You are on page 1of 137

GE Healthcare

DPX-NT and MD+ Service Manual


Date: 06/09/2009
Part No.: LU44246
Revision 1
ECO 2086331

Headquarters Asia/Pacific China


GE Medical Systems Lunar 4-7-127 Asahigaoka No. 19 Changjiang Road
3030 Ohmeda Drive Hino-shi, Tokyo 191-8503 Wuxi, Jiangsu, 214028
Madison, WI 53718-6704 Japan P.R.C.
USA +81-42-585-5111 +86-510-85225888
+1 (800) 437-1171 +81-42-585-3077 (Fax) +86-510-85226688 (Fax)

Europe Germany France


Beethoven Str. 239 GE Healthcare Lunar
D-42655 Solingen 11 Avenue Morane
Germany Saulnier
+49-212-2802-0 78457 Velizy
GE Medical Systems IT GmbH +49-212-2802-390 (Fax) France
Munzinger Strasse 3-5 +33-1-34-49-5365
D-79111 Freiburg +33-1-34-49-5406 (Fax)
Germany
+49-212-2802 -652
+49-761-4543 -233 (Fax)

www.gehealthcare.com

© 2009 by GE Healthcare
This page left blank intentionally.

Page 2 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


This document contains confidential or proprietary information of GE Healthcare.
Neither the document nor the information is to be reproduced, distributed, used
or disclosed, either in whole or in part, except as specifically authorized by GE
Healthcare.

GE Healthcare makes no warranty of any kind with regard to this material, and
shall not be held liable for errors contained herein or for incidental or
consequential damages in connection with the furnishings or use of this manual.

Read through this manual thoroughly before attempting to service any


components. Unauthorized service may void system warranties or service
contracts. Consult the GE Healthcare Customer Support Department prior to
attempting any servicing.

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 3 of 137


Page 4 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)
DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 5 of 137
Page 6 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)
DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 7 of 137
Page 8 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)
DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 9 of 137
Page 10 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)
READ THIS FIRST
Using This Manual

A person who will be performing service work on the DPX-NT / MD+ should use this
manual in the following manner:

Read the Safety and Overview Chapters to familiarize yourself with the scanner
as a whole and with the genera lfunction of the circuit boards.

Chapter 3 should be understood completely as it explains the Diagnostics


Software (built in – requires a password for access).

The Chapter 4 and Chapter 5 contain common procedures and troubleshooting


information and can be read as needed, but are good sources of information.

When a problem arises, Chapter 4 should be referenced. Check the table of


contents for Chapter 4 to see if the problem being experienced is described. If
so, refer to the appropriate page. If not, try to generalize the problem (e.g. the
Detector is repeatedly running into the front of the scanner and reversing and
then running back into the front of the scanner. This is a mechanical problem in
general, specifically with Transverse Mechanics, check that subsection of
Chapter 4 for the subsystem experiencing the fault.

This manual commonly references other Sections and pages of the manual as
needed, so often procedures in the Chapter 5 and the Appendix are referred to
as ways to solve problems described in Chapter 4.

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 11 of 137


This page left blank intentionally.

Page 12 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


Table of Contents

Chapter 1: Safety 15

1.0 General Safety 17


1.1 Symbols and labels found on the DPX-NT / MD+ 17
1.2 Emergency Stop Button 21
1.3 Laser Exposure 22
1.4 Shutter Indicator 23
1.5 Cautions, Warnings, and Notes 23
1.6 Safety Concerns 24
1.7 Scatter Radiation 25
1.8 Controlling Computer and Accessories 28
1.9 Peripheral configurations 28

Chapter 2: System Overview 31

2.0 DPX-NT / MD+ System 35


2.1 Electronics 36
2.2 DPX-NT / MD+ Block Diagrams 38
2.3 DPX-NT / MD+ Fusing 41
2.4 Combined Single Board Controller cSBC 41
2.5 X-ray Source 57
2.6 Display Panel 58
2.7 High Voltage Power Supplies 59
2.8 D-MAX Board (DPX-NT (A) ONLY) 59
2.9 XORB Board (DPX-NT (A) ONLY) 59
2.10 Detector Sub System 60
2.11 X-Ray Collimator Subsystem 60
2.12 DPX-NT / MD+ Specifications 61
2.13 Secondary Calibration / Daily QA 72

Chapter 3: Service Software 79

3.0 Diagnostic Software 81


3.1 The Tools Menu 82

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 13 of 137


3.2 Tools / Diagnostics Menu 83
3.3 Diagnostic Scan Modes 86
3.4 Error Log 86
3.5 Service Options 91

Chapter 4: Troubleshooting 93

4.0 Diagnostic Failure Codes 95


4.1 Transverse Motion failure 95
4.2 Longitudinal Motion failure 98
4.3 Failure of the DCpower supply 102
4.4 Emergency Stop Button 102
4.5 Tube Head Thermostat 103
4.6 Communication Error 103
4.7 Other Diagnostic Failure Codes 103
4.8 Failing Quality Assurance Test 104
4.9 Reference Counts 107
4.10 Arcing 110
4.11 Imaging Problems 112
4.12 Failing Alignment Test Results 114
4.13 Indicator Failures 116
4.14 Communications Failures 118
4.15 Viewing Quality Assurance Trends 118
4.16 MAX Board Troubleshooting 119
4.17 OMI Board Troubleshooting 120
4.18 SBC Troubleshooting 120
4.19 XORB Troubleshooting 121

Chapter 5: Service Procedures 123

5.0 Reloading LUNAR Software 125


5.1 Peaking the Detector 125
5.2 Tube Head Replacement 126
5.3 Lower Cable Bundle Replacement 127
5.4 Upper Cable Bundle Replacement 130
5.5 Tube Head Control Cable Replacement 132
5.6 Tests to Perform after Service 135

Appendix: Associated Service Documentation 137

Page 14 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


1 Safety

This chapter highlights safety devices and features a Service


Engineer should know before servicing a DPX-NT / MD+
system.

Chapter Contents:
1.0 General Safety
1.1 Symbols and labels found on the DPX-NT / MD+
1.1.1 Symbols
1.1.2 Labels
1.2 Emergency Stop Button
1.3 Laser Exposure
1.4 Shutter Indicator
1.5 Cautions, Warnings, and Notes
1.5.1 Caution Statements
1.5.2 Warning Statements
1.5.3 Note Statements
1.6 Safety Concerns
1.7 Scatter Radiation
1.8 Controlling Computer and Accessories
1.8.1 Electrical Safety
1.9 Peripheral configurations
1.9.1 Standard room configuration
1.9.2 Small room configuration

Figure 1-1. The DPX-NT / MD+ Display Panel


Figure 1-2. Laser Warning Label (All systems except Canada)
Figure 1-3. Laser Warning Label (Canadian Systems only)
Figure 1-4. Laser Warning Symbol (on display panel)
Figure 1-5. Source (x-rays) off - Shutter closed (green)
Figure 1-6. Source (x-rays) on - Shutter open (yellow)
Figure 1-7. Potential Pinch Points on the DPX-NT / MD+
Figure 1-8. DPX-NT Iso-Dose Diagram
Figure 1-9. DPX-MD+ Iso-Dose Diagram

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 15 of 137


This page left blank intentionally.

Page 16 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


1.0 General Safety
• DO NOT attempt to service the DPX-NT / MD+ without first reading this
manual.

• DO NOT attempt any repairs without prior instructions from authorized


LUNAR personnel.

• In order to maintain electrical safety and electromagnetic compatibility,


the Lunar DPX-NT / MD+ is only to be connected to a computer, printer,
and peripherals that meet IEC requirements for safety, such as IEC 950/
EN 60950 - “Safety of information technology equipment, including
electrical business equipment” and IEC 601-1-2 - “Medical electrical
equipment”, Part 1: General requirements for safety, 2. Collateral
Standard: Electromagnetic compatibility - Requirements and tests.

1.1 Symbols and labels found on the DPX-NT / MD+


1.1.1 Symbols
• The following symbols are found on the DPX-NT / MD+, in the Operators
manual, and in the Service Manual.

Attention: contains important safety information such as the


location of a pinch point.

Emergency Stop Button: shows the location of the


emergency stop button.

Power On: shows the location of the Power On indicator.

Laser On: shows the location of the Laser On indicator.

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 17 of 137


Shutter Open: shows the location of the Shutter Open indicator.

X-ray On: shows the location of the X-Ray On indicator.

Type B Equipment: shows that the scanner has Type B


protection against electrical shock.

Protective Earth: shows the location of a protective earth


terminal.

Functional Earth: shows the location of a functional earth


terminal.

1.1.2 Labels
• The following labels are found on the DPX-NT / MD+ Scanner.

Laser Caution Label: Shows that the scanner uses


a Class II laser.

Laser Caution Label: Canada only

Page 18 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


Tube Head Assembly Label (All
DPX-MD+ and DPX-NT System
Number 72000 and higher): This
label gives tube head assembly and x-
ray source characteristics information.
It is located on the tube head
assembly and the foot panel of the
scanner.

Tube Head Assembly Label (DPX-


NT System Number 71999 and
lower): This label gives tube head
assembly and x-ray source
characteristics information. It is
located on the tube head assembly
and the foot panel of the scanner.

A definition of each symbol on this label follows:

Inherent Tube X-ray Focal


Filtration Insert Source Point

System Label (DPX-MD+ DPX-NT


System Number 72000 and
higher): This label gives system
input power requirements and
compliance information. It is located
on the foot panel of scanners. The
CE mark shows compliance with
UL/CSA and the Medical Device
Directive 93/42/ EEC.

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 19 of 137


System Label (DPX-NT System
Number 71999 and lower): This label
gives system input power
requirements and compliance
information. It is located on the foot
panel of scanners. The CE mark
shows compliance with UL/CSA and
the Medical Device Directive
93/42/EEC.

High Voltage Power Supply: This


label gives high voltage power supply
(x-ray generator) information. It is
located on the high voltage power
supply(s), and foot panel of the
scanner.

X-ray Controller: This label shows x-


ray controller compliance. It is located
on the foot panel of the scanner.

Collimator Assembly: This label gives


collimator assembly information. It is
located on the collimator and foot panel
of the scanner.

Warning Label and Radiation Symbol: The


Warning label shows that the system uses
ionizing radiation. It is found only on systems
delivered in the United States. Always obey
instructions for safe operation.

Radiation Label: This label shows that the


system uses ionizing radiation.

Page 20 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


Grounding Reliability Label: This label states that
grounding reliability can only be maintained when using a
“Hospital Grade” or “Hospital Only” receptacle. It is only found
on systems delivered in the United States.

Note: When replacing a certified component (x-ray controller, x-ray


tube head, collimator or high voltage power supply(s)) also
install the duplicate certified component label (supplied with
the repalcement certified component) on the foot end panel of
the scanner. The label for the new part should be palced
directly over the laber for the part it is replacing.

Refer to DXSE3001 DPX-NT / MD+ Label Replacement


Procedure (Chapter 5Appendix I - this manual) for specific
instructions.

1.2 Emergency Stop Button


• The Emergency Stop Button is located on the front of the scan arm of the
DPX-NT / MD+ scanner (see Figure 1-1). When pressed, power is
removed from the X-ray tube head, the laser, and the shutter is closed.
Power is also removed from the scan arm motors, allowing the operator/
patient to push the scan arm out of the way.

Figure 1-1. The DPX-NT / MD+ Display Panel

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 21 of 137


1.3 Laser Exposure
• The DPX-NT / MD+ is equipped with a Class II Laser device. This laser is
used for patient positioning. A Class II rating indicates a low power visible
laser that is not normally hazardous to eyesight but has the potential to
be hazardous if viewed directly for an extended period of time. Because
of this potential hazard, DO NOT stare directly into the beam while the
laser is in operation, and DO NOT allow the beam to shine directly into
the patients' eyes. No specific eye protection is required with a Class II
laser.

• A amber laser-on indicator, located on the front of the scan arm, is lit
when the laser is on. The program activates the laser during positioning
for an image acquisition. The program then turns off the laser when you
begin the scan. The emergency stop button will turn off the laser.

• There is a caution label (Figure 1.2) on the scan arm near the Display
Panel.

Figure 1-2. Laser Warning Label (All systems except Canada)

Figure 1-3. Laser Warning Label (Canadian Systems only)

Figure 1-4. Laser Warning Symbol (on display panel)

Note: DO NOT STARE INTO THE BEAM while the laser is


operating.

Page 22 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


1.4 Shutter Indicator
• This symbol is used to indicate an open-shutter condition in accordance
with the safety standards established by the International
Electrotechnical Commission (IEC).

Figure 1-5. Source (x-rays) off - Shutter closed (green)

Figure 1-6. Source (x-rays) on - Shutter open (yellow)

• This symbol appears near the yellow X-ray shutter-open indicator light.
The X-ray shutter-open indicator light is located on the Display Panel on
the scan arm near the front.

Note: When the x-ray on / shutter open symbol appears in literature


associated with the DPX-NT / MD+ scanner, it will be used to
indicate that the procedure being described results in an open-
shutter condition. During these times personnel should
exercise caution to avoid excessive exposure to the X-rays.

1.5 Cautions, Warnings, and Notes


• This manual contains warning and caution statements wherever
appropriate for your safety. The warnings and cautions used throughout
the manual are based on the safety standards established by the
International Electrotechnical Commission (IEC). In addition, the manual
uses notes to attract the reader's attention to important information.

1.5.1 Caution Statements

A caution statement reflects a condition


that, if not avoided, could cause
equipment or property damage.

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 23 of 137


1.5.2 Warning Statements

A warning statement reflects a potentially


hazardous condition that, if not avoided, could
result in serious injury.

1.5.3 Note Statements


Note: This symbol turns the reader's attention to important
information which may otherwise be overlooked.

1.6 Safety Concerns


Because the DPX-NT / MD+ Densitometer
contains moving parts, there are places on the
scanner where there is a danger of pinching.
Operators should be aware of these pinch
points to avoid injury to the patient or
themselves. Labels applied at the LUNAR factory indicate the
location of the pinch points. The pinch points and their labels
are shown in the figure 1-7.

Figure 1-7. Potential Pinch Points on the DPX-NT / MD+

Page 24 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


DO NOT touch live components on the DC power supply - when
the cover is off of the supply some components (such as the
heat sinks) are at line voltage and present a shock hazard

1.7 Scatter Radiation


• There is some scatter radiation from the DPX-NT / MD+ when it is
running. Figure 1-8 shows the radiation dosages while the scanner is
running at 1.50 mA at certain distances.

• These dosages are relatively insignificant as the allowed yearly dosage


for a person working with radiation emitting equipment is 5000 mRem.
Radiation however should be avoided when possible.

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 25 of 137


Figure 1-8. DPX-NT Iso-Dose Diagram

Page 26 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


Figure 1-9. DPX-MD+ Iso-Dose Diagram

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 27 of 137


1.8 Controlling Computer and Accessories
1.8.1 Electrical Safety
• IEC and UL/CSA certification

Note: Not all scanners produced are built to IEC / UL / CSA


standards - IEC / UL / CSA compliant scanners bear the
appropriate mark on the foot end panel when the scanner was
produced and shipped in its compliant form

IEC: To maintain electrical safety, all computer equipment and accessories


connected to the scanner must meet IEC standards for safety, such as IEC
950, “Safety of information technology equipment, including electrical
business equipment,” and IEC 801-5, “EMC Surge Immunity Requirements.”
The computer and all accessories must have “CE” labels.

UL/CSA: To maintain electrical safety, all computer equipment and


accessories connected to the scanner must have saftey agency approvals fo
UL/CSA and comply with these standards.

See Operators manual for host computer / peripheral


configurations

1.9 Peripheral configurations

The correct connection of the computer and all


peripherals is necessary to maintain electrical safety.
The signal cable of the scanner is intended only for
connection to an approved computer. Call LUNAR
Support or your LUNAR distributor before adding
peripherals.

1.9.1 Standard room configuration

Page 28 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


The computer, peripherals, and all other equipment
must be located more than 1.83 m from the scanner.

Use an outlet strip to power the computer and all peripherals.

The outlet strip must be mounted off the floor so that it


does not touch other equipment.

0If the outlet strip was provided by LUNAR, it has a maximum output of 15A,
120VAC.

Only system-related equipment should be powered by the outlet strip.

A modem and/or network connection can be made at any time if you are using
the standard room configuration.

1.9.2 Small room configuration

The computer, peripherals, and all other equipment


must be powered with an isolating transformer if the
room is too small to maintain at least 1.83 m of
separation between the scanner and all other
equipment.

The isolation transformer supplied by LUNAR (part number 8544) has a


maximum output of 400/500VA.

Only system-related equipment should be powered by


the isolation transformer. Failure to use an isolation
transformer can cause leakage currents in excess of
100 microamperes.

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 29 of 137


A modem and/or network connection can only be
made in the small room configuration if all exposed
metal surfaces of the computer and peripherals are
out of the patient environment.

Page 30 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


2 System Overview

This chapter provides an overview of the DPX-NT / MD+ system.


• In addition the chapter contains a brief discussion of major
subsystems.
• This Chapter contains the DPX-NT / MD+ Block Diagrams

2.0 DPX-NT / MD+ System


2.0.1 DPX-NT / MD+ Electronics
2.1 Electronics
2.1.1 Electronics Pan
2.1.2 Power specifications
2.2 DPX-NT / MD+ Block Diagrams
2.2.1 DPX-NT (A) Systems (71999 and lower) Power Distribution
Block Diagram (AC entrance)
2.2.2 DPX-NT (A) System (71999 and lower) Block Diagram
2.2.3 DPX NT(B) / MD+ (systems 72000 and higher) Block Diagram
2.3 DPX-NT / MD+ Fusing
2.4 Combined Single Board Controller cSBC
2.4.1 Motion Control
2.4.2 Patient Positioners
2.4.3 X-ray Source Control
2.4.4 Shutter / Collimator Drive
2.4.5 End of Exposure Alarm
2.4.6 Panel LED's
2.4.7 HVPS Control
2.4.8 ADC
2.4.9 mA Low Range
2.4.10 Detector Interface
2.4.11 Comparators and Reference DAC's
2.4.12 Bias Control
2.4.13 Communications Ports
2.4.14 Diagnostic LED's
2.4.15 Laser Control
2.4.16 Power Distribution
2.5 X-ray Source

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 31 of 137


2.5.1 X-ray generation and Spectrum
2.6 Display Panel
2.7 High Voltage Power Supplies
2.8 D-MAX Board (DPX-NT (A) ONLY)
2.8.1 D-MAX Board Function
2.9 XORB Board (DPX-NT (A) ONLY)
2.10 Detector Sub System
2.10.1 Detector Overview
2.11 X-Ray Collimator Subsystem
2.12 DPX-NT / MD+ Specifications
2.12.1 Component specifications
2.12.2 Functional specifications
2.12.3 Maximum scan area (long x transverse)
2.12.4 Programs
2.12.5 Environmental specifications
2.12.6 Storage and transport environment
2.12.7 X-ray generator specifications
2.12.8 LUNAR 8548 and 8297 X-ray tube housing assemblies
2.12.9 Laser specifications
2.12.10 Compatible components
2.12.11 FDA certified components (US only)
2.13 Secondary Calibration / Daily QA
2.13.1 Secondary Calibration overview
2.13.2 Starting the Daily QA (secondary calibration)
2.13.3 Peak Test
2.13.4 Functional Tests
2.13.5 Reference Value Test
2.13.6 Tissue Value Test
2.13.7 BM Chamber Measurements
2.13.8 Daily QA Results
2.13.9 QA Database

Figure 2-10. DPX-NT (A) Systems (71999 and lower) Power Distribution
Block Diagram
Figure 2-11. DPX-NT (A) Systems (71999 and lower) System Block Diagram
Figure 2-13. DPX-NT / MD+ display panel
Figure 2-14. Reference axis and target angles for tube housing assembly
head assembly
Figure 2-15. Anode heating/cooling curves
Figure 2-16. Cathode emission characteristics

Page 32 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


Figure 2-17. X-ray tube assembly heating / cooling curves

Table 2-1. DPX-NT / MD+ FUSING


Table 2-2. Component specifications
Table 2-3. LUNAR 7681 X-ray generator technical information (system no.
72000 and higher)
Table 2-4. LUNAR 0311 / 0312 or 8531 / 8532 X-ray generator technical
information (NT (A) Systems numbered 71999 and lower)
Table 2-5. LUNAR 8022 X-ray tube technical information
Table 2-6. LUNAR 8548 x-ray tube head assembly (DPX-NT (B) and MD+
systems numbered 72000 and higher) technical information
Table 2-7. LUNAR 8297 x-ray tube housing assembly (DPX-NT (A) systems
71999 and lower) technical information
Table 2-8. Laser specifications
Table 2-9. FDA certified components (DPX-NT and MD+ Systems number
72000 and greater)
Table 2-10. FDA certified components (DPX-NT Systems number 71999 and
lower)

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 33 of 137


This page left blank intentionally.

Page 34 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


2.0 DPX-NT / MD+ System
The DPX-NT has been built as two slightly different revisions, the DPX- NT
(A), system numbers 71999 and lower, and the DPX-NT (B) and MD+ system
numbers are 72000 and greater.

• All of the differences between these systems (A and B / MD+) are in the
electronics pan. NT (A) systems are easily identifiable as they have 3
printed circuit boards in the electronics pan, where as the NT (B) and
MD+ have only 1.

