Professional Documents
Culture Documents
USER’S MANUAL
CONTENTS
1 GENERAL CHARACTERISTICS ....................................................................................................................... 4
1.1 Intended Use ................................................................................................................................................. 4
1.2 Indications ..................................................................................................................................................... 4
1.3 User Profile ................................................................................................................................................... 4
1.4 Operating Environment ................................................................................................................................. 4
1.5 Patient Profile ................................................................................................................................................ 4
1.6 Contraindications........................................................................................................................................... 5
1.7 Possible Side-effects ..................................................................................................................................... 5
2 SAFETY PRECAUTIONS ................................................................................................................................... 6
3 USED SYMBOLS ............................................................................................................................................... 7
4 INSTALLATION AND ASSEMBLY .................................................................................................................... 8
4.1 Installation of BTL CardioPoint Software ....................................................................................................... 8
4.2 Product Protection ......................................................................................................................................... 8
4.3 DEMO Mode.................................................................................................................................................. 8
4.4 Assembly of the BTL CardioPoint-ABPM System ......................................................................................... 8
4.5 Communication Settings ............................................................................................................................... 8
5 OPERATING INSTRUCTIONS ........................................................................................................................... 9
5.1 Start the Examination Using The BTL CardioPoint-ABPM Software ............................................................. 9
5.1.1 Settings Examination Parameters ......................................................................................................... 10
5.1.2 Sending The Examination Setting to the Device .................................................................................... 12
5.2 How to Import the Examination ................................................................................................................... 14
5.2.1 Loading the Examination Data from the Device to a Computer ............................................................. 14
5.3 Evaluation of an Examination ...................................................................................................................... 15
5.3.1 Main Screen Description........................................................................................................................ 15
5.3.2 Blood Pressure Table ............................................................................................................................ 15
5.3.3 Blood Pressure Trend ............................................................................................................................ 17
5.3.4 Statistical Values Table ......................................................................................................................... 18
5.3.5 Medical Findings .................................................................................................................................... 20
5.3.6 Software Settings .................................................................................................................................. 20
6 ADDITIONAL SOFTWARE SETTINGS ........................................................................................................... 27
6.1 Changing the Layout Colour ........................................................................................................................ 27
6.2 Export the Examination Data ....................................................................................................................... 27
7 LIST OF ERROR CODES WITH EXPLANATION ............................................................................................ 28
7.1 Unsuccessful Measurements ...................................................................................................................... 28
7.2 Cuff Related Errors...................................................................................................................................... 28
7.3 Faulty Device............................................................................................................................................... 28
8 FORMULAS USED FOR STATISTICAL CALCULATIONS ............................................................................. 29
8.1 Measured Data ............................................................................................................................................ 29
8.1.1 SYS: Systolic Blood Pressure ................................................................................................................ 29
8.1.2 DIA: Diastolic Blood Pressure ................................................................................................................ 29
8.1.3 Pulse: Heart Rate, Heart Frequency and Pulse Rate ............................................................................ 29
8.2 Calculated Data ........................................................................................................................................... 29
8.2.1 PP: Pulse Pressure................................................................................................................................ 29
8.2.2 MAP: Mean Arterial Pressure ................................................................................................................ 29
8.2.3 Mean...................................................................................................................................................... 29
8.2.4 Night Fall Index (Diurnal Index) [%]: (D/N Index) ................................................................................... 30
8.2.5 PTE [%]: (Percentage Time Elevation) .................................................................................................. 30
8.2.6 LOAD [mmHg]: (Hyperbaric Impact) ...................................................................................................... 30
8.2.7 PTD [%]: (Percentage Time Depression)............................................................................................... 30
8.2.8 Leese [mmHg]: (Hypotensive Load) ...................................................................................................... 31
8.2.9 Morning Surge ....................................................................................................................................... 31
8.2.10 SD (SYS/DIA) [mmHg]: (Standard Deviation) ........................................................................................ 31
8.2.11 CV (SYS/DIA) [%]: (Coeficient of variation) ........................................................................................... 31
9 HOT KEYS........................................................................................................................................................ 32
10 REFERENCES ................................................................................................................................................. 33
11 MANUFACTURER ........................................................................................................................................... 34
The BTL CardioPoint is a software system designed for multiple cardiology and spirometric examinations.
