You are on page 1of 8

Accepted: 11 September 2017

DOI: 10.1111/jocd.12437

BACK TO BASICS

Sensitivity of aspiration as a safety test before injection of


soft tissue fillers

 MD | Job Thuis MD
Jani AJ Van Loghem MD | James J Fouche

Aesthetic Medicine Centre, Doctors Inc.,


Amsterdam, The Netherlands Summary
Background: Embolism due to accidental intra-arterial injection of a soft tissue filler
Correspondence
Jani AJ Van Loghem, Aesthetic Medicine (STF) can lead to serious complications. Physicians practice aspiration as a safety
Centre, Doctors Inc., Amsterdam, The test before injection to rule out intra-arterial placement of the needle tip. The value
Netherlands.
Email: jani@doctors-inc.com of aspiration as a safety test is evaluated in this study, which also considered the
rheological properties of different STFs and their needle dimensions.
Methods: Aspiration with eleven different needle sizes and 24 different STFs was
performed using Ringer’s lactate (RL) colored with blue skin marking ink and sec-
ondly an empty saline bag containing anticoagulated blood, both pressurized to
150 mm Hg to mimic arterial blood pressure. Time between the start of aspiration
and liquid in the needle hub was recorded.
Results: A total of 340 individual aspiration tests with different STFs and needles
were performed, of which 112 yielded true-positive results within 1-s and 128
yielded false-negative results after 10 s. Positive results between one and ten sec-
onds were seen in 101 aspiration tests.
Conclusions: True-positive results were seen in 33% of the tests performed with
STFs within 1 s of aspiration. Aspiration test results are influenced by needle diame-
ter, needle length, and rheological properties of STFs. Additional safety measures
are necessary to further reduce risk of inadvertent intravascular injection of STFs.

KEYWORDS
aspiration, complications, embolism, prevention, soft tissue fillers

1 | INTRODUCTION After intra-arterial injection of gel-based substances, blood flow to a


critical anatomic area may be restricted or blocked, leading to ische-
Botulinum toxin type A and soft tissue filler (STF) treatments are mia, followed by necrosis and sometimes, permanent scarring and
immensely popular, a major reason being the minimal amount of disfigurement or even blindness.3-9 To avoid or limit this risk, many
downtime experienced when compared to surgical aesthetic proce- experts and pharmaceutical companies advocate performing aspira-
dures. Over the past few years, there has been a significant increase tion before injection, as blood in the hub of the needle during aspi-
in the amount of injectable treatments performed. Nonpermanent ration confirms an intravascular needle position, after which injection
STF treatments are second on the list of nonsurgical procedures per- should be avoided. When considering the rise in STF treatments, the
formed in the United States (after botulinum toxin type A treat- incidence of complications is likely to also increase. A critical review
ments), with 2.1 million hyaluronic acid procedures conducted in of current techniques and materials used is therefore imperative.
2015.1 Apart from needle diameter and length, extrusion and aspiration
Various complications after the use of STFs have been docu- of STFs are dependent on differences in physiochemical structure
mented. Among these, the occurrence of an intra-arterial embolism and rheological properties. The elastic quantitative measurement
due to accidental intravascular injection is among the most serious.2 (elasticity) of a STF’s stiffness and its ability to resist deformation

J Cosmet Dermatol. 2017;1–8. wileyonlinelibrary.com/journal/jocd © 2017 Wiley Periodicals, Inc. | 1


2 | VAN LOGHEM ET AL.

