You are on page 1of 48

LCP Compact Hand.

The modular
hand system.

Surgical Technique

This publication is not intended for


distribution in the USA.

Instruments and implants


approved by the AO Foundation.
Image intensifier control

This description alone does not provide sufficient background for direct use
of DePuy Synthes products. Instruction by a surgeon experienced in handling
these products is highly recommended.

Processing, Reprocessing, Care and Maintenance


For general guidelines, function control and dismantling of m ­ ulti-part
instruments, as well as processing guidelines for i­mplants, please contact your
local sales representative or refer to:
http://emea.depuysynthes.com/hcp/reprocessing-care-maintenance
For general information about reprocessing, care and maintenance of Synthes
reusable devices, instrument trays and cases, as well as processing of Synthes
non-sterile implants, please consult the Important Information leaflet
(SE_023827) or refer to:
http://emea.depuysynthes.com/hcp/reprocessing-care-maintenance
Table of Contents

Introduction LCP Compact Hand2

AO Principles4

Intended Use and Indications5

Surgical Techniques Preparation6

Fracture Reduction and Plate Contouring7

Screw Insertion11
– Cortex Screws 12
– Locking Screws 18

Implant Removal23

Additional Information24

Product Information Implants Compact Hand 1.0/1.325

Instruments Compact Hand 1.0/1.327

Implants LCP Compact Hand 1.528

Instruments LCP Compact Hand 1.532

Implants LCP Compact Hand 2.033

Instruments LCP Compact Hand 2.037

Implants LCP Compact Hand 2.438

Instruments LCP Compact Hand 2.441

Sets42

Standard Instrument Set for LCP Compact Hand43

MRI Information 45

LCP Compact Hand  Surgical Technique  DePuy Synthes    1


LCP Compact Hand. The modular
hand system.

Adaptable Locking Plates for


Various Indications

The modular LCP Compact Hand S­ ystem offers a wide range


of ­anatomical and adaptable plates. All implants are available
in titanium and stainless steel.

LCP Compact Hand implants are intended for the following


indications:

Compact Hand 1.0/1.3


–– Fixation of small fragments
–– Avulsion fractures and fractures of distal and middle
­phalanges

LCP Compact Hand 1.5


–– Fractures, osteotomies and arthrodeses of the phalanges
and metacarpals
–– Replantations and reconstructions of the phalanges and
metacarpals

LCP Compact Hand 2.0


–– Fractures of the phalanges, metacarpals, wrist bones and
distal radius (double-plate technique)
–– Osteotomies and arthrodeses of the interphalangeal joints

LCP Compact Hand 2.4


–– Fractures of the metacarpals, wrist bones and distal radius
(double-plate technique)
–– Osteotomies and arthrodeses on the hand

Low profile plates and Anatomically


fully countersunk screws precontoured Rotation Correction Plate LCP combi-hole

Minimizes soft tissue Minimizes need for bending. Allows intraoperative inspec- Intraoperative choice
­irritation. tion of the reduction and ­between compression and
correction of the rotational angular stable locking.
axis.

2    DePuy Synthes  LCP Compact Hand  Surgical Technique


Part of the LCP Compact Hand System

4 modules, 5 dimensions:
–– Adaptable plates for different indications
–– Complete range of implants
1.0 mm
1.3 mm
1.5 mm (new locking)
2.0 mm (locking)
2.4 mm (locking)

–– Broad range of non-locking, locking and LCP plates to


­address various fracture patterns.
–– The use of locking screws allows creating a fixed-angle
­construct, which is particularly advantageous in osteopenic
bone or in fractures near the joint.

LCP Compact Hand  Surgical Technique  DePuy Synthes    3


AO Principles

AO PRINCIPLES

In 1958, the AO formulated four basic principles, which have


In 1958, the guidelines
become AO formulated four basic
for internal principles,
fixation1,2
. which
have become the guidelines for internal fixation1, 2.

4_Priciples_03.pdf 1 05.07.12 12:08

Anatomic
Anatomic reduction fixation
Stable fixation
Fracture reduction
Fracture reduction and
and fixation
fixation to Fracture
Fracture fixation
fixationproviding
providingabsolute
abso- or
to restore
restore anatomical
anatomical relationships.
relationships. relative stability,stability,
lute or relative as required
as by the
­prequired
atient, theby injury, and thethe
the patient, personality
injury,
1 2 of
andthethe
fracture.
personality of the
fracture.

Early,
Early, active mobilization
mobilization Preservationof
Preservation of blood
blood supply
supply
Early and
Early and safe mobilization
mobilizationand
and 4 3 Preservation
Preservationof ofthe
theblood
bloodsupply
supply
rehabilitation the injured partand
rehabilitation of the injured part to
to soft tissues and bone bygentle
soft tissues and bone by
the patient
and as a whole.
the patient as a whole. reduction techniques
gentle reduction and careful
techniques and
handling.
careful handling.

1
Müller ME, M Allgöwer, R Schneider, H Willenegger. Manual of Internal
Fixation. 3rd ed. Berlin Heidelberg New York: Springer. 1991.
2
Rüedi TP, RE Buckley, CG Moran. AO Principles of Fracture Management.
2nd ed. Stuttgart, New York: Thieme. 2007.

