Professional Documents
Culture Documents
RADIOGRAPHY TESTING
1. Introduction:
RADIOLOGY is the general term given to material inspection methods that are based
on the differential absorption of penetrating radiation either electromagnetic
radiation of very short wavelength or particulate radiation by the part or test piece
(object) being inspected.
Because of differences in density and variations in thickness of the part or
differences in absorption characteristics caused by variations in composition,
different portions of a test piece absorb different amounts of penetrating radiation.
These variations in the absorption of the penetrating radiation can be monitored by
detecting the unabsorbed radiation that passes through the test piece.
Radiography can detect only those features that have an appreciable thickness in a
direction parallel to the radiation beam. This means that the ability of the process to
detect planar discontinuities such as cracks depends on proper orientation of the
test piece during inspection.
Discontinuities such as voids and inclusions, which have measurable thickness in all
directions, can be detected as long as they are not too small in relation to section
thickness. In general, features that exhibit a 1% or more difference in absorption
compared to the surrounding material can be detected.
Page 1 of 111
IRONSCAN Institute of Non-Destructive testing and services
Three basic elements of radiography include a radiation source, the test piece or
object being evaluated, and a sensing material. These elements are shown
schematically in Figure.
The test piece in Figure is a plate of uniform thickness containing an internal flaw
that has absorption characteristics different from those of the surrounding material.
Radiation from the source is absorbed by the test piece as the radiation passes
through it; the flaw and surrounding material absorb different amounts of radiation.
The intensity of radiation that impinges on the sensing material in the area beneath
the flaw is different from the amount that impinges on adjacent areas. This produces
an image, or shadow, of the flaw on the sensing material.
There are several characteristics of the photographs and x-ray photographs that are
similar, and yet, there are very distinct differences. Radiography and photography both
often use some type of film to record the image.
The principle difference is in how the image is recorded on the film.
a) In photography, the image on the film is produced due to the reflection of light.
When you want to take a picture, you use a camera, which has a shutter that opens
and closes allowing light to enter and strike the film. The amount of light that
interacts with the film determines how the image appears on the film.
b) In radiography, the process to produce an image is quite different. The camera is
actually a radiation source and it operates quite differently than a photographic
camera. The film is not placed inside the camera but instead is placed on the
opposite side of the object being imaged. The radiation is not reflected to the film,
but rather passes through the object and then strikes the film. The image on the film
is dependent upon how much of the radiation makes it through the object and to the
film. The amount of material that the X-rays must travel through also affects how
many X-rays reach the film. Differences in the type of material and the amount of
material that the X-rays must penetrate are responsible for the details in the image.
In 1895, a German physicist, W. C. Roentgen was working with a cathode ray tube in
his laboratory. He was working with tubes similar to our fluorescent light bulbs. He
evacuated the tube of all air, filled it with a special gas, and passed a high electric
voltage through it. When he did this, the tube would produce a fluorescent glow.
Roentgen shielded the tube with heavy black paper, and found that a green colored
fluorescent light could be seen coming from a screen setting a few feet away from
the tube.
He realized that he had produced a previously unknown "invisible light," or ray, that
was being emitted from the tube; a ray that was capable of passing through the
heavy paper covering the tube. Through additional experiments, he also found that
the new ray would pass through most substances casting shadows of solid objects
on pieces of film.
He named the new ray X-ray, because in mathematics "X" is used to indicated the
unknown quantity.
In his discovery Roentgen found that the X-ray would pass through the tissue of
humans leaving the bones and metals visible. The news of Roentgen’s discovery
spread quickly throughout the world.
In 1896 Henri Becquerel was using naturally fluorescent minerals to study the
properties of x-rays, which had been discovered in 1895 by Wilhelm Roentgen. He
exposed potassium uranyl sulphate to sunlight and then placed it on photographic
plates wrapped in black paper, believing that the uranium absorbed the sun’s
energy and then emitted it as x-rays. This hypothesis was disproved on the 26 th-27th
of February, when his experiment "failed" because it was overcast in Paris.
Becquerel decided to develop his photographic plates with uranium compounds. To
his surprise, the images were strong and clear, proving that the uranium emitted
radiation without an external source of energy such as the sun. Becquerel had
discovered radioactivity.
The term radioactivity was actually coined by Marie Curie, who together with her
husband Pierre, began investigating the phenomenon recently discovered by
Becquerel. The Curies extracted uranium from ore and to their surprise, found that
the leftover ore showed more activity than the pure uranium. They concluded that
the ore contained other radioactive elements. This led to the discoveries of the
elements polonium and radium. It took four more years of processing tons of ore to
isolate enough of each element to determine their chemical properties.
Scientists everywhere could duplicate his experiment because the cathode tube was
very well known during this period. In early 1896, X-rays were being utilized
clinically in the United States for such things as bone fractures and gunshot wounds.
For the discovery of radioactivity, Becquerel was awarded the 1903 Nobel Prize for
physics.
For the work on radioactivity, the Curies were awarded the 1903 Nobel Prize in
physics.
In 1910, Marie Curie was awarded the Nobel Prize in chemistry for her discoveries of
radium and polonium, thus becoming the first person to receive two Nobel Prizes.
Page 4 of 111
IRONSCAN Institute of Non-Destructive testing and services
Since these discoveries, many other radioactive elements have been discovered or
produced. Today many artificial radioactive materials are produced and put to use in
various ways ranging from medical to industrial.
same element can have a different number of neutrons. Atoms want to have
the same number of neutrons and protons but the number of neutrons can
change.
The atomic number of a chemical element (also known as its proton number)
is the number of protons found in the nucleus of an atom of that element,
and therefore identical to the charge number of the nucleus.
It is conventionally represented by the symbol Z.
The atomic number uniquely identifies a chemical element. In an uncharged
atom, the atomic number is also equal to the number of electrons.
The mass number (A), also called atomic mass number or nucleon number, is
the total number of protons and neutrons (together known as nucleons) in an
atomic nucleus.
Because protons and neutrons both are baryons, the mass number A is
identical with the baryon number B as of the nucleus as of the whole atom or
ion.
The mass number is different for each different isotope of a chemical element.
This is not the same as the atomic number (Z) which denotes the number of
protons in a nucleus, and thus uniquely identifies an element.
The difference between the mass number and the atomic number gives the
number of neutrons (N) in a given nucleus: N=A−Z
The mass number is written either after the element name or as a superscript
to the left of an element's symbol.
For example, the most common isotope of carbon is carbon-12, or 12C, which
has 6 protons and 6 neutrons.
The full isotope symbol would also have the atomic number (Z) as a subscript
to the left of the element symbol directly below the mass number: 12
6C.[2] This is technically redundant, as each element is defined by its atomic
number, so it is often omitted.
Page 6 of 111
IRONSCAN Institute of Non-Destructive testing and services
6. Radioactivity:
Page 7 of 111
IRONSCAN Institute of Non-Destructive testing and services
either to a newly created radiation particle within the nucleus or via internal
conversion.
During this process, the radionuclide is said to undergo radioactive decay,
resulting in the emission of gamma ray(s) and/or subatomic particles such as
alpha or beta particles.
These emissions constitute ionizing radiation. Many radio nuclides occur
naturally, and others are produced artificially, for example in nuclear reactors
and cyclotrons.
Radio nuclides with suitable half-lives play an important part in a number of
technologies, for example ionization smoke detectors and nuclear medicine.
A pharmaceutical drug made with radio nuclides is called a
radiopharmaceutical, and an imaging tracer made with radio nuclides is called
a radiotracer.
Nuclear medicine makes use of these drugs and tracers for radiation therapy
such as brachytherapy and imaging.
7. Gamma Radiation
Page 8 of 111
IRONSCAN Institute of Non-Destructive testing and services
The most commonly known chain reactions are nuclear chain reactions. A
nuclear reaction, in general, occurs when two isotopic nuclei collide,
producing molecules different from the reactants, usually different species.
A nuclear chain reaction occurs when consequent nuclear reactions happen
after an initial reaction.
