V1.9.60
Contents
1. System Overview........................................................................................................ 5
1.1 Architecture..................................................................................................... 5
1.2 Applications..................................................................................................... 6
1.3 Imaging features............................................................................................ 6
1.4 Standard features.......................................................................................... 7
1.5 Optional features............................................................................................... 7
1.6 Language support.......................................................................................... 7
2. Ergonomics................................................................................................................. 7
2.1 Keyboard......................................................................................................... 8
2.2 Image display screen.................................................................................... 8
2.3 Touch gestures............................................................................................... 8
2.4 Comments....................................................................................................... 8
2.5 Bodymark........................................................................................................ 9
2.6 Peripherals...................................................................................................... 9
2.7 Dimensions and Weight................................................................................ 9
2.8 Electrical Power..............................................................................................9
2.9 Operating Environment................................................................................. 9
2.10 Storage & Transportation Environment.................................................... 9
3. Transducers................................................................................................................ 9
3.1 Transducer Technology................................................................................. 9
3.2 Transducer types............................................................................................9
3.3 Transducer selection................................................................................... 10
4. Advanced Imaging controls...................................................................................... 13
4.1 VFusion..........................................................................................................13
4.2 VSpeckle....................................................................................................... 13
4.3 VTissue.......................................................................................................... 13
4.4 3D/4D............................................................................................................. 13
4.4.1 HQ (optional)*....................................................................................... 13
4.5 Smart operation............................................................................................13
4.5.1 Free View(optional)............................................................................... 13
4.5.2 Inversion mode*(optional)....................................................................13
4.5.3 Magic Cut(optional)...............................................................................14
4.6 Smart 3D....................................................................................................... 14
4.6.1 Volume Measurement(optional)...........................................................14
4.6.2 Auto Follicle(2D/3D)(optional).............................................................. 14
4.7 Multi-angle M mode *.................................................................................. 14
4.8 Next generation RF-based image processing.........................................14
4.9 Stress Echo(optional)*................................................................................ 14
5. Imaging modes......................................................................................................... 14
2
5.1 2D Imaging....................................................................................................14
5.2 Harmonic Imaging........................................................................................15
5.3 M mode..........................................................................................................15
5.4 Color Doppler mode.................................................................................... 16
5.5 Power Doppler mode...................................................................................16
5.6 Pulsed Wave (PW) Doppler........................................................................17
5.7 Continuous Wave Doppler (CWD)............................................................ 17
5.8 Pview*............................................................................................................ 17
5.9 Tview..............................................................................................................18
5.10 Auto.............................................................................................................. 18
5.11 3D/4D........................................................................................................... 18
6. Touch Panel Interface...............................................................................................19
6.1 2D mode........................................................................................................ 19
6.2 M Mode........................................................................................................... 19
6.3 CF mode........................................................................................................20
6.4 PW/CW mode............................................................................................... 20
6.5 3D mode........................................................................................................ 21
6.6 4D mode........................................................................................................ 21
7. System Feature.........................................................................................................21
7.1 Display modes.............................................................................................. 21
7.2 Display annotation....................................................................................... 22
7.3 Cineloop........................................................................................................ 22
7.4 Compare........................................................................................................22
7.5 Quick save feature........................................................................................... 22
7.6 Physio............................................................................................................ 23
7.7 Archive........................................................................................................... 23
7.8 Report............................................................................................................ 23
7.9 Connectivity.................................................................................................. 23
7.10 Probes/application..................................................................................... 24
7.11 Safety Conformance..................................................................................24
8. Measurement and Analysis...................................................................................... 24
8.1 Measurement in different modes............................................................... 24
8.1.1 Generic Measurement in 2D mode.......................................................24
8.1.2 Generic Measurement in M mode........................................................ 25
8.1.3 Generic Measurement in PW mode......................................................25
8.2 Measurement in different applications......................................................26
8.2.1 Abdominal Measurement..................................................................... 26
8.2.2 Small Part Measurement.......................................................................26
8.2.3 Vessel Measurement............................................................................. 26
8.2.4 Gynecology Measurement.................................................................... 26
8.2.5 Urology Measurement.......................................................................... 26
8.2.6 Pediatric Measurement.......................................................................... 26
3
8.2.7 Obstetrics Measurement.......................................................................26
8.2.8 Cardiac Measurement........................................................................... 26
8.2.9 Auto NT (Nuchal Translucency measurement)(Optional)*................... 27
8.2.10 Smart 3D Volume Measurement(Optional)*...................................... 27
8.2.11 Auto Follicle(2D/3D)(Optional) *.........................................................27
4
Ultrasound System Specifications
probes
● This unique platform is capable of
●Ease of use and ergonomic design
processing multiple data
streams simultaneously
1. System Overview
● Directional-enhanced
1.1 Architecture information compiling for more tissue
detail and reduction of angle-generated
● The revolutionary RF platform, The
artifacts
First In The World, allows for more
● Next generation adaptive image
accurate information. This platform
processing for noise and artifact
transfers all RF data for computing
reduction that improves tissue
without any information loss. It has a
presentation and edge definition
much better advantage in detail
●Fully independent, triplex multiple
imaging than current advanced
mode operation for easy in Doppler
platforms.
