You are on page 1of 25

Galaxy Couch

Interconnect
Drawing

Preliminary

STATOR
SYSTEM
ROTOR
SYSTEM

STATOR
SEE SHEET 1
PATIENT SUPPORT OPERATOR CONSOLE
SYSTEM SYSTEM
GANTRY TILT
SUB SYSTEM
T61-2170
FUSE RELAY C O N T R O L BOARD (FRC)
SCHEMATIC
FRC (Fuse Relay Control) LED's
Standby Norm
LED» Color Signal Name In / Out TPRef Schem Ref Comments
State State
1 Grn +5VDC Supply for RHOST (before F8) On In 2:B3 If not On, check F6 (AC to P/S)
2 Grn +5VDC to RHOST (after F8) On Out 2 2:B3 If not On and LED1 is, check F8.
3 Grn -5VDC Reference for Current Sense On 3_ 3:A3 Made from + 10V (LED4).
Made from +15V (LED 10). Supplies power for -5V
4 Grn +10V for Current Sense On 4 2:D4 (LED3). If not On and LED10 is, 10V P/S (U8) is bad or
loaded down.
Made from +15V (LED10). If not On and LED10 is, 10V
5 Grn +10V Isolated for HSS (via RHOST) On Out 5 2:D2
P/S (U7) is bad or loaded down.
•--

6 Grn + 1 2 V D C t o RSLI (after F11) On Out 6 2:A3 If not On, check F11 and F6.
7 Grn -15VDC Supply for RHOST (before F10) On In 7 2:C6 If not On, check F7 (AC to P/S).
If not On, check F7 (AC to P/S). +15V also supplies
8 Grn +15VDC Supply for RHOST (before F9) On In 8 2:D6
power 10V P/S (LED4) and LED 10 via F9.
9 Grn -15VDC to RHOST (after F10) On Out 9 2:C4 If not on, check F10.
If not On and LED 8 is, check F9. +15V after F9 also
10 Grn +15VDC to RHOST (after F9) On Out 10 2:D4 used to make 10V isolated (LED5) for HSS (via
RHOST).
11 Yel Tube Housing > 60 Degrees Celsius Out 4:A5 To RHOST
12 Red Tube Housing > 80 Degrees Celsius Out 4:C5 To RHOST
- -----

13 Red No Heat Exchanger Oil Flow Out 4:D5 To RHOST


Commands AC to HE Pump and Fans. Commanded
14 Grn Heat Exchanger On On 3:C6
on by RHOST or Tube above 60 Degrees (LED11).

TWS:FRC Service Data Philips Medical Confidential


FRC TEST POINTS
Standby
TP* Name Norm State In/Out LED Ref Schem Ref Comments
State
1 Not Used
2 +5VDC to RHOST (after F8) Out 2 2:C3 Use TP 11 as ref
3 -5VDC for Current Sense 3 3.A3 Use TP 14 as ref.
4 +10V for Current Sense 4 2:D4 Use TP 14 as ref.
5 +10V Isolated for HSS (via RHOST) Out 5 2:D2 Use TP 13 as ref.
6 + 1 2 V D C t o R S L I (after F11) Out 6 2:B2 Use TP12 as ref
7 -15VDC Supply for RHOST (before F10) In 7 2:C6 Use TP 14 as ref.
8 +15VDC Supply for RHOST (before F9) In ' 8 2:D6 Use TP 14 as ref.
9 -15VDC to RHOST (after F10) Out 9 2:C4 Use TP 14 as ref.
10 +15VDC to RHOST (after F9) Out 10 2:D4 Use TP 14 as ref.
11 +5V Return 2:C3 RefforTP2
12 +12V Return 2:A2 RefforTP6
13 +10V Isolated Return 2:D2 Ref for TP5
14 Analog Ground 2:C6 Ref for TP7-10, 4,3
15 HE Pump Current Sense Signal Out 3:D2 To RHOST
16 HE Fans Current Sense Signal Out 3:C2 To RHOST

TWS: FRC Service Data Philips Medical Confidential


FRC CONNECTORS

Conn # # Pins Type Name To / From Schem Ref Comments

J1 25 D sub RHOST Signal Interface RHOST J3 4:C1


J3 3 120VAC to DMS (via F3) DMS 2:B5
J4 3 120VAC to Cathode Pwr Module (via F4) KPM 2:B5
J5 3 120VAC to Anode Pwr Module (via F5) APM 2:A5
P1 8 RHOST Power Interface RHOST P10 2:B1 +5V, + 1 5 V & - 1 5 V
P10 5 Header Tube Temperature Switches Tube 4:B8
P11 Header 5/12 Volt P/S Interface 5 & 12 V P/S 2:C4

CD
P13 5 Header +/-15V P/S Interface +/-15V P/S 2:C7
P19 3 Left Slip Ring Power Slip Ring 2:B8 Not Used?
P2 3 12VDCtoRSLI RSLI 2:A1
P6 5 Header Heat Exchanger Pump Heat Exchanger 4.C8
P7 12 Heat Exchanger Fans Heat Exchanger 4:D4
P8 3 120VACfrom Slip Ring Slip Ring 2:B8

P11-P17 Fast-on Wiring to off-board fuses (120VAC) 2:B7

T W S . FRC Service Data Philips Medical Confidential


FRC FUSES and SWITCHES

Fuses Type Function

Schem
Switches # Poles Settings Ref Comments
T61B-1742

SIGNAL DISTRIBUTION B O A R D

SCHEMATIC
oz
oz
ш
o
Ш
CO
—I
Ш
>
Ш
-J
X
o
X
3
oo
<