3, 1979) 509
A new method is described for hydroxy- for quantitation of HMF for research and regu-
methylfurfural (HMF) in honey; accuracy and latory use: quantitation of earlier color tests (4,
precision are improved over the most used opti- 5), Winkler's specific photometric chemical
cal and chemical methods. With a clarified method (6), and Winkler's optical method us-
honey solution containing 0 . 1 % sodium bisul-
ing the prominent UV absorption band of
fite as reference and a similar solution without
HMF (6-8). Winkler's toluidine-barbituric acid
bisulfite as sample, a difference spectrum is ob-
tained which represents only the HMF in the method (6) has been adopted by the Codex
sample, without the interfering absorption of the Alimentarius; a comparison with his optical
honey. The average recovery was 9 7 . 5 % for 2 4 procedure indicated the former to be more reli-
additions to honey of 0 . 8 - 4 0 mg H M F / 1 0 0 g. able (3). Dhar and Roy (9) have criticized the
Forty honey samples ranging from 0 to 4 0 optical method and described a charcoal column
m g / 1 0 0 g were analyzed by 3 methods with cleanup before analyses. This method is cumber-
the following average results: Winkler optical some and lengthy and not well suited for rou-
method, 7 . 2 5 ; Winkler chemical method, 4 . 8 3 ; tine use.
and new bisulfite method, 5.05 mg H M F / 1 0 0 g
Recent experience at this laboratory with the
honey. Values by the latter 2 methods did not
differ at the P = 0.05 significance level.
Winkler chemical method in the analysis of over
600 samples of honey has emphasized several
The presence in honey of hydroxymethylfur- shortcomings. In addition, a recent AOAC col-
fural (2-hydroxymethyl-5-furaldehyde, HMF) laborative study of the 2 Winkler methods (10)
was originally considered evidence of its adul- has confirmed earlier observations (11) that the
teration with acid-converted invert sirup. Color Winkler UV procedure produces anomalous re-
tests were used for qualitative indication of its sults with certain honeys whose ultraviolet ab-
presence (1) but were soon criticized as it be- sorption spectra do not resemble those studied
came known that authentic but strongly heated by Winkler. The results by the 2 methods dif-
honey gave a positive test. A color test, with fered significantly in the AOAC study (10).
appropriate caveat, is still an official method (2). Interlaboratory precision was considerably bet-
More recently, HMF content has been widely ter for the UV method than for the chemical
used as an indicator of the heating history of procedure. However, the chemical procedure is
honey. Many European honey standards, in- considered to give more accurate values because
cluding the Codex Alimentarius, set maxima for of the specific nature of the reaction of HMF
HMF in honey. with barbituric acid and toluidine (6).
Recognition that HMF could be produced in When the discrepancy became apparent, a
appreciable amounts in honey stored at ambient random selection was made from the sample
temperatures, common in many honey-producing collection noted above and other samples avail-
areas, has further complicated the use of HMF able in this laboratory. The Winkler chemical
as an indicator of adulteration. A semilogarith- method had been applied to all samples. Forty
mic relation has been demonstrated between the of these, representing negligible, low, moderate,
time needed for a honey to accumulate a given and high contents of HMF, were then analyzed
level of HMF and temperatures between 20 and by the Winkler ultraviolet method. The results
80°C (3). Differentiation between heat- or stor- further confirmed the lack of agreement be-
age-abused honey and that containing added in- tween the 2 methods.
vert sugar requires accurate analytical methods.
Several general approaches have been used 1
Present address: 217 Hillside Dr, Navasota, TX 77868.
5756-0004779/6203-0509/06$01.00
© Association of Official Analytical Chemists, Inc.
510 WHITE: J. ASSOC. OFF. ANAL. CHEM. (VOL. 62, NO. 3, 1979)
METHOD
Reagents and Apparatus
(a) Carrez solution I.—Dissolve 15 g potassium
ferrocyanide (K 4 Fe(CN) 6 .3H 2 0) in water and di-
lute to 100 mL.
(b) Carrez solution II.—Dissolve 30 g zinc ace-
tate (Zn(CH3C02)2-2H20) in water and dilute to
100 mL.
(c) Sodium bisulfite (NaHSOs).—0.20% in
water. Prepare fresh daily. Technical grade is ade-
280 340 220 340
quate.
WAVELENGTH, nm
(d) Spectrophotometer. — Bausch and Lomb
FIG. 1—Absorption spectra of Carrez-clarified solutions
of hydroxymethylfurfural in A, water (vs. water); B,
Spectronic 505 recording spectrophotometer was
0.1% NaHS0 3 (vs. 0.1% NaHS0 3 ). Concentrations: 6.26, used. For routine analysis, spectrophotometer pro-
3.13, 1.57, 0.78, 0.31 /ig HMF/mL. viding values at 284 and 336 nm is satisfactory.
