Professional Documents
Culture Documents
Display Unit
Start Stop
Control Section
Electrode A Electrode B
Rest Plane
Arc Section
ON/OFF Switch
Caution!!
Before removing the electrodes switch OFF the power.
Voltage Setting
The high voltage required for the test is 12,500 volts (12.5kV). This
voltage has to be adjusted after turning ON the machine with the help of
auto transformer knob. Rotate the knob and adjust the voltage 12.5kV is
read on the volt meter display. While the test is in progress, if this voltage
happens to vary, adjust it as and when it varies.
Cleaning Electrodes
If excessive accumulations of products of decomposition occur on the
electrodes, clean them gently with laboratory tissue. Switch OFF the
power before starting the cleaning process. Application of a
continuous 40mA Arc for approximately 1 min also can further clean the
electrodes. To do this, place a ceramic plate on the rest plane and select
cycle 7 and turn ON the test. The Arc will automatically stop after one
minute.
Operating modes
Mode Selection
The default setting of the machine is manual OFF mode. Immediately
after the power is turned ON the display shows Auto Off. This means the
machine is in manual OFF mode. Pressing the Start switch displays Auto
Off which means auto off feature is set and the test will be automatically
terminated after the sample forms a conducting track.
At this stage pressing Start push button again and again toggles the test
setting between auto OFF yes and auto OFF no which means Auto OFF
mode and Manual OFF mode respectively.
Note: As per ASTM test specifications the test starts from first cycle
and continues through next cycles until the sample fails. However a
provision is provided in this machine to start the test with any of
the 7 cycles as the starting cycle.
Starting cycle can be selected by Pressing the select switch. Each time
the switch is pressed the cycle is incremented by 1 until cycle 7. The
numbers 1 to 7 are displayed on the display beside the start switch. If the
switch is pressed while in cycle 7 it comes to mode selection. If the
switch is pressed again it comes to cycle 1. In this way pressing the
select switch again and again the selection process passes through all
the cycles and then through mode selection and then again through all
the cycles in a cyclic fashion. The cycle number at which you stop will
get selected.
Note on Auto OFF & Manual OFF modes
Auto OFF mode will work only when specimen fails by forming a clear
conducting path. The specimen may fail in other ways. The person
conducting the test should use his experience in selecting the mode.
Test in progress
While the test is in progress the cycle in progress is displayed (number 1
to 7) and the current in milliampere is also displayed in and the time
elapsed from the start of the test is also displayed. The display is
incremented every second. As long as the test is ON, the red lamp
glows. In case the door is accidentally opened while test is in progress,
the testing process is automatically stopped and the time elapsed is
retained.
T1
T1
RL6 R10-1 R10-2 R10-3 R10-4
T3
T3
T5
T5
RL3 HV Transformer
R40-1 R40-2 R40-3 R40-4
ON PCB
230V AC Input
Electrodes (Tungsten)
T7
Auto Transformer
Fuse
To primary side
Circuit of HV
ON/OFF Switch
To Control Circuitry
3 Pin Socket (Power In)
INPUT / OUTPUT Circuitry
Relay RL2
T8 230V Lamp
T9
0V (Signal Ground)
T7
T6
T1
T2
T3
T4
T12
T5
T13
T14 T8
T9
T15
T10
12V 0V 5V T11
CONTROL SECTION COMPONENTS
TO FAN
CONNECTOR
PCB
POWER SUPPLY
TO
DISPLAY
PCB
TRANFORMER
Right Side View (Calibration)
10mA 20mA
L2cx a
30mA 40mA
CAUTION!
Always ensure that link L is in
place while testing. Removing
L may cause unnecessary
arcing at undesired places
and may cause personal
injury.
This link is meant to be
replaced by a milli ammeter
only at the time of calibration.
Calibration
CAUTION!!!
While calibrating for open circuit voltage, always
make sure that link L is put back in place. Trying to
calibrate without L in place may cause arcing at
undesired locations and can damage the machine
and may also cause injuries to persons near the
machine.
Open circuit Operating Voltage:
Turn OFF the power supply. Put back the link L in place if has been
removed for inserting the ammeter. Remove the tungsten electrodes
(This creates enough gap to avoid arcing while calibrating). Attach the
leads of the high voltage probe to the electrode holders. Attach the
ground lead of the probe to holder of electrode A and attach the High
voltage lead of the probe to the holder of electrode B (Refer to the figure
“Arc resistance Testing Machine” on page No.3). Now turn ON the
power and select cycle 4 and start the test cycle by pressing the start
switch. The red light will glow as usual. Adjust the autotransformer until
12.5KV is obtained on the calibration meter.
If the cycle jumps to 5 while calibrating (which it does after 60seconds),
halt the calibrating process temporarily, stop the test cycle by pressing
Stop switch. Using select switch, select cycle 4 and resume calibration.
Note If any minor Arcing is seen or vibration is felt while holding the high
voltage probe during calibration, then remove one of the hexagonal
electrode holders and attach the probe to the electrode block which
holds the hexagonal holder and restart the calibration process.
20mA Calibration
Refer to the figure “Right side view (Calibration)” on page No.14.
Turn OFF the power supply. Replace the tungsten electrodes in their
place with regular gap. Adjust the gap with the gap gauge. Disconnect
the link L. Connect the Ammeter probes in place of the link L. Place a
Ceramic plate on the rest plane below the electrodes such that the
electrodes rest on it. Now switch ON the power supply. Select cycle 5
and start the test by pressing start switch. Arc will be seen between the
electrodes and the Ammeter will show a reading of approximately 20mA.
Adjust 20mA Knob until the ammeter reads 20mA exactly. Observe for
sometime for stable current reading. Make further fine adjustments of the
knob if necessary and stop the cycle by pressing Stop switch. Maintain
cycle 5 till the end of 20mA calibration.
30mA Calibration
Refer to the figure “Right side view (Calibration)” on page No.14.
Turn OFF the power supply. Make sure the tungsten electrodes are in
place on the Ceramic plate. Make sure the ammeter is connected in
place of link L. Now switch ON the power supply. Select cycle 6 and start
the test by pressing start switch. Arc will be seen between the electrodes
and the Ammeter will show a reading of approximately 30mA. Adjust
30mA Knob until the ammeter reads 30mA exactly. Observe for
sometime for stable current reading. Make further fine adjustments of the
knob if necessary and stop the cycle by pressing Stop switch. Maintain
cycle 6 till the end of 30mA calibration.
40mA Calibration
Refer to the figure “Right side view (Calibration)” on page No.14.
Turn OFF the power supply. Make sure the tungsten electrodes are in
place on the Ceramic plate. Make sure the ammeter is connected in
place of link L. Now switch ON the power supply. Select cycle 7 and start
the test by pressing start switch. Arc will be seen between the electrodes
and the Ammeter will show a reading of approximately 40mA. Adjust
40mA Knob until the ammeter reads 40mA exactly. Observe for
sometime for stable current reading. Make further fine adjustments of the
knob if necessary and stop the cycle by pressing Stop switch. Maintain
cycle 7 till the end of 40mA calibration.
Completion of Calibration
Turn OFF the unit. Disconnect the ammeter and connect the link L into
it’s place and observe if it is properly secured by pulling out one end at a
time with a little force. Securing this link L properly is very essential for
proper and safe functioning of the machine.