1
 Müller ME, Allgöwer M, Schneider R, Willenegger H. Manual of Internal Fixation.
3rd ed. Berlin, Heidelberg, New York: Springer. 1991.
2
 Rüedi TP, Buckley RE, Moran CG. AO Principles of Fracture Management. 2nd ed.
Stuttgart, New York: Thieme. 2007.

4 DePuy Synthes Expert Lateral Femoral Nail Surgical Technique


4    DePuy Synthes  LCP Compact Hand  Surgical Technique
Intended Use and Indications

Intended Use
LCP Compact Hand Implants are intended for temporary
fixation, correction or stabi­lization of bones in anatomical
region of the distal radius and hand

Indications
Compact Hand 1.0/1.3
General:
–– Fixation of small fragments
–– Avulsion fractures and fractures of the distal and
middle phalanges
Straight plate:
–– Shaft fractures of phalanges
Strut plate:
–– Comminuted fractures and replantations of middle or
proximal phalanges
T-plate and Y-plate:
–– Condylar fractures of phalanges
Single cortex screws:
–– Capturing of fragments with lag screw technique

LCP Compact Hand 1.5


–– Fracture fixation of the phalanges and metacarpals
–– Osteotomies
–– Arthrodeses
–– Replantations and reconstructions of phalanges and
metacarpals, particularly in osteopenic bone

LCP Compact Hand 2.0


–– Fractures of the phalanges, metacarpals and wrist bones
–– Osteotomies and arthrodeses of the interphalangeal joints
–– Fractures of distal radius (double-plate technique)
–– Subcapital radial head fractures

LCP Compact Hand 2.4


–– Fractures of the metacarpal and wrist bones
–– Fractures of distal radius (double-plate technique)
–– Osteotomies and arthrodeses on the hand
–– Subcapital radial head fractures

LCP Compact Hand  Surgical Technique  DePuy Synthes    5


Preparation

1
Select implant

Select the plates according to the fracture pattern and anat-


omy of the bone.

Notes:
–– This surgical technique describes the application of the
LCP Compact Hand system using a Locking Condylar
Plate 1.5.
–– The plates of the LCP Compact Hand Module 1.5 are
available either with or without short drill sleeves
­preassembled to the plate. Plates with preassembled drill
sleeves are available sterile only.

6    DePuy Synthes  LCP Compact Hand  Surgical Technique


Fracture Reduction and
Plate Contouring

1
Reduce fracture

Reduce the fracture under image intensification and, if


­necessary, fix with Kirschner wires or reduction forceps.
The reduction method is fracture-specific.

2
Trim plate

Instrument

391.962 Bending/Cutting Pliers

Trim the plate to the desired length using bending/cutting


pliers. Remove the burrs.

LCP Compact Hand  Surgical Technique  DePuy Synthes    7


Fracture Reduction and Plate Contouring

3
1
Contour plate

Instruments

347.901 Pliers, flat-nosed, pointed, for


Plates 1.0 to 2.4

For LCP Compact Hand 1.5

03.114.003 Bending Pin for LCP Plates 1.5, with thread

For LCP Compact Hand 2.0

329.921 Bending Pin for LCP Plates 2.0, with thread

For LCP Compact Hand 2.4

329.922 Bending Pin for LCP Plates 2.4 and


2.7, with thread

If necessary, contour the plate to suit the anatomical


­conditions.

Standard plates
Use bending pins for LCP Plates and thread the pins into the
screw holes to contour the plate. (1)

The plate can also be bent using two flat nosed pliers. (2)

8    DePuy Synthes  LCP Compact Hand  Surgical Technique


Preassembled plates
(LCP Compact Hand 1.5 only) 3
Use the non-threaded ends of the bending pins over the
short threaded drill sleeves. (3)

Precautions:
–– If possible, bend the plate between the locking/LCP holes.
Do not deform the threaded part of the holes or over-
bend the plate as this may adversely affect insertion of
locking screws. Avoid repetitive bending of the plate.
–– Reverse bending or use of the incorrect instrumentation
for bending may weaken the plate and lead to premature
plate failure (e.g. breakage). Do not bend the plate be-
yond what is required to match the anatomy.

LCP Compact Hand  Surgical Technique  DePuy Synthes    9


Fracture Reduction and Plate Contouring

4
Position plate 1

Instrument

399.970 Reduction Forceps with Points,


ratchet lock, length 130 mm

Position the plate over the reduced fracture and, if necessary


fix provisionally with Kirschner wires or reduction forceps. (1)

Preassembled plates
(LCP Compact Hand 1.5 only)
When using reduction forceps with points on a preassembled
plate, insert one branch of the forceps in the cannulation of
the short drill sleeve to secure the plate to the bone. (2)

Optional Instrument

398.819 Holding Forceps with Swivel Foot,


for Plates 1.3 to 2.4, speed lock

When using the holding forceps with swivel foot to hold the 2
plate, remove the drill sleeves from the surrounding holes so
that the holding forceps sit flush on the plate.

