Professional Documents
Culture Documents
ANATOM 64 Clarity
X-ray Computed Tomography System
Add: Block B, LingYa Industrial Zone, Tangtou No.1 Road, Shiyan Sub-district, Bao'an District,
Shenzhen, 518108, P.R.China
Fax: 86-755-26695307
The copyright to this manual is owned by Shenzhen Anke High-tech Co.,Ltd. Any
organization or individual isn’t allowed to abstract, duplicate or translate it in any manner
without the prior written permission of the copyright owner.
All right reserverd.
Together with other Anke trademarks are all the trademarks of Shenzhen Anke
High-tech Co., Ltd. (hereinafter referred to as Anke).
Caution
The purchased products, services or features, etc shall be bound by the commercial
contract and provisions of Anke. All or partial products, services or features described in
this document may be outside of the scope of your purchase or use, and Anke does not
make any express or implied statement or warranty on the contents of this
documentunless otherwise stipulated.
The contents of this document might be updated from time to time due to the product
upgrade or other reasons. Unless otherwise stipulated, this document shall only serve as a
guide, and any statement, information and suggestions in this document will not
constitute any express or implied warranty.
This product complies with the design requirements on environmental protection and
personal safety, and the storage, use and disposal of the product shall be conducted
strictly in accordance with the product manual, relevant agreements and the laws and
regulations of relevant countries.
Document Information
Document No:82-S01809
Version:R1.3
Revision History
Rev. Date Details of Changes
Update 3D Module, system basic information and user layout buttons in Film
R1.2 2019-05-06
module and so on.
R1.3 2019-08-23 View and Patient module add seveal clinical needs.
Feedback
In order to improve the quality of the user document of Anke and provide better services to you, your
suggestions and opinions are expected, please fax them to: 86-755-26695307, or mail them to: Block
B, LingYa Industrial Zone, Tangtou No.1 Road, Shiyan Sub-district, Bao'an District, Shenzhen, 518108,
E-mail: service@anke.com.
Document name ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Product version Document version
Date of installation of the equipment in your organization
Please fill your information below, so that we can contact you in time.
Organization
Name
name
Post code Add
Tel E-mail
Excellent Good Average Poor Terrible
Generally satisfactory
Work guidance
Easy reference
Your comments
Correct contents
on the document
Complete contents
Reasonable structure
Graphical presentation
Easy to understand
Detailed description
Structure of contents
Detailed contents
Your comments Content depth
on improvement Concise expression
of the document Add figures
Add examples
Add FAQ
Others
Other
suggestions on
Anke’s User
Manual
Table of Contents
Chapter 1 Preface.......................................................................................................................................... 1
1.1 About the Manual ................................................................................................................................ 1
1.2 Design Purpose .................................................................................................................................... 1
1.3 Contents ............................................................................................................................................... 2
1.4 Relevant Manuals ................................................................................................................................ 3
1.5 Conventions ......................................................................................................................................... 3
1.6 Compatibility ....................................................................................................................................... 4
1.7 Regulatory Compliance ....................................................................................................................... 5
1.8 Training................................................................................................................................................ 5
-i-
2.9.8 Stray Radiation Pattern.............................................................................................................29
2.10 Electrical Output on the Aspects of Loading Factor/Coefficient......................................................31
2.11 Deterministic Effects ........................................................................................................................32
2.12 Safety of Laser Positioning Lamp ....................................................................................................33
2.13 Biocompatibility ...............................................................................................................................33
2.14 Environmental Pollution...................................................................................................................34
2.15 Safety of Medical Electrical System ................................................................................................34
2.16 Planned Maintenance and Routine Inspection..................................................................................34
2.17 Cleaning, Disinfection ......................................................................................................................37
2.17.1 Cleaning .................................................................................................................................37
2.17.2 Disinfeciton ............................................................................................................................38
2.18 Symbols and Labels .........................................................................................................................39
2.19 Intended Use.....................................................................................................................................44
2.20 Contraindication ...............................................................................................................................44
2.21 Life Cycle .........................................................................................................................................44
2.22 Customer Service .............................................................................................................................45
2.23 Registration Information ..................................................................................................................45
-ii-
3.7 Daily Turn On and Turn Off .............................................................................................................. 66
-iii-
5.6.4 Stop a Scan ...............................................................................................................................95
5.6.5 Add a Scan................................................................................................................................95
5.6.6 Exit a Scan................................................................................................................................96
5.7 Image Setting......................................................................................................................................97
5.7.1 Measure Images........................................................................................................................97
5.7.2 Adjust Images ...........................................................................................................................97
5.8 Images Transmission ..........................................................................................................................98
-iv-
6.7.5 Reset Images .......................................................................................................................... 118
6.7.6 Display/Hide Texts ................................................................................................................ 119
6.7.7 Grid Coordinate Display ........................................................................................................ 119
6.8 Comparison Unit.............................................................................................................................. 119
6.9 Export images .................................................................................................................................. 121
6.10 Images Transmission ..................................................................................................................... 122
-v-
8.3 Create and save printer images .........................................................................................................154
8.4 Select and Arrange Images ...............................................................................................................155
8.5 Image Layout Unit ...........................................................................................................................155
8.6 Image Tools ......................................................................................................................................158
8.7 Image Adjustment ............................................................................................................................159
8.7.1 Display Gray Bar ....................................................................................................................160
8.7.2 Customize Rotation ................................................................................................................161
8.7.3 DisplayTopo Lines and Topo Image .......................................................................................162
8.8 Functions of Right-click Menu.........................................................................................................162
8.8.1 Copy and Paste .......................................................................................................................163
8.8.2 Annotation Font Size ..............................................................................................................163
8.9 Image Print Unit ...............................................................................................................................164
-vi-
10.4 Template Management Unit .......................................................................................................... 184
-vii-
12.3.4 Delete a Scan Protocol .........................................................................................................220
12.4 Modify Scan Items .........................................................................................................................221
12.4.1 Create a Scan Item................................................................................................................221
12.4.2 Copy a Scan Item .................................................................................................................222
12.4.3 Rename a Scan Item .............................................................................................................222
12.4.4 Delete a Scan Item................................................................................................................223
12.5 Add/delete a Recon Item ................................................................................................................223
12.6 Modify Scan Paremeters ................................................................................................................224
12.7 Modify Reconstruction Parameters ................................................................................................226
-viii-
Chapter 1 Preface
Abstract
This chapter mainly gives an introduction to the purpose and format of this Manual. It
provides a brief introduction to the contents of this Manual, including the purpose,
skills necessary for readers and the graphical conventions on all the visual symbols in
the manual.
This Manual shall be always put together with the equipment. It's very important to
regularly go over the process and safety precautions described in the Manual. It's
required to read and understand the contents described in this Manual before using this
product.
The theory of CT is that X-ray attenuates to different extents while various tissues in
the human body are radiated by X-ray. Just by using such feature, CT can indicate
different tissues and their spatial locations in the selected area. For instance, CT may
detect any tissue (e.g. tumor tissue) with abnormal density, and reconstruct its image
by a computer and display it on the monitor after proper processing. A well-trained
doctor can obtain useful diagnosis information from CT images.
1
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
This equipment may be applied to X-ray CT of the head, body, heart and vessels and
other parts of a patients of all ages.
The equipment must be operated strictly in accordance with relevant local laws and
regulations. Users and operators are not allowed to operate the equipment unless they
comply with relevant laws and regulations.
The manufacturer will not undertake or only undertake partial responsibilities if any
user/operator uses the equipment for any other purpose rather than those described by
the manufacturer, or improper use or operation results in any incompatibility,
equipment damage or personal injury.
1.3 Contents
This part describes the main contents of this book, introduces the key points of each
chapter, and guides users to use this book pertinently.
2
Chapter 1 Preface
Belows are relevant manuals matching the equipment, but not delivery attachment
documentation. If clients would like to know more, please contact Anke service.
1.5 Conventions
The following conventions and symbols are adopted in this book.
1. Convention on symbols
“< >” indicates key name, button name and information entered by the operator from
the terminal; “[ ]” indicates man-machine interface, menu bar, data sheet, field name,
etc, and multilevel menu is separated by "→".For example, multilevel menu [File →
Add→ Folder] indicates the item [Folder] under the submenu [Add] under [File].
Format Meaning
Indicate key names and button names. For example, <Enter>,
Characters in angle
<Tab>, <Backspace> and <a> indicate Enter, Tab, Backspace
brackets
and lowercase a respectively.
3
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Format Meaning
Indicates pressing several keys on the keyboard at the same time.
<Key 1+Key 2> For example, <Ctrl+Alt+B> indicates pressing “Ctrl”, “Alt” and
“B” simultaneously.
Indicates pressing the first key, release it, and press the second
<Key 1, Key 2> key. For example, <Ctrl, N> indicates press <Ctrl> first, release it,
and press <N>.
Format Meaning
Single click Rapidly press and release the left mouse button
Double click Rapidly press twice and release the left mouse button
Right click Rapidly press and release the right mouse button
Drag Press and hold the left mouse button, and move the mouse
4. Symbol
The following safety symbol indicating different connotations may appear in this
Manual:
1.6 Compatibility
Except for the compatible devices and components particularly designated by Anke,
the equipment cannot be used together with any other device or component.
The equipment shall not be modified and/or added with parts, unless otherwise
authorized by Anke. The modification and /or addition of parts shall comply with the
local laws and effective regulations. Modification and/or addition of parts by the user
may result in unexpected equipment damage or personal injury.
4
Chapter 1 Preface
If operated properly, the device also complies with international and national laws and
standards on Electromagnetic Compatibility (EMC). These laws and standards set forth
electromagnetic radiation dosage allowable for the equipment and the equipment's
capability in resisting the interference of electromagnetic radiation from the outside.
1.8 Training
This Manual will not give instructions on X-ray CT. You need to have sufficient
knowledge to be competent to perform various diagnostic Imaging processes during the
treatment. Such knowledge can be achieved through various education means,
including clinical work experiences, fundamental subjects of hospitals and some
radiological technology related courses in many universities.
Before operating the equipment, the operator has to receive sufficient training so as to
use the equipment safely and efficiently according to the instructions in the Operation
Manual.
The training contents may be varied in different countries. It's embodied in the
difference in the laws and regulations of different regions, and the operators must be
well trained on these aspects.
5
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Warning:
Do not operate the equipment before sufficient training. Otherwise, it may cause
serious quipment damage, persoanl injury or possible misdiagnose.
Please contact Anke if you expect more information about the training.
6
Chapter 2 CT Safety
Abstract
This chapter mainly presents information about CT safety precautions and process.
It's critical to read carefully and remember all the danger warnings and safety warnings
on the equipment.
Meanwhile, it's also critical to strictly follow the descriptions, all warnings and
cautions provided under “CT Safety” the Operation Manual, so as to ensure the safety
of both the patients and operators.
Caution:
Before examining the patients with this equipment, you're advised to read, understand
and master the contents related to emergency stop described in Chapter 1 "CT safety".
Warning:
Make sure the checking procedures described under the chapter "Daily Maintenance"
in the Maintenance Manual have been satisfactorily completed before any scan, and
preventive maintenance is the latest. If any fault is detected or doubted in the
equipment or system, do not use this system until it's recovered. Operating equipment
or system with defective components may cause unnecessary radiation or other injuries
to the operators or patients.
Before using this equipment, make sure you have carefully read and remembered all
the danger warnings, safety cautions and emergency braking procedures described in
7
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
this section. Failure to observe the safety requirements described in this Manual may
lead to serious equipment damage or personal injury.
Do not operate this equipment unless you have received sufficient and correct training.
Do not operate the equipment if you're not sure of the operating method.
If not properly trained, operating this equipment may lead to equipment damage or
severe personal injury, and it may also result in clinical misdiagnose.
Warning:
Do not try to move, modify, discard or block the safety components in the equipment,
because this may cause serious equipment damage or personal injury.
The guide for the operators and patients provided in this text aims at better image
quality. Failure of fulfillment of the operation procedures (also including Cautions,
Danger, Warnings, etc) described in this text may cause potential hazard to human
body, damage the equipment, and impact the image quality.
After proper installation of the system, addition or replacement with any software or
hardware not authorized by Anke may cause:
Equipment failure
Equipment damage
Anke won't be responsible for the deterioration of image quality, personal injury or
equipment damage caused by unauthorized software installations or hardware
replacements.
8
Chapter 2 CT Safety
Put all the covers back into place before using the equipment. The covers may
protect the users and patients from collision or electric shock by the moving
component, and may also protect the equipment.
Keep the equipment clean. Remove the liquid of the equipment and/or spilt
injection to prevent harm to the health and impact on internal parts. Clean the
equipment with warm soapy water or neutral disinfectant.
When any problem is detected in the equipment, please contact the after-sales
personnel of Anke. Do not handle it by yourself, because Anke will not
undertake any responsibility for any damage caused by human reason.
Do not run any software that not dedicated to this system on the computer.
