Professional Documents
Culture Documents
ADDENDUM
Abstract
This addendum to the code of practice for the determination of absorbed dose
for x-rays below 300 kV has recently been approved by the IPEM and introduces
three main changes: (i) Due to a lack of available data the original code
recommended a value of unity for kch in the very-low-energy range (0.035–
1.0 mm Al HVL). A single table of kch values, ranging from 1.01 to 1.07,
applicable to both designated chamber types is now presented. (ii) For medium-
energy x-rays (0.5–4 mm Cu HVL) methods are given to determine the absorbed
dose to water either at 2 cm depth or at the surface of a phantom depending
on clinical needs. Determination of the dose at the phantom surface is derived
from an in-air measurement and by extending the low-energy range up to 4 mm
Cu HVL. Relevant backscatter factors and ratios of mass energy absorption
coefficients are given in the addendum. (iii) Relative dosimetry: although
not normally forming part of a dosimetry code of practice a brief review of the
current literature on this topic has been added as an appendix. This encompasses
advice on techniques for measuring depth doses, applicator factors for small
field sizes, dose fall off with increasing SSD and choice of appropriate phantom
materials and ionization chambers.
Introduction
Three factors have made it necessary to publish an addendum to the kilovoltage x-ray code
(IPEMB 1996). First, the code did not include values of kch for very-low-energy x-rays. These
have now been determined. Second, it has become apparent that for medium-energy x-rays
in-air is preferable to in-phantom calibration in many circumstances. This is now given as an
alternative option and the relevant data are included. Third, several requests had been received
for advice on techniques for measuring depth doses, applicator factors for small field sizes
and dose fall off with increasing SSD. This would not normally form part of a dosimetry code
of practice. A brief review of the current literature on this topic has therefore been added as
an appendix.
1.1. Background
The IPEMB kV code of practice (1996) recommends that the absorbed dose to water at
the surface1 of a full-scatter water-equivalent phantom (Dw,z=0 ) can be determined from the
reading of a thin-window parallel-plate ionization chamber with its window placed flush with
the surface of the phantom using the following formula:
!" # $
µ̄en
Dw,z=0 = MNK kch (1)
ρ w/air
z=0,φ
where M is the instrument reading corrected to the same ambient conditions as the calibration
factor. NK is the in-air chamber calibration factor which converts the reading to air kerma
under scatter-free conditions at the reference point of the chamber with the chamber assembly
replaced by air. It varies with the radiation quality of the incident beam in air and is valid for
the calibration field size, which is 5.3 cm diameter if the calibration factor is traceable to the
National Physical Laboratory. For a thin-window parallel-plate chamber the reference point
is the centre of the window. [(µ̄en /ρ)w/air ]z=0,φ is the ratio of the mass energy absorption
coefficients of water to air averaged over the photon spectrum at the surface of the water
phantom (z = 0) for field diameter φ. kch is a factor that accounts for the change in response
of the ionization chamber between the calibration in-air and the measurement at the surface of
a full-scatter water phantom. In this addendum, it will be called the chamber correction factor.
The product NK kch corrects the chamber reading to air kerma at the reference point of the
chamber with the chamber assembly replaced by water.
established.
Addendum to the IPEMB code of practice for the determination of absorbed dose for kV x-rays 2741
Table 1. The chamber correction factor (kch) for PTW type M23344 and M23342 chambers
provided with NK in-air calibration factor, at the surface of a full-scatter water phantom irradiated
with an x-ray beam of diameter 3–15 cm. The data are taken from Dierker (1981), Greener (2001),
Ipe (2001) and Perrin et al (2001). The estimated uncertainties are ±2% (SD).
When using equation (1) it should be noted that the value of NK depends not only on the
x-ray beam quality but also on the field size which was used for the calibration of the chamber
in air. This is because of ‘stem scatter’, i.e. radiation scattered into the collecting volume from
the body of the chamber. Values of NK will decrease with increasing in-air calibration field
size to a minimum value when the body of the chamber is fully irradiated.
It follows therefore that the values of kch will also depend on the field size used for the
calibration of the chamber and will increase with increasing in-air calibration field size. This
is important because different primary standard calibration laboratories may employ different
field sizes when calibrating very-low-energy ionization chambers in air. The values given in
table 1 apply to a calibration field size of 5.3 cm diameter, as used at the National Physical
Laboratory.
The value of kch was measured for a limited range of field sizes and source to chamber
distances but it is expected that the values quoted in table 1 are valid for conditions used in
normal clinical practice. However, it is expected that the value of kch will be dependent on
the phantom material. It is therefore recommended that only water, mix-D (Jones et al 1949)
and WT1 (Constantinou et al 1982) are used as phantom materials as these were used by the
above authors.
changed from (8.5 − 3) = 5.5 mg cm−2 to 0; this will affect depth dose measurements, if these
have been made.
Table 2. Ratio of the mean mass energy absorption coefficient, water to air, for the primary
spectrum (free-in-air) for medium-energy x-rays. Uncertainty ±1.5% (1 SD). Reproduced from
AAPM (2001) by kind permission.