Moreover, BTL CardioPoint serves as a joint patient database, provides functionality to support teamwork and
possibility of integration with third party systems. The BTL CardioPoint is consisted up to five modules:
the BTL CardioPoint-ECG, BTL CardioPoint-Holter, BTL CardioPoint-Ergo, BTL CardioPoint-Spiro and
BTL CardioPoint-ABPM. BTL CardioPoint is widely used by cardiologists, pneumologists as well as general
practitioners.
The BTL CardioPoint-ABPM software, together with the compatible ABPM recorder, serves as a system for
ambulatory non-invasive blood pressure measurements and consequent evaluation of the measured data. The
software allows its user to define measuring intervals and select a suitable standard for data evaluation. When the
data is recorded, it is possible to review it, edit it and evaluate it according to the selected standard. The software
displays the data in a well-arranged layout. All examination data including a medical report are digitally stored in
the unified patient database and can be opened or printed anytime later.
The BTL CardioPoint is intended for acquisition, processing, recording, storage, analysis and presentation of
examination-related data.
The BTL CardioPoint-ABPM module is intended for acquisition, processing, recording, analysis and presentation
of long-term blood pressure recording and other related data.
1.2 INDICATIONS
The indications for ambulatory blood pressure monitoring include among others:
Hypertension
Masked hypertension
Suspected white-coat hypertension
Establishment of dipper status
Resistant hypertension
Antihypertensive drug treatment guide
Evaluation of hypotension
Autonomic failure, Collapse
Syncope
The BTL CardioPoint-ABPM shall be used by medically educated personnel. The user shall be familiar with all
safety precautions, operating procedures and maintenance instructions given in this User’s Manual.
The BTL CardioPoint-ABPM is intended to be used primarily in hospitals, but it can also be used in clinics,
medical centres or wherever the ambulatory blood pressure monitoring is performed. The BTL CardioPoint-ABPM
is not intended for home-use.
The BTL CardioPoint-ABPM can be used on all adult or paediatric patients without limitation of age, gender,
weight, height, etc. The patient must not show any symptoms listed within the list of contraindications. The user
should take into account a detailed patient’s medical history and examine the patient thoroughly to determine
whether or not this particular diagnostic method is suitable for the patient.
The contraindications for ambulatory blood pressure monitoring include cardiac rate is irregularity, as in atrial
fibrillation.
There may be technical reasons why ambulatory readings fail in some patients (e.g. problems with cuff fitting in
patients with conical-shaped arms, movement artefact, tremor, weak or irregular pulse, auscultatory gap). ABPM
may not be accurate during exercise or when driving.
Possible side-effects for use of ambulatory blood pressure monitoring may include petechiae, bruising under the
inflating cuff and sleep disturbance.
Read the User’s Manual carefully and become familiar with all its safety requirements and
operating procedures before using the product. It is not allowed to use the product in any
manner other that is in accordance with the User’s Manual. It is prohibited to use the product
in the way that is not in accordance with the Intended use.
Before the use of this product please read carefully User's Manual for a compatible device.
The BTL CardioPoint offers no diagnostic opinions to the user. Instead, it provides interpretive
statements of morphology, rhythm and conduction for the qualified healthcare professionals in
order to render their own medical opinion.
Use only original BTL accessories and compatible devices. Do not use any device or
accessories other than approved by BTL.
Before the start of examination, verify that all parameters that can affect the examination are
set.
In case a remote service of the product is required, disconnect the patient from the system.
Remote service cannot be carried out while a patient is connected to the system.
If the product or its accessories shows any signs of defects or you are in doubt about its
correct operation, stop working with the product immediately. If you cannot detect the source
of your concerns, even after a careful reading through of the User’s Manual, contact
an authorised BTL service without delay. If the product is used contrary to this instruction
manual or it is used in despite of the product showing functional divergences from those
stated in this User’s Manual, then the user is responsible for any damage caused by or
caused to the product.
Caution
CE mark
The installation of the BTL CardioPoint software is described in the BTL CardioPoint User’s Manual.
BTL CardioPoint software users are asked to register their new product. Please ask your
vendor to supply you with the license file. The word “Unregistered” will be displayed next
to the mouse cursor and on all final reports until the product has been officially registered.
