User Manual

Central Infusion Management System


(EN-C7)
Manufacturer: Shenzhen Enmind Technology Co., Ltd.

Room 201,Block A,No.1,Qianhai Road 1,Qianhaishen Port Cooperative District,

Shenzhen, 518000,Guangong,China

Site included: 5th Floor, Block A, Defengsheng Building, No.41 Dabao Road, Bao'an District 23, Shenzhen,

518101, P.R.China

TEL: +86 755 2331 6007

1
Preface
1 Application Scope of the User Manual
Applicable to EN-C7 Central Infusion Management System

This user manual is described based on the software functions of the Central Infusion Management System.
The software needs to work with the designated infusion workstation, syringe pump, and infusion pump. The
functions or accessories mentioned in the user manual may not be available in the products actually used by
the user. For more information, please contact us.

2 Applicable Object of the User Manual


It is applicable to professional trained nurse, doctor, and maintenance technician of this equipment.

3 User Instructions
This user manual covers basic information about the safety and efficacy of the product to guide proper
installation, commissioning, operation, use and maintenance. Please read this user manual carefully before
using the product in order to use the product correctly. Please keep the user manual in a safe place for your
reference.

The company is responsible for the safety, reliability and performance of the product only if all of the
following conditions are met:

(1) Use in accordance with this user manual;

(2) Installation, unloading and testing must be carried out by our professional service personnel;

(3) All software installation packages, upgrade packages and documents are provided by the company.

4 Paraphrase
“ ” Indicates a web page button.

CIMS Central Infusion Management System.

2
Content
Preface ..................................................................................................................................................... 2
1 Application Scope of the User Manual................................................................................................. 2
2 Applicable Object of the User Manual ................................................................................................. 2
3 User Instructions .................................................................................................................................. 2
4 Paraphrase ............................................................................................................................................ 2
Content ........................................................................................................................................................... 3
Chapter 1 Overview ................................................................................................................................... 5
1.1 Application Scope ............................................................................................................................. 5
1.1.1 Intended Use ........................................................................................................................... 5
1.1.2 Intended Use Environment ..................................................................................................... 5
1.1.3 Applicable User ...................................................................................................................... 5
1.2 Working Principle.............................................................................................................................. 5
1.3 Use Limit ........................................................................................................................................... 5
1.4 Contraindications .............................................................................................................................. 5
1.5 Function Introduction ........................................................................................................................ 5
1.6 Specification ...................................................................................................................................... 6
Chapter 2 Safety Instructions ...................................................................................................................... 7
2.1 Warnings............................................................................................................................................ 7
2.2 Cautions............................................................................................................................................. 7
2.3 Symbols ............................................................................................................................................. 8
Chapter 3 Installation ................................................................................................................................ 9
3.1 Installation ......................................................................................................................................... 9
3.2 System configuration....................................................................................................................... 12
Chapter 4 Basic Operation ......................................................................................................................... 13
4.1 Bed Monitoring ............................................................................................................................... 13
4.1.1 Main display frame ............................................................................................................... 13
4.1.1.1 bed management ............................................................................................................ 14
4.1.2 remaining time & alarm display box .................................................................................... 14
4.1.2.1 remaining time ............................................................................................................... 14
4.1.2.2receiving alarm information ........................................................................................... 14
4.2 Drug Library Information Management .......................................................................................... 15
4.3 Patient Information Management .................................................................................................... 15
4.4 Recipe management ........................................................................................................................ 16
4.5 History Records Management ......................................................................................................... 16
4.6 Real layout....................................................................................................................................... 17
4.7 Ward Parameters Settings ................................................................................................................ 17
4.8 Safety management ......................................................................................................................... 17