There are two types of nuclear reactions:
Nuclear fission reactions
Nuclear fusion reactions.
Page 9 of 111
IRONSCAN Institute of Non-Destructive testing and services
Page 10 of 111
IRONSCAN Institute of Non-Destructive testing and services
Page 11 of 111
IRONSCAN Institute of Non-Destructive testing and services
When an atom undergoes radioactive decay, it emits one or more forms of radiation
with sufficient energy to ionize the atoms with which it interacts.
Ionizing radiation can consist of high speed subatomic particles ejected from the
nucleus or electromagnetic radiation (gamma-rays) emitted by either the nucleus or
orbital electrons
Alpha particles consist of two protons and two neutrons bound together into a
particle identical to a helium nucleus.
They are generally produced in the process of alpha decay, but may also be
produced in other ways.
Certain radio nuclides of high atomic mass (Ra226, U238, Pu 239) decay by the
emission of alpha particles.
These alpha particles are tightly bound units of two neutrons and two protons each
(He4 nucleus) and have a positive charge.
Emission of an alpha particle from the nucleus results in a decrease of two units of
atomic number (Z) and four units of mass number (A).
Alpha particles are emitted with discrete energies characteristic of the particular
transformation from which they originate.
All alpha particles from a particular radionuclide transformation will have identical
energies.
Some smoke detectors contain a small amount of the alpha emitter americium-241.
The alpha particles ionize air between a small gap.
A small current is passed through that ionized air. Smoke particles from fire that
enter the air gap reduce the current flow, sounding the alarm.
The isotope is extremely dangerous if inhaled or ingested, but the danger is minimal
if the source is kept sealed
Page 12 of 111
IRONSCAN Institute of Non-Destructive testing and services
8.3.3 Gamma-rays:
A nucleus which is in an excited state may emit one or more photons (packets of
electromagnetic radiation) of discrete energies.
The emission of gamma rays does not alter the number of protons or neutrons in the
nucleus but instead has the effect of moving the nucleus from a higher to a lower
energy state (unstable to stable).
Gamma ray emission frequently follows beta decay, alpha decay, and other nuclear
decay processes.
Page 13 of 111
IRONSCAN Institute of Non-Destructive testing and services
121.8, 344.3,
Europium-152 6660 TBq/kg (180 Ci/g) 13.5 years
1408.0
Page 14 of 111
IRONSCAN Institute of Non-Destructive testing and services
G: 32, 661.6 B:
Caesium-137 3256 TBq/kg (88 Ci/g) 30.1 years Gamma & beta
511.6, 1173.2
Gamma photons have no mass and no electrical charge they are pure
electromagnetic energy.
Because of their high energy, gamma photons travel at the speed of light and
can cover hundreds to thousands of meters in air before spending their
energy. They can pass through many kinds of materials, including human
tissue. Very dense materials, such as lead, are commonly used as shielding to
slow or stop gamma photons.
Their wave lengths are so short that they must be measured in nanometres,
billionths of a meter. They range from 3/100ths to 3/1,000ths of a nanometer.
Each radionuclide decays at its own unique rate which cannot be altered by any
chemical or physical process.
A useful measure of this rate is the half-life of the radionuclide. Half-life is defined as
the time required for the activity of any particular radionuclide to decrease to one-
half of its initial value.
In other words one-half of the atoms have reverted to a more stable state material.
Half-lives of radio nuclides range from microseconds to billions of years.
More exacting calculations can be made for the half-life of these materials; however,
these times are commonly used.
λ = 0.693 / t1/2
where,
t ½ - half life time of radioactive isotope
N= N0 e - λT
Where,
Page 16 of 111
IRONSCAN Institute of Non-Destructive testing and services
N = Decay Rate,
N0 = Initial Number of Atoms,
λ = Decay Constant
T = Time taken
Isotopes Half life time Gamma ray energy Gamma ray Dosage rate
(Mev) (Roentgens per hour at 1 M)
Page 17 of 111
IRONSCAN Institute of Non-Destructive testing and services
13. X-Rays:
Page 18 of 111
IRONSCAN Institute of Non-Destructive testing and services
The form of energy which emitted and absorbed by charged particles is known as
Electromagnetic radiation or EM radiation or EMR.
Electromagnetic radiation has both electric and magnetic field components, which
oscillates perpendicular to each other as well as perpendicular to the direction of
energy and wave propagation.
They travel with the speed of light in vacuum that is 3 x 108 cm/sec. The refraction of
electromagnetic light produced a band of colour like in a rainbow is termed as
electromagnetic spectrum.
The electromagnetic spectrum started from radio wave ended at gamma radiation of
the short-wavelength and high-frequency end; hence it covers the wavelengths from
thousands of kilometres down to a fraction of the size of an atom.
Different electromagnetic radiations with their frequencies and wave length us as follows.
X-rays are electromagnetic radiation; which can travel in a straight line with
the speed of visible light that is 1, 86,000 miles/sec.
X-rays are invisible; it cannot be heard or smelt.
They do not show any deflection, reflection or diffraction in magnetic or
electric field. Just like visible light, they show wave properties like
interference, diffraction and refraction.
They have tendency to produce electric field and magnetic field which is
perpendicular to their path of propagation.
X-rays can travel in vacuum and not required an medium. They have enough
penetration power and can penetrate liquids, solid and gases.
The penetration power depends upon the quality, intensity and wavelength of
light beam.
They can be easily absorbed by matter and absorption power depends on the
atomic structure as well as wavelength of X-rays.
Because of their penetration power, they can ionized any matter and can
emitted visible light when interact with a certain material. This phenomenon
is known as Fluorescence.
Certain substance like Methylene blue gets bleached with X-rays. Due to high
energy, X-rays produced active free radical during any chemical reaction.
X-rays have wide applications in biological field also. They used in the
treatment of malignant lesions. They have germicidal effect and can cause the
destruction of enzymes
Page 20 of 111
IRONSCAN Institute of Non-Destructive testing and services
X-rays are generated when free electrons give up some of their energy when
they interact with the orbital electrons or nucleus of an atom.
The energy given up by the electron during this interaction appears as
electromagnetic energy known as X-radiation.
There are two different atomic processes that can produce x-ray photons.
1. Bremsstrahlung – useful for medical and industrial applications.
2. K-shell emission.
15.1 Bremsstrahlung:
Page 21 of 111
IRONSCAN Institute of Non-Destructive testing and services
Page 23 of 111
IRONSCAN Institute of Non-Destructive testing and services
X-rays generator consists of a tube, the high voltage generator, the control
console, and the cooling system. A stream of high speed electrons strikes at a
target material like tungsten.
The interaction between high speed electron and heavy metal produce X-rays.
This process accomplished in an X-ray tube which is a main component of an X-
ray generator.
The tube contains the cathode filament which is heated with a low-voltage
current. The heating of the filament produces free electrons.
Since there is a large electrical potential between the cathode and the anode
due to high-voltage generator so electrons are strongly attracted to the anode
target. The movement of electrons between the cathode and the anode
creates current in tube.
As the temperature increases, the number of electrons increases in tube.
The control console regulates the filament temperature, which corresponds to
the intensity of the X-ray output.
Page 24 of 111
IRONSCAN Institute of Non-Destructive testing and services
(a) Image enlargement, (b) image distortion, and (c) image overlap for point sources of radiation. (d) Degree of image
unsharpness from a radiation source of finite size. Lo, source-to object distance; Li, source-to-image distance; So, size
of object; Si, size of image; Ug, geometric unsharpness; F, size of focal spot; t, object-to-image distance
Page 25 of 111
IRONSCAN Institute of Non-Destructive testing and services
Any point source which spreads its influence equally in all directions without a limit
to its range will obey the inverse square law.
This comes from strictly geometrical considerations. The intensity of the influence at
any given radius (r) is the source strength divided by the area of the sphere.
Being strictly geometric in its origin, the inverse square law applies to diverse
phenomena. Point sources of gravitational force, electric field, light, sound, and
radiation obey the inverse square law.