It may hurt the patients if you operate this equipment with an implantable or
wearable medical device. Please contact with manufacturer to get more safety
information before using the equipment.
Warning:
Do not disassemble or remove the cover or cables of this equipment. As there are
high-voltage currents inside this equipment, disassembly of the cover or cable may lead
to fatal injury or serious personal injury.
To ensure safe and reliable equipment performance, the room or area for this
equipment must comply with the laws related to the electrical safety of equipment or
laws and regulations with legal force.
Only the after-service personnel of Anke are authorized to install, debug, repair or
modify the equipment. The users are not allowed to disassemble or install the
equipment without permission, and Anke won't undertake the responsibility for the
human-caused damage.
9
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Danger:
When the Gantry tilts or the patient table moves, pay close attention to the patient and
equipment to avoid impacting or squeezing the patient. In the event of accidental
movement or unstoppable movement, press the emergency stop button on the console
or Gantry.
Danger:
To avoid impacting or squeezing the four limbs of the patient, it's advised to separate
the feet and hands from the top/ board of the moving patient table and the interface
equipment, and separate from the base and flank board of the patient table. (Special
attention should be attached when placing any patient with big body size)
1. Before moving the patient table and Gantry, check whether there is any
obstacle around the equipment. When moving the patient table or Gantry,
always observe the movement process.
2. Help the patient get onto and off the patient table or table board in person.
3. When moving the patient in or out, restore the declination of the Gantry to
vertical direction 0°, lock the table board and adjust the patient table to an
appropriate height.
4. When tilting or moving the Gantry or table board, try to avoid contact between
the patient and the Gantry.
10
Chapter 2 CT Safety
6. Before use, check whether any component is damaged. If there is any, please
stop using it.
7. As the maximum weight the patient table can bear is 205kg, please do not place
any patient or object above such weight onto the patient table.
8. Patient table’s movement limit is no more than 10mm and gantry’s movement
angle limit is no more than 0.5º when normally stopped.
9. During operation, if the patient is stuck due to an error, press "Table Lock" of
the keyboard on the gantry, the patient table is unlocked, then push the table out
and set patient free.
10. Position the patients to prevent crash or squeezing due to movements of patient
table or gantry tilting.
Caution:
Ensure that the arms of the patient will not be impacted during scanning, and the
clothes or hair of the patient are not twisted in the internal parts of the device.
It’s extremely important to make sure that the infusion tube, the breathing tube,
the catheter and the ECG cable are not twisted in the gap between the patient table and
the side of gantry. These components can not withstand any tensile stress.
Make sure that the patient mattresses are not twisted in the patient table.
Danger:
If the weight of any patient is above 205kg, do not place the patient onto the
patienttable, because this may cause the patient table to topple and the patient to fall
off.
11
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Press and hold down the e-stop button, and meanwhile, the e-stop indicator is
illuminated, indicating it's currently at emergency stop state.
12
Chapter 2 CT Safety
Press the key and the indicator will be off, and then open the door and power
off the main switch.
Rotate the mergency shutdown key, and then open the door and power on the
main switch.
13
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Danger:
It's forbidden to use flammable or potentially explosive disinfectant spray, because the
generated steam may burn and cause personal injury and/or equipment damage.
It's required to observe and fulfill the regulations on use of such medical facilities. Fire
extinguishers for fire caused by both electricity and non-electricity must be equipped.
They should be labeled clearly and correctly, and be maintained in accordance with the
instructions of the manufacturer and placed at accessible locations.
All medical equipment operators should comprehensively understand and grasp the
rules for using fire extinguishers and other fire devices.
Danger:
In case of electric or chemical fire, special fire extinguishers bearing particular labels
should be used. Use of water or other liquid against electric fire may lead to fatal or
serious injury.
Warning:
If possible, try to switch the equipment off the power grid and other power supply
before fire fighting. this may reduce the danger of electric shock.
Warning:
As this equipment generates, uses and radiates RF energy, it may interfere with other
medical or non-medical devices and wireless communication. This equipment is fully
in compliance with EMC standard and capable to provide reasonable protection
againstsuch interference.
14
Chapter 2 CT Safety
If it's found that this equipment produces interference, the user should settle the
problem by one or more methods listed below:
Increase the distance between this equipment and the affected equipment.
Use a power supply different from that of the affected equipment to provide
power supply to this equipment.
The manufacturer will not undertake any responsibility for the interference caused by
using interconnecting cable not recommended or modifying or changing the equipment
without authorization. Unauthorized modification or change may deprive the user of
the rights to operate the equipment.
1. Control area
The console is located outside of the radiation control area. The patient can be
observed through the lead glass window, and the Gantry and patient table can be
operated by remote control.
Caution:
2. Radiation shielding
15
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
If the scan has to be released from inside the control area, or the accompanying
personnel have to stay within the control area, safety protection measures
(e.g.protective clothes, etc) must be adopted.
3. Exposure indicators
The exposure indicators are located above the Gantry display and the back surface
of the Gantry.
If the scan has been started, the exposure indicators are illuminated. If the
exposure indicators are not illuminated when the scan has been started, please stop
operation immediately and contact the after service personnel of Anke.
The following protective measures should be adopted to protect operators and patients.
In case of hour meter error, the backup hour meter will limit total scan time
within 110% of the pr-eset time. If the scan time exceeds this time length
pre-set in the backup hour meter, the x-ray source will be automatically
shutdown to avoid personal injury by over dose radiation.
2. Personnel
During scan, any person that intends to approach the patient should adopt the
following protective measures.
16
Chapter 2 CT Safety
Warning:
Improper use of X-ray equipment may lead to personal injury. Please read and
understand the guidance provided in this book before operating the equipment. Anke
would assist you to put this system into operation.
3. Patients
Give detailed advice to the patients not to move during scan and observe
the breathing instructions (lest any motion artifact is produced). If
necessary, fix thepatient with a safety belt.
Advise the patient to take off all his/her jewelry, hairpins, etc (lest any
metallic artifact is produced);
Make sure the system is clean, because the remaining contrast agent may
produce artifacts.
Dose phantom is an organic glass cylinder. The diameter of the phantom for head
measurement is 160mm, the diameter of the phantom for body measurement is 320mm,
and its height should not be less than 140mm.
17
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
The phantom should have holes big enough to accommodate the radiation detector,
these holes should be parallel with the symmetry axis of the phantom, and its center
point A should be located at the center of the phantom, the peripheral holes are located
10mm below the phantom surface at 90° intervals, as shown in Figure 2-4.
Measurement methods:
CTDI100(Center): CTDI performs check along the rotation axis of the phantom.
CTDI100 (Periphery): Check the average value of CTDI100 measurement around the
phantom.
The diameter of the phantom is 320mm, mainly used to examine patients with big,
medium and small sizes.
18
Chapter 2 CT Safety
Bowtie Filter
Small Large
CTDI
CTDI100(center) (mGy) 4.85 5.13
CTDI100(Periphery) (mGy) 9.14 10.47
WeightedCTDI100(CTDIw)(mGy) 7.71 8.68
Bowtie Filter
Small Large
kVp
80 0.33 0.34
100 0.63 0.63
120 1.00 1.00
140 1.43 1.42
Bowtie Filter
Small Large
Collimator
20 mm 1.00 1.00
10 mm 1.06 1.06
5 mm 1.19 1.20
2.5 mm 1.36 1.37
1.25 mm 1.85 1.86
The diameter of the phantom is 160mm, mainly used to examine heads and pediatric
patients.
(Technical condition: 120 kVp, 100mA, 4 rotation, large focal spot, collimator
width: 20mm)
Bowtie Filter
Small Large
CTDI
CTDI100(center) (mGy) 16.62 16.97
19
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Bowtie Filter
Small Large
kVp
80 0.36 0.37
100 0.65 0.65
120 1.00 1.00
140 1.39 1.39
Bowtie Filter
Small Large
Collimator
20 mm 1.00 1.00
10 mm 1.07 1.07
5 mm 1.22 1.22
2.5 mm 1.40 1.40
1.25 mm 1.91 1.91
Volume CTDIW(CTDIvol)
Axial scan without increment: CTDIw * [Scan number of patient table at the
fixed position]
CTDI(Air)
20
Chapter 2 CT Safety
Maximum error
21
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Pitch=0.5, Collimation=20mm
p0.5 20mm 0.625,fwhm=0.95686 err=53.0978 p0.5 20mm 1.25,fwhm=1.2292 err=-1.6662
1 1
0.9 0.9
0.8 0.8
0.7 0.7
0.6 0.6
0.5 0.5
0.4 0.4
0.3 0.3
0.2 0.2
0.1 0.1
0 0
-2.5 -2 -1.5 -1 -0.5 0 0.5 1 1.5 2 2.5 -2.5 -2 -1.5 -1 -0.5 0 0.5 1 1.5 2 2.5
p0.5 20mm 2.5,fwhm=2.187 err=-12.5196 p0.5 20mm 5,fwhm=4.639 err=-7.2199 p0.5 20mm 10,fwhm=9.5239 err=-4.761
1 1 1
0 0 0
-2.5 -2 -1.5 -1 -0.5 0 0.5 1 1.5 2 2.5 -15 -10 -5 0 5 10 15 -15 -10 -5 0 5 10 15
Pitch=0.75, Collimation=20mm
p0.75 20mm 0.625,fwhm=0.95849 err=53.359 p0.75 20mm 1.25,fwhm=1.2363 err=-1.0948
1 1
0.9 0.9
0.8 0.8
0.7 0.7
0.6 0.6
0.5 0.5
0.4 0.4
0.3 0.3
0.2 0.2
0.1 0.1
0 0
-2.5 -2 -1.5 -1 -0.5 0 0.5 1 1.5 2 2.5 -2.5 -2 -1.5 -1 -0.5 0 0.5 1 1.5 2 2.5
p0.75 20mm 2.5,fwhm=2.2107 err=-11.5706 p0.75 20mm 5,fwhm=4.6294 err=-7.4123 p0.75 20mm 10,fwhm=9.4056 err=-5.9444
1 1.2 1
0.9 0.9
1
0.8 0.8
0.6 0.6
0.6
0.5 0.5
0.4
0.4 0.4
0.2 0.2
0
0.1 0.1
0 -0.2 0
-2.5 -2 -1.5 -1 -0.5 0 0.5 1 1.5 2 2.5 -15 -10 -5 0 5 10 15 -15 -10 -5 0 5 10 15
Pitch=1.0, Collimation=20mm
22
Chapter 2 CT Safety
0.9 0.9
0.8 0.8
0.7 0.7
0.6 0.6
0.5 0.5
0.4 0.4
0.3 0.3
0.2 0.2
0.1 0.1
0 0
-2.5 -2 -1.5 -1 -0.5 0 0.5 1 1.5 2 2.5 -2.5 -2 -1.5 -1 -0.5 0 0.5 1 1.5 2 2.5
p1.0 20mm 2.5,fwhm=2.1602 err=-13.5933 p1.0 20mm 5,fwhm=4.5943 err=-8.115 p1.0 20mm 10,fwhm=9.5196 err=-4.804
1 1.2 1
0.9 0.9
1
0.8 0.8
0.6 0.6
0.6
0.5 0.5
0.4
0.4 0.4
0.2 0.2
0
0.1 0.1
0 -0.2 0
-2.5 -2 -1.5 -1 -0.5 0 0.5 1 1.5 2 2.5 -15 -10 -5 0 5 10 15 -15 -10 -5 0 5 10 15
Pitch=1.375, Collimation=20mm
p1.375 20mm 0.625,fwhm=0.92747 err=48.3951 p1.375 20mm 1.25,fwhm=1.1988 err=-4.0929
1 1
0.9 0.9
0.8 0.8
0.7 0.7
0.6 0.6
0.5 0.5
0.4 0.4
0.3 0.3
0.2 0.2
0.1 0.1
0 0
-2.5 -2 -1.5 -1 -0.5 0 0.5 1 1.5 2 2.5 -2.5 -2 -1.5 -1 -0.5 0 0.5 1 1.5 2 2.5
p1.375 20mm 2.5,fwhm=2.1612 err=-13.5519 p1.375 20mm 5,fwhm=4.5421 err=-9.1573 p1.375 20mm 10,fwhm=9.311 err=-6.8903
1 1.2 1
0.9 0.9
1
0.8 0.8
0.6 0.6
0.6
0.5 0.5
0.4
0.4 0.4
0.2 0.2
0
0.1 0.1
0 -0.2 0
-2.5 -2 -1.5 -1 -0.5 0 0.5 1 1.5 2 2.5 -15 -10 -5 0 5 10 15 -15 -10 -5 0 5 10 15
23
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Pitch=0.5, Collimation=10mm
p0.5 10mm 0.625,fwhm=0.97068 err=55.3091 p0.5 10mm 1.25,fwhm=1.2451 err=-0.39114
1 1
0.9 0.9
0.8 0.8
0.7 0.7
0.6 0.6
0.5 0.5
0.4 0.4
0.3 0.3
0.2 0.2
0.1 0.1
0 0
-2.5 -2 -1.5 -1 -0.5 0 0.5 1 1.5 2 2.5 -2.5 -2 -1.5 -1 -0.5 0 0.5 1 1.5 2 2.5
p0.5 10mm 2.5,fwhm=2.206 err=-11.7611 p0.5 10mm 5,fwhm=4.6261 err=-7.4779 p0.5 10mm 10,fwhm=9.0286 err=-9.7141
1 1 1
0 0 0