Procedure
The Winkler chemical method is now con- Transfer ca 5 g honey (weighed to 1 mg in small
sidered an undesirable procedure, however, be- beaker) to 50 mL volumetric flask with total of
cause of the following major short-comings: It 25 mL water. Add 0.50 mL Carrez solution I,
mix, add 0.50 mL Carrez solution II, mix, and
requires p-toluidine, a toxic reagent which car-
dilute to volume with water. A drop of alcohol
ries a label warning of carcinogenicity to ani-
may be added to suppress surface foam. Filter
mals, and several AOAC collaborators objected through paper, rejecting first 10 mL filtrate.
to its use; and the color is unstable and tem- Pipet 5 mL filtrate in each of two 18 X 150
perature-dependent in its development. mm test tubes and pipet 5 mL water into one
A search was therefore instituted for a new
procedure. An ideal method should retain the
precision of the optical method and have the
accuracy of the chemical method, without their
deficiencies. The solvent extraction used in sev-
eral other procedures should be avoided also.
This paper describes such a method, in which
the UV absorbance of a clarified aqueous honey
solution is determined against a reference solu-
tion of the same honey in which the 284 nm
chromophore of HMF has been destroyed by
bisulfite. The destruction of the 284 nm chromo-
phore results from the classical addition of a
nucleophile to the a,/?-unsaturated carbonyl sys-
tem.
0.05 0.10 0.20
Amounts of HMF indicated in Table 1 were FIG. 5—A, absorption spectra of Samples 1-6 (Table 3),
prepared for Winkler UV method. Baselines displaced
added to aliquots of 25 ml of a 20% aqueous vertically for clarity. B, difference spectra of same sam-
honey solution. These were analyzed as de- ples (2 g samples) prepared according to bisulfite
method. Baselines as indicated.
scribed above. Five replicate analyses of the
stock honey (1.26, 1.30, 1.30, 1.27, 1.30 mg
HMF/100 g) were averaged to obtain the base Precision
value for calculating recoveries. Per cent re- A sample was analyzed by the bisulfite proce-
covery is the amount found divided by the dure 8 times in 3 days. The average was 1.30
amount originally present plus the amount mg HMF/100 g and the standard deviation was
added. Recoveries were in the 95-100% range, 0.075.
averaging 96.9%, as shown in Table 1.
Six honey samples which showed consider- Comparison of Methods
able discrepancies in HMF content when de- Forty samples of honey were analyzed for
termined by the Winkler chemical and Winkler HMF by the procedure described here. They
UV methods were selected for the second re-
covery test. These are the first 6 samples in Table 2. Recovery of hydroxymethylfurfural
Table 3. Figure 5A shows the variety of shapes added to various honeys (mg/100 g)
of the spectra in the 250-320 nm region, which Hydroxymethylfu rfural
accounts for the discrepancies between the 2
In Total
Winkler methods. Figure 5B shows the differ- No. honey Added present Found Rec, %
ence spectra for these samples. Each has a
1 2.80 0.79 3.59 3.55 98.9
maximum at 284 nm. HMF amounts equivalent 2 0.66 0.79 1.45 1.39 95.9
to about 1 and 10 mg/100 g honey were added 3 0.63 0.79 1.42 1.47 103.5
to two 25 mL aliquots of a 20% solution of each. 1 2.80 13.89 16.69 16.30 97.7
2 0.66 13.89 14.55 13.92 95.7
One aliquot was analyzed without added HMF. 3 0.63 13.89 14.52 14.06 96.8
Each was clarified and analyzed by the bisulfite 4 0.34 1.01 1.35 1.35 100.0
method. Table 2 shows the results, with an 5 1.87 1.01 2.88 2.75 95.5
6 1.26 1.01 2.27 2.27 100.0
average recovery of 98.0%. Recovery is amount 4 0.34 8.10 8.44 8.16 96.7
found divided by the amount originally present 5 1.87 8.10 9.97 9.63 96.6
6 1.26 8.10 9.36 8.89 95.0
plus the amount added.
WHITE: J. ASSOC. OFF. ANAL. CHEM. (VOL. 62, NO. 3, 1979) 513
Table 3. Comparison of methods for hydroxy methyl- the differences between the methods for each
furfural in honey (mg HMF/100 g honey)
sample. The results of 3 paired £-tests showed
Winkler Winkler no evidence of a statistically significant (P —