Professional Documents
Culture Documents
BUREAU OF MINES
Robert C. Horton, Director
Library of Congress Cataloging in Publication Data:
ILLUSTRATIONS
TABLES
cm centimeter mg milligram
ABSTRACT
1Metallu;gist.
2Research chemist.
3Research physicist.
Albany Research Center, Bureau of Mines, Albany, OR.
2
INTRODUCTION
DESCRIPTION OF METHOD
4underlined numbers in parentheses re- --~symbols used in the equations are de-
fer to items in the list of references fined in the nomenclature section at the
preceding the nomenclature section at the end of this report.
end of this report.
3
The mass flow of sample vapor (F) is 5 becomes insignificant as the linear
the sum of a slug-flow term and a dif- velocity of the gas mixture (V 1/A) in-
fusion term so that the mass of the vapor creases, but that care must be taken to
passing any point along the capillary per use flow rates low enough to insure
unit time is given by saturation. For large V,/A,
(3) (7)
.rFurnace
Gas outlet~~ Water out
~20
(
/ --->-
/~g \Condenser shield (Cold finger : Water in
Gas
1rr -
inletJJ~ IWater aut
a! 2 4 6 B 10
KEY ! I ! ! I I I I I I
Scole, em
- Carrier gas
~ Sweep gas (heat up and Heat up and cool down mode
coal dawn)
--- Sweep gas (measurement)
Gas inlet\1
......
LzE: u <-->-
------
/
~
~~.
-- --->- -
-)11
Gas outlet)
Measurement made
FIGURE 2. - Top view of modified transpiration apparatus, showing both unsteady state and
measurement modes.
condensate on one condenser during the If the linear gas velocity (V 1/A) is
heat-up and cool-down periods and to col- great enough to minimize diffusion ef-
lect a sample for measurement on the fects and V1 is small enough to assure
other condenser. By this means precise saturation, the Csat and P sat can be cal-
results may be obtained by collecting culated according to equations 7 and 8,
deposits during a time interval that is respectively. The concentration of the
independent of the heat-up and cool-down sample vapor in the y-intersection of
periods. This allows shorter tests than capillaries can also be calculated ac-
with the conventional method since the cording to
sampling time does not have to be long
compared to the heat-up and cool-down
(9)
periods in order to minimize error.
Therefore, static tests eVe = 0) can be condenser arms and the connecting cap-
performed and the value of D calculated i11ary has no effect on the diffusion
if Csat is obtained from dynamic tests at of gas in the capillary leading to the
the same temperature. This type of test y-intersection.
is valid if the flow of argon through the
APPARATUS
thermometer is used to convert the ther- was not known and the possible small
mocouple electromotive force to degrees error (2 pct) in calibration was
Celsius. Spot checks with a thermocouple neglected. Reproducible flow rates could
placed directy in the molten sample re- be achieved in the range of 2 to 200 cm 3j
vealed that the temperature in the well min at room temperature. The mass flow
is 3 K higher than that of the sample. controller allowed for good control de-
The temperatures presented in the tables spite the variable pressures encountered
have been corrected for this difference in bubbling gas through a liquid melt.
and therefore represent the true tempera-
ture of the sample. REAGENTS
rates. Because of this modification, residue after all experiments were per-
the controller was carefully calibrated formed. This demonstrates that the
by displacing a known amount of argon in vaporization itself is congruent. The
a given time using moving soap films in a carrier gas used in all experiments was
buret. Although the vapor pressure of reagent-grade argon, and flow rates are
water is 19.8 torr (2,633 Pa) at 295 K reported at ambient temperature and
(room temperature), the vapor pressure of pressure.
