GE MEDICAL SYSTEMS DMC - 3 INSTALLATION INSTRUCTIONS
REV 6 DIRECTION 46-030627

REVISION HISTORY

REV DATE REASON FOR CHANGE

0 June 5, 1995 Initial Release


1 April 19, 1996 Add LOGIQ™ 400 and LOGIQ™ 500 Installation
2 July 24, 1997 Add LOGIQ a200 and Null Modem
3 December 18, 1997 Change to New Computer
4 April 22, 1999 Add DMC-3 Y2K
5 May 13, 1999 Change File Name
6 September 17, 1999 Page 14, Paragraph 12

LIST OF EFFECTIVE PAGES

PAGE REVISION PAGE REVISION PAGE REVISION


NUMBER NUMBER NUMBER NUMBER NUMBER NUMBER

Title Page 6

A and B 6

1 - 16 6

Please verify that you are using the latest revision of this document. Information pertaining to
this document is maintained on GPC (Global Product Configuration). If you do not know how
to use GPC, contact your Manager or the National Support Center.

A
GE MEDICAL SYSTEMS DMC - 3 INSTALLATION INSTRUCTIONS
REV 6 DIRECTION 46-030627

This page intentionally left blank.

B
GE MEDICAL SYSTEMS DMC - 3 INSTALLATION INSTRUCTIONS
REV 6 DIRECTION 46-030627

! ! ! IMPORTANT ! ! !

Working with DMC-1 and DMC-3

DMC-1 and DMC-3 can now co-exist on the same computer. No


new data should be entered into the DMC-1 program
(ULTRA.BAT). The data in DMC-1 should be used for archival
or referral purposes only.

All new data being transferred goes into the DMC-3 program
(DMC3.BAT).

If any old data is needed from previous studies, it must be


retrieved from DMC-1. This data can be used to compare against
current DMC-3 data or manually inputted as a new study into
DMC-3.

1
GE MEDICAL SYSTEMS DMC - 3 INSTALLATION INSTRUCTIONS
REV 6 DIRECTION 46-030627

Introduction This manual describes the procedure to install the Data Management Center-3
software, using the 2133397 Installation Kit as well as the steps necessary
to upgrade a previous DMC version to DMC - 3 using the H4400RS (2133398)
and H4400TT (2157714) kits and instructions to have DMC-1 and DMC-3
co-exist on the same computer.

This manual will also include installation instructions for a generic computer.

This software may be used with a RT 3200, RT 3200 Advantage, RT 3200


Advantage II/III, RT 4000, LOGIQ™ 400, LOGIQ™ 500 and LOGIQα200. To
use this software with an RT 3200 or an RT 3200 Advantage, the optional
Serial Port Upgrade, H4008SP, must be installed on the RT 3200.

It is recommended that this entire installation manual be read and


understood before beginning the installation.

Perform each step in sequence to avoid problems. If a problem does occur,


recheck all steps. If rechecking does not solve the problem, contact
Headquarters OnLine Center for assistance.

DMC-3 Computer Processor : Intel based PC with a 386 or better


Requirements Minumum amount of memory (RAM) : 2 Mbytes
Minimum amount of hard disk space : 15 Mbytes
Operating system : MS-DOS 5.0 or better

Parts Required To complete the initial installation of the software and connection to the
Ultrasound system, the 2133397 software installation kit is required. The kit
consists of the following parts:

Part No. Description


2157182-3 Program Disk
2132947 Sentinel Key
2132948 Adapter (94-3855-1)
2157263 Adapter (1613378-1)
46-285365P1 Isolated cable (953576-20)
2156167 Operator Manual
46-030627 Install Manual
E6953EH Box of 10, 3-1/2" disks
TC180 ICD Card
2169205 Null Modem Adapter (LOGIQ a200 only)

To upgrade software from the first DMC version to DMC - 3, the 2133398
software installation kit is required. The kit consists of the following parts:

Part No. Description


2157182-4 Program Disk
2132947 Sentinel Key
2156167 Operator Manual
46-030627 Install Manual
2157263 Adapter (1613378-1)
2169205 Null Modem Adapter (LOGIQ a200 only)

2
GE MEDICAL SYSTEMS DMC - 3 INSTALLATION INSTRUCTIONS
REV 6 DIRECTION 46-030627

Parts Required(continued) To upgrade software from DMC - 2 to DMC - 3, the 2157714 software installation
kit is required. The kit consists of the following parts:

Part No. Description


2157772 Program Upgrade Disk
2157263 Adapter (1613378-1)
2156167 Operator Manual
46-030627 Install Manual
TC180 ICD Card
2169205 Null Modem Adapter (LOGIQ a200 only)

Tools Required Standard Service Personnel Tool Kit

NOTE: If this is a DMC - 3 software installation only, proceed to page 9


DMC - 3 Software Installation.