Table 3. Backscatter factors Bw for open applicators as a function of HVL, field diameter and
SSD. Statistical uncertainty ±0.5% (1 SD). Reproduced from Grosswendt (1990).
in-air calibration with open applicators over the medium-energy range. For convenience these
are reproduced in table 3.
It should be noted that the values given for 4.93 mm Cu HVL are for a mono-energetic
beam of 200 keV and are included only to enable interpolation between 3.358 and 4.0 mm Cu.
Also the values given in the table are raw (unsmoothed) data and contain statistical fluctuations
of about ±0.5% (1 SD).
There is thus a combined uncertainty of about ±1% (1 SD). For this reason, Bw may be
interpolated linearly between field sizes and between HVLs. The uncertainties in the data do
not justify greater precision.
A review of the measured data by Smith (1996) has demonstrated that the backscatter
factors for closed PMMA applicators are larger than those for open ended ones. The increase
is about 1% of the scatter component per millimetre thickness of the PMMA end plate. A
typical end plate is about 3.2 mm thick. It is therefore recommended that for closed ended
PMMA applicators the backscatter factor should be calculated using
Bw,closed = 1 + ((Bw,open − 1) × 1.032) (2)
where Bw,closed is the backscatter factor for the closed applicator and Bw,open is the backscatter
factor for an open-ended applicator taken from the data in table 3.
The uncertainties in the data do not justify using different values for different thicknesses
of PMMA end plate within the range normally encountered.
Several requests have been received for advice on techniques for measuring depth doses,
applicator factors for small field sizes and dose fall off with increasing SSD on kilovoltage
x-ray machines. This probably reflects the fact that although such techniques have been in
use for many years they are much less frequently used, and therefore much less well known,
than those for megavoltage machines. Relative dosimetry has not previously formed part of an
IPEM dosimetry code of practice. However, the working party decided to add a brief review
of the current literature on this topic as an appendix.
The review which follows represents only a summary of the state of the art at the time of
writing. The reader should refer to the original articles and to subsequent publications.
The phantom dimensions should each be sufficiently large that any further increase would
cause no significant change in the instrument reading.
between 300 and 50 kVp. For a 2 mm thick sleeve, the factors vary by 1.9, 3.0 and 3.8%.
There is also a variation with depth. At 100 kVp for a 3 mm thick sleeve the values of these
factors vary by 0.3, 0.5 and 0.7%, respectively, between depths of 0 and 7 cm. No correction
was found to be required at 300 kVp for PMMA, nylon or polystyrene sleeves up to 3 mm
thick at a depth of 2 cm in water.
Latex sleeves often contain a fine powder such as magnesium silicate (talc). Early reports
(e.g. Law and Foster 1987, Hanson et al 1988) suggest that such a powder may contaminate
the chamber and affect its sensitivity and energy response.
It is recommended that waterproof sleeves should be made from PMMA and should not
be more than 1.5 mm thick. Latex sleeves should be avoided.
It is recommended that measurements with small applicators should only be carried out by
centres with sufficient expertise and equipment to confirm the validity of their technique and
the values of any correction factors required. This applies in all cases where measurements
with a conventional ionization chamber orientated perpendicular to the beam axis are not
possible. Even lithium borate TLD and Gafchromic film may require at least dose-linearity
corrections. Where sufficient expertise and equipment are not available the centre concerned
should:
(a) Prohibit the clinical use of smaller applicators and replace them by lead cut-outs used
with larger applicators.
(b) Determine the backscatter factors and depth doses for these cut-outs from standard tables.
References
AAPM 2001 AAPM protocol for 40–300 kV x-ray beam dosimetry in radiotherapy and radiobiology Med. Phys.
28 868–93
Allen Li X, Ma C M and Salhani D 1997 Measurements of percentage depth dose and lateral beam profile for
kilovoltage x-ray beams Phys. Med. Biol. 42 2561–8
Aukett R J, Harrison R M and Rosser K E 1999 The characteristics of ionisation chambers for measurements with
x-rays in the kilovoltage range American Association of Physicists in Medicine Symposium Proceedings No 11:
Kilovoltage X-ray Beam Dosimetry for Radiotherapy and Radiobiology ed C M Ma and J P Seuntjens (Madison:
Medical Physics Publishing) pp 179–94
BIR 1972 Central axis depth dose data for use in radiotherapy British Journal of Radiology Supplement 11 (London:
British Institute of Radiology)
Constantinou C, Attix F H and Paliwal B R 1982 A solid water phantom for radiotherapy x-ray and γ -ray beam
calibration Med. Phys. 9 436–41
Dierker J 1981 Calibration correction with soft x-ray measurements using a phantom Medizinische Physik ed E Bunde
pp 111–6/B1
Greener A G 2001 private communication Medical Physics Department, Guy’s and St Thomas’ NHS Trust, Lambeth
Palace Road, London SE1 7EH, UK
Grosswendt B 1990 Dependence of the photon backscatter factor for water on source-to-phantom distance and
irradiation field size Phys. Med. Biol. 35 1233–45
Hanson W F, Arnold D J, Shalek R J and Humphries L J 1988 Contamination of ionization chambers by talcum
powder Med. Phys. 15 776–7
Harrison R M 1996 X-rays: Half Value Layers 0.01–8.0 mm Al British Journal of Radiology Supplement 25: Central
Axis Depth Dose Data for Use in Radiotherapy (London: British Institute of Radiology) pp 1–23
Harrison R M 1997 Low energy x-ray depth dose data for use in radiotherapy—comments on the review of BJR
Supplement 17 Brit. J. Radiol. 70 946–9
ICRU 1973 Measurement of absorbed dose in a phantom irradiated by a single beam of x or gamma rays ICRU
Report 23 (Bethesda, MD: ICRU)
Ipe N E, Rosser K E, Moretti C J, Manning J W and Palmer M J 2001 Air kerma calibration factors and chamber
correction values for PTW soft x-ray, NACP and Roos ionisation chambers at very low x-ray energies Phys.