water in equilibrium with the soap film
EXPERIMENTAL PROCEDURE
1. The amount of material that dif- 3. The error due to material being
fuses upstream into the analytical con- transferred from the auxiliary to the
denser when the sample flow is directed analytical condenser during a normal run
into the auxiliary was determined. Dur- was determined. Fourteen mg of con-
ing heat-up and after attaining steady- densate was deposited on the auxiliary
state conditions at 870 k for 15 h, less condenser in the usual manner; then the
than 0.005 mg of Sb was deposited on the reaction boat was removed and the reactor
analytical condenser when using Sb 2 S 3 as tube was cleaned. Next, the flow of
the sample. Flow rates of 0.083±0.002 sweep gas through the condensers was
and 0.417±0.008 cm3 /s of argon were used reversed, the furnace was heated, and a
as carrier and sweep gas respectively. 1-h run at 870 K was performed without a
Since 153 mg of Sb would have been depos- sample in the reaction tube. After the
ited during a normal test, the error due run, only 0.07 mg of Sb 2 S 3 was found in
to diffusion into the wrong condenser was the analytical condenser. This test
considered negligible. shows that only a 0.5-pct error would
have occurred in a typical transpiration
2. Particulate matter that may be car- test from reverse deposit transfer. This
ried past the analytical condenser by the possible small error was neglected in
gas stream was determined. An aqua regia subsequent tests.
trap was placed inline just beyond the
Zinc chloride was chosen to study the Several vapor pressure determinations
reliability of the apparatus because it were made in the temperature range of 667
is a halide that can be obtained in pure to 883 K. Equilibrium pressures were
form, its vapor pressure has been well calculated for each determination, assum-
established by Keneshea and Cubicciotti ing that the vapor was entirely mono-
(3), and the temperature range for pres- meric. Pressures calculated on this
sures of 1 to 100 torr (133 to 13,300 Pa) basis are represented by a least-squares
is near that of Sb 2 S 3 • Reagent-grade equation. The set of data obtained is
ZnCl 2 was used in all tests. This mate- presented in table 1 and is compared with
rial was used without prior treatment, the transpiration data of Keneshea and
although water evolved during the heat-up Cubicciotti (3) in figure 5. Both sets
period and was carried out of the system of data are plotted assuming the vapor is
by argon. Deposits collected on the con- monomeric. Very close agreement was ob-
denser were anayzed for zinc by atomic tained between the two sets of data,
absorption spectroscopy, as was done with showing the reliability of the modified
the Sb 2 8 3 deposits discussed earlier. transpiration technique.
TEMPERATURE, K TEMPERATURE, K
800 750 700 650 1,400 1,200 1,000 909
1.5 r-::--------'---.--J----'-rJ--K-E-Y----.'-, 2.6
/'; This study
o Keneshea
2.1
L
L
L
o '-
E 1.6
c.. KEY
01 o Rya zan tsev
- ~ 1.1 o Ustyugov
/), Shendyapin
-.5~-~~'--~~1--~'~~~'~~~~~
1,25 1.30 1.35 1.40 1.45 1.50 1.55
103/ T, K- I
FIGURE 5•• Transpiration data calculated as- F IGU R E 6 •• Compari son of vapor pre ssu re mea sure-
suming monomer only for zinc chloride. mentsmadeover liquid Sb2S3 by other investigators.
9
over the surface of the sample in an were conducted at 870 K with the boat
attempt to obtain vapor pressure data. design that allowed the gas to bubble
through the sample. The data show that
The reaction shown in equation 1 was the vapor density was independent of
assumed to occur, and vapor pressures flow, up to 0.833±0.017 cm 3 /s. A value
were calculated according to equation 8. of 0.083±0.002 cm3 /s argon was chosen for
When the results were compared with data use in subsequent tests in the investiga-
of previous investigations (5-6, 8), our tion. This flow rate was low enough to
data agreed more closely with diose of insure saturation and high enough to make
Shendyapin (6). However, because of poor any effects due to diffusion extremely
reproducibiiity, it was decided to modify small.
the sample boat design to allow for bub-
bling the carrier gas through the molten VAPOR TRANSPORT MEASUREMENTS
sample.
The rate of vapor transport was mea-
Figure 7 shows the calculated vapor sured as a function of temperature using
pressures (assuming equation 1 to be val- the dual condenser apparatus previously
id) for the present study along with described. Vapor pressures as measured
those of Shendyapin. The vapor pressures by Ryazantsev (5) with a quartz manometer
obtained for tests in which the gas is were used in equation 2 to allow calcula-
bubbled through the molten sulfide are tion of V1 and subsequent calculation of
higher than those obtained when flowing
the gas over the sample. Since the data 0.020r-----,-----,-,---,-----,---,
for the unsaturated tests agree with I'"
zE o
those of Shendyapin, it can be concluded
that Shendyapin's vapor pressure data are
oa.
lLJu
ZE
.0 15
61
0 0------.........
