You are on page 1of 6

Original Research

Evaluation of injection force of


three insulin delivery pens
T Asakura†, H Seino, M Kageyama & N Yohkoh
†Department of Clinical Pharmacy, Niigata University of Pharmacy and Applied Life Sciences,
1. Introduction 265-1 Higashizima, Akiha-ku, Niigata, 956-8603 Japan
2. Methods
Objective: Reduced injection force is among the modifications to the Next
3. Results
Generation FlexPen® (NGFP). This force was compared with two other
4. Discussion
prefilled pens: SoloStar® (SS) and KwikPen® (KP). Research design/methods:
5. Conclusions The injection force of the pens was measured, with either a BD Micro-Fine™
Expert Opin. Pharmacother. Downloaded from informahealthcare.com by Selcuk Universitesi on 01/01/15

6. Expert Opinion 31G thin-wall needle, or a NovoFine® 32G Tip extra thin wall needle attached.
Pens of each type were tested with both needles, during injection of 20 U
insulin, at speeds 3.3, 5 and 8.3 mm/s. Results: NGFP had a significantly
(p < 0.05) lower mean injection force than SS or KP, at all injection speeds,
with both needles. Injection forces (mean ± s.d.) with NGFP and the BD Micro-Fine™
31G thin-wall needle were: 8.1 ± 0.7, 10.7 ± 1.4 and 15.6 ± 0.9 N at the three
speeds, respectively. For SS, the corresponding values were: 9.2 ± 0.5, 13.3 ± 0.8
and 20.7 ± 2.4 N. For KP they were: 12.5 ± 1.6, 16.9 ± 1.2 and 24.5 ± 2.6 N.
Attached to the NovoFine® 32G Tip extra thin wall needle, the NGFP injection
forces were: 5.7 ± 0.4, 8.2 ± 0.7 and 12.7 ± 0.5 N; with SS were: 6.7 ± 0.3,
10.4 ± 2.1 and 16.3 ± 1.1 N; and with KP were: 9.1 ± 1.3, 13.1 ± 0.8 and
For personal use only.

21.6 ± 2.0 N. The injection force with NGFP was 12 – 25% lower compared
with SS and 35 – 41% lower compared with KP. Conclusions: This study shows
that NGFP has a significantly lower injection force than SS or KP.

Keywords: injection force, insulin pen, KwikPen, Next Generation FlexPen, SoloStar

Expert Opin. Pharmacother. (2009) 10(9):1389-1393

1. Introduction

Insulin pens have several benefits over the more traditional vials and syringes for
insulin injection in people with diabetes; these include improved accuracy [1-7],
flexibility and convenience [8,9], patient preference [9-14] and increased adher-
ence [8,15-17]. Improved adherence could have benefits in the management of
diabetes by helping in achieving glycaemic control [18-20].
The flexibility and convenience of pens (along with their perceived clinical
efficacy and effect on quality of life) are important factors in user preference [14].
A wide variety of design features of pens contribute to patients’ perceptions of
these factors. The force required to inject insulin (injection force) is one factor
that can influence the perception of insulin pens, as many people with diabetes
have impaired manual dexterity [21-25].
Several pens are now available and different attributes of these pens may deter-
mine which pen is chosen. Next Generation FlexPen® (NGFP; Novo Nordisk
A/S, Bagsværd, Denmark) has been developed to improve the usability of this
prefilled pen [26], and in particular, to reduce the injection force, compared with
the ‘conventional’ FlexPen®. NGFP has an injection force ∼ 30% lower than the
injection force of FlexPen [27]. The low injection force of some marketed pens

10.1517/14656560903018929 © 2009 Informa UK Ltd ISSN 1465-6566 1389


All rights reserved: reproduction in whole or in part not permitted
Injection force of insulin pens

may be perceived as an advantage. NGFP has been shown to values for the SS and KP were compared with the NGFP
have a lower injection force than one of these pens, the pre- (the most widely used prefilled pen in Japan) and calculations
filled pen SoloStar® (SS; Sanofi-Aventis, Paris, France), at of significance were made using the Mann-Whitney U test
several injection speeds [28]. Another prefilled pen with a low and Student t-test.
injection force is KwikPen® (KP; Eli Lilly, Indianapolis, IN,
USA: also marketed as MirioPen). The aim of this study was 3. Results
to make a head-to-head comparison with the three most
frequently used prefilled pens, NGFP, SS and KP. With all three pens and with either needle, the injection
force increased with increased injection speed (Figures 1 and 2).
At all three injection speeds and with both needles, NGFP
2. Methods had a significantly lower mean injection force than either
SS or KP, and KP had the highest mean injection force
2.1 Materials
(Figures 1 and 2). The injection force with all three pens was
The three pens used in this study were NGFP filled with
Expert Opin. Pharmacother. Downloaded from informahealthcare.com by Selcuk Universitesi on 01/01/15