PC Workstation Installation GEMS supplied computer hardware comes tested and loaded with software.
Installation is mainly unpacking the components and attaching the appropri-
ate cable to the correct port on the hardware. Computer hardware changes
from time to time. This procedure does not provide details of hardware
specific installations. It only shows a generic configuration. Refer to the PC
owner's manual for details on specific hardware installations.

(1) ____ Remove the computer and accessories from the cartons and
inspect for physical damage. If damage is apparent, report all
findings to the On Line Center.

(2) ____ Insure the following cables are available:

a) PC Power Cable
b) Keyboard Cable
c) Monitor Power Cable and Signal Cable
d) Printer Power Cable and Signal Cable (if applicable)
e) Isolated serial data cable

GEMS supplied computers have symbols on the respective ports


which the cables go to for the monitor, keyboard and mouse.

(3) ____ Position the computer in place.

CAUTION The computer and printer must be at least six (6) feet away from the
ultrasound system.

3
GE MEDICAL SYSTEMS DMC - 3 INSTALLATION INSTRUCTIONS
REV 6 DIRECTION 46-030627

PC Workstation Installation
(continued)

Illustration 1
PC Keyboard Connection

Adapter Adapter Sentinel Key


2157263 2132948

Null Modem
2169205

(4) ____ Position the keyboard in place. Install the keyboard cable
between the computer and keyboard. See Illustration 2.

Illustration 2
PC Keyboard Connection

Keyboard Connection (Bottom Plug - Green Dot))

4
GE MEDICAL SYSTEMS DMC - 3 INSTALLATION INSTRUCTIONS
REV 6 DIRECTION 46-030627

PC Workstation Installation (4A)____ Position the mouse. Install the plug above the keyboard connec-
(continued) tion. See Illustration 3.

Illustration 3
Monitor Video Connection

Mouse Connection (Top Plug - Blue Dot)

(5) ____ Position the monitor. Install the signal cable between the monitor
and the computer video input. See Illustration 4.

Illustration 4
PC Power Connection

Monitor Video
Connection (Orange Dot)

(6) ____ Make sure the power switches are in the off position on the
computer, monitor and printer. Install the AC power cord between
the computer and the wall outlet.

5
GE MEDICAL SYSTEMS DMC - 3 INSTALLATION INSTRUCTIONS
REV 6 DIRECTION 46-030627

PC Workstation Installation (7) ____ Install the AC power cord between the monitor and the wall outlet.
(continued)

Illustration 5
Monitor Power Connection

Monitor Power
Connection
Monitor Power
Connection

Computer Power
Connection

Interface Cable Installation (1) ____ Install the special Serial Port Cable between the computer
(Yellow Dot) and the ultrasound system using the serial connec-
tors on the back of the units.

The cable adapters are included in the H4400SW software kit. The adapter
accommodates multiple connector combinations with computer hardware. If
necessary, use the 9 pin to 25 pin adapter on the serial data cable going to
the computer.

NOTE: The serial data cable is opto-isolated and limits the ultrasound
system to safe leakage current levels. Do not use any other type
of serial data cable.

NOTE: GEMS hardware comes configured for the program's Sentinel


Key to be installed in the Parallel Printer port (Red Dot). The
isolated data cable from the Ultrasound system is installed in the
computer's serial port labeled COM2. This is the recommended
configuration. If required, this configuration can be changed in
the DMC - 3 software program. Refer to Appendix D of the
DMC - 3 Operator Manual for details.

6
GE MEDICAL SYSTEMS DMC - 3 INSTALLATION INSTRUCTIONS
REV 6 DIRECTION 46-030627

Illustration 6
Serial Data Connection
RT 3200 Advantage-II/III Series

Generic Computer Panel

To Computer Serial Port


COM2
2

RT 3200 Advantage - II/III


Data Port RS-232C Ext TV
In
Ext TV
Out In Out
Camera &
TV Printer
Video Out Shutter
12Vdc
for Camera
Caution
See Operating Manual
Foot Switch

for connection.

7
GE MEDICAL SYSTEMS DMC - 3 INSTALLATION INSTRUCTIONS
REV 6 DIRECTION 46-030627

Illustration 7
Serial Data Connection
LOGIQ α200

Serial Port Connection


Use Null Modem Adapter
2169205

NOTE: The null modem adapter 2169205 is used to connect the special
serial port cable to a LOGIQ™ α200 serial port.