As one of the fields which obey the general inverse square law, a point radiation
source can be characterized by the diagram whether you are talking about
Roentgens, rads, or rems.
All measures of exposure will drop off by the inverse square law. For example, if the
radiation exposure is 100 mR/hr at 1 inch from a source, the exposure will be 0.01
mR/hr at 100 inches.
In industrial radiography, the intensity at one distance is typically known and it is necessary
to calculate the intensity at a second distance. Therefore, the equation takes on the form
of:
Where:
I1 = Intensity 1 at D1
I2 = Intensity 2 at D2
D1 = Distance 1 from source
D2 = Distance 2 from source
Note: This is the commonly found form of the equation. However, for some it is easier to
remember that the intensity time the distance squared at one location is equal to the
intensity time the distance squared at another location. The equation in this form is:
I1 x D12 = I2 x D22
The thickness of any given material where 50% of the incident energy has been
attenuated is known as the half-value layer (HVL). The HVL is expressed in units of
distance (mm or cm).
HVL = ln (2) / μ
Therefore, the HVL are related as follows:
HVL = 0.693 / μ
Where, µ = The linear attenuation coefficient
Page 27 of 111
IRONSCAN Institute of Non-Destructive testing and services
The thickness of any given material where 10% of the incident energy has
been attenuated is known as the tenth -value layer (TVL). The HVL is
expressed in units of distance (mm or cm).
The TVL is inversely proportional to the attenuation coefficient.
TVL = ln (10) / μ
Therefore, the HVL are related as follows:
TVL = 2.3026 / μ
Where, µ = The linear attenuation coefficient
19.2 HVL and TVL Values for X-rays and gamma rays
X-rays:
Values are in centi meters
Gamma rays:
Isotopes Concrete Steel Lead Uranium
HVL TVL HVL TVL HVL TVL HVL TVL
Iridium-192 4.6 1.4 1.25 4.00 0.48 1.60 0.31 1.00
Cobalt -60 6.6 21.8 2.0 6.60 1.20 4.00 0.70 2.20
The linear attenuation coefficient (µ) describes the fraction of a beam of x-rays
or gamma rays that is absorbed or scattered per unit thickness of the
absorber.
This value basically accounts for the number of atoms in a cubic centimetre
volume of material and the probability of a photon being scattered or
absorbed from the nucleus or an electron of one of these atoms.
Using the transmitted intensity equation above, linear attenuation coefficients
can be used to make a number of calculations.
These include:
The intensity of the energy transmitted through a material when the incident
x-ray intensity, the material and the material thickness are known.
The intensity of the incident x-ray energy when the transmitted x-ray
intensity, material, and material thickness are known.
The thickness of the material when the incident and transmitted intensity, and
the material are known.
The material can be determined from the value of µ when the incident and
transmitted intensity, and the material thickness are known.
Page 29 of 111
IRONSCAN Institute of Non-Destructive testing and services
Absorption of x-rays occurs when the x-ray photon is absorbed, resulting in the
ejection of electrons from the outer shell of the atom, and hence the ionization
of the atom. Subsequently, the ionized atom returns to the neutral state with the
emission of an x-ray characteristic of the atom.
This subsequent emission of lower energy photons is generally absorbed and
does not contribute to (or hinder) the image making process. Photoelectron
absorption is the dominant process for x-ray absorption up to energies of about
500 KeV.
Photoelectron absorption is also dominant for atoms of high atomic numbers.
When a beam of x-rays or gamma rays strikes a test piece, secondary radiation
is scattered in all directions.
This causes a haze over all or part of the image, thus reducing contrast and
visibility of detail.
The types of scattering are
1. Compton scattering
2. Thomson scattering
Page 30 of 111
IRONSCAN Institute of Non-Destructive testing and services
It also known as Rayleigh, coherent, or classical scattering, it occurs when the x-ray
photon interacts with the whole atom so that the photon is scattered with no change
in internal energy to the scattering atom, nor to the x-ray photon.
Thomson scattering is never more than a minor contributor to the absorption
coefficient. The scattering occurs without the loss of energy. Scattering is mainly in
the forward direction.
Page 31 of 111
IRONSCAN Institute of Non-Destructive testing and services
Page 32 of 111
IRONSCAN Institute of Non-Destructive testing and services
The three factors controlling unsharpness are source size, source to object
distance, and object to detector distance.
The source size is obtained by referencing manufacturer’s specifications for a
given X-ray or gamma ray source. As the source size decreases, the geometric
unsharpness also decreases.
For a given size source, the unsharpness can also be decreased by increasing
the source to object distance, but this comes with a reduction in radiation
intensity.
The object to detector distance is usually kept as small as possible to help
minimize unsharpness.
However, there are situations, such as when using geometric enlargement,
when the object is separated from the detector, which will reduce the
definition.
The area of varying density at the edge of a feature that results due to
geometric factors is called the penumbra.
These values refer to the degree of penumbra shadow in a radiographic
image. Since the penumbra is not nearly as well defined as shown in the image
it is difficult to measure it in a radiograph.
Therefore it is typically calculated. The source size must be obtained from the
equipment manufacturer or measured. Then the unsharpness can be
calculated using measurements made of the setup.
Page 33 of 111
IRONSCAN Institute of Non-Destructive testing and services
Ug = f * b/a
Where, f = source focal-spot size
a = distance from the source to front surface of the object
b = the thickness of the object
For the case when the detector is not placed next to the sample as shown in
the following figure, such as when geometric magnification is being used, the
calculation becomes:
Ug = f* b/a
Where,f = source focal-spot size.
a = distance from x-ray source to front surface of material/object
b = distance from the front surface of the object to the detector
Page 34 of 111
IRONSCAN Institute of Non-Destructive testing and services
Magnification factor M = (a + b) / a
Where: M = magnification
a = distance from source to object
b = distance from object to detector
Page 35 of 111
IRONSCAN Institute of Non-Destructive testing and services
These are used to improve image contrast, particularly when the radiation
intensity is low or when the radiation energy is high.
The screens are useful at higher energies because the sensitivity of films and
fluorescent screens decreases as the energy of the penetrating radiation
increases.
When solid silver bromide is formed in the manufacture of film, the
silver atoms give up an orbital electron to a bromine atom. Since the
silver atoms have given up an electron, they have a positive electrical charge
and are silver ions (Ag+).
The bromine atoms have acquired this negative electron and have
become bromide ions (Br-).
The silver bromide crystal is a cubical array of the silver and bromine ions.
The cubical crystalline structure of the silver bromide crystal is not perfect; if
it were, the photographic process could not exist.
Within the crystal lattice structure are extra silver ions called interstitial
ions; these do not occupy a lattice position in the crystal.
There are also foreign molecules or dislocations (distortions) of the crystal
array within the crystal, all of which form latent image sites.
It is the combination of filtration and intensification that makes lead screens the
most widely used in industrial radiography.
High-energy (short wavelength) radiation passes through lead much more readily
than low-energy radiation; in other words, low-energy radiation is more readily
absorbed by a lead screen than high-energy radiation.
Because scattered radiation from a test piece is always of a lower energy than the
incident beam passing through a test piece, a lead screen will absorb a relatively high
percentage of unwanted scattered radiation, but will absorb a somewhat lower
percentage of the image forming radiation.
The screen that faces the top of the film is referred to as the front screen, and the
screen behind the film toward the table or floor is referred to as the back screen.
The main function of the beam screen is to absorb unwanted back scatter; back
screens can be of any thickness that performs this function adequately, although
usually they are of the same thickness as the front screen.
In the radiography of thin-gage or low-density materials, in which low photon
energies are used, care must be exercised to ensure that the front screen does not
excessively filter the image-forming radiation.
Page 37 of 111
IRONSCAN Institute of Non-Destructive testing and services
A variation of lead screens is lead oxide screens, which are made by evenly
coating a paper base with lead oxide.
The result is an extremely flexible screen, equivalent to about 0.013 mm
(0.0005 in.) of lead foil that is lightweight and free of antimony segregation.