under applied pressure are known as G prime (G’).10 The higher the the needle hub. When STF product packages contained needles with
G’ of a STF, the less it will deform under pressure and the more dimensions other than our selected needle dimensions, these needles
stored energy it retains. Thickness, or viscosity of a STF in its fluid were also tested with the STF’s with which they were supplied.
form (g*), refers to the measure of its resistance to gradual reforma- The aspiration test was then repeated using the same protocol,
tion by shear stress and how it flows from the needle while being but replacing the dyed ringer’s lactate solution with blood drawn
injected. Therefore, elasticity and viscosity will also determine how a from two of the authors (JvL and JF). Prior to use, the blood was
STF product behaves in the lumen of the needle when aspiration is anticoagulated using tubes containing ethylenediaminetetraacetic
applied. The higher the G’ and g*, the more force is needed to acid (EDTA) with a concentration of 1.8 mg/mL of blood.
obtain a positive aspiration result by emptying a needle filled with
STF material. Most commercial STF products have a viscosity
derm Ultraâ, Allergan Inc.,
between 7.307 centipoises (cP) (Juve 3 | RESULTS
â
Irvine, CA, USA) and 349 830 cP (Radiesse , Merz Pharmaceuticals
GmbH, Frankfurt, Germany).11 When comparing these values to The aspiration results are shown in Table 3. All needles tested in the
water (0.890 cP at 25°C) and blood (between 3.000 and 4.000 cP at 12 saline control tests yielded positive results within 1 s of aspira-
37°C), it becomes clear that the results of an aspiration test may not tion. In total, 340 tests with 24 STFs were performed; 112 tests
always be reliable as the high viscosity of STFs would mean that a yielded true-positive results (within 1 s), 101 tests yielded positive
very high suction pressure must be applied in order to obtain a posi- results between 1 and 10 s after aspiration, and 128 tests yielded
tive aspiration test result.12 negative results beyond 10 s of aspiration.
Our aim of this study was to investigate the value and reliability The reliability of aspiration as a safety test when a true-positive
of aspiration as a medical test to prevent mistakenly placing a pro- result was defined as being positive up to 10 s after the start of
duct in a vascular structure leading to possible complications, while aspiration was 63%. When this time limit was decreased to 1 s after
comparing different STFs and needle dimensions. the start of aspiration, the reliability and amount of true-positive
results dropped to 33%.
When only considering the needles supplied within the packages
2 | METHODS of the different STF products, the reliability of aspiration was 37%
(positive within 1 s) and 74% (positive up to 10 s of aspiration), as
To create an environment where we could be certain of mimicking an seen in Table 4.
intravascular positioning of a needle tip, we first used a 500 mL bag of
Ringer’s Lactateâ (RL) (Baxter Healthcare Corporation, Deerfield, IL,
USA) solution pressurized to 150 mm Hg by means of a Metpakâ 4 | DISCUSSION
pressure infusor (Riester GmbH, Jungingen, Germany) to which was
added 2 mL Bonnie’s Blueâ (Delasco, Council Bluffs, IA, USA) skin Aspiration before injection is in essence a clinical test, to determine
marking ink. The appearance of a blue color in the needle hub was whether the needle tip is inside a blood vessel, and therefore, it is a
considered as a positive aspiration test result (Figure 1). Eleven differ- safety test to reduce the risk of intravascular injection. The speci-
ent sharp point needle sizes were tested on all products as well as the ficity of an aspiration test is high, as blood in the needle hub indi-
needles that were supplied with the products by the manufacturer cates a very high probability of intravascular placement (Table 5).
(Table 1). The syringes and STF products tested are listed in Table 2. However, the sensitivity of aspiration as a diagnostic test is low (de-
All products except the control were produced and supplied by pending on the product and needle tested) as absence of blood in
the respective pharmaceutical companies in the Netherlands. the needle hub does not exclude the possibility of intravascular
Beloteroâ, Radiesseâ, and Etermisâ products were supplied by Merz placement of the needle tip, as can be seen from the false-negative
Pharmaceuticals, Restylaneâ products were supplied by Galderma, test results illustrated in Table 3. As a cutoff time of 1 s after aspira-
â
and Juvederm products were supplied by Allergan. Every product tion revealed fewer true-positive results, one could argue that these
except the control was tested with its originally supplied syringes, all tests are less reliable compared with those with a cutoff time of
of which were 1-mL syringes, with the exception of Radiesseâ which 10 s. However, in clinical practice, physicians usually do not aspirate
was supplied with 1.5 and 0.8-mL syringes. Each product was aspi- for more than 1 s. In this case, the orange cells in Table 3 could also
rated twice with every needle size to determine whether any of the be classified as false-negative test results, making aspiration a highly
blue-dyed Ringer’s lactate could be aspirated. Before testing, the unreliable test for many STFs when no blood is seen in the needle
needles were primed and the syringes emptied of half of their STF hub during aspiration.
contents to allow for sufficient aspiration with the plunger. A mini- Our results may also be slightly more conservative than those
mum of 0.5 mL vacuum was visible in the syringe when the plunger experienced in routine clinical practice in that we used a relatively
was pulled during aspiration, which correlates to a negative pressure high pressure (150 mm Hg) compared with the world population’s
of 50 mm Hg or 6666.118 Pa.13 A timer was used to measure the mean systolic blood pressure of 124 mm Hg as stated by the World
time between start of aspiration and the appearance of blue color in Health Organization.14
VAN LOGHEM ET AL. | 3