procedures
● Thanks to the RF platform, it allows
● Multi-processors
the development of many RF-based
allow simultaneous mode changes and
processing algorithms, which have
support for advanced system
ultra-premium contrast and resolution
functionality
imaging
● World-class design to be thinner and
lighter
●First in class to introduce capacity
8
• Adjustable text size and arrow size
2.9 Operating Environment
• Supports home position
• Covers various application • Ambient temperature:
• User customizable 10-40°C
• Relative humidity: 30-75%
2.5 Bodymark • Atmospheric pressure:
700hPa-1060hPa
• More than 215 bodymarks for
versatile application 2.10 Storage & Transportation
• User customizable
Environment
• Ambient temperature: -5-50°C
2.6 Peripherals
• Relative humidity: 10%-80%
• B&W thermal video printer: Sony (no condensation)
UP-D898MD (optional) • Atmospheric pressure:
• Color thermal video printer: Sony 700hPa-1060hPa
UP-D25MD (optional)
• Memory stick 3. Transducers
• Convex array
9
• Linear array D3-6C broadband curved array volume
• Phase array probe
• 4D probe • Field of view: 75 degree
• Endocavity probe • Convex radius: 40mm
• Micro-convex array • Application: abdomen, OB/Gyn,
urology
• B-mode Frequency range:
3.3 Transducer selection
3.0–6.0MHz
• Electronic switching of • Physical Footprint:
transducers 82mm×53mm
• User customizable imaging • Center frequency: 4.0 MHz
presets for each transducer and • Transducer element: 128
application •Pulsed wave Doppler, color Doppler,
• Automatic dynamic receiving power Doppler, harmonic, 3D/4D
focus in all transducers grayscale and 3D color modes
• Multiple adjustable transmit • Multi-imaging frequency setting in
focal zone, up to 8 focal zoom 2D, Harmonic, color Doppler and Wave
Doppler modes
G2-5C Broadband Curved Array
• Field of view: 66 degree F2-5CE Broadband Curved Array
• Convex radius: 50mm • Field of view: 59 degree
• Application: abdomen, OB/Gyn, • Convex radius: 60mm
urology, pediatric • Application: abdomen, OB/Gyn,
• B-mode Frequency range: urology, pediatric
2.0 -6.0MHz • B-mode Frequency range:
• Physical Footprint: 2.0 -6.5MHz
68.5mm x 27mm • Center frequency: 3.2 MHz
• Center frequency: 3.2 MHz • Physical footprint:
• Transducer element: 128 72mm x 27mm
•Pulsed wave Doppler, color Doppler, • Transducer element: 128
power Doppler, harmonic •Pulsed wave Doppler, color Doppler,
• Multi-imaging frequency setting in power Doppler, harmonic
2D, Harmonic, color Doppler and Wave •Multi-imaging frequency setting in 2D,
Doppler modes Harmonic, color Doppler and Wave
•Reusable biopsy guide available
10
Doppler modes 2D, Harmonic, color Doppler and Wave
•Reusable biopsy guide available Doppler modes
• Reusable biopsy guide available
G4-9M broadband micro convex array
• Field of view: 138 degree F4-9E broadband micro convex
• Convex radius: 12mm endocavity array
• Application: pediatric, abdomen, • Field of view: 156degree
cardiac • Convex radius: 10mm
• B-mode Frequency range: • Application: Ob/Gyn, urology
5.0 – 11.0MHz • B-mode Frequency range:
5.0 - 11.0MHz
• Physical Footprint:
• Center frequency: 3.2MHz
34mm × 29mm
• Physical footprint:
• Center frequency: 7.0MHz
21.0mm (lens) x 19.0mm(lens)
• Transducer element: 128
• Transducer element: 128
•Pulsed wave Doppler, color Doppler,
• Pulsed wave Doppler, color Doppler,
power Doppler, harmonic
power Doppler, harmonic
• Multi-imaging frequency setting in
• Multi-imaging frequency setting in
2D, Harmonic, color Doppler and Wave
2D, Harmonic, color Doppler and Wave
Doppler modes
Doppler modes