-2.5 -2 -1.5 -1 -0.5 0 0.5 1 1.5 2 2.5 -15 -10 -5 0 5 10 15 -15 -10 -5 0 5 10 15
Pitch=0.75, Collimation=10mm
p0.75 10mm 0.625,fwhm=0.96351 err=54.1617 p0.75 10mm 1.25,fwhm=1.233 err=-1.361
1 1
0.9 0.9
0.8 0.8
0.7 0.7
0.6 0.6
0.5 0.5
0.4 0.4
0.3 0.3
0.2 0.2
0.1 0.1
0 0
-2.5 -2 -1.5 -1 -0.5 0 0.5 1 1.5 2 2.5 -2.5 -2 -1.5 -1 -0.5 0 0.5 1 1.5 2 2.5
24
Chapter 2 CT Safety
p0.75 10mm 2.5,fwhm=2.179 err=-12.8411 p0.75 10mm 5,fwhm=4.5669 err=-8.6618 p0.75 10mm 10,fwhm=9.052 err=-9.48
1 1 1
0 0 0
-2.5 -2 -1.5 -1 -0.5 0 0.5 1 1.5 2 2.5 -15 -10 -5 0 5 10 15 -15 -10 -5 0 5 10 15
Pitch=1.0, Collimation=10mm
p1.0 10mm 0.625,fwhm=0.963 err=54.0803 p1.0 10mm 1.25,fwhm=1.2318 err=-1.4525
1 1
0.9 0.9
0.8 0.8
0.7 0.7
0.6 0.6
0.5 0.5
0.4 0.4
0.3 0.3
0.2 0.2
0.1 0.1
0 0
-2.5 -2 -1.5 -1 -0.5 0 0.5 1 1.5 2 2.5 -2.5 -2 -1.5 -1 -0.5 0 0.5 1 1.5 2 2.5
p1.0 10mm 2.5,fwhm=2.1693 err=-13.2298 p1.0 10mm 5,fwhm=4.5277 err=-9.4458 p1.0 10mm 10,fwhm=9.0273 err=-9.727
1 1 1
0 0 0
-2.5 -2 -1.5 -1 -0.5 0 0.5 1 1.5 2 2.5 -15 -10 -5 0 5 10 15 -15 -10 -5 0 5 10 15
Pitch=1.375, Collimation=10mm
p1.375 10mm 0.625,fwhm=0.98587 err=57.7389 p1.375 10mm 1.25,fwhm=1.2639 err=1.11
1 1
0.9 0.9
0.8 0.8
0.7 0.7
0.6 0.6
0.5 0.5
0.4 0.4
0.3 0.3
0.2 0.2
0.1 0.1
0 0
-2.5 -2 -1.5 -1 -0.5 0 0.5 1 1.5 2 2.5 -2.5 -2 -1.5 -1 -0.5 0 0.5 1 1.5 2 2.5
p1.375 10mm 2.5,fwhm=2.2333 err=-10.6666 p1.375 10mm 5,fwhm=4.6653 err=-6.6931 p1.375 10mm 10,fwhm=9.021 err=-9.79
1 1.2 1.2
0.9
1 1
0.8
0.6
0.6 0.6
0.5
0.4 0.4
0.4
0.2
0 0
0.1
0 -0.2 -0.2
-2.5 -2 -1.5 -1 -0.5 0 0.5 1 1.5 2 2.5 -15 -10 -5 0 5 10 15 -15 -10 -5 0 5 10 15
25
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
26
Chapter 2 CT Safety
27
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Half-value layer when measuring X-ray tube voltages 80 kVp, 100 kVp, 120 kVp and
140 kVp under narrow beam condition.
28
Chapter 2 CT Safety
(kVp)
80 100 120 140
Bowtie Filter
Small Bowtie 4.96 6.29 7.25 8.36
Large Bowtie 4.8 5.94 6.9 7.7
Below is the X-ray stray radiation pattern when the system is working. To be
responsible for the users, we put forward appropriate shielding requirements on
operators.
Condition:
29
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Top View
30
Chapter 2 CT Safety
Side View
2. The maximum X-ray tube current is 420mA, the maximal X-ray tube voltage
that can be achieved from the high-voltage generator is 120kV when operating
under the X-ray tube current.
3. The maximum output power is 50kW, the corresponding X-ray tube voltage of
this output power is 120kV, and X-ray tube current is 420mA.
4. The nominal electric power is 50kW, the corresponding X-ray tube voltage of
this output power is 120kV, X-ray tube current is 420mA, and the loading time
is 4s.
31
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
The following table lists the scanning time (s) required for the same scanning position
to reach the CTDI100 1Gy peripheral irradiation dose at 200 mA. The time shown in
the scan time and the product of the 200mA are generated CTDI100 1Gy (outer)
required mAs. Note that for each of the tables in the table, the combination of the mA
and the scan time to achieve or exceed any of the mAs in this table may have a
radiation effect. For obese patients, mA may be larger than the practical example
shown in the table, should be used to get the mAs to determine the patient's exposure.
Test conditions for 200mA, 1 second speed, the use of film scanning mode.
32
Chapter 2 CT Safety
20 26 42 42
Warning:
Tell the patient to close his/her eyes before switching on the laser.
Make sure the laser positioning lamp has been switched off before departure from the
patient.
When the laser positioning key is pressed, the key is highlighted, and
meanwhile, the laser positioning lamp is switched on.
Instruct the patient to keep his/her eyes closed before the laser positioning lamp
is switched off.
When the laser positioning lamp is turned on, it will automatically enable 60s
delay, and it will be turned off automatically in case of failure of switching off
within 60s.
2.13 Biocompatibility
All the components used in this equipment are in compliance with the requirements on
the biocompatibility of components contacting with human body as described in
relevant provisions of ISO10993-1.
33
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
The signal input and output ports in the system must be connected with the equipment
specified by this Company. In case of connecting with any equipment not specified by
this Company, the equipment must comply with IEC 60601-1 Medical Electrical
Equipment Party 1: General Requirements on Safety or equivalent safety standards,
and the system after connection should meet the requirements of IEC60601-1-1
Medical Electrical Equipment Party 1-1: Collateral Standard General Requirements on
the Safety of Medical Electrical System.
During the warranty period, only Anke or the agent authorized by Anke is allowed to
make maintenance and changes on the equipment. In the latter case, there will be a
copy of technical specifications available, including all the changes in nominal
parameters and work conditions that may be used. The names of relevant companies
and maintenance dates are recorded in the technical specifications.
We strongly advise users to keep the maintenance records, including the maintenance
dates, maintenance personnel and any useful descriptions.
Daily maintenance
X-ray tube: Conduct preheating and fast calibration every morning, for the
purpose of protecting the tube and getting better images.
34
Chapter 2 CT Safety
The requirements on temperature and humidity are shown in the table below:
before power on CT system, the scanner room must meets requirements in the
table for more than half an hour and preset humidity before scanning. The
environment requirements which describe in Table 2-1 meet the cooling
requirements for CT system and envirement requirements for submodules.
Items Scope
Temperature Scanner room:20℃~26℃; Console room:18℃~28℃
Scanner room:30%~70% ( No Condensation ); Console
Humidity
room:20%~80%
Disadvantages of dryness
(2) Dusts tend to be gathered onto the machine causing creepage and
mechanical failure.
Ventilation
Normal ventilation is required in the machine room, and it's advised to install a
scavenger fan. The ions produced upon tube exposure may corrode the circuit
board and is harmful to human body.
Electrostatic protection
Plastic flooring or carpet should not be paved in the machine room, and the
humidity in the machine room should not be lower than the required standard.
Any other wireless transmitting and receiving device outside of the system may
affect the system communication. Therefore, the patient is required to turn off
these devices, e.g. pager, mobile phone, radio, etc.
Patient table
Clean and disinfect the table board, accessories of the patients, immobilizing
bandage and table sheets according to the sanitation and healthcare
requirements.
Weekly maintenance
35
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Clean the outer surface of the console with a piece of lint-free cloth dipped with
neutral detergent without any abrasive effect.
Clean the monitor screen with a piece of lint-free cloth dipped with
isopropanol.
Clean the display and the console with a piece of lint-free cloth dipped with
neutral detergent without any abrasive effect.
Make sure all the functions of patient's accessories, e.g. nylon fastener tape for
fixing patients, are in normal condition.
Maintenance Cycle
Maintenance cycle is shown in Table Table 2-1, maintenance may also be done
according to stipulations in the contract if agreed by the user.
36
Chapter 2 CT Safety
Warning:
It's advised to switch off power supply during cleaning and disinfection process to
avoid electric shock occurs.
Caution:
Prevent water and other liquids from entering the equipment to avoid short circuit.
The methods for equipment and room disinfection must comply with the local laws on
equipment use and laws and regulations with legal force.
2.17.1 Cleaning
Components coated with paint and aluminum surface can only be cleaned with a piece
of wet cloths dipped with soft detergent, and then dried with a piece of dry cloth. It's
forbidden to use any corrosive cleaner or any resolvent and corrosive detergent
orpolishing agent. If you have any doubt on any leaner, please do not use it.
37
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Do not use any abrasive polishing agent. Non-abrasive wax can be used to protect the
coating.
2.17.2 Disinfeciton
All the equipment components, including accessories and junction cables, can be
cleaned and disinfected by a piece of cloth dipped with disinfectant.
Do not use any corrosive or resolvent disinfectant. If you have any doubt on a
disinfectant, please do not use it.
Danger:
It's forbidden to use flammable and explosive disinfectant for spraying, because their
steam may burn and cause personal injury and equipment damage.
Warning:
Spray isn't recommended for disinfection of medical equipment room, because the
steam may penetrate into the equipment and lead to electrical short circuit and
corrosion.
To prevent disinfectant moisture from entering the equipment due to convection, do not
spray disinfectant unless the equipment is entirely covered by a piece of plastic.
When the disinfectant has been volatilized, the plastic can be removed and the
equipment can be disinfected by the recommended method.
In case spraying is adopted, the operator has to make sure all the steam has been
volatilized completely before switching on the power supply of the equipment.
The methods for equipment and room disinfection must comply with the local laws on
equipment use and laws and regulations with legal force.
Disclaimer:
38
Chapter 2 CT Safety
Symbols Description
Grounding
Laser radiation
High voltage!
Hand injury
Ionizing radiation
Electric Shock
Max load
Date of manufacture
39
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Symbols Description
Manufacturer
40
Chapter 2 CT Safety
5. Figure 2-10 is a PDU label, pasted on the surface of the PDU Box.
41
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
42
Chapter 2 CT Safety
43
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Scope: People who are suitable for inspection. The clinical trials don’t include children,
and children’s scanning conditions, radiation doses and contrast doses are different
from adults. Be careful with children.
2.20 Contraindication
CT scan for pregnant women should be treated with caution; the need for gonad and
thyroid structure of X-ray sensitive should do X ray protection during CT scan.
In order to get images with high quality, make sure that the patient’s heart rate does not
exceed 65 bpm before cardiac scan. If the patient has a heart rate of more than 65 bpm,
this could generate movement artifacts and cause a certain impact to the diagnosis.
When the patient’s heart rate exceeds 60 bpm, we do not recommand to continue the
cardiac scan.
44
Chapter 2 CT Safety
Warning:
When this CT system’s life cycle ends, its safety and effectivness will degrade in some
degree. To continue to use this CT system will damage operator and patient.
Add: Block B, LingYa Industrial Zone, Tangtou No.1 Road, Shiyan Sub-district,
Bao'an District, Shenzhen, 518108
Add: Block B, LingYa Industrial Zone, Tangtou No.1 Road, Shiyan Sub-district,
Bao'an District, Shenzhen, 518108
Identification Number:
45
Chapter 3 Introduction to CT System
Abstract
This chapter mainly presents the key hardware components of CT system and provides
necessary background information of the system.
3.1 Overview
This chapter mainly presents relevant hardware components, mainly including: PDU,
Gantry (includes detector、HV Generator、tube and collimiator), patient table and
console (includes computer、monitor and intercom). The background information of the
system provided in this chapter is very necessary for understanding other topics
discussed in the following chapters and sections.
3.2 Gantry
The Gantry is as shown in Figure 3-1 and Figure 3-2.
47
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
(1)Gantry display
(4)Gantry keyboard
48
Chapter 3 Introduction to CT System
(3)Gantry keyboard
The Gantry display shows the current states of the Gantry and the patient table, as
shown in Figure 3-3.