F I G U R E 1 Two series of aspiration


tests were performed; the first using
Ringer’s lactate dyed with blue ink and
pressurized to 150 mm mercury to mimic
arterial blood pressure. The second test
was performed in the same way, but with
EDTA-anticoagulated blood that was
pressurized to 150 mm mercury

T A B L E 1 Sharp point needle gauge and length of needles tested A number of other factors can also influence the effectiveness of
in the study aspiration as a safety test. Blood vessels may collapse at negative
Manufacturer (products pressure during aspiration, which could prevent blood from entering
with manufacturer-supplied the syringe. During a preperiosteal injection, there is a chance that
Needle size needles are indicated in brackets) the needle may strike an artery and, instead of piercing through, the
33G, 13 mm TSK Laboratory, Japan needle tip may push the elastic artery to the periosteum. During
30G, 13 mm TSK Laboratory, Japan aspiration, one would see no blood, but an embolism could still arise
30G, 12 mm Terumo Europe N.V. Belgium after injection, while the bevel of the needle may be intra-arterial
(Etermis) even though the needle tip is located on the bone.16
29G, 12 mm Terumo Europe N.V. Belgium In a recent study, where the reliability of the aspiration study
(Restylane)
was examined in a rabbit ear model, results were positive for four
28G, 19 mm Exelint International, USA STF products tested with five different needle sizes.17 However,
(Radiesse)
with in vivo experiments, one can never be certain that the needle
27G, 12 mm Terumo Europe N.V. Belgium
lumen is indeed intra-arterial at the moment of aspiration as hand
(Etermis)
tremors can rapidly effect needle tip position. This can lead to a
27G, 13 mm TSK Laboratory, Japan
true-negative result under the assumption that the needle is placed
27G, 20 mm Terumo Europe N.V. Belgium
intravascularly.
(Radiesse Plus, Etermis)
Technically, aspiration before injection can be challenging. When
25G, 13 mm TSK Laboratory, Japan
one aspirates, the plunger needs to be pulled. After verifying that
25G, 25 mm Terumo Europe N.V. Belgium
there is no blood in the syringe, the fingers are repositioned on the
(Etermis)
plunger so that the STF can be injected. During this repositioning,
23G, 13 mm TSK Laboratory, Japan
the needle tip may change position. As one or two millimeters can
make the difference between intra- and extravascular injection, the
Although blood viscosity is marginally reduced by EDTA, the value of the aspiration test is further reduced. Furthermore, when
shape of the viscosity curve is relatively unchanged when evaluated aspirating with a full syringe, a maximum of 0.1 mL negative pres-
at different shear rates.15 Blood viscosity is a variable in the reliabil- sure can be created due to the limitations of the syringe. Our study
ity of the aspiration test. The higher viscosity of blood (3-4 cP) com- was performed with a minimum of 0.5 mL negative pressure per
pared with Ringer’s lactate solution (0.7-1 cP) slows down the product, indicating that in practice false-negative test results could
intraluminar movement of liquids, as can be seen in the greater time be more likely than in this experimental setup.
interval in the blood assay between aspiration and a positive test When considering bevel length, larger gauge needles have
result. greater bevel lengths compared to smaller gauge needles. When an
4 | VAN LOGHEM ET AL.