'~
also low for the same reason. o
oz'
o:::Q .010
A series of tests was conducted to a:!;J:
determine the maximum flow rate that
could be used to insure saturation of the
~::: °
.0050 0.50 2.0 2.5
gas stream with sample vapor. Figure 8
shows calculated vapor concentrations as ARGON FLOW RATE, cm 3 /s
a function of the flow rate of argon FIGURE 8. " Calculated vapor concentrations of
fed to the reaction tube. All tests
Sb2S3 as a funclion of argon flow rate.
TEMPERATURE, K
1,000 SOO TEMPERATURE, K
I I
1,100 1,000 900 850
o -0.9 I I I
~ 0",,-
I
I.Sr- - Z
o~...o W r<l -1.1 c- 0",,- -
.....
.....
.... 0 o E 0
'0 z
o
+- 1.0 r-
o
"-
....... 0
.... 0
- u
U
0 ........
E -1.3 r- "'-0 -
d
.........0
0"-....6 6
' 0 .......
0 0:::
o " ""'-0
0>
o o
KEY
Saturated tests
".........
6" a.. ~ - 1.5 -
«- ""'-0 -
.2 - "Ucd"""""d tee', ~ . - >!;J: ""-0
o shendyapin .~'-' 0'10:: -1.7 -
~I-
""'-0
-.6 I I
O.S 0.9 1.0 I. I 1.2 1.3 -1.9 I I
0.9 1.0 1.1 1.2
103/T, K- I 103/T, K-I
FIGURE 7. - Calculated vapor pressures for Sb 2 S 3 as* FIGURE 9.· Vapor concentration measurements
suming the reaction Sb 2 S 3 (l) c., 2SbS(g) + 1I2S 2 (g). of Sb2S3 as a function of temperature.
10
Csat according to equation 5. Table 3 Table 4 contains the diffusion data ob-
shows the experimental data for the eight tained in the static tests.
transpiration tests, and figure 9 is a
plot of the logarithm of the vapor con- To demonstrate that diffusion effects
centrations versus 1,000/T. were negligible for the transpiration
measurements, a flow rate of 0.083±0.002
The least-squares regression equation cm 3 /s argon and the value for D at
for data in table 3 is 1,020 K (0.31 cm 2 /s) Were used in equa-
tions 2 and 5. The diffusion component
(3.082±0.070) (the exponential term) of equation 5 was
found to be 3.98 x 10- 14 , which is insig-
- (4,242±66.3)/T. (12) nificant compared to unity. The use of
equations 7 and 8 is therefore justified.
DIFFUSION COEFFICIENT MEASUREMENTS
CALCULATION OF THE AVERAGE
As previously discussed, the modified MOLECULAR WEIGHT
transpiration apparatus makes it possible
to determine the diffusion coefficients Combining our Sb 2 S 3 vapor density data
of compounds with the use of equation 10. with the static vapor pressure mea-
Static tests were run with Vc = 0 at sev- surements of Ryazantsev (5) permitted
eral different temperatures. A flow of calculation of the average molecular
0.167±0.003 cm 3 /s argon was used in the weight of the vapor species at various
condenser. The values used for Csat in temperatures. Average molecular weights
equation 10 were computed by using were calculated by using the vapor
the regression expression (equation 12), density data in table 3 along with vapor
and Co was calculated with equation 9. pressures calculated with equation 11
~
E
"-
01 250 - -
M' = Csat RT (13) 1-'
P
o~
• I
sat
~
w
Table 5 contains the average molecular ~ 200'--- -
weights calculated with equation 13. n::
« o~
-.J
~
:J
TABLE 5. - Calculated average molecular 0 150 -
W -
weight as a function of temperature -.J
0 0 ...............
~ I 0
100 I I
Temperature, P sat Cs at' M, 850 900 950 1,000 1,050
K torr Pa mg/cm 3 glmol TEMPERATURE, K
870 ••••••• 3.04 404 0.0160 286
FIGURE 10. Calculated average molecular
895 ••••••• 4.99 664 .0214 239 0
918 ••••••• 7.71 1,025 .0298 221 weight of gaseous species in equilibrium with
943 ••••••• 12.08 1,607 .0393 191 Sb 2 S3 (l) at various temperatures.