49
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
1. Manufacturer name
3. E-stop indicator: indicating the state of the e-stop, It's in e-stop state when the
indicator is illuminated, and it' will stop the movement of the patient table and
Gantry and as well as the ongoing X-ray exposure.
5. Laser positioning lamp: indicating the state of the laser positioning lamp. It's
ON when the indicator is illuminated.
50
Chapter 3 Introduction to CT System
8. Tube voltage.
9. Tube current.
12. Patient table vertival position: vertical position displayed as absolute position,
the range is over 500mm.
13. Gantry tilt angle: current tile angle. Positive angle means gantry top tilts
towards patient table; negative angle means gantry top moves away from
patient table. unit:°.
Gantry keyboard includes some keys, used to control the movement of the Gantry and
the table. The Gantry keyboard is located on the Gantry, with one arranged on the front
and back respectively, in bilateral symmetry. The Gantry keyboard is shown in Figure
3-4.
51
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Postive angle means gantry top tilt towards the patient table, negative angle
means the gantry top tilt off the patient table. Tilt increases by 0.5 ° ,
ranging :-30°~30°.
4. StopScan
52
Chapter 3 Introduction to CT System
5. StartScan
Press StartScan and scan starts; If delay is pre-set, upon press of StartScan, an
icon of delayed exposure and a countdown hour meter will be displayed, as
shown in Figure 3-5. Time from press of StartScan will be counted down, upon
completion of which x-ray exposure will start automatically.
6. Table Up
Raise the patient table. This key is illuminated at an operable status when the
system is at a static or rotary standby mode.
When the key is held down, the patient table will be raised; Upon release of the
key, the patient table will stop rising.
7. Table In (approaching the Gantry): Make the patient table move towards the
gantry. This key is illuminated at an operable status when the system is at a
static or rotary standby mode.
When the key is held down, it's highlighted, and the patient table will move
forward; The key light will go off upon release, and meanwhile, the patient table
stops moving.
8. Fast
53
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Press this key together with Table Up, Table Down, Table In and TableOut,
gantry tilt in or tilt out, operations will be accelerated. Upon release of this key,
acceleration stops.
Make the patient table descend. This key is illuminated at an operable status
when the system is at a static or rotary standby mode.
When the key is held down, the patient table will descend; Upon release of the
key, the patient table will stop rising. Upon press of the key in the middle of a
scan, the scan will stop and the patient table will descend and come to the
normal operable status.
When the key is held down, the patient table moves to the scan plane and the
illumination of the key is off. If released in the middle before the patient table
moves to the scan plance, the key flashes indicating the movement needs to be
continued. After movement of the patient table to the scan plane, the key
illumination is off and any change of the patient talbe will cause the key to be
illuminated, an operable status.
Alignment Lamp Toggle has two states, i.e. ON and OFF. This key is not
highlighted and laser light is off by default. When pressed, the key is highlighted
54
Chapter 3 Introduction to CT System
and the laser light is switched on, and meanwhile, 60s delay is enabled. Upon
another press, the key light goes off, and meanwhile, the laser light is switched
off. If the laser light isn't switched off within 60s, the laser light will be switched
off automatically after 30s.
Laser alignment lamp is only used for patient positioning. Patient should close
eyes when the laser alighnment lamp is on.
When this key is held down, the patient table is unlocked and this key is
highlighted; When this key is highlighted and the display on the gantry has an
highlighted patient table unlock icon, the table can move freely. Press this key
again, the table is locked and this key as well as the table unlock icon on hte
display is off. This key can not be used when scan loading or scan is ongoing.
When the key is held down, it's highlighted, the patient table moves back to its
original position and descends to the lowest position; Meanwhile, the tilt angle of
the gantry is reset (returns to 0), and it stops moving upon release, and the key
isn't highlighted anymore.
The key is highlighted when it's held down, the patient table rises to the highest
position and more forwards to External Land Mark, and it stops moving upon
release and the key isn't highlighted anymore.
16. Tilt in
Make the Gantry top tilt towards the patient table. This key is illuminated at an
operable status when the system is at a static standby mode.
When held down, the key is highlighted and the Gantry begins to tilt; Upon
release, the key light goes off, and meanwhile, the Gantry stops moving.
55
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Make the Gantry top tilt off the patient table. This key is illuminated at an
operable status when the system is at a static standby mode.
When held down, the key is highlighted and the Gantry begins to tilt; Upon
release, the key light goes off, and meanwhile, the Gantry stops moving.
A breathing indicator is located on the front cover and back cover of the Gantry
respectively. The breathing indicator is as shown in Figure3-6 and Figure3-7.
1. Inhaling indicator
When the automatic voice prompts "Breath in, and hold breath", the green
indicator representing "breath" will be illuminated.
56
Chapter 3 Introduction to CT System
2. Breath countdown
3. Holding-breath indicator
Upon an automatic voice prompt for holding breath, the breathing indicator will
go off, and the amber indicator indicating "holding breath" will be highlighted.
1. Table board
The table board can move into or out of the bore of the gantry.
The foot switch is located below the seat leg of the table. When the operator steps
on “UP” button, the table goes up; when the operator steps on “DOWN” button,
the table falls down.
The assessories are patient table pad, head holder and head holder pad.
The patient table pad and head holder pad are made of PU leather, and the head
holder is made of carbon fiber. All are shown in Figure 3-9.
Warning:
If not fixed well, the head holder may becomes loose and hurt the patient. Make sure
the head holder is well insorted into the socket on the patient table before using it.
Caution:
It is highly recommend wrapping the patient table pad with disposable sheet to prevent
soiling during scanning.
58
Chapter 3 Introduction to CT System
3.4 PDU
PDU provides power supply to the patient table and Gantry. The unit is a separate
hardware, which together with the patient table and Gantry are usually located in the
same room.
1. PDU power indicator, the indicator turns gress when power on.
59
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
The function of PDUs in Figure 3-10 is the same, we will provide the adequate PDU
according to the local power voltage.
60
Chapter 3 Introduction to CT System
Monitors
Computers
Mouse
Keyboard
Intercom
3.5.1 Monitor
3.5.2 Mouse
61
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
operation Function
Roll the mouse wheel View the previous/next page
left click and drag Enlarge or shrink an image
Press the mouse wheel and drag Remove an image
Right click and drag Adjust window width of an image
Double click the left button Max./Current window switch
3.5.3 Intercom
E-stop button is placed on the intercom, used for remote control of the Gantry
and patient table. Upon a press on the e-stop button, the power supply of patient
table and gantry will be switched off, including the ongoing X-ray exposure.
Press again to restore.
62
Chapter 3 Introduction to CT System
Upon a press on the e-stop button, the red emergency shutdown indicator is
illuminated, indicating it's currently under emergency shutdown state.
Press on the key for 3s to start the system, the system switch on indicator is
illuminated and the gantry display shows the process of system switch on.
Press on the key for 3s to shutdown the system, the system switch off indicator
is illuminated and the gantry display shows the process of system shutdown
4. Exposure indicator
Indicate the exposure state of X-ray. It's at exposure state if the amber indicator is
on.
6. Communication indicator
When pressing the key, the speaking indicator is illuminated and release of the
key makes the listening indicator on.
9. Communication key
Press the key to enable the doctor to talk to the patient; Release the key to
enable the patient to talk to the doctor. The key is released by default.
When the key is held down, it's highlighted, the patient table moves back to its
original position and descends to the lowest position; Meanwhile, the tilt angle of
the gantry is reset (returns to 0), and it stops moving upon release, and the key
isn't highlighted anymore.
63
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
When the scan interface indicates the patient table needs to move to the
required place, this key flashes, an operable status. Press and hold this key and
it becomes highlighted. The key light is off when the patient table reaches the
required place; If this key is released before the patient table reaches the
required place, the key flashes and the scan interface indicates the continual of
the operation.
Upon scan completion, the scan interface indicates the gantry needs to move to
a designated angle, meanwhile this key flashes, an operable status. Press and
hold of this key, it becomes highlighted. The key is off when the gantry rotates
to the designated angle. If released before the gantry rotates to the designated
angle, this key flashes and the scan interface indicates the continual of the
operation.
The operating and scan interface shows real time changes of patient table
positions and gantry angles.
Make the patient table move towards the gantry. This key is illuminated at an
operable status when the system is at a static or rotary standby mode.
When the key is held down, the patient table will move forward; when release
this key, the patient table stops moving.
14. StartScan
64
Chapter 3 Introduction to CT System
Press StartScan and it becomes highlighted. The light of the key is off and
exposure indicator is on when exposure starts.
15. StopScan
Caution:
Power on and power off the system can only be used for partical conditions, such as
service engineers or the blackout in hosipital and so on. For operators, please refer to
3.7 Daily turn on and turn off.
3.6.1 Power On
Before power on the system, please make sure that the environment of scannning room
is in an appropriate condition.
1. Open the door of distribution box, switch on the main power switch.
3. Open the door of PDU, put the main switch to “ON”, and the indicator on the
door turns to green, PDU is power on.
4. Press the [system switch on/off] button on the intercom 3 seconds to open the
power of gantry, and the screen shows the process of power on.
65
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Warning:
If the PDU is powered off for more than 30 min, after power on, you need to wait an
hour before calibration and scanning.
1. Press the [system Start/Shutdown] button on the intercom for 3s to switch off
gantry power supply and the display on the gantry shows the process of system
shutdown.
3. Open the door of distribution box, switch off the main power switch.
If you turn off the CT system by one button, you can follow the steps to turn on system.
1. Press the [system switch on/off] button on the intercom 3 seconds to open the
power of gantry, and the screen shows the process of power on.
66
Chapter 4 Operation/Scan Interface
Abstract
This chapter mainly gives a brief introduction to the console software and helps users
achieve a macro-image of the software.
4.1 Overview
The Clearview software is developed on the Windows platform, applicable to the CT
systems.
The software has strict user authorization, and user responsibility and authority are
separated. The system covers patient registration, examination, image 3D, 3D
processing and other procedures, and it also supports film printing and can be
connected to HIS system according to the condition of the hospital.
67
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
4.3.1 Login
Before executing scan in the system, the user needs to log into the system, so that the
system can identify the user type.
68
Chapter 4 Operation/Scan Interface
Enter [Password].
4.3.2 Exit
Quit safely before leaving the system. If it is scanning or has unfinished scanning, you
cannot exit the system. Please close the current patient before exit the system.
69
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
3. Select [Close Computer] to close the computer on the console, this is mainly
used to restart ClearVIEW software.
When switching to another Login account, the current user login should be canceled.
1. Current login account is shown at the bottom of the interface, check if account
switch is necessary.
4.3.5 Help
70
Chapter 4 Operation/Scan Interface
2. Select [Help] and you get the Operation Manual of this system.
1. Click the button at the bottom right corner of the interface, select
[About] to check to expire of code and dog. The code will be alarm 3 days
before expiration, and dog will be alarm 7 days before expiration. Users should
apply for new code or dog updating file from ANKE before expiration.
2. Update code
Copy new code file which Anke provides to U Disk, then click <Update Code>
key, a dialogue box appears, choose the new code from U Disk and you can
activate the new code.
3. Update dog
Click <Update Dog> button, and pops up an interface to update the dog. First,
you have to generate a request file and email to zengkai@ankenj.com, then
update dog after you get the updating file.
71
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
This section describes the system taking [Scanning] interface as an example in Figure
4-6.
2. Tool Bar
4. Set button, Press the botton to exit, log in/off system and view version
information, or get help.
5. Date, Time
72
Chapter 4 Operation/Scan Interface
6. Current user
Warning:
Nothing can be placed in the gantry during warm up or calibration. Anything in the
gantry may cause disturbance to the scan images.
4.6.1 Warm up
73
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Notes:
If the heat capacity in the ClearVIEW is less than 10, do a warm up. If you keep on
scanning, the system will remind you to do a warm up before scanning.
Notes:
If the PDU is powered off for more than 30 min that makes detector cool, you have to
wait about one hour after power on, which can warm up detector to its working
temperature.
74
Chapter 4 Operation/Scan Interface
Fast calibration is one part of the daily operation after warm up, which can ensure the
high quality images. So we suggest that do a fast calibration every 24 hours.
2. Click <Confirm> to enter fast calibration, as shown in Figure 4-8. If the system
hasn’t been done a basic calibration or a full calibration for more than 3 months,
you have to do a full calibration first.
3. Choose parameters of the fast calibration, and then click <Load> to execute
calibration. The system will check the period of validity in this loading.
75
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
4. The interface appears a safety tip box, please follow the tips to make sure
everything is done in the field. Click <Confirm> after completion.
5. After loading, the system reminds you to tilt gantry / move patient table, and
the <Table move/gantry tilt> button on the intercom blinks in blue at the same
time. Press and hold this button until the gantry/table move to the destination,
and the promping box disappears.
6. At the moment, the <StartScan> button on the intercom blinks in green, press
the button to start the calibration. The indicators in the system’s interface and
on the intercom light up simultaneously during exposure.
To get high quality images, we need to do a full calibration if the system has run a long
period.