T A B L E 2 Syringes and soft tissue fillers (STFs) tested


STF Contents
Control (C) Saline in a 1 mL Norm-Ject Tuberculin syringe (Henke Sass Wolf)
Belotero Hydroâ 21 mg/mL glycerol, 18 mg/mL hyaluronate
Belotero Softâ Plus 20 mg/mL HA containing 0.3% lidocaine
Belotero Balanceâ Plus 22.5 mg/mL HA containing 0.3% lidocaine
Belotero Intenseâ Plus 25 mg/mL HA containing 0.3% lidocaine
Belotero Volumeâ Plus 26 mg/mL HA containing 0.3% lidocaine
Undiluted Radiesseâ 0.8 mL syringe CaHA
â
Diluted Radiesse 0.8 mL syringe CaHA with 0.16 mL Xylocaineâ 1% mixed according to the FDA approved protocol22
â
Undiluted Radiesse 1.5 mL syringe CaHA
â
Diluted Radiesse 1.5 mL syringe CaHA with 0.3 mL Xylocaineâ 1% mixed according to the FDA approved protocol22
â
Radiesse Plus 1.5 mL syringe CaHA containing 0.3% lidocaine
â
Etermis 2 20 mg/mL HA
Etermis 3â 23 mg/mL HA
Etermis 4â 24 mg/mL HA
â
derm Volbella
Juve 15 mg/mL HA containing 3 mg/mL lidocaine hydrochloride
derm Voliftâ
Juve 17.5 mg/mL HA containing 3 mg/mL lidocaine hydrochloride
â
derm Voluma
Juve 20 mg/mL HA containing 3 mg/mL lidocaine hydrochloride
derm Ultra 2â
Juve 24 mg/mL HA containing 3 mg/mL lidocaine hydrochloride
derm Ultra 3
Juve 24 mg/mL HA containing 3 mg/mL lidocaine hydrochloride
derm Ultra 4â
Juve 24 mg/mL HA containing 3 mg/mL lidocaine hydrochloride
derm Ultra Smileâ
Juve 24 mg/mL HA containing 3 mg/mL lidocaine hydrochloride
Restylane Kysseâ 20 mg/mL HA containing 3 mg/mL lidocaine hydrochloride
â
Restylane Defyne 20 mg/mL HA containing 3 mg/mL lidocaine hydrochloride
Restylane Refyneâ 20 mg/mL HA containing 3 mg/mL lidocaine hydrochloride
â
Restylane Lyft 20 mg/mL HA containing 3 mg/mL lidocaine hydrochloride
Restylane Lidocaineâ 20 mg/mL HA containing 3 mg/mL lidocaine hydrochloride
Restylane Fynesseâ 20 mg/mL HA containing 3 mg/mL lidocaine hydrochloride
â
Restylane Volyme 20 mg/mL HA containing 3 mg/mL lidocaine hydrochloride

CaHA, calcium hydroxylapatite; HA, hyaluronic acid.