Lead oxide screens are available only in light tight envelopes containing
screens on both sides of the film.
The principle of lead oxide screens is essentially the same as for lead screens.
The main differences are a lesser degree of filtration than lead screens and a
greater intensification below 140 keV as well as slightly lesser intensification
at 300 keV than with 0.13 mm (0.005 in.) thick lead screens.
Although lead oxide screens are particularly advantageous for the radiography
of light alloys or thin material, they can also be used in the radiography of
heavier material.
24.6 Screens of Metals Other Than Lead:
Page 38 of 111
IRONSCAN Institute of Non-Destructive testing and services
Foils of many of the heavier metals can be as effective as lead for radiographic
screens, but usually are not as practical, either because the foils are not as
flexible or because of cost.
Gold screens perform as well or better than lead screens, but gold costs
considerably more and tends to work harder when bent, which eventually
cracks the screen.
Tantalum screens exhibit slightly lower filtration but higher intensification
than lead. However, tantalum foil is stiff and springy, which does not allow it
to be shaped to fit around a test piece. Tantalum foil can be used in solid
cassettes, can be polished, and is quite resistant to minor abrasion, but is
expensive to use.
Depleted-uranium screens exhibit greater filtration than lead screens.
However, uranium is brittle, is somewhat difficult to obtain in thin sheets, and
is also expensive to use.
Copper screens have been used, especially with cobalt-60 radiography.
Copper has a lower degree of filtration than lead and a lower intensification
factor, but copper provides greater radiographic sensitivity.
Composite screens of lead, copper, and aluminum (and sometimes other
metal foils) to control the radiation reaching the film have occasionally been
used with super voltage x-ray machines for the inspection of thick steel test
pieces.
Page 39 of 111
IRONSCAN Institute of Non-Destructive testing and services
Film selection and exposure time are two major factors in film radiography.
The selection of radiographic film for a particular application is generally a
compromise between the desired quality of the radiograph and the cost of exposure
time.
The quality of the radiograph depends mainly on film density, gradient, graininess,
and fog, which are functions of film type and development procedure.
Exposure time depends mainly on film speed and on radiation intensity at the film
surface.
Page 40 of 111
IRONSCAN Institute of Non-Destructive testing and services
Page 41 of 111
IRONSCAN Institute of Non-Destructive testing and services
Although x-ray film is sensitive to light, the characteristics of its emulsion are
different from those of emulsions used in photography.
Industrial x-ray film is of two general types, that used for direct exposure
(called direct-exposure, no-screen, or non screen film) and that used with
fluorescent screens (often called screen-type film).
Most industrial x-ray film is of the direct-exposure type and is available in
various combinations of film speed, gradient, and graininess.
The choice of film for a given application depends on the type of radiography
to be performed.
Screen-type films, which are marketed primarily for medical radiography, are
only occasionally used for industrial applications--for example, when a low-
power x-ray machine is used and exposure time is excessive with direct-
exposure x-ray film.
Screen-type films are more sensitive to light than to x-radiation and are
particularly sensitive to the wavelengths emitted by the fluorescent screen
with which they are used.
Although blue-sensitive emulsion types are the ones most often used for
industrial radiography, other emulsions sensitive to ultraviolet and to green
screen light are available.
Page 42 of 111
IRONSCAN Institute of Non-Destructive testing and services
Film Characteristics
Film Type Speed Gradient Graininess
Type 1 Low Very high Very fine
Type 2 High High Fine
Type 3 Medium Medium Coarse
Type 4 (a) Very high (b) Very High (b) (c)
Very medium (d) Very Medium (d) Medium
(a) Normally used with Fluorescent Screen
(b) When used with Fluorescent Screens
(c) Graininess is mainly a characteristic of the Fluorescent screen
(d)When used for direct exposure or with Lead screens.
Page 43 of 111
IRONSCAN Institute of Non-Destructive testing and services
Where,
D is the density,
I0 is the intensity of light incident on the film,
It is the intensity of light transmitted through the film.
Where,
Dr is the reflection density,
I0 is the light intensity incident on the radiographic image,
Ir is the light intensity reflected from the image.
25.5 Exposure:
It is the intensity of radiation multiplied by the time during which it acts, that is, the amount
of energy that reaches a particular area of the film and that is responsible for producing a
particular density on the developed film.
Exposure can be specified either in absolute units (such as ergs per square centimeter or
roentgens) or in relative units (where one particular exposure is used as a reference and all
others are specified in terms of that reference).
Unless equipment and time are available for making the precise radiation measurements
that are required for defining exposure in absolute units, relative exposure is much more
convenient than absolute and is equally as useful.
Exposure chart for Iridium-192 was as follows,
Where,
S.F.D = Source to film distance
R.H.M = Roentgen / hour/ meter
n = Job Thickness in HVL
Page 45 of 111
IRONSCAN Institute of Non-Destructive testing and services
The term speed is used in radiography refers to the sensitivity of the X-ray film
to X-ray or gamma rays.
It is inversely related to the time required for a given intensity of radiation to
produce a particular density on the film; the shorter the exposure, the faster
the film.
In absolute units, film speed is inversely proportional to the total energy
(roentgens) of a particular radiation spectrum (wavelength distribution at a
given kilo voltage) that produces a given density on the film.
For most practical applications, it is convenient and effective to deal with
relative speeds. In using relative speeds, film speeds are expressed in terms of
the speed of one particular film whose relative speed is arbitrarily assigned a
value, for example, 100.
Thus, if film A requires half the exposure of film B, the slower film (film B) is
chosen as the standard and assigned an arbitrary speed of 100.
Film A, which is twice as fast, would have a relative speed of 200. To avoid
making absolute measurements to determine film speed, it is convenient to
refer to a group of film curves such as those in Figure (a).
Curves positioned to the left of the chart require less exposure for a given
density; those to the right, more exposure. Thus, for a density of 2.0, film C in
Figure (a) is faster than film A, and both are faster than film B.
Their relative speed is calculated by determining the differences in log relative
exposure and converting to the antilog.
Another advantage of using groups of characteristic curves is that the visual
assessment of relative speeds can be easily made.
For example, although films A and B have similarly shaped curves running
almost parallel, the curve for film C is radically different.
As calculated above, film C has a relative speed of 800 at a density of 2.0;
however, at greater densities, the relative speed of film C is lower than 800,
and at lesser densities (down to a density of 1.0), the relative speed of film C is
greater than 800.
Therefore, whenever relative speeds are used, the density at which they were
determined must be given.
Page 46 of 111
IRONSCAN Institute of Non-Destructive testing and services
Film gradient is determined from the characteristic curve by finding the slope
of the curve at a given density.
The slope of the characteristic curve changes continuously over its entire
length, as indicated in Figure (a).
The steeper the slope of the curve over a range of relative exposures, the
greater the difference in density and therefore the greater the resolution of
detail; thus, high gradient is important for good radiography.
In Figure (b) tangents to a characteristic curve for a typical x-ray film have
been drawn at two points, and the two corresponding gradients (a/b and c/d)
have been evaluated.
Page 47 of 111
IRONSCAN Institute of Non-Destructive testing and services
Note that the gradient varies from 0.8 in the shallow portion of the curve
(called the toe) to 5 in the steeper portion of the curve.
Two regions of a test piece that are slightly different in thicknesses will
transmit slightly different intensities of radiation to the film. On the
characteristic curve, this would represent a small difference in the log relative
exposure.
For example, assume that at a given kilo voltage the thinner section transmits
20% more radiation than the thicker section.
The difference in log relative exposure is 0.08 and is independent of tube
current, exposure time, or source-to-film distance.
In the toe of the characteristic curve where, the gradient is only 0.8, the
density difference is only 0.06, as indicated in Figure(c).
However, in the steeper portion of the curve where the gradient is 5, the
density difference is 0.40.
This effect of film gradient is the main reason why it is best to use high
exposures, obtaining the highest density that can be viewed on a given
illuminator for the greatest resolution of minor differences in transmitted
radiation.
