You are on page 1of 2

Technical Specifications

Random Access Clinical Chemistry Analyser System Function


Description Fully Automated, discrete, random access clinical chemistry analyzer
Measuring principle Spectrophotometry
Photometric throughput Up to 240 tests/hour
Photometric system HCFG rear spectrophotometry
EFFICIENT LAUNDRY Methodology End point, Fixed time, Kinetic, Single & Dual reagent chemistries
SYSTEM Mono & Bi chromatic, Linear & non-linear multipoint calibration
Optical System
Ÿ Laundry system adopting 7 stops, 11 steps Light source Halogen - Tungsten lamp (12V/20W)
Ÿ Vacuum draining liquid, detergent and Monochromator Grating Photometry (Holographic Concave Flat Field Grating)
warm water rinsing ensures guaranteed
Wavelength 340nm, 380nm, 405nm, 450nm, 480nm, 505nm, 546nm, 570nm, 600nm, 660nm, 700nm,
accuracy
750 or 800nm
Ÿ Water consumption of 4 L/hour
Linear range 0 ~ 3.3 Abs
Detector Photodiode array
Reagent/Sample Handling
Reagent/Sample tray Multi-functional reagent and sample carousel
Reagent/Sample position Up to 80 positions
Sample cuvette specification Standard cup, original blood tube, multi-specification tube (10~13) x (75~100) mm
SEMI PERMANENT Sample reagent probe Digital liquid level detection and vertical collision protection
Sample dilution Pre & Post dilution facility
REACTION CUVETTE Reagent volume 10~300 µl
Ÿ Rigid semi permanent reaction cuvette, Sample volume 2~35 µl
with good penetrating of Ultraviolet (UV) Dilution vessel UV plastic semi permanent cuvette
Ÿ Reaction cuvettes ensures cost savings Reagent bottle volume 20 & 70 ml
with enhanced durability of 9 to 12 Reaction System
months

ADL/BR/CXLPRO+/R00-06/22 Company reserves the right to change any design and technical features of the product at any time, If needed.
Reaction cuvette 56 positions optical plastic cup
Ÿ 1 Set 7 Pieces | 8 Sets 56 Pieces | Cuvette Reaction volume 100~360 µl
optical path of 5mm
Reaction temperature 37±0.1°C
Reaction disk constant temperature Thermostat air bath
Mixing system Teflon coated stirrer with triple speed mixing mechanism
Laundry system Efficient system adopting 7 stops, 11 steps
Operation Unit
CALIBRATION & PC operation system Windows 7 or Windows 10
QC PROGRAM PC configuration CPU > 2.9 Ghz (dual core processor); RAM > 4 GB; Harddisk ≥ 160 GB
Analysis control Graphical operating software
Ÿ Linear and non-linear calibration with 9 Report printing Supports user-defined mode, QC and state information etc
types of calibration curve
System connection TCP/IP network connection, standard RJ-45
Ÿ 6 different levels of calibration for each item
Calibration & Quality Control
can be programmed
Calibration method Linear (One-point, two-point and multi-point), Logit-Log 4P, Logit-Log 5P, Spline,
Ÿ QC with Westgard multi rules
Ÿ QC plot with L-J and cumulative statistics Exponential, Polynomial
Ÿ Automatic error reporting complaint with Quality control method Real-time, daily, monthly & QC chart | Visual QC assessment | QC histroy checking
lab QC management QC error analysis
Parabola control rules Westgard multi-rule, L-J chart
Working Conditions
Power supply 100~240 VAC | 50/60Hz | Power 600VA
Ambient temperature 15°C~25°C
Relative humidity 40% ~ 85%
Atmospheric pressure 70-106kPa
Water consumption 4L/hour
BETTER TECHNOLOGY FOR BETTER PERFORMANCE Dimensions (LxWxH) 744x703x530 mm
Weight Approx. 100 kg
Random Access Clinical Chemistry Analyser

ISO 9001:2015 “Agappe Hills”, Pattimattom (PO), Dist. Ernakulam, Kerala - 683 562, India.
EN ISO 13485:2016 TEL: + 91 484 2867000 | productcorp@agappe.in | www.agappe.com * Toll Free No: 1800 425 7151
Random Access Clinical Chemistry Analyser Random Access Clinical Chemistry Analyser Random Access Clinical Chemistry Analyser

Mispa CXL Pro Plus is a compact, fully automated clinical chemistry analyzer with
photometric throughput of 240 tests per hour. Mispa CXL Pro Plus has Holographic SAMPLE/REAGENT PIPETTING
Concave Flat Field Grating (HCFG) rear spectrophotometry system designed for best
reliability and maximum accuracy. On-board washing and cleaning system in Mispa
MECHANISM

240
CXL Pro Plus adopting 7 stops 11 steps, ensures minimum carryover and precise result Ÿ Probe with digital liquid level detection and vertical
collision protection
delivery. Mispa CXL Pro Plus is the best in the class with grating, mixing, washing and Ÿ Analyzer has special degassing device to remove air
user-friendly classic features.
Ÿ
dissolved in tube for accurate pipetting
Syringes are made of long life high precision ceramic
piston, ensures minimal maintenance
T/hr
Ÿ 60nm polished probe with nano coating technology
Ÿ Thermostat air bath to ensure temperature of 37±0.1° C

MULTI-FUNCTION SAMPLE &


REAGENT CAROUSEL
Ÿ 40 reagent and 40 sample positions with anywhere
anytime STAT facility
Ÿ 24 hours continuous cooling condition ensures quality
of reagent, control and calibrator
Ÿ Can accommodate 20ml and 70ml reagent bottles
Ÿ Barcoded tailor-made dedicated system reagents
Ÿ Standard cup, original blood tube, multi-specification
tube (10~13) x (75~100) mm

PROVEN HCFG PHOTOMETRY TRIPLE SPEED MIXING


SYSTEM MECHANISM
Ÿ Monochromator with holographic concave flat field
Ÿ Teflon coated stirrer, to avoid liquid
grating (HCFG), rear spectrophotometric 12 ways
suspension and reduce cross
parallel measuring technology, reduces ambient light
contamination
interferences to get accurate result.
Ÿ Triple speed mixing mechanism is
Ÿ Photospot technology to reach super micro analysis.
highly efficient for latex based assays
Ÿ Specially designed lamp placement to reduce signal
Ÿ Flat paddle stirrer design with swirl
attenuation and interference.
rinsing offers homogeneous mixing

PURPOSEFULLY BUILT FOR PERFORMANCE EXPERIENCE THE ULTIMATE PERFECTION STEP UP TO THE BEST QUALITY