Warning:
Before execute every calibration, a safety tip box will be pop up, please make sure that
all the requirements have been met.
2. Select <Air Calibration> button to enter air and scout calibration interface.
(2) The interface appears a safety box, please follow the tips to make sure
everything is done in the field. Click <Confirm> after completion, as shown in
Figure 4-9.
76
Chapter 4 Operation/Scan Interface
(3) After loading, the system reminds you to tilt gantry / move patient table,
and the <Table move/gantry tilt> button on the intercom blinks in blue at the
same time. Press and hold this button till the gantry/table move to the
destination and the promping box disappears.
(4) At the moment, the <StartScan> button on the intercom blinks in green,
press the button to start the calibration. The indicators in the system’s interface
and on the intercom light up simultaneously during exposure.
(5) The that in front of the calibration paremeters turns to green after
the calibration completed.
77
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
(1) Put the phantom on the phantom mounting bracket, as shown in Figure 4-10.
(3) The interface appears a safety box, please follow the tips to make sure
everything is done in the field. Click <Confirm> after completion,
(4) If the height of the table cannot meet the requirements, the interface will pop up
[Moving patient table] box, press and hold <Moving> button until the table
moves to the presetting position.
(5) After loading, the system reminds you to tilt gantry / move patient table, and
the <Table move/gantry tilt> button on the intercom blinks in blue at the same
time. Press and hold this button until the gantry/table move to the destination,
and the promping box disappears.
(6) At the moment, the <StartScan> button on the intercom blinks in green, press
the button to start the calibration. The indicators in the system’s interface and
on the intercom light up simultaneously during exposure.
(7) The that in front of the calibration paremeters turns to green after the
calibration is completed.
78
Chapter 5 Scan Module
Abstract
This mainly describes general procedures for CT scan operation and image viewing &
processing.
79
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
2. Soothe patient by explaining the noise made by the machine during scan.
3. Let patients take depressants as required by the doctor if the patient is not
cooperative, such as a baby or fainted person.
5. Check drug allergy history and diseases that don't go with contrast agent for
patient that needs CT enhancement and determine if allergy test is necessary.
1. Fix headholder or foot pad onto the patient table board at the end towards the
gantry.
2. When the table board is locked, make clear if lying on the back or on the side is
needed, if the head or the feet need move into the gantry and then let the patient
lie down.
80
Chapter 5 Scan Module
Caution:
To avoid patient injury during table motion, more attention should be paid to patient
hands, feet, loose clothes and IV drip pipes and oxygen pipes. Make sure no obejct
becomes an obstacle during table motion.
3. Lift the patient table board as requested by patient body part to scan.
4. Switch on the laser lamp and make it parallel with the anatomy reference and
align with the center of the anatomy part.
Danger:
Always instruct the patient to avoid staring at the laser, the positioning laser may cause
damage to eyes.
5.4 Register
Before scanning, register the patient, through one of the following methods.
Appointment
Scanned patient
1. Click <Register> button in the [Scan] interface and the patient registration
interface pops up as Figure 5-3.
81
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
2. Type patient and doctor information, the words in red are required items. Click
<Start Scan> to register and enter [Select Scan Protocol] interface.
5.4.3 Appointment
Appointment means that you can register patients’ information in advance to save time.
Patients who are already registered but not scanned will be transferred into the
appointment list automatically.
82
Chapter 5 Scan Module
3. Click <Refresh> button to show all appointed patients of the system in the list.
4. After the patient is selected, click <Start Scan> button to load appointed patient
information to [Register] tab.
If you want to connect Worklist server, pre-set the IP address in [Service] to connect
the station server of the hospital.
83
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
3. Select one patient from the list, click <Start Scan> button to load patient
information to [Register] tab.
A registered patient (or HIS) or patient of second visit can be searched as below:
84
Chapter 5 Scan Module
2. Search the specific patient in the [Filter], then choose the patient and click
<Rescan> button to load the scanned patient information to [Register] tab.
85
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Divide patients into [adult] and [Child], and child can also be grouped
according to age and weight.
head
neck
chest
abdomen
lumbar/thoracic vertebrae
pelvic cavity
spine
arm,leg
4. Choose a protocol in the group which divided by the patient’s age. Please see
Chapter 12 Protocol Management Module for more information about protocol
configuration.
86
Chapter 5 Scan Module
Caution:
Patient direction on the screen should be same as direction of patient being scanned.
Select [Replace] when you start a new protocol or replace the existed ones;
Select [Append] when you add more protocols to the existed ones.
Caution:
The interval must be more than 2 minutes during the scans in different patients. Don’t
scan if the tube heat capacity is over 90% which can avoid tube damage.
Caution:
For cardiac scan if there is an unstable heart rate, there will be a poor image quality
dialogue pops up after the completion of scan.
87
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Upon selcetion, scan protocols are automatically displayed on the scanning interface
waiting for loading.
3. Dose information
7. Scan parameters
8. Reconstruction parameters
88
Chapter 5 Scan Module
Scan protocol modification includes: add scan protocols, change scan parameters, edit
scan items.
Both the scan and reconstruction parameters can be edited Please refer to Chapter 12
Protocol Management Module for scan parameters and explanations.
Caution:
If the current is too large or excessive scan times, the interface will pops up a dialogue
to alarm the scan dose is too high.
Replicate a scan item is used to add a scan item in a protocol, or an enhancement scan
is needed for comparison after a helical scan.
2. Click <Copy Scan> in the scan items edit area, and this scan item can be
replicated.
2. Right click the scan item and select [delete] in the pop-up menu to delete the
scan item, as shown in Figure 5-9.
89
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Add a reconstruction
Multi-recons can be pre-set for each scan, the selected reconstruction can be copied
when you choose to <add Recon>.
2. Click <Add Recon> in the scan items edit area, and a reconstruction can be
added. A circle and a reconstructive item will be added in the interface.
Delete a reconstruciton
Right click the recon item and select [Delete Recon] in the pop-up menu, as shown in
Figure 5-9.
2. When the load is complete, the interface indicates to move the gantry/patient
table, and meanwhile the <Table Movement/Gantry tilt> button on the intercom
flashes in blue. Press the button until the gantry/table moves to the desired
position and the prompt box on the interface disappears.
90
Chapter 5 Scan Module
6. When the load is complete, the interface indicates to move the gantry/patient
table, meanwhile the <Table Movement/Gantry tilt> button on the intercom
flashes in blue. Press the button until the gantry/table moves to the desired
position and the prompt box on the interface disappears.
7. The interface indicates to expose, and meanwhile the <StartScan> button on the
intercom flashes in green, Press the button to start a scout scan.
Caution:
Press <Emergency Stop> button on the intercom in an emergency, the scanning will be
stopped immediately, and the whole system will be locked.
91
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
(1) Firstly, perform scout and helical scan, the steps are same as 5.6.2
Workflow of a routine scan.
(2) Click <Load> to continual following enhancement scan. When the load
is complete, the interface indicates to move the gantry/patient table, meanwhile
the <Table Movement/Gantry tilt> button on the intercom flashes in blue. Press
the button until the gantry/table moves to the desired position and the prompt
box on the interface disappears.
(3) Configure the matching injector, when the interface indicates to expose,
the <StartScan> button on the intercom flashes in green, Press this button and
injector button.
(4) Wait for the end of count-down for delay time, the system will start a
enhancement scan automatically.
(1) Firstly, perform scout and helical scan, the steps are same as 5.6.2
92
Chapter 5 Scan Module
(3) When the pre-monitor scan is complete, set the CT threshold in the
[Monitor] tab, the default threshold value is 100, as shown in Figure 5-13, and
then click <Set Track Region>.
(4) Move the mouse to the scanned image, and the mouse pointer changes
93
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
to a green circle which is sizable. Put the green circle into blood vessel to be
monitored.
(5) Press <StartScan> button and injector button at the same time. If the
delay time in the interface is over, the system will begin to monitor the CT
value through exposure which will be over when the scan times set in the
system are completed. When the monitoring to the blood vessels of CT value
reaches the threshold, the system automatically trigger the next scan.
(6) If the real-time curve can’t reache the threshold value, click <Skip
Monitor> in the [Monitor] tab after scan is completed to trigger the next scan
manually.
(7) Tilt gantry/move patient table according to the indicates pops up in the
interface. When the delay time is over, the system will excute next
enhancement scans automatically.
Caution:
Press <Emergency Stop> button on the intercom in an emergency, the scanning will be
stopped immediately, and the whole system will be locked.
94
Chapter 5 Scan Module
In case of emergency, the ongoing scan may be canceled by pressing StopScan button
on the intercom; If scan is completed but recon is ongoing, press StopScan will not
stop recon; If a countdown mechanism is going on before x-ray exposure, press on
StopScan will stop the countdown and exposure preparation.
You can add one or more scan items after the current scan is completed, please see
5.6.1 Modify scan protocols for details.
Repeat a scan means to have a second scan to the patient on same conditions, which is
used if image quality of the previous scan isn't good enough (i.g. artifacts caused by
patient movements).
1. Choose the scan item which need re-scan in the scan item list.
When more than one scan is programmed in the same TOMOGRAPHIC PLANE there
will be a warning in the interface that this mode has been selected, and the
OPERATOR shall confirm that this is to occur before initiating the scan series.
Append a scan
If the scope of the current scan is not enough, you can append a scan in the scanning
direction. Appending scan cannot only be performed until all the renconstrutive tasks
of the current scan item is done.
95
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Add a bloustrack
Right-click the enhancement scan ,select [Add Blous Track] in the right-click menu,
Figure 5-16 shows the added bloustrack.
96
Chapter 5 Scan Module
If image quality meets clinical requirement, enter the shield room to free
patient positioning, instruct patient to leave or wait for the film.
Caution:
After exam completed, the user shall quit scan program to avoid unnecessary tube
heating and rotation. When the user quits scan program, the gantry will let the slip ring
stop as soon as possible. Before quiting scan program, the slip ring will rotate for a
while due to inertia, which should be avoided as the abrasion to slip ring, belt and
carbon brush caused by this rotation is meaningless.
Measurement by angle;
Measurement by distance;
Caution:
Whether images are enlarged or shrunk, the data measured is calculated from the
original pixel.
Adjustments including:
Helical flip
Vertical flip
Rotate 90°clockwise
97
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Rotate 90°anti-clockwise
98
Chapter 6 View Module
Abstract
This chapter mainly describes the displaying, viewing, measurement of 2D data and
interaction with other modules.
Select one or more series in [Patient] interface and click button in the
[Program] area to jump to [View] interface. Please refer to 9.6
Post-reconstruction for detailed operating procedures.
99
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
1. Observation window
2. Patient list
Show all patients in the view and high light the current patient, and you can load 3
patients mostly in the list.
100
Chapter 6 View Module
Select [All], click any place in observation window to select all images of
current page.
Press Ctrl in stacked or tiled mode to select multi series /images in observation
window.
Press Shift in stacked or tiled mode and meanwhile click the first and last series
/ images to select continuous series /images.
Notes:
Border of selected series/images is yellow in observation window.
Exchange series
Press and hold Alt and drag the mouse can exchange the position of any two
series.
Image layout tools are located in the [Layout] area as shown in Figure 6-3.
101
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
The user may view all scan images in tile mode and view each reconstructed images in
stack. Meanwhile, 5 layout modes to view images are provided. The user may also
customize layout, film playing mode is also supported.
Stacked and tiled modes are available to view images. In stacked mode, all
images of a series are displayed in stack only the first image of each series can
be seen.
Press ALT +mouse, you can transfer the location of any two series in stack
mode.
In tiled mode, all images of a series are displayed in tile. When View the last
image of the current series, scroll down the mouse wheel to view the images of
the next series; when View the first image of the current series, scroll up the
mouse wheel to view images of the previous series.
1*1、1*2、2*2、3*3、4*4.
If the fixed layouts fails to meet actural needs of users, you can customize the
layout interface. Press the left mouse button and drag to select the desired row
and column pattern providing up to 10 lines 9 columns of the pattern.
102
Chapter 6 View Module
Mouse wheel
Side bar
Brower buttons
Film play
In stacked mode the mouse is placed on the right side of the series border as shown in
Figure 6-5, which can be viewed by dragging the scroll bar.
103
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
In different image layout, the number of images shown on each page is also different.
You can press browse buttons to view the series/images on next page, till the last one.
The browser buttons are located on the [Browser] tab, as shown in Figure 6-6.
104
Chapter 6 View Module
This funciton allows viewing all images of the same series rapidly.
1. Select one or more series in stacked mode, or select any image in tiled mode.
2. Click on the [Browse] tab and pops up the [Movie Control] dialog.
105
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
106
Chapter 6 View Module
Measurement by angle;
Measurement by distance;
Caution:
Whether images are enlarged or shrunk, the data measured is calculated from the
original pixel.
107
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
3. Click and hold the left key in the image to draw a line as the first side.
4. Pull out another line as the second side and the angle is shown on the image.
5. When the second side is completed, the measured angle value is diaplayed
directly on the image as shown in Figure 6-11.