11    DePuy Synthes  LCP Compact Hand  Surgical Technique


Screw Insertion

Determine screw type LCP combi-hole:


Depending on the individual case, cortex and/or locking Cortex screw in the com-
screws may be inserted. Determine where locking screws pression portion (1), locking
and where cortex screws will be used. screw in the threaded
­portion of the hole (2)
The final screw placement and the use of locking and cortex 1 2
screws are determined by the fracture pattern.

If a locking screw is inserted first, ensure that the plate is


held securely to the bone to prevent the plate from spinning
as the screw locks into the plate.

Locking hole:
Option
Cortex or locking screw
If angularly-stable buttress pins are used, one screw per bone
fragment must be inserted additionally.

Note: For further details regarding the application of


LCP systems consult the corresponding technique guide
(DSEM/TRM/0115/0278).

Compression hole:
Cortex screw only

Elongated compression
hole:
Cortex screw only

Locking screw Cortex screw

LCP Compact Hand  Surgical Technique  DePuy Synthes    11


Screw Insertion
Cortex Screws

1
Pre-drill screw hole for cortex screw

Instruments

For Compact Hand 1.0/1.3

312.133 Double Drill Guide 1.3/1.0

513.005 Drill Bit B 1.0 mm, length 46/34mm,


2-flute, for Mini Quick Coupling

513.035 Drill Bit B 1.3 mm, length 46/34 mm,


2-flute, for Mini Quick Coupling

For LCP Compact Hand 1.5

312.140 Double Drill Guide 1.5/1.1,


for No. 311.150

03.114.007 Drill Bit B 1.1 mm, length 75/61 mm,


2-flute, for Mini Quick Coupling

310.141 Drill Bit B 1.5 mm, length 65/50 mm,


2-flute, for Mini Quick Coupling,
Stainless Steel

For LCP Compact Hand 2.0

323.200 Universal Drill Guide 2.0

310.507 Drill Bit B 1.5 mm, with marking,


length 96/82 mm, 2-flute, for
Mini Quick Coupling

513.140 Drill Bit B 2.0 mm, length 67/55 mm,


2-flute, for Mini Quick Coupling

For LCP Compact Hand 2.4

323.202 Universal Drill Guide 2.4

310.508 Drill Bit B 1.8 mm, with marking,


length 96/82 mm, 2-flute, for
Mini Quick Coupling

317.871 Drill Bit B 2.4 mm, length 80/66 mm,


2-flute, for Mini Quick Coupling

11    DePuy Synthes  LCP Compact Hand  Surgical Technique


Standard plates
Drill the screw hole using the drill bit and the corresponding 1
drill guide / drill sleeve. (1)

Use the following drill diameter for the threaded respectively


the gliding hole:


B Threaded hole B Gliding hole

Compact Hand 1.3 1.0 1.3

Compact Hand 1.5 1.1 1.5

Compact Hand 2.0 1.5 2.0

Compact Hand 2.4 1.8 2.4

When using a LCP hole, drill the screw hole either neutrally
or off-centre (compression).

Preassembled plates
(LCP Compact Hand 1.5 only)
Drill the screw hole using the short LCP drill sleeves that are
preassembled to the plate. To drill for a cortex screw in an
angle other than 90° to the plate, use the double drill guide. 2
Before using the double drill guide, remove the short drill
sleeve. (2)

LCP Compact Hand  Surgical Technique  DePuy Synthes    11


Screw Insertion
Cortex Screws

2
1
Determine screw length of cortex screw

Instruments

For Compact Hand 1.0/1.3

319.003 Depth Gauge for Screws B 1.3 to 1.5 mm,


measuring range up to 24mm

For LCP Compact Hand 1.5

319.003 Depth Gauge for Screws B 1.3 to 1.5 mm,


measuring range up to 24 mm

For LCP Compact Hand 2.0

319.005 Depth Gauge for Screws B 2.0 and


2.4 mm, measuring range up to 40 mm

For LCP Compact Hand 2.4

319.005 Depth Gauge for Screws B 2.0 and


2.4 mm, measuring range up to 40 mm

Standard plates
Determine the screw length with the corresponding depth
gauge. (1)

11    DePuy Synthes  LCP Compact Hand  Surgical Technique


Preassembled plates
(LCP Compact Hand 1.5 only) 2
Screw length may be measured either with of without the
short drill sleeve assembled to the plate (2, 3). When meas­
ured directly through the drill sleeve, subtract 5 mm from the
reading to account for the drill sleeve. (3)

LCP Compact Hand  Surgical Technique  DePuy Synthes    11


Screw Insertion
Cortex Screws

3
1
Insert cortex screw

Instruments

For Compact Hand 1.0/1.3

314.413 Screwdriver Shaft 1.3, cruciform,


with Holding Sleeve, length 62 mm,
for Mini Quick Coupling

311.012 Handle, medium,


with Mini Quick Coupling

For LCP Compact Hand 1.5

03.114.009 Screwdriver Shaft, Stardrive 1.5, with


Holding Sleeve, length 66 mm,
for Mini Quick Coupling

311.012 Handle, medium,


with Mini Quick Coupling

For LCP Compact Hand 2.0

314.676 Screwdriver Shaft Stardrive 2.0,


with Holding Sleeve, length 66 mm,
for Mini Quick Coupling

311.012 Handle, medium,


with Mini Quick Coupling

For LCP Compact Hand 2.4

314.467 Screwdriver Shaft, Stardrive, T8,


self-holding

311.430 Handle with Quick Coupling,


length 110 mm

Standard plates
Insert the self-tapping cortex screw using the corresponding
screwdriver. (1)