3
4.9 Alarm volume adjustment ............................................................................................................... 18
4.10 Help ............................................................................................................................................... 18
4.11 Operation log ................................................................................................................................. 18
4.12 Database backup and recovery ...................................................................................................... 18
4.13 Mobile infusion observation function (optional) ........................................................................... 19
4.14 Others ............................................................................................................................................ 19
4.15 HL7 configuration ......................................................................................................................... 19
Chapter 5 Alarm Prompt and Troubleshooting ........................................................................................ 21
5.1 Introduction to Alarm Level ............................................................................................................ 21
5.2 Multi-level Alarm Rules .................................................................................................................. 21
5.3 Alarm Handle .................................................................................................................................. 21
5.4 Troubleshooting............................................................................................................................... 21
Chapter 6 Maintenance............................................................................................................................ 22
6.1Maintenance instructions.................................................................................................................. 22
6.2 Production Date ............................................................................................................................... 22
6.3 Recycling......................................................................................................................................... 22
Chapter 7 After Sale Service ................................................................................................................... 23
Appendix Alarm and Solution ................................................................................................................ 24

4
Chapter 1 Overview

1.1 Application Scope


1.1.1 Intended Use
The central infusion management system is mainly used to remotely display the infusion station (EN-D7
Smart) with WIFI module produced by our company or the infusion pump with WIFI module (EN-S7 Smart
syringe pump, EN-S5D series syringe pump, Infusion status and alarm information of EN-V7 Smart infusion
pump, EN-V5 series infusion pump, EN-V3 series infusion pump..

1.1.2 Intended Use Environment


Hospital

1.1.3 Applicable User


Applicable to professional nurses, physicians, etc..

1.2 Working Principle


After the central infusion management system is connected to the infusion workstation or the infusion pump,
the infusion workstation or pump transmits the data of the pump to the central infusion management system
via the wireless network, and the central infusion management system displays the collected data on the
interface.

1.3 Use Limit


The central infusion management system restricts the use of some parameter information, including the scope
of the search, and the customer can only search within a limited number of beds. Customers cannot use
unauthorized infusion pumps, syringe pumps, infusion workstations, unauthorized infusion pumps, infusion
workstations will not be allowed to access the central infusion management system.

1.4 Contraindications
N/A.

1.5 Function Introduction


The central infusion management system includes the following features:
1) Remote monitor infusion pump
2) Remote monitor patient bed
3) Receive alarm information function
4) Drug Library Information Management
5) Patient information Management
6) Prescription Information Management
7) History Record Management
5
8) Real layout
9) Ward Parameters Settings
10) Safety Management
11) Alarm Volume Management
12) Help
13) Operation log
14) Database Backup and Recovery
15) Mobile infusion observation function

1.6 Specification
Specifications
Bed No. 100
VIP Bed No. 10
EN-S7 series syringe pump, EN-S5D series syringe pump, EN-V7 series
Applicable pump
infusion pump, EN-V5 series infusion pump, EN-V3 series infusion pump
Applicable workstation EN-D7 Smart Infusion workstation
Network disconnecting response time: ≤ 20 seconds;
efficiency Alarm delay time: ≤ 20 seconds;
Data refresh cycle: ≤ 20 seconds.
Software release version
EN-C7 1
Operating environment
Operating platform Windows10
Memory ≥4GB
Hard Disk 20GB Available
Display resolution 1920*1080
CPU 2.5 GHZ,64 bit
Network CS configuration, Lan network,100Mbps
Compatible with general security software 360 security guards, 360 antivirus,
Tencent computer butler, Kingsoft Internet Security.
Security software
The update of the security software has no effect on the central infusion
management system.
Data interface
Software universal data
Ethernet
interface
Data transfer protocol Support UDP protocol format
User access control
System installation and The key functions of the central infusion management system are password
operation protected and the password can be modified and saved.

6
Chapter 2 Safety Instructions

2.1 Warnings

 The warnings involved indicate that the risk of death or injury may result if not observed.

 The central infusion management system is used for remote monitoring and management of bed
operation status, infusion information, drug information, alarm information, and patient information in
the LAN during clinical use, and must be operated by professional medical personnel.

 Set the system alarm volume according to the actual situation of the patient. The patient should not rely
solely on the audible alarm of the central infusion management system for infusion monitoring of the
patient, and should pay close attention to the actual situation of the patient.

 The maintenance and upgrade of the central infusion management system should be carried out by
maintenance personnel trained and authorized by the company.

2.2 Cautions
 Cautions/Precautions: The precautions involved indicate that the risk of personal injury or property
damage if not observed.