• Reusable biopsy guide available
G4-9E broadband micro convex
endocavity array
F4-12L broadband linear array
• Field of view: 138 degree
• Fine pitch, high resolution
• Convex radius: 11.5mm
•Applications: vascular, small parts
• Application: Ob/Gyn, urology
• Aperture size: 38.4mm
• B-mode Frequency range:
• B-mode Frequency range:
5.0 – 11.0MHz 6.0 -16.0MHz
• Center frequency: 6.9MHz • Center frequency: 7.5MHz
• Physical footprint: • Physical footprint:
24.0mm x 18.8mm 50mm×18.5mm
• Transducer element: 128 • Transducer element: 128
• Pulsed wave Doppler, color Doppler, • Pulsed wave Doppler, color
power Doppler, harmonic Doppler, power Doppler, harmonic
• Multi-imaging frequency setting in • Multi-imaging frequency setting in
11
2D, Harmonic, color Doppler and Wave Doppler modes
Doppler modes
• Reusable biopsy guide available G1-4P phased array
• Applications: cardiac, abdomen
I7-18L broadband linear array (Hock • Field of view 90 degree
stick) • Aperture size: 17.92mm
• Fine pitch, high resolution • B-mode Frequency range:
•Applications: vascular, small parts, 2.0-5.0Mhz
Hock stick • Physical Footprint:
• Aperture size: 25.6mm 34.5mm×28.5mm
• B-mode Frequency range: • Center frequency: 2.5MHz
7.3-18.0 Mhz • Transducer element: 64
• Center frequency: 9.5MHz • Pulsed wave Doppler,
• Physical footprint: continuous wave Doppler, color
31.7mm x 15mm Doppler, power Doppler, harmonic
• Transducer element: 128 • Multi-imaging frequency setting in
•Pulsed wave Doppler, color Doppler, 2D, Harmonic, color Doppler and Wave
power Doppler, harmonic Doppler modes
• Multi-imaging frequency setting in • Reusable biopsy guide available
2D, Harmonic, color Doppler and Wave
Doppler modes G3-10PX phased array
• Application: pediatric cardiology,
I4-11T broadband linear array abdomen
• Fine pitch, high resolution • Aperture size: 15.36 mm
• Aperture size: 25.6mm • Field of view: 90 degree
• B-mode Frequency range: • B-mode Frequency range:
6.0-12.0Mhz 3.0-10.0Mhz
• Physical Footprint: • Physical Footprint:
48mm×15mm 33mmx 33mm
• Center frequency: 7.5MHz • Center frequency: 4.7MHz
• Transducer element: 128 • Transducer element: 96
• Pulsed wave Doppler, color •Pulsed wave Doppler, color Doppler,
Doppler, power Doppler, harmonic power Doppler, harmonic
• Multi-imaging frequency setting in • Multi-imaging frequency setting in
2D, Harmonic, color Doppler and Wave 2D, Harmonic, color Doppler and Wave
12
Doppler modes
4.4 3D/4D
4. Advanced Imaging controls • 3D/4D rotation
• Grayscale imaging controls
• Selectable rendering
4.1 VFusion
approaches
• Available on all transducers and for • Unique high quality rendering
2D, 3D/4D algorithm
• Up to 5 levels of directional imaging • Selectable gray maps
fusion to enrich information • Multi slide cutting
• Operate in conjunction with • Cineloop 3D
VSpeckle, harmonic imaging • Review volume
4.4.1 HQ (optional)*
4.2 VSpeckle
• Amazing high image quality
• Available on all transducers and for • Extreme realistic rendering、
2D, 3D/4D images
• Virtually eliminate speckle noise • Similar operation as normal
artifact and dynamically enhances rendering
tissue margins
• Selectable multiple levels of speckle 4.5 Smart operation
noise reduction and smoothing
• Operates in conjunction with 4.5.1 Free View(optional)
VFusion and harmonic imaging •Provide any plane view to visualize
the internal tissue information
• Improve the contrast resolution to
4.3 VTissue
facilitate the detection of diffuse
• Speical imaging processing to adapt lesions in organs
to the speed of the ultrasound