• Most instructions in this manual apply to both the NT(A) and (B) / MD+.
Where differences apply - they will be noted as follows: NT(A) ONLY or
NT(B) / MD+ ONLY.

The DPX-NT / MD+ includes the patient table and frame, detector, and arm.
Its physical specifications are summarized in section 2.12.1

The DPX-NT (A) and (B) / MD+ have different X-ray generators, Tube
Housing assemblies and cabling. The physical and technical specifications for
each system are summarized in sections 2.7 and 2.12.7

The DPX-NT (A) and (B) / MD+ have a common mechanical design with two
separate motion systems that are capable of simultaneous operation. These
are transverse, and longitudinal. Both motion systems are driven by stepper
motors.

2.0.1 DPX-NT / MD+ Electronics

The internal components of the scanner are safely secured by a number of


panels, including the scanner's tabletop.

• The lower front and side panels are secured by locks.

• The rear panel is secured by screws from the outside.

• The table top is screwed down inside.

Note: Primary Service access to the electronics of the scanner is


through the table top.

• The Detector electronics (in the scan arm) are secured by an upper and
lower shroud, held in place by screws.

• Each metal panel is grounded to the electronics pan.

It is not usually necessary to remove the front and back panels for most
service needs. However, if access is needed to the Front and Rear
Longitudinal Carriages, these can be removed.

The back panel is secured by hex socket head-head screws and must be slid
out of the way, for it is between the Arm Column and the frame.

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 35 of 137


If access is needed to the detector, Transverse Limit Switches or the other
components mounted above in the arm, the covers of the arm must be
removed.

• The lower arm shroud is held in place by four screws, and must be
removed prior to removing the upper scan arm shroud

• The upper scan arm shroud can be removed by loosening the two
screws holding it in place (on the back of the arm column) and tipping it
forward.

2.1 Electronics
2.1.1 Electronics Pan

The electronic components of the DPX-NT / MD+ are mounted on the


grounded Electronics Pan which is horizontally fastened inside the frame.

• There is one switchable low voltage DC power supply (all outputs under
30VDC), and one (DPX-NT (B) / MD+) or two (DPX-NT (A)) high-voltage
DC power supplies (x-ray generator, to supply 76kV to the x-ray tube
housing assembly) on the pan.

• One high-voltage DC power supply (1000VDC) is located in the upper


arm near the X-ray detector and provides power to the Detector.

• In addition to the power supplies, the electronics mounting chassis holds


one (DPX-NT (B) / MD+) or three (DPX-NT (A)) printed circuit boards, a
stepper motor controller, and an AC entrance/line filter/fuse holder (see
appendix 2A for cSBC schematic, see appendix 2B for AC Entrance
wiring diagrams).

2.1.2 Power specifications

Leakage current

• Total System with Isolation Transformer: <100 microamperes.

The isolation transformer supplied by LUNAR has a


maximum output of 400/500VA. Only system-related
equipment should be powered by the isolation
transformer. Failure to use an isolation transformer
can cause leakage currents in excess of 100
microamperes

• Scanner Table alone: <100 microamperes.

Page 36 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


Scanner input power

• The scanner is capable of accepting AC inputs between 100 and 254


VAC. The scanner’s low voltage DC power supply automatically
configures itself for the voltage on site.

• Voltage may fluctuate ±10% from the nominal value without a loss of
scanner performance.

• The nominal input (range of inputs) can be found on the system label.

• The rated power input is 1500 VA.

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 37 of 137


2.2 DPX-NT / MD+ Block Diagrams
The block diagrams for the DPX-NT / MD+ system follow:

2.2.1 DPX-NT (A) Systems (71999 and lower) Power Distribution


Block Diagram (AC entrance)

Figure 2-10. DPX-NT (A) Systems (71999 and lower) Power Distribution Block Diagram

Page 38 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


2.2.2 DPX-NT (A) System (71999 and lower) Block Diagram

Figure 2-11. DPX-NT (A) Systems (71999 and lower) System Block Diagram

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 39 of 137


2.2.3 DPX NT(B) / MD+ (systems 72000 and higher) Block Diagram

Page 40 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


Figure 2-12. DPX-NT (B) and MD+ (systems numbered 72000 and greater)
Block and Power Diagram

2.3 DPX-NT / MD+ Fusing

Table 2-1. DPX-NT / MD+ FUSING

Fuse Rating Type


Condor PS
F3.15 AH 5x20 mm
F1
Condor PS
F3.15 AH 5x20 mm
F2
Condor PS
F3.15 AH 5x20 mm
F3
MAX PCB
F1 (NT (A)
Systems F0.5 AL 1/4 x 1 1/4 in
71999 and
lower only)

All Fuses are 250V

2.4 Combined Single Board Controller cSBC


The cSBC printed wiring board (PWB) is an eight layer rectangular board
measuring 7.400" x 8.100". The board thickness is specified as 0.093"
(standard is 0.062") to increase stiffness and stability for cable insertion. The
PWB is mounted in the via four mounting holes located 1/4" from each corner
and 2 additional interior mounting holes. The components are primarily
surface mount, with board connectors, headers and a few single-style ICs
being the only exceptions.

The cSBC employs an Intel 80C251 micro-controller as its processor. This


processor provides 1K of on-board RAM and no on-board ROM. The
controller is clocked at 16 MHz using a crystal.

cSBC Memory Space

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 41 of 137


The cSBC is designed to support a JEDEC-standard, non-volatile FLASH
memory device up to 512K x 8 bits in size for code and fixed data. The board
supports either 128K or 512K SRAM memory device as needed for program
volatile memory. Complete address decoding is provided via the MAX PLD,
the CBSC bus master, allowing the address space to be arbitrary and
changed via the PLD code. The 80251 can address four 64K segments,
referred to as 0x00, 0x01, 0xFE and 0xFF as per Intel literature. The firmware
has the ability to map any FLASH or SRAM segment to any CPU segment via
SFR’s in the MAX PLD.

FLASH RAM

At startup the CPU executes the boots code which programs the FLEX PLD
and then maps in either NT or Prodigy runtime firmware as appropriate based
on the most significant bit of the CCA REV register. To switch from boot code
to run code the firmware jumps to SRAM and executes a code snippet which
pages the boot code out of 0xFF and the desired firmware into 0xFF. The
snippet then jumps from SRAM back to 0xFF to execute the firmware.

The cSBC contains a single 128K x 8 bit SRAM which provides read/write
memory. The SRAM's segments are arbitrarily mapped to any CPU segment
by the CPU mapping registers.

cSBC Functions

The microprocessor-based Single Board Controller (cSBC) provides overall


operation and control of the scan table.

FLEX PLD Peripherals

The majority of scanner related programmable logic functionality is contained


in the FLEX PLD, an Altera EPF6024AQC208-3 device. On each cold boot
the CPU reprograms the FLEX devices from an image stored in it's FLASH.
As such a firmware download of a new FLEX image is required to
permanently upgrade the PLD code.

The functional components of the programmable logic are discussed in the


following subsections. Polarity of operation can be inferred from bit names
and use of preceding slash for inverted logic bits.

Page 42 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


Note that ports A-F are reloaded with default values at time of CPU reset and
remain in the default state until SCANNER_RESET has been cleared and
new values are written by the firmware. Defaults for port F and all other
registers are invoked at power up only.

PORT A

Bit Name R/ Def. Description


W

0 trans_enable R/W 0 Transverse motor enable – low blocks trans motor pulses and
forces Centent drive to standby current level.

1 /trans_fwd R/W 0 Transverse motor direction control.

2 /shutter_open_ctrl R/W 1 Shutter solenoid control.

3 trans_lsw_override R/W 0 Transverse limit switch override – prevent limit switch contact
from blocking step pulses at hardware level.

4 long_enable R/W 0 Longitudinal motor enable – low blocks trans motor pulses and
forces Centent drive to standby current level.

5 /long_fwd R/W 0 Longitudinal motor direction control.

6 long_lsw_override R/W 0 Longitudinal limit switch override – prevent limit switch contact
from blocking step pulses at hardware level
7 /collimator_open_ctrl R/W 1 Collimator solenoid control.

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 43 of 137


PORT B

Bit Name R/ Def. Description


W

0 /trans_front_lsw R N/A Transverse front limit switch position.

1 /trans_back_lsw R N/A Transverse back limit switch position.

2 /long_foot_lsw R N/A Longitudinal foot limit switch position.

3 /long_head_lsw R N/A Longitudinal head limit switch position.


4 trans_count_eq[0] R N/A Set when transverse step counter equals zero.

5 long_count_eq[0] R N/A Set when longitudinal step counter equals zero.


6 /shutter_open_sense R N/A Shutter limit switch position.

7 /collimator_open_sense R N/A Collimator limit switch position.

PORT C

Bit Name R/ Def. Description


W

0 /long_rev_pos R N/A Patient positioner (joystick) input.

1 /long_fwd_pos R N/A Patient positioner (joystick) input.

2 /trans_rev_pos R N/A Patient positioner (joystick) input.

3 /trans_fwd_pos R N/A Patient positioner (joystick) input.


4 /hvps_ac_relay R/W 1 Enable AC power to X-ray HVPS.
5 /motor_fail_enable R/W 1 Arm logic to shutdown scanner if OMI inputs not sensed.

6 ags_enable R/W 0 Enable detector automatic gain control feedback circuit.


7 /motor_power R/W 1 Enable 24VDC to the stepper motor drives (a.k.a. Centents).

Page 44 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


PORT D

Bit Name R/ Def. Description


W

0 flex_max_i/o_[0] R/W 0 Output signal to MAX PLD (diagnostic use only).


1 flex_max_i/o_[1] R/W 0 Output signal to MAX PLD (diagnostic use only).

2 flex_max_i/o_[2] R/W 0 Output signal to MAX PLD (diagnostic use only).

3 flex_max_i/o_[3] R/W 0 Output signal to MAX PLD (diagnostic use only).


4 flex_diag_3 R/W 0 Firmware controlled diagnostic LED.

5 pit_enable R/W 0 Enable Programmable Interval Timer output pulses.


6 flex_diag_1 R/W 1 Firmware controlled diagnostic LED.

7 /laser_on R/W 1 Patient locator laser control.

PORT E

Bit Name R/ Def. Description


W

0 low_range_dac R/W 0 Switches mA DAC from 2.048V to 0.500V reference.

1 trans_motor_accel R/W 0 Enables motor interrupt on every micro step.

2 low_range_adc R/W 0 Switches ADC from 5.000V to 0.500V reference.

3 long_motor_accel R/W 0 Enables motor interrupt on every micro step.

4 hvps_vendor_id R N/A For 7681 supply, 0 = Spellman, 1 = Bertan.

5 iq_hvps R N/A Set by resistor placement to indicate 0311/0312 supplies.

6 /hvps_enable_status R N/A Enable status monitor from 7681 supply.


7 /power_up R N/A Set to indicate cold boot.

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 45 of 137


PORT F

Bit Name R/ Def. Description


W

0 /motion_fail_enable R/W 1 Arm scanner shutdown if OMI pulses w/o step pulses.

1 long_motor_fail_axis R/W 0 Motor fail circuitry axis control, clear for transverse.

2 /hvps_enable R/W 1 Enable output to 7681 supply.

3 flex_diag_2 R/W 0 Firmware controlled diagnostic LED.


4 /arm_estop_sense R N/A Emergency stop sense bit.

5 spare_jmp_[1] R N/A Unused input, resistor or jumper selectable on CCA.


6 spare_jmp_[0] R N/A Unused input, resistor or jumper selectable on CCA.

7 cpu_p1_2 R N/A Firmware controlled diagnostic LED.

PORT G

Bit Name R/ Def. Description


W

0 adc_mux_[0] R/W 0 ADC analog MUX input selection control bit.

1 adc_mux_[1] R/W 0 ADC analog MUX input selection control bit.

2 adc_mux_[2] R/W 0 ADC analog MUX input selection control bit.

3 adc_mux_[3] R/W 0 ADC analog MUX input selection control bit.


4 adc_mux_enable R/W 0 ADC MUX output enable control.

5 8ms_clock R/W 0 Clock output provided to MAX PLD.

6 unused N/A N/A For expansion.

7 unused N/A N/A For expansion.

TRANS / LONG MOTOR

Dual axis stepper motor control is provided entirely by the FLEX PLD. The
drives provide 10 micro steps per full step. The firmware can track move
status by reading the 16 bit READ register.

As part of the setup for a move the host and/or firmware must enable the
motors via the /motor_power, trans_enable, and long_enable outputs and
setup the trans_lsw_override, long_lsw_override, /motion_fail_enable, /
motor_fail_enable, and long_motor_fail_axis outputs as desired. If the system
is in scanner reset for any reason the FLEX PLD will over-ride the /
motor_power output and prevent 24V power from reaching the motor drives.

AGS ROLL

Page 46 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


This is a read only 8 bit register which returns the count of AGS roll-over
events since the previous read of the register. The AGS roll counter is reset
on read only.

AGS DAC

This port provides R/W access to the AGS circuit's 8 bit U/D counter. The
DAC's analog voltage is tied to the gain control input of the variable gain
amplifier (VGA) used to control gain of the detector input signal. As such the
firmware can read this counter to determine the current DAC voltage level and
hence gain level. If ags_enable is low this port gives the firmware direct
control of the AGS DAC as a parallel R/W device. If ags_enable is high, the
firmware can write to the port but the DAC will continue to respond to UP/
DOWN requests from the AGS DCA circuitry and hence quickly return to the
AGS current operating voltage.

HE/LE COUNTERS

These read only ports provide access to the 16 bit event counters which are
incremented each time the DCA circuitry detects an input pulse within the HE
or LE windows (as defined by the LEL, LEH, HEL, and HEH DAC settings).
These counters are read in two 8 bit bus cycles, MSB then LSB. TPIT MSB/

SCANNER RESET

The scanner reset register is used to provide failsafe shutdown operation of


the scanner. A falling edge on any of the inputs to this register will latch the
current value of the register and drop the /SCAN_FAIL_ANY output. The MAX
PLD latches the master reset register and raises CPU_RESET in response to
the falling edge on /SCAN_FAIL_ANY. The MAX PLD also provides
SCANNER_RESET as the logical OR of CPU_RESET and!/
SCAN_FAIL_ANY. The FLEX PLD uses it's SCANNER_RESET input as the
enable bit to the tri-state buffers used to drive all safety critical output lines
including shutter control, HVPS relay control, motor relay control, etc. As such
the scanner is locked into a fail-safe mode whenever SCANNER_RESET is
asserted.

The MAX's master reset register will remain latched until the next rising edge
on the HOST_RTS input. When the cSBC is latched into reset by a scanner
error it will remain in CPU reset until the host drops the RTS line and re-
asserts it. It will remain in scanner reset until the CPU reads the scanner reset
register following the next raising edge of the RTS line at which the condition
causing the /SCAN_FAIL_ANY has been cleared. The firmware passes the
value of the reset registers to the host to allowing the host to display
appropriate error messages to the operator. The host will be unable to
perform any scanner related operations until the SCANNER_RESET has
been cleared. Red diagnostic LED's are provided for both scanner and CPU
reset lines (D19 and D20). The CPU reset line is tied to the host CTS output
such that the host sees a CTS event when the cSBC enters CPU reset. The
host code provides a CTS event handler which reads the reset registers and
prompts the user accordingly.

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 47 of 137


The firmware can also initiate a reset sequence in response to fatal error
conditions by writing a 'death code' to the suicide reset register. Resets can
also be initiated by the manual push button on the cSBC and by a low 5VDC
power condition as sensed by the MAX705 supervisor. The scanner register is
also latched at the end of read cycles such that current status can be
ascertained by a double read. A bit map of the scanner reset register is
provided below

Bit Name R/ Def. Description


W

0 / R N/A Tube head thermostat over temperature.


thermostat_open_sense

1 /external_estop_sense R N/A Emergency stop input from external options block.

2 /dc_power_fail R N/A Loss of one or more of +5VDC,+12VDC,-12VDC, or +24VDC.

3 /long_motor_fail R N/A Motor failure detected on longitudinal axis.

4 /trans_motor_fail R N/A Motor failure detected on transverse axis.

5 /dmb_error R N/A DMB dropped it’s CTS indicating a DMB reset event.
6 /motion_fail R N/A OMI pulses detected without step pulse (manual arm motion).

7 /watchdog_reset R N/A Watchdog time-out indicates firmware crash.

HVPS Errors

The HVPS error register is used to monitor the status of the LUNAR p/n 7681
(NT (B) and MD+ only) X-ray Generator HVPS. If the register value is not
equal to 0xF when /hvps_enable is low, the FLEX will raise the
HVPS_ERROR_INT output to the MAX PLD. The MAX PLD latches this into
the IIR register and issues an interrupt to the CPU. As such status of the X-ray
Generator is monitored when the unit is enabled. The handler for X-ray
Generator interrupt reads this register to determine the cause of the interrupt.
The X-ray Generator register is also latched at the end of read cycles such
that current status can be ascertained by a double read. A bit map of the
register is provided below.

Bit Name R/ Def. Description


W

0 /hvps_error_0 R N/A Error code bit from 7681 supply.

1 /hvps_error_1 R N/A Error code bit from 7681 supply.

2 /hvps_error_2 R N/A Error code bit from 7681 supply.


3 hvps_enable_status R N/A Set when /hvps_enable == /hvps_eanble_status

4 Unused N/A N/A Expansion room.


5 Unused N/A N/A Expansion room.

Page 48 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


6 Unused N/A N/A Expansion room.

7 Unused N/A N/A Expansion room.

DC FAIL

The DC fail error register latches the status of the DC power monitors at the
time of reset. If scanner reset code indicates /dc_power_fail the firmware can
read this register to identify the specific DC source failure. The register is also
latched at the end of read cycles such that current status can be ascertained
by a double read.

Bit Name R/ Def. Description


W

0 /plus_scanner_fail R N/A Loss of +24V power input.


1 /plus_analog_fail R N/A Loss of +12V power input.

2 /minus_analog_fail R N/A Loss of -12V power input.

3 Unused N/A N/A Expansion room.

4 Unused N/A N/A Expansion room.

5 Unused N/A N/A Expansion room.

6 Unused N/A N/A Expansion room.


7 Unused N/A N/A Expansion room.

DCA / AGS / BIAS DAC's

The cSBC uses a single 10 bit octal DAC, the Linear Technology LTC1660, to
generate the AGS and DAC window reference voltages and the bias program
voltage. KV/mA DAC

The cSBC uses a single 12 bit dual DAC, the Linear Technology LTC1454, to
generate the HVPS kV and mA program voltages.

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 49 of 137


ARC/FIL DAC

The cSBC uses a single 10 bit dual DAC, the Linear Technology LTC1661, to
generate the HVPS filament limit and arc detect threshold voltages.

PEAK DAC

The cSBC uses a 12 bit DAC, the Linear Technology LTC8043, to generate
the detector peak gain voltage.

MAX PLD Peripherals

The programmable logic section is based on an Altera MAX


EPM7128STC100-15 device. The MAX device is FLASH based (non-volatile)
and is programmed at the time of CCA assembly. The functional components
of the programmable logic are discussed in the following subsections.

Interrupts

The CPU's interrupt capacity is effectively increased by running several


interrupt signals to a register in the MAX PLD and tying the register output to
the CPU external INT 0 input. The firmware interrupt handler for INT 0 then
reads this register to identify the source of the interrupt and handles it
accordingly. The firmware then writes a bit masked '1' back to the IIR to clear
the bit of the interrupt it has serviced (the R/C in the table stands for READ/
CLEAR).

Bit Name R/ Def. Description


W

0 HOST_UART_INT R/C N/A Host UART interrupt.

1 DEBUG_UART_INT R/C N/A Debug UART interrupt.

2 DMB_UART_INT R/C N/A DMB UART interrupt.


3 HVPS_ERROR_INT R/C N/A HVPS error interrupt.

4 8mS_CLOCK R/C N/A 8ms clock tick interrupt from FLEX PLD.

5 POWER_FAIL_INT R/C N/A Power down pending in 5ms interrupt from DC supply.
6 Unused N/A N/A Expansion room.

7 Unused N/A N/A Expansion room.

Page 50 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


MASTER RESET

The master reset register will force a CPU and scanner reset condition on the
falling edge of any of its listed inputs. The contents of the register will be
latched at the time of reset such that when the CPU next comes out of reset
the firmware can read the register to determine what caused the preceding
reset and report the appropriate code to the host. If the reset was cause by
the CPU_RST_WR input, the suicide reset register contains the specific error
code. If the reset was cause by the /SCAN_FAIL_ANY input, the scanner
reset register contains the specific error code.

The CPU and scanner resets will remain latched until the next rising edge of
the RTS input. At this time the CPU reset will be cleared if /POWER_RESET
bit is not asserted and the scanner reset will be cleared if /SCAN_FAIL_ANY
is high.

Bit Name R/ Def. Description


W

0 /POWER_RESET R N/A MAX705 supervisor detects VCC < 4.65V.


1 /HOST_RTS R N/A RTS reset request from host via comm line.

2 /MANUAL_RESET R N/A Push button pressed.

3 Unused R N/A Expansion room, reads as ‘1’.


4 Unused R N/A Expansion room, reads as ‘1’.

5 Unused R N/A Expansion room, reads as ‘1’.

6 CPU_RST_WR R N/A Write to the suicide register, read suicide reg for error code.
7 /SCAN_FAIL_ANY R N/A Scanner reset register latched, read scanner reg for error code.

