Results
Kinetex® 3.5 µm PAH is the first core-shell based column made With the additional resolution provided by the Kinetex 3.5 µm PAH
specifically for the analysis of PAHs, bringing the efficiency bene- core-shell media, we can now sacrifice a little efficiency by using
fits of core-shell morphology to the field of PAH analysis. Figure a shorter column. This will allow us to further reduce our analysis
1 is a representative chromatogram obtained using a fully porous times. This is shown in Figure 3, where we have moved to a 100 x
column marketed specifically for PAH analysis (Waters® 5 µm PAH 4.6 mm Kinetex 3.5 µm PAH column and a shorter gradient profile.
C18, 250 x 4.6 mm), the gradient and flow rate were reproduced Under these new conditions, and with this shorter column format,
from the manufacturers product marketing literature. Using that our total cycle time is about 9 minutes, which is about 1/3 of the
column and running conditions, you can see that all 18 analytes original method. That is almost a 3-fold increase in productivity.
in EPA Method 8310 are baseline resolved, with a total cycle time Under these conditions, the minimum Rs value is 1.97 between
of ~28 minutes with re-equilibration. Under these conditions, the 2-methylnaphthalene and acenaphthene, which is still better than
lowest resolution is obtained between peaks 4 and 5, 2-methyl- the original resolution value obtained using the starting method
naphthalene and acenaphthene, with an Rs of 1.92. on the Waters 5 µm PAH C18 250 x 4.6 mm column.
Figure 1. Separation of the 18 EPA Method 8310 PAHs using a fully porous Figure 3. Accelerated analysis of the EPA Method 8310 PAH mixture using
Waters 5 μm PAH C18 250 x 4.6 mm column. a 100 x 4.6 mm column packed with Kinetex 3.5 µm PAH core-shell media.
mAU
mAU
16
60 200
16
50 175
150
40
App ID 24596
15 125
30 11 14
App ID 24598
13 100
20 17 15
11 14
7 18 75
10 56 12
13 17
3 18
2 9 50 7
1 4 10 5
0 8 3 6 12
25 2 9 10
1 4 8
0 2.5 5 7.5 10 12.5 15 17.5 20 22.5 min 0
When run using the identical running conditions, the Kinetex The chromatograms in Figures 1-3 are shown in the same time
3.5 µm PAH, 250 x 4.6 mm core-shell column has a total cycle scale so that you can clearly see how the run time is decreased.
time of about 20 minutes (with re-equilibration), which is already In Figure 4, we have overlaid the original chromatogram obtained
about 25 % faster than the fully porous Waters 5 µm PAH C18 using the Waters 5 µm PAH C18 250 x 4.6 mm and the Kinetex
250 x 4.6 mm column. Using the Kinetex 3.5 µm PAH column, the 3.5 µm PAH 100 x 4.6 mm column, but adjusted the x-axis time
minimum resolution value occurs between the same two peaks, scale so that you can better compare the resolution on the 9
2-methylnaphthalene and acenaphthene, with an Rs of 2.86, minute Kinetex method to the 28 minute original method. You
which is 48 % better than the resolution for those peaks on the can clearly see how well the resolution has been preserved even
Waters 5 µm PAH C18 250 x 4.6 mm column (Rs = 1.92). though the run time is only about 1/3 as long as the original meth-
od. Also note the increase in peak height that is achieved using
Thus, by simply moving from the fully porous Waters 5 µm PAH the Kinetex 3.5 µm PAH column. For instance, the peak height
C18 250 x 4.6 mm column to the Kinetex 3.5 µm PAH 250 x 4.6 mm response for Dibenz[a,h]anthracene, the largest peak, is about 80
core-shell column under identical conditions, we increased our mAU using the Waters 5 µm PAH C18 250 x 4.6 mm column. But
resolution value by 48 % and also reduced our analysis time by that same sample injection yields a peak height response of ~220
~25 %. This is possible because of the high efficiency advantage mAU on the Kinetex 3.5 µm PAH 100 x 4.6 mm column, almost a
of the Kinetex core-shell particle morphology, and is further en- three-fold increase in sensitivity.
hanced by the slightly smaller particle diameter
Figure 2. Separation of the 18 EPA Method 8310 PAHs using a core-shell Kinetex
3.5 μm PAH 250 x 4.6 mm column (same running conditions as Figure 1)
mAU 16
100
80
60 15
11 14
App ID 24597
40 13 17
18
7
20 3 56 12
2 9
1 4 10
0 8
Comparative separations may not be representative of all applications.
0 2.5 5 7.5 10 12.5 15 17.5 min
Page 2 of 4
TN-1228
Conclusions
Figure 4. Comparison of the original method obtained using the Waters ® The new Kinetex 3.5 µm PAH introduces the core-shell efficiency
5 µm PAH C18 250 x 4.6 mm (TOP) with the same sample analyzed using advantage to the field of PAH analysis. The ultra-high efficiency of
the 100 x 4.6 mm column packed with Kinetex® 3.5 µm PAH core-shell me- the core-shell morphology can provide the analyst with the ability
dia (BOTTOM). to dramatically improve their resolving power and their productivi-
ty. In this case, baseline resolution of the 18 component PAH mix-
mAU ture specified in EPA Method 8310 is achieved with a total cycle
200
time of less than 10 minutes, compared to about 30 minutes using
175
150
a typical fully porous alternative column (Waters 5 µm PAH C18).
125
100
75 16
References
50
11 14 15 1. Kuo et al. 1998; Wang et al. 2002
13 17
25 7 18
2 3456 12
1 8 9 10
0
0 2.5 5 7.5 10 12.5 15 17.5 20 22.5 min
mAU 16
200
175
150
125
11 15
100 14
75 13 17
18
50 7
5 12
3 6
25 2 9
1 4 10
8
0
0 2.5 5 7.5 10 12.5 15 17.5 20 22.5 min
APPLICATIONS
Australia Luxembourg
t: +61 (0)2-9428-6444 t: +31 (0)30-2418700
f: +61 (0)2-9428-6445 f: +31 (0)30-2383749
auinfo@phenomenex.com nlinfo@phenomenex.com
Austria Mexico
t: +43 (0)1-319-1301 t: 01-800-844-5226
f: +43 (0)1-319-1300 f: 001-310-328-7768
anfrage@phenomenex.com tecnicomx@phenomenex.com
www.phenomenex.com/TermsAndConditions.
f: +39 051 6327555 f: +1 (310) 328-7768 Trademarks
italiainfo@phenomenex.com info@phenomenex.com Kinetex is a registered trademark and SecurityGuard and MidBore are trademarks of Phe-
nomenex. Waters is a registered trademark of Waters Technologies Corporation.