There is a DEMO patient in the database which allows you to see the software working
even without the protection mentioned above. However, when the DEMO patient is
selected, the BTL-08 ABPM will not communicate with the software, the BTL
CardioPoint-ABPM will only display simulated values.
The BTL CardioPoint-ABPM software only works only with the BTL-08 ABPM.
For communication with the computer, use the special optical/USB interface. This connector can be found on the
right side of the BTL-08 ABPM device.
Connection process:
1) Respect the red mark on the cable – see the picture below.
2) Connect the optical/USB interface to the USB port on the computer.
3) Connect the small ending of the optical/USB interface to the optical port on the BTL-08 ABPM device.
You should hear a soft click.
There are no special settings. The BTL CardioPoint-ABPM will locate the device automatically.
Complete instructions how to operate BTL-08 ABPM unit is described in its own User’s Manual. Below you will
find instructions how to operate the BTL CardioPoint-ABPM only.
After clicking “New ABPM”, the window displaying the parameters for the examination will open.
SUMMARY
Review the Select start time of Save current settings Program the unit. The examination settings
settings for the the BP measuring as default settings for will be entered in the recorder and then
examination. and length of the future examinations. may be disconnected from the computer
examination. and the patient can leave.
BP SETTINGS
You can set measuring intervals for the day and night periods. Default setting is to measure every 15 minutes
during day and every 30 min during night. It is possible to add special or morning measuring intervals as well.
List of all
measurements.
BP SPECIAL SETTINGS
If you change the cuff size or want to change the maximum cuff pressure limit, you can do it on the “BP special
settings” tab.
Be sure to select the proper cuff size each time! If the wrong cuff size is used, inaccurate results may
occur.
Consider, whether it is really necessary to tick the “Display during measurements” option. The changing numbers
can attract the patient’s attention and might influence the blood pressure measuring.
If the patient uses any medications, you can enter them here and their dosages.
Copy the medicines from a previous examination. Select a row in the table
If the medicines have changed, you can add or and remove it by clicking
remove them. this button.
Should there be a record in your ABPM recorder that has not been imported to the BTL CardioPoint yet, a
warning message will pop up. Click “Yes” to erase this record and start new recording.
Note! This message can also appear when starting a new recording with a brand new ABPM recorder. The BTL-
08 ABPM recorders are tested during production and may already contain some data.
You will be notified about successful transmission of the examination plan to the device.
When the computer confirms that the examination settings were sent, you can disconnect the recorder even
though you still see the communication with the computer symbol on the ABPM recorder display.
How to import the examination data from the BTL-08 ABPM to a computer and how to do an evaluation of that
examination data is described here.
If the recorder is not connected (or recognized) during the loading of the
examination data, you will see an error message. In this event, connect the
BTL-08 ABPM recorder and click the “Find ABPM device” button. The
procedure will be repeated.
After successfully finding the ABPM recorder, the big window will change
to the smaller ABPM form. This will be the case if the patient’s data is in
the database and the ID of the examination in the recorder is the same as
on the computer (meaning that the examination was started from the
same computer).
If the examination was started from another computer, the “Load Abpm” dialogue box will opened showing the
possibility to select an already existing patient’s file or to create a new patient file.
All measured values in the To switch between the chart and All measured values in the trend-chart.
table. histograms.
Settings and print buttons. Space for your medical findings. Table with statistical values.
Used to switch the table between listing the data The red highlighted values are over the limit in comparison
measurement by measurement or by the hourly to the selected standard.
averages and events.
The fields in the “note” and “Error” columns are not possible
to edit. These are reserved for system messages.
Used to switch the table between listing measurement by measurement and the hourly averages.
Errors, medication, bedtime indicators and remarks relevant to particular measurement will be hidden
in the table. However, remarks to hourly average can be added.
If you untick one of specific measurements/events, the other ones will disappear from the list and will
not be printed in the final report.
Switches to to
the the Medication, daytime and night- The red line shows the limit of
histogram. time icons. systolic blood pressure.
The blue line shows the limit of
diastolic blood pressure.
Limit values are defined in the
criteria.
Different colors indicate the borderline between Change the graph style and the
the different periods: Morning, day, night and length of the graph.
special.
Time interval used for Periods and their time range which were selected
calculation of these statistics. for the examination.