SUICIDE RESET

The CPU Reset SFR is a byte register into which the CPU can write a failure
code. In response to the write the MAX PLD will store the failure code and
assert the CPU_RESET line. The CPU_RESET line will be released on the
next rising edge of the host RTS, at which time the CPU will be able to read
the bit code from this SFR to determine the cause of the previous reset.

MISC OUT

The misc. output register is used to control the misc. output functions listed in
the following table.

Bit Name R/ Def. Description


W

0 Unused N/A N/A Expansion room.

1 MAX DIAG_2 R/W 0 Firmware controlled diagnostic LED.

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 51 of 137


2 DMB HWPT R/W 0 Enable direct connect of host and DMB XCVR’s, bypassing
UART’s, for maximized scan data bandwidth DMB to host.

3 RESET OVERRIDE R/W 1 Enable override of CPU_RESET signal. Set to 1 on power-up such
that firmware can load the FLEX PLD at power-up regardless of
the host RTS state.

4 Unused N/A N/A Expansion room.

5 Unused N/A N/A Expansion room.

6 Unused N/A N/A Expansion room.

7 Unused N/A N/A Expansion room.

MISC IN

The misc. output register is used to control the misc. input functions listed in
the following table

Bit Name R/ Def. Description


W

0 BOOT JUMPER R N/A JP4, placed to force firmware to remain in boot code.
1 CPU_P1_2 R N/A Input from CPU port 1, pin 2 (diagnostic use only).

2 Unused N/A N/A Expansion room.


3 Unused N/A N/A Expansion room.

4 Unused N/A N/A Expansion room.

5 Unused N/A N/A Expansion room.


6 Unused N/A N/A Expansion room.

7 Unused N/A N/A Expansion room.

Page 52 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


2.4.1 Motion Control

Stepper Motor Control

The stepper motors use the same interface design as used on previous
LUNAR products DPX-IQ and Prodigy. The solid state relay has been
replaced with a FET switch to save cost and board space. Diodes are placed
in series on the Centent power lines to prevent back EMF generated when the
arm is moved manually from reaching the 24V planes and damaging the
cSBC.

OMI Input

The optical motion interrupt (OMI) sub-system connects to the transverse and
longitudinal OMI CCA's. The OMI CCA's are located on the far end (the gear
end farthest from the motor) of each drive axis. When the axis is in motion a
small toothed wheel spins through the OMI opto's beam and pulses are
sensed back on the cSBC. In this manner the system can sense a drive
circuit, motor, or belt failure which might otherwise result is a concentrated
exposure point during a patient scan. As on Prodigy, logic in the FLEX PLD is
used to qualify the CH A and CH B inputs into a single 'valid motion' output.
Based on the phases of the square wave inputs on CH's A&B, the FLEX is
able to sense a change in direction. The circuit provides hysteresis to reject
false motion inputs resulting from scanner vibration when a wheel edge stops
in the center of the opto beam at the end of a move.

2.4.2 Patient Positioners

Four optically isolated inputs are provided for patient positioning. These are
used by the firmware to implement a joystick mode which is used in
conjunction with the laser to position the X-ray beam as desired over the
patient immediately prior to a scan.

Limit Switches

Four optically isolated inputs are provided for limit switches. These are used
by the firmware to define the transverse and longitudinal table limits.

2.4.3 X-ray Source Control

Mechanical Interlocks

The cSBC is designed such that a high on the scanner reset net disables all
scanner functionality and assures a know, fail safe, state. All scanner control
outputs are driven by the FLEX PLD. T. A scanner reset will force all these
FLEX scanner outputs to a high impedance state, de-energizing the opto's
and disabling the scanner.

The FLEX device is SRAM based and hence must be reprogrammed by the
CPU at power up. When the device is not programmed all I/O pins default to
the high impedance state. As such the scanner will also be in a fail safe state
when the FLEX is not programmed.

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 53 of 137


The cSBC provide a failsafe mechanism independent of the programmable
logic via the +5V_IO circuit. If one or more of either the E-stop, external E-
stop, or tube housing assembly thermostat is open, the FET driving the
+5V_IO net from the +5VDC plane will be disabled. The +5V_IO net provides
power to the emitter anode of all opto's which drive critical scanner functions.
As such the scanner will enter a failsafe state in response to these mechanical
interlocks, even in the event of a PLD device failure. The +5V_IO FET will also
be disabled by either a HOST_RTS or CPU_RESET.

2.4.4 Shutter / Collimator Drive

The shutter and collimator solenoid drive circuits are the same as that used on
Prodigy. The first FET is used for an initial 'hard hit' on open commands. It
presents 24V directly to the solenoid for several hundred msec's, resulting in a
large initial current pulse to the solenoid. The second FET provides the 'hold'
current through a pair of current limiting power resistors. The hold FET is tied
directly to the /shutter_open_ctrl bit.

2.4.5 End of Exposure Alarm

An on-board end of exposure alarm is provided. The alarm chosen is the


board mount equivalent of that used on IQ and Prodigy.

2.4.6 Panel LED's

The 4 panel LED's, power on, X-ray on, source exposed, and laser on, are all
driven PS2501-2 opto's through 750R0 / 1W current limiting resistors.

2.4.7 HVPS Control

A single, 16 channel multiplexed, 16 bit, high accuracy, ADC is used in


conjunction with several lower cost, lower bit resolution DAC's. Serial DAC's
and ADC's are chosen to conserve board space and simplify routing. Serial
parts are typically also lower cost as their maximum bandwidth is limited by
the serial baud rate.

A jumper and/or DNP'd resistor pad is provide to drive the IQ_HVPS line,
which the firmware reads to determine which HVPS it is intended to operate.
The PWB provides lemo style connector pads for use with the traditional
0311/0312 (DPX-NT (A)) X-ray Generator and D-MAX CCA. For the 7681
(DPX-NT (B) and MD+) the lemos are DNP'd and a single DB-25 connector is
used to control the X-ray Generator. An opto bank is also provided to support
the digital interface to the 7681 x-ray generator.

2.4.8 ADC

A single LTC1454 12 bit, serial, dual channel, DAC is used to provide the kV
and mA program voltages. The DAC is used in the x2 configuration such that
the full scale output is twice the reference voltage. Voltage outputs feed back
to the ADC MUX such that firmware can calibrate out DAC INL errors.

A single LTC1661 10 bit, serial, dual channel, DAC is used to provide the arc
threshold and filament current limit input voltages to the 7681 x-ray generator
(DPX-NT (B) and MD+).

Page 54 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


2.4.9 mA Low Range

To support the 50uA scan mode required by NT's thinner Cerium filter, an mA
low range circuit is provided. The FLEX PLD provides a control bit by which
the firmware can switch the mA DAC reference voltage from 2.048V to 0.5V,
hence decreasing the LSB size, hence allowing the firmware to take smaller
voltage steps when ramping to low uA settings. A second control bit is
provided to switch the ADC from 5.0 to 0.5V reference.

2.4.10 Detector Interface

The cSBC provides a single channel of detector electronics including analog


gain control, comparator circuitry and DCA/AGS logic within the FLEX PLD.

The cSBC is deigned to work with a traditional DPX IQ AMP 1890 CCA. The
bipolar input from the AMP is received via a lemo connector and sees a 50
Ohm line termination. The Linear Tech LT1228 variable gain amplifier is used
to provide peak and AGS gain control.

Gain Control DAC's

The gain control stage uses cascaded AGS and peak DAC's to generate the
gain control voltage to the LT1228.

2.4.11 Comparators and Reference DAC's

The cSBC uses the same CMP401 high speed comparators to convert the
analog bipolar signal to multiple digital threshold outputs.

Comparator reference voltages are generated by a single LTC1660 10 bit,


serial, octal DAC.

2.4.12 Bias Control

The 7th channel of the DCA octal DAC is used to generate the program
voltage for the PMT bias supply. The 0-3V output it mapped to 0-9V and
driven by the OPA2131 op-amp to the bias supply control cable. Bias program
voltage from the DAC and bias monitor voltage from the supply are both fed to
the ADC MUX such that the firmware has full control over the PMT bias
supply.

2.4.13 Communications Ports

Host RS-232

The host I/O port provides optical isolation per medical leakage requirements
of EN 60601-1-1, Annex BBB, section 7. QT's 6N136 opto's are used to
provide the required 115.2KB operational bandwidth and the required 2500
Vrms standoff. A DB-9 female connector is provided on the host side of the
isolation barrier. RS-232 is supported by populating the XCVR U26. R. All NT
/ MD+ scanners run at 19.2KB.

Debug RS-232 Port

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 55 of 137


The debug port is provided such that the firmware can echo status messages
and other information to a dumb terminal to support debugging, development,
testing, and servicing.

2.4.14 Diagnostic LED's

16 diagnostic LED's are provided by the cSBC. Functionality is listed below

LE Color Function Description (status indicated when lit)


D

D13 Green +24V +24V DC input from supply present.

D14 Green +12V +12V DC input from supply present.

D15 Green -12V -12V DC input from supply present.

D7 Green +5V +5V DC input from supply present.


D16 Red FLEX CONFIG FLEX PLD not programmed.

D20 Red CPU RESET CPU in reset mode.

D19 Red SCANNER RESET Scanner in reset (failsafe) mode.


D27 Red AGS ROLL AGS roll over or under detected.

D29 Amber FLEX DIAG 1 Diagnostic LED for misc use by firmware, control reg in FLEX.
D31 Amber FLEX DIAG 2 Diagnostic LED for misc use by firmware, control reg in FLEX.

D22 Amber MAX DIAG 2 Diagnostic LED for misc use by firmware, control reg in MAX.

D28 Amber TRANS OMI Valid transverse motion sensed by OMI circuitry.
D22 Amber LONG OMI Valid longitudinal motion sensed by OMI circuitry.

D25 Amber HE COUNT Valid high energy photon event sensed by DCA circuitry.

D23 Amber LE COUNT Valid low energy photon event sensed by DCA circuitry.
D21 Amber HVPS ENABLE 7681(NT (B) and MD+ ONLY) HVPS enabled.

2.4.15 Laser Control

The laser control circuit uses a PS2501-2 opto to control a FET which in turn
drives +5V_ANA to the patient locator laser.

Page 56 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


2.4.16 Power Distribution

The DPX NT system reduces scanner cost by using a single switching DC


power supply which provides +5,+/-12, and +24VDC. The supply is connected
directly to the cSBC via J14. J13 is provided as a power out connector to the
D-MAX CCA for NT-A. The PWB is divided into 4 ground plane regions -
DGND for +5V digital returns, AGND for +/-12V analog returns, SCNGND for
+24V returns, and ISO_GND for the host I/O island. The 24V plane contains
primarily power FET switches and power resistors used to control scanner
motors, X-ray source, LED, etc. The +/-12V planes contain analog electronics
used for detector and HVPS control. A +12V regulator is used to produce
12VDC for the Centent drives of off +24VDC, maintaining isolation from the
+12V plane. An additional +5V regulator is used to make 5V for the OMI's of
off +12V motors. 4 opto channels are used to sense the presence of the 4 DC
input voltages. The opto outputs are used to drive 4 green power on indicator
LED's (D7, D13, D14, D15) and to allow the CPU to detect power outs.

2.5 X-ray Source


The Tube Head consists of an oil-filled metal housing which contains a fixed-
anode X-ray Insert (essentially a vacuum tube housing assembly diode), lead-
type shielding, collimating devices, a filament transformer, electrical
connectors, and a Cerium (K-edge) filter.

The X-ray tube housing assembly gets its input from the high voltage power
supply(s) which provide a constant potential of up to 5 mA at 76kV and
operate at a high frequency of 50 kHz. This produces a stream of electrons in
the cathode and accelerates them towards the anode thus producing X-rays
as well as heat.

2.5.1 X-ray generation and Spectrum

• The X-ray Insert converts current into x-rays by accelerating the


electrons across a large potential from the cathode into a tungsten target
on the anode.

• The x-rays produced by the x-ray insert are filtered by the cerium filter to
produce a dual energy spectrum. The x-ray spectrum has peaks at 70
and 38 kV.

• Insert potential is provided by the High Voltage Power Supplies (see


section 2.6)

• The tube housing assembly current is set by the cSBC at 0.1 mA or 1.50
mA (DPX-NT) 0.1mA or 0.750 mA for DPX-MD+ depending on the
acquisition type.

• Tube Current is set by the cSBC and regulated by the D-MAX Board
(DPX-NT (A) NT (B) and MD+ units use the model 7861 X-ray generator
which includes all tube current regulation functions (see section 2.7).

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 57 of 137


2.6 Display Panel
• The Display Panel of the DPX-NT / MD+, located on the front face of the
upper arm, is the main operator interface outside the host computer.

Figure 2-13. DPX-NT / MD+ display panel

• Rocker switches on the panel allow operator control of the beam


position.

• Four indicators display system status.

• A green Light Emitting Diode (LED) indicates a power-on condition; it


lights when the +26VDC power supply comes on.

Power on LED is lit whenever scan table power is on.

• A yellow LED indicates if the laser is on

• An amber X-ray On LED indicates that x-rays are being produced


(current is flowing through the X-ray Insert), though exposure is not
necessarily taking place.

The X-ray on LED is lit when current is flowing through the X-


ray insert.

• The yellow Shutter Open LED comes on is the Shutter open and
exposure possible.

The Shutter open LED is controlled by the cSBC, the cSBC


will allow the shutter to open even if the lamp is not lit (shutter
open indicator is also on the PC controller screen).

Note: It is not recommended that scanner operation be continued if


These LED's signal the operator of exposure to x-rays and
must be replaced as soon as possible.

Page 58 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


2.7 High Voltage Power Supplies
The DPX-NT / MD+ X-ray production system combines a constant potential
generator (consisting of one (DPX-NT(B) and MD+) or two (DPX-NT (A)) high
voltage power supplies) and an X-ray tube housing assembly. The high
voltage power supply(s) (HVPS), operating at a high frequency of 50 kHz,
provide a continuous output of 38 kV each and up to 5 mA to the X-ray tube
housing assembly. The anode is composed of a tungsten alloy. The
specifications of the X-ray generation subsystems are outlined in section 2.x.

• The High Voltage Power Supply(s) provide the anode/cathode potential


to the X-ray insert.

• DPX-NT (A) scanners use two high-voltage power supplies (±40kVDC).


During normal operation, 76kVp is applied (+38kV at the anode and -
38kV at the cathode).

• DPX-NT (B) and MD+ scanners use a single high-voltage power supply
(40kVDC) During normal operation, 76kVp is applied to (+38kV at the
anode and -38kV at the cathode),

• The High Voltage Power Supply(s) are controlled by the cSBC (see
section 2.4).

• The power supply(s) are powered by the AC line voltage and have their
own built in fusing.

• The AC power is routed thru and the supplies are enabled by a solid
state relay(s) which is mounted on the electronics pan (see section
2.1.1).

2.8 D-MAX Board (DPX-NT (A) ONLY)


2.8.1 D-MAX Board Function

• The Tube Head and D-MAX Board work with the DC power supply to
supply tube housing assembly head current.

• The current is set by the cSBC.

• D-MAX board actually regulates current to the filament transformer in the


Tube Head.

• When errors are detected by the cSBC, (for example, loss of arm motion)
the relay will switch off and prevent production of x-rays.

2.9 XORB Board (DPX-NT (A) ONLY)

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 59 of 137


The XORB printed circuit board provides protection to various circuits (low
voltage) from transients within the HVPS and tube housing assembly head
(see 2.1.1 for location).

• There may be occasional static discharges within the Tube Head. The
transients caused by these static discharges (arcs) are shunted to
ground through the array of transorbs present on the XORB board.

• There is no safety hazard to personnel, however, the electronics could


be damaged were no protection provided.

2.10 Detector Sub System


2.10.1 Detector Overview

• The detector has a Sodium Iodide (NaI) crystal that serves as a


scintilator

• The light emitted by the scintilator is passed to the Photo Multiplier Tube
(PMT) which amplifies the signal

• The detector has its own dedicated High Voltage Power Supply, used for
supplying detector (PMT) bias.

• The detector bias is controlled by the cSBC, it is adjusted in the peak


setting test during the daily QA.

2.11 X-Ray Collimator Subsystem


• The Shutter is actuated by a solenoid using +24 VDC, the voltage is
reduced to a holding voltage of 12VDC after the shutter is opened.

• This solenoid is controlled by signals generated on the cSBC.

• A limit switch determines the shutter position and reports it to the cSBC.

• This assembly is located on top of the X-ray Tube Head.

Page 60 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


2.12 DPX-NT / MD+ Specifications
2.12.1 Component specifications

Table 3-2 gives specifications for standard components shipped with the
DPX-NT / MD+ system.

Table 2-2. Component specifications.

Component Specifications
Dimensions: 197.5 cm x 62.3 cm x 132.5 cm
Scanner table* Weight aprox 272kg
Maximum patient weight supported: 136kg
Console table 78.5 cm x 63.3 cm x 48.1 cm
Greater than 266 Mhz Pentium
64MB
Greater than 1GB hard disk
17” SVGA monitor (800x600x16-bit color) #
Computer
LS-120 Super Drive
CD ROM
NT operating system (with Service Pack 6.0 and Internet
Explorer 4.01 or greater with service pack 2.0)
HP DeskJet 930C
Printer
44.7 cm x 35.6 cm x 19.1 cm

*Width is measured from the front edge of the scanner table to the back edge of the
scanner arm. Height is measured from the top of the scanner arm to the bottom of the
scanner arm.

#Some languages may require 1024x768x16-bit color to fit the translated software text
on the screen

2.12.2 Functional specifications

General specifications

Focal spot to image receptor distance is 67 cm.

Attenuation equivalence of patient support table is 0.7 mm Al.

2.12.3 Maximum scan area (long x transverse)

• AP Spine Measurements
40.9 cm x 22cm

• Femur Measurements
20.9 cm x 17.9cm

• Total Body Measurements


195 cm x 60 cm measurement field

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 61 of 137


Forearm Measurements (for Investigational use only)
40.9 cm x 10 cm measurement field

Lateral Measurements (for Investigational use only)


40.9 cm x 10 cm measurement field

2.12.4 Programs

Note: Depending on the number of options purchased, not all of the


options listed below may be included with the DPX-NT / MD+
system software:

• AP spine measurement and analysis

• Femur measurement and analysis

• Total Body measurement and analysis

• Forearm Measurement and analysis

• Lateral Spine measurement and analysis

• Quality Assurance

2.12.5 Environmental specifications

Operational environment

Adhere to the specifications that follow during scanner operation:

• •Ambient Space (Interior Subcomponents)–For scanner operation


and servicing, do not block the area around the scanner table. Make sure
there is a minimum clearance of 30.5 cm at the head and foot ends of the
scanner table, at least 15.2 cm for the arm side, and 45.7 cm for the
operator side.

• Ambient Space (Ventilation)–Do not block the cooling vents on the


computer and scanner table. Make sure there is 15.2 cm from the
console table to the wall for cable clearance and computer plugs.

• Dust, Fumes and Debris–Install the system in a clean, ventilated area.


Dust and other airborne debris can cause the diskette drive heads and
other sensitive mechanical components to malfunction. LUNAR
recommends that smoking is not permitted in the scanner room.

• Humidity–Make sure the humidity for the scanner area is 20%–80%,


non-condensing.

• Static Electricity–Install and operate the system in a static-free area.


Adhere to minimum humidity requirements to prevent malfunctions
caused by static electricity.

Page 62 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


• Shock and Vibration–Make sure the scanner table does not receive
shock greater than 1 G for more than 1 millisecond. Make sure the
scanner table does not receive vibrations greater than 0.25 G at 5 Hz.

• Temperature–Make sure the temperature during system operation is


65°F–81°F (18°C–27°C).

Note: When the system is turned off, or there is a power failure,


allow the system to be on and let it warm for one hour. After
one hour, complete a Quality Assurance procedure.

2.12.6 Storage and transport environment

Adhere to the specifications that follow for scanner storage and


transportation:

• Humidity, 0% to 95% non-condensing.

• Atmospheric pressure, 500 to 1060 hPa.

• Temperature, -40° to 70° C.

2.12.7 X-ray generator specifications

X-ray generator

Table 2-3 contains information about the x-ray generator for DPX-NT (B) and
MD+ series devices and the standard referenced. Table 2-4 contains
information DPX-NT (A) systems.