 This product does not include infusion pumps and infusion workstations. Supporting our company's
infusion workstation with WIFI module (EN-D7 Smart) or infusion pump with WIFI module (EN-S7
series syringe pump, EN-S5D series syringe pump, EN-V7 series infusion pump, EN -V5 series infusion
pump, EN-V3 series infusion pump).

 Users cannot rely solely on this product. This product cannot completely replace the clinical observation
of the infusion pump by medical personnel.

 Please ensure the accuracy of the computer system time, so as not to record the wrong information, or
delay the alarm information.

 The supporting computer hardware and software should meet the operational requirements of this
product.

7
2.3 Symbols
Marks Description Marks Description

Infusion Pump Injection pump

Batch code Serial number

Date of Manufacture Manufacturer

environment-friendly use
Handle with harmless method
period (20 a)

Indicates the wifi connection


Caution
status

8
Chapter 3 Installation

3.1 Installation
1)Unzip the software installation package.

2)Start running the installer center_setup.exe.

The interface after starting the program is as follows:

3)Select the language "English" and click OK.

4)Enter the password 1006 in the password input box of the installation interface that pops up, and click the
“Next” button.

9
5)Select the directory where the software is installed. Use the default directory here and click the “Next”
button.

6)Select the “Create desktop shortcut” checkbox to create a desktop shortcut and click the “Next” button.

10
7)Click the “Install” button.

8)Installing.

11
9)Successful installation.

10)Click the “Finish” button to complete the installation.

11)open the software, obtain the registration code from the manufacturer according to the serial number,
and click "OK" after input to start the operation.

3.2 System configuration


(1) after the central infusion management system is installed successfully, open the network center → local
connection → details to obtain the IPv4 subnet mask, default gateway and server IP parameters.

(2) turn on the WLAN switch for the pump , click “network connection” to input the access point name and
password of wireless route, and click “TCP / IP” to configure the parameters of local IP, subnet mask, default
gateway, server IP and server port “10060”.

Note: the last digit of the local IP must be distinguished from the server IP. If the server IP is 198.168.0.100,
the local IP can be set to 198.168.0.222. When connecting multiple pump ends, the local IP must not conflict.

(3) the system configuration is successful, and the icon is displayed on the pump end screen.

12
Chapter 4 Basic Operation
User open EN-C7, it will prompt an interface as following:

On the left side of the title bar at the top of the interface are company and product information. On the right
is the hospital information, followed by the hospital icon, hospital name and department name.

At the bottom of the interface is the software menu bar:Bed Monitor、Drug Library、Patient Monitor、
Recipe management、History、Setting, and Recipe management、Drug Library、Setting Real Scene Edit、
Department Setting、Mobile infusion observation function is optional.

The function of the software is described in detail below.

4.1 Bed Monitoring


The main interface is bed monitoring interface, which remotely monitors all bed information (100 beds in
total), operation status, patient information, infusion information, drug information (if any), remaining time
and alarm information.

"Bed monitor" window: main display box on the left, remaining time & alarm display box on the right.

4.1.1 Main display frame

The main display box is divided into all bed interfaces and VIP bed interfaces:

a) all bed interface displays all bed in bed information.

b) VIP bed interface displays VIP bed in bed information.

In bed information includes: bed number, patient information, infusion pump type, infusion information,
13
alarm information, VIP information.

c) VIP bed setting:

When the mouse arrow moves over the bed icon, the VIP icon will appear in the upper right corner.
Click the icon to add the bed to the VIP. At the same time, the icon is changed to 10 VIP beds at most.

After joining the VIP bed, the user can view the bed icon with larger characters in the VIP interface.

4.1.1.1 bed management

Double click the bed icon to enter the bed details interface, including patient information, infusion
information, workstation information, and VIP bed settings.

a) patient information: including medical record number, name, gender, age, height and weight settings.

b) infusion information: channel number, drug, status, category, preset quantity, speed, remaining time,
cumulative quantity

Channel number: when the infusion pump is networked through the workstation, the corresponding channel
number in the workstation will be displayed; "--" will be displayed when the single pump is networked;

Drug: display the corresponding drug name;

Status: display the corresponding operation and stop status;

Category: display infusion pump or injection pump;

Preset quantity: display the corresponding preset quantity value;

Speed: display the corresponding speed value;

Remaining time: displays the corresponding remaining time;

Cumulative quantity: display the corresponding cumulative quantity;

4.1.2 remaining time & alarm display box

4.1.2.1 remaining time

The display contents of drug remaining time information include: bed number, remaining time, drug and type.