11    DePuy Synthes  LCP Compact Hand  Surgical Technique


Preassembled plates
(LCP Compact Hand 1.5 only) 2
Remove preassembled short drill sleeves. A tray which can
serve as a tool to help removing the short drill sleeves from
the screwdriver is optionally available. (2)

Insert the self-tapping cortex screw using the corresponding


screwdriver. (3) 3

Notes:
–– It is recommended to count the short LCP drill sleeves to
ensure they are all removed before closure.
–– Find tips regarding how to properly pick-up screws with
Stardrive recess on page 24.

LCP Compact Hand  Surgical Technique  DePuy Synthes    11


Screw Insertion
Locking Screws

1
1
Pre-drill screw hole for locking screw

Instruments

For LCP Compact Hand 1.5

03.114.001 LCP Drill Sleeve 1.5, for Drill Bits B 1.1 mm

03.114.007 Drill Bit B 1.1 mm, length 75/61 mm,


2-flute, for Mini Quick Coupling

For LCP Compact Hand 2.0

323.034 LCP Drill Sleeve 2.0, with Scale,


for Drill Bits B 1.5 mm with marking

310.507 Drill Bit B 1.5 mm, with marking,


length 96/82 mm, 2-flute,
for Mini Quick Coupling

For LCP Compact Hand 2.4

323.029 LCP Drill Sleeve 2.4, with Scale up to


30 mm, for Drill Bits B 1.8 mm

310.508 Drill Bit B 1.8 mm, with marking, 2


length 96/82 mm, 2-flute,
for Mini Quick Coupling

Standard plates
Drill screw hole through the LCP drill sleeve using the
­appropriately sized drill bit. (1)

Preassembled plates
(LCP Compact Hand 1.5 only)
Drill the screw hole using the short LCP drill sleeves that are
preassembled to the plate. (2)

11    DePuy Synthes  LCP Compact Hand  Surgical Technique


2
1
Determine screw length of locking screw

Instruments

For LCP Compact Hand 1.5

319.003 Depth Gauge for Screws B 1.3 to 1.5 mm,


measuring range up to 24 mm

For LCP Compact Hand 2.0

323.034 LCP Drill Sleeve 2.0, with Scale,


for Drill Bits B 1.5 mm with marking

319.005 Depth Gauge for Screws, B 2.0 and


2.4 mm, measuring range up to 40 mm

For LCP Compact Hand 2.4

323.029 LCP Drill Sleeve 2.4, with Scale up to


30 mm, for Drill Bits B 1.8 mm

319.005 Depth Gauge for Screws, B 2.0 and


2.4 mm, measuring range up to 40 mm
2
Standard plates
Determine the screw length either with the corresponding
depth gauge (1) or with the LCP drill sleeve (LCP Compact
Hand 2.0/2.4, marking on drill bit). (2)

When using the LCP drill sleeve check the length directly on
the scale of the drill sleeve. Then remove the drill sleeve.

LCP Compact Hand  Surgical Technique  DePuy Synthes    11


Screw Insertion
Locking Screws

Preassembled plates
(LCP Compact Hand 1.5 only) 3
Screw length may be measured either with of without the
short drill sleeve assembled to the plate. When measured
­directly through the drill sleeve, subtract 5 mm from the
reading to account for the drill sleeve. (3)

22    DePuy Synthes  LCP Compact Hand  Surgical Technique


3
1
Insert locking screw

Instruments

For LCP Compact Hand 1.5

03.114.009 Screwdriver Shaft, Stardrive 1.5,


with Holding Sleeve, length 66 mm,
for Mini Quick Coupling

311.012 Handle, medium,


with Mini Quick Coupling

For LCP Compact Hand 2.0

314.676 Screwdriver Shaft Stardrive 2.0,


with Holding Sleeve, length 66 mm,
for Mini Quick Coupling

311.012 Handle, medium,


with Mini Quick Coupling

511.777 Torque Limiter, 0.4 Nm,


with AO/ASIF Quick Coupling

For LCP Compact Hand 2.4

314.467 Screwdriver Shaft, Stardrive, T8,


self-holding

311.430 Handle with Quick Coupling,


length 110 mm

511.776 Torque Limiter, 0.8 Nm,


with AO/ASIF Quick Coupling

Standard plates
Insert the self-tapping locking screw using the corresponding
screwdriver shaft and the appropriate handle. (1)

LCP Compact Hand  Surgical Technique  DePuy Synthes    22


Screw Insertion
Locking Screws

Preassembled plates
(LCP Compact Hand 1.5 only) 2
Remove preassembled short drill sleeves. Insert the self-tap-
ping locking screw using the corresponding screwdriver. (2)

A tray which can serve as a tool to help removing the short


drill sleeves from the screwdriver is optionally available.