elastic artery is hit and pushed to the periosteum, and then pene- affected, and the chances of alternative oxygenation are reduced.
trated, the chance of the bevel being in the lumen of the artery is This is especially true in high-risk areas such as the nose or the
greater when a larger gauge needle is used, compared to smaller glabellar region where high amounts of end arterioles are present. A
gauge needles with shorter bevel lengths. This is especially relevant recent cadaver study, which sought to determine the amount of filler
in anatomic areas where arteries run very close to the periosteum, necessary to embolize the retina from the supratrochlear artery,
as seen in cadaver dissections at the temporal hollows, the perior- found that while the average volume required was 0.08 mL,
bital, and the frontal area.18 While this may argue in favor of using embolization was possible with amounts as small as 0.04 mL.19 In
small needles, our study has demonstrated that thinner gauge nee- the periorbital area, the maximum injected volume should therefore
dles lead to more false-negative test results and should not be not exceed 0.04 mL per bolus when injecting with a needle, consid-
considered safer on the basis of bevel length. ering the possibility that even though the needle tip might be placed
In the face, the different anatomic regions often have alternative on the periosteum, the lumen might still be intravascular due to the
blood supplies, in case an obstruction arises at one of the arterial bevelled needle tip.
sources. It is the authors’ belief that in practice, many patients are In the current study, the 28-G needle (19 mm) was associated
injected intra-arterially. However, because a retrograde injection with more false-negative results than the 30-G needle (13 mm) in a
technique is generally used, the extent of the embolism is limited number of tests, for example Belotero Balance. The same was true
and an alternative blood supply is sufficient for oxygenation of the for the 23-G (19 mm) vs the 25-G (13 mm) needle tests. When con-
affected capillary bed. When the embolus becomes more voluminous sidering the physics of fluid dynamics, pipe resistance (in this case
due to static injection inside an artery, a larger capillary bed will be needle lumen) is determined by diameter and length and possibly
VAN LOGHEM ET AL. | 5

T A B L E 3 Aspiration test results


Needle gauge (G), length

Product Agent 23, 19 mm 25, 13 mm 27, 13 mm 28, 19 mm 30, 13 mm 33, 13 mm


Control (Saline) RL + ink
Blood + EDTA
Belotero Hydro RL + ink 6s
Blood + EDTA 9s
Belotero Soft RL + ink 5s
Blood + EDTA 4s
Belotero Balance RL + ink 3s
Blood + EDTA 3s 5s 7s
Belotero Intense RL + ink 8s 8s 3s
Blood + EDTA 9s 6s 5s
Belotero Volume RL + ink 2s
Blood + EDTA 5s 8s
Radiesse 0.8 (undiluted) RL + ink
Blood + EDTA
Radiesse 0.8 (diluted) RL + ink 3s
Blood + EDTA
Radiesse 1.5 (undiluted) RL + ink 4s 6s 8s
Blood + EDTA
Radiesse 1.5 (diluted) RL + ink 5s
Blood + EDTA 7s 3s 3s
Radiesse Plus 1.5 RL + ink 4s 3s 13 mm: 4 s
20 mm: negative
Blood + EDTA 6s 3s 13 mm: 6 s
20 mm: negative
Etermis 2 RL + ink 13 mm 13 mm 5s
20 mm 12 mm
Blood + EDTA 13 mm 13 mm 8s
20 mm 12 mm
Etermis 3 RL + ink 13 mm 3s
12 mm
Blood + EDTA 13 mm 3s
12 mm
Etermis 4 RL + ink 13 mm 13 mm 3s
25 mm 12 mm
Blood + EDTA 13 mm 13 mm 4s
25 mm: 3s 12 mm
derm Volbella
Juve RL + ink 4s
Blood + EDTA 6s
derm Volift
Juve RL + ink 4s
Blood + EDTA 5s
derm Voluma
Juve RL + ink
Blood + EDTA
derm Ultra 2
Juve RL + ink 5s 5s 3s
Blood + EDTA 9s 8s 4s

(Continues)
6 | VAN LOGHEM ET AL.