Table 2-3. LUNAR 7681 X-ray generator technical information (system no. 72000 and higher)
Classification Class I Equipment IEC 601-2-7 5.1
Degree of protection against electrical
Type B equipment IEC 601-2-7 5.2
shock
Ordinary medical electrical equip-
Protection against ingress of liquids IEC 601-2-7 5.3
ment
Connection to supply mains Power supply cord IEC 601-2-7 6.1g)
Mode of operation Continuous IEC 601-2-7 6.1m)
Maximum X-ray tube voltage 76 kV IEC 601-2-7 6.1m)
Maximum X-ray tube current 3 mA IEC 601-2-7 6.1m)
Rated mains voltage 100-240 VAC IEC 601-2-7 6.1j)1
Number of phases in mains 1 IEC 601-2-7 6.1j)2
Mains frequency 50/60 Hertz IEC 601-2-7 6.1j)3
Required over-current releases 15 Amp dedicated service IEC 601-2-7 6.1j)5

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 63 of 137


X-ray tube dissipates 243W max.
into surrounding air through
Heat dissipative components forced air convection. Flow rate: IEC 601-2-7 6.1t)
36 m3/h (approx.) Temp. rise of air
stream 25° C (approx.)
Spellman SBD40PN280X2890 or IEC 601-2-7 6.8.1
Allowable high voltage supplies
Bertan 2907. and 50.2.101-102
IEC 601-2-7 6.8.1
Allowable tube head assemblies LUNAR model 8548 or equivalent
and 50.2.101-102
Original language of accompanying docu-
English IEC 601-2-7 6.8.1
ments
Maximum continuous kV, mA at nominal IEC 601-2-7 6.8.2
76 kV, 3 mA
rated kV 1)
Maximum intermittent kV, mA at nominal IEC 601-2-7 6.8.2
76 kV, 3 mA
rated kV 1)
Maximum continuous kV, mA at maximum IEC 601-2-7 6.8.2
76 kV, 3 mA
mA 2)
Maximum intermittent kV, mA at maxi- IEC 601-2-7 6.8.2
76 kV, 3 mA
mum mA 2)
Continuous kV, mA for maximum electric IEC 601-2-7 6.8.2
76 kV, 3 mA
output power 3)
Intermittent kV, mA for maximum electric IEC 601-2-7 6.8.2
76 kV, 3 mA
output power 3)
IEC 601-2-7 6.8.2
Nominal electric power 0.243 kW
4)
0.20 mAs. Parameters: 76 kV,
IEC 601-2-7 6.8.2
Lowest current time product 0.10 mA,
5)
2 seconds.
IEC 601-2-7 6.8.2
Nominal shortest irradiation times 2 seconds.
6)
Voltage divider in high voltage
Method of x-ray tube voltage measure- IEC 601-2-7
power
ment 50.106.1
supply.
Shunt resistor in high voltage sup-
Method of x-ray tube current measure- IEC 601-2-7
ply
ment 50.106.2
return line.
Line normal to the tube port, cen-
IEC 601-2-7
X-ray tube assembly reference axis tered
6.1j)5) 50.107.1
on tube port as shown in Figure 4.

Page 64 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


8.21 x 105 Joules, 3 mA, 76 kV for IEC 601-1-3
Reference loading conditions
1 hour. 29.204.2
IEC 601-1-3
Focal spot to Image Receptor distance 57 cm
29.203.2
Attenuation equivalence of patient support IEC 601-1-3
0.7 mm Al
table. 29.206.2

Table 2-4. LUNAR 0311 / 0312 or 8531 / 8532 X-ray generator technical information (NT (A)
Systems numbered 71999 and lower).
Classification Class I Equipment IEC 601-2-7 5.1
Degree of protection against electrical
Type B equipment IEC 601-2-7 5.2
shock
Ordinary medical electrical equip-
Protection against ingress of liquids IEC 601-2-7 5.3
ment
Connection to supply mains Power supply cord IEC 601-2-7 6.1g)
Mode of operation Continuous IEC 601-2-7 6.1m)
Maximum X-ray tube housing assembly
76 kV IEC 601-2-7 6.1m)
voltage
Maximum X-ray tube housing assembly
5 mA IEC 601-2-7 6.1m)
current
Rated mains voltage 100, 115, 230, 240 volts IEC 601-2-7 6.1j)1
Number of phases in mains 1 IEC 601-2-7 6.1j)2
Mains frequency 50/60 Hertz IEC 601-2-7 6.1j)3
Required over-current releases 20 Amp dedicated service IEC 601-2-7 6.1j)5
X-ray tube housing assembly dissi-
pates 305W max. into
surrounding air through forced air
Heat dissipative components IEC 601-2-7 6.1t)
convection. Flow rate: 36 m3/h
(approx.) Temp. rise of air stream 25º
C (approx.)
Spellman X2112/X2113/ rev. K and
higher. Bertan 2411P and 2411N rev. IEC 601-2-7 6.8.1
Allowable high voltage supplies
A and higher. LUNAR p/n 0311 and and 50.2.101-102
0312.
Allowable tube housing assembly head IEC 601-2-7 6.8.1
LUNAR model 8297or equivalent
assemblies and 50.2.101-102
Original language of accompanying docu-
English IEC 601-2-7 6.8.1
ments
Maximum continuous kV, mA at nominal
76 kV, 4 mA IEC 601-2-7 6.8.2 1)
rated kV
Maximum intermittent kV, mA at maximum
76 kV, 5 mA IEC 601-2-7 6.8.2 1)
kV

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 65 of 137


Table 2-4. LUNAR 0311 / 0312 or 8531 / 8532 X-ray generator technical information (NT (A)
Systems numbered 71999 and lower). (continued)
Maximum continuous kV, mA at maximum
61 kV, 5 mA IEC 601-2-7 6.8.2 2)
mA
Maximum intermittent kV, mA at maximum
76 kV, 5 mA IEC 601-2-7 6.8.2 2)
mA
Continuous kV, mA for maximum electric
76 kV, 4 mA IEC 601-2-7 6.8.2 3)
output power
Intermittent kV, mA for maximum electric
76 kV, 5 mA IEC 601-2-7 6.8.2 3)
output power
Nominal electric power 0.4 kW IEC 601-2-7 6.8.2 4)
7.89 mAs. Parameters: 76 kV, 2.63
Reference current time product mA, IEC 601-2-7 6.8.2 5)
3 seconds.
Nominal shortest irradiation times 3 seconds. IEC 601-2-7 6.8.2 8)
No specific wait period was imposed.
IEC 601-2-7
Repetition rate for loading during tests Time between tests was approxi-
50.104.4
mately 20 seconds.
Method of x-ray tube housing assembly voltage Voltage divider in high voltage power IEC 601-2-7
measurement supply. 50.106.1
Method of x-ray tube housing assembly current Shunt resistor in high voltage supply IEC 601-2-7
measurement return line. 50.106.2
Line normal to the tube housing
IEC 601-2-7 6.1j)5)
X-ray tube housing assembly reference axis assembly port, centered on tube housing
50.107.1
assembly port as shown in figure 3-14.
Reference loading conditions 1.09 x 106 Joules, 4 mA, 76 kV for 1 hour.IEC 601-1-3
Leakage radiation was measured at the
3mA, 76 kV IEC 601-1-3
following loading factors.
Focal spot to Image Receptor distance 57 cm IEC 601-1-3
Attenuation equivalence of patient support
0.7 mm Al IEC 601-1-3
table.

Page 66 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


Figure 2-14. Reference axis and target angles for tube housing assembly
head assembly

2.12.8 LUNAR 8548 and 8297 X-ray tube housing assemblies

• Beam filtration is permanently fixed with a minimum 3.0 mm Aluminum-


equivalent. Refer to table 2-5 for model 8538 x-ray tube assembly or
table 2-6 for model 8297 x-ray tube assembly.

Note: Beam quality has a minimum first half-value layer of 4.7 mm of


Al at 76 kV

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 67 of 137


Figure 2-15. Anode heating/cooling curves

Table 2-5. LUNAR 8022 X-ray tube technical information.


Nominal anode input power 361 Watts IEC 613/1989
Maximum anode heat content 6000 Joules IEC 613/1989
Anode heating and cooling curves Refer to figure 5. IEC 613/1989
Anode target material Tungsten IEC 601-2-28
Reference axis Refer to figure 4. IEC 601-2-28
78° (reference to nor-
Target angle IEC 601-2-28
mal)
Nominal focal spot values 0.5 IEC 336/1982
Not Applica-
Maximum useful voltage 95 kVp
ble
Not Applica-
Maximum filament current 2.2 Amperes
ble

Table 2-6. LUNAR 8548 x-ray tube head assembly (DPX-NT (B) and MD+ systems numbered 72000
and higher) technical information.
IEC 522/
Inherent filtration >3.0 mm Al/70 kV
1976
IEC 613/
Filament characteristics Refer to Figure 6.
1989
76 kV - Anode to Cathode
IEC 613/
Nominal x-ray tube voltage 38 kV - Anode to Earth
1989
38 kV - Cathode to Earth

Page 68 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


IEC 613/
Single load rating 228 W (3 mA, 76 kV) for up to 15 min.
1989
228 W (3 mA, 76 kV) for up to 15 min.
with a IEC 613/
Serial load rating
5 min. cool down time between mea- 1989
surements.
Maximum x-ray tube assembly heat con- IEC 613/
260 kJoules
tent 1989
X-ray tube assembly heating and cooling IEC 613/
Refer to Figure 7.
curves 1989
243 Watts (3mA x 76kV + 15W fila- IEC 613/
Maximum continuous heat dissipation
ment) 1989
IEC 806/
Maximum symmetrical radiation field Diameter = 10 mm
1984
IEC 601-2-
Dimensions 17 cm x 19.4 cm x 11 cm
28
IEC 601-2-
Weight 8.6 kg
28

Table 2-7. LUNAR 8297 x-ray tube housing assembly (DPX-NT (A) systems 71999 and lower)
technical information.
IEC 522/
Inherent filtration >3.0 mm Al/70 kV
1976
IEC 613/
Filament characteristics Refer to Figure 6.
1989
76 kV - Anode to Cathode
IEC 613/
Nominal x-ray tube voltage 38 kV - Anode to Earth
1989
38 kV - Cathode to Earth
361 W (4.75 mA, 76 kV) for up to 4
IEC 613/
Single load rating min.,
1989
59 sec.
361 W (4.75 mA, 76 kV) for up to 4
min., IEC 613/
Serial load rating
59 sec. with a 10 min. cool down time 1989
between measurements.
Maximum x-ray tube assembly heat con- IEC 613/
260 kJoules
tent 1989
X-ray tube assembly heating and cooling IEC 613/
Refer to Figure 7.
curves 1989

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 69 of 137


IEC 613/
Maximum continuous heat dissipation 361 Watts
1989
IEC 806/
Maximum symmetrical radiation field Diameter = 10 mm
1984
IEC 601-2-
Dimensions 17 cm x 19.4 cm x 11 cm
28
IEC 601-2-
Weight 8.6 kg
28

Figure 2-16. Cathode emission characteristics

Figure 2-17. X-ray tube assembly heating / cooling curves

Page 70 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


2.12.9 Laser specifications

Table 3-8 gives the specifications for the LUNAR laser.

Table 2-8. Laser specifications.


Output Power <1mW
Wavelength 635nm
Beam Diameter at
4x1mm
aperture
Aspect Ratio 4.0 to 1
Divergence 24 degrees
Radiant Exposure 0.0001 W
Integrated Radiance 46 W
Current Draw 105 mA
Voltage Input 4-6 VDC
Safety Rating Class II

2.12.10 Compatible components

• For customers located internationally, make sure the computer is


certified to local requirements such as IEC 950. The computer must meet
the minimum requirements that follow:

• Greater than 266 MHz Pentium

• 64 MB RAM

• Greater than 1GB Hard Disk

• Fast serial I/O board (Lunar Part Number 7151)

• 8X CD ROM

• 14” SVGA monitor with at least 800x600x16-bit color (some languages


may require 1024x768 resolution)

• NT operating system (with service pack 6.0 and Internet Explorer 4.01
with Service Pack 2)

• Disk Defragmentation Software

2.12.11 FDA certified components (US only)

Table 2-6 gives components certified to the FDA for use with DPX-NT / MD+
scanners and is updated periodically. Contact LUNAR for a current listing of
compatible components.

Table 2-9. FDA certified components (DPX-NT and MD+ Systems number 72000 and greater)

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 71 of 137


LUNAR
COMPONENT DESCRIPTION
MODEL #
LUNAR DPX-NT (B) / MD+ single
X-ray Controller 7634
board controller

Bertan1 Model: 2907


High Voltage Power 7681
Spellman2 Model:
Supplies 7681
SBD40PN280X2890
Tube Head Assembly LUNAR X-Ray Tube Head Assembly 8548
Collimator DEXA Collimator Assembly 7767
1
Bertan High Voltage Corp., 121 New South Road, Hicksville, NY
2
Spellman High Voltage Electronics Corp., 475 Wireless Boulevard, Hauppauge,
NY

Table 2-10. FDA certified components (DPX-NT Systems number 71999 and lower)

LUNAR Model
Component Description
#
LUNAR DPX-NT single board
X-ray Controller 7844
controller

Bertan1 Models:
High Voltage Power
2411 N 0311 or 8531
Supplies
2411 P 0312 or 8532

Tube Head Assembly LUNAR X-Ray Tube Head Assembly 8297


Collimator DEXA Collimator Assembly 2898
1
Bertan Associates, 121 New South Road, Hicksville, NY

2.13 Secondary Calibration / Daily QA


2.13.1 Secondary Calibration overview

• Daily QA (Quality Assurance) tests the integrity of the scanner so


operator knows it is operating within specifications.

• Daily QA (Secondary Calibration) adjusts the calibration of the scanner to


offset any changes due to the tube housing assembly aging, temperature
varying, or when certain components are changed.

Page 72 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


• A third type of function, QC Phantom, is another method for testing
system performance. A QC phantom is scanned using standard software
and results are stored (aluminum spine in lucite).

• Daily QA ensures the system is operating properly by running 3 types of


tests:

• A peak test to adjust photon counting electronics.

• A functional test that checks the indicator lights, motion system,


shutter, tube housing assembly output, and detector performance.

• A test that measures a secondary calibration object with different


bone chambers and tissue plugs which correct for system aging and
prevents long-term drifts.

2.13.2 Starting the Daily QA (secondary calibration)

Operator presses [F4] to start [Daily QA] which presents a plot of the last used
parameter from the last Daily QA run. (Default display is the medium bone
chamber.) At this point operator can interactively manipulate the data in QA
database to create tables and graphs that can be printed.

Tech presses [Start] to start running Daily QA tests.

Software displays instructions to position calibration block on the table. Tech


places block on tabletop with label up - aligned with laser then click [OK] to
start.

The UI locks out all other operations except Stop button.

Scan arm moves to the "Home" position. In this case, Home is always at head
of scanner for QA independent of user setting for Home.

The scanner performs a scout scan to find the block. If block is not properly
positioned, the scanner will prompt the user to Reposition standard more
accurately.

The software automatically finds landmarks within the secondary phantom.


This reduces the need for exact positioning on the table top.

All tests are run all of the time. The tests are executed and test results
reported as each test completes. Test results include pass/fail and any
quantitative information when appropriate.

2.13.3 Peak Test

This test adjusts the sensitivity of the photon counting electronics. The test
determines the optimal voltage setting for the detector amplifier so the
maximum number of photons are detected.

During this test, the shutter opens with the x-rays on.

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 73 of 137


The system acquires samples sweeping over a voltage range in steps of 50
units for a full range of 500 units which is centered on the last peak setting.

The detector peak test then determines the optimal voltage setting for the
detector amplifier based on these curves. The actual peak setting is taken
from the high energy count rates, the low energy channel is essentially
blocked by the brass spillover piece in the QA standard.

2.13.4 Functional Tests

The next set of functional tests consist of a shutter test and motion tests for
transverse and longitudinal directions.

Beam Stop test - Measures the movement of the source shutter. A functional
test checks the shutter LSW and ensures the shutter stops all photons.

Motor motion and limit switches - Each motor (transverse, longitudinal) will
be run between hard limits and back to home position. If the hard limits do not
engage, this is a failure. Also compare steps against predetermined scan
window size.

Spillover test - Measures the effect of high energy photons registering as low
energy photons in the detector system. The beam passes through a brass
piece that stops all low energy photons. The detector then measures the
number of low energy photons detected as a measure of spillover. Spillover
should be less than 13%. Spillover with the CZT detector will degrade about
0.2% per year.

Spillover Stability test - As part of the secondary calibration, the change in


spillover must be calculated on a per element basis and smoothed over the
last 5 passing QA's. Daily QA measures actual spillover and stores it in QA
database. As part of primary calibration, the spillover value used during that
calibration is stored. The delta spillover calculation using the spillover from
primary calibration and the running average of the last five spillover
measurements. This information is in the scan file.

2.13.5 Reference Value Test

The output of the x-ray tube housing assembly is measured as part of a


Reference Value (old term was air) measurement. This is measured at 1.5
mA.

Reference Value measurement - The HE/LE counts are measured as the


number of photons counted by detector at 1.5 ma. This value is used in all
bone density calculations. It must be stored in the systems data file.

• Reference value measurements will be performed at 150uA and 3000uA.


These reference values will need to be stored separately for all detector
elements.

• Reference Value measurements for 1.5mA are actually made through


lucite sections of the secondary calibration block.

Page 74 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


Reference value ratio test - The HE/LE ratio for each reference value is
calculated for each tube housing assembly current tested in reference value
measurement.

2.13.6 Tissue Value Test

Tissue secondary calibration

• The system measures two tissue plugs (lucite and Acetron).

• Store measured values for each chamber in QA database.

• The system determines the%fat for two different materials with different
compositions.

• Three%fat values are displayed - lean, mid and fat

• These calculated%fat values are compared against expected values and


checked against limits for the mean, standard deviation.

• A slope is calculated and this adjustment is used as a secondary%fat


calibration to offset long term drifts in system performance. This
adjustment is performed at the end of the%fat calculation.

2.13.7 BM Chamber Measurements

The system acquires point measurements on the 3 bone chambers 10 times


using the 1.5 mA medium scan mode. An image is created but not displayed.
Using simple edge detection algorithm, place 3 ROI's and calculate
measurements. The system determines the bone mineral BM and width for
three chamber with different sizes and densities.

• Measured values for each chamber are stored in the QA database.

• These calculated BM values are compared against expected values and


checked against limits for the mean, standard deviation.

• A slope is calculated and this adjustment is used as a secondary BMD


calibration to offset long term drifts in system performance.

2.13.8 Daily QA Results

At the end of Daily QA, a report is automatically printed. If printer is not on line
or fails, post an error to screen.

If Daily QA fails, post an error message that says "Unsuccessful QA -


Recommend repeating procedure."

• If a QA fails the error log will contain detailed information on the


failure.

Test results are saved automatically to a database.

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 75 of 137


Appropriate secondary calibration and Reference Value information is
recalculated from QA database using an average of 5 previous passing test
results. This information is stored in each scan file.

A view with plot of medium bone chamber is generated, and user the user is
allowed to manipulate data interactively.

2.13.9 QA Database

A historical record of system performance is maintained by storing key QA


parameters in a database.

• High (1.5mA) Results - fat is hidden - only bone results are viewable

Page 76 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


This page left blank intentionally.

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 77 of 137


Page 78 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)
3 Service Software

This chapter contains a discussion of the Service Software


and the Error Log.

3.0 Diagnostic Software


3.0.1 To Access the Service Software:
3.1 The Tools Menu
3.1.1 Spectrum
3.1.2 Stability Run
3.1.3 Signal Monitor
3.1.4 Primary Calibration
3.1.5 Secondary Verification
3.1.6 Pileup
3.1.7 Set Download Parameters
3.1.8 enCORE Composer
3.1.9 Error Log
3.1.10 Copy Configuration
3.1.11 User Options
3.1.12 Service Options
3.1.13 Outbox
3.1.14 System File Editor
3.2 Tools / Diagnostics Menu
3.2.1 Pileup
3.2.2 Calibration Pileup
3.2.3 Spillover
3.2.4 Count Rate
3.2.5 Scanner Motion
3.2.6 Scanner Detector
3.2.7 Scanner X-ray
3.2.8 Lin/Rep
3.2.9 Limit Switch Adjustment
3.2.10 Scanner Disconnect
3.3 Diagnostic Scan Modes
3.3.1 Table Top Scan
3.3.2 Alignment Scan

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 79 of 137


3.3.3 Beam Wobble scan
3.3.4 Hacksaw
3.4 Error Log
3.4.1 Printing the Error Log
3.4.2 Troubleshooting Help software
3.4.3 Reading the Error Log
3.5 Service Options
3.5.1 Measure Tab
3.5.2 Analyze Tab
3.5.3 QA Tab
3.5.4 Serial Tab

Figure 3-1. The Tools/Diagnostics Menu


Figure 3-2. The Error Log Dialogue Box
Figure 3-3. Error log entry details
Figure 3-4. enCORE Troubleshooting help screen capture
Figure 3-5. Sample DPX-NT / MD+ Error Log

Page 80 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


3.0 Diagnostic Software
Note: LUNAR Diagnostic software options are not available to the
customer. This software is available to LUNAR authorized
representatives only.

When the software is used in the


diagnostic / service mode, the Engineer
performing service must verify that the
scanner is taken out of service mode
before returning control of the scanner to
the customer. The software will return to operators
mode as soon as the software is exited and restarted.
The operator mode may also be changed under the
Tools/User Options menu.

The Service software gives the Technician


access to all areas of the DPX-NT / MD+
Software, DO NOT adjust AS settings.

• The Diagnostic Software is available through the User Options Tab

3.0.1 To Access the Service Software:


• Start the DPX-NT / MD+ Software by double clicking the DPX-NT / MD+
Icon on the desk top

• Double click the Tools option on the menu bar

• When the drop down menu appears double click on User Options

• Double click on the Systems Tab on the dialogue box

• Click on Service under Display Mode

• When prompted enter the password Award

Note: the password is case sensitive

• The Tools option on the menu bar will now contain Service Options, and
Service scanning options will also be available under the F2 Measure
Option

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 81 of 137


Figure 3-1. The Tools/Diagnostics Menu

3.1 The Tools Menu


3.1.1 Spectrum
Allows the service engineer to view and print the raw energy spectrum for the
detector. The detector can also be repeaked from Spectrum.

3.1.2 Stability Run


Used for the detection of arcing and testing the detectors performance

3.1.3 Signal Monitor


View count rates for low and high-energy channels for the detector.