Note: It is recommended to count the short LCP drill sleeves


to ensure they are all removed before closure.

LCP Compact Hand 2.0/2.4


For mechanical insertion use the corresponding torque
­limiting attachment.

The torque limiter prevents over-tightening and ensures that


the locking screws are securely locked into the plate.

Use manual locking to perform the final locking step for


2.0 mm and 2.4 mm locking screws. Mechanical insertion is
not recommended.

22    DePuy Synthes  LCP Compact Hand  Surgical Technique


Implant Removal

To remove the plate, first unlock all screws then definitely


­remove them in a second step. If the screws are not
­unlocked before removal the plate may rotate while the last
screw is being removed and cause soft tissue damage.

LCP Compact Hand  Surgical Technique  DePuy Synthes    22


Additional Information

Pick-up screws with Stardrive recess


without holding sleeve
Select the screw and position the corresponding screwdriver
perpendicular to the screw.

Pick-up the screw by applying slight axial pressure.

Make sure that the screwdriver remains perpendicular during


screw extraction.

Pick-up screws with holding sleeve


All screws can also be picked-up with help of the correspon­
ding holding sleeves.

Slide the holding sleeve onto the screwdriver, until it clicks


into place. With the holding sleeve jaws open, mount the
­appropriate screw onto the screwdriver, then push the hold-
ing sleeve until it secures the screw.

Note: The holding sleeve covers the head of the screw.

22    DePuy Synthes  LCP Compact Hand  Surgical Technique


Implants Compact Hand 1.0/1.3

Plates

Adaption Plate 1.3, straight


Art. No. Holes Length (mm)
X21.306 6 23
X21.312 12 48

T-Adaption Plate 1.3


Art. No. Holes Length (mm)
X21.333 Shaft 8, head 3 35.5
X21.334 Shaft 8, head 4 35.5

Y-Adaption Plate 1.3


Art. No. Holes Length (mm)
X21.335 Shaft 8, head 3 37

Strut Plate 1.3


Art. No. Holes Length (mm)
X21.320   8 19
X21.321   8 19

X = 2: Stainless Steel


X = 4: Titanium

Implants are available non-sterile or sterile packed.


Add suffix “S” to article number to order sterile product.

LCP Compact Hand  Surgical Technique  DePuy Synthes    22


Implants Compact Hand 1.0/1.3

Screws

Cortex Screws B 1.0 mm, self-tapping


Art. No. Length (mm)
X00.526   6
X00.527   7
X00.528   8
X00.529   9
X00.530 10
X00.531 11
X00.532 12
X00.533 13
X00.534 14

Cortex Screws B 1.3 mm, self-tapping


Art. No. Length (mm)
X00.686   6
X00.687   7
X00.688   8
X00.689   9
X00.690 10
X00.691 11
X00.692 12
X00.693 13
X00.694 14
X00.695 15
X00.696 16
X00.698 18

X = 2: Stainless Steel


X = 4: Titanium

Implants are available non-sterile or sterile packed.


Add suffix “S” to article number to order sterile product.

22    DePuy Synthes  LCP Compact Hand  Surgical Technique


Instruments Compact Hand 1.0/1.3

312.133 Double Drill Guide 1.3/1.0

316.385 Drill Bit B 0.8 mm, length 40/16 mm,


2-flute, for Mini Quick Coupling

513.005 Drill Bit B 1.0 mm, length 46/34 mm,


2-flute, for Mini Quick Coupling

513.035 Drill Bit B 1.3 mm, length 46/34 mm,


2-flute, for Mini Quick Coupling

310.882 Countersink 1.3 to 2.0, length 50 mm,


for Mini Quick Coupling

314.413 Screwdriver Shaft 1.3, cruciform,


with Holding Sleeve, length 62 mm,
for Mini Quick Coupling

314.481 Screwdriver Shaft 1.0, cruciform,


with Holding Sleeve, length 62 mm,
with Mini Quick Coupling

LCP Compact Hand  Surgical Technique  DePuy Synthes    22


Implants LCP Compact Hand 1.5

Plates

LCP Plate 1.5, straight


Art. No. Holes Length (mm)
0X.114.002   4 23
0X.114.004   6 36

Locking Plate 1.5, straight


Art. No. Holes Length (mm)
0X.114.003   6 29
0X.114.005 12 59

Locking T-Plate 1.5


Art. No. Holes Length (mm)
0X.114.006 Shaft 8, head 3 44.5
0X.114.007 Shaft 8, head 4 44.5

Locking Strut Plate 1.5


Art. No. Holes Length (mm)
0X.114.010  Right   8 23
0X.114.009  Left   8 23

Locking Y-Plate 1.5


Art. No. Holes Length (mm)
0X.114.013 Shaft 8, head 3 47

Locking Condylar Plate 1.5


Art. No. Holes Length (mm)
0X.114.014 Shaft 6, head 2 36

X = 2: Stainless Steel


X = 4: Titanium

Implants are available non-sterile or sterile packed.