108
Chapter 6 View Module
3.Select the distance in the image that needs to be measured, then press the left
mouse button to drag out a straight line.
4. When the measurement is complete, the distance value and unit measured are
displayed on the image as shown in Figure 6-12.
Rectangular/Ellipse
109
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
4. When finished, the measurement results are displayed on the image as shown in
Figure 6-13. Current measurement area, max./min. CT value and
avverage/standard CT value can be seen.
Polygon
4. End the polygon measurement with a double-click and the result is displayed
on the image as shown in Figure 6-14. You can view the current measured area,
length, max/min CT value, average/standard CT value.
110
Chapter 6 View Module
Point
3. Click the point that requires measurement of CT value and its current
coordinate and CT value are displayed as shown in Figure 6-15.
111
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
1. In the observation window, choose the measurement data that needs to move
the position, and the pointer changes to .
6.6.4 Marks
112
Chapter 6 View Module
4. If the marks position affects the display of the image, move away the marked
position by mouse.
113
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
After the following operations, the mouse will be on the previous selection status.
Click in the local box in <Window Interaction Mode> area to release the
mouse.
Angle
Distance
CT value
Mark
Zoom in/out
Move
114
Chapter 6 View Module
Choose a measurement record to be cleared in the image, and press [Delete] key on the
keyboard.
Move images
Zoom images
Rotate images
Adjust WW/WL
Display/hide text
Display grid
115
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Image can be enlarged or shrunk in [View] to make the image fit the observation
window. If the enlarged/shrunk image isn't suitable for viewing in the observation
window, move it to center.
2. Select in [Window Interaction Mode] area and move the mouse to the
selected image; or press the left mouse button to let the pointer changes to
3. Click and drag the mouse in the image and set the size needed.
Right drag the mouse to enlarge the image and left drag to shrink.
Notes:
The original image can be enlarged or shrunk up to 10 times.
116
Chapter 6 View Module
The user can rotate images vertically or horizontally, or rotate vertically and
horizontally simultaneously. Images can be rotated 90 ° clockwise or rotate 90 °
counterclockwise.
In order to facilitate the doctor to observe, the system provides the function that can
view different window values in the same reconstruction algorithm.
If you are not satisfying with the preset window values in a specific scanning, adjust it
and can apply to all images during the scanning.
2. Press and hold the left mouse button and drag the mouse up/down to adjust
window width, and rightward/leftward to adjust window level.
The adjusted values are displayed at the bottom right corner of the segment.
117
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
3.Or select window values pre-defined by click on <Default Tissue Window and
Level> as below:
The <Reset> button returns image display to its original status. This function is applied
to zooming, moving or windowing images.
2. Click in [Image] tab to recover the selected image to its original status.
118
Chapter 6 View Module
2. Click <Grid> in [image] tab to display green grid coordinate on the image.
4. The user may set grid distance and grid reference point in drop-down list of
<Grid> as shown in Figure 6-21.
119
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
1. Add series
2. Clear series
3. Series comparison
120
Chapter 6 View Module
4. Quit comparison
121
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
1. Select the series or images in the left column that needs to be sent to 3D, film
and report.
122
Chapter 6 View Module
123
Chapter 7 3D Module
Abstract
This chapter mainly describes viewing, browsing and reconstruction of 3D images
(MPR、CPR、VR、SSD、MIP).
Caution:
If the transmitted series is not suitable for 3D display, or more than one series is
selected to process, the series can not be displayed in 3D module.
7.2 3D Interface
After loading of images, the 3D interface is shown as Figure 7-1. The left side is a
display area, with MPR (Standard axial view, Standard coronal view, Standard sagittal
view ) and VR images loaded by default. The right side is a toolbar.
125
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
7.3 MPR
Mutil-plane Reconstruction (MPR) refers to reconstruction can be carried out along a
plane or curved plane. If the gantry is not tilt when the CT images are generated, the
reconstructed images (cross section) are perpendicular to the sagittal plane and the
coronal plane. Any plane that is not parallel to these planes is called a slope.
The cross lines on the image can indicate the positions of two imaging planes in
the other two segments.
Move the mouse to the middle of one indicator in the segment, and the cursor
changes to moving status. Press and hold the left mouse button up or down or
left or right to move the indicator line, and the image will be changed in real
126
Chapter 7 3D Module
time.
Move the mouse to the sides of one indicator in the segment, and the cursor
changes to rotating status. Press and hold the left mouse button to rotate the
indicator line, and the image will be changed in real time.
Adjust the slice thickness of MPR’s cross plane, sagittal plane and coronal
plane to improve the definition of slice images. Choose the slice thickness from
the right-click menu of MPR images.
Mean:Normal view
127
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
4. If the user clicks [slice distance same as slice thickness] check box, slice
thickness automatically becomes the value of slice distance. If [Slice Distance]
is locked, [Slice Thickness] becomes locked, too.
6. The start plane, end plane and tilt angle of image slice can be adjusted on the
target image as shown in Figure 7-4.
128
Chapter 7 3D Module
7. Scroll mouse wheel or drag slider bar to view images on the new generated
images, and meanwhile the white indicating lines show the path of image
movement on the target image, as shown in Figure 7-4.
CPR function can make curved, overlapping blood vessels, trachea or bowel and other
structures strenched straightly fully in the same plane.
129
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
3. Click to add points among the tissues (blood vessel, airway, bone
marrow, etc) that need observation on any MPR image. The reconstructed CPR
images are displayed on the lower right side.
130
Chapter 7 3D Module
6. Input [Slice Dic] and [Slice Num] for batch processing of CPR images, as
shown in Figure 7-8.
7. Scroll the mouse wheel or drag the slider bar on the new generated image to
view the image, meanwhile the yellow lines on the targeted image show its
relative position, as shown in Figure 7-8.
131
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
7.4 3D
3D reconstruction is an intuitive technology that can see the human anatomical
structure, the shape and size of the lesion, and know the relationship with the
surrounding tissues.
VR
VR (Volume Rendering) is the technology for direct projection of 3D volume data into
2D images by extracting geometric elements from the data.
2. Select VR template
Click <VR Template> to view VR template list, as show in Figure 7-9. The
system provides many display modules in default. Double-click it to apply to
VR displaying interface.
132
Chapter 7 3D Module
3. Edit VR templates
(1) Click [Edit] tab to enter VR template edit interface, as shown in Figure
7-10. The default VR template provided by the system is the result of multiple
authentications and also supports to edit.
133
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
(2) Click <New> to pop up Figure 7-11. Select [Type] and type new VR
template [Name]. Click <OK> and you can see the current VR image as a new
VR template icon in the VR template list.
134
Chapter 7 3D Module
SSD
2. Drag the cursor to adjust the value of SSD, and the results of adjustments can
be seen in the SSD image.
135
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
MIP is often used to show the direction of blood vessels, and MinIP is often used to
reflect the organizational structure of biliary tract, airway and others.
7.4.2 VOI
VOI refers to shield non-interested area with cutting plate removes the excess
components.
1. Select <VOI> in the [Method] area, and a red three-dimensional box will be
displayed on the 3D image to cover the entire image.
2. Select one plane of the VOI box and move it along the plane to the area of
uninterest, as showin in Figure 7-13.
7.4.3 3D Management
1. Select [3D management] in the [Method] area pops up the following interface.
136
Chapter 7 3D Module
2. After checking the three planes, the position of the cull plane can be seen in the
3D image, as shown in Figure 7-15.
137
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
3. Select the positive or negative culling position of each plane, which can be
displayed in real time on the 3D image. If you need to adjust the position or
orientation of the cutting plane, you can move the position and orientation of
the cross-line on each standard view.
7.4.4 3D Position
The system provides 6 positions, as shown below. Click a position and can be
displayed in real time on the 3D image.
A: Anterior
P: Posterior
L: Left
R:Right
H: Head
F: Foot
7.4.5 3D rotating
138
Chapter 7 3D Module
2. Set [Total Rot Angle], [Rot Angle per Step] or [Total no. of Rot] under
[Parameters] to adjust playing speed.
3. Select rotation direction under [Control] and click <Start> to enable 3D image
to rotate and generate series.
4. Save 3D images
139
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
7.4.7 Boning
140
Chapter 7 3D Module
5. The system provides several ways to auto deboning: Head, Abdomen, leg and
CTA. Figure 7-20 are VR and MIP images applied deboning.
6. Click to cancel the deboning operation in the previous step and click
7. Upon completion, right click the mouse to cancel the deboning status.
142
Chapter 7 3D Module
143
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
144
Chapter 7 3D Module
3. After model selection, the brightness, view angle, color, step lengh and position
may still be adjusted according to the interface.
4. Select [ViewPt Ctrl] check box and the cross –line on the image becomes a
controlling point.
5. Click to adjust the controlling point. Click to adjust the position of controlling
point and right click to adjust direction of controlling point as shown in Figure
7-25. The bottom right corner is the endoscop under controlling points.
145
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
146
Chapter 7 3D Module
(2)Or click to add a guided path. User may add several guided paths.
(1) Select [ViewPt Ctrl] check box, the cross-line becomes a controlling
point.
147
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
(3) After defining position and direction, click <Add point> to see
coordinate of new added controlling points in [Control] list.
(5) Select a group of controlling points and click <Delete Point> to delete
the controlling point.
(5) Select the [Sync Image] check box to synchronize each cross-section
image when playing endoscopic images.
148
Chapter 7 3D Module
1. Click to hide the corner information on the image and click again to
recover.
2. The cross-line means the two lines that meet at right angles displayed on
standard MPR. Click to hide the cross-line on MPR images and click
again to recover.
3. Click to highlight the button. Click the cross-line on the image and
drag or rotate, the cross-line maintains a fixed angle, as shown in Figure 7-29.
149
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Open [3D] interface and a prompt box pops up checking if patient table shall be
removed.
The user may also directly click to enable a prompt box to pop up
asking if patient table shall be removed. Click<Yes> to remove the table board
and check if the removal is correct. Shown in Figure 7-30 is the comparisson of
before and after patient table removal.
150
Chapter 7 3D Module
5. Image layout
7. Reset window width and level in the drop-down menu of <WW/WL Def>
Angle Measurement
Distance measurement
151
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Save to database
Save the image window that needs to save. Click to pop up [Save]
interface and save images to existing image series or new created image series,
as shown in Figure 7-33.
Select image windows that need to send, and click , the screen shot of
images in this window are sent to [Film] interface.
Select image windows that need to save, and click , the screen shot of
images in this window are saved to local.
152
Chapter 8 Film Module
Abstract
This chapter mainly introduces image layout and printing.
After finishing transmitting images that need filming, users may view images in [Film].
153
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
1. Image area
3. Toolbar
Click <Save and Close> to save images of current interface, and to start a new
task.
154
Chapter 8 Film Module
Image layout tools are located in [Layout] area, as shown in Figure 8-4.
155
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Images on several pages can be displayed in tabs. Meanwhile, this system provides 3
fixed layouts to view images. Users may also customize image layout, film playing
images is also supported.
When current page can't display all images, tabs are displayed on the upper right
corner, as shown in the following picture. Click <Previous>/<Next> or[1]/[2]tabs.
The red tab means the current page, and the green tab means more pages.
156
Chapter 8 Film Module
3 fixed layouts are provided in views to view series/images: 3*2, 5*4, 7*5.
Click to rearrange the current page, if you select [Apply layout to all pages]
check box, you can apply the new layout to all pages.
(1) if you want to set the current layout as the conventional one, select
[Save Layout as U1/U2], as show in Figure 8-6. The system provides two
conventional layout buttons for users to define.
157
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
View layout can be customized if the fixed layout fails to meet user demand,
with maximum 10 rows and 9 columns. Select [Apply layout to all pages]
check box, you can apply the new layout to all pages.
Measure an image, different from image measurement in [Scan] and [View]: the
measurement data in film can be printed as image notes, but measurement data in scan
and view can only be viewed temporarily.
158
Chapter 8 Film Module
Measure by angle
Measure by distance
Move images
Zoom images
Rotate images
Gray bar
Customize rotation
Topo Line
Topo Image
159
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Refer to 6.7 Adjust Images and the following section for detailed operations.
1. Click in [Image] tab to display on the image, as shown in the left side
in red lines in Figure 8-10. The gray bar can be the reference of CT value.
160
Chapter 8 Film Module
1. Click in [Image] tab and green rotation frame appears on the image, as
shown in Figure 8-11.
2. Drag the small red circle on the upper right corner of the rotation frame and the
angle value is synchroneously displayed on the image as shown in Figure 8-12.
161
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
2. Click in [Image] tab to display scout image on the bottom right corner
of each image and location of the current image in scout image, as shown in
Figure 8-14.
162
Chapter 8 Film Module
The system provides annotation font size adjustment in order to display the
appropritate font in the corner information for user to print images.
Select [Annotation font size] in the right-click menu, and the font size is shown below.