8
GE MEDICAL SYSTEMS DMC - 3 INSTALLATION INSTRUCTIONS
REV 6 DIRECTION 46-030627

Illustration 8
Serial Data Connection
LOGIQ™ 400

Attach DMC cable to either Serial


Port using adapter 2157263. Make
proper selection in the System
Parameter setup menu. Refer to
the Ultrasound System User Manual
Volume 2.

B/W Color

NOTE: The adapter 2157263 is used to connect the special serial port
cable to a LOGIQ™ 400 serial port.

9
GE MEDICAL SYSTEMS DMC - 3 INSTALLATION INSTRUCTIONS
REV 6 DIRECTION 46-030627

Illustration 9
Serial Data Connection
LOGIQ™ 500

Attach DMC cable to any Serial Port.


Make proper selection in the System
Parameter setup menu. Refer to the
Ultrasound System User Manual
Volume 2.

NOTE: The adapter 2157263 is used to connect the special serial port
cable to a LOGIQ™ 500 serial port.

Optional Printer Installation

NOTE: Refer to the Printer Operator Manual for Printer unpacking and
installation.

(1) ____ Position printer in place. Install the AC Power cable to the printer
and the wall outlet. Install the printer data cable (DB25 to 36 pin
Centronics) between the computer parallel port sentinel key and
the printer. Insert the printer paper as necessary.

10
GE MEDICAL SYSTEMS DMC - 3 INSTALLATION INSTRUCTIONS
REV 6 DIRECTION 46-030627

Computer Functional Check Out (1) ____ Once cables are installed, the system should be tested. On
power up, the computer will run self diagnostics and report any
problems encountered. If a problem is encountered, recheck the
work performed. Refer to the DMC 3 Operator Manual Appendix
B, Appendix E and the computer owner's manual for help in
localizing a problem. If a problem persists, call the On Line
Center for support.

GE supplied computer hardware is pre-loaded


with software on the hard drive. The pre-loaded
software includes Disk Operating Software (DOS),
Microsoft Windows and the DMC - 3 program.

DMC Program Configuration Although your computer is already loaded with Data Management Center
software, the program presets need to be configured to select the type of
printer, doctor or hospital name and type of scanner being used.

UConfig Setup:

At the Dos prompt, C:\>, type cd\ultra25 and press Enter.


The screen displays the ultra25 directory, C:\ULTRA25>.
Type uconfig and press Enter.

PRINTER SELECTION
Use the Up/Down arrows to Select Printer (backlight) and press Enter.
Use the Up/Down arrows to select the desire printer type and press Enter.

EDIT HOSPITAL NAME


Use the Up/Down arrows to select (backlight) Edit Hospital Name and press
Enter.
Type in the name of the hospital, institution or doctor that will appear on the
report and press Enter.

SELECT SCANNER
Use the Up/Down arrows to select (backlight) Select Scanner and press
Enter.
Select GE Scanner and press Enter.
Select the commmunications port used in the PC, Comm1 or Comm2. This
depends on where you connected the serial port cable to the PC.

EXIT TO DOS
Use the Up/Down arrows to select (backlight) Exit to DOS and press Enter.

The DMC presets are now stored.

NOTE: This is a limited user setup. See the DMC User Manual for
customer configuration.

11
GE MEDICAL SYSTEMS DMC - 3 INSTALLATION INSTRUCTIONS
REV 6 DIRECTION 46-030627

DMC - 3 Software Installation


Only (Supplemental)
NOTE: The DMC - 3 software program can only be operated with the
serialized Sentinel Key supplied in the software kit.. The
Sentinel Key must be installed on the printer parallel port (LPT1)
of the computer.

The DMC - 3 program disk can be used as many times as needed and can be
installed on other computers. However, the program will not run without the
Sentinel Key.

DOS Installation To install from DOS:

a. Enter the following instructions from the DOS command line prompt,
which is usually "C:>".

A: <Press Enter>
INSTALL <Press Enter>

and then follow the instructions on the computer screen.


OR
b. Reboot the computer with the installation diskette in the "A:" drive, and
follow the instructions on the menu that appears. (The computer must be
able to boot from the "A:" drive for this to work. Some computers are set
to boot from the hard drive first.)

Windows 3.x Installation To install from Windows 3.x:

From the pull down menu of the program manager:

1. Click on "File"
2. Click on "Run"
3. Enter A:INSTALL into the blank area
4. Click on "OK"
5. Follow the DMC installation instructions on the computer screen.