Add suffix “S” to article number to order sterile product.

22    DePuy Synthes  LCP Compact Hand  Surgical Technique


Plates, with Preassembled Short Drill Sleeves

LCP Plate 1.5, straight


Art. No. Holes Length (mm)
0X.114.502S   4 23
0X.114.504S   6 36

Locking Plate 1.5, straight


Art. No. Holes Length (mm)
0X.114.503S   6 29
0X.114.505S 12 59

Locking T-Plate 1.5


Art. No. Holes Length (mm)
0X.114.506S Shaft 8, head 3 44.5
0X.114.507S Shaft 8, head 4 44.5

Locking Strut Plate 1.5


Art. No. Holes Length (mm)
0X.114.510S  Right   8 23
0X.114.509S  Left   8 23

Locking Y-Plate 1.5


Art. No. Holes Length (mm)
0X.114.513S Shaft 8, head 3 47

Locking Condylar Plate 1.5


Art. No. Holes Length (mm)
0X.114.514S Shaft 6, head 2 36

Note: The plates with preassembled short drill sleeves are


available sterile only.

X = 2: Stainless Steel


X = 4: Titanium

LCP Compact Hand  Surgical Technique  DePuy Synthes    22


Implants LCP Compact Hand 1.5

Screws

Cortex Screw Stardrive, B 1.5 mm,


self-tapping
Art. No. Length (mm)
0X.214.106   6
0X.214.107   7
0X.214.108   8
0X.214.109   9
0X.214.110 10
0X.214.111 11
0X.214.112 12
0X.214.113 13
0X.214.114 14
0X.214.115 15
0X.214.116 16
0X.214.118 18
0X.214.120 20
0X.214.122 22
0X.214.124 24

X = 2: Stainless Steel


X = 4: Titanium

Implants are available non-sterile or sterile packed.


Add suffix “S” to article number to order sterile product.

33    DePuy Synthes  LCP Compact Hand  Surgical Technique


Locking Screw Stardrive, B 1.5 mm,
self-tapping
Art. No. Length (mm)
0X.214.006   6
0X.214.007   7
0X.214.008   8
0X.214.009   9
0X.214.010 10
0X.214.011 11
0X.214.012 12
0X.214.013 13
0X.214.014 14
0X.214.015 15
0X.214.016 16
0X.214.018 18
0X.214.020 20
0X.214.022 22
0X.214.024 24

X = 2: Stainless Steel


X = 4: Titanium

Implants are available non-sterile or sterile packed.


Add suffix “S” to article number to order sterile product.

LCP Compact Hand  Surgical Technique  DePuy Synthes    33


Instruments LCP Compact Hand 1.5

03.114.001 LCP Drill Sleeve 1.5, for Drill Bit B 1.1 mm

03.114.003 Bending Pin for LCP Plates 1.5, with thread

03.114.007 Drill Bit B 1.1 mm, lengths 75/61 mm,


2-flute, for Mini Quick Coupling

310.141 Drill Bit B 1.5 mm, length 65/50 mm,


2-flute, for Quick Coupling, Stainless Steel

310.971 Countersink for Screws B 1.3 and 1.5 mm,


Stainless Steel

312.140 Double Drill Guide 1.5/1.1, for


No. 311.150

03.114.009 Screwdriver Shaft Stardrive 1.5,


with Holding Sleeve length 66 mm,
for Mini Quick Coupling

Drill Bits are available non-sterile or sterile packed.


Add suffix “S” to article number to order sterile product.

33    DePuy Synthes  LCP Compact Hand  Surgical Technique


Implants LCP Compact Hand 2.0

Plates

LCP Plate 2.0, straight


Art. No. Holes Length (mm)
X47.344   4 31
X47.345   5 38
X47.346   6 45
X47.347   7 52
X47.348   8 59
X47.360 10 73

Adaption Plate 2.0


Art. No. Holes Length (mm)
X43.091 12 71

LCP Condylar Plate 2.0


Art. No. Holes Length (mm)
X47.349 Shaft 7, head 2 54

LCP Rotation Correction Plate 2.0


Art. No. Holes Length (mm)
X47.354 Shaft 4, head 2 34
X47.355 Shaft 5, head 2 40

LCP T-Plate 2.0


Art. No. Holes Length (mm)
X47.615 Shaft 7, head 3 52.5

X = 2: Stainless Steel


X = 4: Titanium

Implants are available non-sterile or sterile packed.


Add suffix “S” to article number to order sterile product.

LCP Compact Hand  Surgical Technique  DePuy Synthes    33


Implants LCP Compact Hand 2.0

LCP Adaption Plate


Art No. Holes Length (mm)
X47.362 12 81.26

LCP T-Adaption Plate 2.0


Art No. Shaft holes Head holes
X47.351   7   2

LCP Y-Adaption Plate 2.0


Art No. Shaft holes Head holes
X47.350   7   3

Mini Condylar Plate 2.0


Art No.
X43.610 pin left
X43.620 pin left

Mini H-Plate
Art No. Holes
X43.480   4

X = 2: Stainless Steel


X = 4: Titanium

Implants are available non-sterile or sterile packed.