It is often more convenient to express gradient as an average over a given
useful range of densities than for a single density. This is a simple calculation
in which the difference between the two densities is divided by the difference
Page 48 of 111
IRONSCAN Institute of Non-Destructive testing and services
between the log relative exposures for these densities; this ratio, known as
the average gradient, is shown in Figure (d).
25.8 Graininess:
The silver halide grains that are contained in the emulsion of x-ray film are minute
and can be seen only with a high-power microscope, such as an electron microscope.
Even though the emulsion on each side of the film is only about 0.013 mm (0.0005
in.) thick, the grains are piled on top of each other in countless numbers.
When the exposed-and processed radiograph is viewed, these small individual silver
grains appear grouped together in relatively large masses.
This clumping, which is visible to the unaided eye or at low magnification, produces
the visual impression called graininess.
All films exhibit some degree of graininess. In general, slower films have less
graininess than faster films.
This general relation is reflected in Table, which is an old classification scheme
specified by the American Society for Testing and Materials (ASTM) in the earlier
1984 edition of ASTM E 94.
Although this classification scheme is somewhat arbitrary, it is still sometimes
specified in codes and specifications. Type 1 films have the least graininess, type 3
films exhibit the most graininess, and type 2 films are intermediate.
Page 49 of 111
IRONSCAN Institute of Non-Destructive testing and services
Material used for penetrameters is the same as that of the test piece or
has the same absorption characteristics. Suitable penetrameter
Page 50 of 111
IRONSCAN Institute of Non-Destructive testing and services
materials for various metallic test pieces are grouped in ASTM E 1025
and E 142 as shown in Table.
In use, the penetrameter is normally placed directly on the surface of
the test piece that faces the source.
Alternatively, when the test piece is small or when its shape is
unfavorable, the penetrameter is placed on a block or shim of the same
nominal composition and thickness as the test piece, with the upper
surface of the block at the same distance from the recording plane as
the upper surface of the test piece.
In pipe radiography, when permitted by specification, the
penetrameter can be placed on the surface of the pipe that faces the
film.
This film-side placement of the penetrameter is usually read to a
tighter sensitivity requirement (for example, 2-1T instead of 2-2T)
The location of the penetrameter should be such that its projected
image does not coincide with any area of interest in the image of the
test piece
The image of the penetrameter is viewed at the same time as the
image of the test piece; the two images appear simultaneously on the
recording medium.
The image of the penetrameter is evaluated separately from the image
of the test piece and is used solely as a direct check on the quality of
the recorded image
Image-quality levels are usually expressed as the size of the smallest
penetrameter feature (such as hole size or wire size) that is clearly
visible in the processed image
A penetrameter is never used as a size standard against which flaw
sizes are compared
Table : ASTM penetrameter material grades for the material groups in ASTM E 1025 and E 142
There are two general types of plaque penetrameters specified by ASTM and
the American Society of Mechanical Engineers (ASME): rectangular plaque
penetrameters, and circular plaque penetrameters.
The rectangular plaque design is specified by ASTM and ASME for plaque
thicknesses of 0.13 to 1.3 mm (0.005 to 0.050 in.).
The holes in rectangular plaque penetrameters are T, 2T, and 4T in diameter,
where T is the thickness of the plaque.
A notch system is also used to identify the ASTM grade of the penetrameter.
The circular plaque design is larger than the rectangular plaque design and is
specified for plaque thicknesses of 1.5 to 4 mm (0.060 to 0.160 in.).
Sensitivity is usually expressed in terms of penetrameter thickness (as a
percentage of test piece), and resolution is determined by the smallest hole
size visible in the radiograph.
For example, an image-quality level of 2-2T indicates that the thickness of the
penetrameter equals 2% of section thickness and the 2T hole is visible. If
image quality of 1-1T were required, a radiograph would be acceptable if the
outline of a 1% penetrameter were distinguishable.
Alternatively, image quality can be expressed as a percentage only. In the
ASTM or ASME systems, the equivalent sensitivity in percent is based on
visibility of the 2T hole. The following Table lists equivalent sensitivities for
various standard image-quality levels.
Image-quality level Penetrameter thickness, Smallest visible hole size Equivalent sensitivity, %
% of test piece thickness
1 – 1T 1 1T 0.7
1 – 2T 1 2T 1.0
1 – 4T 1 4T 1.4
2 – 1T 2 1T 1.4
2 – 2T 2 2T 2.0
2 – 4T 2 4T 2.8
4 – 1T 4 1T 2.8
4 – 2T 4 2T 4.0
4 – 4T 4 4T 5.6
Page 53 of 111
IRONSCAN Institute of Non-Destructive testing and services
Equivalent sensitivity:
Where,
n = IQI Thickness expressed as % of specimen thickness
T= IQI Thickness
These are widely used in Europe, and a standard design is used in the United
Kingdom, Germany, the Netherlands, and Scandinavia and by the International
Organization for Standardization (ISO) and the International Institute of Welding
(IIW).
In the United States, the penetrameter design specified in ASTM E 747 is widely
used.
The ISO design of wire-type penetrameters has a group (typically seven) of straight
30 mm (1.2 in.) wires made of the same material as the test piece.
This ISO standard is similar to the standard of Deutsche Industrie Norm (DIN 54109),
which consists of sixteen wire sizes of three metals--steel, aluminum, and copper.
The wire sizes on a wire penetrameter consist of different groupings.
In ASTM system, for example, a wire penetrameter may have one of four groupings
of six wire sizes:
1. First group (ASTM set A) contains wire numbers 1 through 6, which
correspond to wire diameters of 0.08 through 0.25 mm (0.0032 through
0.010 in.)
2. The second group (ASTM set B) contains wire numbers 6 through 11, which
correspond to wire diameters of 0.25 through 0.81 mm (0.010 through 0.032
in.)
3. The third group (ASTM set C) contains wire numbers 11 through 16, which
correspond to wire diameters of 0.81 through 2.54 mm (0.032 through 0.100
in.)
4. The fourth group (ASTM set D) contains wire numbers 16 through 21, which
correspond to wire diameters of 2.54 through 8.13 mm (0.100 through 0.320
in.)
Page 56 of 111
IRONSCAN Institute of Non-Destructive testing and services
Page 57 of 111
IRONSCAN Institute of Non-Destructive testing and services
In DIN system, a wire penetrameter may have one of three groupings of seven wire
sizes:
1. First group (DIN set 1) contains wire numbers 1 through 7, which correspond
to wire diameters of 3.20 through 0.80 mm.
2. The second group (DIN set 2) contains wire numbers 6 through 12, which
correspond to wire diameters of 1.00 through 0.25 mm.
3. The third group (DIN set 3) contains wire numbers 10 through 16, which
correspond to wire diameters of 0.40 through 0.10 mm.
Radiographic sensitivity is calculated using the following formula and all the
radiograph shall have a minimum 2% sensitivity.
Page 58 of 111
IRONSCAN Institute of Non-Destructive testing and services
Page 59 of 111
IRONSCAN Institute of Non-Destructive testing and services
Page 60 of 111
IRONSCAN Institute of Non-Destructive testing and services
Page 61 of 111
IRONSCAN Institute of Non-Destructive testing and services
Page 62 of 111
IRONSCAN Institute of Non-Destructive testing and services
27. Collimators:
Side scatter from walls, equipment, and other structures in the x-ray room can
be greatly reduced by improving the directionality of the x-ray beam.
Directionality can be improved by the use of a collimator, which is often a
thick lead diaphragm with a small hole through the middle.
Page 63 of 111
IRONSCAN Institute of Non-Destructive testing and services
28. Filtration:
When the test piece has very thin sections adjacent to thick sections or when
the direct beam can strike the detector after passing around the test piece,
undercutting may be encountered.
If undercutting occurs, additional filtration (that is, more than can be achieved
with conventional lead screens) is necessary.
This may adequately eliminate overexposure in thin regions of the test piece
and also along the perimeter of the test piece.