Label Description
Successfully Readings Represents how many measurements were successfully measured.
All Readings Represents how many measurements were measured, both programmed and manual.
Successfully Ratio [%] Represents the percentages of the measurements that were successfully measured.
SYS/DIA avg [mmHg] Represents the average systolic and diastolic blood pressure through measured
interval. This interval is shown above the table. This pressure does not have to be
measured during the examination. It could be composed from two different pressures.
For more details, see the formula in Chapter 8.2.3.
SBP Interpretation Label Written interpretation of the systolic blood pressure.
DBP Interpretation Label Written interpretation of the diastolic blood pressure.
SBP Borderline values Limit values for systolic blood pressure (it is defined in the criteria).
DPB Borderline values Limit values for diastolic blood pressure (it is defined in the criteria).
MAP avg [mmHg] Represents averaged middle arterial pressure throughout the whole examination. For
more details, see the formulas in Chapters 8.2.2 and 8.2.3.
Pulse avg [bpm] Represents the average pulse. For more details, see the formula in Chapter 8.2.3.
Night fall index SYS/DIA [%] Gives the difference between active (daytime) and passive (night-time) means as a
percentage of the active mean. The DI is abnormally narrow if it is below 10%. For more
details, see the formula in Chapter 8.2.4.
SYS/DIA day/night Represents difference of average values between day and night period.
difference [mmHg]
Morning Surge The morning surge is the difference between the mean systolic blood pressure during
the first two hours after waking and getting up minus the mean systolic blood pressure
of the lowest blood pressure reading during sleep and its two adjacent values (total of
3 values). The software interprets the last nightly measurement at the time when the
patient wakes up. For more details, see the formula in Chapter 8.2.9.
SYS max [mmHg] Represents the maximum systolic blood pressure.
SYS min [mmHg] Represents the minimum systolic blood pressure.
DIA max [mmHg] Represents the maximum diastolic blood pressure.
DIA min [mmHg] Represents the minimum diastolic blood pressure.
PTE SYS/DIA [%] Hypertensive time index SYS/DIA. For more details, see the formula in Chapter 8.2.5.
PTE SYS [%] Hypertensive time index SYS. For more details, see the formula in Chapter 8.2.5.
PTE DIA [%] Hypertensive time index DIA. For more details, see the formula in Chapter 8.2.5.
PTD SYS/DIA [%] Hypotensive time index SYS/DIA. For more details, see the formula in Chapter 8.2.7.
PTD SYS [%] Hypotensive time index SYS. For more details, see the formula in Chapter 8.2.7.
PTD DIA [%] Hypotensive time index DIA. For more details, see the formula in Chapter 8.2.7.
SYS/DIA hyper readings Measuring of increase SYS/DIA blood pressure.
SYS hyper readings Measuring of increase SYS blood pressure.
DIA hyper readings Measuring of increase DIA blood pressure.
SYS/DIA hypo readings Measuring of decrease SYS/DIA blood pressure.
SYS hypo readings Measuring of decrease SYS blood pressure.
DIA hypo readings Measuring of decrease DIA blood pressure.
Time range The time range of the measure intervals (e.g. night, day, special…).
All/Successfully readings Represents the ratio between all readings and the successfully measured values.
SYS/DIA max [mmHg] Represents the maximum systolic and diastolic blood pressure during the measuring.
This pressure did not have to be measured during the examination. This number could
be composed from two different pressures.
SYS/DIA min [mmHg] Represents the minimum systolic and diastolic blood pressure during the measuring.
This pressure did not have to be measured during the examination. This number could
be composed from two different pressures.
SYS/DIA load [mmHg x h] Represents the load of the body with SYS and DIA blood pressure. For more details
see the formula in the Chapter 8.2.6.
SYS load [mmHg x h] Represents the load of the body with SYS blood pressure. For more details, see the
formula in Chapter 8.2.6.
DIA load [mmHg x h] Represents the load of the body with DIA blood pressure. For more details, see the
formula in Chapter 8.2.6.
SYS/DIA leese [mmHg x 24h] Represents systolic/diastolic hypotensive load on the patient's cardiovascular system.
For more details, see the formula in Chapter 8.2.8.
SYS leese [mmHg x 24h] Represents systolic hypotensive load on the patient's cardiovascular system. For more
details, see the formula in Chapter 8.2.8.