Add suffix “S” to article number to order sterile product.

33    DePuy Synthes  LCP Compact Hand  Surgical Technique


Screws

Cortex Screw Stardrive, B 2.0 mm,


self-tapping
Art. No. Length (mm)
X01.356   6
X01.357   7
X01.358   8
X01.359   9
X01.360 10
X01.361 11
X01.362 12
X01.363 13
X01.364 14
X01.366 16
X01.368 18
X01.370 20
X01.372 22
X01.374 24
X01.375 26
X01.376 28
X01.377 30
X01.378 32
X01.379 34
X01.381 38

X = 2: Stainless Steel


X = 4: Titanium

Implants are available non-sterile or sterile packed.


Add suffix “S” to article number to order sterile product.

LCP Compact Hand  Surgical Technique  DePuy Synthes    33


Implants LCP Compact Hand 2.0

Locking Screw Stardrive, B 2.0 mm,


self-tapping
Art. No. Length (mm)
X01.876   6
X01.877   7
X01.878   8
X01.879   9
X01.880 10
X01.881 11
X01.882 12
X01.883 13
X01.884 14
X01.886 16
X01.888 18
X01.890 20
X01.892 22
X01.894 24
X01.896 26
X01.898 28
X01.900 30

X = 2: Stainless Steel


X = 4: Titanium

Implants are available non-sterile or sterile packed.


Add suffix “S” to article number to order sterile product.

33    DePuy Synthes  LCP Compact Hand  Surgical Technique


Instruments LCP Compact Hand 2.0

329.921 Bending Pin for LCP Plates 2.0, with thread

323.034 LCP Drill Sleeve 2.0, with Scale,


for Drill Bits B 1.5 mm with marking

323.200 Universal Drill Guide 2.0

310.507 Drill Bit B 1.5 mm, with marking,


length 96/82 mm, 2-flute,
for Mini Quick Coupling

513.140 Drill Bit B 2.0 mm, length 67/55 mm,


2-flute, for Mini Quick Coupling

310.880 Countersink 1.5 to 2.4, length 52 mm,


for Mini Quick Coupling

314.676 Screwdriver Shaft Stardrive 2.0,


with Holding Sleeve, length 66 mm,
for Mini Quick Coupling

LCP Compact Hand  Surgical Technique  DePuy Synthes    33


Implants LCP Compact Hand 2.4

Plates

LCP Plate 2.4, straight


Art. No. Holes Length (mm)
X47.364 10 83.6
X47.375   5 43.6
X47.377   7 59.6
X49.677   8 68
X49.674   4 36
X49.676   6 52

Adaption Plate 2.4


Art. No. Holes Length (mm)
X49.912 12 72

LCP T-Plate 2.4


Art. No. Holes Length (mm)
X49.615 Shaft 7, head 3 57

LCP Condylar Plate 2.4


Art. No. Holes Length (mm)
X49.679 Shaft 7, head 2 59

LCP Adaption Plate 2.4


Art. No. Holes Length (mm)
X47.366 12 87.9

LCP T-Adaption Plate 2.4


Art. No. Shaft holes Head holes
X49.670   7   2

LCP Y-Adaption Plate 2.4


Art. No. Shaft holes Head holes
X49.669   7   3

X = 2: Stainless Steel


X = 4: Titanium

Implants are available non-sterile or sterile packed.


Add suffix “S” to article number to order sterile product.

33    DePuy Synthes  LCP Compact Hand  Surgical Technique


Screws

Cortex Screw Stardrive, B 2.4 mm,


self-tapping
Art. No. Length (mm)
X01.756   6
X01.757   7
X01.758   8
X01.759   9
X01.760 10
X01.761 11
X01.762 12
X01.763 13
X01.764 14
X01.766 16
X01.768 18
X01.770 20
X01.772 22
X01.774 24
X01.776 26
X01.778 28
X01.780 30
X01.782 32
X01.784 34
X01.786 36
X01.788 38
X01.790 40

X = 2: Stainless Steel


X = 4: Titanium

Implants are available non-sterile or sterile packed.


Add suffix “S” to article number to order sterile product.

LCP Compact Hand  Surgical Technique  DePuy Synthes    33


Implants LCP Compact Hand 2.4

Locking Screw Stardrive, B 2.4 mm,


self-tapping
Art. No. Length (mm)
X12.806   6
X12.807   7
X12.808   8
X12.809   9
X12.810 10
X12.811 11
X12.812 12
X12.813 13
X12.814 14
X12.816 16
X12.818 18
X12.820 20
X12.822 22
X12.824 24
X12.826 26
X12.828 28
X12.830 30

Optional

Buttress Pin (head LCP 2.4) B 1.8 mm


Art No. Length (mm)
X00.190 12
X00.191 14
X00.192 16
X00.193 18

X = 2: Stainless Steel


X = 4: Titanium

Implants are available non-sterile or sterile packed.


Add suffix “S” to article number to order sterile product.