variation in different tissue 4.5.2 Inversion mode*(optional)
• Improved conspicuity of lesions, • This render mode is used to display
such as stone and tendon anechoic structures such as vessels
• It invert the gray values of the
rendered image, such as black image
13
information become white and vice
4.8 Next generation RF-based
versa
image processing
4.5.3 Magic Cut(optional)
• Available on all imaging
• Ability to edit images, make transducers in 2D grayscale modes
possible to cut away structure • Virtually eliminates speckle
obstructing the view in the ROI noise artifact and dynamically
•Several cutting methods available enhance tissue edge
• Operates with other real-time
4.6 Smart 3D processing algorithms
15
• Cineloop review for • Flash reduction:≥ 6 steps
retrospective analysis of M-mode data • Persistence:≥ 20 steps
• 256 gray levels • Color Map:≥ 33types
• Acoustic output:10%-100% • Smooth :≥ 7steps
• Gray filter:≥ 7steps • Sensitivity:≥ 6 steps
• Dynamic range:108db-128db, • Transparency:≥ 6steps
2db/step • Color level:≥ 16 steps
• Vsharpen: ≥ 6steps • Packet size:≥ 9 steps
• Gray Map: ≥ 23types • Reverse function: on/off
• Tint Map: ≥ 24types • Color gain:0-100%
• Gain: 0-100% • Region of interest
• Color M mode:available • Baseline invert
• MultiAngle:available • Simultaneous mode during PW
mode
5.4 Color Doppler mode
• Zoom
• Available on all imaging transducers 5.5 Power Doppler mode
• Automatically adapts transmit and
receive bandwidth processing based on • High sensitive mode for small
the color box position vessel visualization
• Cineloop review with full playback • Available on all transducers
control • Cineloop review
• Steering on linear array transducers • Display format:Single, Dual,
• Selectable in baseline, line density, Quad
flash reduction, persistence, maps, • Individual controls for gain
frequency, PRF, wall filter, packet size, • Selectable line density, flash
color level, sensitivity, focus position, reduction, persistence, maps,
acoustic power, and smooth frequency, PRF, wall filter, packet size,
• FULL screen imaging to larger color level, sensitivity, focus position,
image size acoustic power, and smooth
• L/R flip and U/D flip: on/off • Color maps: ≥ 24 types
• Frequency:≥ 5steps, depend on • Color levels:≥ 11 steps
probes • Sensitivity:≥ 6steps
• Baseline:0-100% • Smooth:≥ 7steps
• Acoustic power:0-100% • Persistence:≥ 20 steps
• Line density:≥ 7 steps • Adjustable region of interest
16
or retrospective (V1/3, V1/2, V2/3,
H1/2, H3/4, full screen)
5.6 Pulsed Wave (PW) Doppler
•Auto function to optimize spectral
• Ultra high resolution spectral FFT Doppler displ
rate • Digitally enhanced stereo
• Angle correction with automatic output
velocity scale adjustment • 256 gray levels
• Normal, invert display around • Post-processing in frozen mode
horizontal zero line includes map, baseline, invert and
• Selectable gray filter, dynamic range, chroma
frequency, PRF, wall filter, baseline, • Simultaneous or duplex mode of
angel correct, sample volume operation
• Selectable sweep speeds: ≥ 10 • Simultaneous 2D, color Doppler,
steps pulsed Doppler
• PW acoustic output: 5%-100% • High PRF capability in all modes
• Gray filter: ≥ 6steps including duplex and triplex
• Dynamic range: 108db-128db
• Baseline: 5%-95% 5.7 Continuous Wave Doppler
• Sample volume: 0.5mm-10mm
• Angle correct:-80°~ 80° (CWD)
• Sensitivity: ≥ 21steps
• Cardiac sector array transducer only
• Audio Volume: 0~20
• Maximum velocity range: 19m/sec
• Spectrum Optimize: ≥ 28steps
• Gray map: ≥ 13types
