You are on page 1of 13

HEALTHCARE Chapter 11

Imaging Services Installation Planning


Document No: DD+DIS148.05E

DRYSTAR 5302
Type 5366/100

Edition 1, Revision 2

This documentation is separately available. Order No: DD+DIS148.05E.

CONFIDENTIALITY NOTE:
Use, dissemination, distribution or reproduction of this document by unauthorized personnel is not permitted and may be unlawful.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

12-2007 printed in Germany Document Node ID: 15602908


eq_11_install-planning_e_template_v04
Agfa Company Confidential Copyright © 2007 Agfa HealthCare N.V.
1
1 2 3
DD+DIS148.05E Installation Planning

2 System Overview

536611JK.CDR

Figure 1

Device Comment
Modality The modality has to support the DICOM print protocol
DRYSTAR 5302 Dry printer for 14x17", 14x14", 14x11", 10x12", 8x10" DT2
blue base or clear base films

NOTE:
Remark to an installation in a 'HIPAA conform environment':
In Figure 1 two additional, optional modules are not listed:
1) Audit logging server: This is a computer in the networking allowing to store all
communication between the different devices in the network.
2) NTP (Network Time Protocol) Server: This is a computer providing time
synchronization. It is useful for audit logging, so that all messages of all
devices have the same time reference.
Both devices are only required in a 'HIPAA' environment, where high data security is
required.
For more information refer to section 6.3.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 DRYSTAR 5302 Chapter 11 / 7


12-2007 5366/100 Agfa Company Confidential
1 2 3
DD+DIS148.05E Installation Planning

3 Scope of Delivery and Accessories

3.1 Scope of delivery


NOTE:
The printer is delivered with all parts, which are required to install it in an existing
Ethernet network.

Device name: DRYSTAR 5302


Type: 5366/100
ABC Code EQOZV

Scope of delivery
x 1 Basic DRYSTAR 5302 including:
- 2 input trays adjustable for all film sizes
- 1 output tray
x 1 Compact Flash card (stored in backup CF card slot in input tray)
x 2 power cables, 3 m long (Europe / US)
x 1 UTP network cable, 5 m long (2 x RJ45 connection) incl. extension piece
x 1 Reference manual (English)
x 3 User Manuals (English; Chinese simplified, Spanish)
x 1 CD with all user manuals and Plug&Play installation manuals in all languages*
x 1 packing list
x 1 Plug and Play Installation manual (English)
x 25 films 14x17" DT 2B for test purposes
x 25 films 8x10" DT 2B for test purposes
x 2 sets film type labels
x 1 Chinese warning sticker
x 1 CE declaration of conformity
x 1 China ROHS declaration
x 1 Installation Report
*available languages: Chinese simplified, Chinese traditional, Czech, Danish, Dutch, English, Estonian,
Finnish, French, German, Greek, Hungarian, Italian, Japanese, Korean, Latvian, Lithuanian, Norwegian,
Polish, Portuguese, Russian, Slovakian, Slovenian, Spanish, Swedish, Thai, Turkish, Vietnamese

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 DRYSTAR 5302 Chapter 11 / 8


12-2007 5366/100 Agfa Company Confidential
1 2 3
DD+DIS148.05E Installation Planning

IMPORTANT:
The printer is delivered with two 3 m power cables. See Figure 9.
x For US: NEMA 5-15P with IEC C13 plug (female)
x For Europe: CEE7/7 with IEC C13 plug (female)
Organize a power cable for your country locally.
See section 8 for “Mains connection in the installation room” specifications.

3.2 Options

Item Order Number Comment


Mobile Kit EX2DV To be used in earthquake areas and in mobile
usage. May not be used in military vessels.

Figure 2

NOTE:
REMOTE SERVICE:
In general it is possible to diagnose and repair the printer to a certain extent remotely.
To enable remote service of the printer remote access has first to be organized.
Remote Access is established via the Agfa own Secure Remote Service System
(SRSS).
For more information to the SRSS refer to MEDNET – GSO Library - Technical
Services - Service Delivery - SRSS

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 DRYSTAR 5302 Chapter 11 / 9


12-2007 5366/100 Agfa Company Confidential
1 2 3
DD+DIS148.05E Installation Planning

4 Machine dimensions

Closed machine:

67,6 cm
.2 m 26.6"
28 5 c
"
.
71

72,
8
28. cm 536611jo.cdr
7"

Figure 3

Opened machine:

76 cm
29,9"
,7 m
45 6 c
"
11

536611jn.cdr

536611jp.cdr

Figure 4 Figure 5

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 DRYSTAR 5302 Chapter 11 / 10


12-2007 5366/100 Agfa Company Confidential
1 2 3
DD+DIS148.05E Installation Planning

5 Minimum free space for operation and maintenance

23 m m
60
0

. 6"
23 mm

0
60
.6"

23 mm
80
0

.6"
31 mm

0
.5"

54
23 m m
60
0
.6"

23 mm
0
60

.6"

536611JL.cdr

Space required for servicing and installation Space required for normal use

Figure 6

NOTE:
In case the minimum free space for installation and repair is not available, the printer
can be carried at a different place. The printer weight is 90 kg (198 lb). It can be
carried by three people.

Caution:
The device is heavy (90 kg;198 lb). Risk of injuries.
x At least three people have to lift the device from one position to the other.
x Use proper foot and hand protection when lifting the device.
x The table might break and cause injuries or damage to the device. The device
weight may be too heavy for the table.
x Communicate to the customer, that the table must be able to carry more
than 90 kg (198 lb)

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 DRYSTAR 5302 Chapter 11 / 11


12-2007 5366/100 Agfa Company Confidential
1 2 3
DD+DIS148.05E Installation Planning

6.5 Example for System Integration

IP Address:192.9.200.155 Digitizer:
CR 75.0
Service
Host

ADC QS

Ultrasound

Nickname: Digitizer 53
63
Printer 11
Modality Brand: AGFA ad
.cd
r
Modality Type: CR QS
Calling AE-Title: QS

Nickname: Ultrasound
Modality Brand: ACUSON
Modality Type: ASPEN
Calling AE-Title: US_ASPEN

IP address: 192.9.200.201
Hostname: ds5302
Subnet mask: 255.255.255.0
Router IP address: -
Printer AE Title: DS5302
Domain name: -

Figure 8

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 DRYSTAR 5302 Chapter 11 / 19


12-2007 5366/100 Agfa Company Confidential
1 2 3
DD+DIS148.05E Installation Planning

7 Ambient conditions
7.1 During operation

Ambient temperature: min.: +15°C (59°F)


max.: +30°C (86°F)

Relative humidity: min.: 20%


(non condensing) max.: 75%

7.2 During transport and storage

NOTE:
x Climate conditions for storage must be in accordance with
EN60721-3-1-class 1K4
x Climate conditions for transport must be in accordance with
EN60721-3-2(1993)-class 2K4

Ambient temperature min.: - 25°C (-13°F)


during storage: max.: + 55°C (131°F)

Relative humidity min.: 10%


during storage: max.: 95%

Absolute humidity min.: 0.1 g/m³


during storage: max.: 35 g/m³

Rate of temperature max.: 1 °C/min.


change during storage

Air pressure during Min. 70 kPa


storage max.: 106 kPa

Ambient temperature min.: - 40°C (-40°F)


during transport: max.: + 70°C (158°F)

Relative humidity min.: 10%


during transport: max.: 95%

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 DRYSTAR 5302 Chapter 11 / 20


12-2007 5366/100 Agfa Company Confidential
1 2 3
DD+DIS148.05E Installation Planning

8 Electrical connections
Installation guidelines Electrical installations have to be done in accordance
with national regulations. Examples:
VDE (Germany) Electrical installations in the
installation room must be in compliance with the
regulations IEC 364, VDE 0100, and VDE 0107.
UL (US) Electrical installations in the installation
room must be in compliance with the regulations:
“National Electrical Code" (NEC) (NFPA70).

IMPORTANT:
A ground fault interrupter (GFI) (IN = 30 mA) must be installed at a suitable point in
the circuit (in compliance with VDE 664).