lower when the NovoFine® 32G Tip extra thin wall needle
Levemir® (Novo Nordisk) (insulin detemir) randomly
was used than when the BD Micro-Fine™ 31G thin-wall
selected from a single lot (VP52202); SS filled with Lantus®
needle was used.
(Sanofi-Aventis) (insulin glargine) randomly selected from a
The mean injection force (± s.d.) for NGFP, SS and KP
single lot (U002); and Humalog KP (insulin lispro) randomly
with the BD needle at an injection speed of 3.3 mm/s was
selected from a single lot (A421438DD). The needles used
8.1 ± 0.7, 9.2 ± 0.5 and 12.5 ± 1.6 N, respectively; at
were BD Micro-Fine™ 31G thin-wall 5mm needle (Becton
5 mm/s the values were 10.7 ± 1.4, 13.3 ± 0.8 and 16.9 ± 1.2
Dickinson, NJ, USA; lot 8094224A) and NovoFine® 32G
N, respectively; and at 8.3 mm/s the values were 15.6 ± 0.9,
Tip extra thin wall 6 mm needle (Nipro-Novo Nordisk
20.7 ± 2.4 and 24.5 ± 2.6 N, respectively (Figure 1). The
Pharma; lot 07H19S). NGFP was imported from Denmark,
maximum injection force (± s.d.) for NGFP, SS and KP with
as it was not launched at the time of the study. SS and
the BD needle at an injection speed of 3.3 mm/s was 9.4 ±
KP were sourced through the hospital pharmacy. All
For personal use only.

0.5, 9.6 ± 0.4 N (p < 0.05 versus NGFP) and 13.0 ± 1.6 N
injections were made into an injection cushion, which is
(p < 0.05), respectively; at 5 mm/s was 11.3 ± 1.5, 14.0 ± 0.7
purported to simulate adipose tissue. All tests were performed
N (p < 0.05) and 17.7 ± 1.1 N (p < 0.05), respectively; and
in a climate-controlled laboratory at 20 ± 2°C, with relative
at 8.3 mm/s was 16.9 ± 0.4, 23.6 ± 1.0 N (p < 0.05) and
humidity 45 ± 7.5%.
26.4 ± 2.4 N (p < 0.05), respectively.
The mean injection force (± s.d.) for NGFP, SS and KP
2.2 Measurement of injection force with the NovoFine® 32G Tip extra thin wall needle at
All three pens were tested separately with both needles an injection speed of 3.3 mm/s was 5.7 ± 0.4, 6.7 ± 0.3
(the same needle was used on each pen to avoid variation and 9.1 ± 1.3 N, respectively; at 5 mm/s the values were
in measured injection force caused by the flow stress of 8.2 ± 0.7, 10.4 ± 2.1 and 13.1 ± 0.8 N, respectively;
different needles). Injection force was measured at three and at 8.3 mm/s the values were 12.7 ± 0.5, 16.3 ± 1.1
different injection speeds: 3.3, 5 or 8.3 mm/s, and and 21.6 ± 2.0 N, respectively (Figure 2). The maximum
10 measurements were made during the delivery of each 20 injection force (± s.d.) for NGFP, SS and KP with the
U dose. All pens were unused, and so before delivering each NovoFine® 32G Tip extra thin wall needle at an injection
dose the pen was prepared by mounting a needle and speed of 3.3 mm/s was 6.6 ± 0.2, 7.1 ± 0.3 N (p < 0.05
expelling air at 2 U steps, until liquid appeared at the tip versus NGFP) and 10.0 ± 1.7 N (p < 0.05), respectively;
of the needle, confirming the delivery of insulin. The at 5 mm/s was 9.1 ± 0.7, 10.9 ± 2.3 N (p < 0.05) and
pen was then mounted on the testing machine (UTC-500: 13.9 ± 0.7 N (p < 0.05), respectively; and at 8.3 mm/s
A & D, Tokyo, Japan) to deliver the dose at the set was 13.2 ± 0.3, 18.0 ± 0.6 N (p < 0.05) and 23.6 ± 1.3 N
push-button speed. (p < 0.05), respectively.
Delivery of the first dose with each pen can require a
higher injection force owing to a running-in effect, caused 4. Discussion
by siliconisation of the glass cartridge during its manufacture,
which causes friction between the rubber piston and the At three different injection speeds for the delivery of 20 U
cartridge. To allow for this, the measurement of the first insulin, using two different needle types, NGFP had a
dose was excluded from the main calculations. significantly lower mean injection force than SS (by 12 – 25%)
Mean and maximum injection force was measured for and KP (by 35 – 41%). This confirms the low injection force
each individual pen and means (± s.d.) were calculated for previously demonstrated with NGFP (when delivering 60 U
each pen type at each injection speed. The injection force insulin) [27,28]. It has been reported that people with diabe-
was measured at the ‘plateau’ that occurs in the typical tes prefer a lower injection force (which is particularly
injection force spectra at a displacement of 5 – 9 mm. The important in people with impaired manual dexterity), but