3.1.4 Primary Calibration


Not used on DPX-NT

3.1.5 Secondary Verification


Used in the manufacturing process

3.1.6 Pileup
Used in the manufacturing process

3.1.7 Set Download Parameters


Download and verify the Firmware versions for the cSBC.

Page 82 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


3.1.8 enCORE Composer
This is not a service tool - (Physical and Business Report Generator - see the
enCORE Operator’s Manual for information).

3.1.9 Error Log


Ability to view and print the contents of the diagnostic error log (see section
3.4)

3.1.10 Copy Configuration


Ability to copy the scanners Error log, Configuration Files and / or QA
Database to a drive on the host PC - useful when a host PC is replaced or
when diagnostic information needs to be sent to LUNAR.

3.1.11 User Options


Set up user configured options for the scanner.

3.1.12 Service Options


Service Mode scanner configurations (see section 3.5).

3.1.13 Outbox
Configure Printing, E-mail and Faxing options.

3.1.14 System File Editor


Edit system settings for the enCORE software - do not alter these settings
without LUNAR Customer Support authorization.

3.2 Tools / Diagnostics Menu


3.2.1 Pileup
Determines Detector parameters for multiple events occurring simultaneously
- not used for service.

3.2.2 Calibration Pileup


Not used for service.

3.2.3 Spillover
Perform and Acquire a Spillover Measurement (test of source spectrum and/
or detector resolution).

3.2.4 Count Rate


Acquire and print Reference Counts (count rate of detector must be
performed through lucite).

3.2.5 Scanner Motion


Control manually and perform scanner motion diagnostics and motion
configuration.

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 83 of 137


•Motion Tests Tab

•Limit Switch Monitor - used to verify limit switch status

•Limit to Limit Test - used to verify limit switch positions

• Configure check box- this box must be checked when limit to limit is
run to establish the scanners range of motion (scan window).

•Cable Stress Test - used to cycle Transverse and Longitudinal


Mechanics - can be used to check new cables / check for binding

•Home scanner - moves mechanics to Home position

•Motion Commands Tab

•Home Position - change Arm Parking position from head to foot -


does not effect where the daily QA (Secondary calibration) is
performed - Move to Home - sends scan arm to home position.

Note: When the Home position is changed from the head end of the
table to the foot or vice versa, the Limit to Limit test (see
Motion Tests tab above) must be run with the Configure box
selected.

•Move Steps - move the scan arm to a specific location on the table -
move Absolute moves an absolute distance from home, Move
Relative - move relative to last position

•Joystick and Laser - Enable and Control Scanner Motion manually,


change motor speeds, toggle laser on and off

•Lower portion of the screen displays the status of limit switches, scan
arm position, and positioning switch (joystick) status.

•Find Block Tab

•Secondary Calibration - can be used to find / verify that the scanner is


able to find the different chambers on the secondary calibration (daily
QA) block. THis tool can be useful for troubleshooting “cannot find
Block errors” and for Transverse binding issues.

3.2.6 Scanner Detector


•Detector Tests Tab

•Gain Control - adjusts the gain on the AGS amplifier used in the
detector peak setting - peak adjusts only the AGS amplifier.

•Bias Control - adjusts the bias of the detector - bias peak adjusts both
the detector bias and the AGS amplifier gain.

Page 84 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


Note: The bias and peak setting of the detector work together to
maximize the number of counts - the bias peak adjusts both
the bias on the detector and the AGS amplifier peak, peak
adjusts only the AGS amplifier peak. When the AGS peak falls
out of range the enCORE software will automatically perform a
bias peak. The operator will not see any difference in
operation, however, the Daily QA may take about 40 seconds
longer.

•DCA window settings - view DCA window settings - these settings


should not be edited.

•Detector status - allows the service engineer to view the detector


status real time - information displayed includes Peak setting Bias
setting, and DCA windows.

•Detector AGS Tab

•AGS DAC control (verify AGS DAC settings), check on detector


rollover, and check on channel count rates.

DO NOT adjust the detector Bias, DCA


windows or AGS DAC settings -
changes in the Detector bias, DCA
windows or AGS DAC’s will change the
scanner’s results.

•AGS Mode - the operate / calibrate signal (disables the AGS systems
so that the detector can be peaked correctly) can be toggled.

3.2.7 Scanner X-ray


•X-ray Tests Tab-

•Manual Test - Control Manually the Shutter, X-ray on Indicator and


patient Positioning laser

•Automatic test - automatically toggle the Shutter, Laser and X-Ray


Relay

•X-Ray Source - Turn on source (ramp and make x-rays), warm up a


tube head at install or replacement.

•X-ray cycle test - cyclically ramp the source - (can be used to test
ramping failures or for arcing)

•X-rays Tab

•Ramp the kV and / or the mA of the x-ray generation system and view
feedback real-time. Faults are displayed and polled real-time in the
window on the right.

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 85 of 137


•Ports Tab - view status of and send commands to cSBC ports (see
section 2.4 for cSBC port definitions).

•Download Tab - identical to Set Download Parameters (section


3.1.7)

•Misc. Tab- not used for service.

3.2.8 Lin/Rep
Not Used for Service

3.2.9 Limit Switch Adjustment


Verify and Adjust Limit switch positions.

Note: When using Limit Switch Adjustment, DO NOT alter the


default settings.

3.2.10 Scanner Disconnect


Disconnect the scanner from the DPX-NT / MD+ software (reset the cSBC).

3.3 Diagnostic Scan Modes -


Performed under F2 Measure - only visible when the scanner is in service
mode.

3.3.1 Table Top Scan -


Not used in Service

3.3.2 Alignment Scan


Verify that the Source and Detector maintain a constant count rate in all areas
of the scan table. Also useful for testing cables (see DXAP3001 Chapter 5
Appendix B)).

3.3.3 Beam Wobble scan


Quantifies beam movement relative to the detector when the scanners
mechanics are moving (see DXAP2010 Chapter 5 Appendices).

3.3.4 Hacksaw
Not used for DPX-NT.

3.4 Error Log


• The DPX-NT / MD+ will maintain an error log file whenever the scan table
is in operation. All errors, fatal or nonfatal will be logged here. The Error
log gives a brief description of the error and if possible the origin of the
fault.

Page 86 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


• The error log is one of the best troubleshooting tools for scanner
malfunction.

Note: To aid in a quick and accurate response when requesting


assistance from LUNAR, please have an error log available for
review by LUNAR Customer Support.

The error log file is named errorlog.bin, this file is the current error log session.
Older sessions are named errorlog.1, errorlog.2 and so on up to errorlog.4.

When requesting help from LUNAR, E-mailing the current error log file and a
description of the symptoms will aid in a rapid diagnosis (see Copy
Configuration in section 3.1.10 for information on copying the error log to an
E-mailable file).

3.4.1 Printing the Error Log.


Note: The scanner does not have to be in Service mode to view or
print the Error Log.

• The Error Log is located under the Tools dropdown menu. Left click on
Tools / and then on Error Log.

Figure 3-2. The Error Log Dialogue Box

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 87 of 137


• Session - The DPX-NT / MD+ software will maintain up to 10 “sessions”
- each session is up to 100 errors.

•A new session is created by shutting down and restarting the


enCORE software. A new session may also be started by clicking on
the button

•To view a session, left click on it, all errors (if present) will be shown
in the box below.

•The Errors list may be printed (click on print errors) or it may be


searched with the Find Error button.

•The details of the failure are shown when the error is highlighted and
then right clicked on.

Figure 3-3. Error log entry details

3.4.2 Troubleshooting Help software


This feature is found in software versions 2.26 and greater. When an error is
highlighted and the Troubleshoot button is clicked on, it opens an online
enCORE troubleshooting guide to assist the service engineer in
troubleshooting the system by error code.

Page 88 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


Figure 3-4. enCORE Troubleshooting help screen capture.

How to use the Troubleshooting Help software:

This Help software is based on Fault Tree Analysis of FTA. The fault trees are
in an outline form. Each heading in the outline represents an “or” choice. For
example, the basic fault below “Shutter will not close” has three possible
causes: “Shutter stuck”, “Solenoid Return Spring not working”, or “Solenoid
powered when it should not be”. Each of these in turn lists their possible
causes.

Sample Fault Tree (from Prodigy Collimator):

Shutter will not close

I.Shutter stuck – move by hand to test

A.Slotted Wheel scraping OMI board

B.Shutter scraping Shutter Plate

II.Solenoid Return Spring not working – move by hand to test

A.Spring broken

B.Spring no longer anchored

III.Solenoid powered when it should not be

A.FOINK problem – Test by disconnecting J1. If the shutter closes, swap the
SBC. If the shutter stays open, swap the FOINK.

B.SBC problem – Test as above.

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 89 of 137


Note: This type of analysis is meant to identify all possible causes,
giving the user a complete list to work from when ruling out
possibilities. If a cause is not listed in the appropriate fault tree
analysis, please e-mail Lunar about the discrepancy.

3.4.3 Reading the Error Log

Figure 3-5. Sample DPX-NT / MD+ Error Log

A Sample DPX-NT / MD+ Error Log with 3 separate error conditions is


illustrated in figure 3-3. An error condition or a subsystem failure does not
typically generate a single error message, rather a cascade of events typically
occurs (see example explanations below).

Error Event 6: The Scanner was unable to ramp the tubehead due to a mA
feedback failure, the error log entry indicates that this was attempt #5 (all five
entries are in the log - only entry 5 is shown to save space). Entry (6) in the
log is Unable to Ramp X-ray current. The status of the x-ray generator(s)
are listed, Program, Expected Values and Feedback are all listed. From the
readings listed the feedback for mA is 0.00 - causing the failure. The failure
was induced by disconnecting the mA feedback at the D-MAX board on an
NT(A) system.

Page 90 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


Error Event 7: follows with Unable to ramp x-ray current - Positive x-ray
current monitor deviation - this is the conclusion of error event 6.

The final entry in the error event is DET SBC DTR Reset this is the cSBC
resetting.

Error Event 8: Shutter failed to open is listed in the error log, the cause of the
event was a disconnected shutter solenoid on a DPX-NT(A). The shutter
failure is the only error log entry for the failure.

Error Events 9-11: This is a cascade of errors - identified as a cascade by the


time stamp - all of the errors occur within seconds of each other.

Error Event 9: Longitudinal motion failure as the log indicates the


longitudinal OMI failed to detect motion during a patient scan. This is followed
by Error Event 10, the cSBC resetting, because the error occurred during a
patient scan the cSBC is reset to interrupt x-ray generation and stop the
motors (fail safe condition). Error Event 11 is the host PC reading the cause of
the interrupt sent to the cSBC - which was that the Longitudinal motor failed to
move.

Note: The cause was actually the loss of the OMI signal - as the
cSBC does not know if the motor moved or if OMI signal is lost
and the probability of OMi failure is lower than a motor failure -
the error message reads “Longitudinal motor failed to move”.

Note: If you are unsure as to which errors belong together, start a


new session in the error log, and then recreate the error
condition. when this is done, only the errors involved in the
error event will be in that log.

3.5 Service Options


3.5.1 Measure Tab
• Centerline Offset - enter the centerline offset in mm - (See DXAP201x
Chapter 5 Appendices)

• Homing end -

• Primary Calibration File - choose the Primary Calibration file - DPX-NT


scanners come with a factory calibration file. The only time the calibration
file should be changed / modified is if the customer needs to run 2.1x
software and 2.2x software on the same system.

•The formats of the two calibration files 2.1x and 2.2x are not
compatible.

3.5.2 Analyze Tab


• Auto Reanalyze all images

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 91 of 137


• The phrase “Image not for Diagnosis' may be toggled off so it does not
appear on the printout.

DO NOT disable the phrase “Image not for


Diagnosis” in the USA.

• Slope Adjustment - to enhance cross calibration between LUNAR


scanners a slope adjustment may be entered. For information on cross
calibration please contact LUNAR Customer Support.

3.5.3 QA Tab
• The lockout times (in hours) for when the user is locked out of the
scanning options and needs to run a QA or secondary calibration may be
modified.

• The warning time (in hours) for when the user is prompted”Reccomend
Daily QA before measuring patients“ warning them that a QA or
Secondary calibration is advised, but not required may be modified.

• The maximum number of sessions may be set - 2000 is the


recommended default.

3.5.4 Serial Tab


• Scanner Communications

•Enable or disable serial communications

•Communications port selection

•Communications tests - test the ability of the host PC and cSBC to


communicate. Stress Communications test is a longer duration test.

• Calibration - DO NOT adjust any entries in the calibration section of the


dialog box with the exception of the QA block batch number. The block
batch number should match the batch number on the block.

Page 92 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


4 Troubleshooting

This chapter contains troubleshooting techniques for


diagnostic failures, failing quality assurance tests, mechanical
failures and imaging problems.

4.0 Diagnostic Failure Codes


4.1 Transverse Motion failure
4.1.1 Operator Induced - switch closed during scan
4.1.2 Mechanical Failures - Unusual noise or irregular motion
4.1.3 Loss of OMI signal
4.2 Longitudinal Motion failure
4.2.1 Limit Switch Tripped During a Scan
4.2.2 Longitudinal Binding
4.2.3 Loss of OMI Signal
4.3 Failure of the DCpower supply
4.4 Emergency Stop Button
4.5 Tube Head Thermostat
4.6 Communication Error
4.7 Other Diagnostic Failure Codes
4.7.1 Reasons For Invalid Diagnostic Failures
4.8 Failing Quality Assurance Test
4.8.1 Block Position
4.8.2 Beam Stop Action
4.8.3 Mean% Spillover
4.8.4 Reference Counts and Ratio
4.8.5 Ratio Fluctuations
4.8.6 Transverse or Longitudinal Mechanics
4.8.7 Tissue Value
4.8.8 Bone Mineral of the Standard Chambers
4.8.9 Symptoms of High and Low KV
4.9 Reference Counts
4.10 Arcing
4.10.1 Limit Switch Tripped During Scan
4.11 Imaging Problems
4.11.1 White, or Grey in the first or second scan line:

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 93 of 137


4.11.2 Femur Scan Problems
4.11.3 AP-Spine Image Problems: Probable causes
4.11.4 Broken Signal Cable
4.11.5 Loss of tube head current
4.11.6 X-Ray Relay Failure
4.11.7 Unstable AC Line
4.11.8 Arcing
4.12 Failing Alignment Test Results
4.12.1 Image
4.13 Indicator Failures
4.13.1 X-ray On LED Blinking
4.13.2 Shutter Open LED Blinking
4.13.3 Shutter Not Operating
4.13.4 End of Exposure Alarm During Scan
4.14 Communications Failures
4.15 Viewing Quality Assurance Trends
4.15.1 What to Look for in the QA History
4.16 MAX Board Troubleshooting
4.17 OMI Board Troubleshooting
4.18 SBC Troubleshooting
4.19 XORB Troubleshooting

Page 94 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


4.0 Diagnostic Failure Codes
The following conditions halt the operations of the scanner, and generate a
diagnostic failure message which is displayed on the monitor screen:

All of these interrupts are detected by the cSBC

• Transverse Mechanics Failure (4.1)

• Longitudinal Mechanics Failure (4.2)

• Emergency Stop Button activated (4.3)

• X-ray source over temperature (4.4)

• DC power supply failure (4.5)

4.1 Transverse Motion failure


Motion Detection

• cSBC / OMI fail to see transverse motion during a patient scan, an


interrupt signal is generated and the cSBC resets and closes the shutter.

• A slotted disk at the end of the arm rotates through an infrared beam
(OMI), and pulses are sent to the cSBC board when there is
transverse motion.

• The transverse and longitudinal motion detection system (on cSBC) is


operational during patient scans and the “find block” test of the daily
QA. If the problem occurs during a quality assurance, it may be high
voltage arcing or a problem with the OMI.

4.1.1 Operator Induced - switch closed during scan


• Cause: When scanning, a defect in the transverse mechanics or a mis-
positioned patient may cause one of the transverse limit switches to
close when the scanner is acquiring data. Closure of a limit switch
prevents further operation of the motor. The shutter will close, the shutter
open lamp will go out and the end of exposure alarm will sound. Seconds
later the error message will appear on the screen.

Solution: If the patient is not centered on the table top or if the region
being scanned is too close to a limit in transverse travel, the limit switch
switch may be close while scanning. Re position the patient on the table,
further away from the limit.

4.1.2 Mechanical Failures - Unusual noise or irregular motion

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 95 of 137


If the Detector / Tube Head motion is irregular, or scraping noises are heard,
the Transverse Mechanics may be binding. This symptom will typically be
detected as a failure by the daily QA Mechanics Test.

• Symptom: If the Detector Array / Tube Head is not moving when


scanning check:

• Mechanics are free to move

• Centent

• Stepper Motor

Troubleshooting Binding

Turn off the power to the scanner and move the affected parts by hand. Feel
the motion for spots where the carriages are more difficult to move. Listen for
unusual noises.

The following are common causes for Transverse Binding

• High Voltage Cable Routing

One of the most common problems is a failure of the Source and Detector to
reach the rear Limit Switch due to the Tube Head running into its own high
voltage cables. These cables must have a hump formed at the Rear
Longitudinal Carriage that allows the lower portion of the Tube Head to pass
under the cables.

The High Voltage Cables can also impede transverse motion toward the front
Limit Switch. This is caused when the cables have been tied down without
enough play for the Tube Head and Detector to move all the way forward.
These problems should be investigated by manually tripping both Limit
Switches while inspecting for cable conflicts, binding, or tension problems.

• Check to see that the bottom of the Tube Head is not hitting the
Transverse Centent

• Wiring

In rare instances, the wires from the Shutter Solenoid and/or Fans can snag
on the bolts that protrude through the frame on the foot end of the scanner.
This is solved by properly tying down these wires.

• Transverse Belt

The Transverse Belt should not be excessively tightened or this will cause
excessive binding in the transverse mechanism. It should be possible to
deflect the belt by 4 cm when it is properly tightened. Sometimes the spare
belt material near the clamp on the Tube Head Carriage comes into contact
with the forward gear and prevents the scanner from going all the way to
Home position.

• Drive Wheels

Page 96 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


The wheels that support the Tube Head and Detector Carriages must be
adjusted so that they come into perfect contact with the Transverse Rails.
These wheel are best inspected with the table top, front panel and arm covers
off so one can sight down the extrusions. However, it is possible to test the
wheels by preventing any wheel from turning and seeing if the carriage will
still move. By preventing any wheel from turning, you should be able to tell
that it slides along the extrusion while the others roll. This indicates that the
wheel has not been excessively tightened down. This is least likely to be the
source of transverse motion problems, as it is unlikely that the adjustment of
the wheels would have become any tighter over time. They would be more
likely to loosen over time and fail alignment tests (see Alignment Test, section
4.12). Adjustments can be made by loosening and rotating the eccentric
bearings of any of the lower wheels.

• Gear And Pulley Positioning

Check all appropriate gears and pulleys. Verify that the set screws are
tightened and the gears and pulleys are not out of position.

• Limit Switch Positioning

If the transverse motion seems to be acceptable, but the number of steps in


the Quality Assurance Test is failing, check the position of the Limit Switches.

On Total Body scans, a limit switch out of position could allow the Source/
Detector Carriages to hit the frame or panels before the Limit Switch is
actuated.

• Transverse Motor

Check the Transverse Motor for a broken wire in one of the internal coils, or a
bad electrical connection to its Centent Motor Controller.

• Transverse Centent

The Centent Motor Controller, if defective, will cause transverse motion


problems. Sometimes the controller works well enough to acquire some
scans, but it will not provide enough torque to complete every fast scan. The
CURRENT SET voltage at terminal 11 on a properly operating controller
should be 14 to 16 VDC (transverse motor wired in parallel).

The longitudinal and transverse Centents are identical, and can be


exchanged. If the problem remains, the Transverse Motor should be
replaced.

4.1.3 Loss of OMI signal


If the error occurs consistently on the first line of a patient scan or during the
find block portion of the daily QA, and the scanner is moving in the transverse
direction, then check the following:

Cause: The Interrupt signal is being lost.

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 97 of 137


Solution 1: The pulses that normally enter the cSBC board at J14 (the
black wire at the center of the connector) may have stopped. These
pulses are necessary to keep the cSBC from resetting. These pulses can
be seen on a cSBC board LED.

Use the DPX-NT service software (Tools/Diagnostics/Scanner Motion /


Motion Commands Tab) to set the joystick speed to 50 steps, enable the
joystick and watch the LED. If the OMI is working the LIght will flash
when the Transverse Motor is run. If the LED flashes when the
mechanics are engaged, but the error still occurs, the interrupt was
invalid. Check for arcing in the high voltage system or replace the cSBC.

If the LED D9 (B in figure 4-1) does not flash,

• Verify that the slotted disk at the front of the lower arm rail is in the
middle of the slot between the photo diode and photo transistor.

• The slotted disk must be completely flat and remain in the center of
the sensor slot during its entire rotation.

• If the slotted disk has been in physical contact with the optical sensor,
the sensor may have debris on it, disassemble this mechanism, and
clean the sensor and the slots of the disk.

• Check the Cable running from the cSBC to the OMI for a cable break
by checking the individual wires for continuity.

4.2 Longitudinal Motion failure


• cSBC/ OMI fail to see longitudinal motion during a patient scan, an
interrupt signal is generated by the cSBC which resets and closes the
shutter.

• A slotted disk at the foot end of the table on the pulley for the
Longitudinal Drive Belt rotates through an infrared beam (OMI), and
pulses are sent to the cSBC board when there is transverse motion.