• Tint map: ≥ 11types 5.8 Pview*
• Trace direction: above, below, both
• Trace type: max, mean, both • Real time extended field of view
• Cardiac cycle: 1-5 composite imaging
• Selectable low frequency signal • Ability to back up and realign the
filtering with adjustable wall filter image during acquisition
settings • Full zoom, cineloop review and
• Selectable grayscale curve for image rotation capabilities
optimal display • User can measure distance and area
• Selectable chroma colorization maps • Measurement can be made on
•Selectable display format prospective individual frames during cineloop
17
review • Image Reference:A,B,C,3D
• Available on linear transducers • Flip: 0°,90°,180°,270°
• View:Front/Back,Back/Front;
Left/Right,Right/Left;Up/Down,
5.9 Tview
Down/Up
• Expand view of scanning • Rotation Direction:X, Y, Z
• Available on linear transducers • 3D Map:≥ 8types
• Tint maps: ≥ 24Types
• Gray maps: ≥ 23Types
5.10 Auto • 2D VSpeckle:≥ 3types
• 3D VSpeckle:≥ 3types
• Intelligent one button
• Render Type:Gray, Graylnv
automatic optimization in 2D and
• Inverse Avaliable
Doppler modes
•Automatically adjust PRF and baseline
• MCUT(optional)
in Doppler
• Slice Number: 2×2, 3×3, 4×4, 5×
5.11 3D/4D 5
• Max Slice Number:25
• 3D/4D rotation • Gray Map: ≥ 23types
• Grayscale imaging controls • Tint Map: ≥ 24types
• Selectable rendering • Cut plane: A,B,C
approaches:HQ Surface, HQ Grad,HQ • Rotation Direction: X, Y, Z
Silhouette, Surf Texture, Surf Smooth, • Volume Angle: 15°~ 75°
Grad Light, Surf HDR, Trans Max, • Interval:1mm-20mm,0.5mm/step
X-ray,Transp Min,Light • Quality:low,mid,good,high
• Unique high quality rendering
algorithm • Free view(optional)
• Volume Angle:15%-75% • Direction: left, right, up, down
• Quality: low,mid,good,high • Route: curve, straight line
• Threshold:256 • Reference image: A,B,C
• Transparency: 0.1-2, 0.1/step • Slice thickness: 0mm-20mm
• Category: Face,Spine,Brain,Heart, • Active line: 1,2,3
Hi speed,Lip&plate,Limbs,Custom • Mix: 10-90
• Display format: • Threshold: 256steps
single,double,triplex,Quad • Transparency: 0.1-2.0, 0.1/step
18
• U/D
• Magic cut(optional) • Center line
•Erase mode: inside lasso, outside • VTissue
lasso,big circle, small circle • VSpeckle
• Erase type: trace, rectangle, ellipse • VFusion
• Rotation direction: X, Y, Z • Gray Filter
• Persistence
• VOCAL • Display Format
• Vocal layers: • Image reference
8, 12,16,20,24,28,32 • Maps
• Frequency
• Display format:single, Quad
• Focus position
• Image reference: A, B, C
• Focus #
• Dynamic Range
• Niche view
• Line density
• Model type: upper, lower
• VSharpen
• Display format: single, quad
• Biopsy
• Rotation direction: X, Y, Z
• Image angle
• Image reference: A, B, C, N
• Focus width
• Smooth
6. Touch Panel Interface
• Acoustic power
6.2 M Mode
6.1 2D mode
• New patient
• New patient
• BodyPattern
• BodyPattern
• Archive
• Archive
• Probe&APP
• Probe&APP
• Comments
• Comments
• End exam
• End exam
• Sys setting
• Sys setting
• Report
• Report
• L/R format
• PView
• U/D format
• Fullscreen
• Maps
• L/R
19
• Dynamic range • Acoustic power
• Acoustic power • Smooth
• Sweep speed
• Gray filter
6.4 PW/CW mode
• VSharpen
• ECG • New patient
• BodyPattern
6.3 CF mode • Archive
• Probe&APP
• New patient • Comments
• BodyPattern • End exam
• Archive • Sys setting
• Probe&APP • Report
• Comments • Invert
• End exam • Triplex
• Sys setting • Display format
• Report • Sweep speed
• Invert • Gray filter
• Full Screen • Dynamic range
• L/R • Trace sensitive
• U/D • Auto trace
• Baseline • Mode/direction