The shorter the remaining time, the earlier the infusion information is displayed.

The list of drug remaining time information only shows the currently running infusion pump.

4.1.2.2receiving alarm information

receive the alarm information alarm display box to display the alarm information of all beds in the ward, and
generate corresponding voice and text prompts according to the high, middle and low level of the alarm. See
Chapter 5 of this manual for details.

14
4.2 Drug Library Information Management
Click “Drug Library” to enter the drug library information management window. The window displays
the drug information in the form of a list.

Click the “Add” button to directly enter the "Edit" window of drugs to add drug information.

Click the“Edit” button to edit the currently selected drug information.

Click“Delete” to delete the currently selected drug information.

Click“Import” button to import the external drug information into the central infusion management system.

Click“Export” button to export the existing drug information of the central infusion management system
to the local file.

Click“Category” button to add drug category information.

Click “Deliver” button, select the infusion pump to be distributed in the pop-up equipment list, click
"confirm", and the drug information of the system will be distributed to the corresponding pump end.

Warning:

● the drug store information management can only be operated by users with permission. Restrict by setting the
password for drug library editing.

4.3 Patient Information Management


Click " Patient Monitor " to enter the patient information management window, including the display of “All
patients”, “VIP patients” and "Add" patients, "Edit" patients, "Change bed" and "Discharge" functions. The
patient information is displayed as a list in the window.

Click the "Add" button to open the new patient information window, and click the "OK" button to save the
patient after editing. All information of the patient must be filled in completely. Click the "reset" button to
roll back the patient information to the state before editing.

The editable patient attribute information includes bed number, patient name, medical record number, gender,
age group, birthday, height and weight. The input of bed number is limited to 1-100; the maximum input of
patient name is 48 English characters or 16 Chinese characters; the maximum input of medical record number
is 24 English characters or numbers or 8 Chinese characters; the input range of height is 0-250; the input
range of weight is 0-300. The date of the birthday must be separated by "-" and the birthday cannot exceed
the date of the day.

Select the patient information to be modified, click "Edit" to open the editing window of the patient, and
click "OK" to save the modified information after modification.

15
Select the patient information to change the bed, click "Change bed", pop up the bed change interface, fill in
an empty bed number, and click "OK" to complete the bed change operation.

Select the patient information to be removed, and click the "Discharge" button to indicate that the patient
information has been deleted. Only after the patient information is deleted can a new patient be added to the
bed. After the patient is discharged, the patient information will disappear from the display window.

The patient information management window can observe up to 100 pieces of patient information. If you
want to observe patients who have been monitored by the central monitoring system in the past, you can view
them through the "history review".

4.4 Recipe management


Click "Recipe management" to enter the prescription information management window. The prescription
information is displayed as a list in the window.

The prescription information management window contains the functions of importing and exporting
prescription information.

Click the "Import" button to get the prescription information.

Click the "Export" button to export the existing prescription information to the local file.

4.5 History Records Management


Click "History" to enter the history management window.

The window includes the functions of viewing infusion record, pump history record, prescription record and
patient record.

Click the "infusion record" button to enter the function of viewing infusion record, select an online patient,
set the viewing time range, and click query to view the infusion record and infusion trend chart of the patient.

Click the "pump history" button to enter the view pump history function, select the SN number of the pump
to be queried, click the "query" button to view the pump history, and click the "export" button to save the
pump history. History information is displayed as a list.

Click the "prescription record" button to enter the view history prescription function, and the system displays
the history prescription information in the form of list. Select the patient, bed number, drug and execution
time, and click "search" to search for a specific prescription. The show all button is used to return to the list
of all historical prescriptions.