44    DePuy Synthes  LCP Compact Hand  Surgical Technique


Instruments LCP Compact Hand 2.4

329.922 Bending Pin for LCP Plates 2.4 and 2.7,


with thread

323.029 LCP Drill Sleeve 2.4, with Scale up to


30 mm, for Drill Bits B 1.8 mm

323.202 Universal Drill Guide 2.4

310.508 Drill Bit B 1.8 mm, with marking,


length 96/82 mm, 2-flute, for Mini Quick
Coupling

317.871 Drill Bit B 2.4 mm, length 80/66 mm,


2-flute, for Mini Quick Coupling

314.467 Screwdriver Shaft, Stardrive, T8,


self-holding

310.880 Countersink 1.5 to 2.4, length 52 mm,


for Mini Quick Coupling

LCP Compact Hand  Surgical Technique  DePuy Synthes    44


Sets

182.480 Compact Hand 1.0/1.3


(Titanium Implants)

182.483 Compact Hand 1.0/1.3


(Stainless Steel Implants)

01.114.104 LCP Compact Hand 1.5


(Titanium Implants)

01.114.102 LCP Compact Hand 1.5


(Stainless Steel Implants)

01.110.014 LCP Compact Hand 2.0


(Titanium Implants)

01.110.012 LCP Compact Hand 2.0


(Stainless Steel Implants)

01.110.024 LCP Compact Hand 2.4


(Titanium Implants)

01.110.022 LCP Compact Hand 2.4


(Stainless Steel Implants)

01.110.003 Standard Instrument Set for


LCP Compact Hand

182.475 Aditional Instrument Set for


Compact Hand

44    DePuy Synthes  LCP Compact Hand  Surgical Technique


Standard Instrument Set for
LCP Compact Hand

311.012 Handle, medium,


with Mini Quick Coupling

311.430 Handle with Quick Coupling,


length 110 mm

319.003 Depth Gauge for Screws B 1.3 to 1.5 mm,


measuring range up to 24 mm

319.005 Depth Gauge for Screws B 2.0 and


2.4 mm, measuring range up to 40 mm

347.901 Pliers, flat-nosed, pointed,


for Plates 1.0 to 2.4

347.985 Holding Forceps for Cortex Screws


B 1.0 to 2.4 mm

391.962 Bending/Cutting Pliers

LCP Compact Hand Surgical Technique DePuy Synthes 43


Standard Instrument Set for
LCP Compact Hand

Optional

03.100.000 Handle, lockable,


with Mini Quick Coupling

311.011 Handle, small, with Mini Quick Coupling

311.013 Handle, large, with Mini Quick Coupling

347.981 Holding Forceps for Plates 1.0 to 2.4

511.776 Torque Limiter, 0.8 Nm,


with AO/ASIF Quick Coupling

511.777 Torque Limiter, 0.4 Nm,


with AO/ASIF Quick Coupling

44    DePuy Synthes  LCP Compact Hand  Surgical Technique


MRI Information

Torque, Displacement and Image Artifacts according


to ASTM F 2213-06, ASTM F 2052-06e1 and
ASTM F 2119-07
Non-clinical testing of worst case scenario in a 3 T MRI
system did not reveal any relevant torque or displacement
of the construct for an experimentally measured local spatial
gradient of the magnetic field of 3.69 T/m. The largest image
­artifact extended approximately 169 mm from the construct
when scanned using the Gradient Echo (GE). Testing was
­conducted on a 3 T MRI system.

Radio-Frequency-(RF-)induced heating according to


ASTM F 2182-11a
Non-clinical electromagnetic and thermal testing of worst
case scenario lead to peak temperature rise of 9.5 °C
with an average temperature rise of 6.6 °C (1.5 T) and a
peak temperature rise of 5.9 °C (3 T) under MRI Conditions
­using RF Coils (whole body averaged specific absorption rate
[SAR] of 2 W/kg for 6 minutes [1.5 T] and for 15 minutes
[3 T]).

Precautions: The above mentioned test relies on non-clini­-


cal testing. The actual temperature rise in the patient will
­depend on a variety of factors beyond the SAR and time of
RF application. Thus, it is recommended to pay particular
­attention to the following points:
–– It is recommended to thoroughly monitor patients under­
going MR scanning for perceived temperature and/or pain
sensations.
–– Patients with impaired thermoregulation or temperature
­sensation should be excluded from MR scanning proce­-
dures.
–– Generally, it is recommended to use a MR system with
low field strength in the presence of conductive implants.
The employed specific absorption rate (SAR) should be
reduced as far as possible.
–– Using the ventilation system may further contribute to
reduce temperature increase in the body.

LCP Compact Hand  Surgical Technique  DePuy Synthes    44


© DePuy Synthes Trauma, a division of Synthes GmbH. 2016.  All rights reserved.  036.000.028 DSEM/TRM/0815/0478(1) 05/16

Synthes GmbH
Eimattstrasse 3 Not all products are currently available in all markets.
4436 Oberdorf
Switzerland This publication is not intended for distribution in the USA.
Tel: +41 61 965 61 11
Fax: +41 61 965 66 00 All surgical techniques are available as PDF files at
www.depuysynthes.com www.depuysynthes.com/ifu 0123

You might also like