1390 Expert Opin. Pharmacother. (2009) 10(9)


Asakura, Seino, Kageyama & Yohkoh

NGFP pens, the injection force with NGFP was 18 – 28% lower
SS than SS [28], a very similar difference to that measured in the
3.3 *
*
KP present study (despite the use of different injection speeds).
It may also have been of interest to capture data on overall
Speed (mm/s)

injection time as reported by Rissler et al. [28]. In this study,


5.0
the time to complete each dose delivery was measured with the
*
* SS and NGFP pens and showed that it took ∼ 2 – 3 s longer to
deliver a 60 U dose with SS than with NGFP at three different
injection speeds. This could be incorporated into the design
8.3 *
of any future studies.
* The study design incorporated the use of needles from two
different manufacturers because the manufacturers of insulin
0 5 10 15 20 25 30 pens recommend different needles. Both recommended needles
Expert Opin. Pharmacother. Downloaded from informahealthcare.com by Selcuk Universitesi on 01/01/15

Force (N) were, therefore, used on all pens. This methodology also allowed
a direct comparison of the three pens without the influence of
needle-related confounding factors.
Figure 1. Mean injection force with NGFP, SS and KP fitted
Three different injection speeds were tested, and
with BD Micro-Fine™ 31G thin-wall needle when delivering
NGFP had the lowest injection force at all three speeds.
20 U at three different injection speeds. Error bars correspond
to s.d.
It is a reasonable supposition that different patients will
* p < 0.05 compared with NGFP. depress the push-button on insulin pens at different
NGFP: Next Generation FlexPen®; SS: SoloStar®; KP: KwikPen®. speeds. Although we are not aware of published data on
the injection speeds commonly used by people with
diabetes, the lower injection force with NGFP across a
NGFP range of injection speeds from 3.3 to 8.3 mm/s suggests
For personal use only.

3.3
SS that most injections in practice would require less force
*
KP with NGFP than with SS or KP. It is interesting to note
*
that at the medium speed (5 mm/s) the injection force
Speed (mm/s)

with the NGFP was similar to that recorded with the SS


5.0 *
at the lower speed (3.3 mm/s). Likewise, the injection
* force with the NGFP at the highest speed (8.3 mm/s)
was similar to that recorded for the KP at the medium
speed. It is possible that in terms of patient preference,
8.3 * the speed of injection would be a deciding factor, rather
* than injection force, by itself.
The injections were all made into an injection cushion,
0 5 10 15 20 25 which is purported to simulate adipose tissue. However, in
Force (N) patients the thickness and distribution of adipose tissue
varies from one individual to another. The use of this
Figure 2. Mean injection force with NGFP, SS and KP fitted
standardised injection medium may be considered to be one
with NovoFine® 32G Tip extra thin wall needle when
of the limitations of this study.
delivering 20 U at three different injection speeds. Error bars
We cannot draw any conclusion on the effects of injection
correspond to s.d. force on patient perception, preference and adherence from
* p < 0.05 compared with NGFP. this technical report. Future studies could assess the role of
NGFP: Next Generation FlexPen®; SS: SoloStar®; KP: KwikPen®. injection force on preference. The significant patient prefer-
ence for NGFP compared with FlexPen may be a result
of this reduced injection force (because this is the main
it is not known how significant these differences in injec- difference between these two pens) [26,27].
tion force (as measured here) would be to the pen-user [22,23,25].
It is clear that further studies to investigate patients’ 5. Conclusions
perceptions of these differences, in a clinical context, would
be beneficial. There is good evidence that NGFP offers some improvements
The only previous comparison of the injection force of to ‘conventional’ FlexPen, such as reduced injection force [27]
NGFP and SS measured a significantly lower injection force and improved colour-coding of the different insulins [26], and
of 18 – 45% with NGFP [28]. However, when the same overall patients prefer NGFP [26]. The proven accuracy of
NovoFine® 32G Tip extra thin wall needle was used on both FlexPen [4,6,7] is also maintained in NGFP [27]. Furthermore,

Expert Opin. Pharmacother. (2009) 10(9) 1391


Injection force of insulin pens

this study and others show that NGFP has advantages used for insulin injection. The impact of improvements to
over SS and KP in terms of improved accuracy and reduced NGFP on outcomes of insulin therapy would be an interesting
injection force [28,29]. area for future research.