Click the "patient record" button to enter the view historical patient function, and the system displays the
historical patient information in the form of a list. Select the patient, bed number, entry time and removal
time, and click "search" to search for a specific patient. The show all button is used to return to the list of all

16
historical patients.

When the central infusion management system is used together with the infusion pump with Wifi module, it
includes the function of viewing and exporting the infusion pump history, infusion trend chart and patient
record.

When the central infusion management system is used with the workstation with Wifi module, only the
patient record viewing function is available.

4.6 Real layout


Click the “Setting”→ “Real Scene Edit” key in the main interface to enter the real scene layout function
and display the hospital bed information according to the real scene layout.

You can select the "room", "vertical door", "horizontal door", "vertical bed" and "horizontal bed" icons to
draw with the mouse, and long press and drag them to the blank area on the right to draw the real picture of
the ward bed.

Click the “Load” button to open the previously saved live view.

Click the“Save” button to save the currently edited live view.

Click the“Use” button to use the selected live view.

Click the“Clear” button to clear the currently drawn realistic picture.

Click the“Not use” to restore the default live view.

Click the “Grid” button to help draw a live view.

You can set the color of the live view through “Color”.

4.7 Ward Parameters Settings


Click the“Setting” → “Department Setting” button in the main interface to enter the department parameter
configuration function.

It includes the display of bubble size, cumulative bubble, KVO rate, occlusion pressure, pressure unit, finish
pre alarm, reminder alarm, consumables setting, system settings and night mode settings.

Click "Edit" and enter the password to enable the editing function. The Save button saves and sends the
current department parameters to the pump.

4.8 Safety management


Click “Setting”→ “Password Setting” in the main interface to enter the security management function.

Password protection is provided for the operation of login and logout of central infusion management system,
17
drug store, real picture editing, Department parameters and database.

service sister director


Software exit √ √ √
scene editor √
User password management √ √ √
Department parameter setting √ √
Database backup and recovery √
Drug Library Edit and distribution √ √

Service password, sister and director password are different.

Service password is a built-in password and cannot be modified. The sister password and the director
password can be modified by yourself.

Default sister password: 6666.

Default director password: 8888.

4.9 Alarm volume adjustment


Click “Setting”→“System settings”→“Volume”in the main interface to adjust the volume of alarm
tone played by the central infusion management system.

Alarm tone: 10 ~ 100, step 10.

4.10 Help
Click “Setting”→“System settings”→“Help Document”,in the main interface to view the description
information and version information of the central infusion management system.

4.11 Operation log


Click “Setting”→“System settings”→“System Log”,in the main interface to view the operation records
of the central infusion management system.

4.12 Database backup and recovery

Click “Setting”→ “Database setting”, in the main interface to backup and restore the database.

Database backup includes manual backup and automatic backup.

Manual backup: after setting the backup path, click the "save" button.

18
Automatic backup: select the "on" radio box for automatic database backup, set the backup time, and
automatically back up to the "database backup" folder in the installation document; turn off the automatic
backup function and select the "off" radio box.

Database restore: search to select the saved database file and click recover.

4.13 Mobile infusion observation function (optional)


The user can view the infusion information of the bed through the browser on the client side, infusion
remaining time and alarm information.

Users can open the browser on the mobile end of the LAN and enter the corresponding web address to view
it.

4.14 Others
Click "setting" → "system settings" in the main interface to set the hospital information, language and date
format.

Hospital information setting: set the hospital name, hospital icon, department name and number of beds.

Language setting: the optional languages are Chinese and English.

Date format setting: three date formats are optional.

4.15 HL7 configuration


There is an config.ini configuration file in the root directory of the software. After opening the configuration
file, you can set the connection type, target IP address, target port number, and local port number. After
properly configuring the relevant network parameters of the HL7 protocol, you can use the HL7 protocol to
communicate with the central station.

Step 1: Open the config.ini file in the installation path

19
Step 2: After opening the config.ini file, there will be 4 parameters that need to be configured. ConnectType
represents the connection method between the central station and the HIS system. ConnectType configuration
1 means the central station is the client and the HIS system is the server; 2 means the central station is the
server and the HIS system is the client; and 3 means communication over UDP mode. DestIp represents the
IP address connected to the HIS system, DestPort represents the port number connected to the HIS system,
and LocalPort represents the port number of the machine.