163
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
164
Chapter 8 Film Module
You can view the failed films in the [Print failed list], as shown in Figure 8-19. And
you can click <reprint> button to print again.
165
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
166
Chapter 9 Patient Management Module
Abstract
This chapter mainly describes interface layout, funcntions and operation instructions of
patient management module.
The main page shows all the patient data of the current data source, Select data
of a patient and the corresponding scan series are displayed in [Series] tab.
Select a series to preview, the images can be zoomed, moved or even the
WW/WL can be adjusted.
The tool bar on the right side is used to manage patient data.
167
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
In the [Data Source] area box choose data source, as shown in Figure 9-2.
Select <Local DB> to display patient data in current database in the left side
list.
DVD disk is inserted in the DVD driver. Select and the patient data in the disk
is displayed in the left side list.
1. Select patient data in the database that needs to be modified in the left side list.
168
Chapter 9 Patient Management Module
(1) Select the patient or a series under the patient in current database that
needs to be deleted in the left side list.
169
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Click in [Data Management] area to restore the patient data which has
been deleted before.
1. Select the patient data to be locked in current database from the left side list.
2. Click in [Data Management] area. A lock on the left end of the patient
column indicates current state.
170
Chapter 9 Patient Management Module
In database, patient information can be sorted by name, patient ID, exam ID,
age or gender, as shown in Figure 9-6.
Type name, patient ID, time or their combination on [Filter] area to filter patient
data.
Search a patient
171
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
(2) The user may search patient by [name], [ID], [Gender], [Age], [Check point] or
their combination
The authority to transfer data is assigned by hospital manager and the general users do
not have the operation authority by default.
2. Select data to be copied in left side list., more patients can be selected.
172
Chapter 9 Patient Management Module
5. Wait until completion of data copy. View local database in [Data Source] to
show successful data copy.
Notes:
The system doesn't repeat copy of existing data in LocalDB. It is advisable to see if the
patient information is already saved in database by checking patient data before copy.
The designated patient images can be uploaded to image servers, the configuration of
servers refers to 11.2.5 Workstation Management.
2. Select image storage device in drop-down list in [Image Upload] area and click
<upload> button to begin uploading.
173
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
3. Or you can send patient data or series to image server by select [send to PACS]
in the right-click menu.
174
Chapter 9 Patient Management Module
Export patient data is to export data to local or other images storage devices.
User can export patient data from any data source to the designated directory.
1. Select the patient data to be exported in current database from the left side list.
4. When data export starts, a progress bar appears and it will disappear when data
export is completed.
5. Select DVD in [Data Source] to enter DVD burning mode, as shown in Figure
9-14.
175
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
[List of data to be burnt]: The green indicating bar shows occupation rate of
the data to be burnt in the DVD capacity.
6. If format is not needed, directly click <Burn> to burn patient information into
DVD.
7. If the burning is complete, the interface prompts the burning is over, and the
DVD is ejected. Uers can insert DVD into the driver to view patients date.
176
Chapter 9 Patient Management Module
9.5 Post-reconstruction
1. Choose a patient to be post-reconstructed in the patients list, and then click
2. Click <Add Recon> to create a new recon, edit these parameters and click
<Start Recon> to execute recon. Please see Figure 9-16.
1. Select patient, series or designated images to be transmitted in the left side list.
177
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
178
Chapter 10 Report Module
Abstrast
This chapter mainly introduces the operation steps and precautions of report module.
Select [Report] tab in [Patient] interface, if the patient already has a report,
select [View report] in the right-click menu.
Click in [Report] interface and pops up Figure 10-2 which lists all the
reports of patients. Choose a report and click <Open>.
179
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
180
Chapter 10 Report Module
2. Choose the patient who need a report in the database, and then select a template,
click <New> button to create a report in the [Report] interface, as shown in
Figure 10-4. The report interface is editable.
181
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
3. The doctor can decide whether to apply the report template to the new blank
report based on the scan results. Select the body part to be examined in [Report
Template] and double click the corresponding template. The template can be
loaded in [Findings] and [Diagnosis] of the new report. Selection of patient
template is shown in Figure 10-5. Users can edit the corresponding template
and click <Update Template> to save it.
182
Chapter 10 Report Module
4. Select [Emergency Report] check box in the upper right corner and a remark
can be added in the report, see Figure 10-6.
Save a report
183
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Review a report
Click to send the current report to the doctor with the authority to
review. The authority to review the report is assigned by the hospital manager.
Print a report
184
Chapter 10 Report Module
2. Right click one template to set it as the default template, as shown in Figure
10-9.
185
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
3. If you choost a template with images, you have to pre-load images in the
database to [Report] interface. Choose images to be sent in the [View] interface,
and then click in the [Patient] area to transmit images to [Image List] in
the [Report] interface, as shown below.
186
Chapter 10 Report Module
4. Choose a blank box in the report, and then double-click the image to be loaded
in the [Image list]. Repeat these steps to load another image, as shown in
Figure10-11.
187
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
6. If you want to change the image in report, select [Delete] in right-click menu
and then repeat Step3 to add a new one.
188
Chapter 11 Service Module
Abstract
This chapter introduces servcie modules of the system. Users may set the operating
system based on demand.
Caution:
Administrator or above has the privilege of configuring the settings while the ordinary
users can only view them.
System basic information includes hospital information and machine information, and
click <Save> button after information is typed.
189
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
[Pacs list]:if you enable [Auto upload images], you can also upload images to
multiple workstations at the same time.
[Recon thin slice in helical scan]: To helical scan, the thickness of 0.625mm can
add to reconstruction under 20mm collimation. It is closed by default and only
superadmin or above can enable it.
190
Chapter 11 Service Module
[ECG module]: the system supports 7800 and iM60 ECG monitors. You need to
select one according to actual ECG monitor during installation.
[Patient name preference in pinyin]: in5.4.4 Select a Patient From Worklist, you
can set whether use Pinyin or Chinese first when search patient in the worklist
server.
Caution:
Administrator or above has the privilege of configuring the settings while the ordinary
users can only view them.
Select [Corner information settings] to enter the interface, as shown in Figure 11-2.
191
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Users can increase, decrease or adjust corner information from actual demanding.
Choose a corner information in the list and click <Merge/Split> button, “**”will be
added in front of this information, which indicates that the information is conbined
with the previous one. Click <Merge/Split> again to split.
Caution:
Administrator or above has the privilege of configuring the settings while the ordinary
users can only view them.
192
Chapter 11 Service Module
Users can customize the window width and level to be display in each module.
1. Select [WW/WL setting] to dispaly the window width and level interface as
shown in Figure 11-3.
193
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
(2) Enter [Name], [Window Width] and [Window Level] value and the type
can be [Slice] or [Topo]. Click <OK> to add a window width and level value in
the list.
3. Modify WW/WL
Choose window width and level value to be edit in Figure 11-3, click <Modify>
to edit all parameters except [Name].
4. Delete WW/WL
Select items that need to be deleted in Figure 11-3 and click <Delete> to delete
corresponding items.
The system provides the following hot keys by default and the user can also define it
by him/her habits.
194
Chapter 11 Service Module
Caution:
Administrator or above has the privilege of configuring the recordings while the
ordinary users can only view or play them.
Voice recording funciton is used to set the prompt sound file by system default. Two
audio prompt messages can be recorded, with one used before and after scanning
respectively. Voice recording is used to provide continuous instructions for breathing
during scanning for precise time counting. A microphone is equipped on the gantry and
on the console for operator - patient communication.
1. Select [Voice Management] to enter Figure 11-6. The default voice is English.
195
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
(2) Type new voice manually and choose add it to voice list, this can make
it easier to choose during scanning.
(3) Choose [Voice before exam] and click <Record>, when speak to
microphone on the intercom. The recording time is controlled in 4 minntes if
the timeout expires.
(4) Click <Stop> and the system will automatically save the recordings.
(5) Record [Voice after exam] as steps described in Steps 2(3) and 2(4).
(6) If you like to play any recorded voice messages, click <Play> to play
back the audition.
Notes:
Start and end voice recording as rapidly as possible to minimize time used on starting
and ending of the audio messages.
196
Chapter 11 Service Module
(1) Choose a voice record in the list and click <Modify> to edit it.
(2) Select the recording you want to modify in the list and click the
<Record> button to re-record the voice. For the convenience of management
please try to keep the recording content and interface display recording
information consistent.
Choose a voice record in the list and click <Delete> to delete it.
Caution:
Administrator or above has the privilege of configuring the settings while the ordinary
users can only view them.
Pre-set transmission and storage mediun of images, restart the ClearView software after
the workstations are configured.
197
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
3. Type [Name], [IP Address] and [Port], the [Connection type] can be
worklistScp and StorageScp. Click <OK> to complete the setting.
5. Choose a workstation in the list and click <Modify> to reset IP address, port,
name and connection type.
6. Choose a workstation information in the list and click <Delete> button,and then
in the prompt box to confirm the deletion of the workstation information.
For management convenience, this system has hospital manager, department manager
and operator. The user may log into the scanner after user configuration. The password
is used for verification when the user logs in. The administrator plans users for the
sytem based on actual needs.
Reset password
Hospital manager has right to modify any user's password. If a user forgets his or her
password, hospital manager can reset a password for the user without knowing the
original password. The department manager also has the right to reset the password for
its subordinate operator.
1. Select any user that need reset password at the user list.
198
Chapter 11 Service Module
Add users
Hospital manager is the administrator of this system, used to manage authority and
system maintenance. In general, a department manager account is created for each
department, and an operator account is created for each doctor to facilitate their
independent operation.
199
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
2. Type [User ID], [User Name], [Password] and [Level], click <OK> to add a
new user.
Level Rights
Service engineer’s exclusive account with the highest authority.
Engineer Engineer has high-level authority of after-sales maintenance besides the
operation of ClearVIEW interface.
Hospital manager has authority to operate on the interface except deleting
Hosiptical Manager
engineer account and modifying its password.
Department Manager Permissions are assigned by hospital manager.
Operator Permissions are assigned by hospital manager.
You can modify user information when you want to change the user privilege or type
the error information.
1. Select the user to be modified at user display area (select one row each time).
the selected user becomes highlights.
200
Chapter 11 Service Module
2. Click <Modify > and an interface pops up as shown in Figure 11-12. Only
[User Name] and [Level] can be modified.
Delete a user
Administrator has the right to delete a user that is not needed. It is advisable to delete
users that are not needed regularly.
Select a user in the list and click <Delete>. Then a dialog box pops up, click <OK> to
delete the user.
Only engineer and hospital manager have the right to assign permission.
201
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
2. The privilege is set by user level. The hospital manager can assign the
permission of department manager and operator in the software respectively.
Scan configuration is shown as Figure 11-14. Users can set some of the scan
configuration according to their habits.
202
Chapter 11 Service Module
[Pitch]: different pitches can be defined by different collimators. Users can set
the values of pitch manually, the range is 0.1-2.
Caution:
Administrator or above has the privilege of configuring the settings while the ordinary
users can only view them.
203
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
3. Type [Printer Name], [IP Address], [Port] and print information. Click <OK> to
add a printer.
5. Choose a printer and click <Modify> to reset IP address, port and print
information.
6. Choose a printer and Click <Delete>, confirm to delete the printer address in
the prompt box.
Caution:
Dapartment manager and operator have only the right to view alert setttings while
engineer and hospital manager can configure them.
In order to reduce the patient’s excessive radiation, the system set dose alert according
to different body parts and ages.
1. Select [Dose Alert Settings] to enter the interface, as shown in Figure 11-17.
204
Chapter 11 Service Module
2. If the user is hospital manager, double-click anyone and pops up dose alert
settings interface, as in Figure 11-18. You can customize the value of dose alert.
205
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
3. If the DLP in a scan series is over the maximun value, there will be pop up a
warning prompt box, click <OK> to continue scanning.
If the accumulated dose in the scanning process is exceed the maximum value, the
following box will pops up as shown in Figure 11-20. The user should describe
the reason for exceeding the dose and re-enter the user name and password to
continue scanning.
206
Chapter 11 Service Module
Caution:
Administrator or above has the privilege of configuring the setting while the ordinary
users can only view them.
2. Choose the patient to be deleted, and then click <Delete> to delete it.
207
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Search logs of the sytem by account, starting and ending time to find out any abnormal
situations in time.
208
Chapter 11 Service Module
3. Click <Search> to display results that meet the searching elements on the left
side. Click <Clear> to clear filters.
209
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
2. Type [Part of Exam] and [Diagnosis]. Edit words for the new template in
[Findings] and [Suggestion], as shown in Figure 11-26.
210
Chapter 11 Service Module
3. Click <Add> to complete creating a template. The new template can be viewed
at the list of body position.
211
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
212
Chapter 11 Service Module
213
Chapter 12 Protocol Management Module
Abstract
This chapter mainly describes how to preset scanning protocol of this system.
Each scan must be started from scan protocol. The pre-set scan protocols may help
save time and conduct scan procedures quickly.