6. Expert Opinion Declaration of interest

As well as the low injection force, NGFP has a number of This study was supported by an unrestricted research grant
attributes that could potentially drive user preference, such from Novo Nordisk to Niigata University of Pharmacy. The
as a high level of accuracy and precision [27,29], simplicity publication was supported by Novo Nordisk A/S (Bagsværd,
and comfort of use [26], and a clear differentiation between Denmark) with editorial assistance from ESP Bioscience
insulin types [26]. With patient preference for the method of (Sandhurst, UK). The authors are wholly responsible for the
insulin injection being an important aspect of adherence in study design, technical measurements, analysis and scientific
diabetes management [15], the attributes of NGFP could have evaluation, and all authors were responsible for the content
an impact on adherence and, therefore, on the choice of pen of this manuscript and approved the submitted draft.
Expert Opin. Pharmacother. Downloaded from informahealthcare.com by Selcuk Universitesi on 01/01/15

Bibliography 9. Korytkowski M, Bell D, Jacobsen C, et al. 15. Lee WC, Balu S, Cobden D, et al.
Papers of special note have been highlighted A multicenter, randomized, open-label, Medication adherence and the associated
as either of interest (•) or of considerable comparative, two-period crossover trial of health-economic impact among patients
interest (••) to readers. preference, efficacy, and safety profiles of a with type 2 diabetes mellitus converting to
prefilled, disposable pen and conventional insulin pen therapy: an analysis of
1. Keith K, Nicholson D, Rogers D. Accuracy
vial/syringe for insulin injection in patients third-party managed care claims data.
and precision of low-dose insulin
with type 1 or 2 diabetes mellitus. Clin Ther 2006;28:1712-25
administration using syringes, pen
Clin Ther 2003;25:2836-48 16. Pawaskar MD, Camacho FT,
injectors, and a pump. Clin Pediatr (Phila)
•• A well-designed trial demonstrating the Anderson RT, et al. Health care
2004;43:69-74
For personal use only.

patient preference for insulin pens over costs and medication adherence
2. Lteif AN, Schwenk WF. Accuracy of pen vial and syringe. associated with initiation of insulin
injectors versus insulin syringes in children
10. Summers KH, Szeinbach SL, Lenox SM. pen therapy in medicaid-enrolled
with type 1 diabetes. Diabetes Care
Preference for insulin delivery patients with type 2 diabetes: a
1999;22:137-40
systems among current insulin retrospective database analysis.
3. Coscelli C, Lostia S, Lunetta M, et al. users and nonusers. Clin Ther Clin Ther 2007;29:1294-305
Safety, efficacy, acceptability of a pre-filled 2004;26:1498-505 17. Lombardo F, Salzano G, Messina MF,
insulin pen in diabetic patients over
11. Albano S. Assessment of quality of et al. Compliance and administration
60 years old. Diabetes Res Clin Pract
treatment in insulin-treated patients with methods in management of type 1
1995;28:173-7
diabetes using a pre-filled insulin pen. The diabetes. Acta Biomed
4. Asakura T. Comparison of the dosing ORBITER Study Group. Acta Biomed 2005;76(Suppl 3):66-9
accuracy of two insulin injection devices. 2004;75:34-9 18. Cramer JA, Pugh MJ. The influence of
J Clin Res 2005;8:33-40
12. Thurman JE. Insulin pen injection insulin use on glycemic control: How well
5. Asakura T, Seino H. Insulin devices for management of patients do adults follow prescriptions for insulin?
dosing accuracy with FlexPen® versus with type 2 diabetes: considerations based Diabetes Care 2005;28:78-83
syringe by healthcare professionals and on an endocrinologist’s practical experience 19. Donnelly LA, Morris AD,
healthy volunteers. Diabetes in the United States. Endocr Pract Evans JM. Adherence to insulin
2005;54:A498-9 2007;13:672-8 and its association with glycaemic control
6. Asakura T, Seino H, Kageyama M, et al. 13. Stockl K, Ory C, Vanderplas A, et al. in patients with type 2. diabetes QJM
Dosing accuracy of two insulin pre-filled An evaluation of patient preference 2007;100:345-50
pens. Curr Med Res Opinion for an alternative insulin delivery 20. Morris AD, Boyle DI, McMahon AD,
2008;24:429-34 system compared to standard vial et al. Adherence to insulin
7. Hänel H, Weise A, Sun W, et al. and syringe. Curr Med Res Opin treatment, glycaemic control, and
Differences in the dose accuracy of 2007;23:133-46 ketoacidosis in insulin-dependent diabetes
insulin pens. J Diabetes Sci Tech 14. Rubin RR, Peyrot M. Quality of life, mellitus. The DARTS/MEMO
2008;2:478-81 treatment satisfaction, and treatment Collaboration. Diabetes Audit and
8. Graff MR, McClanahan MA. Assessment preference associated with use of a pen Research in Tayside Scotland. Medicines
by patients with diabetes mellitus of device delivering a premixed 70/30 insulin Monitoring Unit. Lancet
two insulin pen delivery systems versus aspart suspension (aspart protamine 1997;350:1505-10
a vial and syringe. Clin Ther suspension/soluble aspart) versus alternative 21. Frey MI, Barrett-Connor E, Sledge PA,
1998;20:486-96 treatment strategies. Diabetes Care et al. The effect of noninsulin dependent
2004;27:2495-7 diabetes mellitus on the prevalence of