To establish an icon in Windows 3.x from the program manager for DMC-1 and
DMC 3:

1. Select the Program Group where the new icon should be


2. Click on "File"
3. Click on "New"
4. Click on "Program Item"
5. Enter DMC-3 for "Description"
6. Enter DMC3.BAT for command line
7. Enter C:\ for working directory
8. Click on "OK"
9. Click on "File"
10. Click on "New"
11. Click on "Program Item"
12. Enter DMC-1 for "Description"
13. Enter ULTRA.BAT for command line
14. Enter C:\ for working directory
15. Click on "OK"

12
GE MEDICAL SYSTEMS DMC - 3 INSTALLATION INSTRUCTIONS
REV 6 DIRECTION 46-030627

Windows 95 Installation To install from Windows 95:

From the Windows 95 desktop:

1. Click on "Start" button on the taskbar


2. Click on "Run" from the start menu
3. Enter A:INSTALL into the blank
4. Click on OK
5. Follow the DMC Installation Instructions

To establish an icon in Windows 95:

1. Right click on desktop


2. Click on "New" on the local menu
3. Click on "Shortcut"
4. Enter C:\ULTRA.BAT into blank
5. Click on "Next"
6. Type DMC1
7. Select "Finished"
8. Click on "Next"
9. Select any icon (MS DOS is default)
10. Right click on desktop
11. Click on "New" on local menu
12. Click on "Shortcut"
13. Enter C:\DMC3.BAT into blank
14. Click on "Next"
15. Type DMC3
16. Click on "Next"
17. Select any icon (MS DOS is default)
18. Select "Finished"

The following steps MUST be followed to make the DMC run "full screen" in
Win95:

19. Right click on the DMC1 or DMC3 icon that appears on the desktop
20. Click on "Properties" from the local menu
21. Click on "Screen" tab
22. Click on "Full Screen" under Usage
23. Click on "OK"
24. Click on "Program" tab
25. Click on "Advance", Click on "MS-DOS Mode"
26. Click on "Use Current MS-DOS Config", Click "OK"
27. Ensure that "Warn before entering DOS mode" is not selected
28. Repeat the process for the other newsly created icon.

13
GE MEDICAL SYSTEMS DMC - 3 INSTALLATION INSTRUCTIONS
REV 6 DIRECTION 46-030627

Functional Check Out (1) ____ The DMC - 3 program is started from the Applications menu in
Windows by double clicking the mouse on the DMC - 3 icon. (In
WIN 3.1 it can also be started from DOS by changing to the
ULTRA directory and typing ULTRA.)

(2) ____ To ensure complete and coherent data transfer from the Ultra-
sound system to the computer, enter the OB Report Page on the
Ultrasound system (see the User Manual for details) and the
main menu of the DMC - 3 program.

(3) ____ On the Ultrasound system OB Report Page, enter as much


patient information as possible. The patient ID is required for
data transfer. Activate transfer of the OB data by selecting
RECORD on the LOGIQ 400/500 (CONTROL R on the RT3200
Advantage or CONTROL P on the LOGIQ α200). Observe the
expected acknowledge message from the Ultrasound system
and the DMC - 3 program.

(4) ____ From the DMC - 3 program main menu, enter LIST PATIENTS.
Verify that the patient file transfer was completed.

(5) ____ Select ENTER/EDIT Patient Data in the DMC - 3 program. Use
the patient ID created on the Ultrasound system OB Report
Page.

(6) ____ Ensure that the DMC - 3 patient data is the same as the
Ultrasound system OB report page.

(7) ____ Press Escape to the return to the DMC - 3 main menu. Select
VIEW/PRINT patient data. Select and send the file to the printer.

The previous steps are an abbreviated functional test. Refer to Section 2 of


the DMC - 3 Operator Manual if a more detailed operational check is desired.

(8) ____ Fill out the proper customer information on the Product Locator
Installation Card. See Illustration 10.

(9) ____ Mail the Product Locator Installation Card to Milwaukee.

Working with DMC-1 and DMC-3 DMC-1 and DMC-3 can now co-exist on the same computer. No new data
should be entered into the DMC-1 program (ULTRA.BAT). The data in DMC-1
should be used for archival or referral purposes only.

All new data being transferred goes into the DMC-3 program
(c:\ultra25\ultra.exe).

If any old data is needed from previous studies, it must be retrieved from
DMC-1. This data can be used to compare against current DMC-3 data or
manually inputted as a new study into DMC-3.

14
GE MEDICAL SYSTEMS DMC - 3 INSTALLATION INSTRUCTIONS
REV 6 DIRECTION 46-030627

Illustration 10
Product Locator Installation Cards

DMC - 3 New Installation

Name and Address Date

DMC to DMC - 3
Upgrade Installation

Name and Address Date

DMC - 2 to DMC - 3
Upgrade Installation

Name and Address Date

15