enough to monitor
virtually eliminates most of the measurement covering water treatment to industrial samples with temperatures of up to 70 °C.
and maintenance problems associated with processing. Whatever the application, the SS6 It is also well suited for applications where a
ordinary high-solids turbidimeters. Electrically makes it easier to establish and maintain significant difference between the sample
and mechanically, the SS6 Turbidimeter is built optimum process control. It is the ideal choice temperature and the ambient temperature
to ensure reliability and accuracy in adverse for reliably monitoring: causes condensation and fogging. Constructed
The SS6 drastically reduces interference from • Waste water effluent materials and an innovative moist air removal
stray light. And because the instrument’s system, no other turbidimeter provides
• Flocculation and sedimentation
optical surfaces never contact the sample, the trouble-free performance in such a wide
• Industrial process water
SS6 can monitor even highly turbid samples range of applications:
• Corrosive white and green liquor in
with virtually no maintenance. Sample cell • The SS6/HST monitors turbidity in fluids
paper mills
cleaning and replacement, for example, are ranging from raw water influent to corrosive
• Produced water in oil fields containing white liquor
unnecessary. Hach recommends standardization
oil and hydrogen sulfide
checks monthly and re-calibration every four • High temperature samples – up to 70 °C
months or whenever the light source is • Food processing waste containing starch,
• Wherever temperature differences cause
replaced or adjusted. A calibration cylinder and fat or oil
condensation and fogging
a 500-mL bottle of formazin primary standard
solution are included for convenient calibration
of the SS6. Built to last Backed by a 2-year warranty
*Patent Number 3,309,956 All wetted parts of the SS6 are fabricated with Consistent with its long-standing reputation
corrosion-resistant materials for extended life. for quality and customer service, Hach
Plumbing fittings, tubing, hose clamps and Company warrants the SS6 turbidimeter against
even screws are selected for superior durability. defects in materials or workmanship for two
Special attention is given to the design and years from the date of shipment.
construction of the photo-detector and light-
source assemblies to safeguard them from the
effects of corrosive vapors.
Principle of Operation
The Surface Scatter® 6 Turbidimeter is a PHOTOCELL LENS LAMP
sensitive and precise instrument designed to
measure the light scattered by particles Figure 1: As the sample
suspended in the sample fluid. The sample is overflows the top of the
REFLECTED LIGHT
introduced into the center of an inclined turbidimeter body, a
SCATTERED LIGHT LIGHT BEAM photocell measures the
turbidimeter body at a flow rate of 1 to 2 liters
light scattered by
per minute (1/4 to 1/2 gallon per minute). As
TURBIDIMETER BODY suspended particles.
the fluid spills over the top of the turbidimeter OVERFLOWING SAMPLE
REFRACTED LIGHT
body, a stable, flat surface of liquid forms and
becomes the measuring surface.
A high-intensity beam of light, powered by
a regulated voltage source, is directed to strike
the surface of the liquid at an acute angle
(Fig. 1). Light is scattered by particles
suspended in the sample and is detected by a
photocell carefully positioned directly over the
point where the light enters the liquid. Thanks
to this optical geometry, the light is scattered
at or near the surface and very little is
absorbed by the liquid. The SS6 is able to
measure extremely high levels of turbidity with
excellent accuracy, because the amount of light INSTRUMENT DRAIN SAMPLE IN DRAIN
scattered increases in direct proportion to the
increase in turbidity. Readings are accurate Method of analysis
across the entire range, from 0 to 9999 NTU. Most of the light directed at the surface of the
The inclined turbidimeter body also serves sample is reflected up into the instrument
as a trap for settling out solids that could cabinet and absorbed, or refracted down into
interfere with measurement. A drain equipped the turbidimeter body (Fig. 2). A small amount
with a ball-valve allows the operator to of the light is scattered by the particles
periodically purge accumulated solids. Large- suspended in the fluid. Light scattered at 90°
diameter plumbing and flexible tubing (3/4” from the incident beam is detected by the
for sample inlet and drain, 1” for overflow) photocell assembly. The electronic signal
prevent plugging when monitoring highly generated by the photocell is directly related
turbid samples. Designed for fluids having high to the concentration of particles suspended
levels of suspended solids, the sample flow rate in the sample.
can be increased and the instrument can be
operated with the bottom drain open to
continuously purge solids when necessary.
Operating controls and indicators are PHOTOCELL
located on the keyboard. Twelve numeric keys LIMITS OF DETECTOR Figure 2: Scattered
and nine function keys are used to program the VIEWING WINDOW light is measured at
instrument for recorder output minimums and CENTERLINE OF exactly 90° from the
LIGHT SOURCE
DETECTOR VIEW incident light beam.
maximums, turbidity level alarm set points, 90
°
diagnostic self tests, and programming REF
LE AM
CTED
LIGH GH T BE
operations. Sample turbidity is displayed T LI
T
N CIDEN
continually by a four-digit LED during normal I
WATER SURFACE VOLUME OF ILLUMINATED
operation. Separate system warning and SAMPLE IN DETECTOR VIEW
system alarm features provide automatic self-
testing diagnostics.