Med. Biol. 46 2107–17
IPEMB 1996 The IPEMB code of practice for the determination of absorbed dose for x-rays below 300 kV generating
potential (0.035 mm Al–4 mm Cu HVL; 10–300 kV generating potential) Phys. Med. Biol. 41 2605–25
Jones D E A and Raine H C 1949 Letter to the editor Brit. J. Radiol. 22 549–50
Klevenhagen S C 1989 Experimentally determined back scatter factors for x-rays generated at voltages between 16
and 140 kV Phys. Med. Biol. 34 1871–82
Knight R T 1992 Back scatter factors for low and medium energy x-rays calculated by the Monte Carlo method
Internal Report ICR-PHYS-1/93 (Sutton, Surrey, UK: Royal Marsden NHS Trust)
Knight R T 1996 Absorbed dose conversion factors for therapeutic kilovoltage and megavoltage x-ray beams calculated
by the Monte Carlo method Internal Report ICR-PHYS-1/96 (Sutton, Surrey, UK: Royal Marsden NHS Trust)
Law J and Foster C J 1987 Calibration of radiotherapy dosemeters against secondary standard dosemeters: an
anomalous result Phys. Med. Biol. 32 1039–43
Ma C M, Li X A and Seuntjens J 1999 Study of dosimetry consistency for kilovoltage x-ray beams American
Association of Physicists in Medicine Symposium Proceedings No 11: Kilovoltage X-ray Beam Dosimetry for
Radiotherapy and Radiobiology ed C M Ma and J P Seuntjens (Madison: Medical Physics Publishing) pp 69–88
2748 IPEM Working Party
Ma C M and Seuntjens J P 1997 Correction factors for water-proofing sleeves in kilovoltage x-ray beams Med. Phys.
24 1507–13
Nisbet A, Aukett R J, Davison A, Glendinning A G, Thwaites D I and Bonnett D E 1999 An experimental comparison of
kilovoltage x-ray dosimetry protocols American Association of Physicists in Medicine Symposium Proceedings
No 11: Kilovoltage X-ray Beam Dosimetry for Radiotherapy and Radiobiology ed C M Ma and J P Seuntjens
(Madison: Medical Physics Publishing) pp 43–54
Perrin B A, Whitehurst P, Cooper P and Hounsell A R 2001 The measurement of kch factors for application with the
IPEMB very low energy dosimetry protocol Phys. Med. Biol. 46 1985–95
Rosser K E 1996 Measurement of absorbed dose to water using medium energy x-rays NPL Report CIRA(EXT)006
(Teddington: National Physical Laboratory)
Rosser K E 1998 Investigation of the chamber correction factor (kch) for the UK secondary standard ionisation
chamber (NE2561/NE2611) using medium-energy x-rays Phys Med. Biol. 43 3195–206
Rosser K E, Aukett R J, Greener A G, Harrison R M and Nahum A E 2003 United Kingdom code of practice for
kilovoltage x ray dosimetry IAEA Symposium IAEA-CN-96–22 (Vienna: International Atomic Energy Agency)
Seuntjens J, Aalbers A H L, Grimbergen T W M, Mijnheer B J, Thierens H, Van Dam J, Wittkamper F W, Zoetelief J,
Piessens M and Piret P 1999 Suitability of diamond detectors to measure central axis depth kerma curves for
low and medium energy x-rays American Association of Physicists in Medicine Symposium Proceedings No 11:
Kilovoltage X-ray Beam Dosimetry for Radiotherapy and Radiobiology ed C M Ma and J P Seuntjens (Madison,
WI: Medical Physics Publishing) pp 227–38
Smith C W 1996 Orthovoltage x-ray beams (0.5 mm–4.0 mm Cu HVL) British Journal of Radiology Supplement 25:
Central Axis Depth Dose Data for Use in Radiotherapy (London: British Institute of Radiology) pp 24–38
Wilks R J and Soni K K 2000 An optical method of estimating the relative output of a superficial unit for small
applicators Phys. Med. Biol. 45 N71–6