• The transverse and longitudinal motion detection system (on cSBC) is


operational during patient scans and the find block portion of the daily
QA. If the problem occurs during a quality assurance, for instance, it may
be the high voltage system is arcing or the OMI may be malfunctioning.

4.2.1 Limit Switch Tripped During a Scan

Page 98 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


• Cause: When scanning, a defect in the longitudinal mechanics (binding)
or a mis-positioned patient may cause one of the longitudinal limit
switches to close when the scanner is acquiring data. Closure of a limit
switch prevents further operation of the motor. The shutter will close, the
shutter open lamp will go out and the end of exposure alarm will sound.
Seconds later the error message will appear on the screen.

Solution: If the patient is not centered on the table top (length wise) or if
the region being scanned is too close to a limit in longitudinal travel, the
limit switch may be close while scanning. Re position the patient on the
table, further away from the limit.

4.2.2 Longitudinal Binding


• If the scan arm fails to move when scanning check the stepper motor,
and Centent, if arm motion is irregular, check for binding.

Turn off the power to the scanner and move the affected parts by hand. Feel
the motion for spots where the arm is more difficult to move. Listen for unusual
noises.

When moving the arm longitudinally, only


push on the arm column. Pushing on the
forward parts of the arm can ruin Tube
Head/Detector alignment.

• Cable Track

The major impediment to longitudinal motion is the plastic Cable Track that
runs through the trough at the rear of the scanner.

This track is attached at two points: at the Rear Longitudinal Carriage and to
the scanner frame (low, rear and center) each spot by 4 bolts. Also, adequate
slack must be left in the cables inside the Cable Track or they will stop the arm
from moving fully to the foot end.

Should the Cable Track detach from the scanner frame, it will slide freely in
the trough and will eventually cause trouble. This can allow the Cable Track to
get in between the Rear Longitudinal Carriage and the scanner frame on the
foot end preventing the tripping of the limit switch.

• Front Longitudinal Carriage Dragging

Check the distance between the front longitudinal carriage and the
longitudinal rail with a go/nogo gauge (See installation Procedure DXAP2000
Chapter 5 appendices). The carriage should not rub the front rail, if necessary
insert shims behind the front longitudinal carriage.

• Tube Head Cable Routing

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 99 of 137


Problems at the head end of the scanner can be caused by a limit switch
being out of position (thus the arm runs into the frame before the Limit Switch)
or because the High Voltage Cables are tied down incorrectly where they
snake around from the Rear Longitudinal Carriage to the panel under the
Tube Head (High Voltage Cable Trough). If the cables are tied down too far
toward the head end (at the point where the three tie downs are), they can run
into the Longitudinal Motor Assembly and hold the arm away from the head
end Limit Switch.

• Slip Clutch

A slip clutch is part of the longitudinal motion system to limit torque. This is a
feature to protect the patient should he/she pinch an arm or leg between the
back side of the scanner and the Arm Column. If the Slip Clutch is set too
loose, it will fail to move the belt and will just "slip" as the motor turns. This
may produce the following symptoms:

• failure of Quality Assurance scan Mechanics test

• compressed image in limited areas of the image

• a Longitudinal Mechanics diagnostic failure

• Gear and Pulley Positioning

Check all appropriate gears and pulleys. Verify that the set screws are
tightened and the gears and pulleys are not out of position.

• Limit Switch Positioning

If the longitudinal motion seems to be acceptable, but the number of steps in


the Quality Assurance Test is failing, check the position of the Limit Switches.

On Total Body scans, a limit switch out of position could allow the Source/
Detector Carriages to hit the frame or panels of the scanner before the Limit
Switch.

If the mechanical stop is reached before the Limit Switch is actuated, check
carefully to see which part of the Arm Assembly is in contact with the Table
Assembly. The front part of the Lower Transverse Extrusion is clamped to the
Longitudinal Drive Cable at the front of the scanner. If the Lower Transverse
Extrusion is not clamped in such a way that it forms a 90 degree angle with
the length of the table, the rollers at the front end of the Lower Transverse
Extrusion may strike the end of the scan table before the Limit Switch is
actuated.

• Longitudinal Motor

Check the motor for a broken wire in one of the internal coils, or a bad
electrical connection to the Centent Motor Controller.

• Longitudinal Centent

Page 100 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


The longitudinal Centent Motor Controller may be the cause of a failure. The
current set voltage should be roughly 9-11 volts at terminal 11 on a properly
operating controller. If this voltage is not correct, replace the Centent
Controller.

The longitudinal and transverse Centents are identical, and can be


exchanged. If the problem remains, the Longitudinal Motor should be
replaced.

• Longitudinal Belt

The Longitudinal Belt should not be tightened too much or this will cause the
brackets holding the gears to deform at either end of the scanner. When the
belt is properly tightened, it should be possible to deflect the upper and lower
sides of the belt so that they touch within 8 cm of the gears at either end.

• Drive Wheels

The rollers in front and the wheels in back that support the Arm must be
adjusted so that they come into perfect contact with the Longitudinal Rails.
Test them by preventing any wheel from turning and see if the carriage will still
move. By preventing any wheel from turning, it should be possible to slide the
carriage along the rail with one wheel dragging while the others roll. This
indicates that the wheel has not been excessively tightened down.
Adjustments can be made by loosening and rotating the eccentric bearings of
any of the lower wheels.

4.2.3 Loss of OMI Signal


• If the Scan Arm motion is irregular, or scraping noises are heard, the
Longitudinal Mechanics may be binding. This symptom will typically be
detected as a failure by the daily QA Mechanics Test.

If the error occurs consistently after the first line of a patient scan, and the
scanner is moving in the longitudinal direction, then check the following:

• Cause: The Interrupt signal is being lost.

Solution 1: The pulses that normally enter the FOINK board at J (the
black wire at the center of the connector) may have stopped. These
pulses are necessary to keep the cSBC board from interrupting. These
pulses can be seen on the cSBC.

Use the DPX-NT service software (Tools/Diagnostics/Scanner Motion /


Motion Commands Tab) to set the joystick speed to 50 steps, enable the
joystick and watch the LED. If the OMI / cSBCis working the LIght will
flash when the Longitudinal Motor is run. If the LED flashes when the
mechanics are engaged, but the error still occurs, the interrupt was
invalid. Check for arcing in the high voltage system or replace the cSBC.

If the LED does not flash,

• Verify that the slotted disk at the foot end of the scan table is in the
middle of the slot between the photo diode and photo transistor.

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 101 of 137


• The slotted disk must be completely flat and remain in the center of
the sensor slot during its entire rotation.

• If the slotted disk has been in physical contact with the optical sensor,
the sensor may have debris on it, disassemble this mechanism, and
clean the sensor and the slots of the disk.

• Check the Cable running from the cSBC to the OMI for a cable break
by checking the individual wires for continuity.

4.3 Failure of the DCpower supply


The DC power supply is only enabled when the X-ray tube is ramped, use the
service software (Tools/Diagnostics/ Scanner X-ray) to attempt to ramp the
Tube Head.

The Red and Green LED’s on the MAX board will be lit whenever the DC
power supply is up and running.

If the LED’s illuminate and the go out:

• Measure the output of the DC power supply, and verify that it remains
constant during the voltage ramping and scanning operations.

• Check the High voltage power supplies, insure they are not arcing (Error
Log - see section 3.2) and are ramping.

If the LED’s fail to light:

• This supply is turned on by the X-ray Relay, so verify that the Relay is
closing. If not, then either the Relay is bad or it is not receiving the signal
from the cSBC.

• Check the continuity of the cathode, the filament may have broken, MAX
board TP 4, TP 5 and TP 13 should be continuous with the Tube Head
control cable connected.

• The Tube Head Thermostat is wired in series with the Relay, so if it has
opened, the Relay will not be able to close.

• 28 VDC should also be measured at the Terminal Block. If not, check the
continuity of the wiring and refasten all connections. Also, check the wire
tie-downs for excess tension they may be putting on the wires.

• It may be necessary to check the wiring from the Terminal Block to the
MAX board and to the High Voltage Power Supplies.

4.4 Emergency Stop Button

Page 102 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


The Emergency Stop button is a normally closed circuit, if it opens the
Emergency Stop and the cSBC will generate an interupt.

• Verify that the emergency stop button is out. If it has been pressed in,
press it again to release it.

• If 26VDC can be measured on both pins of cSBC connector J17, the


Emergency Stop Switch and the wires connecting it to the cSBC board
are good and the cSBC board should be replaced.

• If 26 VDC is missing, the Circuit is open between the cSBC and the
Switch.

• Insure the switch is functional

• Check the continuity of the wires from the switch to the cSBC board.

4.5 Tube Head Thermostat


There is a thermostat inside the Tube Head. This thermostat is normally
closed, but opens when the Tube Head temperature is too high.

• The thermostat will close again automatically after a cool down period of
usually less than 30 minutes. If 0 VDC is measured on both pins of cSBC
connector J15, the Thermostat and the wires connecting it to the cSBC
board are good and the cSBC board should be replaced.

4.6 Communication Error


The computer is not communicating with the cSBC board.

Verify communications with the cSBC in the service software (Tools/Service


Options Comm tab - Test Communications).

If this is unsuccessful, verify that the I/O cable from the computer to the cSBC
Board is secure. Also, verify that the comm port is configured correctly (see
DXPC 2000 Chapter 5 appendices) and that all required drivers are present. If
all fails, the cSBC or computer serial port is defective.

4.7 Other Diagnostic Failure Codes


Multiple error codes are possible. These will be displayed one after the other,
but the first one displayed is probably the problem. The Error Log will always
show the failures in the correct order, the others are generated as a result of
the first failure.

4.7.1 Reasons For Invalid Diagnostic Failures

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 103 of 137


Arcing of the x-ray high voltage system releases a large amount of
electromagnetic energy. This energy creates noise in the electronic circuits
and may cause the SBC to do unpredictable things, such as give invalid
diagnostic errors. Arcing is usually accompanied by white, blue or black lines
across patient scans (See Arcing, section 4.10).

4.8 Failing Quality Assurance Test


See chapter 2 section 12, for explanations of the various tests that are
performed during the running of the Daily Quality Assurance.

If any of the QA test results fail, none of the results are considered valid. The
results will be stored in the Quality Assurance History file, but these values will
not be averaged with the other results for calibration purposes. In addition, the
software will prevent patient scans until a passing daily QA has been
completed. Recent valid QA's are necessary for accurate results.

4.8.1 Block Position


After starting a QA, the operator is prompted to place the QA standard n the
table, if the DPX-NT fails to detect the Block in the correct orientation the QA
will not progress.

• Check Block position and orientation

The scanner may not be finding the correct "Home" position. The correct
"Home" position aligns the center of the x-ray beam with the center of the
Brass Piece when the Standard is correctly positioned. The Air Counts must
be obtained outside of the QA standard, next to the Brass Piece. If the
Reference Counts are obtained with the x-ray beam passing through the
Brass Piece, the Quality Assurance Scan will fail.

• Verify that the scanner limit switches are set correctly with the DPX-NT
home position jig (see DXAP2000 - DPX-NT Installation Procedure
Chapter 5 appendices).

4.8.2 Beam Stop Action


The Beam Stop Action test verifies the ability of the lead shutter to attenuate
x-rays from the tube head by verifying that the Background counts when the
shutter is closed are lower than counts taken through the brass piece with the
shutter open.

4.8.3 Mean% Spillover


An increase in the Spillover percentage over time is an indication that the
detector is losing resolution, even if the Spillover test does not fail. A large
change in the Spillover can be explained in some cases by standard
positioning (or a small change in the home position of the scanner). If the
beam is not fully eclipsed by the brass piece during the Spillover test, the
Spillover value increases dramatically.

Spillover stability is a test of detector bias drift. if the detector bias is drifting,
the mean spillover value will also drift.

Page 104 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


4.8.4 Reference Counts and Ratio
During this test the baseline is established which will later be used for
comparing with the values obtained during the scan of the standard. Care
should be taken that the x-ray beam is not missing the standard, nor being
obstructed by the brass piece on the standard during the High and Low Air
Count test.

The Ratio value should remain fairly constant over time.

4.8.5 Ratio Fluctuations


The ratio of High to Low Channel Air Counts is the way the bone density
measurement is calculated by the changes in the ratio of High to Low Channel
counts. The Ratio should remain constant as long as the x-ray beam quality
and the resolution of the detector remain constant.

It is possible to view the Ratio trends in the Quality Assurance History. Check
both the 3mA Ratio and the 150 Ratio to determine if the either Ratio has
changed significantly.

The problems which could cause a change in air ratio are:

· deterioration of the resolution of the detector

· current leakage through the transorbs on the XORB board

· a faulty High Voltage Power Supply (kV unstable over time)

· changes to the x-ray tube insert

These are very difficult to diagnose by a method other than substitution of new
components until the Reference Counts Ratio returns to normal.

Detector deterioration or unstable High Voltage Power Supply, will usually


cause the Alignment Test Scan to fail.

The specification for XORB Board transorbs is that they must allow less than
0.5 micro amperes reverse bias current. At LUNAR, each transorb is
measured by applying a 5 Volt reverse bias to the transorb and a 100 kW 1%
resistor connected in series. The voltage measured across the 100 kW
resistor must then be less than 50 mV.

4.8.6 Transverse or Longitudinal Mechanics


Failure of this test is caused by one or more of the following:

• Incorrect Limit Switch positioning

• Defective Limit Switch

• A mechanical constraint

• A defective Motor

• A defective Centent Motor Controller

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 105 of 137


The values recorded for these tests should remain fairly constant over time.
Variations between QA's of under 25 steps should not be a cause for concern
as 1 transverse step = 0.05 mm and 1 longitudinal step = 0.1 mm, so the
actual variation is only a few millimeters.

If the number of steps continually increases from QA to QA, this could indicate
an impediment to the scanner's motion and should be rectified (see 4.1
(Transverse) or 4.2 (Longitudinal)).

4.8.7 Tissue Value


The tissue value should not be the sole failing parameter. The failure of this
test is usually accompanied by problems with Bone Mineral values (see
below) or Reference Counts Ratio trends (see above). If this is failing alone it
is probably a problem with the counting system or high or low kV (4.8.9).

4.8.8 Bone Mineral of the Standard Chambers


The software contains values which it expects the scanner to measure for the
scan of each chamber. If the mean of the BM measurements made for the
chamber does not fall within the predefined software limits the QA will fail. If
the percent coefficient of variance is over 3%, the QA will fail.

These failures will always occur if the Reference Counts or Reference Ratio
test have deviated severely from normal results. However if the Reference
Count results look normal, and the values are approximately equal to the
numbers obtained during the scanner installation, then the counts may be
unstable.

Another cause of failing on the bone chamber measuremnts can be arcing


during the QA. This can be detected by examining the QA History graphically
and looking for variation of the Large BM values. Note the few data points that
vary. These are the early signs of arcing.

This will also be apparent (but not obvious) on the QA Results printout. The
arc occurred in the third standard scan and elevated the BM values. If a
customer reports a failing QA because of a bone chmaber measurement
being too high, be aware that this could be an early warning of arcing. Obtain
from the site the QA history file and error log for analysis. Look for variation of
the Large BM values.

It is very important to notice these early warning signs of arcing so that the
system can be re-greased before any damage is done to the high voltage
cable connectors or the tube head.

4.8.9 Symptoms of High and Low KV


It has been noted in a small number of cases that defective high voltage
power supplies, or an arcing tube head may produce a voltage other than 76
kV. The DPX-NT monitors kV once per scan sweep.

Page 106 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


Note: :It would be prudent in both of these cases to take a positive
and negative power supply, AND a tube head. Also, the XORB
test points may look normal (indicating 76 KV) in each case as
well. This usually occurs where the power supplies are
defective and so the monitor voltages returned to the XORB
may not show the true voltage. When the tube head arcs
causing the voltage to go too high, then the XORB test points
may indeed show the actual voltages.

4.9 Reference Counts


4.9.1 Unstable Counts
To examine the scanner's ability to count detector pulses consistently over
longer periods of time, run the Alignment Test option from the service
software program.

• If the "End of Exposure Alarm" rings during the time the Alignment Test
scan is running, see "Alarm Pings During Scan" in section 4.13.4. If the
Shutter Open or the X-ray On lamps on the front panel blink, during the
test, see section 4.13 also.

• While the Alignment Test scan is running, measure the voltage at test
points 1, 2, 5 and 6 of the XORB board. They should have approximately
the following values respectively: -0.150VDC, 3.8VDC, 0.150VDC, and
3.8VDC.

• A pure DC voltage on an oscilloscope at test point 1 and 5 of the XORB


will verify that there is constant current through the filament of the X-ray
Insert.

• If there is AC ripple in excess of 0.2 Vpp, the power supplies should be


replaced.

• Start and stop the x-rays several times while observing the voltage at
TP3 and TP7 of the XORB board. This is the programming voltage from
the SBC, and although this voltage is dependent on the feedback
information returned to the SBC, the voltages at the XORB board test
points should be approximately the same each time the x-rays are
produced. The AC ripple on this signal must be less than 0.2 Vpp.
Replace the SBC board if the proper control signal is not present.

• When the x-rays are off between scans, verify that the shutter aperture
lines up with the aperture to the Tube Head beneath. Take some
alignment pictures and verify that the image is a bright, rectangular
image and the entire alignment box is visible. If the alignment box is not
visible, the shutter Tube Head and detector apertures are not aligned.

• Open and close the shutter to make sure that it returns to the same
position each time. Turn the x-rays back on, and make sure that you
obtain nearly the same count rate each time the shutter is opened.

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 107 of 137


4.9.2 No Counts
When the table to the right of the peak graph on the Quality Assurance
Results printout is entirely filled with zeros, use the Signal monitor program in
the Diagnostics (see section 5.1) to create x-rays at 76 kV and 150 uA and
open the Shutter for sampling. Then check the following items:

I.Is the amber X-ray On LED illuminated?

A.If it is, skip to part II.

B.If not, the voltage or current ramping has probably failed. Are both the red
and green LED's on the MAX Board illuminated?

1.Voltage or Current Ramping Failures

If not, is the red LED illuminated?

a.X-ray Relay or cSBC

If not, measure the output of the 28 Volt Power Supply. Check the operation of
the X-ray Relay or the cSBC Board which controls its operation. If all of the
above are working, the red LED may be defective.

b.MAX Board Fuse

If so, the fuse is blown on the MAX Board (see section 4.16).

2.If the red and green MAX Board LED's are illuminated, verify the following
test point voltages:

a.Current Ramping Failure

• XORB TP1 is approximately 0.150 VDC. If incorrect, go to step d. If this


test point is correct, verify that XORB TP5 is also approximately 0.150
VDC. If this voltage is incorrect, substitute a new Positive High Voltage
Power Supply.

b.Voltage Ramping Failure

• XORB TP2 and XORB TP6 are approximately 3.8 VDC. If these voltages
are incorrect, verify that the voltages on XORB TP3 and XORB TP7 are
approximately 3.8 VDC. If TP3 and TP7 are not equal the XORB jumper
at J26 is set in the wrong position. If they are equal but incorrect, test the
cable from the SBC to XORB, or substitute a new SBC Board.

c.X-ray On LED

Test the X-Ray On LED by inserting it into the Power On receptacle. Replace
if defective.

d.High Voltage System

Page 108 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


• Current ramping has failed. If TP1 and TP5 are zero and do not change,
check the polarity of the High Voltage Cables. The X-ray Insert is
essentially a diode, and will not conduct current from the anode to the
cathode.

• If either of the test points is at 1.0 VDC, the High Voltage Power Supply is
delivering as much current as it possibly can, and has automatically
limited the voltage.

• Check TP2 and TP6. If the voltage is approximately 3.8 VDC, the voltage
has been set properly.

Feel the heat sinks on the back of the High Voltage Power Supplies. If one of
the power supplies is cold, this is usually the defective one. If one is warm and
the other is hot, replace the hot one. The best troubleshooting technique may
be to substitute power supplies.

• If either TP2 or TP6 are incorrect, the possibility exists that one of the
High Voltage Cables or the Tube Head is shorted. The short may be
possible to find with an ohm meter, but often it takes several kV to break
down the defective component. It will be difficult to troubleshoot this
problem by any method other than part substitution.

Do not attempt to ramp the power supply above 50


kv without a cable connected, or arcing will occur
near the connector.

• Ramping between 40 and 50 kV is a good way to determine whether the


power supplies are able to increase voltage from 0 to 50 kV.

4.9.3 Decreasing Reference Counts


A long term, continuous decrease in the Reference Counts unaccompanied
by a change in Reference Ratio is a serious problem. A count rate of at least
500,000 Low Channel Reference Counts is necessary to maintain precision of
results when scanning thick patients. Reference Count values should change
by less than 10% from the day of installation. The possible causes are:

I.Lead Filings - Shutter Wear

A rough edge around the shutter aperture may be wearing away on the tube
head below. The lead filings fall off the shutter aperture and gather on the filter
below, causing a reduction in x-rays over time. Any wear should be easily
visible on the underside of the shutter paddle. If there is visible wear, the lead
dust must be removed from the top of the filter. The Collimator assembly must
be removed, the filter cleaned, and then the Collimator assembly reinstalled
and realigned.

II.A Change in Current Control Settings

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 109 of 137


Use the Signal monitor option of the service software program to produce x-
rays (Tools/Diagnostics? Scanner X-Ray). Select 76 kV and 750 µA
operation.

• Verify that the cSBC is properly controlling the current. TP11 of the MAX
board should be approximately 0.75 volts. This voltage is dependent on
feedback information from the power supplies, which makes it difficult to
troubleshoot by any means other than SBC substitution.