Such a filter is particularly useful for reducing undercutting when a lead mask
around the test piece is impractical or when the test piece may be damaged
by masking with liquid absorbers or metallic shot.
Page 64 of 111
IRONSCAN Institute of Non-Destructive testing and services
Page 65 of 111
IRONSCAN Institute of Non-Destructive testing and services
29.1. Development:
The invisible latent image is converted into a visible image by the chemical
process of development.
The developer solution supplies electrons that migrate into the sensitized
grains and convert the other silver ions into black metallic silver.
This causes the grains to become visible black specks in the emulsion.
Radiographic film is generally developed in an automatic processor. A
schematic of a typical processor is shown below.
The four components correspond to the four steps in film processing. In a
conventional processor, the film is in the developer for 20 to 25 seconds. All
four steps require a total of 90 seconds.
a) Reducer:
Chemical reduction of the exposed silver bromide grains is the process that
converts them into visible metallic silver. This action is typically provided by
two chemicals in the solution: Phenidone and hydroquinone. Phenidone is the
more active and primarily produces the mid to lower portion of the gray scale.
Hydroquinone produces the very dense, or dark, areas in an image.
b) Activator:
c) Restrainers:
d) Preservative:
e) Hardener:
Page 66 of 111
IRONSCAN Institute of Non-Destructive testing and services
29.2. Fixing:
After leaving the developer the film is transported into a second tank, which
contains the fixer solution.
The fixer is a mixture of several chemicals that perform the following
functions.
a) Neutralizer:
b) Clearing:
The fixer solution also clears the undeveloped silver halide grains from the
film. Ammonium or sodium thiosulfate is used for this purpose.
The unexposed grains leave the film and dissolve in the fixer solution.
The silver that accumulates in the fixer during the clearing activity can be
recovered.
The usual method is to electroplate it onto a metallic surface within the silver
recovery unit.
c) Preservative:
d) Hardener:
29.4. Drying:
The final step in processing is to dry the film by passing it through a chamber
in which hot air is circulating The final step in processing is to dry the film by
passing it through a chamber in which hot air is circulating
X- ray film should be grasped only at the corners and inserted into the clips of
the film hangers and locked in place. The upper corner of the film grasped and
a Hatched to the top clips. The film now ready to be placed in the developing
tank.
The clips of the film hanger should be cleaned periodically to prevent an
accumulation of chemicals that may rundown on the film during processing
and cause streaks.
Manufacturers specify the correct type.
Two forms of safe-light can be used in the radiographic dark room.
1. Direct: a diffused light shines directly over the work point, such as the
dry and wet bench.
2. Indirect: the filtered light is directed up to the ceiling where it is
reflected over the room.
Safe-light efficiency: Safe-lights should be placed so that the work of the dark
room can be done without fumbling. where the dry and wet benches are
separate , a small direct wall light should be provided for each
The film should be moved to the rinse step as soon as the timer alarm sounds.
There are many variables, such as temperature and chemical activity that can
affect the level of processing that a film receives. Each type of film is designed
and manufactured to have specified sensitivity (speed) and contrast
characteristics.
Under processing: If a film is under processed its sensitivity and contrast will
be reduced below the specified values. The loss of sensitivity can usually be
compensated for by increasing exposure but the loss of contrast cannot be
recovered.
Over processing: Over processing can increase sensitivity. The contrast of
some films might increase with over processing, up to a point, and then
decrease. A major problem with over processing is that it increases fog (base
plus fog density) which contributes to a decrease in contrast.
Radiography is the method most often specified for the inspection of welded
joints between tubes or pipes or between a tube and a pressure vessel when
the weldments is manufactured to ASME Boiler and Pressure Vessel Code
specifications.
Other hollow sections such as cylindrical bosses on castings and forgings,
brazed tube-and-socket joints, large-diameter transmission pipe (line pipe),
and longitudinally welded square or circular structural tubing, are frequently
inspected by radiography.
There are three major inspection techniques for tubular sections:
This method is most commonly used for the inspection of pipelines where the
weld can be radiographed in one exposure, making the technique rapid and
cost effective
Where access to the bore is not possible or the pipe diameter is too small to
permit the use of an internal source then the double wall, single image (DWSI)
technique is used.
Here the film is placed on the outside of the pipe on the farthest side from the
radiographic source, as shown in Figure.
The source may be offset slightly to avoid an image of the upper part of the
weld to be projected onto the film or directly in line.
Page 73 of 111
IRONSCAN Institute of Non-Destructive testing and services
The source may be close to or a substantial distance from the pipe, the
location being a compromise between a less sharp image but short exposure
time for a small stand-off and sharper image but longer exposure time for a
large stand-off.
The need to penetrate two wall thicknesses means that the sensitivity will be
poorer than with the single wall single image technique.
The technique also requires multiple exposures to enable the complete
circumference of the pipe to be examined – specification or contract
requirements frequently specify the minimum numbers of exposures to
ensure complete coverage and images of an acceptable quality.
The technique is generally used on pipes over 80mm in diameter.
This technique is generally used only on pipes less than 75-80mm in diameter.
By offsetting the source from the weld centre line and using a long source to
film distance it is possible to project an image onto the film of both the upper
and the lower parts of the weld as shown in Figure.
As with the DWSI technique multiple exposures are required to achieve
complete coverage.
Page 74 of 111
IRONSCAN Institute of Non-Destructive testing and services
Page 75 of 111
IRONSCAN Institute of Non-Destructive testing and services
The most favorable view for fillet welds is a direction that roughly bisects the
angle between the legs of the section.
Views normal to the surface of the weld are not always satisfactory. Flaws in a
weld such as dense inclusions due to the
deposition of metal from a tungsten electrode, excessive segregation in light
alloys, porosity, and shrinkage can be revealed regardless of view.
However, cracks or non fusion along the edge of the fusion zone can be best
revealed when the viewing direction is parallel to the face of the joint.
The detection of centerline cracks or incomplete penetration requires a
normal view.
Only square-groove butt joints, U-groove joints, and J-groove joints can be
satisfactorily inspected with a normal view; other types of joints, including V-
groove joints, bevel joints, flare joints, and fillets, require two views, one
parallel to each original prepared edge, for correct inspection for cracks or
regions of incomplete fusion.
Page 76 of 111
IRONSCAN Institute of Non-Destructive testing and services
Proper identification of both the radiograph and the test piece, clarity of the
penetrameter, suitability of radiographic techniques, adequacy of coverage,
and the techniques of image processing in real-time systems or the precision
and Uniformity of film processing in film radiography all offset the image that
is being interpreted.
Page 77 of 111
IRONSCAN Institute of Non-Destructive testing and services
Film radiographs and real time radiographs are interpreted similarly with
respect to the recognition of flaws and test piece features.
The primary difference is that film radiographs are negative images, while
real-time radiographs are generally positive images (which may also be
enhanced with image processing).
1. Define the quality of the radiographic image, which includes a critical analysis
of the radiographic procedure and the image-developing procedure
2. Analyze the image to determine the nature and extent of any abnormal
condition in the test piece
3. Evaluate the test piece by comparing interpreted information with standards
or specifications
4. Report inspection results accurately, clearly, and within proper administration
channels
Crescent-shaped light Faulty film handling Keep film flat during handling.
areas Use only clean, dry film
hangers.
Light circular patches Air bubbles on film during Agitate immediately upon
development immersion of film in
developer.
Page 80 of 111
IRONSCAN Institute of Non-Destructive testing and services
Circular or drop shaped Water or fixer splashed on Avoid splashing film with
light patches film water or fixer solution.
before development
Light spots or areas Dust or lint between screens Keep screens clean.
and
film
Page 81 of 111
IRONSCAN Institute of Non-Destructive testing and services
The screen should be illuminated from behind with light of sufficient intensity
to reveal variations in photographic density up to a nominal film density of at
least 3.0.
There may be a need for a smaller, more intensely illuminated viewer for
evaluating small areas of film having densities up to 4.5 or more.
Viewing screens of high-intensity illuminators should be cooled by blowers or
other suitable apparatus to prevent excessive heat from damaging films and
to extend lamp life.