• Flash Reduction • Maps
• Line density • Frequency
• Persistence • PRF
• Display format • Wall filter
• Image reference • Baseline
• Maps • Angle correct
• Frequency • Sample volume
• PRF • Audio Volume
• Wall filter • Spectrum optimize
• Packet size • Acoustic power
• Colorlevel
• Sensitivity
• Focus position
20
• Display format
6.5 3D mode
• Image reference
• Comments • View
• BodyPattern • Gray map
• Back to B • VSpeckle
• Start3D • Quality
• ROI shape • Threshold
• Render • Transparency
• Display format • Volume angle
• Image reference
• View 7. System Feature
• Gray map
• VSpeckle
7.1 Display modes
• Quality
• Threshold • Simultaneous capability
• Transparency • 2D/PW/CW
• Volume angle • 2D/CF or PDI
•Movement step (after data • 2D/M
acquisition) • Dual, 2D/2D
• HQ Light(after data acquisition) • Dual, 2D/2D+CF or PDI
•Rotation angle (after data acquisition) • Dual, duplex and triplex
•Rotation direction (after data • Duplex and Triplex mode
acquisition) • Quad display in 3D/4D
application
• 25 slice images display in
6.6 4D mode
3D/4D application
• Comments • Time line display
• Body Pattern • Independent dual 2D/PW or
• Back to B CW
• Start 4D • Timed based sweep update
• ROI shape mode
• Movement step
• Rotation direction
• Render
21
• PW/CW mode: Doppler
7.2 Display annotation
acoustic power output, Doppler
• Institution/hospital name gain, Doppler frequency, PRF, wall
• Date: 3 types selectable, 9 slice filter, sample depth
images display in 3D/4D • Focus zone marker
• Time: 2 types selectable, 24hours • Body pattern
and 12 hours • PW and CW scale markers:
• Operator identification time/speed
• Patient name, first, last • M scale markers: time/depth, time
• Patient identification: 30 characters • System measurement display
• Gestational age from • System message display
LMP/BBT/DOC/IVF/GA/Avg.US • Biopsy guide line
• VINNO image symbol: Ginkgo leaf • Heart rate
• Power output index 7.3 Cineloop
• MI: mechanical index
• TIS: thermal index soft • Acquisition, storage in memory and
tissue display of up to 1000 seconds long of
• TIC: thermal index cranial 2D, color and PW/CW images for
(Bone) review
• TIB: thermal index bone
• Probe orientation marker: coincide
7.4 Compare
with a probe orientation marking on
the probe • Compare live imaging with stored
• Gray/color bar imaging
• Measurement result window
• Probe type
• Application name 7.5 Quick save feature
• Image depth
• Imaging parameters by mode • The system provides quick save
• 2D/M mode: acoustic power function through USB stick,
output, gain, frequency, frame rate, internal/external HDD during or after
dynamic range exam
• Color mode: color acoustic • Configurable saving file format, VRD
power output, color gain, color flow (VINNO Raw Data), DICOM,JPEG,BMP,
frequency, PRF, wall filter PNG, MP4 and AVI
22
7.6 Physio 7.9 Connectivity
23
•ISO 10993-1:2009 Biological
7.10 Probes/application
evaluation of medical devices
• Selectable multiple applications •IEC 62304:2006 Medical device
• Edit exist application preset software –Software life cycle
• Edit user defined preset processes
• Rename preset •IEC 62366:2007 Medical
• Return to factory preset devices - Application of usability
•Quick save user defined parameters in engineering to medical devices
related application •Council Directive 93/42/EEC on
Medical Device
•WEEE according to 2012/19/EU
7.11 Safety Conformance •RoHS according to 2011/65/EU
• Regulatory Notice:
8. Measurement and Analysis
This device is tested to meet all
applicable requirements in relevant.