• Check the current through the X-ray Insert. The absolute value of the
voltage at test points 1 and 5 on the XORB board is proportional to the
current through the X-ray Insert. 1 millivolt is equal to 1 micro-ampere of
current. A current setting of 750 µA should give a reading of 0.750 VDC
at test points 1 and 5. If either of these voltages vary from the expected
by more than 10 millivolts, the MAX board could be at fault. If TP1 and
TP5 are more than 15 millivolts apart, substitute new high voltage power
supplies.

4.10 Arcing
The X-ray Tube Head Insert is an evacuated glass enclosure. An AC current
is applied to the filament inside the insert. It glows like the filament in a light
bulb, and electrons are boiled off into the evacuated space. A high voltage is
applied between the anode and the cathode causing electrons to rush toward
the anode, striking it and creating x-rays. As long as the insert is properly
evacuated, there can be no internal arc. However, no insert can be totally
evacuated and impurities can be ionized creating a lightning like effect; arcing.
During the arc the resistance of the insert is dramatically decreased and a
large amount of current flows.

• The significant amounts of electromagnetic energy released inside the


scanner can cause problems with the electronics of the scanner system
and may result in abnormal operations of the scanner. An arcing scanner
will have one or more of the following symptoms:

• complete lock up of the system

• a diagnostic error code message which is undefined or inappropriate

• a vertical stripe or artifact in the image of the patient's scan (effects all 16
detectors at once)

• Transverse Motion Failure message.

To specifically locate the source of an arc, it is necessary to find out which


high voltage power supply provides the excessive current. This may be done
by recording the power supply current monitors at TP1 and TP5 of the XORB
board with a storage oscilloscope. An arc typically draws 1.5 mA from the
power supply for about 1 ms. The trigger voltage should therefore be set at
1.5V for TP1 and for TP5.

Page 110 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


• Arcing can also occur inside the high voltage connectors. This will
usually result in a plainly visible black or brown carbon track through the
grease on the connector. In any case, once the high voltage connectors
have been removed from the Tube Head and the power supplies, they
should not be re-connected without first being cleaned and regreased
(see procedure DXSE0002 in the chapter 5 appendices).

• After cleaning the old grease off of the connectors, they should be
carefully inspected for carbon tracks. Look for these tracks on both the
rubber cable connectors and on the phenolic sockets of the Tube Head.
If such tracks are found after cleaning, the following are the options for
returning the scanner to service:

• If carbon tracks are found on the rubber cable connectors, they can be
removed by excising the damaged section with a sharp blade. Severe
tracks can burn quite deep into the rubber, so care must taken to remove
all the damaged rubber. After the carbon tracks have been removed,
additional grease must be used when repacking the connection to fill in
the volume of the removed rubber.

• If the above procedure requires too much rubber to be removed, or the


carbon track looks very severe, another option is to replace the high
voltage cable.

• Arc tracks will be impossible to see on the sockets in the Tube Head.
Therefore, if arc tracks are seen on the rubber cable connectors, the
sockets should be sanded with emory cloth as a precaution. After
sanding the socket, remember to flush the socket with cleaning solution
to remove any particles.

• If the tracking inside the socket is too severe or the carbon track cannot
be removed, another option is to replace the Tube Head.

Note: IT IS VERY IMPORTANT TO REMOVE ANY CARBON


TRACKS ON THESE SURFACES! Carbon tracks that are
simply covered by grease or not completely removed will still
provide a path for arcs to follow.

4.10.1 Limit Switch Tripped During Scan


If a Limit Switch is tripped during an install test, verify the Limit Switches'
location with the Service Software (Tools/Diagnostics/Scanner Motion). If this
problem occurs any time after install, see the problem description below.

If the transverse mechanics of the scanner become imprecise, it may cause a


Limit Switch to be tripped during a scan. This will usually only happen during a
total body scan or during an Alignment Test test since both of these scans
involve full-width scanning. Once the switch is closed, the motors stop, the
End-of-Exposure Alarm sounds, the amber SHUTTER OPEN lamp turns off
and after a few seconds a Diagnostic Failure appears on the screen.

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 111 of 137


• During full-width scanning, the Detector Carriage comes very close to the
Limit Switches, so any imprecision will cause a Limit Switch to be tripped
and the scan will be aborted with a Diagnostic Failure message for
Transverse Motion Failure.

A cause of this problem is a loosening of the first drive Reduction Belt which
connects the Transverse Motor to the first Reduction Pulley. This loosening
causes the belt to "walk" on the pulleys causing enough imprecision in the
motion to trip a switch. To tighten the belt, first remove the Pulley Shroud and
loosen all four nuts that hold the motor in place. Then, while holding the motor
such that the belt is pulled taught, tighten the nuts to secure the motor in
place. Replace the shroud and test the scanner (an Alignment Test test works
well).

If the Tube Head cables come into contact with the cable bundle entering the
cable track, it may be impossible for the scanner to complete all of the
necessary transverse steps away from the operator. Consequently, on each
scan line the detector will move closer to the front transverse limit switch, and
the switch may be eventually closed. Form the cable bundle exiting from the
cable track into an arch such that the Tube Head cables will move under the
arch rather than running into the bundle. This arch must not be too high. If it is,
there will not be sufficient slack in the Tube Head cable bundle to allow the
Tube Head to move to the front transverse limit switch. Also, if the arch is too
small, it may cause too much slack when the Tube Head is at the forward side
of the table. This causes the Tube Head cable bundle to rub against the inside
of the front panel causing a scraping noise to be heard. If the above does not
solve the problem then check the other mechanical components. Electrical
components that could cause such a failure are the Centent Motor Controller,
the Transverse Motor, the FOINK board or the SBC board.

4.11 Imaging Problems


4.11.1 White, or Grey in the first or second scan line:
The software is not perfect in its ability to determine the correct grey level of
the entire scan based on the first line of data. Sometimes the grey levels will
be set incorrectly during acquisition. During analysis the grey level is easily
adjusted to give a good image, and the results are not affected by this imaging
problem. This problem is most often found on very thin patients. Adding extra
tissue equivalent material (rice bag or saline solution) to a very thin patient
may be all that is necessary to cure the symptom.

4.11.2 Femur Scan Problems


Most image discontinuity problems occur in femur images performed at 3 mA
on thin or osteoporotic subjects. These lines usually occur in the trochanter
area where the x-ray beam is least attenuated by tissue.

Page 112 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


These lines are caused by Automatic Gain Stabilizer (AGS) trying to adjust
the input signal which is at an excessively high count rate. However, the AGS
is not at fault, and the correct action is for the customer to attenuate the x-ray
beam. Lunar normally recommends placing a rice bag (on its side to create 12
cm of patient thickness in the x-ray beam) along the side of the patient's thigh
for thin or osteoporotic patients.

4.11.3 AP-Spine Image Problems: Probable causes


• the shutter closes (causes the shutter open lamp to go out, and the End
of Exposure Alarm to sound).

• increase or reduction of high voltage to the detector (causes horizontal


artifacts in the image).

• loss of the signal from the detector (this effects all 16 detectors at once)

• loss of current to the x-ray insert filament (28 volt power supply error
message, the orange lamp to go out, and the End of Exposure Alarm to
sound).

• bad x-ray relay contacts (providing power to tube head power supplies)

• reduction in AC Line voltage

• loss of proper high voltage on the x-ray tube

• faulty operation of the AGS system (white lines are short and found only
where scan lines move from air into tissue).

• arcing (may cause vertical artifacts in the image)

4.11.4 Broken Signal Cable


The cables which carry the voltage to the detector and return the signal from
the detector array are part of the upper and lower cable bundle assemblies.
Any break or significant pinch of the cable can cause reduction in counts in
either or both channels. This could result in horizontal artifacts which effect all
16 detectors at once.

4.11.5 Loss of tube head current


The tube head control cable contains the wires which provide power to the
filament transformer. If the wire to the transformer center tap breaks, the x-
rays will stop, the orange lamp will go off, and the end-of-exposure alarm will
sound. If either of the other two wires break, the MAX board fuse will blow.
Always ask the operator if a "ping" sound (the end of exposure alarm) is
heard, and whether the Shutter Open or X-Ray On lamp turns off. In the case
of a tube head current problem, the X-Ray On light will turn off.

4.11.6 X-Ray Relay Failure


See section 4.16.

4.11.7 Unstable AC Line

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 113 of 137


If the customer's AC line conditions are unstable, voltages 11% below the
scaners AC input configuration will begin to drop the high voltage on the x-ray
tube. The counts will drop significantly with only a change of a few kV.
However, this normally causes black or dark grey lines in the scan image.

4.11.8 Arcing
If the tube head is arcing, the arcs will be more likely to occur at the higher
current settings. Arcing generally also causes other strange symptoms such
as:

• Scanner stops in the middle of a patient or QA scan with no error


message or with transverse mechanics fail

• The QA scan fails intermittently on bone chamber measurements

• The scanner periodically leaves the message on the screen "Starting X-


rays Please Wait" for a longer than normal period of time (more than 10
seconds)

If you see arcing symptoms, try the following service action:

1.View the Quality Assurance History (see section 4.15). In particular note the
trend in BM Values as discussed in the Failing QA's section of this chapter.

If there is evidence of failures in the mean BM value:

2.Repack the high voltage connections as described in the chapter 5


appendices. Note if an arc track is visible in the insulation compound.

3.Replacement of the tube head if no arc tracks are discovered, or if the


repacking of the connectors does not eliminate all arcing symptoms.

Summary

Except for the x-ray relay, the cause of the problem will be difficult to find. Try
to obtain some additional information. If the customer is hearing the end-of-
exposure alarm, or seeing the yellow x-ray on lamp flickering, you can deduce
the tube head control cable is bad. Likewise, it is possible to decide on the
shutter/collimator/fan cable. In the case of no alarm or lamp indications, or in
the case where large deviations occur in the peak, the upper and lower cable
bundles should be replaced, as problems with the coaxial cable will be too
expensive to troubleshoot.

4.12 Failing Alignment Test Results


4.12.1 Image
The Image is a graphic illustration of the counts recorded during the
Alignment Test. A good image has a black field with evenly distributed grey
dots and perhaps a few randomly distributed dots of various colors (green or
blue usually).

Page 114 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


Figure 4-6. Failing Ailigment Test

A poor Deviation Image will have lines, streaks or spots (indicating counts
higher or lower than expected). When the image is analized, click first on
points and then back to results, this will typically point type failing points (they
will have a yellow box around them).

• The information tab (see figure 4-2) will show the test outcome (pass /
fail).

• Lines or streaks going longitudinally in the image indicate an alignment


problem. Recheck the alignment of the scanner including the levelness of
the scanner. Also, check the scanner frame for any bending that may
have occurred at the site or during shipment.

• Lines or streaks going transversely across the Alignment Test image


indicate a problem with the AGS, on the Detector Daughter Board in the
detector array. Again, replace parts until a passing Alignment Test is
obtained.

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 115 of 137


• Troubleshooting tip: if the Alignment test can be run with the Alignment
Aperture on, this will make the test more sensitive as the beam will be
further collimated to the exact size of the detector. Any failures (beam
straying off the detector) will be exaggerated.

4.13 Indicator Failures


4.13.1 X-ray On LED Blinking
The amber X-RAY ON LED should glow steadily once it is illuminated, until
the x-rays are turned off. If the X-RAY ON LED blinks while the shutter is open
for a patient scan, the End of Exposure Alarm should sound.

The amber LED is controlled by a comparator on the FOINK Board. This LED
illuminates when approximately 40µA (or greater) of current is passing
through the X-ray Insert. On Spellman system scanners the amber light may
appear to fade out slowly when x-rays are ramped down. This due to the
Spellman power supplies bleeding off current and is not a problem.

If the voltage ramping fails, the High Voltage Power Supplies must be shut
completely off before a second ramping attempt is made. If the amber X-RAY
ON LED turns on momentarily, then turns off for a few seconds before coming
on steadily, the system is having difficulty ramping the high voltage. One of
the High Voltage Power Supplies may be defective.

The SBC does not monitor the Insert current once the proper level has been
reached. Therefore, if the current to the Insert is interrupted, the SBC will not
recognize the problem and will not alert the computer, or terminate the
exposure.

If the X-RAY ON LED blinks once the x-ray high voltage has been set, there
must be a problem in one of the following areas:

• Tube Head Control Cable-The most likely cause of an intermittent


problem in this circuit is a broken wire in the Tube Head Control Cable.
The Red wire is most likely the broken one, as a broken blue or black
wire should blow the MAX Board Fuse. Turn the scanner off and test for
continuity between MAX Board test points TP4, TP5, and TP13.

• MAX Board-The MAX Board may be operating intermittently. Substitute a


replacement MAX Board.

4.13.2 Shutter Open LED Blinking


The amber Shutter Open LED is controlled by the Limit Switch on the Shutter/
Collimator Assembly. If the Limit Switch is defective or improperly adjusted,
the Shutter Open LED may be switched on and off. This will usually be
accompanied by the sound of the End of Exposure Alarm if an exposure is
underway.

Page 116 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


If the Shutter is actually opening and closing intermittently and uncontrollably,
the problem is either on the FOINK, or a broken wire between the FOINK and
the Shutter/Collimator Assembly. Remove connector J11 from the FOINK
Board and measure the resistance between pin 5 and pin 6. This will be a
measurement of the resistance of the Shutter Solenoid and the cable through
the Cable Track. If the problem occurs on every scan, run a scan while
making this continuity check. If the cable and solenoid seem to be working
properly, substitute a new FOINK Board.

4.13.3 Shutter Not Operating


If the Shutter Solenoid will not open, the problem may be caused by one of the
following:

• The Shutter Open LED-If the Shutter Open LED fails, a properly
operating FOINK board will prevent operation of the Shutter Solenoid.
Check the LED for continuity, replace it, or substitute a working amber
LED temporarily.

• Solenoid Cable-Remove J11 from the FOINK Board and check continuity
between pin 5 and 6 (see Shutter Open LED Blinking above).

• FOINK Board-Replace the FOINK Board.

4.13.4 End of Exposure Alarm During Scan


If the sharp "Ping" sound of the End of Exposure Alarm is heard during a
scan, look at the computer display to see if a Diagnostic Failure Code is being
reported. If so, see section 4.1.

If no Diagnostic Failure Code is reported, the scanner may still be in motion


continuing with the scan. Abort the scan and remove the patient from the
table. In any case, note the status of the SHUTTER OPEN light and the X-
RAY ON light.

• Both the SHUTTER OPEN light and the X-RAY ON light are on and
steady. This would indicate a faulty FOINK board. Replace it

The SHUTTER OPEN light is off. This could indicate the following:

• The LED becomes defective during the scan and since the shutter
solenoid and this light are wired in series, the Shutter closed and the
Alarm sounded.

• The cable to the amber Shutter Open light broke during the scan with the
same result as above.

• The shutter solenoid failed and the Shutter closed followed by the Alarm
sounding.

• The FOINK is faulty.

• The SBC is faulty.

The X-RAY ON light is off. This could indicate the following:

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 117 of 137


• X-ray production has halted. This turns off the X-ray On light and sounds
the Alarm.

4.14 Communications Failures


Should the scanner, host computer and or Detector Mother Board lose
communications with each other, there are a number of things to check:

A good troubleshooting tool is the COMM test in Tools/Service Options -


COMM tab.

Check

• the I/O cable connections at the serial port on the back of the Host
computer and at the cSBC. Be sure both connections are tight and that
the thumb screws are used to hold the connectors firmly together.

• inspect the connector on the cSBC. This connector is very fragile and
may have been damaged. Be very careful when connecting the I/O
cables to the SBC.

• Check and inspect the Signal Cable input at the cSBC

• Check the Connection at the Detector Mother Board and the Bulkhead,
where the two halves of the signal cable meet.

• Verify the power LED’s on the cSBC are lit (See section 4.19)

If the ports are properly configured and all power is present, and
communication still fails, then the probable causes are:

• A faulty I/O cable.

• A faulty cSBC.

• A faulty serial port / serial port card.

Replacement of these parts is the best troubleshooting method.

4.15 Viewing Quality Assurance Trends


DPX-NT software includes the capability to view, print or graph data contained
in the Quality Assurance database. View and print display selected results of
many of the tests run during the Daily Quality Assurance in tabular form. The
graph feature will graph the results of only one of the Quality Assurance
parameters. All features allow for the user to select the time period to be
examined.

4.15.1 What to Look for in the QA History

Page 118 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


All categories should remain steady over time. Check the PASSED column for
failed QA’s. If a failing QA is found, try to determine the cause, including
operator error.

The limit for % Spillover is 13%.

If any sudden jumps are noticed in the values of any of the categories,
determine first if these are due to service work such as a Tube Head
replacement. If service work causes a shift in the QA values, then compare
present values to those obtained only after the service.

4.16 MAX Board Troubleshooting


The MAX board controls the current through the X-ray Insert in the Tube
Head. DPX-NT scanners use MAX version 02B currently.

The current at a given kV across the insert is a function of insert filament


temperature. The MAX board adjusts the current by regulating the filament
temperature to provide the proper current as set by the SBC. The relationship
is Insert µA X 1000 = SBC input (in volts DC), so for an SBC input of 5VDC,
Insert µA = 5000.

The Positive High Voltage Power Supply has a current monitor output (mA
MONITOR); the voltage at this output is proportional to the current passing
through the X-ray Insert. This monitor voltage is connected to the MAX board
as feedback (mA FEEDBACK).

What to look for

• Make sure that all connectors are fully seated.

• Check Lemo connectors for loose parts.

• Check for loose wires on the insulation displacement connectors.

• Check the position of the shorting jumpers on pin headers JB1, JB2, and
JB3 for: Be sure pins 1 and 2 are connected as indicated by silk screen.

When the X-ray Relay is on, both the red and green LED's on the MAX board
should be illuminated. If both are out, no power is getting to the MAX board. If
only the red LED is lit, then the MAX board fuse is burned out or missing.

Electrical tests

When the board is powered up, and x-rays are being generated, the following
measurements may be performed:

With a VOM set on 200 VDC, you should measure about 18 to 21 VDC on
TP's 4, 13 and 5 referenced to TP3 (GND).

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 119 of 137


Note: If an oscilloscope is available, you should see about 18 to 21
VDC on TP13. However, TP 4 & 5 should have a 36 to 42 volt
peak square wave. A VOM will show the average value of the
square wave, i.e. 18 to 21 VDC.

4.17 OMI Board Troubleshooting


The OMI board, being a rather uncomplicated device, has little in the way of
things to watch for in installation and use. There are three such boards, one
on the Front Longitudinal Carriage (transverse), at the foot of the scanner on
the longitudinal idler shaft (longitudinal), and on the collimator (shutter open
detect).

What to look for

• Check the connector to each board for proper seating.

• Examine the area where the wires attach to the board itself; it is possible
the may fray or break.

• Insure that the interrupter disk does not rub against the plastic detector
housing. If it does, adjust by loosening the two machine screws that
attach the plastic detector, reposition the detector and re-tighten the
screws.

Electrical tests

The signal from the OMI board is nominally rectangular in shape. The signal is
generated when the Interrupter Disk alternately passes and occludes a light
beam between the two posts on the photo-transducer. With the beam
occluded by the disk, the output signal should be about 5 VDC (close counts).
When the beam is allowed to pass through a slot in the disk, the signal output
is near ground potential (see FOINK troubleshooting 4.17).

Symptoms of possible OMI failure

• Scanner shuts down at the beginning of a patient scan.

• Scanner shuts down unexpectedly during a scan.

• Either condition above should be accompanied by a Diagnostic Failure


Transverse or Longitudinal (see section 4.0).

4.18 SBC Troubleshooting


The SBC consists of a single circuit board that uses the 8032 microprocessor
with a RS-232 interface to the Host PC

The SBC is located on the electronics pan in the front center.

Page 120 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


What to look for

• Green LED (C in figure 4-5) is on then you have +5VDC to the board.

• Red LED (A in figure 4-5) is on then you have +12VDC to the board.

• Amber LED (B in figure 4-5) is on then you have -12VDC to the board.

• Red LED (E in figure 4-5) is the DTR reset line, when this LED is lit, the
Detector, FOINK, or SBC are detecting an error.

When the DTR Reset Line is lit:

• Check the Signal Cable

Go to Tools - Service Options - Communications Tab - test


communications

• Red LED (D in figure 4-5) the scanner Reset, when this LED is lit, the
communications with the Host PC are in Reset.

• Verify port settings on the Host PC (See DXPC 2000, chapter 5


Appendices)

• Verify communications cable

• Green LEDs (G and H in figure 4-5) flash when data is being received by
the SBC

• Green LED (F in figure 4-5) is not assigned a function at this time.

The reset button near the communications cable connector causes a reset
signal at the SBC microprocessor. Also, disconnecting the serial cable will
cause a reset state until the cable is re-attached and communications
reinstated.

4.19 XORB Troubleshooting


The XORB board routes a number of important signals to the rest of the
system and protects the system from transients by shunting them to ground
with transorbs.

What to look for

• All connectors firmly seated.

• Check to see that all wires and cables are firmly affixed to their
respective connectors. Pay particular attention to Lemo connectors,
since the jam nuts on the connectors are prone to be loose.

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 121 of 137


• Check and tighten, if necessary, all mounting screws for the XORB
board.

Electrical Tests

• Measure impedance between XORB ground plane and the electrical


pan. It should be less than an ohm.