When interpreting paper radiographs or x-ray radiographs, specular light as
from a spotlight or high-intensity lamp should be directed onto the radiograph
from the side at an angle of about 30°.
Background lighting should be heavily subdued.
A densitometer can be provided for accurate evaluation of small variations in
photographic density or for quantitative evaluation of radiographic and
processing techniques.
A transmission densitometer is used with films, and a reflection densitometer
is used with paper radiographs.
32. Discontinuities
a. Flaws in Castings
1. Micro shrinkage
2. Shrinkage porosity
3. Gas porosity
4. Dispersed Discontinuities
Page 82 of 111
IRONSCAN Institute of Non-Destructive testing and services
5. Tearing
6. Cold cracks
7. Cold shunt
8. Misruns
9. Inclusions
10. Un-fused chaplets
11. Core shift
12. Shrinkage cavities
13. Center line shrinkage
14. Segregation
15. Surface irregularities
b. Flaws in Weldments
Most weld flaws consist of abrupt changes in homogeneity and can be readily
detected by radiographic inspection.
Stresses that are induced in the metal by welding but that are not
accompanied by a physical separation of material will not be detected by
radiography.
1. Undercutting
2. Incomplete fusion
3. Cracks
4. Porosity
5. Slag inclusion
6. Incomplete penetration
Cold lap is a condition where the weld filler metal does not properly fuse with the
base metal or the previous weld pass material (inter pass cold lap). The arc does not
melt the base metal sufficiently and causes the slightly molten puddle to flow into
the base material without bonding
Page 83 of 111
IRONSCAN Institute of Non-Destructive testing and services
Porosity is the result of gas entrapment in the solidifying metal. Porosity can take
many shapes on a radiograph but often appears as dark round or irregular spots or
specks appearing singularly, in clusters, or in rows. Sometimes, porosity is elongated
and may appear to have a tail. This is the result of gas attempting to escape while the
metal is still in a liquid state and is called wormhole porosity. All porosity is a void in
the material and it will have a higher radiographic density than the surrounding area.
Cluster porosity is caused when flux coated electrodes are contaminated with
moisture. The moisture turns into a gas when heated and becomes trapped in the
weld during the welding process. Cluster porosity appears just like regular porosity in
the radiograph but the indications will be grouped close together.
Slag inclusions are non metallic solid material entrapped in weld metal or between
weld and base metal. In a radiograph, dark, jagged asymmetrical shapes within the
weld or along the weld joint areas are indicative of slag inclusions.
Page 84 of 111
IRONSCAN Institute of Non-Destructive testing and services
Incomplete penetration (IP) or lack of penetration (LOP) occurs when the weld
metal fails to penetrate the joint. It is one of the most objectionable weld
discontinuities. Lack of penetration allows a natural stress riser from which a crack
may propagate. The appearance on a radiograph is a dark area with well-defined,
straight edges that follows the land or root face down the centre of the weldments.
Incomplete fusion is a condition where the weld filler metal does not properly fuse
with the base metal. Appearance on radiograph: usually appears as a dark line or
lines oriented in the direction of the weld seam along the weld preparation or joining
area.
Internal concavity or suck back is a condition where the weld metal has contracted
as it cools and has been drawn up into the root of the weld. On a radiograph it looks
similar to a lack of penetration but the line has irregular edges and it is often quite
wide in the centre of the weld image.
Page 85 of 111
IRONSCAN Institute of Non-Destructive testing and services
External or crown undercut is an erosion of the base metal next to the crown of the
weld. In the radiograph, it appears as a dark irregular line along the outside edge of
the weld area.
Offset or mismatch are terms associated with a condition where two pieces being
welded together are not properly aligned. The radiographic image shows a
noticeable difference in density between the two pieces. The difference in density is
caused by the difference in material thickness. The dark, straight line is caused by the
failure of the weld metal to fuse with the land area.
Page 86 of 111
IRONSCAN Institute of Non-Destructive testing and services
Excess weld reinforcement is an area of a weld that has weld metal added in excess
of that specified by engineering drawings and codes. The appearance on a
radiograph is a localized, lighter area in the weld. A visual inspection will easily
determine if the weld reinforcement is in excess of that specified by the engineering
requirements.
Cracks can be detected in a radiograph only when they are propagating in a direction
that produces a change in thickness that is parallel to the x-ray beam. Cracks will
appear as jagged and often very faint irregular lines. Cracks can sometimes appear as
"tails" on inclusions or porosity.
Tungsten inclusions. Tungsten is a brittle and inherently dense material used in the
electrode in tungsten inert gas welding. If improper welding procedures are used,
tungsten may be entrapped in the weld. Radiographically, tungsten is more dense
than aluminium or steel, therefore it shows up as a lighter area with a distinct outline
on the radiograph.
Page 87 of 111
IRONSCAN Institute of Non-Destructive testing and services
Oxide inclusions are usually visible on the surface of material being welded
(especially aluminium). Oxide inclusions are less dense than the surrounding material
and, therefore, appear as dark irregularly shaped discontinuities in the radiograph.
Burn-Through results when too much heat causes excessive weld metal to penetrate
the weld zone. Often lumps of metal sag through the weld, creating a thick globular
condition on the back of the weld. These globs of metal are referred to as icicles. On
a radiograph, burn-through appears as dark spots, which are often surrounded by
light globular areas (icicles).
Cavity shrinkage appears as areas with distinct jagged boundaries. It may be produced
when metal solidifies between two original streams of melt coming from opposite directions
to join a common front. Cavity shrinkage usually occurs at a time when the melt has almost
reached solidification temperature and there is no source of supplementary liquid to feed
possible cavities.
Page 88 of 111
IRONSCAN Institute of Non-Destructive testing and services
Dendrite shrinkage is a distribution of very fine lines or small elongated cavities that may
vary in density and are usually unconnected.
Sponge shrinkage shows itself as areas of lacy texture with diffuse outlines, generally
toward the mid-thickness of heavier casting sections. Sponge shrinkage may be dendrite or
filamentary shrinkage. Filamentary sponge shrinkage appears more blurred because it is
projected through the relatively thick coating between the discontinuities and the film
surface.
Cold shuts generally appear on or near a surface of cast metal as a result of two
streams of liquid meeting and failing to unite. They may appear on a radiograph as
cracks or seams with smooth or rounded edges.
Page 89 of 111
IRONSCAN Institute of Non-Destructive testing and services
Inclusions are non metallic materials in an otherwise solid metallic matrix. They may
be less or denser than the matrix alloy and will appear on the radiograph,
respectively, as darker or lighter indications. The latter type is more common in light
metal castings.
Sources of Exposure:
Page 90 of 111
IRONSCAN Institute of Non-Destructive testing and services
Routes of Exposure:
There are two principal routes by which humans are exposed to ionizing
radiation,
1. External Exposure
2. Internal Exposure
External Exposures:
External exposures are those which are due to sources located outside of the
body which emit radiation of a type and energy sufficient to reach and
penetrate the body.
Examples of penetrating radiations include gamma rays, x-rays, and high
energy beta particles (e.g., as from 32P).
External exposures cannot result from sources which emit only alpha particles
or low energy beta particles since these radiations do not penetrate the dead
outer cell layer of the skin.
As with all radiation exposures, the size of the dose resulting from an external
exposure is a function of:
1. The activity of the source;
2. The time or duration of the exposure;
3. The distance from the source to the tissue being irradiated;
4. The amount of shielding between the source and the tissue.
In contrast to the situation for internal exposures, these factors can be
altered for a particular external exposure situation with a resultant increase
or decrease in the dose received.
Internal Exposures
Page 91 of 111
IRONSCAN Institute of Non-Destructive testing and services
Internal exposures are those that result from radioactive material which has
been taken up by the body as a result of ingestion, inhalation, injection, or
absorption through the skin.
Such exposures are of concern for all radioactive materials regardless of the
type of ionizing radiation emitted.