Step 3: Save the file and exit after completing the configuration, the central station will automatically connect
to the HIS system according to the configuration

20
Chapter 5 Alarm Prompt and Troubleshooting

5.1 Introduction to Alarm Level


Alarm connection in the infusion pump and an infusion preparation during the infusion, when reaches or
exceeds the alarm threshold is triggered central fluid management system will display, in voice and text
prompts generated way. According to alarm information of importance, urgency and safety factors to be
processed into high, medium and low alarm levels.

Advanced warning: issued a "Di di di-di di, "Di di di-di di " alarm sound, alarm text on a red background;

Intermediate warning: issued a " Di di di " sound the alarm, the alarm text on a yellow background;

Low-level alarm: issued a " Di " sound the alarm, the alarm text on a yellow background.

5.2 Multi-level Alarm Rules


When a plurality of alarm occurs, the alarm according to the following rules:

The same pump has multiple levels of alarm, display only the most advanced alarm. There are several most
advanced alarm, display more of the most senior police.

When a plurality of alarms occur simultaneously at the same level, the earlier alarm, sorted Echizen

Multi-pump multiple alarms occur simultaneously at different levels, the higher the alarm level, before
ordering more.

5.3 Alarm Handle

Warning ● When there’s alarm, please check the conditions of the patient, remove the reason of

alarm and then continue working.

Please refer to Appendix A for the alarm solution.

5.4 Troubleshooting
In the event of a system crash or crash, force a reboot of the system or computer.

21
Chapter 6 Maintenance

6.1Maintenance instructions
If you find abnormal central transfusion management system, restart the system or the computer is still not
resolved, contact your sales dealer or the manufacturer of this product to you and request repairs.

6.2 Production Date


Please refer to the label of the product.

6.3 Recycling
Please users to process this product in accordance with local laws and regulations.

22
Chapter 7 After Sale Service

After the purchase of this product 1 year free software maintenance services. Fill in the maintenance period
from the "warranty card" that comes with the installation date shown. "Warranty card" is the only evidence
calculate maintenance period, in order to safeguard your interests, please fill out and keep "warranty card"
serious and companies linked to the installers.

The following cases are not within the scope of free maintenance.

1. customer to complete within 30 days of acceptance of the system installed is not complete and return the
warranty card.

2. The impact of the use of computer software operating system instability, system poisoning, illegal
operation.

After the expiration of the maintenance, the company can continue to provide maintenance services, but will
charge a fee based on maintenance costs.

23
Appendix Alarm and Solution

Alarm
Alarm Type Reason Solution
Level
VTBI infused High The preset value infusion Completion. Press【Stop】button to stop alarm.
The liquid medicine in the syringe is
Syringe empty High Press【Stop】button to stop the alarm.
empty.
Manually solve the problem of
1. Line occlusion during infusion. occlusion, Press 【 Start 】 button to
continue infusion.
2. Fluid/drug in the actual infusion line
Pressure high High Rise the alarm Level, Press【Start】
of high viscosity, while the system
button to restart infusion.
occlusion level is set too low.
Please contact the dealer or
3. The pressure sensor is damaged.
manufacturer for repair
Check if the rate regulating adjuster or
The upper part of the line is obstructed
fluid stopping device is open at the
Check upstream High during infusion, and in turn drops the
upper part of the line, Press【Stop】
line pressure intensity.
button to stop alarm.
1.When power is supplied by the built-in
Immediately connect with external
battery only, under low battery, the alarm
power supply.
Battery empty High duration is >30min.
2. Battery aging or the equipment Please contact the dealer or
charging circuit is fault. manufacturer for repair.
Immediately connect with external
Backup battery 1.Backup battery is nearly exhausted
power supply.
power High
Please contact the dealer or
exhaustion 2. Backup battery is detached or aged
manufacturer for repair.
Check
High Syringe drop off during infusion. Reinstall the syringe.
syringe
KVO working time reached 30min,
KVO finished High Press【Stop】button to stop alarm
infusion pump stops working.
Press 【 Stop 】 button to stop alarm,
disconnect the line from the patient,
Single bubble High Air bubble in the infusion line. eliminate the air with purge function, or
open the infusion pump door to
manually remove the air bubbles.