• There should be no low impedance (less than 10 ohms) reading between


any signal or signal ground and chassis ground on the XORB when the
as-associated signal connectors are disconnected.

The following test points should have the following readings:

Table 4-11. XORB Test Points

Test Point kV mA Limits


TP 1 76 .750 0.74 - 0.76
TP 5 76 .750 0.74 - 0.76
TP 1 76 3.00 2.97 - 3.03
TP 5 76 3.00 2.97 - 3.03
TP 2 76 .750 3.76 - 3.84
TP 6 76 .750 3.76 - 3.84

Page 122 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


5 Service Procedures

This chapter contains procedures for testing and replacing


parts.

5.0 Reloading LUNAR Software


5.1 Peaking the Detector
5.1.1 Peaking Procedure
5.2 Tube Head Replacement
5.2.1 Shelf Life
5.2.2 Mounting The Tube Head
5.3 Lower Cable Bundle Replacement
5.3.1 Procedure
5.4 Upper Cable Bundle Replacement
5.4.1 Procedure
5.5 Tube Head Control Cable Replacement
5.5.1 Procedure
5.6 Tests to Perform after Service

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 123 of 137


This page left blank intentionally.

Page 124 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


5.0 Reloading LUNAR Software
It is sometimes recommended in answer to a particular problem that the
software be reloaded. This can be accomplished by locating the latest version
of the Lunar Software. Please note, it may also be necessary to have the
customers system ID number and feature code.

Note: BE SURE OF WHICH VERSION OF SOFTWARE IS


PRESENTLY BEING USED! RELOADING OLDER
SOFTWARE CAN DESTROY IMPORTANT DATA.

• The about tab under the help menu of the Lunar DPX-NT Software
shows which version is presently on the system.

• Insert the Lunar System CDROM into the CDROM: drive

• Double click on the My Computer icon on the Windows NT desktop

• Double click on the CD ROM Drive with the LUNAR DPX-NT CDROM in
the drive

• Double click the Software Installation ICON

• Follow the instructions on the screen until a message is displayed


confirming a successful software load.

Note: When new versions of software are released they may not be
loaded in the above manner. BE SURE TO FOLLOW THE
LOADING INSTRUCTIONS THAT ACCOMPANY ALL
SOFTWARE UPDATES.

5.1 Peaking the Detector


The DPX-NT Detector may need to be repeaked, this is a good diagnostic tool
to use when a detector is suspect.

The actual peak setting of the detector is done in the software.

5.1.1 Peaking Procedure


1) Go to the Service Mode of the DPX-NT Software, open the Spectrum
Option (Tools/Spectrum)

2) Place the brass spillover piece on top of the collimator aperture.

3) Click on the Peak button in the DPX-NT Software toolbar.

4) Verify that all values are 2000 +/- 600.

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 125 of 137


5.2 Tube Head Replacement
When a replacement Tube Head is sent to a site in the field for replacement, it
is not ready for immediate installation into the scanner. Some parts from the
old Tube Head must be transferred to the new Tube Head (see list below and
figure 5-1).

These parts are: 1) tilt bar

2) lead ring (around filter area)

3) lead wrap

4) fan bracket

5) nuts, bolts and washers

6) collimator positioning hardware

7) the collimator assembly.

Figure 5-7. DPX Series Tube Head components

Verify that all necessary parts have been transferred to the new Tube Head
before shipping the defective. LUNAR has received defectives with the lead
ring, for example, still in them.

Page 126 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


5.2.1 Shelf Life
Many distributors have asked about the shelf life on the Tube Heads in their
stock. The shelf life of the Tube Head is not accurately known, but we feel at
this time that it is over two years. Perform the following Tube Head
inspections on all replacements before mounting them:

1.Check the shelf for oil spots which would indicate that the Tube Head has a
leak. If oil is discovered, return the Tube Head for warranty replacement.

2.Shake the Tube Head and listen for the sound of an air pocket or bubble
inside. If a bubble is discovered, return the Tube Head for warranty
replacement.

Note: If a Tube Head has been on the shelf for three weeks or more,
perform the warm up procedure described in Procedure
DXAP2001 DPX-NT Source Detector Alignment.

5.2.2 Mounting The Tube Head


Use procedure DXSE0000 in this chapter's appendix for repacking the new
Tube Head's high voltage connections, and refer to the installation
procedures, DXAP2001 and DXAP2002 (also in this chapter's appendix), for
tube head mounting and alignment procedures.

5.3 Lower Cable Bundle Replacement


Necessary tools:

• 1/8" and 5/32" Hex Wrench,

• Small Screwdriver

• Diagonal Cutters

• 3/8" Wrench

• Large and small wire ties.

5.3.1 Procedure
1.Shut off the power to the scanner.

2.Take off the table top by removing four screws (two at each end).

3.Remove the head and foot end panels. At the foot end of the scan table
unplug the A/C power cable to the scanner.

4.Move the scan arm to the head end of the table.

5. On your new lower cable bundle, use a diagonal cutters to cut the wire ties
holding the cable together, also unfold and read the LUNAR SERVICE
NOTES attached to the new cable bundle.

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 127 of 137


6.Lay the new cable bundle flat out on the floor. Locate the end with the (2) 16
pin connectors and the yellow and orange Lemo connectors, this is the end
that connects to the bulkhead, this is the end to start the install action from.

7.Locate the bulkhead on the scanner. There are (3) wire ties that hold the
cable bundle in place between the bulkhead and the cable track. Before
cutting any wire ties note the way the cable bundle is tied down and the way
the cable lays before it enters the cable track. The new cable must be
fastened in the same position and lay the same way.

8.Carefully cut the (3) wire ties between the bulkhead and the cable track with
your diagonal cutters. Disconnect the large RS232 connector (Detector Signal
Cable) by unscrewing the two screws on the connector.

9.Next disconnect the two white 16 pin molex connectors. These connectors
have a catch lock located in the center of the housing which must be
depressed to disconnect them from the bulkhead.

10.If you try to remove the cable bundle by pulling it back towards the cable
track you will notice a 3 conductor gray wire cable tied to the high voltage
cable bundle. This wire carries the transverse optical interrupt signal from the
front carriage (we will refer to this wire as the OMI cable). To find the location
of the OMI board and the front carriage follow the OMI cable. On your new
Cable bundle the OMI wire is connected through the bulkhead.

11.To remove the OMI cable push the x-ray tube head all the way to the
outside of the scanner (patients left side), then notice the clearance between
the front panel and the high voltage cable bundle connecting to the x-ray tube
head and the way the cables are tied down as you will need to wire tie them
back the same way.

12.Now you can cut the (4) large wire ties and the (2) small wire ties that hold
down the high voltage cable bundle and OMI cable. Then you can pull the
lower cable bundle back to where it enters the cable track.

13.Next take your small screw driver and pry open (20) of the cable track
plastic retaining clips that hold the cables into the track. These clips pivot at
the back of the cable track and can either be removed or left in the open
position so the cables can be removed.

14.Pull the old cable bundle out of the cable track towards the foot end of the
table.

15.Now install the (3) conductor OMI cable from the bulkhead to the OMI
BOARD. To do this you must remove the white 16 pin connector from the
bulkhead labeled bulkhead 2, this connector is a part of the upper cable
bundle and the new OMI cable will connect into this housing. Note the position
of the wire tie that holds the upper cable bundle in place and then cut it in
order to pull the connector out to install the OMI cable. Note the orientation of
the existing terminals in this bulkhead connector as they slide and lock into
the connector in one direction only. Where the wires enter the connector each
hole is numerically labeled. Put the green wire into hole # 11, and the black
wire into hole # 9, and the red wire into hole # 1.

Page 128 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


16.Slide the connector back into the bulkhead bracket the way it came out and
fasten the upper cable bundle with a large wire tie the way it was before the
original wire tie was cut.

17.Route the new OMI cable in the same position as the previous cable and
connect it to the OMI board. Then tie the high voltage cable bundle and the
OMI cable down to the high voltage cable trough,(there are three large wire
ties near the bulkhead, two large wire ties put together to tie the high voltage
cables down at the center of the high voltage cable trough, and two small wire
ties to anchor the OMI cable down to the Bracket before it connects to the
OMI board). Before tightening the wire ties assure that the proper clearances
and no stress is applied at any of the connectors or wires. Tighten the center
cable ties first (two large wire ties put together and located at the center of the
high voltage cable trough). Be sure the high voltage cable bundle does not hit
the front panel when the tube head is all the way toward the outside of the
scanner (patients left). Then tighten the three wire ties near the bulkhead
making sure that the high voltage cables do not hang over into the cut out
space on the high voltage cable trough as they may hit the longitudinal motor
bracket and cause positioning problems.

18.Take the new cable bundle and connect it to the bulkhead by routing the
cable bundle the same way as the replace cable bundle, (note the connectors
are labeled bulkhead 1 and bulkhead 2) and fasten the cable bundle down in
the three spots between the cable track and the bulkhead with wire ties. Lay
the new bundle down into the cable track making sure all the cables lay
parallel to each other. Start replacing the clips to the cable track as far as you
can. Continue the process: remove a few clips, extract the old cable bundle,
add in the new cable bundle, and replace the clips.

19.To finish removing the old and installing the new cable bundle in the lower
portion of the cable track you must remove a screw located at the end of the
cable track. To do this use a 1/8" hex wrench and a 3/8" wrench. Once you
remove the screw you can lift the cable track out of the channel it rides in and
finish prying open the remaining clips. Remove the old cable bundle until it
reaches the cable entrance hole, where all the cables enter the electrical pan
area. Then finish installing the new cable bundle until you reach the cable
entrance hole, finish clipping all the clips and then replace the cable track hold
down screw (do not tighten the screw at this time).

20.Locate where all the wires from the old cable bundle terminate on the
electronics pan. All of the wires are labeled except for the wires that go to the
terminal block. Here is a list of wires that must be disconnected before
removing the cable bundle from the pan area. cSBC(J-5), FOINK(J-17),
FOINK(J-12), FOINK(J-14), FOINK(J-5), FOINK(J-8), PMT/AGS(J-11)(J-11 is
the yellow Lemo connector), the orange Lemo connector (connected to the
H.V. Lemo box), then the last three wires connect to the Terminal block. Write
down the number where the orange, blue and yellow wires go to on the
terminal block and disconnect them with your small screwdriver. Carefully pull
the old cable bundle through the cable entrance hole from the pan area. The
old cable bundle may now be discarded.

21.Carefully feed (one connector at a time) the new cable bundle wires
through the cable entrance hole into the pan area. Connect the wires to their
proper locations.

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 129 of 137


22.Then move the arm all the way down to the foot end of the scan table. At
this time make sure that the Lower cable bundle and the high voltage cable
bundle have some slack where the cables enter the cable track near the
bulkhead area. Then make sure the cable track has at least 1/8" of clearance
from hitting the frame. To adjust this slide the cable track in the trough it rides
in to get the proper clearance, then tighten the screw that holds the cable
track in place with a 1/8" hex wrench and a 3/8" wrench. Move the arm up and
down the scan table to be sure there is nothing wrong with the cable track
motion. Leave the arm at the foot end of the table.

23.Leave a little slack in the cables from the cable track to the pan area to
allow all of the cables to be routed and wire tied exactly like the original
cables.

24.Reconnect the A/C power cord, put the table top, pad, and end panels
back on the scanner.

25.Turn on the scan table, computer, monitor and printer. Perform required
tests per section 6.9 of the DPX-NT Service Manual.

5.4 Upper Cable Bundle Replacement


Tools Necessary:

• Small Phillips Screwdriver

• Small Slot Screwdriver

• Diagonal Cutters

• Hex Wrench Set

• Mini Molex Pin Extractor

• Large and Small Wire Ties.

5.4.1 Procedure
1. Turn off the power to the scan table.

2. Remove the table top by removing the two screws at each end.

3. Remove the front panel of the scan table

4. Lastly, remove the column cover of the arm held in place by screws.

5. Remove the upper and lower arm covers by removing the (8) screws that
hold it in place.

6.Open the cable track and cut the wire ties at the end of it. Disconnect all the
cables coming from the cable track and remove the cable bundle as far as the
top of the column (through the hole).

Page 130 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


7.Place the new cable bundle in the cable track and connect all cables to their
appropriate positions. Close the cable track and wire tie the cables at the end
of it. Route the cable bundle through the hole in the top of the column.

8.Next locate the Laser On light and disconnect it. Remove this cable to the
top of the column cutting any necessary wire ties.

9.Route the new cable to the Laser On light and connect. Replace any wire
ties removed in step 5.

10.From the back of the arm column follow the cables vertically down and cut
all the wire ties holding the upper cable bundle to the other cable assemblies.

11.Carefully pull the upper cable bundle out towards the back of the arm and
down through the holes in the column.

12.Next cut the wire ties that hold the cables down near the bulkhead, these
are located in the lower rear portion of the arm.

13.Make sure the bulkhead connectors are labeled on both sides of the
bulkhead bracket and then disconnect them.

14.Next remove the two (16) pin Molex housing connectors that are attached
to the upper cable assemblies from the bulkhead bracket. Also disconnect the
LEMO connector from the bulkhead bracket. Pull the housings and the Signal
cable connectors out to work on them.

15.Use the mini Molex pin extractor tool and remove from bulkhead 1 pin16
and pin8. From bulkhead 2 (the lower connector) remove pin15, pin16, pin6,
pin7, and pin8.

16.Now remove the old cable bundle by pulling the cable out through the back
of the column. The old cable may now be discarded.

17.Next route the new cable bundle through the column wire tieing as you go.

18.Route the taped end of the cable through the hole at the bottom of the
column to the bulkhead area (follow the other cables that go to the bulkhead).

19.Carefully remove the masking tape from the cable exposing the mini Molex
pins and the Lemo connector. Note the orientation of the existing terminals in
the bulkhead connectors and the numeric labels at each hole of the
connector.

20.Replace the wires one at a time removing the wires from the old cable and
replace them with the wires from the new cable.

21.Insert the connectors into the bulkhead bracket the same way they were
removed and fasten the upper cable bundle with wire ties as original.

22.From the wire tie just fastened, the cables going into the arm column must
not have any slack in the cables (pull the cables from the column area to
remove any slack). Make sure that none of the cables hit any of the moving
parts in the narrow area where the cables come from the back of the column
out to the bulkhead.

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 131 of 137


23.Mount the column cover back in place and replace the top and bottom arm
covers. Replace the table top, table pad and the head and foot end panels
back on to the scanner.

24.Turn on the power to the scan table. Run Required tests as prescribed in
section 6.9 of the DPX-NT service manual.

5.5 Tube Head Control Cable Replacement


Required Tools:

5.5.1 Procedure
1.Switch off the power to the scanner.

2. Remove the tabletop, head and foot end panels. At the foot end of the scan
table unplug the A/C power plug to the scanner.

3.Move the scan arm to the head end of the table.

4.Take the black tape off the high voltage cable bundle at each end of the
sleeving (open any cable track clips that are in the way as in step 8) as well as
any wire ties holding the bundle in place. Next, slide the sleeving away from
the tube head and remove the tape beneath the sleeving. Note the positions
of each cable before removing the tape. The cables must be returned to these
positions.

5.Disconnect the military-style connector and the ground wire (if one exists)
from the tube head and slide the cable out of the black sleeving toward the
cable track.

6.Unplug J3 and J5 on the XORB board. Also, trace the ground wire (if one
exists) from the control cable and disconnect it from the power supply. Cut the
wire ties holding the control cable to the pan.

7.Pry open the plastic retaining clips that hold the cables into the cable track.
These clips pivot at the back of the cable track and can either be removed or
left in the open position to remove the cables. The screw located at the end of
the cable track near the pan must also be removed. Once this screw is
removed the cable track can be lifted out of the channel it rides in and the
remaining clips can be opened. Mark the position of the cable track screw on
the frame to return the track to its original position.

8.Remove the old control cable. Note the position of all the cables in the cable
track. They will need to be returned to the same position during this
procedure.

If the cables are not returned to their


original positions, cable damage will
result.

Page 132 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


9.Feed the new cable, military connector first, through the black sleeving on
the high voltage cable bundle starting near the bulkhead. The sleeving can be
expanded by pushing two parts of it together to better fit the connector
through.

10.Once the cable is through the sleeving, connect it to the tube head and
connect the ground wire (see figures 1 and 2). Position the cables next to the
high voltage cable bracket as in figure 1. The collimator control cable should
come out of the cable bundle between the two high voltage cables toward the
back of the scanner. Tape the cables together for several inches with three
layers of electrical tape starting as close to the bracket as possible. Continue
to spiral the tape around the cables as it was originally and finish by taping
three layers for several inches near the bulkhead. Tape the sleeving near the
bulkhead as it was originally. Pull the sleeving tight along the bundle and tape
near the tube head. Wire tie the cables to the tube head high voltage cable
bracket.

Figure 5-8. Profile Cable Routing Diagram

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 133 of 137


Figure 5-9. Top Down Cable Routing Diagram

11.Lay the control cable in the cable track in the same position as the original
one. Start replacing wire ties between the bulkhead and the tube head. Before
tightening the wire ties assure that there is proper clearance and no stress is
applied at any of the connectors or wires through the entire range of motion.
Tighten the center cable ties first (two large wire ties put together and located
at the center of the high voltage cable trough). Be sure the high voltage cable
bundle does not hit the front panel when the tube head is all the way toward
the front of the scanner. Make sure the high voltage cables do not hang into
the cut out space on the high voltage cable trough. Tighten the three wire ties
near the bulkhead.

12.Replace the clips to the cable track and reattach the track to the scanner
frame.

13.Move the arm to the foot end of the scan table. Check that the lower cable
bundle and the high voltage cable bundle have some slack where the cables
enter the cable track near the bulkhead. Check that the cable track has at
least 1/8" clearance of the frame. To adjust this, slide the track in its trough,
then tighten the screw that holds the track in place. Move the arm up and
down the table to verify that the cable track moves properly.

14.Reconnect the A/C power cord, and replace the table top and end panels.

15.Perform required tests as prescribed in section 6.9 of the DPX-NT Service


Manual.

Page 134 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


5.6 Tests to Perform after Service
Table 5-12. Tests to Be Performed after Service

Part Replaced or Secondary


Primary Phantom Alignment
Procedure Calibration /
Calibration Scans Test
Performed QA

+5DC Power Supply - 1 - -


Detector 1 5 3 1
Centent - 5 3 1
Collimator - 5 3 -
Laser - 1 3 -
Lower Cable Bundle - 5 3 1
MAX, AMP or XORB
- 5 3 1
Board
Motor, Belt, Pulley
- 5 3 1
or Gears
High Voltage Power
1 5 3 1
Supply
Tube Head 1 5 3 1
cSBC - 5 3 1
Tube Head Control
- 5 3 1
Cable
Upper Cable Bundle - 5 3 1
System Move or
1 10 3 1
Installation

5.7 Returning Parts To LUNAR (For International


Distributors)
Return parts prepaid to Lunar. Please list our customs broker as intermediate
consignee:

Burlington Air Express

6023 S. Howell Ave.

Milwaukee, WI 53237-0140

DPX-NT / MD+Service
DPX-NT/MD+ ServiceManual
Manual(Rev.
(Rev.11.1)
- 2009) Page 141
135 of 142
137
This page left blank intentionally.

Page 136 of 137 DPX-NT/MD+ Service Manual (Rev. 1 - 2009)


Appendix

Associated Service Documentation

DXAP3000 - DPX-NT/MD+ Installation Procedure


DXAP3001 - DPX-NT/MD+ Alignment Procedure
DXAP4009 - DPX-NT/Bravo/Duo Beam Wobble Procedure
DXAP5006 - DXA Installation Report Form
DXQC2022 - enCORE Software Aluminum Spine Phantom Scan Procedure
DXAP0010 - DPX-IQ and NT/MD+ Arm Slot Cover Installation Procedure
DXAP0012 - DEXA External Options Installation Procedure
DXSE0000 - High Voltage Repacking Procedure
DXSE3001 - Prodigy and DPX Series Laser Replacement Procedure
DXAP2005 - enCORE Operator Training Outline
DXPC2000 - DPX-Fan/DPX-NT/MD+ PC Configuration Procedure
DXAP3030 - Small Room Kit Installation Instructions
DXAP2010 - enCORE Table Pad Alignment Procedure
DXRA3200 - DPX-NT/MD+ (B) Certified Components
DXAP2023 - DXA U.S. Label Installation Instructions
DXSE0002 - DEXA Label Replacement
DXAP2030 - Teledensitometry Configuration Procedure
DXAP2031 - Importing DPX Database Files into enCORE
DXAP3030 - Small Room Kit Installation Instructions
DXPC2004 - Prodigy/DPX-NT/MD+ Windows XP PC configuration Procedure
DXSE3004 - enCORE Computer Swap Procedure
DXAP0033 - Seismic Installation Procedure
DXSE0012 - DEXA PM Work Instructions
DXSE0038 - DEXA PM Quality Record
Non-USA service personnel: Please omit all steps in the PM Work
Instructions that involve InSite 2.0 until further notice.
DXSE0040 enCORE Based Scanner Transit Tear Down Procedure
DXSE0041 enCORE Series De-Installation Checklist
DXSE0042 OMI/OMD Replacement Procedure
DXSE3005 WinXP Computer Replacement Work Instructions
DXAP5033 InSite 2.0 Installation Check Sheet (pdf)
USA ONLY.

DPX-NT/MD+ Service Manual (Rev. 1 - 2009) Page 137 of 137

You might also like