Mains connection in the VDE Double earthed pin outlet in compliance with
installation room DIN 49441 and with CEE 7 standard cover V II.
UL Earthed contact outlet for the NEMA 5-15 P plug

Mains cable Euro- CEE7/7 - H05VV - F3G - 3 x 2.5 mm² with


(part of the delivery) Version: IEC C13 plug (female); cable length 3 m
US/CAN- NEMA 5-15P - SJT, 3 x AWG 18 with
Version: IEC C13 plug (female); Cable length 3 m

US Europe

IEC C13 CEE7/7

NEMA 5-15P IEC C13


5366_chap11_001.cdr

Figure 9

Network cable UTP; AWG24; Length: 5 m


(part of the delivery)

Main switch Upon machine installation, it must be ensured that either


the mains connector or an all-pole circuit breaker for the
installation on site is located close to the machine and
easily accessible.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 DRYSTAR 5302 Chapter 11 / 21


12-2007 5366/100 Agfa Company Confidential
1 2 3
DD+DIS148.05E Installation Planning

8.1 Electrical connections at the printer

Ground plug. 100 - 240 V AC


6 mm metric thread power supply

Equipotential plug

Network connection.
RJ45 UTP for 100BaseT network,
according to IEC 601

Service Interface.
RS232 Interface male
DTE 9 pin sub D (nullmodem)
536611JM.cdr

Figure 10

9 Specifications

9.1 Type overview

Printer Type
DRYSTAR 5302 5366/100

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 DRYSTAR 5302 Chapter 11 / 22


12-2007 5366/100 Agfa Company Confidential
1 2 3
DD+DIS148.05E Installation Planning

9.2 Functional data

Film sizes 14x17", 14x14", 14x11", 10x12", 8x10"


Throughput (films/hour) 8x10": 140 14x17": 75
Access time1st film approx. (seconds) 77
Printable area(pixels) Refer to 9.2.1
resolution(DPI) 320 (79.4 µm Pixel size)
Bits per pixel / number of grey steps 12 / 4096
Daylight film loading yes
Films per film magazine 100 films
Number of film magazines 2
Image recording type Direct thermography
Densitometer built in yes
Automatic density control yes
Hard disk capacity t 73 GB
RAM Memory Plugged on Themis: 512 MB
Warm-up time Approx. 7 min. (cold printer)

9.2.1 Pixel Matrices


The DRYSTAR 5302 has two film supply trays that can be configured to 14x17",
14x14", 14x11", 10x14" or 8x10" film size. If trim is used around the images (trim is a
small, white frame around each image on the film), the diagnostic area is 6 pixels
smaller at each edge of the film.

DRYSTAR 5302 Diagnostic Printable Area in Pixels

No Annotations Annotations
Film Size
Width Height Width Height
8x10" 2376 3070 2376 2705
10x12" 3070 3653 3070 3288
14x11" 3348 4358 2983 4358
14x14" 4358 4303 4358 3938
14x17" 4358 5232 4358 4867
Table 1: Pixel Matrices

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 DRYSTAR 5302 Chapter 11 / 23


12-2007 5366/100 Agfa Company Confidential
1 2 3
DD+DIS148.05E Installation Planning

9.3 Electrical data

Rated voltage [~V] 100-240 AC (autosensing)


Frequency [Hz] 50/60
Power consumption [W] 70 standby
250 printing average
530 peak

9.4 Climatic data

Heat emission [kJ/h] 252 standby


900 printing average
1908 maximum

9.5 Sound emission data

Noise level: in standby: 53 dB(A) maximum


in operation: 62 dB(A) maximum

9.6 Interface data

Image / control data interface DICOM 3.0 Print SCP


Maximum number of simultaneous 10
DICOM associations

9.7 Service data

InfoCounter yes
Error diagnostics built in yes
Module exchange concept yes
Remote/preventive service capable yes

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 DRYSTAR 5302 Chapter 11 / 24


12-2007 5366/100 Agfa Company Confidential
1 2 3
DD+DIS148.05E Installation Planning

9.8 Dimensions and weight

9.8.1 Dimensions, packed machine

Length / depth (cm / inch) 90 / 35.4


Width (cm / inch) 87 / 34.2
Height (cm / inch) 69 / 27.2
Height incl. Pallet (cm / inch) 84 / 33.1

9.8.2 Dimensions, closed machine

Width (cm / inch) 72,8 / 28.6


Length / depth (cm / inch) 71,5 / 28.2
Height with output tray (cm / inch) 67,6 / 26.6
Height without output tray (cm / inch) 53,6 / 21.1

9.8.3 Dimensions, opened machine

Width (cm / inch) 72,8 / 28.6


Length / depth (cm / inch) 116 / 45.7
with film trays opened
Height (cm / inch) 76 / 29.9
with top cover opened

9.8.4 Weight

Printer alone (kg / lb) 90 / 198


Printer with packing incl. palette (kg / lb) 120,5 / 265,6

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 DRYSTAR 5302 Chapter 11 / 25


12-2007 5366/100 Agfa Company Confidential

You might also like