Interconnect
Drawing
Preliminary
•
STATOR
SYSTEM
ROTOR
SYSTEM
STATOR
SEE SHEET 1
PATIENT SUPPORT OPERATOR CONSOLE
SYSTEM SYSTEM
GANTRY TILT
SUB SYSTEM
T61-2170
FUSE RELAY C O N T R O L BOARD (FRC)
SCHEMATIC
FRC (Fuse Relay Control) LED's
Standby Norm
LED» Color Signal Name In / Out TPRef Schem Ref Comments
State State
1 Grn +5VDC Supply for RHOST (before F8) On In 2:B3 If not On, check F6 (AC to P/S)
2 Grn +5VDC to RHOST (after F8) On Out 2 2:B3 If not On and LED1 is, check F8.
3 Grn -5VDC Reference for Current Sense On 3_ 3:A3 Made from + 10V (LED4).
Made from +15V (LED 10). Supplies power for -5V
4 Grn +10V for Current Sense On 4 2:D4 (LED3). If not On and LED10 is, 10V P/S (U8) is bad or
loaded down.
Made from +15V (LED10). If not On and LED10 is, 10V
5 Grn +10V Isolated for HSS (via RHOST) On Out 5 2:D2
P/S (U7) is bad or loaded down.
•--
6 Grn + 1 2 V D C t o RSLI (after F11) On Out 6 2:A3 If not On, check F11 and F6.
7 Grn -15VDC Supply for RHOST (before F10) On In 7 2:C6 If not On, check F7 (AC to P/S).
If not On, check F7 (AC to P/S). +15V also supplies
8 Grn +15VDC Supply for RHOST (before F9) On In 8 2:D6
power 10V P/S (LED4) and LED 10 via F9.
9 Grn -15VDC to RHOST (after F10) On Out 9 2:C4 If not on, check F10.
If not On and LED 8 is, check F9. +15V after F9 also
10 Grn +15VDC to RHOST (after F9) On Out 10 2:D4 used to make 10V isolated (LED5) for HSS (via
RHOST).
11 Yel Tube Housing > 60 Degrees Celsius Out 4:A5 To RHOST
12 Red Tube Housing > 80 Degrees Celsius Out 4:C5 To RHOST
- -----
CD
P13 5 Header +/-15V P/S Interface +/-15V P/S 2:C7
P19 3 Left Slip Ring Power Slip Ring 2:B8 Not Used?
P2 3 12VDCtoRSLI RSLI 2:A1
P6 5 Header Heat Exchanger Pump Heat Exchanger 4.C8
P7 12 Heat Exchanger Fans Heat Exchanger 4:D4
P8 3 120VACfrom Slip Ring Slip Ring 2:B8
Schem
Switches # Poles Settings Ref Comments
T61B-1742
SIGNAL DISTRIBUTION B O A R D
SCHEMATIC
oz
oz
ш
o
Ш
CO
—I
Ш
>
Ш
-J
X
o
X
3
oo
<