TABLE 3 (Continued)
Needle gauge (G), length

Product Agent 23, 19 mm 25, 13 mm 27, 13 mm 28, 19 mm 30, 13 mm 33, 13 mm


derm Ultra 3
Juve RL + ink 4s 3s 2s
Blood + EDTA 4s 4s 3s
derm Ultra 4
Juve RL + ink 7s 6s
Blood + EDTA 10 s 10 s
derm Ultra Smile
Juve RL + ink 4s 2s 2s 7s
Blood + EDTA 10 s 3s 3s
Restylane Kysse RL + ink 6s 2s
Blood + EDTA 3s 3s
Restylane Defyne RL + ink 9s 10 s 7s 2s 8s
Blood + EDTA
Restylane Refyne RL + ink 6s 4s 4s 3s 9s
Blood + EDTA 5s 8s
Restylane Lyft RL + ink 2s 2s 29G: 4 s
28G: 8 s
Blood + EDTA 29G: 4 s
28G
Restylane Lidocaine RL + ink 29G 2s 8s
28G
Blood + EDTA 29G
28G
Restylane Volyme RL + ink 7s 6s 7s 7s
Blood + EDTA 9s 7s 7s
Restylane Fynesse RL + ink 8s 3s 6s
Blood + EDTA 4s 8s

Green: positive (within 1 s), orange: positive (between 1 and 10 s), red: false-negative test result (when still negative after 10 s).
RL, Ringer’s lactate; EDTA, ethylenediaminetetraacetic acid.

surface roughness of the needle lumen. When comparing the dimen- of the tests were false negative, which is comparable to our
sions of these two needles, the 28-G (or 23-G) needle with its longer results.17
length results in a higher flow resistance during STF aspiration or For optimal results and accurate product placement, anatomic
extrusion than the 30-G (or 25-G) needle, which is in accordance knowledge is a key. In addition, the risk of intra-arterial embolisms
with our results. Test results will therefore be influenced by filler can be reduced by adhering to the following advice:
rheology. For example, the rheology of Belotero Balance differs from
that of Restylane Lidocaine in terms of cohesivity (measured in kPa), • Take caution when practicing aspiration as a negative test result
and the 28-G (19 mm) needle showed positive test results with (no blood in needle hub) might be false negative and give the
Restylane Lidocaine but negative with Belotero Balance, while the injector a false sense of security;
30-G (13 mm) needle showed negative test results with Restylane • When injecting with a sharp point needle, a retrograde technique
Lidocaine and positive for Belotero Balance. should be used when the product is released. In the event that
A few papers have examined the value of aspiration before injec- the needle penetrates an artery, only a limited amount of product
tion. In one study which focused on Radiesse, the authors trans- would then be released intra-arterially;
ferred diluted Radiesse to a 1-mL luer lock syringe and used a 27-G • Avoid large-volume bolus injections with the needle in a static
needle.20 This syringe has different dimensions and therefore differ- position to avoid occlusion of a large area of the arterial system;
ent resistance characteristics compared to the syringe supplied by • Avoid high-pressure injections as these may lead to retrograde
the manufacturer. As seen in our results, both needle and syringe flow against the arterial blood flow into a proximal arterial system
dimensions may contribute to the reliability of an aspiration test. (like the ophthalmic artery);
In another recent publication, where the reliability of aspiration • Although nontraumatic cannulas can penetrate arterial walls,21
as a safety test with different STF materials was also examined, 60% we recommend using cannulas over needles, and with diameters
VAN LOGHEM ET AL. | 7