1392 Expert Opin. Pharmacother. (2009) 10(9)


Asakura, Seino, Kageyama & Yohkoh

clinical osteoarthritis. A population based DiaCAN Multicenter Study Group. 29. Weise A, Pfutzner JW, Borig J,
study. J Rheumatol 1996;23:716-22 Diabet Med 1993;10(Suppl 2):82S-6S et al. Comparison of the dose
22. Savas S, Koroglu BK, Koyuncuoglu HR, 26. Sommavilla B, Jorgensen C, Jensen KH. accuracy of prefilled insulin pens.
et al. The effects of the diabetes related Safety, simplicity and convenience of a J Diabetes Sci Tech 2009;3:149-53
soft tissue hand lesions and the reduced modified prefilled insulin pen. Expert Opin
hand strength on functional disability Pharmacother 2008;9:2223-32 Affiliation
of hand in type 2 diabetic patients. T Asakura†1, H Seino2, M Kageyama1 &
27. Pfützner A, Reimer T, Frokjaer LPF, et al.
Diabetes Res Clin Pract 2007;77:77-83 N Yohkoh1
Prefilled insulin device with reduced
•• Discusses the reduce hand-strength in †Author for correspondence
injection force: patient perception and
some patients with diabetes, which may 1Department of Clinical Pharmacy,
accuracy. Curr Med Res Opin
influence patient preference for lower Faculty of Pharmaceutical Sciences
2008;24:2545-9
injection force devices. Niigata University of Pharmacy and
•• First report on Next Generation FlexPen,
23. Schady W, Abuaisha B, Boulton AJ. which shows the reduced injection force Applied Life Sciences,
Observations on severe ulnar neuropathy 265-1 Higashizima, Akiha-ku,
Expert Opin. Pharmacother. Downloaded from informahealthcare.com by Selcuk Universitesi on 01/01/15

compared with FlexPen and patient


in diabetes. J Diabetes Complications perception of the two pens. Niigata, 956-8603 Japan
1998;12:128-32 Tel: +81 (0)250 25 5287; Fax: +81 (0)250 25 5287;
28. Rissler J, Jorgensen C, Rye Hansen M, E-mail: asa-mac@m7.dion.ne.jp
24. Starkman HS, Gleason RE, Rand LI, et al. et al. Evaluation of injection force 2Seino Clinic
Limited joint mobility (LJM) of the hand
dynamics of a modified prefilled insulin 6-192-2 Kaisei,
in patients with diabetes mellitus: relation pen. Expert Opin Pharmacother Koriyama, 963-8851, Japan
to chronic complications. Ann Rheum Dis 2008;9:2217-22
1986;45:130-5 •• Demonstrates the reduced injection force
25. Ziegler D, Gries FA, Spuler M, et al. of Next Generation FlexPen compared
The epidemiology of diabetic neuropathy. with SoloStar.
For personal use only.

Expert Opin. Pharmacother. (2009) 10(9) 1393


Expert Opin. Pharmacother. Downloaded from informahealthcare.com by Selcuk Universitesi on 01/01/15
For personal use only.

You might also like