972 ••••••• 19.53 2,597 .0511 159
993 ••••••• 27.47 3,654 .0668 151
1,021 • •••••• 42.40 5,639 .0846 127 expressed in equation 1 is 135.8 glmol,
1,045 •••..•• 59.63 7,931 .1025 112 and the molecular weight of Sb 2S 3 is
339.7 glmol, it appears that Sb 2 S 3 dis-
The average molecular weight is plotted sociates to a greater degree as tempera-
as a function of temperature in figure ture increases. This phenomenon was also
10. A regression curve is plotted as the observed qualitatively, as it was noted
solid line in the figure and is repre- that more elemental sulfur was collected
sented by on the condensers as the temperature
increased. Shendyapin also observed the
M = 3,677 - 6.353T + 2.816 higher amount of elemental sulfur at
higher temperatures when doing transpira-
x 10- 3 T2 (870 k < T < 1,045 K). (14) tion experiments. The average molecular
weights listed in table 5 should be use-
Since the average molecular weight of ful in evaluating the results of mass
the gaseous product in the reaction spectrometry studies.
Since this study shows that the average and S2; the third (~) concluded the pri-
molecular weight of gaseous species in mary species to be Sb 2 S 3 • Analysis of
equilibrium with Sb 2S 3 (stibnite) is a the data in the literature indicates that
function of temperature, a discussion of there are several other species present
available mass spectrometric studies is in addition to SbS + S2 and Sb 2S 3 • All
helpful in understanding which vapor spe- previous researchers acknowledged this
cies may be present. While the mass but assumed a simple reaction in order to
spectrometry data available are at tem- calculate vapor pressure values from
peratures below the melting point of their Knudsen cell and transpiration
Sb 2 S 3 , the same species will exist in the data.
vapor at temperatures above the melting
point. Mass spectrometric analysis of Sullivan (7) reported appearance poten-
the vapor species over stibnite has been tials for all the ions found. An evalu-
performed by several researchers (1, 6- ation of the appearance potentials allows
7). Two of these researchers (1, 7)-con- a reasonable sorting of the ions into
eluded that the primary vapor species parents and fragments. The ionization of
below the melting point of Sb 2 S 3 are SbS a molecule without fragmentation requires
12
SUMMARY
A modified transpiration apparatus was apparatus. The use of two condensers and
designed and constructed incorporating a gas valves makes it possible to allow the
two-condenser system. The apparatus was flow of sample to come to steady-state
able to reproduce accepted vapor pressure temperature before transporting vapor to
data for ZnC1 2 • This new device allows the analytical condenser. It was also
more accurate measurements of vapor shown that the device could be used to
transport, since it eliminates errors obtain diffusion coefficient data by
associated with the heat-up and cool-down using zero flow of carrier gas through or
periods when using a conventional over the sample.
13
The vapor concentrations over liquid function of temperature and was shown to
antimony sulfide were measured with the decrease from 286 g/mol at 870 K to 112
modified transpiration apparatus in the g/mol at 1,045 K. Additional mass spec-
temperature range of 870 to 1,045 K. trometry work is needed to determine
Comparison of these data with previously quantitatively the various gaseous spe-
published results disproves the hypoth- cies that exist as a function of
esis that Sb 2 S 3 dissociates completely temperature. Experiments conducted at
to SbS plus S2 on vaporization. The higher temperatures should provide con-
present work indicates that the dissocia- firmation as to whether free antimony
tion is incomplete and increases with exists as a vapor. The average molecular
temperature. weights obtained in this study should be
useful in verifying the mass spectro-
The average molecular weight for the metric results.
gaseous species was calculated as a
REFERENCES
NOMENCLATURE
A Capillary area, cm 2
Volume flow rate of carrier plus sample gas through or over boat, cm 3/s
Volume flow rate of sweep gas through condenser capillary only at reaction
temperature, cm3 /s
V'c Volume flow rate of sweep gas through condenser only at calibration
temperature, cm3 /s