Surface Scatter 6 Specifications* Sample Conditioning Accessories
Range
0 to 9999 Nephelometric Turbidity Units (NTU)
One of many accessory configurations available for the Surface Scatter 6 Turbidimeter
Accuracy
± 5% from 0 to 2000 NTU; ± 10% from 2000 to 9999 NTU
Resolution Electrical
0.01 NTU below 100 NTU; 0.1 NTU between 100-999.9 Connection Bubble
NTU; 1.0 NTU over 1000 NTU 1 Trap
Repeatability 3 3-Way
1.0% or ± 0.04 NTU, whichever is greater Valve
Heat
Response time
Exchanger
Initial response in 30 seconds Cooling
Water In 2
Sample flow required
1.0 to 2.0 L/min (15 to 30 gal/hr)
Sample temperature range
0 to 50 ºC; HST model: 0-70 ºC, intermittent 70-80 ºC; an Cooling
optional heat exchanger is available to reduce sample Water to
temperature Drain
Ambient temperature range
0 to 50 ºC
Humidity specification
5 to 95% non-condensing
3 Conditioned
Recorder output 3-Way
Valve Sample to
Selectable for 0-10 mV, 0-100 mV, 0-1 V, or 4-20 mA
Output span programmable over any portion of the
Analyzer
0-9999 NTU range
Alarms Flow Cooling Water To
Two turbidity set-point alarms, instrument warning and Control Analyzer During
system shutdown alarms are each equipped with an SPDT Valve Auto Flush
relay with unpowered contacts rated for 5A resistive load
at 230 VAC; alarm 2 can be disabled and its contacts used Sample In
to control flush valves Sample Bypass Cooling Water in normal operation
During Auto Flush Cycle Cooling Water during Auto Flush
Power requirements Sample during normal operation
115/230 VAC, 50/60 Hz, switch selectable; 0.5/0.3 A Sample during Auto Flush
Installation category
Category II
Sample inlet fitting Sample Conditioning Accessories 3. Auto Flush Kit
3/4” NPT female
Extreme environments often require additional The programmable Auto Flush Kit initiates and
Overflow drain fitting
1” NPT female sample and pressure conditioning. Several controls automated flushing of sediment from
Body drain fitting optional accessories are available with the the turbidimeter and the sample conditioning
3/4” NPT female SS6 Turbidimeter. panel. It removes sediment that can be flushed
Air purge fitting out with clean water flowing at a rate equal to
3/4” compression fitting; 0-50 SCFH air flow of clean, 1. Bubble Trap/Head Regulator
instrument air the sample flow rate. Sticky materials that
The bubble trap vents air bubbles from the
Control unit case cling to the wall of the sample cell will not be
NEMA-4X plastic instrument enclosure with clear
sample and increases the instrument’s response
removed by the Auto Flush Kit.
polycarbonate cover suitable for indoor installation time to changes in turbidity or concentration.
Sample unit case The head regulator dampens fluctuation in Outputs
NEMA-12 plastic instrument enclosure suitable for indoor
installation flow caused by variations in pump and Serial Interface Kit
Dimensions (width x height x depth) sample pressure. The user-installed serial interface board equips
Control unit: 13.5 x 9 x 7.5 in. (34.3 x 22.8 x 19.0 cm)
Sample unit: 25.3 x 26.6 x 7.5 in. (64.2 x 67.5 x 19.0 cm) 2. Heat Exchanger/Sample Cooler the SS6 to output data directly to an external
Mounting The optional heat exchanger cools samples printer or computer and to receive control
Wall mount commands from the computer.
that exceed 50 °C (122° F) – the SS6 maximum
Shipping Weight temperature specification.
SS6: 35 lbs. (15.8 kg)
SS6/HST: 40 lbs. (18 kg)
*Subject to change without notice.
Installation and maintenance
The SS6 Turbidimeter should be located as ALARM 1
ALARM 2
NTU SYSTEM
WARNING
REC SIG
SCATTER
1 2 3
sunlight. Best performance will result if the
MAX
6
AVG
. + REC SYS
0 -
CLEAR MIN RESET
Light
Source SS6/HST Turbidimeter
Assembly
Flow Multiplier
(Cat. No. 40301-00)
23.820"
(605.03 mm)
Bulkhead Fitting
1" NPT
(Cat. No. 40355-00)
TURBIDIMETER SAMPLE INLET OVERFLOW
BODY DRAIN PORT DRAIN
BOTTOM VIEW
How to Order
Call Hach today to order the Surface Scatter 6 High Optional Accessories
Range Turbidimeter that fits your application. Catalog # Surface Scatter 6
Turbidimeter
45000-43 Upgrade Kit, converts standard Surface
Catalog # Surface Scatter 6
Scatter 6 to Surface Scatter 6 HST
Turbidimeter
44278-00 Serial Interface Kit
45000-10 Surface Scatter 6 Turbidimeter with
46692-12 Auto Flush Kit (120 VAC)
calibration cup and 4000 NTU standard
46692-22 Auto Flush Kit (220 VAC)
solution, installation accessories and
46680-00 Bubble Trap/Head Regulator
instruction manual.
48551-00 Heat Exchanger (sample cooler)
(Note: Power cords must be ordered
40284-00 Flow Meter 100-1600 mL/minute
separately for both models)
23513-00 Standardization Plates
45000-40 Surface Scatter 6/HST Turbidimeter for 46691-00 Tubing Replacement Kit
high temperature samples 45000-19 Manual SS6 and SS6/HST
46306-00 Power Cord, 125 VAC, 10 A, 6' (1.83 m)
46308-00 Power Cord, 250 VAC, 10 A, 6' (1.83 m)