A scan protocol is built by protocol, and it contains all scan parameters. A scan protocol
can be built into adult mode and child mode. The system may provide some typical
scan protocol by default, and user may customize scan protocols based on actual needs.
215
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
(2) Protocol area. The image-quality test protocol is used for factory inspection but
not for clinical diagnoses; only the engineer has the authority to select it in a
scanning.
4. Click procotol and scan items and parameters can be viewed below.
216
Chapter 12 Protocol Management Module
Click <Revert Protocols> in the [Protocol] interface, and then choose local
protocals to recover current ones.
2. Type [Protocol Name], [Age Type], [Scan Organ], [Default Protocol] and [Scan
Type], Click <OK> to complete it. The scan parameters are shown in Table
12-1.
Parameter Description
Protocol Name Named by users
Age Type Adult or child
Scan Organ Head, neck, shoulder, chest, spine, abdomen, pelvic cavity, upper
217
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Parameter Description
limbs, lower limbs
Default Protocol Yes or no
Scan Type Helical
Axial
Perfusion
Two Phase Enhance
Three Phase Enhance
Cadic
2. Click <Modify Protocol> in in scan protocol editing area, and pops up Figure
12-3.
3. [Protocol Name], [Age Type], [Scan Organ], [Default Protocol] are editable.
218
Chapter 12 Protocol Management Module
1. Choose the scan protocol to be copied in the [Protocol] interface in Figure 12-1.
2. Click <Copy Protocol> in in scan protocol editing area, and pops up Figure
12-4.
3. [Protocol Name], [Age Type], [Scan Organ], [Default Protocol] are editable.
4. Click <OK> to copy the protocol, the successed copied protocol highlights in
Figure 12-5.
219
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
2. Click <Delete Protocol> in in scan protocol editing area, and pops up a delete
confirmation box.
220
Chapter 12 Protocol Management Module
Click <Add TOPO>, <Add Helical>, <Add Axial> and <Add BlousTrack> in the scan
items editing area to add a new scan item. The highlighting in Figure 12-8 is the
created axial scan.
221
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Choose the scan item need to be copied in scan items list, and click <Copy Scan> to
copy a scan item. The highlighting in Figure 12-9 is the copied scan item.
2. The [Rename scan] interface pops up, as shown in Figure 12-11. Modify the
scan name.
222
Chapter 12 Protocol Management Module
Choose the scan item to be delete, and then click <Delete Scan> or select [Delete Scan]
in right-click menu to delete it.
Choose the scan item need to add a recon item, and click <Add Recon>. As
shown in Figure 12-12, the red circle is the added recon item.
Choose the recon item need to be deleted, and click <Delete Recon> or select
[Delete Recon] in right-click menu to delete it.
223
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Parameter Description
kV Tube voltage, i.e. 80kV,100 kV,120 kV and 140kV
Tube current, choose a value between 10mA to 420mA, with 10mA as
mA
increment.
Scan Length The scan length of topo scan
The position of tube in 0 degree, 90 degree, 180 degree, 270 degress,or
Tube Position
double scout.
Table direction determines patient table move in or move out gantry when
Direction
scanning, including tableIn and tableOut
Series Description The description of current Topo scan.
WW/Ll provide by the system: TopoHead, TopoNeck, TopoBody,
WW/WL
TopoLimbs, TopoSpine, PediatricBody, PediatricLimbs.
WW/WL Customized by users
Parameter Description
224
Chapter 12 Protocol Management Module
Parameter Description
Scan types includes Helical, Axial, Perfusion, Two phase enhance, Three
Scan type
phase enhance, cadic
Enhancement Whether or not enhanced scanning
kV Tube voltage, i.e. 80kV,100 kV,120 kV and 140kV
Tube current, choose a value between 10mA to 420mA, with 10mA as
mA
increment.
Scan length The length of scanning. The length is a slice in pre-monitor phase.
Refers to the time for tube rotate a round (Unit:s), includes 0.5(optional),
RotSpeed
0.8, 1.0, 1.5, 2.0
Total collimator, determines the minimum thickness, used for
reconstruction.
Coverage
For helical, collimator includes 20mm and 10mm
For axial, collimator includes 20mm, 10mm, 5mm, 2.5mm and 1.25mm
Table direction determines patient table move in or move out gantry when
Direction
scanning, including tableIn and tableOut
Voice Open or close the voice, the default language is English.
Bowtie Bowtie filter can be head or body
The gantry need tilt in lumber disc axial examination, the tilt angel is
Tilt
between ±30°
You can get a series of images which have different title angles with scan
Digital Tilt
plane by post-processing, and the digital tilt range is between 50°.
Pitch is a helical scan paremeter, determins the patient table speed and
acquisition speed.
The bigger pitch value is , the faster table moves, the shorter scan period is.
Pitch
Pitch can be 0.5,0.75,1.0, 1.375or 1.5, bigger pitch value enables larger
coverage area within a given scan time, but sometimes may generate low
quality image. Pitch should be selected based on image quality.
Focus Spot Big or small, the smaller the focus spot, the higher resolution of images.
DFS Flying Focal Spot, none or X/Z axis.
Enable or disable AEC, automatic modulate mA on the body or parts. It is
Adose only related with and can only be seen on the topogram when it is enabled.
The Adose Mode can be selected from Quality, Balance and LowDose.
Whether or not groupscan, if you choose YES, the system will
GroupScan automatically expose after the last scan. Generally, we choose groupscan in
enhanced, perfusion or lumbar disc scan.
The center of topogram, we set a little offset in lumbar disc scan and
Scan Center
thorax-Abdomen scan
The times of scan in one slice, such as once in pre-monitor phase, 30th in
Count
monitor phase and 20th in head perfusion.
ExpInterval The interval in two explosion, the default interval in monitor phase is 2s
225
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Parameter Description
Whether or not trigger the next scan by CT threshold, if you choose yes,
Blous Track
you have to set the value of CT threshold. The default value is 100.
The time from pressing startscan to scan starting, Unit: s. Generally, it
Delay
refers to delay time after using the bolus injector.
Parameter Descripiton
Description To descripe the current reconstruction
Thickness of reconstruction image, 0.3125, 0.625,1.25,2.5,5,7.5,10 can
Thickness
be selected, Unit:mm
The space between two slices, 0.3125, 0.625, 1.25, 2.5, 5, 7.5, 10 can be
Spacing
selected. Unit: mm.
226
Chapter 12 Protocol Management Module
227
Appendix A Guidance and Manufacturer’s
Declaration-IEC60601-1-2
Warning:
Guidance and manufacturer’s declaration-electromagnetic emission for all EQUIPMENT AND SYSTEMS.
RF emissions Group 1 The 64 Clarity uses RF energy only for its internal
CISPR 11 function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
RF emissions Class A The 64 Clarity is suitable for use in all establishments
CISPR 11 other than domestic and those directly connected to the
Harmonic emissions N/A public low-voltage power supply network that supplies
IEC 61000-3-2 buildings used for domestic purposes.
Voltage N/A
fluctuations/flicker
emissions
IEC 61000-3-3
The 64 Clarity is intended for use in the electromagnetic environment specified below. The customer or
the user of the 64 Clarity should assure that it is used in such an environment.
A-1
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
A-2
ANATOM 64 Clarity X-ray System—Operation Manual
d = 1.2
p 80 MHz to 800 MHz
d = 2.3
p 800 MHz to 2,5 GHz
3 V/m
where P is the maximum output
Radiated RF 80 MHz to 3 V/m
power rating of the transmitter in watts
IEC 2,7 GHz 80 MHz to
(W) according to the transmitter
61000-4-3 2,7 GHz
manufacturer and d is the
recommended separation distance in
metres (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey, should be
less than the compliance level in each
frequency range.
Interference may occur in the vicinity
of equipment marked with the following
symbol:
A-3
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
disturbances are controlled. The customer or the user of the 64 Clarity can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the 64 Clarity as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum Separation distance according to frequency of transmitter
output power of m
transmitter 150 kHz to 80 MHz 80 MHz to 800 800 MHz to 2.5
W p MHz GHz
d = 1.16
d = 1.16
p d = 2.33
p
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
A-4
Appendix B Technical Information
1. Gantry
Feature Description
Dimension of gantry(length×width
2313mm×1144mm×1956mm
×heigtht)
Weight of gantry 2120kg
Minimum patient opening 700mm
The distance between focal spot to
540mm
iso-center
Position indicator Transverse, Sagittal
The consistency between the center of
the position light field and the center ≤2mm
of the narrowest tomographic section
Gantry tilt ±30°, in 0.5° increments
Transverse (internal and external)
Scan plane indicators Coronal
Sagittal
Bowtie filter Head, Body and Pediatric, selectable by protocol
Automatic exposure control (mA Yes
modulation) Azimuthal and Z axial modulation provided
External X-ray enable Foot pedal provided
ECG gated Provided
Breath graphical display
Auto-voice management Hold message (Record/Playback)
Breath message (Record/Playback)
Emergency scanning provided
2. Patient table
Feature Description
Minimum height No more than 430mm
Vertical-movable range No less than 500mm
Horizontal-movable range No less than1850mm
Horizontal-scannable range No less than 1800mm
Horizontal - speed 150mm/s maximum
Horizontal speed accuracy +/- 0.5 %
Horizontal – position accuracy +/- 0.25mm
Horizontal – position repeatability +/-0.25mm with a distributed load
B-1
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
3. Detector
Feature Descripiton
Detector material GOS
Width of detector material 1.6mm
Numbers of detector columns 912×32
Coverage (projected to center of
500mm
rotation)
Pixel size (center) 0.625mm×0.6mm
4. PDU
Feature Descripiton
Input voltage 3~,AC380V / AC480V
Output voltage AC220V/AC380V, DC24V/DC48V
5. HVPS
Feature Description
Input power 400VAC, +10/-20% ~3PH, 50/60Hz 190A
kV selections’ range 60kV~140kV
Tube kV accuary ±0.5%
mA selection’ range 10mA~420mA
Tube mA accuary ±2%
Maximun continuous output of HV
50kW
generator
6. Tube
Feature Description
Nominal voltage 140kV
Heat capacity 5.5MHU
Maximum cooling rate 9.6 kW
Small focal spot: 0.6mm×1.2mm
Nominal focal spot(IEC 60336)
Large focal spot: 1.1mm×1.2mm
Target angle 7°
7. Console
Feature Description
B-2
ANATOM 64 Clarity X-ray System—Operation Manual
8. Collimator
9. Scanning
Feature Description
Maximum numbers per rotation 64
Rotate speed 0.5s (Optional), 0.8s, 1s, 1.5s, 2s for selected
Maximum continuous helical scan
120s
duration
Radiation output linear <2.0 %
Scan Field of View 520mm
Scout azimuthal position 0°, 90°, 180° and 270°
Scout speed 150 mm/s
Scout indicator accuracy <1.0 mm
The default pitch is 0.5, 0.75, 1.0, 1.375 or 1.5, user
Pitch can also set pitch manually in the [Sevice] module, the
range is 0.1-2.
Nominal helical reconsturction slice 0.3125mm, 0.625 mm, 1.25 mm, 2.5 mm, 5 mm,
Thickness 7.5 mm 和 10 mm
Cine maximum continuous rotation 120s
Image matrix size 512×512,1024×1024
Range of CT numbers Extended range, from -32767 to 32768
Reconstruction speed Up to 65 frames per second
Operating noise ≤70dB (A weighting)
B-3
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
Feature Description
11. Environment
Operation
Feature Description
Examination room:20℃~26℃;
Temperature
Control room:18℃~28℃
Ambient temperature, rate of change -3 ℃/hr~ 3℃/hr
Humidity @ ambient Examination room:30% ~70% RH
(non-condensing) Control room:20% ~80% RH
Atmospheric pressure 700hPa~1060hPa
Feature Description
Temperature -20℃~55℃
Humidity @ ambient
≤80 % RH
(non-condensing)
Continuous Elapsed time in storage 90 days
12. Accesories
Type Including
Patient table pad
Head holder
Head holder pad
Necessary accessaries
Calibration Phantom(s)
Phantom holder
Tilting head holder
B-4
ANATOM 64 Clarity X-ray System—Operation Manual
13. Lifetime
14. Warranty
1year
B-5
Appendix C Glossary
C
CT Computer Tomography
CTA Computer Tomography Angiography
D
DAS Data Acquisition System
E
EMC Electromagnetic Compatibility
F
FOV Field Of View
H
HDMI High Definition Mulimedia Interface
C-1
ANATOM 64 Clarity X-ray Computed Tomography System Operation Manual
I
IEC International Electrotechnical Commission
IP Internet Protocol
M
MIP Maximum Intensity Projection
P
PC Personal Computer
R
RC Reconstruction Control
C-2
ANATOM 64 Clarity X-ray System—Operation Manual
S
SC Scan Control
U
UPS Un-interruptible Power Supply
V
VGA Video Graphics Array
VR Volume Rendering
C-3