However, are radioactive materials which emit alpha and beta particles, these
radiations cause significant damage to tissue when depositing their energy
along highly localized paths.
In contrast to the situation for external exposures, the source-to-tissue
distance, exposure duration, and source strength cannot be altered for
internal sources.
Instead, once a quantity of radioactive material has been taken up by the
body, an individual is essentially “committed” to the dose which will result
from the quantities and forms of the particular radio nuclides involved.
Page 92 of 111
IRONSCAN Institute of Non-Destructive testing and services
As a result of the chemical changes in the cell caused by the direct or indirect
action of ionizing radiation, large biological molecules may undergo a variety
of structural changes which lead to altered function.
Some of the more common effects which have been observed are inhibition of
cell division, denaturation of proteins and inactivation of enzymes, alteration
of membrane permeability, and chromosome aberrations.
Radio sensitivity
The cell nucleus is the major site of radiation damage leading to cell death.
This is due to the importance of the DNA within the nucleus in controlling all
cellular function.
Damage to the DNA molecule may prevent it from providing the proper
template for the production of additional DNA or RNA.
This hypothesis is supported by research which has shown that cells are most
sensitive to radiation damage during reproductive phases (i.e., during DNA
replication).
In general, it has been found that cell radio sensitivity is directly proportional
to rate of cell division and inversely proportional to the degree of cell
differentiation.
Table 1 presents a list of cells which generally follow this principle.
Page 93 of 111
IRONSCAN Institute of Non-Destructive testing and services
Page 94 of 111
IRONSCAN Institute of Non-Destructive testing and services
Radiation Monitoring:
1. In order to ensure that internal and external doses to radiation are kept
ALARA (As Low As Reasonably Achievable) , it is essential that:
1. All radioactive materials remain confined to designated storage and
work locations.
2. That radiation levels resulting from the storage and use of these
materials are adequately known and controlled.
2. These objectives can only be met through the routine practice of various
radiation monitoring activities as follows,
1. Dosimeter
2. TLD badge
3. Film Badge
4. Barrication
Occupational exposure
Page 95 of 111
IRONSCAN Institute of Non-Destructive testing and services
Some human-made radiation sources affect the body through direct radiation,
known as effective dose (radiation) while others take the form of radioactive
contamination and irradiate the body from within. This latter is known as
committed dose
A legal record of dose is kept.
Examples of activities where occupational exposure is a concern include:
Basic Procedures, Practices, and Rules for the Safe Use of Radioactive
materials
In general, both internal and external exposures to ionizing radiation can be
maintained ALARA through the adherence by radioactive material users to a
number of standard procedures, practices, and rules:
7. Radionuclide work areas shall be clearly designated and should, to the extent
possible, be isolated from the rest of the laboratory. The work area shall be
within a hood if the radioactive material to be used is in a volatile form.
Page 96 of 111
IRONSCAN Institute of Non-Destructive testing and services
8. All work surfaces shall be covered with absorbent paper which should be
changed regularly to prevent the build-up of contamination.
9. Work involving relatively large volumes or activities of liquid radioactive
material should be performed in a spill tray lined with absorbent paper.
12. Dosimeters shall be worn when working with relatively large quantities of
radio nuclides which emit penetrating radiation.
Page 97 of 111
IRONSCAN Institute of Non-Destructive testing and services
15. A radiation survey shall be performed and recorded by the radionuclide user
at the end of each procedure involving radioactive materials (the survey may
be made using a portable survey instrument, wipes, or both depending on the
radio nuclides used). All items found to be contaminated shall be placed either
in the radioactive waste container or an appropriately designated area. Any
Page 98 of 111
IRONSCAN Institute of Non-Destructive testing and services
Page 99 of 111
IRONSCAN Institute of Non-Destructive testing and services
Where:
I1 = Intensity 1 at D1
I2 = Intensity 2 at D2
D1 = Distance 1 from source
D2 = Distance 2 from source
Where:
E1 = Exposure at D1
E2 = Exposure at D2
Where,
S.F.D = Source to film distance
R.H.M = Roentgen / hour/ meter
n = Job Thickness in HVL
Note: This is an approximation and is only valid for the straight line portion of the film characteristic curve. See more info section.
Where:
E1 = Exposure 1
E2 = Exposure 2
FD1 = Film Density at Exposure 1
FD2 = Film Density with Exposure 2
5. Geometric Magnification
Where:
M = Magnification
a = Distance from source to object
b = Distance from object to detector
6. Geometric Unsharpness:
Where:
Ug = Geometric unsharpness
f = Focal spot size
a = Distance from source to object
b = Distance from object to detector
7. Indication Depth:
Where:
Example 1
Use Newton's Inverse Square Law to calculate the intensity of a radioactive source at a
different distance than the distance it was originally measured. If the intensity of a Iridium
192 source was found to be 62 milli roentgen/hour 100 feet, what is the exposure at a distance
of 1 foot.
Where:
I1 = Intensity at D1
I2 = Intensity at D2
D1 = Distance 1
D2 = Distance 2
Example 2
A source is producing an intensity of 456 R/h at one foot from the source. What would be the
distance in feet to the 100, 5, and 2 mR/h boundaries.
Using this equation the 100mR/h boundary would be 68 feet, the 5mR/h boundary would be
301.99 feet, and the 2mR/h boundary would be 477.5 feet.
Example 3
Question 3A:
The half life period of yttrium is 20 min. If there are 1000 gm of yttrium present, how much
will be left after 30 min?
Solution:
Question 3B:
If you are given with 500 gms of radioactive carbon. Calculate the decay rate after 60 min if
the half life period is about 30 min.
Solution:
Question 3C:
If 100.0mg of neptunium -239 (239Np) decays to 73.36mg after 24 hours. Find the value of λ
in the growth decay formula for t expressed in days?
Solution:
Since A = 73.36, Ao = 100.0 and t = 1(day) we have to use the equation
0.7336 = e λ
λ = ln 0.7336
λ = -0.309791358
Thus the daily decay rate is approximately 31%
Question 3D:
Carbon-14 used for archeological dating, has a half-life of 5.730 years. Find the decay rate
for carbon-14?
Solution:
If 100mg of carbon-14 (No = 100) is present, we are given that in 5.730 years (t=5.730) there
will be 50mg present (N=50). We we use the formula.
50 = 100e λ (5.730)
0.5 = e 5.730λ
5,730 λ = ln 0.5
λ = ln0.55.730
λ = -0.525
Example 4
Solve the equation for E2, plug in known values, and solve.
Example 5
Solve for E2 or in this case T2 since only the exposure time will be adjusted. Then plug in the
known values and solve for the new exposure time.
Example 6
An exposure of 5.6 milliamperes with a 30 inch tube to film distance produced a good
radiograph. What would the milliamperes need to be, if the tube to film distance is changed to
24 inches?
Solve for E2 or in this case C 2 since only the exposure current will be adjusted. Then plug in
the known values and solve for the new exposure time.
Example 7
Example 8
If a exposure of 6.2 mA-minutes produces a film density of 1.5, what exposure will produce a
film density of approximately 2.5? Assume that both densities fall on the straightened
portion of the film characteristic curve.
Solve the equation for E2, substitute in known values and solve for E2.
Procedure for Using the Film Characteristic Curve to Adjust the Exposure
Example 9
Calculated the geometric magnification when the source to object distance is 80cm and the
object to detector distance is 30cm.
M = (a + b) / a
Where:
M = magnification
M = (a + b) / a
M = (80cm + 30cm)/80cm
M = 1.375
Example 10
Calculate the geometric un sharpness when using an X-ray generator with a 3mm spot size and the
test component is 100 cm from the x-ray tube and 50 cm from the detector.
Ug = f * t/d
First convert all lengths to like units. Then plug the known information into the equation and
solving for Ug.
Ug = f * b/a
Ug = 0.15cm
Example 11
Where:
D = Defect depth
t = thickness of the component
SD = Shift of defect in the radiographs
SM = Shift of marker in the radiographs
Substitute known values into the equation and solve for defect depth.