Calypsoo Probe Calibration Moving Beyond Just 4 Positions
Calypsoo Probe Calibration Moving Beyond Just 4 Positions
Calypso training includes stylus system calibration and in these classes you art taught to load the MasterProbe and define the reference sphere position. Current Issue
While this article may be redundant for some users, for others you will learn there are any number of angle options to locate the reference sphere.
Past Articles
Training
Recently, I was at a customer's facility and had to use a star probe shown below.
Calypso
CMM Manager
DMIS Corner
General
GD&T
MCOSMOS
PC-DMIS
As you can see the stylus pointing upward would be difficult to calibrate if the sphere position was at one of the four basic positions offered in the Ref. Today I Learned
Sphere Position dialog box. This stylus system has to be measured with the MasterProbe shaft pointing in the +X direction.
CMMQ Publications
When given the four basic sphere positions simply type in the angle of the MasterProbe shaft direction in the Rotation window as shown below
CMM TRAINING
Enter the proper angle for the direction your stylus system requires.
Basic Angles
Masterball at 0 degrees
Masterball at 90 degrees
So don't struggle with trying to change the sphere coverage area to accommodate your stylus angle, just find the correct position that will allow you to
calibrate properly.