DIA leese [mmHg x 24h] Represents diastolic hypotensive load on the patient's cardiovascular system. For more
details, see the formula in Chapter 8.2.8.
SD (SYS/DIA) [mmHg] Standard deviation. For more details, see the formula in Chapter 8.2.10
CV (SYS/DIA) [%] Coeficient of variation. For more details, see formula in Chapter 8.2.11
Here, you can input your medical findings. If this tab is dark, the “Auto Complete” function is enabled.
Input frequently used text formula, select the text and then use
these buttons to add or remove the phrase from the dictionary.
For the acceleration and facilitation of the typing of the final reports, it is possible to utilize the auto-complete
function. This function enables you to pre-set your own templates of the final reports and to save any phrases
which you frequently use.
BP SETTINGS:
You can add or remove a “Morning” or Here, you can view and change the start time and end time for
“Special” period. It will have an impact on statistics calculation.
the statistic calculation. The statistics do not have to be calculated from all of the
examination data.
If you change the “Statistical range”, all values will be recalculated! It is possible to this do after printing the report,
as well! However, in this case, the values will be different.
The recommendations and guidelines for using in the BTL CardioPoint-ABPM software are described here.
The BTL CardioPoint-ABPM contains following recommendations and guidelines for hypertension:
AHA 2008 This recommendation is divided into “Normal” and “Abnormal” values. In these tables, Children
[2] only normal values are mentioned. The abnormal values are opposite to these stated (5 -16
values. years)
Examples:
NORMAL BP, Overall time int. Male Female
5y-5y11m <=116/74 <=115/74
Girls
24h Day Night
Percentile 90th 95th 90th 95th 90th 95th
Evaluation Borderline Abnormal Borderline Abnormal Borderline Abnormal
Age (y)
5 112/72 115/74 118/80 121/82 105/66 108/69
6 114/72 116/74 120/80 122/82 106/65 110/68
7 115/72 118/74 121/80 123/82 107/65 111/67
8 116/72 119/74 122/80 124/82 108/64 112/67
9 117/73 120/74 122/80 125/82 109/64 112/67
10 118/73 121/75 123/79 126/81 110/64 113/67
11 119/73 122/75 124/79 127/81 110/63 114/66
12 120/74 123/76 125/80 128/82 110/63 114/66
ESH 2016 Note: Patient´s height must be entered into the software, either during programming of Boys
pediatrics – the unit, or after loading of examination, otherwise the automatic interpretation will not (120-185
by height work. Go to Examination settings -> tab Patient data. cm)
Girls
(120-175
Boys cm)
24h Day Night
Percentile 90th 95th 90th 95th 90th 95th
Evaluation Borderline Abnormal Borderline Abnormal Borderline Abnormal
Height (cm)
120 114/74 117/77 122/80 125/82 103/61 106/63
125 115/74 118/77 122/80 125/82 105/61 108/63
130 116/74 119/77 122/80 126/82 106/62 110/64
135 117/74 120/77 123/80 126/82 108/63 111/65
140 118/75 121/77 123/80 126/82 109/63 113/65
145 120/75 123/77 124/79 127/81 111/64 114/66
150 121/75 124/77 125/79 128/81 112/64 116/66
155 123/75 126/77 127/79 130/81 113/64 117/66
160 124/75 127/77 129/79 133/81 114/64 118/66
165 126/75 129/78 132/80 135/82 116/64 119/66
170 128/75 131/78 134/80 138/82 117/64 121/66
175 130/75 133/78 136/81 140/83 119/64 122/66
180 131/76 134/78 138/81 142/83 120/64 124/66
185 133/76 136/78 140/81 144/84 122/64 125/66
Girls
24h Day Night
Percentile 90th 95th 90th 95th 90th 95th
Evaluation Borderline Abnormal Borderline Abnormal Borderline Abnormal
Height (cm)
120 112/71 114/72 118/80 120/82 103/63 106/65
125 113/71 116/73 119/80 121/82 104/63 107/66
130 114/72 117/73 120/80 122/82 106/63 108/66
135 115/72 118/74 120/80 123/82 107/63 109/66
140 116/73 119/75 121/80 124/82 108/63 110/66
145 117/73 120/75 123/80 125/82 109/63 112/66
150 119/74 121/76 124/80 127/82 110/63 113/66
155 120/74 123/76 125/80 128/82 111/63 114/66
160 121/74 123/76 126/80 129/82 111/63 114/66
165 122/74 124/76 127/80 130/82 112/63 114/66
170 123/74 125/76 128/80 131/82 112/67 115/71
175 124/75 126/76 129/81 131/82 113/63 115/66
Girls
Height (cm) Overall Day Night
120 113/73 120/84 107/66
130 117/75 124/84 109/66
140 120/76 127/84 111/66
150 122/76 129/84 112/66
160 124/76 131/84 113/66
170 124/76 131/84 113/66
180 124/76 131/84 114/66
Select a
Select a
recommendation for
recommendation/guid
hypertension.