24
Alarm
Alarm Type Reason Solution
Level

Press the【Stop】button to eliminate the


When the bubbles in the infusion alarm, separate the pipe from the
Cumulative
High pipeline within 15 minutes reach the patient, using the purge function to
Bubble
cumulative bubble alarm threshold remove the bubbles, or open the door
manually to remove bubbles.

During infusion, the infusion pump door Close the infusion pump door to stop
Door Open High
is opened. this alarm.

While using drugs in drug library to


Drug dose infuse, alarm will be triggered if max
High Press【Stop】button to stop alarm.
limits exceeded dose in certain time have exceeded the
preset limits.
If system self check failed or internal Restart device to check whether alarm
System error High fault, system error alarm will give with is eliminated, if still exists, contact
code number. maintenance personnel.
The angle of inclination of the drip cup Check the installation of drop sensor or
is too big or drop sensor is installed drip cup fluid level, Press【Stop】button
lower than the drip cup fluid level. to stop alarm.

Check if the infusion apparatus


Drop error High The specification of infusion apparatus
specification is accordant with
is not accordant with the specification
displayed parameters, if it is not
displayed in the interface, which causes
accordant, it shall be modified by
drop rate error.
professional maintenance technician.

The infusion set drip pot was detected Check if there is liquid left in the
Empty bottle High without drops falling within the infusion bag, press【Stop】to cancel the
specified time. alarm.
Infusion pump or infusion station
Adjust the distance between the device
connected to EN-C7 operation exceeds
and the router
WIFI connection distance
WIFI signal is unstable for infusion
pump or infusion station connected to Check the WIFI signal and reconnect
Communication
High EN-C7
interruption
The infusion pump is removed from the
Infusion pump re-access to the infusion
infusion station connected to EN-C7
station
during the infusion
Device WIFI module failure during Please contact the dealer or
infusion manufacturer for repair.

25
Alarm
Alarm Type Reason Solution
Level
Router failure during infusion Repair or replace the router
Computer disconnected during infusion Check computer network
The SN number of the pump is
artificially modified, and the EN-C7 Set the pump end SN number is not
detects the SN conflict of multiple repeatable
SN Confliction High infusion pumps.
Connect the pump and infusion station
Turn off the pump or infusion station
of the infusion station to open the WIFI
WIFI switch to eliminate the alarm
connection EN-C7 at the same time
Occlusion pre Line pressure close to preset occlusion Check if there is occlusion in line and
Medium
alarm pressure level. click OK to eliminate alarm.
Check extension line or patient
Drop in
Middle Pressure in line is dropped in sudden. connection. Click OK to eliminate
pressure
alarm.
Standby time During standby, after reaching the
Medium Press【Stop】button to stop alarm.
expired standby time.
During infusion, the remaining time
This alarm can’t be eliminated, and
VTBI near end Low reached or is less than the set nearing
wait till infusion completes.
completion time.
The syringe is near empty status which
Syringe near is calculated by checking the liquid This alarm can not be eliminated,
Middle
empty medicine remaining in the syringe by waiting until syringe is empty.
current flow rate.
No battery Keep connecting with external power
Middle Battery is removed.
inserted supply, reinstall the battery.
1. When power is supplied only with the The alarm automatically eliminates
built-in battery, under low battery, the after connecting the external power
Battery nearly
Middle alarm duration is >30min supply.
empty
2. Battery aging or the equipment Please contact the dealer or
charging circuit is fault. manufacturer for repair.
After installing infusion tube, under
non-working or alarm state, it is not
Reminder alarm Middle Click any button to stop.
operated within the set time of the
system.
Under ON state, AC power supply is The alarm automatically eliminates
No power
Low adopted, but the AC power wire is after connecting the external power
supply
dropped during the process. supply.
When turning on the drop sensor, the
Drop sensor Connect the drop sensor, or turn off the
Low equipment is not connected with the
connection drop sensor in the menu.
drop sensor.