T A B L E 4 Test results of products with their supplied needles


Aspiration Aspiration
Merz products Agent reliability Merz products Agent reliability
Belotero Hydro (30 G) RL + ink Radiesse 0.8 (undiluted, 28G) RL + ink
Blood + EDTA Blood + EDTA
Belotero Soft (30G) RL + ink Radiesse 0.8 (diluted, 28G) RL + ink
Blood + EDTA Blood + EDTA
Belotero Balance (27G) RL + ink Radiesse 1.5 (undiluted, 28G) RL + ink
Blood + EDTA Blood + EDTA
Belotero Balance (30G) RL + ink Radiesse 1.5 (diluted, 28G) RL + ink
Blood + EDTA Blood + EDTA
Belotero Intense (27G) RL + ink Radiesse Plus 1.5 (27G, 20 mm) RL + ink
Blood + EDTA Blood + EDTA
Belotero Volume (27G) RL + ink Etermis 2 (30G) RL + ink
Blood + EDTA Blood + EDTA
Belotero Volume (30G) RL + ink Etermis 2 (27G) RL + ink
Blood + EDTA Blood + EDTA
ETERMIS 3 (27G) RL + ink
Blood + EDTA
Etermis 4 (27G) RL + ink
Blood + EDTA
Etermis 4 (25G) RL + ink
Blood + EDTA
Aspiration Aspiration
Allergan products Agent reliability Galderma product Agent reliability
derm Volbella (30G)
Juve RL + ink Restylane Kysse (27G) RL + ink
Blood + EDTA Blood + EDTA
derm Volift (30G)
Juve RL + ink Restylane Defyne (30G) RL + ink
Blood + EDTA Blood + EDTA
derm Voluma (27G)
Juve RL + ink Restylane Refyne (30G) RL + ink
Blood + EDTA Blood + EDTA
derm Ultra 2 (30G)
Juve RL + ink Restylane Lyft (29G) RL + ink
Blood + EDTA Blood + EDTA
derm Ultra 3 (27G)
Juve RL + ink Restylane Lidocaine (29G) RL + ink
Blood + EDTA Blood + EDTA
derm Ultra 4 (27G)
Juve RL + ink Restylane Volyme (27G) RL + ink
Blood + EDTA Blood + EDTA
derm Ultra Smile (27G)
Juve RL + ink Restylane Fynesse (30G) RL + ink
Blood + EDTA Blood + EDTA

Green: positive (within 1 s), orange: positive (between 1 and 10 s), red: false-negative test result (when still negative after 10 s).
RL, Ringer’s lactate; EDTA, Ethylenediaminetetraacetic acid.

of 25G and larger to reduce the possibility of penetrating arterial 5 | CONCLUSIONS


walls;
• When choosing periosteal bolus injections with a needle, multiple Aspiration of STFs as a safety test in this study led to many false-
small boluses (0.05 mL per bolus or less) with low pressure negative results. There are many variables that contribute to the
should be used. sensitivity of aspiration as a clinical test including STF rheology, nee-
• When injecting in the periorbital area (eg, close to the dle size (lumen, length), needle brand, syringe dimensions, aspiration
supratrochlear artery), smaller boluses should be used (<0.04 mL) time, negative pressure created during aspiration, and blood pres-
as intravascular injection in this region can cause blindness. sure. Even though a positive test result (blood in the needle hub)
8 | VAN LOGHEM ET AL.