According to 93/42 EEC, it is class IIa 8.1 Measurement in different
medical device.
• Conformity to Standards: modes
•IEC 60601-1 : 2012 Medical
8.1.1 Generic Measurement in 2D
electrical equipment - Part 1: General
requirements for basic safety and mode
essential performance
• Depth
•IEC 60601-1-2:2007
• Distance
Electromagnetic compatibility -
• Perimeter
Requirements and tests
• Length and width method
•IEC 60601-1-6:2010Usability
• Ellipse method
•IEC 60601-2-37:2007
• Polygon method
Medical electrical equipment -
• Spline method
Particular requirements for the safety
• Tracing method
of ultrasonic medical diagnostic and
• Area
monitoring equipment
• Length and width method
•IEC 61157:2007
• Ellipse method
Declaration of acoustic output
• Polygon method
parameters
• Spline method
24
• Tracing method 8.1.3 Generic Measurement in PW
• Volume
• Single line method mode
• Dual line method • Speed (include PV
• Triple line method (Peak Velocity))
• Single ellipse method •Time (include AT (Accelerate Time))
• Single ellipse and single line • Acceleration
method • PS (Peak Speed in systole
• Angle period)
• Stenosis • ED (The speed in the end of
• Diameter method diastole period)
• Square meter method • MD (Minimum speed in diastole
• A and B ratio period)
• Diameter ratio • TAMAX (maximum speed in
Generic Measurement in CFM time average)
mode • TAMEAN (mean speed in time
• CFV average)
• point • TAMIN (minmum speed in time
• profile average)
• PI (Pulsatility Index)
8.1.2 Generic Measurement in M • RI (Resistance Index)
• PS and ED ratio
mode • ED and PS ratio
• Depth • A and B ratio (A/B ratio)
• Distance • Speed ratio
• Time • Time ratio
• Slope • Acceleration ratio
• Heart rate • Square meter ratio
• Stenosis • FLOWVOL (Flow Volume)
• A and B ratio •MaxPG ( maximum pressure gradient)
• Diameter ratio • MeanPG (Mean pressure
• Time ratio gradient)
• Velocity ratio • SV ( Stroke Volume)
• Each volume diameter cardiac
• Time mean speed in each stroke
25
volume 8.2.4 Gynecology Measurement
• Cardiac output
• Uterus and Pelvis
• Heart rate
• Follicle
• SV(LVOT)/SV(RVOT)
• Carotid artery
8.2.8 Cardiac Measurement
• Upper artery
• Upper vein • General
• Lower artery • LV
• Lower vein • MV
• Vessel puncture • Ao
• Transcranial Doppler • AV
• LA
• RV
26
• TV
• PV 8.2.10 Smart 3D Volume
• RA
• System Measurement(Optional)*
• Trace the margin of the
8.2.9 Auto NT (Nuchal Translucency irregular circle in different slices of
volume data in irregular shape
measurement)(Optional)* • Automatically report the
• Automatically detect Nuchal volume of the irregular object
Translucency in interest box
• Automatically report thickness result 8.2.11 Auto Follicle(2D/3D)(Optional)
of NT
Auto IMT (Intima-Media Thickness) *
measurement • Just click on the area of follicle in B
• Automatically detect intima mode, the area of this follicle will be
media thickness in interest box reported automatically
• Automatically report the result of •Report the area of different follicle in
IMT the volume data automatically
• Available in linear probe
27