eline for hypertension.
Configure the
hypotension values. It
is possible to enter
only your values for
hypotensive blood
pressure, because
there are no special
recommendations for
this kind of the blood
Table with predicted values. pressure. Therefore
It is not possible to edit the default recommandation/guidelines values. If you want to no recommendations
edit one of them, it will be saved as user BP criteria. are present.
The rows represent the borderlines for each stage and the columns represent the
blood pressure clasifications during the stages.
(<|=|<=|=<)(-|[0-9]+)/( -|[0-9]+)
Empty spaces and tabs will be
automatically deleted.
Repeat the procedure until you have entered all intervals and blood pressure levels.
Select the pages you want If you select more than 13 items, you will see
to print. this warning message.
You will be able to view all of your selected
parameters, but only the first 13 will be printed.
If you select more than 13 items, you will be able to view all of your selected values, but only the first 13 values
will be printed! You will see a warning message in the window for report settings.
BTL uses two different colour layout defaults for your comfort while using the software. You may change the
layout colours from the “Design” menu:
You may export data from the examination. These data will be exported as *.csv data which will be possible to
open and view in Microsoft Excel.
You can do this by clicking “Examination” => “Export to MS Excel” and then select the target directory for
exporting.
The ERROR codes which you may see in the BTL CardioPoint-ABPM are described here. You will see these
codes in the blood pressure table in the column labelled “Error”. The first number is the error code and the second
one is a number for development. You will find a description of the error codes in the tables below. In the “note”
column, there will be a short written interpretation of the error code.
Here is the description of the formulas and procedures which are used for the statistical analysis and
the calculated data types.
8.2.3 MEAN
Statistics can be the result of simple arithmetic calculations or they can be weighted averages. As blood pressure
is measured sequentially, the profile is not continuous. When using weighted averages, the program will treat an
average like it was a continuous value within a time interval, which is calculated by dividing the time period
weighting for time the program counts with a reading as a valid continuous value within a time range defined
by halving the times to its neighbouring readings.
and
This means that a measurement will represent half of the elapsed time
between its two nearest measurements. To avoid distorting the input,
the program uses the following rules which are valid for both t1 and t2:
Simple means:
Weighted means:
Shows what percentage of the total analysed time blood pressure was above cut-off lines. In other words, it
calculates an elevated time to total time ratio. The 140/90 mmHg cut-off line for casual blood pressure
measurements approximately equals to the ambulatory overall (24-hour) mean cut-off line of 135/85 mmHg. The
PTE for MAP is abnormally high if it is over 25 %.
k: Ranges from 1 to n.
The following is a list of the hotkeys used in the software program to facilitate your work.
Delete: The selected blood pressure will be marked as invalid and the values in the statistical table will
be recalculated.
Backspace: Resets an invalid blood pressure value back to valid and the values in the statistical table will be
recalculated.
Ctrl + Alt + P: Preview the examination report. It is possible print the final report from the preview page as well.
Up Arrow: Jumps to previous row in the table (measured value table and calculated values table) or to the
previous blood pressure reading on the chart.
Down Arrow: Jumps to next row in the table (measured value table and calculated values table) or to the next
blood pressure reading on the chart.
Stevenage
Hertfordshire
SG1 6BU
United Kingdom
E-mail: sales@btlnet.com
th
Date of last revision: 30 April 2019
ID: 591-80MANEN02203
© All rights reserved. No part of this manual may be reproduced, saved or transferred by any means including
electronic, mechanical, and photographic without the prior written approval from BTL Industries Limited.