T A B L E 5 Aspiration a clinical diagnostic test 10. Sundaram H, Cassuto D. Biophysical characteristics of hyaluronic
acid soft-tissue fillers and their relevance to aesthetic applications.
The truth
Plast Reconstr Surg. 2013;132(4 suppl 2):5S-21S.
Aspiration as a diagnostic test Intravascular Extravascular 11. Sundaram H, Voigts B, Beer K, Meland M. Comparison of the rheo-
logical properties of viscosity and elasticity in two categories of soft
Test results Blood True positive False positive
tissue fillers: calcium hydroxylapatite and hyaluronic acid. Dermatol
No blood False negative True negative Surg. 2010;36:1859-1865.
12. Elert G. “Viscosity”. The physics hypertextbook. (http://physics.info/
Only intravascular was measured in this in vitro test resulting in only true-
viscosity/). Retrieved December 2015–February 2016.
positive or false-negative test results. In vivo, this table applies for aspira-
13. Haseler L, Sibbitt RR, Sibbitt Jr WL, Michael AA, Gasparovic CM,
tion as a diagnostic test. Green: true positive test result, dark orange: false
Bankhurst AD. Syringe and needle size, syringe type, vacuum gener-
negative test result (analogue to red and orange in Tables 3,4).
ation, and needle control in aspiration procedures. Cardiovasc Inter-
vent Radiol. 2011;34:590-600.
can be of great value in clinical practice, we caution practitioners not 14. World Health Organization. Global health observatory data. http://
www.who.int/gho/ncd/risk_factors/blood_pressure_prevalence/en/.
to rely on the validity of a negative test result (no blood in the
Accessed April, 19, 2015.
needle hub) due to the high probability of a false-negative result. 15. Peker SM, Helvaci SS. Flow of biological fluids in the circulatory sys-
tem. In: Peker SM, Helvaci SS, eds. Solid liquid two-phase flow, 1st
ed. Amsterdam: Elsevier B.V; 2008:233-234. ISBN: 978-0-444-
CONFLICT OF INTEREST 52237-5.
16. Van Loghem JAJ, Humzah D, Kerscher M. Cannula versus sharp nee-
None.
dle for placement of soft tissue fillers: an observational cadaver
study. Aesthet Surg J. 2016;sjw220, 1-16.
REFERENCES 17. Casabona G. Blood aspiration test for cosmetic fillers to prevent
accidental intravascular injection in the face. Dermatol Surg.
1. The American Society for Aesthetic Plastic Surgery. 2016 Cosmetic 2015;41:841-847.
Surgery National Data Bank Statistics. https://www.surgery.org/med 18. Pessa JE, Rohrich RJ. The temporal fossa. Facial topography: clinical
ia/statistics. Accessed January, 10, 2017. anatomy of the face. St Louis, MO: Quality Medical Publishing, Inc.;
2. Funt D, Pavicic T. Dermal fillers in aesthetics: an overview of 2012:177-218.
adverse events and treatment approaches. Clin Cosmet Investig Der- 19. Khan TT, Colon-Acevedo B, Mettu P, DeLorenzi C, Woodward JA.
matol. 2013;6:295-316. An anatomical analysis of the supratrochlear artery: considerations in
3. Coleman SR. Avoidance of arterial occlusion from the injection of facial filler injections and preventing vision loss. 2017;37:203-208.
soft tissue fillers. Aesthet Surg J. 2002;22:555-557. 20. Aguilera SB, Tivoli YA, Seastrom SJ. How to make calcium hydroxy-
4. Peter S, Mennel S. Retinal branch artery occlusion following injection lapatite injections safer. J Drugs Dermatol. 2014;13:1015.
of hyaluronic acid (Restylane). Clin Exp Ophthalmol. 2006;34:363- 21. Yeh LC, Fabi SG, Welsh K. Arterial penetration with blunt-tipped
364. cannulas using injectables: a false sense of safety? Dermatol Surg.
5. Sung MS, Kim HG, Woo KI, Kim YD. Ocular ischemia and ischemic 2017;43:464-467.
oculomotor nerve palsy after vascular embolization of injectable cal- 22. Marmur E, Green L, Busso M. Controlled, randomized study of pain
cium hydroxylapatite filler. Ophthal Plast Reconstr Surg. 2010;26:289- levels in subjects treated with calcium hydroxylapatite premixed with
291. lidocaine for correction of nasolabial folds. Dermatol Surg.
6. Lazzeri D, Agostini T, Figus M, Nardi M, Pantalone M, Lazzeri S. 2010;36:309-315.
Blindness following cosmetic injections of the face. Plast Reconstr
Surg. 2012;129:994-1012.
7. Park SW, Woo SJ, Park KH, Huh JW, Jung C, Kwon OK. Iatrogenic
retinal artery occlusion caused by cosmetic facial filler injections. Am  JJ, Thuis J.
How to cite this article: Van Loghem JAJ, Fouche
J Ophthalmol. 2012;154:653-662. Sensitivity of aspiration as a safety test before injection of
8. Ozturk CN, Li Y, Tunk R, Parker L, Piliang MP, Zins JE. Complications
soft tissue fillers. J Cosmet Dermatol. 2017;00:1–8.
following injection of soft-tissue fillers. Aesthet Surg J. 2013;33:862-
877. https://doi.org/10.1111/jocd.12437
9. Roberts SA, Arthurs BP. Severe visual loss and orbital infarction fol-
lowing periorbital aesthetic poly-(L)-lactic acid (PLLA) injection. Oph-
thalmic Plast Reconstr Surg. 2012;28:e68-e70.

You might also like